Abstract
Purpose
To develop new encoding and reconstruction techniques for fast multi‐contrast/quantitative imaging.
Methods
The recently proposed Echo Planar Time‐resolved Imaging (EPTI) technique can achieve fast distortion‐ and blurring‐free multi‐contrast/quantitative imaging. In this work, a subspace reconstruction framework is developed to improve the reconstruction accuracy of EPTI at high encoding accelerations. The number of unknowns in the reconstruction is significantly reduced by modeling the temporal signal evolutions using low‐rank subspace. As part of the proposed reconstruction approach, a B 0‐update algorithm and a shot‐to‐shot B 0 variation correction method are developed to enable the reconstruction of high‐resolution tissue phase images and to mitigate artifacts from shot‐to‐shot phase variations. Moreover, the EPTI concept is extended to 3D k‐space for 3D GE‐EPTI, where a new “temporal‐variant” of CAIPI encoding is proposed to further improve performance.
Results
The effectiveness of the proposed subspace reconstruction was demonstrated first in 2D GESE EPTI, where the reconstruction achieved higher accuracy when compared to conventional B 0‐informed GRAPPA. For 3D GE‐EPTI, a retrospective undersampling experiment demonstrates that the new temporal‐variant CAIPI encoding can achieve up to 72× acceleration with close to 2× reduction in reconstruction error when compared to conventional spatiotemporal‐CAIPI encoding. In a prospective undersampling experiment, high‐quality whole‐brain and tissue phase maps at 1 mm isotropic resolution were acquired in 52 seconds at 3T using 3D GE‐EPTI with temporal‐variant CAIPI encoding.
Conclusion
The proposed subspace reconstruction and optimized temporal‐variant CAIPI encoding can further improve the performance of EPTI for fast quantitative mapping.
Keywords: EPI, EPTI, fast imaging, quantitative imaging, spatiotemporal encoding, subspace
1. INTRODUCTION
Echo planar imaging (EPI) 1 is a fast MR sequence that has been widely used for diffusion imaging 2 and functional imaging. 3 , 4 However, the low bandwidth along the phase encoding (PE) direction of single‐shot EPI (ss‐EPI) can lead to severe image distortion and blurring, limiting EPI’s ability to acquire high‐quality images with accurate anatomical structures. To mitigate distortion and blurring, parallel imaging (PI) techniques 5 , 6 have been applied to ss‐EPI by reducing its effective echo spacing (ESP). However, the mitigation of distortion and blurring using PI is limited by the achievable in‐plane acceleration. In the pursuit of higher distortion and blurring reduction, multi‐shot EPI (ms‐EPI) techniques 7 , 8 , 9 , 10 , 11 have been developed to further reduce the effective ESP at the cost of a longer acquisition time. Additionally, a navigator acquisition might be necessary to correct for the shot‐to‐shot B 0 variations due to respiration and physiological noises, 12 , 13 which further increases the scan time.
Point‐Spread‐Function (PSF) mapping 14 , 15 , 16 , 17 is a unique ms‐EPI technique that can achieve high‐quality distortion‐ and blurring‐free images, but requires the use of a large number of acquisition shots. Recently, the tilted‐CAIPI 18 acquisition/reconstruction scheme has been developed to accelerate PSF acquisition, which enables distortion‐ and blurring‐free brain imaging at 1‐mm resolution in ~8 EPI‐shots per slice. PSF mapping with tilted‐CAIPI has also been extended to diffusion imaging, where it was refined to enable self‐navigation of shot‐to‐shot phase variation. 18 To further improve the efficacy of tilted‐CAIPI for multi‐echo multi‐contrast imaging, a new, related ms‐EPI technique, termed Echo Planar Time‐resolved Imaging (EPTI) 19 , 20 was developed. EPTI not only achieves distortion‐ and blurring‐free imaging using a small number of EPI‐shots, but it also resolves hundreds of T2/ ‐weighted images across the EPI readout at different echo times. The high acquisition efficiency and rich signal evolution information provided by EPTI have enabled fast multi‐contrast and quantitative mapping. For example, whole‐brain maps of proton‐density, T2, , B 0, and susceptibility weighted imaging (SWI) at 1.1 × 1.1 × 3 mm3 resolution can be acquired in under 30 seconds. 19
In recent years, a number of advanced reconstruction methods have been developed for accelerated quantitative mapping and multi‐contrast imaging, including compressed sensing, 21 , 22 low‐rank priors, 23 , 24 , 25 , 26 , 27 model‐based approaches, 28 , 29 , 30 and dictionary‐based/pattern‐match methods. 31 , 32 , 33 In particular, the low‐rank subspace methods 24 , 25 , 27 , 34 are able to resolve the temporal evolution of the signal across different contrasts with robustness to partial volume and multi‐compartment effects, and have shown promising results for quantitative mapping. Additionally, the subspace can account for more complicated image artifacts and sequence imperfections such as B 1 inhomogeneity 24 through modeling the effects into the forward signal model. The good performance and flexibility of subspace‐based reconstruction is leveraged in many applications including spectroscopic imaging, 35 , 36 , 37 T2‐weighted knee imaging, 24 , 38 quantitative brain 34 , 39 and cardiac imaging. 40 , 41
In conventional EPTI, a GRAPPA‐like B 0‐informed parallel imaging reconstruction is used to exploit the signal correlation along the temporal, spatial and coil dimensions to recover missing data‐points in k‐t space. In this study, we developed a subspace reconstruction framework tailored to EPTI to improve its reconstruction at very high accelerations. To accurately account for B 0‐inhomogeneity phase accumulation across EPTI’s multi‐echo data, we incorporated a “B 0‐update” algorithm 42 to recover both image magnitude and high‐resolution B 0 information directly from the undersampled EPTI data. With this approach, good reconstruction can be obtained without need for a priori high‐resolution B 0 map that can be lengthy to acquire. Moreover, a data‐driven “shot‐to‐shot B 0‐variation” correction method was also developed to provide robustness to shot‐to‐shot phase variation.
In addition to the outlined reconstruction developments, spatiotemporal CAIPI encoding of EPTI was also extended to 3D k‐space for 3D GE‐EPTI. A new “temporal‐variant” of CAIPI encoding was developed and demonstrated to allow acceleration rates of up to 72× folds in k‐t space. To maintain adequate image SNR for mapping at 3T, a more moderate acceleration factor of 32× was used in a prospectively accelerated acquisition, which enables whole‐brain quantification at 1‐mm isotropic resolution in just 52 seconds. The efficacy of this technique is verified with prospective and retrospective undersampling experiments.
2. THEORY
2.1. Review of EPTI acquisition
In EPTI, continuous EPI readouts are performed using highly accelerated spatiotemporal CAIPI‐sampling to efficiently sample the desired signal in k‐t space. For instance, a 2D gradient‐ and spin‐echo (GESE) EPTI sequence as shown in Figure 1A can be used to acquire images across the T2/ signal decay. The EPTI encoding pattern along with a low‐resolution fully sampled calibration dataset in k‐t space are shown in Figure 1B. With EPTI encoding, neighboring readout lines are spaced apart in time by an ESP ( ) and in PE direction by RPE. Each shot acquires several diagonal ky line‐sections that contain multiple readout lines with different PE encodings, with adjacent line‐sections interleaved along PE to allow for better utilization of coil sensitivity. The sampling pattern is repeated temporally to ensure that the time between two ky line‐sections (Ts) is regular and relatively short. This allows efficient use of temporal correlation during image reconstruction. Rseg is the coverage along PE of each EPTI shot, which also determines the time distance between two adjacent ky line‐sections, Ts, by Ts = Δt × Rseg/RPE. Shorter scan times can be achieved by using a larger Rseg, but this increases the time distance (Ts), leading to increased undersampling along t. Our previous work in Ref. 19 has shown that for brain imaging at 3T with 1 mm in‐plane resolution, 5‐9 EPTI‐shots with Rseg of 20‐40 and Ts range of 5‐10 ms can provide good image quality. As part of the acquisition, a fully sampled low‐resolution k‐t calibration dataset is also acquired to train B 0‐informed GRAPPA kernel for reconstruction. 19
2.2. Subspace reconstruction framework
2.2.1. Subspace reconstruction
Highly undersampled k‐t signals can be acquired by EPTI that contains hundreds of image contrasts that track the signal evolution. To resolve these images from the undersampled data, spatiotemporal correlation in k‐t space could be efficiently used. The key concept of subspace reconstruction 27 is that the signal evolution space is low‐rank, and can be represented by several low‐rank subspace bases. Since the number of bases is much smaller than the number of contrasts or time points, the number of unknowns can be significantly reduced through estimating the coefficients of bases instead of multi‐echo images. The subspace approach utilizes the temporal correlation of signals by reducing the temporal redundancy in the estimation process, with multi‐channel coil sensitivity information also integrated to provide additional spatial encoding information for the reconstruction.
Figure 1C illustrates the proposed subspace reconstruction for EPTI. At first, the signal evolution space (T2/ decay for GESE‐EPTI) within certain quantitative parameter ranges is simulated based on the Bloch equation and the acquisition parameters. Then, several basis vectors are extracted through principle component analysis (PCA) as . These vectors form a low‐dimensional subspace that can closely approximate the entire signal space of interest. Using these bases, the temporal image series can be calculated by , where is the coefficient maps of the bases. In this way, the degrees of freedom of reconstruction is reduced from the number of time points to the number of bases, improving the conditioning of reconstruction and the SNR of images. The subspace‐constrained reconstruction is solved by,
(1) |
where is the phase evolution across different image echoes due to B 0 inhomogeneity, is the coil sensitivity, is the Fourier transform operator, is the undersampling mask, and is the acquired undersampled EPTI dataset. The regularization term can be incorporated to further improve the conditioning and SNR, and is the control parameter of the regularization. After solving for , the time‐series of images can be recovered by .
2.2.2. B 0‐update algorithm
The B 0 inhomogeneity‐induced phase change and coil sensitivity can both be estimated from the low‐resolution k‐t calibration data as shown in Figure 1B. However, low‐resolution B 0 maps do not contain high‐spatial‐frequency information that if not incorporated into the reconstruction can cause image artifacts in areas with strong susceptibility. To avoid a time‐consuming acquisition for a high‐resolution B 0 map, a phase update algorithm named “general phase regularized reconstruction using phase cycling” 42 is incorporated into the proposed subspace reconstruction, to estimate the high‐resolution B 0 map as shown in step 1 and 2 of Figure 2. In step 1, the initial image is reconstructed by subspace reconstruction using the low‐resolution B 0 map obtained from the fast calibration scan. In step 2, a high‐resolution B 0 map is estimated by phase‐cycling using the pre‐reconstructed image and acquired data:
(2) |
where is the multi‐echo image magnitude with background phase estimated from the subspace reconstruction, is the high‐resolution off‐resonance map in Hz to be estimated, , is the gyromagnetic ratio of protons and is the field offset in Tesla. Phase evolution can be calculated by , where is the echo times. A wavelet constraint term is incorporated to improve the conditioning of the B 0 estimation by using L1 regularization on the Daubechies 6 wavelet domain. The updated high‐resolution B 0 map can then be used as input into the subspace reconstruction (step 1) to further improve the accuracy of magnitude images.
2.2.3. Data‐driven shot‐to‐shot B 0 variation correction
Shot‐to‐shot B 0 variation can cause phase inconsistency in the EPTI data, which was shown to result in small local spatial smoothing in the reconstructed images. 19 To mitigate this issue in, 19 a data‐driven approach was developed to estimate and correct for this B 0 variation that can work well with B 0‐informed GRAPPA reconstruction. However, such approach is not compatible with the proposed subspace reconstruction and a new correction approach is developed here as outlined below.
Ref. 43 proposed a navigator‐free data‐driven approach to estimate the mean B 0 change (0th‐order) in each receive coil of each acquisition shot ( ), and correct for these variations in the raw k‐space data to achieve improved reconstruction in the presence of shot‐to‐shot B 0 variation. In this work, a data‐driven correction method that accounts for higher order spatial distribution of the B 0 variation is developed as an extension to this approach, and incorporated into the proposed subspace reconstruction as shown in step 3 of Figure 2. Here, the multi‐echo EPTI data are used to fit the mean B 0 change of each shot to achieve improved accuracy over single echo estimation. A 2nd‐order polynomial B 0 variation map is estimated for each shot that approximates the mean B 0 change across different coils ( ) estimated using the method in Ref. 43.
Specifically, the mean B 0 change of each channel is first calculated as (size of , is the number of coils) for each shot by calculating the phase difference between the raw and the reconstructed k‐space signals as described in, 43 , and is the field variation in Tesla. Then, the coefficients of the polynomial B 0 variation map for one EPTI shot are estimated by solving
(3) |
Here, is the 2nd‐order polynomial matrix, represents the echo‐combined magnitude image obtained from subspace reconstruction, is the coil sensitivity, therefore, is the magnitude‐weighted B 0 variation map (size of ), and operator is used to average the weighted B 0 variation map across all the image pixels to compare with the mean of that shot. Using this approach, we can estimate the polynomial B 0 variation map of each EPTI shot, and correct the corresponding phase by , to mitigate artifacts in the subspace reconstruction. The three reconstruction steps shown in Figure 2 can be repeated several times to improve the accuracy of both the magnitude and phase of the reconstructed image time‐series.
2.3. 3D spatiotemporal encoding
The design of the spatiotemporal sampling in EPTI is critical to achieve accurate signal recovery from highly undersampled data. With the proposed subspace reconstruction, the sampling pattern is no longer constrained to be fixed across k y‐t space as was in the case of B 0‐informed GRAPPA reconstruction (to limit the number of GRAPPA kernels to be trained). In order to design an optimized encoding for 3D EPTI, three types of spatiotemporal EPTI encodings were proposed and evaluated using a 3D GE‐EPTI sequence: (1) CAIPI‐based sampling, (2) temporal‐variant CAIPI sampling, and (3) block‐wise random sampling.
Figure 3A shows the sequence diagram and an example of the encoding pattern of a 3D GE‐EPTI acquisition. Following each excitation pulse, an EPTI readout is acquired with multiple readout lines shown as filled circles in the bottom‐left ky‐kz plot. This forms a spatiotemporal CAIPI pattern within each ky‐kz block (with ky samples interleaving across time) that is acquired during each EPTI readout. Here, for the purpose of illustration, several echoes (6 echoes in Figure 3A) are grouped together shown as an “echo‐section” even if they are acquired at different TEs. The undersampling factor in ky‐kz‐t space is then determined by the block size, R = ky‐blocksize × kz‐blocksize, since only one readout line will be acquired at each TE within the block. For example, the block size of Figure 3A is 12 × 6 (ky‐kz) and the acceleration factor is 72. In each TR, the EPTI readout block covers a portion of the 3D k‐space (blue or red boxes in Figure 3A bottom‐right), and after hundreds of TRs, the full ky‐kz‐t space will be covered.
The details of the three different k‐t sampling trajectories are shown in Figure 3B. The CAIPI‐based sampling (top row) repeats the same block‐wise spatiotemporal encoding pattern in all echo‐sections, whereas the temporal‐variant CAIPI sampling (middle row) shifts all the even echo‐sections along ky and kz by certain amounts to achieve a more complementary sampling with the odd echo‐sections. The block‐wise random sampling (bottom row), on the contrary, randomly samples a ky‐kz point within the block at each echo time. The block‐wise encoding design used in all these three trajectories was chosen to ensure that the PE and partition gradient blips are small enough to avoid increasing ESP by a large amount and/or causing strong nerve stimulation during the acquisition.
Additionally, two variable density sampling (VDS) patterns: (1) VDS temporal‐variant CAIPI and (2) VDS block‐wise random sampling, were also designed and utilized with locally low rank (LLR) constraint to further improve the image SNR. The evaluation of different encoding patterns helps us investigate the factors that will affect the reconstruction performance of 3D EPTI, such as complementary ky‐kz pattern across time and uniform or non‐uniform k‐space sampling pattern.
3. METHODS
All data were acquired with a consented institutionally approved protocol on a Siemens Prisma 3T scanner with a 32‐channel head coil (Siemens Healthineers, Erlangen, Germany). Three subjects were included in the in vivo experiments, one female and two males; mean age: 26 years; range: 25 to 28 years of age), with the retrospective 3D and prospective 2D experiments acquired on the same subject. Spectrally selective fat suppression was used for all the experiments to suppress the fat signal. The proposed subspace reconstruction was implemented using the open source code by Jonathan Tamir (https://github.com/jtamir/t2shuffling‐support) as a start point, while the general phase regularized reconstruction using phase cycling was implemented using the open source code from Frank Ong (https://github.com/mikgroup/phase_cycling).
3.1. Evaluation of reconstruction using 2D GESE EPTI
3.1.1. Subspace reconstruction
A fully sampled 2D GESE EPTI dataset was acquired to evaluate the subspace reconstruction with the following acquisition parameters: FOV = 240 × 240 mm2, in‐plane resolution = 1.1 × 1.1 mm2, slice thickness = 3 mm, number of shots = 216, number of echoes (GE/SE) = 40/80, ESP = 0.93 ms, echo time range of GE / SE = 8.4‐44.7 ms / 70.8‐144.3 ms, TR = 2.5 s. This dataset was then retrospectively undersampled in ky‐t space by Rseg = 24 to synthesize a 9‐shot 2D EPTI acquisition. Subspace reconstruction and B 0‐informed GRAPPA reconstruction were used to reconstruct the 9‐shot data, and the results compared with the 216‐shot fully sampled reference.
The subspace bases were generated using the Bloch equation and a large range of quantitative tissue parameters 44 : range of T2 = 1‐600 ms, range of = 1‐500 ms. To simulate the effect of imperfect refocusing pulse due to B 1 inhomogeneity that will change the magnitude ratio between SE and GE signals, a scaling factor of 0.8‐1.2 between GE and SE was also included. Eight bases were extracted from the simulated signals by PCA, which can approximate the simulated signal space with an error < 0.2%. The subspace reconstruction is solved by the alternating direction method of multipliers (ADMM) 45 algorithm. The stop criterion of the subspace reconstruction was a maximum iteration = 60, and lambda was set to 0.01. The reconstruction parameters were selected based on the exhaustive search method with empirical ranges based on a simulation test.
3.1.2. B 0‐update evaluation
The effectiveness of the B 0 update algorithm was evaluated using the same 2D GESE EPTI dataset. The high‐resolution B 0 map and tissue phase were calculated using the fully sampled data with a matrix size of 216 × 216 × 40 (kx‐ky‐Necho) as reference, while the low‐resolution calibration data with a matrix size = 216 × 49 × 6 (kx‐ky‐Necho) was used to estimate the initial low‐resolution B 0 map. The reconstructed magnitude and phase images from the subspace reconstruction with and without B 0 update were evaluated for the retrospectively undersampled 9‐shot EPTI case. For the B 0 estimation, wavelet constraint was used with a lambda = 0.02 to improve the conditioning. The number of iteration was set to 50 for the B 0 estimation, and 5 for the large loop of subspace reconstruction and B 0 estimation.
3.1.3. Data‐driven B 0‐variation correction
To evaluate the proposed shot‐to‐shot B 0‐variation correction method, 9‐shot 2D GE‐EPTI data were simulated with added B 0‐variation in the same way as in the evaluation performed in Ref. 19 Realistic B 0 maps obtained from an in vivo scan with a spin and gradient echo EPI (SAGE‐EPI) sequence 46 were used, which contain B 0 variations across the 9 shots of up to 3 Hz as shown in Figure 6A. The error maps of the subspace reconstruction with and without B 0‐variation correction were calculated at different echo times and compared. Additionally, PSF analysis was performed to quantify the mitigated local smoothing achieved with B 0‐variation correction. The PSF of 3D GE‐EPTI acquisition was also analyzed with the same level of B 0‐variation (standard deviation = 0‐2 Hz) as in 2D EPTI, to demonstrate that the effect of local smoothing in 3D acquisition is negligible, and therefore, the shot‐to‐shot B 0‐variation correction was only used for 2D EPTI acquisition.
3.2. 3D GE‐EPTI with spatiotemporal encoding
3.2.1. Retrospective undersampling experiment
Fully sampled 3D GE‐EPTI data were acquired and used to evaluate the performance of different spatiotemporal encoding patterns. The acquisition parameters were: FOV = 220 × 220 × 106 mm3, resolution = 1.1 × 1.1 × 1.1 mm3, matrix size = 200 × 200 × 96, number of shots = 200 × 96 (y‐block × z‐block) = 19200, TR = 80 ms. The total acquisition time was 25 minutes and 36 seconds. 50 echoes were acquired in each TR with an ESP of 0.93 ms, covering a TE range of 9.1 ms‐54.7 ms. In the first experiment, the data were retrospectively undersampled in ky‐kz‐t space by 72× folds with a block size of 12 × 6 (ky‐kz) for all the three encoding strategies as shown in Figure 3B, which reduces the acquisition time from >25 minutes to 22 seconds. The reconstruction errors of the three different undersamplings were calculated using the fully sampled data as the reference. Moreover, VDS temporal‐variant CAIPI and VDS block‐wise random sampling were also evaluated at the same net acceleration factor of 72×, with locally low rank (LLR) constraint utilized in the reconstruction. The VDS patterns are using elliptical sampling (no sampling outside the ellipse in ky‐kz), and have two different sampling rates in the ellipse: 32× undersampling at the k‐space center, 72× undersampling at the outer k‐space.
For temporal‐variant CAIPI sampling, reconstruction errors for acquisitions with different ky and kz shifts between odd and even echo‐sections were compared to investigate the key factors in encoding design that can effect performance. Here, the effect of more vs. less complementary ky‐kz patterns between the odd and even echo‐sections on reconstruction performance was analyzed. Using the optimized VDS temporal‐variant CAIPI sampling, quantitative maps and tissue phase maps were calculated at undersampling factors of 72 (block size = 12 × 6) and 32 (block size = 8 × 4). Error maps were calculated using the fully sampled data as reference.
3.2.2. Prospective undersampling experiment
Prospective undersampled 3D GE‐EPTI data were acquired to evaluate the usage of optimized 3D EPTI for fast in vivo quantitative mapping at high spatial resolution. The data were acquired with VDS temporal‐variant CAIPI sampling at an acceleration factor of 32 (block size = 8 × 4). Other acquisition parameters were: FOV = 216 × 216 × 96 mm3, resolution = 1 × 1 × 1 mm3, matrix size = 216 × 216 × 96, ESP = 0.95 ms, number of echoes = 57, TE range = 3.8 ms‐57 ms, number of shots = 27 × 24 (y‐block × z‐block) = 648, flip angle = 20°, TR = 80 ms. The total acquisition time was 52 s. In addition to the imaging data acquisition, a low‐resolution calibration data were also acquired with the same FOV and ESP to obtain sensitivity maps and initial estimation of the B 0 map. The acquisition time of calibration data was ~18 s, where accelerations along ky and kz of 2 × 2 were used and only 8 echoes were acquired to reduce the acquisition time (TR = 30 ms). The key acquisition parameters of calibration data were: matrix size = 216 × 42 × 32, the k‐space center was fully sampled to provide GRAPPA calibration for reconstructing this undersampled calibration data, and the net acceleration factors of ky and kz = 1.6 and 1.5, respectively.
4. RESULTS
Figure 4 shows the comparison of subspace and B 0‐informed GRAPPA reconstructions for the 2D GESE dataset. In Figure 4A, the reconstructed images and error maps at four different TEs are presented. Both methods performed well for the center echoes of the EPTI readout window (TE = 27 and 107 ms), however, the subspace method significantly reduces the reconstruction errors of the edge echoes (TE = 14 and 141 ms) where the k‐t GRAPPA kernel cannot be applied fully to fill in the missing data. Such improvement can also be observed in the reconstructed signal evolution across GE‐SE signals of a representative voxel in the image as shown in Figure 4B, where the subspace reconstruction recovers more accurate signal curve with reduced noise and errors, especially at the edge echoes.
The results of the B 0‐update evaluation are shown Figure 5. The first row shows the high‐resolution B 0 map and tissue phase calculated from the fully sampled calibration dataset with 216 PE and 40 echoes, which took 120 seconds to acquire for 10 slices. The second row shows that a low‐resolution calibration dataset with 49 PE × 6 echoes can reduce the acquisition time to just 10 seconds, but the high‐resolution tissue phase information is missing in this data, causing artifacts in the reconstructed image as can be seen in the error map. Using the B 0‐update algorithm, the high‐resolution tissue phase can be well estimated, and the image artifacts mitigated with a reduced root‐mean square error (RMSE) as shown in the bottom row of the figure.
Figure 6B shows the evaluation results of the proposed B 0‐variation correction, where the reconstructed images and the corresponding error maps are shown for reconstructions with and without B 0 variation correction. Figure 6C shows the standard deviation of the B 0‐variation across the 9 simulated EPTI‐shots, before and after correction. Without the correction, high standard deviations in the B 0 variation can be observed in the anterior portion of the brain, causing noticeable reconstruction errors in this area, particularly at long TEs where the phase accrual from such variation is more pronounced. With the proposed correction, the standard deviations are markedly reduced and this resulted in less image errors with smaller RMSEs. The effects of B 0‐variation correction on the PSF are shown in Supporting Information Figure S1 at the same TE of 70 ms. In the 2D acquisition case (Figure S1A), a B 0 variation with a standard deviation of 1.5 Hz causes local smoothing with a side lobe level of 4.1%, while after correction, the side lobe is reduced to only 2.1%. Supporting Information Figure S1B shows the PSF analysis for 3D EPTI where the smoothing effect of B 0 variation is spread along both z and y direction. In this case, the largest side lobe is only 0.8%, even without any correction, causing negligible smoothing. To further validate the performance of the proposed method, prospective undersampling 2D experiment was also performed as shown in Supporting Information Figure S2.
Figure 7 shows the results of the retrospective undersampling experiment for 3D GE‐EPTI at 72× acceleration. Five different encoding patterns were compared: CAIPI‐based sampling, temporal‐variant CAIPI sampling, block‐wise random sampling, VDS temporal‐variant CAIPI, and VDS random sampling. The ky‐kz views of these patterns in Figure 7 shows the summation of all the k‐t samples along the time dimension. The first three encodings were reconstructed using the proposed subspace approach without LLR constraint. Here, the temporal‐variant CAIPI shows the best performance with the lowest RMSE (6.94%) when compared against CAIPI‐based sampling (11.4%) and block‐wise random sampling (8.56%). The last two encoding patterns which utilize VDS were reconstructed with additional LLR constraint. Here, the reconstruction from VDS temporal‐variant CAIPI shows a lower error (5.89%) when compared against VDS random sampling (6.97%). Based on these results, the temporal‐variant CAIPI sampling achieved better performance than conventional CAIPI‐based sampling and random sampling, and when combing with VDS and LLR, the temporal‐variant CAIPI still shows better results than random sampling.
To further investigate the optimal sampling design based on the temporal‐variant CAIPI scheme, reconstructions from data with different ky and kz shift steps between the odd and the even echo‐sections were analyzed. Figure 8A shows the reconstruction RMSEs at different shift steps for an acquisition block size of 12 × 6 (ky × kz). Here, the encoding pattern with 0‐0 (ky‐kz) shift corresponds to conventional spatiotemporal CAIPI‐sampling, which has the largest error. The sampling patterns in ky‐kz view are shown in Figure 8B,C for two representative shifting patterns: 4‐1 shift (Pattern 1) and 12‐2 shift (equivalent to 0‐2 shift, Pattern 2). The “total pattern” in ky‐kz (summation of the two echo‐sections along t) of Pattern 2 contains more complementary sampling than Pattern 1, and achieved much lower reconstruction RMSE, indicating that a more complementary sampling can improve reconstruction.
The evaluation of quantitative and tissue phase maps obtained from the retrospective undersampling experiment is presented in Figure 9, at two undersampling factors of 72× and 32×, both with VDS temporal‐variant sampling. The scan time of the fully sampled data was 25 minutes at 1.1 mm isotropic resolution, which is reduced to 21 s and 48 s through 72× and 32× accelerations, respectively (with an additional calibration scan of ~18 s). The quality of the tissue phase obtained from the 72× accelerated EPTI is comparable to the full‐sampled case with a small increase in the noise level, while the map shows higher error at higher noise level. With the 32× accelerated EPTI, the map has higher SNR and reduced error (mean percentage error of 32× and 72× maps = 7.66% and 10.5%), while the acquisition time is still less than 1 minute.
The prospective undersampling results are shown in Figure 10, where 3D GE‐EPTI with VDS temporal‐variant sampling was used to acquire whole‐brain ‐weighted images with 57 echoes at 1 mm isotropic resolution in 52 seconds. High‐quality multi‐echo ‐weighted images, along with quantitative and tissue phase maps are presented in three orthogonal views. The results of quantitative ROI analysis on NIST phantom are shown in Supporting Information Figure S3 to further validate the accuracy of the 3D GE‐EPTI.
5. DISCUSSION AND CONCLUSION
A subspace reconstruction framework for EPTI was successfully developed and demonstrated to provide improved reconstruction accuracy over conventional B 0‐informed GRAPPA. Such reconstruction uses prior low‐rank information of the signal model to reduce the number of unknowns and improve the image SNR. In addition to taking advantage of low‐rank modeling, approaches for B 0 estimation and shot‐to‐shot B 0‐variation correction were developed and demonstrated to further improve reconstruction and enable the use of fast low‐resolution calibration scan. Moreover, the EPTI concept was extended successfully to 3D sampling and new 3D spatiotemporal CAIPI encoding schemes were developed to improve acceleration capability.
Through volumetric encoding and improved noise averaging, 3D EPTI can achieve higher SNR than their 2D EPTI counterpart, lending itself well for the application in rapid high‐resolution multi‐contrast and quantitative mapping. The additional partition encoding dimension in 3D EPTI not only provides more flexibility for undersampling, but also adds more complexity for the design of the spatiotemporal sampling pattern. In this work, various 3D spatiotemporal sampling patterns were compared using the subspace reconstruction. Among all of these strategies, the temporal‐variant CAIPI pattern achieves the best performance (Figure 7), and its combination with VDS and LLR constraint was demonstrated to further improve reconstruction accuracy at high accelerations.
The analysis of the encoding patterns in Figure 8 shows that a more complementary sampling between different echo‐sections can improve the reconstruction accuracy. The improvement reflects the benefits of creating more spatial and temporal correlation across the sampled k‐t signals. Specifically, a more complementary sampling in k y‐k z space will create more spatial correlation and allow for better signal recovery using coil sensitivity information and parallel imaging. Therefore, undersampling factor in k y‐k z is chosen to achieve high temporal correlation while keeping within the capability of parallel imaging reconstruction. The proposed temporal‐variant CAIPI encoding achieved the best performance by creating such spatiotemporal correlation in the sampled data, and allows high undersampling of up to 72×. The improved performance of the temporal‐variant CAIPI over random sampling can be attributed to better utilization of coil sensitivity information by taking parallel imaging/coil‐encoding consideration into the sampling design.
The high acceleration and the high acquisition efficiency of EPTI with continuous readout enables ultra‐fast multi‐contrast and quantitative mapping at high isotropic resolution in human brain, as demonstrated by both retrospective (Figure 9) and prospective (Figure 10) undersampling experiments. Interestingly, better reconstruction performance with lower noise was achieved in the prospective undersampling case (Figure 10) when compared to the retrospective case (Figure 9), both at 32× acceleration, even though the retrospective undersampling experiment was performed at a slightly larger voxel size (1.1 mm vs. 1 mm isotropic) and was not affected by varying eddy currents and concomitant gradients due to varying phase‐ and partition‐encoding gradient blips as in the prospective. This is likely to be because of the increased physiological noise and the inevitable motion during the lengthy acquisition of the fully sampled dataset (~25 minutes) used in the retrospective undersampling case. In an even higher acceleration case of 72×, the prospective sampling experiment was able to achieve good reconstruction of the magnitude and phase images. However, the magnitude images under 72× acceleration were relatively noisy at 3T. Application of 3D EPTI at higher field strength, such as at 7T, should help boost the SNR and enable higher performance at such a high acceleration rate. One limitation of the proposed method is that the reconstruction time can be relatively long, currently at ~1 minute per slice for 1 mm resolution brain imaging when perform in MATLAB using a Linux workstation (Intel Xeon, 2.00 GHz 20 Cores). Further optimization using parallel computation as well as implementation in C++ should be helpful to reduce the reconstruction time. The proposed method also did not model the effects of eddy currents, concomitant gradients and unsuppressed lipid signals. Including these effects into the signal model could potentially further improve the accuracy and robustness of the method. In addition, limited number of subjects were included in this study. To further evaluate the EPTI technique and the k‐t trajectory optimization in the future, rigorous multi‐subject experiments and repeatability tests will be performed.
The proposed EPTI acquisition/reconstruction method can achieve high acquisition efficiency to facilitate accurate and rapid quantitative parameter mapping. The use of the continuous EPTI readout improves noise averaging, and the use of temporal‐variant CAIPI and subspace reconstruction improves the conditioning of the reconstruction. The proposed sampling and reconstruction approaches should also be useful for the recently developed inversion‐prep 3D EPTI scheme for rapid quantitative mapping of T1 T2 and , 47 and for the propeller‐based EPTI (PEPTIDE) approach that can achieve motion‐robust fast quantitative imaging. 48 Further improvement in acquisition efficiency should be feasible through further optimization, for example through the use of Cramer‐Rao bound in sequence design 49 and incorporation of the low‐rank tensor modeling. 50
Supporting information
ACKNOWLEDGMENTS
This work was supported by NIH NIBIB (R01‐EB020613, R01‐EB019437, R01‐MH116173, P41‐EB015896, and U01‐EB025162) and by the MGH/HST Athinoula A. Martinos Center for Biomedical Imaging; and was made possible by the resources provided by NIH Shared Instrumentation Grants S10‐RR023401, S10‐RR023043, and S10‐RR019307.
Dong Z, Wang F, Reese TG, Bilgic B, Setsompop K. Echo planar time‐resolved imaging with subspace reconstruction and optimized spatiotemporal encoding. Magn Reson Med. 2020;84:2442–2455. 10.1002/mrm.28295
Funding information
This work was supported by the NIH NIBIB (R01‐EB020613, R01‐EB019437, R01‐MH116173, P41‐EB015896, and U01‐EB025162) and the instrumentation Grants (S10‐RR023401, S10‐RR023043, and S10‐RR019307)
REFERENCES
- 1. Mansfield P. Multi‐planar image formation using NMR spin echoes. J Phys C: Solid State Phys. 1977;10:L55. [Google Scholar]
- 2. Warach S, Gaa J, Siewert B, Wielopolski P, Edelman RR. Acute human stroke studied by whole brain echo planar diffusion‐weighted magnetic resonance imaging. Ann Neurol. 1995;37:231‐241. [DOI] [PubMed] [Google Scholar]
- 3. Biswal B, Zerrin Yetkin F, Haughton VM, Hyde JS. Functional connectivity in the motor cortex of resting human brain using echo‐planar MRI. Magn Reson Med. 1995;34:537‐541. [DOI] [PubMed] [Google Scholar]
- 4. Kwong K. Functional magnetic resonance imaging with echo planar imaging Functional MRI. Milano: Springer; 1996:73‐90. [Google Scholar]
- 5. Pruessmann KP, Weiger M, Scheidegger MB, Boesiger P. SENSE: Sensitivity encoding for fast MRI. Magn Reson Med. 1999;42:952‐962. [PubMed] [Google Scholar]
- 6. Griswold MA, Jakob PM, Heidemann RM, et al. Generalized autocalibrating partially parallel acquisitions (GRAPPA). Magn Reson Med. 2002;47:1202‐1210. [DOI] [PubMed] [Google Scholar]
- 7. Butts K, de Crespigny A, Pauly JM, Moseley M. Diffusion‐weighted interleaved echo‐planar imaging with a pair of orthogonal navigator echoes. Magn Reson Med. 1996;35:763‐770. [DOI] [PubMed] [Google Scholar]
- 8. Dong Z, Wang F, Ma X, et al. Interleaved EPI diffusion imaging using SPIRiT‐based reconstruction with virtual coil compression. Magn Reson Med. 2018;79:1525‐1531. [DOI] [PubMed] [Google Scholar]
- 9. Chen N‐K, Guidon A, Chang H‐C, Song AW. A robust multi‐shot scan strategy for high‐resolution diffusion weighted MRI enabled by multiplexed sensitivity‐encoding (MUSE). NeuroImage. 2013;72:41‐47. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Porter D, Mueller E. Multi‐shot diffusion‐weighted EPI with readout mosaic segmentation and 2D navigator correction. Proc Intl Soc Mag Reson Med. 2004;442. [Google Scholar]
- 11. Holdsworth SJ, Skare S, Newbould RD, Guzmann R, Blevins NH, Bammer R. Readout‐segmented EPI for rapid high resolution diffusion imaging at 3T. Eur J Radiol. 2008;65:36‐46. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12. Dong Z, Wang F, Ma X, Dai E, Zhang Z, Guo H. Motion‐corrected k‐space reconstruction for interleaved EPI diffusion imaging. Magn Reson Med. 2018;79:1992‐2002. [DOI] [PubMed] [Google Scholar]
- 13. Jeong HK, Gore JC, Anderson AW. High‐resolution human diffusion tensor imaging using 2‐D navigated multishot SENSE EPI at 7 T. Magn Reson Med. 2013;69:793‐802. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14. Hu Z, Wang Y, Dong Z, Guo H. Water/fat separation for distortion‐free EPI with point spread function encoding. Magn Reson Med. 2019;82:251‐262. [DOI] [PubMed] [Google Scholar]
- 15. Robson MD, Gore JC, Constable RT. Measurement of the point spread function in MRI using constant time imaging. Magn Reson Med. 1997;38:733‐740. [DOI] [PubMed] [Google Scholar]
- 16. Zaitsev M, Hennig J, Speck O. Point spread function mapping with parallel imaging techniques and high acceleration factors: Fast, robust, and flexible method for echo‐planar imaging distortion correction. Magn Reson Med. 2004;52:1156‐1166. [DOI] [PubMed] [Google Scholar]
- 17. In MH, Posnansky O, Speck O. High‐resolution distortion‐free diffusion imaging using hybrid spin‐warp and echo‐planar PSF‐encoding approach. NeuroImage. 2017;148:20‐30. [DOI] [PubMed] [Google Scholar]
- 18. Dong Z, Wang F, Reese TG, et al. Tilted‐CAIPI for highly accelerated distortion‐free EPI with point spread function (PSF) encoding. Magn Reson Med. 2019;81:377‐392. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Wang F, Dong Z, Reese TG, et al. Echo planar time‐resolved imaging (EPTI). Magn Reson Med. 2019;81:3599‐3615. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20. Wang F, Dong Z, Reese TG, et al. Echo planar time‐resolved imaging (EPTI). In Proceedings of ISMRM, Paris, France, 2018. p. 217.
- 21. Lustig M, Donoho D, Pauly JM. Sparse MRI: The application of compressed sensing for rapid MR imaging. Magn Reson Med. 2007;58:1182‐1195. [DOI] [PubMed] [Google Scholar]
- 22. Huang C, Graff CG, Clarkson EW, Bilgin A, Altbach MI. T2 mapping from highly undersampled data by reconstruction of principal component coefficient maps using compressed sensing. Magn Reson Med. 2012;67:1355‐1366. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23. Zhang T, Pauly JM, Levesque IR. Accelerating parameter mapping with a locally low rank constraint. Magn Reson Med. 2015;73:655‐661. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 24. Tamir JI, Uecker M, Chen W, et al. T2 shuffling: Sharp, multicontrast, volumetric fast spin‐echo imaging. Magn Reson Med. 2017;77:180‐195. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Zhao B, Lu W, Hitchens TK, Lam F, Ho C, Liang ZP. Accelerated MR parameter mapping with low‐rank and sparsity constraints. Magn Reson Med. 2015;74:489‐498. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26. Petzschner FH, Ponce IP, Blaimer M, Jakob PM, Breuer FA. Fast MR parameter mapping using k‐t principal component analysis. Magn Reson Med. 2011;66:706‐716. [DOI] [PubMed] [Google Scholar]
- 27. Liang Z‐P. Spatiotemporal imaging with partially separable functions. IEEE. 2007;988‐991. [Google Scholar]
- 28. Dong Z, Dai E, Wang F, et al. Model‐based reconstruction for simultaneous multislice and parallel imaging accelerated multishot diffusion tensor imaging. Med Phys. 2018;45:3196‐3204. [DOI] [PubMed] [Google Scholar]
- 29. Sumpf TJ, Uecker M, Boretius S, Frahm J. Model‐based nonlinear inverse reconstruction for T2 mapping using highly undersampled spin‐echo MRI. J Magn Reson Imaging. 2011;34:420‐428. [DOI] [PubMed] [Google Scholar]
- 30. Sumpf TJ, Petrovic A, Uecker M, Knoll F, Frahm J. Fast T2 mapping with improved accuracy using undersampled spin‐echo MRI and model‐based reconstructions with a generating function. IEEE Trans Med Imag. 2014;33:2213‐2222. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31. Ma D, Gulani V, Seiberlich N, et al. Magnetic resonance fingerprinting. Nature. 2013;495(7440):187‐192. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Ben‐Eliezer N, Sodickson DK, Block KT. Rapid and accurate T2 mapping from multi‐spin‐echo data using Bloch‐simulation‐based reconstruction. Magn Reson Med. 2015;73:809‐817. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Doneva M, Bornert P, Eggers H, Stehning C, Senegas J, Mertins A. Compressed sensing reconstruction for magnetic resonance parameter mapping. Magn Reson Med. 2010;64:1114‐1120. [DOI] [PubMed] [Google Scholar]
- 34. Pfister J, Blaimer M, Kullmann WH, Bartsch AJ, Jakob PM, Breuer FA. Simultaneous T1 and T2 measurements using inversion recovery TrueFISP with principle component‐based reconstruction, off‐resonance correction, and multicomponent analysis. Magn Reson Med. 2019;81:3488‐3502. [DOI] [PubMed] [Google Scholar]
- 35. Lam F, Ma C, Clifford B, Johnson CL, Liang ZP. High‐resolution 1H‐MRSI of the brain using SPICE: Data acquisition and image reconstruction. Magn Reson Med. 2016;76:1059‐1070. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Ma C, Lam F, Ning Q, Johnson CL, Liang ZP. High‐resolution 1H‐MRSI of the brain using short‐TE SPICE. Magn Reson Med. 2017;77:467‐479. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Peng X, Lam F, Li Y, Clifford B, Liang ZP. Simultaneous QSM and metabolic imaging of the brain using SPICE. Magn Reson Med. 2018;79:13‐21. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38. Tamir JI, Taviani V, Alley MT, et al. Targeted rapid knee MRI exam using T2 shuffling. J Magn Reson Imaging. 2019;49:e195‐e204. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39. Zhao B, Setsompop K, Adalsteinsson E, et al. Improved magnetic resonance fingerprinting reconstruction with low‐rank and subspace modeling. Magn Reson Med. 2018;79:933‐942. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Christodoulou AG, Shaw JL, Nguyen C, et al. Magnetic resonance multitasking for motion‐resolved quantitative cardiovascular imaging. Nat Biomed Eng. 2018;2:215‐226. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41. Shaw JL, Yang Q, Zhou Z, et al. Free‐breathing, non‐ECG, continuous myocardial T1 mapping with cardiovascular magnetic resonance multitasking. Magn Reson Med. 2019;81:2450‐2463. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42. Ong F, Cheng JY, Lustig M. General phase regularized reconstruction using phase cycling. Magn Reson Med. 2018;80:112‐125. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43. Meineke J, Nielsen T. Data consistency‐driven determination of B 0 ‐fluctuations in gradient‐echo MRI. Magn Reson Med. 2019;81:3046‐3055. [DOI] [PubMed] [Google Scholar]
- 44. Schmiedeskamp H, Straka M, Bammer R. Compensation of slice profile mismatch in combined spin‐and gradient‐echo echo‐planar imaging pulse sequences. Magn Reson Med. 2012;67:378‐388. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45. Boyd S, Parikh N, Chu E, Peleato B, Eckstein J. Distributed optimization and statistical learning via the alternating direction method of multipliers. Found Trends Mach Learn. 2011;3:1‐122. [Google Scholar]
- 46. Schmiedeskamp H, Straka M, Newbould RD, et al. Combined spin‐and gradient‐echo perfusion‐weighted imaging. Magn Reson Med. 2012;68:30‐40. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47. Wang F, Dong Z, Reese TG, Wald LL, Setsompop K. 3D‐EPTI for ultra‐fast multi‐contrast and quantitative imaging. In Proc Intl Soc Mag Reson Med. Montreal: 2019; p. 944. [Google Scholar]
- 48. Fair JM, Wang F, Dong Z, Reese TG, Setsompop K. Propeller echo‐planar time‐resolved imaging with dynamic encoding (PEPTIDE). Magn Reson Med. 2020;83:2124‐2137. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49. Zhao B, Haldar JP, Liao C, et al. Optimal experiment design for magnetic resonance fingerprinting: Cramér‐Rao bound meets spin dynamics. IEEE Trans Med Imag. 2018;38:844‐861. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50. He J, Liu Q, Christodoulou AG, Ma C, Lam F, Liang ZP. Accelerated high‐dimensional MR imaging with sparse sampling using low‐rank tensors. IEEE Trans Med Imag. 2016;35:2119‐2129. [DOI] [PMC free article] [PubMed] [Google Scholar]