Skip to main content
PLOS One logoLink to PLOS One
. 2020 Oct 12;15(10):e0240350. doi: 10.1371/journal.pone.0240350

Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients

Thibaud Garcin 1,2,*, Damien Grivet 1, Gilles Thuret 1,2,3, Philippe Gain 1,2
Editor: Ireneusz Grulkowski4
PMCID: PMC7549767  PMID: 33044993

Abstract

The Optical Quality Analysis System (OQAS, Visiometrics) provides objective measurements of image formed onto retina, by combining quantification of ocular media transparency and of optical aberrations. In order to evaluate its contribution in the assessment of age-related cataract, we conducted a monocentric clinical study to determine the relationships between clinical grading of lens opacity, OQAS parameters, and parameters required for cataract surgery by phacoemulsification with ultrasound (called “phacodynamics”). Clinical parameters were: best-corrected visual acuity (BCVA, expressed as Log of minimal angle resolution (logMAR)) and the lens opacity classification system III (LOCS III) as a gold standard determined by two independent observers who graded total cataract and nuclear, cortical and posterior sub capsular components. The OQAS provided an objective scatter index (OSI), a modulation transfer function (MTF, expressed in cycle per degree (cpd)) and a Strehl ratio (SR) used as an aberration marker. Patients were operated on by the same surgeon using a phacoemulsification machine that provided the cumulative dissipated energy (CDE) and total ultrasound time (US time) necessary to extract the lens. Patients with poor compliance, corneal or retinal diseases impairing OSI, or who required surgical settings variation, were excluded. Twenty-one eyes of 21 patients aged 76±8 years were analyzed. They were 11 pure nuclear, 3 pure cortical, and 7 mixed cataracts. Mean LOCS III and OSI were respectively: 4.86 ±2.03 and 6.12 ±3.07 (mean±SD). Medians (10°-90° percentiles) were: for BCVA 0.30 (0.10–0.70) logMAR, for MTF cutoff 9.31 (1.54–30.57) cpd, for SR 0.071 (0.042–0.146), for CDE 8.04 (5.74–23.29) and for US time 58 (39–116) seconds. LOCS III was significantly correlated (spearman r, rs) with BCVA (rs = 0.561, p = 0.008), CDE (rs = 0.457, p = 0.038) and US time (rs = 0.647, p = 0.002). The three OQAS parameters significantly correlated (all rs ≥ 0.526, p<0.05) with BCVA, and LOCS III grading, but the strongest correlations were found with OSI for cortical components and with MTF for nuclear components: only OSI may be used objectively to assess the effect of cortical components on optical quality, and MTF cutoff—integrating scattering and aberrations—seems the best objective parameter for clinical assessment of nuclear cataracts. The three OQAS parameters were also significantly correlated (rs) with CDE, and with US time only for pure nuclear cataracts: OSI had the strongest correlations with phacodynamics (rs = 0.693, p = 0.022 with CDE and rs = 0.703, p = 0.019 US time). OSI increased with cortical components not requiring higher CDE. When measured in optimal conditions (good compliance, no retinal or ocular surface or tear film diseases), the three OQAS parameters are complementary for objective grading of cataract. In the future, they may help to optimize surgical parameters, especially energy distribution, in femtosecond laser assisted cataract surgery.

Introduction

Cataract surgery is a major challenge in the world since cataract is the leading cause of blindness and the second-ranking cause of moderate to severe vision impairment [1]. In countries with high standard of living it is the most common surgical procedure, with a constantly increasing number of operated eyes [2, 3]. In France 830,000 cataract surgeries were performed in 2017 [4]. Cataracts are being operated on earlier and earlier due to demographic changes, changing lifestyles, patients’ expectations and refinement of surgical techniques.

Health authorities [58] and learned societies [9] have long sought to define and regulate decision-making in cataract surgery. Moreover, this surgery is personalized, so it is crucial to define objective criteria that more effectively meet patient needs. In addition, a reliable quantification of lens opacification (accepted as a surrogate criterion for its hardness) could make it possible to optimize the surgical parameters necessary to fragment the lens, whether with ultrasound or with a femtosecond laser (these parameters are hereafter called phacodynamics).

The gold standard gradation system for cataracts remains clinical, based on the Lens Opacities Classification System (LOCS) III [10]. This score, established in 1993 and never revisited, is determined by comparison with a series of reference photographs and separately analyzes the three possible components of a cataract: nucleus (color (NC) and opacification (NO)), cortex (C) and posterior sub-capsular layers (P). Grading should ideally be performed on standardized images, but in practice is most often done by live slit-lamp observation. Despite potential observer and reproducibility biases [11, 12], this is still the scale most widely used to grade cataracts for research or clinical purposes, although there are other systems, such as the Winconsin cataract grading system [13].

New criteria must be explored and integrated in the surgical plan to better meet the expectations of patients with cataract. Various objective techniques have been developed to qualitatively and quantitatively assess lens opacification, which is useful for surgery decision-making. Scheimpflug imaging uses a densitometric analysis that is correlated with LOCS III [1419] and phacodynamics [16, 1923] for nuclear cataracts. Lens nucleus density measured by anterior segment optical coherence tomography is also correlated with LOCS III [23, 24]. Wavefront analyzers, measure higher-order aberrations (HOAs), which increase in age-related cataract patients [2530]. A ray-tracing aberrometry system (iTrace Visual Function Analyzer, Tracey Technologies, Houston, TX) measures a dysfunctional lens index (DLI) in nuclear cataracts [31, 32]. However, none of these techniques measures cataract-induced transparency loss. Shack-Hartmann technology do not consider light scattering [33] and can overestimate the optical quality of an image when scattering affect the eyes, for example in cataract [34].

However, objective measurement of intraocular scattering seems a good way to evaluate the impact of age-related cataract, as this scattering increases with age [3537]. The Optical Quality Analysis System (OQAS) (HD Analyzer II, Visiometrics SL, Terrassa, Spain) is a double pass system (double pass required to collect light focused on the macula using a 780 nm infrared laser diode) [38, 39] that provides objective measurements of the image formed onto the retina, by combining the quantification of optical aberrations and of forward plus backward light scattering caused by loss of ocular transparency [34, 40]. The OQAS provides three measures: the objective scatter index (OSI) measures light scattering and optical aberrations [41] by the ratio between integrated light in the peripheral ring and in the area surrounding the central peak of the double-pass point spread function (PSF) image. The OSI scale ranges from 0 (no scattering) to 25 (maximum scattering). The Modulation Transfer Function (MTF) represents the contrast loss in retinal images at various spatial frequencies [37]. The MTF cutoff, in cycle per degree (cpd), is the highest spatial frequency that the eye can detect: the higher the MTF cutoff value, the better the optical quality [42]. The Strehl ratio (SR), for low optic aberrations, is the ratio between the peak intensity from the point spread function (PSF) of the aberrated eye and the peak intensity from the PSF of the unaberrated eye [37, 43]: a higher value indicates better optical quality. The OQAS presents good reproducibility [44] in everyday practice, and repeatability [45] for patients with cataracts or after refractive surgery.

In this research, we analyzed the relationships between the clinical assessment of cataract severity, QAS measurements, and the surgical parameters (energy and time) required to emulsify the lens, in order to analyze the value of the OQAS in cataract surgery planning.

Material and methods

Patients and ethic statement

All age-related cataract patients scheduled for surgery under topical anesthesia by a senior surgeon (DG) were included for 30 consecutive days in St Etienne University Hospital. All patients were informed of the nature and intent of the study, and their consent was collected. Study was approved by the local Institutional Review Board “Ethics Committee of the St Etienne University Hospital, Research Commission of Terre d’Ethique” (IORG0007394, IRBN172019/CHUSTE) in accordance with the tenets of the Declaration of Helsinki.

Preoperative subjective and objective assessments

Complete preoperative examination comprised: preoperative far best-corrected visual acuity (BCVA) measured with routine Monoyer chart (converted to logMAR for analysis), slit lamp evaluation, Goldmann applanation tonometry, fundus examination and optical coherence tomography measurements to exclude macular impairment. Cataracts were graded using LOCS III [10, 11] by two observers (TG/DG), blind to each other, with a slit lamp, after pupil dilation (pupil ≥5mm). Briefly, the three components of cataract were evaluated separately: nuclear opacity (NO) and nuclear color (NC), with scores ranging from 0.1 (clear or colorless) to 6.9 (very opaque or brunescent cataract); posterior sub-capsular (P) and cortical (C), with scores ranging from 0.1 (clear or colorless) to 5.9 (very opaque). The overall score for each cataract (LOCS III total) obtained by adding up the four sub-scores was considered as a continuous variable. Each cataract type was termed pure when no other component was present. Pure nuclear cataracts (N) were quantified by adding NO+NC.

Measurements by the OQAS were done by a single observer (TG) on both eyes of each patient, before any instillation of eyedrops, as recommended by the manufacturer [40, 46], to avoid any measurement bias. The first step was tear film analysis, with OSI measured every 0.5 second during 20 seconds as recommended by the manufacturer, to objectively identify dry eye disease that could prevent from collection of reliable data [47, 48]. Then, data of interest (OSI, MTF cutoff, SR) were acquired by repeating measurements at least six times, as detailed previously by various authors [45, 49, 50] and by calculating their mean without excluding any data.

Surgical parameters (phacodynamics)

All patients were operated on by the same experienced surgeon (DG) with the same technique, under topical anesthesia by oxybuprocaine chlorhydrate 0,4% and tetracaine 1% (both from Laboratories Théa, Clermont-Ferrand, France) instilled twice in the conjunctival sac, 5 minutes apart, 10 minutes before draping. Surgery was performed through a 2.4mm corneal incision on the 12 o'clock meridian. Dilation was obtained by injecting intracameral Mydrane® 0,2mL (Laboratories Théa). A divide and conquer technique was used with 75% amplitude oZil continuous mode, 10% power phacoemulsification mode, and grade 2 cataract mode on the Infinity® phacoemulsifier (Alcon Laboratories, Fort Worth, Tx, USA). The latter directly provided the cumulative dissipated energy (CDE) and total ultrasound time (US time in seconds). The internal software calculated the CDE in a standardized manner as follows: (torsional amplitude x torsional time x 0.4) + (phaco time x average phaco power).

Inclusion and exclusion criteria

Of the 28 pre-screened patients, four patients did not meet inclusion criteria as we were unable to obtain optimal OSI measurements on the OQAS [46, 48, 51, 52]: two for poor compliance; one for retinal disease, and one ocular surface diseases (poor-quality tear film).

Twenty-four patients and 26 eyes were included for surgery. Five operated eyes were excluded from analysis to avoid any bias: two eyes for poor cooperation during surgery (thus modifying US time) and two with particularly dense cataract requiring phacodynamics mode adjustment; and one chosen at random for a patient operated on both sides (to avoid statistical bias). In total, we analyzed 21 eyes from 10 women and 11 men with a mean age 76±8 years [range 54–90 years].

Statistical analysis

Only operated eyes from patients meeting the inclusion and exclusion criteria were studied. Normality of continuous data distribution was analyzed with the Shapiro-Wilk test, with a non-normality threshold set at 5%. Normally distributed data were described by their mean ± standard deviation (SD). Continuous non-normally distributed variables were summarized as median (10°-90° percentiles).

Spearman (rs) correlation coefficients were calculated according to data normality. The sole independent variable was LOCS III grading for cataract severity. All other variables were dependent. Agreement between the two ophthalmologists for LOCS III grading was determined by calculating the mean difference and the correlation coefficient. Partial correlation was used to analyze the relationship between OQAS measurements and phacodynamics parameters controlling for the LOCS III. The relationship between two variables was represented graphically using linear regression. On the linear regression graphs, the red dots show the 95% confidence interval (95% CI) of the slope and intercept. Statistical significance was based on two-tailed statistical analyses, and probability values <0.05 were considered statistically significant, and the beta risk level was 10%. Statistical analyses and graphs were produced using GraphPad Prism 6.0 and SPSS 25.0 IBM Corp, Armonk, NY, USA.

Results

Baseline population data

For the 21 operated eyes, median far BCVA was 0.30 (0.10–0.70) logMAR, mean LOCS III total score was 4.86 ±2.03 with excellent agreement between observers (mean difference 0.3; r = 0.94 p<0.001, 19 cases with perfect agreement, three cases with a two-point difference).

There were 18 nuclear, 10 cortical and one subcapsular components. Mean OSI was 6.12 ±3.07, median MTF cutoff was 9.31 cpd (1.54–30.57) and median SR was 0.071 (0.042–0.146). All preoperative population data are summarized in Table 1. Distribution of the components is shown in Table 2: note that each eye may have one or more components. We found 11 pure nuclear cataracts and three pure cortical cataracts.

Table 1. Preoperative baseline data (n = 21 eyes).

Data Mean ±SD
Median (10°- 90° percentiles)*
Range
Far BCVA (BCVA) LogMAR 0.30 (0.10–0.70) 0.10–0.80
Spherical Equivalent -0.375 (-5.20 –+1.80) -6.5 to +1.875
LOCS III total score (NO+NC+C+P) 4.86 ±2.03 2.00–10.00
LOCS III N score n = 18 4.39 ±1.94 2–10
LOCS III NO score n = 18 2.22 ±1.00 1–5
LOCS III NC score n = 18 2.17 ±0.99 1–5
LOCS III C score n = 10 2.00 ±1.16 1–4
LOCS III P score n = 1 3 -
OSI 6.12 ±3.07 1.40–11.40
MTF cutoff (cycle/degree) 9.31 (1.54–30.57) 1.54–36.45
Strehl ratio 0.071 (0.042–0.146) 0.042–0.148

*according to data normality distribution. BCVA = best-corrected visual acuity, LOCS III = lens opacity classification system III, NO = nuclear opalescence, NC = nuclear color, C = cortical opacities, P = posterior subcapsular opacities, OSI = objective scatter index, MTF = Modulation Transfer Function, SD = standard deviation.

Table 2. Lens Opacification System III scores distribution, mean values and standard deviation of OSI, median values and percentiles of MTF cutoff and of Strehl ratio (n = 21 eyes).

LOCS III score NO score (n = 18) NC score (n = 18) C score (n = 10) P score (n = 1)
1–1.9 4 (22%) 4 (22%) 5 (50%) -
2–2.9 8 (44%) 9 (50%) 1 (10%) -
3–3.9 5 (28%) 4 (22%) 3 (30%) 1 (100%)
4–4.9 - - 1 (10%) -
5–5.9 1 (6%) 1 (6%) - -
6–6.9 - -
OSI Mean ±SD 5.58 ±2.92 5.58 ±2.92 7.27 ±3.28 3
MTF cutoff Median (10°-90° percentiles) 10.29
(1.54–31.30)
10.29
(1.54–31.30)
6.40
(1.54–34.62)
1.54
Strehl ratio Median (10°-90° percentiles) 0.077
(0.042–0.147)
0.077
(0.042–0.147)
0.059
(0.042–0.144)
0.042

LOCS = lens opacity classification system, NO = nuclear opalescence, NC = nuclear color, C = cortical opacities, P = posterior subcapsular opacities, OSI = objective scatter index, MTF = Modulation Transfer Function, SD = standard deviation.

Regarding phacodynamics, median CDE was 8.04 (5.74–23.29) [2.98–26.47]; median US time was 58 (39–116) seconds [25–122].

Correlations analysis

Main objective, subjective preoperative parameters, and phacodynamics

We found significant correlations between: 1/ BCVA and OSI (rs = 0.526, p = 0.014), BCVA and LOCS III total (rs = 0.561, p = 0.008) as well as in subgroups of different LOCS III components (N, NC, NO with significant rs = 0.633, rs = 0.621, rs = 0.581 respectively) except for LOCS III C (p = 0.696) (Fig 1); 2/ LOCS III total and OSI (rs = 0.586, p = 0.005) as well as in subgroups of different LOCS III components (N, NC, NO, C) (Fig 2); 3/ LOCS III total and CDE (rs = 0.457, p = 0.038) as well as in subgroups of different LOCS III components (N, NC, NO) except for LOCS III C (p = 0.052); LOCS III total and US time (rs = 0.647, p = 0.002) as well as in subgroups of different LOCS III components (N, NC, NO) except for LOCS III C (p = 0.054) (Fig 3).

Fig 1. Linear regressions between BCVA and OSI, and between BCVA and LOCS III total.

Fig 1

Red dots show the 95% CI of the slope and intercept. OSI = objective scatter index, BCVA = far logMAR best-corrected visual acuity, LOCS III = lens opacity classification system III.

Fig 2. Linear regressions between OSI and LOCS III.

Fig 2

Red dots show the 95% CI of the slope and intercept. OSI = objective scatter index, LOCS III = lens opacity classification system III, N = nuclear component = (NO+NC), NO = nuclear opalescence, NC = nuclear color, C = cortical opacities.

Fig 3. Linear regressions between LOCS III and surgical parameters.

Fig 3

Red dots show the 95% CI of the slope and intercept. CDE = cumulative dissipated energy, US = ultrasound, LOCS III = lens opacity classification system III, N = nuclear component = (NO+NC), NO = nuclear opalescence, NC = nuclear color, C = cortical opacities.

For pure nuclear cataracts only, CDE increased significantly with OSI (rs = 0.693, p = 0.022), as well as total US time with OSI (rs = 0.703, p = 0.019) (Fig 4). No significant correlation was found between OSI and CDE if N, C, P components were taken together for analysis. Partial correlation between CDE and OSI, US time and OSI controlling for the LOCS III was negligible (r = 0.013 and r = -0.041 respectively) and not significant (p = 0.993 and p = 0.910 respectively), thus indicating that LOCS III N score greatly influenced the relation between OSI and surgical parameters.

Fig 4. Linear regressions between OSI and surgical parameters for all cataracts and for pure nuclear cataracts.

Fig 4

Red dots show the 95% CI of the slope and intercept. OSI = objective scatter index, US = ultrasound, CDE = cumulative dissipated energy, LOCS III = lens opacity classification system III, N pure = pure nuclear cataract.

Other OQAS parameters, subjective preoperative parameters and phacodynamics

Strong significant correlations were found between: 1/ BCVA and MTF cutoff (rs = -0.685, p<0.001); 2/ LOCS III total and MTF cutoff (rs = -0.711, p<0.001) as well as in subgroups of different LOCS III components (N, NC, NO with significant rs = -0.771, rs = -0.675, rs = -0.762 respectively) except for LOCS III C (p = 0.186); 3/ MTF cutoff and CDE (rs = -0.655, p = 0.034), and also MTF cutoff and US time (rs = -0.648, p = 0.034) only for pure nuclear cataracts (Fig 5).

Fig 5. Linear regressions between MTF cutoff and clinical or surgical parameters.

Fig 5

Red dots show the 95% CI of the slope and intercept. MTF = Modulation Transfer Function, BCVA = far logMAR best-corrected visual acuity, LOCS III = lens opacity classification system III, CDE = cumulative dissipated energy, US = ultrasound.

We also found strong significant correlations between: 1/ BCVA and Strehl ratio (rs = -0.622, p = 0.003); 2/ LOCS III total and Strehl ratio (rs = -0.676, p<0.001) as well as in subgroups of different LOCS III components (N, NC, NO with significant rs = -0.745, rs = -0.640, rs = -0.758 respectively) except for LOCS III C (p = 0.056); 3/ Strehl ratio and CDE (rs = -0.664, p = 0.031), and also Strehl ratio and US time (rs = -0.653, p = 0.032) only for pure nuclear cataracts (Fig 6).

Fig 6. Linear regressions between Strehl ratio and clinical or surgical parameters.

Fig 6

Red dots show the 95% CI of the slope and intercept. BCVA = far logMAR best-corrected visual acuity, LOCS III = lens opacity classification system III, CDE = cumulative dissipated energy, US = ultrasound.

Discussion

In our study, the OQAS was examined to assess its usefulness in clinical practice and try to answer if it may be a reliable tool for clinical objective assessment of different age-related cataract types, but also for prediction of phacodynamics.

Our population sample was representative of general French population who undergo cataract surgery [4]. Even if BCVA may be less considered now, as the latest recommendations not include it in surgery decision-making, our mean BCVA respected the previous cutoff established by the competent Health Authorities, like Hwang et al. [53], or contrary to some studies evaluating the OQAS on patients with lower preoperative logMAR BCVA at earlier stages age-related cataract [21, 49, 50, 5456].

As a gold standard for cataract grading, we used the LOCS III classification, as in the initial publication [10] and obtained a strong agreement between both observers and reliable classification [1012]. The level of LOCS III for surgery decision-making has decreased over time since 1993 [50, 5658] which limits studies’ comparability. In addition, in the first OQAS study, Artal et al. [40] interpreted LOCS III classification by redefining three LOCS III subgroups for nuclear cataracts, as Vilaseca et al. [59] did for the three cataract components (N/C/P), limiting comparability between papers on clinical grading of cataracts.

As nuclear and cortical components were well represented in our population, correlations between the LOCS III total and BCVA confirmed results from previous studies [14, 21, 50]. LOCS III remains convenient, and cost-effective to assess the impact of nuclear cataract on visual acuity. This is not the case for the cortical components because all opacities can be graded since there are located at the periphery, whereas BCVA decreased only when central cortex (generally the central 4.0mm) is concerned by moderate to advanced cortical cataracts.

As previously described [23, 60, 61], we logically found moderate to strong correlations only between LOCS III grading of nuclear components and phacodynamics. Contrary to the lens nucleus that becomes harder when it loses transparency with age (due to centripetal compaction of lens fibers), the cortex, composed of loose fibers, remains soft when it becomes opaque and does not require increased ultrasound energy nor time (only irrigation-aspiration is usually used).

Previous studies on the OQAS showed no consensus on population selection: most authors chose to include cataracts with different components (N/C/P) [49, 50, 53, 55, 56, 59], but one selected only nuclear cataracts [21]. As cortical components cause disability in daily life without necessarily greatly impairing visual acuity [57, 62], we chose to consider every type of cataract (N/C/P) for both population selection and LOCS III grading, to prevent potential bias: Artal et al. [40] did not consider this point in selecting their population, and mentioned it as a limitation. Other authors analyzed correlations for nuclear cataracts only [50], or considered all components as nuclear cataracts, judging that other types of cataracts were poorly represented in their study [54].

The principle of OQAS measurement requires a healthy eye surface. We have therefore excluded dry eye syndromes that affect the passage of light, to provide reliable objective data as recommended [45, 46, 49, 51].

We found moderate correlation between OSI and BCVA, like other teams [49, 56], while other authors found a stronger correlation [21, 50, 59]. Differences can be explained by centers' different visual acuity thresholds for surgery decision-making, and, as explained above for correlation between LOCS III and BCVA, by the different proportions of nuclear and cortical cataracts [59], cortical components being well represented with different grades in our population.

Similarly, we found moderate to strong correlations between OSI and different components of LOCS III. Again, variations exist with previous studies which found stronger [21] or weaker correlations [50, 54, 56] probably because of differences in spectrum of cataract severity.

Until now, among the three measurements provided by the OQAS, OSI is the most frequently studied and seemed the most effective one for cataract surgery decision-making with high specificity and sensibility [53, 55], with OSI from 3 to 7 deemed a good indication for surgery when all cataract components (N/C/P) are considered [40, 50]. Notably, OSI is influenced by low or high order aberrations [40], which explains why at least spheric and cylindric ametropias must be corrected before OQAS measurement. In the present study, we also analyzed the MTF cutoff and the SR.

The MTF cutoff and the SR values decreased significantly as OSI increased, even more than previously reported [50]. The MTF cutoff was strongly correlated with LOCS III N, with higher association than previously described [50, 63]. Cortical components are known to increase OSI [49, 55, 59]. Interestingly, we found no correlation between MTF cutoff and LOCS III C or SR and LOCS III C (more sensitive to pupil diameter), whereas OSI increased strongly with LOCS III C. We hypothesize that variations in cataract severity between studies could explain this difference, as some of our patients, had visible cortical opacities in the 4.0mm pupillary area. In total, by combining the three parameters, OQAS thus makes it possible to separate the impact of moderate to severe cortical cataract components on visual quality. In contrast, compared to OSI or SR, MTF cutoff had the strongest correlations with BCVA, LOCS III N, NO, NC. As MTF cutoff simultaneously considers scattering and aberrations [49, 63], MTF cutoff may be the best objective OQAS parameter for clinical assessment of nuclear cataracts.

Reliable prediction of phacodynamics would considerably help to optimize cataract surgery devices (ultrasound or femtosecond laser), avoiding over- and underestimations due to preset parameters, which are both detrimental. We chose CDE as the main criterion because it is the most relevant data currently used to quantify US energy as used to emulsify the lens [64].

Importantly, our study reports a new key point: when considering mixed cataract types, no correlation exists between OQAS measurements (OSI, MTF cutoff, SR) and phacodynamics (CDE or US time). OSI increased with cortical components that not requiring more CDE or US time. After removing cortical and sub capsular components, for pure nuclear cataracts only, OSI ranged from 1.4 to 8.3 and strong correlations were found between CDE and OSI. The correlation was even stronger than between CDE and LOCS III N, suggesting that OQAS could better predict phacodynamics than LOCS III for pure nuclear cataracts. The same findings were found for US time and OSI, and for US time and LOCS III N. Partial correlation analysis clearly demonstrates that LOCS III acts as a confounding factor to explain the strong correlation between OSI and phacodynamics. This is because LOCS III and the OQAS both measures the same physical phenomenon -lens opacification- if there are no further ocular pathologies. This demonstrates OQAS reliability as objective and automatic measurement. In addition, OSI had stronger correlations with phacodynamics than MTF cutoff or SR, suggesting that OSI may be the OQAS parameter for optimizing surgical plans with energy modulation. Correlation between OSI and CDE was also previously studied [21] as secondary endpoint, with several limitations including: selection only of cataracts with a predominant nuclear component without stating whether there were pure nuclear cataracts, contrary to Artal et al. [40]; absence of details on surgical protocol, suggesting possibly variation in preset program and thus in energy levels; and no clear exclusion of ocular surface diseases, which may change OQAS measurements [47, 48, 52].

Our research also has limitations: 1/ although our sample is representative of patients typically operated for age-related cataract in many centers worldwide, it is small, limiting the range of LOCS III scores. No statistic could be generated for subcapsular posterior cataracts (n = 1). Lack of statistical significance for correlation between LOCS III C and CDE as well as US time may be due to the insufficient number of patients in this subgroup; 2/ despite being the clinical gold standard, LOCS III does not differentiate peripheral from central opacities, resulting in discordance between symptoms and the LOCS III score [62]. Therefore, it may be more suitable to precisely assess and consider cortical components with another device or with another clinical scale, such as the Wisconsin score system.

Conclusions

The OQAS improves preoperative assessment of age-related cataract patients. We confirm that OSI, measured in optimal conditions (good compliance, no retinal or ocular surface or tear film diseases), is correlated with clinical parameters, but MTF cutoff, incorporating scattering and aberrations, seems the better objective parameter to assess nuclear cataracts. Only OSI may be used to assess objectively the effect of cortical components on optical quality. OSI may also predict the phacodynamics (US energy and time) needed, but only for pure nuclear cataracts, as OSI increased with cortical components that do not require more CDE or US time. The OQAS provides objective and automatic measurements that can be used to personalize cataract surgery parameters.

Data Availability

All relevant data are within the paper.

Funding Statement

The authors received no specific funding for this work.

References

  • 1.Flaxman SR, Bourne RRA, Resnikoff S, Ackland P, Braithwaite T, Cicinelli MV, et al. Global causes of blindness and distance vision impairment 1990–2020: a systematic review and meta-analysis. Lancet Glob Health. 2017;5(12):e1221–e34. Epub 2017/10/17. 10.1016/S2214-109X(17)30393-5 . [DOI] [PubMed] [Google Scholar]
  • 2.Prokofyeva E, Wegener A, Zrenner E. Cataract prevalence and prevention in Europe: a literature review. Acta ophthalmologica. 2013;91(5):395–405. Epub 2012/06/22. 10.1111/j.1755-3768.2012.02444.x . [DOI] [PubMed] [Google Scholar]
  • 3.Kauh CY, Blachley TS, Lichter PR, Lee PP, Stein JD. Geographic Variation in the Rate and Timing of Cataract Surgery Among US Communities. JAMA ophthalmology. 2016;134(3):267–76. Epub 2016/01/01. 10.1001/jamaophthalmol.2015.5322 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.Daien V, Le Pape A, Heve D, Carriere I, Villain M. Incidence and Characteristics of Cataract Surgery in France from 2009 to 2012: A National Population Study. Ophthalmology. 2015;122(8):1633–8. Epub 2015/06/09. 10.1016/j.ophtha.2015.04.017 . [DOI] [PubMed] [Google Scholar]
  • 5.O'Day DM. Management of cataract in adults. Agency for Health Care Policy and Research. American family physician. 1993;47(6):1421–30. . [PubMed] [Google Scholar]
  • 6.O'Day DM. Management of cataract in adults. Quick reference guide for clinicians. The Cataract Management Guideline Panel of the Agency for Health Care Policy and Research. Arch Ophthalmol. 1993;111(4):453–9. 10.1001/archopht.1993.01090040045026 . [DOI] [PubMed] [Google Scholar]
  • 7.O'Day DM. A new guideline for patients with cataract. Arch Ophthalmol. 1993;111(3):317–8. 10.1001/archopht.1993.01090030035030 . [DOI] [PubMed] [Google Scholar]
  • 8.O'Day DM, Steinberg EP, Dickersin K. Systematic literature review for clinical practice guideline development. Transactions of the American Ophthalmological Society. 1993;91:421–36; discussion 37–8. Epub 1993/01/01. . [PMC free article] [PubMed] [Google Scholar]
  • 9.Olson RJ, Braga-Mele R, Chen SH, Miller KM, Pineda R 2nd, Tweeten JP, et al. Cataract in the Adult Eye Preferred Practice Pattern(R). Ophthalmology. 2017;124(2):P1–P119. Epub 2016/10/18. 10.1016/j.ophtha.2016.09.027 . [DOI] [PubMed] [Google Scholar]
  • 10.Chylack LT Jr., Wolfe JK, Singer DM, Leske MC, Bullimore MA, Bailey IL, et al. The Lens Opacities Classification System III. The Longitudinal Study of Cataract Study Group. Arch Ophthalmol. 1993;111(6):831–6. 10.1001/archopht.1993.01090060119035 . [DOI] [PubMed] [Google Scholar]
  • 11.Karbassi M, Khu PM, Singer DM, Chylack LT Jr. Evaluation of lens opacities classification system III applied at the slitlamp. Optometry and vision science: official publication of the American Academy of Optometry. 1993;70(11):923–8. Epub 1993/11/01. 10.1097/00006324-199311000-00009 . [DOI] [PubMed] [Google Scholar]
  • 12.Tan AC, Loon SC, Choi H, Thean L. Lens Opacities Classification System III: cataract grading variability between junior and senior staff at a Singapore hospital. Journal of cataract and refractive surgery. 2008;34(11):1948–52. Epub 2008/11/14. 10.1016/j.jcrs.2008.06.037 . [DOI] [PubMed] [Google Scholar]
  • 13.Wong WL, Li X, Li J, Cheng CY, Lamoureux EL, Wang JJ, et al. Cataract conversion assessment using lens opacity classification system III and Wisconsin cataract grading system. Invest Ophthalmol Vis Sci. 2013;54(1):280–7. Epub 2012/12/13. 10.1167/iovs.12-10657 . [DOI] [PubMed] [Google Scholar]
  • 14.Pei X, Bao Y, Chen Y, Li X. Correlation of lens density measured using the Pentacam Scheimpflug system with the Lens Opacities Classification System III grading score and visual acuity in age-related nuclear cataract. Br J Ophthalmol. 2008;92(11):1471–5. Epub 2008/07/01. 10.1136/bjo.2007.136978 . [DOI] [PubMed] [Google Scholar]
  • 15.Grewal DS, Brar GS, Grewal SP. Correlation of nuclear cataract lens density using Scheimpflug images with Lens Opacities Classification System III and visual function. Ophthalmology. 2009;116(8):1436–43. 10.1016/j.ophtha.2009.03.002 . [DOI] [PubMed] [Google Scholar]
  • 16.Gupta M, Ram J, Jain A, Sukhija J, Chaudhary M. Correlation of nuclear density using the Lens Opacity Classification System III versus Scheimpflug imaging with phacoemulsification parameters. Journal of cataract and refractive surgery. 2013;39(12):1818–23. Epub 2013/11/30. 10.1016/j.jcrs.2013.05.052 . [DOI] [PubMed] [Google Scholar]
  • 17.Lim DH, Kim TH, Chung ES, Chung TY. Measurement of lens density using Scheimpflug imaging system as a screening test in the field of health examination for age-related cataract. Br J Ophthalmol. 2015;99(2):184–91. Epub 2014/09/11. 10.1136/bjophthalmol-2014-305632 . [DOI] [PubMed] [Google Scholar]
  • 18.Dominguez-Vicent A, Birkeldh U, Laurell CG, Nilson M, Brautaset R. Objective Assessment of Nuclear and Cortical Cataracts through Scheimpflug Images: Agreement with the LOCS III Scale. PloS one. 2016;11(2):e0149249 Epub 2016/02/20. 10.1371/journal.pone.0149249 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19.Faria-Correia F, Ramos I, Lopes B, Monteiro T, Franqueira N, Ambrosio R Jr. Correlations of Objective Metrics for Quantifying Dysfunctional Lens Syndrome With Visual Acuity and Phacodynamics. Journal of refractive surgery. 2017;33(2):79–83. Epub 2017/02/14. 10.3928/1081597X-20161206-05 . [DOI] [PubMed] [Google Scholar]
  • 20.Kim JS, Chung SH, Joo CK. Clinical application of a Scheimpflug system for lens density measurements in phacoemulsification. Journal of cataract and refractive surgery. 2009;35(7):1204–9. 10.1016/j.jcrs.2009.02.032 . [DOI] [PubMed] [Google Scholar]
  • 21.Lim SA, Hwang J, Hwang KY, Chung SH. Objective assessment of nuclear cataract: comparison of double-pass and Scheimpflug systems. Journal of cataract and refractive surgery. 2014;40(5):716–21. 10.1016/j.jcrs.2013.10.032 . [DOI] [PubMed] [Google Scholar]
  • 22.Faria-Correia F, Lopes BT, Ramos IC, Monteiro T, Franqueira N, Ambrosio R Jr. Application of different Scheimpflug-based lens densitometry methods in phacodynamics prediction. Clinical ophthalmology. 2016;10:609–15. Epub 2016/04/23. 10.2147/OPTH.S95993 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Makhotkina NY, Berendschot T, van den Biggelaar F, Weik ARH, Nuijts R. Comparability of subjective and objective measurements of nuclear density in cataract patients. Acta ophthalmologica. 2018;96(4):356–63. Epub 2018/03/27. 10.1111/aos.13694 . [DOI] [PubMed] [Google Scholar]
  • 24.Wong AL, Leung CK, Weinreb RN, Cheng AK, Cheung CY, Lam PT, et al. Quantitative assessment of lens opacities with anterior segment optical coherence tomography. Br J Ophthalmol. 2009;93(1):61–5. Epub 2008/10/08. 10.1136/bjo.2008.137653 . [DOI] [PubMed] [Google Scholar]
  • 25.Kuroda T, Fujikado T, Maeda N, Oshika T, Hirohara Y, Mihashi T. Wavefront analysis of higher-order aberrations in patients with cataract. Journal of cataract and refractive surgery. 2002;28(3):438–44. Epub 2002/04/26. 10.1016/s0886-3350(01)01176-2 . [DOI] [PubMed] [Google Scholar]
  • 26.Applegate RA, Marsack JD, Ramos R, Sarver EJ. Interaction between aberrations to improve or reduce visual performance. Journal of cataract and refractive surgery. 2003;29(8):1487–95. Epub 2003/09/05. 10.1016/s0886-3350(03)00334-1 . [DOI] [PubMed] [Google Scholar]
  • 27.Donnelly WJ 3rd, Pesudovs K, Marsack JD, Sarver EJ, Applegate RA. Quantifying scatter in Shack-Hartmann images to evaluate nuclear cataract. Journal of refractive surgery. 2004;20(5):S515–22. Epub 2004/11/05. . [PubMed] [Google Scholar]
  • 28.Packer M, Fine IH, Hoffman RS. Wavefront technology in cataract surgery. Current opinion in ophthalmology. 2004;15(1):56–60. Epub 2004/01/27. 10.1097/00055735-200402000-00011 . [DOI] [PubMed] [Google Scholar]
  • 29.Rocha KM, Nose W, Bottos K, Bottos J, Morimoto L, Soriano E. Higher-order aberrations of age-related cataract. Journal of cataract and refractive surgery. 2007;33(8):1442–6. Epub 2007/07/31. 10.1016/j.jcrs.2007.03.059 . [DOI] [PubMed] [Google Scholar]
  • 30.Lee J, Kim MJ, Tchah H. Higher-order aberrations induced by nuclear cataract. Journal of cataract and refractive surgery. 2008;34(12):2104–9. Epub 2008/11/26. 10.1016/j.jcrs.2008.08.029 . [DOI] [PubMed] [Google Scholar]
  • 31.Faria-Correia F, Ramos I, Lopes B, Monteiro T, Franqueira N, Ambrosio R Jr. Comparison of Dysfunctional Lens Index and Scheimpflug Lens Densitometry in the Evaluation of Age-Related Nuclear Cataracts. Journal of refractive surgery. 2016;32(4):244–8. Epub 2016/04/14. 10.3928/1081597X-20160209-01 . [DOI] [PubMed] [Google Scholar]
  • 32.Li Z, Yu L, Chen D, Chang P, Wang D, Zhao Y, et al. Dysfunctional Lens Index Serves as a Novel Surgery Decision-Maker for Age-Related Nuclear Cataracts. Current eye research. 2019;40(7):733–8. Epub 2019/03/02. 10.1080/02713683.2019.1584676 . [DOI] [PubMed] [Google Scholar]
  • 33.Artal P, editor Retinal Image Quality (aberration and scatter) from the double-pass PSF. 7th International Wave-Front Congress, Bahamas; 2006.
  • 34.Diaz-Douton F, Benito A, Pujol J, Arjona M, Guell J, Artal P. Comparison of the retinal image quality with a Hartmann-Shack wavefront sensor and a double-pass instrument. Invest Ophthalmol Vis Sci. 2006;47:1710–6. 10.1167/iovs.05-1049 [DOI] [PubMed] [Google Scholar]
  • 35.Artal P, Ferro M, Miranda I, Navarro R. Effects of aging in retinal image quality. J Opt Soc Am A. 1993;10(7):1656–62. Epub 1993/07/01. 10.1364/josaa.10.001656 . [DOI] [PubMed] [Google Scholar]
  • 36.Elliott DB. Evaluating visual function in cataract. Optometry and vision science: official publication of the American Academy of Optometry. 1993;70(11):896–902. Epub 1993/11/01. 10.1097/00006324-199311000-00006 . [DOI] [PubMed] [Google Scholar]
  • 37.Guirao A Ga C, Redondo M, Geraghty E, Norrby S, Artal P. Average optical performance of the human eye as a function of age in a normal population. Invest Ophthalmol Vis Sci. 1999;40:203–13. [PubMed] [Google Scholar]
  • 38.Guell JL, Pujol J, Arjona M, Diaz-Douton F, Artal P. Optical Quality Analysis System; Instrument for objective clinical evaluation of ocular optical quality. Journal of cataract and refractive surgery. 2004;30(7):1598–9. Epub 2004/06/24. 10.1016/j.jcrs.2004.04.031 . [DOI] [PubMed] [Google Scholar]
  • 39.Tian M, Miao H, Shen Y, Gao J, Mo X, Zhou X. Intra- and Intersession Repeatability of an Optical Quality and Intraocular Scattering Measurement System in Children. PloS one. 2015;10(11):e0142189 Epub 2015/11/05. 10.1371/journal.pone.0142189 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Artal P, Benito A, Perez GM, Alcon E, De Casas A, Pujol J, et al. An objective scatter index based on double-pass retinal images of a point source to classify cataracts. PloS one. 2011;6(2):e16823 10.1371/journal.pone.0016823 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41.Nam J, Thibos LN, Bradley A, Himebaugh N, Liu H. Forward light scatter analysis of the eye in a spatially-resolved double-pass optical system. Optics express. 2011;19(8):7417–38. Epub 2011/04/20. 10.1364/OE.19.007417 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.Alio JL, Pinero DP, Ortiz D, Montalban R. Clinical outcomes and postoperative intraocular optical quality with a microincision aberration-free aspheric intraocular lens. Journal of cataract and refractive surgery. 2009;35(9):1548–54. Epub 2009/08/18. 10.1016/j.jcrs.2009.03.055 . [DOI] [PubMed] [Google Scholar]
  • 43.Martinez-Roda JA, Vilaseca M, Ondategui JC, Giner A, Burgos FJ, Cardona G, et al. Optical quality and intraocular scattering in a healthy young population. Clinical & experimental optometry. 2011;94(2):223–9. Epub 2010/11/19. 10.1111/j.1444-0938.2010.00535.x . [DOI] [PubMed] [Google Scholar]
  • 44.Hu AL, Qiao LY, Zhang Y, Cai XG, Li L, Wan XH. Reproducibility of optical quality parameters measured at objective and subjective best focuses in a double-pass system. Int J Ophthalmol. 2015;8(5):1043–50. Epub 2015/11/12. 10.3980/j.issn.2222-3959.2015.05.34 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Saad A, Saab M, Gatinel D. Repeatability of measurements with a double-pass system. Journal of cataract and refractive surgery. 2010;36(1):28–33. 10.1016/j.jcrs.2009.07.033 . [DOI] [PubMed] [Google Scholar]
  • 46.Benito A, Perez GM, Mirabet S, Vilaseca M, Pujol J, Marin JM, et al. Objective optical assessment of tear-film quality dynamics in normal and mildly symptomatic dry eyes. Journal of cataract and refractive surgery. 2011;37(8):1481–7. Epub 2011/07/26. 10.1016/j.jcrs.2011.03.036 . [DOI] [PubMed] [Google Scholar]
  • 47.Tan CH, Labbe A, Liang Q, Qiao L, Baudouin C, Wan X, et al. Dynamic change of optical quality in patients with dry eye disease. Invest Ophthalmol Vis Sci. 2015;56(5):2848–54. Epub 2015/02/05. 10.1167/iovs.14-15757 . [DOI] [PubMed] [Google Scholar]
  • 48.Herbaut A, Liang H, Rabut G, Trinh L, Kessal K, Baudouin C, et al. Impact of Dry Eye Disease on Vision Quality: An Optical Quality Analysis System Study. Translational vision science & technology. 2018;7(4):5 Epub 2018/07/17. 10.1167/tvst.7.4.5 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 49.Cabot F, Saad A, McAlinden C, Haddad NM, Grise-Dulac A, Gatinel D. Objective assessment of crystalline lens opacity level by measuring ocular light scattering with a double-pass system. American journal of ophthalmology. 2013;155(4):629–35, 35.e1–2. 10.1016/j.ajo.2012.11.005 . [DOI] [PubMed] [Google Scholar]
  • 50.Pan AP, Wang QM, Huang F, Huang JH, Bao FJ, Yu AY. Correlation among lens opacities classification system III grading, visual function index-14, pentacam nucleus staging, and objective scatter index for cataract assessment. American journal of ophthalmology. 2015;159(2):241–7.e2. Epub 2014/12/03. 10.1016/j.ajo.2014.10.025 . [DOI] [PubMed] [Google Scholar]
  • 51.Habay T, Majzoub S, Perrault O, Rousseau C, Pisella PJ. [Objective assessment of the functional impact of dry eye severity on the quality of vision by double-pass aberrometry]. J Fr Ophtalmol. 2014;37(3):188–94. Epub 2014/02/25. 10.1016/j.jfo.2013.07.010 . [DOI] [PubMed] [Google Scholar]
  • 52.Herbaut A, Liang H, Denoyer A, Baudouin C, Labbe A. [Tear film analysis and evaluation of optical quality: A review of the literature (French translation of the article)]. J Fr Ophtalmol. 2019;42(3):226–43. Epub 2019/03/19. 10.1016/j.jfo.2018.10.004 . [DOI] [PubMed] [Google Scholar]
  • 53.Hwang JS, Lee YP, Bae SH, Kim HK, Yi K, Shin YJ. Utility of the optical quality analysis system for decision-making in cataract surgery. BMC ophthalmology. 2018;18(1):231 Epub 2018/09/05. 10.1186/s12886-018-0904-1 . [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 54.Rzemyk V, Cochener B. [Quality of vision studied by comparative measurement of light scattering]. J Fr Ophtalmol. 2014;37(7):540–7. 10.1016/j.jfo.2014.01.006 . [DOI] [PubMed] [Google Scholar]
  • 55.Galliot F, Patel SR, Cochener B. Objective Scatter Index: Working Toward a New Quantification of Cataract? Journal of refractive surgery. 2016;32(2):96–102. 10.3928/1081597X-20151222-02 . [DOI] [PubMed] [Google Scholar]
  • 56.Cochener B, Patel SR, Galliot F. Correlational Analysis of Objective and Subjective Measures of Cataract Quantification. Journal of refractive surgery. 2016;32(2):104–9. 10.3928/1081597X-20151222-01 . [DOI] [PubMed] [Google Scholar]
  • 57.Pesudovs K, Coster DJ. Cataract surgery reduces subjective visual disability. Australian and New Zealand journal of ophthalmology. 1997;25 Suppl 1:S3–5. Epub 1997/05/01. 10.1111/j.1442-9071.1997.tb01742.x . [DOI] [PubMed] [Google Scholar]
  • 58.Skiadaresi E, McAlinden C, Pesudovs K, Polizzi S, Khadka J, Ravalico G. Subjective quality of vision before and after cataract surgery. Arch Ophthalmol. 2012;130(11):1377–82. Epub 2012/11/13. 10.1001/archophthalmol.2012.1603 . [DOI] [PubMed] [Google Scholar]
  • 59.Vilaseca M, Romero MJ, Arjona M, Luque SO, Ondategui JC, Salvador A, et al. Grading nuclear, cortical and posterior subcapsular cataracts using an objective scatter index measured with a double-pass system. Br J Ophthalmol. 2012;96(9):1204–10. 10.1136/bjophthalmol-2011-301055 . [DOI] [PubMed] [Google Scholar]
  • 60.Davison JA, Chylack LT. Clinical application of the lens opacities classification system III in the performance of phacoemulsification. Journal of cataract and refractive surgery. 2003;29(1):138–45. Epub 2003/01/29. 10.1016/s0886-3350(02)01839-4 . [DOI] [PubMed] [Google Scholar]
  • 61.Bencic G, Zoric-Geber M, Saric D, Corak M, Mandic Z. Clinical importance of the lens opacities classification system III (LOCS III) in phacoemulsification. Collegium antropologicum. 2005;29 Suppl 1:91–4. Epub 2005/10/01. . [PubMed] [Google Scholar]
  • 62.Brown NA. The morphology of cataract and visual performance. Eye. 1993;7 (Pt 1):63–7. Epub 1993/01/01. 10.1038/eye.1993.14 . [DOI] [PubMed] [Google Scholar]
  • 63.Ortiz D, Alio JL, Ruiz-Colecha J, Oser U. Grading nuclear cataract opacity by densitometry and objective optical analysis. Journal of cataract and refractive surgery. 2008;34(8):1345–52. Epub 2008/07/29. 10.1016/j.jcrs.2008.04.022 . [DOI] [PubMed] [Google Scholar]
  • 64.Chen M, Anderson E, Hill G, Chen JJ, Patrianakos T. Comparison of cumulative dissipated energy between the Infiniti and Centurion phacoemulsification systems. Clinical ophthalmology. 2015;9:1367–72. Epub 2015/08/01. 10.2147/OPTH.S88225 . [DOI] [PMC free article] [PubMed] [Google Scholar]

Decision Letter 0

Ireneusz Grulkowski

7 May 2020

PONE-D-20-08844

Using Optical quality analysis system for cataract preoperative assessment & phacodynamics prediction in age-related cataract patients

PLOS ONE

Dear Dr. Garcin,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

We would appreciate receiving your revised manuscript by Jun 21 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

Ireneusz Grulkowski, PhD

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following financial disclosure:

"NO

The funders had no role in study design, data collection and analysis, decision to

publish, or preparation of the manuscript."

At this time, please address the following queries:

a) Please clarify the sources of funding (financial or material support) for your study. List the grants or organizations that supported your study, including funding received from your institution.

b) State what role the funders took in the study. If the funders had no role in your study, please state: “The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.”

c) If any authors received a salary from any of your funders, please state which authors and which funders.

d) If you did not receive any funding for this study, please state: “The authors received no specific funding for this work.”

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

Reviewer #2: Yes

Reviewer #3: Partly

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: No

Reviewer #2: Yes

Reviewer #3: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: No

Reviewer #2: Yes

Reviewer #3: No

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: The study evaluated the association between OSI, LOCS III and phacodynamics, and there are some questions need to be considered.

1. The title used the Optical quality analysis system, but the author seems to focus on the OSI. The criterion was only related to OSI, so I am not clear why the relationships between MTF cutoff, LOCS III and phacodynamics were studied here. The abstract also did not mention the results of MTF cutoff.

2. The English used in the paper need to be improved. Occasionally, it is difficult to understand some statements. A professional editing service is urgently needed.

3. The Introduction part is relatively long, and the author should simplify the text.

4. The cataract grading score (LOCS III) used in the study were grade variables, which should not use mean±SD, and the correlation analysis should be changed to spearman correlation. In addition, the author mentioned the regression analysis in the Figure lengends, but not occurred in the main manuscript. And the independent variables and dependent variables should be defined in the whole study.

5. The author analyzed the correlation between OSI and LOCS III, LOCS III and phacodynamics, then analyze the association between OSI and phacodynamics. The Partial correlation need to be used and the LOCS III should be considered in the association between OSI and phacodynamics.

6. The “95%CI” in Figure legends should be cleared.

Reviewer #2: Dear Thibaud Garcin,

First of all, I would like to thank for submitting in PLOS ONE your work “Using Optical quality analysis system for cataract preoperative assessment & phacodynamics prediction in age-related cataract patients”. I have to admit that this work might be very useful for ophthalmologists and consequently, it might improve the life quality of our society. However, there are some doubts, that I would like to ask you:

Material and Methods

a) Involved eligible patients.

• In relation to the topic “involved eligible patients”, it is indicated that all patients were under topical anaesthesia. Is it possible to indicate from which brand was used this anaesthesia?

• In this section, it might appropriate to include information about the subjects who participated in this clinical trial (initial number of patients and eyes, age and gender, refractive error, etc.) instead of the results section.

b) Preoperative subjective and objective assessment.

• In relation to the gold-standard LOCSIII test, I would like to know the reason why the nuclear opacity (NO) and nuclear colour (NC) are scaled from 1 to 6 meanwhile the sub-capsular (SCP) and cortical cataract (C) are scaled from 1 to 5. Moreover, it might be very useful to indicate it in the manuscript.

• OQAS measurements were done in both eyes where some subjects showed that one eye was undergone to phacoemulsification. This data allowed to check the reliable and consistent of the OQAS. It might be very interested to include these to show the advantage of the OQAS.

c) Statistical analysis

• Regarding to the statistical analysis section, it is not indicated which type of test is used to compare the proposal parameters. I would recommend to indicate that all parameters were compared by linear correlations in order to avoid misunderstanding with another tests like T – student test or ANOVA.

• In the line 210, it is indicated “mean + SD”. Is it possible to indicate in the manuscript what it means SD.? I think, it is Standard Deviation.

Results

a) Baseline population data

• LOCSIII test was done by two observers and no differences were appreciated between them. I would like to ask you how it was analysed the reproducibility of the LOCSIII between these 2 observers to obtain a p-value = 0.51.

• In table 1, is it possible to indicate which parameters are following a normal and not normal distribution after doing the Sapphire – Wilk test. In addition, in the title box Mean +- SD, it should be included median +- percentiles.

b) Correlations

• In relation to figure 2, I am quite surprised about the dispersion when it is correlated the BCVA with the OSI and with LOCSIII test. I am wondering if this dispersion might be associated to the low influence of the visual acuity in the decision to cataract surgery as it was indicated in the introduction. It would be fantastic to discuss this point in the discussion section.

• Another interesting point that I would like to ask you in relation to the figure 3 is about the slopes of each graphic. It seems that the correlation LOCSIII C vs OSI shows the higher slope (with r=0.715, which it is the highest), but as well the most dispersive graphic (p-value = 0.02 although <0.05). How clinically might it be interpreted? It is a pity that the sample population is too low (n=10) to analyse better this behaviour. Moreover, according to these graphics, it is observed that the scattering is higher in nuclear cataracts, but the strong OSI change is caused in the cortical cataracts? Do you know why?

• Regarding to figure 4, how is it possible that the CDE hasn´t got any influence in the cortical cataract? According to your correlations, it is not significance (p-value = 0.058). I think, that it is caused by the sample population n=10. What do you think?

• According to figure 6, I think it does not have any sense to include it because the MTF and OSI are specific parameters from OQAS which are strongly related between them by the PSF. Thus, the OQAS must give strong correlation to have a high reproducibility. Consequently, this correlation does not give any significant information for a clinical purpose. I would recommend to remove the data and graphics between MTF and OSI and prepare another figure with correlations between MTF with BCVA, MTF with LOCSIII, MTF with CDE and MTF with US.

Discussion

• In the discussion about the BCVA vs OQAS, it is explained that the differences between your research with other ones might be caused by the sample size or the visual acuity decision-making surgery after consulting people. However, there is not any references about the type of visual test was used and how it might impact in the correlated results.

Reviewer #3: COMMENT:

This article discusses the usefulness of the OSI parameter in the clinical management of age-related cataracts. Furthermore, the relationship of preoperative parameters with photodynamic parameters involved in surgery is interesting and original. However, the sample of patients included in the study is very small. In addition, as several works recommend, the statistical analysis must be corrected to include 1 eye per patient. In fact, only one case in this work does not meet this criterion and the results of the study are unlikely to be different if the correction is made. Likewise, the discussion should be improved for a better understanding. For example, when the author removes cases with a cortical component, he consequently removes cases with OSI> 9. This should be added in the discussion. The influence of aberrations on the overestimation of OSI should be discussed as well.

As far as language and style are concerned, the use of punctuation marks does need some revising as on certain occasions it interferes with clarity. See, for example, lines 341 – 345 and 402 - 403. Similarly, the paper shows some segments lacking in clear sentence structure that should be rewritten (lines 21 – 23; 159 -163; 207 – 208; 222 – 224). Other aspects that should be revised include unnecessary use of passive forms - particularly of the verb “correlate” -, the use of contractions, vocabulary (“similar as” in lines 364 – 385 should be “similar to”), missing words (“decision-making surgery”) and verb tense consistency (lines 160 – 161).

SPECIFIC COMMENTS:

Page 8. Lines 159-163: this sentence should be rewritten for a better understanding.

Page 9. Lines 180-183: the definition of SR does not exactly correspond to the one used by the double-pass system, which computes the SR in two dimensions as the ratio between the areas under the MTF curve of the measured eye and that of the aberration-free eye. Ref.: DOI:10.1111/j.1444-0938.2010.00535.x

Page 9. Lines 194-195: missing parenthesis open and close.

Page 11. Line 225: in the statistical analysis it is recommended to include an eye per patient in all cases due to strong correlation between the two eyes of a subject.

Page 14. Lines 298-303: correlations with SR should be included in the results since this parameter is the one most related to loss of image quality and very sensitive to aberrations.

Page 14. Line 305: correlation between the MTF and OSI is evident, especially considering that they have been obtained with the same instrument. Therefore, Figure 6 is unnecessary.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files to be viewed.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email us at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Oct 12;15(10):e0240350. doi: 10.1371/journal.pone.0240350.r002

Author response to Decision Letter 0


8 Jul 2020

PONE-D-20-08844

Using Optical quality analysis system for cataract preoperative assessment & phacodynamics prediction in age-related cataract patients

PLOS ONE

Saint-Etienne, July 4th 2020

Dear Ireneusz Grulkowski,

Please find attached a revised version of our article titled “Using Optical quality analysis system for cataract preoperative assessment & phacodynamics prediction in age-related cataract patients” with a new title "Using Optical Quality Analysis System (OQAS) for predicting surgical parameters in age-related cataract patients", to make it more explicit for a wider audience.

We updated all the points concerning Journal Requirements. The authors received no specific funding for this work.

We have carefully considered and responded to all the points addressed by the reviewers. Per your instructions, all substantive amendments in the revised version are stated in our point-by-point response, and are marked in red in the article.

We greatly hope that this new version will meet the reviewers' expectations and comply with your editorial policy.

Yours sincerely,

Dr. Thibaud GARCIN, M.D., Ph.D., FEBO

t.garcin@univ-st-etienne.fr

Dear Dr. Garcin,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

We would appreciate receiving your revised manuscript by Jun 21 2020 11:59PM. When you are ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter.

To enhance the reproducibility of your results, we recommend that if applicable you deposit your laboratory protocols in protocols.io, where a protocol can be assigned its own identifier (DOI) such that it can be cited independently in the future.

Please include the following items when submitting your revised manuscript:

• A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). This letter should be uploaded as separate file and labeled 'Response to Reviewers'.

• A marked-up copy of your manuscript that highlights changes made to the original version. This file should be uploaded as separate file and labeled 'Revised Manuscript with Track Changes'.

• An unmarked version of your revised paper without tracked changes. This file should be uploaded as separate file and labeled 'Manuscript'.

Please note while forming your response, if your article is accepted, you may have the opportunity to make the peer review history publicly available. The record will include editor decision letters (with reviews) and your responses to reviewer comments. If eligible, we will contact you to opt in or out.

We look forward to receiving your revised manuscript.

Kind regards,

Ireneusz Grulkowski, PhD

Academic Editor

PLOS ONE

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming.

Updates were done.

2. Thank you for stating the following financial disclosure:

"NO

The funders had no role in study design, data collection and analysis, decision to

publish, or preparation of the manuscript."

At this time, please address the following queries:

a) Please clarify the sources of funding (financial or material support) for your study. List the grants or organizations that supported your study, including funding received from your institution.

No funding was received for this work.

b) State what role the funders took in the study. If the funders had no role in your study, please state: “The funders had no role in study design, data collection and analysis, decision to publish, or preparation of the manuscript.”

c) If any authors received a salary from any of your funders, please state which authors and which funders.

d) If you did not receive any funding for this study, please state: “The authors received no specific funding for this work.”

Please include your amended statements within your cover letter; we will change the online submission form on your behalf.

“The authors received no specific funding for this work.”

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Partly

Reviewer #2: Yes

Reviewer #3: Partly

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: No

Reviewer #2: Yes

Reviewer #3: Yes

3. Have the authors made all data underlying the findings in their manuscript fully available?
 The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: No

Reviewer #2: Yes

Reviewer #3: No

We have considered the corrections advised by Reviewers 1 & 3, and the manuscript has now been reviewed by a native English speaker.

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

The questions and comments of the 3 reviewers helped us to rewrite a significant part of the article in particular by simplifying the abstract, shortening the introduction and completing the results for the 3 OQAS parameters.

In addition, we would also like to slightly modify the title to make it more explicit for a wider audience.

The new title we propose is now : Using Optical Quality Analysis System (OQAS) for predicting surgical parameters in age-related cataract patients.

Reviewer #1: The study evaluated the association between OSI, LOCS III and phacodynamics, and there are some questions need to be considered.

1. The title used the Optical quality analysis system, but the author seems to focus on the OSI. The criterion was only related to OSI, so I am not clear why the relationships between MTF cutoff, LOCS III and phacodynamics were studied here. The abstract also did not mention the results of MTF cutoff.

Initially we had chosen to present only OSI to simplify but your remark is correct. We have therefore decided to present all the results of the OQAS. The text and the abstract have been modified accordingly.

2. The English used in the paper need to be improved. Occasionally, it is difficult to understand some statements. A professional editing service is urgently needed.

The manuscript has now been reviewed by a native English speaker.

3. The Introduction part is relatively long, and the author should simplify the text.

We have shortened the introduction considerably.

4. The cataract grading score (LOCS III) used in the study were grade variables, which should not use mean±SD, and the correlation analysis should be changed to spearman correlation.

The LOCSIII score is perfectly comparable to a continuous variable as mentioned in princeps publication1, and used by numerous authors in different published studies in the field2-7.

Each of the 4 items (Nuclear opacity/Nuclar Color/Cortical cataract/Posterior cataract) that compose it is defined to the decimal from 0.1 to 5.9 5C&P) or 6.9 (NC&NO). The overall score thus varies from 0.1 to 0.1 between 0.4 and 25.6. Consequently, mean±SD can be used for LOCSIII grading. We detailed the LOCSIII grading system in material and methods section for clarification.

References:

1. Chylack LT, Jr., Wolfe JK, Singer DM, Leske MC, Bullimore MA, Bailey IL, et al. The Lens Opacities Classification System III. The Longitudinal Study of Cataract Study Group. Arch Ophthalmol. 1993;111(6):831-6. PubMed PMID: 8512486.

2. Karbassi M, Khu PM, Singer DM, Chylack LT, Jr. Evaluation of lens opacities classification system III applied at the slitlamp. Optometry and vision science: official publication of the American Academy of Optometry. 1993;70(11):923-8. Epub 1993/11/01. PubMed PMID: 8302528.

3. Wong WL, Li X, Li J, Cheng CY, Lamoureux EL, Wang JJ, et al. Cataract conversion assessment using lens opacity classification system III and Wisconsin cataract grading system. Invest Ophthalmol Vis Sci. 2013;54(1):280-7. Epub 2012/12/13. doi: 10.1167/iovs.12-10657. PubMed PMID: 23233255.

4. Cabot F, Saad A, McAlinden C, Haddad NM, Grise-Dulac A, Gatinel D. Objective assessment of crystalline lens opacity level by measuring ocular light scattering with a double-pass system. American journal of ophthalmology. 2013;155(4):629-35, 35 e1-2. doi: 10.1016/j.ajo.2012.11.005. PubMed PMID: 23317652.

5. Pan AP, Wang QM, Huang F, Huang JH, Bao FJ, Yu AY. Correlation among lens opacities classification system III grading, visual function index-14, pentacam nucleus staging, and objective scatter index for cataract assessment. American journal of ophthalmology. 2015;159(2):241-7 e2. Epub 2014/12/03. doi: 10.1016/j.ajo.2014.10.025. PubMed PMID: 25448993.

6. Wong AL, Leung CK, Weinreb RN, Cheng AK, Cheung CY, Lam PT, et al. Quantitative assessment of lens opacities with anterior segment optical coherence tomography. Br J Ophthalmol. 2009;93(1):61-5. Epub 2008/10/08. doi: 10.1136/bjo.2008.137653. PubMed PMID: 18838411.

7. Galliot F, Patel SR, Cochener B. Objective Scatter Index: Working Toward a New Quantification of Cataract? Journal of refractive surgery. 2016;32(2):96-102. doi: 10.3928/1081597X-20151222-02. PubMed PMID: 26856426.

For correlation, we therefore used Spearman or Pearson according to normality of the data distribution.

In addition, the author mentioned the regression analysis in the Figure legends, but not occurred in the main manuscript.

Thank you for that comment. We have added in M&M that linear regressions have been done.

And the independent variables and dependent variables should be defined in the whole study.

The only independent variable is the grade of the cataract (LOCSIII score). A priori, all the others can be considered as dependent.

Logically, the hypotheses are that the parameters of the OQAS must depend on LOCSIII and that it is the same for the surgical parameters (CDE and US time being very good indicators of the hardness of the crystalline lens). However, the establishment of LOCSIII remains subjective and has never been automated since 1993. What we have therefore sought to determine is whether one or more objective parameters of the OQAS were correlated with the surgical parameters and could thus make it possible to predict the latter.

5. The author analyzed the correlation between OSI and LOCS III, LOCS III and phacodynamics, then analyze the association between OSI and phacodynamics. The Partial correlation need to be used and the LOCS III should be considered in the association between OSI and phacodynamics.

We thank you for this suggestion. Indeed, we have now calculated partial correlation coefficients between OQAS and surgical parameters, taking into account LOCSIII.

For paragraph results

“Partial correlation between CDE and OSI, US time and OSI controlling for the LOCSIII was negligible (r = 0.013 and r = -0.041 respectively) and not significant (p=0.993 and p=0.910 respectively), thus indicating that LOCSIII N score greatly influenced the relation between OSI and surgical parameters.”

For paragraph discussion

“Partial correlation analysis clearly demonstrates that LOCSIII acts as a confounding factor to explain the strong correlation between OSI and phacodynamics. This is explained by the fact that LOCSIII and OQAS both measures the same physical phenomenon: lens opacification. This is a demonstration of OQAS reliability as objective and automatic measurement.”

6. The “95%CI” in Figure legends should be cleared.

We do not understand why you wanted us to remove the confidence intervals. We think it is necessary and honest considering our sample size. We precised in material and methods paragraph that “The Red dots show the 95% confidence interval (95%CI) of the slope and intercept on the linear regression graphs”. So everything is clearer and we don’t note iteratively the same thing for each linear regression graph.

Reviewer #2: Dear Thibaud Garcin,

First of all, I would like to thank for submitting in PLOS ONE your work “Using Optical quality analysis system for cataract preoperative assessment & phacodynamics prediction in age-related cataract patients”. I have to admit that this work might be very useful for ophthalmologists and consequently, it might improve the life quality of our society. However, there are some doubts, that I would like to ask you:

Material and Methods

a) Involved eligible patients.

• In relation to the topic “involved eligible patients”, it is indicated that all patients were under topical anaesthesia. Is it possible to indicate from which brand was used this anaesthesia?

We have now made that clear in the text.

One drop of Oxybuprocaine chlorhydrate 0,4% (Théa, Clermont-Ferrand, France). and one drop of Tetracaine 1% (Théa, Clermont-Ferrand, France) were instilled twice in the conjunctival sac, 5 minutes apart, 10 minutes before draping.

• In this section, it might appropriate to include information about the subjects who participated in this clinical trial (initial number of patients and eyes, age and gender, refractive error, etc.) instead of the results section.

We moved these data from results section to material and methods section, and deleted figure 1 caption as well as figure 1 not so necessary considering already details in text.

b) Preoperative subjective and objective assessment.

• In relation to the gold-standard LOCSIII test, I would like to know the reason why the nuclear opacity (NO) and nuclear colour (NC) are scaled from 1 to 6 meanwhile the sub-capsular (SCP) and cortical cataract (C) are scaled from 1 to 5. Moreover, it might be very useful to indicate it in the manuscript.

The LOCSIII grading system was published by Chylack et al. in 19931 and is conceived like this, and this gold standard did not change. But often, authors referring to this grading tend to adapt their own “scale” by simplifying categories/range/scales2,3.

References:

1. Chylack LT, Jr., Wolfe JK, Singer DM, Leske MC, Bullimore MA, Bailey IL, et al. The Lens Opacities Classification System III. The Longitudinal Study of Cataract Study Group. Arch Ophthalmol. 1993;111(6):831-6. PubMed PMID: 8512486.

2. Karbassi M, Khu PM, Singer DM, Chylack LT, Jr. Evaluation of lens opacities classification system III applied at the slitlamp. Optometry and vision science: official publication of the American Academy of Optometry. 1993;70(11):923-8. Epub 1993/11/01. PubMed PMID: 8302528.

3. Wong WL, Li X, Li J, Cheng CY, Lamoureux EL, Wang JJ, et al. Cataract conversion assessment using lens opacity classification system III and Wisconsin cataract grading system. Invest Ophthalmol Vis Sci. 2013;54(1):280-7. Epub 2012/12/13. doi: 10.1167/iovs.12-10657. PubMed PMID: 23233255.

• OQAS measurements were done in both eyes where some subjects showed that one eye was undergone to phacoemulsification. This data allowed to check the reliable and consistent of the OQAS. It might be very interested to include these to show the advantage of the OQAS.

As the aim was to analyze the relationships between OQAS parameters, surgical parameters and LOCSIII we included in the statistical analysis only the operated eyes.

As you suggest, we have added, for your information, this descriptive paragraph below of the OQAS data of non-operated eyes, only this file and not in the manuscript to make it clearer and simpler.

“We used the data acquired for each fellow eye (phakic or pseudophakic) to check if the data concerning each operated eye was reliable and consistent.

For the fellow eyes, 2 were long-standing visually impaired (2.0 logMAR) with corneal scars, 14 were pseudophakic with median far BCVA 0.10 (0.00 – 0.30) logMAR and mean OSI 1.27 ±0.41 [0.5-2.1], 5 were phakic (3 pure early cortical cataracts rated LOCSIII, 2 pure early nuclear cataracts) with median far BCVA 0.2 (0.10 – 0.30) logMAR and mean OSI 5.54 ±1.83 [3.9-8.4]. The 5 patients with phakic fellow eye were not really bothered compared to their other eye operated and included for analysis.

Regarding the fellow eyes, OQAS provided reliable data for phakic and pseudophakic eyes with a strong correlation between BCVA and OSI: respectively Spearman r=0.797 p<0.001 and Spearman r=0.888 p=0.044.”

c) Statistical analysis

• Regarding to the statistical analysis section, it is not indicated which type of test is used to compare the proposal parameters. I would recommend to indicate that all parameters were compared by linear correlations in order to avoid misunderstanding with another tests like T – student test or ANOVA.

We reviewed the statistical analysis and clarified the methods used.

• In the line 210, it is indicated “mean + SD”. Is it possible to indicate in the manuscript what it means SD.? I think, it is Standard Deviation.

This has been done

Results

a) Baseline population data

• LOCSIII test was done by two observers and no differences were appreciated between them. I would like to ask you how it was analysed the reproducibility of the LOCSIII between these 2 observers to obtain a p-value = 0.51.

Thanks for your note. As LOCSIII is decimal score, the agreement between the 2 ophthalmologists for LOCSIII grading was determined by calculating the mean difference and the correlation coefficient. The mean LOCSIII total score was 4.86 ±2.03 with excellent agreement between observers (mean difference 0.3; r=0.94 p<0.001, 19 cases with perfect agreement, 3 cases with differences of 2 points).

We added the points in methods & results section.

• In table 1, is it possible to indicate which parameters are following a normal and not normal distribution after doing the Sapphire – Wilk test. In addition, in the title box Mean +- SD, it should be included median +- percentiles

We reconsidered the statistical analysis and, given the numbers involved, preferred to display more robust non-parametric tests. Nevertheless, we show in the table the data with a normal distribution (Shapiro-Wilk test and Kolmogorov-Smirnov test). We have performed in parallel the 2 types of tests, parametric and non-parametric. In 95% of cases, both tests gave the same result, which is a good indicator of robustness.

b) Correlations

• In relation to figure 2, I am quite surprised about the dispersion when it is correlated the BCVA with the OSI and with LOCSIII test. I am wondering if this dispersion might be associated to the low influence of the visual acuity in the decision to cataract surgery as it was indicated in the introduction. It would be fantastic to discuss this point in the discussion section.

In our sample of population, LOCSIII C is well represented with different grades, and it may impact on dispersion of data. OSI is more dispersed as well as BCVA when we consider Cortical components (as reported by Vilseca et al. 10.1136/bjophthalmol-2011-301055).

In our department the surgical indications for cataract surgery usually follow the national recommendations (= or >0.3 logMAR).

We added this point to the discussion.

“We found moderate correlation between OSI and BCVA as other teams [51, 56], while other authors found a stronger correlation [21, 52, 59]. Differences can be explained by different visual acuity threshold for decision-making surgery between the different centers, and, as explained above for correlation between LOCSIII and BCVA, by the differences in proportion of nuclear and cortical cataracts [59], cortical components are well represented with different grades in our population.”

• Another interesting point that I would like to ask you in relation to the figure 3 is about the slopes of each graphic. It seems that the correlation LOCSIII C vs OSI shows the higher slope (with r=0.715, which it is the highest), but as well the most dispersive graphic (p-value = 0.02 although <0.05). How clinically might it be interpreted? It is a pity that the sample population is too low (n=10) to analyse better this behaviour.

Despite the small size of our population with cortical cataract (n=10 indeed), we find results consistent with those of Vilaseca et al. : They found that BCVA and OSI are significantly more scattered in groups with cortical and posterior subcapsular cataract. They also showed that the best agreement between OSI and LOCSIII was found for cortical cataracts. (Vilaseca M, Romero MJ, Arjona M, Luque SO, Ondategui JC, Salvador A, et al. Grading nuclear, cortical and posterior subcapsular cataracts using an objective scatter index measured with a double-pass system. (Br J Ophthalmol. 2012;96(9):1204-10. doi: 10.1136/bjophthalmol-2011-301055. PubMed PMID: 22790434.).

Moreover, according to these graphics, it is observed that the scattering is higher in nuclear cataracts, but the strong OSI change is caused in the cortical cataracts? Do you know why?

It is demonstrated that cortical cataract influences OSI more than nuclear cataract as explained above. Dispersion is not greater in the nuclear group. We have redrawn all the diagrams with the same scales to facilitate visual comparisons.

Here one part of the new graph, with same scale on X axis.

Regarding to figure 4, how is it possible that the CDE hasn´t got any influence in the cortical cataract? According to your correlations, it is not significance (p-value = 0.058). I think, that it is caused by the sample population n=10. What do you think?

The cortex always remains soft even when it becomes opaque. Regardless of the level of opacity the cortex can always be extracted without the use of ultrasound, with an irrigation aspiration cannula. It is therefore completely logical not to find any correlation between cortical LOCSIII and CDE or US Time. Therefore, a larger number of patients would probably not change anything. There are 2 other articles that did not find a relationship between LOCSIIIC and CDE or US Time 1,2.

References

1. Davison JA, Chylack LT. Clinical application of the lens opacities classification system III in the performance of phacoemulsification. Journal of cataract and refractive surgery. 2003;29(1):138-45. Epub 2003/01/29. PubMed PMID: 12551681.

2. Bencic G, Zoric-Geber M, Saric D, Corak M, Mandic Z. Clinical importance of the lens opacities classification system III (LOCS III) in phacoemulsification. Collegium antropologicum. 2005;29 Suppl 1:91-4. Epub 2005/10/01. PubMed PMID: 16193685.

• According to figure 6, I think it does not have any sense to include it because the MTF and OSI are specific parameters from OQAS which are strongly related between them by the PSF. Thus, the OQAS must give strong correlation to have a high reproducibility. Consequently, this correlation does not give any significant information for a clinical purpose. I would recommend to remove the data and graphics between MTF and OSI and prepare another figure with correlations between MTF with BCVA, MTF with LOCSIII, MTF with CDE and MTF with US.

We followed your suggestion and deleted fig 6. Instead we added a graph for relationship between MTF and clinical and surgical parameters (and the same for SR as suggested in specifics comments)

Discussion

• In the discussion about the BCVA vs OQAS, it is explained that the differences between your research with other ones might be caused by the sample size or the visual acuity decision-making surgery after consulting people. However, there is not any references about the type of visual test was used and how it might impact in the correlated results.

French ophthalmologists use the decimal Monoyer chart for routine far vision. This was added in the revised text. Visual acuity was converted in LogMar to allow performing statistics.

Reviewer #3: COMMENT:

This article discusses the usefulness of the OSI parameter in the clinical management of age-related cataracts. Furthermore, the relationship of preoperative parameters with photodynamic parameters involved in surgery is interesting and original. However, the sample of patients included in the study is very small.

In addition, as several works recommend, the statistical analysis must be corrected to include 1 eye per patient. In fact, only one case in this work does not meet this criterion and the results of the study are unlikely to be different if the correction is made.

We have updated all statistical analysis on 21 eyes, with one eye per subject. As expected the results and their interpretation did not change.

Likewise, the discussion should be improved for a better understanding.

The discussion was rewritten and simplified

For example, when the author removes cases with a cortical component, he consequently removes cases with OSI> 9. This should be added in the discussion.

We added comment in discussion.

“Therefore, After removing cortical and sub capsular components, for pure nuclear cataracts only, OSI ranged from 1.4 to 8.3 and strong correlations were found between CDE and OSI. The correlation was even stronger than between CDE and LOCSIII N, suggesting that OQAS could better predict phacodynamics than LOCSIII for pure nuclear cataracts. The same findings were found for US time and OSI, or US time and LOCSIII N.”

The influence of aberrations on the overestimation of OSI should be discussed as well.

This time we have integrated the detailed analysis of the SR parameter which is the most sensitive to high-order aberrations.

“We found also strong significant correlations between: 1/ Strehl ratio and OSI (rs = -0.948, p<0.001); 2/ BCVA and Strehl ratio (rs = -0.622, p=0.003); 3/ LOCSIII total and Strehl ratio (rs = -0.676, p<0.001) as well as in subgroups of different LOCSIII components (N, NC, NO) except for LOCSIII C (p=0.056); 4/ Strehl ratio and CDE (Spearman r = -0.664, p=0.031), and also Strehl ratio and US time (Spearman r = -0.653, p=0.032) only for pure nuclear cataracts (Fig.6).”

As far as language and style are concerned, the use of punctuation marks does need some revising as on certain occasions it interferes with clarity. See, for example, lines 341 – 345 and 402 - 403. Similarly, the paper shows some segments lacking in clear sentence structure that should be rewritten (lines 21 – 23; 159 -163; 207 – 208; 222 – 224). Other aspects that should be revised include unnecessary use of passive forms - particularly of the verb “correlate” -, the use of contractions, vocabulary (“similar as” in lines 364 – 385 should be “similar to”), missing words (“decision-making surgery”) and verb tense consistency (lines 160 – 161).

The manuscript has now been reviewed by a native English speaker.

SPECIFIC COMMENTS:

Page 8. Lines 159-163: this sentence should be rewritten for a better understanding.

Sentence was rewritten as suggested by reviewer N°3.

“We used the data acquired for each fellow eye (phakic or pseudophakic) to check if the data concerning each operated eye was reliable and consistent.“

Page 9. Lines 180-183: the definition of SR does not exactly correspond to the one used by the double-pass system, which computes the SR in two dimensions as the ratio between the areas under the MTF curve of the measured eye and that of the aberration-free eye. Ref.: DOI:10.1111/j.1444-0938.2010.00535.x

Thank you for this comment. We updated the definition.

Page 9. Lines 194-195: missing parenthesis open and close.

Missing parenthesis were added.

Page 11. Line 225: in the statistical analysis it is recommended to include an eye per patient in all cases due to strong correlation between the two eyes of a subject.

We updated all statistical analysis on 21 eyes, with one eye per subject.

Page 14. Lines 298-303: correlations with SR should be included in the results since this parameter is the one most related to loss of image quality and very sensitive to aberrations.

We added data concerning SR in results and discussion sections.

“We found also strong significant correlations between: 1/ Strehl ratio and OSI (rs = -0.948, p<0.001); 2/ BCVA and Strehl ratio (rs = -0.622, p=0.003); 3/ LOCSIII total and Strehl ratio (rs = -0.676, p<0.001) as well as in subgroups of different LOCSIII components (N, NC, NO) except for LOCSIII C (p=0.056); 4/ Strehl ratio and CDE (Spearman r = -0.664, p=0.031), and also Strehl ratio and US time (Spearman r = -0.653, p=0.032) only for pure nuclear cataracts (Fig.6).”

Page 14. Line 305: correlation between the MTF and OSI is evident, especially considering that they have been obtained with the same instrument. Therefore, Figure 6 is unnecessary.

We deleted this Figure 6 (of the R0 version) for the R1 version.

Attachment

Submitted filename: Vend_040720TG__R1_Reponse to reviewers.docx

Decision Letter 1

Ireneusz Grulkowski

30 Jul 2020

PONE-D-20-08844R1

Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients

PLOS ONE

Dear Dr. Garcin,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Sep 13 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Ireneusz Grulkowski, PhD

Academic Editor

PLOS ONE

Journal Requirements:

Additional Editor Comments (if provided):

Please, address additional comments of the reviewers.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #2: Dear Thibaud Garcin,

Thank you for submitting in PLOS ONE your research work described in the manuscript “Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients”. The new version of the manuscript has improved considerable in clearly and comprehension terms. However, I would like to comment some small details that I have been able to observe in this review:

• First of all, I would like to suggest a deep revision of the manuscript in order to improve some expressions and correct grammar mistakes. Although, in general terms the text is comprehensible, there are some sections where there are incoherent expressions and not well-written. I do recommend to send the manuscript to an expert or native speaker to improve the writing quality.

Introduction

• I think that the definition of the Strehl ratio (SR) is not correct. The SR does not depend on the Modulate Transfer Function (MTF). The SR is the ration between the peak intensity from the Point Spread Function (PSF) of the aberrated eye and the peak intensity from the PSF of the non-aberrated eye. Please, correct lines 113 to 115.

Material and Methods

a) Patients and ethic statement

• The 2 last paragraph are not well located in this section and it makes confusion to the reader. In lines 132 – 135, it is described that the only inclusion criteria is to obtain good OSI measurements. However, it is not known any exclusion criteria and ophthalmic exam until next section. Similarly, it is observed in the last paragraph (lines 136 – 141), where some subjects were excluded because of poor cooperation and dense cataracts, without knowing any details from phacodynamics mode adjustment. I would recommend to move this section before the section “statistical analysis” to describe previously inclusion and exclusion methods.

Results

a) Correlation Analysis: Main objective, subjective preoperative parameters and phacodynamics.

• In order to avoid misunderstanding, I do recommend strongly to indicate “LogMar scale” in the axis Best-Corrected Visual Acuity (BCVA) in figures 1, 5 and 6. The current values can confuse to the reader with decimal scale and the interpretation of the Spearman coefficient can be opposite.

b) Correlation Analysis: Other OQAS parameters, subjective preoperative parameters and phacodynamics.

• Relations between Optical Scattering Index (OSI) with Modulate Transfer Function (MTF) in lines 273 – 274; and between Strehl ratio (SR) with MTF in lines 281 – 282, should not be included in the manuscript. These correlations were well developed and thoroughly analysed by the company in order to determine the performance of the device. Consequently, these data are not going to give any particular and novel information.

Discussion

• I do not really agree with the idea that OQAS and LOCSIII measures the same physical phenomenon. Specially, because OQAS measures the quality of the whole eye, and a simple uveitis or a corneal scar can affect completely the measurement. I would be more specific in this sentence (lines 390 – 392) explaining that both methods will be measuring the same parameters if there are no more ocular pathologies.

Reviewer #3: Thank you for your answers and comments. You have solved most of my concerns successfully. Regarding language, it has improved significantly and it makes the text much easier to follow now. On the other hand, I personally think that IQR=Q-Q3 provides an easier explanation for the data dispersion and allows the comparison with SD (line 187 and tables).

Some specific comments:

- Line 90: Repetition of a fragment

- Lines 210 – 211: The sentence needs revising for a better understanding.

- Lines 240 – 242: Lack of statistical significance for correlation between LOCSIII C and CDE as well as US is probably due to insufficient number of patients in this subgroup (LOCSIII C). Notice that rs is > 0,580. You should comment it here and explain it in discussion.

- Figures 5 and 6: The top right and bottom left graphs seem to be wrong e.g. n = 21?

- I would be very grateful if you could send me the data rs and p for the different subgroups of LOCSIII components because note (lines 235 – 237) that they are not shown in Figure 1.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Oct 12;15(10):e0240350. doi: 10.1371/journal.pone.0240350.r004

Author response to Decision Letter 1


29 Aug 2020

Saint-Etienne, August 29th 2020

Dear Dr. Grulkowski,

Please find attached a revised version R2 of our article "Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients".

We updated all the points concerning Journal Requirements. The authors received no specific funding for this work.

We have carefully considered and responded to all the points addressed by the reviewers.

We greatly hope that this new version will meet the reviewers' expectations and comply with your editorial policy.

Yours sincerely,

Dr. Thibaud GARCIN, M.D., Ph.D., FEBO

t.garcin@univ-st-etienne.fr

PONE-D-20-08844R1

Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients

PLOS ONE

Dear Dr. Garcin,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

Please submit your revised manuscript by Sep 13 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Ireneusz Grulkowski, PhD

Academic Editor

PLOS ONE

Journal Requirements:

Additional Editor Comments (if provided):

Please, address additional comments of the reviewers.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Partly

Reviewer #3: Yes

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: No

Reviewer #3: Yes

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: (No Response)

Reviewer #2: Dear Thibaud Garcin,

Thank you for submitting in PLOS ONE your research work described in the manuscript “Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients”. The new version of the manuscript has improved considerable in clearly and comprehension terms. However, I would like to comment some small details that I have been able to observe in this review:

• First of all, I would like to suggest a deep revision of the manuscript in order to improve some expressions and correct grammar mistakes. Although, in general terms the text is comprehensible, there are some sections where there are incoherent expressions and not well-written. I do recommend to send the manuscript to an expert or native speaker to improve the writing quality. The manuscript has now been reviewed again, this time by a native English speaker.

Introduction

• I think that the definition of the Strehl ratio (SR) is not correct. The SR does not depend on the Modulate Transfer Function (MTF). The SR is the ration between the peak intensity from the Point Spread Function (PSF) of the aberrated eye and the peak intensity from the PSF of the non-aberrated eye. Please, correct lines 113 to 115. Thank you for this comment. We updated the definition.

Material and Methods

a) Patients and ethic statement

• The 2 last paragraph are not well located in this section and it makes confusion to the reader. In lines 132 – 135, it is described that the only inclusion criteria is to obtain good OSI measurements. However, it is not known any exclusion criteria and ophthalmic exam until next section. Similarly, it is observed in the last paragraph (lines 136 – 141), where some subjects were excluded because of poor cooperation and dense cataracts, without knowing any details from phacodynamics mode adjustment. I would recommend to move this section before the section “statistical analysis” to describe previously inclusion and exclusion methods. Thank you for this comment. We create an “inclusion & exclusion criteria” section just before statistical analysis section.

Results

a) Correlation Analysis: Main objective, subjective preoperative parameters and phacodynamics.

• In order to avoid misunderstanding, I do recommend strongly to indicate “LogMar scale” in the axis Best-Corrected Visual Acuity (BCVA) in figures 1, 5 and 6. The current values can confuse to the reader with decimal scale and the interpretation of the Spearman coefficient can be opposite. Thank you for this comment. We have updated figures 1, 5 and 6.

b) Correlation Analysis: Other OQAS parameters, subjective preoperative parameters and phacodynamics.

• Relations between Optical Scattering Index (OSI) with Modulate Transfer Function (MTF) in lines 273 – 274; and between Strehl ratio (SR) with MTF in lines 281 – 282, should not be included in the manuscript. These correlations were well developed and thoroughly analysed by the company in order to determine the performance of the device. Consequently, these data are not going to give any particular and novel information. Thank you for this comment. The data in question have been deleted.

Discussion

• I do not really agree with the idea that OQAS and LOCSIII measures the same physical phenomenon. Specially, because OQAS measures the quality of the whole eye, and a simple uveitis or a corneal scar can affect completely the measurement. I would be more specific in this sentence (lines 390 – 392) explaining that both methods will be measuring the same parameters if there are no more ocular pathologies. Thank you for this comment. The clarification has been made.

Reviewer #3: Thank you for your answers and comments. You have solved most of my concerns successfully. Regarding language, it has improved significantly and it makes the text much easier to follow now. On the other hand, I personally think that IQR=Q-Q3 provides an easier explanation for the data dispersion and allows the comparison with SD (line 187 and tables). Thank you for this comment.

Some specific comments:

- Line 90: Repetition of a fragment. Thank you for this comment. The repetition has been removed.

- Lines 210 – 211: The sentence needs revising for a better understanding. The sentence was revised.

- Lines 240 – 242: Lack of statistical significance for correlation between LOCSIII C and CDE as well as US is probably due to insufficient number of patients in this subgroup (LOCSIII C). Notice that rs is > 0,580. You should comment it here and explain it in discussion. As section results must be factual, we have now explained it in the discussion, per your suggestion.

- Figures 5 and 6: The top right and bottom left graphs seem to be wrong e.g. n = 21? Thank you for this comment. The titles of the top-right and bottom-left graphs were inverted. This mistake has now been corrected.

- I would be very grateful if you could send me the data rs and p for the different subgroups of LOCSIII components because note (lines 235 – 237) that they are not shown in Figure 1.

We found significant correlations between BCVA and LOCSIII total (rs = 0.561 ; p=0.008 ; Y = 6.258*X + 2.575) (Fig 1) as well as in subgroups of different LOCSIII components

N rs = 0.633 ; p=0.005 ; Y = 4.738*X + 2.678

NC rs = 0.621 ; p=0.006 ; Y = 2.294*X + 1.338

NO rs = 0.581 ; p=0.012 ; Y = 2.444*X + 1.340

except for LOCSIII C rs = 0.007 ; p=0.696 ; Y = -0.746*X + 2.284

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

Attachment

Submitted filename: R2_Reponse to reviewers.docx

Decision Letter 2

Ireneusz Grulkowski

25 Sep 2020

Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients

PONE-D-20-08844R2

Dear Dr. Garcin,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Ireneusz Grulkowski, PhD

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Please address the comment of the 3rd reviewer during proof.

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: All comments have been addressed

Reviewer #2: All comments have been addressed

Reviewer #3: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

Reviewer #3: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: the study uses Optical Quality Analysis System to predicte surgical parameters in age-related cataract patients.

Reviewer #2: (No Response)

Reviewer #3: Thank you for your satisfying answers and comments. The new version of the manuscript has improved considerably. However, I would like to comment on a small detail that I have been able to observe in this review:

In the Introduction (lines 97-99) the last sentence in the paragraph states, ““ Scheimpflug camera or Shack-Hartman technology do not consider light scattering [33] and can overestimate the optical quality of an image when scattering affect the eyes, for example in cataract [34].”

This statement is true for Shack-Hartmann technology (by the way, the spelling of the surname needs correcting) but it is not for the Scheimpflug camera. I suggest removing the Scheimpflug camera from the sentence.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

Reviewer #3: No

Acceptance letter

Ireneusz Grulkowski

1 Oct 2020

PONE-D-20-08844R2

Using Optical Quality Analysis System for predicting surgical parameters in age-related cataract patients

Dear Dr. Garcin:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Ireneusz Grulkowski

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    Attachment

    Submitted filename: Vend_040720TG__R1_Reponse to reviewers.docx

    Attachment

    Submitted filename: R2_Reponse to reviewers.docx

    Data Availability Statement

    All relevant data are within the paper.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES