Skip to main content
Data in Brief logoLink to Data in Brief
. 2020 Aug 29;32:106232. doi: 10.1016/j.dib.2020.106232

Data on biodistribution and dose calculation of 99mTechnetium -Dimercaptosuccinic acid in pediatric patients using a hybrid planar/single emission computed tomography method

Mahmoud Bagheri a,b,, Masoumeh DorriGiv c, Marjaneh Hejazi a,b, Mohammad Reza Fouladi a,b, Ali Asghar Parach d,⁎⁎
PMCID: PMC7479315  PMID: 32939384

Abstract

The Biodistribution and absorbed dose data from the administration of radiopharmaceuticals are necessary to analyze the risk-benefit of the procedure. It has particular significance in children, as their metabolism is very different from adults. 99mTc-DMSA scintigraphy is the golden standard imaging technique for the assessment of renal involvement in febrile urinary tract infection and renal sequels. However, 99mTc-DMSA biodistribution data for children are scarce and usually outdated which have been obtained by older methods. In this data article, we analysed the biodistribution of 99mTc-DMSA in 12 pediatric patients using planar/SPECT method. In addition, the radiation absorbed doses were calculated by MIRDOSE software.

Keywords: Radiation absorbed dose, Biodistribution, 99mTc-DMSA, Planar/SPECT method

Specifications Table

Subject Nuclear medicine, clinical research
Specific subject area Biodistribution analysis and absorbed dose calculation of 99mTc-DMSA (Technetium-99m-dimercaptosuccinic acid) in pediatric patients.
Type of data Tables, Figures
How data were acquired Direct collection of tissues from pediatric patients at different time-points using planar, SPECT (single emission computed tomography), and MRI (magnetic resonance imaging).
Data format Raw, Analyzed.
Parameters for data collection Each patient underwent 3 to 5 planar scans, and also single SPECT scan after 99mTc-DMSA injection with a dual-head gamma camera system (a parallel hole and LEHR [low energy high resolution] collimator). In addition, each patient imaged by MRI before injection.
Description of data collection All acquisition data were stored on the computer, including count-rates and measurement times. For all images, the count-rates were determined using suitable ROIs (region of interests), as well as a region surrounding each ROI was used for background correction. The cumulative activity and residence times for each source organ were calculated from count-rates with planar/SPECT method.
Data source location Shahid Sadoughi Hospital of Yazd, Iran
Data accessibility Raw and processed data are available with the article.

Value of the Data

  • These data present the biodistribution and absorbed dose of 99mTc-DMSA for children as sensitive to ionizing radiation.

  • Our data provide important information on the value of hybrid planar/SPECT and MRI techniques for biodistribution measurement and will be useful to calculate in absorbed dose more accurately.

  • The date can be used for children patients in renal scintigraphy with 99mTc-DMSA for the best/optimize time during the imaging process especially in busy nuclear medicine departments.

  • These data will be of interest to all those scientists who have access the biodistribution and absorbed dose data from the administration of radiopharmaceuticals which are necessary to analyze the risk-benefit of the procedure.

  • The data can be used to further improve the standardization of children's dosimetric assessments and recommendations for activity administration for future studies on risk prognostication in clinical practice.

1. Data Description

In previous data, for calculating the absorbed dose of 99mTc-DMSA in pediatric, a planar method was used [1] at a short time period acquisition after injection [2]. Furthermore, in the past researches, the lateral planar images have used to obtain the organ and patient body thicknesses for self-attenuation and background corrections, and also transmission factor [2]. However, we used MRI method to calculate these corrections which is more accurate than the planar method as well as without unnecessarily patient's absorbed dose compared to computed tomography (CT) images [3].

In this data article, the aim was to obtain biodistribution data with planar/SPECT method [3] from children at various ages and degrees of renal dysfunction after the administration of 99mTc-DMSA in order to look for evidence of age-dependency. Herein we have provided the biodistribution in different time periods ranging from 30 min to 19 h. In addition, the percentage of 99mTc-DMSA uptake in source organs and reminders are separately calculated for each patient.

2. Experimental design, materials and methods

2.1. Patient studies

Twelve pediatric patients including 4 males and 8 females, aged from 3 to 12 years old have participated in this data article. Informed consent was obtained from all participants after the procedures were fully explained and the study was approved by Shahid Sadoughi University of Medical Sciences (Yazd, Iran) with the registration number of “4137″. The patients had the genitourinary abnormalities problem. They were injected with 86–170 MBq with the mean value ± standard deviation of 116.7 ± 26.7, 99mTc-DMSA for acquisition Nuclear Medicine imaging. The patients’ demography, including height, weights, and ages has been shown in Table 1.

Table 1.

Demographic data for the patients and administered activity to each patient are also included.

Patient number Age (yr) Sex Weight (kg) Height (cm) Administered activity (MBq)
1 7 F 23 112 98
2 5 M 19 119 115
3 8 M 25 130 107
4 5 F 20 110 106
5 7 F 20 114 170
6 4 F 18 105 102
7 12 F 43 145 169
8 3 F 15 98 115
9 7 F 21 108 86
10 4 F 13 100 127
11 4 M 15 100 107
12 4 M 14 106 98

2.2. Imaging procedures

2.2.1. Planar and SPECT images

The injection activity measurements were obtained using a calibrated ‘dose calibrator’ (Capintec, Inc., Ramsey, New Jersey, USA). A dual-head gamma camera system (Philips ADAC, forte) with a parallel hole, LEHR (low energy high resolution) collimator, was used for recording the patients’ imaging. After the injection of 99mTc-DMSA, each patient underwent 3–5 planar scans (30 min-19 h), and also a single SPECT scan (2 h after injection). The time duration for each planar scan was approximately 300 s. The views of abdominal and pelvic regions including kidneys, bladder, liver, and spleen were acquisitioned so that the organs have the predominant uptake compared to the rest organs.

A triple energy window scatter correction method was used for both planar and SPECT scans. In this method, a 15% main energy window centered on the 99mTc photo-peak and two 7% windows positioned on each side of the emission photo-peak. In the planar method a matrix size of 256 × 256 (pixel size = 1.75 mm) was used. For the SPECT scans, the step-and-shoot mode was utilized to acquire 40 projections over 360° with a circular orbit. The time per SPECT projection was 30 s. The SPECT images were reconstructed to a 128 × 128 matrix (resolution = 4.75 × 4.75 mm2 and slice thickness = 4.75 mm).

2.2.2. MRI parameters

A Siemens Avanto MRI machine (Siemens Healthineers, Germany) with the magnetic field power of 1.5-T, was used for measuring the diameter of the patient's body and organ thicknesses for each patient (Fig. 1). The parameters of the MRI were set as repetition time (TR) =3.48 ms, echo time (TE) =1.39 ms, slice thickness 1.7 mm, and flip angle =10˚. The patient's body and organs thicknesses were measured by ITK-SNAP (version 3.6.0-RC1; http://www.itksnap.org), a free open-source segmentation software.

Fig. 2.

Fig. 2

A sample of picture from patient 12 shown the biodistrubution of 99mTc-DMSA based on time. Anterior planar images acquired at 2.1 h (a), 2.86 h (b), and 19 h (c) after injection.

Fig. 1.

Fig. 1

A sample of axial MRI image shown the diameter and thickness.

2.2.3. Calibration factor

The calibration experiment was performed to convert the measured SPECT and planar image count rates to absolute values. To evaluate the SPECT calibration factor, a point of 99mTc source (a small insulin syringe), 37 MBq, prepared and placed in air [2]. Then, the SPECT image acquired using the same parameters of the patients for the point source.

2.3. Activity quantification

The following equation was used to quantify the activity in each source organ A (j), MBq:

A(j)=R(j)K.T×f (1)

Where R(j) is the count rate in the drawn volume of interest, T represents the transmission factor across patient thickness and linear attenuation coefficient (0.15 cm−1) based on the MIRD (Medical Internal Radiation Dose) pamphlet No. 16 [4], f is the source organ self-absorption coefficient (f=[(μjdj/2)/sinh(μjdj/2)]) (μjand djare source organ attenuation coefficient and thickness, respectively) and K is gamma camera calibration factor (cps/MBq). The differences in tissue composition and density were not included in calculations and the mean effective attenuation coefficient was used for all body organs and tissues. These values are illustrated in Table 2. In this table, the raw data of self-attenuation correction factor and thickness values for source organs and whole body (in pelvic and abdomen), also the transmission factor for whole body measured for all patients.

Table 2.

Self-attenuation correction factor and thickness values for source organs and whole body (in pelvic and abdomen), also transmission factor for whole body measured for all patients.

Patient Transmission factor Self-attenuation correction factor Thickness (cm)
Whole body (in pelvic region) Whole body (in abdomen region) Kidney Liver Spleen Bladder Whole body (in pelvic region) Whole body (in abdomen region) Kidney Liver Spleen Bladder Whole body (in pelvic region) Whole body (in abdomen region)
1 0.192 0.153 0.986 0.972 0.992 0.991 0.895 0.867 3.8 5.5 3 3.1 11 12.5
2 0.165 0.142 0.986 0.968 0.992 0.990 0.877 0.857 3.9 5.9 3 3.2 12 13
3 0.192 0.138 0.984 0.972 0.995 0.992 0.895 0.854 4.2 5.5 2.2 2.9 11 13.2
4 0.237 0.149 0.988 0.975 0.992 0.992 0.919 0.863 3.6 5.2 2.9 3 9.6 12.7
5 0.165 0.156 0.990 0.977 0.993 0.989 0.877 0.869 3.3 5 2.8 3.5 12 12.4
6 0.237 0.195 0.990 0.977 0.996 0.989 0.919 0.897 3.3 5 2 3.5 9.6 10.9
7 0.091 0.067 0.982 0.942 0.992 0.985 0.795 0.750 4.4 8 3 4 16 18
8 0.223 0.149 0.985 0.977 0.993 0.990 0.912 0.863 4 5 2.8 3.2 10 12.7
9 0.173 0.160 0.989 0.976 0.994 0.988 0.882 0.873 3.4 5.1 2.7 3.6 11.7 12.2
10 0.237 0.173 0.989 0.967 0.993 0.980 0.919 0.882 3.4 6 2.5 4.7 9.6 11.7
11 0.259 0.201 0.984 0.977 0.996 0.992 0.928 0.900 4.2 5 2.8 3 9 10.7
12 0.237 0.186 0.990 0.981 0.992 0.994 0.919 0.891 3.2 4.5 2.1 2.5 9.6 11.2

2.4. Calculation of cumulative activity

2.4.1. Time activity curves

To calculate the cumulative activity for each source organ, the hybrid planar/SPECT approach was employed. For each planar image series of patients, the ROIs were drawn around the border of the organs in the first image. Then, these ROIs were registered in the rest of image series. It is notable that the spatial distribution of activity changes during the time has decreased with this method [5,6].

A series of planar images, count rates in each ROI, were plotted against time. Then, an appropriate exponential function fit for each time-count rate curve was obtained [[1], [2], [3], 6].

Background correction was used for estimating the count rates following the equation below [3]:

I=I(1(dj/D))IBG (2)

In this formula, the I and I′ are the background-corrected and uncorrected counts of each ROI, in that order. The dj and D are the diameter of organ and patient body thickness by MR image in the axial view, and IBG is the background counts. To obtain the IBG value, the mean count of pixels in the background region multiplied by the number of pixels in the source organs [3].  The geometric mean of counts, (IAIP)1/2, was used to obtain the time-count rates curve (IA= anterior counts, IP = posterior counts).

A sample of picture shown the biodistribution of 99mTc-DMSA based on time (after injection) in Fig. 1. Also, the biodistribution in different time periods ranging from 30 min to 19 h has shown in Table 3. According to this table, the pharmacokinetic behavior of 99mTc-DMSA uptake in whole-body, liver, bladder, and spleen was decreased immediately followed by a clearance phase, while, the kidneys had an opposed behavior compared to the above-mentioned organs with initial uptake phase to a maximum value. The curves were fitted with two-exponential and mono-exponential functions following their correlation coefficient values [3].

Table 3.

Count rates (count/second) of source organs at various time after 99mTc-DMSA.

Patient Time (h) Kidneys liver Spleen bladder Whole body
1 1.9 1104 99 13.5 153 2303
2.5 1206 48 7.1 159 2197
15 249 13 3 47 289
2 1.2 1123.6 205 94 45.3 2989
2.7 1356 167 92 56 2446
4.5 980 145 55.8 47 2100
16.5 229 43 10 4 450
3 2.71 712.4 50.2 40.3 15.7 1316
3.65 697.6 42.8 31.2 12.2 1249
4.15 677.9 25.5 15 13 1173
4 0.5 315.8 59.3 49.9 14.3 1517
2.53 709.4 51.2 42.3 15.7 1316
3.9 655.6 44.8 29.2 14.2 1249
7.15 355.9 15.5 10 9 1173
5 1.28 909 72 11.2 21 2300.5
2.10 936.5 73.8 8.8 16.3 2127.1
2.50 948.8 65 7.7 74.9 2138.1
3.16 966.1 59.1 8.1 26.9 1954.5
3.5 935.7 63.7 7.2 56 1992.6
6 1.13 572 126.6 18.3 134 1743
1.63 629 111.6 12 146 1683
2.71 633 65.7 9.8 57.1 1359
3.33 639 42 8.1 17.5 1238
3.7 619.6 50.3 7.3 21.4 1255
7 1.9 595.8 117 16.1 347.8 2015
2.50 600.7 69 13.9 380 1944
3.83 581 54.7 9.5 6.9 1254
4.9 518 53.8 8.8 5.3 1113
8 2.2 1263.7 195 95 14.3 2819
2.8 1312 160 88 35 2452
4.2 987 151 59.1 8.3 2051
17.3 219 33 9 3 436
9 1.3 435 59.9 12.8 68.9 1533
2.4 596.5 43.2 8.6 98.6 1254
16 153 9 3 19 168
10 1.88 1687 186.9 12.9 131.9 3306
2.43 2105.2 106.8 15.7 138.8 3424.2
3.00 1656.3 115.2 13.1 182.1 2927
7.08 1183.7 87.9 7.2 22.2 1892
11 1.33 1400 124 13.3 128 2989
2.50 1456 113 13 25 2278
5.30 1101 65 7.6 55 2010
12 2.1 1209 95 11.8 123 2203
2.86 1162 51 9.1 144 2152
19 244 12 1 8 295

2.4.2. Estimation of effective half lives

The effective half-lives (λeff) was obtained by the planar image acquisitions. In this method [6], λeff used estimates the cumulated activity (Ã) for each organ of interest:

A=ASPECT×eλefftSPECTλeff (3)

In this formula, ASPECT and tSPECT are the activity in each source region acquired from the SPECT image and the time of the acquisition, respectively. Actually each count rate in the planar image acquisitions rescaling by each SPECT image which provides an estimate of the time-cumulated (integrated) activities [6]. The cumulated activity was calculated for the kidneys, liver, spleen, and bladder, and for the remainder of the body it was obtained by subtracting the above-mentioned organs from the whole body activity. The percentage of 99mTc-DMSA uptake in source organs and the reminders are separately shown for each patient in Table 4. For obtained the percentage, the cumulative activity has calculated for source organs and whole body for each patient and then the cumulated activity of each source organ and remainder of the body divided in whole body cumulated activity.

Table 4.

The relative percentage of 99mTc-DMSA uptake calculated for each patient's organ. Also, the mean and standard deviation (SD) in source organs and the remainders are described.

Patient
Organs 1 2 3 4 5 6 7 8 9 10 11 12 mean SD
Kidneys 22.6 23.7 21.2 21.9 39.9 13.4 8.0 23.9 11.5 34.2 22.9 22.4 22.1 8.8
Liver 3.6 1.8 3.7 0.9 6.7 2.7 3.9 1.8 5.2 2.6 2.9 4.4 3.3 1.6
Spleen 0.7 0.9 3.1 0.7 1.0 0.6 0.8 0.9 1.0 0.2 0.5 0.6 0.9 0.7
UB contents 2.8 1.2 0.7 0.7 4.1 6.5 20.3 1.2 3.6 5.5 2.8 2.9 4.4 5.3
Remainder 70.3 72.4 71.3 75.8 48.3 76.8 67.0 72.2 78.7 57.5 70.9 69.7 69.3 8.5

The time integrated activities were normalized to the administered activity for calculating the residence time (Table 5). Post-processing of reconstructed planar and SPECT data was performed by ITK-SNAP software.

Table 5.

The residence time along with average (±SD) number of source organs and the remainder of the body (MBq × h/MBq).

Patient number Organ Residence Time (MBq.h/MBq)
Kidney Liver Spleen Urinary Bladder contents Remainder of the body
1 0.69 0.05 0.04 0.05 1.01
2 2.68 0.21 0.16 0.08 3.98
3 1.90 0.17 0.16 0.03 3.03
4 2.32 0.04 0.15 0.04 3.88
5 1.93 0.16 0.01 0.10 1.15
6 0.97 0.09 0.05 0.23 2.73
7 0.39 0.09 0.04 0.49 1.58
8 2.97 0.27 0.17 0.07 4.02
9 0.31 0.07 0.03 0.05 1.06
10 3.13 0.22 0.09 0.25 2.23
11 2.06 0.14 0.07 0.13 3.17
12 0.62 0.04 0.06 0.04 0.97
Mean ± SD 1.66 ± 1.02 0.13 ± 0.08 0.08 ± 0.06 0.13 ± 0.13 2.40 ± 1.22

2.5. Dosimetry

The organ absorbed dose and effective dose (equivalents) were estimated for various organs of the patients (mGy/MBq) using MIRDOSE 3.1 software (Oak Ridge Institute for Science and Education, Oak Ridge, TN 37,831) shown in Table 6. The input of MIRDOSE software was residence times in source organs including kidneys, liver, spleen, and remaining body calculated in 2.4.2 section.

Table 6.

The organ absorbed dose, effective dose (ED), and effective dose equivalents (EDE) per administered activity (mGy/MBq) for each patient using phantom based on the patient demography.

Patient (phantom used)
Organ dose 1 (5) 2 (5) 3 (10) 4 (5) 5 (5) 6 (5) 7 (15) 8 (5) 9 (5) 10 (5) 11 (5) 12 (5)
Adrenals 7.37E-03 2.88E-02 1.43E-02 2.52E-02 1.75E-02 1.25E-02 2.75E-03 3.13E-02 4.51E-03 2.92E-02 2.20E-02 6.82E-03
Gallbladder wall 4.91E-03 1.92E-02 9.32E-03 1.64E-02 1.04E-02 9.56E-03 2.36E-03 2.08E-02 3.74E-03 1.75E-02 1.47E-02 4.56E-03
Kidneys 8.15E-02 3.18E-01 1.58E-01 2.76E-01 2.27E-01 1.16E-01 2.39E-02 3.52E-01 3.76E-02 3.69E-01 2.44E-01 7.38E-02
Liver 4.14E-03 1.66E-02 9.00E-03 1.03E02 1.08E-02 7.44E-03 2.43E-03 1.92E-02 3.78E-03 1.68E-02 1.21E02 3.64E-03
Pancreas 5.44E-03 2.13E-02 1.12E-02 1.87E-02 1.13E-02 1.00E-02 2.48E-03 2.30E-02 3.84E-03 1.96E-02 1.59E-02 5.24E-03
Spleen 1.63E-02 6.46E-02 3.91E-02 5.93E-02 1.60E-02 2.25E-02 6.66E-03 6.93E-02 1.12E-02 4.66E-02 3.52E-02 2.13E-02
Urinary bladder wall 7.29E-03 1.56E-02 5.62E-03 1.07E-02 1.36E-02 2.99E-02 2.65E-02 1.47E-02 7.22E-03 3.24E-02 1.98E-02 6.09E-03
Gonads 2.99E-03 6.29E-03 2.88E-03 1.02E-02 4.95E-03 7.47E-03 2.69E-03 1.15E-02 2.66E-03 9.19E-03 5.35E-03 1.64E-03
ED 5.17E-03 1.93E-02 9.56E-03 1.68E-02 1.13E-02 1.06E-02 3.42E-03 2.07E-02 3.65E-03 1.97E-02 1.54E-02 4.75E-03
EDE 8.79E-03 3.36E-02 1.71E-02 2.97E-02 1.99E-02 1.55E-02 5.07E-03 3.65E-02 5.37E-03 3.48E-02 2.54E-02 8.39E-03

The biodistribution variation at 3 patients’ accrued different functions which introduce uncertainty in absorbed dose has been shown in Fig. 3.

Fig. 3.

Fig. 3

Anterior images for (a) patient 5, (b) patient 7, and (c) patient 11 at 2.50 h after injection illustrating differences in relative uptake of 99mTc DMSA in the kidneys, urinary bladder contents, and liver.

Declaration of Competing Interest

The Authors declare that there is not any competing of interest regarding this article.

Acknowledgments

This study was approved by Shahid Sadoughi University of Medical Sciences (Yazd, Iran) with the registration number of “4137″. The authors would like to thank the Nuclear Medicine Department at the Shahid Sadoughi Hospital for their sincere cooperation.

Footnotes

Supplementary material associated with this article can be found, in the online version, at doi:10.1016/j.dib.2020.106232.

Contributor Information

Mahmoud Bagheri, Email: Mahmoudbagheri68@gmail.com.

Ali Asghar Parach, Email: aliparach@ssu.ac.ir.

Appendix. Supplementary materials

mmc1.xml (312B, xml)

Reference

  • 1.Smith T., Evans K., Lythgoe M.F., Anderson P.J., Gordon I. Radiation dosimetry of technetium-99m-DMSA in children. J. Nucl. Med. 1996;37:1336–1342. [PubMed] [Google Scholar]
  • 2.Teles P., Costa J., Costa D., Matela N., Vaz P. Activity estimation and biokinetic analysis of 99mTc-DMSA in renal infant patients using a gamma camera. Phys. Med. 2018;52:9–17. doi: 10.1016/j.ejmp.2018.06.010. [DOI] [PubMed] [Google Scholar]
  • 3.Bagheri M., Parach A.A., Razavi-Ratki S.K., Nafisi-Moghadam R., Jelodari M.A. Patient-specific dosimetry for pediatric imaging of 99mTc-dimercaptosuccinic acid with gate Monte Carlo code. Radiat. Prot. Dosim. 2018;178:213–222. doi: 10.1093/rpd/ncx101. [DOI] [PubMed] [Google Scholar]
  • 4.Siegel J.A., Thomas S.R., Stubbs J.B., Stabin M.G., Hays M.T., Koral K.F., Robertson J.S., Howell R.W., Wessels B.W., Fisher D.R. MIRD pamphlet no. 16: techniques for quantitative radiopharmaceutical biodistribution data acquisition and analysis for use in human radiation dose estimates. J. Nucl. Med. 1999;40:37S–61S. [PubMed] [Google Scholar]
  • 5.Ebrahimnejad Gorji K., Abedi Firouzjah R., Khanzadeh F., Abdi-Goushbolagh N., Banaei A. Estimating the absorbed dose of organs in pediatric imaging of 99mTc-DTPATc-DTPA radiopharmaceutical using MIRDOSE software. J. Biomed. Phys. Eng. 2019;9:285. doi: 10.31661/jbpe.v0i0.984. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Grimes J., Celler A., Birkenfeld B., Shcherbinin S., Listewnik M.H., Piwowarska-Bilska H., Mikołajczak R., Zorga P. Patient-specific radiation dosimetry of 99mTc-HYNIC-Tyr3-octreotide in neuroendocrine tumors. J. Nucl. Med. 2011;52:1474–1481. doi: 10.2967/jnumed.111.088203. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

mmc1.xml (312B, xml)

Articles from Data in Brief are provided here courtesy of Elsevier

RESOURCES