A novel skin-stretching device for closing large skin-soft tissue defects after soft tissue sarcoma resection

Qiang Wu 1,#, Zengwu Shao 1,#, Yubin Li 2, Saroj Rai 3, Min Cui 1, Ying Yang 4, Baichuan Wang 1,
PMCID: PMC7499963  PMID: 32943050

Abstract

Background

Closure of large skin-soft tissue defects following soft tissue sarcoma (STS) resection has been a great challenge. The objective of this study was to evaluate the effectiveness of a novel, simple, and cheap skin-stretching device (bidirectional regulation-hook skin closure system, BHS) for closing large skin-soft tissue defects resulting from the removal of STS and the complications associated with the use of the BHS.

Methods

From January 2017 to September 2018, 25 patients with STS underwent BHS therapy after tumor resection. BHS was used for two main clinical applications: securing wound closure after high-tension suture closure and delayed wound closure. We described a detailed reconstruction procedure regarding this therapy. Wound closure and complications associated with BHS therapy were recorded. We also analyzed tumor recurrence and metastases.

Results

All patients were observed for 16–36 months with an average follow-up of 25.6 months. During the follow-up period, no significant functional restriction was observed and the final scar was aesthetically acceptable. Superficial wound infection occurred in six patients, wound edge ischemia in two patients, and small skin tears in two patients. Two patients developed pulmonary metastasis, two patients had a local recurrence, and one patient died of pulmonary metastasis.

Conclusions

BHS therapy can effectively close large skin-soft tissue defects following STS resection and obtain acceptable functional results, without severe complications. However, larger studies are required to further evaluate the effectiveness, indications, and complications of BHS therapy.

Keywords: Soft tissue sarcoma, Radical resection, Tissue expansion, Tension suture, Complications

Introduction

Soft tissue sarcoma (STS) is an uncommon malignant tumor that originates from primitive mesenchymal tissue. At present, treatment of STS is a multimodal therapy that involves surgical resection, chemotherapy, and radiotherapy [1]. Among these, surgical resection is the mainstay of therapy. However, it frequently results in large skin-soft tissue defect and exposure of vital structures because of the wide margins that need to be taken.

Closure of large skin-soft tissue defects following STS resection is a common reconstructive challenge. Primary closure of wounds with conventional suturing techniques is usually the optimal solution because of its simplicity and acceptable outcomes, yet it is often limited, particularly with regard to large wounds [2]. The most commonly used methods for closing large skin-soft tissue defects following STS resection are local or regional flaps and free tissue flaps [1, 36]. However, these methods are often associated with relatively complex surgical procedures, long operation time, significant donor site morbidity, extended hospitalization, and increased costs. In recent years, various external skin-stretching devices, as a promising alternative, have been used to close different types of large skin-soft tissue defects, and acceptable results had been obtained [715]. These types of devices take full advantage of viscoelastic properties of the skin (such as mechanical creep and stress relaxation) [1618] to provide immediate or delayed primary wound closure while avoiding the shortcomings of the aforementioned techniques. However, the application of many devices may be greatly restricted due to a variety of reasons such as different countries and high cost. In this study, we introduced a novel, simple, and cheap skin-stretching device (bidirectional regulation-hook skin closure system, also called BHS; Tianjin Xinzhong Medical Devices CO., LTD, China) to close large skin-soft tissue defects following STS resection.

The objective of this study was to evaluate the effectiveness of the novel skin-stretching device BHS for closing large skin-soft tissue defects resulting from the removal of STS. Moreover, we also evaluated the complications associated with the use of the device, and tumor recurrence and metastases.

Patients and methods

Clinical data

From January 2017 to September 2018, following approval by our hospital ethical committee, 25 patients who underwent a novel skin-stretching device (BHS) therapy after STS resection were enrolled in this study. All participating patients gave written informed consent. Inclusion criteria were an open skin defect wound on the limbs or trunk in which direct closure was impossible or primary closure was achieved but with a high tension. Patients with severe cardiopulmonary diseases, vascular diseases, or uncontrolled diabetes mellitus; patients with damaged skin (e.g., severe edema skin and deeply scarred wound edges); and patients refusing the procedure were excluded from this study. BHS was used for two main clinical applications: (a) securing wound closure after high-tension suture closure and (b) delayed wound closure.

The pathological diagnoses of 20 patients were obtained by preoperative needle biopsy, while those of the other 5 patients were obtained by excisional biopsy (done in other institutions). Nine patients were diagnosed with fibrosarcoma, 5 with synovial sarcoma, 3 with undifferentiated sarcoma, 3 with melanoma, 2 with skin protuberant fibrosarcoma, 2 with clear cell sarcoma, and 1 with leiomyosarcoma. The five patients with excisional biopsy had to undergo secondary wide resection due to positive resection margins.

Viscoelastic properties of the skin vary in different sites, and therefore, the site, width, and length of the wound are critical for successful wound closure. These wound parameters were recorded in detail. In addition, information including postoperative wound dressing, skin stretching, wound closure, and follow-up were also recorded (Table 1). The scar tissues were assessed by the following aspects: color, thickness, relief, pliability, and surface area [19].

Table 1.

Summary data for all patients

Variables Value
Biopsy (n, %)
 Needle biopsy 20 (80%)
 Excisional biopsy 5 (20%)
Pathological diagnoses (n, %)
 Fibrosarcoma 9 (36%)
 Synovial sarcoma 5 (20%)
 Undifferentiated sarcoma 3 (12%)
 Melanoma, 3 (12%)
 Skin protuberant fibrosarcoma 2 (8%)
 Clear cell sarcoma 2 (8%)
 Leiomyosarcoma 1 (4%)
Wound location (n, %)
 Upper extremity Forearm 3 (12%); upper arm 3 (12%)
 Lower extremity Thigh 8 (32%), calf 6 (24%), feet 2 (8%)
 Trunk Back 2 (8%), abdomen 1 (4%)
Wound size (cm)
 Length: range, mean (SD) cm 3.5–19, 10.1 (4.1)
 Width: range, mean (SD) cm 1.5–11, 6.1 (2.9)
Skin stretching (n, %)
 Intraoperative stretching 7 (28%)
 Postoperative stretching 18 (72%)
Time to start postoperative chemo and/or radio
Range, mean (SD) days 18–31, 23.2 (4.5)
Follow-up time: range, mean (SD) months 16-36, 25.6 (3.7)
Complications (n, %)
 Minor superficial wound infection 6 (24%)
 Wound edge ischemia 2 (8%)
 Small skin tears 2 (8%)
Survival status (n, %)
 DFS 20 (80%)
 AWD 4 (16%; 2 with pulmonary metastasis, 2 with local recurrence)
 DOD 1 (4%)

SD standard deviation, DFS disease-free survival, AWD alive with disease, DOD died of disease

Skin-stretching device

BHS is composed of two hook holder modules and a compression module. The hook holder module is made of polyetheretherketone (PEEK) material, and the compression module is made of stainless steel. Each holder module contains a threaded plunger hole, two guide holes, and three hook holder holes with sharp hooks at their tips facing each other. The compression module is formed of a threaded rod, a spring, and two guide rods. The threaded rod is passed through the threaded plunger holes on the hook holder modules with the aid of the two guide rods. The two hook holder modules can be approximated together by rotating the threaded rod and pushing the spring along the guide rods (Fig. 1). The price of each BHS device is about $ 80.

Fig. 1.

Fig. 1

The novel skin-stretching device, BHS. a Threaded rod cap, b spring, c hook holder module, d sharp hook, e threaded plunger hole, f threaded rod, and (g) guide rod. The two hook holder modules can be approximated together by rotating the threaded rod and pushing the spring along the guide rods

Surgical technique

After complete removal of the tumor in accordance with the principle of STS resection, large skin-soft tissue defects were treated by using BHS. After cleaning the wound, 2.0-mm diameter Kirschner wires (KWs) were inserted into the skin and subcutaneous tissue approximately 1 cm away from wound margins. They passed subcutaneously for some distance, and then out of the skin to form two or more equal-sized bridges as required. Then, the two hook holder modules were placed on the KWs bridges (or the skin and subcutaneous tissue directly) by using hooks. The threaded rod was passed through the threaded plunger hole and gently advanced until the wound begins to resist closure. In order to avoid overstretching of the skin, the color of the wound edges was carefully observed during the period of gradual approximation. For wounds with immediate closure under high tension, the device was still in place until the stretching force significantly reduced. For wounds with delayed closure, the wound edges were approximated one to two times 1 day by the patient him/herself. The device was removed when the wound edges were completely together, and then the wound was sutured (Figs. 2 and 3). The number of skin-stretching devices required depended on the size of the wound.

Fig. 2.

Fig. 2

Reconstruction of a large skin-soft tissue defect in the right calf by using BHS. a Wound after fibrosarcoma resection measuring about 13 × 8 cm. b Gradual approximation of wound edges. c Wound was closed 8 days after gradual stretching. d Three months following reconstruction, uneventful recovery, with acceptable aesthetic result and minor widening of the scar

Fig. 3.

Fig. 3

Reconstruction of a large skin-soft tissue defect in the right thigh by using BHS. a Recurred fibrosarcoma in the right thigh. b Wound after tumor resection measuring about 19 × 11 cm. c, d Gradual approximation of wound edges. e Wound was sutured 12 days after gradual stretching. f Eighteen months following reconstruction, the photograph showed acceptable aesthetic result and mild scar

Postoperative treatments

Patients with open wounds received an intravenous injection of broad-spectrum antibiotic both preoperatively and postoperatively. For wounds with immediate closure, conventional dressing change was performed every 2 days until the wound healed completely. For wounds with delayed closure, the open wounds were covered with vaseline gauze and silver ion surgical dressing, and the dressings were changed every 2 days. In order to avoid the necrosis and laceration of the skin, the color of the wound edge was carefully observed and the status of pain was assessed during the progress of gradual approximation. Wound biopsy for tissue culture was performed to evaluate wound infection before suturing. Once the wound was sutured, patient follow-up took place at 7, 14, and 21 days and every 3 months thereafter. All patients received postoperative chemotherapy and/or radiotherapy or biological immunization therapy after wound healing.

Results

There were 15 male and 10 female patients, and their average age was 56.1 years (range, 25–71 years). The rate of wound closure was determined according to the size of the defect. Of the 25 patients, 7 patients underwent immediate intraoperative wound closure with the aid of BHS and 18 patients underwent postoperative gradual approximation to secure wound closure (Table 1). For wounds with delayed closure, the average time required for wound closure was 7.8 days (range, 5–15 days). The average time to start postoperative chemotherapy and/or radiotherapy is 23.2 days (range, 18–31 days).

The mean follow-up time was 25.6 months (range, 16–36 months). Twenty patients were alive without evidence of disease, 2 patients with fibrosarcoma were alive with local recurrence and had to undergo secondary resection, 2 patients with synovial sarcoma and fibrosarcoma were alive with pulmonary metastasis, and 1 patient with fibrosarcoma died of pulmonary metastasis 25 months after surgery. All patients showed scarring according to the length of the wound. Two patients complained of mild pain and itching at the time of final follow-up. In all patients, no significant functional restriction of adjacent joints was observed and the final scars were aesthetically acceptable.

None of our patients experienced severe complications in this study. Six of 18 patients with delayed wound closure developed minor superficial wound infection. However, these wounds were closed directly after thorough debridement and completely healed after 2 weeks of dressing change. Wound edge ischemia occurred in 2 patients with immediate wound closure, which was successfully treated by adjusting the threaded rod and/or removing some sutures to reduce the tension of the wound edge. Small skin tears surrounding the wound occurred in 2 patients with the hooks directly hooked on the skin. However, none of them required any secondary surgical procedure for these small skin tears.

Discussion

STS is uncommon malignant tumor accounting for approximately 1% of all malignancies [20, 21]. Radical surgical resection is the most important approach in the curative multimodal therapy of STS [1, 20]. However, large skin-soft tissue defect coverage following tumor resection presents a challenge for reconstructive surgeons, especially for surgeons with the absence of reconstructive expertise.

Closure of large skin-soft tissue defects following tumor resection as well as the closure of large trauma or chronic wounds could often not be achieved by conventional suturing technique due to high tension [2]. The difference is that reconstruction of skin-soft tissue defects following tumor resection not only requires consideration of functional and aesthetic aspects, but also oncological factors [8, 10, 22]. Currently, many surgical methods such as skin graft, local flap, free tissue transfer procedure, tissue expansion, or various combinations of the aforementioned options have been used to close these large defects following tumor resection [68, 1013, 2226]. Of these methods, tissue expansion procedure may be a promising, cost-effective treatment option [22, 2426]. It significantly downgrades surgical complexity, reduces operating time, and shortens hospital stay. Although internal tissue expansion technique has been widely used in the cosmetology plastic surgery and obtained good clinical outcomes [27, 28], it may be greatly limited in the treatment of soft-tissue tumors, especially malignant tumors, due to long expansion period and potential stimulating effects on tumors.

Recently, external tissue expansion technique was developed with the intention to close large skin-soft tissue defects that cannot be sutured directly, reduce preoperative preparation time, and decrease complications. In 1993, Hirschowitz et al. firstly described an external skin-stretching device which harnesses the viscoelastic properties of the skin to close open wounds [13]. Since then, many modifications, such as SureClosure®, Wiseband®, and TopClosure®, have been introduced and reported by different authors and obtained acceptable outcomes [710, 12, 15, 25, 29] (Table 2). However, due to geographical, economical, or other special factors, many of them are not used in many hospitals or countries. To expand the application of this technique, we here introduced a novel, simple, and cheap skin-stretching device to close large skin-soft tissue defects following STS resection and evaluated its efficacy and safety as well as the related complications.

Table 2.

Representative skin-stretching devices for the closure of large skin-soft tissue wounds

Study Devices No. of wounds Aetiology of the wounds Manner of wound closure Follow-up (months) Complications
Kanjoor et al. [7] SureClosure 4 Trauma (50%), burns (50%)

Immediate closure, 2

Delayed closure, 2

24; 12; 12; NC Minimal scar widening, 1; compartment syndrome, 1
Barnea et al. [10] Wisebands device 22 Trauma (35%), surgery (35%), tumors (25%), burns (5%)

Immediate closure, 6

Delayed closure, 16

12 (mean) Intractable pain, 1; serious wound infection, 1
Ismavel et al. [30] Kirschner wires 10 Trauma (40%), fasciotomy wounds (30%), surgery (30%) Delayed closure, 10 9.9 (mean) Wound infection, 4; partial wire cutout, 3; partial wound dehiscence, 2
Santiago et al. [9] DermaClose RC 14 Blast-related injury Delayed closure, 12 NC Blistering, 2; maceration of the wound edges, 3
Verhaegen et al. [15] Skin Stretch system 8 Burns Immediate closure, 8 12.1 (mean) Wound infection, 2
Topaz et al. [8]

TopClosure®

TRS

3 Malignant soft tissue tumors Immediate closure, 3 6; 10; 18 Wound dehiscence, 1; blistering, 1; Scar widening, 1; limitation in ROM, 1
Aboelatta et al. [31] Home-made wound approximation device 34 Posttraumatic alopecia or ulcers (79%); post burn scars (21%)

Immediate closure, 25

Delayed closure, 9

Over a period of 48 months Wound infection, 21; wound dehiscence, 6; edge necrosis, 2; pressure necrosis, 1; adherent scars, 6
Karkos et al. [32] ETE Blomqvist 8 Trauma Delayed closure, 8 Over a period of 144  months Wound infection, 1
Current study BHS device 25 Malignant soft tissue tumors

Immediate closure, 7

Delayed closure, 18

25.6 (mean) Superficial wound infection, 6; wound edge ischemia, 2; small skin tears, 2.

NC not clear, ROM range of movement, ETE external tissue extender, BHS bidirectional regulation-hook skin closure system

Compared with many previous skin-stretching devices, the novel BHS device has many significant advantages: (a) it has a very simple design and very cheap; (b) it is made of PEEK material and stainless steel that can be sterilized and therefore can be reused; (c) it can be regulated bidirectionally, which not only ensures the closure of large skin-soft tissue wound, but also maintains appropriate tension at the edge of the wound; (d) the assisted use of KWs along the entire length of the wound edge ensures relatively uniform distribution of the stretching force across the wound edge. In addition, it is an easy-to-use device and could be applied under local anesthesia in the out-patient department in most patients with delayed wound approximation (e.g., closure of large chronic wounds), although more clinical evidence of BHS therapy is still needed.

Our results found that the novel BHS device provided the phenomenal ability of stretching the skin for immediate or delayed primary closure of large skin-soft tissue defects following STS resection. As a topical tension-relief platform, the BHS device alleviates the typical tearing and scarring, traditionally inflicted by tension sutures. In our cases, BHS therapy provides acceptable functional and cosmetic wound closure, which is comparable with many previous results [8, 31]. We would like to emphasize that undermining of the skin edges and adjacent tissues is not required for successful execution of wound closure with this method. Although undermining can provide a small additional tension decrease [13], it may result in dead space, hematoma, and skin edge necrosis.

Wound dehiscence, wound edge necrosis, and skin tears are the most reported complications [14, 30, 31] (Table 2). Studies showed that the wound approximation technique was associated with relatively high wound dehiscence (20–33%) [8, 30, 31]. However, no patient developed wound dehiscence in this small series. This might be related with the difference in defect size, wound edge tension, stretching time, and time of suture removal in different studies. Although no wound necrosis was observed in our patients, wound edge ischemia occurred in 2 patients with immediate wound closure. They were successfully treated by adjusting the threaded rod and removing some sutures to reduce the tension of the wound edge. Skin tears occurred in 2 patients with the hooks directly hooked on the skin. However, the similar problems did not arise when the KWs were used. Of note, local infection is a relatively rare complication [710, 32]. However, 6 patients were diagnosed with minor superficial wound infection in this study. This may be greatly attributed to the fact that some patients in our study had infected wounds or vaseline gauze alone was used at the early stage. On the whole, the expertise and experience of the surgeons are critical to reduce these complications.

Admittedly, this study has some limitations. The main limitation is that it had no appropriately matched control group. Another limitation of this study is that this device was only used for closure of large skin-soft tissue defects following STS resection in the present study. Hence, the indications and contraindications for BHS therapy are still unclear. A wider scope of application (such as traumatic and chronic wounds) is required to further evaluate this technique. Finally, it is limited by the small total number of patients included. More cases need to be studied before widely using this device.

Conclusion

In this study, we introduced a novel, simple, and cheap skin-stretching device BHS and presented our experiences using BHS therapy. The device is technically simple to apply and permits bidirectional control depending on the situation of the wound edges. BHS therapy facilitates the closure of large skin-soft tissue defects following STS resection, and acceptable functional and cosmetic results are obtained, without severe complications. In summary, BHS provides the surgeon an important tool for the closure of large skin-soft tissue defects, although a larger scale remains necessary to evaluate the effectiveness, indications, and complications of BHS therapy.

Acknowledgements

Not applicable

Abbreviations

STS

Soft tissue sarcoma

BHS

Bidirectional regulation-hook skin closure system

KWs

Kirschner wires

Authors’ contributions

Baichuan Wang, Qiang Wu, and Zengwu Shao contributed to the research design and manuscript writing. Yubin Li, Min Cui, and Saroj Rai carried out the data collection and statistical analysis. Ying Yang provided some advices. The authors read and approved the final manuscript.

Funding

Not applicable

Availability of data and materials

The data that support the findings of this study are available from the corresponding author upon reasonable request. The data are not publicly available due to privacy or ethical restrictions.

Ethics approval and consent to participate

This study was approved by our hospital ethical committee. All participating patients gave written informed consent.

Consent for publication

All authors have agreed with submission to WORLD J SURG ONCOL.

Competing interests

The authors declare no potential conflicts of interest.

Footnotes

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Qiang Wu and Zengwu Shao contributed equally to this work.

References

  • 1.Fujiki M, Miyamoto S, Kobayashi E, Sakuraba M, Chuman H. Early detection of local recurrence after soft tissue sarcoma resection and flap reconstruction. Int Orthop. 2016;40(9):1975–1980. doi: 10.1007/s00264-016-3219-y. [DOI] [PubMed] [Google Scholar]
  • 2.Al-Mubarak L, Al-Haddab M. Cutaneous wound closure materials: an overview and update. J Cutan Aesthet Surg. 2013;6(4):178–188. doi: 10.4103/0974-2077.123395. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3.Lohman RF, Nabawi AS, Reece GP, Pollock RE, Evans GR. Soft tissue sarcoma of the upper extremity: a 5-year experience at two institutions emphasizing the role of soft tissue flap reconstruction. Cancer-Am Cancer Soc. 2002;94(8):2256–2264. doi: 10.1002/cncr.10419. [DOI] [PubMed] [Google Scholar]
  • 4.Townley WA, Mah E, O'Neill AC, Wunder JS, Ferguson PC, Zhong T, et al. Reconstruction of sarcoma defects following pre-operative radiation: free tissue transfer is safe and reliable. J Plast Reconstr Aesthet Surg. 2013;66(11):1575–1579. doi: 10.1016/j.bjps.2013.06.029. [DOI] [PubMed] [Google Scholar]
  • 5.Wang CM, Zhang R, Luo P, Wu Z, Zheng B, Chen Y, et al. Reconstruction of extensive thoracic wall defect using the external oblique myocutaneous flap: an analysis on 20 Chinese patients with locally advanced soft tissue sarcoma. J Surg Oncol. 2018;117(2):130–136. doi: 10.1002/jso.24823. [DOI] [PubMed] [Google Scholar]
  • 6.Darrach H, Kokosis G, Bridgham K, Stone JP, Lange JR, Sacks JM. Comparison of keystone flaps and skin grafts for oncologic reconstruction: a retrospective review. J Surg Oncol. 2019;119(7):843–849. doi: 10.1002/jso.25394. [DOI] [PubMed] [Google Scholar]
  • 7.Kanjoor JR, Bang RL, Ebrahim MK. Role of skin stretching device for wound closure. Eur J Plast Surg. 2002;25(6):323–326. doi: 10.1007/s00238-002-0371-y. [DOI] [Google Scholar]
  • 8.Topaz M, Carmel NN, Topaz G, Li M, Li YZ. Stress-relaxation and tension relief system for immediate primary closure of large and huge soft tissue defects: an old-new concept: new concept for direct closure of large defects. Medicine (Baltimore) 2014;93(28):e234. doi: 10.1097/MD.0000000000000234. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Santiago GF, Bograd B, Basile PL, Howard RT, Fleming M, Valerio IL. Soft tissue injury management with a continuous external tissue expander. Ann Plast Surg. 2012;69(4):418–421. doi: 10.1097/SAP.0b013e31824a4584. [DOI] [PubMed] [Google Scholar]
  • 10.Barnea Y, Gur E, Amir A, Leshem D, Zaretski A, Shafir R, et al. Our experience with Wisebands: a new skin and soft-tissue stretch device. Plast Reconstr Surg. 2004;113(3):862–869. doi: 10.1097/01.PRS.0000105339.41838.D2. [DOI] [PubMed] [Google Scholar]
  • 11.Molea G, Schonauer F, Blasi F. Progressive skin extension: clinical and histological evaluation of a modified procedure using Kirschner wires. Br J Plast Surg. 1999;52(3):205–208. doi: 10.1054/bjps.1997.3044. [DOI] [PubMed] [Google Scholar]
  • 12.Narayanan K, Futrell JW, Bentz M, Hurwitz D. Comparative clinical study of the sure-closure device with conventional wound closure techniques. Ann Plast Surg. 1995;35(5):485–491. doi: 10.1097/00000637-199511000-00007. [DOI] [PubMed] [Google Scholar]
  • 13.Hirshowitz B, Lindenbaum E, Har-Shai Y. A skin-stretching device for the harnessing of the viscoelastic properties of skin. Plast Reconstr Surg. 1993;92(2):260–270. doi: 10.1097/00006534-199308000-00010. [DOI] [PubMed] [Google Scholar]
  • 14.Katzengold R, Topaz M, Gefen A. Tissue loads applied by a novel medical device for closing large wounds. J Tissue Viability. 2016;25(1):32–40. doi: 10.1016/j.jtv.2015.12.003. [DOI] [PubMed] [Google Scholar]
  • 15.Verhaegen PD, Bloemen MC, van der Wal MB, Vloemans AF, Tempelman FR, Beerthuizen GI, et al. Skin stretching for primary closure of acute burn wounds. Burns. 2014;40(8):1727–1737. doi: 10.1016/j.burns.2014.03.014. [DOI] [PubMed] [Google Scholar]
  • 16.Wilhelmi BJ, Blackwell SJ, Mancoll JS, Phillips LG. Creep vs. stretch: a review of the viscoelastic properties of skin. Ann Plast Surg. 1998;41(2):215–219. doi: 10.1097/00000637-199808000-00019. [DOI] [PubMed] [Google Scholar]
  • 17.Saulis AS, Lautenschlager EP, Mustoe TA. Biomechanical and viscoelastic properties of skin, SMAS, and composite flaps as they pertain to rhytidectomy. Plast Reconstr Surg. 2002;110(2):590–600. doi: 10.1097/00006534-200208000-00035. [DOI] [PubMed] [Google Scholar]
  • 18.Samis AJ, Davidson JS. Skin-stretching device for intraoperative primary closure of radial forearm flap donor site. Plast Reconstr Surg. 2000;105(2):698–702. doi: 10.1097/00006534-200002000-00034. [DOI] [PubMed] [Google Scholar]
  • 19.Verhaegen PDHM, Wal MBVD, Middelkoop E, Zuijlen PPMV. Scar assessment. In: Kamolz LP., Jeschke M.G., Horch R.E., Küntscher M., Brychta P. (eds). Hand book of burns. 2012:69-89.
  • 20.Vinals JM, Rodrigues TA, Sildenikova DP, Payro JM, Porte JA, Sune CH, et al. Indications of microsurgery in soft tissue sarcomas. J Reconstr Microsurg. 2012;28(9):619–625. doi: 10.1055/s-0032-1326740. [DOI] [PubMed] [Google Scholar]
  • 21.Siegel RL, Miller KD, Jemal A. Cancer statistics, 2019. CA Cancer J Clin. 2019;69(1):7–34. doi: 10.3322/caac.21551. [DOI] [PubMed] [Google Scholar]
  • 22.Cheng LF, Lee JT, Hsu H, Wu MS. Simple skin-stretching device in assisted tension-free wound closure. Ann Plast Surg. 2017;78(3 Suppl 2):S52–S57. doi: 10.1097/SAP.0000000000001006. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Hanada M, Kadota H, Yoshida S, Takeuchi N, Okada T, Matsumoto Y, et al. Large-defect resurfacing: a comparison of skin graft results following sarcoma resection and traumatic injury repair. Wounds. 2019;31(7):184–192. [PubMed] [Google Scholar]
  • 24.Lei Y, Liu L, Du SH, Zong ZW, Zhang LY, Guo QS. The use of a skin-stretching device combined with vacuum sealing drainage for closure of a large skin defect: a case report. J Med Case Rep. 2018;12(1):264. doi: 10.1186/s13256-018-1779-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Huahui Z, Dan X, Hongfei J, Hang H, Chunmao H, Haitao R, et al. Evaluation of a new tension relief system for securing wound closure: a single-Centre, Chinese cohort study. Plast Surg (Oakv) 2016;24(3):177–182. doi: 10.1177/229255031602400306. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Wooten KE, Ozturk CN, Ozturk C, Laub P, Aronoff N, Gurunluoglu R. Role of tissue expansion in abdominal wall reconstruction: a systematic evidence-based review. J Plast Reconstr Aesthet Surg. 2017;70(6):741–751. doi: 10.1016/j.bjps.2017.02.018. [DOI] [PubMed] [Google Scholar]
  • 27.Manconi A, De Lorenzi F, Chahuan B, Berrino V, Berrino P, Zucca-Matthes G, et al. Total breast reconstruction with fat grafting after internal expansion and expander removal. Ann Plast Surg. 2017;78(4):392–396. doi: 10.1097/SAP.0000000000000833. [DOI] [PubMed] [Google Scholar]
  • 28.Kim BJ, Lee Y. Use of scar and capsule flap for preventing the contour deformity after applying tissue expansion in patients with scar. J Craniofac Surg. 2014;25(2):e151–e154. doi: 10.1097/SCS.0000000000000412. [DOI] [PubMed] [Google Scholar]
  • 29.Song M, Zhang Z, Liu T, Liu S, Li G, Liu Z, et al. EASApprox((R)) skin-stretching system: a secure and effective method to achieve wound closure. Exp Ther Med. 2017;14(1):531–538. doi: 10.3892/etm.2017.4539. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Ismavel R, Samuel S, Boopalan PR, Chittaranjan SB. A simple solution for wound coverage by skin stretching. J Orthop Trauma. 2011;25(3):127–132. doi: 10.1097/BOT.0b013e318206f556. [DOI] [PubMed] [Google Scholar]
  • 31.Aboelatta YA, Elshahm A, Saleh MA, Kamel IH, Aly HM. Intraoperative and delayed wound approximation in closure of skin defects in different areas. J Wound Care. 2015;24(12):600–605. doi: 10.12968/jowc.2015.24.12.600. [DOI] [PubMed] [Google Scholar]
  • 32.Karkos CD, Papoutsis I, Giagtzidis IT, Pliatsios I, Mitka MA, Papazoglou KO, et al. Management of postfasciotomy wounds and skin defects following complex vascular trauma to the extremities using the external tissue extender system. Int J Low Extrem Wounds. 2018;17(2):113–119. doi: 10.1177/1534734618765343. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request. The data are not publicly available due to privacy or ethical restrictions.


Articles from World Journal of Surgical Oncology are provided here courtesy of BMC

RESOURCES