To the Editor,
Tremendous increase in workload due to coronavirus disease 2019 (COVID‐19) pandemic has caused intense strain on laboratory service. To enhance testing capacity, installation of high‐throughput platform with short hands‐on time becomes essential. The Panther Fusion severe acute respiratory syndrome coronavirus 2 (SARS‐CoV‐2) assay (PF assay) (Hologic, Inc.) operated in Panther Fusion system (PF system) is one of the options. It is validated with nasopharyngeal (NP), oropharyngeal and bronchoalveolar lavage (BAL) specimens while other sample types like deep throat saliva (DTS) and lower respiratory tract (LRT) specimens besides BAL are not validated. This study aims to evaluate the diagnostic performance of PF assay for detection of SARS‐CoV‐2 in comparison to the TIB‐Molbiol LightMix SarbecoV E‐gene assay (TIB‐Molbiol assay) (TIB‐Molbiol) using DTS, NP, and LRT specimens.
One hundred and fifyt‐eight specimens (87 positive and 71 negative specimens) collected from 142 patients with suspected COVID‐19 disease were tested. These included 60 NP, 59 DTS, and 39 LRT specimens (including 26 sputum, 12 tracheal aspirate, and 1 BAL). The median age of patients was 49 (interquartile range, 32–70) with 47.2% of female and proportion of in‐patients was 75.4%. One hundred and nine of them were prospective samples and the remaining ones were archived samples stored at −70°C. All positive samples were confirmed by reference laboratory as published previously. 1 , 2 Results of PF assay were compared with TIB‐Molbiol assay. Discordant results were resolved with Xpert Xpress SARS‐CoV‐2 assay (Xpert Xpress assay) (Cepheid).
NP samples were collected in 3 ml viral transport medium. Both DTS and LRT specimens were collected in sterile plain bottles. The 87 samples tested positive by TIB‐Molbiol assay spanned the full range of cycle threshold (C t) scores ranged from 13.57 to 38.58 (n = 21 with C t < 20, n = 35 with C t 20–30 and n = 31 with C t > 30).
DTS and LRT specimens were pretreated as published previously 2 while NP specimens were tested without pretreatment. TIB‐Molbiol assay was performed as described previously 1 and PF assay was performed according to manufacturer's instruction except supernatant was used for DTS or LRT specimens.
Positive percent agreement (PPA), negative percent agreement (NPA) and Cohen's kappa were determined by using consensus results as reference method. Statistical analysis was performed by MedCalc 19.4.1. Cohen's kappa value greater than 0.81 was interpreted as very good agreement.
Results of PF assay were compared to those of TIB‐Molbiol assay, there were eight samples with discrepancy and were resolved by testing with Xpert Xpress assay. Using consensus results as gold standard, PPA of NP, LRT, and DTS specimens were 96.43%, 100%, and 96.00%, respectively. The overall PPA and NPA were 97.53% and 100%, respectively and overall Cohen's kappa value was 0.97 (Table 1).
Table 1.
Panther Fusion | Consensus result | PPA (95% CI) | NPA (95% CI) | Cohen's kappa value (95% CI) | |
---|---|---|---|---|---|
SARS‐CoV‐2 assay | Detected | Not detected | |||
Nasopharyngeal | |||||
Detected | 27 | 0 | 96.43% | 100% | 0.97 |
Not detected | 1 | 32 | (81.65%–99.91%) | (89.11%–100%) | (0.90–1.00) |
Lower respiratory tract | |||||
Detected | 28 | 0 | 100% | 100% | 1.00 |
Not detected | 0 | 11 | (87.66%–100%) | (71.51%–100%) | (1.00–1.00) |
Deep throat saliva | |||||
Detected | 24 | 0 | 96.00% | 100% | 0.97 |
Not detected | 1 | 34 | (79.65%–99.90%) | (89.72%–100%) | (0.90–1.00) |
Overall sample types | |||||
Detected | 79 | 0 | 97.53% | 100% | 0.97 |
Not detected | 2 | 77 | (91.36%–99.70%) | (95.32%–100%) | (0.94–1.00) |
Abbreviations: CI, confidence interval; NPA, negative percent agreement; PPA, positive percent agreement; SARS‐CoV‐2, severe acute respiratory syndrome coronavirus 2.
Among the eight samples with discrepancy, six of them initially tested positive by TIB Miobiol assay were tested negative by both PF and Xpert Xpress assay. These samples were hence regarded as false‐positive. Despite the manufacturer of TIB Miobiol regarded results with C t < 36 as positive, when use as an initial screening test in our laboratory, sample with any C t value will be tested supplementary with Xpert Xpress assay. The high C t values of two discrepant samples (36.49 and 37.90, respectively) suggested that their viral loads were low.
Studies have verified the LoD of PF assay ranged from 62.5 to 100 copies/ml. 3 , 4 , 5 When tested with SARS‐CoV‐2 synthetic quantified standard from Exact Diagnostics (BioRad), we found that the LoD of both TIB‐Molbiol and PF assays were 100 copies/ml while that of Xpert Xpress assay was 50 copies/ml. 2
Detection of other respiratory viruses with the respiratory panels on PF System for upper and LRT samples including BAL have been evaluated in various studies. 6 , 7 , 8 For SARS‐CoV‐2, several studies have assessed the diagnostic performance of PF assay with NP specimens. 4 , 9 , 10 , 11 Performance of PF assay with DTS and LRT samples other than BAL has not been reported. In this study, we demonstrated that in addition to NP specimens, performance of PF assay for DTS and various LRT specimens was good and comparable to reference method.
Presence of mucus in the lysis tube will lead to pipetting errors and invalid results. Similar problem has been reported by Szymczak et al. 3 To resolve such problem, before adding sample into the lysis tube, we recommend to centrifuge the homogenized DTS & LRT specimens according to the pretreatment protocol reported in our previous study. 2
In conclusion, the Panther Fusion SARS‐CoV‐2 assay has comparable performance with the reference method. With its capability of random access and high‐throughput, it maximizes the flexibility of SARS‐CoV‐2 testing in clinical laboratories.
CONFLICT OF INTERESTS
The authors declare that there are no conflict of interests.
AUTHOR CONTRIBTUONS
River Chun‐Wai Wong: planned and conducted experiments, performed data analysis, and drafted the manuscript. Ann Han Wong: supervision, review, and editing the manuscript. Yolanda Iok‐Ieng Ho: methodology, review, and editing the manuscript. Eddie Chi‐Man Leung: coordination, review, and editing the manuscript. Raymond Wai‐Man Lai: project administration and supervision. All authors discussed the results and contributed to the final manuscript.
DATA AVAILABILITY STATEMENT
Data sharing not applicable to this article as no datasets were generated or analyzed during the current study.
REFERENCES
- 1. Leung EC, Chow VC, Lee MK, Lai RW. Deep throat saliva as an alternative diagnostic specimen type for the detection of SARS‐CoV‐2. J Med Virol. 2020:1‐4. 10.1002/jmv.26258 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Wong RC, Wong AH, Ho YII, Leung EC, Lai RWM. Evaluation on testing of deep throat saliva and lower respiratory tract specimens with Xpert Xpress SARS‐CoV‐2 assay. J Clin Virol. 2020;131:104593. 10.1016/j.jcv.2020.104593 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Szymczak WA, Goldstein DY, Orner EP, et al. Utility of stool PCR for the diagnosis of COVID‐19: comparison of two commercial platforms. J Clin Microbiol. 2020;58(9):e01369‐20. 10.1128/JCM.01369-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Smith E, Zhen W, Manji R, Schron D, Duong S, Berry GJ. Analytical and clinical comparison of three nucleic acid amplification tests for SARS‐CoV‐2 detection. J Clin Microbiol. 2020;58(9):e01134‐20. 10.1128/JCM.01134-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5. Fung B, Gopez A, Servellita V, et al. Direct comparison of SARS‐CoV‐2 analytical limits of detection across seven molecular assays. J Clin Microbiol. 2020;58(9):e01535‐20. 10.1128/JCM.01535-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Cordes AK, Heim A. Rapid random access detection of the novel SARS‐coronavirus‐2 (SARS‐CoV‐2, previously 2019‐nCoV) using an open access protocol for the Panther Fusion. J Clin Virol. 2020;125:104305. 10.1016/j.jcv.2020.104305 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. Ganzenmueller T, Kaiser R, Baier C, et al. Comparison of the performance of the Panther Fusion respiratory virus panel to R‐gene and laboratory developed tests for diagnostic and hygiene screening specimens from the upper and lower respiratory tract. J Med Microbiol. 2020;69(3):427‐435. 10.1099/jmm.0.001133 [DOI] [PubMed] [Google Scholar]
- 8. Sam SS, Caliendo AM, Ingersoll J, Abdul‐Ali D, Hill CE, Kraft CS. Evaluation of performance characteristics of Panther Fusion assays for detection of respiratory viruses from nasopharyngeal and lower respiratory tract specimens. J Clin Microbiol. 2018;56(8):e00787‐18. 10.1128/JCM.00787-18 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9. Hogan CA, Sahoo MK, Huang C, et al. Comparison of the Panther Fusion and a laboratory‐developed test targeting the envelope gene for detection of SARS‐CoV‐2. J Clin Virol. 2020;127:104383. 10.1016/j.jcv.2020.104383 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10. Craney AR, Velu PD, Satlin MJ, et al. Comparison of two high‐throughput reverse transcription‐PCR systems for the detection of severe acute respiratory syndrome coronavirus 2. J Clin Microbiol. 2020;58(8):e00890‐20. 10.1128/JCM.00890-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11. Stevens B, Hogan CA, Sahoo MK, et al. Comparison of a point‐of‐care assay and a high‐complexity assay for detection of SARS‐CoV‐2 RNA. J Appl Lab Med. 2020. 10.1093/jalm/jfaa135 [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
Data sharing not applicable to this article as no datasets were generated or analyzed during the current study.