Abstract
Three-dimensional (3D) spheroidal cell cultures are now recognised as better models of cancers as compared to traditional cell cultures. However, established 3D cell culturing protocols and techniques are time-consuming, manually laborious and often expensive due to the excessive consumption of reagents. Microfluidics allows for traditional laboratory-based biological experiments to be scaled down into miniature custom fabricated devices, where cost-effective experiments can be performed through the manipulation and flow of small volumes of fluid. In this study, we characterise a 3D cell culturing microfluidic device fabricated from a 3D printed master. HT29 cells were seeded into the device and 3D spheroids were generated and cultured through the perfusion of cell media. Spheroids were treated with 5-Fluorouracil for five days through continuous perfusion and cell viability was analysed on-chip at different time points using fluorescence microscopy and Lactate dehydrogenase (LDH) assay on the supernatant. Increasing cell death was observed in the HT29 spheroids over the five-day period. The 3D cell culturing microfluidic device described in this study, permits on-chip anti-cancer treatment and viability analysis, and forms the basis of an effective platform for the high-throughput screening of anti-cancer drugs in 3D tumour spheroids.
Subject terms: Cancer, Cell biology, Physiology, Engineering, Nanoscience and technology
Introduction
The interest in and practicality of microfluidic systems for biochemical analysis of living cells and tissues has dramatically increased over the past decade1,2. ‘Lab-on-a-chip’ technologies carry many advantages, including the convenience of culturing cells and performing analytical techniques on small microfabricated platforms. Consequently, these devices consume only a fraction of the amount of reagents, tissue culturing media and precious biological specimens as do conventional in vitro experiments that are carried out in large commercial plates and flasks3. Another advantage of microfluidic devices for biomedical research is their ability to streamline complex assay protocols4. For instance, fluid flow enables the culture and treatment of cells to be spatially and temporally controlled with a great degree of accuracy5. Specifically for anti-cancer research, microfluidic systems have the potential to quickly and easily screen drugs in real time and diagnose disease6.
Concurrent with the increasing practical application of microfluidics in the biomedical sciences, three-dimensional (3D) cell culturing is increasingly being adopted in anti-cancer research, especially for the purpose of drug development. In comparison to conventional two-dimensional (2D) monolayer cell cultures, 3D spheroids are recognised as better in vitro models of cancer due their inherent properties, making them more comparable to tumour in vivo7. Cells in 2D cultures are uniform in their rate of proliferation, access to oxygen and nutrients and response to treatment, whereas the architectural arrangement of cells in 3D spheroids, together with the physiological gradients of oxygen, nutrients, growth factors and catabolites results in the uneven growth and diverse survival rates of cells. In addition, the layered structure of cells influences the ability of anti-cancer therapeutics to penetrate through and induce cytotoxicity in the different regions of the spheroid. Cells cultured in a 3D structure also experience different mechanical and topographical forces, than those growing on a 2D planar surface8. The overall benefit of 3D spheroidal models is that they offer a better understanding of in vivo molecular mechanisms involved in tumour growth and drug resistance9.
Taken together, 3D spheroids offer a better representation of in vivo environments and microfluidic devices offer convenient and cost-effective platforms for high-throughput screening applications. A common limitation among established 3D cell culturing techniques is the manual effort required to grow and maintain 3D spheroid cultures in non-specialised conventional commercial tissue culturing equipment. An appropriately designed microfluidic platform can improve the success rate of establishing and characterising 3D spheroids by reducing the amount of effort required for cell culturing, administering cytotoxic treatments and evaluating the outcomes of treatment—conveniently within a single device5,10–16.
In this article, we describe the development and application of a 3D cell culturing microfluidic flow platform and the use of 3D printing to create a master from which polydimethylsiloxane (PDMS) flow chips can be conveniently cast. We demonstrate the platform by culturing 3D spheroids, treating with the anti-cancer drug, 5-Fluorouracil (5-FU) and evaluating cytotoxicity through flow. We show that the 3D cell culturing microfluidic platform we have developed can accommodate the generation of 3D spheroids from cell aggregates ‘on-chip’, administer cytotoxic treatment under limited- and continuous perfusion conditions and perform quantitative and qualitative cell viability analysis.
Results
The microfluidic device described in this study utilises the non-cell adherent properties of the PDMS flow chip located between two layers of Poly(methyl methacrylate) (PMMA), to generate 3D spheroids (Fig. 1A)17. Placing a cell seeding port directly above the cell culturing wells was vital to improving the efficiency of introducing cells to the wells. The port was added after it was identified that perfusing a HT29 cell suspension through the device to seed the wells was not effective (Fig. 2A,B). Seeding cells through direct deposition via the cell seeding port, results in 3D spheroids forming in 100% of the wells, as compared to 68% using flow-driven cell seeding (p = 0.02) (Fig. 2C). An equivalent number of HT29 cells, (mean = 377 cells, min. = 361 cells, max. = 386 cells), were deposited into each well to generate 3D spheroids (Fig. 2D). The central region of the PDMS flow chip was designed with a 5 × 5 array of concave-shaped spheroid culturing wells (Fig. 1B). The concave-shaped wells were designed to aid in the cell seeding process, by assisting cells in initially forming cell aggregates, which is vital for successfully generating 3D spheroids (Fig. 3).
As depicted in Fig. 4A, after seeding the HT29 cells into the wells of the PDMS flow chip, cell aggregates formed within a day (Day 0) and subsequent culturing under static conditions, resulted in 3D spheroids forming by Day 2. The HT29 spheroids then continued to mature in density, roundness in spheroidal structures and grew in overall size. A 148% (p < 0.0001), 89% (p < 0.0001) and 58% (p < 0.0001) increase in spheroid volumes were found between Day 2 versus Day 4, Day 4 versus Day 7 and Day 7 versus Day 10 respectively (Fig. 4B). Following 10 days of culture, scanning electron microscopy confirmed the spheroidal structures of the HT29 spheroids (Fig. 4C).
Using the microfluidic platform, further experiments were carried out to highlight the (1) administration of an anti-cancer drug to 3D HT29 spheroids and (2) evaluate cytotoxicity induced by the anti-cancer drug through the microfluidic device. For limited-perfusion experiments, a single syringe pump was connected to the microfluidic devices and was used to drive flow (Fig. 5A,B). Briefly, 3D HT29 spheroids were generated on chip and subjected to 5-FU treatment (500 µM) through perfusion. After 24 h under non-flow static conditions, cell viability analysis showed that 5-FU treated HT29 spheroids were 70% less viable, as compared to untreated spheroids (p = 0.0001) (Fig. 5C). Fluorescent imaging highlighting cell death confirmed cytotoxicity in 5-FU treated spheroids, as indicated by the uptake of propidium iodide and lack of dye uptake in untreated HT29 spheroids (Fig. 5D). Quantifying propidium iodide fluorescence also confirmed cytotoxicity in 5-FU treated spheroids with no toxicity detected in untreated spheroids (p < 0.0001) (Fig. 5E). In addition, a 20% reduction in Hoechst 33342 fluorescence in treated spheroids, as compared to untreated spheroids (p = 0.028) indicated the loss and shedding of cells following 5-FU treatment (Fig. 5E).
Microfluidic device-cultured HT29 spheroids were also subjected to 5-FU treatment (200 µM) through continuous perfusion. The microfluidic devices were connected in parallel to reagent-loaded syringes through a peristaltic pump (Fig. 6A,B). Cell viability was monitored over 5 days and a time dependant reduction in cell viability was observed (p < 0.0001) (Day 0 = 100% cell viability and Day 5 = 15% cell viability, p < 0.0001) (Fig. 6C). Fluorescent imaging confirmed results of the LDH assay, as indicated by the increasing uptake of propidium iodide over time (Fig. 6D). In addition, quantifying propidium iodide fluorescence also showed significant cytotoxicity over the 5 days (Day 0 vs. Day 5: p < 0.0001) (Fig. 6E). A 16% reduction in Hoechst 33342 fluorescence was also observed at Day 5, as compared to Day 0, indication the loss and shedding of cells (p = 0.0003) (Fig. 6E).
Discussion
Three-dimensional in vitro spheroidal models of cancer are recognised as more accurate representations of the tumour microenvironment as compared to 2D monolayers, and responses to treatment in 3D cultured cells are more reflective of responses observed in vivo8. Combined microfluidics and 3D spheroid culturing has allowed conventional laborious 3D cell culturing protocols to be scaled down onto custom micro-fabricated devices. This allows the generation, culturing, treatment and analysis of 3D spheroids to be performed conveniently on-chip through fluid perfusion18–23, with the additional practical advantages of reduced experimental and consumables costs and reduced manual labour23.
The 3D cell culturing microfluidic device described in this paper, builds upon previous work and is based upon the ‘non-cell adherent’ technique for generating 3D spheroids10,24–27. Cells are seeded on a non-adherent substrate, preventing them from adhering to the surface and instead forming cell aggregates26. The PDMS flow chip used for our microfluidic device is a non-cell adherent substrate, biocompatible, facilitates the exchange of gases and has been used extensively for fabricating 3D cell culturing microfluidic platforms5,11,12,14–16,18–21,27–41. SU-8 based microfluidic platforms are also commonly used for 3D cell culturing systems, however the chemical treatment of SU-8 to produce hydrophobic surfaces requires dangerous and toxic chemicals to be used42. We employed 3D printing and PDMS casting, which does not require surface treatment or bonding to other materials, making the fabrication of the flow chips inherently simple. The flow chips were cast from a mould designed with concave-shaped spheroid-culturing wells to facilitate the aggregation of cells and assist in the generation of 3D spheroids, as has been described previously for 3D cell culturing platforms43,44. Due to the nature of the photocuring method for forming the 3D prints, the master mould is built in layers forming ‘steps-like’ structures as an approximation to the convex well to make the 3D cell culturing wells (Fig. 7). These structures were nevertheless still successful in generating spheroids in the polygonal concave-shaped wells. Our method for making 3D spheroid culturing chips is effective, yet simple enough to allow biologists and non-microfluidics specialists to produce and operate these chips in their laboratories without the need of a cleanroom or delicate microfabrication/assembly steps. The PDMS-based spheroid culturing chip was designed to accommodate the culture of twenty-five spheroids (Fig. 1B). As a prototype design, we limited culturing to twenty-five spheroids, yet retain reliability and simplicity in generating 3D spheroids, as compared to conventional 96-well plates. However, the advantage of our design is such that the 5 × 5 array of spheroid-culturing wells, could easily be expanded to accommodate the simultaneous culturing of larger numbers of 3D spheroids.
A limitation of previously reported 3D cell culturing microfluidic platforms is that they have relied on external 3D cell culturing methods, such as the use of ultra-low attachment plates or the hanging drop method, to generate spheroids off-chip before being introduced into the microfluidic platforms5,13,35,36,40. The microfluidic platform described in this study eliminates the need for generating 3D spheroids beforehand and allows non-aggregated cells to be seeded directly into the device, with equivalent numbers of cells seeded into each well (Fig. 2D). We found the seeding of cells through flow, to be inefficient in generating spheroids. For biomedical applications, the use of a pipette to seed cells in an exact location on a microfluidic chip is an advantage when accurately producing spheroids. This user-friendly approach is also compatible in conventional biology laboratories. For our experiments, the human colorectal adenocarcinoma cell line, HT29, was used to generate 3D spheroids. HT29 has been used extensively in 3D cell culturing based studies, and from previous multiple studies, we have found HT29 to form spheroidal cultures with relative ease45–48. As demonstrated in Figs. 2A and 8, we found the microfluidic platform presented here, enables HT29 cells to form uniformly sized spheroidal structures. 3D spheroids grew to a maximum of around 250 μm in diameter, dictated by the diameters of the spheroid-culturing wells. Typically, HT29 3D spheroids with ≥ 300 μm diameters, possess hypoxic and necrotic cores with physiological gradients of cell proliferation (outer rapidly proliferating and inner non-proliferating cells)49. It could therefore be postulated, that the HT29 3D spheroids grown in this study did not possess the in vivo like physiological gradients or hypoxic/necrotic cores.
Our device was additionally designed to streamline high-throughput chemosensitivity experiments and improve the overall efficiency of 3D cell culturing. To demonstrate this, we used the chemotherapeutic agent, 5-Fluorouracil, to treat the HT29 spheroids. As 5-FU is currently the primary anti-cancer drug used to treat colorectal cancers50, it makes for an excellent model system and demonstrates the relevant real-world application of our device. Following the culturing of 3D spheroids, treatment and analysis could be performed conveniently on-chip, through the perfusion of anti-cancer drugs and other culture media. Fluorescent imaging (qualitative analysis) using propidium iodide to visually highlight cell death in spheroids was complimented by the leakage and quantification of LDH in the supernatant (quantitative analysis), an indicator of cell death. Our study makes use of a single 3D cell culturing microfluidic platform, to administer an anti-cancer treatment and permit subsequent on-chip viability imaging confirmed by an LDH assay.
In our study, we demonstrated two different setups for performing chemosensitivity experiments in the microfluidic devices. The first utilised a single syringe pump to drive flow through one device at a time (Fig. 5). This limited-perfusion setup permits simple experiments to be conducted with a single experimental endpoint. However, experiments can also be performed with continuous flow as demonstrated by the second setup utilising a multichannel pump (Fig. 6). An advantage of continuous flow setup is that it better models the physiological flow of systemic circulation in the body. This second setup also enables multiple devices to be run simultaneously in parallel. Each channel is controlled independently, therefore flow through one microfluidic device can be switched off without disrupting flow through the rest of the channels or devices. In addition, the continuous flow setup was designed to be placed into a conventional tissue culturing incubator, permitting prolonged experiments (lasting multiple days) to be performed using existing laboratory infrastructure and environmental standards.
Conclusion
In this study, we have developed and evaluated a 3D cell culturing microfluidic device. The device comprises a layer of PDMS with concave 3D cell culturing wells sandwiched between two layers of transparent PMMA, allowing for direct visualization. HT29 colorectal cancer cells in suspension were seeded into the wells through the central port on the top PMMA layer and subsequently formed into 3D spheroids. These spheroids were treated through one-time or continuous perfusion of the anti-cancer drug, 5-Fluorouracil, and the resulting cell viability was analysed using fluorescent microscopy and an LDH assay. The two perfusion regimes simulated different physiological flow conditions and enabled the evaluation of distinct methods of administering anti-cancer treatment. Finally, the microfluidic device described here offers a convenient, efficient and quick platform for generating, culturing, treating and analysing 3D cancer spheroids and will facilitate high-throughput screening of anti-cancer drugs on complex 3D in vitro models.
Methods
Designing and fabricating the microfluidic devices
A single microfluidic device consists of a central PDMS chip consisting of a single straight flow channel joining the inlet to outlet, with an array of 3D cell culturing wells located towards the centre. This is sandwiched between two layers of transparent PMMA allowing for direct optimal imaging. Schematic designs for the different parts of the flow device were drawn in SolidWorks (Dassault Systèmes, Vélizy-Villacoublay, France). To create the PDMS flow chips, the inverse design was printed as a reusable master mould on an EnvisionTEC 3D printer (EnvisionTEC GmbH, Gladbeck, Germany) in a photocurable acrylonitrile butadiene styrene (ABS) thermoplastic polymer. The well-forming hemispherical “pimples” in the master mould are 250 µm in diameter located in 5 × 5 array, on a pitch of 500 µm × 500 µm (Fig. 7). Chips were then cast from the master mould using PDMS with a release agent to aid removal. The PDMS flow chips were fixed between two layers of PMMA. The top layer of PMMA was fabricated with fluid inlet and outlet ports and a cell seeding port in the centre, while the bottom layer of PMMA serves as the device base (Fig. 1). The PMMA layers were machined from cast sheet, with inlet and outlet ports positioned to match the ends of the PDMS chip and machined to take standard UNF 1/4-28 flangeless fittings. An additional port was machined into the top layer of PMMA, located directly above the 3D cell culturing wells. The microfluidic device was fashioned together using four bolts, one in each corner, to hold the assembly together. Inlet and outlet ports were connected to tubing using flangeless fittings and a blanking plug was fitted to the central port to allow direct access to the cell culturing wells (Fig. 1). Following assembly, the microfluidic device was sterilised using a steam and high-pressured autoclave prior to the addition of cells.
Preparation of cancer cells
The human colorectal adenocarcinoma cell line, HT29, was obtained from the European Collection of Authenticated Cell Cultures (Salisbury, UK). Cells were cultured in Roswell Park Memorial Institute (RPMI) 1640 Medium plus GlutaMAX (Gibco by Life Technologies, Paisley, UK) (without sodium pyruvate) supplemented with 10% (v/v) Fetal Bovine Serum (FBS) (Sigma-Aldrich) and maintained at 37 °C, 5% CO2 and 95% relative humidity (RH). Cell cultures were passaged weekly and seeded into 75 cm2 tissue culture flasks (Corning Inc., New York, USA). Complete cell media was changed once per week. 80–90% confluent cell cultures were used for the experiments.
Culturing 3D tumour spheroids in a microfluidic device
Cells were washed and trypsinised to create a cell suspension of 2 × 107 cells/mL in cell medium. 20 μL of cell suspension was deposited into the wells of assembled PDMS chips. The chips were then incubated and maintained under static conditions at 37 °C/5% CO2/95% RH. All cell media was changed daily using a pipette. Manual cell counting and acquisition of the images of spheroids was performed using the EVOS FL Imaging System (Life Technologies). For scanning electron microscopy (SEM), spheroids were cultured for 10 days then briefly washed with DPBS, fixed in 2.5% (v/v) glutaraldehyde for 20 min and subjected to increasing gradients of ethanol dehydration (25%, 40%, 60%, 80%, 90% and 100%). The top surface of the PDMS flow chip was then coated with 4 nm of iridium and imaged using a FEI Nova NanoSEM450 scanning electron microscope (Thermo Fisher).
Anti-cancer treatment of 3D tumour spheroids in microfluidic devices
Culturing 3D HT29 spheroids
HT29 cells in suspension were deposited through the central port of the top PMMA layer into the wells of the PDMS flow chip in an assembled microfluidic device. Cell media was then flowed through the microfluidic device at 10 µL/min using an Aladdin Single-Syringe Pump (AL-2000, World Precision Instruments, Florida, USA). Following complete cell media perfusion through the device, the device was incubated (37 °C/5% CO2/95% RH) under static conditions.
5-FU treatment (Limited-perfusion conditions)
Two different flow conditions were used to study chemosensitivity in 3D HT29 spheroids. For the first (limited-perfusion) condition, 48 h after seeding cells, 500 µM 5-FU was prepared in cell media and perfused through the microfluidic device at 20 µL/min using the single-syringe pump. This approach can only accommodate one device at a time. Following the perfusion of 500 µL of 5-FU (20 µL/min) completely through the microfluidic device, lasting 25 min, flow was stopped and the device was detached from the pump and placed back into incubation (37 °C/5% CO2/95% RH). The spheroids were incubated with 5-FU under static conditions for an additional 24 h.
5-FU treatment (Continuous perfusion conditions)
For the second (continuous perfusion) condition, 48 h after seeding cells, 250 µM 5-FU was prepared in cell media and perfused continuously through five microfluidic devices in parallel at 20 µL/min for 5 days. An Ismatec IPC Low-Speed Digital Peristaltic Pump (Cole-Parmer Instrument Company Ltd, St. Neots, UK) with 8 independent channels, was used to drive continuous fluid flow. Microfluidic devices under continuous perfusion were kept in incubation (37 °C/5% CO2/95% RH) for the duration of the experiments.
Evaluating cytotoxicity in 3D tumour spheroids
Lactate dehydrogenase (LDH) assay
Spent cell culture media supernatant was collected from the microfluidic device and stored at − 20 °C until the end of the continuous flow 5-FU treatment period. The LDH cytotoxicity assay was performed using the Pierce LDH Cytotoxicity Assay Kit (Thermo Fisher) and carried out as per the manufacturer’s instructions and guidelines. This method has been previously applied to evaluate the cell viability of tumour tissue specimens cultured on a microfluidic platform51–53.
Fluorescent imaging
5 μg/mL Hoechst 33342 and 1.3 μg/mL Propidium Iodide was prepared in cell media, and flowed into the device at 20 µL/min. Spheroids were incubated with the dyes for 30 min, washed with DPBS through flow and imaged using the EVOS FL Imaging System (Life Technologies). Imaging was carried out directly on-chip. We note that the spheroids were relatively well held in the individual wells. This is ideal when moving between different laboratory and measurement platforms.
Image and statistical analysis
ImageJ (National Institute of Health, Maryland, USA) was used to analyse the transillumination and fluorescent images of 3D spheroids. The student t-test and one-way ANOVA were used to perform data statistical analyses using GraphPad Prism 8 (GraphPad Software, Inc., California, USA). p < 0.05 was considered to be statistically significant. Data is presented as the mean ± standard deviation.
Acknowledgements
We would like to acknowledge and thank Matthew Bourn and Sally A. Peyman (School of Physics and Astronomy, University of Leeds) for making the PDMS layers of the microfluidic devices. We would also like to thank Alexander Kulak (School of Chemistry, University of Leeds) for assisting in preparing samples and taking the SEM images.
Author contributions
M.I.K., N.K. and D.G.J. conceptualised and designed the research project. M.I.K., M.A.L., G.d.B., N.K. and D.G.J. were involved in designing, optimising and fabricating the microfluidic devices. M.I.K., G.A., T.M., H.A. and S.L.P carried out the experiments and data analysis. M.I.K., M.A.L., H.S.S., N.K. and D.G.J wrote, edited and revised the manuscript. All authors approve the final version of the manuscript.
Funding
This work was funded by a University of Leeds 110 Anniversary Research Scholarship and Wellcome Trust Institutional Strategic Support Fund Fellowship to M.I.K. This work was supported by the National Institute for Health Research (NIHR) infrastructure at Leeds. The views expressed are those of the authors and not necessarily those of the National Health Service, the NIHR, or the Department of Health. It was also supported by a GSK/Royal Academy of Engineering Research Chairmanship to N.K. and an NIHR Research Professorship and NIHR Senior Investigator Award to D.G.J.
Competing interests
The authors declare no competing interests.
Footnotes
Publisher's note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
References
- 1.Whitesides G. The lab finally comes to the chip. Lab Chip. 2014;14:3125. doi: 10.1039/c4lc90072c. [DOI] [PubMed] [Google Scholar]
- 2.Kovarik ML, et al. Micro total analysis systems for cell biology and biochemical assays. Anal. Chem. 2012;84:516–540. doi: 10.1021/ac202611x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Vadivelu R, Kamble H, Shiddiky M, Nguyen N-T. Microfluidic technology for the generation of cell spheroids and their applications. Micromachines. 2017;8:94. [Google Scholar]
- 4.Sackmann EK, Fulton AL, Beebe DJ. The present and future role of microfluidics in biomedical research. Nature. 2014;507:181–189. doi: 10.1038/nature13118. [DOI] [PubMed] [Google Scholar]
- 5.Frey O, Misun PM, Fluri DA, Hengstler JG, Hierlemann A. Reconfigurable microfluidic hanging drop network for multi-tissue interaction and analysis. Nat. Commun. 2014;5:1–11. doi: 10.1038/ncomms5250. [DOI] [PubMed] [Google Scholar]
- 6.Valente KP, et al. Microfluidic technologies for anticancer drug studies. Drug Discov. Today. 2017;22:1654–1670. doi: 10.1016/j.drudis.2017.06.010. [DOI] [PubMed] [Google Scholar]
- 7.Breslin S, O’Driscoll L. Three-dimensional cell culture: the missing link in drug discovery. Drug Discov. Today. 2013;18:240–249. doi: 10.1016/j.drudis.2012.10.003. [DOI] [PubMed] [Google Scholar]
- 8.Edmondson R, Broglie JJ, Adcock AF, Yang L. Three-dimensional cell culture systems and their applications in drug discovery and cell-based biosensors. Assay Drug Dev. Technol. 2014;12:207–218. doi: 10.1089/adt.2014.573. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Pickl M, Ries CH. Comparison of 3D and 2D tumor models reveals enhanced HER2 activation in 3D associated with an increased response to trastuzumab. Oncogene. 2009;28:461–468. doi: 10.1038/onc.2008.394. [DOI] [PubMed] [Google Scholar]
- 10.Patra B, Peng C-C, Liao W-H, Lee C-H, Tung Y-C. Drug testing and flow cytometry analysis on a large number of uniform sized tumor spheroids using a microfluidic device. Sci. Rep. 2016;6:21061. doi: 10.1038/srep21061. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Patra B, et al. A microfluidic device for uniform-sized cell spheroids formation, culture, harvesting and flow cytometry analysis. Biomicrofluidics. 2013;7:1–11. doi: 10.1063/1.4824480. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Shin CS, Kwak B, Han B, Park K. Development of an in Vitro 3D tumor model to study therapeutic efficiency of an anticancer drug. Mol. Pharm. 2013;10:2167–2175. doi: 10.1021/mp300595a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Ayuso JM, et al. Study of the chemotactic response of multicellular spheroids in a microfluidic device. PLoS ONE. 2015;10:1–16. doi: 10.1371/journal.pone.0139515. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Aijian AP, Garrell RL. Digital microfluidics for automated hanging drop cell spheroid culture. J. Lab. Autom. 2015;20:283–295. doi: 10.1177/2211068214562002. [DOI] [PubMed] [Google Scholar]
- 15.McMillan KS, Boyd M, Zagnoni M. Transitioning from multi-phase to single-phase microfluidics for long-term culture and treatment of multicellular spheroids. Lab Chip. 2016;16:3548–3557. doi: 10.1039/c6lc00884d. [DOI] [PubMed] [Google Scholar]
- 16.Sabhachandani P, et al. Generation and functional assessment of 3D multicellular spheroids in droplet based microfluidics platform. Lab Chip. 2016;16:497–505. doi: 10.1039/c5lc01139f. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Zhang W, Choi DS, Nguyen YH, Chang J, Qin L. Studying cancer stem cell dynamics on PDMS surfaces for microfluidics device design. Sci. Rep. 2013;3:2332. doi: 10.1038/srep02332. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Sun D, Lu J, Chen Z, Yu Y, Li Y. A novel three-dimensional microfluidic platform for on chip multicellular tumor spheroid formation and culture. Microfluid. Nanofluidics. 2014;17:831–842. [Google Scholar]
- 19.Ziółkowska K, et al. Long-term three-dimensional cell culture and anticancer drug activity evaluation in a microfluidic chip. Biosens. Bioelectron. 2013;40:68–74. doi: 10.1016/j.bios.2012.06.017. [DOI] [PubMed] [Google Scholar]
- 20.Kwapiszewska K, Michalczuk A, Rybka M, Kwapiszewski R, Brzózka Z. A microfluidic-based platform for tumour spheroid culture, monitoring and drug screening. Lab Chip. 2014;14:2096–2104. doi: 10.1039/c4lc00291a. [DOI] [PubMed] [Google Scholar]
- 21.Chen Y-C, Lou X, Zhang Z, Ingram P, Yoon E. High-throughput cancer cell sphere formation for characterizing the efficacy of photo dynamic therapy in 3D cell cultures. Sci. Rep. 2015;5:12175. doi: 10.1038/srep12175. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Zuchowska A, Jastrzebska E, Chudy M, Dybko A, Brzozka Z. 3D lung spheroid cultures for evaluation of photodynamic therapy (PDT) procedures in microfluidic Lab-on-a-Chip system. Anal. Chim. Acta. 2017;990:110–120. doi: 10.1016/j.aca.2017.07.009. [DOI] [PubMed] [Google Scholar]
- 23.Khot MI, Levenstein MA, Kapur N, Jayne DG. A review on the recent advancement in “tumour spheroids-on-a-chip”. J. Cancer Res. Pract. 2019;6:55–63. [Google Scholar]
- 24.Lim JH, Kim DE, Kim EJ, Ahrberg CD, Chung BG. Functional graphene oxide-based nanosheets for photothermal therapy. Macromol. Res. 2018;26:557–565. doi: 10.1007/s13233-018-6067-3. [DOI] [Google Scholar]
- 25.Moshksayan K, et al. Spheroids-on-a-chip: recent advances and design considerations in microfluidic platforms for spheroid formation and culture. Sens. Actuators B Chem. 2018;263:151–176. [Google Scholar]
- 26.Costa EC, de Melo-Diogo D, Moreira AF, Carvalho MP, Correia IJ. Spheroids formation on non-adhesive surfaces by liquid overlay technique: considerations and practical approaches. Biotechnol. J. 2018;13:1–12. doi: 10.1002/biot.201700417. [DOI] [PubMed] [Google Scholar]
- 27.Okuyama T, et al. Preparation of arrays of cell spheroids and spheroid-monolayer cocultures within a microfluidic device. J. Biosci. Bioeng. 2010;110:572–576. doi: 10.1016/j.jbiosc.2010.05.013. [DOI] [PubMed] [Google Scholar]
- 28.Yu L, Chen MCW, Cheung KC. Droplet-based microfluidic system for multicellular tumor spheroid formation and anticancer drug testing. Lab Chip. 2010;10:2424. doi: 10.1039/c004590j. [DOI] [PubMed] [Google Scholar]
- 29.Liu T, Lin B, Qin J. Carcinoma-associated fibroblasts promoted tumor spheroid invasion on a microfluidic 3D co-culture device. Lab Chip. 2010;10:1671. doi: 10.1039/c000022a. [DOI] [PubMed] [Google Scholar]
- 30.Jin HJ, et al. A multicellular spheroid formation and extraction chip using removable cell trapping barriers. Trans. Korean Soc. Mech. Eng. A. 2011;35:131–134. doi: 10.1039/c0lc00134a. [DOI] [PubMed] [Google Scholar]
- 31.Lee S-A, et al. Spheroid-based three-dimensional liver-on-a-chip to investigate hepatocyte–hepatic stellate cell interactions and flow effects. Lab Chip. 2013;13:3529. doi: 10.1039/c3lc50197c. [DOI] [PubMed] [Google Scholar]
- 32.Toley BJ, Tropeano Lovatt ZG, Harrington JL, Forbes NS. Microfluidic technique to measure intratumoral transport and calculate drug efficacy shows that binding is essential for doxorubicin and release hampers Doxil. Integr. Biol. 2013;5:1184. doi: 10.1039/c3ib40021b. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Yoon S, Kim JA, Lee SH, Kim M, Park TH. Droplet-based microfluidic system to form and separate multicellular spheroids using magnetic nanoparticles. Lab Chip. 2013;13:1522. doi: 10.1039/c3lc41322e. [DOI] [PubMed] [Google Scholar]
- 34.Chan HF, et al. Rapid formation of multicellular spheroids in double-emulsion droplets with controllable microenvironment. Sci. Rep. 2013;3:3462. doi: 10.1038/srep03462. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Albanese A, Lam AK, Sykes EA, Rocheleau JV, Chan WCW. Tumour-on-a-chip provides an optical window into nanoparticle tissue transport. Nat. Commun. 2013;4:2718. doi: 10.1038/ncomms3718. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Ruppen J, et al. A microfluidic platform for chemoresistive testing of multicellular pleural cancer spheroids. Lab Chip. 2014;14:1198–1205. doi: 10.1039/c3lc51093j. [DOI] [PubMed] [Google Scholar]
- 37.Wang Y, Wang J. Mixed hydrogel bead-based tumor spheroid formation and anticancer drug testing. Analyst. 2014;139:2449–2458. doi: 10.1039/c4an00015c. [DOI] [PubMed] [Google Scholar]
- 38.Yu L, Ni C, Grist SM, Bayly C, Cheung KC. Alginate core-shell beads for simplified three-dimensional tumor spheroid culture and drug screening. Biomed. Microdevices. 2015;17:33. doi: 10.1007/s10544-014-9918-5. [DOI] [PubMed] [Google Scholar]
- 39.Prabhakarpandian B, et al. Synthetic tumor networks for screening drug delivery systems. J. Control. Release. 2015;201:49–55. doi: 10.1016/j.jconrel.2015.01.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Kim J-YJ-Y, Fluri DA, Kelm JM, Hierlemann A, Frey O. 96-Well format-based microfluidic platform for parallel interconnection of multiple multicellular spheroids. J. Lab. Autom. 2015;20:274–282. doi: 10.1177/2211068214564056. [DOI] [PubMed] [Google Scholar]
- 41.Jeong SY, Lee JH, Shin Y, Chung S, Kuh HJ. Co-culture of tumor spheroids and fibroblasts in a collagen matrix-incorporated microfluidic chip mimics reciprocal activation in solid tumor microenvironment. PLoS ONE. 2016;11:1–17. doi: 10.1371/journal.pone.0159013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Ayuso JM, et al. SU-8 based microdevices to study self-induced chemotaxis in 3D microenvironments. Front. Mater. 2015;2:37. [Google Scholar]
- 43.Lee D, Pathak S, Jeong J-H. Design and manufacture of 3D cell culture plate for mass production of cell-spheroids. Sci. Rep. 2019;9:13976. doi: 10.1038/s41598-019-50186-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Zhang B, et al. Fabrication of agarose concave petridish for 3D-culture microarray method for spheroids formation of hepatic cells. J. Mater. Sci. Mater. Med. 2018;29:49. doi: 10.1007/s10856-018-6058-0. [DOI] [PubMed] [Google Scholar]
- 45.Fan X, Ouyang N, Teng H, Yao H. Isolation and characterization of spheroid cells from the HT29 colon cancer cell line. Int. J. Colorectal Dis. 2011;26:1279–1285. doi: 10.1007/s00384-011-1248-y. [DOI] [PubMed] [Google Scholar]
- 46.Courau T, et al. Cocultures of human colorectal tumor spheroids with immune cells reveal the therapeutic potential of MICA/B and NKG2A targeting for cancer treatment. J. Immunother. Cancer. 2019;7:74. doi: 10.1186/s40425-019-0553-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Hoffmann OI, et al. Impact of the spheroid model complexity on drug response. J. Biotechnol. 2015;205:14–23. doi: 10.1016/j.jbiotec.2015.02.029. [DOI] [PubMed] [Google Scholar]
- 48.Khot MI, et al. Inhibiting ABCG2 could potentially enhance the efficacy of hypericin-mediated photodynamic therapy in spheroidal cell models of colorectal cancer. Photodiagnosis Photodyn. Ther. 2018;23:221–229. doi: 10.1016/j.pdpdt.2018.06.027. [DOI] [PubMed] [Google Scholar]
- 49.Däster S, et al. Induction of hypoxia and necrosis in multicellular tumor spheroids is associated with resistance to chemotherapy treatment. Oncotarget. 2017;8:1725–1736. doi: 10.18632/oncotarget.13857. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Gustavsson B, et al. A review of the evolution of systemic chemotherapy in the management of colorectal cancer. Clin. Colorectal Cancer. 2015;14:1–10. doi: 10.1016/j.clcc.2014.11.002. [DOI] [PubMed] [Google Scholar]
- 51.Riley A, et al. A novel microfluidic device capable of maintaining functional thyroid carcinoma specimens ex vivo provides a new drug screening platform. BMC Cancer. 2019;19:259. doi: 10.1186/s12885-019-5465-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Kennedy R, et al. A patient tumour-on-a-chip system for personalised investigation of radiotherapy based treatment regimens. Sci. Rep. 2019;9:6327. doi: 10.1038/s41598-019-42745-2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Cheah R, et al. Measuring the response of human head and neck squamous cell carcinoma to irradiation in a microfluidic model allowing customized therapy. Int. J. Oncol. 2017;51:1227–1238. doi: 10.3892/ijo.2017.4118. [DOI] [PubMed] [Google Scholar]