Abstract
Purpose
Multileaf collimator (MLC) tracking is being clinically pioneered to continuously compensate for thoracic and pelvic motion during radiotherapy. The purpose of this work was to characterize the performance of two MLC leaf‐fitting algorithms, direct optimization and piecewise optimization, for real‐time motion compensation with different plan complexity and tumor trajectories.
Methods
To test the algorithms, both in silico and phantom experiments were performed. The phantom experiments were performed on a Trilogy Varian linac and a HexaMotion programmable motion platform. High and low modulation VMAT plans for lung and prostate cancer cases were used along with eight patient‐measured organ‐specific trajectories. For both MLC leaf‐fitting algorithms, the plans were run with their corresponding patient trajectories. To compare algorithms, the average exposure errors, i.e., the difference in shape between ideal and fitted MLC leaves by the algorithm, plan complexity and system latency of each experiment were calculated.
Results
Comparison of exposure errors for the in silico and phantom experiments showed minor differences between the two algorithms. The average exposure errors for in silico experiments with low/high plan complexity were 0.66/0.88 cm2 for direct optimization and 0.66/0.88 cm2 for piecewise optimization, respectively. The average exposure errors for the phantom experiments with low/high plan complexity were 0.73/1.02 cm2 for direct and 0.73/1.02 cm2 for piecewise optimization, respectively. The measured latency for the direct optimization was 226 ± 10 ms and for the piecewise algorithm was 228 ± 10 ms. In silico and phantom exposure errors quantified for each treatment plan demonstrated that the exposure errors from the high plan complexity (0.96 cm2 mean, 2.88 cm2 95% percentile) were all significantly different from the low plan complexity (0.70 cm2 mean, 2.18 cm2 95% percentile) (P < 0.001, two‐tailed, Mann–Whitney statistical test).
Conclusions
The comparison between the two leaf‐fitting algorithms demonstrated no significant differences in exposure errors, neither in silico nor with phantom experiments. This study revealed that plan complexity impacts the overall exposure errors significantly more than the difference between the algorithms.
Keywords: fitting algorithm, MLC tracking, radiotherapy, real‐time
1. Introduction
One of the main advantages of radiation therapy as opposed to other types of cancer treatment is that the treatment is noninvasive and highly targeted to the tumor. Despite strong evidence that the ITV‐based planning technique (Internal Target Volume planning, ICRU 621) provides safe radical treatment for stage I nonsmall cell lung carcinoma, there are no guarantees that the tumor will remain within the planned aperture throughout the entire treatment.2, 3
New delivery approaches have been introduced to improve the targeting of the tumor during treatment. These techniques come in various forms, either by shifting the therapeutic beam to the tumor using a robotic arm CyberKnife,4, 5 a gimballed linear accelerator (Vero),6, 7 or the multileaf collimator (MLC)8, 9, 10 or by adjusting the patient couch (couch tracking).11
Real‐time MLC tracking is a novel technique that optimizes the leaf positions within the head of the linear accelerator to shift the radiation beam multileaf collimator leaves according to tumor motion. It has been implemented preclinically in several institutions on commercial linear accelerators12, 13, 14 or developed into in‐house control software and leaf‐fitting algorithms.8, 10, 12, 15, 16, 17 Real‐time MLC tracking has been clinically pioneered with three clinical trials leading to the first MLC tracking treatment for prostate18, 19, 20 and stereotactic lung21 with results reported in previous publications.19, 21
The current clinically used version of MLC tracking relies on a leaf‐fitting optimization algorithm (also known as “MLC tracking algorithm”) named “direct optimization” algorithm.22 A recent publication by Moore et al.17 introduced an alternative MLC tracking algorithm named “piecewise optimization algorithm”. With the current design of the piecewise algorithm, Moore et al. investigated its performances in silico using standard tumor motion (three patients) and intensity‐modulated radiation therapy (IMRT) plans. However, in silico tests do not always reflect the real‐life clinical situation. For that reason, their respective performances should be tested utilizing a linear accelerator with a broad range of tumor motions and MLC plan complexity.
To allow a thorough performance comparison between both algorithms in a clinical setting, the piecewise algorithm was implemented in the clinical version of the MLC tracking software. The aim of this work was to characterize the performance of two MLC leaf‐fitting algorithms used in real‐time motion compensation. This will be done both in silico and experimentally, spanning a range of tumor motions and treatment plans with varying degree of MLC modulation.
The significance of this paper is that it is the first to investigate and experimentally compare two MLC tracking algorithms in the identical clinical setting on a linear accelerator.
2. Method
2.A. Principle of multileaf collimator tracking algorithms
Multileaf collimator tracking is operated via an optimization algorithm tasked with finding the best‐fitted leaf positions given a set of various constraints (finite leaf width and speed), or constraints setup by the user prior to treatment delivery, such as prescribing various tolerances or radiobiological properties to the organs‐at‐risk to avoid excessive overdosing.
The mechanism for managing these setup constraints differs between the direct and piecewise optimization algorithm. The different components of the direct optimization algorithm can be found in Ruan et al.22, while more extensive explanations on the piecewise algorithm can be found in Moore et al.17 Although both algorithms allow the MLC leaf positions to be optimized according to the radio‐sensitivity factor attributed to different OAR (connoted as and constraints in the respective papers17, 22), each algorithm deals with spatial variance differently. The optimization process is operated for the direct optimization on a pixel basis within the beam's eye view, therefore relying on a two‐dimensional map of the organs.
The main difference between the two algorithms is that the piecewise algorithm deals with spatial variance by having an arbitrary number of volumetric ROI (Regions of Interest), hence accounting for the radio‐sensitivity in three dimensions. In both cases, this implies that an a priori knowledge of the position and volume of OAR is available prior to treatment, or that each OAR is being localized in real‐time during the treatment delivery.
The second difference is the way each algorithm deals with the quantification of exposure area that is sought to be minimized. For the direct optimization, the cost function is integrated both along and perpendicular to the leaf motion, as opposed to the piecewise algorithm where the algorithm resolves the integration linearly in one direction, solely along the leaf motion direction. Linear integral implicates that the algorithm is expected to converge faster toward a solution with the piecewise algorithm given equivalent set of constraints.
2.B. Experiments to assess and compare the algorithm performances
To characterize the performances of the algorithms, a series of in silico simulations and phantom experiments were performed. Both algorithms were tested under identical conditions assuming homogenous dose conditions: the target is considered as a rigid, nondeformable body and the underdose and overdose weights are set to be equal. Variables included the tumor motion, treatment site, and plan complexity. Comparison of algorithm performance was based on exposure errors, plan complexity, and the system latency. Figure 1 provides an overview of the method to assess the performance of each algorithm both in silico and experimentally on a linear accelerator. Further details are provided below.
2.B.1. In silico and phantom experiments
The in silico experiments were performed on a Latitude E7450 i7 2.60 GHz Dell 16 Gb RAM using an MLC simulator.23 The tumor motion traces were imported into the simulator as text files. The DICOM plan was read by the software and the treatment delivery was simulated. The simulator leaf speed was limited to 3.6 cm/s being the leaf speed of the actual linear accelerator.
The phantom experiments were performed on a Trilogy (Varian, Palo Alto, CA, USA) linear accelerator. Tumor motion traces were loaded into the HexaMotion programmable motion platform (Scandidos, Uppsala, Sweden) and triggered to start 10 s before the beginning of the beam delivery to allow training of the prediction algorithm. Calypso electromagnetic transponders (Varian Medical System, Palo Alto, CA, USA) were embedded into the HexaMotion platform, with a research version of the Calypso system sending the target position to the MLC tracking system. The kernel density estimation algorithm24 currently used was clinically used for the lung trajectories.
2.B.2. Tumor motion
To span the type of tumor motion observed during radiation therapy, thoracic and pelvic tumor motion traces were selected from published databases to be characteristic three‐dimensional (3D) motion patterns for those sites. Four types of motion were chosen for the lung25 from a CyberKnife study, and four motion patterns for the prostate26 obtained from a study with patients implanted with Calypso electromagnetic transponders. These tumor motion traces were categorized and named according to their characteristic pattern in previous study. Thoracic motion patterns were categorized as typical tumor motion, high‐frequency breathing, a predominantly lateral motion, and characterized baseline shift. The represented prostate motion patterns were continuous drift, high‐frequency excursions, erratic tumor motion, and stable tumor position.
2.B.3. Treatment plans
For each clinical site (lung and prostate), a selection of treatment plans used for previous MLC tracking experiments27 were delivered that differed in MLC modulation to span the plan complexity expected during clinical practice. Two plans, low and highly modulated VMAT plan, were selected for each site, by varying the set of competitive objectives on the target and OARs. All arcs spanned a 358° revolution with the collimator set at 90° (i.e., with the leaves parallel to longitudinal target motion). All plans were prescribed to deliver 2 Gy to 95% of the Planning Target Volume.
2.B.4. Evaluation of plan complexity
With MLC tracking, the plan complexity is known to complicate the task of the algorithm for the leaves to reach the desired positions.28, 29 Therefore, for each of the four plans lung/prostate and modulation high/low modulations, their complexity needed to be quantified. The plan complexity was evaluated based on four parameters:
MU weighted Average adjacent Leaf Distance (ALDw29), previously shown to correlate with MLC tracking performance28
The average leaf travel for each plan, considering solely the leaves that contribute to the open leaves aperture.28
The average area over circumference 28 with formula
The VMAT modulation score (MCs) by Masi et al.30
2.B.5. Measuring the system latency
Multileaf collimator tracking latency represents the inherent time delay between the tumor motion and the finished movement of the leaves to align the beam and the tumor. While execution of both MLC tracking algorithms possesses some inherent amount of latency, it is expected that a faster algorithm will be able to reduce the overall system latency.
The latency was evaluated using the setup described in Sawant et al.31 A ball bearing was moving in a superior–inferior direction along the parallel motion of a circular shape radiation field during which EPID images were acquired at 15 Hz operated on the computer console equipped with a 2.27 GHz Intel Xeon E5520 processor and 4 GB RAM. The ball bearing was placed onto the HexaMotion platform embedded with the Calypso electromagnetic beacons. For each optimization algorithm, EPID projections were obtained over 10 periods. Since both the ball bearing and the leaves move in a sinusoidal motion, the two structures were segmented from the EPID and a sinusoidal fit was used to calculate the temporal offset between the centroid of the ball and the MLC aperture. The latency was then calculated as the time delay between the ball position and the segmented MLC aperture.
2.C. Comparing MLC tracking algorithm performances based on leaf‐fitting exposure errors
To compare both performances, the exposure errors (overdose + underdose) were quantified in the beam's eye view using a framework developed by Poulsen et al.32
The mismatched area between the actual and planned MLC aperture represents the total amount of exposure errors which can be separated into individual sources of errors, namely the exposure errors due to width of the leaves, their speed and prediction algorithm errors when in use.
For each experiment, the exposure errors were computed using the fitted MLC positions obtained from the MLC tracking software. The fitted MLC positions corresponded to the given MLC positions fitted by the algorithm, thereby accounting for the width of the leaves but regardless of their physical speed. Focusing solely on the fitted MLC position dismisses any potential source of uncertainties allowing for a more direct comparison between the algorithms.
For each paired experiment, the exposure errors throughout the treatment arc were compared between each other using the Pearson correlation coefficient and root‐mean‐square error to evaluate the differences in exposure errors for each control point. Figure 2 provides an example of the exposure errors for a “paired experiment”, representing identical experimental conditions (same plan and tumor). For each experiment, these exposure errors were computed using the resulting tumor tracking logs and fitted MLC position updated at 30 Hz into text files. Exposure error computation was achieved using MATLAB (R2017a, Math Works).
3. Results
3.A. Quantification of exposure errors for each optimization algorithm
The average exposure errors for in silico low/high modulation were 0.66/0.88 cm2 for direct optimization and 0.66/0.88 cm2 for piecewise optimization. For the phantom experiment, it was 0.73/1.02 cm2 for direct and 0.73/1.02 cm2 for piecewise optimization. The side‐by‐side exposure errors displayed in Fig. 3 suggests that both algorithms performed equivalently spanning a large range of tumor motion, plan complexity, and treatment site.
The analysis of the in silico experiments demonstrated that the Pearson correlation coefficient for both algorithms is higher than r = 0.96 for all sets of organs and trajectories. The similar data obtained during linac experiments also showed strong correlation (r > 0.9) in most cases. The mean root‐mean‐square errors (RMSE) between paired algorithms were 0.10 cm2 for the in silico and 0.18 cm2 for the phantom experiments. High correlation and small RMSE error suggest strong relationship between paired experiment results for all types of trajectory and plan complexity, indicating that both algorithms performed equivalently.
3.B. Relationship between plan complexity and exposure errors
The quantified simulated and phantom experiment exposure errors for each treatment plan established that the exposure errors from the high modulation plan (0.96 cm2, 2.88 cm2 95% percentile) were all significantly different from the low modulation (0.70 cm2, 2.18 cm2 95% percentile) (P < 0.001, two‐tailed, Mann–Whitney statistical test). The descriptive metrics used to quantify the plan complexity are summarized in Table 1.
Table 1.
Lung | Prostate | |||
---|---|---|---|---|
High modulation | Low modulation | High modulation | Low modulation | |
Field MU | 596 | 342 | 737 | 422 |
ALDw | 0.71 cm | 0.20 cm | 1.40 cm | 0.70 cm |
Leaf travel | 0.19 cm | 0.04 cm | 0.30 cm | 0.22 cm |
AoC | 0.34 | 0.75 | 0.49 | 0.92 |
MCS | 0.07 | 0.17 | 0.15 | 0.28 |
AoC, Area over Circumference; ALDw, Average adjacent Leaf Distance; MCs, modulation score; MU weighted.
The average distance to adjacent leaves and leaf travel distance was shown to increase with plan complexity while the modulation score (MCs) and relative area over circumference decreases with plan complexity. These results provide further evidence of the impact of treatment complexity on the exposure errors.
3.C. Latency
The latency for the direct optimization was 226 ± 10 ms and for the piecewise algorithm 228 ± 10 ms. These physical latencies can be compared with the fitting latency within the software. Across all the plans and tumor motion, the in silico fitting latency for the direct optimization algorithm was 12.2 ± 5 ms, compared with the piecewise algorithm computed as 3.1 ± 1 ms. Despite these differences, the fitting time between algorithms did not impact the overall latency of the experimental setup, only capable of detecting uncertainties within ±10 ms.
4. Discussion
The goal of this study was to characterize the performance of two MLC tracking algorithms for radiotherapy in a realistic simulated and clinical environment. Both algorithms were tested alternatively in silico and experimentally on a linear accelerator for the range of organ motion and plan complexity that may be expected during clinical practice.
This is the first time that two MLC tracking algorithms were experimentally compared in the identical clinical setting on a linear accelerator. Moore et al.1 tested the performances of the piecewise algorithm in silico with IMRT plans as a proof of concept. However, leaf‐fitting is one part of the larger MLC tracking framework, and while in silico validation is a valuable tool to demonstrate proof of concept, the ultimate test is experimental investigation. Experimental investigation captures the impact of the leaf‐fitting algorithm with other software and hardware subsystems (e.g., compatibility issues with the Calypso tracking system, error catching, beam‐hold assertion, or constant rotating gantry during VMAT). For these reasons, this paper presents the first empirical comparison between the two algorithms.
We found that the plan complexity and tumor motion patterns have a much larger impact on dosimetric fidelity than the leaf‐fitting algorithms. The implication is that there are bigger gains to be made by improved planning than developing more complex or faster algorithms.
The implementation and development of faster MLC tracking algorithms is therefore potentially marginalized by the prerequisite to reduce plan complexity or improve the hardware capabilities. Hardware enhancement has been investigated under diverse forms. Pommer et al.28 investigated the dosimetric impact of finer leaves by testing alternatively a Varian Novalis Tx with Millennium MLC (5 mm leaf with) and High‐Definition MLC (2.5 mm leaf width). Using reflective markers and the ExacTrac (Brainlab, Germany) to provide positional input to the tracking system, they found that finer leaves improved the tracking accuracy compared with 5 mm leaf width. The Varian TrueBeam system equipped with High‐Definition MLC also provides MLC tracking capabilities in developer mode, but no performance analysis or dosimetric comparisons with other systems have been published to date.
Falk et al.29 found that leaf position constraints can be setup within the treatment planning system during planning optimization to limit the movement of the leaves during planning. Other hardware enhancement, such as dynamic alignment of the collimator angle,33 hybrid couch‐MLC tracking strategies34 improves MLC tracking accuracy by reducing the exposure errors for both prostate and lung.
Using a 2D time‐resolved framework for performance analysis provides a fast and reliable comparison of exposure errors. This method offers a point‐by‐point analysis that conceptually facilitates the search of exposure errors and allows a straightforward comparison between multiple plan parameters within a single fixed analysis framework. Also, the analysis of exposure errors for MLC tracking has been shown to be correlated with dosimetric errors for lung and prostate32, 35 using gamma failure and root‐mean‐square errors.
An application where MLC tracking is uniquely capable of motion compensation is tracking deforming targets and deforming systems, e.g., a primary tumor and regional nodes for locally advanced lung and prostate cancer radiotherapy. Preliminary studies using the direct optimization algorithm have investigated experimental target deformation and multitarget tracking.36 These experiments have been carried out on a linear accelerator using phantoms by mapping the deformation field in the linear accelerator beam's eye view and optimizing the fitting process accordingly.
The treatment plans and the tumor motion traces are included as supplementary materials (Data S1) to allow other groups to benchmark their algorithms against the results shown here.
5. Conclusion
The performance of two MLC tracking algorithms was characterized and compared using a 2D time‐resolved framework in a clinical realistic scenario. The comparison was based on the quantification of fitted exposure errors attributed by the optimization algorithm solely, regardless of the speed of the leaves. Our results showed that the two algorithms performed similarly and provide equivalent quality‐of‐fit for the scenarios evaluated. The main source of error can be attributed to the complexity of the plan, quantified prior to plan delivery, which was shown to greatly impact on the MLC tracking accuracy.
Conflicts of interests
Paul J Keall and Jeremy Booth are investigators on one completed and two ongoing MLC tracking clinical trials that have been partially supported by Varian Medical Systems. Paul J Keall and Amit Sawant are inventors on one licensed patent and one unlicensed patent related to MLC tracking. Paul J Keall, Ricky O'Brien, and Vincent Caillet gratefully acknowledge funding from the Australian Cancer Research foundation. Paul J Keall acknowledges funding from an Australian Government NHMRC Senior Principal Research Fellowship.
Supporting information
Acknowledgment
Paul J Keall and Ricky O'Brien gratefully acknowledge funding from the Australian Cancer Research Foundation. Paul J Keall acknowledges funding from an Australian Government NHMRC Senior Principal Research Fellowship. Jeremy Booth thanks Varian Medical Systems for provision of research systems used in this study.
References
- 1. ICRU . International Commission on Radiation Units and Measurements. Prescribing, recording and reporting photon beam therapy (Supplement to ICRU Report 50). ICRU Report 62. Oxford: Oxford University Press; 1999. [Google Scholar]
- 2. Wang L, Hayes S, Paskalev K, et al. Dosimetric comparison of stereotactic body radiotherapy using 4D CT and multiphase CT images for treatment planning of lung cancer: evaluation of the impact on daily dose coverage. Radiother Oncol. 2009;91:314–324. [DOI] [PubMed] [Google Scholar]
- 3. Koto M, Takai Y, Ogawa Y, et al. A phase II study on stereotactic body radiotherapy for stage I non‐small cell lung cancer. Radiother Oncol. 2007;85:429–434. [DOI] [PubMed] [Google Scholar]
- 4. Adler JR Jr, Chang SD, Murphy MJ, et al. The Cyberknife: a frameless robotic system for radiosurgery. Stereotact Funct Neurosurg. 1997;69:124–128. [DOI] [PubMed] [Google Scholar]
- 5. Prévost J‐B, Hoogeman MS, Praag J, et al. Stereotactic radiotherapy with real‐time tumor tracking for non‐small cell lung cancer: clinical outcome. Radiother Oncol. 2009;91:296–300. [DOI] [PubMed] [Google Scholar]
- 6. Depuydt T, Poels K, Verellen D, et al. Initial assessment of tumor tracking with a gimbaled linac system in clinical circumstances: a patient simulation study. Radiother Oncol. 2013;106:236–240. [DOI] [PubMed] [Google Scholar]
- 7. Mukumoto N, Nakamura M, Yamada M, et al. Intrafractional tracking accuracy in infrared marker‐based hybrid dynamic tumour‐tracking irradiation with a gimballed linac. Radiother Oncol. 2014;111:301–305. [DOI] [PubMed] [Google Scholar]
- 8. Fast MF, Nill S, Bedford JL, Oelfke U. Dynamic tumor tracking using the Elekta Agility MLC. Med Phys. 2014;41:111719. [DOI] [PubMed] [Google Scholar]
- 9. Falk M, af Rosenschöld PM, Keall P, et al. Real‐time dynamic MLC tracking for inversely optimized arc radiotherapy. Radiother Oncol. 2010;94:218–223. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Keall P, Kini VR, Vedam SS, Mohan R. Motion adaptive x‐ray therapy: a feasibility study. Phys Med Biol. 2001;46:1. [DOI] [PubMed] [Google Scholar]
- 11. Lang S, Zeimetz J, Ochsner G, Schmid Daners M, Riesterer O, Klöck S. Development and evaluation of a prototype tracking system using the treatment couch. Med Phys. 2014;41:021720. [DOI] [PubMed] [Google Scholar]
- 12. Tacke MB, Nill S, Krauss A, et al. Real‐time tumor tracking: automatic compensation of target motion using the Siemens 160 MLC. Med Phys. 2010;37:753–761. [DOI] [PubMed] [Google Scholar]
- 13. Krauss A, Nill S, Tacke M, Oelfke U. Electromagnetic real‐time tumor position monitoring and dynamic multileaf collimator tracking using a Siemens 160 MLC: geometric and dosimetric accuracy of an integrated system. Int J Radiat Oncol Biol Phys. 2011;79:579–587. [DOI] [PubMed] [Google Scholar]
- 14. Hansen R, Ravkilde T, Worm ES, et al. Electromagnetic guided couch and multileaf collimator tracking on a TrueBeam accelerator. Med Phys. 2016;43:2387–2398. [DOI] [PubMed] [Google Scholar]
- 15. Davies G, Poludniowski G, Webb S. MLC tracking for Elekta VMAT: a modelling study. Phys Med Biol. 2011;56:7541. [DOI] [PubMed] [Google Scholar]
- 16. Davies G, Clowes P, Bedford JL, Evans PM, Webb G, Poludniowski G. An experimental evaluation of the Agility MLC for motion‐compensated VMAT delivery. Phys Med Biol. 2013;58:4643. [DOI] [PubMed] [Google Scholar]
- 17. Moore D, Ruan D, Sawant A. Fast leaf‐fitting with generalized underdose/overdose constraints for real‐time MLC tracking. Med Phys. 2016;43:465–474. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. Keall PJ, Colvill E, O'Brien R, et al. The first clinical implementation of electromagnetic transponder‐guided MLC tracking. Med Phys. 2014;41:020702. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19. Colvill E, Booth JT, O'Brien RT, et al. Multileaf collimator tracking improves dose delivery for prostate cancer radiation therapy: results of the first clinical trial. Int J Radiat Oncol Biol Phys. 2015;92:1141–1147. [DOI] [PubMed] [Google Scholar]
- 20. Keall P, Nguyen DT, O'Brien R, et al. Stereotactic prostate adaptive radiotherapy utilising kilovoltage intrafraction monitoring: the TROG 15.01 SPARK trial. BMC Cancer. 2017;17(1):180. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Caillet V, Colvill E, Szymura K, Stevens M, Booth J, Keall P. SU‐G‐JeP1‐05: Clinical Impact of MLC Tracking for Lung SABR. Med Phys. 2016;43:3648–3649. [Google Scholar]
- 22. Ruan D, Keall P. Dynamic multileaf collimator control for motion adaptive radiotherapy: An optimization approach. in Power Engineering and Automation Conference (PEAM), 2011 IEEE. 2011. IEEE.
- 23. Poulsen PR, Ravkilde T, O'Brien RT, Keall PJ. Experimentally validated simulator of dynamic MLC tracking treatments: a tool for tracking QA. Int J Radiat Oncol Biol Phys. 2013;87:S45. [Google Scholar]
- 24. Ruan D, Keall P. Online prediction of respiratory motion: multidimensional processing with low‐dimensional feature learning. Phys Med Biol. 2010;55:3011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25. Suh Y, Dieterich S, Cho B, Keall PJ. An analysis of thoracic and abdominal tumour motion for stereotactic body radiotherapy patients. Phys Med Biol. 2008;53:3623. [DOI] [PubMed] [Google Scholar]
- 26. Langen KM, Willoughby TR, Meeks SL, et al. Observations on real‐time prostate gland motion using electromagnetic tracking. Int J Radiat Oncol Biol Phys. 2008;71:1084–1090. [DOI] [PubMed] [Google Scholar]
- 27. Keall PJ, Sawant A, Cho B, et al. Electromagnetic‐guided dynamic multileaf collimator tracking enables motion management for intensity‐modulated arc therapy. Int J Radiat Oncol Biol Phys. 2011;79:312–320. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28. Pommer T, Falk M, Poulsen PR, Keall PJ, O'Brien RT, af Rosenschöld PM. The impact of leaf width and plan complexity on DMLC tracking of prostate intensity modulated arc therapy. Med Phys. 2013;40:111717. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29. Falk M, Larsson T, Keall P, et al. The dosimetric impact of inversely optimized arc radiotherapy plan modulation for real‐time dynamic MLC tracking delivery. Med Phys. 2012;39:1588–1594. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30. Masi L, Doro R, Favuzza V, Cipressi S, Livi L. Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med Phys. 2013;40:071718. [DOI] [PubMed] [Google Scholar]
- 31. Sawant A, Dieterich S, Svatos M, Keall P. Failure mode and effect analysis‐based quality assurance for dynamic MLC tracking systems. Med Phys. 2010;37:6466–6479. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32. Poulsen PR, Fledelius W, Cho B, Keall P. Image‐based dynamic multileaf collimator tracking of moving targets during intensity‐modulated arc therapy. Int J Radiat Oncol Biol Phys. 2012;83:e265–e271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33. Murtaza G, Toftegaard J, Ullah Khan E, Poulsen PR. Volumetric modulated arc therapy with dynamic collimator rotation for improved multileaf collimator tracking of the prostate. Radiother Oncol. 2017;122:109–115. [DOI] [PubMed] [Google Scholar]
- 34. Toftegaard J, Hansen R, Ravkilde T, Macek K, Poulsen PR. An experimentally validated couch and MLC tracking simulator used to investigate hybrid couch‐MLC tracking. Med Phys. 2017;44:798–809. [DOI] [PubMed] [Google Scholar]
- 35. Ravkilde T, Keall PJ, Grau C, Høyer M, Poulsen PR. Time‐resolved dose distributions to moving targets during volumetric modulated arc therapy with and without dynamic MLC tracking. Med Phys. 2013;40:111723. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36. Ge Y, O'Brien RT, Sheih C‐C, Booth JT, Keall PJ. Toward the development of intrafraction tumor deformation tracking using a dynamic multi‐leaf collimator. Med Phys, 2014;41:061703. [DOI] [PMC free article] [PubMed] [Google Scholar]