Mammographic compression practices of force‐ and pressure‐standardisation protocol: A scoping review

Elizabeth Serwan 1, Donna Matthews 1, Josephine Davies 2, Minh Chau 1,
PMCID: PMC7476195  PMID: 32420700

Abstract

Introduction

As an efficient, effective and moderately inexpensive modality, mammography has been implemented as a cancer screening tool and in diagnostic management. However, appropriate breast compression is necessary for optimal outcomes. Current key measures of compression force are subjective and variable, giving rise to the concept of a ‘personalised’ pressure‐standardisation protocol.

Methods

A scoping review of the literature was performed using the Arksey and O’Malley framework to explore the existing force‐ and pressure‐standardisation protocols in clinical application. A comprehensive search strategy and standardised study selection and evaluation were completed. This synthesis of existing knowledge can lead to the implementation of mechanically standardised mammographic compression pressure as a feasible tailored approach to clinical practice. Four databases (PubMed, MEDLINE, Embase and Scopus) were searched from the databases’ inception to 13 December 2019 for relevant information, and eighteen articles were selected for analysis.

Results

In addition to current protocol comparison, emerging key concepts include the reasoning behind standardisation, the benefits of improved diagnostic outcomes/decreased pain with negligible change in image quality and average glandular dose (AGD), and the recommendation of a 10kPa (approximate) pressure‐standardisation protocol. Research to date is largely based abroad (Netherlands), with a strong focus on screening practices. Consequently, several gaps in the current literature were identified as potential directions for future investigation.

Conclusions

As a suggested mammographic guideline, compression pressures of approximately 10kPa aid in image acquisition reproducibility both within and between women; pain levels decrease, with minimal variations to breast thickness, AGD and image quality.

Keywords: breast, mammographer, mammography, review, screening


Current methods of mammographic compression are subjective and variable. Research supports an alternate standardised compression protocol founded on pressure; this approach accounts for individual breast characteristics in a ‘personalised’ manner. A compression pressure of approximately 10kPa was found to decrease pain, with a negligible effect on breast thickness, average glandular dose, and resultant image quality.

graphic file with name JMRS-67-233-g002.jpg

Introduction

As breast cancer is one of the most prevalent female cancers, it is unsurprising that mammographic screening programs have been implemented worldwide. 1 Such programs are estimated to result in an approximate annual reduction in breast cancer mortality by 30%. 2 Additionally, mammography is a cost‐effective and gold standard practice. 3 Sources report high sensitivity and the highest specificity in mammographic screening relative to MRI; higher MRI sensitivity does not translate into improved clinical outcomes due to the ineffective detection of ductal carcinoma in situ. 3 Furthermore, mammography is the preliminary imaging modality used in most diagnostic settings (depending on patient age); calcium deposits, most frequently benign, can be indicative of potentially cancerous pathological change, which are usually seen clearly on a mammogram. 4 , 5 Breast compression is necessary for any successful mammographic examination; many sources discuss the effect of a uniform breast tissue thickness on improving image contrast and quality, with an associated reduction in radiation dose, geometric/motion blurring and tissue superimposition. 1 , 5 , 6 , 7 , 8 , 9 , 10 , 11 , 12 , 13 , 14 Whilst ultimately aiding in the detection of pathology, the discomfort attributed to breast compression remains the highest reason for patient non‐compliance with screening programs and lack of diagnostic lesion visibility. 9 , 15 , 16 According to the literature, the key factors affecting compression can be attributed to the patient, the mammographer and/or the equipment. 17 However, compression standardisation is difficult due to the lack of explicit criteria to assess consistency; the predictability and reproducibility of mammographic examinations are thus difficult to determine. 13 This is increasingly apparent when comparing the mammography protocols utilised on a global scale, in conjunction with international approaches to clinical practice – the specification of concrete parameters and/or guidelines is brief and minimal, with sufficient compression seemingly based on mammographer opinion and experience. The only objective measures currently obtainable in real time are those of compression force and breast thickness; these are gained mechanically from the mammography system, yet do not account for individual variations in breast size and elasticity. 13 This has led to recent investigation of pressure (expressed in kilopascals – kPa) as a ‘personalised’ adaptation of force. Since pressure considers both the force and contact area simultaneously (pressure = compression force/breast contact area), it has been reported as a more physiologically appropriate compression parameter than force. 14 In comparison, an opinion‐based application of force is variable and unreliable, whilst a mechanical application of target pressure is consistent and reproducible. Consequently, recent research has stemmed into developing an objective pressure‐standardisation protocol for clinical implementation, which allows for real‐time appraisal. Therefore, the aim of this review is to assess the feasibility of mechanically standardised mammographic compression in clinical practice. This will be evaluated in comparison with current techniques – both nationally and internationally – with the perceived benefits, recommendations and limitations outlined.

Methods

Method of review

A scoping, or ‘mapping’, review was performed to assess the current standing of mammographic compression practices available in existing literature. The organisational framework, derived as described by Arksey and O’Malley, was chosen as a means of evaluating the extent of available evidence for a clinical overview. 18 As information from existing evidence was extracted and summarised for this scoping review, rather than methodologically appraised and statistically analysed, a systematic review was deemed unsuitable. To establish the context of this research in the wider mammographic community, the following research questions were pondered:

  • Would mammographic practices benefit from mechanical standardisation of objective parameters?

  • In what ways can compression force and compression pressure be justified as an appropriate measure of standardisation?

  • How are outcomes relevant from both a clinical and patient‐centred perspective?

Search strategy

Four electronic databases (PubMed, MEDLINE, Embase and Scopus) were searched for English‐language articles discussing mammographic compression practices from the databases’ inception to 13 December 2019; most sources were found to be relatively recent due to the nature of the topic. Structured search strategies were devised to optimise findings; the initial search began with the Medical Subject Heading (MeSH) term ‘mammography’, which was altered to ‘mammograph*’ as a database entry. Further keywords added to the search strategy included (AND) ‘compress*’ OR ‘force’ OR ‘pressure’, AND ‘standard*’. Although no relevant MeSH terms exist for such keywords, these were deemed relevant to the aims of the research. Search fields were set at ‘all fields’ for each database, except for Scopus, which was limited to ‘article title, abstract, keywords’. The number of hits from each database was 53, 35, 43 and 60 articles from Scopus, PubMed, MEDLINE and Embase, respectively.

Following the addition of hand‐sought records from expert contribution and duplicate removal by a single researcher, titles and abstracts were screened independently by two researchers. The independent screening and reviewing of eligible studies were in line with the 2005 scoping review framework by Arksey and O’Malley, 18 as well as the Preferred Reporting Items for Systematic Review and Meta‐Analysis (PRISMA) 19 guidelines. Specified eligibility criteria for each step of the selection process were adhered to systematically, with eligible articles then selected by reading the full text. Eligible titles were those relevant to ‘mammography’ or ‘breast’, and eligible abstracts discussed at least one aspect of ‘compression’, ‘force’ or ‘pressure’. Review articles, case reports, conference reports, letters, editorial comments, opinions (including qualitative reports) and non‐English articles were excluded. Other exclusion criteria included articles purely discussing quality assurance/physics without reference to clinical applicability. Any disagreement was discussed and resolved by consensus among the entire research team. The team comprised of researchers with a medical imaging background, including an experienced breast imaging specialist with post‐graduate qualification in breast imaging. The entire research team also had extensive experience in conducting scoping reviews, systematic reviews and meta‐analyses.

Literature analysis

A data extraction form was developed using Microsoft Excel to collect information related to (i) country of origin and current practices, (ii) the perceived need for standardisation, (iii) benefits, (iv) recommendations and (v) limitations. Information compiled in the extraction form was reviewed by two researchers, with data analysed for key concepts and frequencies. These included overall benefits of the standardisation protocols, such as diagnostic outcomes, image quality, AGD and patient experience. A total of 18 articles were included for analysis in this scoping review; the PRISMA flow chart (Figure 1) details the review process, whilst Table 1 presents the summary of the reviewed studies, arranged alphabetically:

Figure 1.

Figure 1

PRISMA flow diagram for scoping review data evaluation 19

Table 1.

This table presents the data extraction and analysis of the selected studies.

  Authors Country of origin Protocol used Study details Why standardise? Benefits of pressure standardisation Recommendations Limitations
Diagnostic outcome Image quality AGD Patient experience
1 Branderhorst et al. 13 Netherlands (Netherlands vs USA)

Netherlands: Dutch screening QA protocol – 12daN.

USA: no target – taut breast, unless painful

Retrospective cohort:

Netherlands: screening asymptomatic patients, aged 50–75; CC + MLO

USA: screening/diagnostic (symptomatic/ asymptomatic), aged 50–75; CC + MLO

Account for contact area; contact area decreases, average pressure/variance increase, despite lower force       Reduced pain 10kPa standard pressure; reduced variation and clear standards Screening vs diagnostic cohort
2 Branderhorst et al. 20 Netherlands  

Prospective cohort:

Real‐time and retrospective methods/software studied on asymptomatic patients; CC + MLO

Ensure contact area between breast/paddle can be determined for accurate pressure readings         Current systems give accurate/ precise pressure readings Positioning discrepancies; diagnostic application
3 de Groot et al. 8 Netherlands Compress to 18daN or until pain threshold Prospective observational: Hospital setting excluding interventions, patients aged 30–88; CC + MLO Force‐controlled compression causes large pressure variation; pressure‐controlled accounts for contact area as predictive parameter for severe pain   Not compromised Unchanged Reduced pain 10–12kPa standard pressure; increased thickness of 9% (small) and 2% (large) breasts Clinical only – not representative of screening population
4 de Groot et al. 21 Netherlands 18daN compression protocol

Prospective observational:

Real‐time examination of post‐intervention (unilateral) patients

Protocol applicable to post‐conservation therapy population       Reduced pain post‐intervention Support for 10kPa standard pressure Post‐intervention only
5 de Groot et al. 10 Netherlands (comparing Netherlands with USA)

Netherlands: Dutch screening QA protocol – 12daN.

USA: no target – taut breast, unless painful

Retrospective cohort:

Netherlands: screening asymptomatic patients, aged 50–75; CC + MLO

USA: screening/diagnostic (symptomatic/ asymptomatic), aged 50–75; CC + MLO

Account for contact area; contact area decreases, average pressure/variance increase, despite lower force   Potentially unchanged Potentially unchanged Potentially reduced pain 10kPa standard pressure Screening vs diagnostic cohort
6 de Groot et al. 14 Netherlands 14daN force standardisation, to be compared with 10kPa pressure standardisation

Prospective cohort:

Screening asymptomatic patients aged 50–75; CC + MLO

Force applied in proportion to contact area (account for breast size)   Not compromised Similar between protocols Reduced pain: 1/2 cohort = less pain, 1/3 = unchanged 10kPa standard pressure (between venous/arterial blood pressure) Screening only – not extend conclusions to lesion detection
7 de Groot et al. 5 Netherlands 18daN force standardisation, compared to 10kPa standard pressure

Case–control:

Case pairing of patients over time, with alternate protocol used for each; CC + MLO

Ensure lesion appearance is consistently satisfactory   Unchanged   Reduced unnecessary pain Support for 10kPa standard pressure Changes in practice between examinations
8 den Boer et al. 7 Netherlands Compress to 100–150N or until tolerance

Retrospective cohort:

Screening follow‐up of asymptomatic patients, aged 36–74: CC + MLO (separate)

Compression as a function of contact area, to apply pressure independent of breast size   Not compromised, more consistent   Reduced pain 9.6–12kPa standard pressure Time delay between successive acquisitions
9 Holland et al. 12 Netherlands  

Retrospective cohort:

Screening asymptomatic patients, aged 50–75; MLO only

Possible improvement to screening program performance Moderate pressures best (statistically insignificant)       Quantitative protocol; high compression worse on lesion visibility than low No CC view
10 Holland et al. 6 Netherlands   Retrospective cohort: Screening asymptomatic patients, aged 50–75; MLO views only Possible improvement to screening program performance High pressure decreases sensitivity; low pressure decreases specificity   Unchanged Reduced pain No optimal pressure; high compression worse on lesion visibility than low No CC view; international screening differences; pectoral interference
11 Jeukens et al. 23 Netherlands 10kPa standard compression

Retrospective cohort:

Screening follow‐up of asymptomatic patients, mean age 59; CC view

Reduce pain due to compression     No clinically relevant difference No clinically relevant difference None of clinical relevance Psychological impact on post‐intervention/ follow‐up examinations
12 Lau et al. 24 Malaysia Compress until taut or intolerable pain (no target force) Retrospective cohort: Screening/diagnostic imaging of symptomatic/asymptomatic patients, aged 35–80 (clinical study); CC + MLO Compression based on contact area; optimise protocols for Asian women (assumed to have smaller breasts)   Not compromised Unchanged Potentially reduced pain Compression reduction of 32.5% from 12–9daN Assume Asian women have small breasts, and study phantom = human tissue
13 Mercer et al. 17 United Kingdom Subjective mammographer judgement

Retrospective cohort:

Evaluating mammographer practice in screening service

Minimise compression practice variance between/within mammographers         Focus on training process Variation due to inconsistent examinations/patient modification
14 Mercer et al. 15 United Kingdom Subjective mammographer judgement

Retrospective cohort:

Evaluating mammographer practice in screening service

Minimise compression practice variance between/within mammographers   Improved consistency Improved consistency Varied force impact screening attendance Consistently applied force per patient, regardless of mammographer Small, single‐centre study
15 Mercer et al 16 United Kingdom Subjective mammographer judgement

Retrospective cohort:

Evaluating mammographer practice in screening service

Minimise compression practice variance between/within mammographers Potential improvement in cancer detection Potential positive impact Potential reduction Stabilisation may increase re‐attendance Establish guidelines for cessation of compression force Geographical proximity implies similar training
16 Moshina et al. 1 Norway  

Retrospective cohort:

Screening asymptomatic patients aged 50–69; CC + MLO

Optimise the performance of screening measures High force (proportional to pressure) decreases sensitivity and specificity Possible improvement (inconclusive)     Force of at least 130N, pressure less than 9.8kPa Patient positioning of questionable quality
17 Moshina et al. 22 Norway 10kPa standard compression

Prospective cohort:

Screening asymptomatic patients aged 50–69; CC + MLO

Minimise pain experienced during screening mammography   Reduced with pressure‐standardised paddle (inconclusive) Increased with pressure‐standardised paddle Higher pain with fixed paddle – clinically inconclusive None of clinical relevance Subjectivity of data acquisition
18 Poulos & McLean 9 Australia ‘breast is taut at the sides'; 'skin blanches'

Prospective cohort:

Based on participants in screening setting

More objective criteria for application of compression force         Compression practice focus: minimise breast thickness  

Results

The retained articles consisted of a mixture of quantitative study methods. There were six prospective 8 , 9 , 14 , 20 , 21 , 22 and eleven retrospective studies, 1 , 6 , 7 , 10 , 12 , 13 , 15 , 16 , 17 , 23 , 24 with one case–control study. 5 Publication dates ranged from 2004 to 2019, with the last publication being a 2019 retrospective cohort study. 24 Eleven studies were sourced from the Netherlands, 5 , 6 , 7 , 8 , 10 , 12 , 13 , 14 , 20 , 21 , 23 with one remaining publication originating from Malaysia, 24 two from Norway, 1 , 22 three from the United Kingdom 15 , 16 , 17 and one from Australia. 9 Twelve studies focussed on an asymptomatic screening population; 1 , 6 , 7 , 9 , 12 , 14 , 15 , 16 , 17 , 20 , 22 , 23 however, three studies investigated a diagnostic/post‐intervention population; 5 , 8 , 21 and three studies considered both screening/diagnostic populations simultaneously. 10 , 13 , 24 Twelve studies 1 , 5 , 7 , 8 , 10 , 13 , 14 , 15 , 16 , 20 , 22 , 24 investigated both standard mammographic views (cranio‐caudal (CC) and mediolateral oblique (MLO)), two studies investigated only the MLO view, 6 , 12 and one study investigated only the CC view. 23 Two studies 22 , 23 were conducted from an equipment‐based perspective, focussing mainly on the compression paddle.

Discussion of findings

This study aimed to provide a structured insight into the potential benefits of implementing mechanically standardised mammographic compression pressure in clinical practice. Findings are assessed in the following discussion alongside current practices and directions for suggested future research.

Current practices

The subjectivity and lacking consistency of current compression guidelines are key concerns raised in much of the current literature; many studies describe the variation in applied compression force among mammographers, screening centres and countries. 1 , 5 , 6 , 7 , 8 , 9 , 10 , 13 , 14 , 15 , 16 , 20 , 24 Clinically, force is typically adjusted in an experience‐based manner according to breast size, elasticity and pain threshold; 8 , 10 , 13 , 20 , 21 , 24 Mercer et al. suggest that variation in compression may be associated with technique adaptation, as opposed to inconsistent practice. 17 These parameters will differ significantly in a given population; thus, whilst conventional methods attempt to minimise disparity with the implementation of a target force, this is independent of individual breast characteristics. 6 , 8 , 13 Further to this, it is noted that the compression forces used during mammographic examinations are attributed to the mammographer, as opposed to the patient. 15 , 16 Current European guidelines propose that ‘the breast should be properly compressed, but no more than is necessary to achieve a good image quality’; 11 this corresponds with a maximum compressive force of 200N. 10 Likewise, the United States of America (USA) recommendations suggest that the breast be compressed until ‘taut’ or ‘just less than painful’, 10 , 13 whilst studies conducted in Malaysia utilise similar subjective techniques. 24 Dutch screening programs operate with a force compression protocol, with targets ranging between 12–18 daN. 5 , 8 , 10 , 13 , 14 , 21 A study completed by Branderhorst et al. comparing the mammographic compression practices within the USA and Netherlands found targets (mean ± standard deviation) to be 13.7 ± 5.9kPa versus 8.1 ± 4.1kPa, and 13.8 ± 2.7 daN versus 7.4 ± 3.1 daN for the Netherlands and USA, respectively. 11 Overall, large variations in force, and even larger variations in pressure, were listed within and between both data sets, with distinctly higher compression limits in the Dutch data. 10 , 13 The variations in applied force and pressure during routine mammographic compression are supported with a phantom study completed by Lau et al.. 24 Regarding Australian practices, no formalised documentation of national protocols could be found in the literature, aside from stating the maximum operable force on a mammographic unit; this is 200N for motorised force and 300N for manual force. 25 However, the subjective criteria of breast tautness and skin blanching are widely accepted. 9 Overall, this seems to highlight a clinical need for standardised protocol.

The need for standardisation

Pressure standardisation operates on the premise of compression force applied by the paddle to the breast, divided by the contact area between the paddle and the breast. 6 , 7 , 8 , 13 , 14 By definition, it is most probable that pain experienced upon compression is more closely correlated with pressure than force. 6 For completeness, Branderhorst et al. confirm that the contact area between the breast and compression paddle can be accurately and precisely determined with existing technology. 20 Current literature suggests pressure is a ‘personalised’ version of force, 14 with relevance to physiological factors such as tissue composition and blood pressure. 8 , 14 The theoretical dependence on contact area is of particular note; the same force results in a higher pressure when applied to a small contact area (i.e. small breast) than a large contact area (i.e. large breast). 1 , 6 , 7 , 8 , 10 , 13 , 14 , 20 , 21 , 24 The Dutch study conducted by Branderhorst et al. described a trend between force and contact area; average pressure and variance increases as contact area decreases. 13 This was found to correlate with pain severity; patients with small breasts experience severe pain more often than large breasts when force standardisation is used. 14 A Dutch study conducted by de Groot et al. reached similar conclusions; an intra‐individual comparison of 14daN force standardisation and 10kPa pressure standardisation was found to decrease pain severity without compromising image quality. 14 A Norwegian study conducted by Moshina et al. extended these findings, concluding that pressure standardisation could optimise the performance of early screening measures. 1 Ultimately, pressure‐standardisation protocols satisfy the need for objective criteria in mammographic breast compression; this would offer concrete mammographer guidelines, whilst consistently adjusting for individual breast parameters, such as size and elasticity.

Benefits of pressure standardisation

The diagnostic outcomes of pressure‐standardised breast compression have been statistically evaluated in recent research. Generally, it was found that excessive pressure decreases mammographic sensitivity, whilst insufficient pressure decreases mammographic specificity. 6 This reflects the findings described by Moshina et al., whereby a high compression force (directly proportional to pressure) decreased both sensitivity and specificity. 1 The apparent conflict between the specificity outcome can be attributed to factors influencing resultant image contrast at the extreme ends of the compression scale. Also, Moshina et al. based conclusions on force alone, independent of contact area. 1 This is important to note as results cannot be aligned with measures of compression pressure.

Current literature concludes that image quality remains unchanged with a pressure‐standardisation protocol; 5 , 7 , 8 , 10 , 14 , 24 no studies have shown otherwise. De Groot et al. investigated this trend explicitly, noting that although breast compression is milder with a pressure‐standardisation protocol, the visibility, contrast and sharpness of stable lesions remain virtually unchanged. 5 Another study even proposed that a force reduction of as much as one third is possible with minimal impact on image quality. 24 As noted by recent research, a large range of pressures result in diagnostically acceptable images in the digital setting; 10 , 14 a minimum must exist, though, as image quality is obviously degraded without compression. Nevertheless, Holland et al. suggest over‐compression may diminish image quality to a greater extent than under‐compression; lesion detectability is compromised as suspicious densities are dissolved. 6 However, this primarily correlates with the high pressures required for spot compression. The opposite can likewise be reasoned; too little compression degrades image quality in that the potential for small cancers to remain undetected increases. 9

The amount of radiation dose received by the patient with a shift in compression protocol was also investigated. Overall, it could be concluded that pressure standardisation has a negligible impact on AGD, despite the slight increase in breast thickness. 6 , 8 , 10 , 14 , 24 Whilst the literature does demonstrate variation in data surrounding the actual dose received, this may possibly be attributed to different system types and automatic exposure control (AEC) settings. 13 The proportion of image repeats was also found to remain unchanged. 14

Although another recurring issue associated with breast compression is the patient experience of pain, a pressure‐based protocol was found to reduce unnecessary pain. 5 , 6 , 7 , 10 , 13 , 14 , 21 This has the potential to positively impact compliance with screening programs 9 , 16 , 17 and benefit the post‐intervention patients for whom follow‐up mammograms are mandatory. 21 De Groot et al. revealed that the implementation of pressure standardisation resulted in half the cohort reporting less pain and a third reporting no difference when compared to force standardisation. 14 However, this appears to apply to protocols in entirety; studies explicitly comparing force‐ and pressure‐controlled compression paddles found no preferential basis for either in clinical practice. 22 , 23 Moshina et al. suggest that pressure‐standardised paddles may reduce image quality and increase AGD, 22 although this requires additional research beyond the scope of this review to validate.

Recommendations

Current literature offers a pressure‐standardised compression protocol of approximately 10kPa, with some variation. 1 , 7 , 8 , 10 , 13 , 14 , 24 Although many studies recorded higher average values in the actual results, this can be attributed to subjective force standardisation as the foundation of data collection. One study indicated positive outcomes occur with a pressure reduction from 12–9kPa, 24 another proposed an ideal force of at least 130N with associated pressure less than 9.8kPa, 1 another recommends a standard pressure between 9.6 and 12kPa, 7 and yet others suggest 10kPa results in an ideal tissue pressure between that of normal venous and arterial blood pressure. 14 , 21 Research conducted by de Groot et al. similarly concluded that a compression of 10–12kPa corresponds with breast arterial pressure, although an increase in thickness of 9% for small and 2% for large breasts was also recorded. 8 Whilst not ideal, this increase in thickness may be considered negligible due to the large dynamic range of digital mammography systems, as discussed previously. This notion is supported by Lau et al; a reduction in compression force of 32.5% has minimal effect on image quality and AGD. 24 Low pressure and high breast volume are considered ideal for enhanced cancer detection, 1 with pressures between 9.2 and 10.7kPa resulting in the highest detection rate. 10 , 14 Further to this, high compression has been shown to reduce lesion visibility to a greater extent than insufficient compression. 6 , 10 , 12 Mercer et al. also evaluate this from a mammographer’s perspective, suggesting that extending mammographer training may aid performance and technique consistency. 15 , 16 , 17 Another recommendation includes the need for peer observation to allow for comparison of a mammographer’s own force compression practice to that of colleagues. 26

Limitations

Several drawbacks arose when analysing the literature, many of which are applicable across multiple studies. Firstly, generalising conclusions to an external population is unwise, as screening policies and populations differ across countries. 6 This also raises the issue of extending results between screening and diagnostic mammography; as suggested by the literature, a diagnostic or post‐intervention cohort may experience increased sensitivity to pain and hence are less tolerant to compression. 8 , 10 , 13 , 14 , 20 , 21 A study conducted by de Groot et al. supports this concept; the odds for severe pain in women post‐breast‐conserving therapy was 5.3 times higher than normal, largely due to changes in breast composition (elasticity). 21 Accounting for differences between standard mammographic views (CC and MLO) is also important; the inherent discrepancies resulting from alternating the positioning technique limit the comparability of recorded values, both within and between studies. 6 , 12 Additionally, the impact of pectoral muscle inclusion in the MLO contact area was reported; due to its presence, the computed pressure does not accurately reflect the pressure exerted on the breast tissue. 6 Clinically, the appearance of specific lesion types experiencing over‐ or under‐compression is inconclusive. 6 Furthermore, variations across study types are useful to note; parameters available for acquisition and analysis differ depending on whether data collection was prospective or retrospective.

Limitations inherent in the review process itself must also be acknowledged. Timeframe limits were not applied, although dates were consciously noted throughout analysis; most studies were relatively recent regardless (i.e. all 2013 onwards, excluding two studies), given this field is currently evolving. Applying language‐limiting parameters resulted in the omission of non‐English sources; hence, scholarly evidence in foreign languages is not captured in this review. Upon analysis, most information originated from the Netherlands; this may demonstrate selection bias in that evidence for favoured outcomes is strengthened, yet similar conclusions were reached in the other literature. An extended search of more databases and ‘grey’ literature would confirm this. In fact, the distribution of resource origins serves to highlight the identified lack of Australian evidence, whilst reinforcing this review’s objective of pressure‐standardisation protocol in a national context. Although potentially indicative of publication bias, this could not be assessed as no accepted method exists for its evaluation of diagnostic test accuracy studies. 27 Furthermore, the selection of sources may display potential subjectivity as only two reviewers were involved in the process; reproducible criteria were adhered to in order to minimise this. Similarly, it is possible that data were missed due to keyword specifications and terminology issues.

Conclusion

The notion of mechanical standardisation techniques for breast compression is central to recent mammographic advances, with existing literature strongly supporting its implementation. Based on the current data available, several conclusions can be deduced. Key mammographic measures are obtained subjectively; the resultant disparity thus infers that optimal compression force values are not present in published research guidelines. Research supports an alternate standardised compression protocol founded on pressure; this approach accounts for individual breast characteristics in a ‘personalised’ manner. 7 , 8 As well as an objective measure, a compression pressure of approximately 10kPa was found to decrease pain, with a negligible effect on breast thickness, AGD and resultant image quality. 1 , 6 , 7 , 8 , 10 , 13 , 14 , 24 This also aids the reproducibility of image acquisition between and within women, whilst offering suggested guidelines for mammographers. 7 , 10 , 13 It is therefore suggested that patient compliance would increase in accordance with the perceived benefits of a standardised technique, which would ultimately aid in the detection of early‐stage breast cancer. 9 , 16 , 17 However, recent literature also presents several limitations which may ideally be considered before widespread clinical application. This is particularly relevant as current Australian data are lacking; hence, the generalisation of evidence‐based conclusions to a specific population requires additional research.

Conflict of Interest

The authors declare no conflict of interest.

J Med Radiat Sci 67(2020) 233–242

References

  • 1. Moshina N, Sebuødegård S, Hofvind S. Is breast compression associated with breast cancer detection and other early performance measures in a population‐based breast cancer screening program? Breast Cancer Res Treat 2017; 163: 605–13. [DOI] [PubMed] [Google Scholar]
  • 2. Verbeek AL. Mammographic screening: Keeping women alive. Womens Health (Lond). 2011; 7: 631–3. [DOI] [PubMed] [Google Scholar]
  • 3. Morrow M, Waters J, Morris E. MRI for breast cancer screening, diagnosis, and treatment. Lancet. 2011; 378: 1804–11. [DOI] [PubMed] [Google Scholar]
  • 4. Jensen A, Geller BM, Gard CC, et al. Performance of diagnostic mammography differs in the United States and Denmark. Int J Cancer 2010; 127: 1905–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5. de Groot JE, Hopman IGM, van Lier MGJTB, Branderhorst W, Grimbergen CA, den Heeten GJ. Pressure‐standardised mammography does not affect visibility, contrast and sharpness of stable lesions. Eur J Radiol 2017; 86: 289–95. [DOI] [PubMed] [Google Scholar]
  • 6. Holland K, Sechopoulos I, Mann RM, den Heeten GJ, van Gils CH, Karssemeijer N. Influence of breast compression pressure on the performance of population‐based mammography screening. Breast Cancer Res 2017; 19: 126. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. den Boer D, Dam‐Vervloet LAJ, Boomsma MF, de Boer E, van Dalen JA, Poot L. Clinical validation of a pressure‐standardized compression mammography system. Eur J Radiol 2018; 105: 251–4. [DOI] [PubMed] [Google Scholar]
  • 8. de Groot JE, Broeders MJM, Branderhorst W, den Heeten GJ, Grimbergen CA. A novel approach to mammographic breast compression: Improved standardization and reduced discomfort by controlling pressure instead of force. Med Phys. 2013; 40: 081901. [DOI] [PubMed] [Google Scholar]
  • 9. Poulos A, McLean D. The application of breast compression in mammography: a new perspective. Radiography 2004; 10: 131–7. [Google Scholar]
  • 10. De Groot JE, Branderhorst W, Grimbergen CA, et al. Mammographic compression: A need for mechanical standardisation In: A Holistic Approach, Digital Mammography. Springer, Berlin, Germany, 2015; 189–94. [Google Scholar]
  • 11. Perry N, Broeders M, de Wolf C, Törnberg S, Holland R. von Karsa L. European guidelines for quality assurance in breast cancer screening and diagnosis. Fourth edition—summary document. Ann Oncol 2008; 19: 614–22. [DOI] [PubMed] [Google Scholar]
  • 12. Holland K, Sechopoulos I, den Heeten G, Mann RM, Karssemeijer N. Performance of breast cancer screening depends on mammographic compression In: Tingberg A, Lång K. Timberg P. (eds). Breast Imaging. Springer International Publishing, Cham, 2016; pp. 183–9. [Google Scholar]
  • 13. Branderhorst W, de Groot JE, Highnam R, et al. Mammographic compression–a need for mechanical standardization. Eur J Radiol 2015; 84: 596–602. [DOI] [PubMed] [Google Scholar]
  • 14. de Groot JE, Branderhorst W, Grimbergen CA, den Heeten GJ, Broeders MJM. Towards personalized compression in mammography: a comparison study between pressure‐ and force‐standardization. Eur J Radiol. 2015; 84: 384–91. [DOI] [PubMed] [Google Scholar]
  • 15. Mercer CE, Hogg P, Szczepura K, Denton ER. Practitioner compression force variation in mammography: A 6‐year study. Radiography 2013; 19: 200–6. [Google Scholar]
  • 16. Mercer CE, Szczepura K, Kelly J, et al. A 6‐year study of mammographic compression force: Practitioner variability within and between screening sites. Radiography 2015; 21: 68–73. [Google Scholar]
  • 17. Mercer CE, Hogg P, Lawson R, Diffey J, Denton ERE. Practitioner compression force variability in mammography: A preliminary study. Br J Radiol 2013; 86: 20110596. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Arksey H, O'Malley L. Scoping studies: Towards a methodological framework. Int J Soc Res Methodol 2005; 8: 19–32. [Google Scholar]
  • 19. Arksey DLA, Tetzlaff J, Altman DG, The PRISMA Group . Preferred reporting items for systematic reviews and meta‐analyses: The PRISMA statement. PLoS Med 2009; 6(7): e1000097. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20. Branderhorst W, de Groot JE, van Lier MGJTB, Highnam RP, den Heeten GJ, Grimbergen CA. Technical note: Validation of two methods to determine contact area between breast and compression paddle in mammography. Med Phys 2017; 44: 4040–4. [DOI] [PubMed] [Google Scholar]
  • 21. de Groot JE, Broeders MJ, Branderhorst W, den Heeten GJ, Grimbergen CA. Mammographic compression after breast conserving therapy: controlling pressure instead of force. Med Phys. 2014; 41: 023501. [DOI] [PubMed] [Google Scholar]
  • 22. Moshina N, Sebuødegård S, Evensen KT, Hantho C, Iden KA, Hofvind S. Breast compression and experienced pain during mammography by use of three different compression paddles. Eur J Radiol 2019; 115: 59–65. [DOI] [PubMed] [Google Scholar]
  • 23. Jeukens CRLPN, van Dijk T, Berben C, Wildberger JE. Lobbes MBI. Evaluation of pressure‐controlled mammography compression paddles with respect to force‐controlled compression paddles in clinical practice. Eur Radiol 2019; 29: 2545–52. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Lau S, Aziz YFA, Ng KH. Mammographic compression in Asian women. PLoS ONE 2017; 12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Australia B. National Accreditation Standards. Australian Government Department of Health, Australia, 2015. [Google Scholar]
  • 26. Murphy F, Nightingale J, Hogg P, Robinson L, Seddon D, Mackay S. Compression force behaviours: An exploration of the beliefs and values influencing the application of breast compression during screening mammography. Radiography 2015; 21: 30–5. [Google Scholar]
  • 27. McInnes MDF, Bossuyt PMM. Pitfalls of Systematic Reviews and Meta‐Analyses in Imaging Research. Radiology 2015; 277: 13–21. [DOI] [PubMed] [Google Scholar]