Diabetes Technology in the Inpatient Setting for Management of Hyperglycemia

Georgia M Davis 1, Rodolfo J Galindo 1, Alexandra L Migdal 1, Guillermo E Umpierrez 1,*
PMCID: PMC7453786  NIHMSID: NIHMS1619667  PMID: 31980123

INTRODUCTION

Advances in diabetes technology over the past decades have revolutionized patient care and diabetes management. The use of this technology, including continuous glucose monitoring (CGM) and continuous subcutaneous insulin infusion (CSII), in patients with type 1 (T1) and type 2 (T2) diabetes mellitus (DM) continues to grow in the ambulatory setting. Although substantial data support outpatient use of this technology for improvement in glycemic control and diabetes outcomes, there are limited data regarding its use in the inpatient setting.1 There is consensus among experts and medical societies that, compared with intermittent capillary blood glucose testing, CGM technology offers benefits in the prevention of severe hyperglycemia and hypoglycemia by identifying glucose trends and allowing insulin doses to be adjusted more accurately.24 In addition, several diabetes clinical guidelines support the continued use of outpatient CSII in patients who are physically able to continue using their insulin pump during hospitalization.3,5 However, randomized controlled trials are needed to determine whether use of CGM and CSII systems in the hospital can improve clinical outcomes compared with intermittent capillary blood glucose monitoring and conventional insulin treatment. As CGM and CSII technologies continue to advance, the use of artificial pancreas technology is being evaluated in the inpatient setting. Use of computerized decision support systems for glycemic control in the hospital is also expanding. Here we review current available diabetes technology as it relates to the management of hospitalized patients.

CONTINUOUS GLUCOSE MONITORING IN THE HOSPITAL

The use of capillary point-of-care (POC) glucose testing has been the mainstay for monitoring and treatment of hospitalized patients with diabetes. POC blood glucose (BG) testing is commonly performed 3 to 4 times daily in hospitalized patients for monitoring of glycemic control and adjustment of diabetes treatment regimens. However, the need for frequent BG testing in certain inpatient populations, the intermittent nature of testing, and the associated time burden for nursing and ancillary hospital staff are significant limitations of POC BG testing. The desire for closer monitoring of glucose values led to the development of CGM beginning in the late 1970s, with approval of the first CGM by the Food and Drug Administration occurring in 1999.6 The ability of CGM to provide estimated glucose values every 5 to 15 minutes with information regarding glucose trends allows for a more comprehensive assessment of glycemic control, an attractive feature in both inpatient and outpatient settings. Rapid improvements in accuracy and increased commercial availability of CGM technology has led to its widespread use in ambulatory diabetes management. However, use of CGM in the hospital remains investigational, with ongoing studies evaluating its use in diverse inpatient populations.

In the hospital setting, several studies have shown improvement in the detection of both hyperglycemia and hypoglycemia in critically ill and non-critically ill patients.2,79 In contrast to the exclusive use of subcutaneous CGM sensors in the outpatient realm, CGM use in the hospital can also include invasive (intravascular) or noninvasive (transdermal) CGM in intensive care unit (ICU) settings.

Most studies in ICU populations using CGM have focused on accuracy and reliability, and have in general been small and underpowered to detect changes in patient-centered outcomes (ie, incidence of hypoglycemia or complications) (Table 1). Of the studies assessing glycemic control, most did not show significant differences in average glycemic control with CGM versus POC glucose testing. Kopecky and colleagues12 investigated the use of CGM combined with an enhanced model predictive control (eMPC) insulin titration algorithm in postoperative cardiac ICU patients. The use of the eMPC algorithm combined with CGM was compared with standard use of the algorithm with interval POC glucose testing (control group). Overall, there were no significant differences in glycemic control between the eMPC-CGM system with more frequent input of glucose values compared with the eMPC algorithm with intermittent BG values alone.12 In a larger study, Holzinger and colleagues11 observed a significant reduction in hypoglycemia (defined as glucose <40 mg/dL or 2.2 mmol/L) with real-time CGM (RT-CGM) use versus POC testing (1.6 vs 11.5%; P = .031) in ICU patients requiring mechanical ventilation. A subgroup analysis in this population revealed improved glycemic control and time in target glucose range (defined as glucose <110 mg/dL or 6.1 mmol/L) in patients with higher illness severity indices. Despite these findings, there were no differences in hospital length of stay (LOS) or mortality between CGM and POC BG testing groups.11 A recent expert consensus meeting acknowledged that use of CGM in critical care populations appears to be accurate and reliable, but there continues to be a need for larger studies powered to determine efficacy in improving glycemic control, detection and reduction of hypoglycemic events, and impact on hospital stay and clinical outcomes.2

Table 1.

ICU CGM outcomes studies

Authors CGM Type Patient Population Study Design Outcomes Results
Logtenberg et al,10 2009 Paradigm (Medtronic MiniMed) Cardiac ICU; non-DM patients undergoing elective cardiothoracic surgery (n = 30) Prospective RCT: RT-CGM with alerts at 72 mg/dL and 180 mg/dL vs RT-CGM without alerts Accuracy and glycemic control between alert groups No significant difference in glycemic control with use of CGM alerts.
Holzinger et al,11 2010 Guardian (Medtronic MiniMed) DM and non-DM ICU patients requiring mechanical ventilation (n = 124) Prospective RCT: RT-CGM (no alerts) vs standard of care (with blinded CGM) Glycemic control, LOS, mortality No significant difference in mean glucose or time with glucose <150 or <110 mg/dL; marked reduction in hypoglycemia with RT-CGM (1.6 vs 11.5%; P = .031). No differences in LOS or hospital mortality.
Kopecky et al,12 2013 Guardian (Medtronic MiniMed) Cardiac ICU; patients with DM and non-DM undergoing elective major cardiac surgery (n = 24) Prospective RCT: eMCP algorithm for BG management alone vs eMPC combined with CGM values every 15 min Glycemic control No significant differences in glycemic control between groups; 2 hypoglycemic events in control eMPC vs no events in eMPC-CGM group.
Umbrello et al,13 2014 OptiScanner 5000 (OptiScan Biomedical Corporation) DM and non-DM ICU patients with sepsis (n = 6) Pilot prospective observational study: use of CGM to evaluate glycemic control protocol Glycemic control Adequate glycemic control ([93% of cohort with median time in target range [BG 80–150 mg/dL]). No episodes of severe hypoglycemia (<40 mg/dL).
De Block et al,14 2015 GlucoDay S (A. Menarini Diagnostics) ICU patients (n = 35) Prospective RCT: RT-CGM vs standard of care and blinded CGM Accuracy and glycemic control No significant difference in glycemic control, variability, or hypoglycemic events. No significant difference in LOS or hospital mortality.

Conversion: mg/dL × 0.0555 = mmol/L.

Abbreviations: CGM, continuous glucose monitoring; DM, diabetes mellitus; eMPC, enhanced model predictive control; ICU, intensive care unit; LOS, length of stay; RCT, randomized-controlled trial; RT-CGM, real-time continuous glucose monitoring.

Data from Refs.1014

In non-critically ill hospitalized patients, studies with the use of CGM have been mostly observational (Table 2). Observational data have provided important insight into glycemic control patterns in hospitalized patients, emphasizing improved detection of hypoglycemia using CGM. Gomez and Umpierrez7 compared the use of blinded/professional CGM versus POC glucose testing in non-critically ill hospitalized patients with T2DM treated with basal bolus insulin regimens. Even though there was no significant difference in mean daily blood glucose concentration between groups, CGM detected a higher number of hypoglycemic events (52 vs 12, P = .0001). More than 50% of the hypoglycemic events occurred between dinner and breakfast, highlighting the utility of CGM use for detection of nocturnal hypoglycemia in this population.7 A recent pilot study by Spanakis and colleagues19 in a population of non-ICU hospitalized older adult patients (mean age 70.8 ± 6.2 years) demonstrated feasibility of using CGM data transmitted to nursing personnel in a telemetry-type method, with an alert for hypoglycemia set at a sensor glucose value of 85 mg/dL (4.7 mmol/L). In this group of 5 patients on insulin therapy, there were no episodes of severe hypoglycemia (<54 mg/dL or 3 mmol/L) and potential hypoglycemic episodes were prevented in 2 patients using the CGM alert system during the hospital stay.19

Table 2.

Non-ICU CGM studies

Authors CGM Type Patient Population Study Design Outcomes Results
Burt et al,15 2013 CGMS Gold (Medtronic MiniMed) T1DM and T2DM admitted to general wards on basal bolus insulin (n = 26) Observational Prospective Cohort: blinded CGM vs POC glucose testing Accuracy and glycemic control No significant difference in glycemic control; most hyperglycemic episodes were postprandial and hypoglycemic episodes more likely to occur between midnight and 7 AM
Schaupp et al,16 2015 iPro2 system (Medtronic MiniMed) T2DM admitted to general wards on basal bolus insulin (n = 84) Observational Prospective Cohort: blinded CGM vs POC glucose testing Accuracy Good agreement between CGM and POC glucose values; Clarke Error Grid analysis with 98.7% of values in Zone A or Zone B (88.2% within Zone A); 15-fold increase in detection of nocturnal hypoglycemia with CGM.
Gómez et al,17 2015 iPro2 system (Medtronic MiniMed) T2DM or hyperglycemia admitted to general wards on basal bolus insulin (n = 38) Prospective Pilot RCT: blinded CGM vs POC glucose testing Accuracy Good agreement between CGM and POC glucose values; Clarke Error Grid analysis with 91.9% of values in Zone A or Zone B. Increased detection of hypoglycemic episodes with CGM (55 vs 12, P<.01).
Gu et al,18 2017 Paradigm 722 or CGMS Gold (Medtronic MiniMed) T2DM admitted to general wards for glycemic control (n = 81) Prospective RCT: SAP vs MDI with blinded CGM Glycemic control, time to target glucose 21 SAP vs 6 MDI patients achieved glycemic targets within 3 d. SAP vs MDI had less hypoglycemia (sensor glucose <50 mg/dL: 0.04% vs 0.32%; P<.05) and hyperglycemia (sensor glucose >180 mg/dL: 21.56% vs 35.03%; P<.05).
Spanakis et al,19 2018 DEXCOM G4 CGM with Share2 application (DEXCOM) T2DM admitted to general wards on insulin therapy (n = 5) Single-arm pilot trial of consecutive patients using glucose telemetry system (CGM alert at 85 mg/dL) Glucose telemetry system feasibility Prevention of potential hypoglycemia (CGM BG <70 mg/dL for >20 min) captured by alarm occurred in 2 patients (3 events). No patients had CGM glucose value <54 mg/dL.

Conversion: mg/dL × 0.0555 = mmol/L.

Abbreviations: BG, blood glucose; CGM, continuous glucose monitoring; CGMS, continuous glucose monitoring system; ICU, intensive care unit; MDI, multiple daily injections; POC, point of care; SAP, sensor-augmented pump; T1DM, type 1 diabetes mellitus; T2DM, type 2 diabetes mellitus;

Data from Refs.1519

CGM technology continues to evolve, with investigations expanding to more diverse inpatient populations with diabetes. More accuracy data are needed in specific hospital patient populations, including those with severe dehydration and volume depletion, anasarca, and end-stage renal disease on hemodialysis or peritoneal dialysis. Although CGM metrics (including time in target glucose range, time in hyperglycemia, time in hypoglycemia, and glycemic variability) continue to be updated and defined for targeting improved glycemic control and outcomes in ambulatory patients,20 no current consensus exists regarding CGM metrics in hospitalized patients. Further information is needed in this setting on standardization of glycemic control metrics, appropriate target glucose ranges, and impact of CGM use on hospital outcomes and costs to understand how to safely and effectively implement this technology. In addition to targeting improved glycemic control, emerging CGMs that no longer require calibration with POC glucose testing (factory-calibrated) have the potential to decrease both nursing and patient burden associated with frequent POC glucose testing in the hospital. Ongoing studies (NCT03832907) with factory-calibrated CGM are testing its accuracy in diverse inpatient populations, and the use of CGM data by nursing and ancillary staff to detect and prevent glycemic excursions (NCT03877068).

CONTINUOUS SUBCUTANEOUS INSULIN INFUSION IN THE HOSPITAL

With the use of CSII increasing in the ambulatory setting, the importance of guidelines for its continued use in the inpatient setting for health care providers has been addressed by several professional societies, including the American Diabetes Association, the American Association of Clinical Endocrinologists, and The Endocrine Society. These societies advocate for continuation of CSII therapy in appropriate hospitalized patients, with the support of (1) implemented hospital policies on CSII, (2) inpatient endocrinology or diabetes management teams, and (3) a signed agreement from the patient acknowledging responsibilities of CSII therapy.3,5,21 Continuation of CSII is not recommended in critically ill and hemodynamically unstable patients, as well as in those who are not able to demonstrate appropriate use of their insulin pump.1,3,5 For instance, a retrospective analysis of 50 patients by Kannan and colleagues22 demonstrated that 24% of patients admitted to the hospital using CSII in the outpatient setting were unable to correctly demonstrate use of critical pump skills during hospitalization. A recent review by Umpierrez and Klonoff1 illustrates a proposed algorithm for decision-making regarding hospital continuation of CSII therapy (Fig. 1), as well as contraindications to its use (Box 1).

Fig. 1.

Fig. 1.

Algorithm for inpatient continuation of CSII-CGM therapy hospitalization. EGD, upper endoscopy; IV, intravenous. (From Umpierrez GE, Klonoff DC. Diabetes Technology Update: Use of Insulin Pumps and Continuous Glucose Monitoring in the Hospital. Diabetes Care. 2018;41(8):1579–1589; with permission.)

Box 1. Contraindications to insulin pump therapy in the hospital.

Impaired level of consciousness (except during short-term anesthesia)

Patient’s inability to correctly demonstrate appropriate pump settings

Critical illness requiring intensive care

Psychiatric illness that interferes with a patient’s ability to self-manage diabetes

Diabetic ketoacidosis and hyperosmolar hyperglycemic state

Refusal or unwillingness to participate in self-care

Lack of pump supplies

Lack of trained health care providers, diabetes educators, or diabetes specialist

Patient at risk for suicide

Health care decision

From Umpierrez GE, Klonoff DC. Diabetes Technology Update: Use of Insulin Pumps and Continuous Glucose Monitoring in the Hospital. Diabetes Care. 2018;41(8):1579–1589; with permission.

Most of the data regarding hospital use of CSII have been retrospective and focused on the continuation of outpatient CSII during admission. These studies have suggested that with appropriate patient selection and hospital guidelines, patients on preexisting CSII (with or without concurrent CGM) can safely maintain glycemic control during hospitalization. Review of 125 hospitalizations of 65 patients on insulin pump therapy by Nassar and colleagues23 showed an increase in prevalence of insulin pump use during hospitalization over a 3-year period. During this time, there were no significant differences in mean hospital glucose levels between patients who continued CSII versus those transitioned to multiple daily insulin injections (MDI).23 An additional review by Cook and colleagues24 of 253 hospitalizations of 136 patients on outpatient CSII over a 6-year period showed similar results regarding glycemic control between CSII use and subcutaneous insulin therapy during hospitalization, but there were fewer severe hyperglycemic (BG >300 mg/dL or 16.7 mmol/L) and hypoglycemic (BG <50 mg/dL or 2.8 mmol/L) events in patients remaining on CSII in the hospital. In addition, a recent prospective pilot trial by Levitt and colleagues25 investigated the feasibility of CSII, both with and without CGM technology, in hospitalized patients with T2DM. Patients were randomized to 3 groups: (1) basal bolus insulin therapy with blinded CGM, (2) CSII with blinded CGM, or (3) CSII with RT-CGM. Although there were no significant differences in time in target glucose range or time in hypoglycemia between groups, this study showed the feasibility of combined CSII-CGM therapy in the inpatient setting. In addition, they reported that CGM detected more episodes of hypoglycemia than POC glucose testing alone (19 vs 12 episodes).25 A study by Gu and colleagues18 in China evaluated CSII with CGM versus MDI in non-acutely ill patients with T2DM hospitalized for glycemic optimization over a 2-week period. In 81 patients (40 on CSII-CGM vs 41 on MDI), more patients using CSII-CGM were able to achieve target glucose values between 70 and 180 mg/dL (3.9–10.0 mmol/L) within a 3-day period compared with those on MDI (53% vs 15%, respectively). Overall, those on CSII-CGM also had significantly less hypoglycemia (glucose <50 mg/dL [2.8 mmol/L]; 0.02 vs 0.31%, P<.05), and less severe hyperglycemia (glucose >250 mg/dL [13.9 mmol/L]; 3.9 vs 8.3%, P<.05).18

Although CSII does continue to be used in the inpatient setting, current investigations are moving toward the use of combined CSII-CGM technology with the ability to provide automated modulation of subcutaneous insulin infusion rates (ie, closed-loop or hybrid closed-loop technology).

ARTIFICIAL PANCREAS TECHNOLOGY IN THE HOSPITAL

The artificial pancreas system, also referred to as a “closed-loop” system, “automated insulin delivery” system, or “autonomous system for glycemic control,” is composed of a CGM and insulin infusion pump for CSII. Insulin delivery is regulated by a computer algorithm that determines the amount of insulin to administer in response to a given sensor glucose concentration, thereby more closely approximating physiologic insulin action. Initial studies evaluating the use of a closed-loop system in the hospital setting focused on critically ill patients. These small, randomized trials demonstrated good efficacy data with improvement in time in target glucose range, and lower mean glucose levels without an increased risk of hypoglycemia.2628

More recent studies by Hovorka and colleagues2932 evaluating the use of a closed-loop system in the non-critically ill hospital setting have shown promising safety and efficacy data. In one study, patients with T2DM were randomized to receive conventional insulin treatment or insulin delivery based on a closed-loop system with ad lib meal intake and activity. Findings from the initial pilot study of 40 patients demonstrated a greater percentage of time spent with sensor glucose in target range (100–180 mg/dL or 5.6–10.0 mmol/L) in the closed-loop intervention compared with control, 59.8% versus 38.1%, and less time in hyperglycemia (difference of 19%).31 There was no significant difference in mean BG between groups or hypoglycemia rates, and no significant differences in insulin doses. Glucose variability was decreased in the closed-loop group. Patients were overwhelmingly satisfied with the use of the closed-loop system.31 Similar findings were seen in a larger multicenter study implementing the same protocol with improvement in time in target glucose range in the closed-loop group compared with controls (65.8% ± 16.8% vs 41.5% ± 16.9%, 95% confidence interval [CI], 18.6 to 30.0; P<.0013).29 A post hoc analysis in patients with T2DM admitted to the hospital on hemodialysis found that patients randomized to the closed-loop system spent 37.6% more time with sensor glucose in target range compared with standard-of-care POC glucose monitoring without an increased risk of hypoglycemia.32

Nutritional support, either through parenteral or enteral routes, frequently results in hyperglycemia in patients with and without a prior diagnosis of diabetes.33 Unique challenges exist in this population in that unplanned interruptions in feedings can place patients at risk for hypoglycemia. Use of a closed-loop system in the setting of nutritional support resulted in improved glycemic control with a higher proportion of time with glucose in target range (68.4% [standard deviation (SD) 15.5] vs 36.4% [SD 26.6], P<.0001), lower mean glucose (153 mg/dL [SD 1.2] vs 205 mg/dL [SD 3.4], 8.5 vs 11.4 mmol/L, P = .001), lower rates of hyperglycemia (32.6% less time with glucose >180 mg/dL or 10.0 mmol/L [95% CI 17.8–47.3], P<.0001), although no difference in hypoglycemia when compared with conventional insulin treatment.30

A recent observational study reported glycemic control of patients with T1DM (N = 27) who participated in randomized crossover trials during pregnancy using closed-loop during labor, delivery, and postpartum.34 Use of closed-loop was associated with 82.0% (interquartile range [IQR] 49.3, 93.0) time in target glucose range during labor and delivery and a mean glucose of 124 ± 25 mg/dL (6.9 ± 1.4 mmol/L). Closed-loop resulted in good glycemic control throughout vaginal, elective, and emergency cesarean deliveries. After delivery, women spent 83.3% of time in target glucose range (70–180 mg/dL or 3.9–10.0 mmol/L).34

Potential advantages to using a closed-loop system in the hospital setting include the ability to continually adapt insulin administration to changing glucose levels with minimal input from nursing or support staff. Use of a closed-loop system means less active management for nursing staff and therefore less risk of dosing errors compared with MDI or intravenous insulin administration, with improved glycemic control outcomes and a lower risk of iatrogenic hypoglycemia. Previous randomized controlled trials have been limited to patients with T2DM, and have mostly excluded patients with T1DM, which may be a more vulnerable population. There is concern with regard to pump or sensor failure and the need for device removal given the associated potential for diabetic ketoacidosis. Among closed-loop studies, up to 27% of patients had devices removed at least once during hospitalization.29 It is also important to note that limited data exist evaluating the impact of a closed-loop system on clinical outcomes and hospital costs. These data will be essential to justify widespread adoption of such technology, given high implementation costs and need for specialized training for health care staff.

COMPUTERIZED DECISION SUPPORT SYSTEMS FOR GLYCEMIC CONTROL IN THE HOSPITAL

The need for frequent hospital glucose monitoring and insulin titration to maintain glycemic control while avoiding hypoglycemia with the use of intravenous insulin infusion has triggered the emergence of computerized insulin dosing systems, also known as computerized decision support systems for glycemic control.35

Several systems have become commercially available to assist with glycemic management in critically ill patients with hyperglycemia, such as Glucommander (Glytec, Greenville, SC), EndoTool System (MD Scientific LLC, Charlotte, NC), and GlucoStabilizer (Medical Decision Network, Charlottesville, VA). In addition, several institutions have developed their own computerized insulin protocols and have integrated these systems into their electronic medical record (EMR), including, among others Vanderbilt University Hospital,36 Medical University of South Carolina,37 Tuft Medical Center,38 and Kaiser Sunnyside Medical Center39 (Table 3).

Table 3.

Computerized glycemic management systems

Authors Glucose Management System Design Glycemic Outcome (mg/dL)a Results
Commercially available computerized glycemic management systems
Juneja et al,40 2009 Glucostabilizer (ICU) Retrospective, descriptive %Time in target range (80–110)
%BG <70
%BG <40
73.4%
0.31%
0.1%
Newton et al,41 2010 Glucommander (ICU) Multicenter, randomized controlled trial, paper-based vs computerized Mean BG
%BG in target Range (80–120)
Patients with BG <60
117 vs 103c
51% vs 71%c
31.9 vs 42.9
Tanenberg et al,42 2017 EndoTool (ICU) Retrospective, descriptive %BG <70
%BG <40
CV%
0.93%
0.03%
26.5%
John et al,43 2018 EndoTool (ICU) Retrospective, comparative analysis of paper-based vs computerized Time in targetb
%BG <70
Hours on IV insulin infusion
47.3% vs 45.2%
0.36 vs 0.007
2.39 vs 20.9
Ullal et al,44 2018 Glucommander (DKA) Retrospective, comparative analysis of paper-based vs computerized %Patients with BG <60
%Patients with BG <40
11%−14% vs 33%−39%
0.2%−0.4% vs 6%−7.5%
Institution-specific computerized glycemic management systems
Hermayer et al,37 2007 Medical University of South Carolina Retrospective, comparative analysis of paper-based vs computerized Mean BG
%BG in target range (80–120)
%BG <70
163 vs 154c
82% vs 54%c
1.42 vs 1.14
Dortch et al,45 2008 Vanderbilt University Hospital, TN Retrospective, comparative analysis of paper-based vs computerized %BG 80–110
%BG >150
%BG ≤40
41.8% vs 34%c
12.8% vs 15.1%c
0.2% vs 0.5%c
Pachler et al,46 2008 Medical University Graz, Austria Randomized controlled trial, paper-based vs computerized Mean BG
Hyperglycemic index
106 vs 133c
0.4 vs 1.6c
Lee et al,47 2012 University of California San Diego – Burn Unit Retrospective, descriptive %BG in target range (90–150)
%BG <50
75.8%
0.07%
Saur et al,38 2013 Tufts Medical Center, MA Retrospective, comparative analysis of paper-based vs computerized Mean BG
%Time in target range (95–135)
%Time BG <70
117 vs 135c
68% vs 52%c
0.51% vs 1.44%c

Abbreviations: BG, blood glucose; CV, coefficient of variation; DKA, diabetic ketoacidosis; ICU, intensive care unit; IV, intravenous; SQ, subcutaneous.

a

BG presented in mg/dL (conversion: mg/dL × 0.0555 = mmol/L).

b

Range not described.

c

(P<.05).

Data from Refs.37,38,4047

These systems aim to direct the nursing staff on adjusting insulin infusion rates and frequency of glucose testing to optimize inpatient glycemic control and alleviate some of the increased burden of nursing care associated with titrating insulin infusions in medical or surgical critical care units. The software considers previous glucose values and recommends changes in insulin infusion based on a dynamic insulin sensitivity multiplier derived from glucose changes after insulin dose adjustments.48 Most of the software is based on proportional-integral-derivative algorithms.

Several prospective and observational studies in critically ill patients,4043,49 burn unit patients,47 and patients with diabetic ketoacidosis44 have reported that use of these systems resulted in improved glycemic control with low rates of hypoglycemia, and also less glycemic variability, when compared with standard paper-based algorithms.

Some systems also include algorithms for the management of hyperglycemia in non-critically ill patients treated with basal bolus insulin regimens, such as the Gluco-Tab (Joanneum Research GmbH [Graz, Austria] and Medical University of Graz) and Glucommander.5052 As shown in the critically ill population, these computerized decision support systems can improve protocol adherence and glycemic control without increased rates of hypoglycemia.52

Still, most institutions use standard paper-based, nursing-driven protocols, likely due to the added licensing and implementation costs associated with these systems. There are also considerations regarding compatibility requirements for integration with the electronic medical records system at each individual institution. These devices may be useful in hospitals without diabetes management teams or diabetes experts on staff; however, considerations need to be given to the potential added costs and implementation needs.

SUMMARY

The rapid evolution of diabetes technology during the past decades has led to increased use of CGM and CSII in the ambulatory setting for management of both T1DM and T2DM. In this volume of the Endocrine Clinics, experts have extensively reviewed benefits of outpatient diabetes technology use and the development of new CGM-derived glycemic control metrics. Expanding use of CGM and CSII technology has emphasized the need for more evidence regarding the continuation of these therapies during hospitalization. Recent data in hospitalized patients have shown remarkable progress in the use of diabetes technology in the hospital, including (1) improved accuracy and reliability of CGM, (2) safety of CSII in appropriate hospital populations, (3) improvement of glycemic control with computerized glycemic management systems in ICU and non-ICU settings, and (4) feasibility of inpatient CGM-CSII closed-loop systems for inpatient glycemic control. Ongoing studies are focusing on continued translation of this technology to improve glycemic control and outcomes in hospitalized patients.

KEY POINTS.

  • Ambulatory use of diabetes technology, including continuous glucose monitoring (CGM), continuous subcutaneous insulin infusion (CSII), and closed-loop systems, has rapidly expanded during the past decades, with more recent studies evaluating its translation to the hospital setting.

  • Preliminary data show improvement in detection of both hyperglycemia and hypoglycemia with use of CGM in the hospital.

  • Recent studies have tested the use of closed-loop systems in diverse populations of hospitalized patients.

  • Further investigation is needed regarding the inpatient use of diabetes technology and how it pertains to glycemic control and patient-centered outcomes.

DISCLOSURE

G.M. Davis and A.L. Migdal have no disclosures. R.J. Galindo is partly supported by a grant from the National Institute of Diabetes and Digestive and Kidney Diseases of the National Institutes of Health under Award Number P30DK11102. R.J. Galindo has received research grant support to Emory University for investigator-initiated studies from Novo Nordisk, and consulting fees from Abbott Diabetes Care, Sanofi, Valeritas, and Novo Nordisk. G.E. Umpierrez is partly supported by research grants from the National Center for Advancing Translational Sciences of the National Institutes of Health under Award Number UL1TR002378 from the Clinical and Translational Science Award program and a National Institutes of Health (NIH) grant U30, P30DK11102, and has received research grant support to Emory University for investigator-initiated studies from Sanofi, Novo Nordisk, and Dexcom.

REFERENCES

  • 1.Umpierrez GE, Klonoff DC. Diabetes technology update: use of insulin pumps and continuous glucose monitoring in the hospital. Diabetes Care 2018;41(8): 1579–89. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Wallia A, Umpierrez GE, Rushakoff RJ, et al. Consensus statement on inpatient use of continuous glucose monitoring. J Diabetes Sci Technol 2017;11(5): 1036–44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Peters AL, Ahmann AJ, Battelino T, et al. Diabetes technology-continuous subcutaneous insulin infusion therapy and continuous glucose monitoring in adults: an Endocrine Society clinical practice guideline. J Clin Endocrinol Metab 2016; 101(11):3922–37. [DOI] [PubMed] [Google Scholar]
  • 4.Fonseca VA, Grunberger G, Anhalt H, et al. Continuous glucose monitoring: a consensus Conference of the American Association of Clinical Endocrinologists and American College of Endocrinology. Endocr Pract 2016;22(8):1008–21. [DOI] [PubMed] [Google Scholar]
  • 5.Grunberger G, Abelseth JM, Bailey TS, et al. Consensus statement by the American Association of Clinical Endocrinologists/American College of Endocrinology insulin pump management task force. Endocr Pract 2014;20(5):463–89. [DOI] [PubMed] [Google Scholar]
  • 6.Clarke SF, Foster JR. A history of blood glucose meters and their role in self-monitoring of diabetes mellitus. Br J Biomed Sci 2012;69(2):83–93. [PubMed] [Google Scholar]
  • 7.Gomez AM, Umpierrez GE. Continuous glucose monitoring in insulin-treated patients in non-ICU settings. J Diabetes Sci Technol 2014;8(5):930–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Torimoto K, Okada Y, Sugino S, et al. Determinants of hemoglobin A1c level in patients with type 2 diabetes after in-hospital diabetes education: a study based on continuous glucose monitoring. J Diabetes Investig 2017;8(3):314–20. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Levitt DL, Silver KD, Spanakis EK. Mitigating severe hypoglycemia by initiating inpatient continuous glucose monitoring for type 1 diabetes mellitus. J Diabetes Sci Technol 2017;11(2):440–1. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Logtenberg SJ, Kleefstra N, Snellen FT, et al. Pre- and postoperative accuracy and safety of a real-time continuous glucose monitoring system in cardiac surgical patients: a randomized pilot study. Diabetes Technol Ther 2009;11(1):31–7. [DOI] [PubMed] [Google Scholar]
  • 11.Holzinger U, Warszawska J, Kitzberger R, et al. Real-time continuous glucose monitoring in critically ill patients: a prospective randomized trial. Diabetes Care 2010;33(3):467–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Kopecky P, Mraz M, Blaha J, et al. The use of continuous glucose monitoring combined with computer-based eMPC algorithm for tight glucose control in cardiosurgical ICU. Biomed Res Int 2013;2013:186439. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Umbrello M, Salice V, Spanu P, et al. Performance assessment of a glucose control protocol in septic patients with an automated intermittent plasma glucose monitoring device. Clin Nutr 2014;33(5):867–71. [DOI] [PubMed] [Google Scholar]
  • 14.De Block CE, Gios J, Verheyen N, et al. Randomized evaluation of glycemic control in the medical intensive care unit using real-time continuous glucose monitoring (REGIMEN Trial). Diabetes Technol Ther 2015;17(12):889–98. [DOI] [PubMed] [Google Scholar]
  • 15.Burt MG, Roberts GW, Aguilar-Loza NR, et al. Brief report: comparison of continuous glucose monitoring and finger-prick blood glucose levels in hospitalized patients administered basal-bolus insulin. Diabetes Technol Ther 2013;15(3):241–5. [DOI] [PubMed] [Google Scholar]
  • 16.Schaupp L, Donsa K, Neubauer KM, et al. Taking a closer look-Continuous glucose monitoring in non-critically ill hospitalized patients with type 2 diabetes mellitus under basal-bolus insulin therapy. Diabetes Technol Ther 2015;17(9): 611–8. [DOI] [PubMed] [Google Scholar]
  • 17.Gómez AM, Umpierrez GE, Munoz OM, et al. Continuous glucose monitoring versus capillary point-of-care testing for inpatient glycemic control in type 2 diabetes patients hospitalized in the general ward and treated with a basal bolus insulin regimen. J Diabetes Sci Technol 2015;10(2):325–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Gu W, Liu Y, Chen Y, et al. Multicentre randomized controlled trial with sensor-augmented pump vs multiple daily injections in hospitalized patients with type 2 diabetes in China: Time to reach target glucose. Diabetes Metab 2017;43(4): 359–63. [DOI] [PubMed] [Google Scholar]
  • 19.Spanakis EK, Levitt DL, Siddiqui T, et al. The effect of continuous glucose monitoring in preventing inpatient hypoglycemia in general wards: the glucose telemetry system. J Diabetes Sci Technol 2018;12(1):20–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Battelino T, Danne T, Bergenstal RM, et al. Clinical targets for continuous glucose monitoring data interpretation: recommendations from the international consensus on time in range. Diabetes Care 2019;42(8):1593–603. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Association AD. 15. Diabetes care in the hospital: standards of medical care in diabetes—2019. Diabetes Care 2019;42(Supplement 1):S173–81. [DOI] [PubMed] [Google Scholar]
  • 22.Kannan S, Satra A, Calogeras E, et al. Insulin pump patient characteristics and glucose control in the hospitalized setting. J Diabetes Sci Technol 2014;8(3): 473–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Nassar AA, Partlow BJ, Boyle ME, et al. Outpatient-to-inpatient transition of insulin pump therapy: successes and continuing challenges. J Diabetes Sci Technol 2010;4(4):863–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Cook CB, Beer KA, Seifert KM, et al. Transitioning insulin pump therapy from the outpatient to the inpatient setting: a review of 6 years’ experience with 253 cases. J Diabetes Sci Technol 2012;6(5):995–1002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Levitt DL, Spanakis EK, Ryan KA, et al. Insulin pump and continuous glucose monitor initiation in hospitalized patients with type 2 diabetes mellitus. Diabetes Technol Ther 2018;20(1):32–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Leelarathna L, English SW, Thabit H, et al. Feasibility of fully automated closed-loop glucose control using continuous subcutaneous glucose measurements in critical illness: a randomized controlled trial. Crit Care 2013;17(4):R159. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Chee F, Fernando T, van Heerden PV. Closed-loop glucose control in critically ill patients using continuous glucose monitoring system (CGMS) in real time. IEEE Trans Inf Technol Biomed 2003;7(1):43–53. [DOI] [PubMed] [Google Scholar]
  • 28.Yatabe T, Yamazaki R, Kitagawa H, et al. The evaluation of the ability of closed-loop glycemic control device to maintain the blood glucose concentration in intensive care unit patients. Crit Care Med 2011;39(3):575–8. [DOI] [PubMed] [Google Scholar]
  • 29.Bally L, Thabit H, Hartnell S, et al. Closed-loop insulin delivery for glycemic control in noncritical care. N Engl J Med 2018;379(6):547–56. [DOI] [PubMed] [Google Scholar]
  • 30.Boughton CK, Bally L, Martignoni F, et al. Fully closed-loop insulin delivery in inpatients receiving nutritional support: a two-centre, open-label, randomised controlled trial. Lancet Diabetes Endocrinol 2019;7(5):368–77. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 31.Thabit H, Hartnell S, Allen JM, et al. Closed-loop insulin delivery in inpatients with type 2 diabetes: a randomised, parallel-group trial. Lancet Diabetes Endocrinol 2017;5(2):117–24. [DOI] [PubMed] [Google Scholar]
  • 32.Bally L, Gubler P, Thabit H, et al. Fully closed-loop insulin delivery improves glucose control of inpatients with type 2 diabetes receiving hemodialysis. Kidney Int 2019;96(3):593–6. [DOI] [PubMed] [Google Scholar]
  • 33.Pasquel FJ, Spiegelman R, McCauley M, et al. Hyperglycemia during total parenteral nutrition: an important marker of poor outcome and mortality in hospitalized patients. Diabetes Care 2010;33(4):739–41. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Stewart ZA, Yamamoto JM, Wilinska ME, et al. Adaptability of closed loop during labor, delivery, and postpartum: a secondary analysis of data from two randomized crossover trials in type 1 diabetes pregnancy. Diabetes Technol Ther 2018; 20(7):501–5. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35.Eslami S, Abu-Hanna A, de Jonge E, et al. Tight glycemic control and computerized decision-support systems: a systematic review. Intensive Care Med 2009; 35(9):1505–17. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 36.Boord JB, Sharifi M, Greevy RA, et al. Computer-based insulin infusion protocol improves glycemia control over manual protocol. J Am Med Inform Assoc 2007;14(3):278–87. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Hermayer KL, Neal DE, Hushion TV, et al. Outcomes of a cardiothoracic intensive care Web-based online intravenous insulin infusion calculator study at a medical university hospital. Diabetes Technol Ther 2007;9(6):523–34. [DOI] [PubMed] [Google Scholar]
  • 38.Saur NM, Kongable GL, Holewinski S, et al. Software-guided insulin dosing: tight glycemic control and decreased glycemic derangements in critically ill patients. Mayo Clin Proc 2013;88(9):920–9. [DOI] [PubMed] [Google Scholar]
  • 39.Olinghouse C Development of a computerized intravenous insulin application (AutoCal) at Kaiser Permanente Northwest, integrated into Kaiser Permanente HealthConnect: impact on safety and nursing workload. Perm J 2012;16(3): 67–70. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Juneja R, Roudebush CP, Nasraway SA, et al. Computerized intensive insulin dosing can mitigate hypoglycemia and achieve tight glycemic control when glucose measurement is performed frequently and on time. Crit Care 2009; 13(5):R163. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Newton CA, Smiley D, Bode BW, et al. A comparison study of continuous insulin infusion protocols in the medical intensive care unit: computer-guided vs. standard column-based algorithms. J Hosp Med 2010;5(8):432–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Tanenberg RJ, Hardee S, Rothermel C, et al. Use of a computer-guided glucose management system to improve glycemic control and address national quality measures: a 7-year, retrospective observational study at a tertiary care teaching hospital. Endocr Pract 2017;23(3):331–41. [DOI] [PubMed] [Google Scholar]
  • 43.John SM, Waters KL, Jivani K. Evaluating the implementation of the EndoTool Glycemic Control Software System. Diabetes Spectr 2018;31(1):26–30. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Ullal J, Aloi JA, Reyes-Umpierrez D, et al. Comparison of computer-guided versus standard insulin infusion regimens in patients with diabetic ketoacidosis. J Diabetes Sci Technol 2018;12(1):39–46. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Dortch MJ, Mowery NT, Ozdas A, et al. A computerized insulin infusion titration protocol improves glucose control with less hypoglycemia compared to a manual titration protocol in a trauma intensive care unit. JPEN J Parenter Enteral Nutr 2008;32(1):18–27. [DOI] [PubMed] [Google Scholar]
  • 46.Pachler C, Plank J, Weinhandl H, et al. Tight glycaemic control by an automated algorithm with time-variant sampling in medical ICU patients. Intensive Care Med 2008;34(7):1224–30. [DOI] [PubMed] [Google Scholar]
  • 47.Lee J, Fortlage D, Box K, et al. Computerized insulin infusion programs are safe and effective in the burn intensive care unit. J Burn Care Res 2012;33(3):e114–9. [DOI] [PubMed] [Google Scholar]
  • 48.Rattan R, Nasraway SA. The future is now: software-guided intensive insulin therapy in the critically ill. J Diabetes Sci Technol 2013;7(2):548–54. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Umpierrez G, Cardona S, Pasquel F, et al. Randomized controlled trial of intensive versus conservative glucose control in patients undergoing coronary artery bypass graft surgery: GLUCO-CABG trial. Diabetes Care 2015;38(9):1665–72. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Mader JK, Neubauer KM, Schaupp L, et al. Efficacy, usability and sequence of operations of a workflow-integrated algorithm for basal-bolus insulin therapy in hospitalized type 2 diabetes patients. Diabetes Obes Metab 2014;16(2):137–46. [DOI] [PubMed] [Google Scholar]
  • 51.Neubauer KM, Mader JK, Holl B, et al. Standardized glycemic management with a computerized workflow and decision support system for hospitalized patients with type 2 diabetes on different wards. Diabetes Technol Ther 2015;17(10): 685–92. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 52.Aloi J, Bode BW, Ullal J, et al. Comparison of an electronic glycemic management system versus provider-managed subcutaneous basal bolus insulin therapy in the hospital setting. J Diabetes Sci Technol 2017;11(1):12–6. [DOI] [PMC free article] [PubMed] [Google Scholar]

RESOURCES