Abstract
Purpose
Recently, we introduced a quantum coherence based method (ge‐HSQC) for indirect 13C‐MRS in the liver to track 13C‐labeled lipids into the hepatic lipid pool in vivo. This approach is more robust in case of respiratory motion, however, inherently leads to a signal loss of 50% when compared with a conventional J‐difference editing technique (JDE). Here, we intend to improve the robustness of a regular JDE (STEAM‐ACED) with the use of a BIlinear Rotation Decoupling (BIRD) filter to achieve 100% higher signal gain when compared with ge‐HSQC.
Methods
To determine the efficiency of the BIRD filter 1H‐[13C] lipid spectra were acquired on 3T from a peanut oil phantom, with three different MR sequences: ge‐HSQC, STEAM‐ACED, and the BIRD filter together with STEAM‐ACED (BIRD‐STEAM‐ACED). Finally, our proposed method is tested in vivo in five healthy volunteers with varying liver fat content. In these subjects we quantified the 1H‐[13C]‐signal from the hepatic lipid pool and determined 13C enrichment, which is expected to be 1.1% according to the natural abundance of 13C.
Results
The application of the proposed BIRD filter reduces the subtraction artifact of 1H‐[12C] lipid signal efficiently in JDE experiments, which leads to a signal gain of 100% of 1H‐[13C]‐lipid signals when compared with the ge‐HSQC. Phase distortions in vivo were minimal with the use of BIRD compared with STEAM‐ACED, which enabled us to robustly quantify the 13C‐enrichment in all five subjects.
Conclusion
The BIRD‐STEAM‐ACED sequence is an efficient and promising tool for 13C‐tracking experiments in the human liver in vivo.
Keywords: 13C MRS, 13C tracking, BIRD filter
1. INTRODUCTION
The application of 13C‐MRS together with the administration of specific 13C‐labeled substrate enables to measure a large number of metabolic fluxes noninvasively in vivo. 1 , 2 , 3 When compared with direct 13C‐MRS, the use of indirect 13C‐MRS (i.e., the detection of 1H signals that are coupled to a 13C‐nuclei) leads to a large signal gain and allows localization of the MR signal with standard 1H‐MRS localization techniques. 4 , 5 As the natural abundance of 13C is only 1.1%, the efficacy of the spectral editing sequences used is crucial for robust suppression of the large 1H‐[12C] signals.
Editing of the 1H‐[13C] signal (e.g., indirect 13C) is mainly performed in two different ways; either by regular J‐difference editing (JDE) methods or by making use of approaches that allow single‐shot acquisitions by applying multiple quantum‐coherence (MQC) based sequences. 5 Recently, we demonstrated that it is possible to noninvasively follow the retention of labeled dietary 13C‐lipids into the hepatic lipid pool in vivo using gradient enhanced heteronuclear single quantum coherence (ge‐HSQC). 6 However, the drawback of applying ge‐HSQC is that it inherently leads to a signal loss of 50% when compared with a conventional JDE sequence. Earlier studies have demonstrated the applicability of JDE techniques for 13C tracking applications especially to study glucose metabolism in brain. 7 , 8 Such JDE techniques are quite challenging to apply directly in liver to track 13C lipids, as respiratory motion will result in imperfect subtraction due to the (large) overlapping 1H‐[12C] lipid signal at 1.3 ppm. In a previous study, a JDE sequence was applied on skeletal muscle and liver of rats to study lipid handling; 9 however, it has not been used in humans yet due to the problem of respiratory motion leading to subtraction artifacts. Therefore, the aim of this study was to increase the robustness of a regular JDE technique to increase the edited 1H‐[13C] signals when compared with the ge‐HSQC methodology, which is required to be able to track the low incorporation rates of 13C lipids in the human hepatic fat pool in vivo.
A STEAM‐ACED sequence (STEAM localization with Adiabatic Carbon Editing and Decoupling) has been proposed because of its short TE to minimize the effect of 1H‐1H coupling on the acquired signal intensity. 10 A BIRD filter 11 is basically a heteronuclear spin echo with delays equal to 1/2JCH to align 1H spins coupled to 12C and 13C spins in the opposite phase (–y and y axes), which allows a 90° pulse to align these vectors parallel and antiparallel to the main magnetic field (z and –z axes) respectively. After a 90° pulse, 1H attached to 12C spins are inverted and will recover based on T1 relaxation, while 1H attached to 13C spins are along the positive z‐axis. By choosing an optimal inversion time (TI), it is possible to suppress the 1H‐[12C] signal. Here, we hypothesize that pre‐suppression of the strong 1H‐[12C] lipid signal, by introducing a BIlinear Rotation Decoupling (BIRD) filter prior to a STEAM‐ACED sequence, optimized for detection of 1H‐[13C]‐lipid signals, will lead to an accurate and robust detection of the 1H‐[13C] lipid signal irrespective of respiratory motion, with higher signal yield as compared to ge‐HSQC.
2. METHODS
2.1. General
All MR experiments were performed on a 3T MR system (Achieva 3T‐X Philips Healthcare, Best, Netherlands) using a double tuned 1H/13C surface coil (Rapid Biomedical GmbH). The coil was specifically designed for indirect 13C experiments in the liver, with two 13C elements in quadrature arrangement (size of the elements: 13.2 × 21 cm, covering 13.2 × 23.9 cm) and two slightly larger 1H elements in quadrature arrangement (18 × 20 cm and 15 × 15 cm, covering 18 × 20 cm) to achieve optimal sensitivity (Expected at a depth of at least 8‐12 cm). All in vivo experiments were performed according to the protocol approved by the Institutional Human Medical Ethics Committee of Maastricht University Medical Centre (MUMC), Maastricht, Netherlands. A written informed consent was obtained from all the subjects prior to participating in this study.
2.2. MR sequences
For 1H‐[13C] editing of the lipid signals, a STEAM‐ACED 10 sequence was used with a TE of 7.9 ms (TE1 = TE2 = 3.95 ms = 1/2JCH). Spredrex pulses (Standard Philips 90° excitation pulse) with duration of 2.75 ms were used on the 1H channel for both excitation and selective refocusing of 1H signals. During the TM (Mixing Time) period, an adiabatic inversion pulse with a bandwidth of 1600 Hz was placed on the 13C channel to selectively invert the 1H‐[13C] spins. The 13C inversion pulse was used with OFF/ON condition on alternate scans. Subtraction of the acquired data with and without 13C inversion pulse provided MR signals only from 1H attached to 13C nuclei (two times), whereas the sum provided MR signals from all compounds containing 1H but not attached to 13C nuclei (shown in Supporting Information Figure S1, which is available online).
In a next experiment, the BIRD 11 filter was placed prior to the STEAM‐ACED sequence. The TI was optimized in vivo to achieve pre‐suppression of the large hindering lipid‐CH2 signal (1H‐[12C] spins) at 1.3 ppm in order to minimize the subtraction artifact. Block pulses were used in the BIRD filter on the 1H channel for both excitation and selective refocusing of 1H signals. A MLEV inversion pulse was used on the 13C channel as shown in Figure 1, to selectively invert the 13C–CH2 resonances (Figure 1). In ge‐HSQC sequence, the editing was performed with a MQC pulse sequence placed during the TM period. 6
Figure 1.

Application of a BIRD filter in front of a STEAM‐ACED sequence. The filter is basically a spin echo sequence with a 13C inversion pulse (MLEV) added to avoid J‐refocusing of the 1H‐[13C] signals, which will acquire opposite phase when compared with 1H‐[12C] in the xy‐plane by choosing the appropriate timings. The second 90° pulse is used to flip the 1H‐[12C] spins along the –z‐axis and the 1H‐[13C] along the +z‐axis. The TI is chosen according to the lipid CH2‐T1 relaxation time in order to null the 1H‐[12C] spins. As a result, only 1H‐[13C] spins will be excited by the excitation pulse of the localization sequence (STEAM‐ACED)
2.3. Phantom experiments
All MR sequence parameters and timings were optimized using a peanut oil phantom. Scout MR images were acquired with a turbo spin echo (TSE) sequence with repetition time/echo time (TR/TE) = 2000/100 ms to place a voxel in the center of the phantom. Power calibration was performed using MEGA‐sLASER sequence to achieve maximal 1H sensitivity in the volume of interest and the calculated power was kept same for both 1H and 13C channel throughout the experiment. First, STEAM‐ACED spectral editing sequence was used to acquire lipid 1H‐[13C] MR spectra, once with and without 13C inversion with following parameters TR/TE = 4000/7.9 ms, NSA = 128, voxel = 30 × 30 × 30 mm, data points = 2048. In a second experiment, the BIRD filter was used together with STEAM‐ACED to acquire lipid spectra using identical parameters as above, but now with a TI of 240 ms in which the 1H‐[12C] lipid signal was maximally suppressed. In a third experiment, ge‐HSQC was used as previously described, 6 to acquire 1H‐[13C] lipid spectra with a t1 of 4 ms. This set of experiments was used to compare the signal intensity and signal‐to‐noise ratio (SNR) of 1H‐[13C] lipid signal and therefore to determine the efficiency of the BIRD filter. SNR was determined by calculating the ratio of the 13C mean peak height and the standard deviation of noise in the edited spectrum.
2.4. In vivo MR experiments
Five healthy male subjects (mean age 34 ± 13 years; mean body mass index [BMI] 28.6 ± 3.7 kg/m2) were included in this study to determine the in vivo feasibility of our proposed approach. The subjects were positioned in the supine position. The 1H/13C surface coil was placed on the right side of the torso to cover the liver region. A fast scout MR image (slice thickness = 15 mm, flip angle = 15°, field of view [FOV] = 450 × 450 × 115 mm) was acquired in three orientations (axial, coronal, and transversal) with four breath holds to verify the positioning of the RF coil in the middle of the liver. A voxel of 45 × 45 × 45 mm was positioned in the liver by avoiding edges of the liver, diaphragm, but also the biliary and vascular structures. STEAM‐ACED was used to acquire hepatic lipid 1H‐[13C] MR spectra, once with and without 13C inversion with the same MR parameters as used in the phantom. With the same experimental setup, additionally a separate experiment was performed using a BIRD filter prior to the STEAM‐ACED sequence (TI = 240 ms) to selectively pre‐suppress the 1H‐[12C] lipid signal. The TI = 240 ms was chosen based on the average T1 relaxation time of lipid‐CH2 in the liver (340 ms), predicting that the signal would be nulled after TI of 240 ms. The long TR (4000 ms) was chosen to allow the subjects to breath in the rhythm of the measurement (MR acquisition at end of expiration) and to avoid saturation effects due to T1 relaxation.
2.5. Spectral analysis
All obtained MRS data were post‐processed with a custom‐built Matlab (version R2017b; The Mathworks Inc.) script. The developed script allowed us to select both series of spectra acquired with and without editing pulses at once for each subject and allowed for automatic phasing, eddy current correction, and frequency alignment of all spectra individually. Moreover, the script automatically removed spectra with poor quality based on a correlation approach, in which all individual spectra were ranked based on the correlation with all other spectra. An equal number of bad quality spectra were removed in both series acquired with and without editing pulse. Finally, the 1H‐[13C] lipid resonances obtained in the edited spectrum were fitted as two identical peaks based on iterative based algorithm developed within the Matlab script, using prior knowledge on the known frequency shift of 127 Hz (1H‐[13C] coupling for lipid). The best fit for the respective target lipid resonances were selected automatically when there was low residual (difference between fit and actual obtained spectrum) signal.
2.6. Calculation of 13C natural abundance and hepatic fat %
The total 1H lipid signal coupled to both 12C and 13C spins were determined from the fitted peak of lipid‐CH2 resonance of data acquired only with a conventional STEAM sequence (without 13C inversion pulse in STEAM‐ACED). The area of the 1H‐[13C] lipid signal (13C natural abundance) was calculated by fitting and adding the two obtained 1H‐[13C] lipid‐CH2 resonances in the edited spectrum acquired with and without 13C inversion pulse of STEAM‐ACED. The percentage of 13C enrichment was determined using the equation as shown below. Similarly, the 13C lipid signal was also calculated from the data acquired with the BIRD‐STEAM‐ACED sequence. Then, we compared the calculated percentage of 13C enrichment in both sequences to determine the percentage of signal loss from the expected value of 1.1% (natural abundance). The hepatic fat % was calculated after T2 correction (water T2 = 25.1 ms; lipid T2 = 56.7 ms), using the ratio of CH2/(CH2+H20) from the two spectra obtained with and without water suppression, acquired at TE = 20 ms using conventional STEAM sequence.
where S[total] represents the total 1H lipid‐CH2 signal coupled to both 12C and 13C spins (peak at 1.3 ppm in STEAM spectra) and S[C13] indicates only the addition of two signal intensities of 1H‐[13C] lipid resonances obtained in the edited spectrum.
3. RESULTS
3.1. Phantom
The lipid spectra acquired from the peanut oil phantom using the three different MR sequences are depicted in Figure 2. As expected, the subtraction artifact of 1H‐[12C] lipid signal was largely reduced with the ge‐HSQC sequence when compared with a regular JDE sequence (STEAM‐ACED; Figure 2A,B). Moreover, as expected, the measured 13C% enrichment with STEAM‐ACED was found to be 0.94% in an oil phantom, indicating only 15% signal loss, as the obtained value is close to the known value of 1.1% (13C natural abundance).
Figure 2.

1H‐[13C] lipid spectra obtained from the peanut oil phantom with three different sequences. A, ge‐HSQC. B, STEAM‐ACED. C, BIRD filter together with STEAM‐ACED
Interestingly, the proposed sequence with the addition of the BIRD filter reduces the subtraction artifact of the 1H‐[12C] lipid signal in the oil phantom when compared with a conventional STEAM‐ACED sequence. Moreover, the obtained 1H‐[13C] lipid signal intensity with the use of the BIRD filter was approximately two‐fold higher SNR (173 vs 91) compared with the previously proposed ge‐HSQC method (Figure 2C). The measured 13C natural abundance with the BIRD‐STEAM‐ACED sequence in an oil phantom was found to be 0.74%, indicating 21% signal loss in 1H‐[13C] lipid signal intensity due to the BIRD filter compared with the STEAM‐ACED sequence alone or if SNR is considered, we find a 16% loss in SNR (206 vs 173).
3.2. In vivo
Our new method was successfully applied in vivo and demonstrated the feasibility to detect 13C natural abundance in subjects with wide range of liver fat content (2‐21%, Table 1). As expected, a large subtraction artifact was apparent in in vivo when a conventional STEAM‐ACED sequence was used (Figure 3B). This artifact leads to phase distortions in the edited spectrum, which hampers the accurate and robust quantification of the 1H‐[13C] lipid signal (Figure 3B). The addition of the BIRD‐filter leads to pre‐suppression of the huge 1H‐[12C]‐lipid signal during MR acquisition and, thereby, leads to marked reduction in the subtraction artifact of the 1H‐[12C] lipid signal (Figure 3C).
Table 1.
Measured 1H‐[13C] lipid signal (natural abundance) with the BIRD‐STEAM‐ACED sequence in five different subjects with varying hepatic fat content (%)
| Fat % | 13C%‐BIRD‐STEAM‐ACED | |
|---|---|---|
| Subject 1 | 13.8 | 0.58 |
| Subject 2 | 3.5 | 0.63 |
| Subject 3 | 3.3 | 0.54 |
| Subject 4 | 20.6 | 0.53 |
| Subject 5 | 2.4 | 0.63 |
| Mean | 0.58 ± 0.05 | |
Figure 3.

A, Screenshot of scout MR image showing the voxel position in three different orientations of a volunteer. 1H‐[13C] hepatic lipid spectra obtained from the voxel with two different MR sequences. B, STEAM‐ACED. C, BIRD filter together with STEAM‐ACED
Due to the implementation of the BIRD filter, we were able to visualize our target 1H‐[13C] lipid signal clearly as two distinct peaks with known coupling constant of 127 Hz. We were able to quantify the 1H‐[13C] lipid signal successfully in all five subjects. The SNR was sufficient, even in the subject with very low hepatic lipid content. Interestingly, the phase distortions in vivo were minimal with the use of the BIRD filter when compared with STEAM‐ACED (Figure 3B,C). The measured 1H‐[13C] lipid signal in all five subjects with the use of the BIRD‐STEAM‐ACED sequence are shown in the Table 1.
4. DISCUSSION
In this study, we improved the robustness of a regular JDE (STEAM‐ACED sequence) technique for the application in human liver with the use of the BIRD filter and demonstrated the in vivo feasibility of our proposed approach. We were successful in the determination of the 13C (natural abundance) lipids in hepatic fat pool in all five subjects. The subtraction artifact of the 1H‐[12C] lipid signal is efficiently reduced with the use of the BIRD filter compared with the conventional STEAM‐ACED. Therefore, the two distinct peaks of the 1H‐[13C] lipid‐CH2 signal were clearly visible with the use of the BIRD filter together with STEAM‐ACED spectral editing sequence. Interestingly, with the BIRD filter, there is a minimum signal loss in 13C lipid signal intensity in phantom compared with the conventional STEAM‐ACED.
The addition of the BIRD filter leads to inherent 1H‐[13C] signal loss, probably due to 1H‐1H couplings. However, due to T1 recovery, the 1H‐[13C] signal losses due to the BIRD filter will be less with longer TI (when TI > actual null TI of lipid‐CH2), which results in higher 1H‐[13C] signal intensity. We aimed at nulling the 1H‐[12C] signal with the chosen TI. A longer TI would provide higher 1H‐[13C] lipid signal intensity, but simultaneously the 1H‐[12C] residue will also be increased. Thus, there is a tradeoff between 12C and 13C lipid signals. Due to differences in T1 relaxation time between in vivo and the oil phantom (340 vs 270 ms), the degree of recovery is slightly different; therefore, the oil phantom provides higher enrichment (0.74%) than in vivo (0.58%) while using identical TI (shown in Supporting Information Figure S2).
Moreover, both phantom and in vivo measurements yield lower 13C enrichment than expected (13C natural abundance of 1.1%). This indicates that there is (systemic) inherent signal loss, which might be due to 1H‐1H couplings. However, this is not a problem for 13C tracking applications in liver when indirect 13C MRS is generally used to measure changes in the 13C% enrichment, as the baseline enrichment can be set to 1.1%. This inherent signal loss will be taken into account to calculate the absolute 13C% enrichment.
Interestingly, the in vivo measured 13C lipid signal enrichment (natural abundance) with the BIRD‐STEAM‐ACED sequence was very similar among all subjects with varying liver fat content, suggesting that the present approach is very robust. Due to the insertion of the B1‐insensitive MLEV pulse on the 13C channel in the BIRD filter, signal loss is minimized, irrespective of B1 inhomogeneity while using a surface coil. In contrast, some B1 inhomogeneity is expected due to the use of standard pulses on the 1H channel, and while this can lead to some signal loss in absolute terms, it will not affect the estimation of 13C enrichment, as both the 1H‐[13C] signal and the reference signal are equally affected. Due to phase distortions and a large residual of 12C artifact, approaches to quantify the 1H‐[13C] lipid signal intensity with solely STEAM‐ACED were not successful in vivo. Thus, our proposed sequence with the BIRD filter was designed to minimize 1H‐[12C] signal and was successfully applied in vivo in subjects with varying total liver fat content and showed higher SNR for 13C lipids (natural abundance), even in the subject with low liver fat content.
Also the ge‐HSQC provides excellent editing of the 1H‐[13C] lipid‐CH2 peak in an oil phantom, but the signal loss of >50% compared with a conventional STEAM‐ACED is unavoidable. Our proposed sequence with the BIRD filter leads to a large signal gain in 1H‐[13C]‐lipid signal intensity and SNR (almost two‐fold) when compared with ge‐HSQC. Next to this, the use of JDE in combination with BIRD also has other advantages when compared with ge‐HSQC. Due to the relatively long T1, the hepatic water signal intensity is not much affected by the optimized TI of 240 ms. Therefore, the residual water signal can be used to perform phase and frequency alignment of individual spectra, whereas in ge‐HSQC, this is not possible due to the gradient de‐phasing of all 1H‐[12C] lipid signals and also water signals. 6 Also, a signal gain is expected when compared with ge‐HSQC in practice due to the insertion of B1 insensitive pulses in our approach (MLEV pulse in the BIRD filter and adiabatic hyperbolic secant pulse in the TM period of the STEAM‐ACED sequence). In contrast to ge‐HSQC, miscalibration of the 13C transmit power does not per se lead to lower a 13C‐enrichment, due to the B1 insensitivity of the pulses used in the proposed approach. The higher 13C signal (compared with ge‐HSQC) together with minimal 12C artifact with the use of the BIRD filter will enable to track even the smaller contribution of labeled glucose 12 to form 13C lipids (compared with the contribution from 13C‐labeled fatty acids) in the hepatic fat pool. Thus, our proposed approach can now be used as an efficient and promising tool for 13C tracking applications in vivo.
5. CONCLUSION
The proposed sequence with the BIRD filter reduces the subtraction artifacts of 1H‐[12C] lipid signal efficiently without much signal loss in the target 1H‐[13C] lipid signal; thus, it can be used as an indirect 13C method for 13C tracking applications in liver.
CONFLICT OF INTEREST
The authors declare no conflict of interest.
Supporting information
FIGURE S1 Sequence diagram of STEAM‐ACED sequence. 10 An adiabatic inversion pulse with bandwidth of 1600 Hz is only applied on 13C channel on alternate scans for editing purpose
FIGURE S2 1H‐MR lipid spectra acquired with different TI values from the peanut oil phantom (A) and from a healthy volunteer (B) using an inversion recovery with STEAM sequence. Plotted the signal intensity of lipid‐CH2 vs TI in the respective figure. The null TI (indicated by arrow) for the lipid‐CH2 was found to be 186 ms and 240 ms in oil phantom and in vivo respectively. Therefore T1 of lipid‐CH2 was estimated to be 270 ms in this oil phantom and 340 ms in in vivo
Veeraiah P, Brouwers K, Wildberger JE, Schrauwen‐Hinderling VB, Lindeboom L. Application of a BIlinear Rotation Decoupling (BIRD) filter in combination with J‐difference editing for indirect 13C measurements in the human liver. Magn Reson Med. 2020;84:2911–2917. 10.1002/mrm.28394
This research was in part financed by the Ministry of Economic Affairs and Climate Policy by means of the PPP Allowance made available by the Top Sector Life Sciences & Health to stimulate public‐private partnerships and by Unilever Foods Innovation Centre Wageningen. LL was a recipient of a veni grant from ZonMW (016.veni.188.036) and a junior fellowship (2017.81.004) from the Diabetes Fonds (Dutch Diabetes Research Foundation). VS‐H was supported by an ERC starting grant (grant no. 759161 “MRS in diabetes”).
REFERENCES
- 1. Rothman DL, De Feyter HM, de Graaf RA, Mason GF, Behar KL. 13C MRS studies of neuroenergetics and neurotransmitter cycling in humans. NMR Biomed. 2011;24:943‐957. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2. Beckmann N, Turkalj I, Seelig J, Keller U. 13C NMR for the assessment of human brain glucose metabolism in vivo. Biochemistry. 1991;30:6362‐6366. [DOI] [PubMed] [Google Scholar]
- 3. Petersen KF, Befroy DE, Dufour S, Rothman DL, Shulman GI. Assessment of hepatic mitochondrial oxidation and pyruvate cycling in NAFLD by (13)C magnetic resonance spectroscopy. Cell Metab. 2016;24:167‐171. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4. Chen H, De Feyter HM, Brown PB, Rothman DL, Cai S, de Graaf RA. Comparison of direct 13C and indirect 1H‐[13C] MR detection methods for the study of dynamic metabolic turnover in the human brain. J Magn Reson San Diego Calif. 1997;2017:33‐44. [DOI] [PubMed] [Google Scholar]
- 5. de Graaf RA, Rothman DL, Behar KL. State of the art direct 13C and indirect 1H‐[13C] NMR spectroscopy in vivo. A practical guide. NMR Biomed. 2011;24:958‐972. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Lindeboom L, de Graaf RA, Nabuurs CI, et al. Quantum coherence spectroscopy to measure dietary fat retention in the liver. JCI Insight. 2016;1:e84671. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 7. de Graaf RA, Mason GF, Patel AB, Behar KL, Rothman DL. In vivo 1H‐[13C]‐NMR spectroscopy of cerebral metabolism. NMR Biomed. 2003;16:339‐357. [DOI] [PubMed] [Google Scholar]
- 8. Fitzpatrick SM, Hetherington HP, Behar KL, Shulman RG. The flux from glucose to glutamate in the rat brain in vivo as determined by 1H‐observed, 13C‐edited NMR spectroscopy. J Cereb Blood Flow Metab Off J Int Soc Cereb Blood Flow Metab. 1990;10:170‐179. [DOI] [PubMed] [Google Scholar]
- 9. Jonkers RAM, Geraedts TR, van Loon LJC, Nicolay K, Prompers JJ. Multitissue assessment of in vivo postprandial intracellular lipid partitioning in rats using localized 1H‐[13C] magnetic resonance spectroscopy. Magn Reson Med. 2012;68:997‐1006. [DOI] [PubMed] [Google Scholar]
- 10. Pfeuffer J, Tkác I, Choi IY, et al. Localized in vivo 1H NMR detection of neurotransmitter labeling in rat brain during infusion of [1‐13C] D‐glucose. Magn Reson Med. 1999;41:1077‐1083. [DOI] [PubMed] [Google Scholar]
- 11. Garbow JR, Weitekamp DP, Pines A. Bilinear rotation decoupling of homonuclear scalar interactions. Chem Phys Lett. 1982;93:504‐509. [Google Scholar]
- 12. Silva JCP, Marques C, Martins FO, et al. Determining contributions of exogenous glucose and fructose to de novo fatty acid and glycerol synthesis in liver and adipose tissue. Metab Eng. 2019;56:69‐76. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
FIGURE S1 Sequence diagram of STEAM‐ACED sequence. 10 An adiabatic inversion pulse with bandwidth of 1600 Hz is only applied on 13C channel on alternate scans for editing purpose
FIGURE S2 1H‐MR lipid spectra acquired with different TI values from the peanut oil phantom (A) and from a healthy volunteer (B) using an inversion recovery with STEAM sequence. Plotted the signal intensity of lipid‐CH2 vs TI in the respective figure. The null TI (indicated by arrow) for the lipid‐CH2 was found to be 186 ms and 240 ms in oil phantom and in vivo respectively. Therefore T1 of lipid‐CH2 was estimated to be 270 ms in this oil phantom and 340 ms in in vivo