. 2020 Sep 11;99(1):115205. doi: 10.1016/j.diagmicrobio.2020.115205

Pooling samples: a testing option for SARS-CoV-2 during a supply shortage

Hossein Salimnia a,b, Robert Mitchell a, Angela Gundel a, Alicia Cambell a, Fadi Gammou a, Teena Chopra a,c, Marilynn Fairfax a,b,
PMCID: PMC7483063  PMID: 33190092

Abstract

Pooling of 1 positive sample with up to 5 negative samples prior to testing with the Cepheid GenXpert SARS-CoV-2 assay did not adversely impact detection of positive samples. At our current prevalence of 2%, it could save up to 70% of the test kits.

Abbreviations: CoV-2, SARS CoV-2; NA, not available

Keywords: SARS-CoV-2, COVID-19, Sample pooling, Molecular diagnosis, Reverse transcriptase PCR

Highlights

  • Validation of sample pooling to conserve reagents during COVID-19 testing

  • Benefits and limitations of sample pooling during a COVID-19 test reagent shortage

  • Low SARS coronavirus-2 samples detected in pools with 3 negative samples

  • Pooling 4 COVID-19 samples saves 63% of test kits when the prevalence is 3%.


Shortages of reagents and kits for SARS-CoV-2 (CoV-2) tests may limit testing as hospitals reopen to patients at low risk of COVID-19. We wished to validate pooling nasopharyngeal samples collected in viral transport medium from such patients to allow large-scale CoV-2 testing while conserving reagents (Abdalhamid et al., 2020; Hogan et al., 2020). US blood collection agencies routinely pool 16 donor samples to perform molecular-based screening for hepatitis B and C and human immunodeficiency viruses (Dwyre et al., 2011). The US Food and Drug Administration has not issued emergency use authorization for pooling samples for CoV-2, although they say that “we realize that pooling and asymptomatic testing are critical to ending this pandemic and we want to ensure that recommended validation approaches are appropriately designed to provide sufficient data in a least burdensome manner” [email from Yvonne Shea, on 06/12/2020 (yvonne.shea@fda.hhs.gov)]. In addition, they have recommended a comment to be added if sample pooling is utilized (https://www.fda.gov/emergency-preparedness-and-response/coronavirus-disease-2019-covid-19/covid-19-frequently-asked-questions 6/16/2020). We wanted to validate the use of pooling using the Cepheid GenXpert SARS-CoV-2 assay (CoV-assay), a reverse transcriptase polymerase chain reaction (PCR) assay, in an atmosphere of low positivity rates. This assay detects 2 targets genes: E (envelope) and N2 (nucleocapsid). A result is interpreted as positive if N2 is positive regardless of whether E is detected (Cepheid, 2020).

Our medical system (7 hospitals; 1 each specializing in pediatrics and cancer; 2000 beds) experienced a CoV-2 positivity rate > 50% at the peak of the Detroit epidemic when we were testing up to 150 samples per day from symptomatic patients. This rate has fallen to around 2% as our hospitals reopen to patients at lower risk of COVID-19, but we are testing approximately 400 samples per day. Testing 100 samples, in pools of 5, as opposed to individually would require 20 test kits followed by 10 more to test individual samples in the 2 anticipated positive pools, conserving 70 kits. This may prove essential as Cepheid is limiting the amount of testing supplies they will send us.

We determined the impact of sample pooling on detection by retesting 15 previously frozen CoV-2–positive samples initially submitted for patient testing. The samples were chosen based on the presumed concentrations of virus, reflected by their initial cycle or crossing threshold (Ct) value with the E target. Ct values decrease as the viral load decreases. Five had Ct <25 (high virus concentration), 5 had Ct 25–33 (intermediate), and 5e had Ct >33 (low). The manufacturer provides a Ct value for positive but not negative samples, and the cutoff value is proprietary (Cepheid, 2020).

Each sample was first retested to ensure that storage and 1 freeze/thaw cycle had not altered the positivity. The apparent viral load occasionally increased (the Ct decreased), possibly due to random fluctuations or to disaggregation of sample and tissue clumps in the freeze/thaw process. Samples were deidentified and given a testing number. Each sample was then retested repeatedly after pooling with 2–5 negative samples (Table 1 ). To prepare a pool of 3 samples, 0.5 mL from a positive sample was pooled with an equal volume taken from each of 2 negative samples. Pools containing 1 positive and up to 5 added negative samples were prepared similarly. Negative samples were used only once. Each pool was vortexed for 5 s, and 300 μL was used for CoV-2 testing according to the manufacturer's instructions (Cepheid, 2020).

Table 1.

Crossing thresholds and interpretations of 15 samples when pooled with various numbers of negative samples.

Sample number, concentration, initial Ct with E and N2 Targets
Repeat result
Pool of 3 samples
Pool of 5 samples
Pool of 6 samples
E N2 Result E N2 Result E N2 Result E N2 Result E N2
1 High 22.5 24.9 POS 23.3 25.6 POS 25.1 27.4 POS 25.1 27.5 POS 25.4 27.6
2 High 14.6 17.9 POS 13.4 16.2 POS 15.7 18.1 POS 16.1 18.4 POS 16.5 18.7
3 High 15.6 18.4 POS 14.9 17.4 POS 16.4 18.6 POS 17.1 19.4 POS 17.7 19.9
4 High 20.8 23.7 POS 17.7 20.2 POS 19.6 21.9 POS 20.2 22.7 POS 20.5 22.9
5 High 19.3 21.7 POS 18.1 20.5 POS 20.1 22.5 POS 20.7 23.3 POS 21.1 23.4
6 Int 26.2 28.8 POS 26.1 28.8 POS 28.0 30.4 POS 28.4 30.9 POS 28.5 31.2
7 Int 28.3 31.3 POS 31.6 34.4 POS 33.4 36.6 POS 34.6 37.1 POS 30.3 33.3
8 Int 29.6 32.5 POS 30.2 32.8 POS 32.0 35.0 POS 34.2 35.8 POS 30.2 32.9
9 Int 23.0 25.2 POS 21.6 23.7 POS 23.6 25.7 POS 24.3 26.4 POS 24.4 26.4
10 Int 31.2 34.0 POS 30.2 32.8 POS 32.2 35.4 POS 32.7 36.0 POS 32.7 35.8
11 Low 33.4 36.7 POS 31.1 33.8 POS 32.9 36.3 POS 32.4 35.1 POS 34.8 36.8
12 Low 33.2 36.4 POS 27.5 30.2 POS 33.4 36.6 POS 36.2 37.9 POS 35.6 39.3
13 Low 38.5 41.9 POS 33.7 36.9 POS 0.0 42.6 POS 0.0 41.2 POS NA 41.0
14 Low 35.8 38.0 POS 38.5 38.3 POS 41.6 42.3 POS 37.8 43.2 POS NA 42.2
15 Low 35.7 39.3 POS 35.8 39.4 POS 39.9 42.1 POS 40.5 41.4 POS NA 41.9

The initial sample was pooled with negative samples to give the indicated pool sizes. The Ct values of the samples (based on the initial value obtained with the E target) were as follows: high, >25; int, 25–32; low, <33. NA = not available; the instrument does not provide Ct values for samples deemed negative, and the manufacturer's cutoff values are proprietary.

Each pool remained positive regardless of initial Ct or of whether the pool contained 2, 4, or 5 negative samples. However, the E target became undetectable in the 3 samples with the highest initial Ct values (lowest virus titers) when they were pooled with 5 negative samples (pool size 6 samples; Table 1). This suggested that further dilution could cause samples with low titers of CoV-2 to become undetectable.

There are 2 general groups of CoV-2 molecular diagnostic assays from multiple manufacturers. The first requires extraction and purification of the nucleic acid prior to the assay, which could theoretically facilitate concentration and pooling of larger numbers of specimens but may exhibit prolonged turnaround time (TAT). The second group, which includes our assay, is designed for direct from the sample testing and should have a shorter TAT. Sample concentration is not an option, so pooling requires sample dilution and may be more consequential. Before pooling is utilized, initial studies should be performed with each assay to determine how many samples can be pooled without impacting the detection of positive samples. A theoretical calculation by Abdalhamad et al. (2020) concluded that with a sensitivity of 95% or 100%, a specificity of 100%, a lower limit of detection of 1–3 copies/μL, and a prevalence of 5%, the optimal pool size was 5 samples, and their experimental validation supported this. Hogan et al. pooled 10 samples, but our data suggest that such a pool size may be too great for samples with low viral loads tested in our assay. A further consideration is that approximately half of our current testing is ordered stat, as the person being tested is going for a procedure or to the operating room or is in the process of delivering a baby. If such a sample is in a pool that exhibits a positive result, the required retesting would double the in-lab testing time. Thus, pooling of samples may not be appropriate for stat tests.

Acknowledgments

Acknowledgments

This laboratory-based quality control validation using deidentified samples submitted for the diagnosis of SARS CoV-2 infection did not require consent of the patients, did not require approval by the Human Investigation Committee, and received no support outside of the laboratory budget. The authors declare no competing interests.

Author statement

None of the authors of this paper has a conflict of interest or competing interest to declare.

References

  1. Cepheid SARS-CoV-2 package insert queid on line on 6/25/2020 at https://www.cepheid.com/Package%20Insert%20Files/Xpress-SARS-CoV-2/Xpert%20Xpress%20SARS-CoV-2%20Assay%20ENGLISH%20Package%20Insert%20302-3562%20Rev.%20C.pdf.
  2. Abdalhamid B., Bilder C.R., McCutchen E.L., Hinrics S.H., Koebsell S.A., Irwen P.C. Assessment of specimen pooling to conserve SARS CoV-2 testing resources. Am J. Clin Path. 2020;153:6. doi: 10.1093/ajcp/aqaa064. Queried on line 20 may 2020. [DOI] [PMC free article] [PubMed] [Google Scholar]
  3. Dwyre D.M., Fernando L.P., Holland P.V. Hepatitis B, hepatitis C and HIV transfusion-transmitted infections in the 21st century. Vox Sang. 2011;100:92–98. doi: 10.1111/j.1423-0410.2010.01426.x. [DOI] [PubMed] [Google Scholar]
  4. Hogan C.A., Sahoo M.K., Pinsky B.A. 2020. Sample pooling as a strategy to detect community transmission of SARS-CoV-2. 2020; pp. 1967–1969. [DOI] [PMC free article] [PubMed] [Google Scholar]