Value of liver iron concentration in healthy volunteers assessed by MRI

Marzanna Obrzut 1,2, Vitaliy Atamaniuk 1,, Kevin J Glaser 2, Jun Chen 2, Richard L Ehman 2, Bogdan Obrzut 3, Marian Cholewa 1, Krzysztof Gutkowski 4
PMCID: PMC7577999  PMID: 33087836

Abstract

Iron overload is a relatively common clinical condition resulting from disorders such as hereditary hemochromatosis, thalassemia, sickle cell disease, and myelodysplasia that can lead to progressive fibrosis and eventually cirrhosis of the liver. Therefore, it is essential to recognize the disease process at the earliest stage. Liver biopsy is the reference test for the assessment of liver fibrosis. It also allows for quantifying liver iron concentration (LIC) in patients. However, this is an invasive method with significant limitations and possible risks. Magnetic resonance imaging (MRI) and evaluation of the R2* relaxation rate can be an alternative to biopsy for assessing LIC. However, it causes a need for accurate R2* data corresponding to standard value for further comparison with examined patients. This study aimed to assess the normative values of liver R2* in healthy individuals. A total of 100 volunteers that met established criteria were enrolled in the study: 36 (36%) men and 64 (64%) women. The mean age was 22.9 years (range 20 to 32 years). R2* was estimated by an MRI exam with a 1.5 T clinical magnetic resonance scanner. Images for measuring the LIC and liver fat concentration were obtained using the IDEAL-IQ technique for liver imaging. The Mean (SD) liver R2* was 28.34 (2.25) s−1 (95% CI, 27.78–28.90, range 23.67–33.00 s−1) in females, 29.57 (3.20) s−1 (95% CI, 28.49–30.66, range 23.93–37.77 s−1) in males, and 28.72 (2.69) s−1 (range 23.67–37.77 s−1) in the whole group. R2* value in this particular population with a high proportion of young women did not exceed 38 s−1. In the absence of fibrosis or steatosis, liver stiffness and fat fraction did not show any relationship with R2*.

Subject terms: Liver, Magnetic resonance imaging

Introduction

Iron overload is a relatively common clinical condition, resulting from disorders such as hereditary hemochromatosis, thalassemia, sickle cell disease, and myelodysplasia1. Iron accumulation can result from excessive iron supply or absorption. The most common cause of excessive iron supply is secondary to red blood cell (RBC) transfusions used to treat chronic anemia (e.g., thalassemia, sickle cell disease, inherited bone marrow failure syndrome, and myelodysplastic syndrome). A less common cause is the use of supplements or therapeutic infusions rich in iron, such as hemin used to treat some types of porphyria.

The most common cause of increased iron absorption in adults is congenital hemochromatosis caused by gene mutation, inefficient erythropoiesis in the course of thalassemia, sideroblastic and other hereditary anemias, and liver diseases (in particular, acute liver disease (ALD) and chronic diseases like nonalcoholic fatty liver disease (NAFLD) and chronic viral hepatitis). Usually, patients have more than one reason for iron overload. The mechanism for iron absorption in the liver is not fully understood and is most likely associated with a decrease in hepcidin production in the diseased liver2.

Irrespective of the causes of iron overload, the liver is the organ that is most affected by iron overload. In the case of increased intestinal iron absorption due to the accumulation and damage of hepatocytes, this can lead to progressive fibrosis and eventually cirrhosis of the liver, resulting in portal hypertension, liver failure, and hepatocellular carcinoma3,4. Therefore, it is extremely important to recognize the disease process in as early a stage as possible. Liver biopsy is considered to be the definitive test for the assessment of liver fibrosis. In the setting of iron overload, a biopsy can be used to quantify liver iron concentration (LIC)5. However, this is an invasive method with substantial sampling error. A significant limitation of this diagnostic method is the fact, that the distribution of iron in the liver is inherently very heterogeneous6. Besides, this method carries a risk of complications, such as postinterventional hemorrhage and bleeding7,8.

Due to the known systematic and accidental errors, costs, and invasiveness, a biopsy cannot be considered a gold standard for LIC assessment and should not be used for diagnostics except for disease processes requiring histological assessment of the organ to establish a diagnosis and correct treatment6.

Magnetic resonance imaging (MRI) techniques have been advocated as an alternative to biopsy for assessing LIC. One approach is to measure the R2* relaxation rate of liver tissue by using a multipoint Dixon sequence such as IDEAL-IQ9. Studies have shown that R2* increases systematically with hepatic iron concentration10. This parameter has a high correlation with biopsy measurements and could be a reliable, noninvasive method for measuring LIC10. For MRI to be used for the rapid routine assessment of LIC, it is necessary to establish ranges of R2* values for the healthy population. To our knowledge, there is little data in the literature assessing the liver iron level in healthy adult volunteers11. Published studies have qualified the absence of liver disease in cohorts by the participant's declaration of the absence of liver disease without the confirmation of this fact, even using basic laboratory tests. There is a need to obtain objective R2*data based on representative groups of healthy subjects who undergo more tests to confirm their health status, especially to confirm their lack of liver inflammation and fibrosis.

Objectives

To assess the normative values of liver R2* measured by the IDEAL-IQ MRI sequence in healthy volunteers.

Study enrollment

115 subjects were primarily enrolled in the study. The inclusion criteria were as follows: no contraindications to MRI, no history of liver disease, no risk factors for chronic liver disease, negative family history of chronic liver disease, normal liver enzyme levels, normal ferritin (13–150 ng/ml for women, 30–400 ng/ml for men) and iron values (37–170 mcg/dl for women, 49–181 mcg/dl for men), no drug or medication use of any kind and absence of liver fibrosis or steatosis on MRI. As exclusion criteria were used following values of parameters: liver stiffness < 2.9 kPa and fat fraction < 5%12. All subjects declared they consumed less than 30 g of alcohol per day. The participants were not required to change their normal diet to be included in this study but required 6 h of fasting before the MRI scans. The research was carried out in the magnetic resonance laboratory of the Center for Medical and Natural Sciences Research and Innovation of the University of Rzeszow, Poland in the period from December 2018 until July 2019.

Study protocol

Subjects were examined on a 1.5-T whole-body OPTIMA MR360 Advance MR scanner (GE Healthcare, Milwaukee, Wisconsin, United States) with software version SV23. This device is equipped with a gradient system with a maximum amplitude of 33 mT/m and a 120 mT/m/ms slew rate. For liver MRI and MRE examinations, all patients underwent the scan in the feet-first supine position with an 8-channel torso phased-array coil around the abdomen. As part of the MRE setup, a drum-like passive liver driver was secured to the right side of the chest wall with its center at the superior-inferior (SI) level of the xiphoid process. There was a 29-foot long PVC tube (3/4-inch diameter) connecting the passive driver to an active driver (the source of the vibrations used during MRE acquisitions) located in the adjacent equipment room.

Images for measuring the LIC and liver fat concentration were obtained using the IDEAL-IQ technique for liver imaging (GE Healthcare, Waukesha, WI, USA) and was performed using a gradient-echo multi-echo MR sequence with the following imaging parameters: FOV = 39 cm, TR = 12.4 ms, acquisition matrix = 160 × 160, slice thickness = 10 mm, slice number = 28. Six echo times were collected and R2* and fat-fraction images were generated automatically on the scanner. The acquisition was performed in a single breath-hold of 17 s.

Liver stiffness images were obtained using the MR-Touch MRE technique (GE Healthcare, Waukesha, WI, USA) using a 2D gradient-echo pulse sequence with a 60-Hz motion frequency. The imaging parameters were as follows: FOV = 40 cm, TR = 33.3 ms, TE = 20.5 ms, acquisition matrix = 224 × 64, parallel imaging factor of 2, reconstructed matrix size = 256 × 256, slice thickness = 10 mm, flip angle = 30°, number of time offsets = 4, time of acquisition = 1 min 8 s. Nine slices were obtained through the liver in three or four breath holds, and the subjects held their breath at the end of expiration. Images of the liver stiffness (elastograms) were generated automatically on the scanner and were transferred offline for additional analysis and measurements.

Image analysis

To measure liver R2* and fat concentration, four irregular polygon regions of interest (ROIs) were defined in the liver on four consecutive R2* images and were applied to the fat content images as well. ROIs were drawn to be as large as possible while excluding large vessels and the central biliary tree. Besides, the position of the ROIs was chosen in areas far from the organ boundaries. Liver R2* was calculated as the mean of the four R2* values and was used to estimate LIC according to Wood formula ([Fe] mg/g = R2* × 0.0254 + 0.202)6. Since not all our patients had saved individual eight echoes in the clinic scans for calculating the offline classic R2* results by using the magnitude-based monoexponential model6, we used the scanner-generated the complex-based IDEAL-IQ R2* results. A newly published study13 found the classic R2* and IDEAL-IQ R2* were significantly correlated to each other (r = 0.9231, p < 0.0001), and the patient R2* range covered our patients R2*.

Liver stiffness was measured by placing a total of four irregular polygon ROIs in the right hepatic lobe on the elastograms, with large vessels and regions of low confidence excluded. Liver stiffness was calculated as the mean of the mean liver stiffness values obtained from each of the four middle slices, weighted by the ROI area.

Statistical analysis

Liver stiffness and liver R2* values were presented as mean (SD). To assess the correlation of liver R2* with liver stiffness, liver fat fraction, and ferritin level, the Pearson correlation test was performed. The Mann–Whitney U test was used to identify differences between male and female volunteers. Statistical significance was assumed at a p value of less than 0.05. MedCalc software (version 18.9.1; Mariakerke, Belgium) was used to perform the statistical analysis.

Ethics

The study protocol was approved by the Ethics Committee of the Medical Department of Rzeszow University (Resolution No. 8/10/2016) and conformed to the ethical guidelines of the 1975 Declaration of Helsinki (6th revision, 2008). After receiving an explanation of the study, each subject provided written informed consent to enroll in the study.

Results

Of the 115 volunteers, 15 were excluded because of elevated (abnormal) values of aspartate aminotransferase (AspAT) and/or alanine aminotransferase (AlAT). The remaining 100 were considered eligible and were included in the study. All MRE exams were successfully completed. None of the volunteers reported any discomfort during the procedure. Of these 100 subjects, 36 (36%) were male and 64 (64%) were female and the mean age was 22.9 years (range 20 to 32 years). The mean BMI for whole group was 21.33 kg/m2 (SD 2.19 kg/m2, range: 16.85–25.76 kg/m2. For female volunteers the mean BMI was 20.64 kg/m2 (SD: 1.96 kg/m2, range: 16.85–25.59 kg/m2. For men, the mean BMI was 22.56 kg/m2 (SD: 2.05 kg/m2, range: 17.96–25.76 kg/m2.

To measure liver R2*, the average size of the ROIs was 5 to 7 cm2 for the liver. The mean (SD) liver R2* for the population was 28.72 (2.69) s−1 (95% CI, 28.26–29.32, range 23.67–37.77 s−1) (Fig. 1). The hepatic iron load was assessed using the Wood formula ([Fe] mg/g = R2* × 0.0254 + 0.202). This equates to a mean LIC of 0.93 mg/g (range 0.80–1.16 mg/g)6.

Figure 1.

Figure 1

Liver R2* values in normal volunteers. Box-and-whisker plot graph showing the distribution of the liver R2* values in the 100 normal healthy volunteers. Dots represent the value of each volunteer.

Mean (SD) liver R2* for the females was 28.34 (2.25) s−1 (95% CI, 27.78–28.90, range 23.67–33.00 s−1). For males, the mean was 29.57 (3.20) s−1 (95% CI, 28.49–30.66, range 23.93–37.77 s−1) (Fig. 2). This difference was statistically significant (p = 0.04).

Figure 2.

Figure 2

Liver R2* values separated by sex. Boxplot graph showing the distribution of liver R2* values in the female and male volunteers.

Figure 3 shows an example of an MRE scan. The liver stiffness in healthy volunteers ranged from 1.84 to 2.82 kPa, with a mean (SD) of 2.30 (0.23) kPa (95% CI, 2.25–2.34). Mean (SD) liver stiffness for females was 2.31 (0.23) kPa (95% CI, 2.26–2.37, range 1.84–2.82 kPa) and for males was 2.27 (0.23) kPa (95% CI, 2.19–2.35, range 1.92–2.79 kPa). These values did not differ significantly (p = 0.35). There was no correlation between the liver stiffness and gender (p = 0.19). There was also no correlation between liver R2* and liver stiffness (r = − 0.04, p = 0.68) (Fig. 3).

Figure 3.

Figure 3

An example of MRE scan showing (a) magnitude, (b) elastogram, and (c) wave images.

Liver fat fraction ranged from 1.40 to 4.90%. Mean value (SD) was 2.51 (0.75)% (95% CI, 2.36–2.66%). For males, the mean was 2.69 (0.88) % (95% CI, 2.39–2.99%; range 1.70–4.90%). For females, the mean was 2.41 (0.66) % (95% CI, 2.25–2.58%; range 1.40–4.87%). There was no correlation between the liver fat fraction and gender (p = 0.19). Liver R2* did not correlate with liver fat fraction (r = 0.13, p = 0.20).

Ferritin level ranged from 3.74 to 260.10 ng/ml. Mean value (SD) was 53.29 (52.15) ng/ml (95% CI, 42.94–63.64 ng/ml). For males, the mean was 93.31 (64.56) ng/ml (95% CI, 71.46–115.16 ng/ml; range 6.22–260.10 ng/ml). For females, the mean was 30.78 (22.98) ng/ml (95% CI, 25.04–36.52 ng/ml; range 3.74–113.1 ng/ml). This difference was statistically significant (p < 0.0001).

There was no correlation between liver R2* and ferritin level neither for males: r = 0.28, p = 0.096 (Fig. 4), nor for females: r = − 0.01, p = 0.91 (Fig. 5).

Figure 4.

Figure 4

Relationship between liver stiffness and R2*. Scatterplot diagram showing the relation between liver stiffness and liver R2* in the 100 healthy volunteers.

Figure 5.

Figure 5

Relationship between ferritin level and liver R2* for males. Scatterplot diagram showing the relation between ferritin level and liver R2* in the 36 healthy male volunteers.

Discussion

The normal iron content in the human body is 40 mg/kg in women by menopause and 50 mg/kg in men14,15. About 80% of total iron in the human body occurs in the red blood cells and in proteins other than hemoglobin. The remaining 20% is largely stored in the ferritin and hemosiderin16,17. Ferritin acts as a deposit storing excess iron and occurs primarily in the liver, spleen, and bone marrow17.

A smaller amount of iron is associated with transferrin in plasma18. In women, the amount of stored iron is lower and additionally depends on the severity of menstruation, pregnancy, lactation as well as the amount of iron supplements taken. Premenopausal women lose iron due to menstruation at a rate of about 0.5–1.0 mg/day.

There are no physiological mechanisms to eliminate iron from the body when there is excess2. If the supply of iron from the various sources is too high, then systemic iron overload occurs in the body. Initially, ferritin and hemosiderin (which is a partially denatured form of ferritin) store the excess iron. If the iron-binding capacity of the substances is exceeded, free iron is deposited in the cells of other organs.

Even if the level of ferritin in the blood can be used as a rough indicator of liver iron overload in a highly overloaded population19, it is considered that it does not allow the diagnosis of liver iron overload and should not be used solely for clinical decision making2022.

In the past 20 more years, many MRI methods have been developed to measure hepatic iron concentration, including SIR methods at 0.5 T23, 1.5 T24,25, and 3 T26, R2 calculation at 1.5 T1 and R2* calculations at 1.5 T6,21,2729 and at 3 T30.

Due to whole body iron metabolism31, MRI is the most practical and accurate method to monitor the concentration of iron in various human organs. The literature describes attempts to assess iron content in the body using R2* or T2*magnetization relaxation values (R2* = 1000/T2*). The majority of the studies showed a good correlation between iron concentration in the liver and T2 or T2* in patients with iron-overload diseases in the course of congenital hemochromatosis, thalassemia, and sickle cell disease1,6,19,20,23,25,3237.

The MRI measurements hepatic R2* or T2* is a fast, widely available, and relatively inexpensive technique that some institutions are willing to use as an alternative to liver biopsies. However, for studies of T2*/R2* in healthy individuals, the literatures only reported a small number of participants6,35. Maris et al. measured the T2* value for the liver in 21 healthy volunteers35. In the Wood study, for comparative purposes, data were obtained in a 13-person control group6. The only more notable work assessing the levels of T2* in the liver and spleen in healthy volunteers and comparing these values with the level of ferritin in blood11 is of limited value due to the non-objective criterion of qualification: only a participant health declaration without basic corroborating laboratory tests.

The results from our study fill this existing gap in the literature. In our work, we included healthy volunteers with no history of liver disease and negative blood tests that excluded inflammatory processes in the body. The absence of fibrosis processes in the liver was additionally confirmed by MR elastography exam. The stiffness of the liver was in the range of 1.84–2.82 kPa with an average of 2.30 (0.23) kPa, which was within the normal range38.

In our study, the average value of R2* was 28.751 s−1 (T2* = 34.78 ms). For comparison, Anderson gives a T2* value of 33 ± 7 ms for the liver and 56 ± 22 ms for the spleen27. Maris et al. reported the average T2* for the liver as 24.2 ± 3.0 ms in 21 healthy volunteers35. The recommended upper limit of LIC is < 2 mg Fe/g dry liver tissue39. Another study by Nuttall et al., based on a series of 141 liver biopsies, suggests reference limit 1.8 mg Fe/g dry liver tissue in healthy adults40. The LIC increases with age, an average of 14 mcg Fe/g per year40. As expected, the LIC values in women are lower than in men41. The above-mentioned LIC value of 2 mg/g is equivalent to the R2* value of about 70 s−1 29. The R2* value measured in our healthy volunteers corresponds well with the proposition of classification of iron overload severity offered by Henninger et al.39. In the 13 healthy volunteers (9 male, 4 female) with ages of 29.3 ± 12.3 years (range, 12–50 years) in the Wood’s study, the normal R2* was 39.9 ± 2.8 s−1 (HIC = 1.17 mg/g dry weight)6.

In the presented series, there was no correlation between R2* and liver stiffness, neither for women nor for men. This is consistent with the results of other authors42.

Our study showed a clear difference in the mean liver R2* between men and women. Women had a statistically significantly lower R2*value (28.34 s−1) compared to men (29.57 s−1), p = 0.040. This was expected as premenopausal women lose iron due to menstruation. However, premenopausal women lose iron due to menstruation so our women-dominated population of young healthy volunteers should has a lower iron level than a more general population that includes more elderly people and men.

Also, in this study, we only used the R2* images on the scanner, generated by the manufacturer IDEAL-IQ complex-based fat-water R2* model. There are many other magnitude-based R2* models, for example, Yokoo et al.43 studied the effect of Rician and non-Rician noise on the accuracy of R2* estimation in 7 different magnitude-based models, and found, the agreement of the R2* estimation was excellent in patients with low ferritin but poor in patients with high ferritin, as expected. Tipirneni-Sajja et al.44 simulated 5 different magnitude-based R2* models (GRE-A, GRE-B, GRE-C, UTE-A, and UTE-A), and found, the constant offset monoexponential model overestimated the R2* at the lowest simulated SNR, and the monoexponential fitting with noise subtraction is the most robust and accurate R2* model for ultrashort TE (UTE) sequence.

A limitation of the presented work is the narrow age range of the studied group. Therefore, it was not possible to analyze the correlation of R2* with age. Further studies should include individuals within a wider age range to investigate such potential dependence. Equally interesting would be the analysis of LIC in menstruating and postmenopausal women.

The lack of liver biopsies as an alternative measure of iron content can also be considered a weakness of this study, but the design of the work a priori excluded such a reference point.

By measuring liver R2* during routine MR testing and referring it to the reference values, patients with mild elevation of ferritin blood levels and no iron overload can avoid invasive diagnostics (e.g., liver biopsy) or invasive blood-letting tests. It should be remembered that an increase in ferritin levels might occur in the course of other disease processes such as hepatitis, obesity, or alcohol consumption45.

Also, the MR method used in this study to assess the hepatic iron level has the additional advantage of being able to evaluate the total amount of iron in the whole liver, as opposed to the localized assessment from biopsy, which can cause a significant diagnostic error in the case of an inhomogeneous distribution of iron in the liver46.

In summary, MRI R2* has been developed to accurately measure hepatic iron concentration. Our study has shown a normative hepatic R2* value of less than 38 s−1 in a women-dominated (64%) and relatively large population of healthy young people, which is a valuable reference for the medical community (Fig. 6).

Figure 6.

Figure 6

Relationship between ferritin level and liver R2* for females. Scatterplot diagram showing the relation between ferritin level and liver R2* in the 64 healthy female volunteers.

Author contributions

M.O. initiated and conceived the concept of the study, co-collected the data, co-led the data analysis and interpretation, wrote the first draft of the paper, coordinated draft revisions, and wrote the final manuscript. V.A. co-collected the data, led the statistical analysis, and co-wrote the first draft of the paper. J.Ch. and K.J.G. contributed to the data analysis and interpretation, and draft revisions. R.L.E., B.O., and M.Ch. contributed to data interpretation and draft revisions, K.G. contributed to the study conception, data analysis and interpretation, and draft revisions. All authors read and approved the final version of the manuscript.

Data availability

The data that support the findings of this study are available from the corresponding author upon reasonable request.

Competing interests

The Mayo Clinic, RLE, KJG and JCh have intellectual property rights and a financial interest in magnetic resonance technology. The remaining authors declare no conflict of interest.

Footnotes

Publisher's note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.St Pierre TG, et al. Noninvasive measurement and imaging of liver iron concentrations using proton magnetic resonance. Blood. 2005;105(2):855–861. doi: 10.1182/blood-2004-01-0177. [DOI] [PubMed] [Google Scholar]
  • 2.Bacon, B. R., Kwiatkowski, J. L. Approach to the Patient with Suspected Iron Overload. Post TW, ed. UpToDate. Waltham, MA: UpToDate Inc. https://www.uptodate.com. Accessed on 30 Sept 2019.
  • 3.Ghavamzadeh A, Mirzania M, Kamalian N, Sedighi N, Azimi P. Hepatic iron overload and fibrosis in patients with beta thalassemia major after hematopoietic stem cell transplantation: a pilot study. Int. J. Hematol. Oncol. Stem Cell Res. 2015;9(2):55–59. [PMC free article] [PubMed] [Google Scholar]
  • 4.Arthur MJ. Iron overload and liver fibrosis. J. Gastroenterol. Hepatol. 1996;11(12):1124–1129. doi: 10.1111/j.1440-1746.1996.tb01840.x. [DOI] [PubMed] [Google Scholar]
  • 5.Piperno A. Classification and diagnosis of iron overload. Haematologica. 1998;83:447–455. [PubMed] [Google Scholar]
  • 6.Wood JC, et al. MRI R2 and R2* mapping accurately estimates hepatic iron concentration in transfusion-dependent thalassemia and sickle cell disease patients. Blood. 2005;106:1460–1465. doi: 10.1182/blood-2004-10-3982. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Strassburg CP, Manns MP. Approaches to liver biopsy techniques-revisited. Semin. Liver Dis. 2006;26:318–327. doi: 10.1055/s-2006-951599. [DOI] [PubMed] [Google Scholar]
  • 8.Garcia-Compean D, Cortes C. Transjugular liver biopsy. An update. Ann. Hepatol. 2004;3:100–103. doi: 10.1016/S1665-2681(19)32101-5. [DOI] [PubMed] [Google Scholar]
  • 9.Yu H, et al. Multiecho reconstruction for simultaneous water-fat decomposition and T2* estimation. J. Magn. Reson. Imaging. 2007;26(4):1153–1161. doi: 10.1002/jmri.21090. [DOI] [PubMed] [Google Scholar]
  • 10.Queiroz-Andrade M, et al. MR imaging findings of iron overload. RadioGraphics. 2009;29:1575–1589. doi: 10.1148/rg.296095511. [DOI] [PubMed] [Google Scholar]
  • 11.Schwenzer NF, et al. T2* relaxometry in liver, pancreas, and spleen in a healthy cohort of one hundred twenty-nine subjects-correlation with age, gender, and serum ferritin. Invest. Radiol. 2008;43(12):854–860. doi: 10.1097/RLI.0b013e3181862413. [DOI] [PubMed] [Google Scholar]
  • 12.Gu J, et al. Diagnostic value of MRI-PDFF for hepatic steatosis in patients with non-alcoholic fatty liver disease: a meta-analysis. Eur. Radiol. 2019;29:3564–3573. doi: 10.1007/s00330-019-06072-4. [DOI] [PubMed] [Google Scholar]
  • 13.Rostoker G, et al. Histological scores validate the accuracy of hepatic iron load measured by signal intensity ratio and R2* relaxometry MRI in dialysis patients. J. Clin. Med. 2020;9(1):17. doi: 10.3390/jcm9010017. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 14.Alustiza JM, et al. Iron overload in the liver diagnostic and quantification. Eur. J. Radiol. 2007;61:499–506. doi: 10.1016/j.ejrad.2006.11.012. [DOI] [PubMed] [Google Scholar]
  • 15.Brittenham GM, Badman DG. Noninvasive measurement of iron: report of an NIDDK workshop. Blood. 2003;101:15–19. doi: 10.1182/blood-2002-06-1723. [DOI] [PubMed] [Google Scholar]
  • 16.Siegelman ES, Mitchell DG, Semelka RC. Abdominal iron deposition: metabolism, MR findings, and clinical importance. Radiology. 1996;199:13–22. doi: 10.1148/radiology.199.1.8633135. [DOI] [PubMed] [Google Scholar]
  • 17.Sirlin CB, Reeder SB. Magnetic resonance imaging quantification of liver iron. Magn. Reson. Imaging Clin. N. Am. 2010;18:359–381. doi: 10.1016/j.mric.2010.08.014. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18.Princiotto JV, Zapolski EJ. Difference between the two iron-binding sites of transferrin. Nature. 1975;255:87–88. doi: 10.1038/255087a0. [DOI] [PubMed] [Google Scholar]
  • 19.Alexopoulou E, et al. R2 relaxometry with MRI for the quantification of tissue iron overload in beta-thalassemic patients. J. Magn. Reson. Imaging. 2006;23:163–170. doi: 10.1002/jmri.20489. [DOI] [PubMed] [Google Scholar]
  • 20.Brittenham GM, et al. Hepatic iron stores and plasma ferritin concentration in patients with sickle cell anemia and thalassemia major. Am. J. Hematol. 1993;42:81–85. doi: 10.1002/ajh.2830420116. [DOI] [PubMed] [Google Scholar]
  • 21.Hankins JS, et al. R2* magnetic resonance imaging of the liver in patients with iron overload. Blood. 2009;113(20):4853–4855. doi: 10.1182/blood-2008-12-191643. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22.Karam LB, et al. Liver biopsy results in patients with sickle cell disease on chronic transfusions: poor correlation with ferritin levels. Pediatr. Blood Cancer. 2008;50:62–65. doi: 10.1002/pbc.21215. [DOI] [PubMed] [Google Scholar]
  • 23.Gandon Y, et al. Hemochromatosis: diagnosis and quantification of liver iron with gradient-echo MR imaging. Radiology. 1994;193:533–538. doi: 10.1148/radiology.193.2.7972774. [DOI] [PubMed] [Google Scholar]
  • 24.Gandon Y, et al. Non-invasive assessment of hepatic iron stores by MRI. Lancet. 2004;363(9406):357–362. doi: 10.1016/S0140-6736(04)15436-6. [DOI] [PubMed] [Google Scholar]
  • 25.Alustiza JM, et al. MR quantification of hepatic iron concentration. Radiology. 2004;230:479–484. doi: 10.1148/radiol.2302020820. [DOI] [PubMed] [Google Scholar]
  • 26.Paisant A, et al. Assessment of liver iron overload by 3 T MRI. Abdom. Radiol. 2017;42(6):1713–1720. doi: 10.1007/s00261-017-1077-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 27.Anderson LJ, et al. Cardiovascular T2-star (T2*) magnetic resonance for the early diagnosis of myocardial iron overload. Eur. Heart J. 2001;22:2171–2179. doi: 10.1053/euhj.2001.2822. [DOI] [PubMed] [Google Scholar]
  • 28.Garbowski MW, et al. Biopsy-based calibration of T2* magnetic resonance for estimation of liver iron concentration and comparison with R2 Ferriscan. J. Cardiovasc. Magn. Reson. 2014;16:40. doi: 10.1186/1532-429X-16-40. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29.Henninger B, et al. R2* relaxometry for the quantification of hepatic iron overload: biopsy-based calibration and comparison with the literature. Rofo. 2015;187:472–479. doi: 10.1055/s-0035-1550887. [DOI] [PubMed] [Google Scholar]
  • 30.d’Assignies G, et al. (2018) Non-invasive measurement of liver iron concentration using 3-Tesla magnetic resonance imaging: validation against biopsy. Eur. Radiol. 2018;28(5):2022–2030. doi: 10.1007/s00330-017-5106-3. [DOI] [PubMed] [Google Scholar]
  • 31.Wood JC. Use of magnetic resonance imaging to monitor iron overload. Hematol. Oncol. Clin. N. Am. 2014;28(4):747–764. doi: 10.1016/j.hoc.2014.04.002. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32.Acton RT, et al. Relationships of serum ferritin, transferrin saturation, and HFE mutations and self-reported diabetes in the Hemochromatosis and Iron Overload Screening (HEIRS) study. Diabetes Care. 2006;29:2084–2089. doi: 10.2337/dc05-1592. [DOI] [PubMed] [Google Scholar]
  • 33.Ernst O, et al. Hepatic iron overload: diagnosis and quantification with MR imaging. Am. J. Roentgenol. 1997;168:1205–1208. doi: 10.2214/ajr.168.5.9129412. [DOI] [PubMed] [Google Scholar]
  • 34.Papakonstantinou O, et al. The pancreas in beta-thalassemia major: MR imaging features and correlation with iron stores and glucose disturbances. Eur. Radiol. 2007;17:1535–1543. doi: 10.1007/s00330-006-0507-8. [DOI] [PubMed] [Google Scholar]
  • 35.Maris TG, et al. Myocardial and liver iron status using a fast T*2 quantitative MRI (T*2qMRI) technique. Magn. Reson. Med. 2007;57:742–753. doi: 10.1002/mrm.21204. [DOI] [PubMed] [Google Scholar]
  • 36.Argyropoulou MI, et al. Liver, bone marrow, pancreas and pituitary gland iron overload in young and adult thalassemic patients: a T2 relaxometry study. Eur. Radiol. 2007;17:3025–3030. doi: 10.1007/s00330-007-0683-1. [DOI] [PubMed] [Google Scholar]
  • 37.Chezmar JL, et al. Hepatic iron overload: diagnosis and quantification by noninvasive imaging. Gastrointest. Radiol. 1990;15:27–31. doi: 10.1007/BF01888728. [DOI] [PubMed] [Google Scholar]
  • 38.Obrzut M, et al. Normative values for magnetic resonance elastography-based liver stiffness in a healthy population. Pol. Arch. Intern. Med. 2019;129(5):321–326. doi: 10.20452/pamw.4456. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 39.Henninger B, Alustiza J, Garbowski M, Gandon Y. Practical guide to quantification of hepatic iron with MRI. Eur. Radiol. 2020;30(1):383–393. doi: 10.1007/s00330-019-06380-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 40.Nuttall KL, Palaty J, Lockitch G. Reference limits for copper and iron in liver biopsies. Ann. Clin. Lab. Sci. 2003;33:443–450. [PubMed] [Google Scholar]
  • 41.Bassett ML, Halliday JW, Powell LW. Value of hepatic iron measurements in early hemochromatosis and determination of the critical iron level associated with fibrosis. Hepatology. 1986;6:24–29. doi: 10.1002/hep.1840060106. [DOI] [PubMed] [Google Scholar]
  • 42.Serai SD, Trout AT. Can MR elastography be used to measure liver stiffness in patients with iron overload? Abdom. Radiol. 2019;44(1):104–109. doi: 10.1007/s00261-018-1723-9. [DOI] [PubMed] [Google Scholar]
  • 43.Yokoo T, et al. Quantitative R2* MRI of the liver with rician noise models for evaluation of hepatic iron overload: simulation, phantom, and early clinical experience. JMRI. 2015;42(6):1544–1559. doi: 10.1002/jmri.24948. [DOI] [PubMed] [Google Scholar]
  • 44.Tipirneni-Sajja A, et al. Ultrashort echo time imaging for quantification of hepatic iron overload: Comparison of acquisition and fitting methods via simulations, phantoms, and in vivo data. JMRI. 2019;49(5):1475–1488. doi: 10.1002/jmri.26325. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 45.Adams PC, et al. HFE mutations in Caucasian participants of the hemochromatosis and iron overload screening study with serum ferritin level <1000 μg/L. Can. J. Gastroenterol. 2013;27(7):390–392. doi: 10.1155/2013/493170. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46.Sarigianni M, et al. Accuracy of magnetic resonance imaging in diagnosis of liver iron overload: a systematic review and meta-analysis. Clin. Gastroenterol. Hepatol. 2015;13(1):55–63. doi: 10.1016/j.cgh.2014.05.027. [DOI] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Data Availability Statement

The data that support the findings of this study are available from the corresponding author upon reasonable request.


Articles from Scientific Reports are provided here courtesy of Nature Publishing Group

RESOURCES