Home dialysis machine use for emergency dialysis during the COVID-19 pandemic

Cyril Mousseaux s1,s2,s3, Valentin Mayet s1, Anselme Poda s1, Chloé Schwarz s1,s2,s3, Samir Saheb s4, Jérôme Tourret s5,s6, Pierre Galichon s2,s3,s5, Nadia Arzouk s5, Inna Mohamadou s2,s5, Maud Cazenave s5, Hugo Garcia s1,s5, Eric Rondeau s1,s2,s3, Laurent Mesnard s1,s2,s3, Yosu Luque s1,s2,s3,
PMCID: PMC7543322  PMID: 33117529

The coronavirus disease 2019 (COVID-19) pandemic resulted in a reduced number of beds available for patients with other pathologies, such as acute kidney injury (AKI) [1]. We describe the adaptation of our renal intensive care unit (RICU) during the pandemic and novel use of home dialysis machines to provide emergency dialysis.

In late March 2020, in order to increase the critical care capacity of our hospital for COVID-19 patients, we converted our 15-bed RICU in Tenon Hospital (Paris, France) to intensive care beds, transferring the unit to another hospital (Pitié-Salpêtrière, Paris, France) with capacity for 6 monitored beds. The nursing staff within this new RICU was neither dialysis-trained nor was there an ultrapure water supply to provide haemodialysis. Although haemodialysis was feasible during the day in the hospital dialysis unit, emergency dialysis could only be provided at night in the intensive care department, which was already overburdened with COVID-19 patients. As a solution, we acquired a home dialysis machine (Physidia S3, Physidia, Saint-Barthélémy d’Anjou, France) on which the medical staff were trained by the manufacturer in a half-day course. This machine allows 2–3 h of dialysis via a central venous catheter or arteriovenous fistula. Instead of the use of a reverse osmosis water circuit, pre-prepared bagged sterile fluid is used. The dialysis potassium concentration is fixed at 1 mmol/L. The maximum hourly ultrafiltration is 1 L and no anticoagulation is needed. Within 2 days, four residents and three nephrologists could set up the machine.

From 30 March to 5 May, 34 patients were admitted in the relocated RICU. The reasons for hospitalization were AKI (n = 24), end-stage renal disease (ESRD; n = 6) and electrolyte disorders (n = 4). Causes of AKI were acute tubular necrosis (n = 7), cardiorenal syndrome (n = 5), anti-neutrophil cytoplasmic antibody (ANCA) vasculitis (n = 3), anti-glomerular basement membrane (GBM) disease (n = 1), malignant hypertension with thrombotic microangiopathy (n = 2), scleroderma renal crisis (n = 1), hemolysis elevated liver enzymes, low platelets syndrome (n = 1), graft rejection (n = 1) and post-renal (n = 2) and myeloma cast nephropathy (n = 1). For eight patients, at least one emergency haemodialysis session with the home dialysis machine was performed. There were 21 dialysis sessions in total during this period using the new device. The patients' characteristics are shown in Table 1. The most common first emergency dialysis indication was fluid overload (n = 6), followed by hyperkalaemia (n = 1) and ESRD with uraemic symptoms (n = 1). Blood flow and dialysate rates were low (150–220 mL/min and 150–200 mL/min, respectively). Only one patient was dialysed using an arteriovenous fistula and the remainder required emergency central venous catheterization. Small molecule clearance was estimated using the urea reduction ratio: the median rate was 26% (interquartile range 25–75:23–31). Patient 5 had 10 sessions of combined haemodialysis and plasma exchange (PEX) for anti-GBM (anti-glomerular basement membrane) disease, as we were able to connect both the haemodialysis and the PEX machines in parallel on her catheter. Haemodynamic tolerance was optimal for all the patients and we noted no catheter-associated infection or symptomatic hypokalaemia.

Table 1.

Characteristics of patients requiring emergency dialysis with home dialysis machine

Patient Age (years) Gender Other pathologies Cause of kidney failure Session indication Vascular access Session length (hours) Blood flow rate (mL/min) Dialysate rate (mL/min) Ultrafiltration (L) Number of sessions Complications URR (%) K+ before K+ after
1 55 M HIV ATN Fluid overload Catheter 2 220 150 1.8 1 None 31 3.9 3.0
2 59 M

Diabetes mellitus,

hypertension

Nephroangiosclerosis and diabetic nephropathy Fluid overload Catheter 2 220 150 1.8 1 None 25 5.1 3.4
3 74 M

HIV,

hypertension

Unknown HyperK Fistula 2 220 200 1.5 1 High venous pressure 32 6.1 5.2
4 74 M

Diabetes mellitus,

hypertension

Cardiorenal syndrome Fluid overload Catheter 2 220 150 1.8 1 None 17.8 4.5 3.7
5 68 F Obesity Goodpasture disease Fluid overload Catheter 2 220 150 1.7 10 None 37.3 4 3.5
6 35 M Hypertension ANCA vasculitis Fluid overload Catheter 2 220 150 1.7 2 High venous pressure 15.7 4.9 4.1
7 57 F None ANCA vasculitis Fluid overload Catheter 2 220 150 1.9 3 Low arterial pressure 37.7 3.7 2.9
8 74 M

Cardiac transplantation,

diabetes mellitus

CNI/ cardio renal syndrome Fluid overload Catheter 2 150 150 1 2 None 26.1 4.9 3.7

Kalaemia (K+) values and URR were calculated on the first session. ATN: acute tubular necrosis; CNI: calcineurin inhibitor; HyperK: hyperkalaemia; URR: urea reduction ratio.

In order to ensure continuity of nephrology critical care (for non-COVID-19 patients) and reduce pressure on other emergency departments, our renal RICU was maintained but relocated during the pandemic. To our knowledge, this technique has never been used in a critical care context. The use of home dialysis machines has been associated with increased survival [2] and quality of life [3] in ESRD patients. We demonstrated that the technique can be useful for AKI treatment in a time of crisis, with a short training period for medical staff with good results. Our experience shows that it is easy to handle and safe.

In conclusion, we describe the use of a home dialysis machine for emergency dialysis with good results and tolerance during the COVID-19 pandemic. We believe this innovative technique could be of interest in other contexts, such as emergency dialysis in low-income countries or disaster sites with no facilities for water treatment.

AUTHORS’ CONTRIBUTIONS

The research idea and study design were carried out by Y.L., L.M. and E.R. Data acquisition was performed by C.M., V.M., A.P., C.S. and H.G. Data analysis/interpretation was performed by C.M. and Y.L. Supervision or mentorship was provided by L.M. and E.R. J.T., I.M., P.G., S.S., M.C. and N.A. provided critical feedback and helped shape the research, analysis and manuscript. Each author contributed important intellectual content during manuscript drafting or revision, accepts personal accountability for the author’s own contributions and agrees to ensure that questions pertaining to the accuracy or integrity of any portion of the work are appropriately investigated and resolved.

CONFLICT OF INTEREST STATEMENT

The authors report that they have no relevant financial interests. The results presented in this article have not been published previously in whole or part.

ACKNOWLEDGEMENTS

We want to thank Physidia for providing one home dialysis machine in the relocated RICU, technical support and training for the medical staff.

DATA AVAILABILITY

The data that support the findings of this study are available from the corresponding author, YL, upon reasonable request.

REFERENCES

  • 1. Verelst F, Kuylen E, Beutels P.. Indications for healthcare surge capacity in European countries facing an exponential increase in coronavirus disease (COVID-19) cases, March 2020. Eurosurveillance 2020; 25: 2020323. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Kjellstrand CM, Buoncristiani U, Ting G. et al. Short daily haemodialysis: survival in 415 patients treated for 1006 patient-years. Nephrol Dial Transplant 2008; 23: 3283–3289 [DOI] [PubMed] [Google Scholar]
  • 3. Wyld M, Morton RL, Hayen A. et al. A systematic review and meta-analysis of utility-based quality of life in chronic kidney disease treatments. PLoS Med 2012; 9: e1001307. [DOI] [PMC free article] [PubMed] [Google Scholar]