Cuffless Blood Pressure Monitoring

Promises and Challenges

Jay A Pandit 1,, Enrique Lores 2, Daniel Batlle 2,
PMCID: PMC7536750  PMID: 32680913

Abstract

Current BP measurements are on the basis of traditional BP cuff approaches. Ambulatory BP monitoring, at 15- to 30-minute intervals usually over 24 hours, provides sufficiently continuous readings that are superior to the office-based snapshot, but this system is not suitable for frequent repeated use. A true continuous BP measurement that could collect BP passively and frequently would require a cuffless method that could be worn by the patient, with the data stored electronically much the same way that heart rate and heart rhythm are already done routinely. Ideally, BP should be measured continuously and frequently during diverse activities during both daytime and nighttime in the same subject by means of novel devices. There is increasing excitement for newer methods to measure BP on the basis of sensors and algorithm development. As new devices are refined and their accuracy is improved, it will be possible to better assess masked hypertension, nocturnal hypertension, and the severity and variability of BP. In this review, we discuss the progression in the field, particularly in the last 5 years, ending with sensor-based approaches that incorporate machine learning algorithms to personalized medicine.

Keywords: Blood Pressure Monitoring, Ambulatory, blood pressure, Masked Hypertension, Heart Rate, Blood Pressure Determination, Machine Learning, Blood Pressure Monitors

Introduction

Hypertension is the primary modifiable risk factor for coronary artery disease, cerebrovascular disease, and CKD (1,2). The US Preventative Services Taskforce (USPSTF) has deemed the office-based cuff BP measurements as insufficient, requiring additional measurements outside of the clinical setting before diagnosing and treating elevated BP (3). Despite this recommendation, the diagnosis and management of hypertension continue to be largely on the basis of office BP cuff measurements. The USPSTF statement, along with recent debates about BP targets and thresholds for hypertension diagnoses, have led to an increasing interest in better and alternative approaches to BP measurement (46). It is also recognized that BP management with ambulatory BP monitors, which provide a more comprehensive picture, offers better correlation with outcomes (79). Despite these developments, reducing the global burden of hypertension will only happen with truly transformative approaches for both diagnosis and management (10).

The traditional method of measuring BP has always been a cuff to compress the brachial artery and listening to Korotkoff sounds as the cuff is slowly released. This cuff-based approach has not changed much since Scipione Riva-Rocci introduced it almost 125 years ago (11). Understandably, there is increasing excitement for newer methods on the basis of cuffless methods. Mukkamala et al. (12) provided a detailed narrative review on the theories of various continuous BP monitoring approaches. This excellent review covered the different theoretical approaches up to 2015. This, however, was before the Systolic Blood Pressure Intervention Trial (5), the USPSTF statement (3) on screening for BP, and the change to the hypertension clinical guidelines (6). Since 2015, there have been high levels of interest and urgency in the search for a better way to measure BP. There is currently a race to develop a noninvasive device that can more frequently and accurately derive systolic and diastolic BP using noninvasive cuffless techniques.

With our review, we wish to examine recent attempts to accomplish the transition to cuffless BP monitoring. We conducted a search of PubMed, Medline, SCOPUS, and Web of Science with mesh words of ambulatory BP, cuffless BP, continuous BP, pulse transit time (PTT), pulse wave velocity (PWV), and pulse wave analysis. We selected, at our discretion, the articles that we felt were more relevant and compelling. This review is not a systematic review because there were no randomized clinical trials or even outcomes-based clinical studies using cuffless devices at this time in the literature. This review in itself cannot address the validation and failings of each novel approach present in the literature. Rather, we have structured this article to provide an overview of current methods and the advancement toward a passive accurate continuous and cuffless BP measurement, and we have pointed out some shortcomings that limit its current clinical use.

BP Measurement in Common Practice

Office-Based Cuff Measurement

Most outpatient clinics use a cuff to acquire either an auscultated or an automated measurement of BP. According to the recent American Heart Association (AHA) statement on BP measurement (13), the available devices are reliable only when multiple physician- and patient-related variables are done according to standard.

Even when accurate, the fundamental issue with the office-based cuff BP measurement is that it is still a snapshot (14). The recommended averaging of several readings during a visit, usually three, is an improvement, but only a small piece of the full picture of the patient’s BP. This is apparent when considering the increasingly recognized importance of white coat and masked hypertension (1518) and interest in chronotherapy (19). Masked hypertension, in particular, is now considered a very strong predictor of cardiovascular disease (17,2022). As suggested by the USPSTF, multiple measurements out of clinic are needed to support the office-based BP measurement for it to be clinically relevant (3).

Ambulatory BP Monitors

These are BP cuffs that are powered by a wearable pump that can be programmed to provide measurements as frequently as every 15 minutes for 24–72 hours. Ambulatory BP monitoring (ABPM) has been around for over 30 years (23). In more recent years, several groups (79,17,24) have argued that ABPM should replace the office measurement for diagnosis of hypertension given that it provides an average over a longer period. Studies have shown that ABPM is superior to the clinic and home BP measurement approaches in predicting clinical outcomes (6,9,25,26). Cost-effectiveness analyses demonstrate that adopting an ambulatory approach for hypertension diagnosis would reduce hypertension costs despite the cost of the devices (9). ABPM allows for measurement of both white coat and masked hypertension (27) in a noninvasive manner. However, even with the latest wearable Omron HeartGuide (28) ambulatory BP monitor, design continues to hinder adherence. It still requires arterial compression, causing sleep disruption, which is a common reason for patient nonadherence (29). Also, the devices are significantly more expensive than the BP cuffs, and this has been a barrier for widespread clinic adoption (7,29). Ultimately, although more data clearly provide a better picture than the office-based BP, reimbursement had been an issue in the United States, although the documentation for this procedure has recently been relaxed by the Centers for Medicare and Medicaid Services (CMS) (30). Importantly, we are clearly due for more innovation to come.

Home BP Measurement

This approach relies on patients using a cuff-based device of their choice and keeping a log for the physician to review. There is a validation methodology for such devices that seek to measure BP with an agreed upon level of accuracy (31), and there are guidelines for patients to follow. For instance, it has been recommended that home BP measurement should be done in a quiet room after 5 minutes of rest in the seated position, with the back and arm supported (13,32). Given that this measurement is patient driven, it is difficult for physicians to confirm that the measurements were obtained correctly. Guidelines also suggest measuring home BP four times a day (twice a day with a repeated measurement) (32), which is difficult for patients to adhere to. Literature reviews of home BP measurement suggest that there is conflicting evidence for its value compared with other approaches (32,33). Advantages of the approach are that it is low cost and easily accessible, it gets patients involved in their own care, and many of the new devices are also compatible with smartphone applications (33). Disadvantages are concerns about accuracy of existing devices, validation of available devices by a recognized body, and reliance on patient adherence to proper measurement techniques (13). Regardless, this continues to be a commonly used method of follow-up for elevated BPs measured in the clinic to identify white coat hypertension and masked hypertension.

Beyond Brachial Cuff–Based Approaches

The Omron HeartGuide and the relaxation of the CMS reimbursement for ABPM reflect a movement toward more comprehensive BP measurements over longer time periods. The objective is to record a patient’s BP through activities of daily living without the patient being aware of the measurement. Much of this drive comes from the remarkable expansion of the wearables market, which now has the ability to continuously measure almost every single vital sign except BP. Acknowledging this trend, we feel that it is particularly important to confirm or rule out the interindividual variability in BP by a continuous measurement that is reproducible. As it stands right now, often we do not know if the variability is real or operator and device dependent. Notably, even the widely used BP cuff measurements lack a true gold standard for comparison.

Theoretically, invasive arterial BP measurement could provide a platform for comparison, but this is usually not done due to practical reasons in the ambulatory context. Although this has been used for comparison by new devices in some studies, we think that it is not the ideal gold standard because it requires a hospital setting for monitoring and is subject to its own set of limitations, such as calibration error, over- or underdampening of the waveform, and reliance on pulse volume (34). Further, a properly done, AHA-approved, seated brachial cuff measurement using a validated device may not be the perfect standard, but it is the standard used so far and what has been used in all BP trials of importance. Therefore, at the moment, any effort to develop a cuffless continuous BP device is stuck with either the cuff as a gold standard (as required by the Food and Drug Administration [FDA]) or the arterial line (as expected by some in the medical community). We surmise that there has to be a more practical and representative control and that ABPM may be a preferable gold standard.

For the forgoing discussion, we have separated the cuffless continuous BP approaches into active and passive. Active approaches require the need for a noticeable obtrusive intervention, such as compression or applanation pressure, for the sensor to derive a BP measurement. Passive approaches require only sensor application, with no further intervention needed to derive a BP measurement. Although we do not have enough data to suggest that continuous measurements are better than frequent intermittent measurements, the ability to obtain and store electronically a patient’s BP measurement continuously and without the patient noticing during sleep and diverse activities has become the ultimate target.

Active Approaches

Vascular Unloading Time Technique.

In this approach, the cuff is moved from the arm to the finger to obtain an indirect measurement of beat-to-beat systolic and diastolic pressure (Figure 1). The finger cuff is used as a counterpressure to keep a constant volume of blood in the finger. Photoplethysmography (PPG) is used at the tip of the finger to obtain a pulse waveform signal that is fed into the circuit to control the finger cuff pressure to maintain a constant pulse waveform amplitude. The assumption is that, because the amplitude of the pulse waveform is largely generated by pulse volume and the blood volume in the digit is finite, then any changes required to maintain a constant volume should mostly be due to peripheral arterial resistance. Hence, the deduction is that the finger cuff pressure required to counter the pressure should correlate (with correction factors) to BP (35,36).

Figure 1.

Figure 1.

Vascular unloading technique. In this approach on the basis of photoplethysmography, a pulse volume signal is fed into a control system, which then reacts by providing a counterpressure through the finger cuff in order to maintain a constant pulse amplitude. The counterpressure required is used to derive a systolic BP and a diastolic BP. Redrawn from www.cnsystems.com.

Studies have validated this approach in the perioperative (36,37) and ambulatory setting (38). However, compressing the finger has led to concerns about digital ischemia, and therefore, this approach is not as practical for frequent measurements in the ambulatory setting.

Applanation Pulse Tonometry.

In this approach, a pressure sensor is placed directly onto the artery of interest to obtain a measurement (Figure 2). It allows for noninvasive recordings of the arterial pressure waveform and amplitudes in both central and peripheral arteries (39,40). However, as implied by this statement, applanation tonometry can be limited by different pulse waveforms for each artery and difficulty with direct applanation for a given body habitus (41). There has been success in integrating tonometry with other modalities in order to get PWV as a measure of arterial stiffness (42). Devices using this approach still require continuous calibration with a brachial BP cuff.

Figure 2.

Figure 2.

Applanation tonometry. In this approach, the sensor is placed on the radial artery and calibrated to the brachial blood pressure. Used with permission from Healthstats International.

Passive Approaches

Given the interest in identifying an unobtrusive continuous approach, the most recent work in the space has largely focused on the pulse waveform obtained by various sensors, such as optical, seismic (43), or radar (44), with the most common being PPG-acquired optical signals. PPG waveforms are easily obtainable and do not require an active intervention. Most of these approaches are variations of the PWV theory.

Pulse Wave Velocity.

PWV is defined as the velocity at which the pressure waves, generated by the systolic contraction of the heart, propagate along the arterial tree (45) (Figure 3). PWV is a known way to estimate vascular stiffness (46) and central arterial pressures (47) using the Moens–Kortweg and Hughes equations (12). The ability of PWV to predict future hypertension (48) and link to BP autoregulation (49) adds to its relevance.

Figure 3.

Figure 3.

Pulse wave velocity (PWV). In this particular image, the PWV is calculated using ΔL/Δt, where ΔL is the measured distance between the carotid artery and the femoral artery. In most cases, this is done by using a measure tape and measuring the direct length between the two sites where the pulse waveform signal is acquired, disregarding the body contour; Δt is the time difference between the foot of the carotid pulse waveform and the foot of the femoral pulse waveform. From Schillaci G, Sarchielli P, Corbelli I, Pucci G, Settimi L, Mannarino MR, Calabresi P, Mannarino E: Aortic stiffness and pulse wave reflection in young subjects with migraine. Neurology 75, 11: 960-966, 2010, https://n.neurology.org. Reprinted with permission.

Essentially, these approaches are on the basis of a variation on the velocity equation. Every patient’s BP changes, however, differently to hemodynamic maneuvers, and therefore, calibration is necessary and usually has to be done with a BP cuff (50). PWV has a strong physiologic underpinning; however, the challenge initially was to derive pulse travel distance and pulse arrival time (in order to calculate velocity) and to calibrate to each individual. The calibration challenge has led to the development of multiple algorithm improvements, and several studies have evaluated PWV-based approaches that are promising (5153). The drawbacks to the PWV approach that have been acknowledged include site specificity and better prediction of central BPs than peripheral BPs (47).

In summary, PWV is generally considered the best connector of pulse waveforms to BP. As mentioned, in order to derive PWV, a distance and a time are required. In the early stages of developing a cuffless continuous BP measurement approach, a constant was used as a surrogate for distance in almost all approaches, and most of the innovation was focused on deriving an accurate time. This gave rise to the focus on PTT, which is discussed below.

Pulse Transit Time.

PTT refers to the amount of time required for a pulse that stems from each cardiac cycle to travel from the heart to a distal point in the cardiovascular system with a known distance (12,54) using an electrocardiogram (ECG) and a PPG sensor (Figure 4). The goal with PTT is to obtain this time and correlate it to BP or use it to calculate velocity, which in turn could be used to derive BP. Although the list of references on PTT-based approaches is by no means exhaustive, there has been and continues to be an explosion of PTT-based research (5561). Since 2015 alone, there have been a myriad of conference papers not yet developed into full publications.

Figure 4.

Figure 4.

Pulse transit time (PTT). PTT is calculated here by using the time difference between the peak of the QRS (lower) from the electrocardiogram (ECG) and the peak of the photoplethysmography (PPG) waveform (upper) acquired at the radial artery distally. Modified from ref. 99, which is available under the terms of the Creative Commons Attribution License.

There are, however, consistent limitations to overcome in this approach. Inherent limitations of the PTT approach are that the method makes some indefensible assumptions or approximations regarding important parameters, such as the elasticity of the arterial wall, the pre-ejection period of the heart (time between peak of QRS on the ECG and actual systolic contraction of the left ventricle), and blood density (viscosity). Technical limitations of the approach are that in order to acquire an accurate derivation, the ECG and PPG signals need to be noise free, and the patient needs to be immobile to minimize motion artifact, suggesting that PTT can be inaccurate during activity. These challenges are known, and several groups are working to overcome them using curve-fitting techniques (6269).

Device form factors include wrist wearables, ear wearables, armchair sensors, and bathroom scale–type sensors (56,70). Many of these approaches have demonstrated mean errors that fall within standards of recognized validation organizations, suggesting that these approaches are waiting for full validation and clinical trials.

Pulse Wave Analysis and Machine Learning

Much of the recent work has now focused on improving raw signal accuracy, incorporating machine learning into the derivation algorithms and creative modeling techniques (43,44,6268,7192). It is important to note that, although the majority of the publications are using optical light-emitting diode sensors to obtain a pulse waveform, there have been other attempts, such as using seismic (43) and radar (44) sensors. Ultimately, the goal has been to obtain a pulse waveform and use this waveform to derive BP. We have provided some differences between PTT and pulse wave analysis in Table 1.

Table 1.

Notable features of pulse transit time and pulse wave analysis

Features Pulse Transit Time Pulse Wave Analysis
Sites of signal acquisition Two or more One or more
Raw signal required ECG and PPG PPG, seismocardiography, or radar
Pre-ejection period Source of inaccuracy Not a source for inaccuracy
Generalizability Needs calibration Needs calibration

ECG, electrocardiogram; PPG, photoplethysmography.

The pulse waveform was initially felt to be directly correlated to the opening and closing of the heart valves and the elasticity of the artery (46). There are, however, other factors that need consideration. The shape of the pulse waveform also depends on the pressure of blood in arteries itself and the amplitude of reflected waves from curves and takeoffs in the arterial highway, as well as the contractile force of the heart, among other things. Because pulse wave analysis takes all of these variables into account, it is likely a more comprehensive source of vascular health information than the isolated systolic and diastolic numbers derived from arterial compression.

Our group (93), along with many others (92,9496), has attempted to collect a dataset of these pulse waveforms in order to develop an algorithm using machine learning to derive BP. There is tremendous activity in curve fitting and algorithm development such that The Handbook of Cuffless Blood Pressure Monitoring: A Practical Guide for Clinicians was published in 2019 (97,98). Despite the activity, large-scale clinical trials in patients with varying systolic and diastolic BPs are missing.

Challenges

Despite the search for an accurate cuffless continuous BP monitoring approach, all current noninvasive BP measurement devices on the basis of any method will likely not give the same exact absolute BP value. This is compounded by the inherent dynamicity of BP. There are consistent challenges that have faced attempts to move away from the cuff, and we present a few of them below.

Accuracy of Control

Developing a novel cuffless and continuous approach to BP requires comparison with an accepted standard. The arterial line remains the clinical gold standard but is not appropriate for ambulatory settings. There are some FDA-approved volume clamping devices that provide beat-to-beat BP, but all have their own limitations and are still not widely accepted in the clinical community. Currently, any effort to develop a cuffless continuous BP device is stuck with the cuff as the gold standard required by the FDA and the arterial line, neither of which provide an ambulatory beat-to-beat measurement. There has to be a more practical and representative model for comparisons.

Need for Validation Standards

The FDA requires continuous BP monitors to be compared with snapshots of a BP cuff. IEEE Standard for Wearable, Cuffless Blood Pressure Measuring Devices—Amendment 1 has now been published (98), and there is discussion of a new set of International Organization of Standardization guidelines to assist companies; however, none of these have been informed by clinical trials. This is direly needed to match the speed of innovation with the speed of regulation because patients already have access to the many cuffless BP devices that did not require FDA clearance due to being classified as recreational.

Technical Issues

BP, although simplified into a systolic and diastolic reading, is a complex physiologic parameter affected by age, sex, position, internal physiology, external context, time of day, and equipment used, among other things. Sensors, just like absolute measurements from BP cuffs, are affected by the structural and functional properties of both large and small arteries, which are quite different from patient to patient. For example, an optical-based device still might not be accurate in patients with severely stiffened arteries. We know that functional stiffness (determined by BP) and structural stiffness (dependent on compliance of vessel wall) are different from patient to patient. We know this to be a fact by seeing the inaccuracy of the BP cuff–based measurements in calcified vessels, but we need more studies to test optical approaches in these patients. They could be inaccurate, or they could be a novel solution. Development of a prototype is at the intersection of multiple disciplines, including physiology, signal processing, hardware and software engineering, and data analytics, just to name a few. This team needs to work together to iterate their prototype, addressing additional questions that come up at each stage.

Calibration

Finally, and probably the most important, is that any attempt at continuous measurement of BP requires constant calibration to an existing acceptable standard. There have been myriad attempts to get around the issue of calibration, but none have been able to replace the need for intermittent calibration with a cuff. With larger datasets to learn from, even this problem will be one that can potentially be overcome using machine learning.

These are the challenges that are faced while developing a cuffless continuous BP monitoring device. After the device is developed, there will be many more challenges in terms of incorporation into current workflows, including infrastructure, privacy, and liability in dealing with continuous BP data, just to mention a few.

Despite the undisputable value of BP using a cuff, a singular snapshot of BP just cannot tell the whole BP story (14). Frequent and, ideally, continuous measurements should better capture important aspects of BP dysfunction that underpin, among others, the clinical syndromes of masked hypertension and nocturnal hypertension. Even if continuous measurements by cuffless approaches will not necessarily be more precise than the cuff-based method, if reproducible for a single individual, they provide the unique advantage of continuity in assessing BP in that individual easily and under diverse conditions. Any newer cuffless continuous approach for BP measurement should have the intrinsic “potential,” justified by clinic considerations, to be user friendly and provide many more readings over time possibly prognostically better than the existing methods (i.e., office BP, home BP, and ABPM). Clearly, there is a long way to go, and many studies will be needed for cuffless BP to become the standard of care. On the other hand, there is no reason why the field could not move faster than in the past given the need for transformative approaches to hypertension diagnosis and care.

Despite the challenges, all key stakeholders (clinicians, medical device companies and patients, to name a few) recognize the need for continuous BP monitoring and its promise. This should also facilitate involvement of patients in their medical care. Having a large database for continuous BPs in the ambulatory setting with machine learning algorithms could help identify novel biomarkers for cardiovascular disease prediction and subclinical forms of multiple diseases and, ideally, further knowledge on the natural progression of existing diseases as a start. With our review, we hope to send the message that, technically speaking, the solution is not on the horizon but may be around the corner.

Disclosures

D. Batlle is a coinventor of the patent “Active Low Molecular Weight Variants of Angiotensin Converting Enzyme 2” and founder of Angiotensin Therapeutics Inc. J. Pandit is a cofounder of Bold Diagnostics Inc. and is coinventor of the patent “Ambulatory Blood Pressure and Vital Signs Monitoring System and Method” (US20180279965A1). The remaining author has nothing to disclose.

Funding

D. Batlle is funded by National Institute of Diabetes and Digestive and Kidney Diseases grant RO1DK104785.

Footnotes

Published online ahead of print. Publication date available at www.cjasn.org.

References

  • 1.Lackland D, Weber M: Global burden of cardiovascular disease and stroke: Hypertension at the core. Can J Cardiol 31: 569–571, 2015 [DOI] [PubMed] [Google Scholar]
  • 2.Poulter N, Prabhakaran D, Caulfield M: Hypertension. Lancet 386: 801–812, 2015 [DOI] [PubMed] [Google Scholar]
  • 3.Siu A; U.S. Preventive Services Task Force : Screening for high blood pressure in adults: U.S. Preventive Services task force recommendation statement. Ann Intern Med 163: 778–786, 2015 [DOI] [PubMed] [Google Scholar]
  • 4.Gerstein H, Miller M, Byington R, Goff D Jr., Bigger J, Buse J, Cushman W, Genuth S, Ismail-Beigi F, Grimm R Jr., Probstfield J, Simons-Morton D, Friedewald W; Action to Control Cardiovascular Risk in Diabetes Study Group : Effects of intensive glucose lowering in type 2 diabetes. N Engl J Med 358: 2545–2559, 2008 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Wright J Jr., Williamson J, Whelton P, Snyder J, Sink K, Rocco M, Reboussin D, Rahman M, Oparil S, Lewis C, Kimmel P, Johnson K, Goff D Jr., Fine L, Cutler J, Cushman W, Cheung A, Ambrosius W, Ambrosius W; SPRINT Research Group : A randomized trial of intensive versus standard blood-pressure control [published correction appears in N Engl J Med 377: 2506, 2017]. N Engl J Med 373: 2103–2116, 2015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Whelton P, Carey R, Aronow W, Casey D Jr., Collins K, Dennison Himmelfarb C, DePalma S, Gidding S, Jamerson K, Jones D, MacLaughlin E, Muntner P, Ovbiagele B, Smith S Jr., Spencer C, Stafford R, Taler S, Thomas R, Williams K Sr., Williamson J, Wright J Jr.: 2017 ACC/AHA/AAPA/ABC/ACPM/AGS/APhA/ASH/ASPC/NMA/PCNA guideline for the prevention, detection, evaluation, and management of high blood pressure in adults: Executive summary: A report of the American College of Cardiology/American Heart Association task force on clinical practice guidelines [published correction appears in Hypertension 72: e33, 2018]. Hypertension 71: 1269–1324, 2018 [DOI] [PubMed] [Google Scholar]
  • 7.Head G: Ambulatory blood pressure monitoring is ready to replace clinic blood pressure in the diagnosis of hypertension: Pro side of the argument. Hypertension 64: 1175–1181, 2014 [DOI] [PubMed] [Google Scholar]
  • 8.Redon J, Lurbe E: Ambulatory blood pressure monitoring is ready to replace clinic blood pressure in the diagnosis of hypertension: Con side of the argument. Hypertension 64: 1169–1174, 2014 [DOI] [PubMed] [Google Scholar]
  • 9.Banegas J, Ruilope L, de la Sierra A, Vinyoles E, Gorostidi M, de la Cruz J, Ruiz-Hurtado G, Segura J, Rodríguez-Artalejo F, Williams B: Relationship between clinic and ambulatory blood-pressure measurements and mortality. N Engl J Med 378: 1509–1520, 2018 [DOI] [PubMed] [Google Scholar]
  • 10.Dzau V, Balatbat C: Future of hypertension. Hypertension 74: 450–457, 2019 [DOI] [PubMed] [Google Scholar]
  • 11.O’Brien E, Fitzgerald D: The history of blood pressure measurement. J Hum Hypertens 8: 73–84, 1994 [PubMed] [Google Scholar]
  • 12.Mukkamala R, Hahn J, Inan O, Mestha L, Kim C, Töreyin H, Kyal S: Toward ubiquitous blood pressure monitoring via pulse transit time: Theory and practice. IEEE Trans Biomed Eng 62: 1879–1901, 2015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Muntner P, Shimbo D, Carey R, Charleston J, Gaillard T, Misra S, Myers M, Ogedegbe G, Schwartz J, Townsend R, Urbina E, Viera A, White W, Wright J Jr.: Measurement of blood pressure in humans: A scientific statement from the American Heart Association. Hypertension 73: e35–e66, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Pandit J, Batlle D: Snapshot hemodynamics and clinical outcomes in hypertension: Precision in the measurements is key. Hypertension 67: 270–271, 2016 [DOI] [PubMed] [Google Scholar]
  • 15.Hansen T, Kikuya M, Thijs L, Björklund-Bodegård K, Kuznetsova T, Ohkubo T, Richart T, Torp-Pedersen C, Lind L, Jeppesen J, Ibsen H, Imai Y, Staessen J, Investigators I; IDACO Investigators : Prognostic superiority of daytime ambulatory over conventional blood pressure in four populations: A meta-analysis of 7,030 individuals. J Hypertens 25: 1554–1564, 2007 [DOI] [PubMed] [Google Scholar]
  • 16.Gorostidi M, Vinyoles E, Banegas J, de la Sierra A: Prevalence of white-coat and masked hypertension in national and international registries. Hypertens Res 38: 1–7, 2015 [DOI] [PubMed] [Google Scholar]
  • 17.Shimbo D, Abdalla M, Falzon L, Townsend R, Muntner P: Studies comparing ambulatory blood pressure and home blood pressure on cardiovascular disease and mortality outcomes: A systematic review. J Am Soc Hypertens 10: 224–234 e217, 2016 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Agarwal R: Albuminuria and masked uncontrolled hypertension in chronic kidney disease. Nephrol Dial Transplant 32: 2058–2065, 2017 [DOI] [PubMed] [Google Scholar]
  • 19.Hermida R, Crespo J, Dominguez-Sardina M, Otero A, Moya A, Rios M, Sineiro E, Castineira M, Callejas P, Pousa L, Salgado J, Duran C, Sanchez J, Fernandez J, Mojon A, Ayala D; Hygia Project Investigators : Bedtime hypertension treatment improves cardiovascular risk reduction: The Hygia Chronotherapy Trial [published online ahead of print October 22, 2019]. Eur Heart J [DOI] [PubMed] [Google Scholar]
  • 20.Anstey D, Booth J 3rd, Abdalla M, Spruill T, Min Y, Muntner P, Shimbo D: Predicted atherosclerotic cardiovascular disease risk and masked hypertension among blacks in the Jackson Heart Study. Circ Cardiovasc Qual Outcomes 10: e003421, 2017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21.Fagard R, Cornelissen V: Incidence of cardiovascular events in white-coat, masked and sustained hypertension versus true normotension: A meta-analysis. J Hypertens 25: 2193–2198, 2007 [DOI] [PubMed] [Google Scholar]
  • 22.Tientcheu D, Ayers C, Das S, McGuire D, de Lemos J, Khera A, Kaplan N, Victor R, Vongpatanasin W: Target organ complications and cardiovascular events associated with masked hypertension and white-coat hypertension: Analysis from the Dallas Heart Study. J Am Coll Cardiol 66: 2159–2169, 2015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Batlle D, von Riotte A, Lang G: Delayed hypotensive response to dialysis in hypertensive patients with end-stage renal disease. Am J Nephrol 6: 14–20, 1986 [DOI] [PubMed] [Google Scholar]
  • 24.Parati G, Ochoa JE, Bilo G: Moving beyond office blood pressure to achieve a personalized and more precise hypertension management: Which way to go? [published online ahead of print July 31, 2017]. Hypertension doi:10.1161/HYPERTENSIONAHA.117.08250 [DOI] [PubMed] [Google Scholar]
  • 25.Lurbe E, Redon J, Kesani A, Pascual J, Tacons J, Alvarez V, Batlle D: Increase in nocturnal blood pressure and progression to microalbuminuria in type 1 diabetes. N Engl J Med 347: 797–805, 2002 [DOI] [PubMed] [Google Scholar]
  • 26.Yang W, Melgarejo J, Thijs L, Zhang Z, Boggia J, Wei F, Hansen T, Asayama K, Ohkubo T, Jeppesen J, Dolan E, Stolarz-Skrzypek K, Malyutina S, Casiglia E, Lind L, Filipovský J, Maestre G, Li Y, Wang J, Imai Y, Kawecka-Jaszcz K, Sandoya E, Narkiewicz K, O’Brien E, Verhamme P, Staessen J; International Database on Ambulatory Blood Pressure in Relation to Cardiovascular Outcomes (IDACO) Investigators : Association of office and ambulatory blood pressure with mortality and cardiovascular outcomes. JAMA 322: 409–420, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Benjamin E, Muntner P, Alonso A, Bittencourt M, Callaway C, Carson A, Chamberlain A, Chang A, Cheng S, Das S, Delling F, Djousse L, Elkind M, Ferguson J, Fornage M, Jordan L, Khan S, Kissela B, Knutson K, Kwan T, Lackland D, Lewis T, Lichtman J, Longenecker C, Loop M, Lutsey P, Martin S, Matsushita K, Moran A, Mussolino M, O’Flaherty M, Pandey A, Perak A, Rosamond W, Roth G, Sampson U, Satou G, Schroeder E, Shah S, Spartano N, Stokes A, Tirschwell D, Tsao C, Turakhia M, VanWagner L, Wilkins J, Wong S, Virani S; American Heart Association Council on Epidemiology and Prevention Statistics Committee and Stroke Statistics Subcommittee : Heart disease and stroke statistics-2019 update: A report from the American Heart Association [published correction appears in Circulation 141: e33, 2020]. Circulation 139: e56–e528, 2019 [DOI] [PubMed] [Google Scholar]
  • 28.Kuwabara M, Harada K, Hishiki Y, Kario K: Validation of two watch-type wearable blood pressure monitors according to the ANSI/AAMI/ISO81060-2:2013 guidelines: Omron HEM-6410T-ZM and HEM-6410T-ZL. J Clin Hypertens (Greenwich) 21: 853–858, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.van der Steen M, Lenders J, Thien T: Side effects of ambulatory blood pressure monitoring. Blood Press Monit 10: 151–155, 2005 [DOI] [PubMed] [Google Scholar]
  • 30. Jensen TS, Chin J, Ashby L, Hakim R, Hutter J, Caplan S, McKesson R; CMSgov Medicare & Medicaid Service: Decision memo for ambulatory blood pressure monitoring (ABPM) (CAG-00067R2), Baltimore, MD, US Centers for Medicare & Medicaid Services, 2019.
  • 31.Stergiou G, Alpert B, Mieke S, Asmar R, Atkins N, Eckert S, Frick G, Friedman B, Graßl T, Ichikawa T, Ioannidis J, Lacy P, McManus R, Murray A, Myers M, Palatini P, Parati G, Quinn D, Sarkis J, Shennan A, Usuda T, Wang J, Wu C, O’Brien E: A universal standard for the validation of blood pressure measuring devices: Association for the Advancement of Medical Instrumentation/European Society of Hypertension/International Organization for Standardization (AAMI/ESH/ISO) collaboration statement. Hypertension 71: 368–374, 2018 [DOI] [PubMed] [Google Scholar]
  • 32.Shimbo D, Abdalla M, Falzon L, Townsend R, Muntner P: Role of ambulatory and home blood pressure monitoring in clinical practice: A narrative review. Ann Intern Med 163: 691–700, 2015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 33.Sheikh S, Sinha A, Agarwal R: Home blood pressure monitoring: How good a predictor of long-term risk? Curr Hypertens Rep 13: 192–199, 2011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 34.Shinozaki T, Deane R, Mazuzan J: The dynamic responses of liquid-filled catheter systems for direct measurements of blood pressure. Anesthesiology 53: 498–504, 1980 [DOI] [PubMed] [Google Scholar]
  • 35.Yamakoshi K, Shimazu H, Togawa T: Indirect measurement of instantaneous arterial blood pressure in the human finger by the vascular unloading technique. IEEE Trans Biomed Eng 27: 150–155, 1980 [DOI] [PubMed] [Google Scholar]
  • 36.Ilies C, Bauer M, Berg P, Rosenberg J, Hedderich J, Bein B, Hinz J, Hanss R: Investigation of the agreement of a continuous non-invasive arterial pressure device in comparison with invasive radial artery measurement. Br J Anaesth 108: 202–210, 2012 [DOI] [PubMed] [Google Scholar]
  • 37.Biais M, Vidil L, Roullet S, Masson F, Quinart A, Revel P, Sztark F: Continuous non-invasive arterial pressure measurement: Evaluation of CNAP device during vascular surgery. Ann Fr Anesth Reanim 29: 530–535, 2010 [DOI] [PubMed] [Google Scholar]
  • 38.Imholz B, Langewouters G, van Montfrans G, Parati G, van Goudoever J, Wesseling K, Wieling W, Mancia G: Feasibility of ambulatory, continuous 24-hour finger arterial pressure recording. Hypertension 21: 65–73, 1993 [DOI] [PubMed] [Google Scholar]
  • 39.Van Bortel L, Balkestein E, van der Heijden-Spek J, Vanmolkot F, Staessen J, Kragten J, Vredeveld J, Safar M, Struijker Boudier H, Hoeks A: Non-invasive assessment of local arterial pulse pressure: Comparison of applanation tonometry and echo-tracking. J Hypertens 19: 1037–1044, 2001 [DOI] [PubMed] [Google Scholar]
  • 40.Benetos A, Laurent S, Hoeks A, Boutouyrie P, Safar M: Arterial alterations with aging and high blood pressure. A noninvasive study of carotid and femoral arteries. Arterioscler Thromb 13: 90–97, 1993 [DOI] [PubMed] [Google Scholar]
  • 41.Vlachopoulos C, O’Rourke M, Nichols W: McDonald’s Blood Flow in Arteries: Theoretical, Experimental and Clinical Principles, Boca Raton, FL, CRC Press, 2011 [Google Scholar]
  • 42.Nelson M, Stepanek J, Cevette M, Covalciuc M, Hurst R, Tajik A: Noninvasive measurement of central vascular pressures with arterial tonometry: Clinical revival of the pulse pressure waveform? Mayo Clin Proc 85: 460–472, 2010 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 43.Yousefian P, Shin S, Mousavi A, Kim C, Mukkamala R, Jang D, Ko B, Lee J, Kwon U, Kim Y, Hahn J: The potential of wearable limb ballistocardiogram in blood pressure monitoring via pulse transit time. Sci Rep 9: 10666, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Johnson J, Shay O, Kim C, Liao C: Wearable millimeter-wave device for contactless measurement of arterial pulses. IEEE Trans Biomed Circuits Syst 13: 1525–1534, 2019 [DOI] [PubMed] [Google Scholar]
  • 45.Pereira T, Correia C, Cardoso J: Novel methods for pulse wave velocity measurement. J Med Biol Eng 35: 555–565, 2015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Davies J, Struthers A: Pulse wave analysis and pulse wave velocity: A critical review of their strengths and weaknesses. J Hypertens 21: 463–472, 2003 [DOI] [PubMed] [Google Scholar]
  • 47.Ma Y, Choi J, Hourlier-Fargette A, Xue Y, Chung H, Lee J, Wang X, Xie Z, Kang D, Wang H, Han S, Kang S, Kang Y, Yu X, Slepian M, Raj M, Model J, Feng X, Ghaffari R, Rogers J, Huang Y: Relation between blood pressure and pulse wave velocity for human arteries. Proc Natl Acad Sci U S A 115: 11144–11149, 2018 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 48.Koivistoinen T, Lyytikäinen L, Aatola H, Luukkaala T, Juonala M, Viikari J, Lehtimäki T, Raitakari O, Kähönen M, Hutri-Kähönen N: Pulse wave velocity predicts the progression of blood pressure and development of hypertension in young adults. Hypertension 71: 451–456, 2018 [DOI] [PubMed] [Google Scholar]
  • 49.Tripathi A, Obata Y, Ruzankin P, Askaryar N, Berkowitz D, Steppan J, Barodka V: A pulse wave velocity based method to assess the mean arterial blood pressure limits of autoregulation in peripheral arteries. Front Physiol 8: 855, 2017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 50.Wang Y, Chen C, Sue C, Lu W, Chiou Y: Estimation of blood pressure in the radial artery using strain-based pulse wave and photoplethysmography sensors. Micromachines (Basel) 9: 556, 2018 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 51.Nabeel P, Jayaraj J, Mohanasankar S: Single-source PPG-based local pulse wave velocity measurement: A potential cuffless blood pressure estimation technique. Physiol Meas 38: 2122–2140, 2017 [DOI] [PubMed] [Google Scholar]
  • 52.Liu Z, Liu J, Wen B, He Q, Li Y, Miao F: Cuffless blood pressure estimation using pressure pulse wave signals. Sensors (Basel) 18: 4227, 2018 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 53.Fung P, Dumont G, Ries C, Mott C, Ansermino M: Continuous noninvasive blood pressure measurement by pulse transit time. Conf Proc IEEE Eng Med Biol Soc 2006: 738–741, 2004 [DOI] [PubMed] [Google Scholar]
  • 54.Sola J, Bertschi M, Krauss J: Measuring pressure: Introducing oBPM, the optical revolution for blood pressure monitoring. IEEE Pulse 9: 31–33, 2018 [DOI] [PubMed] [Google Scholar]
  • 55.Kachuee M, Kiani M, Mohammadzade H, Shabany M: Cuffless blood pressure estimation algorithms for continuous health-care monitoring. IEEE Trans Biomed Eng 64: 859–869, 2017 [DOI] [PubMed] [Google Scholar]
  • 56.Martin S, Carek A, Kim C, Ashouri H, Inan O, Hahn J, Mukkamala R: Weighing scale-based pulse transit time is a superior marker of blood pressure than conventional pulse arrival time. Sci Rep 6: 39273, 2016 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 57.Kim C, Carek A, Mukkamala R, Inan O, Hahn J: Ballistocardiogram as proximal timing reference for pulse transit time measurement: Potential for cuffless blood pressure monitoring. IEEE Trans Biomed Eng 62: 2657–2664, 2015 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 58.Gao M, Olivier N, Mukkamala R: Comparison of noninvasive pulse transit time estimates as markers of blood pressure using invasive pulse transit time measurements as a reference. Physiol Rep 4: e12768, 2016 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 59.Lin W, Wang H, Samuel O, Liu G, Huang Z, Li G: New photoplethysmogram indicators for improving cuffless and continuous blood pressure estimation accuracy. Physiol Meas 39: 25005, 2018 [DOI] [PubMed] [Google Scholar]
  • 60.Huynh T, Jafari R, Chung W: Noninvasive cuffless blood pressure estimation using pulse transit time and impedance plethysmography. IEEE Trans Biomed Eng 66: 967–976, 2019 [DOI] [PubMed] [Google Scholar]
  • 61.Sharifi I, Goudarzi S, Khodabakhshi M: A novel dynamical approach in continuous cuffless blood pressure estimation based on ECG and PPG signals. Artif Intell Med 97: 143–151, 2019 [DOI] [PubMed] [Google Scholar]
  • 62.Xing X, Sun M: Optical blood pressure estimation with photoplethysmography and FFT-based neural networks. Biomed Opt Express 7: 3007–3020, 2016 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 63.Ding X, Zhang Y, Tsang H: Impact of heart disease and calibration interval on accuracy of pulse transit time-based blood pressure estimation. Physiol Meas 37: 227–237, 2016 [DOI] [PubMed] [Google Scholar]
  • 64.Tang Z, Tamura T, Sekine M, Huang M, Chen W, Yoshida M, Sakatani K, Kobayashi H, Kanaya S: A chair-based unobtrusive cuffless blood pressure monitoring system based on pulse arrival time. IEEE J Biomed Health Inform 21: 1194–1205, 2017 [DOI] [PubMed] [Google Scholar]
  • 65.Zhang Q, Chen X, Zhen F, Xue Y, Xia S: Cuff-less blood pressure measurement using pulse arrival time and a Kalman filter. J Micromech Microeng 27: 24002, 2017 [Google Scholar]
  • 66.Esmaili A, Kachuee M, Shabany M: Nonlinear cuffless blood pressure estimation of healthy subjects using pulse transit time and arrival time. IEEE Trans Instrum Meas 66: 3299–3308, 2017 [Google Scholar]
  • 67.Ding X, Yan B, Zhang Y-T, Liu J, Zhao N, Tsang H: Pulse transit time based continuous cuffless blood pressure estimation: A new extension and a comprehensive evaluation. Sci Rep 7: 11554, 2017 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 68.Esmaili A, Kachuee M, Shabany M: Nonlinear cuffless blood pressure estimation of healthy subjects using pulse transit time and arrival time. IEEE Trans Instrum Meas 66: 3299–3308, 2017 [Google Scholar]
  • 69.Huttunen J, Kärkkäinen L, Lindholm H: Pulse transit time estimation of aortic pulse wave velocity and blood pressure using machine learning and simulated training data. PLOS Comput Biol 15: e1007259, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 70.Zhang Q, Zeng X, Hu W, Zhou D: A machine learning-empowered system for long-term motion-tolerant wearable monitoring of blood pressure and heart rate with Ear-ECG/PPG. IEEE Access 5: 10547–10561, 2017 [Google Scholar]
  • 71. Shimazaki S, Kawanaka H, Ishikawa H, Inoue K, Oguri K: Cuffless blood pressure estimation from only the waveform of photoplethysmography using CNN. Presented at the 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, July 23–27, 2019. [DOI] [PubMed] [Google Scholar]
  • 72.Huang K-H, Tan F, Wang T-D, Yang Y-J: A highly sensitive pressure-sensing array for blood pressure estimation assisted by machine-learning techniques. Sensors (Basel) 19: 848, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 73.Zhang B, Ren H, Huang G, Cheng Y, Hu C: Predicting blood pressure from physiological index data using the SVR algorithm. BMC Bioinformatics 20: 109, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 74.Chen S, Ji Z, Wu H, Xu Y: A non-invasive continuous blood pressure estimation approach based on machine learning. Sensors (Basel) 19: 2585, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 75.Ribas Ripoll V, Vellido A: Blood pressure assessment with differential pulse transit time and deep learning: A proof of concept. Kidney Dis 5: 23–27, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 76.Rastegar S, Gholamhosseini H, Lowe A, Mehdipour F, Lindén M: Estimating systolic blood pressure using convolutional neural networks. Stud Health Technol Inform 261: 143–149, 2019 [PubMed] [Google Scholar]
  • 77. Yan C, Li Z, Zhao W, Hu J, Jia D, Wang H, You T: Novel deep convolutional neural network for cuff-less blood pressure measurement using ECG and PPG signals. Presented at the 41st Annual International Conference of the IEEE Engineering in Medicine and Biology Society (EMBC), Berlin, July 23–27, 2019. [DOI] [PubMed] [Google Scholar]
  • 78.Maher N, Elsheikh G, Anis W, Emara T: Non-invasive calibration-free blood pressure estimation based on artificial neural network In: Hassanien A, Azar A, Gaber T, Bhatnagar R, Tolba M. (eds) International Conference on Advanced Machine Learning Technologies and Applications, Berlin, Springer, 2019, pp 701–711 [Google Scholar]
  • 79.Miao F, Liu Z, Liu J, Wen B, Li Y: Multi-sensor fusion approach for cuff-less blood pressure measurement. IEEE J Biomed Health Inform 24: 79–91, 2020 [DOI] [PubMed] [Google Scholar]
  • 80.Tanveer M, Hasan M: Cuffless blood pressure estimation from electrocardiogram and photoplethysmogram using waveform based ANN-LSTM network. Biomed Signal Process Control 51: 382–392, 2019 [Google Scholar]
  • 81.Mousavi S, Firouzmand M, Charmi M, Hemmati M, Moghadam M, Ghorbani Y: Blood pressure estimation from appropriate and inappropriate PPG signals using A whole-based method. Biomed Signal Process Control 47: 196–206, 2019 [Google Scholar]
  • 82.Hasanzadeh N, Ahmadi M, Mohammadzade H: Blood pressure estimation using photoplethysmogram signal and its morphological features. IEEE Sens J 20: 4300–4310, 2019 [Google Scholar]
  • 83.Wang B, Huang Z, Wu J, Liu Z, Liu Y, Zhang P: Continuous blood pressure estimation using PPG and ECG signal In: Advances in Body Area Networks I, edited by Fortino G, Wang Z, Berlin, Springer, 2019, pp 61–74 [Google Scholar]
  • 84.Landry C, Peterson S, Arami A: Nonlinear dynamic modeling of blood pressure waveform: Towards an accurate cuffless monitoring system. IEEE Sens J 20: 5368–5378, 2020 [Google Scholar]
  • 85.Kei Fong M, Ng E, Er Zi Jian K, Hong T: SVR ensemble-based continuous blood pressure prediction using multi-channel photoplethysmogram. Comput Biol Med 113: 103392, 2019 [DOI] [PubMed] [Google Scholar]
  • 86.Hassani A, Foruzan A: Improved PPG-based estimation of the blood pressure using latent space features. Signal Image Video Process 13: 1141–1147, 2019 [Google Scholar]
  • 87.Zhang B, Ren J, Cheng Y, Wang B, Wei Z: Health data driven on continuous blood pressure prediction based on gradient boosting decision tree algorithm. IEEE Access 7: 32423–32433, 2019 [Google Scholar]
  • 88.Fujita D, Suzuki A, Ryu K: PPG-based systolic blood pressure estimation method using PLS and level-crossing feature. Appl Sci (Basel) 9: 304, 2019 [Google Scholar]
  • 89.Lazazzera R, Belhaj Y, Carrault G: A new wearable device for blood pressure estimation using photoplethysmogram. Sensors (Basel) 19: 2557, 2019 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 90.Mukkamala R: Blood pressure with a click of a camera? Circ Cardiovasc Imaging 12: e009531, 2019 [DOI] [PubMed] [Google Scholar]
  • 91.Sun B, Wang C, Chen X, Zhang Y, Shao H: PPG signal motion artifacts correction algorithm based on feature estimation. Optik (Stuttg) 176: 337–349, 2019 [Google Scholar]
  • 92.Luo H, Yang D, Barszczyk A, Vempala N, Wei J, Wu S, Zheng P, Fu G, Lee K, Feng Z-P: Smartphone-based blood pressure measurement using transdermal optical imaging technology. Circ Cardiovasc Imaging 12: e008857, 2019 [DOI] [PubMed] [Google Scholar]
  • 93.Pandit J, Miller K, Wu J, Connell S, Shah S: Abstract 12960: Differential pulse arrival time: A novel approach to continuous cuff-less blood pressure monitoring. Circulation 134: A12960, 2016 [Google Scholar]
  • 94.Jeong I, Finkelstein J: Introducing contactless blood pressure assessment using a high speed video camera. J Med Syst 40: 77, 2016 [DOI] [PubMed] [Google Scholar]
  • 95.Secerbegovic A, Bergsland J, Halvorsen P, Suljanovic N, Mujcic A, Balasingham I: Blood pressure estimation using video plethysmography. Presented at the IEEE 13th International Symposium on Biomedical Imaging (ISBI), Prague, Czech Republic, April 13–16, 2016 [Google Scholar]
  • 96.Sugita N, Yoshizawa M, Abe M, Tanaka A, Homma N, Yambe T: Contactless technique for measuring blood-pressure variability from one region in video plethysmography. J Med Biol Eng 39: 76–85, 2019 [Google Scholar]
  • 97.Solà J, Delgado-Gonzalo R: The Handbook of Cuffless Blood Pressure Monitoring: A Practical Guide for Clinicians, Heidelberg, Germany, Springer Nature, 2019 [Google Scholar]
  • 98.Wearable Cuffless Blood Pressure Monitors Working Group : IEEE Standard for Wearable, Cuffless Blood Pressure Measuring Devices—Amendment 1, New York, IEEE, 2019 [Google Scholar]
  • 99.Ma H: A blood pressure monitoring method for stroke management. Biomed Res Int 2014: 571623, 2014 [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Clinical Journal of the American Society of Nephrology : CJASN are provided here courtesy of American Society of Nephrology

RESOURCES