Abstract
Purpose
It is widely accepted that atrial fibrillation (AF) accounts for half of arrhythmia recurrences following endocardial catheter ablation of AF. An epicardial-endocardial approach (hybrid) has emerged as an alternative to endocardial ablation alone for the treatment of AF, yet recurrence after a hybrid procedure has not been well characterized. This retrospective study is aimed at characterizing recurrence following hybrid ablation for patients with persistent AF.
Methods
Patients with persistent AF (N=108) received both endocardial and epicardial ablation of the posterior left atrial wall using catheter ablation and a small midline surgical approach (hybrid). Presence of atrial flutter or AF was determined with ambulatory monitoring (n=22) or electrocardiogram analysis (n=86) at each follow-up visit. Recurrence mode was confirmed by electrophysiology study for those patients undergoing subsequent catheter ablation after hybrid ablation.
Results
Patients were followed for a mean ± standard deviation of 25 ± 14 months. Of patients who had a recurrence, 53% (n=33) were in atrial flutter and 47% (n=29) were in AF. Of those who had a recurrence with atrial flutter, 14 received repeat ablation for either left (n=11) or left/right (n=3) atrial flutter and 3 received AF ablation. Half of ablations for atrial flutter recurrence following the hybrid procedure involved the mitral isthmus.
Conclusions
Atrial flutter accounts for about half of arrhythmia recurrences post-hybrid ablation. If catheter ablation of the mitral isthmus is considered during the hybrid procedure to prevent subsequent occurrence of perimitral flutter, bidirectional block must be performed to ensure a complete line of block.
Keywords: atrial fibrillation, atrial flutter, catheter ablation, hybrid, surgical ablation, mitral isthmus, recurrence
It is estimated that there are 33 million people worldwide living with atrial fibrillation (AF).1 AF has been shown to increase the risk of stroke,2 sudden death,3 and heart failure,4 and patients with AF suffer from a lower quality of life.5 Antiarrhythmic drug therapy is an option for treating patients with paroxysmal AF, but for patients in persistent or long-standing persistent AF, catheter ablation is effective6–8 and may reverse left atrial (LA) electrophysiological remodeling and improve left ventricular ejection fraction (LVEF).9
Studies have shown that pulmonary vein (PV) isolation (PVI) is effective in treating paroxysmal AF by targeting the AF triggers in the PV region.10,11 Yet, it has been shown that patients in persistent and long-standing persistent AF respond more favorably to an approach that includes PVI plus further substrate modification of the posterior LA wall.12
The surgical Cox-maze procedure was developed in the late 1980s as a technique to interrupt electrical impulses and restore sinus rhythm,13,14 and evolution of this approach has led to the most recent version of the procedure, Cox-maze IV, which uses bipolar radiofrequency energy to simplify and shorten the procedure.15 While these procedures are highly effective in treating AF, they typically require pleural access and lack the option of using electrophysiology study to uncover the presence of atrial tachycardias (AT) that may exist in the patient.
Alternatively, Wolf et al16 and Edgerton et al17 described effective and minimally invasive modified surgical ablation techniques using pleural access epicardial isolation of the PVs; however, the FAST trial demonstrated that, although the procedure is highly effective, surgical ablation had a significantly higher procedural adverse event rate than did catheter ablation.18
A minimally invasive convergent, or hybrid, approach has been developed that utilizes both epicardial ablation of the posterior surface of the LA as well as transvenous endocardial catheter ablation and has been demonstrated to be effective in the treatment of AF.19 The hybrid approach was shown to be successful in the challenging setting of persistent or long-standing persistent AF, and a study by our center demonstrated that patients receiving the hybrid procedure had fewer repeat ablations and less recurrence at 16 months.20 It has been shown that atrial flutter (AFL) encompasses approximately half of recurrences when simple endocardial ablation techniques are used;21–23 furthermore, treating recurring AF in this population can be relatively challenging.24 Thus, our center sought to characterize arrhythmia recurrence after hybrid ablation in the setting of persistent and long-standing persistent AF.
METHODS
Study Population
The local institutional review board reviewed this protocol and determined it to be exempt status. Data from 108 consecutive patients undergoing the hybrid procedure with and without prior endocardial catheter ablation from June 2010 through March 2016 were retrospectively collected. All patients had persistent or long-standing persistent AF, as defined by the 2017 expert consensus statement on AF ablation,21 for a median of 12 (interquartile range: 6–34) months. Mean age was 61 ± 10 years, with a mean follow-up of 25 ± 14 months. Baseline characteristics are summarized in Table 1.
Table 1.
Baseline Characteristics
Characteristic | Overall (N=108) | With prior ablation (n=60) | Without prior ablation (n=48) | P |
---|---|---|---|---|
Age | 61.1 ± 9.7 years | 60.7 ± 8.5 years | 61.6 ± 10.9 years | 0.448 |
Sex | 0.654 | |||
Female | 14 (13.0%) | 7 (11.7%) | 7 (16.6%) | |
Male | 94 (87.0%) | 53 (88.3%) | 41 (85.4%) | |
Body mass index | 34.6 ± 6.2 kg/m2 | 33.2 ± 5.3 kg/m2 | 36.2 ± 6.9 kg/m2 | 0.016 |
Hypertension | 70 (64.8%) | 36 (60.0%) | 34 (70.8%) | 0.241 |
Coronary artery disease | 22 (20.4%) | 9 (15.0%) | 13 (27.1%) | 0.121 |
Diabetes mellitus | 15 (13.9%) | 7 (11.7%) | 8 (16.7%) | 0.455 |
Prior stroke, TIA, or thromboembolism | 10 (9.3%) | 5 (8.3%) | 5 (10.4%) | 0.711 |
CHA2DS2-VASc score | 1.8 ± 1.2 | 1.5 ± 1.1 | 2.1 ± 1.3 | 0.010 |
LVEF | 54.7% ± 10.8% | 55.5% ± 9.6% | 53.8% ± 12.2% | 0.707 |
Left atrial size | 5.2 ± 0.7 cm | 5.0 ± 0.7 cm | 5.3 ± 0.7 cm | 0.060 |
Duration of persistent AF | 12 (6–34) months | 12 (6–36) months | 12 (6–34) months | 0.582 |
Values are reported as n (%), mean ± standard deviation, or median (interquartile range).
AF, atrial fibrillation; LVEF, left ventricular ejection fraction; TIA, transient ischemic attack.
Anesthesia and Monitoring
Anesthesia and monitoring were performed as previously described.20 General anesthesia was induced in all patients in the electrophysiology operating room. Hemodynamic monitoring was performed throughout the procedure, and transesophageal echocardiography was used to exclude thrombus in the LA appendage. Esophageal temperature monitoring was used throughout the entire procedure.
Hybrid Ablation
Patients undergoing the hybrid procedure had epicardial ablation concomitant with endocardial ablation, as described hereafter. Patients underwent epicardial ablation of the posterior LA wall using a transabdominal endoscopic approach followed by endocardial catheter ablation, with the goal of posterior wall isolation.20 Briefly, a 12-mm port through a 2-cm midline incision and two 5-mm ports were placed for instruments and carbon dioxide insufflation. A 30- or 40-cm cannula (Subtle, AtriCure, Inc.) was placed through a 2-cm opening in the central tendon of the diaphragm and guided into the oblique sinus with a 6.5-mm scope. A preablation epicardial map of the posterior LA and proximal PV trunks was created for all patients after March 2014 using a 3D mapping system (CARTO 3®, Biosense Webster, Inc.) under direct visualization. Each lesion was delivered with a 3-cm probe (Numeris®, AtriCure), which was saline-irrigated and vacuum-attached, at 30 watts for 90 seconds. All lesions were delivered with cold saline infusion within the pericardial sac. Lesions were repeated, if necessary, to achieve a cumulative delivery of 2700 Joules. Esophageal temperature was monitored to avoid an increase of more than 0.5°C during any lesion. In the event of temperature rise exceeding 0.5°C, the power was interrupted sufficiently long for the temperature to return to baseline, often with the instillation of additional cold saline into the pericardial sac. The entire posterior LA from the right PV trunk to the left PV trunk was ablated with 2 rows of parallel overlapping lesions (Figure 1). The anterior left inferior PV trunk was ablated. Lesions were placed in the recess between the inferior vena cava and right inferior PV in some patients. A new acquisition of voltage signals was completed to create a post-epicardial ablation voltage map in those patients with pre-epicardial ablation mapping. Areas that failed to show electrical silence underwent repeat ablation to achieve a silent final postablation map. A 19-mm drain (BLAKE®, Ethicon, Inc.) was placed into the pericardial cavity. The abdominal incisions were closed in a standard fashion.
Figure 1.
Cutaway view of the left and right atria of the heart. CTI, cavotricuspid isthmus; LIPV, left inferior pulmonary vein; LSPV, left superior pulmonary vein; RIPV, right inferior pulmonary vein; RSPV, right superior pulmonary vein.
Immediately following epicardial ablation, endocardial ablation was performed as previously described by Kress et al.20 In brief, after obtaining venous access, a 20-pole diagnostic catheter (St. Jude Medical) was advanced under fluoroscopic guidance into the coronary sinus. Diagnostic quadripolar catheters were positioned at the right atrial appendage and His bundle positions. Upon completion of all sheath insertions, 5000 units of heparin was given. An additional 5000 units was given via the transseptal sheath after puncture. During the catheter ablation procedure, infusion of heparin was maintained to achieve an activated clotting time of 300 to 350 seconds.
Fluoroscopy and either transesophageal or intracardiac echocardiography were performed to guide the sheath (PREFACE®, Medtronic) for transseptal puncture. An esophageal probe was placed to monitor temperature throughout the entire endocardial ablation procedure.
Using the CARTO 3D mapping system, maps of the left and right atria were created for identifying fractionated electrograms and any AT or AFL.
The PVs were checked for electrical activity and isolated as needed. If the PVs were silent from previous ablation, PVI was not performed. PVI, when needed, was achieved using either conventional catheter ablation or the cryoballoon (Medtronic), and isolation of the PVs was confirmed with a mapping and ablation catheter. Radiofrequency catheter ablation was used to ablate complex fractionated atrial electrograms in the LA if they were present, and linear lesions were applied at the discretion of the physician performing the catheter ablation if a reentrant tachycardia occurred during the procedure. Linear ablation involved some or all of the following: a line connecting the mitral annulus to the left superior PV, the roofline connecting the superior veins, the mitral annulus to the right superior PV, and the cavotricuspid isthmus line on the right side of the atrium. Ablation in the coronary sinus was performed in patients not initially demonstrating conduction block across the mitral isthmus line or having fractionated electrogram recordings in the coronary sinus. Sinus rhythm was the procedural endpoint in all cases. Ablation of the most common lesions is illustrated in Figure 1.
Demonstration of conduction block was attempted in all linear lesions and confirmed during pacing from the LA appendage for both mitral isthmus and rooflines25 as well as through CARTO 3D analysis of the scar. If AT or AFL developed at any point during the initial ablation procedure, mapping and ablation of the AT/AFL was performed and conduction block confirmed across the linear lesion. For all patients, if tachycardia persisted following delivery of all lesions, patients were converted to sinus rhythm using intravenous infusion of ibutilide (1 mg over 10 minutes) and/or direct-current cardioversion. Heparin was partially reversed with protamine following the procedure.
Follow-up Visits
Patient follow-up was conducted as part of the standard of care, as previously described.20 Data related to procedure complications occurring within 30 days of ablation, as defined by 2017 expert consensus,21 were collected. Cardiac symptom data and need for antiarrhythmic drugs were collected from each patient follow-up visit. Nonambulatory and/or ambulatory data were collected from each follow-up visit or device interrogation following the hybrid procedure. Typically, patients were followed at 1, 3, 6, 9, and 12 months, or any time symptoms were noted.
Detectable AF/AFL/AT was evaluated with a minimum of a 30-second electrocardiogram recording or rhythm strip or with remote and/or in-person ambulatory monitoring, including pacemaker interrogation, loop recorder, activated clotting time monitor, and Holter monitor. Recurrence was defined as documented AF, AFL, or AT occurring after the 3-month blanking period or a repeat ablation or atrioventricular junction ablation with permanent pacemaker at any time throughout the follow-up period, regardless of antiarrhythmic drug use. Initial intervention for recurrence was treated at the physician’s discretion based on symptoms, patient history, and clinical condition.
Characterizing Recurrence and Repeat Ablation Following the Hybrid Procedure
If repeat ablation was not the treatment for recurrence, diagnosis of AF, AFL, or AT was based off monitoring device records reviewed at the time of recurrence. For patients demonstrating both AF and AFL, recurrence was treated statistically as AF for the purposes of this study. For patients having an electrophysiology study/repeat ablation following the hybrid procedure, cardiac mapping, entrainment, and pacing maneuvers were used to characterize the anatomical location of the AFL during the procedure. Upon repeat ablation, the PVs were checked for electrical isolation using mapping and catheter techniques for every repeat ablation procedure. During repeat ablation, the PVs found to have electrical activity were isolated by either radiofrequency or cryoballoon.
Statistical Analysis
SAS 9.4 software (SAS Institute Inc.) was used for statistical analysis. Continuous variables are reported as mean ± standard deviation or median (interquartile range), as appropriate. Categorical variables are presented as frequency count (percentage). Comparisons of characteristics for patients who had prior ablation versus no prior ablation were made using the Wilcoxon rank-sum test for continuous variables and chi-squared or Fisher’s exact tests for categorical variables. Comparisons of the time to recurrence for patients with and without implantable devices were made with a Student’s paired t-test.
RESULTS
Procedural Results
Procedural results are outlined in Table 2. In summary, the median total hybrid procedure time for both epicardial and endocardial ablation was 308 (278–352) minutes. Overall, 85.2% of patients required endocardial PVI during the hybrid procedure, and those with prior ablation required significantly less endocardial PVI than those without prior ablation (75.0% vs 97.9%; P<0.001). Patients with ablation prior to the hybrid procedure also required cardioversion to restore sinus rhythm during the procedure significantly less often than did those with no prior ablation (26.7% vs 45.8%; P=0.038). Endocardial linear lesions, including the mitral isthmus line (67.6%), roofline (52.8%), mitral annulus to right superior PV (19.4%), and cavotricuspid isthmus line (38.0%), were performed in 95 (88%) patients.
Table 2.
Procedural Characteristics
Procedural data | Overall (N=108) | With prior ablation (n=60) | Without prior ablation (n=48) | P |
---|---|---|---|---|
Endocardial PVI required | 92 (85.2%) | 45 (75.0%) | 47 (97.9%) | <0.001 |
Cryoballoon ablation used | 78 (72.2%) | 36 (60.0%) | 42 (87.5%) | 0.002 |
Linear lesions | 95 (88.0%) | 53 (88.3%) | 42 (87.5%) | 0.894 |
Roofline | 57 (52.8%) | 32 (53.3%) | 25 (52.1%) | 0.897 |
Mitral isthmus | 73 (67.6%) | 40 (66.7%) | 33 (68.8%) | 0.818 |
Cavotricuspid isthmus line | 41 (38.0%) | 22 (36.7%) | 19 (39.6%) | 0.756 |
Mitral annulus to RSPV | 21 (19.4%) | 11 (18.3%) | 10 (20.8%) | 0.744 |
DCCV required | 38 (35.2%) | 16 (26.7%) | 22 (45.8%) | 0.038 |
Ibutilide required | 33 (30.6%) | 16 (26.7%) | 17 (35.4%) | 0.327 |
Procedural complications | 12 (11.1%) | 7 (11.7%) | 5 (10.4%) | 0.837 |
Radiation dose, mGy | 573.5 (289–932) | 469.0 (260–886) | 597.0 (322–1129) | 0.190 |
Time, minutes | ||||
Cryoballoon ablation | 26.5 (0–32) | 23.5 (0–32) | 27.5 (21–32) | 0.164 |
Radiofrequency ablation | 24.6 (16–43.2) | 28.6 (14.5–50.8) | 24.2 (16.7–38.3) | 0.208 |
Total procedure | 308.0 (278–352) | 308.0 (268–346) | 308.0 (278.5–355.5) | 0.694 |
Fluoroscopy | 54.0 (43.2–73.8) | 53.0 (40.2–76.6) | 55.6 (44.2–68.1) | 0.861 |
Values are reported as n (%) or median (interquartile range).
DCCV, direct-current cardioversion; PVI, pulmonary vein isolation; RSPV, right superior pulmonary vein.
Time to Recurrence
Patients were monitored for recurrence following the hybrid procedure using either ambulatory (n=22) or nonambulatory (n=86) methods. Of the ambulatory methods used, 18 patients had continuous monitoring for more than 48 hours with either permanent pacemaker (n=9), implantable loop recorder (n=6), or implantable cardioverter-defibrillator (n=3).
In the overall population, mean days to recurrence was 396 ± 324. Patients who received continuous ambulatory monitoring for their arrhythmia following the hybrid ablation were also studied separately to help aid the analysis. Of the 18 patients who had continuous monitoring with permanent pacemaker, implantable loop recorder, or implantable cardioverter-defibrillator, 14 had recurrence as defined by more than 30 seconds of arrhythmia at some point in their follow-up.
Patients with continuous implantable devices (n=18) recurred, on average, at 265 ± 226 days, and those without continuous monitoring from an implantable device (n=90) recurred at a mean of 434 ± 340 days. There was no significant difference between the time to recurrence in patients with and without an implantable device (P=0.9).
Rate of 12-Month Recurrence and Effect of Prior Ablation on Outcome
Overall, 64.2% of patients had arrhythmia-free survival at 12 months, as defined by no AF, AFL, or AT. Of all the patients who underwent the hybrid procedure, 57.4% were on an antiarrhythmic drug at 12 months. There was no significant difference in arrhythmia-free survival among patients with and without ablation prior to the hybrid procedure (Figure 2). The 12-month arrhythmia-free survival estimate following hybrid ablation and up to 1 additional ablation for AFL was 70.0% (Figure 3). There also was no significant difference in estimated AF-free survival among patients with and without prior ablation (83.0% vs 82.0%). Overall, AF-free survival was estimated to be 82.6% at 12 months (Figure 4).
Figure 2.
Kaplan-Meier curve for 12-month arrhythmia-free (atrial fibrillation/flutter/tachycardia) survival for a single hybrid ablation procedure. HR, hazard ratio.
Figure 3.
Kaplan-Meier curve for 12-month arrhythmia-free survival following hybrid ablation, allowing for 1 additional atrial flutter ablation. The first occurrence of flutter was not counted as an event if it was successfully ablated in a single additional procedure. HR, hazard ratio.
Figure 4.
Kaplan-Meier curve for 12-month atrial fibrillation (AF)-free survival following a single hybrid ablation procedure. HR, hazard ratio.
Mode of Recurrence and Initial Intervention Following Hybrid Ablation
Following hybrid ablation and spanning the entirety of the follow-up period, 53.2% (n=33) of patients who recurred at any time were in AFL and 46.8% (n=29) were in AF. Of these 62 total patients who had recurrence over some period during their follow-up, 66.1% (n=41) were on an antiarrhythmic drug at the time of recurrence. Looking at the subpopulation of those who were monitored with implantable devices, of the patients who recurred at any time (n=14), 8 (57.1%) recurred with AF and 6 (42.9%) with AFL. Similarly, of those without implantable devices who had recurrence (n=48), 21 (43.8%) had AF and 27 (56.3%) had AFL.
Initial intervention for recurrence of both AF and AFL is outlined in Figure 5. The initial intervention for AF and AFL patients, respectively, was as follows: 37.9% (n=11) and 24.2% (n=8) medication change; 20.7% (n=6) and 30.3% (n=10) direct-current cardioversion; 6.9% (n=2) and 27.3% (n=9) repeat ablation; 3.4% (n=1) and 6.1% (n=2) rate control; 10.4% (n=3) and 0% atrioventricular junction ablation; and 20.7% (n=6) and 12.1% (n=4) no intervention.
Figure 5.
Initial intervention for recurrences following the hybrid ablation procedure. AF, atrial fibrillation; AFL, atrial flutter; AVJ, atrioventricular junction; DCCV, direct-current cardioversion.
Repeat Ablations/Electrophysiology Study for AFL
Of the 14 patients undergoing a repeat ablation for their AFL at any time following the hybrid procedure, all were found to have a left-sided flutter component (Table 3). Of the 14 confirmed AFL recurrences, 9 (64.3%) occurred in patients with a catheter ablation prior to the hybrid procedure. All the catheter ablations prior to the hybrid procedure were for AF and had a PVI component except for one, which was performed owing to right-sided AFL. Seven of the confirmed 14 flutters involved the perimitral isthmus (as defined by Matsuo et al26). Of the 7 confirmed AFL recurrences that contained a perimitral component, 6 (85.7%) had the mitral isthmus line as part of the hybrid ablation procedure. Other areas of AFL occurrence included the LA roof (n=3), the left superior PV and base of the LA appendage (n=1), and the LA posterior wall (n=1); 1 case had multiple locations in the LA.
Table 3.
Mode of Recurrent AFL Ablation in Patients With and Without Ablation Prior to Hybrid Procedure Who Required Repeat Ablation Subsequent to the Hybrid Procedure
Repeat AFL ablation location* | Overall (N=14) | With prior ablation (n=9) | Without prior ablation (n=5) |
---|---|---|---|
Perimitral, n (%) | 4 (29%) | 3 (75%) | 1 (25%) |
Left atrial roof, n (%) | 3 (22%) | 3 (100%) | 0 (0%) |
Perimitral and right atrium, n (%) | 2 (14%) | 0 (0%) | 2 (100%) |
Perimitral and left atrial roof, n (%) | 1 (7%) | 1 (100%) | 0 (0%) |
LSPV and LAA base, n (%) | 1 (7%) | 1 (100%) | 0 (0%) |
Posterior wall, n (%) | 1 (7%) | 1 (100%) | 0 (0%) |
Multiple left, n (%) | 2 (14%) | 0 (0%) | 2 (100%) |
Location of AFL was determined by electrophysiology study conducted during the repeat AFL ablation. This table represents AFL ablations that followed the hybrid ablation, regardless of timing of the follow-up intervention.
AFL, atrial flutter; LAA, left atrial appendage; LSPV, left superior pulmonary vein.
Two patients (14.3%) requiring repeat ablation and electrophysiology study for AFL required some additional radiofrequency ablation to re-isolate PVs that demonstrated electrical activity, and 12 patients (85.7%) demonstrated silent PVs upon repeat ablation and electrophysiology study. The posterior wall was silent in 8 of these 14 patients (57.1%), as demonstrated by CARTO 3D, and 6 patients (42.9%) demonstrated areas of fractionated electrograms on their posterior wall that required catheter ablation.
Repeat Ablations/Electrophysiology Study for AF
Of the 3 patients who underwent a repeat ablation for their AF at any time following the hybrid procedure, 1 patient required radiofrequency ablation of the right inferior PV to re-isolate the vein. Other veins previously isolated in this patient showed silence according to CARTO 3D. This same patient required additional ablation in the left septal region anterior to the right superior PV and LA appendage.
The PVs of the other 2 patients who had repeat AF ablation and electrophysiology study demonstrated silence. In one case, the posterior wall demonstrated mainly scar, but the LA required catheter ablation for the roofline, lateral isthmus line, and multiple fractionated areas. This patient had no ablation prior to the hybrid procedure. The roofline and mitral isthmus line were completed as part of the hybrid procedure. The other repeat ablation also had electrically silent PVs from hybrid ablation but required ablation of the mitral isthmus line, right superior PV region, roofline, anterior mitral line from the right superior PV to the roof, and the cavotricuspid isthmus line. This patient had no ablation prior to the hybrid procedure.
Complications
There were 8 complications noted in the study cohort. Three patients suffered from groin complication as follows: 1 received a right femoral embolectomy, 1 had a right groin bleed within 48 hours of the procedure, and 1 had a pseudoaneurysm of the groin requiring thrombin injection. One patient suffered from pericarditis and a nonocclusive pulmonary embolus in 1 of the left pulmonary branches. There were 2 late tamponade/pericardial effusions, 1 cerebrovascular accident, and 1 death due to a gastrointestinal bleed 2 weeks following the procedure. The patient was on warfarin at the time of the bleed.
DISCUSSION
In this retrospective analysis, we aimed to describe the manner with which recurrence occurred following the minimally invasive convergent, or hybrid, ablation procedure for persistent and long-standing persistent AF. This series was performed in a high-volume center with experienced operators. The population included patients both with and without catheter ablation prior to the hybrid procedure. Although data are reported as a heterogeneous population using either radiofrequency or cryoballoon ablation to isolate the PVs as part of the endocardial procedure, in a randomized multicenter study, the cryoballoon was found noninferior to radiofrequency ablation in PVI.27 All baseline characteristics were similar except the patients who had no prior ablation had a significantly higher CHA2DS2-VASc score, putting them at a higher risk of stroke prior to the hybrid ablation compared with patients who did undergo prior ablation.
Time to Recurrence
Overall mean time to recurrence was slightly over 12 months for the entire population. Although not statistically significant, patients with continuous monitoring of >48 hours had a shorter time to recurrence compared with those with no ambulatory monitoring, most likely owing to the ability of continuous monitoring to pick up small disturbances in rhythm. Since this study defines arrhythmia of >30 seconds as recurrence, there is a greater likelihood of having a reportable recurrence with continuous monitoring, although this number was not statistically significant. Analysis, especially that which is retrospective in nature, is oftentimes limited by follow-up monitoring tools, and there may be concern regarding nonambulatory monitoring results. Yet, there is also a challenge when studying those with continuous monitoring, owing to the incessant nature of persistent AF. While the 2017 expert consensus statement recommends centers report recurrence as anything exceeding 30 seconds of AF, they also encourage centers to report other endpoints, such as AF burden reduction or quality of life. This is due to the oftentimes noncurative nature of persistent AF and the fact that Kaplan-Meier curve analysis of this strict definition could underestimate the benefit of ablation for persistent AF.21
Arrhythmia-Free Survival Outcomes
Patients with persistent or long-standing persistent AF may be a challenging population to treat with ablation. Wokhlu et al have described that persistent AF is a predictor of overall recurrence and very late recurrence.24 In a large randomized trial of 589 patients, Verma et al described 49% and 46% freedom from AF with PVI and complex fractionated atrial electrograms or PVI and linear lesions, respectively.28 At our center, the overall AF-free survival at 12 months was 82.6% and the 12-month arrhythmia-free survival, including both AF and AFL recurrence, following hybrid ablation was 64.2%. Allowing for 1 additional ablation for AFL following the hybrid procedure, the 12-month arrhythmia-free survival was 70% in this cohort. Our results are noteworthy for persistent and long-standing persistent AF patients given the lower predictive value of nonpersistent AF and lower rates of ablation success in prior studies.29,30 Our analysis of 12-month outcomes also aligns with recent similar studies by de Asmundis et al and Maesen et al.22,23 This is important, as we are a large-volume center with experienced operators performing the hybrid procedure and our population reported is roughly double that of both similar papers, further validating their results. We also have the additional benefit of prior ablation Kaplan-Meier curve analysis, but we caution that the numbers may not be high enough to see a difference statistically between the two groups, as discussed henceforth.
Effect of Prior Ablation on Acute Hybrid Procedure Outcome
Those patients with prior ablation required significantly less frequent endocardial PVI during the hybrid procedure. This observation is most likely due to previously isolated PVs from the prior ablation or from the epicardial ablation portion of the hybrid procedure — most likely the isolation occurred with the former since epicardial ablation using this technique does not aim to isolate the PVs. Since pre- and postablation epicardial mapping did not occur until later in the data set, it is difficult to establish whether the PVs were isolated from previous ablation or with epicardial ablation itself. Patients with prior ablation also converted to sinus rhythm with ablation during the hybrid procedure significantly more frequently than those without prior ablation. This may suggest that prior ablation has a cumulative effect on outcomes, as seen with Maesen et al,23 but this requires further analysis.
Results of this study show that ablation prior to the hybrid procedure does not affect the probability of recurrence-free outcome at 12 months in this population. This finding should be studied further in a larger cohort to assess the importance of prior ablation on hybrid procedure outcomes.
Mode of Recurrence and Initial Intervention Following Hybrid Ablation
We found that slightly over half of all recurrences following hybrid ablation were AFL, which is similar to the statement made in the 2017 expert consensus document: “ATs of new onset make up to 50% of all arrhythmias observed following catheter-based ablation of AF.”21 From our results, it seems that recurrence following hybrid ablation does not largely involve reconnection of the PVs.
Studies have shown that AFL recurrences following endocardial ablation often involve focal AT originating from PV ostial reconnection or macroreentrant flutters occurring around anatomical scarring of the LA.31,32 In the case of macroreentrant flutters, Chae and colleagues found that 42% of recurrent ATs could be ablated at the mitral isthmus, with the LA roof and septum accounting for 20% and 13% of reentrant tachycardia, respectively.32 It is suggested that these common postablation recurrences could be reduced or prevented by demonstration of complete block or avoidance of linear lesions altogether.32 A recent similar study from de Asmundis and colleagues reported that about half of the AFL recurrences involve the mitral isthmus line.22 Similarly, at our center, AFL accounted for 53% of recurrences, and of those, 50% were perimitral or had a perimitral component, verifying further that the mitral isthmus is a difficult area to ablate.
The subpopulation analysis of those who had a recurrence of AFL or AF with or without continuous monitoring looked like that of the overall population, suggesting that although the study lacks a large population with continuous monitoring, results may have been similar.
Overall, initial intervention for AF recurrence relied more heavily on medication changes, atrioventricular junction ablation, or no intervention than did the AFL recurrences. AFL recurrences had more direct-current cardioversion, repeat ablation, and rate control methods as their initial intervention. This may be due to the stringent definition of recurrence that was used in the study and the patient’s ability to convert spontaneously, with a dose change or with a change in antiarrhythmic medication, while in AF for a very short period. Ablation of AFL may have been used more often because of the greater ability to treat the recurrence with a repeat ablation using linear lesions. Most recurrences following the hybrid procedure did not require PVI because the PVs were deemed silent. Many patients receiving an ablation following the procedure received radiofrequency ablation for complex fractionated atrial electrograms seen in the LA on electrophysiology study and/or AFL.
Our data suggest that during the hybrid procedure, if the endocardial mitral isthmus line is created, conduction block must be confirmed to optimize its postprocedural scar transmurality and gap absence. This will reduce the incidence of developing new perimitral AFL. If there is a history of previous ablation of the mitral isthmus, the scar should be evaluated for conduction gaps, and ablation to close the gaps should be performed during the endocardial portion of the procedure. There were multiple flutters seen in 6 of the 14 patients with repeat AFL ablation (42.9%) following the hybrid procedure, emphasizing the importance of entrainment mapping of multiple flutters upon recurrence as suggested by Chae et al.32
Limitations
The study was retrospective in nature, and ambulatory monitoring was used in a minority of the patients. The electrophysiological evaluation of postablation arrhythmias was limited to patients deemed symptomatic enough to undergo a follow-up catheter ablation and therefore did not include less symptomatic patients. Detection of postablation arrhythmias was done at follow-up intervals that were based on standard of care, emergency department, or acute care visits.
CONCLUSIONS
Atrial flutter accounts for about half of arrhythmia recurrences in the first year following a hybrid ablation procedure. Knowledge of the recurrence mechanism is important for physicians to consider when treating a patient with this procedure so that perimitral and other types of flutter might be avoided.
Patient-Friendly Recap.
Hybrid ablation, or ablation of the heart’s left atrium using both catheter and surgical techniques, has emerged as an option for treating persistent atrial fibrillation, a complicated cardiac condition.
After analyzing past cases, the authors found that in patients who experience a recurrent arrhythmia following the hybrid procedure, half have what is called atrial flutter and half of those flutters stem from the mitral isthmus region of the left atrium.
Having knowledge of the anatomical site where recurrence is likely to originate allows physicians to consider that region when performing hybrid ablation, especially in those patients who have undergone a previous ablation procedure.
Acknowledgments
The authors thank Jennifer Pfaff and Susan Nord of Aurora Cardiovascular and Thoracic Services for editorial preparation of the manuscript and Brian Schurrer and Brian Miller of Aurora Cardiovascular and Thoracic Services for assistance with the figures.
Footnotes
Author Contributions
Study design: Kress, Erickson, Mengesha, Sra. Data acquisition or analysis: Erickson, Mengesha. Manuscript drafting: Kress, Erickson, Krum, Sra. Critical revision: Kress, Erickson.
Conflicts of Interest
Dr. Kress served as chair of the Data and Safety Monitoring Board for the Converge trial (sponsor: nContact Inc. and, subsequently, AtriCure) during the years 2014–2018. He was a consultant with Clinipace Inc. to serve in this role, not of the sponsors.
References
- 1.Chugh SS, Havmoeller R, Narayanan K, et al. Worldwide epidemiology of atrial fibrillation: a Global Burden of Disease 2010 Study. Circulation. 2014;129:837–47. doi: 10.1161/CIRCULATIONAHA.113.005119. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Kannel WB, Wolf PA, Benjamin EJ, Levy D. Prevalence, incidence, prognosis, and predisposing conditions for atrial fibrillation: population-based estimates. Am J Cardiol. 1998;82(8A):2N–9N. doi: 10.1016/s0002-9149(98)00583-9. [DOI] [PubMed] [Google Scholar]
- 3.Chen LY, Benditt DG, Alonso A. Atrial fibrillation and its association with sudden cardiac death. Circ J. 2014;78:2588–93. doi: 10.1253/circj.CJ-14-0814. [DOI] [PubMed] [Google Scholar]
- 4.Wang TJ, Larson MG, Levy D, et al. Temporal relations of atrial fibrillation and congestive heart failure and their joint influence on mortality: the Framingham Heart Study. Circulation. 2003;107:2920–5. doi: 10.1161/01.CIR.0000072767.89944.6E. [DOI] [PubMed] [Google Scholar]
- 5.Dorian P, Jung W, Newman D, et al. The impairment of health-related quality of life in patients with intermittent atrial fibrillation: implications for the assessment of investigational therapy. J Am Coll Cardiol. 2000;36:1303–9. doi: 10.1016/S0735-1097(00)00886-X. [DOI] [PubMed] [Google Scholar]
- 6.Parkash R, Tang AS, Sapp JL, Wells G. Approach to the catheter ablation technique of paroxysmal and persistent atrial fibrillation: a meta-analysis of the randomized controlled trials. J Cardiovasc Electrophysiol. 2011;22:729–38. doi: 10.1111/j.1540-8167.2011.02010.x. [DOI] [PubMed] [Google Scholar]
- 7.Oral H, Pappone C, Chugh A, et al. Circumferential pulmonary-vein ablation for chronic atrial fibrillation. N Engl J Med. 2006;354:934–41. doi: 10.1056/NEJMoa050955. [DOI] [PubMed] [Google Scholar]
- 8.Mont L, Bisbal F, Hernández-Madrid A, et al. Catheter ablation vs. antiarrhythmic drug treatment of persistent atrial fibrillation: a multicentre, randomized, controlled trial (SARA study) Eur Heart J. 2014;35:501–7. doi: 10.1093/eurheartj/eht457. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Pump A, Di Biase L, Price J, et al. Efficacy of catheter ablation in nonparoxysmal atrial fibrillation patients with severe enlarged left atrium and its impact on left atrial structural remodeling. J Cardiovasc Electrophysiol. 2013;24:1224–31. doi: 10.1111/jce.12253. [DOI] [PubMed] [Google Scholar]
- 10.Haïssaguerre M, Jaïs P, Shah DC, et al. Spontaneous initiation of atrial fibrillation by ectopic beats originating in the pulmonary veins. N Engl J Med. 1998;339:659–66. doi: 10.1056/NEJM199809033391003. [DOI] [PubMed] [Google Scholar]
- 11.Packer DL, Kowal RC, Wheelan KR, et al. Cryoballoon ablation of pulmonary veins for paroxysmal atrial fibrillation: first results of the North American Arctic Front (STOP AF) pivotal trial. J Am Coll Cardiol. 2013;61:1713–23. doi: 10.1016/j.jacc.2012.11.064. [DOI] [PubMed] [Google Scholar]
- 12.Haïssaguerre M, Hocini M, Sanders P, et al. Catheter ablation of long-lasting persistent atrial fibrillation: clinical outcome and mechanisms of subsequent arrhythmias. J Cardiovasc Electrophysiol. 2005;16:1138–47. doi: 10.1111/j.1540-8167.2005.00308.x. [DOI] [PubMed] [Google Scholar]
- 13.Cox JL, Canavan TE, Schuessler RB, et al. The surgical treatment of atrial fibrillation. II. Intraoperative electrophysiologic mapping and description of the electrophysiologic basis of atrial flutter and atrial fibrillation. J Thorac Cardiovasc Surg. 1991;101:406–26. doi: 10.1016/S0022-5223(19)36723-6. [DOI] [PubMed] [Google Scholar]
- 14.Cox JL, Schuessler RB, Boineau JP. The surgical treatment of atrial fibrillation. I. Summary of the current concepts of the mechanisms of atrial flutter and atrial fibrillation. J Thorac Cardiovasc Surg. 1991;101:402–5. doi: 10.1016/S0022-5223(19)36722-4. [DOI] [PubMed] [Google Scholar]
- 15.Gaynor SL, Diodato MD, Prasad SM, et al. A prospective, single-center clinical trial of a modified Cox maze procedure with bipolar radiofrequency ablation. J Thorac Cardiovasc Surg. 2004;128:535–42. doi: 10.1016/j.jtcvs.2004.02.044. [DOI] [PubMed] [Google Scholar]
- 16.Wolf RK, Schneeberger EW, Osterday R, et al. Video-assisted bilateral pulmonary vein isolation and left atrial appendage exclusion for atrial fibrillation. J Thorac Cardiovasc Surg. 2005;130:797–802. doi: 10.1016/j.jtcvs.2005.03.041. [DOI] [PubMed] [Google Scholar]
- 17.Edgerton JR, Jackman WM, Mack MJ. A new epicardial lesion set for minimal access left atrial maze: the Dallas lesion set. Ann Thorac Surg. 2009;88:1655–7. doi: 10.1016/j.athoracsur.2009.05.046. [DOI] [PubMed] [Google Scholar]
- 18.Boersma LV, Castella M, van Boven W, et al. Atrial fibrillation catheter ablation versus surgical ablation treatment (FAST): a 2-center randomized clinical trial. Circulation. 2012;125:23–30. doi: 10.1161/CIRCULATIONAHA.111.074047. [DOI] [PubMed] [Google Scholar]
- 19.Pison L, La Meir M, van Opstal J, Blaauw Y, Maessen J, Crijns HJ. Hybrid thoracoscopic surgical and transvenous catheter ablation of atrial fibrillation. J Am Coll Cardiol. 2012;60:54–61. doi: 10.1016/j.jacc.2011.12.055. [DOI] [PubMed] [Google Scholar]
- 20.Kress DC, Erickson L, Choudhuri I, et al. Comparative effectiveness of hybrid ablation versus endocardial catheter ablation alone in patients with persistent atrial fibrillation. JACC Clin Electrophysiol. 2017;3:341–9. doi: 10.1016/j.jacep.2016.10.010. [DOI] [PubMed] [Google Scholar]
- 21.Calkins H, Hindricks G, Cappato R, et al. 2017 HRS/EHRA/ECAS/APHRS/SOLAECE expert consensus statement on catheter and surgical ablation of atrial fibrillation. Heart Rhythm. 2017;14:e275–444. doi: 10.1016/j.hrthm.2017.05.012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.de Asmundis C, Varnavas V, Sieira J, et al. Two-year follow-up of one-stage left unilateral thoracoscopic epicardial and transcatheter endocardial ablation for persistent and long-standing persistent atrial fibrillation. J Interv Card Electrophysiol. 2019 Sep 13; doi: 10.1007/s10840-019-00616-w. [Epub ahead of print]. [DOI] [PubMed] [Google Scholar]
- 23.Maesen B, Pison L, Vroomen M, et al. Three-year follow-up of hybrid ablation for atrial fibrillation. Eur J Cardiothorac Surg. 2018;53(suppl_1):i26–32. doi: 10.1093/ejcts/ezy117. [DOI] [PubMed] [Google Scholar]
- 24.Wokhlu A, Hodge DO, Monahan KH, et al. Long-term outcome of atrial fibrillation ablation: impact and predictors of very late recurrence. J Cardiovasc Electrophysiol. 2010;21:1071–8. doi: 10.1111/j.1540-8167.2010.01786.x. [DOI] [PubMed] [Google Scholar]
- 25.Jaïs P, Hocini M, Hsu LF, et al. Technique and results of linear ablation at the mitral isthmus. Circulation. 2004;110:2996–3002. doi: 10.1016/j.accreview.2005.02.014. [DOI] [PubMed] [Google Scholar]
- 26.Matsuo S, Wright M, Knecht S, et al. Peri-mitral atrial flutter in patients with atrial fibrillation ablation. Heart Rhythm. 2010;7:2–8. doi: 10.1016/j.hrthm.2009.09.067. [DOI] [PubMed] [Google Scholar]
- 27.Kuck KH, Brugada J, Furnkranz A, et al. Cryoballoon or radiofrequency ablation for paroxysmal atrial fibrillation. N Engl J Med. 2016;374:2235–45. doi: 10.1056/NEJMoa1602014. [DOI] [PubMed] [Google Scholar]
- 28.Verma A, Jiang CY, Betts TR, et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372:1812–22. doi: 10.1056/NEJMoa1408288. [DOI] [PubMed] [Google Scholar]
- 29.Hummel J, Michaud G, Hoyt R, et al. Phased RF ablation in persistent atrial fibrillation. Heart Rhythm. 2014;11:202–9. doi: 10.1016/j.hrthm.2013.11.009. [DOI] [PubMed] [Google Scholar]
- 30.Bassiouny M, Saliba W, Hussein A, et al. Randomized study of persistent atrial fibrillation ablation: ablate in sinus rhythm versus ablate complex-fractionated atrial electrograms in atrial fibrillation. Circ Arrhythm Electrophysiol. 2016;9(2):e003596. doi: 10.1161/CIRCEP.115.003596. [DOI] [PubMed] [Google Scholar]
- 31.Ouyang F, Antz M, Ernst S, et al. Recovered pulmonary vein conduction as a dominant factor for recurrent atrial tachyarrhythmias after complete circular isolation of the pulmonary veins: lessons from double Lasso technique. Circulation. 2005;111:127–35. doi: 10.1161/01.CIR.0000151289.73085.36. [DOI] [PubMed] [Google Scholar]
- 32.Chae S, Oral H, Good E, et al. Atrial tachycardia after circumferential pulmonary vein ablation of atrial fibrillation: mechanistic insights, results of catheter ablation, and risk factors for recurrence. J Am Coll Cardiol. 2007;50:1781–7. doi: 10.1016/j.jacc.2007.07.044. [DOI] [PubMed] [Google Scholar]