. 2020 Aug 1;32:106107. doi: 10.1016/j.dib.2020.106107

Comprehensive data on the mechanical properties and biodegradation profile of polylactide composites developed for hard tissue repairs

Abraham K Aworinde a,, Samson O Adeosun a,b, Festus A Oyawale a, Eyere Emagbetere c, Felix A Ishola a, Obafemi Olatunji d, Stephen A Akinlabi a,e, Sunday O Oyedepo a, Oluseyi O Ajayi a, Esther T Akinlabi a,d
PMCID: PMC7426527  PMID: 32817869

Abstract

Polylactide (PLA), a biopolymer, was reinforced with three fillers (two organic reinforcements and one inorganic filler). The processing technique used to fabricate the composites was the melt-blending technique. The composites and the unreinforced PLA were subjected to microhardness, compression and biodegradation characterisations. Data obtained are presented in this article as raw data. Data from microhardness and compression tests were used to predict the fracture toughness. The biodegradation of the composites was also examined, and the data obtained reported in this article. The data presented in this article allow for a comprehensive understanding of the mechanical behaviour and the biodegradation profile of three composites of PLA with respect to their applications as biodegradable implants. It also helps in the selection of fillers for biopolymers such as PLA.

Keywords: Compressive properties, Hard tissue regeneration, Melt-blending technique, Predicted fracture toughness, Vickers microhardness


Specifications table

Subject area Material science
More specific subject area Biomaterials
Type of data Table
How data were acquired Vickers microhardness tester, double column Instron universal testing machine with model number 3369 equipped with Bluehill software, and immersion in phosphate buffer solution
Data format Raw
Analysed
Parameter for data collection Geometric progression, with the first term and the common ratio being 0.5 g and 2 respectively, was used to obtain the fillers’ weight percentages. Parameters (such as strength, modulus, fracture toughness, biodegradation, etc.) for the design of a biodegradable implant were considered in the characterisations.
Description of data collection Melt-blending technique was used to develop three different PLA composites with chitin, chitosan and titanium (Ti-6Al-2Sn-2Mo-2Cr-0.25Si) powders as the reinforcements. Cylinders were produced from the composites and subjected to microhardness, compression and biodegradation tests.
Data source location Centre for Energy Research and Development, Obafemi Awolowo University, Ile-Ife, Nigeria.
University of Johannesburg, Auckland Park campus, South Africa.
Covenant University, Ota, Nigeria.
Data accessibility With the article
Related research article Aworinde, A. K., Adeosun, S. O., Oyawale, F. A., Akinlabi, E. T., & Akinlabi, S. A. (2019). The Strength characteristics of Chitosan- and Titanium- Poly (L-lactic) Acid Based Composites. Journal of Physics: Conference Series, 1378, 022061. https://doi.org/10.1088/1742-6596/1378/2/022061

Value of the data

  • These data are significant because they present the mechanical competencies and biodegradation profile of polylactide (PLA) reinforced with organic fillers (i.e., chitin and chitosan) and compared with PLA loaded with inorganic powder (i.e., Ti-6Al-2Sn-2Mo-2Cr-0.25Si) for applications in osteologic repairs

  • Researchers, orthopaedists and institutions that are interested in the application of accelerated biodegradation can benefit from these data because the data help to understand the extent to which organic and inorganic fillers could influence the mechanical properties of PLA and its biodegradation tendencies

  • These data can be reused for further insights and development of experiments by examining the influence of greater weight percentages of the fillers, organic ones in particular, on the biomechanical properties of PLA.

  • The dataset can be applied in the short or long term because bone fracture and issues associated with the need for biodegradable materials are not one-off issues

1. Data description

The unreinforced polylactide, chitosan reinforced polylactide, chitin reinforced polylactide, and titanium reinforced polylactide have been abbreviated to PLA, PLA/Ch, PLA/Ct and PLA/Ti respectively for the purpose of terseness. The choice of chitin and chitosan, as reinforcements, is based on their hydrophobicity-reduction tendencies [1] and track record in the biomedical applications [2], [3], [4]. Ti-6Al-2Sn-2Mo-2Cr-0.25Si was considered as a filler because all its alloying elements are biocompatible. Besides, a similar alloy of titanium (i.e., Ti–6Al–2Sn–2Zr–2Mo–2Cr–0.25Si) has shown evidence of high corrosion resistance [5]. The mechanical properties of pure PLA and its composites are presented in Section 1.1, while Section 1.2 presents the analyses of variance (two-way ANOVA) of all the mechanical properties considered in this article. Section 1.3 is a record of the data got from the biodegradation test. The negative values in Sections 1.1 and 1.2 are indicative of the percentage decrease.

The Vickers microhardness values (HV), are presented in Table 1. Table 2 depicts half diameter of the indented section (a), and the largest lateral extension of cracks (C) observed during the microhardness test. These Tables comprehensively describe the microhardness properties of all the samples considered in this article as well as the measurements of parameters relating to the observed cracks during indentations. Table 1, for instance, gives the measure of the resistance of the samples to plastic deformation during indentation while Table 2 details the quantitative values of half of the diameter of the impression made by the indenter and the largest lateral extension (C) of the observed cracks. C was obtained by the addition of a to the measured length of microcracks [6] observed through the measuring microscope of the Vickers microhardness tester. These values (i.e., HV, a and C) allude to the ductility of PLA and its composites and are precursors to the fracture toughness of the samples.

Table 1.

Vickers microhardness values.

Filler (wt%) Hardness value (HV)
% increase in hardness of
PLA/Ch PLA/Ct PLA/Ti PLA/Ct over PLA/Ch PLA/Ti over PLA/Ct PLA/Ti over PLA/Ch
0.00 68.6 68.6 68.6 0.0000 0.0000 0.0000
1.04 76.1 134 151.8 76.0841 13.2836 49.8683
2.08 90.2 135.5 161.8 50.2217 19.4096 44.2522
4.17 107.2 154.5 180.5 44.1231 16.8285 40.6094
8.33 148.4 213.1 268.1 43.5984 25.8095 44.6475
16.67 167.7 210.9 165.2 25.7603 -21.6690 -1.5133

Table 2.

Half-diameter of the indented section, a (mm) and largest lateral extension of the crack, C (mm).

Filler (wt%) a (mm)
C (mm)
PLA/Ch PLA/Ct PLA/Ti PLA/Ch PLA/Ct PLA/Ti
0.00 25.7750 25.7750 25.7750 53.0917 53.0917 53.0917
1.04 20.7667 17.5917 17.0083 59.9333 58.3250 59.8917
2.08 21.8417 19.0000 15.5667 60.5250 57.0167 56.7833
4.17 20.7167 17.9333 18.7917 73.0500 62.9500 63.5083
8.33 18.0000 15.0750 16.9083 61.9833 59.1250 59.9750
16.67 16.0750 17.0583 18.2250 64.0250 55.4917 62.2750

Tables 3 and 4 are data obtained from the compression test. Table 3 shows the ultimate compressive strength of the samples. The values were obtained at the maximum compressive load and are necessary values in the determination of the ability of the samples to withstand compressive loads. Compressive moduli are presented in Table 4. The values were the slope of the stress-strain curves within the elastic region. These values summarised the stress-strain behaviour of PLA and its developed composites. The modulus of any material for hard tissues repairs, e.g., bone internal fixations, can help to understand if there would be a modulus mismatch. Table 5 collates the predicted fracture toughness values using Eqs. (2) and (3). The fracture toughness values obtained using Eq. (2) only used data from Vickers microhardness test, whereas Eq. (3) used data obtained from both microhardness and compression tests. These values reveal the quantitative ability of the samples to resist fracture in the presence of cracks.

Table 3.

Ultimate compressive strength (UCS).

Filler (wt%) UCS (MPa)
% increase in UCS of
PLA/Ch PLA/Ct PLA/Ti PLA/Ct over PLA/Ch PLA/Ti over PLA/Ct PLA/Ti over PLA/Ch
0.00 24.75 24.75 24.75 0.0000 0.0000 0.0000
1.04 12.53 23.46 22.84 87.2565 −2.6450 45.1464
2.08 6.91 9.32 18.52 34.9998 98.6323 62.7079
4.17 13.82 4.89 13.44 −64.6299 174.8765 −2.8552
8.33 21.93 21.14 23.94 −3.6250 13.2707 8.3952
16.67 28.83 6.44 16.66 −77.6749 158.7551 −73.1084

Table 4.

Compressive modulus (Ec).

Filler (wt%) Ec (MPa)
% increase in Ec of
PLA/Ch PLA/Ct PLA/Ti PLA/Ct over PLA/Ch PLA/Ti over PLA/Ct PLA/Ti over PLA/Ch
0.00 522.18 522.18 522.18 0.0000 0.0000 0.0000
1.04 376.34 444.61 546.13 18.1400 22.8342 31.0898
2.08 357.24 400.90 424.42 12.2212 5.8685 15.8298
4.17 419.96 513.47 416.79 22.2680 −18.8294 −0.7601
8.33 536.29 709.73 637.59 32.3418 −10.1646 15.8884
16.67 841.94 812.21 863.77 −3.5312 6.3484 2.5275

Table 5.

Predicted fracture toughness.

Filler (wt. %) Fracture toughness from Eq. (2)
Fracture toughness from Eq. (3)
PLA/Ch PLA/Ct PLA/Ti PLA/Ch PLA/Ct PLA/Ti
0.00 0.1840 0.1840 0.1840 0.0410 0.0410 0.0410
1.04 0.1534 0.1598 0.1536 0.0352 0.0402 0.0434
2.08 0.1512 0.1654 0.1663 0.0345 0.0383 0.0409
4.17 0.1140 0.1425 0.1407 0.0340 0.0405 0.0366
8.33 0.1459 0.1566 0.1533 0.0416 0.0498 0.0447
16.67 0.1390 0.1722 0.1449 0.0503 0.0519 0.0502

1.1. Mechanical properties

Table 1, Table 2, Table 3, Table 4, Table 5.

1.2. Analysis of variance on the mechanical properties

Table 6, Table 7, Table 8, Table 9, Table 10 are the analyses (using two-way ANOVA) of the data on the Vickers hardness, ultimate compressive strength, compressive modulus, fracture toughness values obtained from Eq. (2), and fracture toughness values obtained using Eq. (3) respectively. The summary of the analyses of variance on the mechanical properties of the samples is in Table 12. These analyses relate two factors (i.e., the variation in the weight per cent of the fillers and differences in the structural strengths of the fillers) to the mechanical properties of PLA and its composites.

Table 11.

Analysis of variance on the fracture toughness values of composites obtained using Eq. (3).

Summary Count Sum Average Variance
0 3 0.123044 0.041015 0
1.04 3 0.11876 0.039587 1.73E−05
2.08 3 0.11367 0.03789 1.02E−05
4.17 3 0.111018 0.037006 1.07E−05
8.33 3 0.136021 0.04534 1.69E−05
16.67 3 0.152404 0.050801 9.66E−07
PLA/Ch 6 0.236597 0.039433 3.95E−05
PLA/Ct 6 0.261624 0.043604 3.28E−05
PLA/Ti 6 0.256696 0.042783 2.07E−05
ANOVA
Source of Variation SS df MS F P-value F crit
Rows 0.000412 5 8.23E-05 15.38525 0.000203 3.325835
Columns 5.86E−05 2 2.93E-05 5.47389 0.024792 4.102821
Error 5.35E−05 10 5.35E-06
Total 0.000524 17

Table 6.

Analysis of variance on the hardness properties of composites.

Summary Count Sum Average Variance
0 3 205.8 68.6 0
1.04 3 361.9 120.6333 1566.623
2.08 3 387.5 129.1667 1311.723
4.17 3 442.2 147.4 1381.03
8.33 3 629.6 209.8667 3589.863
16.67 3 543.8 181.2667 660.1633
PLA/Ch 6 658.2 109.7 1613.272
PLA/Ct 6 916.6 152.7667 2951.391
PLA/Ti 6 996 166 4068.268
ANOVA
Source of Variation SS df MS F P-value F crit
Rows 36544.94 5 7308.989 11.04125 0.000811 3.325835
Columns 10399.1 2 5199.549 7.854649 0.008903 4.102821
Error 6619.709 10 661.9709
Total 53563.75 17

Table 7.

Analysis of variance on the compressive strength of composites.

Summary Count Sum Average Variance
0 3 74.25 24.75 0
1.04 3 58.83 19.61 37.6909
2.08 3 34.75 11.58333 37.54003
4.17 3 32.15 10.71667 25.49863
8.33 3 67.01 22.33667 2.084033
16.67 3 51.93 17.31 125.6449
PLA/Ch 6 108.77 18.12833 69.71578
PLA/Ct 6 90 15 82.41636
PLA/Ti 6 120.15 20.025 20.50551
ANOVA
Source of Variation SS df MS F P-value F crit
Rows 483.5401 5 96.70802 2.547307 0.097684 3.325835
Columns 77.26888 2 38.63444 1.017638 0.396022 4.102821
Error 379.6481 10 37.96481
Total 940.4571 17

Table 8.

Analysis of variance on the compressive modulus of composites.

Summary Count Sum Average Variance
0 3 1566.526 522.1754 0
1.04 3 1367.086 455.6952 7299.444
2.08 3 1182.562 394.1872 1162.256
4.17 3 1350.219 450.0731 3017.178
8.33 3 1883.609 627.8698 7591.635
16.67 3 2517.926 839.3085 669.8638
PLA/Ch 6 3053.942 508.9903 32065.26
PLA/Ct 6 3403.1 567.1833 25588.82
PLA/Ti 6 3410.886 568.481 27675.96
ANOVA
Source of Variation SS df MS F P-value F crit
Rows 401024 5 80204.79 31.29792 8.49E−06 3.325835
Columns 13854.51 2 6927.255 2.703188 0.115212 4.102821
Error 25626.24 10 2562.624
Total 440504.7 17

Table 9.

Analysis of variance on the energy at maximum strength of composites.

Summary Count Sum Average Variance
0.00 3 4.71 1.5700 0.0000
1.04 3 1.68 0.5600 0.1101
2.08 3 0.84 0.2800 0.0553
4.17 3 1.24 0.4133 0.1332
8.33 3 2.69 0.8967 0.1204
16.67 3 1.77 0.5900 0.2457
PLA/Ch 6 4.20 0.7000 0.3489
PLA/Ct 6 3.13 0.5217 0.3073
PLA/Ti 6 5.60 0.9333 0.1577
ANOVA
Source of Variation SS df MS F P-value F crit
Rows 3.2515 5 0.6503 7.9489 0.0029 3.3258
Columns 0.5114 2 0.2557 3.1257 0.0882 4.1028
Error 0.8181 10 0.0818
Total 4.5811 17

Table 10.

Analysis of variance on the fracture toughness values of composites obtained using Eq. (2).

Summary Count Sum Average Variance
0 3 0.552485 0.184162 0
1.04 3 0.46719 0.15573 1.33E−05
2.08 3 0.483273 0.161091 7.22E−05
4.17 3 0.397511 0.132504 0.000255
8.33 3 0.456081 0.152027 3.01E−05
16.67 3 0.456375 0.152125 0.000315
PLA/Ch (Eq 2) 6 0.888166 0.148028 0.000516
PLA/Ct (Eq 2) 6 0.981267 0.163545 0.000201
PLA/Ti (Eq 2) 6 0.94348 0.157247 0.000252
ANOVA
Source of Variation SS df MS F P-value F crit
Rows 0.004204 5 0.000841 13.13114 0.000397 3.325835
Columns 0.000731 2 0.000365 5.706376 0.022215 4.102821
Error 0.00064 10 6.4E-05
Total 0.005576 17

Table 12.

Summary of the analysis of variance on the mechanical properties of the developed composites.

Mechanical property Effect of fillers’ weight per cent
Effect of fillers’ structural strengths
F P-value F crit F P-value F crit
Hardness 11.04 8.11×104 3.33 7.85 8.90×103 4.10
Compressive strength 2.55 9.77×102 3.33 1.02 3.96×101 4.10
Compressive modulus 31.30 8.49×106 3.33 2.70 1.15×101 4.10
Fracture toughness (Eq. 2) 13.13 3.97×104 3.33 5.71 2.22×102 4.10
Fracture toughness (Eq. 3) 15.39 2.03×104 3.33 5.47 2.48×102 4.10

1.3. Biodegradation Profile of the composites

In Table 13, the changes in mass observed during the biodegradation test are recorded. The initial masses of the samples, masses after four weeks and eventual masses after ten weeks of immersion of the samples in the phosphate buffer solution are all contained in Table 13. The masses gained or lost after four and ten weeks of immersion are shown in Table 14 in percentage terms. These data describe the onset of biodegradation by hydrolytic degradation [7].

Table 13.

Mass change during biodegradation.

Filler (wt%) Initial mass (g)
Mass (g) after 4 weeks
Mass (g) after 10 weeks
PLA/Ch PLA/Ct PLA/Ti PLA/Ch PLA/Ct PLA/Ti PLA/Ch PLA/Ct PLA/Ti
0.00 1.2473 1.2473 1.2473 1.3935 1.3935 1.3935 1.3251 1.3251 1.3251
1.04 1.2094 1.3011 1.1973 1.3607 1.4064 1.2164 1.2928 1.4163 1.2776
2.08 1.0173 1.3290 1.1813 1.1878 1.4636 1.2716 1.1709 1.4110 1.3653
4.17 1.1857 1.1184 1.2092 1.2473 1.2592 1.2991 1.2585 1.2248 1.3696
8.33 1.3278 1.3618 1.1859 1.4026 1.4878 1.2553 1.4511 1.4790 1.3485
16.67 1.2272 1.4951 1.3791 1.3857 1.6863 1.4607 1.3722 1.8330 1.5073

Table 14.

Percentage changes in mass after immersion in Phosphate Buffer Solution (PBS).

Filler (wt%) Mass (g) after 4 weeks
Mass (g) after 10 weeks
PLA/Ch PLA/Ct PLA/Ti PLA/Ch PLA/Ct PLA/Ti
0.00 11.7213 11.7213 11.7213 6.2375 6.2375 6.2375
1.04 12.5103 8.0932 1.5953 6.8960 8.8540 6.7068
2.08 16.7600 10.1279 7.6441 15.0988 6.1700 15.5761
4.17 5.1952 12.5894 7.4347 6.1398 9.5136 13.2650
8.33 5.6334 9.2525 5.8521 9.2860 8.6063 13.7111
16.67 12.9156 12.7884 5.9169 11.8155 22.6005 9.2959

2. Experimental design, materials and methods

Polylactide (PLA) with the molecular weight of 144 g/mol and the overall lactide purity ≥99.5% was purchased from NatureWorks, China. Chitin and chitosan were obtained via chemical extraction processes from shrimp shells while titanium powder (Ti–6Al–2Sn–2Mo–2Cr–0.25Si) was purchased from TLS Technik GmbH & Co. Spezialpulver KG, Bitterfeld-Wolfen, Germany. While PLA served as the matrix, chitin, chitosan and titanium powder served as the reinforcements. The matrix was melt-blended with each of the fillers at the weight percentages shown in Table 15. Although there are several polymer composites processing technique [8], [9], [10], [11], the melt-blending technique was used because it is environmentally benign, cost-effective, best for mass production, toxin-free [12] and allows for the addition of higher weight per cent of fillers. The weight per cent formulation of the fillers was obtained from the mass of the fillers (Table 15) according to Eq. (1). The stirring speed of 60 rpm was used to ensure a fairly homogenous mix of the fillers with the matrix. Each of the molten composites was mould-pressed at the pouring temperature of 290 °C to form solid cylinders with 12.5 mm diameter and 7.0 mm length.

Mn=arn1 (1)

where:

  • Mn = mass of the nth term

  • a = the starting mass of the filler (i.e., 0.5 g)

  • r = common ratio (i.e., 2)

  • n = nth term of mass of the filler

Table 15.

The formulation of the fillers’ weight per cent.

n PLA (g) PLA (wt. %) Filler (g) Filler (wt%)
0 88.0 100.00 0.0 0.00
1 87.5 98.96 0.5 1.04
2 87.0 97.92 1.0 2.08
3 86.0 95.83 2.0 4.17
4 84.0 91.67 4.0 8.33
5 80.0 83.33 8.0 16.67

2.1. Mechanical characterisation

The unreinforced and developed solid composites were subjected to Vickers microhardness test with indentation load of 100 kgf for 10 s dwell time (except for PLA/Ti at 16.67 wt% which took 15 s). The microhardness machine used was located at the Mechanical Engineering Science Department, University of Johannesburg, Auckland Park Campus, South Africa. The Vickers hardness values and lateral extended micro-cracks were measured and recorded. The compression test was done using a double column Instron universal testing machine with model number 3369 (equipped with Bluehill software for data acquisition) located at Centre for Energy Research and Development (CERD) at Obafemi Awolowo University, Ile-Ife, in Nigeria. Fracture toughness was predicted from the data obtained from compression and Vickers microhardness tests using Eqs. (2) and (3) [6].

KIC=0.0726PC32 (2)
KIC=0.0089(EHV)25(PaC12) (3)

where

  • KIC = fracture toughness (MPa.m0.5)

  • P = indentation loading (N)

  • E = Young's modulus (GPa)

  • HV = Vickers hardness (GPa)

  • a = half-diameter of the indented section (mm)

  • C = largest lateral extension of the crack (mm)

2.2. Biodegradation Test

20 ml of Phosphate Buffer Solution (PBS) with 7.4 pH was measured into each test tube and kept in an oven with a preset temperature of 36.5 °C. The test tubes were left in the oven for about 30 min to ensure the conditioning of the PBS to 36.5 °C [13]. The weighed samples were then immersed in 20 ml of PBS. The test tubes were returned into the oven, and the temperature maintained at 36.5 °C.

The biodegradation test was left on for ten (10) weeks. Changes in mass, which is considered as the progress of biodegradation [7], were measured after the first four (4) weeks and at the end of the tenth week. The percentage change in mass after weeks of immersion, MΔ, was calculated for every sample using Eq. (4).

MΔ=MfMiMi×100% (4)

where

  • MΔ=percentagechangeinmass

  • Mi=initialmassbeforeimmersion

  • Mf=finalmassafterimmersion

Declaration of Competing Interest

The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.

Acknowledgement

We are grateful for the financial support received from Covenant University, Ota, Ogun State in Nigeria.

Footnotes

Supplementary material associated with this article can be found, in the online version, at doi:10.1016/j.dib.2020.106107.

Appendix. Supplementary materials

mmc1.xml (398B, xml)

References

  • 1.Aworinde A.K., Adeosun S.O., Oyawale F.A., Akinlabi E.T., Akinlabi S.A. Comparative effects of organic and inorganic bio-fillers on the hydrophobicity of polylactic acid. Results Eng. 2020;5:1–3. doi: 10.1016/j.rineng.2020.100098. [DOI] [Google Scholar]
  • 2.Deepthi S., Sundaram M.N., Kadavan J.D., Jayakumar R. Layered chitosan-collagen hydrogel/aligned PLLA nanofiber construct for flexor tendon regeneration. Carbohydr. Polym. 2016;153:492–500. doi: 10.1016/j.carbpol.2016.07.124. [DOI] [PubMed] [Google Scholar]
  • 3.Aworinde A.K., Adeosun S.O., Oyawale F.A., Akinlabi E.T., Akinlabi S.A. The strength characteristics of chitosan- and titanium- poly (L-lactic) acid based composites. J. Phys. Dec. 2019;1378 doi: 10.1088/1742-6596/1378/2/022061. [DOI] [Google Scholar]
  • 4.Singh S.K. Solubility of lignin and chitin in ionic liquids and their biomedical applications. Int. J. Biol. Macromol. Jul. 2019;132:265–277. doi: 10.1016/j.ijbiomac.2019.03.182. [DOI] [PubMed] [Google Scholar]
  • 5.Ikumapayi O.M., Akinlabi E.T., Majumdar J.D., Akinlabi S.A. Influences of 17-4PH stainless steel and α + β titanium alloy powders for corrosion susceptibility on friction stir-processed AA7075-T651 aluminium matrix composites. J. Bio- Tribo-Corros. 2019;5(4):1–11. doi: 10.1007/s40735-019-0294-9. [DOI] [Google Scholar]
  • 6.Moradkhani A., Baharvandi H., Tajdari M., Latifi H., Martikainen J. Determination of fracture toughness using the area of micro-crack tracks left in brittle materials by Vickers indentation test. J. Adv. Ceram. 2013;2(1):87–102. doi: 10.1007/s40145-013-0047-z. [DOI] [Google Scholar]
  • 7.H. Azevedo and R. Reis, “Understanding the enzymatic degradation of biodegradable polymers and strategies to control their degradation rate,” in Biodegradable Systems in Tissue Engineering and Regenerative Medicine, 2004, pp. 177–202.
  • 8.Aworinde A.K., Adeosun S.O., Oyawale F.A., Akinlabi E.T., Emagbetere E. Proceedings of the International Conference on Industrial Engineering and Operations Management. 2018. Mechanical strength and biocompatibility properties of materials for bone internal fixation: a brief overview; pp. 2115–2126. [Google Scholar]
  • 9.Aworinde A.K., Adeosun S.O., Oyawale F.A., Akinlabi E.T., Akinlabi S.A. Parametric effects of fused deposition modelling on the mechanical properties of polylactide composites: a review. J. Phys. 2019;1378 doi: 10.1088/1742-6596/1378/2/022060. [DOI] [Google Scholar]
  • 10.Akpan E.I., Gbenebor O.P., Igogori E.A., Aworinde A.K., Adeosun S.O., Olaleye S.A. Electrospun porous bio-fibre mat based on polylactide/natural fibre particles. Arab J. Basic Appl. Sci. 2019;26(1):225–235. doi: 10.1080/25765299.2019.1607995. [DOI] [Google Scholar]
  • 11.Gbenebor O.P., Atoba R.A., Akpan E.I., Aworinde A.K., Adeosun S.O., Olaleye S.A. Minerals, Metals and Materials Series. M. & Materials Society; Cham: 2018. Study on polylactide-coconut fibre for biomedical applications; pp. 263–273. The Minerals, Ed. [Google Scholar]
  • 12.Xue B. Fabrication of super-tough ternary blends by melt compounding of poly(lactic acid) with poly(butylene succinate) and ethylene-methyl acrylate-glycidyl methacrylate. Composites Part B. 2019;172:743–749. doi: 10.1016/j.compositesb.2019.05.098. [DOI] [Google Scholar]
  • 13.Kokubo T., Takadama H. How useful is SBF in predicting in vivo bone bioactivity? Biomaterials. May 2006;27(15):2907–2915. doi: 10.1016/j.biomaterials.2006.01.017. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

mmc1.xml (398B, xml)

Articles from Data in Brief are provided here courtesy of Elsevier

RESOURCES