. 2020 Sep 17;15(9):e0237694. doi: 10.1371/journal.pone.0237694

Evaluation of performance of two SARS-CoV-2 Rapid IgM-IgG combined antibody tests on capillary whole blood samples from the fingertip

Thierry Prazuck 1,*, Mathilda Colin 1, Susanna Giachè 1, Camélia Gubavu 1, Aymeric Seve 1, Vincent Rzepecki 1, Marie Chevereau-Choquet 1, Catherine Kiani 1, Victor Rodot 1, Elsa Lionnet 1, Laura Courtellemont 2, Jérôme Guinard 2, Gilles Pialoux 3, Laurent Hocqueloux 1
Editor: Kwok Hung Chan4
PMCID: PMC7498027  PMID: 32941461

Abstract

Background

The SARS-CoV-2 (Severe Acute Respiratory Syndrome CoronaVirus 2) is responsible for the infectious respiratory disease called COVID-19 (COronaVIrus Disease 2019). In response to the growing COVID-19 pandemic, point-of-care (POC) tests have been developed to detect specific antibodies, IgG and IgM, to SARS-CoV-2 virus in human whole blood. We conducted a prospective observational study to evaluate the performance of two POC tests, COVID-PRESTO® and COVID-DUO®, compared to the gold standard, RT-PCR (real-time reverse transcriptase polymerase chain reaction).

Methods

RT-PCR testing of SARS-Cov-2 was performed from nasopharyngeal swab specimens collected in adult patients visiting the infectious disease department at the hospital (Orléans, France). Capillary whole blood (CWB) samples from the fingertip taken at different time points after onset of the disease were tested with POC tests. The specificity and sensitivity of the rapid test kits compared to test of reference (RT-PCR) were calculated.

Results

Among 381 patients with symptoms of COVID-19 who went to the hospital for a diagnostic, 143 patients were RT-PCR negative. Results of test with POC tests were all negative for these patients, indicating a specificity of 100% for both POC tests.

In the RT-PCR positive subgroup (n = 238), 133 patients were tested with COVID-PRESTO® and 129 patients were tested with COVID-DUO® (24 patients tested with both). The further the onset of symptoms was from the date of collection, the greater the sensitivity. The sensitivity of COVID-PRESTO® test ranged from 10.00% for patients having experienced their 1st symptoms from 0 to 5 days ago to 100% in patients where symptoms had occurred more than 15 days before the date of tests. For COVID-DUO® test, the sensitivity ranged from 35.71% [0–5 days] to 100% (> 15 days).

Conclusion

COVID-PRESTO® and DUO® POC tests turned out to be very specific (none false positive) and to be sensitive enough after 15 days from onset of symptom. These easy to use IgG/IgM combined test kits are the first ones allowing a screening with CWB sample, by typing from a finger prick. These rapid tests are particularly interesting for screening in low resource settings.

Introduction

At the end of 2019, a pneumonia of unknown cause detected in Wuhan, China was first reported to the WHO Country Office in China. On January 9th, 2020, the Chinese health authorities and the World Health Organization (WHO) officially announced the discovery of a novel coronavirus, first named 2019-nCoV, then officially termed SARS-CoV-2 (Severe Acute Respiratory Syndrome CoronaVirus 2). This virus, belonging to the coronavirus family, differs from the viruses SARS-CoV, responsible for the SARS outbreak in 2003, and MERS-CoV, responsible for an ongoing outbreak that began in 2012 in the Middle East.

The SARS-CoV-2 virus causes the infectious respiratory disease called COVID-19 (COronaVIrus Disease 2019). This infection mainly results in pneumonia and upper/lower respiratory tract infection. The symptoms of COVID-19 infection appear after an incubation period of approximately 5.2 days [1]. The most common symptoms at onset of COVID-19 illness are fever, cough, and fatigue, but others include headache, sore throat, and even acute respiratory distress syndrome, leading to respiratory failure.

Since the emergence of COVID-19 in China at the end of last year, the SARS-CoV-2 virus has caused a large global outbreak and has become a major worldwide public health issue. The WHO has declared this outbreak a global health emergency at the end of January 2020. On April 12th, 2020, the World Health Organization (WHO) announced that the total global deaths from COVID-19 has surpassed 100 000. Globally, by April 28th, 2020, 2,892,688 cases of COVID-19 have been confirmed and 210,193 patients have died. An estimated 1.7 billion people have been ordered to remain at home as governments take extreme measures to protect their populations.

Due to the rapid spread and increasing number of COVID-19 cases caused by this new coronavirus SARS-CoV-2, rapid and accurate detection of virus and/or disease is increasingly vital to control the sources of infection and prevent the progression of the disease.

Besides the main priority, which is finding an efficient treatment, one of the most important research questions targets the diagnosis of COVID-19. Currently, the real-time RT-PCR (real-time reverse transcriptase polymerase chain reaction) assay is the gold-standard method to detect SARS-CoV-2 [2]. This diagnostic test aims at detecting nucleic acid (RNA) from SARS-CoV-2 in upper and lower respiratory specimens such as nasopharyngeal or oropharyngeal swabs or broncho-alveolar lavage.

In response to the growing COVID-19 pandemic, antibody tests have been developed to detect specific antibodies, IgG and IgM, to SARS-CoV-2 virus in human whole blood, serum or plasma. Two kinds of antibody tests are currently available [3]: quantitative laboratory tests with antibodies titration by enzyme-linked immunosorbent assay (ELISA) and easy-to-use point-of-care (POC) tests, mainly based on lateral flow chromatographic immunoassays.

COVID-PRESTO® and COVID-DUO® are two POC tests products with CE marking which are marketed by AAZ-LMB. In line with the recommendations of the health authorities, we conducted a prospective observational study to evaluate the performance of both AAZ COVID 19 IgM/IgG POC tests compared to the gold standard, RT-PCR.

Methods and materials

Ethics approval

The study was approved by the Orleans Regional Hospital Ethics and Research Committee on March 17th, 2020, and informed consent was obtained from each participant.

Study population

The study population consisted of adult patients visiting the infectious disease department (Centre Hospitalier Regional Orléans, France) from March, 18th, 2020 to April 10th, 2020. This department receives patients whose symptoms, such as headache, fatigue, fever or respiratory signs suggest a COVID infection, and for whom a diagnosis is requested. Date of onset of symptoms as declared by the patient and age were collected at inclusion. According to severity of disease, patients RT-PCR positive were either hospitalized in the infectious diseases ward, only devoted to treat COVID-19 infected patients, or invited to have regular medical visits in the outpatient consultation. Capillary whole blood (CWB) samples from the fingertip were taken at various stages of the follow-up, even after clinical cure, in order to collect samples from convalescent patients.

Specimen collection

Nasopharyngeal (NP) swab specimens were collected from patients by trained surveillance officers. A polyester-tipped flexible aluminum-shafted applicator (Microtest M4RT, Remel) was inserted into one of the nostrils until resistance was felt at the nasopharynx, then rotated 180 degrees and withdrawn. After swabbing, the swab applicator was cut off, and each absorbent swab was placed into a vial containing 3 mL of viral transport media. Vials were immediately shipped via a triple packaging system to the virology unit located in the same building of the hospital, then stored if necessary at 4°C for up to 24 hours until testing.

For CWB samples taken at the fingertip, a lancet was used to prick the side of the fingertip to let a large drop of suspended blood form. This blood sample was collected with a 10 μl capillary micropipette that filled automatically. The sample was then expelled by squeezing the micropipette bulb to deposit the blood on the appropriate well of the test cassette. Retesting was performed in a same patient only if the previous test was negative.

Real-time RT-PCR assays for the detection of SARS-CoV-2

RT-PCR testing of SARS-CoV-2 was performed in Unit of Virology, CHR Orléans. Nucleic acid extraction was performed with automated EZ1 (Qiagen). Specific real-time RT-PCR assays targeting two RNA‐dependent RNA polymerases (IP2 and IP4) and E genes were used to detect the presence of SARS‐CoV‐2 following the instructions in the protocols of the Institut Pasteur and Corman et al., respectively [4, 5]. Amplification was performed on an ABI 7900 Sequence Detection System (Applied Biosystem).

Point-of-care tests to be assessed

The SARS-CoV-2 IgG/IgM antibody test kits, COVID-PRESTO® and COVID-DUO®, are targeting on the antibodies specific to N-protein of SARS-CoV-2. They are manufactured and marketed by AAZ-LMB.

Tests were conducted at the site by clinical staff, physicians or nurses, according to manufacturers’ instructions. Health workers involved in the study received a two-hours training session for each type of test prior to the beginning of the study.

Both COVID-PRESTO® and COVID-DUO® are lateral flow immune-chromatographic assays (Figs 1 and 2). These tests use anti-human IgM antibody (test line IgM), anti-human IgG antibody (test line IgG) and rabbit IgG (control line C) immobilized on a nitrocellulose strip. The Conjugate (recombinant COVID-19 antigens labeled with colloidal gold) is also integrated into the strip. When a specimen is added to the sample well, followed by assay buffer, IgM and IgG antibodies, if present, will bind to COVID-19 conjugates forming an antigen-antibodies complex.

Fig 1. Interpretation of results for COVID-PRESTO®.

Fig 1

Fig 2. Interpretation of results for COVID-DUO®.

Fig 2

This complex migrates through nitrocellulose membrane by capillary action. When the complex meets the line of the corresponding immobilized antibody (anti-human IgM and/or anti-human IgG), the complex is trapped, forming a burgundy colored band which confirms a reactive test result. The result has to be read within 10 minutes by two independent operators. When the control line is the only to be burgundy, the sample is negative. If the control line does not appear, the test is invalid and should be repeated with a new cassette.

Data analysis

Population were described in terms of %, mean, standard deviation, range and median values.

The test data was analyzed in the Department of Infectiology. The specificity and sensitivity of the rapid test kits compared to test of reference (RT-PCR) were calculated according to the following formulas:

Specificity (%) = 100 x [Negative / (Negative + Positive)].

Sensitivity (%) = 100 x [Positive/ (Positive + Negative)]

Confidence intervals for sensitivity were produced with the Wilson score method [6].

Results

Overall, 381 patients with symptoms of COVID-19 who went to the hospital for a diagnostic, were included in the study.

RT-PCR was performed in all patients: 62.47% were positive (n = 238). Based on these results, two sub-groups were defined: 143 patients with negative and 238 patients with positive RT-PCR results (Fig 3).

Fig 3. Number of samples screened with RT-PCR and Point-of-care (POC) tests.

Fig 3

In the negative RT-PCR subgroup, the mean age was 48.20 years (SD: 17.00; range 19–72), median at 46 years. Among these patients, 72 and 71, respectively, were tested with COVID-PRESTO® and COVID-DUO® tests between 24 hours to 8 days from onset of symptoms (median 2 days; range 1–8 days). All results were negative indicating a specificity of 100% for both POC tests.

In the RT-PCR positive subgroup, the mean age of patients was 53.68 years ± 20.18 (median 54; range 19–96).

For COVID-PRESTO® test, CWB samples from the fingertip were collected from 133 patients, only once (n = 133) or at two (n = 16) or three different times (n = 1). Overall 150 samples used to evaluate the sensitivity of this test. The further the onset of symptoms was from the date of collection, the greater the sensitivity (Table 1): 69.23% [CI95%: 53.58–81.43%] for patients with symptoms that occurred from 11 to 15 days before the date of test and 100% [CI95%: 92.59–100%] in patients who experienced first symptoms more than 15 days before the test. Interestingly, among patient with samples collected at two different times, an elderly woman, 75 years of age, with multiple cancer treated by chemotherapy was negative at Day 15 and positive at Day 19, both for IgM and IgG.

Table 1. Evaluation of the sensitivity of the COVID-PRESTO® test.

Number of days since the onset of symptoms
0–5 days 6–10 days 11–15 days >15 days
Positive 2 25 27 48
Negative 18 18 12 0
Sensitivity 10.00% 58.14% 69.23% 100%
[CI 95%] [2.79–30.10%] [4.33–71.62%] [53.58–81.43%] [92.59–100%]

For COVID-DUO® test, 129 patients were screened with one (n = 129), two (n = 4) or three samples (n = 1) at different times. The sensitivity was assessed based on 134 conducted tests (Table 2). The sensitivity ranged from 35.71% [CI95%: 16.34–61.24%] for patients having experienced their 1st symptoms from 0 to 5 days ago, to 100% [CI95%: 89.85–100%] in patients where symptoms had occurred more than 15 days before the date of tests.

Table 2. Evaluation of the sensitivity of the COVID-DUO® test.

Number of days since the onset of symptoms
0–5 days 6–10 days 11–15 days >15 days
Positive 5 23 36 34
Negative 9 19 8 0
Sensitivity 35.71% 54.76% 81.82% 100%
[CI 95%] [16.34–61.24%] [39.95–68.78%] [68.04–90.49%] [89.85–100%]

When considering the distribution of IgM+ and IgG+ patterns among patients with a positive POC test, the IgM were systematically present in the few positive patients with an onset of symptoms from 0 to 5 days ago (n = 2 in COVID-PRESTO® population; n = 5 in COVID-DUO®). The IgM stayed prevalent until 15 days after viral infection while IgG increased over time and became more prevalent after 15 days (Figs 4 and 5).

Fig 4. Patients with a positive COVID-PRESTO® test: Distribution of IgM+ and IgG+ patterns.

Fig 4

Fig 5. Patients with a positive COVID-DUO® test: Distribution of IgM+ and IgG+ patterns.

Fig 5

Discussion

This prospective observational study aimed at evaluating the performance of two POC tests designed to detect SARS-CoV-2 antibodies IgG and IgM from a CWB sample from the fingertip. We investigated the quick detection approach of COVID-PRESTO® and COVID-DUO® in comparison with RT-PCR testing.

The performance analysis was conducted in 381 patients. The results showed that the sensitivity of both POC tests increases with the duration from symptoms onset, reaching 100% in patients experiencing first symptoms of COVID-19 more than 15 days ago. The specificity of both POC tests was found to be 100%, no false positive results having been obtained.

The sensitivity and specificity of such strip assays based on immuno-chromatography have been recently estimated in several studies performed with venous blood samples. In a retrospective study, serum from 179 patients was used to detect SARS-CoV-2 IgG/IgM antibodies [7]. Patients were stratified by the time from symptoms onset to sample collection: 0–7 days, 8–15 days and >15 days. Sensitivities of 18.8%, 100% and 100% were reported, respectively, for the three groups with very few patients (n = 8) in the 8–15 days group. The specificity was 77.8%, 50% and 64.3%, respectively, with numerous reported cases of “false positives”. In a second prospective study, the sensitivity of a strip assay investigated in 86 patients was 11.1%, 92.9% and 96.8% at the early stage (1–7 days after onset), intermediate stage (8–14 days after onset), and late stage (more than 15 days), respectively [8]. In another prospective study with 397 PCR confirmed COVID-19 patients and 128 negative patients, the performance of another lateral flow immunoassay test product was evaluated [9]. Overall, the sensitivity was 88.66% and the specificity 90.63%. Although this study was performed with more patients (n = 525) than in our study, the evaluation of performance was limited because no information was collected about the period over which each patient had experienced symptoms at the time of blood sample collection. Furthermore, to date, no performance study has been reported based on capillary blood samples.

Although COVID-PRESTO® and COVID-DUO® are only qualitative tests, the reported sensitivities and specificities are closed to those of quantitative assays such as enzyme linked immunosorbent assay (ELISA). Zhao et al. collected blood samples from 173 patients with a confirmed infection with SARS-CoV-2 (acute respiratory infection syndromes and/or abnormalities in chest CT images accompanied by detectable SARS-CoV-2 RNA) at different times after onset of COVID-19: <7 days since onset (early phase), 8–14 days after onset (middle phase) and 13–39 days after onset (later phase) [10]. The detection of IgM and IgG against SARS-CoV-2 in this study was performed using ELISA kits. The sensitivities of IgM assays were 38.3%, 73.3% and 94.3% successively, among samples from patients in early, middle and later phases, respectively. For IgG, the values were 38.3%, 54.1% and 79.8%. Interestingly, the RNA test (RT-PCR on samples from the respiratory tract) had the highest sensitivity (66.7%) in the early phase of illness while RNA was only detectable in 45.5% of samples of day 15–39. From a methodological point of view, the performance study presented here was more robust to that of Zhao et al. because the positive population used as reference to evaluate the sensitivity of POC tests was only based on positive RT-PCR results, and not a mix between syndromes, imagining findings and RNA detection.

The results of the present study highlight two major points. First, similar to ELISA tests the sensitivity of POC tests increases when the sample is collected further from the symptom onset. Second, these tests (either qualitative or quantitative) can help to diagnose a past infection after elimination of the virus by the immune system. Therefore, rapid POC testing with CWB samples can provide similar epidemiological information as immunoassay tests [11], but with a lower cost and easier implementation, thus facilitating a larger coverage.

Currently, the extent and the time kinetics of humoral response against SARS-CoV-2 are not known. It is widely accepted that IgM is usually the first responded antibody providing the first line of defense during viral infections, prior to the generation of adaptive, high affinity IgG responses serving as the more robust long term immunity. We were not able to study the humoral response at the individual level because too few patients could have been tested more than once. At the population level, the patterns of IgM/IgG results obtained for positive tests with COVID-DUO® made it possible to perceive a constant dominant presence of IgM which was surpassed by the progressive appearance of IgG from 15 days after symptoms onset. This coincided with our observations with the COVID-PRESTO®. One of the reasons could lie on the high proportion (90%) of false negative results during the early phase of infection, directly linked to the low titers of antibodies during the first days after infection. Both IgM and IgG titers were found to be low or undetectable 4 days after infection [12, 13]. It was also shown that the presence of antibodies was less than 40% among patients within 1 week since onset, and rapidly increased to 94.3% (IgM) and 79.8% (IgG) from day-15 after onset [10]. The presence of IgM and IgG antibodies against SARS-CoV-2 within 2 weeks from the onset of symptoms was confirmed by others [12, 14]. Recently, in 41 COVID-19 patients confirmed by RT-PCR, it was shown by chemiluminescent immunoassay that the median time of seroconversion was 11 days after disease onset for IgG and 14 days for IgM [15]. The time required to have detectable levels of antibodies explains the poor performance (sensitivity 18.4%) reported for antibody tests evaluated in acute patients enrolled from the emergency room, of which only 7 of 38 RT-PCR-positive samples gave positive results for a COVID‐19 IgM/IgG Rapid Test [16]. From this study, Cassaniti et al. concluded that the Rapid Test lateral Flow Immuno Assay was not recommended for triage of patients with suspected COVID‐19 as the disease cannot be excluded when viral serological testing is negative. Although slightly lower than the specificity obtained for COVID-PRESTO® and–DUO®, the specificity demonstrated by Cassaniti et al. at early stages was also high (91.7%), with only one false negative result among 12 tests on RT-PCR negative samples.

This study has several limitations. First, the date of onset of symptoms related to SARS-CoV-2 infection implied recall of facts from memory. This recall bias could lead to some imprecise classification when stratifying the samples by days between onset of symptoms and date of blood samples. Second, few patients with a negative serology could have been re-tested with a second blood sample. In these conditions, we were not able to study the dynamics of seroconversion on individual level. Third, there were still negative tests in RT-PCR positive patients up to 15 days after onset. The reasons are multiple and include the relatively low titers of antibody in the early stages of infection as reported by others [17] and the difference in individual immune response antibody production. Lastly, the strength of antibody response depends on several factors, including age, severity of disease, and certain conditions like immunodeficiency disorders. Therefore it would have been interesting to stratify the population depending on immune health. Indeed, we had few subjects with profound immunosuppression who were still negative 15 days after onset. We know, however, that seroconversion could occur later in such patients [18, 19]. Future studies should focus on seroconversion from Day 15 to Day 30 in highly immunocompromised patients infected with COVID-19. However, the highly immunosuppressed patient in this study was well documented to seroconvert between day 15 and day 19, which provides reassurance of the performance of the POC tests, even in this population. Furthermore, CWB samples collected from patients with other respiratory infections could also be investigated in future studies, in order to further investigate the specificity of the two POC tests and exclude any cross-reactivity with other virus infections, particularly of the coronavirus family.

Despite these limitations, COVID-PRESTO® and DUO® POC tests turned out to be very specific (none false positive) and to be sensitive enough after 15 days from onset of symptom. These easy to use IgG/IgM combined test kits are the first ones allowing an epidemiological screening using CWB samples from the fingertip in order to determine the seroprevalence in a large asymptomatic population. The tests are simple, qualitative, visually interpretable, and give a result within 15 minutes. A positive serology allows to determine whether a person has already been infected by SARS-CoV-2. Serologic tests will be needed to assess the response to vaccine candidates and to map levels of immunity in communities. These rapid tests are particularly interesting for low resource settings such as at the bedside or any other locations where lab tests are less obvious.

Supporting information

S1 Raw data

(XLSX)

Acknowledgments

The authors would like to thank the technical staff of the Department of Infectious diseases for excellent assistance. Furthermore, the authors thank Angèle Guilbot of Clinact, France for providing medical writing support/editorial support in accordance with Good Publication Practice (GPP3) guidelines.

Data Availability

All relevant data are within the manuscript and its Supporting Information files.

Funding Statement

The study was funded by CHR Orleans (Orléans Regional Hospital Centre), a public hospital with no-profit status, of which all authors are employees. The point-of-care tests were provided free of charge by AAZ-LMB.

References

  • 1.Rothan HA, Byrareddy SN. The epidemiology and pathogenesis of coronavirus disease (COVID-19) outbreak. Journal of Autoimmunity. 2020;109: 102433 10.1016/j.jaut.2020.102433 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2.Gorbalenya AE, Baker SC, Baric RS, de Groot RJ, Drosten C, Gulyaeva AA, et al. Severe acute respiratory syndrome-related coronavirus: The species and its viruses–a statement of the Coronavirus Study Group. Microbiology; 2020. February 10.1101/2020.02.07.937862 [DOI] [Google Scholar]
  • 3.World Health Organization 2020. Advice on the use of point-of-care immunodiagnostic tests for COVID-19. Scientific brief. 8 April 2020. Available from: https://www.who.int/news-room/commentaries/detail/advice-on-the-use-of-point-of-care-immunodiagnostic-tests-for-covid-19. Accessed 9 May 2020.
  • 4.World Health Organization 2020—Real-time RT-PCR assays for the detection of SARS-CoV-2, Institut Pasteur, Paris. Available from: https://www.who.int/docs/default-source/coronaviruse/real-time-rt-pcr-assays-for-the-detection-of-sars-cov-2-institut-pasteur-paris.pdf?sfvrsn=3662fcb6_2. Accessed 9 May 2020.
  • 5.Corman VM, Landt O, Kaiser M, Molenkamp R, Meijer A, Chu DK, et al. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Eurosurveillance. 2020;25 10.2807/1560-7917.ES.2020.25.3.2000045 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Newcombe RG. Two-sided confidence intervals for the single proportion: comparison of seven methods. Statistics in Medicine. 1998;17: 857–872. [DOI] [PubMed] [Google Scholar]
  • 7.Liu Y, Liu Y, Diao B, Ren F, Wang Y, Ding J, et al. Diagnostic Indexes of a Rapid IgG/IgM Combined Antibody Test for SARS-CoV-2. Infectious Diseases (except HIV/AIDS); 2020. March 10.1101/2020.03.26.20044883 [DOI] [Google Scholar]
  • 8.Pan Y, Li X, Yang G, Fan J, Tang Y, Zhao J, et al. Serological immunochromatographic approach in diagnosis with SARS-CoV-2 infected COVID-19 patients. Journal of Infection. 2020. 10.1016/j.jinf.2020.03.051 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Li Z, Yi Y, Luo X, Xiong N, Liu Y, Li S, et al. Development and Clinical Application of A Rapid IgM‐IgG Combined Antibody Test for SARS‐CoV‐2 Infection Diagnosis. Journal of Medical Virology. 2020. 10.1002/jmv.25727 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Zhao J, Yuan Q, Wang H, et al. Antibody responses to SARS-CoV-2 in patients of novel coronavirus disease 2019. Clin Infect Dis. 2020. March 28 pii: ciaa344 10.1093/cid/ciaa344 [Epub ahead of print] [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 11.Pollán M, Pérez-Gómez B, Pastor-Barriuso R, Oteo J, Hernán MA, Pérez-Olmeda M, et al. Prevalence of SARS-CoV-2 in Spain (ENE-COVID): a nationwide, population-based seroepidemiological study. The Lancet. 2020; S0140673620314835. 10.1016/S0140-6736(20)31483-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Haveri A, Smura T, Kuivanen S, Österlund P, Hepojoki J, Ikonen N, et al. Serological and molecular findings during SARS-CoV-2 infection: the first case study in Finland, January to February 2020. Eurosurveillance. 2020;25 10.2807/1560-7917.ES.2020.25.11.2000266 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Wölfel R, Corman VM, Guggemos W, Seilmaier M, Zange S, Müller MA, et al. Virological assessment of hospitalized patients with COVID-2019. Nature. 2020. 10.1038/s41586-020-2196-x [DOI] [PubMed] [Google Scholar]
  • 14.Nisreen M.A. Okba, Marcel A. Müller, Wentao Li, Chunyan Wang, Corine H. GeurtsvanKessel, Victor M. Corman, et al. Severe Acute Respiratory Syndrome Coronavirus 2−Specific Antibody Responses in Coronavirus Disease 2019 Patients. Emerging Infectious Disease journal. 2020;26 10.3201/eid2607.200841 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15.Qu J, Wu C, Li X, Zhang G, Jiang Z, Li X, et al. Profile of IgG and IgM antibodies against severe acute respiratory syndrome coronavirus 2 (SARS-CoV-2). Clinical Infectious Diseases. 2020. 10.1093/cid/ciaa489 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16.Cassaniti I, Novazzi F, Giardina F, Salinaro F, Sachs M, Perlini S, et al. Performance of VivaDiag COVID‐19 IgM/IgG Rapid Test is inadequate for diagnosis of COVID‐19 in acute patients referring to emergency room department. Journal of Medical Virology. 2020. 10.1002/jmv.25800 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Du Z, Zhu F, Guo F, Yang B, Wang T. Detection of antibodies against SARS-CoV-2 in patients with COVID-19. Journal of Medical Virology. 2020. 10.1002/jmv.25820 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Brousseau N, Murphy DG, Gilca V, Larouche J, Mandal S, Tedder RS. Acute hepatitis B virus infection with delayed appearance of hepatitis B core antibody in an immunocompromised patient: a case report. Journal of Medical Case Reports. 2017;11 10.1186/s13256-017-1264-9 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Lehrnbecher T, Koehl U, Wittekindt B, Bochennek K, Tramsen L, Klingebiel T, et al. Changes in host defence induced by malignancies and antineoplastic treatment: implication for immunotherapeutic strategies. The Lancet Oncology. 2008;9: 269–278. 10.1016/S1470-2045(08)70071-8 [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Kwok Hung Chan

Transfer Alert

This paper was transferred from another journal. As a result, its full editorial history (including decision letters, peer reviews and author responses) may not be present.

1 Jul 2020

PONE-D-20-15374

Evaluation of performance of two SARS-CoV-2 Rapid whole-blood finger-stick IgM-IgG Combined Antibody Tests

PLOS ONE

Dear Dr. Guilbot,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Aug 15 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Kwok Hung Chan, Ph.D

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2.Thank you for including your ethics statement:   "The study was approved by the local Ethics Committee on March 17th 2"020, and informed consent was obtained from each participant."

Please amend your current ethics statement to include the full name of the ethics committee/institutional review board(s) that approved your specific study.

Once you have amended this/these statement(s) in the Methods section of the manuscript, please add the same text to the “Ethics Statement” field of the submission form (via “Edit Submission”).

For additional information about PLOS ONE ethical requirements for human subjects research, please refer to http://journals.plos.org/plosone/s/submission-guidelines#loc-human-subjects-research.

3.Thank you for stating the following financial disclosure:

 [Rapid Diagnostic Tests were provided free of charge by AAZ-LMB.].

At this time, please address the following queries:

  1. Please clarify the sources of funding (financial or material support) for your study. List the grants or organizations that supported your study, including funding received from your institution.

  2. State what role the funders took in the study. If the funders had no role in your study, please state: “The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.”

  3. If any authors received a salary from any of your funders, please state which authors and which funders.

  4. If you did not receive any funding for this study, please state: “The authors received no specific funding for this work.”

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

4.We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. Please see http://www.bmj.com/content/340/bmj.c181.long for guidelines on how to de-identify and prepare clinical data for publication. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

5. Please amend your authorship list in your manuscript file to include author Angèle Guilbot.

Additional Editor Comments (if provided):

The specificity of the POCT (RDTs) which depends on COVID-19 negative patients by RT-PCR is not enough. The specificity test must be included patients with other respiratory virus infection including human coronaviruses.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Review

This manuscript compares performance of two Point-of-care rapid whole blood finger stick antibody tests for COVID-19. Authors were able to show the performance of these antibody tests (sensitivity and specificity) is good when used after 15 days from onset of symptom

Major Concerns:

1. I have concerns with authors calling COVID-19 antibody tests as “Rapid Diagnostic Tests” (RDTs). WHO, USA CDC, and USA FDA have clearly discouraged using antibody tests as diagnostic tests because of test performance issues and low prevalence of disease. The appropriate way to address these lateral flow antibody tests would be to call them “Point-of-care (POC) tests”.

2. In my opinion, this study is not a “real-life study”, the authors are not studying the effectiveness of intervention (lab results) in “real-life” in the enrolled patients. Authors do not describe how the results of antibody tests were used for providing patient care in this manuscript to qualify as “real-life” study. This study is better characterized as a “Prospective Observational Study”.

Minor Concerns

1. Cover page lists “Angèle Guilbot, Ph.D” as corresponding author and page 1 lists “Thierry Prazuck” as corresponding author. Please verify corresponding author.

2. Please use abbreviations consistently. First use of abbreviation should be spelled out in both abstract and main body separately. Eg. SARS-CoV-2 never spelled out as “Severe acute respiratory syndrome coronavirus 2”. COVID-19 should be spelled out as “coronavirus disease 2019”. RT-PCR should be spelled out as “real-time reverse transcriptase polymerase chain reaction” during first use.

3. Page 3, Line # 92-95:

a. Consider using “antibody tests” instead of “serologic tests” because serologic tests also include antigen tests.

b. Two types of antibody tests are “Point-of-care (POC) tests” (generally lateral flow devices) and “Laboratory tests” (ELISA or chemiluminescent immunoassay tests). Please refer to CDC website for antibody testing classification, https://www.cdc.gov/coronavirus/2019-ncov/lab/resources/antibody-tests-guidelines.html.

c. Consider changing “Rapid Diagnostic Tests (RDTs)” to “Point-of-care (POC) tests” (throughout the manuscript).

4. Page 4, line # 113, consider changing “Fingertip blood samples” to “capillary whole blood (WB) sample from fingertip” (throughout the manuscript).

5. Page 4, Line # 136: Consider changing “Rapid Diagnostic Tests” to “Point-of-care (POC) tests” (throughout the manuscript).

6. Page 6, line # 172-175: Consider providing details of average day from symptoms onset (at least average and range).

7. Page 6, line # 192-196: Not sure if authors can draw the conclusion “the IgM were the first antibodies to be detected” from the data provided in Figure 4 and 5. IgG antibodies were present at 0-5 days along with IgM for both COVID-PRESTO® and COVID-DUO®. The % was lower but IgG were not absent at 0-5 days. Also, the number of specimens tested was too small to come to such conclusions.

8. Page 8, line # 239-245:

a. Consider rephrasing the statement “Firstly, as for the assessed RDT, the sensitivity of ELISA tests increases with the duration from symptom onset. Moreover, we showed that direct antibody typing with whole blood is as sensitive as immunoassay performed with serum in a retrospective way.” My understanding is that the authors are trying to say, similar to ELISA tests, the sensitivity of Point of care (lateral flow) antibody tests increases when specimen is collected further from the symptom onset.

b. I disagree with the second point that authors are making “Secondly, these diagnostic tests (either qualitative or quantitative) can help to diagnose a past infection after elimination of the virus by the immune system.” This study does not show diagnostic application of antibody tests. This generalized statement is premature without defining boundaries and criteria for using antibody tests to “diagnose past infections” of COVID-19. Clinical utility of diagnosing past COVID-19 infections is very limited (eg. negative molecular tests in patients presenting late in illness and to establish diagnosis of late complications of COVID-19 illness like Multisystem Inflammatory Syndrome in children) and should be clearly defined before calling antibody tests as “diagnostic tests”.

c. Authors do try to explain use of antibody tests by saying “Thus, combining RT-PCR and antibody detection allows to largely diagnose COVID-19 people regardless of the delay between infection and diagnosis”, but this is till too vague description of its use. This statement is true if antibody tests are used in symptomatic patients only.

9. Page 8, line # 251-253: The results of the study do not show “the switch between the first production of IgM and the later onset of IgG.” As shown in Figure 4 and 5, both IgG and IgM were present throughout the duration from onset of symptoms (0-5 days to >15 days). There was no data provided which showed “seroconversion” or “switch” in patients tested.

10. Page 8, line # 267-269: Consider rephrasing statement “Although slightly lower than the specificity obtained for COVID-PRESTO® and –DUO®, the specificity at early stages was high (91.7%), preventing false positive diagnoses.”

a. Not sure what authors are trying to convey here. Because of high specificity, antibody tests will have low “false positive” diagnosis but more importantly these antibody tests will have significantly high numbers of “false negatives” because of low sensitivity in early stages of disease.

b. Please provide more details of calculations to understand how authors came up with specific of 91.7%. Please also define what does “early stage” means (0-5 days or 0-10 days)?

11. Page 9, line 289-291: Authors say that the antibody tests can be used for “screening” but do not explain screening for what purpose? Screening for diagnosis or immunity or epidemiologic studies, etc.?

12. Page 9, line 292-296: the conclusion is completely different than rest of the manuscript. Throughout the manuscript authors are presenting antibody tests as “diagnostic tests” and then at the end they appropriately limit the use to “assess the response to vaccine candidates and to map levels of immunity in communities”. These two uses are Epidemiologic use of antibody tests and are not considered as “diagnostic use”.

Reviewer #2: This is well written manuscript.

Please change verb responsible (line 68) to causing or similar to omit repeating (previous sentence already has responsible)

My preference would be to use first, second, third etc.(lines 239, 242, 270, 273, 275).

Figure 3 can be omitted

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Gagan Mathur, MD, MBA

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Sep 17;15(9):e0237694. doi: 10.1371/journal.pone.0237694.r002

Author response to Decision Letter 0


17 Jul 2020

We would like to thank the editor and the reviewers for their thorough and constructive review of our manuscript. We considered all of your requests very carefully and provide our responses in blue below as well as, if applicable, directly in the manuscript.

Answers to Editor’s comments:

1. PLOS ONE style requirements: The style requirement have been verified.

2. Updated ethics statement (also corrected in the manuscript): “The study was approved by the Orleans Regional Hospital Ethics and Research Committee on March 17th 2020, and informed consent was obtained from each participant.”

3. Financial disclosure statement: “The study was funded by CHR Orleans (Orléans Regional Hospital Centre), a public hospital with no-profit status, of which all authors are employees. The point-of-care were provided free of charge by AAZ-LMB.”

4. Data availability: The data is available via APREMIT45, Direction de la recherche, Centre Hospitalier Regional Orleans, 14 avenue de l’hôpital 45067 Orleans

5. List of authors/role of Angèle Guilbot: As the submission of this manuscript has been delegated to Clinact. Therefore Angèle Guilbot will act as primary contact person regarding the manuscript processing in the journal, i.e. serve as corresponding “author” for the submission procedure only. She will not appear as corresponding author on the manuscript.

6. The specificity test must be included patients with other respiratory virus infection including human coronaviruses: This is indeed relevant when evaluating performances on serum. In real life, when testing patients’ blood samples from the fingertip, it is actually not possible to find any patient with current active corononavirus infection different from COVID19.

Answers to Reviewer#1 comments:

1. Concerns regarding “Rapid Diagnostic Tests” (RDTs): We agree and have renamed these “Point-of-care (POC) tests” throughout the manuscript according to your suggestion.

2. Concerns regarding “real-life study”: “Prospective Observational Study”: We agree and have renamed this “Prospective Observational Study” throughout the manuscript according to your suggestion.

3. List of authors/role of Angèle Guilbot: As the submission of this manuscript has been delegated to Clinact. Therefore Angèle Guilbot will act as primary contact person regarding the manuscript processing in the journal, i.e. serve as corresponding “author” for the submission procedure only. She will not appear as corresponding author on the manuscript.

4. Consistency of abbreviations: We verified all abbreviations to make sure they are spelled out the first time used.

5. Page 3, Line # 92-95 :

a. Concerns regarding “serologic tests” Renamed to “antibody tests”.

b. Definition of two types of antibody tests: Rephrased.

c. Concerns regarding “Rapid Diagnostic Tests (RDTs)”: Renamed.

6. Page 4, line # 113, Concerns regarding “Fingertip blood samples” Renamed.

7. Page 4, Line # 136: Concerns regarding “Rapid Diagnostic Tests” Renamed.

8. Page 6, line # 172-175: Details of average day from symptoms onset Added median (2 days) and range (1-8 days).

9. Page 6, line # 192-196: Concerns regarding conclusion “the IgM were the first antibodies to be detected”: We agree, the conclusion was rephrased.

10. Page 8, line # 239-245:

a. Concerns regarding statement of the sensitivity: We agree, the sentence was rephrased.

b. Concerns regarding statement of diagnostic application We agree, it was deleted.

c. Authors do try to explain use of antibody tests by saying “Thus, combining RT-PCR and antibody detection allows to largely diagnose COVID-19 people regardless of the delay between infection and diagnosis”, but this is till too vague description of its use. This statement is true if antibody tests are used in symptomatic patients only. We agree, the paragraph was rephrased.

11. Page 8, line # 251-253: Concerns regarding description of IgG and IgM detection: We agree, the sentence was rephrased.

12. Page 8, line # 267-269: Concerns regarding statement of specificity: We agree, it was rephrased.

a. Not sure what authors are trying to convey here. Because of high specificity, antibody tests will have low “false positive” diagnosis but more importantly these antibody tests will have significantly high numbers of “false negatives” because of low sensitivity in early stages of disease. We agree, the paragraph was rephrased.

b. Please provide more details of calculations to understand how authors came up with specific of 91.7%. Please also define what does “early stage” means (0-5 days or 0-10 days)? More details were included.

13. Page 9, line 289-291: Request for more information regarding “screening”: The claim has been specified.

14. Page 9, line 292-296: the conclusion is completely different than rest of the manuscript. Throughout the manuscript authors are presenting antibody tests as “diagnostic tests” and then at the end they appropriately limit the use to “assess the response to vaccine candidates and to map levels of immunity in communities”. These two uses are Epidemiologic use of antibody tests and are not considered as “diagnostic use”. We agree and addressed the “diagnostic application” throughout the rest of the manuscript.

Answers to Reviewer#2 comments:

1. Please change verb responsible (line 68) to causing or similar to omit repeating (previous sentence already has responsible): Done.

2. My preference would be to use first, second, third etc.(lines 239, 242, 270, 273, 275): Agreed and changed.

3. Figure 3 can be omitted: In this case we disagree. In our opinion the figure summarizes and visualizes the design of the study and we would like to keep it. As several journals request this type of study flow-chart, we leave the final decision of the editor.

Decision Letter 1

Kwok Hung Chan

27 Jul 2020

PONE-D-20-15374R1

Evaluation of performance of two SARS-CoV-2 Rapid IgM-IgG Combined Antibody Tests on capillary whole blood samples from the fingertip

PLOS ONE

Dear Dr. Guilbot,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Sep 10 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Kwok Hung Chan, Ph.D

Academic Editor

PLOS ONE

Additional Editor Comments (if provided):

Serum samples collected from other respiratory infections can be used to replace blood samples to show the specificity of these two antibody rapid kits. So, the lack of fingertip blood samples from current patients with other respiratory virus infections is a major limitation of this manuscript and should be included in the discussion. This indicates cross-reactivity with other virus infections particularly coronavirus family can’t be excluded.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: All concerns satisfactorily addressed by the authors in the revision. Ok to proceed with publication

Reviewer #2: Thank you for addressing all comments. I would suggest final proofread by the native English speaker to address minor things throughout manuscript. Please change COronaVirus disease to coronavirus disease (lines 27, 72)

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: Yes: Gagan Mathur, MD, MBA

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Sep 17;15(9):e0237694. doi: 10.1371/journal.pone.0237694.r004

Author response to Decision Letter 1


30 Jul 2020

A cover letter responding to the editor's comment and an excel file with the raw data as supporting information have been uploaded.

Decision Letter 2

Kwok Hung Chan

3 Aug 2020

Evaluation of performance of two SARS-CoV-2 Rapid IgM-IgG Combined Antibody Tests on capillary whole blood samples from the fingertip

PONE-D-20-15374R2

Dear Dr. Guilbot,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Kwok Hung Chan, Ph.D

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Kwok Hung Chan

8 Sep 2020

PONE-D-20-15374R2

Evaluation of performance of two SARS-CoV-2 Rapid IgM-IgG Combined Antibody Tests on capillary whole blood samples from the fingertip

Dear Dr. Prazuck:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Kwok Hung Chan

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Raw data

    (XLSX)

    Data Availability Statement

    All relevant data are within the manuscript and its Supporting Information files.