Taller staff occupationally exposed to less radiation to the temple in cardiac procedures, but risk higher doses during vascular cases

Kelly S Wilson-Stewart 1,2,, Davide Fontanarosa 2,3, Dan Li 4,5,6, Chris C Drovandi 4,5,6, Rebecca K Anderson 1, Jamie V Trapp 2
PMCID: PMC7528059  PMID: 32999372

Abstract

This study aimed to evaluate the effect of nurse and doctor height on occupational dose to the temple during fluoroscopically guided cardiovascular procedures. Additionally, an evaluation of the relationship between doctor height and table height was performed. Staff exposed during fluoroscopic procedures may be at elevated risk of cardiovascular damage or oncogenesis and have demonstrated a higher incidence of subscapular cataracts. The heads of taller staff may be exposed to reduced levels of radiation due to the increased distance from the area of highest intensity X-ray scatter. Limited research has been performed investigating height as a predictor of head dose to nursing staff. The level of radiation dose at the level of the temple to the doctor (n = 25), scrub (n = 28), and scout nurse (n = 29) was measured in a prospective single-center, observational study using Philips DoseAware badges. Procedural characteristics were recorded for vascular and cardiac cases performed in three dedicated angiography suites. Data were also collected to investigate relationships between doctor height and table height. Data were collected for 1585 cardiac and 294 vascular procedures. Staff height was a statistically significant predictor of temple dose for doctors, scrub, and scout nurses when considering the full data sample. The log temple dose demonstrated an inverse relationship to staff height during cardiac procedures, but a positive relationship for scrub and scout nurses during vascular studies. This observational study has demonstrated that taller staff are exposed to less cranial exposure dose during fluoroscopically guided cardiac examinations but has revealed a positive correlation between height and temple dose during vascular procedures. It was also determined that doctor height was correlated with average procedural table height and that vascular access point influences the choice of table elevation.

Subject terms: Health occupations, Cardiology, Interventional cardiology

Introduction

Catheter-based procedures have transformed the way patients are diagnosed and treated in recent decades1. Diagnostic and interventional cases require the use of X-ray imaging to allow the doctor to visualize the anatomical location and appropriate deployment of interventional devices. As implantable products and imaging technology improve, there has been an increase in the complexity and number of procedures undertaken, which creates additional concerns regarding radiation dose to the patient and staff1,2.

Any exposure to ionizing radiation carries an inherent risk of adverse effects on tissue at a molecular level and may result in tissue reactions not only to the patient but exposed staff as well2. To maintain adherence to the ALARA (as low as reasonably achievable) principle, it is essential to understand the factors which contribute to occupational dose so that effective mitigation strategies can be implemented to minimize exposure3. This observational study investigates the effect of staff height on occupational dose to the temple during fluoroscopically guided cardiovascular procedures.

Background

In the United States, cardiovascular procedures contribute approximately 45% of the total cumulative effective dose to patients from all medical sources, excluding radiotherapy4. Given the assumed benefit to the patient and correct application of dose optimization techniques, this exposure can be justified2. In-room staff are also inadvertently exposed, receive no medical benefit, and may subsequently experience detrimental health effects. Resultant tissue effects following exposure to high levels of ionizing radiation have been well documented, but the impact of protracted, low dose radiation is less well understood5. Studies of occupationally exposed staff during fluoroscopic procedures have reported cardiovascular damage6, early vascular aging7 and an increased risk of oncogenesis8,9. Staff exposure to low levels of radiation has also resulted in a significantly higher incidence of subscapular cataracts when compared to unexposed populations1012.

The intensity of scattered radiation is highest on the X-ray beam entrance side of the patient. Consequently, the X-ray tube is positioned under the operating table in modern fluoroscopy suites to reduce dose to the head, torso and upper extremities of staff13. One underpinning principle of radiation physics is the inverse square rule, which states that the intensity of radiation will decrease proportionally according to the inverse square of the distance from the radiation source14. Although scattered radiation from patient procedures is not a point-source and does not strictly follow the inverse square law, this dictum has implications for doctors and other in-room staff2,15, and may lead to the presumption that the increased vertical distance between the area of highest scatter and a taller staff member’s head may result in decreased dose.

There is limited research that investigates the effect of doctor height on occupational radiation exposure. Among the existing studies, there is almost universal agreement that height has an inverse relationship to dose (Table 1). To date, there is a notable lack of literature explicitly investigating the relationship between the height of staff other than the doctors, such as scrub and scout nurses, and occupational exposure during fluoroscopic procedures.

Table 1.

Methodology and conclusion summary of previous investigations of staff height as a predictor of occupational dose during fluoroscopic procedures.

Author Clinical (C)/phantom (P) Doctor Other staff Measured location Specialty Result
Albayati et al.16 C (n = 17) ✓ (n = 3) Assistant doctor Radiologist (n = 2) Head Thorax Peripheral vascular/radiology

Inverse relationship between doctor height and head dose

No relationship between doctor height and thorax dose

No relationship between assistant doctor height and measured doses

Kuon et al.17 P Thorax Cardiology Inverse relationship between doctor height and dose
Principi et al.18

C (n = 13)

P

✓ (n = 6) Eye Radiology /cardiology

Inverse relationship between doctor height and dose

A reduction factor up to 2 is observed for a 10-cm-taller doctora

Rigatelli et al.19

C (n = 2630)

P

✓ (n = 4) Nurses (n = 9) Technicians (n = 7) Thorax Peripheral vascular/cardiology

Inverse relationship between staff height and dose for both clinical and phantom measurements independent on procedure type

Result was consistent with/without the use of a CMLSa

Sciahbasi et al.20 C (n = 2028) ✓ (n = 6)

Thorax

Head

Left wrist

Cardiology Inverse relationship between height and dose measured at thorax
Willard et al.31 P Head Urology Shorter surgeons receive higher scattered radiation exposure to the brain and eyes during fluoroscopy

CMLS ceiling-mounted lead shield; cm centimetre; n number of procedures in clinical studies, or the number of individual staff included in the study; ✓, staff role included in the study sample; ✗, staff role not included in the study sample; C, clinical study, P, phantom study

aNoteworthy.

Aim

This prospective observational study aimed to evaluate the effect of nurse and doctor height on occupational temple dose during fluoroscopically guided cardiovascular procedures. A secondary aim assessed the legitimacy of the assumption that taller doctors operate using higher patient table levels during angiographic procedures.

Methodology

Radiation dose levels were measured in a prospective single-center, observational study to evaluate occupational radiation exposure for the staff performing the roles of scrub nurse (n = 28), scout nurse (n = 29), cardiologist (n = 22) and vascular surgeon (n = 3) between February 2017 and August 2019. Approval was granted by the Ramsay Health Care QLD Human Research Ethics Committee (Protocol number—16/67) and informed, written consent was obtained from staff participants. Written informed consent was also obtained to publish identifying images in an online open-access publication. The research was conducted in accordance with the National Health and Medical Research Council guidelines.

Procedural characteristics and occupational dose data were collected for cardiovascular procedures performed in three dedicated angiography theatres using Philips Allura Xper equipment (Philips Healthcare, Best, Netherlands). Two of the systems had Clarity dose reducing software installed, and one did not (Room 1). As Room 3 was equipped with a larger detector and dedicated vascular software applications, the majority of the non-cardiac procedures, subsequently referred to as ‘vascular’, were performed in this room.

Case characteristics such as staff category, room, procedure, contrast type and volume, nature of any intervention, and access point were recorded. Cumulative procedural values for radiological parameters including fluoroscopy time, number of cine acquisitions, and patient dose information such as Dose Area Product (DAP) and Air Kerma (AK) were retrieved from the generated procedural dose reports and also documented. All of these variables were incorporated for possible selection in the regression model so that more accurate statements about the variables of interest could be obtained after accounting for other potential variables. Procedures were categorized as cardiac or vascular, diagnostic or interventional, and allocated depending on patient body part imaged.

Staff height and temple dose

Discrete measurements of occupational radiation exposure were prospectively collected for the doctor (cardiologist or vascular surgeon), the scrub and scout nurse using Dose Aware badges (Philips Healthcare, Best, Netherlands). DoseAware is an active personal dosimeter (APD) that features a real-time exposure monitoring system. Previous research has stated that DoseAware responds satisfactorily in realistic fields in interventional cardiology scattered radiation fields21,22. In addition to individual calibration certificates supplied by the manufacturer, DoseAware readings have been verified when compared to thermoluminescent dosimetry by the Institut de Tècniques Energètiques—Universitat Politècnica de Catalunya, Barcelona21. Individual badge calibration was also tested before the commencement of data collection.

DoseAware dosimeter badges have an operational quantity of Hp(10) and an X-ray dose rate range and linearity of ± 10% 40μSv/h–150 mSv/h (Sievert)23. Ideally, eye dose should be measured using Hp(3) dosimeters, unfortunately, at the time of data collection, commercially available dosimeters specifically calibrated to Hp(3) were uncommon. It has been noted that DoseAware worn close to the eye can provide appropriate dose estimates21, but the response may present an overestimation of 5–15%22.

Given the varying rotational position of the X-ray tube during procedures, the International Electrotechnical Commission recommends an angular response from 00 to 600 for the energy range 20–100 kev24. Philips states an angular response of ± 5% within ± 5°, ± 30% within ± 50° and + 200%/− 100% within ± 90° for an energy range up to 120 kev23.

When scrub nurse or doctor were located with their left side closest to the X-ray tube, as in a coronary angiogram, badges were worn near the left temple which has previously been demonstrated to be the preferred badge position for measuring eye dose25. For procedures where the staff worked on the opposite side of the patient, such as insertion of permanent pacemakers (PPM), badges were worn on the right temple. Staff wore the badge according to where they anticipated they would be located for the majority of the case. The most common position adopted by the scout nurse during procedures was nearest the doors to the room; generally 2–3 m behind the scrub nurse and consequently badges were typically situated on the left temple. Dosimeter badges were located on the left temple of staff in > 97% of procedures.

Coronary cases, including left ventriculograms (LV) and aortograms, were performed with a cine and fluoroscopy rate of 15 frames per second (fps). PPM and electrophysiology (EP) procedures had a standard low dose acquisition protocol of 7.5 fps for both cine and fluoroscopy. Endovascular cases used a fluoroscopic frame rate of 7.5 fps. Digital subtraction angiography (DSA) acquisitions varied from 6 fps for endovascular aneurysm repairs (EVAR), 3 fps for abdominal and pelvic imaging, and reduced to 0.5 fps for distal leg vessels.

All doctors had a minimum of 15 years’ clinical experience in catheter-based cardiovascular procedures. All staff wore lead gowns and thyroid shields during cases, and most doctors and scrub nurses opted to wear lead glasses. Scrub and scout nurses routinely utilized leaded skull caps, while less than 10% of doctors chose to wear them. Depending on staff preference, the badge was either clipped external to the glasses, as demonstrated in Fig. 1 or fixed to the outside of the skull or theatre cap. Both ceiling-mounted lead shields (CMLS) and table-mounted lead shields (TMLS) were consistently used, and an extra movable lead shield was regularly utilized to provide additional protection to the scout nurse as shown in Fig. 2a. The relationship between the height of scrub nurse, scout nurse and doctor, and the levels of occupational temple dose was evaluated.

Figure 1.

Figure 1

Position of DoseAware badge when worn attached to glasses.

Figure 2.

Figure 2

(a, b) Typical room setup and staff location. CMLS ceiling-mounted lead shield, TMLS table-mounted lead shield, D doctor, SN scrub nurse, ST scout nurse.

The impact of doctor height on table height

To determine whether doctor height was correlated with average procedural table height (Fig. 2b), data were collected in 202 cases. The procedural average was determined by documenting table height data from the imaging system during individual fluoroscopic or cine events and calculating the mean table height for each procedure. Other potentially influential variables such as room number, access point and patient body mass index (BMI) were also recorded. At the author’s institution, the radiographer controls the X-ray tube geometry and the table height, and hence the effect of individual radiographers on table height was also investigated. Occupational dose was not examined.

Statistical analysis

Staff height and temple dose

Linear regression models were applied to the data with the staff temple dose (log-transformed) as the response variable with the potential explanatory variables, such as staff role/height, patient BMI, room number, procedure type, imaged body part, entry point, and patient/case dose parameters. Two-way interactions between covariates were also considered. If a covariate appeared in a significant interaction, it was also included as a main effect regardless of whether the main effect was significant or not. The Akaike Information Criterion (AIC) was used to select among competing regression models26.

The impact of doctor height on table height

A linear regression analysis was performed to determine the potential effect of dr height, room number, radiographer, vascular access point and patient BMI on the procedural table height.

Results

Staff height and temple dose

Data were collected for 1585 cardiac and 294 vascular procedures. Males constituted approximately 70% of patients, and mean patient BMI was 29.6 for coronary and 28.4 for vascular procedures. Basic procedural information is presented in Table 2. Routine tube angles utilized for both coronary and vascular procedures are provided in Table 3. Staff height ranged from 155 to 190 cm for nurses and 167.5–192 cm for doctors, with the mean and quartiles represented in Table 4. The average dose per procedure was 3.14 μSv for the cardiologist and 6.6 μSv for the vascular surgeon. The scrub nurse received a mean dose of 2.8 μSv/5.6 μSv and the scout nurse a dose of 2.0/0.50 μSv for cardiac and vascular procedures, respectively.

Table 2.

Procedure type and case numbers.

Procedure category Procedure type Number
Coronary Coronary angiography—diagnostic 866
Coronary angiography—interventional 534
Total 1400
TAVI TAVI workups 21
TAVI/Balloon valvuloplasty 24
Total 45
Implantables PPM 67
Watchman/amulet 14
ASD/PFO closure 9
Balloon pump 1
Total 91
EP Total Diagnostic and interventional 48
Vascular Abdominal angiography—diagnostic 20
Abdominal angiography—interventional 42
Abdominal + lower limb/s—diagnostic 28
Abdominal + lower limb/s—interventional 48
Lower limb diagnostic 25
Lower limb interventional 68
Subclavian/fistulogram/carotids—diagnostic 5
Subclavian/fistulogram/carotids—interventional 20
EVAR/FEVAR 38
Total 294

TAVI transcatheter aortic valve implantation, PPM permanent pacemaker, ASD atrial septal defect, PFO patent foramen ovale, EP electrophysiology, EVAR endovascular aneurysm repair, FEVAR fenestrated endovascular aneurysm repair.

Table 3.

Routinely utilized procedural tube angles.

Procedure LAO ↑ (35°/25°) LAO ↓ (40°/35°) RAO ↓ (20°/20°) RAO ↑ (30°/30°) AP ↑ (10°/35°) AP (0°/0°) RAO (30°) LAO (30°) LAO ↑ (25°/10°) RAO (20°) LAO
(20°)
LAO (40°)
Cardiac
Coronary angiogram
LCA
RCA
LV
TAVI workup
TAVI a
PPM
EP
Vascular
EVAR √ (10° ↓) √ (10° ↓)
Abdominal aorta
leg
subclavian
fistulograms

LCA left coronary artery, RCA right coronary artery, LV left ventriculogram, TAVI transcatheter aortic valve implantation, PPM permanent pacemaker, EP electrophysiology, EVAR endovascular aneurysm repair, LAO left anterior oblique, RAO right anterior oblique, AP anterior–posterior, ↑ cranial angulation, ↓ caudal angulation.

aIndividually customized implantation angle.

Table 4.

Staff height per procedural category.

Procedure type Staff role N Mean Std Dev Min 1st quartile Median 3rd quartile Max
Coronary Doctor 1234 174 6.2 168 169 172 178 192
Scrub 1147 173 10.0 157 165 170 181 190
Scout 677 171 9.3 157 165 167 178 190
Vascular Doctor 194 179 6.5 168 176 176 185 185
Scrub 231 170 9.6 155 165 166 178 190
Scout 143 173 9.9 160 166 174 180 190

Measured in centimeters (cm) expressed as mean, standard deviation, minimum, 1st Quartile, median, 3rd Quartile and maximum.

Staff height was a statistically significant predictor of temple dose for doctors, scrub and scout nurses when considering the full data sample. To quantify the impact of staff height on the temple dose with respect to a specific categorical variable that has significant interaction with staff height, other significantly interactive variables were considered. For different selected values of the interactive continuous variable, contrast volume, the impact was averaged over other interactive categorical variables.

The imaged area of the patient’s body was found to influence temple dose with a decrease in exposure for taller staff (doctor, scrub and scout) within procedures performed on both the abdomen and extremities (predominantly aorto-femoral angiograms) and the extremities (Fig. 3a–c). For abdominal imaging, including endovascular aortic repairs (EVAR), and chest imaging, including transcatheter aortic valve implantation (TAVI) procedures, there was a positive relationship between staff height and temple dose for both scrub and scout nurses regardless of contrast volume. This relationship was also demonstrated for doctors in procedures within the chest (Fig. 3d) when contrast usage was 146 ml or 266 ml (75% and 95% quantile, respectively), but the impact of doctor height on temple dose was negative for procedures in which the contrast volume equalled 15 ml, 60 ml or 95 ml/60/95 ml (5%, 25% and 50% quantile, respectively).

Figure 3.

Figure 3

Impact of staff height on log temple dose with respect to body part imaged and different staff roles, doctor (a), scrub nurse (b), and scout nurse (c), for the median value of contrast volume. (d) demonstrates doctor dose with 95% quantile of contrast volume.

Coronary and vascular procedures were separately analyzed, and the relationship between occupational temple dose and the doctor, scrub nurse and scout nurse height are presented in Fig. 4. As expected, taller staff are exposed to less cranial exposure during cardiac procedures, but surprisingly this trend was reversed for scrub and scout nurses during vascular cases.

Figure 4.

Figure 4

Impact of staff height on log temple for coronary (a) and vascular (b) procedures and different staff roles, for the median value of contrast volume.

Table 5 demonstrates the changes for log temple dose for different staff height during coronary and vascular procedures, again highlighting the trend of increasing temple dose to taller staff members during vascular procedures.

Table 5.

Log temple dose change for different levels of staff height.

Procedure type Contrast (ml) Staff height (reference level: 160 cm)
170 cm 180 cm 190 cm
Coronary 15 − 0.55 − 1.11 − 1.66
60 − 0.52 − 1.02 − 1.53
95 − 0.48 − 0.97 − 1.45
146 − 0.44 − 0.88 − 1.32
266 − 0.34 − 0.67 − 1.01
Vascular 15 0.11 0.23 0.34
60 0.15 0.31 0.46
95 0.18 0.37 0.55
146 0.23 0.46 0.69
266 0.33 0.67 1.00

The impact of doctor height on table height

Procedural table height was collected in all three rooms for 21 doctors, and for a total of 202 procedures (coronary n = 192; vascular n = 10) as shown in Table 6.

Table 6.

Procedures, doctors and average table height.

Room procedures No. doctors Mean TH Std Dev TH Min TH Max TH
1 62 4 92.5 cm 19.9 86 110
2 88 16 91.6 2.2 86 99
3 52 16 91 2.09 85 95

Measured in centimeters (cm).

TH table height.

There is strong evidence of a positive linear relationship between doctor height and procedural table height in room 1(Fig. 5a). The positive relationship in rooms 2 and 3 could become negative with a high patient BMI such as the 95% quantile BMI (Fig. 5b). There was little evidence of any impact on table height by individual radiographers.

Figure 5.

Figure 5

Impact of doctor height (cm) on table height for the median patient BMI (a) of 28.40 and the 95% quantile BMI (b) of 38.61.

The vascular access point had a statistically significant effect on procedural table height, and the average table heights for different vascular access points are demonstrated in Fig. 6.

Figure 6.

Figure 6

Average table height (cm) for different vascular access points.

Discussion

Staff height and temple dose

It stands to reason that the taller the staff member, the larger the distance between their head and the source of greatest scatter and consequently should be exposed to lower levels of cranial radiation than their shorter counterparts (Fig. 7).

Figure 7.

Figure 7

The heads of taller staff (a) may be exposed to less scattered radiation than shorter staff (b). D doctor, SN scrub nurse.

There has been limited research investigating the effect of doctor height on occupational head dose, but of the available studies, the majority have demonstrated an inverse relationship between height and dose (Table 1). Kuon et al.27 found that scatter entrance skin dose decreased with greater body height, with taller doctors having a reduction in mean doctor head dose of approximately 1%/cm. Albayati et al. report a fivefold increase in measured dose for a shorter doctor (170 cm) when compared to taller colleagues, despite procedural parameters being similar16. Comparable to previous studies, analysis of the combined cardiac dataset reveals an overall inverse relationship between staff height and radiation dose to the head.

There is a notable paucity of literature explicitly evaluating the height of additional in-room personnel as a potential predictor of dose, and this may be due to the assumption that the proximity of the doctor to the primary source of X-ray scatter results in an additional dose. Rigatelli et al. measured dose in the clinical setting at the chest pocket level and classified doctors, nurses and technologists into two groups: those > 165 cm and those < 165 cm and reported significantly lower doses to taller individuals than their shorter colleagues, independent on procedure type, staff position or the use of a CMLS19. The effect of height on doses to individual staff roles in the clinical setting was not specifically articulated or discussed. Albayati el al. demonstrated no relationship between height and dose to an assistant doctor (radiologist)16. If it were assumed that the procedural positioning of an assistant doctor is similar to that of a scrub nurse, this investigation reveals important differences.

The results of this study indicate a strong positive height-temple dose relationship for scrub nurses after averaging over other categorical variables that have significant interactions with height for a collection of sample quantiles values of contrast volumes of 15 ml, 60 ml, 95 ml, 146 ml, and 266 ml (i.e. 5%, 25%, 50%, 75%, and 95% quantile, respectively), which contrasts with the previous (albeit limited) studies. BMI was found to affect dose levels to staff. This is unsurprising as the greater thickness of tissue an X-ray beam has to traverse, the higher number of X-rays are required, with a proportionally more substantial amount of scattered radiation incident on staff, which has been demonstrated previously28. When relatively large volumes of contrast were used, this was also shown to alter the trends of occupational dose as a predictor of temple dose, and this effect is understandable as larger volumes of contrast administered during an angiogram may be an indicator of greater procedural complexity.

Scrub nurses are generally positioned further away from the most significant source of x-ray scatter, which suppositionally reduces their level of exposure compared to the doctor. Still, it has been reported that nursing staff may be exposed to higher levels of radiation than the doctor due to the shelter provided by the CMLS2830, and this may be a contributing factor in the increased dose to scrub nurses. The presence of the X-ray detector has also been identified as providing an incidental protective barrier to doctors’ heads by absorbing X-ray photons, depending on the position of the doctor18,31. While these elements may rationalize higher doses to the nurses, they do not explain the positive height-dose relationship for nurses, or why the effect is more pronounced for procedures which image the chest or abdomen.

One possibility is that the CMLS, while not providing the same level of protection afforded to the doctor, nevertheless offers a degree of cover from scattered radiation. In this case, the heads of shorter nurses would more likely be protected by this shield regardless of the doctor's height, and related positional height of the CMLS. In contrast, the heads of taller scrub nurses may extend above the lead glass, especially when the doctor may position it lower, allowing for scattered radiation to project above the shield and impinge on the cranium of the scrub nurse. It has been demonstrated that doses during endovascular procedures may be higher than that of coronary32,33 and this may explain why there is a marked difference between both the height-dose relationship when comparing coronary procedures to vascular (Fig. 4), but again fails to account for the positive height-dose relationship for scrub and scout nurses during procedures imaging the chest, which largely constitutes cardiac procedures.

It should be acknowledged that there is a wide range of scatter radiation at the level of the doctors’ eyes due to the variability of tube position34 and routine angles may differ between institutions or due to doctor preference. Procedures routinely utilizing steeper left anterior obliques (LAO) or LAO cranial tube angles provided in Table 3 may increase the risk of occupational exposure17,20. Additionally, these measurements were taken at a single point, and due to the changing nature of the X-ray scatter profile may not reflect dose levels to other areas of the head.

When extrapolated to anticipated yearly caseloads, the mean procedural doses for all staff categories during both vascular and cardiac procedures were < 2 mSv/year, which is well below the ICRP recommended yearly limits of 20 mSv/year1.

The impact of doctor height on table height

Some authors have made the assumption that taller doctors would utilize a higher table position for ergonomic preference20,35,36, and this was corroborated by Faroux et al.37. Although this presumption seems logical, the interdependence was examined in the local clinical environment before this premise was applied to the findings of this study.

As expected, doctor height positively influenced table height when the patient’s BMI was not extremely high (Fig. 3a), with the strongest interrelationship demonstrated in room 1. The vascular access point was found to influence table height, with the highest average table height found to be during the insertion of pacemaker wires via the subclavian vein (Fig. 6). This association would be anticipated, as the shoulder is generally a much thinner body part than the hip, and the superficial location of the subclavian vein would lead itself to a higher table elevation. No relationship was demonstrated between individual radiographers and table height.

The positive relationship between doctor height and table height was mitigated with an increase in patient BMI; for instance, the previous positive relationship between table height and doctor height was negative when the patient’s BMI was at the 95% quantile level (BMI = 38.61). It is postulated that this may be due to larger patients having a skin access point higher above the surface of the table than that of a thinner patient, requiring a lower table height for doctor comfort.

Studies with similar case numbers20, as well as comparable procedures19, report an inverse relationship between height and dose measured at the level of the thorax, which is similar to the findings for the cardiac cases included in this investigation. In contrast to this, the results of this study indicate that nursing staff receive higher levels of occupational radiation dose to the left temple during procedures which image the chest and the abdomen and show a positive height-temple dose relationship during vascular procedures. This study investigated the dose to 28 nursing staff and 25 doctors, as opposed to Rigatelli et al.’s four doctors, nine nurses and seven technicians, and Sciahbasi et al.’s six doctors, which may also explain the discrepancy in results due to the increased number of variables. The measurement of dose at the height of the temple, as opposed to the thorax in this study, may have also contributed to the disparity in results.

Limitations

This study was an observational design based on data from a single-center, which represents the most significant limitation. Specific procedural protocols, imaging equipment, and staff operational preferences are potentially limiting to the generalisability of the results.

Another limitation is the assessment of the effect of staff height on occupational doses at a single anatomical location only, and results may have been strengthened by data from other areas of the body. Dosimeters were also worn external to protective apparel, which results in an overestimation of actual temple dose. It should be noted, though, that this study intended to investigate dose trends with increasing staff height, rather than quantitatively reporting on discrete temple doses.

Conclusion

It is acknowledged that measuring the effect of one variable during fluoroscopically controlled procedures is problematic. Phantom measurements are useful for investigating parameters in isolation, but findings may be simplistic and may not reflect the complicated clinical reality. Conversely, results that are based on procedural data may have limited transferability due to the many variables that cannot realistically be considered in isolation, given a large number of uncontrollable procedural differences.

This observational study has demonstrated an inverse relationship between combined staff height and occupational temple dose during fluoroscopically guided cardiac examinations but has revealed an important positive relationship between height and temple dose during vascular procedures. Taller nurses performing the role of scrub or scout nurse during vascular angiography may be at greater risk of occupational radiation dose to the head than their shorter counterparts. This is contrary to previous findings in studies with similar procedural parameters investigating non-doctor dose, and the results may reflect the variability of differences in clinical practice. While increased patient and staff dose have been previously noted for vascular procedures, it is unclear the reasons why height in this setting influences dose, and additional research is required to investigate possible explanations. It was also determined that doctor height was correlated with average procedural table height and that vascular access point influences the choice of table elevation.

Acknowledgements

We thank our nursing, radiography and medical colleagues from the cardiovascular suites at Greenslopes Private Hospital for their contributions towards data collection during the course of this research.

Author contributions

K.W.S.: conceptualization and design, methodology, data collection, software, investigation, writing, project administration. D.F.: review and editing, supervision. D.L.: resources (analysis tools), data analysis, review and editing. C.D.: resources (analysis tools), data analysis, review and editing. R.A.: design, data collection, software, project administration. J.T.: review and editing, supervision.

Data availability

The datasets generated during the current study are available from the corresponding author on a reasonable request.

Competing interests

The authors declare no competing interests.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.López PO, et al. ICRP Publication 139: Occupational radiological protection in interventional procedures. Ann. ICRP. 2018;47:1–118. doi: 10.1177/0146645317750356. [DOI] [PubMed] [Google Scholar]
  • 2.Hirshfeld JW, et al. 2018 ACC/HRS/NASCI/SCAI/SCCT expert consensus document on optimal use of ionizing radiation in cardiovascular imaging: best practices for safety and effectiveness: a report of the American College of Cardiology task force on expert consensus decision pathways. J. Am. Coll. Cardiol. 2018;71:e283–e351. doi: 10.1016/j.jacc.2018.02.016. [DOI] [PubMed] [Google Scholar]
  • 3.The 2007 Recommendations of the International Commission on Radiological Protection ICRP publication 103. Ann. ICRP. 2007;37:1. doi: 10.1016/j.icrp.2007.10.003. [DOI] [PubMed] [Google Scholar]
  • 4.Picano E, Vano E. The radiation issue in cardiology: The time for action is now. Cardiovasc. Ultras. 2011;9:35. doi: 10.1186/1476-7120-9-35. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.McLean AR, et al. A restatement of the natural science evidence base concerning the health effects of low-level ionizing radiation. Proc. Biol. Sci. 2017 doi: 10.1098/rspb.2017.1070. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Picano E, Vano E, Domenici L, Bottai M, Thierry-Chef I. Cancer and non-cancer brain and eye effects of chronic low-dose ionizing radiation exposure. BMC Cancer. 2012;12:157. doi: 10.1186/1471-2407-12-157. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Andreassi MG, et al. Subclinical carotid atherosclerosis and early vascular aging from long-term low-dose ionizing radiation exposure: A genetic, telomere, and vascular ultrasound study in cardiac catheterization laboratory staff. JACC. Cardiovasc. Interv. 2015;8:616–627. doi: 10.1016/j.jcin.2014.12.233. [DOI] [PubMed] [Google Scholar]
  • 8.Ho TL, Shieh SH, Lin CL, Shen WC, Kao CH. Risk of cancer among cardiologists who frequently perform percutaneous coronary interventions: A population-based study. Eur. J. Clin. Invest. 2016;46:527–534. doi: 10.1111/eci.12628. [DOI] [PubMed] [Google Scholar]
  • 9.Roguin A, Goldstein J, Bar O, Goldstein JA. Brain and neck tumors among physicians performing interventional procedures. Am. J. Cardiol. 2013;111:1368–1372. doi: 10.1016/j.amjcard.2012.12.060. [DOI] [PubMed] [Google Scholar]
  • 10.Jacob S, et al. Interventional cardiologists and risk of radiation-induced cataract: Results of a French multicenter observational study. Int. J. Cardiol. 2013;167:1843–1847. doi: 10.1016/j.ijcard.2012.04.124. [DOI] [PubMed] [Google Scholar]
  • 11.Vano E, Kleiman NJ, Duran A, Romano-Miller M, Rehani MM. Radiation-associated lens opacities in catheterization personnel: Results of a survey and direct assessments. J. Vasc. Interv. Radiol. 2013;24:197–204. doi: 10.1016/j.jvir.2012.10.016. [DOI] [PubMed] [Google Scholar]
  • 12.Ciraj-Bjelac O, et al. Radiation-induced eye lens changes and risk for cataract in interventional cardiology. Cardiology. 2012;123:168–171. doi: 10.1159/000342458. [DOI] [PubMed] [Google Scholar]
  • 13.Miller D, et al. Occupational radiation protection in interventional radiology: A joint guideline of the Cardiovascular and Interventional Radiology Society of Europe and the Society of Interventional Radiology. Cardiovasc. Intervent. Radiol. 2010;33:230–239. doi: 10.1007/s00270-009-9756-7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Carroll QB. Radiography in the Digital Age: Physics-Exposure-Radiation Biology. Springfield: Charles C Thomas Publisher; 2018. [Google Scholar]
  • 15.Massalha S, et al. Strategies for minimizing occupational radiation exposure in cardiac imaging. Curr. Cardiol. Rep. 2019;21:1–6. doi: 10.1007/s11886-019-1157-1. [DOI] [PubMed] [Google Scholar]
  • 16.Albayati MA, et al. Angulation of the c-arm during complex endovascular aortic proceduresincreases radiation exposure to the head. Eur. J. Vasc. Endovasc. Surg. 2015;49:396–402. doi: 10.1016/j.ejvs.2014.12.032. [DOI] [PubMed] [Google Scholar]
  • 17.Kuon E, Empen K, Reuter G, Dahm JB. Personal operator dose in invasive cardiology as a function of body height and tube angulation. Rofo. 2004;176:739–745. doi: 10.1055/s-2004-813010. [DOI] [PubMed] [Google Scholar]
  • 18.Principi S, et al. The influence of operator position, height and body orientation on eye lens dose in interventional radiology and cardiology: Monte Carlo simulations versus realistic clinical measurements. Phys Med. 2016;32:1111–1117. doi: 10.1016/j.ejmp.2016.08.010. [DOI] [PubMed] [Google Scholar]
  • 19.Rigatelli G, et al. Impact of operators’ height on individual radiation exposure measurements during catheter-based cardiovascular interventions. J. Interv. Cardiol. 2016;29:83–88. doi: 10.1111/joic.12263. [DOI] [PubMed] [Google Scholar]
  • 20.Sciahbasi A, et al. Determinants of operator radiation exposure during percutaneous coronary procedures. Am. Heart J. 2017;187:10–18. doi: 10.1016/j.ahj.2017.02.012. [DOI] [PubMed] [Google Scholar]
  • 21.Chiriotti S, et al. Performance of several active personal dosemeters in interventional radiology and cardiology. Radiat. Meas. 2011;46:1266–1270. doi: 10.1016/j.radmeas.2011.05.073. [DOI] [Google Scholar]
  • 22.Principi S, et al. Influence of dosemeter position for the assessment of eye lens dose during interventional cardiology. Radiat. Prot. Dosim. 2015;164:79–83. doi: 10.1093/rpd/ncu359. [DOI] [PubMed] [Google Scholar]
  • 23.Harrysson, H. DoseAware Base Station Package User Manual (ed. Philips Healthcare). https://fccid.io/XWK8603021/User-Manual/User-manual-Dose-aware-1228983 (2010).
  • 24.International Electrotechnical Commission. Radiation protection instrumentation - measurement of personal dose equivalent Hp(10) and Hp(0.07) for X, gamma, neutron and beta radiations - direct reading personal dose equivalent and/or dose equivalent rate dosemeters. International Standard IEC 61526 (IEC, Geneva, 2010).
  • 25.Geber T, Gunnarsson M, Mattsson S. Eye lens dosimetry for interventional procedures—Relation between the absorbed dose to the lens and dose at measurement positions. Radiat. Meas. 2011;46:1248–1251. doi: 10.1016/j.radmeas.2011.07.028. [DOI] [Google Scholar]
  • 26.Akaike H. Selected Papers of Hirotugu Akaike. New York: Springer; 1998. pp. 199–213. [Google Scholar]
  • 27.Kuon E, Birkel J, Schmitt M, Dahm JB. Radiation exposure benefit of a lead cap in invasive cardiology. Heart. 2003;89:1205–1210. doi: 10.1136/heart.89.10.1205. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28.Wilson-Stewart K, Hartel G, Fontanarosa D. Occupational radiation exposure to the head is higher for scrub nurses than cardiologists during cardiac angiography. J. Adv. Nurs. 2019 doi: 10.1111/jan.14085. [DOI] [PubMed] [Google Scholar]
  • 29.Omar A, et al. Assessment of the occupational eye lens dose for clinical staff in interventional radiology, cardiology and neuroradiology. J. Radiol. Prot. 2017;37:145–159. doi: 10.1088/1361-6498/aa559c. [DOI] [PubMed] [Google Scholar]
  • 30.Božović P, Ciraj-Bjelac O, Petrović J. Occupational eye lens dose estimated using whole-body dosemeter in interventional cardiology and radiology: A Monte Carlo study. Radiat. Prot. Dosim. 2019;185:135–142. doi: 10.1093/rpd/ncy283. [DOI] [PubMed] [Google Scholar]
  • 31.Willard J, et al. Does the surgeon's height and distance from the c-arm affect brain and eye radiation exposure during fluoroscopy? J. Urol. 2018;199:E841–E841. [Google Scholar]
  • 32.Vano E, Fernandez JM, Resel LE, Moreno J, Sanchez RM. Staff lens doses in interventional urology. A comparison with interventional radiology, cardiology and vascular surgery values. J. Radiol. Prot. 2016;36:37–48. doi: 10.1088/0952-4746/36/1/37. [DOI] [PubMed] [Google Scholar]
  • 33.McLean D, Hadaya D, Tse J. Eye dose to staff involved in interventional and procedural fluoroscopy. J. Phys. Conf. Series. 2016 doi: 10.1088/1742-6596/694/1/012054. [DOI] [Google Scholar]
  • 34.Leyton F, Nogueira MS, Gubolino LA, Pivetta MR, Ubeda C. Correlation between scatter radiation dose at height of operator's eye and dose to patient for different angiographic projections. Appl. Radiat. Isot. 2016;117:100–105. doi: 10.1016/j.apradiso.2016.01.013. [DOI] [PubMed] [Google Scholar]
  • 35.Omar A, Marteinsdottir M, Kadesjo N, Fransson A. On the feasibility of utilizing active personal dosimeters worn on the chest to estimate occupational eye lens dose in X-ray angiography. J. Radiol. Prot. 2015;35:271–284. doi: 10.1088/0952-4746/35/2/271. [DOI] [PubMed] [Google Scholar]
  • 36.Prater S, Rees C, Bruner A, Savage C. Determination of minimum effective height of transparent radiation face shielding for fluoroscopy. Health Phys. 2011;101:S135. doi: 10.1097/HP.0b013e31821ec5bf. [DOI] [PubMed] [Google Scholar]
  • 37.Faroux L, et al. Impact of the table height and the operator's height on the level of radiation delivered to interventional cardiologists. Radiat. Prot. Dosim. 2019;187:21–27. doi: 10.1093/rpd/ncz131. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The datasets generated during the current study are available from the corresponding author on a reasonable request.