Abstract
Whole-room indirect calorimeters have been used to study human metabolism for over a century. These studies have contributed substantial knowledge to the assessment of nutritional needs, and the regulation of energy expenditure and substrate oxidation in humans. However, comparing results from studies conducted at different sites is challenging due to lack of consistency in reporting technical performance, study design, and results. In May 2019, an expert Panel was convened to consider minimal requirements for conducting and reporting the results of human whole-room indirect calorimeter studies. We propose Room Indirect Calorimetry Operating and Reporting Standards, version 1.0 (RICORS 1.0) to provide guidance to ensure consistency and facilitate meaningful comparisons of human energy metabolism studies across publications, laboratories, and clinical sites.
Keywords: Energy expenditure, metabolism, human, respiration chamber
1. Introduction
The idea that the human body produces heat was a topic of fascination of the ancient Greeks, who believed that “vital heat” was produced by the heart, maintained by breathed air, and circulated through the blood. Aristotle postulated a link between energy intake, heat production, and breathing nearly perfectly capturing the concepts of direct and indirect calorimetry (1). In the late 1700’s, Antoine-Laurent Lavoisier made the first measurements oxygen consumption in humans at rest, during exercise, and during exposure to different temperatures (2). Around the same time, Lavoisier and Laplace reasoned that the amount of ice melted by a hot object as it cooled down to 0° C would be proportional to heat production. They subsequently demonstrated this using a guinea pig (3). In 1862, building on the progression of technology to measure expired gases, Pettenkofer constructed the first whole room indirect calorimeter to measure respiratory gas exchange in humans (Figure 1) (3). In 1905, Atwater and Benedict built a calorimeter that measured both heat exchange (i.e., direct calorimetry) and respiratory gas exchange simultaneously, and confirmed that both methods could accurately quantify human energy expenditure (4). With advances in technology, the use of whole-room indirect calorimeters expanded. Groups at Cambridge, UK (5, 6) and in Lausanne, Switzerland (7, 8, 9) set the early standard for adapting new techniques to measurement of energy expenditure in humans and even extended its use in rural Gambia, Africa (8). The principles of room calorimetry have been extended to specialized calorimeters to measure metabolic rates in infants. (10, 11). Many other prominent groups have built on their expertise over the years, incorporating the latest technological advances into room calorimeter systems which now employ indirect calorimetry almost exclusively.
Figure 1.

The first human open-circuit calorimeter built by Pettenkofer.
Whole-room indirect calorimetry allows subjects to move freely and the extended time of measurement (24 hours to several days) permits differentiation of the components of energy expenditure (e.g. sleeping, resting, physical activity, thermic effect of food). These systems are open-circuit as the room is ventilated (diluted) with “fresh air” and the expired gases by the study participant are mixed and the excurrent air is sampled to determine oxygen consumption (V̇O2) and carbon dioxide production (V̇CO2) (Figure 2). Most calorimeter systems allow for a broad range of temperature settings which can be useful to ensure thermal comfort or to study the effects of ambient temperature on energy metabolism. However, in most settings temperature and relative humidity are kept relatively constant during a measurement period. Participants’ physical activity is often monitored using wall- or ceiling-mounted Doppler-based motion sensing systems (12). Alternatively, wearable devices, such as accelerometers or heart rate sensors, and force platforms (13), can also quantify the intensity of participants’ activity.
Figure 2. (A).

Schematic layout of a room calorimeter. (B) Calorimeters located at the University of Colorado Anschutz Medical Campus (upper left), Pennington Biomedical Research Center (upper right), Maastricht University (lower left), and National Institutes of Health, Bethesda (lower right).
Due to the challenges of detecting small changes of gas concentrations in a large volume of air, early calorimeters had slow response times, with V̇O2 and V̇CO2 measured in intervals of 10–15 minutes (14, 15). Improvements in technology and equipment, combined with the application of sophisticated signal processing techniques, have contributed to the development of systems with a response time on the order of one to several minutes (16, 17, 18). Most systems either push air into or pull air from the room, while some employ a “push-pull” that minimizes variability in CO2 concentration in the calorimeter (16). Regardless of approach, leakage of air into or out of the room must be minimized. It is ideal to have a single air entry point to the room or at least confidence that all air entering is of the same composition. All systems also require complete mixing of air so the “sampled” air gas concentrations are accurate representations of the entire room. For comprehensive reviews on the history and technical advances in open circuit respirometry, the reader is referred to the excellent textbooks by Paul Webb (15) and John Lighton (19) as well as a recent review by Schoffelen and Plasqui (20)
There are currently more than 40 research laboratories on four continents which house whole-room indirect calorimeters for human studies (Table 1). The limited availability is due to expense and the technical expertise required to support their operation and maintenance. Types of study include assessments of energy need for various populations, evaluation of wearable sensors, and interventions of diet, sleep, exercise, temperature, and pharmacological agents. However, there are currently no accepted standards for calibration and operation, quality control, protocol design, signal processing, or data reporting. As a result, comparing results between laboratoriesis often difficult and the ability to perform multicenter trials is limited.
Table 1.
Known research-based room calorimeters currently in operation.
| Center | Location | Number of Rooms | Configuration | |
|---|---|---|---|---|
| North America | ||||
| 1. Baylor College of Medicine | Houston, TX, USA | 4 | Push-pull | |
| 2. Columbia University Medical Center | New York, NY, USA | −3 | Pull | |
| 3. NIDDK Metabolic Clinical Research Unit | Bethesda, MD, USA | 3 | Pull | |
| 4. NIDDK Phoenix Epidemiology and Clinical Research Branch | Phoenix, AZ, USA | 2 | Push | |
| 5. Pennington Biomedical Research Center | Baton Rouge, LA, USA | 4 | Pull | |
| 6. Mount Sinai St. Luke’s Hospital | New York, NY, USA | 1 | Pull | |
| 7. Translational Research Institute for Metabolism and Diabetes | Orlando, FL, USA | 4 | Push-pull | |
| 8. University of Alabama Birmingham | Birmingham, AL, USA | 1 | Pull | |
| 9. University of Colorado Anschutz Medical Campus | Denver, CO, USA | 1 | Pull | |
| 10. University of Massachusetts Amherst | Amherst, MA, USA | 2 | Push-pull | |
| 11. University of North Carolina Kannapolis | Kannapolis, NC, USA | 1 | Pull | |
| 12. USDA Beltsville Human Nutrition Research Center | Beltsville, MD, USA | 2 | Pull | |
| 13. USDA Western Human Nutrition Research Center | Davis, CA, USA | 1 | Pull | |
| 14. USDA Grand Forks Human Nutrition Research Center | Grand Forks, ND, USA | 1 | Push-pull | |
| 15. University of Wisconsin Madison | Madison, WI, USA | 1 | Pull | |
| 16. Vanderbilt University Medical Center | Nashville, TN, USA | 1 | Pull | |
| 17. Virginia Commonwealth University | Richmond, VA, USA | 2 | Push-pull | |
| Europe | ||||
| 18. NIHR Cambridge Clinical Research Facility | Cambridge, UK | 2 | Push | |
| 19. University of Warwick | Warwick, UK | 2 | Pull | |
| 20. Metabolic Research Unit Maastricht | Maastricht, The Netherlands | 5 | Pull | |
| 21. University Hospital of Pisa | Pisa, Italy | 1 | Push/Pull | |
| 22. Christian-Albrechts University | Kiel, Germany | 2 | Pull | |
| 23. University of Hohenheim | Stuttgart, Germany | 2 | Pull | |
| 25. University of Copenhagen | Frederiksberg, Denmark | 2 | Pull | |
| Asia | ||||
| 25. National Institute of Health and Nutrition | Tokyo, Japan | 2 | Pull | |
| 26. University of Tsukuba | Tsukuba, Japan | 4 | Pull | |
| 27. KAO Corporation | Tokyo, Japan | 2 | Pull | |
| 28. Sendai University | Sendai, Japan | 1 | Pull | |
| 29. Doshisha University | Kyotanabe, Japan | 1 | Pull | |
| 30. Hirosaki University | Hirosaki, Japan | 1 | Pull | |
| 31. Fukuoka University Institute for Physical Activity | Fukuoka, Japan | 1 | Pull | |
| 32. Ritsumeikan University | Kusatsu City, Japan | 1 | Pull | |
| 33. Juntendo University | Tokyo & Inzai, Japan | Pull | ||
| 34. Hosei University | Machida, Japan | 1 | Pull | |
| 35. Daikin Industries, Ltd. | Settsu, Japan | 1 | Pull | |
| 36. Singapore Institute for Clinical Sciences | Singapore, Singapore | Pull | ||
| 37. BGI Group | Shenzhen, China | 1 | Pull | |
| 38. Shanghai University of Sport | Shanghai, China | 1 | Pull | |
| 39. Shanghai Jiao Tong University | Shanghai, China | Pull | ||
| 40. Capital Medical University | Beijing, China | 1 | Pull | |
| 41. St John’s Medical College | Bangalore, India | 1 | Push | |
| Australia | ||||
| 42. University of Wollongong | Wollongong, Australia | 2 | Push | |
In May 2019, an international Panel of calorimetry experts was convened at the Translational Research Institute for Metabolism & Diabetes in Orlando Florida. The objective of this meeting was to establish minimum requirements for reporting 1) technical specifications; 2) data reduction and analytical approaches; 3) performance standards including accuracy, precision, stability, and repeatability; 4) elements of study design, including the basis for power analyses, inclusion/exclusion criteria, and appropriate considerations of potential confounders (e.g., pre-study dietary intake and activity); and 5) outcome measures. The Panel agreed this effort would improve the ability to compare data among studies, educate and guide new researchers to the field, and facilitate preparations and reviews of new publications. The result is the present Room Indirect Calorimetry Operating and Reporting Standards, version 1.0 (RICORS 1.0), which provides guidance to ensure consistency and facilitate meaningful comparisons and reproducibility of human energy metabolism studies across publications and clinical sites.
2. Reporting Technical Specifications
The performance of a calorimeter system is defined by its accuracy and precision. Accuracy describes the proximity of measurements to traceable standards. Precision describes the variability in repeated measures of the same quantity. Knowledge of these factors, along with biological variability, are necessary to appropriately power a study. This section focuses on determining and reporting the technical accuracy and precision of the calorimetry measurement systems. Minimizing confounding biological variabilities is a matter of protocol design, compliance, and supervision which is addressed in Section 5.
System accuracy and variability
The Panel recommends regular calibration of all calorimeter system components to traceable standards, including gas analyzers, flow meters, and environmental sensors. Once calibrated, system accuracy and precision can be determined using various validation tests. We recommend reporting gas exchange rates (V̇O2 and V̇CO2, in l/min, ml/min, or l/day) as primary outcomes for these validation tests, since energy expenditure, macronutrient oxidation rates, and respiratory exchange ratio (RER, the ratio of V̇CO2 to V̇O2) are all derived from gas exchange rates. It should be noted that although RER reflects measurements made at the mouth (i.e., whole body level) and respiratory quotient (RQ) reflects gas exchange at the tissue or organ level, the two terms are often used synonymously.
To obtain accurate measurements of V̇O2 and V̇CO2, determining the room volume is vital. One approach is to use room dimensions to estimate volume, accounting for the volume of all impermeable objects in the room (e.g. furniture, sink, toilet, equipment) (17). Room volume can also be derived from the exponential change in gas concentrations in a closed room calorimeter after a “pulse” of gas is introduced into the room (16). This can be done using a gas infusion or combustion, or even after a participant exits the room at the end of a study.
We recommend that performance of a calorimeter system be regularly verified using a combination of approaches. First, a “zero test” (i.e., recording of gas exchange rates in an empty, undisturbed room over a typical measurement period) will provide basic information about bias, drift, and inherent noise of the system (21). Second, system performance should be evaluated against traceable standards using these two common approaches: 1) combustion of a flammable substance such as ethanol, propane, or butane; or 2) infusion of dry gases (nitrogen and CO2) to simulate V̇O2 and V̇CO2. With combustion, expected V̇O2 and V̇CO2 are calculated based on stoichiometry from the amount of fuel burned (e.g., measured by a calibrated scale). Combustion has the advantages of closely simulating fuel oxidation with heat and water vapor as products, is relatively easy to setup (Figures 3a, 3b), and thus has been widely practiced (17, 18). Limitations of this approach include safety concerns with an open flame, assumed combustion efficiencies and purity (e.g., free of water and other contaminants), a resultant RER that is below the physiological range (0.667 for alcohols and 0.60 for alkanes), and the expense of high-grade fuels (e.g., a bottle of 99.2% propane is about $50 which lasts two 24-h burns). Furthermore, even though the burning rate is manually set for a combustion test and typically not modifiable during the test (Figures 3a–b), progressive cooling of the liquefied gas as it evaporates which gradually reduces the actual burning rate. As such, programmable devices may be used to set and control combustion rates more precisely. In contrast, gas infusion systems (Figure 3c) have the advantage of flexibility, allowing both independent and combined gas exchange simulations to mimic rises and falls in EE and RER. This is achieved by using a pair of calibrated mass flow controllers to blend high-grade (>99%) nitrogen (N2) and CO2 mixtures to simulate EE and RER profiles over the physiological range (16, 22, 23). Compared to weighted combustion approach for validations of calorimetry systems, the technical requirements are more complex for gas infusions. For example, because mass flow controllers are prone to drift, we recommend verification of their gas delivery rate during each infusion, independent of the stated manufacturer calibration. Gas infusions may also be done gravimetrically, by observing the weight loss of a cylinder as gas flows from it or volumetrically by using a drum, bellows, or wet-type gas meter. These meters offer a highly accurate (±0. 2%) reference for volumetric gas flow measurements (14).
Figure 3.

Room calorimeter validation methods. Combustion of propane (A) and alcohol (B) are well-established validation approaches that provide relatively constant rates of CO2 production (VCO2) and O2 consumption (VO2), verified by weight. The ratio of VCO2/VO2 is constant and stoichiometrically-derived for each fuel source (0.6 for propane and 0.67 for alcohol). (C) Gas blenders infuse N2 and CO2 at user-defined rates governed by mass-flow controllers to mimic variable, independent rates of VO2 (via oxygen dilution) and VCO2. Gas blender images provided by MEI Research Ltd.
Gas infusions do not completely mimic human calorimeter studies due to a lack of heat and water vapor production. A participant will add water content (upwards of 1g min−1) to the room air through respiration and perspiration. As a result, ingoing and outgoing air samples will have different water vapor contents. To measure V̇O2 and V̇CO2 accurately, the water vapor pressure must be accounted for, either by employing a system to remove water vapor from the samples of both the incurrent and excurrent, or by precisely measuring water vapor pressure and mathematically removing the effects of water vapor on gas pressures (14, 16, 18, 24, 25). Inadequate drying is most clearly evident as an over-estimate of V̇O2 (25). A variety of approaches have been employed to remove water vapor content in room calorimeter systems including cooling systems that condense most of the water, chemical dryers, and counter-flow devices that promote exchange of water vapor from the sample gasses using a continuous flow of dry gas around the liquid-permeable sample tubes. Depending on the drying approach, seasonal variations in ambient conditions may interfere with the performance (25). To more closely approximate human studies, gas infusion tests may be modified by introducing water vapor (e.g. evaporation on a boiling plate) and/or heat (e.g. electric heater with known wattage) into the calorimeter room. To identify issues caused by ineffective water vapor removal or temperature control, the results of infusion tests with these additional elements can then be compared to infusions with dry gas alone.
The Panel recommends that the first manuscript following establishment or modification (major changes to hardware and/or software) of a room calorimeter, include sufficient detail regarding instrumentation, system characteristics (e.g., room volume, analyzer and flow meter specifications, procedures for eliminating water vapor pressure effects on gas concentrations, and temperature and humidity control), and results of validation tests (zero tests, combustion tests, infusion tests, and repeated measures on human participants). External validations against metabolic carts may also be informative and can provide additional data that can be used to “fine tune” the performance of the system (25). The Panel also recommends regular system validation using combustion as a minimum standard. The Panel did not have consensus of what would constitute “regular” but recommends combustion tests be performed at least quarterly. Other tests (infusion tests, zero tests, water evaporation tests) can be performed less frequently (e.g., after a software/hardware modifications). The Panel also recommends that the length of the validations be close to the length of the human studies (e.g. 24 h) with rates of V̇O2 and V̇CO2 within the range expected when studying human participants. Accuracy should be reported as the mean (or median) difference of expected versus observed gas exchange rates. Precision (variability) should be reported as the standard deviation of the expected versus observed differences for a single test or as the confidence limits of these differences for a group of tests performed over the duration of the protocol.
Signal processing
Early room calorimeters employed chemical or gravimetric approaches to measure V̇O2 and V̇CO2 (4, 20). These methods were time and labor intensive and, thus, studies were limited to steady states of metabolism or the aggregate of several metabolic states over a matter of hours. Gravimetric instrumentation and hand-written records have now been replaced with digital instrumentation and computers capable of measuring gas concentrations at rates less than 1 Hz (20). However, because the volume of most room calorimeters is approximately 50,000 to 100,000 times larger than the volume of V̇O2 and V̇CO2 each minute at rest, the practical limit to detect a change in gas concentration is on the order of one minute (19) and software-based algorithms are typically implemented to enhance the signal-to-noise ratio of room calorimeter systems.
The minimum detectable change in gas concentration for a room calorimeter is related to both instrumentations and physical dimensions. Modern gas analyzers (e.g. paramagnetic, infra-red, fuel cell, mass-spectrometry) differ in their physical properties that govern their relative sensitivity to different noise sources (e.g. vibration, water vapor content, electromagnetic interference, changes in flow and pressure), lifespan, and propensity to drift, amongst other things (19). However, their response time (milliseconds to seconds) and resolution (<0.01%) to changes in gas concentration fall within a similar range (19), and thus the type of analyzer employed is not the major determinant of the response time, accuracy, and precision of a room calorimeter system.
The time between a sudden change in room gas concentration, such as the onset of exercise, and its detection by gas analyzers, is defined by the calorimeter “time constant” (i.e. response time), which is calculated as the room volume divided by excurrent flow rate (14, 19, 26). The time constant is incorporated into the equations to determine V̇O2 and V̇CO2, which include a closed-system (first derivative) term, i.e. the change in room gas concentration over time multiplied by the room volume, either directly (14, 15, 16, 17, 24) or using the Z-transformation method (19). The time constant can be reduced by decreasing room volume and/or increasing flow rate. However, because room volume is approximately 100–500 times larger than excurrent flow rate in a typical room calorimeter it plays the greater role. Reducing room volumes can limit subject comfort and range of viable activities while increasing flow rates can increase noise from the device used to control incurrent or excurrent flow.
While incorporating closed-system terms into equations for V̇O2 and V̇CO2 can analytically improve response time, it also amplifies noise inherent to O2 and CO2 measurements (19, 24). Thus, some laboratories have developed software-based approaches to minimize this noise while retaining the improved response time using dynamic gas-infusion protocols as the validation criteria for their signal-processing techniques. Early noise-suppression efforts included averaging the gas concentration signals over time-windows of various lengths prior to computing a discrete time derivative (16, 18, 24). Choosing the duration of the averaging window amounts to a tradeoff between a longer window with a less noisy steady-state and slower response and a shorter window with a quicker response and noisier steady-state (27). Others proposed fitting sliding windows of gas concentration data with two exponentials (26, 28) or quadratic splines (24) prior to computing the derivative. These methods assume that changes in metabolic state will take a specific form and occur over a predetermined minimum duration. While this approach yields fast responses and improves the signal-to-noise ratio when these assumptions are met (26, 28), shorter-than-expected state changes may produce inaccurate calculations of V̇O2 and V̇CO2 (22). More complex signal processing techniques, such as Kalman filtering (29), deconvolution (29), wavelet denoising (22), and total variation denoising (27) have demonstrated promising improvements for noise suppression but are more complex to implement and require independent validations on multiple room calorimeter systems before field-wide adoption. It should be noted that zero tests and validations (infusions and combustions) conducted without software filters can provide information about the statistical properties of noise and are helpful in the design of signal processing algorithms.
The Panel recommends troubleshooting and optimization of instrumentation and operational setup prior to development and implementation of noise suppression methods, as software-based techniques should enhance system performance rather than correct for system inadequacies. Signal processing algorithms should be validated against traceable standards, not only to demonstrate improved short-term-precision (e.g. reduced minute-to-minute error in V̇O2 and V̇CO2) but also to ensure that they do not compromise longer-term accuracy (e.g. unchanged total V̇O2 and V̇CO2 over hours and days). The Panel also recommends that the first manuscript from a newly commissioned room calorimeter or following hardware/software modification to an existing system include sufficient signal processing details and performance comparisons to published approaches. Subsequent studies from the same system should then reference the original paper. This is particularly important if the reported calorimeter outcomes are of short durations or are dynamic in nature (such as V̇O2 during an exercise bout or slope of a postprandial RER); only reporting the validity of 24hr performances using a static combustion test would not be sufficient under these circumstances.
3. Reporting Data Analysis Approaches
While indirect calorimeters measure V̇O2 and V̇CO2, energy expenditure and/or substrate oxidation are the desired outcomes for clinical or physiological studies. Thus, it is important to report the analytical approaches used to determine the physiological determinants of energy metabolism. Components of 24 h EE measured using room calorimetry include (30): 1) sleeping metabolic rate (SMR), 2) resting metabolic rate (RMR); 3) thermic effect of food (TEF); and 4) physical activity related energy expenditure (PAEE) (Figure 4). With proper experimental design, it is also possible to measure additional components of 24 h EE such as excess post-exercise oxygen consumption (31). When comparing groups of disparate body size or composition, proper normalization procedures should be employed.
Figure 4.

Components of total daily energy expenditure. Physical activity energy expenditure (PAEE) has a linear relationship with concurrent activity measured by radar. The energy expenditure at zero activity, i.e. the y-intercept, represents the sum of the resting metabolic rate (RMR) and thermic effect of feeding (TEF). The RMR is comprised of the sleeping metabolic rate (SMR) and the energy expenditure from arousal. Adapted from Ravussin, et. al., 1986 (17).
Components of 24 h energy expenditure
The BMR is the rate of EE at complete rest, and represents the EE required to maintain vital functions, such as basic chemical reactions of the body and is the primary contributor (60–75%) of 24 h EE measured in a room calorimeter (30). The BMR is typically measured using a ventilated hood. The strictest definition of BMR requires that an individual be in a post-absorptive state (10–12 hours without consumption of caloric food or beverages), physically undisturbed, awake, in a thermally neutral environment (32). The measurement should be obtained in the morning following an overnight stay in a testing facility. The BMR can be measured in a room calorimeter if the conditions described above are met, although a longer measurement period is recommended (e.g., minimum of 30 minutes). If the resting measurements are obtained under less strict conditions, this is referred to as the RMR or resting energy expenditure (REE).
In room calorimeter studies, sleeping metabolic rate (SMR) is sometimes used as an alternative measure of BMR/RMR. The SMR has the lowest intra-subject coefficient of variation and, consequently, is the most reproducible component of 24 h EE (21, 33). There are several challenges that must be considered when determining SMR. For example, EE during sleep decreases as the night progresses because TEF is sustained for at least four to six hours after consuming a meal (34). The RMR also varies by circadian phase; RMR is lowest late in the biological night at a time (~4–5 AM) that corresponds to the nadir in endogenous core body temperature (35). Thus, it is likely that SMR varies during the biological night, such that SMR measured during the early part of the biological night is higher than during later phases of the biological night. There is no clear consensus on how to define SMR. Some investigators have defined SMR as the lowest observed EE during the night over three consecutive hours (33). whereas others have defined SMR as the average EE during the three consecutive hours in the nighttime period with the lowest observed physical activity (body movement) (33, 36, 37). The latter approach requires a measure of body movement such as accelerometry or in-room motion sensor system. Theoretically, the lowest body movement could be earlier in the night where TEF is still present. Thus, a more robust definition in this case would be the three consecutive hours with the lowest body movement that occur at least six hours after consumption of the last meal or the EE averaged from minutes of minimal activity over a prespecified post-TEF time period, for example 0100–0500 h (38) or 0200–0500 h (39). Finally, another factor that should be considered when determining SMR is the “first night” effect. It has long been known that when individuals sleep in an environment they are not accustomed to, sleep stages are disrupted such that the first night of “laboratory” sleep contains more awake periods, less rapid eye movement (REM) sleep, and delay in the onset of deep sleep stages. These disruptions in sleep quality rapidly dissipate during the second night of laboratory sleep (40). A first night effect on SMR has been shown in some studies (41), but not others (33), which may be partly attributed to the method of SMR determination. Similarly, some studies have shown energy expenditure to differ between sleep stages (42, 43), while others have not (44). Nonetheless, the Panel recommends that the first night effect should be considered when measuring SMR and may be particularly important for cross-over design studies.
The TEF is the increase in EE above BMR/RMR due to food intake. The TEF is dependent on both the size and macronutrient composition of the meal (45). When healthy subjects are in approximate energy balance (energy expenditure ≈ energy intake) and consuming a mixed diet, TEF is ~10% of 24 h EE. In a room calorimeter study, it is difficult to independently attribute increases in EE above BMR/RMR to TEF and PAEE, unless one of these components is minimized by design. For example, with a participant studied awake and under complete bed rest, the increase in EE above RMR represents TEF. Conversely, in a participant studied under fasted conditions, the increase in EE above RMR represents PAEE. Some rigorous studies of TEF take advantage of the day-long measurement capability of room calorimeters by fasting participants for one day and feeding on another day while minimizing activity each day (46, 47, 48). Since this protocol requires two separate days and burdens participants with a prolonged fast, alternative methods to estimate TEF using a measurement of physical activity have been proposed (46, 47, 49, 50). With this approach, physical activity is regressed against energy expenditure per unit of time (Figure 4). The y-interecept is the EE in the inactive state, and the difference between this value and RMR is considered to represent the cumulative TEF. However, TEF computed by this method had poor reproducibility when participants were studied on more than one occasion (47). Thus, methodology to measure TEF using whole-room calorimeters is an area ripe for development.
Macronutrient oxidation rates
Stoichiometric equations of different substrates reacting with oxygen to produce carbon dioxide can be used to estimate macronutrient oxidation. The earliest equations published by Zuntz (51) assumed the amount of protein oxidized was negligible. Because room calorimetry measures are typically carried out over many hours, the contribution of protein oxidation cannot be ignored, and thus, more current calculations of macronutrient oxidation also include protein oxidation determined by the quantification of nitrogen excretion in the urine. Many sites use Weir’s equations to derive macronutrient oxidation (52), which account for the fraction of protein oxidation (assumed or measured), although alternative equations have been presented (53, 54, 55). Depending on the choice of equation, fat oxidation can vary by ~3% whereas carbohydrate oxidation can vary by up to 6% (56). Thus, the Panel recommends that the equations used to calculate macronutrient oxidation be specified, and strongly recommends that protein oxidation be determined from measured urinary nitrogen excretion rather than using an assumed value (52).
As eloquently detailed by Livesey and Elia (57), there is more uncertainty when using indirect calorimetry to determine macronutrient oxidation than in the uncertainty of using indirect calorimetry to measure energy expenditure. There are several sources of potential error in calculating macronutrient oxidation from gas exchange rates and nitrogen excretion. First, small errors in the measurement of V̇O2 and/or V̇CO2 can lead to large errors in carbohydrate and fat oxidation (57). Second, estimates of protein oxidation from nitrogen excretion can vary substantially (57). Part of this is due to variability in nitrogen assay performance. Thus, the accuracy of the methods used to determine urinary nitrogen concentrations should be reported. There are a number of assays available for performing urine nitrogen analyses including colorimetric and chemiluminescence assays (58). Although not commonly employed, urinary urea nitrogen is ~80–100% of total urinary nitrogen, and thus can be used as a means of verifying total urine nitrogen. It is essential that all urine be collected over the 24 h period, including a void obtained at the end of the calorimeter study.
Respiratory exchange ratio (RER)
It is common practice to report RER as an index of relative macronutrient oxidation, either over 24 h or during discrete periods (e.g., during exercise or sleep). Although not discouraging this practice, the Panel recommends that care be taken in determining the RER values, particularly over short periods of time. RER is a dimensionless ratio defined as V̇CO2/V̇O2. Noise inherent to the measurement of CO2 and O2 results in minute-to-minute variation in V̇CO2 and V̇O2. Because noised-based variations in V̇CO2 and V̇O2 are not covariant this can lead to spuriously high or low values when RER is computed over short durations (i.e., minutes). Assuming noise is normally distributed and biological state is constant, using total V̇CO2 and V̇O2 over longer durations (i.e., hours) to calculate RER will minimize the effects of this noise, whereas calculating RER on a minute-to-minute basis results in a greater likelihood of incorporating values near zero or infinity into the average.
Normalization procedures
The major determinant of BMR/RMR and 24h EE in humans is fat-free mass (FFM) (17), although fat mass (FM) contributes to a small but perhaps important proportion (59). Thus, EE data must be adjusted or “normalized” to compare values between individuals of different body size or composition. Dividing EE by body weight or fat free mass may lead to misinterpretation due to a “non-zero intercept artifact” (30, 60), as illustrated in Figure 5. In this example, RMR is plotted as a function of FFM for both men (open circles) and women (closed circles). Men and women fall on a shared regression line with a non-zero intercept, indicating the relationship between RMR and FFM is independent of sex. However, dividing RMR by FFM results in a significant difference between women and men (0.143±0.012 and 0.128±0.080 MJ/kg for women and men, respectively, P < 0.0001) (61). Regression-based methods are considered to be the preferred approach when comparing EE data from groups of different body composition (62) or when assessing changes in EE resulting from interventions that alter body composition.
Figure 5.

Body composition adjustments for energy expenditure. (A) Plotting resting metabolic rate (RMR) as a function of fat-free mass for 17 women (closed symbols) and 20 men (open symbols) results in a shared linear regression (RMR (MJ/d) = 2.27 + 0.091 fat-free mass (kg), r2 = 0.78) suggesting a sex-independent relationship between RMR and fat-free mass. (B) Dividing RMR by fat-free mass, however, yields a greater “normalized” RMR for women due to the non-zero intercept in (A). Adapted from Westerterp, 2000 (61).
For reporting data analysis approaches, the Panel recommends that the methods used to determine the components of energy expenditure be reported. If BMR or RMR is measured in the calorimeter, procedures to ensure that the measure was obtained under standard conditions should be reported, including whether participants were monitored to be sure that they remained awake and motionless during the period of measurement. Physical activity may be measured using a variety of methods, but recent advances have made using wearable monitors more feasible. This also has the advantage that the same wearable device can be worn outside of the calorimeter (e.g., during preceding study days), which permits a comparison of the physical activity level in the calorimeter to habitual physical activity levels. Although TEF is not commonly measured, the Panel suggests that this is an area for further methodological development and investigation. If SMR is measured, the period of time and additional criteria used to determine sleep vs. wakefulness should be reported. If macronutrient oxidation rates or RER are measured over discreet periods of time (e.g., during a bout of activity), data should be presented or referenced to demonstrate the ability of the system to accurately and reliably measure these outcomes over such short intervals. Finally, when normalizing data to permit comparisons of groups of disparate size/body composition, the Panel recommends regression-based approaches that consider both FFM and FM as covariates.
4. Reporting Elements of Study Design and Conduct
Appropriate study design for room calorimeter studies requires consideration of confounding factors that may affect energy expenditure and/or substrate oxidation (e.g., preceding diet, physical activity and sleep; weight stability; menstrual cycle phase; state of energy balance). There are generally three types of studies performed in room calorimeters: observational, longitudinal, and interventional. (63). Observational studies involve measuring participants under one or more conditions (e.g., high fat vs low fat diet; exercise vs. no exercise) within a short period of time. Longitudinal studies measure changes over time, with measurements performed at two or more time points. Interventional studies (e.g., diet, exercise training, weight loss, overfeeding) require measurements pre-, post-, and sometimes during the intervention. Almost all published room calorimeter studies to date are based on studies performed at a single site. To our knowledge, only one multi-site study has been published (63). Multi-site studies must take careful consideration to remove as many site-specific variables as possible to align the entire dataset.
In a well-performing system, biological variability should comprise the largest source of total variability. To estimate the number of study participants required, the variability in repeated measurements (intra-individual variability) should be determined (Figure 6). In many studies, EE is the primary study outcome. The coefficient variability (CV) of 24 h EE measured by indirect room calorimeters have been reported to be 1–5% (17, 21, 53, 64, 65, 66), whereas the CV of macronutrient oxidation are reported to be 5–25% (65, 66). In our review of published studies, in many cases the primary study outcomes were not clearly stated. The Panel strongly recommends that calorimeter studies clearly define the primary outcome, and that information to justify the sample size be presented.
Figure 6.

Repeated 23h participant measurements to assess biological intra-individual variability. Bland-Altman plots of the difference between day 1 and day 2 versus their average for oxygen consumption (VO2, A), carbon dioxide production (VCO2, B), energy expenditure (EE, C), and respiratory exchange ratio (RER, D) for 10 participants fed a eucaloric, energy balanced diet and instructed to approximate their same daily routine both days. Average differences for the group are VO2: 11.2±13.3 L/day (2.6±2.9%); VCO2: 5.6±12.8 L/day (1.5±3.4%); EE: 50.4±62.7 kcal/day (2.4±2.9%); RQ: −0.01±0.02 (−1.3±2.4%). (E) Minute-by-minute tracings over 23 hours for EE (solid lines, left y-axis) and RER (broken lines, right y-axis) on day 1 (black) and day 2 (red) for the subject with the lowest daily EE and activity. Data provided by KYC (unpublished).
A checklist for reporting participant characteristics, pre-study controls, and participant preparation is presented in Table 2. Where appropriate, these are listed according to type of study. The minimum information to report includes participant age, sex, ethnicity, weight, height, and body composition (including method of determination). Additional inclusion and exclusion criteria such as weight stability and medication, alcohol, and nicotine use should be reported. Studies of female participants should report menstrual status because ovarian hormones impact energy expenditure (67) and perhaps physical activity (68). When pre-menopausal women are studied, whether studies were performed in a specific menstrual cycle phase (luteal vs. follicular) should be reported, along with the methods used to determine cycle phase (69). Scheduling all pre-menopausal female participants in the same phase of the cycle presents many logistical challenges. Thus, when this cannot be achieved, the Panel recommends measuring ovarian hormone concentrations (e.g., estrogen, progesterone) and using this as covariates in the statistical models. Whether pre-menopausal women using hormonal contraceptives have been included or excluded should also be reported. When post-menopausal women are studied, the methods used to determine menopausal status should be reported (e.g., self-report based on cessation of menstrual cycle or confirmation by ovarian hormone), as well as whether women using hormonal replacement therapy were included or excluded.
Table 2.
Summary of Recommendations
| Participant Characteristics | Recommendation |
|---|---|
| Age, sex, ethnicity, race, weight, (height), BMI |
|
| Body composition |
|
| Prescription and over the counter medications |
|
| Weight stability |
|
| Tobacco/nicotine use |
|
| Female participants |
|
| Technical Specifications | Recommendation |
| Oxygen consumption (VO2) and carbon dioxide production (VCO2) |
|
| Signal Processing |
|
| Data analysis | Recommendation |
| 24 h expenditure |
|
| Components of total daily energy expenditure |
|
| Physical activity |
|
| Respiratory exchange ratio |
|
| Macronutrient oxidation |
|
| Normalization of energy expenditure |
|
| Study Design | Recommendation |
| Sample size and power |
|
| Pre-study control |
|
| Procedures for orienting participants to study design and safety procedures |
|
| Meals during calorimeter study |
|
| Physical activity monitoring |
|
| Safety procedures/participant monitoring |
|
| Bedtime/lights out time/napping |
|
In many calorimeter studies, a “run-in” period is included where exercise and energy intake are prescribed or controlled (70). Although not commonly practiced, a consistent sleep schedule may also be prescribed prior to calorimeter studies (71). The purpose of this run-in phase is to minimize the impact of variations in these parameters on energy stores, energy expenditure, and substrate oxidation (72). Any pre-study restrictions (e.g., diet, physical activity, sleep, etc.) should be reported, along with the prescribed duration of these restrictions and how they were verified. In some studies, energy and macronutrient intake is strictly controlled by providing food to participants. In these cases, formulae used to estimate energy and macronutrient requirements and methods to track compliance should be reported. If participants were instructed to complete an overnight fast prior to the study day, the length of this fast should also be reported.
Prior to the start of a study, consideration should be given to orienting the participant to the room calorimeter and study protocol. The Panel recommends that a scripted orientation be provided to each study participant including a review of the room layout and facilities, such as access to food ports, toilets, safety (e.g. emergency call buttons), and privacy features (e.g., video monitors and windows that view into the room, and whether the participant has any control over these). A written script of the study protocol should be provided including mealtimes, sample collection procedures (e.g., urine and/or blood) and schedule, and any prescribed exercise, sleep, and wake times. Participants should also be given instructions if meals are to be consumed within a prescribed period of time (e.g., 30 minutes), whether they should consume all food provided, and what activities they are permitted or required to perform during the study. These practices should minimize potential sources of biological variability, enhancing the ability to detect significant differences. The Panel recommends that regardless of study design, that physical activity during the calorimeter study be monitored and reported. Finally, participants should be instructed whether the use of cellular phones and other personal electronic devices is permitted in the calorimeter, and the time limits if they are permitted. Personal device use after bedtimes should be discouraged as this may interfere with sleep.
As we were finalizing this manuscript for submission, the COVID19 pandemic has become a serious health issue in many countries. Considering the seriousness of its impact, we believed it prudent to provide recommendations to reduce the risk of accidental transmission of viruses in calorimetry-related clinical research. It is recommended that users of calorimeters implement rigorous cleaning procedures prior to and following each participant study. Some calorimeters are located in hospital facilities or affiliated with medical centers, and procedures are often performed according to hospital standards. However, some calorimeters are not affiliated with hospital or medical centers, and these locations should adopt appropriate procedures. The Panel recommends that, before and after each human study, all surfaces inside the calorimeter chamber (sinks, toilets, tables), as well as devices such as computer keyboards, televisions, etc. should be thoroughly wiped with a disinfectant. Floors should be mopped with a disinfecting solution. Cleaning should be performed using nitrile gloves. Trash cans should be lined with plastic disposable bags that are changed after each study. Any linens provided should be changed and cleaned. In many calorimeters, a portion of the excurrent air is vented into the anterior chamber areas, which could promote the transmission of viruses into areas outside of the calorimeter. In these settings, the Panel recommends that a suitable filter be placed in-line of the venting system.
Another consideration for reducing the risk of accidental transmission is screening individuals for recent illnesses. The need to protect spread of viruses during the process of metabolic measurements is an often-overlooked area. In most situations, participants will perform the consent process and any health screening days or weeks prior to their scheduled study visits. Thus, the Panel recommends that sites employ procedures to screen individuals immediately prior to the study visit to inquire about recent illnesses. Participants be screened about recent (e.g., previous 2–3 weeks) incidents of fever, cough, shortness of breath, or any other illness or symptom, and that these cases be reviewed by appropriate medical personnel to determine whether the study visit should be rescheduled.
5. Summary and future directions
The future of room calorimetry is bright: compared to 20 years ago, technological advancements are making instruments more affordable, and the acquisition, processing. and analysis of data is more streamlined and efficient. New room calorimeters are being constructed around the world, particularly in Asia and North America. Most calorimeters are custom built with variability in room dimensions, instrumentation, environmental controls, and data processing. Thus, standardizing performance testing and reporting can facilitate comparison of different sites.
Because of the specialized nature of building, operating, and maintaining room calorimeters, there are currently limited opportunities for comprehensive training. Furthermore, unlike metabolic carts, there are few companies that specialize in the design and construction of room calorimeters. We can offer several recommendations for those that wish to pursue learning opportunities in this field. First, Table 1 lists the sites known to maintain operating room calorimeters. Second, although not exclusively focused on room calorimetry, the triannual conference entitled “Recent Advances and Controversies in Measuring Energy Expenditure” (www.RACMEM.org) devotes numerous sessions on technical and operational components that are covered in this document.
We propose RICORS 1.0 criteria (Table 2) as the minimum guidance for performing, analyzing, and reporting human whole-room calorimetry studies. There are many decisions and assumptions made during study design and conduct, data acquisition, and data analysis that should be considered and reported when performing such studies. The RICORS recommendations are intended to provide consistency in reporting to permit proper evaluation and interpretation of results, as well as providing consistency so that results from different studies can be compared. Moreover, the RICORS recommendations can guide further improvements in calorimetry technological developments and practical operations, such as developing new algorithms to increase sensitivity and reduce response time, and measuring TEF, excess post-exercise oxygen consumption (EPOC) and sedentary energy expenditure. The RICORS criteria were developed with the consultation of many room calorimetry experts, some of whom have performed studies in this area for several decades. As the first consensus Panel report, RICORS should be evaluated by operators in the field and updated as needed.
The authors welcome suggestions from the research community to refine these recommendations (please send suggestions to the corresponding author). Finally, we envision that this document will serve as a template for other fields where standardization in reporting is lacking (e.g., animal calorimetry studies).
What is known about this topic?
Studies of humans using whole-room indirect calorimetry have advanced our understanding of energy metabolism in humans
There are a limited number of operational calorimeters around the world, but a number of new calorimeters have opened in recent years
There are no agreed upon standards for reporting the technical aspects or study results using whole-room indirect calorimeters
What are the new findings?
Most calorimeters are custom built with variability in room dimensions, instrumentation, environmental controls, and data processing.
Guidance is needed to improve the ability to compare data among studies, educate and guide new researchers to the field, and facilitate preparations and reviews of new publications
To address these issues, an expert Panel was convened to consider minimal requirements for conducting and reporting the results of human whole-room indirect calorimeter studies.
How might your results change the direction of research?
Recommending standards performance testing and reporting will facilitate comparison of study results from different sites
This document can also serve as a guide to reviewers to ensure that appropriate information is reported to enhance the rigor and reproducibility of research using whole-room indirect calorimetry
The RICORS recommendations can guide further improvements in calorimetry technological developments and practical operations
FUNDING:
Z01 DK071013, Z01 DK071014, P30 DK048520, P30 DK072476
DISCLOSURES: Dr. Melanson is supported by resources from the Geriatric Research, Education, and the Clinical Center at the Denver VA Medical Center. The contents do not represent the views of the U.S. Department of Veterans Affairs or the United States Government
References
- 1.Ogle W Aristotle on Youth & Old Age, Life & Death: And Respiration. Longmans, Green, 1897. [Google Scholar]
- 2.Lavoisier AL. Expériences sur la respiration des animaux, et sur les changemens qui arrivent à l’air en passant par leur poumon. Académie des sciences 1780. [Google Scholar]
- 3.Lavoisier AL, marquis de Laplace PS. Mémoire sur la chaleur: Lû à’Académie royale des sciences, le 28 juin 1783. De l’Imprimerie royale, 1783. [Google Scholar]
- 4.Atwater WO, Benedict FG. A respiration chamber, with appliances for the direct determination of oxygen. In: Washington CIo (ed). Carnegie Institution of Washington: Washington, D.C., 1905. [Google Scholar]
- 5.Dauncey MJ, Murgatroyd PR. A direct and indirect calorimeter for studies on energy expenditure in man over 24 h periods [proceedings]. J Physiol 1978;284: 7P–8P. [PubMed] [Google Scholar]
- 6.Dauncey MJ, Murgatroyd PR, Cole TJ. A human calorimeter for the direct and indirect measurement of 24 h energy expenditure. Br J Nutr 1978;39: 557–566. [DOI] [PubMed] [Google Scholar]
- 7.Jequier E, Schutz Y. Long-term measurements of energy expenditure in humans using a respiration chamber. Am J Clin Nutr 1983;38: 989–998. [DOI] [PubMed] [Google Scholar]
- 8.Minghelli G, Schutz Y, Charbonnier A, Whitehead R, Jequier E. Twenty-four-hour energy expenditure and basal metabolic rate measured in a whole-body indirect calorimeter in Gambian men. Am J Clin Nutr 1990;51: 563–570. [DOI] [PubMed] [Google Scholar]
- 9.Jequier E Measurement of energy expenditure in clinical nutritional assessment. JPEN J Parenter Enteral Nutr 1987;11: 86S–89S. [DOI] [PubMed] [Google Scholar]
- 10.Cole CR, Rising R, Hakim A, Danon M, Mehta R, Choudhury S, et al. Comprehensive assessment of the components of energy expenditure in infants using a new infant respiratory chamber. J Am Coll Nutr 1999;18: 233–241. [DOI] [PubMed] [Google Scholar]
- 11.Jensen CL, Butte NF, Wong WW, Moon JK. Determining energy expenditure in preterm infants: comparison of 2H(2)18O method and indirect calorimetry. Am J Physiol 1992;263: R685–692. [DOI] [PubMed] [Google Scholar]
- 12.Schutz Y, Ravussin E, Diethelm R, Jequier E. Spontaneous physical activity measured by radar in obese and control subject studied in a respiration chamber. Int J Obes 1982;6: 23–28. [PubMed] [Google Scholar]
- 13.Sun M, Hill JO. A method for measuring mechanical work and work efficiency during human activities. J Biomech 1993;26: 229–241. [DOI] [PubMed] [Google Scholar]
- 14.Schoffelen PF, Westerterp KR, Saris WH, Ten Hoor F. A dual-respiration chamber system with automated calibration. J Appl Physiol (1985) 1997;83: 2064–2072. [DOI] [PubMed] [Google Scholar]
- 15.Webb P, Saris WH, Schoffelen PF, Van Ingen Schenau GJ, Ten Hoor F. The work of walking: a calorimetric study. Med Sci Sports Exerc 1988;20: 331–337. [DOI] [PubMed] [Google Scholar]
- 16.Moon JK, Vohra FA, Valerio Jimenez OS, Puyau MR, Butte NF. Closed-loop control of carbon dioxide concentration and pressure improves response of room respiration calorimeters. J Nutr 1995;125: 220–228. [DOI] [PubMed] [Google Scholar]
- 17.Ravussin E, Lillioja S, Anderson TE, Christin L, Bogardus C. Determinants of 24-hour energy expenditure in man. Methods and results using a respiratory chamber. J Clin Invest 1986;78: 1568–1578. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Sun M, Reed GW, Hill JO. Modification of a whole room indirect calorimeter for measurement of rapid changes in energy expenditure. J Appl Physiol (1985) 1994;76: 2686–2691. [DOI] [PubMed] [Google Scholar]
- 19.Lighton JR. Measuring metabolic rates: a manual for scientists. Oxford University Press, 2018. [Google Scholar]
- 20.Schoffelen PFM, Plasqui G. Classical experiments in whole-body metabolism: open-circuit respirometry-diluted flow chamber, hood, or facemask systems. Eur J Appl Physiol 2018;118: 33–49. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Murgatroyd PR, Davies HL, Prentice AM. Intra-individual variability and measurement noise in estimates of energy expenditure by whole body indirect calorimetry. Br J Nutr 1987;58: 347–356. [DOI] [PubMed] [Google Scholar]
- 22.Brychta RJ, Rothney MP, Skarulis MC, Chen KY. Optimizing energy expenditure detection in human metabolic chambers. Conf Proc IEEE Eng Med Biol Soc 2009;2009: 6864–6868. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Chakravarthy M, Parsons S, Lassman ME, Butterfield K, Lee AY, Chen Y, et al. Effects of 13-Hour Hyperglucagonemia on Energy Expenditure and Hepatic Glucose Production in Humans. Diabetes 2017;66: 36–44. [DOI] [PubMed] [Google Scholar]
- 24.Brown D, Cole TJ, Dauncey MJ, Marrs RW, Murgatroyd PR. Analysis of gaseous exchange in open-circuit indirect calorimetry. Med Biol Eng Comput 1984;22: 333–338. [DOI] [PubMed] [Google Scholar]
- 25.Melanson EL, Ingebrigtsen JP, Bergouignan A, Ohkawara K, Kohrt WM, Lighton JR. A new approach for flow-through respirometry measurements in humans. Am J Physiol Regul Integr Comp Physiol 2010;298: R1571–1579. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Henning B, Lofgren R, Sjostrom L. Chamber for indirect calorimetry with improved transient response. Med Biol Eng Comput 1996;34: 207–212. [DOI] [PubMed] [Google Scholar]
- 27.Chen S, Wohlers E, Ruud E, Moon J, Ni B, Celi FS. Improving temporal accuracy of human metabolic chambers for dynamic metabolic studies. PLoS One 2018;13: e0193467. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Nguyen T, de Jonge L, Smith SR, Bray GA. Chamber for indirect calorimetry with accurate measurement and time discrimination of metabolic plateaus of over 20 min. Med Biol Eng Comput 2003;41: 572–578. [DOI] [PubMed] [Google Scholar]
- 29.Tokuyama K, Ogata H, Katayose Y, Satoh M. Algorithm for transient response of whole body indirect calorimeter: deconvolution with a regularization parameter. J Appl Physiol (1985) 2009;106: 640–650. [DOI] [PubMed] [Google Scholar]
- 30.Ravussin E, Bogardus C. Relationship of genetics, age, and physical fitness to daily energy expenditure and fuel utilization. Am J Clin Nutr 1989;49: 968–975. [DOI] [PubMed] [Google Scholar]
- 31.Ohkawara K, Tanaka S, Ishikawa-Takata K, Tabata I. Twenty-four-hour analysis of elevated energy expenditure after physical activity in a metabolic chamber: models of daily total energy expenditure. Am J Clin Nutr 2008;87: 1268–1276. [DOI] [PubMed] [Google Scholar]
- 32.UNU, WHO. Human Energy Requirements: Report of a Joint FAO/WHO/UNU Expert Consultation: Rome, 17–24 October 2001. Food & Agriculture Org, 2004. [Google Scholar]
- 33.Schoffelen PF, Westerterp KR. Intra-individual variability and adaptation of overnightand sleeping metabolic rate. Physiol Behav 2008;94: 158–163. [DOI] [PubMed] [Google Scholar]
- 34.Reed GW, Hill JO. Measuring the thermic effect of food. Am J Clin Nutr 1996;63: 164–169. [DOI] [PubMed] [Google Scholar]
- 35.Zitting KM, Vujovic N, Yuan RK, Isherwood CM, Medina JE, Wang W, et al. Human Resting Energy Expenditure Varies with Circadian Phase. Curr Biol 2018;28: 3685–3690 e3683. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Plasqui G, Kester AD, Westerterp KR. Seasonal variation in sleeping metabolic rate, thyroid activity, and leptin. Am J Physiol Endocrinol Metab 2003;285: E338–343. [DOI] [PubMed] [Google Scholar]
- 37.van Dale D, Schoffelen PF, ten Hoor F, Saris WH. Effects of addition of exercise to energy restriction on 24-hour energy expenditure, sleeping metabolic rate and daily physical activity. Eur J Clin Nutr 1989;43: 441–451. [PubMed] [Google Scholar]
- 38.Piaggi P, Krakoff J, Bogardus C, Thearle MS. Lower “awake and fed thermogenesis” predicts future weight gain in subjects with abdominal adiposity. Diabetes 2013;62: 4043–4051. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Heilbronn LK, de Jonge L, Frisard MI, DeLany JP, Larson-Meyer DE, Rood J, et al. Effect of 6-month calorie restriction on biomarkers of longevity, metabolic adaptation, and oxidative stress in overweight individuals: a randomized controlled trial. JAMA 2006;295: 1539–1548. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Agnew HW, Webb WB, Williams RL. The first night effect: an EEG study of sleep. Psychophysiology 1966;2: 263–266. [DOI] [PubMed] [Google Scholar]
- 41.Shannon JR, Gottesdiener K, Jordan J, Chen K, Flattery S, Larson PJ, et al. Acute effect of ephedrine on 24-h energy balance. Clin Sci (Lond) 1999;96: 483–491. [PubMed] [Google Scholar]
- 42.Fontvieille AM, Rising R, Spraul M, Larson DE, Ravussin E. Relationship between sleep stages and metabolic rate in humans. Am J Physiol 1994;267: E732–737. [DOI] [PubMed] [Google Scholar]
- 43.Kayaba M, Park I, Iwayama K, Seya Y, Ogata H, Yajima K, et al. Energy metabolism differs between sleep stages and begins to increase prior to awakening. Metabolism 2017;69: 14–23. [DOI] [PubMed] [Google Scholar]
- 44.Gonnissen HK, Drummen M, Rosique Esteban N, Schoffelen PF, Westerterp-Plantenga MS. Overnight energy expenditure determined by whole-body indirect calorimetry does not differ during different sleep stages. Am J Clin Nutr 2013;98: 867–871. [DOI] [PubMed] [Google Scholar]
- 45.Westerterp KR. Diet induced thermogenesis. Nutr Metab (Lond) 2004;1: 5. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Ogata H, Kobayashi F, Hibi M, Tanaka S, Tokuyama K. A novel approach to calculating the thermic effect of food in a metabolic chamber. Physiol Rep 2016;4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Tataranni PA, Larson DE, Snitker S, Ravussin E. Thermic effect of food in humans: methods and results from use of a respiratory chamber. Am J Clin Nutr 1995;61: 1013–1019. [DOI] [PubMed] [Google Scholar]
- 48.Thearle MS, Pannacciulli N, Bonfiglio S, Pacak K, Krakoff J. Extent and determinants of thermogenic responses to 24 hours of fasting, energy balance, and five different overfeeding diets in humans. J Clin Endocrinol Metab 2013;98: 2791–2799. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Schutz Y, Bessard T, Jequier E. Diet-induced thermogenesis measured over a whole day in obese and nonobese women. Am J Clin Nutr 1984;40: 542–552. [DOI] [PubMed] [Google Scholar]
- 50.Westerterp KR, Wilson SA, Rolland V. Diet induced thermogenesis measured over 24h in a respiration chamber: effect of diet composition. Int J Obes Relat Metab Disord 1999;23: 287–292. [DOI] [PubMed] [Google Scholar]
- 51.Zuntz N, Schumberg W. Studien zu einer Physiologie des Marsches Bibliothek von Coler Bd. 6 Hirscwald: Berlin, 1901. [Google Scholar]
- 52.Weir JB. New methods for calculating metabolic rate with special reference to protein metabolism. J Physiol 1949;109: 1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Astrup A, Thorbek G, Lind J, Isaksson B. Prediction of 24-h energy expenditure and its components from physical characteristics and body composition in normal-weight humans. Am J Clin Nutr 1990;52: 777–783. [DOI] [PubMed] [Google Scholar]
- 54.Brouwer E On simple formulae for calculating the heat expenditure and the quantities of carbohydrate and fat oxidized in metabolism of men and animals, from gaseous exchange (Oxygen intake and carbonic acid output) and urine-N. Acta Physiol Pharmacol Neerl 1957;6: 795–802. [PubMed] [Google Scholar]
- 55.Consolazio CF, Johnson RE, Pecroa LJ. Physiological measurements of metabolic functions in man. McGraw-Hill: New York, 1963. [PubMed] [Google Scholar]
- 56.Jeukendrup AE, Wallis GA. Measurement of substrate oxidation during exercise by means of gas exchange measurements. Int J Sports Med 2005;26 Suppl 1: S28–37. [DOI] [PubMed] [Google Scholar]
- 57.Livesey G, Elia M. Estimation of energy expenditure, net carbohydrate utilization, and net fat oxidation and synthesis by indirect calorimetry: evaluation of errors with special reference to the detailed composition of fuels. Am J Clin Nutr 1988;47: 608–628. [DOI] [PubMed] [Google Scholar]
- 58.Skogerboe KJ, Labbe RF, Rettmer RL, Sundquist JP, Gargett AM. Chemiluminescent measurement of total urinary nitrogen for accurate calculation of nitrogen balance. Clin Chem 1990;36: 752–755. [PubMed] [Google Scholar]
- 59.Weyer C, Snitker S, Rising R, Bogardus C, Ravussin E. Determinants of energy expenditure and fuel utilization in man: effects of body composition, age, sex, ethnicity and glucose tolerance in 916 subjects. Int J Obes Relat Metab Disord 1999;23: 715–722. [DOI] [PubMed] [Google Scholar]
- 60.Heymsfield SB, Thomas D, Bosy-Westphal A, Shen W, Peterson CM, Muller MJ. Evolving concepts on adjusting human resting energy expenditure measurements for body size. Obes Rev 2012;13: 1001–1014. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Westerterp KR. Control of Energy Expenditure in Humans In: Feingold KR, Anawalt B, Boyce A, Chrousos G, Dungan K, Grossman A, et al. (eds). Endotext: South Dartmouth (MA), 2000. [Google Scholar]
- 62.Watson LP, Raymond-Barker P, Moran C, Schoenmakers N, Mitchell C, Bluck L, et al. An approach to quantifying abnormalities in energy expenditure and lean mass in metabolic disease. Eur J Clin Nutr 2014;68: 234–240. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Hall KD, Chen KY, Guo J, Lam YY, Leibel RL, Mayer LE, et al. Energy expenditure and body composition changes after an isocaloric ketogenic diet in overweight and obese men. Am J Clin Nutr 2016;104: 324–333. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Rumpler WV, Seale JL, Conway JM, Moe PW. Repeatability of 24-h energy expenditure measurements in humans by indirect calorimetry. Am J Clin Nutr 1990;51: 147–152. [DOI] [PubMed] [Google Scholar]
- 65.Toubro S, Christensen NJ, Astrup A. Reproducibility of 24-h energy expenditure, substrate utilization and spontaneous physical activity in obesity measured in a respiration chamber. Int J Obes Relat Metab Disord 1995;19: 544–549. [PubMed] [Google Scholar]
- 66.White MD, Bouchard G, Buemann B, Almeras N, Despres JP, Bouchard C, et al. Reproducibility of 24-h energy expenditure and macronutrient oxidation rates in an indirect calorimeter. J Appl Physiol (1985) 1996;80: 133–139. [DOI] [PubMed] [Google Scholar]
- 67.Melanson EL, Gavin KM, Shea KL, Wolfe P, Wierman ME, Schwartz RS, et al. Regulation of energy expenditure by estradiol in premenopausal women. J Appl Physiol (1985) 2015;119: 975–981. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Melanson EL, Lyden K, Gibbons E, Gavin KM, Wolfe P, Wierman ME, et al. Influence of Estradiol Status on Physical Activity in Premenopausal Women. Med Sci Sports Exerc 2018;50: 1704–1709. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 69.Janse DEJX, Thompson B, Han A. Methodological Recommendations for Menstrual Cycle Research in Sports and Exercise. Med Sci Sports Exerc 2019;51: 2610–2617. [DOI] [PubMed] [Google Scholar]
- 70.Melanson EL, Sharp TA, Seagle HM, Horton TJ, Donahoo WT, Grunwald GK, et al. Effect of exercise intensity on 24-h energy expenditure and nutrient oxidation. J Appl Physiol (1985) 2002;92: 1045–1052. [DOI] [PubMed] [Google Scholar]
- 71.Melanson EL, Ritchie HK, Dear TB, Catenacci V, Shea K, Connick E, et al. Daytime bright light exposure, metabolism, and individual differences in wake and sleep energy expenditure during circadian entrainment and misalignment. Neurobiol Sleep Circadian Rhythms 2018;4: 49–56. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 72.Hall KD, Bemis T, Brychta R, Chen KY, Courville A, Crayner EJ, et al. Calorie for Calorie, Dietary Fat Restriction Results in More Body Fat Loss than Carbohydrate Restriction in People with Obesity. Cell Metab 2015;22: 427–436. [DOI] [PMC free article] [PubMed] [Google Scholar]