Role of Internal Jugular Vein Ultrasound Measurements in the Assessment of Central Venous Pressure in Spontaneously Breathing Patients

A Systematic Review

Nicola Parenti 1,, Marco Scalese 2, Carmela Palazzi 1, Federica Agrusta 1, Jane Cahill 3, Giancarlo Agnelli 4
PMCID: PMC7440374  PMID: 32995230

Abstract

Background

Few studies have tested the role of the internal jugular vein (IJV) ultrasonographic (US) diameters in the assessment of central venous pressure (CVP) in spontaneously breathing patients. No review or meta-analysis is currently available on the role of IJV assessment in this setting. The aim of this systematic review is to check the reliability and accuracy of IJV US diameters in predicting CVP and to evaluate its correlation with CVP in spontaneously breathing patients.

Methods

This systematic review was based on the Preferred Reporting Items for Systematic Review and Meta-Analysis (PRISMA) guidelines. We included studies on the accuracy and reliability of the IJV ultrasound measures and studies exploring its correlation with CVP in adult spontaneously breathing patients. The studies’ report quality was assessed by Standards for Reporting of Diagnostic Accuracy (STARD) and Quality Assessment of Diagnostic Accuracy Studies (QUADAS)-2 scales.

Results

A total of five studies was eligible for final analysis. The studies on IJV ultrasound measures showed a good quality in reporting. The anterior-posterior diameter maximum of IJV (AP-IJV Dmax) showed the best correlation with the CVP with a good inter-rater reliability and validity in predicting CVP. All measures showed good inter-rater reliability and validity in predicting CVP, but only the AP-IJV Dmax showed good correlation with CVP.

Conclusions

The AP-IJV Dmax could be a potential surrogate of CVP because of its good reliability and validity in predicting CVP value and its fair-moderate correlation with CVP. Anyway, further research should confi rm these conclusions.

Keywords: systematic review, internal jugular vein ultrasound, reliability, validity

Introduction

The assessment of intravascular volume status (IVS) is one of the great challenges in the care of critically ill patients, particularly if presenting with shock. An accurate IVS evaluation is essential in this setting to properly manage fl uid therapy and responsiveness to treatment. Vital signs, biochemical markers, invasive and non-invasive tests are all theavailable tools to monitor IVS, even though according to many experts dynamic measures better predict fluid responsiveness. [1-4] Although the central venous pressure (CVP) does not seem to predict the volume status, [4] many studies used it as a gold standard to manage fluid therapy. However, the changes of CVP values should not be used to test the fluid responsiveness. [4]

In the last years, the ultrasound has been suggested as a valid method to guide the diagnosis and therapy in shock patients. [3] Among the ultrasound measures used as IVS surrogates, the diameter of inferior vena cava has been proposed as a feasible measureof fluid responsiveness in critically ill patients. [2-6] The ultrasound assessment of IVC diameter and respiratory variations was included in the 2008 American College of Emergency Physicians (ACEP) Policy Statement on Emergency Ultrasound Guidelines asa tool for IVS and CVP evaluation. The value of this assessment is supported by the pathophysiological observation that total body fluid volume correlate with absolute IVC diameter, while the variations in intra-thoracic and intra-abdominal pressure duringspontaneous breathing are able to induce IVC collapse and dilatation with inspiration and exhalation, respectively. The 2008 ACEP Policy Statement on Emergency Ultrasound Guidelines suggest measuring IVC inspiratory and expiratory diameter at 2 cm from itsjunction with the right atrium using a low-frequency probe (3.5–5 MHz) at the trans-abdominal subcostal view. The degree of inspiratory and expiratory diameters change is generally expressed as a percentage by the caval or IVC collapse index (CI) obtained by theIVC maximum diameter (IVC Dmax) minus the minimum (IVC Dmin) diameter of the IVC divided by the maximum diameter: IVC-CI = (IVC Dmax – IVC Dmin)/IVC Dmax. All these measures are obtained during one respiratory cycle (http://www.acep.org).

According to the ACEP Policy Statement on Emergency Ultrasound Guidelines, a “hypovolemic” patient presented a small IVC diameter and an IVC CI greater than 50%. In a recent review, Schmidt et al. suggested that a very small IVC diameter and a small and hyperkinetic left ventricle could be predictive of shock. [3]

Several studies have tested the relationship between the ultrasound measure of IVC diameters and the CVP, [7-11] but few reports are available in spontaneously breathing patients. [9,12,13] In particular, the IVC Dmax, IVC Dmin, the IVC distensibility, or cavalindex (IVC Dmax − IVC Dmin/IVC Dmax × 100, expressed as a percentage), IVC ratio (IVC maximum transversal/IVC maximum longitudinal) have been evaluated to estimate CVP mostly in spontaneously breathing critically ill patients bedside, during their admission to the emergency department or in patients undergoing right heart catheterization, but only few studies describe their scanning protocol and the inter-rater reliability. [9,12-17]

To our knowledge, few reports have been published on the role of internal jugular vein (IJV) ultrasonographic (US) measures as a tool to estimate the volume status and fluid responsiveness and to test their relation with the CVP. [9,11,18-20] The IJV ultrasoundmeasures should be obtained according to the methods described by previous researchers [9,11,18] with the patient in the supine position and a high frequency (5–10 MHz, linear array) linear transducer lightly placed on the neck in a transverse plane over the IJV 2 cm above the clavicle. Using a B-mode cine loop is possible to obtain the anterior-posterior of IJV (APIJV) and the transverse diameter of IJV (T-IJV). The following ultrasound measures of IJV have been tested: AP-IJV Dmax; the IJV area (maximal area in transversal section); the IJV ratio: IJV maximum transversal/IJV maximum longitudinal. [9,10,18-20]

To summarize, very few reports have tested the effectiveness of IJV ultrasound measures among spontaneously breathing patients.

To our there are no reviews on the role of the IJV in the assessment of volume status. Moreover, the conclusions of the reports published on the reliability, validity, and correlation with the CVP of ultrasound measures of IVC and IJV are divergent. For these reasons, we decided to conduct this systematic review.

The primary aim of this review was to check the available studies exploring one of the following measures: (1) the reliability of IJV ultrasound measures; (2) the validity of IJV ultrasound measures in predicting CVP; (3) the IJV ultrasound measures correlation with the CVP in spontaneously breathing critically ill patients. Another purpose was to assess the quality of selected studies published on this topic.

Methods

This systematic review was conducted according to the Preferred Reporting Items for Systematic Review and Meta-Analysis (PRISMA) statement forreporting systematic reviews and meta-analyses. [21] The flow diagram in Figure 1 schematically shows each step of the review process.

Figure 1. Review selection process.

Figure 1

A broad search of the literature was initially performed by an expert in literature searching using PubMed, Cochrane Library, Web of Science, and Scopus. All the articles available, starting from the year 1941 until the time of search (June 30, 2015), were considered.

An update of the literature search has been conducted from the June 1, 2016.

The literature search strategy and criteria are shown in Supplement 1.

A total of 322 records was retrieved (Figure 1).

Three researchers independently and in a blinded manner reviewed the four lists from the literature database (PubMed, Cochrane Library, Web of Science, and Scopus), and by the records’ title and abstract inspection, removed 254 duplicates and other 50 reports because they were not relevant to the aims of the review.

Eighteen full text articles were finally selected and independently reviewed by three researchers to identify those potentially relevant.

Only five studies met the inclusion criteria for the systematic review, exploring the correlation of IJV ultrasound measures with CVP and/or accuracy and/or reliability of US exams. The reasons for exclusion were justifi ed in each phase of the selection.

An appraisal of the reporting quality of the five remaining studies selected for the analysis, was independently conducted by three authors using the Standards for Reporting of Diagnostic Accuracy (STARD) [22] and the Quality Assessment of Diagnostic Accuracy Studies (QUADAS)-2 [23] guidelines.

A narrative summary was used to synthetize the data to provide a narrative description and ordering of the evidence, with commentary and interpretation.

The three reviewers’ yes/no level of agreement for each study was entered into an Excel 2010 (Microsoft Corporation, Redmond, WA, USA) spreadsheet, and Fleiss’ kappa for observed agreement was performed. We obtained a Fleiss’ kappa score of κ = 0.72, equating to a high level of agreement between the raters.

Outcome measures were the following: (1) inter-rater and intra-rater reliability coefficients (tested using kappa coefficient, weighted and/or un-weighted, intraclass correlation coefficient, Pearson correlation coeffi cient, and Spearman’s rank correlation coeffi cient);(2) the validity of IJV ultrasound measures in predicting CVP (tested using the following accuracy indexes: accuracy, sensitivity, specificity measure; receiver operator characteristic (ROC) curves with areas under the ROC curves [AUCs]). The prediction, measured by AUC(ROC curve), is the ability of the test (IVC and IJV ultrasound) to correctly classify those with and without the CVP (i.e., CVP < 10 mmHg = 0; CVP > 10 mmHg = 1); (3) coefficient of determination of regression analysis (R2); Pearson correlation coefficients (r) or Spearmanrank correlation coefficients (rho). Correlation is a statistical technique that can show whether and how strongly pairs of continuous variables (i.e., IJV ultrasound diameters or CVP values) are related.

Inclusion criteria were studies which tested the following measures: (1) the reliability of IJV ultrasound measures; (2) the validity of IJV ultrasound measures in predicting CVP; (3) the IJV ultrasound measures’ correlation with the CVP in spontaneously breathing critically ill patients.

We included studies conducted on all ages of adult patients (> 18 years) in all languages.

We excluded the duplicate studies.

Results

Out of a total pool of 322 collected records, five [9,11,18-20] studies were considered eligible for the analysis.

In this review, 233 patients have been enrolled, with an age mean range from 51 to 66 years; a percentage of male range from 45 to 63% (Table 1).

Table 1. Characteristics of studies on internal jugular veina.

AP-IJV: anterior-posterior of internal jugular vein; AUC: areas under the receiver operator characteristic curves; CI: confidence interval; CSA: cross-sectional area; CVP: central venous pressure; ICC: intraclass correlation coefficient; IJV: internal jugular vein; RAP: right atrial pressure; RIJV: right internal jugular vein; ROC: receiver operator characteristic; US: ultrasonographic.aTest characteristic: sensitivity, specificity, positive and negative predictive values.bThe IJV ratio: IJV maximum transversal/IJV maximum longitudinal.cThe IJV area: maximal area in transversal section.

Author Study design Target population US measures Outcome Results Conclusion
Prekker et al., 2013 [9] USA Cross-sectional 44 adultspontaneously breathing patients IJV ratiob ROC curvesLinear regression analysis R2 = 0.21 for IJV ratiobAccuracy of IJV in predicting CVP < 10mmHg: AUC = 0.76 (95% CI = 0.65–0.89) Poorly correlationModerate validity
Keller et al.,2009 [11]USA Prospectiveobservational 44 adultspontaneouslybreathing patients IJV ratiob Intra and inter observeragreement (Bland-Altmanplots) ROC curves Accuracy of IJV in predicting CVP < 8mmHg: AUC = 0.84 (95% CI = 0.72–0.96) Inter and intra-raterreliability moderateIJV ratiob accuratelyestimates CVP = 8 mmHg
Donahue et al., 2009 [18] USA Prospectiveobservational 34 adultspontaneouslybreathing patients AP-IJVIJV areac Inter-rater reliabilityCorrelation betweenCVP and ultrasound IJVmeasures (in severaldecubitus and respiratorycycle) Correlation of CVP and IJV supineposition and at the end of Expiration and IJV areac: r = 0.82 and 0.69IJV interclass correlation coefficientbetween the 2 observers = 0.92 Reliability excellentIf AP-IJV diameter < 7 mm then CVP < 10 cmH2O
Siva et al.,2012 [19]Australia 44 adultspontaneouslybreathing patients AP-IJV ROC curvesPearson correlation Significant correlation of anteriorposterior RIJV with CVPAccuracy of RIJV in predicting CVP, for overvolumed AUC = 0.73 (95% CI = 0.59–0.86), for undervolumed AUC =0.83 (95% CI = 0.70–0.96) Significant correlation and validity
Simon et al., 2010 [20] USA 67 adultspontaneouslybreathing patients IJV areac ROC curvesReliability (inter-intrarater)Pearson correlation Accuracy of RIJV area (with Valsalva)in predicting high RAP (> 12 mmHg):AUC = 0.86 (95% CI = 0.75–0.97)No significant correlation betweenresting RAP and RIJV CSA (r = 0.14)RIJV area: ICC intra = 0.88; ICC inter = 0.93 Intra- and inter-raterreliability very good

The study conducted by Prekker et al. [9] tested IVC and IJV ultrasound measures.

Many studies on IJV ultrasound measures showed good quality in reporting according toSTARD guidelines (Figure 2) and all records had similar quality according to QUADAS-2 tool (Table 2).

Figure 2. Comparison of quality of reporting among studies on internal jugular vein (IJV) (Standards for Reporting of Diagnostic Accuracy [STARD] results). The vertical axis shows the percentage of items from STARD score met by each study according to the judgement of three researchers who checked each report using the STARD tool.

Figure 2

Table 2. Quality Assessment of Diagnostic Accuracy Studies-2 results.

HR: high risk; LR: low risk; UR: unknown risk.

Study Risk of bias Applicability concerns
Patientselection Index test Referencestandard Flow andtiming Patientselection Index test Referencestandard
Prekker et al., 2013 [9] HR LR HR LR LR LR LR
Keller et al., 2009 [11] LR LR HR LR LR LR LR
Donahue et al., 2009 [18] LR UR HR LR LR LR LR
Siva et al., 2012 [19] UR UR LR LR LR LR LR
Simon et al., 2010 [20] LR LR HR LR LR HR HR

Among the reports included: four tested the correlation with CVP of IJV ultrasound diameters; five tested IJV validity in predicting CVP; three the reliability of IJV. All the studies on IJV enrolled spontaneously breathing patients.

The following ultrasound measures of IJV have been tested in the included studies: AP-IJV diameter (two studies); IJV area (two studies); IJV ratio (two studies).

The comparison of effectiveness of IJV ultrasoundmeasures is shown in Table 3.

Table 3. Comparison of internal jugular vein ultrasound measures effectivenessa.

AP-IJV: anterior-posterior of internal jugular vein; AUC: areas under the receiver operator characteristic curves; CI: confi dence interval; CVP: central venous pressure; ICC: intraclass correlation coeffi cient; IJV: internal jugular vein.aThe numbers near each value are the studies’ reference number which report the variables shown in the table.bThe IJV ratio: IJV maximum transversal/IJV maximum longitudinal.cThe IJV area: maximal area in transversal section.dThe authors tested the IJV’s validity in predicting CVP using several cut-off: CVP < or > 10 mmHg; CVP < or > 12 mmHg; < or > 8 mmHg. In particular the AP-IJV has been tested for undervolume (AUC = 0.83) and overvolume (AUC = 0.73);[19] the IJV ratio has been tested for CVP < 10 mmHg (AUC = 0.76) and CVP < 8 mmHg (AUC = 0.84);[9,11] the IJV area for CVP >12 mmHg (AUC = 0.86).[20]

AP-IJV IJV ratiob IJV areac
Reliability Inter-rater [18] ICC = 0.92 [18,20] ICC = 0.92–0.93
Reliability Intra-rater [20] ICC = 0.88
Correlation with CVP [18] r = 0.82 [9] R2 = 0.21 [18] r = 0.69
Validity in predictingCVP or overvolume orundervolumed [19] AUC = 0.73 (95% CI = 0.59–0.86)[19] AUC = 0.83 (95% CI = 0.70–0.96) [9] AUC = 0.76 (95% CI = 0.65–0.89)[11] AUC = 0.84 (95% CI = 0.72–0.96) [20] AUC = 0.86 (95% CI = 0.75–0.97)

There are no reports on the reliability of IJV ratio (Table 3).

The AP-IJV Dmax showed the best correlation with the CVP: r = 0.82 (Table 3).

All ultrasound measures of IJV showed good validity in predicting CVP; the AP-IJV Dmax and IJV area showed good inter-rater reliability.

Discussion

Few studies have been published on the relationship between the ultrasound measures of IJV diameters and the CVP as well as their reliability.

In this systematic review, the AP-IJV Dmax and IJV area showed the best correlation with the CVP, the best validity in predicting its values and a verygood inter-rater reliability.

However, we should be prudent when we use the CVP and the IJV ultrasound measures to manage fluid resuscitation because a recent meta-analysis [4] showed that the CVP does not seem to predict the volume status.

Furthermore, we should be cautious in interpreting the IJV ultrasound measures in some clinical contexts.

For example in patients with cor pulmonale, a high IJV diameter and a reduced caval index does not exclude fluid responsiveness.

In this clinical context, even CVP has the same problems of IJV, of being unreliable in predicting fluid responsiveness. For this reason, the specificity of IJV would have been lower if considering prediction of fluid responsiveness compared to CVP.

Our findings suggest that the AP-IJV Dmax and its area could be two new indexes useful in the management of fluid in resuscitation. In fact, both show good reliability and validity in predicting CVP values. Few recommend that further research should confirm these conclusions because existing research on these ultrasound measures is insufficient.

In particular further studies are needed assessing IJV in predicting fluid responsiveness.

To our knowledge, there are not reviews on the role of IJV ultrasound measures in the assessment of volume status and fluid responsiveness in critically ill patients.

However, it is difficult to compare the results of studies collected on this topic because of their different designs and statistical methodologies used to test the outcomes.

In our opinion, because the authors of studies included have chosen different thresholds to test the accuracy of IJV in predicting the CVP (Table 1), it is not possible to combine their results using a forest plot method.

Even though the studies included in the systematic review showed a moderate-good quality in reporting, especially according to QUADAS-2 guidelines.

Several limitations should be noted in the results of this review.

Firstly, a very small number of patients was enrolled in all the studies collected and secondly there was significant heterogeneity among patient populations, inclusion and exclusion criteria, design, and statistical methodology employed in each study: this could be a very important limitation in terms of reaching conclusions from this review.

Moreover, only few studies supported the findings and conclusions on the AP-IJV and IJV area.

Finally, because there are very few data on the reliability of the IJV ultrasound measurements and there could be a great operator-dependent variability, we should be very cautious before using these US indexes in clinical practice.

On the other hand, this systematic review is original because for the first time, data on the role of IJV in fluid management in critically ill patients has been collected. Moreover, to our knowledge, this is the first systematic review on the reliability of ultrasound measures of IJV. Finally, there are no existing reviews or meta-analyses on the comparison of performance of these ultrasound measures.

Although recent literature [4,24,25] suggests that CVP measurements do not accurately predict volume responsiveness, many International Society of Emergency and Critical Care suggest to use the CVP to manage the patients in shock [26,27] and the CVP remains the most frequently used variable to guide fluid resuscitation in critically ill patients. [28]

Moreover, many studies used CVP as gold standard to compare other indices to predict volume responsiveness.

Finally, according to De Backer and Vincent’s opinion, the CVP values provide important information about the cardiocirculatory status of the patient and should not be abandoned. [29]

For these reasons, we think that our findings on the effectiveness of IJV as a surrogate of CVP could impact the actual clinical practice; as they support the role of ultrasound to guide clinical management of patient in shock [3] and support further research on newultrasound measures as AP-IJV which could be a useful, fast, and safe tool to provide rapid fluid resuscitation early in the course of shock. The IJV ultrasound examination could be used in several settings by the emergency department physician, specifically when the IVC ultrasound method is not feasible because of poor acoustic windows (obesity, abdominal air interposition, surgical wounds).

Moreover, both IJV and IVC ultrasound measures are a very convenient method not requiring extensive training or expensive equipment.

However, the volaemic status is probably better assessed by an integrated sonographic approach with IVC and IJV, cardiac and pulmonary ultrasound.

In conclusion, the findings of this systematic review seem to suggest that the AP-IJV Dmax could be a good surrogate of CVP because of its good validity in predicting CVP but its correlation with CVP value among the studies collected is fair-moderate. Furtherstudies should test its reliability.

Finally, the AP-IJV Dmax and its area could be a promising alternative to be confirmed by future research on their effectiveness.

Conflicts of Interest Statement

The authors declare that they have not had conflicts of interest within the 36 months of submission.

Funding

For this work, we have not received funding from any of the following organizations: National Institutes of Health (NIH), Wellcome Trust, or Howard Hughes Medical Institutes (HHMI). The authors have not received funding from any other European or Italian organizations.

Supplement 1. Literature search: strategy and criteria

The following key words and strategy were used for the literature search.

(1) One search was performed using PubMed

Search field title: ((jugular veins[mesh] OR “internal jugular vein”) AND ultrasonography) AND (blood volume[mesh] OR “blood volume” OR “volume status” OR “volemic status” OR central venous pressure[Mesh] OR “central venous pressure”): 87 records.

(2) One search was performed using Web of Science

Search field title: Internal jugular vein AND (blood volume OR volume status OR volemic status OR central venous pressure): 117 records.

(3) One search using Scopus

Search field title: ((TITLE-ABS-KEY(internal jugular vein AND ultrasonography) AND TITLE-ABSKEY(blood VOLUME))) OR ((TITLE-ABS-KEY(internal jugular vein AND ultrasonography) AND TITLE-ABS-KEY(central venous pressure))) OR ((TITLE-ABS-KEY(internal jugular vein AND ultrasonography) AND TITLE-ABS-KEY(volemic status))) OR ((TITLE-ABS-KEY(internal jugular vein AND ultrasonography) AND TITLE-ABS-KEY(VOLUME status))): 116 records.

(4) One search using Cochrane Library

Search field title: internal jugular vein ultrasound: 2 records.

A total of 322 records were selected at the end of this research.

References

  • 1.Michard Frédéric, Teboul Jean-Louis. Predicting Fluid Responsiveness in ICU Patients. Chest. 2002 Jun;121(6) doi: 10.1378/chest.121.6.2000. [DOI] [PubMed] [Google Scholar]
  • 2.Bendjelid Karim, Romand Jacques-A. Fluid responsiveness in mechanically ventilated patients: a review of indices used in intensive care. Intensive Care Medicine. 2003 Jan 21;29(3) doi: 10.1007/s00134-002-1615-9. [DOI] [PubMed] [Google Scholar]
  • 3.Schmidt Gregory A., Koenig Seth, Mayo Paul H. Shock. Chest. 2012 Oct;142(4) doi: 10.1378/chest.12-1297. [DOI] [PubMed] [Google Scholar]
  • 4.Marik Paul E., Baram Michael, Vahid Bobbak. Does Central Venous Pressure Predict Fluid Responsiveness?*: A Systematic Review of the Literature and the Tale of Seven Mares. Chest. 2008 Jul;134(1) doi: 10.1378/chest.07-2331. [DOI] [PubMed] [Google Scholar]
  • 5.Zhang Zhongheng, Xu Xiao, Ye Sheng, Xu Lei. Ultrasonographic Measurement of the Respiratory Variation in the Inferior Vena Cava Diameter Is Predictive of Fluid Responsiveness in Critically Ill Patients: Systematic Review and Meta-analysis. Ultrasound in Medicine & Biology. 2014 May;40(5) doi: 10.1016/j.ultrasmedbio.2013.12.010. [DOI] [PubMed] [Google Scholar]
  • 6.Dipti Agarwal, Soucy Zachary, Surana Alok, Chandra Subhash. Role of inferior vena cava diameter in assessment of volume status: a meta-analysis. The American Journal of Emergency Medicine. 2012 Oct;30(8) doi: 10.1016/j.ajem.2011.10.017. [DOI] [PubMed] [Google Scholar]
  • 7.Barbier Christophe, Loubi�res Yann, Schmit Christophe, Hayon Jan, Ric�me Jean-Louis, Jardin Fran�ois, Vieillard-Baron Antoine. Respiratory changes in inferior vena cava diameter are helpful in predicting fluid responsiveness in ventilated septic patients. Intensive Care Medicine. 2004 Mar 18;30(9) doi: 10.1007/s00134-004-2259-8. [DOI] [PubMed] [Google Scholar]
  • 8.Feissel Marc, Michard Fr�d�ric, Faller Jean-Pierre, Teboul Jean-Louis. The respiratory variation in inferior vena cava diameter as a guide to fluid therapy. Intensive Care Medicine. 2004 Mar 25;30(9) doi: 10.1007/s00134-004-2233-5. [DOI] [PubMed] [Google Scholar]
  • 9.Prekker Matthew E., Scott Nathaniel L., Hart Danielle, Sprenkle Mark D., Leatherman James W. Point-of-Care Ultrasound to Estimate Central Venous Pressure. Critical Care Medicine. 2013 Mar;41(3) doi: 10.1097/ccm.0b013e31827466b7. [DOI] [PubMed] [Google Scholar]
  • 10.Deol Gur Raj, Collett Nicole, Ashby Andrew, Schmidt Gregory A. Ultrasound Accurately Reflects the Jugular Venous Examination but Underestimates Central Venous Pressure. Chest. 2011 Jan;139(1) doi: 10.1378/chest.10-1301. [DOI] [PubMed] [Google Scholar]
  • 11.Keller A. Scott, Melamed Roman, Malinchoc Michael, John Reverly, Tierney David M., Gajic Ognjen. Diagnostic accuracy of a simple ultrasound measurement to estimate central venous pressure in spontaneously breathing, critically ill patients. Journal of Hospital Medicine. 2009 Jul;4(6) doi: 10.1002/jhm.503. [DOI] [PubMed] [Google Scholar]
  • 12.Nagdev Arun D., Merchant Roland C., Tirado-Gonzalez Alfredo, Sisson Craig A., Murphy Michael C. Emergency Department Bedside Ultrasonographic Measurement of the Caval Index for Noninvasive Determination of Low Central Venous Pressure. Annals of Emergency Medicine. 2010 Mar;55(3) doi: 10.1016/j.annemergmed.2009.04.021. [DOI] [PubMed] [Google Scholar]
  • 13.Kircher Barbara J., Himelman Ronald B., Schiller Nelson B. Noninvasive estimation of right atrial pressure from the inspiratory collapse of the inferior vena cava. The American Journal of Cardiology. 1990 Aug;66(4) doi: 10.1016/0002-9149(90)90711-9. [DOI] [PubMed] [Google Scholar]
  • 14.Brennan J. Matthew, Blair John E., Goonewardena Sascha, Ronan Adam, Shah Dipak, Vasaiwala Samip, Kirkpatrick James N., Spencer Kirk T. Reappraisal of the Use of Inferior Vena Cava for Estimating Right Atrial Pressure. Journal of the American Society of Echocardiography. 2007 Jul;20(7) doi: 10.1016/j.echo.2007.01.005. [DOI] [PubMed] [Google Scholar]
  • 15.Simonson Jay S, Schiller Nelson B. Sonospirometry: A new method for noninvasive estimation of mean right atrial pressure based on two-dimensional echographic measurements of the inferior vena cava during measured inspiration. Journal of the American College of Cardiology. 1988 Mar;11(3) doi: 10.1016/0735-1097(88)91531-8. [DOI] [PubMed] [Google Scholar]
  • 16.Akkaya Arif, Yesilaras Murat, Aksay Ersin, Sever Mustafa, Atilla Ozge Duman. The interrater reliability of ultrasound imaging of the inferior vena cava performed by emergency residents. The American Journal of Emergency Medicine. 2013 Oct;31(10) doi: 10.1016/j.ajem.2013.07.006. [DOI] [PubMed] [Google Scholar]
  • 17.Saul Turandot, Lewiss Resa E., Langsfeld Alexis, Radeos Michael S., Del Rios Marina. Inter-rater Reliability of Sonographic Measurements of the Inferior Vena Cava. The Journal of Emergency Medicine. 2012 May;42(5) doi: 10.1016/j.jemermed.2011.05.095. [DOI] [PubMed] [Google Scholar]
  • 18.Donahue Sean P., Wood Joseph P., Patel Bhavesh M., Quinn James V. Correlation of sonographic measurements of the internal jugular vein with central venous pressure. The American Journal of Emergency Medicine. 2009 Sep;27(7) doi: 10.1016/j.ajem.2008.06.005. [DOI] [PubMed] [Google Scholar]
  • 19.Siva Brian, Hunt Austin, Boudville Neil. The sensitivity and specificity of ultrasound estimation of central venous pressure using the internal jugular vein. Journal of Critical Care. 2012 Jun;27(3) doi: 10.1016/j.jcrc.2011.09.008. [DOI] [PubMed] [Google Scholar]
  • 20.Simon Marc A., Kliner Dustin E., Girod John P., Moguillansky Diego, Villanueva Flordeliza S., Pacella John J. Detection of elevated right atrial pressure using a simple bedside ultrasound measure. American Heart Journal. 2010 Mar;159(3) doi: 10.1016/j.ahj.2010.01.004. [DOI] [PubMed] [Google Scholar]
  • 21.Liberati A., Altman D. G, Tetzlaff J., Mulrow C., Gotzsche P. C, Ioannidis J. P A, Clarke M., Devereaux P J, Kleijnen J., Moher D. The PRISMA statement for reporting systematic reviews and meta-analyses of studies that evaluate healthcare interventions: explanation and elaboration. BMJ. 2009 Jul 21;339(jul21 1) doi: 10.1136/bmj.b2700. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Bossuyt Patrick M., Reitsma Johannes B., Bruns David E., Gatsonis Constantine A., Glasziou Paul P., Irwig Les M., Moher David, Rennie Drummond, de Vet Henrica C.W., Lijmer Jeroen G. The STARD Statement for Reporting Studies of Diagnostic Accuracy: Explanation and Elaboration. Annals of Internal Medicine. 2003 Jan 07;138(1) doi: 10.7326/0003-4819-138-1-200301070-00012-w1. [DOI] [PubMed] [Google Scholar]
  • 23.Whiting Penny F. QUADAS-2: A Revised Tool for the Quality Assessment of Diagnostic Accuracy Studies. Annals of Internal Medicine. 2011 Oct 18;155(8) doi: 10.7326/0003-4819-155-8-201110180-00009. [DOI] [PubMed] [Google Scholar]
  • 24.Kumar Anand, Anel Ramon, Bunnell Eugene, Habet Kalim, Zanotti Sergio, Marshall Stephanie, Neumann Alex, Ali Amjad, Cheang Mary, Kavinsky Clifford, Parrillo Joseph E. Pulmonary artery occlusion pressure and central venous pressure fail to predict ventricular filling volume, cardiac performance, or the response to volume infusion in normal subjects. Critical Care Medicine. 2004 Mar;32(3) doi: 10.1097/01.ccm.0000114996.68110.c9. [DOI] [PubMed] [Google Scholar]
  • 25.Magder Sheldon, Bafaqeeh Fahad. The Clinical Role of Central Venous Pressure Measurements. Journal of Intensive Care Medicine. 2007 Jan;22(1) doi: 10.1177/0885066606295303. [DOI] [PubMed] [Google Scholar]
  • 26.Rhodes Andrew, Evans Laura E., Alhazzani Waleed, Levy Mitchell M., Antonelli Massimo, Ferrer Ricard, Kumar Anand, Sevransky Jonathan E., Sprung Charles L., Nunnally Mark E., Rochwerg Bram, Rubenfeld Gordon D., Angus Derek C., Annane Djillali, Beale Richard J., Bellinghan Geoffrey J., Bernard Gordon R., Chiche Jean-Daniel, Coopersmith Craig, De Backer Daniel P., French Craig J., Fujishima Seitaro, Gerlach Herwig, Hidalgo Jorge Luis, Hollenberg Steven M., Jones Alan E., Karnad Dilip R., Kleinpell Ruth M., Koh Younsuck, Lisboa Thiago Costa, Machado Flavia R., Marini John J., Marshall John C., Mazuski John E., McIntyre Lauralyn A., McLean Anthony S., Mehta Sangeeta, Moreno Rui P., Myburgh John, Navalesi Paolo, Nishida Osamu, Osborn Tiffany M., Perner Anders, Plunkett Colleen M., Ranieri Marco, Schorr Christa A., Seckel Maureen A., Seymour Christopher W., Shieh Lisa, Shukri Khalid A., Simpson Steven Q., Singer Mervyn, Thompson B. Taylor, Townsend Sean R., Van der Poll Thomas, Vincent Jean-Louis, Wiersinga W. Joost, Zimmerman Janice L., Dellinger R. Phillip. Surviving Sepsis Campaign Critical Care Medicine. 2017 Mar;45(3) doi: 10.1097/ccm.0000000000002255. [DOI] [Google Scholar]
  • 27.Emergency Ultrasound Guidelines. Annals of Emergency Medicine. 2009 Apr;53(4) doi: 10.1016/j.annemergmed.2008.12.013. [DOI] [PubMed] [Google Scholar]
  • 28.Cecconi Maurizio, Hofer Christoph, Teboul Jean-Louis, Pettila Ville, Wilkman Erika, Molnar Zsolt, Della Rocca Giorgio, Aldecoa Cesar, Artigas Antonio, Jog Sameer, Sander Michael, Spies Claudia, Lefrant Jean-Yves, De Backer Daniel. Fluid challenges in intensive care: the FENICE study. Intensive Care Medicine. 2015 Jul 11;41(9) doi: 10.1007/s00134-015-3850-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.De Backer Daniel, Vincent Jean-Louis. Should we measure the central venous pressure to guide fluid management? Ten answers to 10 questions. Critical Care. 2018 Feb 23;22(1) doi: 10.1186/s13054-018-1959-3. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Acute Medicine are provided here courtesy of Taiwan Society of Emergency Medicine