Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2020 Aug 25;15(8):e0238196. doi: 10.1371/journal.pone.0238196

Interlaboratory comparison for the Filovirus Animal Nonclinical Group (FANG) anti-Ebola virus glycoprotein immunoglobulin G enzyme-linked immunosorbent assay

Michael S Anderson 1,*, Nancy A Niemuth 1, Carol L Sabourin 1,¤, Christopher S Badorrek 2, Callie E Bounds 2, Thomas L Rudge Jr 1
Editor: John Schieffelin3
PMCID: PMC7447032  PMID: 32841291

Abstract

The need for an efficacious vaccine against highly pathogenic filoviruses was reinforced by the devastating 2014–2016 outbreak of Ebola virus (EBOV) disease (EVD) in Guinea, Sierra Leone, and Liberia that resulted in over 28,000 cases and over 11,300 deaths. In addition, the 2018–2020 outbreak in the Democratic Republic of the Congo currently has over 3,400 cases and over 2,200 deaths. A fully licensed vaccine and at least one other investigational vaccine are being deployed to combat this EVD outbreak. To support vaccine development and pre-clinical/clinical testing a Filovirus Animal Nonclinical Group (FANG) human anti-EBOV GP IgG ELISA was developed to measure anti-EBOV GP IgG antibodies. This ELISA is currently being used in multiple laboratories. Reported here is a characterization of an interlaboratory statistical analysis of the human anti-EBOV GP IgG ELISA as part of a collaborative study between five participating laboratories. Each laboratory used similar method protocols and reagents to measure anti-EBOV GP IgG levels in human serum samples from a proficiency panel consisting of ten serum samples created by the differential dilution of a serum sample positive for anti-GP IgG antibodies (BMIZAIRE105) with negative serum (BMI529). The total assay variability (inter- and intra-assay variability) %CVs observed at each laboratory ranged from 12.2 to 30.6. Intermediate precision (inter-assay variability) for the laboratory runs ranged from 8.9 to 21.7%CV and repeatability (intra-assay variability) %CVs ranged from 7.2 to 23.7. The estimated slope for the relationship between log10(Target Concentration) and the log10(Observed Concentration) across all five laboratories was 0.95 with a 90% confidence interval of (0.93, 0.97). Equivalence test results showed that the 90% confidence interval for the ratios for the sample-specific mean concentrations at the five individual labs to the overall laboratory consensus value were within the equivalence bounds of 0.80 to 1.25 for each laboratory and test sample, except for six test samples from Lab D, two samples from Lab B1, and one sample from Lab B2. The mean laboratory concentrations for Lab D were less than those from the other laboratories by 20% on average across the serum samples. The evaluation of the proficiency panel at these laboratories provides a limited assessment of assay precision (intermediate precision, repeatability, and total assay variability), dilutional linearity, and accuracy. This evaluation suggests that the within-laboratory performance of the anti-EBOV GP IgG ELISA as implemented at the five laboratories is consistent with the intended use of the assay based on the acceptance criteria used by laboratories that have validated the assay. However, the assessment of between-laboratory performance revealed lower observed concentrations at Lab D and greater variability in assay results at Lab B1 relative to other laboratories.

Introduction

The filoviruses (family Filoviridae) from the genera Ebolavirus and Marburgvirus are etiologic agents of sporadic viral hemorrhagic fever outbreaks in humans with high mortality rates. An unprecedented outbreak of Ebola virus (EBOV; species Zaire ebolavirus) disease that began in Guinea during December 2013 [1] subsequently spread into neighboring West African countries of Sierra Leone and Liberia, prompting the World Health Organization (WHO) to declare the epidemic a public health emergency of international concern (http://www.who.int/mediacentre/news/statements/2014/ebola-20140808/en/). Phylogenetic analysis of viral isolates from this epidemic suggests a single transmission event introduced the virus, named the EBOV Makona variant [2], from an undetermined natural reservoir into humans in Guinea, followed by transmission between humans to spread the virus throughout Guinea and into Sierra Leone and Liberia [3]. Implementation of containment measures such as patient isolation and improved burial practices eventually controlled the epidemic, which resulted in 28,616 reported cases with a mortality rate of approximately 40% (http://www.who.int/csr/disease/ebola/en/).

The severity of this epidemic and principle transmission from human to human underscored the need for efficacious vaccines (and therapeutics) against EBOV, accelerating the placement of candidate EBOV vaccines into clinical safety trials [46]. This need for safe and efficacious vaccines was again evident with the onset of the 10th and largest outbreak in the Democratic Republic of the Congo (DRC) from 2018–2020. The 11th outbreak of EVD continues in the Western DRC.

The characteristics of filovirus infection, where infected patients are contagious only after manifestation of symptoms, allows one to use a ring vaccination strategy for disease containment. Ring vaccination strategy relies on the combination of contact tracing for case identification and a rapid effective vaccine for use in contacts and contacts of contacts of infected patients. The application of this strategy led to the approval of rVSV-ZEBOV (ERVEBO®), a single dose vaccine, using the safety and efficacy data from the clinical trial during the 2014 outbreak in West Africa by the Food and Drug Administration in December 2019 (https://www.fda.gov/news-events/press-announcements/first-fda-approved-vaccine-prevention-ebola-virus-disease-marking-critical-milestone-public-health). The effectiveness of ERVEBO in a ring vaccination response provides an important countermeasure for public health but does not address all unresolved questions in filovirus vaccine utilization including duration of protection, alternate dosing regimens, and the effectiveness of filovirus vaccines based on other viral platforms or alternative strategies. The development of multiple countermeasures against a disease necessitates the use of a common assay based on a surrogate of protection which can be used to compare the elicited immune response between vaccines and provide valuable information as to the effectiveness and durability of protection. Ideally, this assay is not only informative but simple, reproducible, species independent, and transferrable between labs. For example, during the development of countermeasures against anthrax, a lethal toxin neutralization assay was developed and used by many laboratories [7].

The development of vaccine candidates for Ebola virus disease prophylaxis [8] continues today, including deployment of a heterologous prime boost vaccine with European Commission Market Authorization during the last outbreak. However, the demonstration of efficacy for new filovirus vaccines will be complicated in the absence of a large outbreak and may require evaluation under the FDA Animal Rule or via non-inferiority trials against ERVEBO. Regulatory evaluation using these approaches is only possible with a correlate of protection and a well-developed assay that can measure the response in well-characterized animal challenge models as well as in human clinical trials. The species-neutral ELISA is ideal for bridging data between humans and animal models. Also, since the assay likely will be utilized in multiple experiments at many sites, it is important to demonstrate that the assay is reproducible among different laboratories.

In order to facilitate the development of additional vaccine countermeasures and to address such questions as the durability of immunity, the FANG has supported the development of a human anti-EBOV GP IgG ELISA. This study describes the FANG efforts to determine if the performance of the human anti-EBOV GP IgG ELISA [9] is acceptable for sample evaluation across five participating laboratories. Each laboratory used an anti-EBOV GP IgG ELISA to measure levels of binding in human serum samples from a FANG designed human proficiency panel. The panel consisted of ten human serum samples created by the differential dilution of human serum lot number BMIZAIRE105 (pool of serum with an approximate anti-GP IgG concentration of 1,000 ELISA units/mL) with control human serum (BMI529) without antibody activity. The concentration of the proficiency panel samples ranged from 0.00 ELISA units/mL to approximately 800 ELISA units/mL.

Each participating laboratory received sufficient volume of the proficiency panel for initial testing plus repeats and used their own anti-EBOV GP IgG ELISA established assay. The assay was validated at some laboratories and qualified at others [9]. Data from the participating laboratories were compared by statistical analysis. Both intra-laboratory and inter-laboratory analyses were performed to evaluate repeatability, intermediate precision, dilutional linearity, and accuracy. This paper summarizes both the intra- and inter-laboratory analysis of the results generated in the five separate laboratories. Results from the laboratories are de-identified in the analysis and reported as Laboratory A through E. The repeatability estimate for Laboratory B was greater than the acceptance criteria as established in laboratories that validated the anti-EBOV GP IgG ELISA with human serum, and, as a result, the proficiency panel assay runs were repeated. Results from both the original and repeated runs were included in the analysis and labeled as being from Laboratory B1 and B2, respectively.

Assay method

A common assay method [9] was tech-transferred to the participating laboratories, but there were minor variations in equipment/materials/procedures between laboratories. The analysis of the proficiency panel in the ELISA was performed similarly at Labs A, B1, and B2. All three used two separate operators on separate days. Samples were analyzed using a starting dilution of 1:62.5 and followed the plate layout as illustrated in Table 1. These plate layouts represent 15 plates with specific proficiency panel samples on each plate. All 15 plates were run twice for a total of 30 plates for each of Labs A, B1, and B2.

Table 1. Plate layout used at Laboratories A, B1, and B2.

Sample ID Plate Number
1 2 3 4 5 6 7 8
BMI-ZPP-11 X (3) X X X
BMI-ZPP-12 X X X X X (3)
BMI-ZPP-13 X (3) X X X X X
BMI-ZPP-14 X X X X X X X (3)
BMI-ZPP-15 X X (3) X X
BMI-ZPP-16 X X X X X X
BMI-ZPP-17 X X (3) X X X
BMI-ZPP-18 X X X X X
BMI-ZPP-19 X X (3) X X
BMI-ZPP-20 X X X X
Sample ID Plate Number
9 10 11 12 13 14 15
BMI-ZPP-11 X X X X X X
BMI-ZPP-12 X X X X X
BMI-ZPP-13 X X X X
BMI-ZPP-14 X X X
BMI-ZPP-15 X X X X X X
BMI-ZPP-16 X (3) X X X
BMI-ZPP-17 X X X X X
BMI-ZPP-18 X (3) X X X X
BMI-ZPP-19 X X X X X X
BMI-ZPP-20 X X (3) X X X X

An “X” indicates that sample was analyzed on the indicated plate.

An “X (3)” (shaded) indicates that sample was analyzed on the indicated plate three times.

The analysis of the proficiency panel in the ELISA was performed at Lab C by two separate operators over three days and at Lab D by two separate operators over five days. Samples were analyzed using a starting dilution of 1:50 and followed the plate layout as illustrated in Table 2. These plate layouts represent 12 plates with specific proficiency panel samples on each plate. All 12 plates were run at least twice for a total of 24 plates for each of Labs C and D.

Table 2. Plate layout used at Laboratories C and D.

Sample ID Plate Number
1 2 3 4 5 6
BMI-ZPP-11 X (3) X X X
BMI-ZPP-12 X X X X X X
BMI-ZPP-13 X (3) X X X
BMI-ZPP-14 X X X X X X
BMI-ZPP-15 X X (3) X X
BMI-ZPP-16 X X X X X X
BMI-ZPP-17 X X X (3) X
BMI-ZPP-18 X X X X X X
BMI-ZPP-19 X X X (3) X
BMI-ZPP-20 X X X X X X
Sample ID Plate Number
7 8 9 10 11 12
BMI-ZPP-11 X X X X X X
BMI-ZPP-12 X (3) X X X
BMI-ZPP-13 X X X X X X
BMI-ZPP-14 X (3) X X X
BMI-ZPP-15 X X X X X X
BMI-ZPP-16 X X (3) X X
BMI-ZPP-17 X X X X X X
BMI-ZPP-18 X X X (3) X
BMI-ZPP-19 X X X X X X
BMI-ZPP-20 X X X (3) X

An “X” indicates that sample was analyzed on the indicated plate.

An “X (3)” (shaded) indicates that sample was analyzed on the indicated plate three times.

The analysis of the proficiency panel in the ELISA was performed at Lab E by two separate operators over four days. Samples were analyzed using a starting dilution of 1:50 and followed the plate layout as illustrated in Table 3. This plate layout represents six plates with specific proficiency panel samples on each plate. The six plates were each run four times for a total of 24 plates. For all laboratories, some samples were analyzed three times on the same plate [denoted with “X (3)” in the plate layouts]. These contributed to assay repeatability.

Table 3. Plate layout used at Laboratory E.

Sample ID Plate Number
1 2 3 4 5 6
BMI-ZPP-11 X X X X X X
BMI-ZPP-12 X (3) X X X
BMI-ZPP-13 X X X X X X
BMI-ZPP-14 X (3) X X X
BMI-ZPP-15 X X X X X X
BMI-ZPP-16 X X (3) X X
BMI-ZPP-17 X X X X X X
BMI-ZPP-18 X X X (3) X
BMI-ZPP-19 X X X X X X
BMI-ZPP-20 X X X (3) X

An “X” indicates that sample was analyzed on the indicated plate.

An “X (3)” (shaded) indicates that sample was analyzed on the indicated plate three times.

Samples on a given plate were excluded from analysis if the within-assay CV of at least three dilution-adjusted concentrations determined for that sample was greater than 20%. Samples were also excluded if the plate including that sample failed to meet system suitability criteria. Some samples and plates that failed to meet the sample suitability criteria or system suitability criteria were repeated on later days. The ELISA concentrations of each qualification test sample by laboratory are provided in the supplemental information (S1S6 Tables).

This study, and specifically the use of human serum samples, was approved in writing by the Battelle Institutional Review Board in April of 2015 (approval number HSRE 0223–100062052). Human serum samples were collected from subjects by the sponsor (Crucell Holland) via written consent according to their IRB-approved protocol. These samples were not specifically collected for this interlaboratory study but rather for a different study. Battelle nor any authors were affiliated with this initial study. The sponsor subsequently provided Battelle volumes of these samples for the purposes of conducting the study described in this manuscript. Throughout its analysis of human biological materials and reporting, Battelle had no access to volunteer subjects’ identifiers nor any access to any code-key that would allow Battelle researchers to attribute any results of analysis to the original volunteer human research subjects.

Statistical methods

Inter-laboratory analysis was performed using the combined results across all laboratories. A mixed-effects analysis of variance (ANOVA) model was fitted to the base-10 log-transformed concentrations to evaluate both inter-laboratory precision (i.e., between lab precision) and intra-laboratory precision (i.e., within-laboratory precision). The model included a fixed effect for test sample and random effects for laboratory, test date nested within laboratory, and plate nested within day. Here, test operator was excluded as a random effect because this variable was indistinguishable from test day in most laboratories. Because of this confounding of effects, any variability attributable to test day may also be due to the different test operators.

Results were screened for outliers within each laboratory separately. Deleted studentized residuals were computed for each observation. If the absolute value of the deleted studentized residual was greater than four, then the observation was considered a statistical outlier and removed from the inter-laboratory analysis.

Variability associated with the random effects as well as intermediate precision, repeatability, and total assay variability were estimated separately for each lab using model-based percent coefficient of variation (CV). The percent CV for each source of variance was calculated using Tan’s [10] relative standard deviation as

100×eln(10)2×σ21

where σ2 is the model-estimated variance for the specific variance source. The percent CV associated with the residual variance served as an estimate for the assay repeatability. The percent CV associated with the test day and plate effects served as an estimate for the intermediate precision of the assay. Total assay variability was estimated using all variance components from the model (both inter- and intra-run variability).

The model intercept was obtained for each test sample from the mixed effects ANOVA model to serve as test sample consensus values across the laboratories. Agreement among laboratories was evaluated by comparing individual assay results from each laboratory to the consensus values. Boxplots were produced for each test sample to show the distribution of concentrations by laboratory in relation to the corresponding consensus value. The ratio of individual test results to consensus values was calculated by test sample to evaluate the level of agreement among laboratories based on two one-sided tests (TOST) of equivalence.

To assess dilutional linearity, a random coefficients linear regression model was fitted to the log-transformed observed concentrations versus the log-transformed target concentrations. The model included both a random intercept and slope effect for each laboratory, along with random effects for laboratory, test day nested within laboratory, and plate nested within laboratory. The random slope coefficients were modeled as laboratory-specific differences from the overall slope. The overall slope was used to assess the dilutional linearity based on a test of equivalence (TOST) and random slope coefficients were used to evaluate the level of agreement among the laboratories.

Results

Across all six laboratory runs, there were some false positive observations for Sample 18, a sample with a known negative concentration. All reportable values from Sample 18 were excluded from the statistical models. Table 4 lists five outliers that were removed from their respective intra-laboratory analyses that were also removed from this inter-laboratory analysis. One outlier each were removed from Laboratories B1 and B2. Three outliers were removed from Laboratory C. In the final analysis, Lab A contributed 204 reportable values, Lab B1 had 179 reportable values, Lab B2 had 214 reportable values, Lab C had 268 reportable values, Lab D had 216 reportable values, and Lab E had 218 reportable values.

Table 4. Statistical outliers identified during analysis of intra-laboratory data.

Laboratory Test Sample Observed Concentration (ELISA Units/mL) Target Concentration (ELISA Units/mL) Studentized Residual
B1 BMI-ZPP-17 4.28 200 -9.48
C BMI-ZPP-13 896.47 300 5.34
C BMI-ZPP-16 236.02 500 -4.74
B2 BMI-ZPP-19 51.20 100 -4.39
C BMI-ZPP-14 1845.88 700 4.33

These observations were deleted from both intra- and inter-laboratory analyses.

Table 5 presents ANOVA variance estimates and %CV for each source of variability, intermediate precision, and total assay variability by laboratory. For Laboratory A, the %CV for test date and plate nested within test date were 0.0 and 9.8, respectively. For Laboratory B1, the %CV for test date and plate nested within test date were 10.8 and 15.3, respectively. For Laboratory B2, the %CV for test date and plate nested within test date were 4.5 and 8.9, respectively. For Laboratory C, the %CV for test date and plate nested within test date were 9.8 and 8.5, respectively. For Laboratory D, the %CV for test date and plate nested within test date were 18.9 and 10.5, respectively. Finally, for Laboratory E, the %CVs for test date and plate nested within test date were 7.3 and 5.0, respectively. Laboratory E had the lowest %CV for intermediate precision (8.9) while Laboratory A had the lowest %CV for repeatability (7.2) and total assay variability (12.2). Laboratory B1 had the highest repeatability and total assay variability (23.7%CV and 30.6%CV, respectively) while Laboratory D had the highest %CV for intermediate precision (21.7).

Table 5. Summary of variance components obtained from mixed ANOVA model fit to data from all laboratories (results shown by laboratory).

Laboratory A
Source of Variability Variance %CV
Test Date 0.0000 0.0
Plate Nested in Test Date 0.0018 9.8
Intermediate Precision1 0.0018 9.8
Residual (Repeatability) 0.0010 7.2
Total Assay Variability2 0.0028 12.2
Laboratory B1
Source of Variability Variance %CV
Test Date 0.0022 10.8
Plate Nested in Test Date 0.0044 15.3
Intermediate Precision1 0.0065 18.8
Residual (Repeatability) 0.0103 23.7
Total Assay Variability2 0.0169 30.6
Laboratory B2
Source of Variability Variance %CV
Test Date 0.0004 4.5
Plate Nested in Test Date 0.0015 8.9
Intermediate Precision1 0.0019 9.9
Residual (Repeatability) 0.0033 13.3
Total Assay Variability2 0.0052 16.7
Laboratory C
Source of Variability Variance %CV
Test Date 0.0018 9.8
Plate Nested in Test Date 0.0014 8.5
Intermediate Precision1 0.0031 13.0
Residual (Repeatability) 0.0027 11.9
Total Assay Variability2 0.0058 17.7
Laboratory D
Source of Variability Variance %CV
Test Date 0.0066 18.9
Plate Nested in Test Date 0.0021 10.5
Intermediate Precision1 0.0087 21.7
Residual (Repeatability) 0.0023 11.2
Total Assay Variability2 0.0110 24.6
Laboratory E
Source of Variability Variance %CV
Test Date 0.0010 7.3
Plate Nested in Test Date 0.0005 5.0
Intermediate Precision1 0.0015 8.9
Residual (Repeatability) 0.0015 9.0
Total Assay Variability2 0.0030 12.7

1. Comprised of test date and plate nested within test date sources of variability.

2. Comprised of repeatability and intermediate precision.

Table 6 shows the consensus values (geometric means) along with 95% confidence intervals for each test sample generated from the mixed model ANOVA fitted to the data. Boxplots by sample of the reportable values from each laboratory, with each plot including a horizontal line for the consensus value estimate for the given sample, are provided in the supplemental information (S1S9 Figs).

Table 6. Consensus values by test sample generated from intercept of mixed ANOVA model fit to data from all laboratories.

Sample ID Target Concentration Consensus Value 95% CI Consensus Value
BMI-ZPP-11 600 695.93 (677.31, 715.06)
BMI-ZPP-12 400 475.20 (462.47, 488.27)
BMI-ZPP-13 300 325.47 (316.72, 334.46)
BMI-ZPP-14 700 844.08 (821.24, 867.55)
BMI-ZPP-15 800 871.34 (847.91, 895.41)
BMI-ZPP-16 500 561.81 (546.71, 577.33)
BMI-ZPP-17 200 226.25 (220.18, 232.49)
BMI-ZPP-19 100 110.63 (107.66, 113.68)
BMI-ZPP-20 50 70.81 (68.89, 72.79)

Table 7 shows the ratio of the mean concentration for each of the six individual laboratory runs to the consensus value for a given sample along with a 90% confidence interval for the ratio. Agreement among laboratories implies that these ratios should be close to one, indicating that the average concentrations are about the same as the consensus values. The ratios range from 0.95 to 1.08 for Laboratory A; from 0.96 to 1.19 for Laboratory B1; from 0.83 to 1.12 for Laboratory B2; from 0.96 to 1.16 for Laboratory C; from 0.71 to 0.97 for Laboratory D; and from 0.90 to 1.06 for Laboratory E. Fig 1 shows a graph of the mean ratio and 90% confidence interval for each test sample by laboratory.

Table 7. Ratio of laboratory mean concentration to overall consensus value with 90% confidence intervals for each test sample.

Sample ID Laboratory A Laboratory B1 Laboratory B2
Ratio 90% Confidence Interval Ratio 90% Confidence Interval Ratio 90% Confidence Interval
BMI-ZPP-11 1.08 (1.04, 1.13) 1.09 (0.97, 1.23) 0.83 (0.81, 0.85)
BMI-ZPP-12 1.02 (0.98, 1.08) 0.97 (0.87, 1.09) 1.12 (1.08, 1.16)
BMI-ZPP-13 1.02 (0.98, 1.06) 1.15 (1.02, 1.30)* 0.93 (0.89, 0.97)
BMI-ZPP-14 0.99 (0.96, 1.02) 0.96 (0.87, 1.05) 1.09 (1.04, 1.13)
BMI-ZPP-15 1.01 (0.97, 1.06) 1.08 (0.98, 1.18) 0.88 (0.84, 0.91)
BMI-ZPP-16 1.00 (0.96, 1.04) 1.07 (1.00, 1.14) 1.05 (1.00, 1.09)
BMI-ZPP-17 1.06 (1.01, 1.11) 1.01 (0.88, 1.15) 1.12 (1.06, 1.18)
BMI-ZPP-19 0.95 (0.91, 0.98) 0.98 (0.87, 1.09) 0.83 (0.79, 0.87)*
BMI-ZPP-20 0.98 (0.95, 1.02) 1.19 (1.02, 1.39)* 0.92 (0.88, 0.97)
Sample ID Laboratory C Laboratory D Laboratory E
Ratio 90% Confidence Interval Ratio 90% Confidence Interval Ratio 90% Confidence Interval
BMI-ZPP-11 1.10 (1.05, 1.15) 0.85 (0.81, 0.88) 0.90 (0.87, 0.94)
BMI-ZPP-12 1.08 (1.03, 1.12) 0.74 (0.72, 0.77)* 0.96 (0.92, 0.99)
BMI-ZPP-13 1.09 (1.04, 1.14) 0.79 (0.75, 0.83)* 0.95 (0.92, 0.99)
BMI-ZPP-14 1.00 (0.96, 1.05) 0.75 (0.71, 0.80)* 1.06 (1.02, 1.10)
BMI-ZPP-15 1.10 (1.04, 1.15) 0.90 (0.84, 0.97) 0.98 (0.95, 1.01)
BMI-ZPP-16 1.06 (1.02, 1.11) 0.76 (0.73, 0.79)* 1.01 (0.97, 1.06)
BMI-ZPP-17 0.96 (0.93, 1.00) 0.71 (0.69, 0.74)* 1.02 (0.99, 1.05)
BMI-ZPP-19 1.16 (1.10, 1.22) 0.97 (0.94, 1.01) 0.92 (0.89, 0.96)
BMI-ZPP-20 1.10 (1.02, 1.19) 0.77 (0.73, 0.82)* 1.02 (0.98, 1.06)

* 90% confidence interval is outside the acceptance bounds of (0.80, 1.25). Therefore, the concentrations for this test sample are not equivalent to those of other laboratories.

Fig 1. Graph of ratio of laboratory mean concentration to consensus value with 90% confidence intervals for each test sample by laboratory.

Fig 1

Dotted lines show equivalence region (0.80 to 1.25) and perfect agreement with consensus value (1.00). All means and confidence bounds are entirely within equivalence region for Laboratories A, B2, C, and E.

An equivalence test was conducted to determine if the mean test sample concentrations for each laboratory were equivalent to the corresponding test sample consensus value. An equivalence interval of 0.80 to 1.25 (representing a difference of 20% on the log scale) for the ratio of laboratory mean concentration to consensus concentration was used. The mean laboratory concentration for a given test sample is said to be equivalent to the consensus value for that sample if the 90% confidence interval for the ratio of these two values falls completely within the interval (0.80, 1.25).

Following this equivalence criteria: two intervals from Laboratory B1 (corresponding to BMI-ZPP-13 and BMI-ZPP-20) had an upper bound greater than the upper acceptance limit of 1.25 (1.30 and 1.39); one interval from Laboratory B2 (corresponding to BMI-ZPP-19) had a lower bound less than the lower acceptance limit of 0.80 (0.79); and six intervals from Laboratory D (corresponding to BMI-ZPP-12, BMI-ZPP-13, BMI-ZPP-14, BMI-ZPP-16, BMI-ZPP-17, and BMI-ZPP-20) had a lower bound less than the lower acceptance limit of 0.80. Furthermore, three of the six intervals are entirely below the lower acceptance bound of 0.80. These findings indicate that mean concentrations observed at Laboratory D are not equivalent to the other laboratories for six of the nine test samples.

Table 8 presents the estimated slope across the five laboratories and the corresponding 90% confidence interval obtained from the random regression model fit to assess the relationship between log10(observed concentration) and log10(target concentration). The overall slope was estimated to be 0.95 with a 90% confidence interval of (0.93, 0.97). An equivalence test was conducted to determine if the overall slope was equivalent to 1.00 (perfect dilutional linearity). An equivalence interval of 0.80 to 1.25 for the overall slope was used. Because the 90% confidence interval for the overall slope was completely within the interval (0.80, 1.25), the concentrations were found to be dilutionally linear across the laboratories. The slope estimates specific to each laboratory ranged from 0.94 to 0.96 (Table 8) and were consistent with the overall slope.

Table 8. Estimated slope and lower and upper 90% confidence interval bounds by laboratory from random coefficients regression model fit to all data.

Laboratory Slope Estimate 90% Confidence Interval#
Overall (All Labs) 0.95 (0.93, 0.97)
A 0.96 (0.90, 1.02)
B1 0.94 (0.88, 1.01)
B2 0.96 (0.90, 1.02)
C 0.96 (0.90, 1.02)
D 0.95 (0.89, 1.00)
E 0.95 (0.90, 1.01)

# 90% confidence interval is within the acceptance bounds of (0.80, 1.25). Therefore, the concentrations were dilutionally linear across the laboratories.

Discussion

The value of an assay as a regulatory tool is dependent on its accuracy, consistency, simplicity, and reproducibility. An assay that is relevant, is species independent, and replicable among laboratories is a powerful tool for product development. The data from a number of clinical trials utilizing ERVEBO strongly suggest that the anti-EBOV GP IgG ELISA provides data that correlate with product efficacy against Ebola infection. The development of new vaccines, or the evaluation of durability or alternative dosing regimens will be based on interpretation of data using the human anti-EBOV GP IgG ELISA. Our ability to use, or trust the data generated from non-clinical studies in different laboratories and clinical trials carried out with sera evaluated at different sites will require an understanding regarding the consistency and reproducibility of the assay among laboratories. In particular, assays using material from animal studies may be performed in laboratories different from those where the assay was performed to evaluate clinical trials. If the assay performance is not consistent among species and across laboratories, then data interpretation will not be possible. This interlaboratory study provided a direct head-to-head comparison of the ELISA performance in five different laboratories. The results from this study confirm the assay can be a universal tool for Ebola virus vaccine evaluation since results were similar when using the assay at multiple labs. However, the small differences in assay performance reinforce that for regulatory purposes, it is still ideal to rely on only one test site where the assay is fully validated.

Intermediate precision for the six laboratory runs ranged from 8.9 to 21.7%CV and repeatability ranged from 7.2 to 23.7%CV. The total assay variability %CVs range from 12.2 to 30.6. As a point of reference, laboratories that validated the anti-EBOV GP IgG ELISA have used the following precision acceptance criteria: (1) The intermediate precision of the assay for samples within the analytic range of the assay must be no larger than 25% CV; and (2) the repeatability of the assay for samples within the analytic range of the assay must be no larger than 20% CV. The repeatability estimate for Laboratory B1 was greater than the upper acceptance bound as established in laboratories that validated the anti-EBOV GP IgG ELISA with human serum. However, a repeat of the proficiency panel run at this laboratory following additional training of laboratory staff resulted in a repeatability estimate less than the upper acceptance bound; thus, illustrating the importance of rigorous training of laboratory staff and the strict adherence to assay procedures to ensure consistent results between runs.

Similarly, laboratories that validated the anti-EBOV GP IgG ELISA have used the following dilutional linearity (relative accuracy) acceptance criteria: the 90% confidence interval for the slope from the random regression model fit to data between the limits of quantitation and relating log10(concentration) to log10(spike level) will be entirely within (-1.20, -0.80). The interlaboratory study models dilutional linearity as log10(observed concentration) to log10(target concentration) resulting in a positive relationship between the two variables. Therefore, to conclude that dilutional linearity is acceptable in relation to the validation in human serum, the 90% confidence interval for the slope should be positive and fall entirely between 0.80 and 1.20. The overall slope was 0.95 and has a 90% confidence interval estimate of (0.93, 0.97); thus, the dilutional linearity is within the acceptance criteria as established in the assay validation with human serum.

Agreement among laboratories implies that the ratios of the mean concentration for the five individual labs to the overall laboratory consensus value for a given sample should be close to one. The ratios range from 0.95 to 1.08 for Laboratory A; from 0.96 to 1.19 for Laboratory B1; from 0.83 to 1.12 for Laboratory B2; from 0.96 to 1.16 for Laboratory C; from 0.71 to 0.97 for Laboratory D; and from 0.90 to 1.06 for Laboratory E. Equivalence test results showed that the 90% confidence interval for the ratio were within the equivalence bounds of 0.80 to 1.25 for each laboratory except for samples BMI-ZPP-13 and BMI-ZPP-20 in Laboratory B1, BMI-ZPP-19 in Laboratory B2, and six samples in Laboratory D.

The assessment of between-laboratory performance revealed lower observed concentrations at Lab D and greater variability in assay results at Lab B1 relative to the other laboratories. The lower observed concentrations at Lab D illustrate the importance of monitoring assay performance and harmonizing across laboratories. Given the inherent differences from subject-to-subject in clinical trials and animal-to-animal in non-clinical studies, these differences observed at Lab D relative to the other laboratories are not likely to affect interpretation of study results. The variability in assay results at Lab B1 was mitigated by additional laboratory staff training.

The evaluation of the proficiency panel at these laboratories provides a limited assessment of assay precision (intermediate precision, repeatability, and total assay variability), dilutional linearity, and accuracy. This limited evaluation suggests that the within-laboratory performance of anti-EBOV GP IgG ELISA as implemented at the five laboratories is performing consistently with the intended use of the assay based on the acceptance criteria used by laboratories that have validated the assay.

Supporting information

S1 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-11.

(Consensus Concentration = 695.93). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S2 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-12.

(Consensus Concentration = 475.20). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S3 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-13.

(Consensus Concentration = 325.47). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S4 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-14.

(Consensus Concentration = 844.08). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S5 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-15.

(Consensus Concentration = 871.34). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S6 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-16.

(Consensus Concentration = 561.81). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S7 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-17.

(Consensus Concentration = 226.25). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S8 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-19.

(Consensus Concentration = 110.63). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

(TIF)

S9 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-20.

(Consensus Concentration = 70.81). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentration.

(TIF)

S1 Table. ELISA concentration of each test sample—Laboratory A.

(XLSX)

S2 Table. ELISA concentration of each test sample—Laboratory B1.

(XLSX)

S3 Table. ELISA concentration of each test sample—Laboratory B2.

(XLSX)

S4 Table. ELISA concentration of each test sample—Laboratory C.

(XLSX)

S5 Table. ELISA concentration of each test sample—Laboratory D.

(XLSX)

S6 Table. ELISA concentration of each test sample—Laboratory E.

(XLSX)

Acknowledgments

Opinions, interpretations, conclusions, and recommendations are those of the authors and are not necessarily endorsed by the U.S. Army. We acknowledge the members of the FANG EBOV ELISA community and the participating laboratories for their insightful questions, comments, and recommendations during the performance of this interlaboratory evaluation of the assay.

Data Availability

All relevant data are within the manuscript and its Supporting Information files.

Funding Statement

This project has been funded in whole or in part with funds from the U.S. Department of Defense (DOD) Joint Program Executive Office for Chemical, Biological, Radiological, and Nuclear Defense (JPEO-CBRND) Chemical Biological Radiological and Nuclear Medical (JPM-CBRN Medical) under Battelle contract number GS00Q140ADU402, delivery order W911QY-16-F0074. The funding body contributed to study design and in the decision to submit the article for publication. The views expressed here are those of the authors and do not necessarily represent the views or official position of the DOD.

References

  • 1.Baize S, Pannetier D, Oestereich L, Rieger T, Koivogui L, Magassouba N, et al. Emergence of Zaire Ebola virus disease in Guinea. The New England journal of medicine. 2014;371(15):1418–25. 10.1056/NEJMoa1404505 [DOI] [PubMed] [Google Scholar]
  • 2.Kuhn JH, Andersen KG, Baize S, Bao Y, Bavari S, Berthet N, et al. Nomenclature- and database-compatible names for the two Ebola virus variants that emerged in Guinea and the Democratic Republic of the Congo in 2014. Viruses. 2014;6(11):4760–99. 10.3390/v6114760 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Gire SK, Goba A, Andersen KG, Sealfon RS, Park DJ, Kanneh L, et al. Genomic surveillance elucidates Ebola virus origin and transmission during the 2014 outbreak. Science (New York, NY). 2014;345(6202):1369–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Chappell KJ, Watterson D. Fighting Ebola: A Window for Vaccine Re-evaluation? PLoS pathogens. 2017;13(1):e1006037 10.1371/journal.ppat.1006037 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Keshwara R, Johnson RF, Schnell MJ. Toward an Effective Ebola Virus Vaccine. Annual review of medicine. 2017;68:371–86. 10.1146/annurev-med-051215-030919 [DOI] [PubMed] [Google Scholar]
  • 6.Banadyga L, Marzi A. Closer than ever to an Ebola virus vaccine. Expert review of vaccines. 2017:1–2. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Omland KS, Brys A, Lansky D, Clement K, Lynn F. Clinical And Vaccine Immunology. 2008;15:946–953 10.1128/CVI.00003-08 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.O’Donnell K, Marzi A. The Ebola virus glycoprotein and its immune responses across multiple vaccine platforms. Expert Rev Vaccines. 2020;19(3):267–277. 10.1080/14760584.2020.1738225 [DOI] [PubMed] [Google Scholar]
  • 9.Rudge TL Jr., Sankovich KA, Niemuth NA, Anderson MS, Badorrek CS, Skomrock ND, et al. (2019). Development, qualification, and validation of the Filovirus Animal Nonclinical Group anti-Ebola virus glycoprotein immunoglobulin G enzyme-linked immunosorbent assay for human serum samples. PLoS ONE 14(4): e0215457 10.1371/journal.pone.0215457 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Tan CY. RSD and Other Variability Measures of the Lognormal Distribution. Pharmacopeial Forum. 31:653–655. [Google Scholar]

Decision Letter 0

John Schieffelin

3 Jun 2020

PONE-D-20-10369

Interlaboratory comparison for the Filovirus Animal Nonclinical Group (FANG) anti-Ebola virus glycoprotein immunoglobulin G enzyme-linked immunosorbent assay

PLOS ONE

Dear Dr. Anderson,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Jul 18 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

John Schieffelin, MD

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Please provide additional details regarding human serum samples used in your study. In the ethics statement in the Methods and online submission information, please ensure that you have specified (1) the sources, e.g. are they collected from human subjects specifically recruited for your study? (2) if so, whether consent was informed and what type you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed). If the need for consent was waived by the ethics committee, please include this information.”

3.  Thank you for stating the following in the Competing Interests section:

"The authors have declared that no competing interests exist.".

We note that one or more of the authors are employed by a commercial company: 'Battelle'

1.     Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.

Please also include the following statement within your amended Funding Statement.

“The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.”

If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.

2. Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc.  

Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and  there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

* Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf.

Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Line 22: first sentence of the abstract is long, consider breaking up into 2 points about the West African outbreak and then additionally the ongoing outbreak in DRC

Line 27: mention that the anti-EBOV GP IgG ELISA is the FANG assay here

Line 54: again I think at least initially (even though your cite it) you should say that the ELISA is the FANG assay.

Some of the information in the Introduction sounds more like study design which should be part of the Methods. I would like to see more of a background here in the Introduction section perhaps on the development of the assay initially, why this assay was chosen, rationale for doing the study, etc. I think a bit more context for the study is needed in this section.

I would like to see more expansion on the implications of the results in the discussion section. For example spell out the implications of the lower observed concentrations at Lab D and the variability of results in Lab B1. What do these results mean for the use of this assay moving forward?

This manuscript provides important information on assay validation of the FANG assay, an anti-EBOV GP IgG ELISA. These studies are needed in order to ensure robust assays and provide confidence in interpreting data between and within laboratories. My main concern with this manuscript is that more information is needed in both the Introduction and the Discussion to better understand the context of this assay and why the study is warranted. I think it is very much warranted but would like to see more of this information presented in the text.

Reviewer #2: Anderson et al provide data examining the performance of the FANG IgG EBOV ELISA across multiple sites. The study was intended to assess inter and intra-laboratory performance of the assay. The resulting data suggest that the assay is relatively robust, but does highlight a few limitations that the authors claim tend to be laboratory specific. The study has value for assessing preclinical/clinical samples for vaccine development across multiple institutions and clinical sites; where uniformity and agreement among analytical methods are critical for evaluating performance of a given countermeasure, in this case, vaccines targeting the EBOV GP. The paper was digestible and fairly straight forward.

A point of clarification:

Abstract, line 26: One vaccine is fully licensed and not considered investigational, the other is investigational at present.

Please go through the article and verify issues with spelling and grammar throughout to ensure clean readability.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Aug 25;15(8):e0238196. doi: 10.1371/journal.pone.0238196.r002

Author response to Decision Letter 0


15 Jul 2020

1. Comment: Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming.

Response: We have verified that the manuscript adheres to PLOS ONE’s style requirements.

2. Comment: Please provide additional details regarding human serum samples used in your study. In the ethics statement in the Methods and online submission information, please ensure that you have specified (1) the sources, e.g. are they collected from human subjects specifically recruited for your study? (2) if so, whether consent was informed and what type you obtained (for instance, written or verbal, and if verbal, how it was documented and witnessed). If the need for consent was waived by the ethics committee, please include this information.”

Response: Human serum samples were collected from subjects by the sponsor (Crucell Holland) via written consent according to their IRB-approved protocol. These samples were not specifically collected for this interlaboratory study but rather for a different study (“A Phase 1, Randomized, Placebo-Controlled, Observer-Blind Study to Evaluate the Safety, Tolerability and Immunogenicity of Heterologous Prime-Boost Regimens Using MVA-BN®-Filo and Ad26.ZEBOV Administered in Different Doses, Sequences, and Schedules in Healthy Adult Subjects”). Battelle was not involved in this study. The sponsor subsequently provided Battelle volumes of these samples for the purposes of conducting the study described in this manuscript. Throughout its analysis of human biological materials and reporting, Battelle had no access to volunteer subjects’ identifiers nor any access to any code-key that would allow Battelle researchers to attribute any results of analysis to the original volunteer human research subjects. The ethics statement in the Methods Section and online submission form has been updated with the requested information.

3. Comment: Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.

Response: We have provided the following amended Funding Statement in our cover letter:

“This project has been funded in whole or in part with funds from the U.S. Department of Defense (DOD) Joint Program Executive Office for Chemical, Biological, Radiological, and Nuclear Defense (JPEO-CBRND) Medical Countermeasure Systems Joint Vaccine Acquisition Program (MCS-JVAP) under Battelle contract number GS00Q140ADU402, delivery order W911QY-16-F-0074. The funding body provided support in the form of salaries for some of the authors (MSA, NAN, CLS, TLR) and contributed to study design and in the decision to submit the article for publication. Our (MSA, NAN, TLR) affiliation with Battelle, a non-profit contract research organization, did not play a role in the study. The specific roles of these authors are articulated in the ‘author contributions’ section. The funding body contributed to study design and in the decision to submit the article for publication. The views expressed here are those of the authors and do not necessarily represent the views or official position of the DOD.”

4. Comment: Please also include the following statement within your amended Funding Statement: “The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.” If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.

Response: Our new funding statement contains this language.

5. Comment: Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc. Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Response: We have provided the following amended Competing Interests Statement in our cover letter:

All authors listed on the manuscript have read the manuscript and approve its submission to PLOS ONE. We have no issues relevant to your journal policies for publication, and the authors have no potential competing interests to declare. Our (MSA, NAN, TLR) employment by Battelle, a non-profit contract research organization, does not alter our adherence to PLOS ONE policies on sharing data and materials nor played a role in the study. This manuscript has not been published previously and is not under consideration for publication in any other scientific journal.

6. Comment: Line 22: first sentence of the abstract is long, consider breaking up into 2 points about the West African outbreak and then additionally the ongoing outbreak in DRC.

Response: The first sentence has been broken into two separate statements.

7. Comment: Line 27: mention that the anti-EBOV GP IgG ELISA is the FANG assay here.

Response: The mention has been added to the manuscript.

8. Comment: Again I think at least initially (even though your cite it) you should say that the ELISA is the FANG assay. Some of the information in the Introduction sounds more like study design which should be part of the Methods. I would like to see more of a background here in the Introduction section perhaps on the development of the assay initially, why this assay was chosen, rationale for doing the study, etc. I think a bit more context for the study is needed in this section.

Response: We added language to clarify that the ELISA is the FANG assay. Text has also been added to the Introduction section to include some background of the assay development.

9. Comment: I would like to see more expansion on the implications of the results in the discussion section. For example spell out the implications of the lower observed concentrations at Lab D and the variability of results in Lab B1. What do these results mean for the use of this assay moving forward?

Response: The requested information has been added to the Discussion section.

10. Comment: This manuscript provides important information on assay validation of the FANG assay, an anti-EBOV GP IgG ELISA. These studies are needed in order to ensure robust assays and provide confidence in interpreting data between and within laboratories. My main concern with this manuscript is that more information is needed in both the Introduction and the Discussion to better understand the context of this assay and why the study is warranted. I think it is very much warranted but would like to see more of this information presented in the text.

Response: The requested additional information has been added to the Introduction and Discussion sections.

11. Comment: Abstract, line 26: One vaccine is fully licensed and not considered investigational, the other is investigational at present.

Response: This clarification has been made in the manuscript.

12. Comment: Please go through the article and verify issues with spelling and grammar throughout to ensure clean readability.

Response: We have gone through and verified there are no issues with spelling and grammar issues that would hinder clean readability of the manuscript.

Attachment

Submitted filename: Response to Reviewers.docx

Decision Letter 1

John Schieffelin

12 Aug 2020

Interlaboratory comparison for the Filovirus Animal Nonclinical Group (FANG) anti-Ebola virus glycoprotein immunoglobulin G enzyme-linked immunosorbent assay

PONE-D-20-10369R1

Dear Dr. Anderson,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

John Schieffelin, MD

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: (No Response)

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: (No Response)

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: (No Response)

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: (No Response)

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Acceptance letter

John Schieffelin

14 Aug 2020

PONE-D-20-10369R1

Interlaboratory comparison for the Filovirus Animal Nonclinical Group (FANG) anti-Ebola virus glycoprotein immunoglobulin G enzyme-linked immunosorbent assay

Dear Dr. Anderson:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr, John Schieffelin

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-11.

    (Consensus Concentration = 695.93). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S2 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-12.

    (Consensus Concentration = 475.20). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S3 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-13.

    (Consensus Concentration = 325.47). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S4 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-14.

    (Consensus Concentration = 844.08). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S5 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-15.

    (Consensus Concentration = 871.34). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S6 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-16.

    (Consensus Concentration = 561.81). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S7 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-17.

    (Consensus Concentration = 226.25). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S8 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-19.

    (Consensus Concentration = 110.63). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentrations.

    (TIF)

    S9 Fig. Observed concentration (ELISA Units/mL) by laboratory for sample BMI-ZPP-20.

    (Consensus Concentration = 70.81). Center line in the box depicts the median concentration while the height of the box represents the 25th and 75th percentile of the concentration distribution. Vertical lines extending above and below the box represent the maximum and minimum concentration values for the laboratory. Open circles show the observed concentration.

    (TIF)

    S1 Table. ELISA concentration of each test sample—Laboratory A.

    (XLSX)

    S2 Table. ELISA concentration of each test sample—Laboratory B1.

    (XLSX)

    S3 Table. ELISA concentration of each test sample—Laboratory B2.

    (XLSX)

    S4 Table. ELISA concentration of each test sample—Laboratory C.

    (XLSX)

    S5 Table. ELISA concentration of each test sample—Laboratory D.

    (XLSX)

    S6 Table. ELISA concentration of each test sample—Laboratory E.

    (XLSX)

    Attachment

    Submitted filename: Response to Reviewers.docx

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting Information files.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES