Skip to main content
Wiley Open Access Collection logoLink to Wiley Open Access Collection
. 2020 May 9;84(5):2636–2644. doi: 10.1002/mrm.28308

MR fingerprinting for rapid simultaneous T1, T2, and T 1 ρ relaxation mapping of the human articular cartilage at 3T

Azadeh Sharafi 1,, Marcelo V W Zibetti 1, Gregory Chang 1, Martijn Cloos 1, Ravinder R Regatte 1
PMCID: PMC7396294  NIHMSID: NIHMS1612377  PMID: 32385949

Abstract

Purpose

To implement a novel technique for simultaneous, quantitative multiparametric mapping of the knee articular cartilage.

Methods

A novel MR fingerprinting pulse sequence is proposed and implemented for simultaneous measurements of proton density, T1, T2, and T 1 ρ relaxation times at 3T. The repeatability and reproducibility of the proposed technique were assessed in model phantoms. Institutional review board‐approved MR fingerprinting imaging sequence was performed on healthy volunteers and patients with mild knee osteoarthritis. The Wilcoxon test was used to compare healthy controls and patients. The intra‐ and intersubject repeatability were assessed with coefficient of variation and the RMS coefficient of variation, respectively

Results

The Bland‐Altman plots demonstrated an average difference of 4.67 ms, −0.09 ms, and 0.05 ms between 2 scans in the same scanner; and 9.68 ms, 0.29 ms, and −0.72 ms between the scans acquired on 2 different scanners for T1, T2, and T 1 ρ, respectively. The in vivo knee study showed excellent repeatability with RMS coefficient of variation less than 3%, 6%, and 5% for T1, T2, and T 1 ρ, respectively. The Wilcoxon test showed a significant difference between control and mild osteoarthritis patients for T1 (P = .04), T2 (P = .01), and T 1 ρ (P = .02) relaxation time in medial tibial cartilage, as well as for T2 relaxation time (P = .02) in medial femoral cartilage.

Conclusion

The proposed MRF sequence is fast and can simultaneously measure the T1, T2, T 1 ρ, and B1+ maps in a single scan. It is able to discriminate between mild osteoarthritis patients and healthy volunteers.

Keywords: knee joint, magnetic resonance fingerprinting, multiparametric mapping, osteoarthritis, T1ρ

1. INTRODUCTION

Osteoarthritis (OA) of the knee, the most common joint disease, 1 is a degenerative heterogeneous musculoskeletal disease that is mainly recognized by the progressive loss of hyaline articular cartilage. 2 Although the standard MRI techniques 3 , 4 for imaging cartilage can quantify morphological changes such as cartilage volume and thickness as a result of OA progression, 5 , 6 they are still insensitive in detecting early‐stage OA. 7 On the other hand, quantitative MRI can detect early changes in cartilage biochemical components that occur before the morphological change. More specifically, early‐stage OA is associated with loss of proteoglycans, an increase in water content, and changes in the structure of collagen fibers. 2 Yao et al 8 showed the sensitivity of T1 and T2 relaxation to early‐stage OA, and it has been shown that T1 mapping can be used to calculate the water content of the cartilage. 9 In addition to the longitudinal T1 8 and transverse T2 relaxation times, 8 the spin‐lattice relaxation in the rotating frame (T 1 ρ) can be used to quantify the biochemical changes in cartilage. 2 , 7 , 10 , 11 , 12 Both T 1 ρ and T2 have been consistently reported to be elevated with enzymatic cartilage degradation. 2 , 7 , 10 , 11 , 12 T2 is sensitive to the collagen fibril orientation and anisotropy, 13 , 14 whereas T 1 ρ is found to be sensitive to the slow motional interactions between local macromolecular environments and bulk water, 15 and as a result, to proteoglycan content in articular cartilage. 16 Moreover, quantitative evaluation of knee joint pathologies is often dependent not only on a single tissue property but also a combination of properties coregistered and evaluated together.

Conventional quantitative cartilage MRI (e.g. T1, T2, and T 1 ρ) approaches are relatively slow and sensitive to both noise and B1+ variations. Moreover, they generally provide only a single parameter at a time, therefore necessitating long data acquisition times if all biochemical information is desired.

Magnetic resonance fingerprinting (MRF) is a recently developed technique 17 that can estimate multiple MR parameters simultaneously (e.g., T1 and T2) using dynamic signal patterns. It has improved scan efficiency and robustness for the brain, 18 , 19 prostate, 20 abdomen, 21 , 22 hip, 23 and cardiac 24 applications. The MRF framework relies on a deliberate and continuous variation of multiple MR sequence parameters throughout the acquisition to provide unique fingerprint‐like evolutions for different combinations of tissue properties. In addition, incoherent undersampled k‐space data is acquired during this evolution, making the scan time much shorter than traditional quantitative mapping. The undersampling generates uncorrelated artifacts on the signal evolutions that are easily ignored by the dictionary‐matching task, using a dictionary constructed with predicted signal evolutions for different combinations of properties. The tissue property values associated with the best dictionary match are assigned as the measured properties for that voxel.

The main aim of this study is to implement a novel MR fingerprinting sequence for simultaneous T1, T2, and T 1 ρ relaxation mapping of the human knee joint at 3T. We evaluated the feasibility and performance of our MRF‐sequence on model systems, healthy and early knee OA patients.

2. METHODS

2.1. MRF‐sequence design

We developed an MRF‐sequence based on the method described in Ref. 23 by adding T 1 ρ preparation modules to estimate proton density, T1 and T2, T 1 ρ, and B1+ on 6 slices with 0.6 mm × 0.6 mm × 4 mm voxel resolution in less than 8 min. The proposed MRF‐sequence started with an adiabatic inversion pulse followed by 2 fast imaging with steady‐state precession segments, which predominantly encode T1/T2; and 2 fast FLASH segments, which encode T1 and B1+ (Figure 1). Each segment contains 250 RF excitations with a time‐bandwidth product of 3 followed by a single radial readout. The flip angle (Figure 1) varies in a sinusoidal fashion to smoothly vary the transient state of the magnetization, ranging from 0° to 60° in the first segment and from 0° to 20° in the second segment of both fast imaging with steady‐state precession and FLASH readouts. 23 The radial rotation was incremented between excitation by the golden angle. 25 There was a delay equal to 50 repetition times between segments for partial recovery of the magnetization and enhancing T1 encoding. To encode the T 1 ρ, N = 6 paired self‐compensated spin‐lock preparation modules 26 with different spin‐lock duration (Tsl = 2.0, 3.7, 6.9, 12.9, 24.1, 45 ms) were added at the end of the train. The most basic T 1 ρ preparation module starts with a 90° RF pulse along the x‐axis (90x) to flip the magnetization to the transverse plane. Afterward, a continuous spin‐lock pulse was applied along the y‐axis (αy) to lock the magnetization in the transverse plane. The magnetization will decay during the lock time (Tsl) according to T 1 ρ relaxation time. Applying a tip‐up 90° RF pulse in the opposite direction of the first (90 x) will tip the magnetization back to the longitudinal direction. Crusher gradients were then applied to destroy the remaining magnetization in the transverse plane before readout. In practice, the actual spin‐lock field strength and direction is affected by the field inhomogeneities. 27 The paired self‐compensated spin‐lock, 26 which was used in this study, tries to compensate for the B1+ inhomogeneities by altering the phase of the spin‐lock pulse. In this method, the spin‐lock pulse was divided into 4 segments with an alternative phase, and a refocusing pulse is applied in the middle between 2 pairs to compensate for the B0 inhomogeneities. 27 Each of the preparation modules was followed by a FLASH readout segment consisting of 125 RF excitations, with variable flip angle ranging from 0° to 20° (Figure 1). The complete train of 1750 excitations constitutes 1 shot. We acquired 4 shots per slice, uniformly distributed within k‐space. 23

FIGURE 1.

FIGURE 1

MRF sequence timing diagram for multiparametric mapping. The adiabatic inversion pulse is followed by 2 FISP and 2 FLASH segments to encode T1/T2 and T1/ B1+, respectively. Each segment contains 250 RF excitations followed by a single radial readout. There was a delay equal to 50 TR between segments for partial recovery of the magnetization and enhancing T1 encoding. The train is followed by T 1 ρ preparation modules with different spin‐lock durations. The spin‐lock pulse (αy) was divided into 4 segments with an alternative phase, and a refocusing pulse was applied in the middle between 2 pairs to compensate for the B0 inhomogeneities. The crusher gradients were applied after the T 1 ρ module to destroy any remnant of transverse magnetization. Each of the preparation modules was followed by a FLASH readout segment with radial readout consisting of 125 RF excitations with variable FA ranging from 0° to 20°. FA, flip angele; FISP, fast imaging with steady‐state precession; RF, radiofrequency

2.2. Model phantoms study

A model phantom consisting of 3%, 4%, and 8% agarose gels and cross‐linked bovine serum albumin was scanned on a 3T MRI scanner (Magnetom Prisma, Siemens Healthcare GmbH, Germany) with our proposed MRF sequence as well as the conventional T1 (MPRAGE), T2 (spin‐echo), and T 1 ρ (customized turbo‐flash 28 ) mapping sequences. The imaging parameters are shown in Table 1. The conventional T1 map was calculated using the inline MapIt feature on the Siemens scanner. Scans acquired with different TEs, including 25, 24, 36, 48, 60, 72, 84, 144, 278 ms, were fitted voxel by voxel to the monoexponential model to estimate the T2 relaxation time. The same method was used to estimate the T 1 ρ relaxation time from scans with different Tsl, including 2.0, 3.7, 6.9, 12.9, 24.1, 45, and 85 ms.

TABLE 1.

Imaging parameters of the proposed MRF and the conventional T1, T2, and T 1 ρ sequences

  MRF IR‐MP2RAGE (T1‐Conv) SE (T2‐Conv) T1rTFL (T 1 ρ‐Conv)
FOV (mm2) 160 220 160 160
TR (ms) 7.5 5000 6000 1500
TE (ms) 3.4 3.4 25, 24, 36, 48, 60, 72, 84, 144, 278 3.4
Tsl (ms) 2.0, 3.7, 6.9, 12.9, 24.1, 45 2.0, 3.7, 6.9, 12.9, 24.1, 45, 85
FA (°) 0‐60 4,5 90 8
Slice thickness (mm) 4 4 4 4
Matrix size 192 × 192 256 × 256 128 × 128 128 × 128
Receiver bandwidth (Hz/px) 390 240 150 390
TA (min) 7:06 4:00 25:44/TE 3:00/Tsl

Conv, conventional; FA, flip angle; IR, inversion recovery; MRF, MR fingerprinting; Tsl, spin‐lock duration, TA, total acqusition time; TFL, turbo‐FLASH.

In addition,  the MRF sequence was repeated on the same scanner to evaluate the repeatability and tested on another 3T MRI scanner (Magnetom Skyra, Siemens Healthcare GmbH, Germany) to assess the reproducibility.

2.3. In vivo study

Institution review board‐approved MRF imaging was performed on healthy volunteers (n = 5, mean age: 38 ± 12 years) and patients (n = 3) with mild knee OA (MOA) (KL score 1‐2, mean age: 63 ± 5 years) using a 15 channel Rx/Tx knee coil (Quality Electrodynamics, Mayville, OH). Six sagittal images were acquired with 0.6 × 0.6 mm2 in‐plane resolution, 4.0 mm slice thickness, 224 × 224 matrix size, TE/TR = 3.5/7.5 ms, bandwidth = 420 Hz/pixel, number of shots = 4, spin‐power fsl = 500 Hz. The acquisition time was 7:06 min. Three control subjects were scanned twice in the same session to assess in vivo repeatability.

2.4. MRF‐dictionary construction for multiparametric mapping

All the algorithms were implemented in MatLab (R2018a, MathWorks, Natick, MA). A dictionary of simulated MR fingerprints for possible T1, T2, T 1 ρ, and B1+ values was computed based on the extended phase graphs. 29 The T 1 ρ preparation module was simulated as a mono‐exponential decay of longitudinal magnetization because the application of the crusher gradients after the tip‐up 90° pulse destroyed any remnant transverse magnetizations. Assuming the initial magnetization of Mz 1 at the beginning of the T 1 ρ, the final magnetization, Mz 2 was estimated as:

Mz2=Mz1expTslT1ρ. (1)

The slice profile was incorporated into the simulation based on the Fourier transform of the RF waveform. 21 The T1, T2 (same range for T 1 ρ), and B1+ were ranged from 50 to 3000 ms, 2 to 200 ms, and 30° to 90° with the 5% increments, respectively. The dictionary and the measured fingerprints were then compressed using the singular value decomposition approach 21 , 30 and matched with each other using an iterative approach 31 , 32 , 33 , 34 to generate 5 multiparametric maps (proton density, T1, T2, T 1 ρ, and B1+) simultaneously.

2.5. Statistical analysis

The average of each parameter over all slices was calculated in phantom for each agarose gel and for bovine serum albumin, and for each volunteer over 5 regions of interest (medial and lateral femoral cartilage, medial and lateral tibial cartilage, and patellar cartilages).

The Bland‐Altman analysis was performed to assess the reproducibility and repeatability in the model phantoms. The Wilcoxon test was used to compare healthy control and MOA patients. The mean and SD of each parameter was calculated for each subject across the test–retest scan, and the coefficient of variation (CV) is calculated as CV = SD/Mean to assess intrasubject repeatability.

The intersubject repeatability was evaluated using the RMS CV:

RMS CV=i=1NCVi2N, (2)

where CVi is the CV of an individual subject and N (=5) is the total number of subjects.

3. RESULTS

Figure 2A shows the representative T1, T2, T 1 ρ relaxation and B1+ maps of the model agar gel phantoms. Analysis of the Bland‐Altman plot demonstrated an average difference of 4.67 ms, −0.09 ms, and 0.05 ms between 2 scans from the same scanner (Figure 2B); and 9.68 ms, 0.29 ms, and −0.72 ms between the scans acquired on 2 scanners for T1, T2, and T 1 ρ (Figure 2C), respectively.

FIGURE 2.

FIGURE 2

Representative PD, T1, T2, and T 1 ρ and B1+ maps of the agar gel model phantoms. The gels with higher agar concentrations have lower relaxation times (A). Regression and Bland‐Altman plots show excellent repeatability (B) and reproducibility (C) in estimating PD, T1, T2, and T 1 ρ in the agar gel model phantoms. PD, proton density

Comparisons between the proposed MRF relaxation times and the conventional techniques are shown in Figure 3. We observed average absolute difference of 328.0 ± 123.8 ms with a correlation coefficient of r = 0.96, 2.4 ± 2.1 ms with r = 0.99, and 2.3 ± 2.0 ms with r = 0.99 between the conventional and our proposed MRF technique for T1, T2, and T 1 ρ, respectively. The high correlation coefficient between our proposed method and the conventional techniques confirmed the agreement between MRF and conventional relaxation mappings.

FIGURE 3.

FIGURE 3

Regression and Bland‐Altman plots indicate good agreement between our proposed MRF method and the conventional mapping techniques. The average absolute difference of 328.0 ± 123.8 ms with a correlation coefficient of r = 0.96, 2.4 ± 2.1 ms with r = 0.99, and 2.3 ± 2.0 ms with r = 0.99 was observed between the conventional and our proposed MRF technique for T1, T2, and T 1 ρ, respectively. MRF, MR fingerprinting

The representative maps of the knee joint (medial and lateral cartilages) for control and MOA patient are shown in Figure 4A. The in vivo study showed excellent interscanner repeatability with CV smaller than 5%, 10%, and 7% for all individuals; and RMS CV less than 3%, 6%, and 5% across all ROIs for T1, T2, and T 1 ρ, respectively. The global average of T1 = 778.33 ± 48.46 ms, T2 = 23.25 ± 4.39 ms, and T 1 ρ = 32.42 ± 5.52 ms was observed in the control group, whereas the numbers were increased to T1 = 834.31 ± 37.79 ms, T2 = 26.05 ± 2.53 ms, and T 1 ρ = 33.88 ± 4.41 ms in the MOA patients. The boxplot comparisons of 3 relaxation parameters between control and MOA patients are shown in Figure 4B. An overall increase was observed in the patients. The Wilcoxon test showed a significant difference between control and MOA patients for T1 (P = .04), T2 (P = .01), and T 1 ρ (P = .02) relaxation parameters in medial tibial cartilage, as well as for T2 relaxation time (P = .02) in medial femoral cartilage.

FIGURE 4.

FIGURE 4

(A) Representative PD, T1, T2, and T 1 ρ relaxation maps of the knee joint (medial, lateral, and patellar cartilages) for healthy control and MOA of knee patient. (B) Boxplot comparison of T1, T2, and T 1 ρ relaxation times in different cartilage compartments (LFC, LTC, MFC, MTC, and PC) between healthy controls and MOA patients. A significant increase was observed in T2 and T 1 ρ values in MOA patients. LFC, lateral femoral cartilage; LTC, lateral tibial cartilage; MFC, medial femoral cartilage; MTC, medial tibial cartilage; MOA, mild osteoarthritis; PC, patellar cartilage

4. DISCUSSION

In this work, we implemented an MRF sequence and showed the feasibility of rapid simultaneous acquisition of accurate proton density, T1, T2, B1+, and T 1 ρ maps of the human knee joint. The proposed MRF method is fast, reproducible, and robust to B1+ inhomogeneity. Integrating the T 1 ρ module allowed us to perform voxel‐wise comparisons between different relaxation times. Moreover, when measured separately, each of these quantities can easily be biased by residual effects from the others. Measuring all of them at once and reconstructing multiparametric maps in 1 single step provides an elegant way to disentangle the individual contrast mechanisms and eliminate experimental imperfections such as B1+ inhomogeneities.

The estimated T 1 ρ was higher than T2, which is in agreement with the findings of previous literature. 7 , 35 , 36 The range was also comparable to the other studies. 35 , 36 , 37 , 38 Li et al 37 reported RMS CV of 6.0% and 6.3% for the knee T2 and T 1 ρ estimation, which is in agreement with our repeatability studies. Our relaxation times estimation in phantoms showed a good agreement with conventional techniques. The difference in T1 values could be due to the magnetization transfer effect, 39 which we did not include in our dictionary. We plan to use a more accurate simulation for our dictionary in the future.

Our proposed technique yields T1, T2, and T 1 ρ maps of the 6 sections of the knee cartilage with 0.6 × 0.6 × 4.0 mm3 resolution in 7:06 min (1:12 min per slice). Previous work reported 9:30 min to acquire T2 and T 1 ρ maps of 26 slices with 0.54 × 1.1 × 4.0 mm3 resolution, 37 and ~15 min to acquire a 3D biexponential T2 40 or T 1 ρ 28 map of the cartilage with 0.54 × 1.1 × 2.0 mm3 resolution. The T1 map was not provided in other studies. We note that to acquire T1, T2, and T 1 ρ data separately via traditional methods would require ~45 min of scan time.

The MRF sequence can be sensitive to motion. Whereas Ma et al 17 suggested the MRF is relatively robust to in‐plane motion at the end of the scan, Yu et al 41 showed a substantial decrease in T2 values when a through‐plane motion happened in the middle of the scan. They suggested that the difference in sensitivity to 2 different motions is due to the difference in the gradient used in in‐plane (balanced) and thorough‐plane (gradient spoiled), which leads to different phase accumulation for each coherence pathways when a through‐plane motion occurs. 41 Moreover, when using a 2D‐MRF implementation, the spin history is biased by fresh magnetization entering the excitation plane due to motion. Compared to T2 values, T1 values are shown to be less affected by motion, although some blurring in the fine details was observed. 41 We used extensive padding to avoid motion during scan time. Further studies are warranted to investigate motion sensitivity during the T 1 ρ acquisition.

Our study has some limitations. First, there were a small number of volunteers in each group, which is not enough to run reliable statistics. Moreover, the dictionary matching used in the proposed method is computationally costly and memory consuming, preventing online reconstruction. However, fast reconstruction and matching methods that can alleviate this problem are under investigation. Second, although the total scan time (~7 min) is shorter compared to acquiring T2, T1, and T 1 ρ data separately, the per slice scan time of our 2D‐multislice MRF is relatively long (~7 min for the 6 slices with 4 shots). To reduce scan time and improve the SNR, a 3D‐sequence with stack of stars 43 , 44 , 45 , 46 could be used. Future work could also explore the optimal design of experiments (e.g., flip angle train and optimal Tsls) based on Crammer‐Rao lower bound 47 and deep learning‐based reconstruction 48 instead of template matching. Finally, we simulated the slice profile based on the Fourier transform of the RF waveform. This method is valid for small tip angle approximation and hence may not be accurate for large flip angle. A general approach proposed by Ma et al 42 can be used to increase accuracy. This problem could also be addressed by using a 3D image‐encoding strategy. 43 , 44

5. CONCLUSION

In summary, we implemented and demonstrated the feasibility of an MRF sequence that can simultaneously measure the T1, T2, T 1 ρ, and B1+ maps in a single scan. Nevertheless, the in vivo results showed that it could distinguish the mild OA patients from the healthy controls and hence has the potential to be used for the quantitative assessment of the cartilage for the early detection of OA.

ACKNOWLEDGMENT

The authors thank James Babb, Ph.D., and Ding Xia, M.Sc, for their assistance in data analysis.

Sharafi A, Zibetti MVW, Chang G, Cloos M, Regatte RR. MR fingerprinting for rapid simultaneous T1, T2, and T 1 ρ relaxation mapping of the human articular cartilage at 3T. Magn Reson Med. 2020;84:2636–2644. 10.1002/mrm.28308

Funding information

This study was supported by NIH grants R21‐AR075259‐01A, R01 AR070297, R01 AR067156, and R01 AR068966, and was performed under the rubric of the Center of Advanced Imaging Innovation and Research (CAI2R), and NIBIB Biomedical Technology Resource Center (NIH P41 EB017183)

REFERENCES

  • 1. Vos T, Flaxman AD, Naghavi M, et al. Years lived with disability (YLDs) for 1160 sequelae of 289 diseases and injuries 1990–2010: A systematic analysis for the global burden of disease study 2010. Lancet. 2012;380:2163‐2196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Li X, Benjamin Ma C, Link TM, et al. In vivo T1ρ and T2 mapping of articular cartilage in osteoarthritis of the knee using 3 T MRI. Osteoarthr Cartil. 2007;15:789‐797. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Guermazi A, Alizai H, Crema MD, Trattnig S, Regatte RR, Roemer FW. Compositional MRI techniques for evaluation of cartilage degeneration in osteoarthritis. Osteoarthr Cartil. 2015;23:1639‐1653. [DOI] [PubMed] [Google Scholar]
  • 4. Gray ML, Eckstein F, Peterfy C, Dahlberg L, Kim Y‐J, Sorensen AG. Toward imaging biomarkers for osteoarthritis. Clin Orthop Relat Res. 2004;427:S175‐S181. [DOI] [PubMed] [Google Scholar]
  • 5. Cohen ZA, McCarthy DM, Kwak SDaniel, et al. Knee cartilage topography, thickness, and contact areas from MRI: In‐vitro calibration and in‐vivo measurements. Osteoarthr Cartil. 1999;7:95‐109. [DOI] [PubMed] [Google Scholar]
  • 6. Cicuttini F, Wluka A, Wang Y, Stuckey S. The determinants of change in patella cartilage volume in osteoarthritic knees. J Rheumatol. 2002;29:2615‐2619. [PubMed] [Google Scholar]
  • 7. Regatte RR, Akella SVS, Lonner JH, Kneeland JB, Reddy R. T1ρ relaxation mapping in human osteoarthritis (OA) cartilage: Comparison of T1ρ with T2. J Magn Reson Imaging. 2006;23:547‐553. [DOI] [PubMed] [Google Scholar]
  • 8. Yao W, Qu N, Lu Z, Yang S. The application of T1 and T2 relaxation time and magnetization transfer ratios to the early diagnosis of patellar cartilage osteoarthritis. Skeletal Radiol. 2009;38:1055‐1062. [DOI] [PubMed] [Google Scholar]
  • 9. Shiguetomi‐Medina JM, Ramirez‐GL JL, Stødkilde‐Jørgensen H, Møller‐Madsen B. Systematized water content calculation in cartilage using T1‐mapping MR estimations: Design and validation of a mathematical model. J Orthop Traumatol. 2017;18:217‐220. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Nishioka H, Nakamura E, Hirose J, Okamoto N, Yamabe S, Mizuta H. MRI T1ρ and T2 mapping for the assessment of articular cartilage changes in patients with medial knee osteoarthritis after hemicallotasis osteotomy. Bone Joint Res. 2016;5:294‐300. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11. Nishioka H, Hirose J, Nakamura E, et al. T1ρ and T2 mapping reveal the in vivo extracellular matrix of articular cartilage. J Magn Reson Imaging. 2012;35:147‐155. [DOI] [PubMed] [Google Scholar]
  • 12. Wang L, Chang G, Xu J, et al. T1rho MRI of menisci and cartilage in patients with osteoarthritis at 3T. Eur J Radiol. 2012;81:2329‐2336. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13. Gray ML, Burstein D, Xia Y. Biochemical (and functional) imaging of articular cartilage. Semin Musculoskelet Radiol. 2001;5:329‐344. [DOI] [PubMed] [Google Scholar]
  • 14. Zheng S, Xia Y. Multi‐components of T2 relaxation in ex vivo cartilage and tendon. J Magn Reson. 2009;198:188‐196. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Regatte RR, Akella SVS, Borthakur A, Kneeland JB, Reddy R. Proteoglycan depletion‐induced changes in transverse relaxation maps of cartilage: Comparison of T2 and T1ρ. Acad Radiol. 2002;9:1388‐1394. [DOI] [PubMed] [Google Scholar]
  • 16. Akella SVS, Regatte RR, Borthakur A, Kneeland JB, Leigh JS, Reddy R. T1ρ MR imaging of the human wrist in vivo. Acad Radiol. 2003;10:614‐619. [DOI] [PubMed] [Google Scholar]
  • 17. Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature. 2013;495:187‐192. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Badve C, Yu A, Rogers M, et al. Simultaneous T1 and T2 brain relaxometry in asymptomatic volunteers using magnetic resonance fingerprinting. Tomography. 2015;1:136‐144. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Badve C, Yu A, Dastmalchian S, et al. MR fingerprinting of adult brain tumors: initial experience. Am J Neuroradiol. 2017;38:492‐499. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Yu AC, Badve C, Ponsky LE, et al. Development of a combined MR fingerprinting and diffusion examination for prostate cancer. Radiology. 2017;283:729‐738. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Cloos MA, Knoll F, Zhao T, et al. Multiparametric imaging with heterogeneous radiofrequency fields. Nat Commun. 2016;7:1‐10. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Chen Y, Jiang Y, Pahwa S, et al. MR fingerprinting for rapid quantitative abdominal imaging. Radiology. 2016;279:278‐286. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23. Cloos MA, Assländer J, Abbas B, et al. Rapid radial T1 and T2 mapping of the hip articular cartilage with magnetic resonance fingerprinting. J Magn Reson Imaging. 2019;50:810‐815. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Hamilton JI, Jiang Y, Chen Y, et al. MR fingerprinting for rapid quantification of myocardial T 1, T 2, and proton spin density. Magn Reson Med. 2017;77:1446‐1458. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Winkelmann S, Schaeffter T, Koehler T, Eggers H, Doessel O. An optimal radial profile order based on the golden ratio for time‐resolved MRI. IEEE Trans Med Imaging. 2007;26:68‐76. [DOI] [PubMed] [Google Scholar]
  • 26. Mitrea BG, Krafft AJ, Song R, Loeffler RB, Hillenbrand CM. Paired self‐compensated spin‐lock preparation for improved T1ρ quantification. J Magn Reson. 2016;268:49‐57. [DOI] [PubMed] [Google Scholar]
  • 27. Yuan J, Wang YXJ. T1rho MR imaging principle, technology, and application In: Farncombe T, Iniewski K, eds. Medical Imaging. Boca Raton: CRC Press; 2014;565‐592 [Google Scholar]
  • 28. Sharafi A, Xia D, Chang G. Regatte RR. Biexponential T1ρ relaxation mapping of human knee cartilage in vivo at 3 T. NMR Biomed. 2017;30:e3760. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Weigel M. Extended phase graphs: Dephasing, RF pulses, and echoes—Pure and simple. J Magn Reson Imaging. 2015;41:266‐295. [DOI] [PubMed] [Google Scholar]
  • 30. McGivney DF, Pierre E, Ma D, et al. SVD compression for magnetic resonance fingerprinting in the time domain. IEEE Trans Med Imaging. 2014;33:2311‐2322. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31. Assländer J, Cloos MA, Knoll F, Sodickson DK, Hennig J, Lattanzi R. Low rank alternating direction method of multipliers reconstruction for MR fingerprinting. Magn Reson Med. 2018;79:83‐96. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Zhao BO, Setsompop K, Adalsteinsson E, et al. Improved magnetic resonance fingerprinting reconstruction with low‐rank and subspace modeling. Magn Reson Med. 2018;79:933‐942. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33. Doneva M, Amthor T, Koken P, Sommer K, Börnert P. Matrix completion‐based reconstruction for undersampled magnetic resonance fingerprinting data. Magn Reson Imaging. 2017;41:41‐52. [DOI] [PubMed] [Google Scholar]
  • 34. Mazor G, Weizman L, Tal A, Eldar YC. Low‐rank magnetic resonance fingerprinting. Med Phys. 2018;45:4066‐4084. [DOI] [PubMed] [Google Scholar]
  • 35. Zarins ZA, Bolbos RI, Pialat JB, et al. Cartilage and meniscus assessment using T1rho and T2 measurements in healthy subjects and patients with osteoarthritis. Osteoarthr Cartil. 2010;18:1408‐1416. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36. Stahl R, Luke A, Li X, et al. T1rho, T2 and focal knee cartilage abnormalities in physically active and sedentary healthy subjects versus early OA patients—A 3.0‐Tesla MRI study. Eur Radiol. 2009;19:132‐143. [DOI] [PubMed] [Google Scholar]
  • 37. Li X, Wyatt C, Rivoire J, et al. Simultaneous acquisition of T1ρ and T2 quantification in knee cartilage: Repeatability and diurnal variation. J Magn Reson Imaging. 2014;39:1287‐1293. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38. Schooler J, Kumar D, Nardo L, et al. Longitudinal evaluation of T1ρ and T2 spatial distribution in osteoarthritic and healthy medial knee cartilage. Osteoarthr Cartil. 2014;22:51‐62. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39. Hilbert T, Xia D, Block KT, et al. Magnetization transfer in magnetic resonance fingerprinting. Magn Reson Med. 2020;84:128‐141. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40. Sharafi A, Chang G, Regatte RR. Biexponential T2 relaxation estimation of human knee cartilage in vivo at 3T. J Magn Reson Imaging. 2018;47:809‐819. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Yu Z, Zhao T, Assländer J, Lattanzi R, Sodickson DK, Cloos MA. Exploring the sensitivity of magnetic resonance fingerprinting to motion. Magn Reson Imaging. 2018;54:241‐248. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42. Ma D, Coppo S, Chen Y, et al. Slice profile and B1 corrections in 2D magnetic resonance fingerprinting. Magn Reson Med. 2017;78:1781‐1789. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43. Ma D, Jones SE, Deshmane A, et al. Development of high‐resolution 3D MR fingerprinting for detection and characterization of epileptic lesions. J. Magn Reson Imaging. 2019;49:1333‐1346. [DOI] [PubMed] [Google Scholar]
  • 44. Ma D, Jiang Y, Chen Y, et al. Fast 3D magnetic resonance fingerprinting for a whole‐brain coverage. Magn Reson Med. 2018;79:2190‐2197. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45. Liao C, Bilgic B, Manhard MK, et al. 3D MR fingerprinting with accelerated stack‐of‐spirals and hybrid sliding‐window and GRAPPA reconstruction. Neuroimage. 2017;162:13‐22. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Fujimoto K, Cloos MA, Urushibata Y, Okada T. Cortical T1 mapping with 3D MR fingerprinting at 7T using a single transmit channel. In Proceedings of the 26th Annual Meeting of ISMRM, Paris, France, 2018. p. 1135. [Google Scholar]
  • 47. Zhao BO, Haldar JP, Liao C, et al. Optimal experiment design for magnetic resonance fingerprinting: Cramér‐rao bound meets spin dynamics. IEEE Trans Med Imaging. 2019;38:844‐861. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48. Cohen O, Zhu B, Rosen MS. MR fingerprinting deep reconstruction network (DRONE). Magn Reson Med. 2018;80:885‐894. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Magnetic Resonance in Medicine are provided here courtesy of Wiley

RESOURCES