Seizure Localization Using EEG Analytical Signals

Mark H Myers a, Akaash Padmanabha b, Gavin M Bidelman a,c,d, James W Wheless e
PMCID: PMC7437564  NIHMSID: NIHMS1608830  PMID: 32682240

Abstract

Objective:

Localization of epileptic seizures, usually characterized by abnormal hypersynchronous wave patterns from the cortex, remains elusive. We present a novel, robust method for automatic localization of seizures on the scalp from clinical electroencephalogram (EEG) data.

Methods:

Seizure patient EEG data was decomposed via the Hilbert Transform and processed through the following methodology: sorting the analytic amplitude (AA) in the time instance, locating the maximum amplitude within the vector of channels, cross-correlating amplitude values in the time index with the channel vector. The channel with highest AA value in time was located.

Results:

Our approach provides an automated way to isolate the epi-genesis of seizure events with 93.3% precision and 100% sensitivity. The method differentiates seizure-related neural activity from other common EEG noise artifacts (e.g., blinks, myogenic noise).

Conclusions:

We evaluated performance characteristics of our source location methodology utilizing both phase and energy of EEG signals from patients who exhibited seizure events. Feasibility of the new algorithm is demonstrated and confirmed.

Significance:

The proposed method contributes to high-performance scalp localization for seizure events that is more straightforward and less computationally intensive than other methods (e.g., inverse source modeling). Ultimately, it may aid clinicians in providing improved patient diagnosis.

Keywords: analytic amplitude (AA), analytic phase (AP), seizure onset zone (SOZ), electroencephalograph (EEG)

1. Introduction

Epileptic seizures are a dysfunction in the brain that are characterized by simultaneous and periodic firing of large neural populations (Shoeb et al., 2004). Rapid localization of seizures is critical in medical environments to notifying medical staff and prevent further patient complications. Automatic localization of seizure events from EEG signals has been examined in several previous studies using a variety of approaches including machine learning (Shoeb and Guttag, 2010), frequency transforms (Hartmann et al., 2011), and nonlinear association analysis (Westmijse et al., 2009).

Knowledge about the exact seizure onset zone (SOZ) is of utmost importance for planning surgical treatment of epilepsy (Boon and D’Have, 1995; Boon et al., 1997). This task can be accomplished using either invasive or non-invasive methods. The problem is finding the exact location of the source of the seizure from the projections of electrodes scattered on the scalp. Although invasive methods (e.g., intracranial recordings) are generally more accurate and reliable because of the higher signal-to-noise ratio (SNR) of the signals, the higher risk of infection and the high cost of performing the required operations necessitates the exploration of alternative methods like the scalp EEG. A comprehensive review of the available methods is presented in Michel et al. (2004). The two most common approaches to EEG source analysis are the parametric beamformer method and non-parametric LORETA method.

The beamformer enables source localization by suppressing interfering activity of neighboring neural regions from the region of interest through spatial filtering. Source locations are found through dipole modelling, where the parameters used are location and orientation. A series of dipoles represents the spatial alignment across the brain. There are several types of beamformer methods (e.g., quiescent beamformer, linearly constrained minimum variance (LCMV) beamformer, and eigenspace-based beamformer). A mathematical computation of the LCMV is provided in Appendix A. Essentially, the LCMV beamformer applies a covariance matrix in order to suppress any interference from neighboring sources. Isolating regions of interest while suppressing interference from neighboring sources have also been utilized in other approaches. For example, bandpass filtering in the Fourier domain can isolate frequencies of interest within the EEG, while applying spatial low pass filters have been used to reduce interference from neighboring sources.

A related solution is to estimate the intracranial current densities associated with the electric potentials generated by active cortical pyramidal neurons. An example of source localization under this approach is the widely used low-resolution brain electromagnetic tomography (LORETA) and improved standardized low-resolution electromagnetic tomography (sLORETA) solutions (Pascual-Marqui et al., 2011; see Appendix A for details). sLORETA applies a statistical standardization of the current density values to scale cortical activity within a range of values. Groups of electrodes exhibiting similar cortical activity are clustered and displayed within a range of standardized values instead of a single source location. Despite their utility in estimating interictal spikes associated with SOZ foci (Song et al., 2015; Duez et al., 2019), a major limitation of both the LCMV and sLORETA approaches are that they (i) are computationally expensive and (ii) represent inverse estimates with non-unique solutions (i.e., source locations cannot be definitively determined using scalp data). Moreover, gradations in intensity from specific sources are difficult to view under these approaches because inverse methods only provide broad clusters of neural activity.

Here, we focus on monitoring both phase and energy of EEG signals in order to perform seizure localization. This method also has the potential to pin-point seizure occurrences on the cortex by the projected power of EEG signals and assuming a single source of power at a given instance in time. Existing methods of seizure location rely on the raw EEG analysis from an experienced clinician. Moreover, a robust automatic seizure location method should be sensitive, i.e., have high accuracy in locating all clinically locatable seizures and be reliable, i.e., have low false errors and not erroneously label non-seizure events (e.g., common EEG artefacts like eye-blinks, movements, etc.) as seizure occurrences. Thus, we compared our approach in its ability to distinguish true seizure events from other common artifacts of the EEG.

Our approach promotes the concept of quantitative EEG analysis (qEEG) to provide automated location based on neural markers. Spatial-temporal activity can be viewed as seizure activity producing an epicenter at the height of neural activity, as well as width varying gradations from the epicenter. The proposed method focuses on focal seizures that exhibit ~3 Hz activity and exhibit 5–10 times greater increase in amplitude from basal neural activity. Our method exploits the continuous analytic amplitude and phase information across EEG channels to robustly localize seizure foci on the scalp.

2. Methods

2.1. EEG data and preprocessing

This study used an existing, de-identified patient EEG database provided by the last author (J.W.W.), a board-certified pediatric neurologist specializing in epilepsy and its treatment, and Chief of Pediatric Neurology University of Tennessee Health Science Center (UTHSC). Original data collection protocols were approved by the UTHSC IRB (Protocol# 13–02783-XM). Raw EEGs were collected on an Electrical Geodesics Inc. (EGI) high-density EEG system. The dataset consisted of EEGs from N=15 patients (6 females, 9 males), who were, on average, ~20 years old. Patients in the sample were being treated at UTHSC for medically intractable epilepsy characterized by focal seizures consisting of abnormal hypersynchronous wave patterns.

EEG recordings included normal and abnormal (seizure) brain activity monitored for 60–90 minutes during examination, recorded at 128-channels according to the 10–20 system. Data were average referenced and sampled at 1000 Hz. Preprocessing included notch filtering to remove 60 Hz line noise and artifact removal whereby segments contaminated by myogenic noise and blinks (±50 μV) were automatically discarded using the algorithm described in the manuscript. Preprocessing was conducted using custom routines coded in MATLAB. In some instances, patient EEGs did not contain any seizure occurrences. This made it possible to first detect, and when present, locate seizure instances across channels. Confirmation of seizure location was provided by clinician judgments of J.W.W.’s neurology group.

Epileptic seizure occurrences were projected as synchronized oscillations with a specific frequency that was visible within a wide range of EEG channels. Figure 1 shows a sample of how this type of seizure was observed in raw EEG data.

Fig. 1.

Fig. 1.

An example of a seizure EEG waveform

In order to find the most effective frequency band characterizing seizure episodes, we first calculated the Fast Fourier Transform (FFT) of the EEG signals within the time limit of a sample seizure occurrence. FFTs showed peak power at ~3 Hz and additional harmonic energy extending up to 11 Hz. We then determined the dominant frequency of seizure oscillations across all patients, which was between 2.7 to 2.9 Hz. Consequently, we band-pass filtered data to include this nominal frequency range that described seizure activity (1–15 Hz; transition bandwidth=1Hz; N=370). Peak-to-peak pass-band ripple was < 0.02dB, which guaranteed signal distortion of <2%. A ±2% gain (0–20 Hz) and ±2% distortion was achieved within the passband. Filtering was implemented in the EEGLAB MATLAB toolbox (http://sccn.ucsd.edu/eeglab/) (Delorme and Makeig, 2004) using the Parks-McClellan optimal finite impulse response (FIR) filter design method (McClellan and Parks, 1973).

2.2. Analytical Amplitude Method

Our seizure localization routines followed three main steps: (1) Hilbert Transform, (2) calculating the analytic amplitude and phase differences per channel, and (3) comparing these metrics across time/channels to determine the most likely seizure epicenter.

The proposed method first employed the Hilbert Transform to calculate the power and phase of the EEG signals. The Hilbert transform is defined as (Eq. 1):

H(uj(t))=1πPV+Vj(t)(tt)dt (1)

where PV signifies the Cauchy Principal Value. The Hilbert transform is a linear operator like the FFT and is particularly appropriate for analyzing non-stationary signals by expressing frequency as a rate of change in phase. Calculating the Hilbert on all EEG signals yielded the analytical signals of each channel. The vector length at each digitizing step, t is the state variable for analytic amplitude. For each channel j, denoted as vj(t) from the recorded EEG, the analytical signal,Vj(t) is a time series which consists of a real part, vj(t) and an imaginary part uj (t)which together form the complex number:Vj(t) = vj(t)+iuj(t) The real part of the Hilbert transform is decomposed from the EEG signal to reflect the spatial ensemble average of the amplitude activity, vj(t), representing excitatory neurons, v(t) (Freeman, 2004). The Hilbert transform provides the imaginary part,uj(t) of the signal representing an estimate of the output of the activity of inhibitory neurons (Freeman, 2004). The sum of squares of the real and imaginary parts of the signal produces the instantaneous power required by the excitatory and inhibitory neurons in local areas of the cortex from dendritic currents (Freeman, 2004). The analytic amplitude is calculated through the square root of that sum of values.

From the Hilbert output, we then calculated the time-varying Analytical Amplitude [Aj(t) ] and Phase signals [Pj(t)]) of the EEG for each channel j during the time period of seizure using equations (2) and (3).

Aj(t)=vj2(t)+uj2(t) (2)

The square root of the sum of the squares from the real and imaginary parts of the signal produces the Analytical Amplitude Aj(t).

Pj(t)=tan1(uj(t)vj(t)) (3)

The arctangent of the real and imaginary parts of the signal produce the Analytical Phase Pj(t).

The phase difference at every instant was then computed by differentiating the consecutive phase values at every sample (Eq. 4).

ΔPj(t)=Pj(t)-Pj(t-1) (4)

We could not rely solely on the instantaneous values of Analytical Amplitude and Phase to locate the occurrence of seizures, due to spurious noise, movement and artefacts that may produce false-positives. Thus, a weighted summation function of the Analytical Amplitude and Phase Differences within a fixed time window acted as a memory function to predict if the specific period of data was representative of a seizure occurrence. Two indices were introduced to quantify the resemblance of signal to the seizure. The first index operates on the Analytical Amplitude (SAAj) and the second on the Analytical Phase Difference (SAPj):

SAAj(t)=k=0MeλkAj(tk)SAPj(t)=k=0MeλkΔPj(tk) (5)

where λ is the forgetting factor, M is the length of the window and their values determine the degree in which past data would affect the indices. The received signal t is subtracted from the transmitted signal k. These values were selected to filter out the transient effect of common artefacts on the analytical amplitude and phase signals that would typically last for several seconds. The weighted summation function was based on the recursive least-squares algorithm with an adaptive forgetting factor for optimal filtering. If the forgetting factor λ is large, the function has a long time series length to reduce the effects of spurious noise; if the forgetting factor λ is small then the function is adapted to the channel dynamics and can identify artefacts such as seizure instances (Zhuang, 1998). The weighted summation function is applied to the Analytical Amplitude [Aj] and Phase Differences [ΔPj]). Values of both indices were then compared to data driven patient specific thresholds to locate the seizure occurrences.

Because differences between seizure occurrences exist across patients, a fixed threshold could not be used for all individuals. We used the following formulation to scale individual threshold values:

SAthreshold=X×SDpatientSPthreshold=Y (6)

The values of coefficients (X, Y) were chosen experimentally by analyzing and comparing the patterns of changes for different occurrences of seizures and artefacts and selecting the most appropriate value. In order to compensate for cross-patient differences in signal amplitudes, we measured the standard deviation of a period of EEG data for each subject SDpatient ) and used that value to scale the thresholds for each patient. It should be noted that ( only the amplitude threshold (SAthreshold ) was variable between patients. In addition, differences between the two coefficients were due to the fact that SPthreshold was being compared to the phase differences, which tended to be small as compared to amplitude values. The algorithm was set to recognize seizure occurrences only when both conditions were met for a region of interest (ROI) of EEG channels:

SAAj>SAthresholdSAPj>SPthreshold (7)

The Analytical Amplitude (SAAj) and the Analytical Phase Difference (SAPj) values greater than (the threshold would signify a vector of channels that exhibit synchronous, low frequency (~3Hz), and high amplitude behavior.

2.3. Seizure localization

Having located seizure occurrences within the time dimension (Figure 2a), our next step was to find the most probable location of its occurrence. Assuming that a seizure is caused by a single source located on the cortex of the brain, we reasoned we could localize seizure activity by finding a constant value appearing through several consecutive channels across a phase difference diagram (see Figure 2c). Initially, we examined the analytic phase of all channels within a time sequence spanning the occurrence of a seizure event, and expected to see unified low frequency, high amplitude EEG energy during seizure states. We measured the corresponding maximum analytical amplitude values within the period of seizure (see Figure 2b). We used the analytical amplitude values from that instant to find the channel with the maximum value.

Fig. 2.

Fig. 2.

(a) Seizure activity from multiple channels, (b) analytic amplitude and (c) phase.

Analytic amplitudes were then sorted in time and cross-correlated to the channel vector which holds a row of analytic amplitudes per time segment. For all time segments, the maximum analytic amplitude was found, which was taken as the estimated epicenter of seizure intensity at that time instance.

The following steps summarize our seizure localization routine:

  1. Select EEG data with seizure occurrences for processing

  2. Apply Eqs. 27.

  3. Sort AA values with time, in order to associate associating highest AA to a time instance. iv. Find max AA value among all channels

  4. Cross-correlate AA index value in channel vector/electrode value with time instance.

The final output provided the location (channel) and time instance with highest AA value associated with seizure activity.

3. Results

3.1. Analytic amplitude and phase descriptions of seizure EEG

When present, seizure activity was identified as a clear boundary from low-amplitude, asynchronous activity to high-amplitude, rhythmic activity, where small changes in excitatory synaptic gain led to strong oscillatory activity. Among 128 electrodes, we found seizure activity were robust in ≥ 25 channels (~ 20% of electrodes). All 15 patients had focal seizures whose abnormal neural behavior was observable in their EEG. Figure 2b shows how the analytic amplitude and phase difference (Figure 2c) of the signal changed during the seizure period across all channels and time. Seizure oscillations manifested as an almost sinusoidal signal with a frequency of ~3 Hz and abnormal changes in power. Phase changes of the signal also became more organized, causing the phase difference diagram to transform from completely random changing values to small constant values that lingered for as long as the seizure event. The standard deviation across channels of the phase differences provided plateaus of low values punctuated by spikes of the analytic amplitude.

Figure 2a demonstrates the localized power of a seizure event across all electrodes at a single time instance. The phase of a signal that is approaching periodic behavior could be represented as continually increasing, and if there were two signals having the same sign, they would be reinforcing each other, whereas the phase difference between them would approach zero. We see the phase differences approached zero in Figure 2c during the seizure instance due to the semi-periodic behavior of the EEG during aberrant events. This instance is concentrated due to the phase differences of neighboring semi-periodic behaving signals.

3.2. Distinguishing seizure activity from EEG artifact

Another important challenge was to distinguish between EEG artefacts and true seizure events. Our method could effectively rule out common artefacts like bad electrode contacts and muscle or eye movements because these artefacts are usually time limited, are aperiodic, and do not induce phase synchronization. Figure 3 depicts this difference projected on the calculated Analytical Phase Index from Eq. 5. Figure 3a shows a typical artefact (i.e., voluntary head movement). Similarly, Figure 3b demonstrates the analytic amplitude and Figure 3c demonstrates the phase differences observed in artefact instances. The phase differences of seizure instances approach zero (depicted in blue in Figure 2c) since neighboring electrodes exhibit the same semi-periodic behavior as opposed to the phase differences of the artefacts which exhibit non-zero random values in Figure 3c.

Fig. 3.

Fig. 3.

(a) Artefact activity from multiple channels, (b) analytic amplitude and (c) phase.

The Analytical Phase Index (Eq. 5) shows a clear period of synchronization for the seizure event across all channels, while an artefact exhibits a sudden change in the amplitude synchronization observed briefly, and dissipates much faster than a seizure event. By choosing the appropriate threshold (Eq. 6) we could successfully rule out the artefacts by the location method itself. This capability of our method removes the need for further preprocessing for artefact removal which becomes an important limitation when the algorithm is desired to work in real-time.

Figure 4 shows changes in the analytic amplitude of the signal during the occurrence of seizures. The relative distance of the source from each channel is apparent. Figure 4a illustrates changes through time of the peak of the seizure location temporally. A 2-dimensional display of the analytic amplitude with time on the y-axis and channel/electrodes on the x-axis provides an overview of the intensity of the seizure as the higher amplitudes cluster around a group of channels. A magnified view shown in Figure 4b demonstrates how peak seizure occurrence concentrates into concentric circles and then slowly dissipates. Varying gradients of amplitude intensities can be seen in this magnified view.

Fig. 4.

Fig. 4.

Visualization of the Analytical Amplitude changes within the time period of the seizure occurrence. (a) complete time range, (b) zoomed view. The peak seizure occurrence is found in in the Analytical Amplitude which concentrates into concentric circles and then slowly dissipates over time.

3.3. Comparison of algorithm performance with other methods

Our EEG data consisted of time instances of seizures and artefacts. The number of artefacts existing in the data and the percentage of false positive seizure detections were assessed to examine how robust our method is in deciphering seizure occurrences vs. artefacts. EEG datasets were separated into possible seizure events and processed through our methodology. Most patients exhibited frontal lobe seizure activity (9/15). In some instances, more than one seizure event was detected in their EEG time series. The average duration of seizures was 10.8 sec. Our blinded source location results were then confirmed against expert clinical judgements of Dr. Wheless’ group.

We found the proposed algorithm achieved 93.3% precision and accuracy and 100% sensitivity in seizure detection. The specificity (true negative rate) was 0%, i.e., the algorithm produced no true negatives. Additionally, only 6% of EEG data resulted in false locations. Results from the localization analysis are summarized in Table 1. Chi Squared Goodness of Fit was performed to determine how empirically determined electrode locations compared to physician observed seizure locations. Calculated electrode locations showed excellent agreement with clinical judgements (X2=0.044, P=1.000). Our seizure location algorithm could not precisely locate one of the seizure locations (Patient 6). The false positive the algorithm produced may be due to the cluster occurrences in multi-focal locations, such that seizure activity was found in a cluster (F1, F3, F5, FC1, FC3 and FC5) vs. an epicenter.

Table 1.

Results from the seizure localization

Patient Gender Age Seizure Location TP/FP

1 M 15 T9 TP
2 F 21 F5 TP
3 M 17 P5 TP
4 M 22 CP1, FT10 TP
5 M 20 P8 TP
6 M 20 FP
7 F 15 Oz TP
8 F 14 FT9, F10 TP
9 F 38 FT8 TP
10 F 19 C4 TP
11 F 21 POz, FT7, P4, Cz TP
12 M 15 F8, T4 TP
13 M 17 Fp2 TP
14 M 21 F3 TP
15 M 20 Fp1 TP
*

TP (true positive) and FP (false positive), M (male), F (female)

As a further validation test, we compared our proposed algorithm of seizure localization to the well-known beamformer approach (for details, see Appendix A). The eigenspace-based beamformer algorithm achieved 73.3% precision and accuracy and 100% sensitivity. Only 11/15 seizures were located. The specificity or true negative rate was 0% (i.e. the algorithm produced no true negatives). Additionally, only 27% of the EEG seizure data set resulted in false locations. Results from the localization analysis are summarized in Table 2. The empirically determined electrode foci agreed with clinical judgments (X2=1.298, P=0.999). The false positive of the algorithm may again be due to cluster occurrences in multi-focal locations in some patients including Patient 6, Patient 4 (left centroparietal), Patient 5 (right posterior temporal), and Patient 11 (POz, FT7, P4, Cz).

Table 2.

Results from the seizure localization - Beamformer.

Patient Seizure Location TP/FP

1 T9 TP
2 F5 TP
3 PS TP
4 FP
5 FP
6 FP
7 Oz TP
8 rre. F10 TP
9 FT8 TP
10 C4 TP
n FP
12 F8.T4 TP
13 Fp2 TP
14 F3 TP
15 Fpl TP

3.3. Epicenter localization

A 3D visualization of the topography of analytical amplitude across the scalp is shown in Figure 5. Maximum analytic amplitude was taken as the epicenter of seizure intensity at a time instance. As seen in the representative patient in Fig. 5, localized seizure instances are observed with varying degrees of lesser intensities over the scalp. Results show that channel 22 (Fp1) is the source of the highest seizure activity, with neighboring channels 21 and 26. Neighboring electrodes exhibit varying gradations of color intensity, highlighting the focal point of seizure activity (dark red), and varying levels of lower intensities around the focal point.

Fig. 5.

Fig. 5.

Map of localization results using the Analytical Amplitude of EEG signals (Patient 15).

4. Discussion

We present a new approach to localizing seizure-related neural activity from EEG recordings based on analysis of analytic amplitude and phase differences over time and across channels. Results revealed localized epicenters of low frequency, high energy neural activity. Our method was robust in locating relatively short seizure instances (8–12 seconds) and achieved 93.3% precision and accuracy and 100% sensitivity in detecting true seizure activity from ongoing EEG and common artifacts.

EEG data showed that seizure activity was largely restricted to a bandwidth of periodic activity at ~ 3 Hz. Internally modulated neural activity is maintained through the selforganization amongst excitatory cortical neurons and inhibitory neurons (Freeman, 1975; 2004). When mutual inhibition exceeds that of mutual excitation, an oscillation repeats a complex pattern about every 0.3 s (Freeman 1972). Freeman has found through electrical stimulation of the lateral olfactory tract (LOT) in the prepyriform cortex (PC) of the cat, rat and rabbit models that seizures are induced that exhibit dominant localized oscillatory behavior (Freeman 1962, 1972). Biologically based models have been developed to capture the dynamics of localized seizure behavior and explain the dynamics of local and global neural networks (Freeman 1972, Kozma and Freeman 2001, Myers and Kozma, 2018). Dominant delta neural activity (~3 Hz) has been exhibited in which some inhibitory neurons become more disinhibited (excited) and others more inhibited (less active) (Freeman, 1986). If inhibitory neurons gain in intensity, thereby inhibiting their neighbors, excited regenerative feedback produces an explosive discharge. This result in an inhibitory post-synaptic potential in the excitatory neurons to which they are projected (Freeman et al. 2006). Seizure instances are sustained losses of mutual competitive activity that can manifest into explosive growth of local activity (Freeman et al. 2006).

Our technique reveals that localized neural behavior can be captured through signal decomposition (Hilbert amplitude and phase). The Hilbert transform provides temporal resolution of the near-instantaneous amplitude and phase (Freeman, 2004) and differentiated localized electrode intensities with the appropriate choice of band pass filter to isolate seizure activity. The inverse relationship demonstrated in Figure 2 of the phase differences and analytic amplitude confirms previous studies using the Hilbert and Fourier techniques (Freeman, 2006).

Another important property of our approach is that it cleanly differentiated aberrant neural activity of interest (seizure events) from common EEG artifacts. Our algorithm performed with 93.3% precision and accuracy and 100% sensitivity in segregating true seizure events. When random movements are produced that contribute to artefacts, phase differences produce non-zero random values (Fig. 3) and there is little coherence across neural neighboring areas. Compared to seizure events, we found that phase differences can distinguish seizure instances that are relatively short term due to the high coherence of neural activity among neighboring electrodes. The high-temporal resolution of our approach resolves the semi-periodic behavior of seizure instances, as brief as several seconds.

The epicenter of seizure activity is also highlighted through implementation of our novel algorithm. As neural activity transitions to seizure behavior, neural spiking activity increases across spatial and temporal instances. These events may also involve electromyographic activity (e.g., appendage and face movement) that can confound EEG recordings. In this regard, implementing a weighted summation function [Eq. 5] helped eliminate true negatives and reduced artefact occurrences. The analytic phase and amplitude approach compared favorably to the beamformer methodology in terms of artefact and noise suppression. Seizure thresholds were applied (Eq. 67) in order to distinguish seizure activity from background neural activity. The temporal changes of the epicenter of the seizure intensity are also visible using our method (Figs. 45). Our approach shows that maximum analytic amplitude can be used to determine the epicenter of seizure intensity on the scalp. Still, the effects of high-amplitude seizure instances are both localized and have gradient effects that are far reaching across the head (Fig. 5). As an analogy, similar effects would be expected during earthquake events near the epicenter.

Other source localization methods have been proposed in the literature (see Appendix A) and have been used to localize seizure activity with variable success. The eigenspace-based beamformer, for example, is superior to both the quiescent and the LCMV beamformer, as it combines a low white noise gain with a high suppression of the interfering source. However, the quiescent beamformer does not succeed in suppressing interfering 15 Hz source activity (Muthuraman et al., 2014). The eigenspace-based beamformer strongly suppresses interfering sources, yet it still maintains a low white noise gain. Additionally, the output signal of the eigenspace beamformer has slightly lower amplitude than the original source activity (Muthuraman et al., 2014). Similarly, LORETA source estimate errors can be measured by taking the distance between the location of the absolute maximum value (|zi | or |[J]i |) and the actual location of the test source (Pascual-Marqui, 2011). The related sLORETA algorithm shows zero localization error under ideal (no-noise) conditions as compared to the LORETA method (Pascual-Marqui, 2002; 2011). Still, a major limitation of both the LCMV and LORETA approaches are that they are computationally expensive and represent inverse estimates with nonunique solutions (i.e., source locations cannot be definitively determined using scalp data). Our “analytic signal” method performs well under ‘ideal’ and realistic noise conditions, considering that it was tested in noisy, biologically-based systems (i.e., real patients). Yet, our approach maintains a high precision, accuracy, and sensitivity rate while robustly localizing seizure activity at the scalp—which does not require the assumptions assumed by the aforementioned inverse source methods.

5. Conclusions

We developed a novel method to localize seizure-related activity using scalp EEG based on the Hilbert Transform and analytic amplitude/phase of neural signals. Empirical analysis on clinical seizure EEG data showed the proposed method is effective for fast and reliable localization of seizure activity. The method is applicable to seizures located in the cortex vs. seizure occurrences due to substructures within the brain where EEG and MRI techniques have been utilized. The algorithm achieved high precision, accuracy and sensitivity rates of localization due to optimal filtering and the introduction of a weighted summation function that reduced extraneous artefacts from the signal. The method was found to be robust in locating relatively short seizure instances (8–12 seconds), usually considered as intractable types of seizures where surgery and medication cannot be utilized. Our method may offer a new approach to aid clinicians in localization of seizure activity. Future studies are needed to determine the utility of the method to offer seizure diagnostics in real-time.

Supplementary Material

1

Highlights.

  • The work focuses on continuous monitoring of phase and energy of EEG signals in order to perform source location in real-time.

  • This method was found to be robust in locating relatively short seizure instances (8–12 seconds).

  • The algorithm achieved 93.3% precision and accuracy and 100% sensitivity.

Acknowledgements

G.M.B. was partially supported by NIH/NIDCD R01DC016267

Footnotes

Conflict of Interest

None of the authors have potential conflicts of interest to be disclosed.

Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.

References

  1. Boon P, D’Havé M, Adam C, Vonck K, Baulac M, Vandekerckhove T, Reuck JD. Dipole modeling in epilepsy surgery candidates. Epilepsia. 1997; 38(2):208–18. [DOI] [PubMed] [Google Scholar]
  2. Boon P, D’Havé M. Interictal and ictal dipole modelling in patients with refractory partial epilepsy. Acta Neurol Scand. 1995; 92(1):7–18. [DOI] [PubMed] [Google Scholar]
  3. Chavez M, Le Van Quyen M, Navarro V, Baulac M, Martinerie J. Spatio-temporal dynamics prior to neocortical seizures: amplitude versus phase couplings. IEEE Trans Biomed Eng. 2003; 50(5):571–583. [DOI] [PubMed] [Google Scholar]
  4. Delorme A, Makeig S. EEGLAB: an open source toolbox for analysis of single-trial EEG dynamics including independent component analysis. J Neurosci Methods. 2004; 134(1):9–21. [DOI] [PubMed] [Google Scholar]
  5. Duez L, Tankisi H, Hansen PO, Sidenius P, Sabers A, Pinborg LH, et al. Electromagnetic source imaging in presurgical workup of patients with epilepsy: A prospective study. Neurology. 2019;92:e576–e86. [DOI] [PMC free article] [PubMed] [Google Scholar]
  6. Freeman WJ. Alterations in prepyriform evoked potential in relation to stimulus intensity. Exp Neurol. 1962; 6:70–84. [DOI] [PubMed] [Google Scholar]
  7. Freeman WJ. Waves, pulses and the theory of neural masses. Progr Theor Biol. 1972; 2:87–165. [Google Scholar]
  8. Freeman WJ. Mass Action in the Nervous System. Academic Press, New York, 1975. [Google Scholar]
  9. Freeman WJ. Petit mal seizure spikes in olfactory bulb and cortex caused by runaway inhibition after exhaustion of excitation. Brain Research Reviews. 1986; 11(3):259–84. [DOI] [PubMed] [Google Scholar]
  10. Freeman WJ. Origin, structure, and role of background EEG activity. Part 1. Analytic amplitude. Clinical Neurophysiology. 2004; 115(9):2077–88. [DOI] [PubMed] [Google Scholar]
  11. Freeman WJ. Origin, structure, and role of background EEG activity. Part 4 Neural Frame Simulation. Clin Neurophysiol. 2006; 117:572–589. [DOI] [PubMed] [Google Scholar]
  12. Freeman WJ, Holmes MD, West GA, Vanhatalo S. Dynamics of human neocortex that optimizes its stability and flexibility. Int J Intell Syst. 2006; 21:1–21. [Google Scholar]
  13. Hartmann MM, Fürbaß F, Perko H, Skupch A, Lackmayer K, Baumgartner C, Kluge T. EpiScan: online seizure detection for epilepsy monitoring units. In: 2011. Annual International Conference of the IEEE Engineering in Medicine and Biology Society 2011 (pp. 6096–6099). IEEE. [DOI] [PubMed] [Google Scholar]
  14. Kozma R, Freeman WJ. Chaotic resonance—method and applications for robust classification of noisy and variable patterns. Int J Bifurc Chaos. 2001; 11:1607–1629. [Google Scholar]
  15. McClellan J, Parks T. A unified approach to the design of optimum FIR linear-phase digital filters. IEEE Trans Circuit Theory. 1973; 20(6):697–701. [Google Scholar]
  16. Michel CM, Murray MM, Lantz G, Gonzalez S, Spinelli L, de Peralta RG. EEG source imaging. Clin Neurophysiol. 2004; 115(10):2195–222. [DOI] [PubMed] [Google Scholar]
  17. Muthuraman M, Hellriegel H, Hoogenboom N, Anwar AR, Mideksa JG, Krause H, Raethjen J. Beamformer source analysis and connectivity on concurrent EEG and MEG data during voluntary movements. PLoS One. 2014; 9(3):e91441. [DOI] [PMC free article] [PubMed] [Google Scholar]
  18. Myers MH, Kozma R. Mesoscopic neuron population modeling of normal/epileptic brain dynamics. Cogn Neurodyn. 2018; 12(2):211–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
  19. Pascual-Marqui RD, Lehmann D, Koukkou M, Kochi K, Anderer P, Saletu B, Tanaka H, Hirata K, John ER, Prichep L, Biscay-Lirio R. Assessing interactions in the brain with exact low-resolution electromagnetic tomography. Philos Trans A Math Phys Eng Sci. 2011; 369(1952):3768–84. [DOI] [PubMed] [Google Scholar]
  20. Pascual-Marqui RD. Standardized low-resolution brain electromagnetic tomography (sLORETA): technical details. Methods Find Exp Clin Pharmacol. 2002; 24(Suppl D):5–12. [PubMed] [Google Scholar]
  21. Rao CR, Mitra SK. Theory and application of constrained inverse of matrices. SIAM J Appl Math. 1973; 24(4):473–88. [Google Scholar]
  22. Shoeb A, Edwards H, Connolly J, Bourgeois B, Treves ST, Guttag J. Patient-specific seizure onset detection. Epilepsy Behav. 2004; 5(4):483–98. [DOI] [PubMed] [Google Scholar]
  23. Shoeb AH, Guttag JV. Application of machine learning to epileptic seizure detection. InProceedings of the 27th International Conference on Machine Learning (ICML-10) 2010. (pp. 975–982). [Google Scholar]
  24. Song J, Davey C, Poulsen C, Luu P, Turovets S, Anderson E, et al. EEG source localization: Sensor density and head surface coverage. J Neurosci Meth. 2015;256:9–21. [DOI] [PubMed] [Google Scholar]
  25. Tikhonov AN, Goncharsky AV, Stepanov VV, Yagola AG. Numerical Methods for Solving Ill-Posed Problems Translation: Hoksbergen RAM (Kluwer, Boston: ). 1995. [Google Scholar]
  26. Van Hoey G, Van de Walle R, Vanrumste B, D’Havse M, Lemahieu I, Boon P. Beamforming techniques applied in EEG source analysis. Proc. ProRISC99. 1999;10:545–9. [Google Scholar]
  27. Varela F, Lachaux J-P, Rodriguez E, Martinerie J. The brainweb: phase synchronization and large-scale integration. Nat Rev Neurosci 2001; 2:229–239. [DOI] [PubMed] [Google Scholar]
  28. Westmijse I, Ossenblok P, Gunning B, Van Luijtelaar G. Onset and propagation of spike and slow wave discharges in human absence epilepsy: a MEG study. Epilepsia. 2009; 50(12):2538–48. [DOI] [PubMed] [Google Scholar]
  29. Zhuang W RLS Algorithm with Variable Fogetting Factor for Decision Feedback Equalizer over Time-Variant Fading Channels. Wirel Pers Commun. 1998; 8(1):15–29. [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

1

RESOURCES