. 2020 Sep 8;10:14781. doi: 10.1038/s41598-020-71757-6

Asymmetry analysis of optical coherence tomography angiography macular perfusion density measurements in preperimetric and perimetric glaucoma

Pei-Yao Chang 1,2,3,4,, Jiun-Yi Wang 5,6, Jia-Kang Wang 1,2,3,4, Shih-Cheng Yeh 7, Shu-Wen Chang 1,2
PMCID: PMC7479593  PMID: 32901074

Abstract

Macular retinal layer thickness asymmetry indices, particularly for the ganglion cell layer, are promising early indicators of glaucomatous damage. We evaluated macular perfusion density asymmetry (MPDA) among normal, preperimetric glaucoma (PPG), and perimetric glaucoma (PG) eyes, and we tested the performance of MPDA in differentiating between control and glaucoma eyes with or without visual field (VF) defects. In this study, 116 eyes (39 normal, 27 PPG, and 50 PG eyes) with optical coherence tomography angiography images of the macula were analysed. No significant difference was found in outer and inner MPDA between the control and PPG groups. However, outer MPDA was significantly higher in the PG group than in the PPG group (p = 0.009). Asymmetry of perfusion density and structural parameters was compared; no significant difference was found between controls and glaucoma patients. Outer MPDA had significantly higher discrimination ability between PPG and PG than did macular ganglion cell layer–inner plexiform layer thickness asymmetry (p = 0.039). In conclusion, the discriminant capability of MPDA for discriminating between glaucoma patients with and without VF defects is significantly higher than that of structural asymmetry. MPDA may be helpful in monitoring glaucoma progression in clinical practice.

Subject terms: Biomarkers, Diseases

Introduction

Evaluation of the circumpapillary retinal nerve fibre layer (cpRNFL) is useful for glaucoma diagnosis; however, more than 50% of retinal ganglion cell somas are located within the macula. Increasing evidence implicates early macular involvement in glaucomatous damage13, and it has been highlighted that cpRNFL analysis alone is insufficient for detecting macular damage4. Kim et al.2 reported that macular ganglion cell layer–inner plexiform layer (mGCIPL) changes are detected even before corresponding cpRNFL changes in the optical coherence tomography (OCT) deviation map.

Optical coherence tomography angiography (OCTA) is a newly developed noninvasive imaging modality that provides for qualitative and quantitative evaluation of the microvasculature of different retinal layers. Researchers have used OCTA to quantify the density of perfused vessels in peripapillary and macular regions in glaucoma patients, and increasing evidence has shown strong correlations between vessel density measurements and cpRNFL defects, visual field (VF) deficits, and glaucoma severity58. Studies have even revealed that OCTA vascular density parameters have stronger correlations with VF defects than do OCT structural parameters7,8.

The thicknesses of the cpRNFL, mGCIPL, and ganglion cell complex (GCC) in normal eyes are highly symmetrical in the upper and lower retinal hemispheres9,10. The glaucoma hemifield test (GHT) of the VF, which measures disparity in retinal sensitivity between the superior and inferior hemifields, is regarded as a useful tool in glaucoma assessment11,12. Macular retinal layer thickness asymmetry indices, particularly for the ganglion cell layer, are promising early indicators of glaucomatous retinal damage10,13. However, limited information is available on differences in superficial macula microvasculature damage between superior and inferior hemifields in glaucoma patients. It is desirable to have asymmetry indices in OCTA measurement for the early detection of glaucomatous change. Therefore, this study evaluated differences in hemifield macula vascular density among normal, preperimetric glaucoma (PPG), and perimetric glaucoma (PG) groups and tested its performance in distinguishing between controls and glaucoma patients with or without VF defects.

Methods

This retrospective cross-sectional study was conducted between January and December 2018. Patients with primary open-angle glaucoma (POAG) and healthy individuals (control group) who visited the glaucoma clinic of Far Eastern Memorial Hospital were recruited. The study protocol adhered to the tenets of the Declaration of Helsinki and was approved by the Institutional Review Board and Ethics Committee of Far Eastern Memorial Hospital, which waived the need for informed consent, as this study was based on retrospective data.

The inclusion criteria for all groups were as follows: age more than 20 years, best-corrected visual acuity ≥ 20/30, and refractive errors >  − 6 diopters. Patients with systemic hypertension and diabetes mellitus were included unless they had hypertensive or diabetic retinopathy. The exclusion criteria for all groups were as follows: presence of any other retinal or optic nerve disease, previous ocular surgery, or presence of any media opacities that prevented high-quality OCT scans. Diagnostic criteria for glaucoma were a normal anterior segment in slit-lamp examination, an open angle in gonioscopy, presence of glaucomatous optic nerve head changes (i.e., neuroretinal rim thinning, notching, cupping, and optic disc haemorrhage), or an RNFL defect in red-free fundus photography6. Eyes with glaucomatous VF defects were defined as those with a GHT result outside normal limits or a pattern standard deviation outside 95% of normal limits. Additionally, a cluster of three points with probabilities of 5% on the pattern deviation map in at least one hemifield, including at least one point with a probability of 1%, or a cluster of two points with a probability of 1% was needed10. To include preperimetric glaucoma eyes, VF defect was not considered as a diagnostic criterion in the present study.

Each subject underwent complete ophthalmic examination that included the following assessments: slit lamp biomicroscopy, intraocular pressure (IOP) measurement through noncontact tonometry, refractive error measurement through autorefraction (Auto Refractometer AR-610; Nidek Co, Ltd., Tokyo, Japan), gonioscopy, and dilated fundus examination with simultaneous stereophotography of the optic disc and red-free RNFL photography. VF testing was performed using a Humphrey Field Analyzer (SITA full threshold programs 30-2; Carl Zeiss Meditec, Inc., Dublin, CA, USA).

Peripapillary (Optic Disc Cube 200 × 200 protocol) and macular (Macular Cube 512 × 128 protocol) scans (collectively referred to as ganglion cell analysis) were acquired using the Cirrus 5000 HD-OCT (Carl Zeiss Meditec, Inc.). Software released by the manufacturer was used to calculate RNFL and GCIPL thicknesses, as previously described14. OCTA imaging of the macula was performed using the Cirrus HD-OCT (version 10.0.0.14618). The procedure for OCTA imaging using the Cirrus HD-OCT has been detailed previously15. Angiographic images were generated through OCT-based microangiography (OMAG), and the macula was imaged using a 6 × 6 mm2 scan pattern. Angiometric software of the Cirrus HD-OCT automatically calculates two parameters from the superficial retinal layer slab: vessel length density (defined as the total length of perfused vasculature per unit area in the region of measurement) and perfusion density (defined as the total area of perfused vasculature per unit area in the region of measurement). This software calculates the vessel length and perfusion density parameters across four inner and four outer sectors of the Early Treatment Diabetic Retinopathy Study (ETDRS) grid over the macula (Fig. 1). Image quality was assessed for all OCTA and OCT scans. Poor quality images, which were defined as those with a signal strength of < 7, poor centration, or motion artefacts and segmentation errors, were excluded from analysis. Outer macular perfusion density asymmetry (MPDA) was defined as the absolute value of the difference between outer superior and outer inferior parafoveal perfusion density, and the same definition was applied to inner MPDA. Outer macular vessel length density asymmetry (MVDA) was defined as the absolute value of the difference between outer superior and outer inferior parafoveal vessel length density, and the same definition was applied to inner MPDA. For mGCIPL asymmetry, it was defined as the absolute value of the difference between superior and inferior mGCIPL thicknesses; average mGCIPL asymmetry was defined as the absolute value of the difference between the average superior (i.e., superior + superotemporal + superonasal mGCIPL thicknesses/3) and average inferior (i.e., inferior + inferotemporal + inferornasal mGCIPL thicknesses/3) mGCIPL thicknesses. For cpRNFL asymmetry, it was defined as the absolute value of the difference between superior and inferior cpRNFL thicknesses.

Figure 1.

Figure 1

Inferior disc excavation with inferior nerve fibre layer wedge defect of right eye of a 52-year-old woman with POAG (a). Visual field results revealed defects at the corresponding superonasal site (b). OCT angiography image (Cirrus HD-OCT , version 10.0.0.14618) of a 6 × 6 mm scan exhibited the superficial vessels at the macula. The grids represented the sectors across which the vessel length densities (/mm) (c) were calculated. The area between the inner two circles in the 6 × 6 mm scan represented the inner sectors, and the area between the outer two circles represented the outer sectors. Combined RNFL and GCIPL deviation maps indicated structural glaucomatous damage (d). OCT scan showed the GCIPL thickness (μm) provided by the ganglion cell analysis report (e, left) and RNFL thickness (μm) (e, right) in each sector. GCIPL ganglion cell-inner plexiform layer; OCT optical coherence tomography. I inferior; IN inferonasal; IT inferotemporal; N nasal; S superior; SN superonasal; ST superotemporal; T temporal.

Statistical analysis

Descriptive statistics included mean and standard deviation for normally distributed variables and median and interquartile range for nonnormally distributed variables. The Shapiro–Wilk test was used to examine the normality distribution of continuous variables. If both eyes of a patient met the inclusion criteria, one eye was randomly selected. Depending on data distribution, ANOVA or the Kruskal–Wallis test with a post hoc test was conducted to compare baseline characteristics. The area under the receiver operating characteristic curve (AUC) was used to evaluate the diagnostic ability of asymmetry of OMAG and structural thickness parameters between the groups. A comparison of two AUCs was conducted in the package of pROC in R-language with the bootstrap method16. For each parameter of the direct measurement and asymmetry of OMAG and structural thickness, three post hoc tests or pairwise comparisons were conducted, and the significance level was set at 0.017 according to Bonferroni correction17. For other tests, a p value of < 0.05 was considered statistically significant. Due to the limited sample size available in the hospital, G-Power 3.1 was used to estimate the power for critical parameters given the observed sample size. Powers of 99.7%, 99.5%, and 61.4% were estimated for outer MVDA, outer MPDA, and average mGCLPL thickness asymmetry, given the sample sizes were 27 and 50 for the PPG and PG groups, respectively, at the significance level of 0.017, and the estimated effect sizes based on the observed data.

Results

In total, 151 individuals (50 normal and 101 POAG patients) underwent OCTA measurements of the macula. Among them, 30 eyes (19.9%) with motion artefacts in OMAG scans and 4 eyes (3.31%) with media opacity such as vitreous floaters were excluded. Finally, 116 individuals (116 eyes), comprising 39 healthy subjects, 27 PPG patients, and 50 PG patients, were included in the analysis. The demographic and ophthalmic characteristics of the three groups are summarised in Table 1. All participants in the glaucoma group were treated with at least one type of ocular antihypertensive medication at the time of OCTA imaging.

Table 1.

Baseline characteristics of control and glaucoma groups.

Median (25% Quartile,75% Quartile) P Post hoc Test
Control (n = 39) PPG (n = 27) PG (n = 50) Control vs PPG PPG vs PG Control vs PG
Age (y) 51.0 ± 10.7 49.1 ± 12.3 56.5 ± 9.4 0.007 0.770 0.016 0.057
Gender (male:female) 21:18 11:16 36:14 0.02 0.30 0.007 0.08
Eye (OD:OS) 20:19 16:11 16:34 0.05
Diabetes (yes:no) 3:36 3:24 4:46 0.87
Hypertension (yes:no) 5:34 4:23 6:44 0.94
Spherical equivalence (D) − 1.00 (− 2.50 to 0.25) − 4.00 (− 5.00 to − 1.50) − 2.25 (− 4.38 to 0.00) 0.008 0.006 0.28 0.25
Visual field index (%) 99.0 (97.8–100.0) 99.0 (98.0–99.0) 87.5 (75.0–94.3)  < 0.001 1.00  < 0.001  < 0.001
Mean deviation (dB) − 0.1 (− 2.3 to 0.6) − 0.5 (− 2.2 to 0.0) − 4.9 (− 8.1 to − 2.6)  < 0.001 1.00  < 0.001  < 0.001
Optic disc area (mm2) 2.09 (1.81–2.27) 1.84 (1.72–2.15) 1.93 (1.56–2.22) 0.13
Rim area (mm) 1.17 (1.11–1.31) 0.97 (0.83–1.13) 0.78 (0.67–0.93)  < 0.001 0.003 0.001  < 0.001
Average C/D ratio 0.65 (0.54–0.71) 0.67 (0.58–0.75) 0.76 (0.69–0.82)  < 0.001 0.26 0.002  < 0.001
IOP at the scanning visit (mm Hg) 17.0 (15.7–20.7) 15.0 (13.8–18.3) 13.7 (12.4–15.6)  < 0.001 0.31 0.03  < 0.001

PPG preperimetric glaucoma; PG perimetric glaucoma; IOP intraocular pressure.

No significant differences were observed in the vessel length or perfusion density in the macular area between the control and PPG groups, regardless of the sector. However, mGCIPL thickness in every sector was significantly lower in the PPG group than in the control group (all p < 0.017). In addition, cpRNFL in the superior and inferior sectors was significantly lower in the PPG group than in the control group (both p < 0.001; Table 2).

Table 2.

Optical coherence tomography angiography vessel density parameters and structural measurements of participants.

Median (25% Quartile, 75% Quartile) Pairwise comparison
Control (n = 39) PPG (n = 27) PG (n = 50) Control vs PPG PPG vs PG Control vs PG
Signal strength of 6 × 6 mm OMAG
SS (macular scan) 9 (8–10) 9 (8–10) 8 (7–9) 0.759 0.920 0.176
Macular vessel length density (/mm)
Outer superior 18.2 (16.6–19.1) 17.7 (16.4–18.8) 15.3 (13.5–17.7) 0.897 0.124 0.009
Outer inferior 18.1 (16.0–19.2) 16.8 (15.7–17.9) 12.8 (10.1–16.1) 0.093 0.031  < 0.001
Outer temporal 17.2 (14.4–18.3) 15.9 (12.8–17.8) 13.9 (10.9–16.0) 0.885 0.251 0.004
Outer nasal 19.7 (18.7–20.2) 19.8 (18.9–20.4) 18.6 (16.3–19.8) 0.230 0.888 0.184
Inner superior 18.0 (15.1–19.0) 16.8 (15.1–18.8) 16.6 (12.6–17.8) 0.785 0.870 0.246
Inner inferior 18.1 (15.8–19.2) 17.5 (14.4–19.0) 15.4 (12.1–17.2) 0.369 0.559 0.036
Inner temporal 17.4 (15.6–18.8) 17.5 (14.1–19.1) 15.3 (12.7–17.9) 0.531 0.147 0.013
Inner nasal 18.1 (15.3–19.0) 16.3 (14.9–19.4) 17.2 (14.1–18.8) 0.187 0.235 0.710
Inner mean 18.1 (15.6–19.1) 16.4 (15.3–19.1) 16.2 (13.0–17.4) 0.374 0.627 0.029
Outer mean 18.1 (16.7–19.0) 17.7 (16.0–18.4) 15.6 (12.9–17.1) 0.459 0.304 0.006
Full mean 17.5 (16.3–18.8) 17.2 (15.4–18.5) 15.5 (12.8–16.8) 0.217 0.348 0.081
Macular perfusion density
Outer superior 0.45 (0.41–0.47) 0.44 (0.40–0.46) 0.38 (0.34–0.43) 0.681 0.180 0.005
Outer inferior 0.44 (0.39–0.47) 0.42 (0.38–0.44) 0.32 (0.25–0.41) 0.230 0.119  < 0.001
Outer temporal 0.42 (0.33–0.45) 0.38 (0.30–0.44) 0.34 (0.26–0.39) 0.981 0.211 0.005
Outer nasal 0.48 (0.45–0.49) 0.49 (0.47–0.50) 0.45 (0.39–0.49) 0.229 0.763 0.071
Inner superior 0.42 (0.36–0.46) 0.39 (0.35–0.47) 0.39 (0.29–0.44) 0.657 0.492 0.286
Inner inferior 0.43 (0.37–0.46) 0.40 (0.35–0.46) 0.37 (0.28–0.42) 0.561 0.950 0.111
Inner temporal 0.41 (0.37–0.44) 0.39 (0.32–0.45) 0.36 (0.29–0.43) 0.917 0.165 0.144
Inner nasal 0.43 (0.35–0.44) 0.38 (0.35–0.45) 0.41 (0.32–0.45) 0.077 0.148 0.788
Inner mean 0.42 (0.37–0.45) 0.38 (0.35–0.46) 0.38 (0.31–0.42) 0.327 0.609 0.096
Outer mean 0.44 (0.41–0.47) 0.43 (0.39–0.45) 0.38 (0.30–0.42) 0.620 0.431 0.017
Full mean 0.44 (0.40–0.46) 0.41 (0.37–0.45) 0.38 (0.31–0.41) 0.431 0.417 0.076
mGCIPL (μm)*
Superior thickness 85.1 ± 6.3 77.6 ± 4.8 70.1 ± 9.9  < 0.001 0.006  < 0.001
Inferior thickness 81.6 ± 4.8 71.7 ± 5.7 64.0 ± 10.3  < 0.001 0.004  < 0.001
Superotemporal thickness 83.4 ± 5.7 74.4 ± 5.2 66.1 ± 12.0  < 0.001 0.008  < 0.001
Inferotemporal thickness 83.5 ± 5.0 73.3 ± 6.1 61.4 ± 12.6  < 0.001  < 0.001  < 0.001
Superonasal thickness 86.1 ± 6.8 79.6 ± 7.0 75.1 ± 10.6 0.001 0.068  < 0.001
Inferonasal thickness 84.3 ± 5.5 77.1 ± 5.3 70.8 ± 10.7  < 0.001 0.012  < 0.001
Average superior thickness 84.9 ± 5.9 77.2 ± 4.8 70.4 ± 9.9  < 0.001 0.008  < 0.001
Average inferior thickness 83.1 ± 4.6 74.0 ± 4.5 65.4 ± 10.5  < 0.001 0.001  < 0.001
Average mGCIPL thickness 84.0 ± 5.2 75.6 ± 3.9 67.9 ± 9.5  < 0.001 0.001  < 0.001
cpRNFL (μm)*
Superior thickness 122.1 ± 12.3 101.7 ± 16.7 90.9 ± 17.8  < 0.001 0.032  < 0.001
Inferior thickness 126.1 ± 12.1 99.7 ± 18.0 78.4 ± 22.1  < 0.001 0.001  < 0.001
Temporal thickness 72.9 ± 13.9 71.0 ± 10.5 60.8 ± 13.7 0.052 0.064  < 0.001
Nasal thickness 71.4 ± 10.6 68.6 ± 9.2 65.0 ± 8.4 0.640 0.033 0.003
Average cpRNFL thickness 99.0 ± 7.6 85.3 ± 6.4 73.8 ± 10.3  < 0.001 0.000  < 0.001

All values represent median with interquartile range in parentheses unless specified * (mean ± SD).

Age, gender, and spherical equivalence were adjusted in pairwise comparisons.

PPG preperimetric glaucoma; PG perimetric glaucoma; mGCIPL macular ganglion cell-inner plexiform layer; cpRNFL circumpapillary retinal nerve fibre layer asymmetry; OMAG optical microangiography.

Table 3 summarises the asymmetry of OMAG parameters and structural thickness parameters in glaucoma patients and controls. In structural analysis, cpRNFL thickness asymmetry was significantly higher in the PPG group than in the control group (p = 0.007). However, outer and inner MVDA exhibited no significant difference between the control and PPG groups. However, outer MPDA was significantly higher in the PG group than in the control and PPG groups (both p < 0.01).

Table 3.

Asymmetry of OMAG parameters and structural thickness parameters in glaucoma and control groups.

Median (25% quartile, 75% quartile) Pairwise comparison
Control (n = 39) PPG (n = 27) PG (n = 50) Control vs PPG PPG vs PG Control vs PG
Macular outer vessel length density asymmetry (/mm) 0.7 (0.2–1.4) 1.0 (0.5–1.7) 2.7 (1.7–4.6) 0.580 0.051  < 0.001
Macular inner vessel length density asymmetry (/mm) 0.7 (0.2–1.5) 0.6 (0.4–2.0) 1.5 (0.5–2.5) 0.743 0.528 0.199
Macular outer perfusion density asymmetry 0.02 (0.01–0.03) 0.02 (0.01–0.04) 0.08 (0.05–0.11) 0.582 0.009  < 0.001
Macular inner perfusion density asymmetry 0.02 (0.01–0.04) 0.02 (0.01–0.06) 0.03 (0.01–0.06) 0.865 0.355 0.235
mGCIPL thickness asymmetry (μm) 5.0 (3.0–6.0) 5.0 (3.0–9.0) 8.5 (2.0–14.0) 0.814 0.054 0.004
Average mGCIPL thickness asymmetry (μm) 2.7 (1.7–3.7) 4.0 (1.0–6.0) 7.5 (2.5–11.0) 0.959 0.018 0.001
cpRNFL thickness asymmetry (μm) 7.0(4.0–11.8) 19.0(7.0–33.0) 20.5(11.0–33.3) 0.007 0.366 0.002

All values represent median with interquartile range in parentheses.

Age, gender, and spherical equivalence were adjusted in pairwise comparisons.

PPG preperimetric glaucoma; PG perimetric glaucoma; mGCIPL macular ganglion cell-inner plexiform layer; cpRNFL circumpapillary retinal nerve fibre layer asymmetry.

Figure 2 demonstrated the diagnostic accuracy of vessel density and structural thickness asymmetry for differentiating between the following: (1) control and PPG, (2)PPG and PG, and (3) control and PG groups. For discriminating between healthy and PPG eyes, the highest AUC was obtained for cpRNFL thickness asymmetry (AUC = 0.81). Outer MPDA had the highest AUC for discriminating between PPG and PG eyes (AUC = 0.86), and outer MVDA had the highest AUC for discriminating between healthy and PG eyes (AUC = 0.88). AUC for asymmetry of all parameters decreased progressively with decreasing severity of glaucoma. We compared differences in AUC for vessel density asymmetry parameters and thickness asymmetry parameters between the groups (Table 4). No significant difference was found in the AUC of the asymmetry index of vessel density parameters and thickness parameters in terms of differentiating healthy eyes from those with glaucoma, regardless of whether they had PPG or PG. However, the AUC of outer MPDA was significantly higher than the AUC of average mGCIPL thickness asymmetry (p = 0.039) for differentiating between the PPG and PG groups.

Figure 2.

Figure 2

ROC curves of asymmetry of OMAG parameters and structural thickness parameters for differentiating between the groups.

Table 4.

Comparisons of the diagnostic ability of asymmetry of OMAG parameters and structural thickness parameters for differentiating between the groups.

P-value
Control vs PPG PPG vs PG Control vs PG
Macular outer vessel length density asymmetry vs Average mGCIPL thickness asymmetry 0.982 0.077 0.202
Macular outer vessel length density asymmetry vs cpRNFL thickness asymmetry 0.546 0.118 0.266
Macular outer perfusion density asymmetry vs verage mGCIPL thickness asymmetry 0.685 0.039 0.369
Macular outer perfusion density asymmetry vs cpRNFL thickness asymmetry 0.415 0.060 0.384

The p value: comparisons of AUC of the vessel density asymmetry parameters and structural thickness asymmetry parameters.

AUC area under the receiver operation characteristic curve, adjusted for age, gender, and spherical equivalence.

Discussion

Through a simple method for defining asymmetry in this study, we found that OMAG macular asymmetry parameters were comparable with structural asymmetry parameters (mGCIPL) in differentiating either PPG or PG patients from controls. However, we found the discriminant capability of outer MPDA for differentiating between PPG and PG to be significantly better than that of average mGCIPL asymmetry. Macular vessel density has been shown to be reduced in glaucoma eyes in comparison with healthy eyes5,18. Rao et al. reported significantly lower diagnostic abilities for macular vessel density parameters in POAG than for peripapillary vessel density19. Most studies have concluded that the diagnostic ability of macular vessel density parameters in OCTA is moderate18,20,21. In support of such findings, our results revealed no significant difference in the macular vessel length or perfusion density between the control and PPG groups, regardless of the sector; however, the macular vessel length or perfusion density was only significantly lower in the PG group than in the control group. Previous studies have concluded that VF MD has a stronger correlation with OCTA parameters than with OCT parameters7,8. In this study, the superiority of outer MPDA to the asymmetry index of structural thickness in distinguishing PG from PPG may relate to the stronger correlation between VF and OCTA parameters.

Glaucomatous damage usually occurs in only one horizontal hemifield; therefore, the GHT of the VF, which measures disparity in retinal sensitivity between superior and inferior hemifield, is regarded as very useful in early glaucoma assessment11,12. However, the VF test is still a subjective and time-consuming test. We found that compared with structural asymmetry, outer MPDA had significantly higher diagnostic ability in PG than in PPG, whether through cpRNFL or mGCIPL measurement. To the best of our knowledge, no study has explored intraeye asymmetry of macular retinal vessel density between the upper and lower hemispheres. Although intraeye differences in macular retinal vessel density were not sensitive enough to identify very early glaucoma, such as PPG, in this study, they were able to differentiate PPG from PG. When PPG progresses to PG, only longitudinal follow-up with a VF test can confirm the diagnosis; thus, supplementary parameters such as vessel densities should be used in monitoring PPG. A longitudinal study found progressive macular vessel density loss in glaucoma eyes through serial OCTA measurements22. OCTA has even been reported to be useful in detecting changes in retinal microvasculature before VF damage becomes detectable in POAG patients23. Future studies with a larger sample size are needed to elucidate the clinical implications of macular vessel density asymmetry parameters measured using OCTA.

Measuring the inner retinal layers in the macular region is another strategy for the early diagnosis of glaucoma. No definite conclusion has been reached regarding whether the inner macular layer or cpRNFL is superior in differentiating glaucoma eyes from normal eyes. Mwanza et al. reported that average mGCIPL was comparable to average cpRNFL24, whereas Mahdavi et al. reported that average cpRNFL measurements were superior to global GCIPL thickness values for the detection of early glaucoma25. The AUC value from GCC thickness analysis was reported to range from 0.86 to 0.955,6 for discriminating between normal and glaucoma eyes, and the range decreased to 0.6–0.83,26 when the analysis was restricted to glaucoma eyes such as PPG eyes. The discriminant capability would not be the same in differentiating different stages of glaucoma. In our study, direct measurement of cpRNFL or mGCIPL thickness demonstrated significant differences between the control and PPG groups. However, cpRNFL thickness asymmetry was significantly higher in the PPG group than in the control group (p = 0.007), whereas mGCIPL thickness asymmetry showed no significant difference between the two groups. This might be because the definition of glaucoma was based on optic nerve appearance, which is in favour of peripapillary parameters. Studies have yielded inconsistent results regarding whether measurement of macular vessel density outperforms structural measurement of inner macular thickness in the diagnosis of glaucoma5,6,20. Leung et al.27 standardised the retinal layers for segmentation and the region of interest for analysis to enable a fair comparison between OCTA and OCT measurements. They concluded that the OCTA of the macula may have a limited role in the diagnostic evaluation of glaucoma. In this study, no significant difference was found in the macular vessel length or perfusion density between the control and PPG groups; however, mGCIPL thickness was significantly lower in the PPG group than in the control group. Our results are consistent with Leung’s finding that OCT structural measurement of inner macular thickness exhibited higher diagnostic performance in detecting glaucoma than did OCTA measurement of inner macular vessel density.

Macular hemifield differences in full retinal or inner retinal layer thickness have been reported to help in in early glaucoma diagnosis10,13,28,29. Posterior pole asymmetry analysis nstalled in the Spectralis SD-OCT (Heidelberg Engineering, Heidelberg, Germany) used a retinal full thickness value of 64 cells obtained from a macular area equivalent to a central 20° VF30. Um et al.10 and Seo et al.29 have used the Spectralis OCT with different approaches and attained similar results; specifically, that detecting asymmetry in macular hemifield full thickness may be a useful diagnostic aid in the detection of early glaucoma. Regarding inner retinal layers, Yamada et al.13 defined the asymmetry index based on a logarithmic ratio of upper to lower thickness and found that the ganglion cell layer asymmetry index for PPG (AUC = 0.81 for GCL; 0.79 for GCC) is a promising early indicator. In the present study, mGCIPL asymmetry was simply defined based on the ganglion cell analysis report provided by the Cirrus OCT, which measured an elliptical annulus centred at the fovea with outer vertical and horizontal axes of 4.0 and 4.8 mm, respectively. Although our measured area was smaller than that in the Spectralis or Optovue OCT machines, the AUC of average mGCIPL asymmetry for PPG (AUC = 0.76) and PG (AUC = 0.84) was still obtained. Furthermore, the commercially available analysis report by the Cirrus machine automatically divides the measured area into six parts. The superior or inferior mGCIPL that we used for asymmetry analysis was only related to the central part of the upper or lower hemisphere. In this study, by definition, average mGCIPL asymmetry covered larger areas than mGCIPL asymmetry; therefore, we used average mGCIPL thickness asymmetry instead of mGCIPL in the pairwise comparison. However, localised thinning of mGCIPL in PPG eyes might not have a marked effect on hemifield asymmetry as compared with control eyes. When mGCIPL progresses to loss in PG, this might substantially affect the mGCIPL asymmetry, which could explain why we found only mGCIPL thickness asymmetry to be significantly different between the control and PG groups but not between the control and PPG groups. This study has some limitations. First, because of its retrospective design, we could not investigate the blood pressure of patients or their antihypertensive medications. Most glaucoma patients were using IOP-lowering medications, and the effect of different IOP-lowering medications on OCT angiography imaging was unclear. Second, version 10.0 of our software did not provide for analysis of peripapillary vessel density. Additional details of optic disc vessel density would provide a comprehensive understanding of vascular asymmetry in glaucoma. Moreover, the ETDRS grid, which was used for OMAG macular sectoral measurements, was primarily developed for diabetic retinopathy evaluation rather than glaucoma evaluation. The GCIPL sectors were developed for the evaluation of structural changes in the macula in glaucoma. This may have biased the results in favour of the GCIPL measurements but our finding was that MPDA had a stronger ability to differentiate between PPG and PG comparing to GCIPL. Furthermore, the sector definitions of OMAG measurements used in the present study were the ones automatically demarcated by the software; hence, the asymmetry results would be applicable to all clinical settings when using this commercial software21. Third, we included patients with only mild to moderate VF defects in the PG group. Generalised loss, whether in nerve thickness or vessel density in both hemifields, usually occurs during glaucoma progression. Our findings might not apply to all glaucoma patients, especially not to those with severe VF defects. Fourth, we only excluded patients with diabetic retinopathy, and eight diabetic patients (three in the control group, three in the PPG group, and four in the PG group) did not have diabetic retinopathy in our study. A previous study reported that the density of vascular perfusion was reduced with increasing severity of diabetic retinopathy, even in non-diabetic retinopathy (NDR)31; however, Carnevali et al. investigated both superficial and deep macular vascular perfusion and reported only the deep capillary plexus in NDR was reduced32. Although only the superficial macular vascular perfusion was investigated in this study, the subclinical macular microvascular change in NDR might still be a confounding factor.

In conclusion, this study demonstrated that macular perfusion density asymmetry is not an aid in early glaucoma diagnosis for differentiating between healthy and PPG eyes; however, the discrimination ability of macular perfusion density asymmetry was significantly higher than structural asymmetry for differentiating between glaucoma patients with or without VF defects. Macular OCT angiography imaging could be included in the imaging algorithm for serial observation of patients with glaucoma. Longitudinal studies are needed to determine whether OCTA measurement can improve the detection or prediction of glaucoma progression and to elucidate the role of OCTA and VF in long-term glaucoma follow-up.

Acknowledgements

The authors had research funding (FEMH- 2019-C-032) from Far Eastern Memorial Hospital for article processing charges.

Author contributions

P.Y.C wrote the main manuscript text; J.Y.W and S.C.Y did data acquisition; S.W.C. and J.K.W. did the supervision. All authors reviewed the manuscript.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Anctil JL, Anderson DR. Early foveal involvement and generalized depression of the visual field in glaucoma. Arch. Ophthalmol. 1984;102:363–370. doi: 10.1001/archopht.1984.01040030281019. [DOI] [PubMed] [Google Scholar]
  • 2.Kim YK, Ha A, Na KI, Kim HJ, Jeoung JW, Park KH. Temporal relation between macular ganglion cell-inner plexiform layer loss and peripapillary retinal nerve fiber layer loss in glaucoma. Ophthalmology. 2017;124:1056–1064. doi: 10.1016/j.ophtha.2017.03.014. [DOI] [PubMed] [Google Scholar]
  • 3.Kotera Y, Hangai M, Hirose F, Mori S, Yoshimura N. Three-dimensional imaging of macular inner structures in glaucoma by using spectral-domain optical coherence tomography. Invest. Ophthalmol. Vis. Sci. 2011;52:1412–1421. doi: 10.1167/iovs.10-5572. [DOI] [PubMed] [Google Scholar]
  • 4.Wang DL, Raza AS, de Moraes CG, Chen M, Alhadeff P, Jarukatsetphorn R, et al. Central glaucomatous damage of the macula can be overlooked by conventional OCT retinal nerve fiber layer thickness analyses. Transl. Vis. Sci. Technol. 2015;4:4. doi: 10.1167/tvst.4.6.4. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Chen HS, Liu CH, Wu WC, Tseng HJ, Lee YS. Optical coherence tomography angiography of the superficial microvasculature in the macular and peripapillary areas in glaucomatous and healthy eyes. Invest. Ophthalmol. Vis. Sci. 2017;58:3637–3645. doi: 10.1167/iovs.17-21846. [DOI] [PubMed] [Google Scholar]
  • 6.Chung JK, Hwang YH, Wi JM, Kim M, Jung JJ. Glaucoma diagnostic ability of the optical coherence tomography angiography vessel density parameters. Curr. Eye Res. 2017;42:1458–1467. doi: 10.1080/02713683.2017.1337157. [DOI] [PubMed] [Google Scholar]
  • 7.Richter GM, Madi I, Chu Z, Burkemper B, Chang R, Zaman A, et al. Structural and functional associations of macular microcirculation in the ganglion cell-inner plexiform layer in glaucoma using optical coherence tomography angiography. J. Glaucoma. 2018;27:281–290. doi: 10.1097/IJG.0000000000000888. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Yarmohammadi A, Zangwill LM, Diniz-Filho A, Saunders LJ, Suh MH, Wu Z, et al. Peripapillary and macular vessel density in patients with glaucoma and single-hemifield visual field defect. Ophthalmology. 2017;124:709–719. doi: 10.1016/j.ophtha.2017.01.004. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Ooto S, Hangai M, Tomidokoro A, Saito H, Araie M, Otani T, et al. Effects of age, sex, and axial length on the three-dimensional profile of normal macular layer structures. Invest. Ophthalmol. Vis. Sci. 2011;52:8769–8779. doi: 10.1167/iovs.11-8388. [DOI] [PubMed] [Google Scholar]
  • 10.Um TW, Sung KR, Wollstein G, Yun SC, Na JH, Schuman JS. Asymmetry in hemifield macular thickness as an early indicator of glaucomatous change. Invest. Ophthalmol. Vis. Sci. 2012;53:1139–1144. doi: 10.1167/iovs.11-8373. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Asman P, Heijl A. Glaucoma Hemifield Test. Automated visual field evaluation. Arch. Ophthalmol. 1992;110:812–819. doi: 10.1001/archopht.1992.01080180084033. [DOI] [PubMed] [Google Scholar]
  • 12.Susanna R, Jr, Nicolela MT, Soriano DS, Carvalho C. Automated perimetry: a study of the glaucoma hemifield test for the detection of early glaucomatous visual field loss. J. Glaucoma. 1994;3:12–16. doi: 10.1097/00061198-199400310-00005. [DOI] [PubMed] [Google Scholar]
  • 13.Yamada H, Hangai M, Nakano N, Takayama K, Kimura Y, Miyake M, et al. Asymmetry analysis of macular inner retinal layers for glaucoma diagnosis. Am. J. Ophthalmol. 2014;158:1318–1329. doi: 10.1016/j.ajo.2014.08.040. [DOI] [PubMed] [Google Scholar]
  • 14.Kotowski J, Folio LS, Wollstein G, Ishikawa H, Ling Y, Bilonick RA, et al. Glaucoma discrimination of segmented cirrus spectral domain optical coherence tomography (SD-OCT) macular scans. Br. J. Ophthalmol. 2012;96:1420–1425. doi: 10.1136/bjophthalmol-2011-301021. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Rosenfeld PJ, Durbin MK, Roisman L, Zheng F, Miller A, Robbins G, et al. ZEISS angioplex spectral domain optical coherence tomography angiography: technical aspects. Dev. Ophthalmol. 2016;56:18–29. doi: 10.1159/000442773. [DOI] [PubMed] [Google Scholar]
  • 16.DeLong ER, DeLong DM, Clarke-Pearson DL. Comparing the areas under two or more correlated receiver operating characteristic curves: a nonparametric approach. Biometrics. 1988;44:837–845. doi: 10.2307/2531595. [DOI] [PubMed] [Google Scholar]
  • 17.Armstrong RA. When to use the Bonferroni correction. Ophthalmic. Physiol. Opt. 2014;34:502–508. doi: 10.1111/opo.12131. [DOI] [PubMed] [Google Scholar]
  • 18.Rao HL, Dasari S, Riyazuddin M, Puttaiah NK, Pradhan ZS, Weinreb RN, et al. Diagnostic ability and structure-function relationship of peripapillary optical microangiography measurements in glaucoma. J. Glaucoma. 2018;27:219–226. doi: 10.1097/IJG.0000000000000873. [DOI] [PubMed] [Google Scholar]
  • 19.Rao HL, Pradhan ZS, Weinreb RN, Reddy HB, Riyazuddin M, Dasari S, et al. Regional comparisons of optical coherence tomography angiography vessel density in primary open-angle glaucoma. Am. J. Ophthalmol. 2016;171:75–83. doi: 10.1016/j.ajo.2016.08.030. [DOI] [PubMed] [Google Scholar]
  • 20.Triolo G, Rabiolo A, Shemonski ND, Fard A, Di Matteo F, Sacconi R, et al. Optical coherence tomography angiography macular and peripapillary vessel perfusion density in healthy subjects, glaucoma suspects, and glaucoma patients. Invest. Ophthalmol. Vis. Sci. 2017;58:5713–5722. doi: 10.1167/iovs.17-22865. [DOI] [PubMed] [Google Scholar]
  • 21.Rao HL, Riyazuddin M, Dasari S, Puttaiah NK, Pradhan ZS, Weinreb RN, et al. Diagnostic abilities of the optical microangiography parameters of the 3x3 mm and 6x6 mm macular scans in glaucoma. J. Glaucoma. 2018;27:496–503. doi: 10.1097/IJG.0000000000000952. [DOI] [PubMed] [Google Scholar]
  • 22.Shoji T, Zangwill LM, Akagi T, Saunders LJ, Yarmohammadi A, Manalastas PIC, et al. Progressive macula vessel density loss in primary open-angle glaucoma: a longitudinal study. Am. J. Ophthalmol. 2017;182:107–117. doi: 10.1016/j.ajo.2017.07.011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Yarmohammadi A, Zangwill LM, Manalastas PIC, Fuller NJ, Diniz-Filho A, Saunders LJ, et al. Peripapillary and macular vessel density in patients with primary open-angle glaucoma and unilateral visual field loss. Ophthalmology. 2018;125:578–587. doi: 10.1016/j.ophtha.2017.10.029. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Mwanza JC, Durbin MK, Budenz DL, et al. Glaucoma diagnostic accuracy of ganglion cell-inner plexiform layer thickness: comparison with nerve fiber layer and optic nerve head. Ophthalmology. 2012;119:1151–1158. doi: 10.1016/j.ophtha.2011.12.014. [DOI] [PubMed] [Google Scholar]
  • 25.Nouri-Mahdavi K, Nowroozizadeh S, Nassiri N, et al. Macular ganglion cell/inner plexiform layer measurements by spectral domain optical coherence tomography for detection of early glaucoma and comparison to retinal nerve fiver layer measurements. Am. J. Ophthalmol. 2013;156:1297–1307. doi: 10.1016/j.ajo.2013.08.001. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Lisboa R, Paranhos A, Jr, Weinreb RN, Zangwill LM, Leite MT, Medeiros FA. Comparison of different spectral domain OCT scanning protocols for diagnosing preperimetric glaucoma. Invest. Ophthalmol. Vis. Sci. 2013;54:3417–3425. doi: 10.1167/iovs.13-11676. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Wan KH, Lam AKN, Leung CK. Optical coherence tomography angiography compared with optical coherence tomography macular measurements for detection of glaucoma. JAMA Ophthalmol. 2018;136:866–874. doi: 10.1001/jamaophthalmol.2018.1627. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Sullivan-Mee M, Ruegg CC, Pensyl D, Halverson K, Qualls C. Diagnostic precision of retinal nerve fiber layer and macular thickness asymmetry parameters for identifying early primary open-angle glaucoma. Am. J. Ophthalmol. 2013;156:567–577. doi: 10.1016/j.ajo.2013.04.037. [DOI] [PubMed] [Google Scholar]
  • 29.Seo JH, Kim TW, Weinreb RN, Park KH, Kim SH, Kim DM. Detection of localized retinal nerve fiber layer defects with posterior pole asymmetry analysis of spectral domain optical coherence tomography. Invest. Ophthalmol. Vis. Sci. 2012;53:4347–4353. doi: 10.1167/iovs.12-9673. [DOI] [PubMed] [Google Scholar]
  • 30.Asrani S, Rosdahl JA, Allingham RR. Novel software strategy for glaucoma diagnosis: asymmetry analysis of retinal thickness. Arch. Ophthalmol. 2011;129:1205–1211. doi: 10.1001/archophthalmol.2011.242. [DOI] [PubMed] [Google Scholar]
  • 31.Dimitrova G, Chihara E, Takahashi H, Amano H, Okazaki K. Quantitative retinal optical coherence tomography angiography in patients with diabetes without diabetic retinopathy. Invest. Ophthalmol. Vis. Sci. 2017;58:190–196. doi: 10.1167/iovs.16-20531. [DOI] [PubMed] [Google Scholar]
  • 32.Carnevali A, Sacconi R, Corbelli E, Tomasso L, Querques L, Zerbini G, et al. Optical coherence tomography angiography analysis of retinal vascular plexuses and choriocapillaris in patients with type 1 diabetes without diabetic retinopathy. Acta Diabetol. 2017;54:695–702. doi: 10.1007/s00592-017-0996-8. [DOI] [PubMed] [Google Scholar]

Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES