Abstract
Exacerbations of chronic heart failure (HF) with the necessity for hospitalisation impact hospital resources significantly. Despite all of the achievements in medical management and non-pharmacological therapy that improve the outcome in HF, new strategies are needed to prevent HF-related hospitalisations by keeping stable HF patients out of the hospital and focusing resources on unstable HF patients. Remote monitoring of these patients could provide the physicians with an additional tool to intervene adequately and promptly. Results of telemonitoring to date are inconsistent, especially those of telemonitoring with traditional non-haemodynamic parameters. Recently, the CardioMEMS device (Abbott Inc., Atlanta, GA, USA), an implantable haemodynamic remote monitoring sensor, has shown promising results in preventing HF-related hospitalisations in chronic HF patients hospitalised in the previous year and in New York Heart Association functional class III in the United States. This review provides an overview of the available evidence on remote monitoring in chronic HF patients and future perspectives for the efficacy and cost-effectiveness of these strategies.
Keywords: Remote monitoring, Heart failure, CardioMEMS, Telemedicine, Telemonitoring, eHealth
Introduction
The management of patients with chronic heart failure (HF) places a high burden on health care resources due to the frequent follow-up visits combined with recurrent hospitalisations due to cardiac decompensation [1]. Early detection of HF deterioration is crucial to prevent HF-related hospitalisations, potentially improve overall survival and quality of life and lower the burden on health care resources. Remote monitoring of chronic HF patients can aid in the detection of HF deterioration; therefore several remote monitoring strategies have been developed. In this review, we provide an overview of available evidence on remote monitoring of chronic HF patients and provide further perspectives of anticipated developments in the remote care of HF.
Non-haemodynamic remote monitoring
Over the last few decades, several studies have investigated the use of non-haemodynamic remote monitoring. However, the results have been largely inconsistent. A recently updated Cochrane review included 41 randomised controlled trials (RCTs) investigating the use of structured telephone support (25 studies, 9332 patients) or non-invasive telemonitoring (18 studies, 3860 patients) compared with standard HF care [2]. This review showed a modest beneficial effect of remote monitoring on all-cause mortality and HF-related hospitalisations, although no effect on the overall hospitalisation rates was observed. However, the quality of the evidence of this review is limited by the many different inclusion and exclusion criteria for patients included in the studies and considerably heterogeneity of compared data. Also, the studies included used different intervention therapies, ranging from telephone calls only, weight monitoring to complex multiple-variable telemonitoring strategies making it difficult to conclude which component drives the effect. Additionally, the majority of selected individual studies (more than twenty) were neutral.
Multiple large multi-centre prospective clinical studies and RCTs have investigated multiple non-invasive remote monitoring strategies, ranging from symptom and body weight monitoring to complex and intensive strategies including body weight, blood pressure, electrocardiography and peripheral capillary oxygen saturation. The landmark studies of high quality design and well specified intervention show no consistent beneficial effect of non-haemodynamic remote monitoring in HF patients (Tab. 1; [3–11]). Of specific note and most promising are the recent results of TIM-HF2 trial showing a benefit on all-cause mortality and cardiovascular hospitalizations of a well structured but labour intensive 24/7 telemonitoring strategy, but remarkably showed no effect on quality of life [6]. Also, ‘real-world’ data, such as those from the Medicare database, did not show consistent benefits of non-haemodynamic remote monitoring strategies on mortality or hospitalisation rates [12]. Our conclusion is that although results are inconsistent for non-invasive telemonitoring, the simplicity makes it potentially useful for larger groups of HF patients at relatively lower risk or less symptomatic, where invasive telemonitoring may have more impact in sicker patients.
Table 1.
Trial/study | Author; journal; year | No. of patients | Parameter | Endpoint | Impact on HF hospitalisation |
---|---|---|---|---|---|
TEN-HMS [5] |
Cleland et al.; J Am Coll Cardiol; 2005 |
426 | Signs/symptoms, daily weights, BP, nurse telephone calls | HF hospitalisation | Non-significant |
TELE-HF [4] |
Chaudhry et al.; N Engl J Med; 2010 |
1653 | Signs/symptoms, daily weights | HF hospitalisation | Non-significant |
TIM-HF [7] |
Koehler F et al.; Circulation; 2011 |
710 | Signs/symptoms, daily weights | HF hospitalisation | Non-significant |
INH [3] |
Angermann et al.; Circ Heart Fail; 2012 |
715 | Signs, symptoms, telemonitoring nurse coordinated | HF hospitalisation | Non-significant |
WISH [10] |
Lynga et al.; Eur J Heart Fail; 2012 |
344 | Daily weights | HF hospitalisation | Non-significant |
CHAT [9] |
Krum et al.; Cardiovasc Ther; 2013 |
405 | Monthly telephone‐based automated telemedicine system | HF hospitalisation | Non-significant |
BEAT-HF [11] |
Ong et al.; JAMA Intern Med; 2016 |
1437 | Signs, symptoms, daily weights, nurse communications | HF hospitalisation | Non-significant |
TIM-HF2 [6] |
Koehler F et al.; Lancet; 2018 |
1571 | Web-based remote monitoring on daily weight, BP, pulse, ECG, peripheral capillary oxygen saturation, a self-related health status. ECG and BP machine at home | Reduction in the weighted average of ‘the % of days lost due to unplanned CV hospital admissions or death’ | HR 0.80; 95% CI 0.65–1.00 |
BP blood pressure, CV cardiovascular, ECG electrocardiography
aDemonstrating the landmark trials only, sample size >250 patients, discounting studies with only phone calls as intervention
Remote monitoring using pacemaker/ICD devices
Multiple studies have investigated the remote monitoring abilities of implantable cardioverter defibrillator/cardiac resynchronisation therapy (ICD/CRT) devices in chronic HF patients to improve HF-related hospitalisation rates (Tab. 2). The MORE-CARE multi-centre RCT showed that remote monitoring of advanced diagnostics via CRT‑D did not reduce mortality or hospitalisation rates, although the health care resource utilisation was reduced due to a reduction in outpatient follow-up visits [13]. Additionally, the DOT-HF, OptiLink and REM-HF trials investigated the use of remote monitoring using ICD/CRT devices, but all failed to show a reduction in HF-related hospitalisation rates [14–16]. The DOT-HF trial even showed an increase in the number of HF hospitalisations in the remotely monitored groups[16]. The EFFECT study, a multi-centrer clinical trial, showed that remote monitoring of ICD in HF patients reduced mortality and cardiovascular hospitalisations [17], and the COMMIT-HF trial showed that remote monitoring of ICD/CRT HF patients significantly reduces long-term mortality but not HF-related hospitalisations [18].
Table 2.
Trial/study | Author; journal; year | No. of patients | Parameter | Endpoint | Impact on HF hospitalisation |
---|---|---|---|---|---|
DOT-HF [16] |
Van Veldhuisen et al.; Circulation; 2011 |
335 | Intrathoracic impedance | HF hospitalisation | Increased |
OptiLink [14] |
Brachmann et al.; Eur J Heart Fail; 2011 |
1002 | Intrathoracic impedance | HF hospitalisation | Non-significant |
EFFECT [17] |
De Simone et al.; Europace; 2015 |
987 | Remote monitoring via ICD, or CRT | HF hospitalisation |
Reduced (IRR 0.54; 95% CI 0.24–0.62) |
MORE-CARE [13] |
Boriani et al.; Eur J Heart Fail; 2017 |
865 | Remote monitoring of advanced diagnostics via CRT‑D | HF hospitalisation | Non-significant |
REM-HF [15] |
Morgan et al.; Eur Heart J; 2017 |
1650 | Remote monitoring via ICD, or CRT | HF hospitalisation | Non-significant |
COMMIT-HF [18] |
Kurek et al.; J Cardiovasc Electrophysiol; 2017 |
574 | Remote monitoring via ICD, or CRT |
HF hospitalisation/ All-cause mortality |
Non-significant/ Reduced all-cause mortality (HR 0.24; 95% CI 0.14–0.41) |
IN-TIME [19] |
Hindricks et al; Lancet; 2014 |
716 | Remote monitoring via ICD, or CRT | HF worsening score | OR 0.63 95% Ci 0.43–0.90 |
Other patient outcomes have been investigated as well, with mixed results. The IN-TIME RCT showed that using the remote monitoring abilities of the ICD and CRT devices leads to a reduction of a combined endpoint of all-cause death, overnight HF-related hospitalisation, change in New York Heart Association (NYHA) class, and change in patient global self-assessment [19]. However, other trials found no significant effect on patient outcomes [20, 21]. The effect of remote monitoring using ICD/CRT devices has recently been investigated in a meta-analysis, including 11 RCTs (5702 patients). This meta-analysis showed a reduction in the number of outpatient visits in remotely monitored patients, although remote monitoring with an ICD/CRT device had no effect on mortality or HF-related hospitalisations rates in these patients [22].
The MultiSENSE algorithm aims to predict the individualised risk for worsening of HF based on first and third heart sounds, thoracic impedance, respiration rate, the ratio of respiration rate to tidal volume, heart rate and patient activity. This could aid in the timely detection of HF worsening with the threshold retrospectively calculated by the algorithm. However, the overall sensitivity is only 70% [23]. Another algorithm with a similar aim is the HeartLogic algorithm [24]. To date, no clinical endpoint data or trial data are available and the technique is limited to certain ICD types and brands only.
The shift in remote HF care: haemodynamic (invasive) remote monitoring
In HF patients cardiac filling pressures rise weeks before an exacerbation of HF leading to a related hospitalisation. Symptoms of clinical congestion such as gain in body weight will occur about 2 weeks later, usually shortly before hospitalisation (Fig. 1; [25]). Monitoring of cardiac filling pressures can be an effective strategy to detect upcoming HF decompensation, as it might provide a window of opportunity to intervene adequately and promptly, which is not possible with previous remote monitoring strategies. Therefore multiple implantable haemodynamic monitoring devices have been developed over the last few years. The ePAD (Medtronic, Dublin, Irland) device, an estimate pulmonary artery (PA) end-diastolic pressure device, can be implanted in the right ventricle and has been investigated in the COMPASS-HF trial. In this trial, NYHA class III/IV chronic HF patients were included and investigated as to whether remote haemodynamic monitoring using the ePAD could reduce HF-related hospitalisation, emergency or urgent care visits requiring intravenous therapy. This study did not find significant differences in its endpoint, although it was underpowered due to a lower inclusion rate. Furthermore, clinicians did not receive guidance on how to react to pressure changes [26].
Left atrium pressures (LAP) can be directly measured using a LAP device. The tip of this device is implanted transvenously into the atrial septum oriented towards the left atrium, enabling remote LAP monitoring. This device was used only in the LAPTOP-HF trial, which aimed to investigate the safety and effectiveness of this sensor. However, the enrollment was stopped early due to a perceived excess of procedure-related complications. This is an important issue as the procedure needs an interatrial septum puncture and is placed in the left side of the heart with the risk of arterial side complications. However, in the patients already included in this trial, and followed for 12 months, a 41% reduction of HF-related hospitalisations was observed in the patients with a LAP device [27]. Currently, the V‑LAP™ Left Atrium Monitoring systEm for Patients With Chronic sysTOlic and Diastolic Congestive heaRt Failure (VECTOR-HF) trial is investigating a new LAP device (V-LAP; Vectorious Medical Technologies Ltd., Tel Aviv, Israel) to assess the safety, performance and usability of this device in NYHA class III HF patients (NCT03775161).
Off all the remote monitoring strategies currently available, remote haemodynamic monitoring using the CardioMEMS HF system device (Abbott Inc., Atlanta, GA, USA) (Fig. 2) appears to be the most promising with respect to safety, durability and ability to prevent HF-related hospitalisations. The CardioMEMS is implanted into the PA and enables daily pulmonary artery pressure (PAP) readings. Treating physicians can react to these changes in PA trend data to maintain normal PAP levels, as a sign of a stable clinical status. Furthermore, these daily PAP readings can be used as a feedback mechanism after treatment changes, providing feedback on whether the treatment changes led to a sufficient decline of PAPs. These strategies can lead to individualised HF therapy.
The CardioMEMS consists of a coil combined with a pressure-sensitive capacitor sealed in a capsule, forming an electrical circuit that resonates at a specific frequency when it is electromagnetically coupled with an external antenna [28]. This antenna provides the power for the device, so the device is completely free from batteries or leads. At both sides of the capsule, a loop is placed to ensure that the CardioMEMS remains at the implanted position until the endothelialisation is complete, approximately 3–4 weeks after implantation. When pressure is applied, the resonant frequency changes via a characteristic pattern and is received by the external antenna. The antenna converts this signal into a pressure waveform and sends it to a secure website, where it can be monitored. The device is implanted during a right heart catheterisation, with access via the femoral vein. An appropriate target location, based on vessel size and location, is identified on a pulmonary arteriogram. The CardioMEMS delivery system is advanced to the target location over a guidewire, where the CardioMEMS is released. After implantation, the device is calibrated using PAP obtained with a Swan Ganz catheter.
Two studies have validated the PAP measured by the CardioMEMS, with cardiac filling pressures measured by Swan-Ganz catheterisation or echocardiography directly after implantation and after 6 months of follow-up [29, 30]. Swan-Ganz measurements showed a good correlation with mean PAP assessed by CardioMEMS (r2 = 0.90 at implantation and r2 = 0.94 at follow-up, p < 0.01) [30]. Furthermore, a good correlation (r2 = 0.80 at implantation and r2 = 0.75, both p < 0.01 at follow-up) was found between PAP measurements by the CardioMEMS and estimated pressure measurements by echocardiography [29].
Safety
The safety of the CardioMEMS has been investigated in the CHAMPION trial. A total of 15 serious adverse events occurred during all implantation attempts in the CHAMPION trial [31]. In total, 1% (n = 8) of patients developed a device-related adverse event, and 1% (n = 7) developed a procedure-related adverse event. The following events were reported: four bleeding events, three anticoagulation-related hospitalisations, two pre-existing atrial dysrhythmia exacerbations during implantation, two febrile illnesses, one pulmonary in situ thrombus during implantation that was treated with anticoagulation, one cardiogenic shock, one case of atypical chest pain, and one delivery-system failure requiring a snare to remove the delivery system [32]. An analysis of the post-marketing data of more than 5500 CardioMEMS implantations showed that 2.8% of all CardioMEMS patients experienced an adverse event [33]. Most adverse events were a recalibration of the system (n = 35) or access-site-related bleeding (n = 15). The reported adverse event rates are comparable with those of a standard right heart catheterisation, which is considered a safe procedure [34]. The recent US Post Approval Study (PAS) reported a device- or system-related complication in 0.3% of all patients, and a sensor failure in only 0.1% of all patients, which confirms the safety and durability of this technique.
Clinical efficacy
The CardioMEMS was investigated for the first time in the CHAMPION trial [32]. In this trial, 550 patients with NYHA class III HF and at least one hospitalisation in the last year received a CardioMEMS and were randomised. Of the patients in the intervention group, the haemodynamic information was available to the treating physicians, and the physicians were instructed to react on pressure changes. In the control group, the CardioMEMS readings were not available to the physicians, and these patients received only the standard care. Using the haemodynamic feedback in the intervention group led to a significantly higher number of medication changes, especially diuretics and vasodilator changes, compared to the control group [35]. Furthermore, remote monitoring with the CardioMEMS device led to a significant reduction in mean PAP [32, 36]; similar results were observed in a real-world setting [37].
The effectiveness of the CardioMEMS in preventing HF-related hospitalisations has been investigated in multiple studies ([32, 36, 38, 39]; Fig. 3). During the first 6 months of remote monitoring of HF patients, the HF hospitalisation rates declined by approximately 30% [32, 38] compared with standard care. During the long-term follow-up, the sustained reduction was approximately 33% [31, 38, 39]. Also, all-cause hospitalisation rates were reduced: 45% at 6 months [38] and 16% at 18 months of follow-up [31]. None of these studies were powered to observe mortality differences; however, the CHAMPION trial showed a strong trend towards survival benefit in HF and reduced ejection fraction (HFrEF) patients monitored with the CardioMEMS system (p = 0.06) [40].
The recently presented PAS results confirm the consistent treatment benefit with CardioMEMS in chronic HF patients, reducing the number of HF hospitalisations in a more contemporary setting. The PAS study showed a 58% reduction in HF-related hospitalisation in the first year after CardioMEMS implantation compared with 1 year before implantation. Furthermore, a reduction in HF hospitalisations, mortality and all-cause mortality was observed after CardioMEMS implantation. However, patients included in the PAS study were their own historical controls and there has been no randomised comparison to standard care without PA monitoring.
CardioMEMS and evidence in HFpEF patients
In a real-world setting, remote monitoring using the CardioMEMS leads to a similar reduction in mean PAP in both HFrEF and HFpEF patients [37]. Interestingly, in the CHAMPION trial, a larger reduction of HF-related hospitalisations in HFpEF patients compared with HFrEF patients was observed after at least 6 months of remote monitoring [40, 41]. Besides the alleged benefit of spironolactone in the United States (US) and European participants of the TOPCAT trial [42] with spironolactone, this is the first evidence of a treatment or tool to improve the outcome in HFpEF patients.
Standard care in chronic HF
Recently two large HF registries have been published, the CHAMP-HF registry [43] from the USA and the CHECK-HF Registry [44] from The Netherlands. These two registries showed the differences in standard care between the USA and Western Europe. The prescription rates of RAS inhibitors (82.3% vs 59.9%), beta blockers (80.6% vs 66.8%) and mineralocorticoid receptor antagonist (MRA) (54.8% vs 33.1%) in HFrEF patients were much higher in the CHECK-HF registry compared to the CHAMP-HF registry (Fig. 4a). Furthermore, the prescribed dosages differed between the two registries, with higher prescribed dosages for RAS inhibitors in the CHECK-HF registry and higher dosages for MRA in the CHAMP-HF registry (Fig. 4b) Differences in the HF readmission rates were observed between the USA and Europe [45, 46]. The generalisability of the US findings in terms of clinical effectiveness when using the CardioMEMS device in addition to standard care is therefore limited, and additional costs cannot be directly extrapolated between the two different health care structures. Additional research is needed in the European setting. In Germany, the CardioMEMS European Monitoring Study for Heart Failure (MEMS-HF) study was set up as a post-marketing study to test the safety and clinical effectiveness in a European setting but lacks a control group [47].
Cost-effectiveness of CardioMEMS
The cost-effectiveness of remote monitoring using the CardioMEMS is highly relevant. Using the US CHAMPION trial data the incremental cost-effective ratio (ICER, cost per quality-adjusted life-year) for the US setting has been calculated [32, 48–50]. These studies estimated an increase in the quality-adjusted life-years in the CardioMEMS group of between 0.28 and 0.58, with incremental costs between $4282 and $20,079, compared with standard care patients. This results in an estimated ICER in the USA of between $13,379 and $71,462, which are additional costs in order to gain one quality-adjusted life-year in patients monitored with the CardioMEMS device. Sensitivity analyses demonstrated that the cost-effectiveness of the CardioMEMS is highly influenced by device costs, costs of routine outpatient care, hospitalisation rates, mortality rates and duration of remote monitoring using the CardioMEMS.
There are no patient-level data for cost-effectiveness analyses in Western Europe. With assumptions and estimations based on extrapolating data from the CHAMPION trial and despite the large differences in standard care and financial systems, Cowie et al. [51] calculated the ICER in the European setting, which was approximately between €22,555 (for the Netherlands) and €23,814 (for Germany). However, all these analyses used data on the reduction of HF hospitalisation from the CHAMPION trial and used different estimated mortality rates from population-based cohorts for the cost-effectiveness analyses.
Health care utilisation
Two studies investigated the potential reduction of health care utilisation achieved by using the CardioMEMS [38, 52]. In a real-world Medicare database, 1‑year remote monitoring with the CardioMEMS led to an $11,260 cost reduction for HF hospitalisations compared with 1 year before the CardioMEMS implantation [38]. Based on the effects reported in the CHAMPION trial, and the expected prevalence and hospitalisation costs in Germany, remote monitoring with the CardioMEMS could lead to an overall cost reduction of €106,000,000 in Germany in 2021 [52].
As shown above, remote monitoring of PAP with the CardioMEMS in chronic HF patients leads to more medication changes and a larger reduction of PAP compared with patients receiving standard care, indicating that these patients receive more individualised HF care. In the US, this strategy was effective in reducing the number of HF-related and all-cause hospitalisations. It was suggested that this strategy could improve mortality rates and has been shown to be cost-effective. However, as discussed earlier, some important differences in HF care exist between the USA and Europe.
Recommendation of ESC 2016 guidelines on remote monitoring
The 2016 ESC guidelines report on the lack of consistent evidence for non-haemodynamic telemonitoring or remote monitoring in HF patients. The guidelines state that remote monitoring may be considered in selected patients to improve HF outcome with individual approaches such as CardioMEMS to reduce the risk of HF admissions and multi-parameter monitoring with ICD (in-time approach) to improve outcome in HFrEF patients with a level IIb class B recommendation [53].
Conclusion
In recent years, many remote monitoring strategies have been developed, and development continues at a rapid rate. Non-invasive remote monitoring of symptoms and signs, as well as weight, has not been proven to be effective in improving outcome measurements. Also, the monitoring of biomarkers or thoracic impedance has not been shown to be beneficial. Invasive or haemodynamic measures of remote monitoring are developed with right-sided (CardioMEMS) and left-sided (LA devices) sensors. The LAPTOP-HF trial with LA devices was stopped early for safety reasons. The CardioMEMS is the most promising (invasive) remote monitoring tool currently available. The haemodynamic information allows for a window of timely and adequate intervention based upon raised PAP, preventing an upcoming HF decompensation. Additionally, its safety and durability have been tested and confirmed in post-marketing studies. However, important information on the effect on the quality of life and cost-effectiveness is still lacking in a Western European setting, which is currently being investigated in the MONITOR-HF study.
References
- 1.Farre N, Vela E, Cleries M, et al. Real world heart failure epidemiology and outcome: a population-based analysis of 88,195 patients. PLoS ONE. 2017;12(2):e0172745. doi: 10.1371/journal.pone.0172745. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Inglis SC, Clark RA, Dierckx R, Prieto-Merino D, Cleland JG. Structured telephone support or non-invasive telemonitoring for patients with heart failure. Cochrane Database Syst Rev. 2015 doi: 10.1002/14651858.CD007228.pub3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Angermann CE, Stork S, Gelbrich G, et al. Mode of action and effects of standardized collaborative disease management on mortality and morbidity in patients with systolic heart failure: the Interdisciplinary Network for Heart Failure (INH) study. Circ Heart Fail. 2012;5(1):25–35. doi: 10.1161/CIRCHEARTFAILURE.111.962969. [DOI] [PubMed] [Google Scholar]
- 4.Chaudhry SI, Mattera JA, Curtis JP, et al. Telemonitoring in patients with heart failure. N Engl J Med. 2010;363(24):2301–2309. doi: 10.1056/NEJMoa1010029. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Cleland JG, Louis AA, Rigby AS, et al. Noninvasive home telemonitoring for patients with heart failure at high risk of recurrent admission and death: the Trans-European Network-Home-Care Management System (TEN-HMS) study. J Am Coll Cardiol. 2005;45(10):1654–1664. doi: 10.1016/j.jacc.2005.01.050. [DOI] [PubMed] [Google Scholar]
- 6.Koehler F, Koehler K, Deckwart O, et al. Efficacy of telemedical interventional management in patients with heart failure (TIM-HF2): a randomised, controlled, parallel-group, unmasked trial. Lancet. 2018;392(10152):1047–1057. doi: 10.1016/S0140-6736(18)31880-4. [DOI] [PubMed] [Google Scholar]
- 7.Koehler F, Winkler S, Schieber M, et al. Impact of remote telemedical management on mortality and hospitalizations in ambulatory patients with chronic heart failure: the telemedical interventional monitoring in heart failure study. Circulation. 2011;123(17):1873–1880. doi: 10.1161/CIRCULATIONAHA.111.018473. [DOI] [PubMed] [Google Scholar]
- 8.Koehler F, Winkler S, Schieber M, et al. Telemedical Interventional Monitoring in Heart Failure (TIM-HF), a randomized, controlled intervention trial investigating the impact of telemedicine on mortality in ambulatory patients with heart failure: study design. Eur J Heart Fail. 2010;12(12):1354–1362. doi: 10.1093/eurjhf/hfq199. [DOI] [PubMed] [Google Scholar]
- 9.Krum H, Forbes A, Yallop J, et al. Telephone support to rural and remote patients with heart failure: the Chronic Heart Failure Assessment by Telephone (CHAT) study. Cardiovasc Ther. 2013;31(4):230–237. doi: 10.1111/1755-5922.12009. [DOI] [PubMed] [Google Scholar]
- 10.Lynga P, Persson H, Hagg-Martinell A, et al. Weight monitoring in patients with severe heart failure (WISH). A randomized controlled trial. Eur J Heart Fail. 2012;14(4):438–444. doi: 10.1093/eurjhf/hfs023. [DOI] [PubMed] [Google Scholar]
- 11.Ong MK, Romano PS, Edgington S, et al. Effectiveness of remote patient monitoring after discharge of hospitalized patients with heart failure: the Better Effectiveness After Transition—Heart Failure (BEAT-HF) randomized clinical trial. JAMA Intern Med. 2016;176(3):310–318. doi: 10.1001/jamainternmed.2015.7712. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Blum K, Gottlieb SS. The effect of a randomized trial of home telemonitoring on medical costs, 30-day readmissions, mortality, and health-related quality of life in a cohort of community-dwelling heart failure patients. J Card Fail. 2014;20(7):513–521. doi: 10.1016/j.cardfail.2014.04.016. [DOI] [PubMed] [Google Scholar]
- 13.Boriani G, Da Costa A, Quesada A, et al. Effects of remote monitoring on clinical outcomes and use of healthcare resources in heart failure patients with biventricular defibrillators: results of the MORE-CARE multicentre randomized controlled trial. Eur J Heart Fail. 2017;19(3):416–425. doi: 10.1002/ejhf.626. [DOI] [PubMed] [Google Scholar]
- 14.Brachmann J, Bohm M, Rybak K, et al. Fluid status monitoring with a wireless network to reduce cardiovascular-related hospitalizations and mortality in heart failure: rationale and design of the OptiLink HF Study (Optimization of Heart Failure Management using OptiVol Fluid Status Monitoring and CareLink) Eur J Heart Fail. 2011;13(7):796–804. doi: 10.1093/eurjhf/hfr045. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Morgan JM, Kitt S, Gill J, et al. Remote management of heart failure using implantable electronic devices. Eur Heart J. 2017;38(30):2352–2360. doi: 10.1093/eurheartj/ehx227. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.van Veldhuisen DJ, Braunschweig F, Conraads V, et al. Intrathoracic impedance monitoring, audible patient alerts, and outcome in patients with heart failure. Circulation. 2011;124(16):1719–1726. doi: 10.1161/CIRCULATIONAHA.111.043042. [DOI] [PubMed] [Google Scholar]
- 17.De Simone A, Leoni L, Luzi M, et al. Remote monitoring improves outcome after ICD implantation: the clinical efficacy in the management of heart failure (EFFECT) study. Europace. 2015;17(8):1267–1275. doi: 10.1093/europace/euu318. [DOI] [PubMed] [Google Scholar]
- 18.Kurek A, Tajstra M, Gadula-Gacek E, et al. Impact of remote monitoring on long-term prognosis in heart failure patients in a real-world cohort: results from all-comers COMMIT-HF trial. J Cardiovasc Electrophysiol. 2017;28(4):425–431. doi: 10.1111/jce.13174. [DOI] [PubMed] [Google Scholar]
- 19.Hindricks G, Taborsky M, Glikson M, et al. Implant-based multiparameter telemonitoring of patients with heart failure (IN-TIME): a randomised controlled trial. Lancet. 2014;384(9943):583–590. doi: 10.1016/S0140-6736(14)61176-4. [DOI] [PubMed] [Google Scholar]
- 20.Bohm M, Drexler H, Oswald H, et al. Fluid status telemedicine alerts for heart failure: a randomized controlled trial. Eur Heart J. 2016;37(41):3154–3163. doi: 10.1093/eurheartj/ehw099. [DOI] [PubMed] [Google Scholar]
- 21.Buchta P, Tajstra M, Kurek A, et al. The impact of remote monitoring of implanted cardioverter-defibrillator (ICD) and cardiac resynchronisation therapy device (CRT-D) patients on healthcare costs in the Silesian population: three-year follow-up. Kardiol Pol. 2017;75(6):573–580. doi: 10.5603/KP.a2017.0019. [DOI] [PubMed] [Google Scholar]
- 22.Klersy C, Boriani G, De Silvestri A, et al. Effect of telemonitoring of cardiac implantable electronic devices on healthcare utilization: a meta-analysis of randomized controlled trials in patients with heart failure. Eur J Heart Fail. 2016;18(2):195–204. doi: 10.1002/ejhf.470. [DOI] [PubMed] [Google Scholar]
- 23.Boehmer JP, Hariharan R, Devecchi FG, et al. A multisensor algorithm predicts heart failure events in patients with implanted devices: results from the MultiSENSE study. Jacc Heart Fail. 2017;5(3):216–225. doi: 10.1016/j.jchf.2016.12.011. [DOI] [PubMed] [Google Scholar]
- 24.Gardner RS, Singh JP, Stancak B, et al. HeartLogic multisensor algorithm identifies patients during periods of significantly increased risk of heart failure events: results from the MultiSENSE study. Circ Heart Fail. 2018;11(7):e004669. doi: 10.1161/CIRCHEARTFAILURE.117.004669. [DOI] [PubMed] [Google Scholar]
- 25.Zile MR, Bennett TD, Sutton StJM, et al. Transition from chronic compensated to acute decompensated heart failure: pathophysiological insights obtained from continuous monitoring of intracardiac pressures. Circulation. 2008;118(14):1433–1441. doi: 10.1161/CIRCULATIONAHA.108.783910. [DOI] [PubMed] [Google Scholar]
- 26.Bourge RC, Abraham WT, Adamson PB, et al. Randomized controlled trial of an implantable continuous hemodynamic monitor in patients with advanced heart failure: the COMPASS-HF study. J Am Coll Cardiol. 2008;51(11):1073–1079. doi: 10.1016/j.jacc.2007.10.061. [DOI] [PubMed] [Google Scholar]
- 27.Abraham WTAPB, Costanzo MR, Eigel N, Gold M, Klapholz M, Maurer M, Saxon L, Singh J, Troughton R. Hemodynamic monitoring in advanced heart failure: results from the LAPTOP-HF trial. J Card Fail. 2016;22(11):940. [Google Scholar]
- 28.Adamson PB, Abraham WT, Aaron M, et al. CHAMPION trial rationale and design: the long-term safety and clinical efficacy of a wireless pulmonary artery pressure monitoring system. J Card Fail. 2011;17(1):3–10. doi: 10.1016/j.cardfail.2010.08.002. [DOI] [PubMed] [Google Scholar]
- 29.Tolia S, Khan Z, Gholkar G, Zughaib M. Validating left ventricular filling pressure measurements in patients with congestive heart failure: CardioMEMS pulmonary arterial diastolic pressure versus left atrial pressure measurement by transthoracic echocardiography. Cardiol Res Pract. 2018;2018:8568356. doi: 10.1155/2018/8568356. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Verdejo HE, Castro PF, Concepcion R, et al. Comparison of a radiofrequency-based wireless pressure sensor to swan-ganz catheter and echocardiography for ambulatory assessment of pulmonary artery pressure in heart failure. J Am Coll Cardiol. 2007;50(25):2375–2382. doi: 10.1016/j.jacc.2007.06.061. [DOI] [PubMed] [Google Scholar]
- 31.Abraham WT, Stevenson LW, Bourge RC, et al. Sustained efficacy of pulmonary artery pressure to guide adjustment of chronic heart failure therapy: complete follow-up results from the CHAMPION randomised trial. Lancet. 2016;387(10017):453–461. doi: 10.1016/S0140-6736(15)00723-0. [DOI] [PubMed] [Google Scholar]
- 32.Abraham WT, Adamson PB, Bourge RC, et al. Wireless pulmonary artery haemodynamic monitoring in chronic heart failure: a randomised controlled trial. Lancet. 2011;377(9766):658–666. doi: 10.1016/S0140-6736(11)60101-3. [DOI] [PubMed] [Google Scholar]
- 33.Vaduganathan M, DeFilippis EM, Fonarow GC, Butler J, Mehra MR. Postmarketing adverse events related to the CardioMEMS HF system. JAMA Cardiol. 2017;2(11):1277–1279. doi: 10.1001/jamacardio.2017.3791. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Hoeper MM, Lee SH, Voswinckel R, et al. Complications of right heart catheterization procedures in patients with pulmonary hypertension in experienced centers. J Am Coll Cardiol. 2006;48(12):2546–2552. doi: 10.1016/j.jacc.2006.07.061. [DOI] [PubMed] [Google Scholar]
- 35.Costanzo MR, Stevenson LW, Adamson PB, et al. Interventions linked to decreased heart failure hospitalizations during ambulatory pulmonary artery pressure monitoring. Jacc Heart Fail. 2016;4(5):333–344. doi: 10.1016/j.jchf.2015.11.011. [DOI] [PubMed] [Google Scholar]
- 36.Adamson PB, Abraham WT, Stevenson LW, et al. Pulmonary artery pressure-guided heart failure management reduces 30-day readmissions. Circ Heart Fail. 2016;9(6):e002600. doi: 10.1161/CIRCHEARTFAILURE.115.002600. [DOI] [PubMed] [Google Scholar]
- 37.Heywood JT, Jermyn R, Shavelle D, et al. Impact of practice-based management of pulmonary artery pressures in 2000 patients implanted with the CardioMEMS sensor. Circulation. 2017;135(16):1509–1517. doi: 10.1161/CIRCULATIONAHA.116.026184. [DOI] [PubMed] [Google Scholar]
- 38.Desai AS, Bhimaraj A, Bharmi R, et al. Ambulatory hemodynamic monitoring reduces heart failure hospitalizations in “real-world” clinical practice. J Am Coll Cardiol. 2017;69(19):2357–2365. doi: 10.1016/j.jacc.2017.03.009. [DOI] [PubMed] [Google Scholar]
- 39.Jermyn R, Alam A, Kvasic J, Saeed O, Jorde U. Hemodynamic-guided heart-failure management using a wireless implantable sensor: infrastructure, methods, and results in a community heart failure disease-management program. Clin Cardiol. 2017;40(3):170–176. doi: 10.1002/clc.22643. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Givertz MM, Stevenson LW, Costanzo MR, et al. Pulmonary artery pressure-guided management of patients with heart failure and reduced ejection fraction. J Am Coll Cardiol. 2017;70(15):1875–1886. doi: 10.1016/j.jacc.2017.08.010. [DOI] [PubMed] [Google Scholar]
- 41.Adamson PB, Abraham WT, Bourge RC, et al. Wireless pulmonary artery pressure monitoring guides management to reduce decompensation in heart failure with preserved ejection fraction. Circ Heart Fail. 2014;7(6):935–944. doi: 10.1161/CIRCHEARTFAILURE.113.001229. [DOI] [PubMed] [Google Scholar]
- 42.Pfeffer MA, Claggett B, Assmann SF, et al. Regional variation in patients and outcomes in the Treatment of Preserved Cardiac Function Heart Failure With an Aldosterone Antagonist (TOPCAT) trial. Circulation. 2015;131(1):34–42. doi: 10.1161/CIRCULATIONAHA.114.013255. [DOI] [PubMed] [Google Scholar]
- 43.Greene SJ, Butler J, Albert NM, et al. Medical therapy for heart failure with reduced ejection fraction: the CHAMP-HF registry. J Am Coll Cardiol. 2018;72(4):351–366. doi: 10.1016/j.jacc.2018.04.070. [DOI] [PubMed] [Google Scholar]
- 44.Brunner-La Rocca HP, Linssen GC, Smeele FJ, et al. Contemporary drug treatment of chronic heart failure with reduced ejection fraction: the CHECK-HF registry. Jacc Heart Fail. 2019;7(1):13–21. doi: 10.1016/j.jchf.2018.10.010. [DOI] [PubMed] [Google Scholar]
- 45.Cheng RK, Cox M, Neely ML, et al. Outcomes in patients with heart failure with preserved, borderline, and reduced ejection fraction in the Medicare population. Am Heart J. 2014;168(5):721–730. doi: 10.1016/j.ahj.2014.07.008. [DOI] [PubMed] [Google Scholar]
- 46.Crespo-Leiro MG, Anker SD, Maggioni AP, et al European Society of Cardiology Heart Failure Long et al. European Society of Cardiology Heart Failure Long-Term Registry (ESC-HF-LT): 1-year follow-up outcomes and differences across regions. Eur J Heart Fail. 2016;18(6):613–625. doi: 10.1002/ejhf.566. [DOI] [PubMed] [Google Scholar]
- 47.Angermann CE, Assmus B, Anker SD, et al. Safety and feasibility of pulmonary artery pressure-guided heart failure therapy: rationale and design of the prospective CardioMEMS Monitoring Study for Heart Failure (MEMS-HF) Clin Res Cardiol. 2018;107(11):991–1002. doi: 10.1007/s00392-018-1281-8. [DOI] [PubMed] [Google Scholar]
- 48.Ollendorf DA, Sandhu AT, Pearson SD. CardioMEMS HF for the management of heart failure-effectiveness and value. JAMA Intern Med. 2016;176(10):1551–1552. doi: 10.1001/jamainternmed.2016.5467. [DOI] [PubMed] [Google Scholar]
- 49.Sandhu AT, Goldhaber-Fiebert JD, Owens DK, et al. Cost-effectiveness of implantable pulmonary artery pressure monitoring in chronic heart failure. Jacc Heart Fail. 2016;4(5):368–375. doi: 10.1016/j.jchf.2015.12.015. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Schmier JK, Ong KL, Fonarow GC. Cost-effectiveness of remote cardiac monitoring with the CardioMEMS heart failure system. Clin Cardiol. 2017;40(7):430–436. doi: 10.1002/clc.22696. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 51.Cowie MR, Simon M, Klein L, Thokala P. The cost-effectiveness of real-time pulmonary artery pressure monitoring in heart failure patients: a European perspective. Eur J Heart Fail. 2017;19(5):661–669. doi: 10.1002/ejhf.747. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Kolominsky-Rabas PL, Kriza C, Djanatliev A, et al. Health economic impact of a pulmonary artery pressure sensor for heart failure telemonitoring: a dynamic simulation. Telemed J E Health. 2016;22(10):798–808. doi: 10.1089/tmj.2015.0226. [DOI] [PubMed] [Google Scholar]
- 53.Ponikowski P, Voors AA, Anker SD, et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failure: the Task Force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC). Developed with the special contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J. 2016;37(27):2129–2200. doi: 10.1093/eurheartj/ehw128. [DOI] [PubMed] [Google Scholar]
- 54.Emani S. Remote monitoring to reduce heart failure readmissions. Curr Heart Fail Rep. 2017;14(1):40–47. doi: 10.1007/s11897-017-0315-2. [DOI] [PubMed] [Google Scholar]