Abstract
The success of in vivo neural interfaces relies on their long-term stability and large scale in interrogating and manipulating neural activity after implantation. Conventional neural probes, owing to their limited spatiotemporal resolution and scale, face challenges for studying the massive, interconnected neural network in its native state. In this review, we argue that taking inspiration from biology will unlock the next generation of in vivo bioelectronic neural interfaces. Reducing the feature sizes of bioelectronic neural interfaces to mimic those of neurons enables high spatial resolution and multiplexity. Additionally, chronic stability at the device-tissue interface is realized by matching the mechanical properties of bioelectronic neural interfaces to those of the endogenous tissue. Further, modeling the design of neural interfaces after the endogenous topology of the neural circuitry enables new insights into the connectivity and dynamics of the brain. Lastly, functionalization of neural probe surfaces with coatings inspired by biology leads to enhanced tissue acceptance over extended timescales. Bioinspired neural interfaces will facilitate future developments in neuroscience studies and neurological treatments by leveraging bidirectional information transfer and integrating neuromorphic computing elements.
Keywords: Biomimetics, brain-machine interfaces, nanoelectronics, tissue-like electronics, neural recording, neural stimulation, neural probes, brain implants, chronic stability, spatiotemporal resolution
eTOC blurb:
Bioelectronic neural interfaces have afforded unprecedented opportunities to understand information processing in the brain and treat neurological disorders. Despite decades of effort, the design of conventional neural interfaces generally bears little resemblance to the properties of the neural tissue they are designed to interface. In this review, we emphasize the importance of leveraging bioinspired properties at the brain-device interface. Specifically, we classify four properties for next-generation bioelectronic neural interfaces to emulate: size and morphology, mechanics, topology, and biochemical functionalization.
Materials innovations have constantly driven the development of neural interfaces that allow neuroscientists to advance our understanding of the brain and enable neurologists to treat a variety of brain disorders.1–4 Scientific inquiry on the relationship between behaviors and their underlying neural circuits arguably began in the late 18th century, when Galvani demonstrated the ability of electrical stimuli to evoke motor output.5 His experiment, which laid the foundation for electrophysiology, routed a frog’s spinal cord to its thigh with an iron and bronze arch, producing the first electrical interface to stimulate the nerve. In the 19th century, scientists established direct connection between the brain and motor output by stimulating a dog’s cerebrum with a platinum wire and observing the corresponding limb movement.6 The field initiated by these early experiments has since produced electrical therapies such as deep-brain stimulation (DBS), arguably one of the most successful neurotechnologies for translational medicine, offering relief for debilitating neurological conditions such as Parkinson’s disease, epilepsy, and dystonia.7, 8 Despite its profound clinical impact, modern DBS electrodes are relatively large and rigid rods comprising platinum-iridium (PtIr) or stainless steel electrodes.9 Therefore, conventional neuromodulation electrodes are similar in material design and properties to the metallic arch used in Galvani’s experiment more than two centuries ago,10 while bearing little resemblance to the structure and properties of the neural tissue they are designed to interface.
Besides the neural interfaces that stimulate activity in neural tissue, neural recording interfaces, which afford information flow in the opposite direction, have traditionally been constructed with similar materials. In 1957, Hubel developed an electrolytically sharpened tungsten wire for direct recording of extracellular action potentials at the single neuron level, demonstrating reliable recording in a feline brain for up to 1 hour.11 These tungsten electrodes have since received widespread use in both neuroscience and neurology studies.12 Shortly thereafter, the patch clamp technique was developed to resolve neuronal behavior at the level of individual ion channels.13 By taking advantage of glass thermal drawing methods, the patch clamp technique involves elongating glass into a tapered micropipette, with a tip fine enough to perturb a small number of ion channels without affecting the entire cell.14 However, the patch clamp technique requires substantial insulation from vibration and reduction in shunt conductance to resolve neural signals.15 It is noteworthy that intracellular recordings by patch clamp are usually obtained from one cell at a time under non-physiological conditions, such as cultured neurons, tissue slices, and anesthetized brains.16 In vivo extracellular recording electrodes, such as the tungsten wire electrodes pioneered by Hubel, provide a proxy of the intracellular membrane potential waveform, which has been confirmed by simultaneous recordings of both intracellular and extracellular signals using separate electrodes16 or the same neural probe.17 In the 1970’s and 1980’s, advancements in semiconductor fabrication provided the foundation for silicon-based microelectrode arrays (MEAs).18 The Michigan array, with multiple recording electrodes per silicon shank, and the Utah array, featuring a two-dimensional (2D) array with micron-sized metal-plated silicon tips as recording sites, enabled scientists to probe neural circuitry with multi-site throughput and single-unit resolution, significantly improving neurotechnology standards.19–22 Similar to DBS electrodes, these recording devices are designed and fabricated with large and rigid non-biological features that prevent intimate interfacing over extended time periods with the endogenous neural tissue.
These foundational neural interfaces, both for stimulation and recording of neural activity, laid the framework for emergent bioelectronic neural probes with more desirable, bioinspired design features.18, 23, 24 The efficacy of neural probes is contingent on how well the device conforms to and interfaces with the local brain environment – both with its feature sizes and with its mechanical, topological, and biochemical properties.25, 26 Mismatch in these properties, where poor device design and material choices aggravate the inherent device-brain interaction, results in failure modes such as irreversible device degradation, inflammation-induced gliosis, neuron death and the subsequent loss of recorded neural signals and stimulation capability over a chronic timescale.27, 28 For example, DBS electrodes suffer from significant glial scarring at the device-tissue interface as a result of implant size and mechanical mismatch.29 Conventional neural recording interfaces, such as the Michigan and Utah arrays, also suffer from chronic recording instability as a result of glial scarring and mechanical mismatch with the neural tissue.27, 28
We argue that the challenges faced by existing bidirectional neural interfaces are a direct consequence of designs that have been historically developed without an a priori understanding of the nervous system.30 It has been demonstrated that neurotechnology design principles should match the brain’s known properties to optimize recording stability, resolution, and signal-to-noise ratio (SNR).2, 31 By taking inspiration from the brain and biology, there exists a rich opportunity to develop bidirectional neural interfaces which seamlessly integrate with the brain environment. In particular, the focus of this review is to define and demonstrate several principal properties of modern neurotechnology to optimize bidirectional transfer of neural information, based on fundamentally new structural, mechanical, topological, and biochemical design principles inspired by biology. Specially, we highlight the following key features of the nervous system (Fig. 1A) that have constantly inspired the evolution of emerging neural interfacing technologies (Table 1 & Fig. 2):
Table 1.
Year of appearance | Bioelectronic neural interface | Reference |
---|---|---|
1781 | Galvani bimetal arch wire | 5 |
1870 | Fritsch lead wire | 6 |
1957 | Hubel microwire | 11 |
1970 | Michigan array | 19 |
1972 | DBS electrode | 40 |
1989 | Tetrode | 41 |
1990 | Utah array | 20 |
2001 | Polyimide probe | 42 |
2007 | Michigan MEA with anti-inflammatory surface coating | 43 |
2007 | Dexamethane-coated Michigan MEA | 44 |
2008 | Neurotrophic electrode | 45 |
2011 | Neural probes coated with neuron-adhesion molecules | 46 |
2011 | Fish-bone-shaped neural probe | 47 |
2012 | Microthread electrode (MTE) | 48 |
2013 | Sub-10 nm BIT-FET | 49 |
2015 | Multifunctional fiber | 50 |
2015 | e-dura | 51 |
2015 | NeuroGrid array | 52 |
2015 | Subretinal photovoltaic array | 53 |
2016 | Endovascular stent-electrode | 54 |
2016 | Bioresorbable silicon electronics | 55 |
2017 | Ultraflexible nanoelectronic probes (“nanoelectronic threads”: NET-50, NET-10) | 56 |
2017 | Neuropixels | 57 |
2017 | Highly scalable mesh electronics | 58 |
2018 | Silicon nanoscale membrane | 59 |
2018 | Nanowire-based artificial photoreceptors | 60 |
2018 | NeuroRoots | 61 |
2018 | Carbon nanotube fiber microelectrodes (CNTf) | 62 |
2019 | Neuron-like electronics (NeuE) | 63 |
2019 | Neurotassels | 64 |
2019 | Flexible electrode threads (Neuralink) | 65 |
2019 | Integrated glass microelectrodes | 17 |
2019 | Polymer electrode array | 66 |
2019 | Flexible nanopipette | 67 |
2019 | U-shaped silicon nanowire transistor | 68 |
2020 | CMOS interfaced microwire array | 69 |
2020 | Neural Matrix | 70 |
2020 | Enhancement-mode, internal ion-gated organic electrochemical transistor (e-IGT) | 71 |
2020 | Liquid retina prosthesis | 72 |
2020 | The Argo (a 65,536-channel neural recording system) | 73 |
Bioinspired Size and Morphology
In the late 19th century, Ramón y Cajal utilized the Golgi staining method to illuminate the underlying complexity of the composition and connections within the brain.74 The stunningly exquisite drawings of Ramón y Cajal revealed the largely heterogenous structure of neurons (unlike many other somatic cells), comprising feature sizes that span many orders of magnitude and structures vastly different in their shapes and aspect ratios. For example, neuron somata range from 1 to 20 μm in size,32 axonal diameters range from 0.16 to 9 μm (with a high aspect ratio),33 and synaptic cleft distances range from 20 to 40 nm.75 Early-stage bioelectronic neural interfaces were limited by their large feature sizes, which showed great disparity from the characteristic sizes of neurons and sub-neuronal structures. Consequently, in vertebrate nervous systems, these devices were exclusively capable of measuring collective neuronal behavior, such as LFPs common in synchronous brain function.76 Single-neuron activity was measured only in invertebrate nervous systems with these devices, such as squid giant axons with millimeter diameters.77 Since the 1950’s, neural probes have established a paradigm for minimizing size (Fig. 1B) and increasing packing density of electrodes (Fig. 1C) through advancements in device fabrication strategies, such as photolithography, electron-beam lithography (EBL), and micromachining. For example, deep ultraviolet (DUV) photolithography is able to produce feature sizes down to 50 nm,78 EBL has been reported to achieve feature sizes down to 2 nm,79 and micromachining can achieve high precision typically better than 1 μm.80 This strategy has enabled the measurement of extracellular single-unit neuron activity from the mammalian brain by making the recording electrodes similar in size and density to the neurons they are designed to measure. Our ability to understand the connection between brain activity and behavior ultimately relies on improving the resolution and fidelity of next-generation bioelectronic neural probes inspired by the size and morphology of individual neurons.
Reducing the feature sizes of bioelectronic neural interfacing devices improves the resolution of measured neuronal behavior by decreasing the likelihood of signal attenuation though spatial averaging.81 For example, a simulated signal measured 1 μm away by a 1×1 μm2 electrode was reduced by 95% in signal amplitude when measured by a 100×100 μm2 electrode at the same distance.81 However, reduction of feature sizes is simultaneously demanding of applicable materials and fabrication methods. This challenge directly results from the fundamental difference in the bottom-up and top-down approaches that nature and engineers use to create functional devices, respectively.82 Traditionally, top-down fabrication technologies are used in engineering of neural interfaces, such as the electropolishing method used for Hubel’s tungsten microwire electrodes,11 and micromachining and photolithography for making silicon-based Michigan/Utah arrays.19, 20 The same top-down approaches have been used for fabricating the state-of-the-art high-density Michigan array, Neuropixels.57 In contrast, although widely found in biology, bottom-up approaches were employed to a lesser degree in engineering until recent years. For example, bioelectronic neural interfaces have been demonstrated with synthesized kinked silicon nanowires for extracellular and intracellular in vitro recordings of electrogenic cells.83, 84 In addition, free standing kinked nanowires and U-shaped nanowires, which integrate both bottom-up (nanowire synthesis) and top-down (lithographic patterning and assembly) approaches, have been demonstrated for enhanced cell targeting and multiplexing for in vitro intracellular recording of cardiomyocytes and neurons.68, 85, 86 Despite these examples demonstrating the feasibility of bottom-up bioelectronic neural devices, bottom-up processing generally limits the complexity, reproducibility, and scalability of devices with small feature sizes.87 Therefore, it is essential to judiciously design bioelectronic neural probes with reduced feature sizes and improved spatial resolution.
The maturation of fabrication technologies and materials discovery significantly refined the resolution of modern neural interface design to approach the feature sizes of individual neurons and their subcellular components (Fig. 3A). Modern neural interfaces now use a range of manufacturing techniques, from EBL to polymer-based thermal drawing processes used to create multifunctional probes.50, 88, 89 EBL offers a higher patterning resolution than photolithography, as the feature resolution of photolithography is limited by UV diffraction (approximately the wavelength of light used, ranging from 50 nm to ~1 μm),78 while the resolution of electron-beam lithography is often limited by the secondary electrons generated by the interaction between photoresist and the electron beam (typically ~tens of nm).90 In comparison, thermal drawing processing allows for high-throughput manufacturing of bioelectronic devices with up to a 200× reduction in cross-sectional area, producing electrode cross-sectional areas down to 5 μm via manufacturing of materials with similar glass and melting temperatures (Fig. 2A & Fig. 3B, i).50 Below, we highlight two bioinspired approaches to address such challenges: bioelectronic neural interfaces with feature sizes approaching individual neurons and with morphology mimicking neurites.
Feature sizes approaching individual neurons
By constructing electronic devices with densely packed electrodes of similar size to that of a neuron soma, ranging from 1 – 20 μm, an intimate neural tissue-probe interface can be realized at the single-neuron level with a significantly increase in electrode density and multiplexity.57, 73 This strategy has been demonstrated in a range of modern neural probes recently, including the multifunctional fibers (Fig. 2A, Fig. 3B, i),50 NeuroGrid (Fig. 2B),52 ultraflexible nanoelectronic probe (Fig. 2F, Fig. 3B, ii),56 Neuropixels (Fig. 2E, Fig. 3B, iii),57 NeuroRoots (Fig. 2G, Fig. 3B, iv),61 and neuron-like electronics (NeuE, Fig. 2H, Fig. 3B, v).63 For example, NeuroGrid offers an approach to record both collective and single-unit neural signals from the brain’s cortical surface in an electrocorticography (ECoG) setting (Fig. 2B).52 Inspired by the size and spacing of neurons in the cortex, NeuroGrid is designed with 10×10 μm2 electrodes and 30 μm interspacing, resolving both LFPs and action potentials from single neurons without brain penetration. In a further development, the enhancement-mode ion-gated transistor (e-IGT) with a channel length of 5 μm provides high speed and high gain for electrophysiological recordings performed with low power consumption owing to the default OFF state at zero gate potential. The combination of the enhancement mode and the conventional depletion mode achieves internal filtering in a wide voltage range to suit electromyography, electrocardiography, intracranial encephalography, and action potential measurements.71 In a recent example of a flexible, multiplexed electrode array called “Neural Matrix”, ECoG signals were recorded from 1,008 channels over a year in duration and a centimeter-scale brain region in space.70
The strategy of reducing electrode size to approach individual neurons and thus affording a high channel count of single-unit recording has also been adapted in the implantable neural probes. For example, the Neuropixels array (Fig. 2E, Fig. 3B, iii) represents a significant upgrade in channel count to similar silicon-based intracranial neural interfaces, such as the Michigan array. The Neuropixels array utilizes 130 nm CMOS technology91 and is now a consumer standard for high electrode density and small feature size neural interfaces with its 960 12×12 μm2 electrodes placed on a 70×20 μm2 shank.57 Meanwhile, the Argo system, a highly multiplexed Utah-like array, provides the highest channel count in vivo neural recording system to date, affording simultaneous recording from its 65,536 channels.73
Finally, building devices with inspiration from synapses and single ion channels wields the power to elucidate the behavior of subthreshold neural activity, such as transmembrane dynamics of individual ion channels. Ultrashort-channel nanowire field-effect transistors (FETs) with a 50 nm channel length and sub-10 nm branched intracellular nanotube field-effect transistors (BIT-FETs), mimicking the size of single ion channels, offer an alternative to patch clamp for recording from individual ion channels with subcellular resolution (Fig. 3C, vi & vii).49, 68, 86, 92 The impedance of ultrashort FETs and BIT-FETs with sizes down to <10 nm is significantly lower than would be anticipated with decreasing micropipette size in the patch clamp technique. This is enabled by the devices’ independence from interfacial impedances, allowing higher signal-to-noise recording of transmembrane potential with the capability to increase the multiplexity (i.e., channel count) of subcellular recording.93 Progress has also been made to the conventional patch clamp techniques. In a recent example, flexible quartz nanopipettes offer an alternative approach to conventional glass microelectrodes by featuring a reduced tip diameter of ~ 10 – 20 nm, which affords improved resolution and SNR for intracellular recordings (Fig. 3C, viii).67
Mimicking morphology of neurites
With a different perspective, neural probes inspired by the morphology of neurites and non-neuronal cells have also shown improved acceptance into the brain. This strategy has been demonstrated earlier in non-electronic neural tissue scaffolds, recognizing the importance of myelination of displaced axonal fibers in improving the longevity of neural probes in the brain.94, 95 For example, inspired by the elongated morphology of axons, Lee et al. optimized a neural scaffold containing axon-like polystyrene nanofibers to optimize oligodendrocyte myelination.96 The study uncovered the impact of polystyrene nanofiber diameters ranging from 0.2 – 4.0 μm on myelination efficiency, and unearthed the optimal range of diameters of polystyrene nanofiber for promoting myelination (greater than or equal to 0.5 μm). Typical axonal diameters fall within the range discovered in their results, suggesting that mirroring the morphology of neurites improves both the restoration of the endogenous neural tissue environment and the device-tissue acceptance. In addition, imitating the structural features of neurites has been demonstrated to improve the interface of neural tissue with electrophysiological neural probes in chronic studies. For example, NeuroRoots, a thread-like bioelectronic neural interface with small feature sizes, was designed with inspiration from the size and distribution of axonal bundles within the brain (Fig. 2G, Fig. 3B, iv).61 With 10 μm diameter circular electrode pads and 7×1.5 μm2 leads, NeuroRoots affords chronic recording stability up to 7 weeks and represents a promising technology to interrogate and modulate the firing of specific subpopulations of neurons in behaving animals. Similarly, Neurotassels offers a scalable, high electrode-density, and small-footprint approach to high-resolution and chronic recording stability devices, with sizes approaching those of neurites.64 With a bioresorbable polymer coating comprising polyethylene glycol (PEG), Neurotassels self-assembles prior to insertion through elastocapillary interactions with a molten, tissue-dissolvable polymer. Another demonstration of a bioelectronic neural interface which is indistinguishable from the endogenous neural tissue is the bioinspired NeuE. By mirroring both the morphology and size of neuron somata and neurites, NeuE affords exceptional chronical stability for long-term brain interfacing (Fig. 2H, Fig. 3B, v). With recording electrodes mimicking the soma of typical pyramidal neurons (8 – 20 μm) and interconnect lines along with insulating ribbons mimicking axons and dendrites (width = 1 – 4 μm), NeuE allows high-resolution recording and encourages neural progenitor cell (NPC) migration.63 The rostral migratory stream (RMS) is an endogenous structure in the rodent brain comprised of highly aligned vascular and glial cells which is dedicated to encouraging NPC migration.97 As its role and structure become increasingly understood, there is a significant opportunity for next-generation bioelectronic devices to explicitly mimic the role of the RMS, as has been demonstrated in the NeuE probe, to further persuade brain-device compliance. Therefore, mimicking the cellular and subcellular anatomy of neurons and other cells has demonstrated increased bioelectronic neural interface acclimation and improved recording stability and resolution.
To conclude this section, we posit that feature size reduction and size- and morphology-mimicking of neurons and neurites imparts the following advantages on neural interface functionality:
High resolution of interfacing: Neural interfaces are capable of measuring single-unit activity in addition to LFPs, as electrodes approach the size of neural somata (Fig. 1B).
High density multiplexing: Due to modern manufacturing techniques and as a direct consequence of smaller electrodes, neural interfaces are constructed with orders of magnitude more electrodes while occupying the same, or even smaller, volume (Fig. 1C).
Improved axon myelination and progenitor cell migration: By reducing interconnect widths and mimicking the morphology of neurites and glial cells, neural interfaces encourage acceptance into the brain by promoting their interactions with neurons and non-neuronal cells.
Bioinspired Mechanical Properties
Failure modes of electrical neural probes generally arise from design flaws such as material-degradation or shorting,98 and/or from device-brain interactions.99 While design limitations of bioelectronic neural interfaces are largely addressed by modern fabrication and materials technologies, deleterious device-brain interactions remain a significant challenge to chronic interfacing with neural tissue. Effects of device-brain interactions may emerge as early as insertion (e.g., puncturing blood vessels and displacing neural tissue), or later during chronic interfacing due to an inherent mechanical property mismatch (e.g., neuronal death during relative motion and micromotion, or device insulation due to glial sheath).100 We argue that the failure modes of implantable bioelectronic neural devices fundamentally arise from the mechanical property mismatch between neural interfaces and surrounding neural tissue.
Neural tissue is extremely soft, with its Young’s modulus ranging from 100 Pa to 10 kPa.101 This is a direct result of the composition of the neural tissue and the functions the tissue must fulfill: in order to transmit neural signals (in the form of membrane potentials) between different brain regions, the cable theory of neurites prescribes that the interior of the neurites must be conductive, while the exterior must be insulating.102 Ionic solution is an ideal conductive material a neurite can use for its interior, while myelinating glial cells with rich membrane lipid content are a readily available insulating material for its exterior.34 Both water and lipids are very fluidic, thus leading to the exceptional softness of the neural tissue.
Neural devices, however, are usually fabricated with much stiffer materials than the neural tissue, leading to mechanical mismatch which becomes the main contributor to the limited longevity and stability of a neural interface.101 Two physical properties evaluate the mechanical compliance of a neural probe: the Young’s modulus of the materials, and the bending stiffness of the structure. On one hand, Young’s modulus is an intrinsic material property that measures the ability of a material to resist a change in length under tensile or compressive forces.103 Typical materials used in neural interfaces are rigid, with Young’s moduli 8 – 9 orders of magnitude larger than that of the brain.1, 18 On the other hand, the bending stiffness measures the ability of a structure to resist bending deformation to applied force. Although the Young’s modulus defines a material’s intrinsic mechanical properties, the bending stiffness highlights the mechanical interaction between devices with certain structures and the surrounding neural tissue (Fig. 1D), which affects the magnitude of neural tissue damage and the evoked chronic immune responses.1, 18
When the bending stiffness of bioelectronic neural implants exceeds that of neural tissue, a cascade of coupled chronic immune responses occur, ranging from the development of a local insulating glial sheath that renders electronic devices inoperable, to complete neuronal death.27 Determining the origin of chronic immune responses is a complex problem, but it is known that a major source derives from the inertial difference between neural implants and surrounding tissue during micromotion (e.g., due to local blood flow and global motion).2 Studies have demonstrated that softer and smaller devices have the potential to minimize chronic immune responses;104 therefore, next-generation bioelectronic neural interfaces must be judiciously designed with materials and features with reduced mechanical mismatch to optimize chronic interfacing.
To improve mechanical compliance via a reduction of bending stiffness, which is proportional to the dominant material’s Young modulus and feature sizes (Eqn. 1), next-generation bioelectronic device design must incorporate small size with softer materials. This design principle is governed by the dependence of effective bending stiffness (K; normalized against the width of the neural probe) on the Young’s modulus and thickness of the neural probe:
#(1) |
where E and d are the Young’s modulus and dimension in the direction of bending, respectively. Neural implants with significantly reduced bending stiffness can thus be achieved by reducing Young’s modulus and/or reducing feature sizes (e.g., thickness). However, challenges remain in the design and fabrication of bioelectronic neural interfaces with similar mechanics to those of the neural tissue, since conventional technologies for making high-performance electronic systems usually use materials that are not mechanically compliant with the nervous tissue. Therefore, advances are needed in the constituent materials, architectural dimensions, and integration methods to impart bioinspired mechanical properties in next-generation neural interfaces.4
Softer materials reduce bending stiffness
Compared to hard materials such as silicon, polymers are ideal materials for neural probes owing to their low Young’s modulus and thus high flexibility (Fig. 4A). For example, polyimide provides a flexible material for the insulating layer of neural probes and can be patterned via standard photolithography methods105 and a novel laser-engraving method.106 By encapsulating gold electrodes and interconnects in a polyimide substrate, Rousche et al. introduced a new direction for neural interfaces by combining flexibility, biocompatibility, and chronic recording stability. The first generation of polyimide probes demonstrated chronic recording stability of multiunit neural activity, with a SNR of ~5.42 A multitude of neural interfaces with various polymer-based materials, shapes, and topologies have since been developed and demonstrated evidence for improved chronic compliance. NeuroGrid combines ultrathin and highly conformable materials (4 μm, parylene C) with its bioinspired architecture, described in the previous section, and demonstrates stable recordings for up to 10 days.52 Similarly, a device designed to mimic the properties of the dura mater, called the e-dura, combines silicone, gold interconnects, and soft Pt-silicone composite electrodes into a flexible package (Fig. 2C).51 By comparing the e-dura with a rigid device, the authors noted significant improvements in motor performance, decreased spinal cord damage, and lessened microglial and astrocytic accumulation after 6 weeks. In addition, multifunctional fibers, which leverage polymers alone to form optical waveguides, conductive electrodes, and microfluidic channels, demonstrated reduced glial response three months after implantation.107
It is noteworthy that the polymer materials used in these examples still have Young’s moduli orders of magnitude higher than that of the neural tissue, thus necessitating reliance on structural designs, such as ultrathin and macroporous electronics,108, 109 to further improve the neural tissue interface, as will be discussed below. A recent demonstration of ‘elastronics’, which consists of a highly conductive soft hydrogel as the conductor and an elastic fluorinated photoresist as the insulating layer, successfully produced a bioelectronics neural implant with Young’s modulus (104 – 105 Pa) similar to that of the neural tissue for the first time (Fig. 4A).110 Owing to the tissue-like Young’s moduli of the constituent materials, the soft elastronic arrays enabled ultralow voltage electrical stimulation of the mouse sciatic nerve. Furthermore, by employing tissue-like polymers with viscoplastic and self-healing properties, “morphing electronics” affords a reconfigurable and seamless neural interface that grows together with the sciatic nerve, thus allowing stable electrical stimulation during the fastest growth period in rats.111
Reduction of critical dimensions decreases bending stiffness
Besides replacement of harder materials with softer counterparts, it has also been demonstrated that reduction of critical dimensions can make nominally rigid materials mechanically compliant. For example, by varying the diameter and tip geometry of microwires, Obaid et al. demonstrated that a significantly tempered interfacial force, suggesting less tissue damage, upon insertion was measured for smaller-diameter microwires.113 Similarly, microthread electrodes (MTEs) are ultrasmall composite electrodes, made with polymer-coated 7 μm diameter carbon fibers, which produce over 4 orders of magnitude increase in flexibility and over twice the fracture strength compared to silicon. Owing to these properties, Kozai et al. demonstrated chronic recordings in rats for up to 5 weeks with reduced chronic immune responses at implantation site.48 Additionally, further reduction of feature sizes from micron-sized carbon fibers to 30 nm diameter multi-walled carbon nanotubes (CNTs) has been offered as a solution to improve the mechanical compliance and reduce the chronic immune responses of hard materials. By spinning CNTs into 10 μm diameter yarns, McCallum et al. demonstrated a 10-fold increase in conductivity and flexibility, compared to PtIr electrodes of the same diameter, as well as up to 16 weeks of recording stability in rat autonomic nerves.114 Finally, a combination of softer materials with reduced feature sizes has led to the unprecedented tissue-like mechanical properties of NeuE which has demonstrated few to none chronic immune responses and seamless integration with neuronal circuitry (Fig. 4B).63, 115, 116
Strategies to address the challenges of delivery
However, there remains an intrinsic dilemma in creating neural interfaces that seamlessly integrate with the brain while minimizing the impact of brain penetration and insertion. Flexible and ultrathin neural probes conform to neural tissue and minimize chronic immune responses, yet these probes require a specialized insertion technique to avoid mechanical deformation during the implantation process, which is the prominent failure mode for these devices.18, 26 Stiffeners and shape-memory polymers, which are inspired by stimulus-responsive materials found in biology, offer solutions for device implantation without buckling.117, 118 By coating thin film polymer probes with bioresorbable PEG or silk fibroin, an otherwise flexible device may be inserted without crumpling.119, 120 Freezing a highly flexible macroporous probe in liquid nitrogen has also been shown to demonstrate environment-induced softening, simply due to the thawing of ice without requiring an additional material-based coating.121 In addition, by utilizing shape memory polymers, neural interfaces have been shown to soften in vivo, hosting a similar elastic modulus to neural tissue after insertion.118 In an example inspired by twining plants, shape memory polymers have been employed to afford self-climbing of flexible electrodes onto peripheral nerves for in vivo stimulation and recording of nerve activity in rabbits.122 Finally, by taking inspiration from the reversible hardening of sea cucumber dermis, bioinspired neural devices are capable of actively changing its stiffness via chemoresponsive and thermoresponsive composites, which can complement preexisting polymeric systems.123
Besides the approaches that involve a dynamic change of stiffness to facilitate insertion, mesh electronics are delivered precisely into the targeted brain regions via syringe-assisted injection.124 Mesh electronics can be suspended in a biocompatible solution of phosphate buffered saline (PBS), loaded into a syringe needle, and delivered into the brain region of interest with stereotaxic syringe injection like pharmaceuticals.124 Inspired by the shuttle device-assisted delivery of ultraflexible nanoelectronic probes, a scalable, automatable alternative to injection is a ‘sewing machine’-like rapid insertion, which guides devices into precise brain regions while evading essential vasculature to minimize the impact of insertion.56, 65 Moreover, taking inspiration from the mechanisms behind the mosquito’s skin-piercing fascicles allows neural interfaces to be inserted with increased critical buckling force, and therefore greater precision free of crumpling (Fig. 4C).112 In addition, distributing the load of flexible devices via microfluidic actuation offers an alternative to inserting neural interfaces that would otherwise buckle. By coupling a flexible microelectrode to a microfluidic channel, the fluid applies a viscous drag force to the microelectrode due to the velocity differential between the two, thus holding the device under tension in preparation for insertion.62 Therefore, taking inspiration from biology, both within the brain and beyond, has the potential to improve the seamless integration of neural interfaces via improved mechanical compliance and optimized delivery methods to ultimately obtain a chronic interface for recording and modulating neural circuitry.
Bioinspired Topological Properties
In addition to revealing the intricate size and scale of neurons and neurites, the workings of Ramon y Cajal and Golgi also highlighted the elaborate connectivity between neurons and the complex circuits they comprise.125 Recent advances in electron microscopy and two-photon fluorescence microscopy have enabled the reconstruction of the three-dimensional (3D) network comprised of neurons and non-neuronal cells in the mouse brain with exceedingly convoluted connections and interactions.126, 127 The brain is comprised of highly organized and interconnected networks of neurons and non-neuronal cells, such as astrocytes and microglia, which significantly contribute to neuronal function, that neural interfaces should ideally leave intact.128 Despite knowledge of the underlying complexity of neural networks established in the late 19th century, early-stage bioelectronic neural interfaces were limited by available materials and techniques for fabrication.11 Owing again to the technological advancements since the 1950’s, 2D neural interfaces were developed as a direct result of planar microfabrication techniques.129 As discussed in the section “Bioinspired Size and Morphology”, these 2D structures allow for both a larger electrode packing density and the new capability of simultaneously measuring single-unit action potentials from multiple neurons, giving rise to knowledge of neural connectivity and circuitry.130–132 By mimicking the structure of neurons and the topology of their corresponding circuitry in three dimensions, next-generation neural interfaces will provide an unprecedented perspective on the dynamics and evolution of neural function.
Neural interface topology pertains to both the global device shape and the local distribution of its electrodes and/or sensors. Generally, the global and local topologies of devices can be identical but are not necessarily so. For example, the tetrode is a one-dimensional (1D) bundle of four microwires, where four exposed electrodes lay on the end, thereby measuring signals from a small plane of neural tissue in 2D.133 In contrast, the Utah-type MEA (Fig. 5, top middle panel) has its recording electrodes distributed laterally in 2D over a larger area (5×5 mm2) on a planar silicon substrate.20 This difference comes from distinct needs for neural signal measurement: the densely packed microwires in the tetrode are used for oversampled measurement and triangulation of action potentials from the same neurons,134 while the Utah array is designed for sampling a larger volume of neural tissue with many neurons.135 The historical evolution of neural interface topology, owing to materials discovery and modern processing techniques, is summarized to reveal a clear trend from lower-dimensional neural interfaces to 3D bioinspired topology (Fig. 1E).
Owing to the complex and dynamic structure of neural networks, we argue that there exist two principal opportunities for neural interface topology mimicry at different scales, which depend on the spatial distribution pattern of neurons.25 On one hand, the retina hosts a curvilinear interface where the ganglion cells reside for visual processing and integration. Because of this, the optimal interface topology extends primarily to two dimensions (Fig. 5, top middle column). A key technology for retinal interfaces is the hexagonally packed 2D silicon-based photovoltaic array that can convert incoming light into electrical stimulus as retinal prosthetics.53, 72, 136 On the other hand, the brain displays a high degree of interpenetration between the neuronal and glial networks in 3D;126 therefore, an ideal bioelectronic neural interface implanted within the brain should be designed to afford a similar degree of interpenetration between the implanted electronic network and the endogenous neuronal and glial networks by leaving sufficient open space for interpenetration to occur. On a global scale, the cooperation of glial cells directs functional connectivity and evolution of neuronal networks in the brain.137, 138 Therefore, the device design should ensure an absence of or, at worst, minimal disturbance of the endogenous distribution of neurons and non-neuronal cells. Despite these design principles of an ideal neural probe, conventional probes necessarily exclude a solid volume of neural tissue permanently, thus not only prohibiting 3D interpenetration with the neuronal and glial networks but also disrupting the endogenous distribution of cells139 and 3D diffusion of important molecular and macromolecular signaling species such as the inflammatory cytokines.140 We note that both principal opportunities have driven the development from lower-dimensional neural interfaces to 3D bioinspired topology, and below we will discuss this trend in the topological evolution of neural probes for rigid and flexible bioelectronic neural interfaces, respectively.
Topological constraints for rigid neural interfaces
Rigid bioelectronic neural interfaces, which inherit the conventional designs and materials of neural electrophysiological probes, are progressing toward an increased density of electrodes, refined fabrication processes, and accommodating layouts of recording and stimulation sites in the topological evolution of neural probes. For example, while 1D devices were originally limited by available materials and processing techniques, such as Hubel’s tungsten microwire (Fig. 5, top left panel),11 modern devices have made a resurgence in low dimensions with a variety of fabrication and processing techniques. MTEs are created by coating 7 μm diameter 1D carbon fibers with polymers via chemical vapor deposition (CVD) and electrochemical deposition for the insulating layer and the recording site, respectively, resulting in an advanced 1D interface with reduced footprint and chronic immune responses as discussed above.48 Similarly, CNTf microelectrodes utilize a clever fluidic microdrive insertion mechanism to deliver small carbon nanotube-based electrodes into the neural tissue, producing a 1D interface with neural tissue.62 Additionally, the Behnke-Fried electrodes are comprised of bundled platinum-iridium microwires that spread out into the target brain region like a cone, thus maximizing the interfaced volume of neural tissue in the brain with a limited number of 1D microwire electrodes.141 Meanwhile, multifunctional fibers have a topology similar to that of microwire electrode arrays such as the tetrode, where the device contains multiple electrodes on the end of a cylindrical structure to afford a 2D neural interface.50, 142 Conventional 2D MEA configurations such as the Michigan and Utah type silicon arrays (Fig. 5, top middle panel) have also received improvements in the topology of neural interfacing. For example, the Neuropixels probe is built on the conventional 2D layout of a Michigan-type MEA with a much larger number of recording sites, benefiting from high resolution planar processing techniques (e.g., 130-nm CMOS technology) which allow recordings from a large population of neurons.57 Similarly, Roukes et al. developed a 3D Michigan-type array comprising 16 silicon shanks in a 4×4 layout, affording up to 1024 electrodes in a 0.6 mm3 footprint (Fig. 5, top right panel).89 Lastly, the recent CMOS-integrated massively parallel microwire array integrates hundreds to thousands of fine micromachined microwires into a CMOS readout system for enhanced 3D interfacing with neural circuitry and readout technologies.69
Topological constraints for flexible neural interfaces
The popularization of flexible materials in neural interfaces allows for flexibility in device topology by reducing mechanical load and increasing device conformability to surrounding tissue.143 The ultraflexible nanoelectronic probes (Fig. 2F, Fig. 3B, ii), for example, host a linear array of four or eight electrodes on a 1D flexible thread, and offer chronically stable recording of single-unit action potentials for four months (Fig. 5, bottom left panel).56 Similarly, Neuralink offers 32-channel Michigan-style flexible threads that are implanted individually with a ‘sewing machine’.65 Comprising a 2D array of recording electrodes, NeuroGrid is a planar structure that, due to its flexibility, conforms to the irregularly curved three-dimensional structure of the cortical surface, offering measurement of both low-frequency ECoG signal and high-frequency single-unit action potentials from an extended area of the cortex (Fig. 5, bottom middle panel).52, 144 Besides the ECoG-type interrogation of single-unit action potentials from the cortical surface, a flexible epicortical device, which is comprised of a distributed mesh of thin Si membrane and a polydimethylsiloxane (PDMS) substrate enables precise activation of limb movement via light-induced capacitive and Faradaic effect.59, 145 Additionally, the ultraflexible and macroporous properties of mesh electronics allows for enhanced conformability to the curvilinear surface of the retina. Upon injection into the vitreous body of the eye, mesh electronics is capable of unfolding and coating the retinal surface to seamlessly interface with retinal ganglion cells (RGCs).146 In addition to its ultraflexible properties, the macroporosity of mesh electronics allows for unimpeded light to interact with the retinal ganglion cells, preserving visual acuity and light perception during chronic in vivo retinal experiments.146 This chronic 2D retinal interface enabled longitudinal study of RGCs after optic nerve crush, revealing dramatic differences in the physiological responses of distinct RGC types and their resilience to injury.147 Further, by interfacing a fully organic device comprising a flexible silk substrate and layers of photoactive conjugated polymers, Maya-Vetencourt et al. demonstrated a fully-autonomous bioelectronic neural prosthesis that restores vision in both a rat model of degenerative blindness and of retinal dystrophy.72, 136
In addition, flexible low-dimensional devices introduce the opportunity for higher dimensional occupancy. Neurotassels incorporates flexible and independent 1D fibers that can span two- to three-dimensions upon insertion into the brain.64 Similarly, NeuroRoots is a modular set of 1D electrodes that occupy a volume of neural tissue after being guided by a microwire (Fig. 5, bottom right panel).61 Mesh electronics, by incorporating multiple ribbons of recording electrodes into a porous, 3D structure, enables neurite interpenetration108 and prevents accumulation of inflammatory molecules.121, 140 As a result, mesh electronics exhibits few to none chronic immune responses, evidenced by immunohistology, for up to one year post-injection.115, 116 Likewise, NeuE is designed with a similar architecture, but with an increased porosity above 99% to further encourage neural circuitry interpenetration and NPC migration in 3D (Fig. 5, bottom right panel).63
By combining the advantages of flexible materials, advanced insertion techniques, and modern interfacing methodologies with higher-dimensional topologies mimicking those of the endogenous neural network, modern neural interfaces significantly reduce volume occupancy compared to traditional devices. Therefore, these topological design features contribute to a much larger number of interfaced neurons, a significant reduction of nearby chronic immune responses and the maintenance of a close-to-native local environment of the neural tissue. We argue that bioinspired neural interface topologies, made possible by flexible materials and modern processing techniques, result in unprecedented chronic recording stability and high-dimensional conformability, and, ultimately, represent the next step towards seamless integration with endogenous neural circuitry.
Bioinspired Coatings and Functionalization
Neuroinflammation, the immune system’s response to implanted neural interfaces within the central nervous system, is characterized by the accumulation of activated microglia and astrocytes at the implanted probe interface and exists in all stages of the device-brain interaction. A few weeks following implantation, activated microglia attach to the surface of the device and release proinflammatory and neurotoxic factors, eventually leading to neuronal death and device degradation.28, 148–150 Around 6–8 weeks post-implantation, a glial scar, comprising microglia and astrocytes, will form an encapsulating layer around the implant, limiting axonal regeneration and preventing restoration of damage done to the blood-brain barrier (BBB) during implantation.27, 149, 151 Because neuroinflammation comprises various cellular and chemical pathways, we argue that the local biochemical environment of the implanted bioelectronic neural interface can be modified via bioinspired surface coating and functionalization with a clear trend of surface-functionalized neural probes in recent years (Fig. 1F). These surface modification approaches will help suppress inflammation and promote neuronal growth, thus encouraging the regeneration of a native cellular environment at the neural tissue-implant interface.27, 139, 152
Anti-inflammatory coatings
Chronically stable neural interfaces have been developed by integrating anti-inflammatory coatings comprising endogenous molecules (Fig. 6, right). For example, alpha melanocyte-stimulating hormone (α-MSH) is an endogenous tridecapeptide, which is naturally produced by cells in the pituitary gland153 and inhibits pro-inflammatory cytokines and neurotoxic nitric oxide production by microglia.154 Additionally, α-MSH can directly mitigate microglial responses to foreign brain implants.155 By depositing α-MSH onto Michigan-type silicon MEAs, a significantly attenuated glial response was found for at least four weeks post-implantation.43 Similarly, dexamethasone is a synthetic glucocorticoid hormone used to treat many inflammatory responses via local delivery.156 By either coating silicon-based Michigan arrays or retrofitting a flexible polyimide-based device for controlled release of dexamethasone, both an attenuated inflammatory response and a reduction of neuronal loss are achieved.44, 157 Integrating minocycline, a broad synthetic tetracycline, to a bioelectronic neural interface can provide an additional pathway to the reduction of deleterious immune responses.158 Combining a silicon oxide substrate, to mimic the surface of silicon-based MEAs (i.e. Utah and Michigan-type MEAs), with an engineered polymer film for sustained release of minocycline, resulted in a reduction of biomarkers associated with inflammation and retention of neuron viability within 46 days in comparison to an uncoated silicon oxide substrate.158 Similarly, integrating a flexible microfluidic channel to silicon-parylene neural probes to administer minocycline into the local environment produced a reduction in microglial reaction when compared to a plain silicon shank.159 Another bioinspired approach to inhibit neuroinflammatory response is via targeting the receptor of a biomarker associated with the cascade of immune responses with the receptor’s antagonist. IL-1 is a family of cytokines that is known for inducing scars and contributing to the pro-inflammatory response in the brain.160 Coating silicon-based MEAs with IL1-receptor antagonist (IL1-ra), which blocks the receptor from the IL-1 cytokine molecules, produced significantly less glial scarring 4 weeks post-implantation compared to the non-coated MEA.161
Neuron-promoting coatings
In the 1950s, Levi-Montalcini and Cohen observed that chick embryos, after being injected with purified snake venom, exhibited an overgrowth of sensory and sympathetic ganglia and sympathetic nerve bundles.162 By isolating nerve growth factor (NGF), a pro-growth and regenerative biological signaling molecule, Levi-Montalcini and Cohen produced the first of many neurotrophic factors, resulting in their joint Nobel Prize in Physiology or Medicine in 1986. Since then, numerous additional neurotrophic factors have been discovered and implemented in vivo, demonstrating both anti-inflammatory and pro-growth regeneration of neuronal cells.100 The discovery of NGF and introduction to neurotrophic factors established an additional dimension to designing and implementing next-generation neural interfaces.163, 164
Neural regeneration-based long-term recording was first demonstrated in vivo in the late 1980’s by implanting a 75 μm diameter Teflon-insulated gold wire inside a hollow glass tube containing a piece of sciatic nerve, which contained active NGF.165 By inducing neurite growth, Kennedy et al. recorded single and multiple unit activity for up to twelve months in the motor cortex and demonstrated histological evidence for in vivo neurite growth. Additionally, extracellular matrix (ECM) materials can mitigate an inflammatory response due to their biocompatibility and biodegradability and, if incorporated in a scaffold-like framework, they have the potential to encourage neural regrowth.166, 167 In addition to providing mechanical support, ECM materials encourage regrowth of endogenous tissue through their intrinsic properties (e.g., biochemical and mechanical cues). ECM materials have found significant application in stroke recovery models and have the potential to translate in other models of neural interface technologies.168 For example, laminin is a key ECM regulator that plays crucial roles in cell adhesion, differentiation, migration and axonal pathfinding.169 By coating eight bilayers of laminin on a silicon-based neural probe, there was a ~50% reduction in glial fibrillary acid protein (GFAP), an indicator for astrocytes, four weeks post-implantation.170 Meanwhile, introducing a neural cell adhesion molecule may encourage neural regeneration while diminishing neuroinflammatory response (Fig. 6, center). L1, which is a transmembrane protein member of the L1 protein family, is a neuronal cell adhesion molecule expressed in the majority of neurons in the central nervous system (CNS), and is known to promote neuronal survival and migration.46 By comparing L1-coated with non-coated four-shank Michigan-type MEAs, Cui et al. noted significantly less neural loss and a lower activation of microglia and astrocytes on the L1-coated device.46 Inspired by the use of endogenous biological factors with the neurotrophic electrode and the mechanical and biochemical cues provided by the ECM, Lee et al. developed the Matrigel coating, which is a biologically-derived hydrogel containing basement membrane proteins as well as growth factors.171 By integrating Matrigel on a flexible parylene-C shank (optionally with the addition of dexamethasone), the authors demonstrated reduced electrical impedance during the first week and enhanced SNR at 14 weeks of recording relative to the uncoated shank.
While anti-inflammatory and neuron-promoting molecular coatings demonstrate the potential to curtail the chronic cascade of neuroinflammatory responses at the neural interface, most have been demonstrated on large and stiff devices (e.g., Michigan arrays). On one hand, this is advantageous as results can be interpreted to demonstrate some sort of “worst-case scenario”. However, on the other hand, we envision that combining modern small and flexible neural interfaces with the advantages of anti-inflammatory coatings can realize enhanced chronic recording stability. We imagine that, by mimicking interfacial biochemical and mechanical properties, neural interfaces can ‘trick’ the immune system into accepting what might have normally been flagged as a foreign body. For example, bioresorbable silicon-based flexible electronics globally integrate bioresorbable materials to encourage conformal interfacing and complete dissolution to minimize chronic impacts (Fig. 2D).55, 172, 173 By integrating a sheet of poly(lactic-co-glycolic) acid (PLGA) and silicon nanomembranes, the ECoG-like array serves as a bioinspired solution for minimally invasive integration with the cortical surface for up to 4–5 weeks.55 Additionally, the incorporation of a PLGA fiber with a bioresorbable photodetector (comprised of a doped monocrystalline Si nanomembrane) and zinc electrodes yielded a bioresorbable spectrometer capable of measuring neural activity via calcium indicator transients, cerebral temperature, and cerebral oxygenation.172
Conclusion and Outlook
The fields of neuroengineering and neuroscience have intimately connected roots, yet their corresponding objectives have largely been independent. We argue that there exists a remarkable opportunity for these fields to engage and refine the means of interrogating and manipulating neural activity at an unprecedented resolution and scale through both a fundamental understanding of endogenous tissue, including its size, mechanical, topological, and biochemical properties, and the development of next-generation devices that allow for seamless integration with surrounding tissue (Fig. 7).
Because neural interface properties are fundamentally coupled in complex and seemingly mercurial ways, it is difficult, and even inappropriate, to argue the importance of one property above all others. Instead, to counsel the direction of this exceptionally interdisciplinary field, we highlight the impact that varying key properties, inspired by the brain and beyond, has on the device’s ability to interface with endogenous neuronal circuitry. Therefore, we emphasize a modern trend of neural interface design (Fig. 1), where design principles are taken with an a priori knowledge of the properties of endogenous neural tissue and unlock unprecedented spatiotemporal resolution and scale. In particular, we reiterate key features of bioinspired neural interfaces:
Size and morphology: The feature sizes and morphologies of neural interfaces afford the opportunity for greater spatial resolution, substantial multiplexing, and endogenous neural tissue acceptance.
Mechanical properties: The device-brain interface improves as the bending stiffness of neural interfaces approaches that of adjacent neural tissue, thereby affording greater chronic recording stability.
Topological properties: As the topology and interpenetration of neural interfaces increases, the opportunity for improved interfacing and neuroscience study in its native cellular environment is afforded.
Biochemical properties: Coating or functionalizing neural interfaces with anti-inflammatory or neuron-promoting molecules influences the extent and magnitude of an inherent immune response to neural interface implantation.
With the progressing trend towards bioinspired bioelectronic neural interfaces, there exists a further opportunity for bidirectional information transfer. Stimulating neural devices have been demonstrated with significant clinical potential, such as the DBS electrode,8 yet stimulating devices have not evolved to the same great extent as recording devices. We assert that, in addition to the potential for closed-loop feedback between neuroscience and neuroengineering advances (Fig. 7), there also exists a fortuitous opportunity for closed-loop bidirectional information transfer from the brain and the user, which could take the form beyond the conventional implantable electrodes and fiber optics. For example, significant strides have been made in developing wirelessly operating bioelectronic devices, evidenced by the recording capabilities of Neural Dust.174 Additionally, by coupling a custom integrated circuit and energy storage capacitor to a piezoelectric transducer, StimDust affords low-power, wireless neural stimulation via ultrasonic power, with a minimal device volume footprint.175 Similarly, the recent development of ‘sono-optogenetics’ from our lab has demonstrated a minimally invasive method of delivering nanoscopic light sources deep inside the brain via intravenous injection of mechanoluminescent nanoparticles and a brain-penetrant, ultrasound-based neural stimulation interface.176 Besides the endogenous circulatory system in the body that has offered unique opportunities for delivery and formation of neural stimulation interfaces,176, 177 there are plenty of bioinspired opportunities for establishing a bidirectional neural interface with the desired properties in Fig. 7. Bidirectional information transfer via a seamless neural interface has the potential to elucidate function and dysfunction of neural circuitry, as well as the ability to better inform more efficient, less invasive, and longer lasting neural therapies.
Beyond the ways in which the field can improve the front-end interface between the brain and the bioelectronic device, additional opportunities exist to refine back-end connections with recording and stimulation instruments. First, there is a need to improve the downstream connection between the bioelectronic neural interface and data acquisition and storage systems. This requirement is emphasized by understanding the trend towards larger electrode multiplexity (as seen in Fig. 1C). As electrode multiplexity of bioelectronic neural interfaces increases, standard connection methods (e.g., flat flexible cables) are no longer relevant and new, reliable systems are required. Recent efforts have tackled this looming problem by taking inspiration from silicon-based technologies. For example, Neuropixels utilizes a customized CMOS-based connector and user-programmable switches to address 384 out of the 960 total electrodes simultaneously.57 Taking a step further, a massively paralleled microwire array has been developed recently to mate to a variety of CMOS chips, allowing further opportunities for scalability and customizability.69 The Argo system, taking a similar approach as the Neuropixels recording system and that of Obaid et al., affords the highest channel count of any in vivo bioelectronic neural interface with simultaneous recording of up to 65,536 channels at 32 kHz and 12-bit resolution.73 By designing a custom, on-chip CMOS-based system capable of standard preprocessing (e.g., analog to digital conversion and amplification), the Argo system provides a high throughput end-to-end solution to high density recording.73 Unlike spatially separated recording sites and data multiplexers, the implementation of on-shank CMOS multiplexing circuits on high electrode density probes (e.g. NeuroSeeker178) causes heat dissipation in the brain. The maximum power of these circuits must be curtailed to prevent excessive heating of tissue, limiting further increases in electrode multiplexity.18, 179, 180 While placing all active electronics on the probe base can reduce tissue heating, this technique has its own limitation as it increases the size of the headstage.18 Thus, further work is needed towards the optimal management of heat dissipation in high electrode density probes.
Furthermore, we envision the future of neural interfaces will surpass the conventional aim of achieving a physically seamless and high-throughput interface, additionally achieving a reliable and direct computational interface between the exogenous neural implants and the endogenous neural network that can process and analyze ultra-dense neural recording data efficiently.181 Fortunately, this is already galvanizing with Neuralink’s custom on-chip integrated circuit for real-time spike thresholding and wireless data transfer.65 Going beyond this, we envision the potential to interface endogenous neuronal circuitry with neuromorphic devices.182, 183 While neuromorphics were initially implemented to mimic desirable dynamics of neuronal computation, such as short-term and long-term potentiation involved in neuroplasticity, with silicon-based technology, these systems encompass much more. Now formed from inorganic and organic,184 low-dimensional and layered (e.g., van der Waals) materials,185 neuromorphic devices are envisioned to be efficient in memory computation via massive parallelism, capable of supervised and unsupervised learning, and strip the need for analog to digital (ADC) converters.186 The integration of sensing elements into neuromorphic devices will enable the construction of artificial central and peripheral nervous systems, which could repair or augment existing neural tissue.187 To unlock the great potential of chronic neuromorphic implants, this bio-influenced computational concept must be realized in bioinspired devices. By replacing energy demanding electronics with a system inspired by the tissue it measures, we envision a fundamental shift in what it means to listen to and mediate neural matter.
Progress and Potential Statement:
In the past decade, significant strides towards optimizing in vivo neural interfaces have been achieved with continuous materials advances. While modern materials discovery and fabrication techniques unlocked a fundamental avenue for remarkable progress of bioelectronic neural interfaces, a significant dimension impeding further progress emerges from the lack of an a priori understanding of the properties of the neural tissue. By taking inspiration from the brain and biology, there exists a rich opportunity to develop bidirectional neural interfaces which seamlessly integrate with the endogenous tissue. This review focuses on an exciting new dimension by demonstrating how several principal properties of modern neurotechnology may optimize bidirectional transfer of information, based on fundamentally new structural, mechanical, topological, and biochemical design principles inspired by biology.
Highlights:
Bioinspired sizes and morphologies enhance spatial resolution and multiplexing
Reducing the interfacial mechanical mismatch increases chronic recording stability.
Bioinspired topologies allow for conformal interfacing of various neural structures.
Biochemical functionalization tempers immune response, improving chronic interfacing.
Acknowledgements
We thank R. Yin, Y. Chen, and T. Guo for helpful discussions. This work was supported by a National Institutes of Health (NIH) Pathway to Independence Award (National Institute on Aging 5R00AG056636-04) and Wu Tsai Neurosciences Institute of Stanford University.
Footnotes
Publisher's Disclaimer: This is a PDF file of an unedited manuscript that has been accepted for publication. As a service to our customers we are providing this early version of the manuscript. The manuscript will undergo copyediting, typesetting, and review of the resulting proof before it is published in its final form. Please note that during the production process errors may be discovered which could affect the content, and all legal disclaimers that apply to the journal pertain.
Declaration of interests
The authors declare no competing interests.
References
- 1.Rivnay J, Wang H, Fenno L, Deisseroth K, and Malliaras GG (2017). Next-generation probes, particles, and proteins for neural interfacing. Sci. Adv 3, e1601649. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Chen R, Canales A, and Anikeeva P (2017). Neural recording and modulation technologies. Nat. Rev. Mater 2, 16093. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Patel SR, and Lieber CM (2019). Precision electronic medicine in the brain. Nat. Biotechnol 37, 1007–1012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Song E, Li J, Won SM, Bai W, and Rogers JA (2020). Materials for flexible bioelectronic systems as chronic neural interfaces. Nat. Mater 19, 590–603. [DOI] [PubMed] [Google Scholar]
- 5.Galvani L (1791). D viribus electricitatis in motu musculari: Commentarius. Bologna: Tip. Istituto delle Scienze, 1791; 58p.: 4 tavv. ft; in 4.; DCC. f. 70. [Google Scholar]
- 6.Fritsch G, and Hitzig E (2009). Electric excitability of the cerebrum (Über die elektrische Erregbarkeit des Grosshirns). Epilepsy Behav. 15, 123–130. [DOI] [PubMed] [Google Scholar]
- 7.Kringelbach ML, Jenkinson N, Owen SLF, and Aziz TZ (2007). Translational principles of deep brain stimulation. Nat. Rev. Neurosci 8, 623–635. [DOI] [PubMed] [Google Scholar]
- 8.Lozano AM, Lipsman N, Bergman H, Brown P, Chabardes S, Chang JW, Matthews K, McIntyre CC, Schlaepfer TE, Schulder M, et al. (2019). Deep brain stimulation: current challenges and future directions. Nat. Rev. Neurol 15, 148–160. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Cogan SF (2008). Neural Stimulation and Recording Electrodes. Annu. Rev. Biomed. Eng 10, 275–309. [DOI] [PubMed] [Google Scholar]
- 10.Cagnan H, Denison T, McIntyre C, and Brown P (2019). Emerging technologies for improved deep brain stimulation. Nat. Biotechnol 37, 1024–1033. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Hubel DH (1957). Tungsten microelectrode for recording from single units. Science 125, 549–550. [DOI] [PubMed] [Google Scholar]
- 12.Patel SR, Sheth SA, Martinez-Rubio C, Mian MK, Asaad WF, Gerrard JL, Kwon C-S, Dougherty DD, Flaherty AW, Greenberg BD, et al. (2013). Studying task-related activity of individual neurons in the human brain. Nat. Protoc 8, 949–957. [DOI] [PubMed] [Google Scholar]
- 13.Neher E, and Sakmann B (1976). Single-channel currents recorded from membrane of denervated frog muscle fibres. Nature 260, 799–802. [DOI] [PubMed] [Google Scholar]
- 14.Sakmann B, and Neher E (1984). Patch Clamp Techniques for Studying Ionic Channels in Excitable Membranes. Annu. Rev. Physiol 46, 455–472. [DOI] [PubMed] [Google Scholar]
- 15.Hamill OP, and McBride DW (1997). Induced Membrane Hypo/Hyper-Mechanosensitivity: A Limitation of Patch-Clamp Recording. Annu. Rev. Physiol 59, 621–631. [DOI] [PubMed] [Google Scholar]
- 16.Henze DA, Borhegyi Z, Csicsvari J, Mamiya A, Harris KD, and Buzsáki G (2000). Intracellular Features Predicted by Extracellular Recordings in the Hippocampus In Vivo. J. Neurophysiol 84, 390–400. [DOI] [PubMed] [Google Scholar]
- 17.Hunt DL, Lai C, Smith RD, Lee AK, Harris TD, and Barbic M (2019). Multimodal in vivo brain electrophysiology with integrated glass microelectrodes. Nat. Biomed. Eng 3, 741–753. [DOI] [PubMed] [Google Scholar]
- 18.Hong G, and Lieber CM (2019). Novel electrode technologies for neural recordings. Nat. Rev. Neurosci 20, 330–345. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Wise KD, Angell JB, and Starr A (1970). An Integrated-Circuit Approach to Extracellular Microelectrodes. IEEE Trans. Biomed. Eng BME-17, 238–247. [DOI] [PubMed] [Google Scholar]
- 20.Campbell PK, Jones KE, and Normann RA (1990). A 100 electrode intracortical array: Structural variability. Biomed. Sci. Instrum 26, 161–165. [PubMed] [Google Scholar]
- 21.Drake KL, Wise KD, Farraye J, Anderson DJ, and BeMent SL (1988). Performance of Planar Multisite Microprobes in Recording Extracellular Single-Unit Intracortical Activity. IEEE Trans. Biomed. Eng 35, 719–732. [DOI] [PubMed] [Google Scholar]
- 22.Campbell PK, Jones KE, Huber RJ, Horch KW, and Normann RA (1991). A Silicon-Based, Three-Dimensional Neural Interface: Manufacturing Processes for an Intracortical Electrode Array. IEEE Trans. Biomed. Eng 38, 758–768. [DOI] [PubMed] [Google Scholar]
- 23.Shoffstall AJ, and Capadona JR (2018). Bioinspired materials and systems for neural interfacing. In Current Opinion in Biomedical Engineering, Volume 6 (Elsevier Ltd), pp. 110–119. [Google Scholar]
- 24.Li H, Wang J, and Fang Y (2020). Bioinspired flexible electronics for seamless neural interfacing and chronic recording. Nanoscale Adv, DOI: 10.1039/d1030na00323a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Hong G, Viveros RD, Zwang TJ, Yang X, and Lieber CM (2018). Tissue-like Neural Probes for Understanding and Modulating the Brain. Biochemistry 57, 3995–4004. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Obidin N, Tasnim F, and Dagdeviren C (2020). The Future of Neuroimplantable Devices: A Materials Science and Regulatory Perspective. Adv. Mater 32, 1901482. [DOI] [PubMed] [Google Scholar]
- 27.Polikov VS, Tresco PA, and Reichert WM (2005). Response of brain tissue to chronically implanted neural electrodes. J. Neurosci. Methods 148, DOI: 10.1016/j.jneumeth.2005.1008.1015. [DOI] [PubMed] [Google Scholar]
- 28.Salatino JW, Ludwig KA, Kozai TDY, and Purcell EK (2017). Glial responses to implanted electrodes in the brain. Nat. Biomed. Eng 1, 862–877. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Cicchetti F, and Barker RA (2014). The glial response to intracerebrally delivered therapies for neurodegenerative disorders: Is this a critical issue? Front. Pharmacol 5, 139. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Acarón Ledesma H, Li X, Carvalho-de-Souza JL, Wei W, Bezanilla F, and Tian B (2019). An atlas of nano-enabled neural interfaces. Nat. Nanotechnol 14, 645–657. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Frank JA, Antonini M-J, and Anikeeva P (2019). Next-generation interfaces for studying neural function. Nat. Biotechnol 37, 1013–1023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Oyster CW, Takahashi ES, and Hurst DC (1982). Analysis of neuronal soma size distributions. J. Neurosci. Methods 6, 311–326. [DOI] [PubMed] [Google Scholar]
- 33.Liewald D, Miller R, Logothetis N, Wagner HJ, and Schüz A (2014). Distribution of axon diameters in cortical white matter: an electron-microscopic study on three human brains and a macaque. Biol. Cybern 108, 541–557. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Sterling P, and Laughlin S (2015). Principles of Neural Design, (MIT Press; ). [Google Scholar]
- 35.Herculano-Houzel S, Mota B, and Lent R (2006). Cellular scaling rules for rodent brains. Proc. Natl. Acad. Sci. USA 103, 12138–12143. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Hinsch K, and Zupanc GKH (2007). Generation and long-term persistence of new neurons in the adult zebrafish brain: A quantitative analysis. Neuroscience 146, 679–696. [DOI] [PubMed] [Google Scholar]
- 37.Lagercrantz H, Hanson MA, Ment LR, and Peebles DM eds. (2010). The Newborn Brain: Neuroscience and Clinical Applications, 2nd Edition (Cambridge University Press; ). [Google Scholar]
- 38.Cash D, and Carew TJ (1989). A quantitative analysis of the development of the central nervous system in juvenile Aplysia californica. J. Neurobiol 20, 25–47. [DOI] [PubMed] [Google Scholar]
- 39.Bode H, Berking S, David CN, Gierer A, Schaller H, and Trenkner R (1973). Quantitative analysis of cell types during growth and morphogenesis in Hydra. Wilhelm Roux Arch. Entwickl. Mech. Org 171, 269–285. [DOI] [PubMed] [Google Scholar]
- 40.Bechtereva NP, Bondartchuk AN, and Smirnov VM (1972). Therapeutic electrostimulations of the deep brain strucutres. Vopr. Neirokhir 1, 7–12. [PubMed] [Google Scholar]
- 41.Recce ML, and O’Keefe J (1989). The tetrode: a new technique for multi-unit extracellular recording. Soc. Neurosci. Abstr 15, 1250. [Google Scholar]
- 42.Rousche PJ, Pellinen DS, Pivin DP, Williams JC, Vetter RJ, and Kipke DR (2001). Flexible polyimide-based intracortical electrode arrays with bioactive capability. IEEE Trans. Biomed. Eng 48, 361–371. [DOI] [PubMed] [Google Scholar]
- 43.He W, McConnell GC, Schneider TM, and Bellamkonda RV (2007). A novel anti-inflammatory surface for neural electrodes. Adv. Mater 19, 3529–3533. [Google Scholar]
- 44.Zhong Y, and Bellamkonda RV (2007). Dexamethasone-coated neural probes elicit attenuated inflammatory response and neuronal loss compared to uncoated neural probes. Brain Res. 1148, 15–27. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Bartels J, Andreasen D, Ehirim P, Mao H, Seibert S, Wright EJ, and Kennedy P (2008). Neurotrophic electrode: Method of assembly and implantation into human motor speech cortex. J. Neurosci. Methods 174, 168–176. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Azemi E, Lagenaur CF, and Cui XT (2011). The surface immobilization of the neural adhesion molecule L1 on neural probes and its effect on neuronal density and gliosis at the probe/tissue interface. Biomaterials 32, 681–692. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Wu F, Im M, and Yoon E (2011). A flexible fish-bone-shaped neural probe strengthened by biodegradable silk coating for enhanced biocompatibility. In 16th International Solid-State Sensors, Actuators and Microsystems Conference (IEEE), pp. 966–969. [Google Scholar]
- 48.Kozai TDY, Langhals NB, Patel PR, Deng X, Zhang H, Smith KL, Lahann J, Kotov NA, and Kipke DR (2012). Ultrasmall implantable composite microelectrodes with bioactive surfaces for chronic neural interfaces. Nat. Mater 11, 1065–1073. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 49.Fu TM, Duan X, Jiang Z, Dai X, Xie P, Cheng Z, and Lieber CM (2014). Sub-10-nm intracellular bioelectronic probes from nanowire-nanotube heterostructures. Proc. Natl. Acad. Sci. USA 111, 1259–1264. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Canales A, Jia X, Froriep UP, Koppes RA, Tringides CM, Selvidge J, Lu C, Hou C, Wei L, Fink Y, et al. (2015). Multifunctional fibers for simultaneous optical, electrical and chemical interrogation of neural circuits in vivo. Nat. Biotechnol 33, 277–284. [DOI] [PubMed] [Google Scholar]
- 51.Minev IR, Musienko P, Hirsch A, Barraud Q, Wenger N, Moraud EM, Gandar J, Capogrosso M, Milekovic T, Asboth L, et al. (2015). Electronic dura mater for long-term multimodal neural interfaces. Science 347, 159–163. [DOI] [PubMed] [Google Scholar]
- 52.Khodagholy D, Gelinas JN, Thesen T, Doyle W, Devinsky O, Malliaras GG, and Buzsáki G (2015). NeuroGrid: Recording action potentials from the surface of the brain. Nat. Neurosci 18, 310–315. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Lorach H, Goetz G, Smith R, Lei X, Mandel Y, Kamins T, Mathieson K, Huie P, Harris J, Sher A, et al. (2015). Photovoltaic restoration of sight with high visual acuity. Nat. Med 21, 476–482. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 54.Oxley TJ, Opie NL, John SE, Rind GS, Ronayne SM, Wheeler TL, Judy JW, McDonald AJ, Dornom A, Lovell TJH, et al. (2016). Minimally invasive endovascular stent-electrode array for high-fidelity, chronic recordings of cortical neural activity. Nat. Biotechnol 34, 320–327. [DOI] [PubMed] [Google Scholar]
- 55.Yu KJ, Kuzum D, Hwang SW, Kim BH, Juul H, Kim NH, Won SM, Chiang K, Trumpis M, Richardson AG, et al. (2016). Bioresorbable silicon electronics for transient spatiotemporal mapping of electrical activity from the cerebral cortex. Nat. Mater 15, 782–791. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Luan L, Wei X, Zhao Z, Siegel JJ, Potnis O, Tuppen CA, Lin S, Kazmi S, Fowler RA, Holloway S, et al. (2017). Ultraflexible nanoelectronic probes form reliable, glial scar–free neural integration. Sci. Adv 3, e1601966. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Jun JJ, Steinmetz NA, Siegle JH, Denman DJ, Bauza M, Barbarits B, Lee AK, Anastassiou CA, Andrei A, Aydin Ç, et al. (2017). Fully integrated silicon probes for high-density recording of neural activity. Nature 551, 232–236. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58.Fu TM, Hong G, Viveros RD, Zhou T, and Lieber CM (2017). Highly scalable multichannel mesh electronics for stable chronic brain electrophysiology. Proc. Natl. Acad. Sci. USA 114, E10046–E10055. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 59.Jiang Y, Li X, Liu B, Yi J, Fang Y, Shi F, Gao X, Sudzilovsky E, Parameswaran R, Koehler K, et al. (2018). Rational design of silicon structures for optically controlled multiscale biointerfaces. Nat. Biomed. Eng 2, 508–521. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Tang J, Qin N, Chong Y, Diao Y, Yiliguma, Wang Z, Xue T, Jiang M, Zhang J, and Zheng G (2018). Nanowire arrays restore vision in blind mice. Nat. Commun 9, 786. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 61.Ferro MD, Proctor CM, Gonzalez A, Zhao E, Slezia A, Pas J, Dijk G, Donahue MJ, Williamson A, Malliaras GG, et al. (2018). NeuroRoots, a bio-inspired, seamless Brain Machine Interface device for long-term recording. bioRxiv, 460949. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Vitale F, Vercosa DG, Rodriguez AV, Pamulapati SS, Seibt F, Lewis E, Yan JS, Badhiwala K, Adnan M, Royer-Carfagni G, et al. (2018). Fluidic Microactuation of Flexible Electrodes for Neural Recording. Nano Lett. 18, 326–335. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Yang X, Zhou T, Zwang TJ, Hong G, Zhao Y, Viveros RD, Fu TM, Gao T, and Lieber CM (2019). Bioinspired neuron-like electronics. Nat. Mater 18, 510–517. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Guan S, Wang J, Gu X, Zhao Y, Hou R, Fan H, Zou L, Gao L, Du M, Li C, et al. (2019). Elastocapillary self-assembled neurotassels for stable neural activity recordings. Sci. Adv 5, eaav2842. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Musk E (2019). An Integrated Brain-Machine Interface Platform With Thousands of Channels. J. Med. Internet. Res 21, e16194. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Chung JE, Joo HR, Fan JL, Liu DF, Barnett AH, Chen S, Geaghan-Breiner C, Karlsson MP, Karlsson M, Lee KY, et al. (2019). High-Density, Long-Lasting, and Multi-region Electrophysiological Recordings Using Polymer Electrode Arrays. Neuron 101, 21–31.e25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 67.Jayant K, Wenzel M, Bando Y, Hamm JP, Mandriota N, Rabinowitz JH, Plante IJL, Owen JS, Sahin O, Shepard KL, et al. (2019). Flexible Nanopipettes for Minimally Invasive Intracellular Electrophysiology In Vivo. Cell Rep. 26, 266–278. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Zhao Y, You SS, Zhang A, Lee J-H, Huang J, and Lieber CM (2019). Scalable ultrasmall three-dimensional nanowire transistor probes for intracellular recording. Nat. Nanotechnol 14, 783–790. [DOI] [PubMed] [Google Scholar]
- 69.Obaid A, Hanna ME, Wu YW, Kollo M, Racz R, Angle MR, Müller J, Brackbill N, Wray W, Franke F, et al. (2020). Massively parallel microwire arrays integrated with CMOS chips for neural recording. Sci. Adv 6, eaay2789. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 70.Chiang C-H, Won SM, Orsborn AL, Yu KJ, Trumpis M, Bent B, Wang C, Xue Y, Min S, Woods V, et al. (2020). Development of a neural interface for high-definition, long-term recording in rodents and nonhuman primates. Sci. Transl. Med 12, eaay4682. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 71.Cea C, Spyropoulos GD, Jastrzebska-Perfect P, Ferrero JJ, Gelinas JN, and Khodagholy D (2020). Enhancement-mode ion-based transistor as a comprehensive interface and real-time processing unit for in vivo electrophysiology. Nat. Mater 19, 679–686. [DOI] [PubMed] [Google Scholar]
- 72.Maya-Vetencourt JF, Manfredi G, Mete M, Colombo E, Bramini M, Di Marco S, Shmal D, Mantero G, Dipalo M, Rocchi A, et al. (2020). Subretinally injected semiconducting polymer nanoparticles rescue vision in a rat model of retinal dystrophy. Nat. Nanotechnol, DOI: 10.1038/s41565-41020-40696-41563. [DOI] [PubMed] [Google Scholar]
- 73.Sahasrabuddhe K, Khan AA, Singh AP, Stern TM, Ng Y, Tadić A, Orel P, LaReau C, Pouzzner D, Nishimura K, et al. (2020). The Argo: A 65,536 channel recording system for high density neural recording in vivo. bioRxiv, 2020.2007.2017.209403. [Google Scholar]
- 74.Ramón y Cajal S, DeFelipe J, Jones EG, and May RM (1991). Cajal’s degeneration and regeneration of the nervous system, (Oxford University Press; ). [Google Scholar]
- 75.Savtchenko LP, and Rusakov DA (2007). The optimal height of the synaptic cleft. Proc. Natl. Acad. Sci. USA 104, 1823–1828. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 76.Buzsáki G, Anastassiou CA, and Koch C (2012). The origin of extracellular fields and currents - EEG, ECoG, LFP and spikes. Nat. Rev. Neurosci 13, 407–420. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 77.Hodgkin AL, and Huxley AF (1939). Action potentials recorded from inside a nerve fibre. Nature 144, 710–711. [Google Scholar]
- 78.Anselme K, Davidson P, Popa AM, Giazzon M, Liley M, and Ploux L (2010). The interaction of cells and bacteria with surfaces structured at the nanometre scale. Acta Biomater. 6, 3824–3846. [DOI] [PubMed] [Google Scholar]
- 79.Manfrinato VR, Zhang L, Su D, Duan H, Hobbs RG, Stach EA, and Berggren KK (2013). Resolution Limits of Electron-Beam Lithography toward the Atomic Scale. Nano Lett. 13, 1555–1558. [DOI] [PubMed] [Google Scholar]
- 80.Piljek P, Keran Z, and Math M (2014). Micromachining–review of literature from 1980 to 2010. Interdiscip. Description Complex Syst 12, 1–27. [Google Scholar]
- 81.Viswam V, Obien MEJ, Franke F, Frey U, and Hierlemann A (2019). Optimal Electrode Size for Multi-Scale Extracellular-Potential Recording From Neuronal Assemblies. Front. Neurosci 13, 385. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 82.(2018). Bottom-up biology. Nature 563, 171. [DOI] [PubMed] [Google Scholar]
- 83.Tian B, Xie P, Kempa TJ, Bell DC, and Lieber CM (2009). Single-crystalline kinked semiconductor nanowire superstructures. Nat. Nanotechnol 4, 824–829. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 84.Tian B, Cohen-Karni T, Qing Q, Duan X, Xie P, and Lieber CM (2010). Three-Dimensional, Flexible Nanoscale Field-Effect Transistors as Localized Bioprobes. Science 329, 830–834. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 85.Qing Q, Jiang Z, Xu L, Gao R, Mai L, and Lieber CM (2014). Free-standing kinked nanowire transistor probes for targeted intracellular recording in three dimensions. Nat. Nanotechnol 9, 142–147. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 86.Zhang X (2019). Nanowires Pin Neurons: a Nano “Moon Landing”. Matter 1, 560–562. [Google Scholar]
- 87.Iqbal P, Preece JA, and Mendes PM (2012). Nanotechnology: The “Top-Down” and “Bottom-Up” Approaches In Supramol. Chem, Gale PA and Steed JW, eds. (John Wiley and Sons; ). [Google Scholar]
- 88.Scholvin J, Kinney JP, Bernstein JG, Moore-Kochlacs C, Kopell N, Fonstad CG, and Boyden ES (2016). Close-Packed Silicon Microelectrodes for Scalable Spatially Oversampled Neural Recording. IEEE Trans. Biomed. Eng 63, 120–130. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 89.Rios G, Lubenov EV, Chi D, Roukes ML, and Siapas AG (2016). Nanofabricated Neural Probes for Dense 3-D Recordings of Brain Activity. Nano Lett. 16, 6857–6862. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 90.Broers AN, Hoole Andrew ACF, and Ryan JM (1996). Electron beam lithography - Resolution limits. Microelectron. Eng 32, 131–142. [Google Scholar]
- 91.Mora Lopez C, Putzeys J, Raducanu BC, Ballini M, Wang S, Andrei A, Rochus V, Vandebriel R, Severi S, Van Hoof C, et al. (2017). A Neural Probe with Up to 966 Electrodes and Up to 384 Configurable Channels in 0.13 μm SOI CMOS. IEEE Trans. Biomed. Circuits Syst 11, 510–522. [DOI] [PubMed] [Google Scholar]
- 92.Cohen-Karni T, Casanova D, Cahoon JF, Qing Q, Bell DC, and Lieber CM (2012). Synthetically encoded ultrashort-channel nanowire transistors for fast, pointlike cellular signal detection. Nano Lett. 12, 2639–2644. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 93.Zhang A, and Lieber CM (2016). Nano-Bioelectronics. Chem. Rev 116, 215–257. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 94.Wellman SM, Li L, Yaxiaer Y, McNamara I, and Kozai TDY (2019). Revealing Spatial and Temporal Patterns of Cell Death, Glial Proliferation, and Blood-Brain Barrier Dysfunction Around Implanted Intracortical Neural Interfaces. Front. Neurosci 13, 493. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 95.Wellman SM, and Kozai TDY (2018). In vivo spatiotemporal dynamics of NG2 glia activity caused by neural electrode implantation. Biomaterials 164, 121–133. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 96.Lee S, Leach MK, Redmond SA, Chong SYC, Mellon SH, Tuck SJ, Feng ZQ, Corey JM, and Chan JR (2012). A culture system to study oligodendrocyte myelination processes using engineered nanofibers. Nat. Methods 9, 917–922. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 97.James R, Kim Y, Hockberger PE, and Szele FG (2011). Subventricular zone cell migration: Lessons from quantitative two-photon microscopy. Front. Neurosci 5, 30. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 98.Wellman SM, Eles JR, Ludwig KA, Seymour JP, Michelson NJ, McFadden WE, Vazquez AL, and Kozai TDY (2018). A Materials Roadmap to Functional Neural Interface Design. Adv. Funct. Mater 28, 1701269. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 99.Kozai TDY, Marzullo TC, Hooi F, Langhals NB, Majewska AK, Brown EB, and Kipke DR (2010). Reduction of neurovascular damage resulting from microelectrode insertion into the cerebral cortex using in vivo two-photon mapping. J. Neural Eng 7, 046011. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 100.Jorfi M, Skousen JL, Weder C, and Capadona JR (2015). Progress towards biocompatible intracortical microelectrodes for neural interfacing applications. J. Neural Eng 12, 011001. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 101.Lacour SP, Courtine G, and Guck J (2016). Materials and technologies for soft implantable neuroprostheses. Nat. Rev. Mater 1, 16063. [Google Scholar]
- 102.Rall W (2011). Core Conductor Theory and Cable Properties of Neurons In Comprehensive Physiology, Terjung R, ed. (American Cancer Society; ), pp. 39–97. [Google Scholar]
- 103.Hibbeler RC (2011). Mechanics of Materials, 8th Edition, (Pearson; ). [Google Scholar]
- 104.Subbaroyan J, Martin DC, and Kipke DR (2005). A finite-element model of the mechanical effects of implantable microelectrodes in the cerebral cortex. J. Neural Eng 2, 103–113. [DOI] [PubMed] [Google Scholar]
- 105.Torrente-Rodríguez RM, Tu J, Yang Y, Min J, Wang M, Song Y, Yu Y, Xu C, Ye C, IsHak WW, et al. (2020). Investigation of Cortisol Dynamics in Human Sweat Using a Graphene-Based Wireless mHealth System. Matter 2, 921–937. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 106.Yang Y, Song Y, Bo X, Min J, Pak OS, Zhu L, Wang M, Tu J, Kogan A, Zhang H, et al. (2020). A laser-engraved wearable sensor for sensitive detection of uric acid and tyrosine in sweat. Nat. Biotechnol 38, 217–224. [DOI] [PubMed] [Google Scholar]
- 107.Park S, Guo Y, Jia X, Choe HK, Grena B, Kang J, Park J, Lu C, Canales A, Chen R, et al. (2017). One-step optogenetics with multifunctional flexible polymer fibers. Nat. Neurosci 20, 612–619. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 108.Liu J, Fu TM, Cheng Z, Hong G, Zhou T, Jin L, Duvvuri M, Jiang Z, Kruskal P, Xie C, et al. (2015). Syringe-injectable electronics. Nat. Nanotechnol 10, 629–635. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 109.Hong G, Yang X, Zhou T, and Lieber CM (2018). Mesh electronics: a new paradigm for tissue-like brain probes. Curr. Opin. Neurobiol 50, 33–41. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 110.Liu Y, Liu J, Chen S, Lei T, Kim Y, Niu S, Wang H, Wang X, Foudeh AM, Tok JBH, et al. (2019). Soft and elastic hydrogel-based microelectronics for localized low-voltage neuromodulation. Nat. Biomed. Eng 3, 58–68. [DOI] [PubMed] [Google Scholar]
- 111.Liu Y, Li J, Song S, Kang J, Tsao Y, Chen S, Mottini V, McConnell K, Xu W, Zheng Y-Q, et al. (2020). Morphing electronics enable neuromodulation in growing tissue. Nat. Biotechnol, DOI: 10.1038/s41587-41020-40495-41582. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 112.Shoffstall AJ, Srinivasan S, Willis M, Stiller AM, Ecker M, Voit WE, Pancrazio JJ, and Capadona JR (2018). A Mosquito Inspired Strategy to Implant Microprobes into the Brain. Sci. Rep 8, 122. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 113.Obaid A, Wu Y-W, Hanna M, Nix W, Ding J, and Melosh N (2018). Ultra-sensitive measurement of brain penetration with microscale probes for brain machine interface considerations. bioRxiv, 454520. [Google Scholar]
- 114.McCallum GA, Sui X, Qiu C, Marmerstein J, Zheng Y, Eggers TE, Hu C, Dai L, and Durand DM (2017). Chronic interfacing with the autonomic nervous system using carbon nanotube (CNT) yarn electrodes. Sci. Rep 7, 11723. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 115.Fu TM, Hong G, Zhou T, Schuhmann TG, Viveros RD, and Lieber CM (2016). Stable long-term chronic brain mapping at the single-neuron level. Nat. Methods 13, 875–882. [DOI] [PubMed] [Google Scholar]
- 116.Zhou T, Hong G, Fu TM, Yang X, Schuhmann TG, Viveros RD, and Lieber CM (2017). Syringe-injectable mesh electronics integrate seamlessly with minimal chronic immune response in the brain. Proc. Natl. Acad. Sci. USA 114, 5894–5899. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 117.Lewitus D, Smith KL, Shain W, and Kohn J (2011). Ultrafast resorbing polymers for use as carriers for cortical neural probes. Acta Biomater. 7, 2483–2491. [DOI] [PubMed] [Google Scholar]
- 118.Ware T, Simon D, Arreaga-Salas DE, Reeder J, Rennaker R, Keefer EW, and Voit W (2012). Fabrication of responsive, softening neural interfaces. Adv. Funct. Mater 22, 3470–3479. [Google Scholar]
- 119.Felix SH, Shah KG, Tolosa VM, Sheth HJ, Tooker AC, Delima TL, Jadhav SP, Frank LM, and Pannu SS (2013). Insertion of flexible neural probes using rigid stiffeners attached with biodissolvable adhesive. J. Vis. Exp, e50609. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 120.Kim T. i., McCall JG, Jung YH, Huang X, Siuda ER, Li Y, Song J, Song YM, Pao HA, Kim R-H, et al. (2013). Injectable, Cellular-Scale Optoelectronics with Applications for Wireless Optogenetics. Science 340, 211. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 121.Xie C, Liu J, Fu TM, Dai X, Zhou W, and Lieber CM (2015). Three-dimensional macroporous nanoelectronic networks as minimally invasive brain probes. Nat. Mater 14, 1286–1292. [DOI] [PubMed] [Google Scholar]
- 122.Zhang Y, Zheng N, Cao Y, Wang F, Wang P, Ma Y, Lu B, Hou G, Fang Z, Liang Z, et al. (2019). Climbing-inspired twining electrodes using shape memory for peripheral nerve stimulation and recording. Sci. Adv 5, eaaw1066. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 123.Capadona JR, Shanmuganathan K, Tyler DJ, Rowan SJ, and Weder C (2008). Stimuli-responsive polymer nanocomposites inspired by the sea cucumber dermis. Science 319, 1370–1374. [DOI] [PubMed] [Google Scholar]
- 124.Hong G, Fu TM, Zhou T, Schuhmann TG, Huang J, and Lieber CM (2015). Syringe Injectable Electronics: Precise Targeted Delivery with Quantitative Input/Output Connectivity. Nano Lett. 15, 6979–6984. [DOI] [PubMed] [Google Scholar]
- 125.Swanson LW, Newman E, Araque A, and Dubinsky JM (2017). The Beautiful Brain: The Drawings of Santiago Ramon y Cajal, (Abrams; ). [Google Scholar]
- 126.Kasthuri N, Hayworth KJ, Berger DR, Schalek RL, Conchello JA, Knowles-Barley S, Lee D, Vázquez-Reina A, Kaynig V, Jones TR, et al. (2015). Saturated Reconstruction of a Volume of Neocortex. Cell 162, 648–661. [DOI] [PubMed] [Google Scholar]
- 127.Winnubst J, Bas E, Ferreira TA, Wu Z, Economo MN, Edson P, Arthur BJ, Bruns C, Rokicki K, Schauder D, et al. (2019). Reconstruction of 1,000 Projection Neurons Reveals New Cell Types and Organization of Long-Range Connectivity in the Mouse Brain. Cell 179, 268–281.e213. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 128.Allen NJ, and Barres BA (2009). Glia - more than just brain glue. Nature 457, 675–677. [DOI] [PubMed] [Google Scholar]
- 129.(2018). 60 Years of Integrated Circuits. Nat. Electron 1, 483. [Google Scholar]
- 130.Rousche PJ, and Normann RA (1998). Chronic recording capability of the Utah Intracortical Electrode array in cat sensory cortex. J. Neurosci. Methods 82, DOI: 10.1016/S0165-0270(1098)00031-00034. [DOI] [PubMed] [Google Scholar]
- 131.Fujisawa S, Amarasingham A, Harrison MT, and Buzsáki G (2008). Behavior-dependent short-term assembly dynamics in the medial prefrontal cortex. Nat. Neurosci 11, 823–833. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 132.Barthó P, Hirase H, Monconduit L, Zugaro M, Harris KD, and Buzsáki G (2004). Characterization of neocortical principal cells and interneurons by network interactions and extracellular features. J. Neurophysiol 92, 600–608. [DOI] [PubMed] [Google Scholar]
- 133.Jog MS, Connolly CI, Kubota Y, Iyengar DR, Garrido L, Harlan R, and Graybiel AM (2002). Tetrode technology: Advances in implantable hardware, neuroimaging, and data analysis techniques. J. Neurosci. Methods 117, 141–152. [DOI] [PubMed] [Google Scholar]
- 134.McNaughton BL, O’Keefe J, and Barnes CA (1983). The stereotrode: A new technique for simultaneous isolation of several single units in the central nervous system from multiple unit records. J. Neurosci. Methods 8, 391–397. [DOI] [PubMed] [Google Scholar]
- 135.Gilja V, Pandarinath C, Blabe CH, Nuyujukian P, Simeral JD, Sarma AA, Sorice BL, Perge JA, Jarosiewicz B, Hochberg LR, et al. (2015). Clinical translation of a high-performance neural prosthesis. Nat. Med 21, 1142–1145. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 136.Maya-Vetencourt JF, Ghezzi D, Antognazza MR, Colombo E, Mete M, Feyen P, Desii A, Buschiazzo A, Di Paolo M, Di Marco S, et al. (2017). A fully organic retinal prosthesis restores vision in a rat model of degenerative blindness. Nat. Mater 16, 681–689. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 137.Eroglu C, and Barres BA (2010). Regulation of synaptic connectivity by glia. Nature 468, 223–231. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 138.Clarke LE, and Barres BA (2013). Emerging roles of astrocytes in neural circuit development. Nat. Rev. Neurosci 14, 311–321. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 139.Biran R, Martin DC, and Tresco PA (2005). Neuronal cell loss accompanies the brain tissue response to chronically implanted silicon microelectrode arrays. Exp. Neurol 195, 115–126. [DOI] [PubMed] [Google Scholar]
- 140.Saxena T, and Bellamkonda RV (2015). A sensor web for neurons. Nat. Mater 14, 1190–1191. [DOI] [PubMed] [Google Scholar]
- 141.Fried I, Wilson CL, Maidment NT, Engel J, Behnke E, Fields TA, Macdonald KA, Morrow JW, and Ackerson L (1999). Cerebral microdialysis combined with single-neuron and electroencephalographic recording in neurosurgical patients. J. Neurosurg 91, 697–705. [DOI] [PubMed] [Google Scholar]
- 142.Yan W, Dong C, Xiang Y, Jiang S, Leber A, Loke G, Xu W, Hou C, Zhou S, Chen M, et al. (2020). Thermally drawn advanced functional fibers: New frontier of flexible electronics. Mater. Today, DOI: 10.1016/j.mattod.2019.1011.1006. [DOI] [Google Scholar]
- 143.Kozai TDY, Jaquins-Gerstl AS, Vazquez AL, Michael AC, and Cui XT (2015). Brain tissue responses to neural implants impact signal sensitivity and intervention strategies. ACS Chem. Neurosci 6, 48–67. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 144.Khodagholy D, Gelinas JN, and Buzsáki G (2017). Learning-enhanced coupling between ripple oscillations in association cortices and hippocampus. Science 358, 369–372. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 145.Park S, Frank JA, and Anikeeva P (2018). Silicon biointerfaces for all scales. Nat. Biomed. Eng 2, 471–472. [DOI] [PubMed] [Google Scholar]
- 146.Hong G, Fu TM, Qiao M, Viveros RD, Yang X, Zhou T, Lee JM, Park HG, Sanes JR, and Lieber CM (2018). A method for single-neuron chronic recording from the retina in awake mice. Science 360, 1447–1451. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 147.Tran NM, Shekhar K, Whitney IE, Jacobi A, Benhar I, Hong G, Yan W, Adiconis X, Arnold ME, Lee JM, et al. (2019). Single-Cell Profiles of Retinal Ganglion Cells Differing in Resilience to Injury Reveal Neuroprotective Genes. Neuron 104, 1039–1055.e1012. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 148.Barron KD (2003). Microglia: History, cytology, and reactions. J. Neurol. Sci 207, 98. [DOI] [PubMed] [Google Scholar]
- 149.McConnell GC, Rees HD, Levey AI, Gutekunst CA, Gross RE, and Bellamkonda RV (2009). Implanted neural electrodes cause chronic, local inflammation that is correlated with local neurodegeneration. J. Neural Eng 6, 056003. [DOI] [PubMed] [Google Scholar]
- 150.Schwartz M (2003). Macrophages and microglia in central nervous system injury: Are they helpful or harmful? J. Cerebr. Blood F. Met 23, 385–394. [DOI] [PubMed] [Google Scholar]
- 151.Tsui C, Koss K, Churchward MA, and Todd KG (2019). Biomaterials and glia: Progress on designs to modulate neuroinflammation. Acta Biomater. 83, 13–28. [DOI] [PubMed] [Google Scholar]
- 152.Kim YT, Hitchcock RW, Bridge MJ, and Tresco PA (2004). Chronic response of adult rat brain tissue to implants anchored to the skull. Biomaterials 25, 2229–2237. [DOI] [PubMed] [Google Scholar]
- 153.Kastin A, and Schally A (1967). Autoregulation of Release of Melanocyte Stimulating Hormone from the Rat Pituitary. Nature 213, 1238–1240. [Google Scholar]
- 154.Galimberti D, Baron P, Meda L, Prat E, Scarpini E, Delgado R, Catania A, Lipton JM, and Scarlato G (1999). α-MSH peptides inhibit production of nitric oxide and tumor necrosis factor-α by microglial cells activated with β-amyloid and interferon γ. Biochem. Biophys. Res. Commun 263, 251–256. [DOI] [PubMed] [Google Scholar]
- 155.Zhong Y, and Bellamkonda RV (2005). Controlled release of anti-inflammatory agent α-MSH from neural implants. J. Control. Release 106, 309–318. [DOI] [PubMed] [Google Scholar]
- 156.Chao CC, Hu S, Close K, Choi CS, Molitor TW, Novick WJ, and Peterson PK (1992). Cytokine Release from Microglia: Differential Inhibition by Pentoxifylline and Dexamethasone. J. Infect. Dis 166, 847–853. [DOI] [PubMed] [Google Scholar]
- 157.Boehler C, Kleber C, Martini N, Xie Y, Dryg I, Stieglitz T, Hofmann UG, and Asplund M (2017). Actively controlled release of Dexamethasone from neural microelectrodes in a chronic in vivo study. Biomaterials 129, 176–187. [DOI] [PubMed] [Google Scholar]
- 158.Zhang Z, Nong J, and Zhong Y (2015). Antibacterial, anti-inflammatory and neuroprotective layer-by-layer coatings for neural implants. J. Neural Eng 12, 046015. [DOI] [PubMed] [Google Scholar]
- 159.Liu B, Kim E, Meggo A, Gandhi S, Luo H, Kallakuri S, Xu Y, and Zhang J (2017). Enhanced biocompatibility of neural probes by integrating microstructures and delivering anti-inflammatory agents via microfluidic channels. J. Neural Eng 14, 026008. [DOI] [PubMed] [Google Scholar]
- 160.Shaftel SS, Griffin WST, and O’Banion MK (2008). The role of interleukin-1 in neuroinflammation and Alzheimer disease: an evolving perspective. J. Neuroinflammation 5, 7. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 161.Taub AH, Hogri R, Magal A, Mintz M, and Shacham-Diamand Y (2012). Bioactive anti-inflammatory coating for chronic neural electrodes. J. Biomed. Mater. Res. A 100A, 1854–1858. [DOI] [PubMed] [Google Scholar]
- 162.Levi-Montalcini R (1987). The Nerve Growth Factor 35 Years Later. Science 237, 1154–1162. [DOI] [PubMed] [Google Scholar]
- 163.Kennedy P (2012). A Prosthesis for Restoration of Speech in Locked-In Patients. Neurology 78, P02.003. [Google Scholar]
- 164.Kennedy PR, Andreasen DS, Bartels J, Ehirim P, Wright EJ, Seibert S, and Cervantes AJ (2018). Validation of Neurotrophic Electrode Long-Term Recordings in Human Cortex In Brain–Computer Interfaces Handbook: Technological and Theoretical Advances, Nam CS, Nijholt A and Lotte A, eds. (CRC Press; ), pp. 279–296. [Google Scholar]
- 165.Kennedy PR (1989). The cone electrode: a long-term electrode that records from neurites grown onto its recording surface. J. Neurosci. Methods 29, 181–193. [DOI] [PubMed] [Google Scholar]
- 166.Chen S, and Allen MG (2012). Extracellular matrix-based materials for neural interfacing. MRS Bull. 37, 606–613. [Google Scholar]
- 167.Hussey GS, Dziki JL, and Badylak SF (2018). Extracellular matrix-based materials for regenerative medicine. Nat. Rev. Mater 3, 159–173. [Google Scholar]
- 168.Ghuman H, Gerwig M, Nicholls FJ, Liu JR, Donnelly J, Badylak SF, and Modo M (2017). Long-term retention of ECM hydrogel after implantation into a sub-acute stroke cavity reduces lesion volume. Acta Biomater. 63, 50–63. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 169.Heiduschka P, Romann I, Ecken H, Schöning M, Schuhmann W, and Thanos S (2001). Defined adhesion and growth of neurones on artificial structured substrates. Electrochim. Acta 47, 299–307. [Google Scholar]
- 170.He W, McConnell GC, and Bellamkonda RV (2006). Nanoscale laminin coating modulates cortical scarring response around implanted silicon microelectrode arrays. J. Neural Eng 3, 316–326. [DOI] [PubMed] [Google Scholar]
- 171.Lee CD, Hara SA, Yu L, Kuo JTW, Kim BJ, Hoang T, Pikov V, and Meng E (2016). Matrigel coatings for Parylene sheath neural probes. J. Biomed. Mater. Res. B 104, 357–368. [DOI] [PubMed] [Google Scholar]
- 172.Bai W, Shin J, Fu R, Kandela I, Lu D, Ni X, Park Y, Liu Z, Hang T, Wu D, et al. (2019). Bioresorbable photonic devices for the spectroscopic characterization of physiological status and neural activity. Nat. Biomed. Eng 3, 644–654. [DOI] [PubMed] [Google Scholar]
- 173.Kang S-K, Murphy RKJ, Hwang S-W, Lee SM, Harburg DV, Krueger NA, Shin J, Gamble P, Cheng H, Yu S, et al. (2016). Bioresorbable silicon electronic sensors for the brain. Nature 530, 71–76. [DOI] [PubMed] [Google Scholar]
- 174.Seo D, Neely RM, Shen K, Singhal U, Alon E, Rabaey JM, Carmena JM, and Maharbiz MM (2016). Wireless Recording in the Peripheral Nervous System with Ultrasonic Neural Dust. Neuron 91, 529–539. [DOI] [PubMed] [Google Scholar]
- 175.Piech DK, Johnson BC, Shen K, Ghanbari MM, Li KY, Neely RM, Kay JE, Carmena JM, Maharbiz MM, and Muller R (2020). A wireless millimetre-scale implantable neural stimulator with ultrasonically powered bidirectional communication. Nat. Biomed. Eng 4, 207–222. [DOI] [PubMed] [Google Scholar]
- 176.Wu X, Zhu X, Chong P, Liu J, Andre LN, Ong KS, Brinson K Jr., Mahdi AI, Li J, Fenno LE, et al. (2019). Sono-optogenetics facilitated by a circulation-delivered rechargeable light source for minimally invasive optogenetics. Proc. Natl. Acad. Sci. USA 116, 26332–26342. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 177.Opie NL, John SE, Rind GS, Ronayne SM, Wong YT, Gerboni G, Yoo PE, Lovell TJH, Scordas TCM, Wilson SL, et al. (2018). Focal stimulation of the sheep motor cortex with a chronically implanted minimally invasive electrode array mounted on an endovascular stent. Nat. Biomed. Eng 2, 907–914. [DOI] [PubMed] [Google Scholar]
- 178.Raducanu BC, Yazicioglu RF, Lopez CM, Ballini M, Putzeys J, Wang S, Andrei A, Rochus V, Welkenhuysen M, van Helleputte N, et al. (2017). Time multiplexed active neural probe with 1356 parallel recording sites. Sensors 17, 2388. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 179.Steinmetz NA, Koch C, Harris KD, and Carandini M (2018). Challenges and opportunities for large-scale electrophysiology with Neuropixels probes. Curr. Opin. Neurobiol 50, 92–100. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 180.Owen SF, Liu MH, and Kreitzer AC (2019). Thermal constraints on in vivo optogenetic manipulations. Nat. Neurosci 22, 1061–1065. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 181.Fang Y, Meng L, Prominski A, Schaumann EN, Seebald M, and Tian B (2020). Recent advances in bioelectronics chemistry. Chem. Soc. Rev, DOI: 10.1039/D1030CS00333F. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 182.Keene ST, Lubrano C, Kazemzadeh S, Melianas A, Tuchman Y, Polino G, Scognamiglio P, Cinà L, Salleo A, van de Burgt Y, et al. (2020). A biohybrid synapse with neurotransmitter-mediated plasticity. Nat. Mater, DOI: 10.1038/s41563-41020-40703-y. [DOI] [PubMed] [Google Scholar]
- 183.Cramer T (2020). Learning with brain chemistry. Nat. Mater, DOI: 10.1038/s41563-41020-40711-y. [DOI] [PubMed] [Google Scholar]
- 184.Fuller EJ, Keene ST, Melianas A, Wang Z, Agarwal S, Li Y, Tuchman Y, James CD, Marinella MJ, Yang JJ, et al. (2019). Parallel programming of an ionic floating-gate memory array for scalable neuromorphic computing. Science 364, 570–574. [DOI] [PubMed] [Google Scholar]
- 185.Sangwan VK, and Hersam MC (2020). Neuromorphic nanoelectronic materials. Nat. Nanotechnol 15, 517–528. [DOI] [PubMed] [Google Scholar]
- 186.Xia Q, and Yang JJ (2019). Memristive crossbar arrays for brain-inspired computing. Nat. Mater 18, 309–323. [DOI] [PubMed] [Google Scholar]
- 187.Park H-L, Lee Y, Kim N, Seo D-G, Go G-T, and Lee T-W (2020). Flexible Neuromorphic Electronics for Computing, Soft Robotics, and Neuroprosthetics. Adv. Mater 32, 1903558. [DOI] [PubMed] [Google Scholar]