Abstract
Hip dislocation is a common complication after total hip arthroplasty surgery. Newer prosthetic implants aim to reduce the risk of dislocation. The new dual mobility implant has a unique design that may result in intra‐prosthetic dislocation. We report a case of a recurrently missed iatrogenic intra‐prosthetic dislocation following closed reduction efforts in the emergency department (ED). Emergency physicians must be aware of the design, pitfalls, and management of this new prosthetic hip design.
Keywords: arthroplasty, dislocation, dual‐mobility, hip, iatrogenic
1. INTRODUCTION
It is projected that by the year 2030, >572,000 total hip arthroplasties will be performed each year in the United States alone.1 One of the most common complications following a hip replacement is dislocation. 2 Implants have been developed that aim to reduce rates of dislocation, one of which is the dual mobility acetabular component or dual mobility implant. In the dual mobility implant, a comparatively small femoral head is press‐fit into a larger polyethylene liner, which in turn articulates and rotates within the acetabular shell (Figure 1A).
FIGURE 1.

Dual mobility system. (A) 1: acetabular liner; 2: acetabular shell; 3: femoral head; 4: femoral stem. (B) Demonstrating dislocation of the liner post reduction. Adapted from O'Neill et al 11
A complication unique to the dual mobility design is that of intra‐prosthetic dislocation, in which the small and radiopaque femoral head dislocates from the surrounding, radio‐lucent polyethylene liner, which in turn dislocates from the acetabular shell (Figure 1B). This report presents a case of a missed iatrogenic intra‐prosthetic dislocation following closed reduction in the emergency department setting.
2. CASE REPORT
A 82‐year‐old female presented to the ED following a mechanical fall in which she sustained a right sub capital neck of femur fracture. She had a background history of poorly controlled hypertension and early stage vascular dementia. Prior to this fall, she was independently mobile and active. She underwent a posterior approach right total hip arthroplasty 2 days later. Only 4 months later, the patient represented to the ED after an unwitnessed fall sustaining a superior dislocation of the right hip. The hip was reduced in the department under sedation (Figures 2A and 2B). A week later, the patient presented again after her right leg “gave way” when she bent over, sustaining another superiorly dislocated right hip, and reduction was performed by the emergency physicians, with the post‐reductions films being reported as satisfactory. A further 6 days following her second dislocation, the patient presented to the ED with a painful and restricted range of movement of her right hip; imaging confirmed her third dislocation of her right hip. Her hip was relocated by the emergency physicians under sedation and she was discharged home.
FIGURE 2.

(A) Pre‐reduction image with superior dislocation. (B) Post‐reduction x‐ray demonstrating eccentric positioning of the femoral head in the acetabular shell, and a subtle circular opacity or the “bubble sign” (orange arrow)
Just 4 days later, the patient had ongoing atraumatic hip pain and presented to the ED with her fourth superior dislocation of the right hip. Enlocation was unsuccessful under sedation and the orthopedic services were consulted. A computed tomography (CT) scan of the hip was performed, and although retrospectively a circular opacity is appreciable in the right gluteal musculature (Figure 3), it remained undetected by both the reporting radiologists and treating orthopedic team. The decision was made to attempt closed reduction in the operating theatre under direct imaging guidance, with a plan to convert to a revision if the closed reduction was unsuccessful. The closed reduction in theatres failed and the procedure was converted into a revision right total hip arthroplasty.
FIGURE 3.

Pre‐operative computed tomography demonstrating the MDM liner lying within the gluteal musculature
The patient was weight‐bearing the following day, and after a short duration in rehabilitation with protected weight‐bearing as to protect soft tissue repair, she returned to her premorbid function, ambulating independently without any aids. She was followed up routinely by the orthopedic surgeon at 2 and 6 weeks postoperatively, then again at 6 and 12 months. There were no further complications or dislocations.
3. DISCUSSION
With the increase in numbers of total hip arthroplasties, so comes a rapid increase in the utilization of the dual mobility system. 3 , 4 , 5 Closed reductions of traditional prosthetic hip dislocations are commonly and successfully performed within the ED. 6 However, only limited case reports reported worldwide describe the iatrogenic intra‐prosthetic dislocation and outline the difficulty identifying complications radiologically. The likely mechanism for this iatrogenic intra‐prosthetic dislocation is engagement of the outer polyethylene liner on the rim of the metal cup or on pelvic bony prominences, with subsequent dissociation of the inner bearing couple during the closed reduction maneuver. Loubignac and Boissier 7 termed this the “bottle‐opener” effect.
Radiological features suggestive of intra‐prosthetic dislocation include a halo around the prosthesis, coined the “bubble sign” or “crescent sign” and/or the eccentric position of the femoral head bearing within the acetabular shell. 6 , 8 The eccentric positioning of the femoral head in the correct clinical context has been shown to be the most sensitive of diagnosis of intra‐prosthetic dislocation. 9
Reducing the risk of iatrogenic intra‐prosthetic dislocation includes appropriate sedation with adequate muscle relaxation and to consider fluoroscopic guidance for real‐time feedback that may reduce the risk of intra‐prosthetic dislocation and more dynamically confirm the position of the femoral head at the end of the procedure. 8 , 10 Gentle reduction maneuvers rather than forceful levering have been suggested to avoid the bottle opener effect. 6 We also support performing a CT if there is any doubt of a persistent sensation of instability after reduction. 10 One recent analysis found that 71% of dual mobility dislocations resulted in an intra‐prosthetic dislocation after closed reduction. 8
This case represents a scenario where the dual mobility prosthesis was not recognized, and a closed reduction resulted in an iatrogenic intra‐prosthetic dislocation. The liner is now dissociated from the prosthesis resulting in multiple atraumatic dislocations. It is a unique scenario of a recurrently missed intra‐prosthetic dislocation despite clear radiological evidence to suggest liner dissociation. This case demonstrates a unique but likely typical history of an initial traumatic hip dislocation followed by multiple atraumatic hip dislocations in quick succession. We hope that patients with similar presentations will alert the emergency physician to have a strong suspicion of polyethylene liner dissociation.
We suggest that polyethylene liners be marked or impregnated with a radio‐lucent ring, so that intra‐prosthetic dislocation does not go unnoticed in prosthetic dislocations. However, until that change is implemented, we suggest the current management algorithm for emergency physicians: review medical record for the type of prosthesis, if a dual mobility implant is suspected, to then either (1) discuss with the orthopedic surgeon prior to reduction, or (2) perform the reduction with adequate sedation and relaxation, then only using gentle reduction maneuvers under fluoroscopy. A post reduction x‐ray should be thoroughly reviewed for evidence of the “bubble sign” or “Crescent sign” in addition to the eccentric position of the femoral head. If any concern, a low threshold for a CT of the hip is strongly suggested.
4. CONCLUSIONS
Patients with dual mobility hip prostheses or recurrent hip dislocations should warrant suspicion for intra‐prosthetic dislocation with the emergency physician taking care with reduction techniques and attention to imaging. Orthopedic services should be made aware of these patients as closed reduction is often not adequate and may require revision surgery.
CONFLICTS OF INTEREST
The authors declare no conflict of interest.
ACKNOWLEDGMENT
No financial support to declare.
Fox OJK, Cairns SL, Jarman PG. Recurrent and radiographically unrecognized iatrogenic intra‐prosthesis hip dislocations. JACEP Open. 2020;1:419–422. 10.1002/emp2.12050
Supervising Editor: Alexander X. Lo, MD, PhD
Funding and support: By JACEP Open policy, all authors are required to disclose any and all commercial, financial, and other relationships in anyway related to the subject of this article as per ICMJE conflict of interest guidelines (see www.icmje.org). The authors have stated that no such relationships exist.
REFERENCES
- 1. Nho S, Kymes S, Callaghan J, Felson D. The burden of hip osteoarthritis in the United States: epidemiologic and economic considerations. J Am Acad Orthop Surg. 2013;21(Suppl 1):S1‐S6. [DOI] [PubMed] [Google Scholar]
- 2. Johnson A, Loving L, Herrera L, Delanois R, Wang A, Mont M. Short‐term wear evaluation of thin acetabular liners on 36‐mm femoral heads. Clin Orthop Relat Res. 2014;472(2):624‐629. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. McKee G, Watson‐Farrar J. Replacement of arthritic hips by the McKee‐Farrar prosthetesis. J Bone Joint Surg Br. 1966;48:245‐259. [PubMed] [Google Scholar]
- 4. Philippot R, Camilleri J, Boyer B, Adam P, Farizon F. The use of a dual‐articulation acetabular cup system to prevent dislocation after primary total hip arthroplasty: analysis of 384 cases at a mean follow‐up of 15 years. Int Orthop. 2009;33(4):927‐932. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. De Martino I, D'Apolito R, Waddell B, McLawhorn A, Sculco P, Sculco T. Early intraprosthetic dislocation in dual‐mobility implants: a systematic review. Arthroplast Today. 2017;3(3):197‐202. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. De Martino I, D'Apolito R, Waddell BS, McLawhorn AS, Sculco PK, Sculco TP. Early intraprosthetic dislocation in dual‐mobility implants: a systematic review. Arthroplast Today. 2017;3(3):197‐202. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Loubignac F., Boissier F. Cup dissociation after reduction of a dislocated hip hemiarthroplasty. Rev Chir Orthop Reparatrice Appar Mot. 1997;83(5):469‐472. [PubMed] [Google Scholar]
- 8. Addona JL, Gu A, De Martino I, Malahias M, Sculco TP, Sculco PK. High rate of early intraprosthetic dislocations of dual mobility implants: a single surgeon series of primary and revision total hip replacements. J Arthroplasty. 2019;34(11):2793‐2798. [DOI] [PubMed] [Google Scholar]
- 9. Endo Y, Geannette C, Chang WT. Imaging evaluation of polyethylene liner dissociation in total hip arthroplasty. Skeletal Radiol. 2019;48(12):1933‐1939. [DOI] [PubMed] [Google Scholar]
- 10. Piette N, Guyen O, Moerenhout K. Intra‐prosthetic dislocation of dual mobility hip prosthesis: an original and unusual complication. J Radiol Case Rep. 2019;13(5):15‐23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. O'Neill CKJ, Napier RJ, Diamond OJ, O'Brein S, Beverland DE. Acetabular liner dissociation following total hip arthroplasty: a rare but serious complication that may be easily misinterpreted in the emergency department. Case Rep Emerg Med. 2015;2015:802753. [DOI] [PMC free article] [PubMed] [Google Scholar]