Abstract
The uneven morphology and the trapped charges at the surface of the traditionally used supporting substrate-based 2D biosensors produces a scattering effect, which leads to a irregular signals from individually fabricated devices. Though suspended 2D channel material has the potential to overcome scattering effects from the substrates but achieving reliability and selectivity, have been limiting the using of this biosensor technology. Here, we have demonstrated nanogap electrodes fabrication by using the self-assembly technique, which provides suspension to the 2D-MoS2. These nano-spacing electrodes not only give suspension but also provide robustness strength to the atomic layer, which remains freestanding after coating of the Hafnium oxide (HfO2) as well as linkers and antibodies. For evaluating the electrical characteristics of suspended MoS2 FET, gating potential was applied through an electrolyte on the suspended MoS2 transistor. This helped in achieved a lower subthreshold swing 70 mV/dec and ON/OFF ratio 107. Later, pH detection was conducted at room temperature, which showed an impressive sensitivity of ~880 by changing 1 unit of pH. We have also successfully shown Escherichia coli (E. coli) bacteria sensing from the suspended MoS2 transistor by functionalizing dielectric layer with E. coli antibodies. The reported biosensor has shown the ~9% of conductance changes with a lower concentration of E. coli (10 CFU/mL; colony-forming unit per mL) as well as maintain the constant sensitivity in three fabricated devices. The obtained enhancement in the sensitivity of devices and its effect on biomolecules detection can be extened to other biomolecules and this type of architecture has the potential to detect COVID-19 viruses based biomolecules.
Keywords: Biosensors, Field effect transistors, Escherichia coli, MoS2, Nanogap
Highlights
-
•
A suspended MoS2 atomic layer field-effect transistors is reported.
-
•
Self-assembled nanogap electrodes provide remarkable robust strength to the MoS2 atomic layer.
-
•
Suspended MoS2 field-effect transistor device exhibits impressive pH sensitivity and Escherichia coli bacteria detection.
-
•
The sensing mechanism highly selective to Escherichia coli bacteria.
1. Introduction
Due to their potential of providing continuous and real-time information, biosensors are prominent devices in the medical field and are garnering substantial interest in other domains such as forensic industries, national security, and ecological monitoring (Mehrotra and research, 2016). Specifically, electrochemical field-effect transistor (FET)-based biosensors have been recognized as one of the most promising candidates because of their compatibility with electronic devices, low power consumption, label-free detection of specific biomolecules, and low-cost mass production (Kim et al., 2016; Kuriyama and Kimura 1991; Suvarnaphaet and Pechprasarn 2017). In FET based biosensors, the charged biomolecules generate the electrostatic effect and bring the variation in the conductivity of the channel, which can be easily measured by transistor characteristics like source to drain current (Kaisti 2017; Sarkar et al., 2014a). To obtain better electrostatic effects from charged biomolecules, many semiconducting materials have been employed (Holzinger et al., 2014; Li et al., 2017; Zhu et al., 2012). Among the various materials, two-dimensional (2D) semiconducting materials holds great promise because of their tunable bandgap and a high surface to volume ratio that provides higher sensitivity for detection of biomolecules.
In the 2D domain, transition metal dichalcogenides (TMDs) like MoS2 have gained much attention in the field of FET-biosensor due to their direct bandgap, biocompatibility, and high mobility (Kalantar-zadeh and Ou 2015b). Despite these merits, the performance and consistency of these atomic layers are easily affected by supporting substrate interaction (Chen et al., 2018). Such an interaction of supporting substrate with atomic layers of MoS2 degrades the transport properties due to scattering, which implies that, interface control is vital for the performance and reliability of biosensor devices (Du et al., 2008). For instance, the surface of silicon dioxide (SiO2) substrate is highly disordered as well as chemically active due to the trapped atmospheric gases, chemical adsorbates, and unknown functional groups (Zhang et al., 2009). Therefore, transferring another layer of MoS2 on the top of SiO2 or any other insulating substrate cannot contribute to carrier charge transport distinctly, which leads to an unreliable output from every single device. In recent years, many efforts have been employed to enhance the quality of the substrate, such as placing an active layer interface by using Poly (methyl methacrylate) (PMMA) and polymer electrolytes [9]. These layers elude chemical bonding or surface roughness and improve carrier transport properties, but they cannot be employed in biosensor applications due to the difficulty in fabrication and reproducibility issues. The other approach is creating suspension of an atomic layer in between electrodes to enhance the carrier transport and remove scattering effect by wet etching of the SiO2 layer underneath the monolayer MoS2. Freestanding MoS2 has shown better performance than the supporting on the SiO2 substrate in terms of back gating electronic conduction (Jin et al., 2013). However, the existing SiO2 requires hazardous chemical etchants such as hydrofluoric acid (HF), which is difficult to handle and affects the 2D film structure and purity (Zhang et al., 2017). Secondly, freestanding MoS2 sags between the two electrodes because of the large spacing (~2 μm), which makes it impossible to coat dielectric layers such as hafnium oxide (HfO2) and antibodies. Therefore, this structure impedes making top gate FET biosensors and allows only back gating. However, the back-gate FET requires more power (input gate voltage) to turn ON the device than the top gate, which hinders making a low power and highly sensitive biosensor. In FET based biosensor, the sensitivity is inversely proportional to the subthreshold swing (SS), which is defined as a change in the current with respect to dielectric surface potential (; dVgate and Idrain are a change in gate potential and drain current respectively). Thus, back gate requires higher input voltage to turn on the device than the top gate, which reduces the sensitivity of detection (Sarkar et al. 2014a, 2014b).
Given the importance of reliability and sensitivity of FET based biosensor, in this work, chemical vapor deposition (CVD) grown MoS2 is transferred by using novel dry stamping method on self-assembled photolithographically patterned nano-gaps to achieve suspension. Fabricated nanogap electrodes have 70–90 nm spacing, which provides mechanical strength to the MoS2 monolayer and does not allow to sag between the two electrodes. This robustness in the monolayer film permits the coating of another thin dielectric layer as well as linkers/antibodies without interacting, supporting the substrate. HfO2 as a dielectric material was considered, which is easily functionalized by a linker (3-Aminopropyl)triethoxysilane (APTES), glutaraldehyde, and E. coli antibodies to bring variation to the suspended 2D MoS2 by targeting a charged E. coli bacteria. The suspended FET architecture lowered down the subthreshold swing (SS) to 70 mV/dec by terminating the scattering phenomenon and enhance the sensitivity of the pH as well as E. coli detection (Liao et al., 2018). Freestanding nature of 2D channel material i.e., MoS2, retains its intrinsic properties and provides reliability and consistency in every single device.
2. Experimental section
2.1. CVD synthesis
Experimental procedures of CVD (In-house built) grown MoS2 are as follows: quartz boat consists of MoO3 powder, and SiO2/Si pieces (top of the powder) were positioned in the middle of furnace whereas sulphur boat was placed at the upstream of lower temperature zone. In the typical process flow of CVD growth of MoS2, Argon gas was passed for 20 min at 250 sccm to wash out all environmental gases and then ramp up the temperature till 750 °C in 15 min. Afterward, the temperature was maintained at 750 °C for 20 min. In this condition, sulphur starts evaporating and reduced MoO3 to converts into metallic MoO2. Further, more sulphurization of MoO2 leads towards the triangular shape of MoS2 flakes on the SiO2/Si pieces.
2.2. Photolithography of nanogaps
In this process, the self-assembly technique was adapted to achieve nanogaps between the metal electrodes. Here, few parameters were optimized while fabrication such as lift-off resists thickness, rpm rate of spinning photoresist, baking, developing the photoresist, avoiding the residues during fabrication, and most importantly the optimization of thickness of metal electrode layer. The optical photolithography technique of patterning the substrate was done by coating the first positive photoresist (MICROPOSIT S1800) and expose (SUSS MicroTec Mask Aligner MA6) it for 80 mJ/cm2 (flood exposure). Then another layer of photoresist 1811 is coated on the top and patterned with design chrome mask followed by evaporation of the Ti/Au/Cr (thickness is 10 nm/50 nm/80 nm). After patterning three metal layers Ti/Au/Cr, the wafer is kept for 12 Hrs. to control thermal oxidation of the chromium to achieve nanogap. Nano-spacing between electrodes rely on the rate of oxidation of the Cr layer, where temperature controls the rate of the oxidation. Therefore, it is essential to maintain the temperature as well as environmental gases of the wafer constant throughout the oxidation process. To achieve the proper pattern of the second layer, the thickness of the Ti/Au (second layer, whose thickness is 10 nm/50 nm) should not exceed more than 50% of chromium thickness. Liftoff the chromium as well as the second Au electrode is processed in chromium etchant solution (Chromium etchant, Sigma Aldrich).
2.3. Electrical characterization
Before transferring the 2D material, it was characterized by the current-voltage method. I–V curve between two electrodes was measured by the source meter (Keysight B2912A) to discern its electrical characteristics.
2.4. Transferring of 2D material and coating of the HfO2
After achieving the optimum nanogap electrodes, triangular shape MoS2 atomic layers was transferred on the top of the spacing by using a novel dry stamping method. This method provides proper placement of CVD grown MoS2 in the desired area. Atomic layer deposition (ALD), University of Michigan, Ann Arbor (Veeco Fiji ALD), was performed to deposit 25 nm of the high dielectric layer.
2.5. Raman and AFM characterization
Prepared samples of MoS2 are characterized by Raman Spectroscopy using 514 nm green (Nd-YaG) laser, whereas AFM (XE NSOM) mapping was performed by using a customized tip purchased from App Nano.
2.6. Functionalization of the HfO2
3-Aminopropyltriethoxysilane (99%) (Sigma Aldrich), glutaraldehyde solution (50 wt% in H2O, Bio Basic), E. coli antibodies (Rabbit polyclonal to E. coli, Abcam) was utilized for functionalization of the HfO2. The suspended device was absorbed into APTES for 2 Hrs. in the ethanol/water mixture followed by the cleaning of the chip by ethanol and then blow-dried. Subsequently, APTES was functionalized on chip by immersing them in a glutaraldehyde solution for approximately 1 Hr. Afterward, E. coli antibodies were incubated to the suspended chip in 0.05 M PBS solution for 2 Hrs. at room temperature.
2.7. Bacteria growth
Bacterial specimens used in the study were prepared from culture plates. An isolated singer colony was added to 5 ml of media in a 14 ml culture tube. The culture tube was placed on a shaker in a 37 °C incubator and incubated overnight (18 h). The overnight culture was centrifuged at room temperature for 5 min at 3500 rpm. The supernatant was removed, and the bacteria pellet was re-suspended in 5 ml of filtered sterilized tap water. The bacteria were centrifuged, and the washing process was repeated once. After the final wash, filtered tap water was added to the bacteria pellet until the optical density (measured at a wavelength of 600 nm using a Spectro Max PLUS spectrophotometer) of the solution reached the desired value. After the optical density at 600 nm was measured for the stock solution, serial dilutions were employed.
3. Results and discussions
3.1. Suspended biosensor fabrications
Our approach for patterning the nanogap electrode uses a self-assembly photolithography technique, as shown in Fig. 1 (a). The metal lift-off method has been used for engraving the first electrode gold (Ti/Au) and chromium (Cr), the thickness of the Au was kept less than the Cr because it acts as a sacrificial layer as represented in Fig. 1(a) step I. After achieving the first electrode, the wafer was kept at ambient temperature for an overnight where Cr gets oxidized and forms a thin layer of CrxOy. Due to the formation of CrxOy, the overall volume of the Cr increases by a few nanometers and overhangs on the edges of the bottom electrode, as demonstrated in Fig. 1(a) step II. It is crucial to keep the wafer in a controlled environment to achieve uniform expansion of the Cr to avoid unevenness in the nanogap and connection between two electrodes (Fursina et al., 2008). In the next step, Au is deposited and patterned for the second electrode. However, enlargement of the Cr layer in a few nanometers protects the second electrode from meeting the first metal layer and creates the nanometer spacing, as shown in Fig. 1 (a) step III. At IV step, Cr was stripped out left behind two Au electrodes, which are separated by few nanometers. In this self-assembly technique, the lateral resolution of the gap between two gold electrodes achieved nearly 70–90 nm at room temperature. The nanogap spacing between two electrodes is reproducible, constant throughout the wafer, and feasible for bulk fabrication due to the involvement of the wafer-scale lithography. To examine the fabricated electrode, which is not electrically connected, the voltage sweep has been performed, and the resulting current was monitored. This electrical characteristic demonstrated open circuit or short circuit between two nanogap electrodes (Supplementary Fig. 1). Throughout the wafer scale fabrication process of the nanogaps, there are several factors involves for achieving high yield such as chromium thickness, thermal oxidation, and thickness of the second layer Ti/Au. In our optimized process flow, the yield of the constant spacing was 70%–80% (±4 nm), which are mostly found in the center of the Silicon wafer.
For channel material in FET, CVD grown MoS2 atomic crystals used in this work, which was characterized by using Raman spectroscopy and atomic force microscopy (AFM) (Supplementary Fig. 2). After achieving nanogap electrodes, a single atomic layer of MoS2 flake was transferred on the top of nano spaced electrode by using a novel dry stamping method (Supplementary Fig. 3). The PMMA acts a sacrificial layer to transfer 2D material onto the nano gap electrode. After the transfer is done the PMMA is removed by placing the device in acetone bath, this will dissolve the PMMA without affecting 2D material transferred onto the nano gaps. This dry stamping method allows transferring a single crystalline MoS2 triangle precisely on the top of nanogaps by using a micromanipulator. Nano-spacing between the electrode (90 nm) does not allow the MoS2 membrane (0.65 nm) to touch the supporting substrate, and it is suspended in between the two electrodes. To confirm the freestanding nature of MoS2, AFM imaging was performed on the nanogap region with and without MoS2 to elucidate the suspension of the atomic layer on the electrodes. Fig. 1 (b) shows the optical image and AFM mapping with the line profile of the nanogap area without MoS2 flake. AFM line profile of nanogap electrodes demonstrates the step height of 75 nm in between the two Au electrodes, which confirms the two metal layers are separated by nanometer spacing. Fig. 1(c) represents the optical image and AFM mapping of transferred MoS2. AFM line profile of transferred MoS2 on the top of nanogap indicates the flat line and no step height, which verify that the atomic membrane is suspended and is remarkably stiff as well as robust. This stiffness of the atomic layer confirms that it doesnot interacted with supporting substrates and possibly eliminates the scattering phenomena of the interface between the 2D material and the holder substrate. This mechanical strength provided by nano spacing does not allow the monolayer sagging between the electrodes, which permits depositing another layer, i.e., HfO2 and biological receptor for selective sensing. To further validate the rigid suspension of the MoS2 monolayer, a scanning electron microscope (SEM) and dark field image of the MoS2 triangle on the top of the nanogaps were performed, as shown in Fig. 1 (d) & (e). These images clearly illustrate that the MoS2 triangle does not sag in between the nanogap spacing; it is suspended and flat on the top of two Au electrodes.
After confirming the rigidness of the MoS2, it is desirable to eliminate the density of defects by covering the bare channel material with insulating material and then functionalizing linkers and antibodies on the top of the insulator (Kalantar-zadeh and Ou 2015a; Sarkar et al., 2014a). If biomolecule directly functionalized on the semiconductor layer without insulation, it might attach to the metal electrodes and change the work function as well as the contact resistance, which leads to the reliability issue and bring the electrical variation in the sensor output. Therefore, the dielectric layer is essential to passivate the metal electrodes and eliminate direct contact with electrolyte or biomolecules. Therefore, a 25 nm HfO2 dielectric layer is deposited on the top of the suspended device using atomic layer deposition (ALD). Afterward, macro fluidic storage is fabricated using Polydimethylsiloxane (PDMS) for encompassing electrolyte droplets, which is placed precisely on the top MoS2 triangle covered with HfO2. A reference electrode (Ag/AgCl) is used to apply bias through the electrolyte for stabilizing the process and better control of the operation under the biosensor regime, as shown in the optical image of Fig. 1 (f) (Sarkar and Banerjee 2012). Fig. 1 (g) represents the schematic structure of the suspended MoS2 FET biosensor by selective detection of the E. coli bacteria.
3.2. Electrical characterization
The electrostatic effect of MoS2 FET was investigated to compare the electrical performance of the supported and suspended devices. Fig. 2 represents the current-voltage (I–V) characteristics by applying different gate voltages (VBG) of fabricated devices in a dry and wet environment. Fig. 2 (a) and (b) illustrates the ID-VBG curve of supported and free-standing MoS2 at 100 mV bias voltage (VDS), where supported channel length is 2 μm and suspended is ~90 nm. The suspended device represents excellent ON/OFF ratio 107 and threshold voltage (VT) 3.9 V as compared to the supported one, which is equal to 106 and -9 V at room temperature. The improvement of the VT, i.e., switching of the device at low voltage, is due to the elimination of the supporting substrate, which confines the electrons mean free path by scattering as well as from trapped charges on the surface (Masurkar et al., 2020). Fig. 2 (a’) and (b’) represents the ID-VDS characteristics at different VBG (−10 V–30 V). In both cases, the VDS varies from the 0 mV–100 mV, and the output current shows the linearity with the input voltage, which confirms that MoS2 has ohmic contact with the Au/Ti contacts. Suspended FET displays better VT and an ohmic contact with metal in back gate voltage characteristics. Still, due to the high power consumption, it is not feasible for use in the biosensor domain. Therefore, we have used a thin dielectric layer of the HfO2 on the top of the channel and studied the ionic gating effect on the suspended MoS2 channel, as shown in Fig. 2 (c) and (c’). Phosphate buffer solution as an electrolyte has been used for the ionic gating measurements by transporting a drop in macro fluidic storage. ID-VLG analysis of suspended ionic gating FET displayed impressive Subthreshold swing (SS) 70 mV/dec, which is near to ideal at room temperature as well as the mobility (13.5 cm2 V−1 s−1) ON/OFF ratio (107) and VT (− 0.29 V). The mobility of all three types of the devices are calculated and shown in the Supplementary Table S1 (Supplementary Fig. S4). Low SS, threshold voltage, high mobility and elimination of substrate scattering make these types of the FET structure more sensitive as well as consistency with better gate controllability in a wet environment.
(1) |
3.3. pH sensing
After achieving excellent performance in the ionic gating effect, the suspended device was investigated by changing the pH of the electrolyte solution. The pH sensing strongly depends upon the proton (H+) present in the electrolyte solution which is react with the OH group present on the HfO2 surface and causes the protonation and deprotonation reaction, as shown in Fig. 3 (a) (Shadman et al., 2016). Depending upon the pH value of the electrolyte solution, the dielectric surface charge of the suspended FET varies, a lower pH value protonates the surface by generating OH2 + on the HfO2 whereas the higher pH value deprotonates the surface by extracting H+ (Ameri et al., 2015; Dinar et al., 2019). Fig. 3 (b) represents the drain current (ID) (100 mV bias voltage) at different pH values (3, 5, 7, and 9) by changing the electrolyte gate voltage (VLG). It is found that the VT of ID-VLG curves shifts to the positive side from low to high pH due to the negative charge developed on the dielectric surface. The change in ΔVT found to be 59.1 mV/pH, which is satisfied the Nernst limit at room temperature, i.e., 59.3 mV/pH and agree with early studies of ion-sensitive field-effect transistor (ISFET) characterization on HfO2 (Van Hal et al., 1995; Zafar et al., 2011). Therefore, HfO2 does not require any functionalization for pH sensing as compared to the SiO2, whose change in threshold voltage found to be 30–40 mV/pH (Bergveld and Physical 1996; Gao et al., 2009). Fig. 3 (c) represents the sensitivity of the pH at three different regions (Subthreshold, saturation, and linear), which is defined by the equation (1), where Sn is the sensitivity of pH, IpH1 and IpH2 are the transistor current measured from two different pH values (where pH1>pH2).
The sensitivity at the subthreshold region was found to be much higher because the drain current is an exponential function of the gate voltage. In contrast, the saturation and linear parts are quadratic and linear with the change in gate potential. In a suspended device, the sensitivity at the subthreshold region was found to be ~876 from pH 5 to 6 and ~880 from pH 7 to 8 respectively, which is far better than the previous reports of MoS2 FET biosensor on the supporting substrate (Sarkar et al., 2014b). The suspended MoS2 FET possesses higher sensitivity than supported one because of the lower SS, where SS defined as the subthreshold swing. This equation elucidates that the change in subthreshold current by one decade is a function of the applied gate voltage, whereas the consistency in the sensitivity of the suspended device at two different ΔpH ranges is due to the elimination of the external scattering from the supporting substrate.
3.4. Bacteria sensing from suspended FET
Finally, suspended FETs are fabricated as described in the fabrication section to detect the biomolecules. To specifically detect the E. coli bacteria from the buffer solution (pH = 7.4), linkers were immobilized on the top of the HfO2. For evaluating the bonding of the linkers with oxide surface, di-thiol and (3-Aminopropyl)triethoxysilane (APTES) have used, followed by glutaraldehyde treatment and antibodies immobilization (Gunda et al., 2014; Lee et al., 2006). From these two molecules, APTES has shown promising ID-VLG curve shifts while detecting bacteria as compared to the di-thiol (Supplementary Fig. 5). It is interesting to see that after the functionalization of the APTES/glutaraldehyde linkers on the surface of the HfO2, there is a small shift of the ID-VLG curve as illustrated (Supplementary Fig. 6). The overall process flow of the linkers binding, antibodies immobilization on suspended MoS2 FET through macro storage fluidic channel to detect E. coli as shown in the schematic of Fig. 4 (a) (Gunda et al., 2014; Wu et al., 2013). The ID-VLG transfer curve of the linkers, antibodies, PBS buffer solution, and the 100 CFU/mL of E. coli bacteria was performed, as demonstrated in Fig. 4 (b). It is essential to introduce a buffer solution again after antibodies immobilization to confirm there is no change of conductance (Cyan curve in Fig. 4 (b)). A shift found in the ID-VLG curve on the left side after incubation of 100 CFU/mL of E. coli (Magenta curve in Fig. 4 (b)).This illustrated that the current in the MoS2 channel deteriorated due to the increment of hole concentration induced on the dielectric surface by highly negative charged bacteria wall (Huang et al., 2011).
After achieving considerable performance from the suspended MoS2 FET device, different E. coli concentrations ranging from 0 to 103 CFU/mL were prepared and used for detection. The increase in the concentration of the bacteria leads to the decrement of the MoS2 channel conductance. For buffer solution, ID at (VLG = 0V) is illustrated 0.3 μA, whereas 10 CFU/mL shows the current response 0.274 μA, the change in the conductance of freestanding MoS2 was found to be ~9% with a low concentration of the bacteria. The change in conductance for 100 CFU/mL and 103 CFU/mL with buffer solution illustrated 18% and 25%, respectively, as shown in Fig. 4 (c). Another essential factor in the biosensor domain is selectivity, which attributes that other bacterial strain should not be attached with E. coli antibodies to achieve false signals. Therefore, Pseudomonas aeruginosa (P. aeruginosa) bacteria (100 CFU/mL) were considered to measure the ID-VLG signal by keeping VDS = 100 mV. P. aeruginosa bacteria did not show any change in conductance as compared to the E. coli (100 CFU/mL), as shown in Fig. 4(d). Fig. 4 (e) illustrates the sensitivity of the biosensor, which is defined as shown in equ (2),
(2) |
where Sn(CFU/mL) sensitivity, Buffer is a buffer solution current, In(CFU/mL) is current after bacteria bind to the FET biosensor. It was found that the shift in ID-VLG is more in the subthreshold region as compared to saturation and linear regions. The sensitivity of the sensor in the subthreshold region achieved 83.3 for 100 CFU/mL of bacteria at −0.35 V. After obtaining higher sensitivity, the other most crucial factor in the 2D FET biosensor is the reliability in every fabricated sensor, which can be easily affected by external scattering from supporting substrate or the contact resistance from metal. In our case, we have maintained the constant work function between the electrode and the MoS2 in every single device as well as eliminate the supporting substrate interaction, which attributes to the constant sensitivity in three suspended MoS2 FET at the subthreshold region (100 CFU/mL and −0.35 V) as demonstrated in Fig. 4 (f).
The sensitivity presented in this report is found to be impressive as compare to the previous reports of biomolecule detection via 2D FET based technology. Graphene-based FET biosensor has shown the conductance change of 3.25% at 10 CFU/mL of E. coli, which is almost three times smaller than the presented value (9% at 10 CFU/mL) (Chang et al., 2013; Huang et al., 2011). The detection limits of the graphene are due to the zero band gap in nature, which leads to an increase in the off-state leakage current and the SS value (SS is inversely proportional to the sensitivity) (Schwierz 2010). On the other hand, FET based MoS2 biosensor has shown promising sensitivity in biomolecule detection due to ~1.8 eV bandgap. However, the uneven morphology from the supporting substrate leads to an unreliable output from every single device due to different interface resistances caused majorly by trapped charges, chemical adsorbates, and unknown functional groups (Jin et al., 2013; Kaushik et al., 2018). This is a significant barrier in proliferating this technology in biosensor industries. Therefore, elimination of the scattering effect in the proposed biosensor shows that the overall change of conductance in the 2D film is entirely due to the change in concentration of the biomolecule. Table S2 in Supporting Information represents a comparison of different 2D material based FET biosensor in terms of channel materials, detection limits, biomolecules, and sensitivity, whereas Table S3 in Supporting Information illustrates the comparison between different dimension and materials with respective their sensitivity, feasibility in bulk fabrication, and reliability. Due to these merits i.e., the involvement of the CVD method, direct transfer technique, and optical lithography open the new avenues for the bulk fabrication of suspended MoS2 biosensor.
4. Conclusion
In summary, we provide an comprehensive approach for fabrication highly sensitive and reliable biosensor based on suspended MoS2 FET to detect the pH as well as E. coli bacteria. CVD grown MoS2 channel material was transferred on photolithographically patterned nano-gaps to achieve suspension and covered with Hafnium oxide (HfO2) as a dielectric material to eliminate direct functionalization on the channel material. The proposed biosensor demonstrated the reliable and quantitative detection of pH and biomolecule by eliminating the scattering effect from the supporting substrate. A control experiment of pH sensing confirms that the high sensitivity (~880/pH) as well as the constant ID-VLG curve shift at different pH range (3–9). Suspended MoS2 FET biosensor provides selective detection of the E. coli bacteria as low as 10 CFU/mL through changes in the conductance by 9% of MoS2 channel at ultralow power (VDS = 100 mV). This architecture can detect the low concentration of biomolecule because of the atomic layer as a channel material, and is scalable due to the involvement of optical photolithography, dry stamping, and CVD technique as well as reliable and reproducible due to the elimination of the supporting substrate. These merits of suspended MoS2 FET biosensor makes it promising next-generation sensing device for on-spot detection of biomolecules and can be extened to COVID-19 viruses detection and bioelectronics applications.
CRediT authorship contribution statement
Nirul Masurkar: Conceptualization, and designed the experiments. Naresh Kumar Thangavel: performed the biological experiments, Writing - original draft. Sally Yurgelevic: performed the biological experiments, Writing - original draft. Sundeep Varma: fabricated and characterized materials and devices, electrical characterization, Formal analysis. Gregory W. Auner: Conceptualization, and designed the experiments, Formal analysis. Leela Mohana Reddy Arava: Conceptualization, and designed the experiments; electrical characterization, performed the biological experiments, Formal analysis, Writing - original draft.
Declaration of competing interest
The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Acknowledgments
This article was supported in part by the National Science Foundation under grant no 1751472 and ACS Petroleum Research Fund (ACS PRF: 57647-DNI10).
Footnotes
Supplementary data to this article can be found online at https://doi.org/10.1016/j.bios.2020.112724.
Appendix A. Supplementary data
The following is the Supplementary data to this article:
References
- Ameri S., Singh P., Sonkusale S. IEEE; 2015. Three Dimensional Monolayer Graphene Foam for Ultra-sensitive pH Sensing. 2015 Transducers-2015 18th International Conference on Solid-State Sensors, Actuators and Microsystems (TRANSDUCERS) pp. 1378–1380. [Google Scholar]
- Bergveld P.J.S., Physical A.A. The future of biosensors. 1996;56(1–2):65–73. [Google Scholar]
- Chang J., Mao S., Zhang Y., Cui S., Zhou G., Wu X., Yang C.-H., Chen J.J.N. Ultrasonic-assisted self-assembly of monolayer graphene oxide for rapid detection of Escherichia coli bacteria. 2013;5(9):3620–3626. doi: 10.1039/c3nr00141e. [DOI] [PubMed] [Google Scholar]
- Chen H., Li J., Chen X., Zhang D., Zhou P.J.S.S., Technology Dramatic switching behavior in suspended MoS2 field-effect transistors. 2018;33(2) [Google Scholar]
- Dinar A.M., Zain A.M., Salehuddin F.J., Engineering C. Comprehensive identification of sensitive and stable ISFET sensing layer high-k gate based on ISFET/electrolyte models. 2019;9(2):926. [Google Scholar]
- Du X., Skachko I., Barker A., Andrei E.Y.J.N.n. Approaching ballistic transport in suspended graphene. 2008;3(8):491. doi: 10.1038/nnano.2008.199. [DOI] [PubMed] [Google Scholar]
- Fursina A., Lee S., Sofin R., Shvets I., Natelson D.J.A.P.L. Nanogaps with very large aspect ratios for electrical measurements. 2008;92(11) [Google Scholar]
- Gao X.P., Zheng G., Lieber C.M.J.N.l. Subthreshold regime has the optimal sensitivity for nanowire FET biosensors. 2009;10(2):547–552. doi: 10.1021/nl9034219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Gunda N.S.K., Singh M., Norman L., Kaur K., Mitra S.K. Optimization and characterization of biomolecule immobilization on silicon substrates using (3-aminopropyl) triethoxysilane. APTES) and glutaraldehyde linker. 2014;305:522–530. [Google Scholar]
- Holzinger M., Le Goff A., Cosnier S.J.F.i.c. Nanomaterials for biosensing applications: a review. 2014;2:63. doi: 10.3389/fchem.2014.00063. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Huang Y., Dong X., Liu Y., Li L.-J., Chen P.J. Graphene-based biosensors for detection of bacteria and their metabolic activities. 2011;21(33):12358–12362. [Google Scholar]
- Jin T., Kang J., Su Kim E., Lee S., Lee C. Suspended single-layer MoS2 devices. J. Appl. Phys. 2013;114(16) [Google Scholar]
- Kaisti M. Detection principles of biological and chemical FET sensors. Biosens. Bioelectron. 2017;98:437–448. doi: 10.1016/j.bios.2017.07.010. [DOI] [PubMed] [Google Scholar]
- Kalantar-zadeh K., Ou J.Z. Biosensors based on two-dimensional MoS2. ACS Sens. 2015;1(1):5–16. [Google Scholar]
- Kalantar-zadeh K., Ou J.Z.J.A.S. Biosensors based on two-dimensional MoS2. 2015;1(1):5–16. [Google Scholar]
- Kaushik N., Ghosh S., Lodha S. Low-frequency noise in supported and suspended MoS 2 transistors. IEEE Trans. Electron. Dev. 2018;65(10):4135–4140. [Google Scholar]
- Kim S.K., Bhatia R., Kim T.-H., Seol D., Kim J.H., Kim H., Seung W., Kim Y., Lee Y.H., Kim S.-W.J.N.E. Directional dependent piezoelectric effect in CVD grown monolayer MoS2 for flexible piezoelectric nanogenerators. 2016;22:483–489. [Google Scholar]
- Kuriyama T., Kimura J. FET-based biosensors. Bioproc. Technol. 1991;15:139–162. [PubMed] [Google Scholar]
- Lee L.M., Heimark R.L., Baygents J.C., Zohar Y.J.N. Self-aligned immobilization of proteins utilizing PEG patterns. 2006;17(4):S29. doi: 10.1088/0957-4484/17/4/006. [DOI] [PubMed] [Google Scholar]
- Li X., Shan J., Zhang W., Su S., Yuwen L., Wang L.J.S. Recent advances in synthesis and biomedical applications of two‐dimensional transition metal dichalcogenide nanosheets. 2017;13(5) doi: 10.1002/smll.201602660. [DOI] [PubMed] [Google Scholar]
- Liao W., Wei W., Tong Y., Chim W.K., Zhu C.J.A.a.m., interfaces Low-frequency noise in layered ReS2 field effect transistors on HFO2 and its application for pH sensing. 2018;10(8):7248–7255. doi: 10.1021/acsami.8b00193. [DOI] [PubMed] [Google Scholar]
- Masurkar N., Varma S., Mohana Reddy Arava L. Supported and suspended 2D material-based FET biosensors. Electrochemistry. 2020;1(3):260–277. [Google Scholar]
- Mehrotra P.J., research c. Biosensors and their applications–A review. 2016;6(2):153–159. doi: 10.1016/j.jobcr.2015.12.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Sarkar D., Banerjee K.J.A.P.L. Proposal for tunnel-field-effect-transistor as ultra-sensitive and label-free biosensors. 2012;100(14) [Google Scholar]
- Sarkar D., Liu W., Xie X., Anselmo A.C., Mitragotri S., Banerjee K. MoS2 field-effect transistor for next-generation label-free biosensors. ACS Nano. 2014;8(4):3992–4003. doi: 10.1021/nn5009148. [DOI] [PubMed] [Google Scholar]
- Sarkar D., Liu W., Xie X., Anselmo A.C., Mitragotri S., Banerjee K.J.A.n. MoS2 field-effect transistor for next-generation label-free biosensors. 2014;8(4):3992–4003. doi: 10.1021/nn5009148. [DOI] [PubMed] [Google Scholar]
- Schwierz F.J.N.n. Graphene transistors. 2010;5(7):487. doi: 10.1038/nnano.2010.89. [DOI] [PubMed] [Google Scholar]
- Shadman A., Rahman E., Khosru Q.D.J.S., Research B.-S. Monolayer MoS2 and WSe2 double gate field effect transistor as super nernst pH sensor and nanobiosensor. 2016;11:45–51. [Google Scholar]
- Suvarnaphaet P., Pechprasarn S.J.S. Graphene-based materials for biosensors: a review. 2017;17(10):2161. doi: 10.3390/s17102161. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Van Hal R., Eijkel J.C., Bergveld P.J.S., Chemical A.B. A novel description of ISFET sensitivity with the buffer capacity and double-layer capacitance as key parameters. 1995;24(1–3):201–205. [Google Scholar]
- Wu S.-M., Chen J., Ai X.-X., Yan Z.-Y.J., analysis b. Detection of Escherichia coli in drugs with antibody conjugated quantum dots as immunofluorescence probes. 2013;78:9–13. doi: 10.1016/j.jpba.2013.01.005. [DOI] [PubMed] [Google Scholar]
- Zafar S., D'Emic C., Afzali A., Fletcher B., Zhu Y., Ning T.J.N. Optimization of pH sensing using silicon nanowire field effect transistors with HfO2 as the sensing surface. 2011;22(40) doi: 10.1088/0957-4484/22/40/405501. [DOI] [PubMed] [Google Scholar]
- Zhang F., Erb C., Runkle L., Zhang X., Alem N. Etchant-free transfer of 2D nanostructures. Nanotechnology. 2017;29(2) doi: 10.1088/1361-6528/aa9c21. [DOI] [PubMed] [Google Scholar]
- Zhang Y., Brar V.W., Girit C., Zettl A., Crommie M.F. Origin of spatial charge inhomogeneity in graphene. Nat. Phys. 2009;5(10):722–726. [Google Scholar]
- Zhu Z., Garcia-Gancedo L., Flewitt A.J., Xie H., Moussy F., Milne W.I.J.S. A critical review of glucose biosensors based on carbon nanomaterials. carbon nanotubes and graphene. 2012;12(5):5996–6022. doi: 10.3390/s120505996. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.