Clinical Utility of a Highly Sensitive Lateral Flow Immunoassay as determined by Titer Analysis for the Detection of anti-SARS-CoV-2 Antibodies at the Point-of-Care

Amanda Haymond, Claudius Mueller, Hannah Steinberg, K Alex Hodge, Caitlin W Lehman, Shih-Chao Lin, Lucia Collini, Heather Branscome, Tuong Vi Nguyen, Sally Rucker, Lauren Panny, Rafaela Flor, Raouf Guirguis, Richard Hoefer, Giovanni Lorenzin, Emanuel Petricoin, Fatah Kashanchi, Kylene Kehn-Hall, Paolo Lanzafame, Lance Liotta, Alessandra Luchini
PMCID: PMC7402046  PMID: 32766594

Abstract

Coronavirus disease 2019 (COVID-19), caused by the severe acute respiratory syndrome coronavirus-2 (SARS-CoV-2), became a pandemic in early 2020. Lateral flow immunoassays for antibody testing have been viewed as a cheap and rapidly deployable method for determining previous infection with SARS-CoV-2; however, these assays have shown unacceptably low sensitivity. We report on nine lateral flow immunoassays currently available and compare their titer sensitivity in serum to a best-practice enzyme-linked immunosorbent assay (ELISA) and viral neutralization assay. For a small group of PCR-positive, we found two lateral flow immunoassay devices with titer sensitivity roughly equal to the ELISA; these devices were positive for all PCR-positive patients harboring SARS-CoV-2 neutralizing antibodies. One of these devices was deployed in Northern Italy to test its sensitivity and specificity in a real-world clinical setting. Using the device with fingerstick blood on a cohort of 27 hospitalized PCR-positive patients and seven hospitalized controls, ROC curve analysis gave AUC values of 0.7646 for IgG. For comparison, this assay was also tested with saliva from the same patient population and showed reduced discrimination between cases and controls with AUC values of 0.6841 for IgG. Furthermore, during viral neutralization testing, one patient was discovered to harbor autoantibodies to ACE2, with implications for how immune responses are profiled. We show here through a proof-of-concept study that these lateral flow devices can be as analytically sensitive as ELISAs and adopted into hospital protocols; however, additional improvements to these devices remain necessary before their clinical deployment.

Full Text Availability

The license terms selected by the author(s) for this preprint version do not permit archiving in PMC. The full text is available from the preprint server.


Articles from medRxiv are provided here courtesy of Cold Spring Harbor Laboratory Preprints

RESOURCES