Preliminary Clinical Evaluation of Intrafraction Prostate Displacements for Two Immobilization Systems

Farshad Mostafaei 1, Shona T Dougherty 2, Russell J Hamilton 3,
Editors: Alexander Muacevic, John R Adler
PMCID: PMC7532867  PMID: 33033682

Abstract

Immobilization systems and their corresponding set-up errors influence the clinical target volume to the planning target volume (CTV-PTV) margins, which is critical for hypofractionated prostate stereotactic body radiotherapy (SBRT). This preliminary study evaluates intrafraction prostate displacement for two immobilization systems (A and B). Six consecutive patients having localized prostate cancer and implanted prostate marker seeds were studied. Planar X-ray images were acquired pre- and post-treatment to find the intrafraction prostate displacement. The average absolute displacements (lateral, longitudinal, vertical) were 0.9 ± 0.4 mm, 1.7 ± 0.1 mm, 1.3 ± 0.3 mm (system A), and 0.5 ± 0.2 mm, 0.6 ± 0.1 mm, 0.8 ± 0.3 mm (system B), with average three-dimensional displacements of 2.6 ± 0.2 mm (system A) and 1.3 ± 0.2 mm (system B). The computed CTV-PTV margins (lateral, longitudinal, vertical) were 2.5 mm, 2.5 mm, 3.6 mm and 1.4 mm, 1.6 mm, 2.4 mm for systems A and B, respectively. This suggests that the immobilization system influences intrafraction prostate displacement and, therefore, the margins applied. However, the margins found for both systems are comparable to the margins used for hypofractionated prostate SBRT.

Keywords: prostate, immobilization, hypofractionation, sbrt, intrafraction motion, intrafraction

Introduction

Prostate carcinoma is a prevalent malignant disease in men [1]. In general, prostate cancer can be treated by external beam radiotherapy while sparing organs-at-risk (OARs), such as the rectum, small bowel, bladder, bilateral femoral heads, and penile bulb, from high-dose radiation [2-3].

Studies have emphasized the importance of immobilization to reduce set-up uncertainties [4]. By reducing set-up uncertainties, the likelihood of covering the target improves and the likelihood of irradiating normal tissues is reduced [5]. For prostate cancer treatment, improved set-up uncertainties will result in smaller field margins, higher local control rates, and reduce the occurrence of normal tissue toxicities such as diarrhea and cystitis [2,6].

Ideally, patient positioning and immobilization would place a patient in the same position during treatment as during simulation when the data used for treatment planning was collected. This requires that a patient’s position would be reproducible for every treatment fraction (no interfraction variation) and that, once positioned, the patient’s position would not change during treatment (no intrafraction variation). Immobilization systems are used to improve the accuracy of the initial patient placement and to restrict patient motion during treatment, seeking to minimize both interfraction and intrafraction variations, which are inevitable in practice [7-9]. Intensity-modulated radiation therapy (IMRT) (which involves high-dose gradients, often close to critical organs) places increased demands upon precise, reproducible positioning and rigid immobilization. Without careful training in the fabrication of custom treatment devices, significant geometric errors in radiotherapy treatment can occur. Immobilization systems and their corresponding set-up errors influence the CTV to PTV margins and, therefore, may result in undesirable treatment outcomes [7-9].

The aim of this study was to determine the magnitudes of the intrafraction prostate displacements obtained with the two immobilization systems being used in our hospital. These results would then provide guidance on the CTV-PTV margins needed during treatment planning. It is well-known that internal organ motion creates intrafraction prostate displacements [10]. Thus, intrafraction prostate displacement depends on the immobilization system and internal organ motion. The secondary aim of this study was to determine if the intrafraction prostate displacement differed between the two immobilization systems since both are commercially available and in wide clinical use.

Materials and methods

Six consecutive patients (5 fractions per patient) with localized prostate cancer and implanted fiducial marker seeds in the prostate were selected. Patients were treated on a Novalis® (BrainLab AG, Munich, Germany) linear accelerator and were positioned using the ExacTrac6D Robotics (ETR) (BrainLab AG, Munich, Germany) system by seed matching. Three patients were immobilized with the CombiFixTM (Civco Medical Solutions, Orange City, IA) (system A) and three patients were immobilized with the BodyFIX® (Medical Intelligence GmbH, Schwabmünchen, Germany) (system B). Planar X-ray images were acquired pre-treatment once the patient was in position for treatment and post-treatment following treatment delivery. The time interval between the image acquisitions was determined by reviewing the timestamps in the record and verify system. Digitally reconstructed radiographs (DRRs) reconstructed from the computed tomography (CT) simulation and ExacTrac X-ray images were aligned either manually or automatically by matching the implanted fiducial marker seeds. The residual differences between planning DRRs and X-ray images in the vertical, longitudinal, and lateral directions were recorded for pre-treatment and post-treatment images. Differences between the pre-treatment and post-treatment positions indicate the intrafraction displacement of the prostate.

Figure 1 shows systems A and B, which have been used for prostate immobilization in our hospital. System A is a baseplate system providing enhanced positioning for the pelvic region and lower extremities. The system combines two cushions with optional elevation blocks into a fixed and indexable position. System B is a custom bag molded to the patient and a plastic covering that secures the patient to the bag using adjustable vacuum pressure. This enables accurate, precise patient positioning and immobilization.

Figure 1. Immobilization systems.

Figure 1

a) Immobilization system A, and b) Immobilization system B

Prostate cancer treatment delivery with the Novalis® linear accelerator takes several minutes, and during this time, the prostate may move even though the patient is in an immobilization device. For a treatment fraction, the intrafraction displacement of the prostate is equal to the difference of the post-treatment values and the pre-treatment values. The displacements were found in the lateral, longitudinal, and vertical directions. In addition, the total magnitude of the displacement was computed from these values. The mean displacements for a particular patient over the course of five fractions are found by averaging the displacements from all fractions. If there is not a systematic error, the mean displacement is expected to be zero but may be positive or negative. The average absolute displacements for a particular patient are found by averaging the absolute displacements from all fractions, which will always be positive, and similarly for the average total displacement.

The van Herk formula (Inline graphic) was used to calculate the CTV-PTV margins (lateral, longitudinal, vertical) based on intrafraction displacements, where S is the standard deviation of the systematic error and s is the standard deviation of the random error [11]. Inline graphic was estimated by the standard deviation of all the patient means, Inline graphic is the standard deviation of random error, which equals the root mean square (RMS) of the standard deviation of each patient’s mean displacements.

Results

The average absolute displacements for patients who were immobilized with system A were 0.9 ± 0.2 mm, 1.7 ± 0.4 mm, and 1.3 ± 0.3 mm, and for patients who were immobilized with system B, they were 0.5 ± 0.2 mm, 0.6 ± 0.2 mm, and 0.8 ± 0.2 mm in the lateral, longitudinal, and vertical directions, respectively. Averaged over all patients and treatment fractions, the total three-dimensional intrafraction displacements were 2.8 ± 0.4 mm for system A and 1.3 ± 0.3 mm for system B. Table 1 shows the average absolute displacements in all directions for five fractions. The average of the total patient displacement distance from the treatment position for six patients in five fractions each is also shown in Table 1. The average time interval between imaging pre- and post-treatment varied from 10.0 to 12.8 minutes (Table 1). 

Table 1. Intrafraction prostate displacements.

Average absolute displacement in all directions and average total patient displacement distance from the treatment position for six patients over five fractions, including immobilization system used and image interval (treatment) time mean and standard deviation (SD)

  System Image Interval = Treatment Time (min) Average Absolute Displacement (mm) Average Total Displacement Distance (mm)
    Mean ± SD Lateral Longitudinal Vertical  
Patient #1 A 12.2 ± 0.9 0.4 ± 0.4 1.7 ± 0.8 1.6 ± 0.7 2.4 ± 1.0
Patient #2 A 10.0 ± 0.6 0.7 ± 0.3 1.6 ± 0.9 1.3 ± 0.8 2.4 ± 0.9
Patient #3 A 12.8 ± 0.4 1.5 ± 1.0 1.9 ± 0.9 1.0 ± 0.8 2.9 ± 0.7
Patient #4 B 11.6 ± 0.5 0.7 ± 0.5 0.6 ± 0.3 0.5 ± 0.1 1.1 ± 0.4
Patient #5 B 10.6 ± 0.8 0.3 ± 0.2 0.7 ± 0.6 1.0 ± 0.6 1.3 ± 0.7
Patient #6 B 12.0 ± 0.0 0.6 ± 0.4 0.5 ± 0.4 1.1 ± 0.6 1.5 ± 0.4

The averages of all the individual means that were used to calculate the systematic error are provided in Table 2. In addition, Table 2 demonstrates a higher variability of intrafraction displacements for system A, and, therefore, a lower CTV-PTV margin for system B.

Table 2. Intrafraction displacements of two immobilization systems.

Mean standard deviation (SD Σ), root mean square (RMS σ) deviations, and margin (CTV-PTV margin Δ) for intrafraction displacements of two immobilization systems

  System A   System B
  Lateral Longitudinal Vertical   Lateral Longitudinal Vertical
Mean (mm) -0.2 -1.3 -0.8   -0.1 0.3 -0.3
SD (Inline graphic) (mm) 0.5 0.3 1.0   0.3 0.4 0.5
RMS (Inline graphic) (mm) 1.7 2.4 1.8   1.0 1.0 1.6
Margin (Inline graphic) (mm) 2.5 2.5 3.6   1.4 1.6 2.4

Discussion

The intrafraction prostate motion of two immobilization systems for prostate patients was investigated. A major limitation of this report is that only six patients were accrued. Therefore, the power of statistical inference from the results is limited. However, the results are informative, as they provide insight into the performance one might expect to achieve with these systems. In order to obtain statistically significant conclusions, further investigation with more patients is necessary. Despite this limitation of having few patients, it is encouraging that the intrafraction displacements and calculated margins found here are similar, and the standard deviations of 0.3-0.5 mm for prostate displacements are smaller than those found in a large study of SBRT patients [10] using the same van Herk formulation [11]. Furthermore, the observed mean displacement in the present study is the largest in the longitudinal and vertical directions, which is also consistent with previous results [10]. These results are consistent with internal anatomical variations resulting from rectal gas, bladder filling, or coughing, creating prostate displacements that are predominantly in the longitudinal and vertical directions.

Our results indicate that there is less intrafraction displacement for patients immobilized with system B than system A. In fact, for all patients immobilized with system B, the average total displacement distance observed was less than for any patient immobilized with system A. This suggests that the observed intrafraction displacement is not entirely a result of internal organ motion. There might be an observable and significant component contributed by the immobilization system.

Although not rigorously evaluated, there were not any noticeable dosimetric differences in the treatment plans between patients immobilized with the two systems. The treatment couch is modeled in the treatment planning system and the immobilization system is included in the external contour and accounted for in the calculations.

It is important to consider a few possible confounding factors, which prohibit drawing stronger conclusions from the results. Patients were immobilized using either system A or system B, so the different systems were not compared on the same patient. The immobilization assignment was made consecutively; patients were not randomized. Each patient was observed for only five fractions. Finally, the total sample size was only three patients with each system.

The intrafraction displacements observed for patients immobilized with system B are comparable to those found during real-time tracking of the prostate [12], suggesting that for these patients, internal organ motion might be the primary cause of the displacements. If so, the performance of this immobilization system is optimal.

Several studies have investigated the necessity of immobilization devices for patient position reproducibility [13-17]. By considering average isocenter shifts, Fiorino et al. showed that an immobilization system that fixes the legs is more reproducible than an immobilization system that fixes the pelvic region [15]. In addition, patient positioning depends on the flexibility and rigidity of the immobilization system materials [18]. System A positions both the pelvis and lower legs. However, the patient is not rigidly fixed in place. System B exhibited negligible changes in shape and, in addition to positioning the pelvis and the lower legs, fixes the patient in place with pressure from a vacuum system. The observed intrafraction displacements for patients in system A show values larger than observations of intrafraction organ motion alone, while patients in system B show displacements comparable to intrafraction organ motion. This is consistent with previous immobilization studies given the physical properties of these immobilization systems [13-18].

SBRT prostate clinical trial protocols specify the CTV to PTV margins to be used in treatment planning [19-21]. While there is no consensus for standardized margin values for prostate SBRT across all trials, clinical sites participating in a given trial should have positioning and immobilization techniques that meet the specified CTV-PTV margins in the study protocol. A phase II study of prostate SBRT (35 Gy in five fractions) using volumetric modulated arc therapy and flattening free 10 MV photons, which specified CTV to PTV margins of 3-5 mm in each direction, found that the treatments were feasible and tolerated in an acute setting [19]. The PROMETHEUS study protocol, exploring the safety, efficacy, and feasibility of a two fraction SBRT boost, specifies a CTV to PTV margin of 5 mm everywhere except posterior where it is 3 mm [20]. The Novel Integration of New prostate radiation schedules with adJuvant Androgen deprivation (NINJA) clinical trial is a randomized study comparing two emerging SBRT regimens (a five fraction and a two fraction) for efficacy specifies a 3 mm uniform CTV to PTV expansion for SBRT treatments [21]. The margins found for the two systems analyzed in our study are comparable to those required in these SBRT protocols. Depending on the protocol, one or both would be acceptable for use.

Conclusions

Patients immobilized with system A require a larger CTV-PTV margin than those with system B. Since both systems position the hips and lower legs, this suggests that rigid fixation of the patient is an important feature of an immobilization system to reduce intrafraction prostate displacement. However, the margins found for both systems are comparable to the margins used for hypofractionated prostate SBRT protocols. The margins found for system B are comparable to previously reported intrafraction internal prostate organ displacements in the literature. A real-time tracking system would be needed to reduce the margins further.

The content published in Cureus is the result of clinical experience and/or research by independent individuals or organizations. Cureus is not responsible for the scientific accuracy or reliability of data or conclusions published herein. All content published within Cureus is intended only for educational, research and reference purposes. Additionally, articles published within Cureus should not be deemed a suitable substitute for the advice of a qualified health care professional. Do not disregard or avoid professional medical advice due to content published within Cureus.

The authors have declared that no competing interests exist.

Human Ethics

Consent was obtained by all participants in this study

Animal Ethics

Animal subjects: All authors have confirmed that this study did not involve animal subjects or tissue.

References

  • 1.Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. Wolff D, Stieler F, Welzel G, et al. Radiother Oncol. 2009;93:226–233. doi: 10.1016/j.radonc.2009.08.011. [DOI] [PubMed] [Google Scholar]
  • 2.Hoskin P. Oxford: Oxford University Press. Oxford: Oxford University Press; 2012. External Beam Therapy. [Google Scholar]
  • 3.Health outcomes after prostatectomy or radiotherapy for prostate cancer: results from the prostate cancer outcomes study. Potosky AL, Legler J, Albertsen RC, et al. J Natl Cancer Inst. 2000;92:1582–1592. doi: 10.1093/jnci/92.19.1582. [DOI] [PubMed] [Google Scholar]
  • 4.Set-up error in supine-positioned patients immobilized with two different modalities during conformal radiotherapy of prostate cancer. Fiorino C, Reni M, Bolognesi A, Bonini A, Cattaneo GM, Calandrino R. Radiother Oncol. 1998;49:133–141. doi: 10.1016/s0167-8140(98)00127-3. [DOI] [PubMed] [Google Scholar]
  • 5.Assessment of three-dimensional set-up errors in conventional head and neck radiotherapy using electronic portal imaging device. Gupta T, Chopra S, Kadam A, Agarwal JP, Devi PR, Ghosh-Laskar S, Dinshaw KA. Radiat Oncol. 2007;2:1–8. doi: 10.1186/1748-717X-2-44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.The effect of breathing and set-up errors on the cumulative dose to a lung tumor. Engelsman M, Damen MF, Jaeger KD, van Ingen KM, Mijnheer BJ. Radiother Oncol. 2001;60:95–105. doi: 10.1016/s0167-8140(01)00349-8. [DOI] [PubMed] [Google Scholar]
  • 7.McDermott PN, Orton CG. Medical Physics Publishing, Madison, Wisconsin. Vol. 14. Madison, Wisconsin: Medical Physics Publishing; 2010. The Physics & Technology of Radiation Therapy. [Google Scholar]
  • 8.Washington CM, Leaver D. St. Louis, Missouri: Mosby; 2004. Principles and Practice of Radiation Therapy. [Google Scholar]
  • 9.Khan FM, Potish RA. Treatment Planning in Radiation Oncology. Philadelphia, Pennsylvania: Lippincott Williams & Wilkins; 1998. Treatment Planning in Radiation Oncology. [Google Scholar]
  • 10.Errors and margins in radiotherapy. van Herk M. Semin Radiat Oncol. 2004;14:52–64. doi: 10.1053/j.semradonc.2003.10.003. [DOI] [PubMed] [Google Scholar]
  • 11.Clinical assessment of prostate displacement and planning target volume margins for stereotactic body radiotherapy of prostate cancer. Levin-Epstein R, Qiao-Guan G, Juarez JE, et al. Front Oncol. 2020;10:539. doi: 10.3389/fonc.2020.00539. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 12.Target localization and real-time tracking using the Calypso 4D localization system in patients with localized prostate cancer. Willoughby TR, Kupelian PA, Pouliot J, et al. Int J Radiat Oncol Biol Phys. 2006;65:528–534. doi: 10.1016/j.ijrobp.2006.01.050. [DOI] [PubMed] [Google Scholar]
  • 13.The effectiveness of immobilization during prostate irradiation. Bentel GC, Marks LB, Sherouse GW, Spencer DP, Anscher MS. Int J Radiat Oncol. 1995;31:143–148. doi: 10.1016/0360-3016(94)00351-K. [DOI] [PubMed] [Google Scholar]
  • 14.Immobilization improves the reproducibility of patient positioning during six-field conformal radiation therapy for prostate carcinoma. Rosenthal SA, Roach M, Goldsmith GW, et al. Int J Radiat Oncol. 1993;27:921–926. doi: 10.1016/0360-3016(93)90469-c. [DOI] [PubMed] [Google Scholar]
  • 15.Set-up error in supine-positioned patients immobilized with two different modalities during conformal radiotherapy of prostate cancer. Fiorino C, Reni M, Bolognesi A, Bonini A, Cattaneo GM, Calandrino R. Radiother Oncol. 1998;49:133–141. doi: 10.1016/s0167-8140(98)00127-3. [DOI] [PubMed] [Google Scholar]
  • 16.A randomised study of the use of a customised immobilisation system in the treatment of prostate cancer with conformal radiotherapy. Nutting CM, Khoo VS, Walker V, McNair H, Beardmore C, Norman A, Dearnaley DP. Radiother Oncol. 2000;54:1–9. doi: 10.1016/s0167-8140(99)00181-4. [DOI] [PubMed] [Google Scholar]
  • 17.A comparison of four patient immobilization devices in the treatment of prostate cancer patients with three-dimensional conformal radiotherapy. Song PY, Washinton M, Vaida F, et al. Int J Radiat Oncol. 1995;34:213–219. doi: 10.1016/0360-3016(95)02094-2. [DOI] [PubMed] [Google Scholar]
  • 18.Evaluation of daily online set-up errors and organ displacement uncertainty during conformal radiation treatment of the prostate. Osei EK, Jiang R, Barnett R, Fleming K, Panjwani D. Brit J Radiol. 2009;82:49–61. doi: 10.1259/bjr/58088207. [DOI] [PubMed] [Google Scholar]
  • 19.Linac based SBRT for prostate cancer in 5 fractions with VMAT and flattening filter free beams: preliminary report of a phase II study. Alongi F, Cozzi L, Arcangeli S, et al. Radiat Oncol. 2013;8:171. doi: 10.1186/1748-717X-8-171. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 20.Prostate multicentre external beam radiotherapy using a stereotactic boost: the PROMETHEUS study protocol. Richardson M, Sidhom M, Gallagher S, et al. BMC Cancer. 2018;18:588. doi: 10.1186/s12885-018-4511-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.TROG 18.01 phase III randomised clinical trial of the novel integration of new prostate radiation schedules with adjuvant androgen deprivation: NINJA study protocol. Martin J, Keall P, Siva S, et al. BMJ Open. 2019;9:0. doi: 10.1136/bmjopen-2019-030731. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Cureus are provided here courtesy of Cureus Inc.

RESOURCES