Skip to main content
PLOS ONE logoLink to PLOS ONE
. 2020 Oct 27;15(10):e0241100. doi: 10.1371/journal.pone.0241100

A comparison of health care worker-collected foam and polyester nasal swabs in convalescent COVID-19 patients

Brian Hart 1,*, Yuan-Po Tu 2, Rachel Jennings 3, Prateek Verma 1, Leah R Padgett 4, Douglas Rains 4, Deneen Vojta 1, Ethan M Berke 1
Editor: Eric HY Lau5
PMCID: PMC7591034  PMID: 33108384

Abstract

Both polyester and foam nasal swabs were collected from convalescent COVID-19 patients at a single visit and stored in viral transport media (VTM), saline or dry. Sensitivity of each swab material and media combination were estimated, three by three tables were constructed to measure polyester and foam concordance, and cycle threshold (Ct) values were compared. 126 visits had polyester and foam swabs stored in viral transport media (VTM), 51 had swabs stored in saline, and 63 had a foam swab in VTM and a polyester swab stored in a dry tube. Polyester and foam swabs had an estimated sensitivity of 87.3% and 94.5% respectively in VTM, 87.5% and 93.8% respectively in saline, and 75.0% and 90.6% respectively for dry polyester and foam VTM. Polyester and foam Ct values were correlated, but polyester showed decreased performance for cases with a viral load near the detection threshold and higher Ct values on average.

Introduction

In the months since the Centers for Disease Control and Prevention began tracking cases of coronavirus disease 2019 (COVID-19), more than 6.5 million people have officially tested positive in the United States alone, resulting in more than 193,000 deaths [1]. The exponential spread of SARS-CoV-2 has resulted in an enormous demand for testing for the SARS-CoV-2 virus that causes the disease. This increased demand has put a strain on every level of the system from personal protective equipment (PPE) worn by health care workers while administering the test, to swabs for collecting the samples, to viral transport media (VTM) to store the sample until tested, to laboratory capacity [24]. Efforts to provide viable alternatives for each of these supply shortages are crucial to keeping the healthcare workforce safe and the testing throughput high.

The shortages of PPE and swabs are driven, in part, because the originally recommended nasopharyngeal (NP) test cannot be self-administered, requiring a change in PPE for each test. A recent study showed that patient-collected foam nasal swabs were comparable to health care worker-collected NP swabs for collecting SARS-CoV-2 virus, providing a safer and less invasive sampling method [5]. Similarly, another independent study demonstrated that patient-collected nasal swabs were comparable to health care worker-collected oropharyngeal or nasal swabs [6]. These results validate findings for influenza testing [7, 8]. While these results open the possibility of using patient-collected foam nasal swabs for sample collection, reducing the risk of viral exposure for health care workers, several issues remain. In the United States, foam nasal and nylon flocked swabs are not as readily available or mass produced as polyester swabs [9]. Additionally, swabs are typically stored and transported in VTM under refrigeration at or below 4°C. With VTM and swab supplies running low, and difficulties in obtaining sufficient refrigeration space for the massive number of samples arriving at the labs to be tested, testing requirements must be reevaluated to see if they can be safely altered [10, 11]. Recent work has shown that saline may be a suitable replacement for VTM [12]. Since these findings were released, the Food and Drug Administration (FDA) has updated their testing recommendations to allow for a wider variety of substances to be used for viral transport and stabilization and allow for self-collected foam nasal swabs when VTM and NP swabs are not available [13].

To address the dwindling supply of recommended swabs, we compared the relative performance of polyester and foam nasal swabs for collecting SARS-CoV-2, stored and transported either in VTM, saline, or in a dry tube.

Methods

Population and sample collection

Patients who have previously tested positive for SARS-CoV-2 at any site from the original study in Washington state were approached to return for additional testing [5]. A cohort of 63 positive patients returned 7–9, 14–18, and 28–31 days after their initial positive diagnosis for longitudinal sample collection, with swabbing continuing until all results for samples from a single visit were negative. A second cohort consisted of participants who also previously tested positive. They were contacted by ambulatory clinic staff and asked to return for a single visit as soon as possible.

People who agreed to participate in either study were consented by medical staff. Participants were evaluated at a designated ambulatory clinic site that only saw patients that were confirmed with a prior positive SARS-CoV-2 reverse transcription polymerase chain reaction (RT-PCR) test. Inclusion criteria included a previous positive SARS-CoV-2 test and the ability to consent and agree to participate in the study. People who were not able to demonstrate understanding of the study, not willing to commit to having all samples collected, had a history of nosebleed in the past 24 hours, nasal surgery in the past two weeks, chemotherapy treatment with documented low platelet and low white blood cell counts, or acute facial trauma were excluded from the study. Health care workers used a written consent form to explain the study and give eligible patients the opportunity to decline. This study protocol was deemed to be part of a minimal risk research protocol by the Office of Human Research Affairs at UnitedHealth Group.

In the first cohort of 63 SARS-CoV-2 positive individuals, three nasal swabs were collected by a health care worker from each patient: a foam swab and two polyester swabs. The first swab was gently inserted into the right nostril until resistance was met at the level of the turbinate (less than one inch into the nostril) and gentle pressure was applied to the outside nasal wall and the swab was rotated several times against the nasal wall and then slowly removed from the nostril. The second swab was then gently inserted into the left nostril and sampling was obtain in a similar manner. Next, the first swab was inserted into the left nostril and sampling was obtained in a similar manner. Finally, the second swab was inserted into the right nostril and sampling was obtained in a similar manner.

Left and right nostril randomization was based on the patient’s year of birth. For patient born in an odd numbered year, the left nostril was initially sampled with the foam swab followed by their right nostril being sampled by the polyester swab. Conversely, for patients born in an even numbered year, the right nostril was initially sampled by the foam swab followed by their left nostril being sampled by the polyester swab. The foam and polyester swabs were eluted in viral transport media, stored at 4°C, and sent to a reference laboratory for immediate testing.

After obtaining and storing the foam and first polyester swabs, a second polyester swab was used to sample both the right and left nostril in a similar manner to the previous swabs. The second polyester swab was then inserted in a dry tube (10 mL BD Vacutainer), stored at 4°C, and sent to a reference laboratory for immediate testing. Note that this sample collection procedure put the polyester swab stored in the dry tube at a disadvantage as it was always the third swab in a nostril and thus potentially had less virus available to sample. Due to participant testing constraints, the polyester swab to be stored in a dry tube was not collected from all patients after the first visit. The three swabs will be referred to as the foam, VTM polyester, and dry polyester swabs.

Patients in the second cohort were also convalescing from COVID-19. They were recruited immediately after their initial positive test result for SARS-CoV-2. During a single visit (between two and seven days after initial diagnosis), these patients were swabbed twice: once with a polyester swab and once with a foam swab. The order of sample collection was randomized and collected in the same manner as the first two swabs in the first cohort. Both samples were stored at 4°C and immediately shipped to the reference laboratory for testing. All nasal swab samples were collected by health care workers and marked with a patient study number, swab type, and media type (if any).

RT-PCR was performed by a reference laboratory using the TaqPathTM COVID-19 Combo Kit (ThermoFisher; A47814). Cycle threshold (Ct) values for all samples were reported back to the clinical sites based on three different RNA targets: S Gene, ORF1ab, and N Gene. A higher Ct value corresponds to a lower viral load. Per the reference laboratory standard protocols, a sample was considered positive if two or three Ct values were less than or equal to 37, negative if all three Ct values were greater than 37, and inconclusive otherwise. All inconclusive samples were retested and the results of the retest were used, per manufacturer protocol. While a Ct value of 37 was used to determine the clinical results of the test, RT-PCR continued until 40 cycles were completed. Therefore, the maximum Ct value in the data is 40. Foam swabs were stored in 1 mL of VTM or 2 mL of saline and vortexed for three to five seconds. Polyester swabs were stored in 3 mL of VTM or 2 mL of saline and vortexed for three to five seconds. Dry polyester swabs were eluted in 1 mL of phosphate-buffered saline (PBS) and vortexed for 30 seconds followed by a ten-minute incubation at room temperature.

Because of the volumetric differences among the three types of collection and/or elution media, mathematical adjustments to some Ct values were made in order to account for the differences in dilution of the various swab samples. By scaling (i.e., normalizing) Cts for both the 2 mL saline samples and the 3 mL VTM samples to match the 1 mL sample volume (VTM or 1 mL PBS dry swab elution), the sensitivity comparisons among the three sample collection volumes are more legitimate than using unnormalized data. Values were normalized as follows: to account for the 2-fold dilution difference between saline (2 mL) and dry polyester (1 mL PBS) or foam (1 mL VTM), 1 Ct was subtracted from all qPCR results for the former; similarly, to account for the 3-fold sample volume difference between the 3 mL VTM polyester swab samples and dry polyester (1 mL PBS) or foam (1 mL VTM), 1.585 Ct was subtracted from the former. Note that the specific Ct adjustments follow the formula 2-dCt = F, where dCt is the necessary Ct adjustment and F is the fold increase in volume. This formula assumes the assays perform at approximately 100% PCR efficiency during the geometric phase of amplification [14].

Statistical analysis

We report the 3x3 confusion matrix showing the results of the polyester swabs against the results of the foam swabs. We estimate the sensitivity of the polyester and foam swabs, considering a patient to be a true positive if either swab returned a positive result and negative otherwise, and report the 95% confidence intervals for the sensitivity estimates. Note that this definition treats all positives as true positives and, thus, implies a specificity of 100% for all swabs. While this is a strong assumption, it allows for the estimation of the sensitivity of each swab without assuming one swab type is a gold standard truth. Confusion matrices and sensitivity estimates are broken out by visit number where applicable. Fisher’s exact tests are used to calculate p-values to compare the sensitivity of foam and polyester swabs in different media types. P-values were not calculated for individual visits because of small sample sizes.

For each RNA target, we calculated the correlation of the normalized Ct values between the foam and polyester swabs for all pairs where at least one of the Ct values was < 40, indicating enough of the RNA target was detected to stop the RT-PCR machine short of the maximum Ct value. For the same subset of polyester/foam swab pairs with at least one Ct < 40 we also took the difference between the two Ct values and plotted the results in a box plot. Positive difference values indicate a higher Ct (and thus lower viral load) in the polyester samples and negative values indicate a higher Ct in the foam samples. All statistical analysis was performed using R version 3.6.1 [15].

Results

Of the 63 patients in the first cohort, 29 were positive at the 7–9 day visit (Visit 1), five were positive at the 14–18 day visit (Visit 2), and zero were positive at the 28–31 day visit (Visit 3). One patient’s swabs from the 28–31 day visit leaked in transport and were not tested. The second cohort consisted of 76 participants, of which 51 had their swab samples stored in saline and 25 had their swab samples stored in VTM. A summary of the patient demographics in the two groups can be found in Table 1.

Table 1. Cohort demographics.

Cohort 1 (N = 63) Cohort 2 (N = 76)
Age, median (IQR) 46 (36, 56.5) 45 (34.75, 51.25)
Gender
    Female, n (%) 34 (54.0%) 36 (47.3%)
    Male, n (%) 29 (46.0%) 40 (52.6%)
Days since diagnosis, mean (sd) 13.8 (5.1) 10.0 (6.4)
Days since first symptoms, mean (sd) 7.8 (1.6) 3.7 (3.1)

A summary of the demographics of the two cohorts. Cohort 1 consists of 63 patients who returned for follow-up visits until all swabs tested negative for up to three visits. Cohort 2 consists of 71 patients with a single visit each. All dry polyester swabs were collected from Cohort 1. All saline swabs were collected from Cohort 2. VTM swabs were collected from both cohorts.

Tests run on the polyester nasal swabs detected four fewer positive cases than foam swabs and had an estimated sensitivity (95% confidence interval) of 87.3% (74.9%, 94.3%) when stored in VTM compared to 94.5% (83.9%, 98.6%) for the foam swabs in VTM (p-value: 0.32, Table 2). Additionally, there were eight visits with an inconclusive VTM foam swab and a negative VTM polyester swab, but no visits with an inconclusive VTM polyester swab and a negative VTM foam swab. In saline, tests run on the polyester swabs detected two fewer positive cases and had an estimated sensitivity of 87.5% (70.1%, 95.9%) compared to 93.8% (77.8%, 98.9%) for foam (p-value: 0.67, Table 3). Tests run on the dry polyester swabs detected five fewer cases than foam swabs in VTM and had an estimated sensitivity of 75.0% (56.2%, 87.9%) compared to 90.6% (73.8% to 97.5%) for VTM foam from the same visits (p-value: 0.18, Table 4).

Table 2. VTM results.

VTM Polyester Polyester Sens. (95% CI) Foam Sens. (95% CI)
Visit 1 (7–9 Days) Positive Negative Inconclusive Total
Positive 42 5 0 47 45/50 47/50
Negative 3 32 0 35 90.0% 94.0%
Inconclusive 0 6 0 6 (77.4%, 96.2%) (82.5%, 98.4%)
Total 45 43 0 88
Visit 2 (14–18 Days) Positive Negative Inconclusive Total
Positive 3 1 1 5 3/5 5/5
Negative 0 27 0 27 60.0% 100%
Inconclusive 0 2 0 2 (17.0%, 92.7%) (46.2%, 100%)
VTM Foam
Total 3 30 1 34
Visit 3 (28–31 Days) Positive Negative Inconclusive Total
Positive 0 0 0 0
Negative 0 4 0 4 NA NA
Inconclusive 0 0 0 0
Total 0 4 0 4
All Time Points Positive Negative Inconclusive Total
Positive 45 6 1 52 48/55 52/55
Negative 3 63 0 66 87.3% 94.5%
Inconclusive 0 8 0 8 (74.9%, 94.3%) (83.9%, 98.6%)
Total 48 77 1 126 p-value: 0.32

A 3x3 table and estimated sensitivities for each visit and all visits combined of the test results for the foam and polyester nasal samples, both stored in VTM.

Table 3. Saline results.

Saline Polyester Polyester Sens. (95% CI) Foam Sens. (95% CI)
Positive Negative Inconclusive Total
Saline Foam Positive 26 4 0 30 28/32 30/32
Negative 2 17 1 20 87.5% 93.8%
Inconclusive 0 1 0 1 (70.1%, 95.9%) (77.8%, 98.9%)
Total 28 22 1 51 p-value: 0.67

A 3x3 table of the test results and estimated sensitivities with 95% confidence intervals for the foam and polyester nasal samples, both stored in saline. Note that the saline results were from a cohort with a single visit and are, thus, not broken out by visit.

Table 4. Dry results.

Dry Polyester Polyester Sens. (95% CI) Foam Sens. (95% CI)
Positive Negative Inconclusive Total
VTM Foam Positive 21 5 3 29 24/32 29/32
Negative 2 26 2 30 75.0% 90.6%
Inconclusive 1 3 0 4 (56.2%, 87.9%) (73.8%, 97.5%)
Total 24 34 5 63 p-value: 0.18

A 3x3 table and estimated sensitivities with 95% confidence intervals for the VTM foam and dry polyester nasal samples. Note that the dry swabs were only collected at the first visit, so results are not broken out by visit.

Correlations between foam and polyester Ct values were highest in the saline samples, followed by the VTM samples and then the dry polyester versus VTM foam samples (Fig 1). While Ct correlations were highest in saline, the foam Ct value was lower than the polyester Ct value more than 63% of the time for the saline swabs, indicating the increased viral loads estimated using foam swabs (Fig 2). Similar biases toward lower Ct values appeared in the VTM polyester versus VTM foam and dry polyester versus VTM foam comparisons. The tendency toward lower Ct values for foam swabs is also apparent in the paired Ct plots (Fig 3). Fig 3 is further broken out by samples taken less than ten days from symptom onset and samples taken at least 10 days from symptom onset in S1 and S2 Figs.

Fig 1. Ct correlation plots.

Fig 1

Plots showing the cycle threshold (Ct) values for each of the three RNA targets and three transport media. The black line represents the best fitting linear regression, the dashed blue line represents a perfect one-to-one relationship.

Fig 2. Ct difference boxplots.

Fig 2

Plots showing the difference in cycle threshold (Ct) of the polyester and foam swabs collected at the same visits. Positive values represent higher Ct values in the polyester swab. The dashed red line represents equivalent Ct values. The percentage of samples for which the foam swab has a lower Ct value is shown below each sub-plot.

Fig 3. Paired Ct plots.

Fig 3

Paired Ct plots showing the polyester and foam Ct values for each transport media and target gene combination considered. Swabs collected at the same visit are connected by a black line.

Discussion

Direct comparison of foam and polyester nasal swabs stored in VTM, in saline, or dry demonstrated decreased, but potentially adequate performance of polyester swabs in times of shortages. The estimated sensitivity of the polyester swabs was 87.3% and 87.5% in VTM and saline respectively compared to 94.5% and 93.8% for foam swabs. These calculations were conservative because they counted inconclusive polyester results as negatives for the purposes of sensitivity estimation, although no positive VTM foam swabs had a corresponding inconclusive VTM polyester swab. The assumption used for sensitivity estimation that any positive result was a true positive is reasonable when viewed in the light that all patients had previously tested positive for SARS-CoV-2.

The comparison of polyester and foam swabs did not differ between saline and VTM storage. While the estimated sensitivity above 87% may be deemed sufficient in times of a public health emergency, the strength of the findings is diminished for cases with a viral load near the positive/negative threshold of 37 cycles. This issue can be most clearly seen via the cluster of points along the upper edge of the right-hand side of the correlation plots. These points represent cases where tests run on foam swabs detected virus and the polyester did not. This finding is also reflected in the eight previously mentioned inconclusive VTM foam results that were negative for the VTM polyester swab. All three polyester versus foam swab comparisons exhibited a tendency for the foam swab to have lower Ct values than the polyester swab, indicating the foam swab’s superior ability to collect virus. Despite these reductions in performance, polyester swabs in VTM or saline may be an adequate sample collection method in cases where foam nasal and NP swabs have been entirely exhausted, a situation which exists in many locations.

While the dry polyester swabs appeared to show poor performance, they were put at a disadvantage by the study design. Because all dry polyester swabs were the third swab collecting a sample from a nostril at a given visit, the decrease in performance cannot be attributed solely to the swab type and lack of storage media. Dry swabs were only collected for the first follow-up visit in the cohort of 63 patients that also had swabs stored in VTM, resulting in a smaller sample size and different time since symptom onset than the other comparisons. Dry swabs have been found to have high sensitivity in other diagnostic settings and should not be ruled out entirely based on this study [16].

The current study has several limitations. All participants were convalescing COVID-19 patients, and the time from first symptoms and first diagnosis to test date varied from test to test. As patients progressed further from their diagnosis date their viral loads dropped, creating more cases where the Ct values were near the border of detection. Other research has suggested viral load may have already peaked at the time of diagnosis followed by a slow decline over time [17]. The samples from this study may not be representative of testing in newly infected patients who are seeking their first SARS-CoV-2 test. Additionally, because no NP swabs were obtained for this study, performance of the polyester swab cannot be directly compared to the FDA’s preferred swabbing method [13]. Although final Ct values were adjusted for varying amounts of transport media, imprecision in these adjustments could confound data interpretation; specifically, comparisons between the mathematically corrected samples (2 mL and 3 mL) and the non-corrected samples (1 mL) present the greatest risk for error. Finally, the study was not designed to compare the performance of VTM and saline or to estimate an interaction of swab type and transport media type.

Despite these limitations, polyester swabs stored in VTM or saline may be a viable sample collection method for COVID-19 testing, especially in light of the shortages of other swab types. The viability of polyester swabs is most clearly demonstrated via the high correlation between polyester and foam Ct values from the same visit. Any recommendation for polyester swab usage should bear in mind that the decrease in performance near the border of detection may lead to false negatives in patients with low viral loads.

Supporting information

S1 Fig. Paired Ct plots: Newly symptomatic.

Paired Ct plots showing the polyester and foam Ct values for each transport media and target gene combination considered for swabs from patients less than 10 days from symptom onset at the time of sample collection. Swabs collected at the same visit are connected by a black line.

(TIF)

S2 Fig. Paired Ct plots: Not newly symptomatic.

Paired Ct plots showing the polyester and foam Ct values for each transport media and target gene combination considered for swabs from patients at least 10 days from symptom onset at the time of sample collection. Swabs collected at the same visit are connected by a black line.

(TIF)

Acknowledgments

The authors acknowledge the contributions of James S Elliott and Lauren A. Kennington from Quantigen Biosciences, Lindsay Nelson from UnitedHealth Group, and Garrett Galbreath, the health care workers, and staff from The Everett Clinic.

Data Availability

The data cannot be made publicly available due to patient privacy concerns arising from the small sample size of patients collected in a specific timeframe and region of the country. Deidentified data will still be shared upon reasonable request by contacting the UnitedHealth Group Office of Human Research affairs at ohra_uhg@uhg.com or Tracy Ziolek at Tracy_Ziolek@uhg.com. All legal and ethical questions regarding the sharing of this data can also be directed to the same sources.

Funding Statement

Laboratory testing was conducted with financial support from Thermo Fisher Scientific. Brian Hart, Rachel Jennings, Prateek Verma, Deneen Vojta, and Ethan Berke were employees of UnitedHealth Group during the design and analysis of the study and initial drafting of the manuscript. Yuan-Po Tu is an employee of The Everett Clinic, which is a subsidiary of UnitedHealth Group. Leah Padgett and Douglas Rains are employees of Quantigen Biosciences and have performed contract services for Thermo Fisher Scientific. UnitedHealth Group, The Everett Clinic, and Quantigen Biosciences provided support in the form of salaries for the authors, but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.

References

  • 1.Coronavirus Disease 2019 (COVID-19) in the US. Centers for Disease Control and Prevention, accessed September 15, 2020. https://www.cdc.gov/coronavirus/2019-ncov/cases-in-us.html
  • 2.Seelye KQ, Jacobs A, Becker J, Arango T. Doctors and Governors Vie for Masks in Cloak- and Dagger Deals. New York Times. April 20, 2020. [Google Scholar]
  • 3.Artenstein AW. In Pursuit of PPE. New England Journal of Medicine. 2020. April 17:e46 10.1056/NEJMc2010025 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4.A joint statement by the Association of State and Territorial Health Officials, Association of Public Health Laboratories, and Council of State and Territorial Epidemiologists. COVID-19 Testing Needs to Be Limited to Priority Groups Until Sufficient Testing Supplies and Personal Protective Equipment is Available Nationwide. Association of State and Territorial Health Officials. March 20, 2020. [Google Scholar]
  • 5.Tu YP, Jennings R, Hart B, Cangelosi GA, Wood RC, Wehber K, et al. Swabs Collected by Patients or Health Care Workers for SARS-CoV-2 Testing. New England Journal of Medicine. 2020. June 3. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Altamirano J, Govindarajan P, Blomkalns AL, Kushner LE, Stevens BA, Pinsky BA, et al. Assessment of Sensitivity and Specificity of Patient-Collected Lower Nasal Specimens for Sudden Acute Respiratory Syndrome Coronavirus 2 Testing. JAMA Network Open. 2020. June 1;3(6):e2012005–. 10.1001/jamanetworkopen.2020.12005 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7.Dhiman N, Miller RM, Finley JL, Sztajnkrycer MD, Nestler DM, Boggust AJ, et al. Effectiveness of patient-collected swabs for influenza testing. In Mayo Clinic Proceedings 2012. June 1 (Vol. 87, No. 6, pp. 548–554). Elsevier. 10.1016/j.mayocp.2012.02.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 8.Larios OE, Coleman BL, Drews SJ, Mazzulli T, Borgundvaag B, Green K, et al. Self-collected mid-turbinate swabs for the detection of respiratory viruses in adults with acute respiratory illnesses. PloS one. 2011;6(6). 10.1371/journal.pone.0021335 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9.Webber L and Jewett C. Testing Swabs Run In Short Supply As Makers Try To Speed Up Production. NPR. March 18, 2020. [Google Scholar]
  • 10.Viral Transport Medium (VTM) Supplies Needed. American Academy of Dermatology. https://assets.ctfassets.net/1ny4yoiyrqia/5gCfl5bSWjHwl3ZLGEg2Xl/0be62e6c8a876254b98f55c85c08489a/VTM_Supplies_Needed.pdf
  • 11.Perrone M. US virus testing faces new headwind: Lab supply shortages. The Associated Press; March 21, 2020. [Google Scholar]
  • 12.Rodino KG, Espy MJ, Buckwalter SP, Walchak RC, Germer JJ, Fernholz E, et al. Evaluation of saline, phosphate buffered saline and minimum essential medium as potential alternatives to viral transport media for SARS-CoV-2 testing. Journal of clinical microbiology. 2020. March 30 10.1128/JCM.00590-20 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.FAQs on Diagnostic Testing for SARS-CoV-2. The Food and Drug Administration, accessed on April 16, 2020. https://www.fda.gov/medical-devices/emergency-situations-medical-devices/faqs-diagnostic-testing-sars-cov-2
  • 14.Schmittgen TD, Livak KJ. Analyzing real-time PCR data by the comparative C T method. Nature protocols. 2008. June;3(6):1101 10.1038/nprot.2008.73 [DOI] [PubMed] [Google Scholar]
  • 15.R: The R Project for Statistical Computing [Internet]. [cited 2020 Mar 21]. (https://www.r-project.org/)
  • 16.Campbell AP, Kuypers J, Englund JA, Guthrie KA, Corey L, Boeckh M. Self-collection of foam nasal swabs for respiratory virus detection by PCR among immunocompetent subjects and hematopoietic cell transplant recipients. Journal of clinical microbiology. 2013. January 1;51(1):324–7. 10.1128/JCM.02871-12 [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 17.Wölfel R, Corman VM, Guggemos W, Seilmaier M, Zange S, Müller MA, et al. Virological assessment of hospitalized patients with COVID-2019. Nature. 2020. April 1:1–0. 10.1038/s41586-020-2196-x [DOI] [PubMed] [Google Scholar]

Decision Letter 0

Eric HY Lau

2 Sep 2020

PONE-D-20-13951

A comparison of health care worker-collected foam and polyester nasal swabs in convalescent COVID-19 patients

PLOS ONE

Dear Dr. Hart,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

The Authors are expected to address all the criticisms by all Reviewers. In particular, please discuss if different volume of VTM or saline used, or normalization of Ct values would affect the results (Reviewer #1) and provide basic clinical information of the two cohorts (Reviewer #2). In additional to the above comments, please address,

  1. Please present the number of samples collected by day of visit group

  2. Tables 1-3 have pooled samples collected at different days after diagnosis. However, viral shedding decreased over time and hence test sensitivity is unlikely constant over sampling dates. Please further assess the sensitivity by the time of sampling.

  3. The authors have assessed sensitivity of different sampling methods. Could the authors explain why specificity was not assessed?

Please submit your revised manuscript by Oct 17 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Eric HY Lau, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments:

The Authors are expected to address all the criticisms by all Reviewers. In particular, please discuss if different volume of VTM or saline used, or normalization of Ct values would affect the results (Reviewer #1) and provide basic clinical information of the two cohorts (Reviewer #2). In additional to the above comments, please address,

1. Please present the number of samples collected by day of visit group

2. Tables 1-3 have pooled samples collected at different days after diagnosis. However, viral shedding decreased over time and hence test sensitivity is unlikely constant over sampling dates. Please further assess the sensitivity by the time of sampling.

3. The authors have assessed sensitivity of different sampling methods. Could the authors explain why specificity was not assessed?

Journal Requirements:

When submitting your revision, we need you to address these additional requirements.

1. Please ensure that your manuscript meets PLOS ONE's style requirements, including those for file naming. The PLOS ONE style templates can be found at

https://journals.plos.org/plosone/s/file?id=wjVg/PLOSOne_formatting_sample_main_body.pdf and

https://journals.plos.org/plosone/s/file?id=ba62/PLOSOne_formatting_sample_title_authors_affiliations.pdf

2. Thank you for stating the following in the Financial Disclosure section:

"Laboratory testing was conducted with financial support from Thermo Fisher Scientific."

We note that one or more of the authors are employed by a commercial company: UnitedHealth Group and Quantigen Biosciences.

2.1. Please provide an amended Funding Statement declaring this commercial affiliation, as well as a statement regarding the Role of Funders in your study. If the funding organization did not play a role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript and only provided financial support in the form of authors' salaries and/or research materials, please review your statements relating to the author contributions, and ensure you have specifically and accurately indicated the role(s) that these authors had in your study. You can update author roles in the Author Contributions section of the online submission form.

Please also include the following statement within your amended Funding Statement.

“The funder provided support in the form of salaries for authors [insert relevant initials], but did not have any additional role in the study design, data collection and analysis, decision to publish, or preparation of the manuscript. The specific roles of these authors are articulated in the ‘author contributions’ section.”

If your commercial affiliation did play a role in your study, please state and explain this role within your updated Funding Statement.

2.2. Please also provide an updated Competing Interests Statement declaring this commercial affiliation along with any other relevant declarations relating to employment, consultancy, patents, products in development, or marketed products, etc.  

Within your Competing Interests Statement, please confirm that this commercial affiliation does not alter your adherence to all PLOS ONE policies on sharing data and materials by including the following statement: "This does not alter our adherence to  PLOS ONE policies on sharing data and materials.” (as detailed online in our guide for authors http://journals.plos.org/plosone/s/competing-interests) . If this adherence statement is not accurate and  there are restrictions on sharing of data and/or materials, please state these. Please note that we cannot proceed with consideration of your article until this information has been declared.

Please include both an updated Funding Statement and Competing Interests Statement in your cover letter. We will change the online submission form on your behalf.

Please know it is PLOS ONE policy for corresponding authors to declare, on behalf of all authors, all potential competing interests for the purposes of transparency. PLOS defines a competing interest as anything that interferes with, or could reasonably be perceived as interfering with, the full and objective presentation, peer review, editorial decision-making, or publication of research or non-research articles submitted to one of the journals. Competing interests can be financial or non-financial, professional, or personal. Competing interests can arise in relationship to an organization or another person. Please follow this link to our website for more details on competing interests: http://journals.plos.org/plosone/s/competing-interests

3. We noted in your submission details that a portion of your manuscript may have been presented or published elsewhere.

"A preprint of the manuscript has been posted on medRxiv, but the manuscript has not other wise been published. " ext-link-type="uri" xlink:type="simple">https://www.medrxiv.org/content/10.1101/2020.04.28.20083055v1"

Please clarify whether this  publication was peer-reviewed and formally published. If this work was previously peer-reviewed and published, in the cover letter please provide the reason that this work does not constitute dual publication and should be included in the current manuscript.

4. We note that you have indicated that data from this study are available upon request. PLOS only allows data to be available upon request if there are legal or ethical restrictions on sharing data publicly. For information on unacceptable data access restrictions, please see http://journals.plos.org/plosone/s/data-availability#loc-unacceptable-data-access-restrictions.

In your revised cover letter, please address the following prompts:

a) If there are ethical or legal restrictions on sharing a de-identified data set, please explain them in detail (e.g., data contain potentially identifying or sensitive patient information) and who has imposed them (e.g., an ethics committee). Please also provide contact information for a data access committee, ethics committee, or other institutional body to which data requests may be sent.

b) If there are no restrictions, please upload the minimal anonymized data set necessary to replicate your study findings as either Supporting Information files or to a stable, public repository and provide us with the relevant URLs, DOIs, or accession numbers. Please see http://www.bmj.com/content/340/bmj.c181.long for guidelines on how to de-identify and prepare clinical data for publication. For a list of acceptable repositories, please see http://journals.plos.org/plosone/s/data-availability#loc-recommended-repositories.

We will update your Data Availability statement on your behalf to reflect the information you provide.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

2. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: Yes

Reviewer #2: Yes

**********

4. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

5. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: A timely trial of polyester vs foam nasal swabs stored and/or tested using VTM, saline, or dry tubes with buffered saline.

Comment only: It is unfortunate that you could not have treated the samples more similarly. I’m not sure why you chose to use different amount of VTM and saline, for instance, at the lab. Correcting for/normalizing the Ct values is adequate but not as convincing.

Did you try to correct for the first/second collection of swabs in the analysis? It is not necessary, just a thought.

Swabs do not detect virus, but rather “collect” them. Consider changing the wording.

Discussion: line 186, please add, “in times of shortages” or something to that effect.

Reviewer #2: In the context of the COVID-19 pandemic, the simple, convenient and effective detection methods are worth promoting. The polyester swabs stored in VTM or saline may be a viable sample collection method for COVID-19.

But in the RT-PCR part, the author should show more details.

In addition, the basic clinical information of the two cohorts should be presented, such as sex, BMI, age, symptoms and so on.

**********

6. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Oct 27;15(10):e0241100. doi: 10.1371/journal.pone.0241100.r002

Author response to Decision Letter 0


16 Sep 2020

The authors would like to thank the reviewers and editor for their careful review and consideration of our manuscript titled “A comparison of health care worker-collected foam and polyester nasal swabs in convalescent COVID-19 patients”. Below we provide a point by point response to the reviewer comments below:

• Please discuss if different volume of VTM or saline used, or normalization of Ct values would affect the results (Reviewer #1).

Yes, using different sample volumes and applying mathematical corrections is not ideal and could affect the results. However, considering the urgent need for these studies, the standard 1-3 mL volumes used for clinical sample collection, and the valid mathematical approach for performing fold change calculations on qPCR, we felt normalization of the data was the best approach. For clarification, we have added text to the methods section and to the discussion.

Methods:

“Because of the volumetric differences among the three types of collection and/or elution media, mathematical adjustments to some Ct values were made in order to account for the differences in dilution of the various swab samples. By scaling (i.e., normalizing) Cts for both the 2 mL saline samples and the 3 mL VTM samples to match the 1 mL sample volume (VTM or 1 mL PBS dry swab elution), the sensitivity comparisons among the three sample collection volumes are more legitimate than using unnormalized data. Values were normalized as follows: to account for the 2-fold dilution difference between saline (2 mL) and dry polyester (1 mL PBS) or foam (1 mL VTM), 1 Ct was subtracted from all qPCR results for the former; similarly, to account for the 3-fold sample volume difference between the 3 mL VTM polyester swab samples and dry polyester (1 mL PBS) or foam (1 mL VTM), 1.585 Ct was subtracted from the former. Note that the specific Ct adjustments follow the formula 2-dCt=F, where dCt is the necessary Ct adjustment and F is the fold increase in volume. This formula assumes the assays perform at approximately 100% PCR efficiency during the geometric phase of amplification.”

Discussion:

“Although final Ct values were adjusted for varying amounts of transport media, imprecision in these adjustments could confound data interpretation; specifically, comparisons between the mathematically corrected samples (2 mL and 3 mL) and the non-corrected samples (1 mL) present the greatest risk for error.”

• Provide basic clinical information of the two cohorts (Reviewer #2).

Thank you for the suggestion. We have added in a new table one with basic clinical/demographic information available on the two cohorts.

• Please present the number of samples collected by day of visit group

These values can now be found in the expanded version of Tables 2 (previously Table 1).

• Tables 1-3 have pooled samples collected at different days after diagnosis. However, viral shedding decreased over time and hence test sensitivity is unlikely constant over sampling dates. Please further assess the sensitivity by the time of sampling.

Table 2 (previously called Table 1) has been expanded to breakout the results by the three different visits. Note that Tables 3 and 4 contain results which were collected at a single visit, so no similar table breakout was necessary/possible.

• The authors have assessed sensitivity of different sampling methods. Could the authors explain why specificity was not assessed?

When estimating sensitivity we assume that all positives were true positives. The primary advantage of this approach is that it allows for the estimation of sensitivity without assuming a single method is a gold standard “truth”. The disadvantage of our approach is that assuming all positives are true positives implies that all methods have a specificity of 100%. One could alternatively treat a single method (e.g. foam VTM) as the truth and compare the other methods when estimating sensitivity. This approach also involves strong assumptions and, while it allows for the estimation of both sensitivity and specificity of the non-reference methods, it implies 100% sensitivity and specificity of the reference method. Based on these tradeoffs, we chose to treat any positive as a true positive even though it precludes accurate specificity estimation.

In addition to the edits made in response to reviewer comments we have made some additional changes to the description of the procedures. These changes were made so that the manuscript accurately describes how the study was carried out in practice. The sample collection procedures are now more accurately and clearly described in the manuscript. We have also changed the method used to calculate the 95% confidence intervals from the Normal approximation to the Wilson Score method, which has been shown to have better coverage. This change was motivated by the need for more stable confidence interval estimates given the low sample sizes, especially when broken out by visit.

Attachment

Submitted filename: Response to Reviewers_9_15.docx

Decision Letter 1

Eric HY Lau

8 Oct 2020

PONE-D-20-13951R1

A comparison of health care worker-collected foam and polyester nasal swabs in convalescent COVID-19 patients

PLOS ONE

Dear Dr. Hart,

Thank you for submitting your manuscript to PLOS ONE. After careful consideration, we feel that it has merit but does not fully meet PLOS ONE’s publication criteria as it currently stands. Therefore, we invite you to submit a revised version of the manuscript that addresses the points raised during the review process.

The Authors have addressed most of the reviewers’ comments satisfactorily, but are expected to address the remaining criticisms by Reviewer #1.

Please submit your revised manuscript by Nov 22 2020 11:59PM. If you will need more time than this to complete your revisions, please reply to this message or contact the journal office at plosone@plos.org. When you're ready to submit your revision, log on to https://www.editorialmanager.com/pone/ and select the 'Submissions Needing Revision' folder to locate your manuscript file.

Please include the following items when submitting your revised manuscript:

  • A rebuttal letter that responds to each point raised by the academic editor and reviewer(s). You should upload this letter as a separate file labeled 'Response to Reviewers'.

  • A marked-up copy of your manuscript that highlights changes made to the original version. You should upload this as a separate file labeled 'Revised Manuscript with Track Changes'.

  • An unmarked version of your revised paper without tracked changes. You should upload this as a separate file labeled 'Manuscript'.

If you would like to make changes to your financial disclosure, please include your updated statement in your cover letter. Guidelines for resubmitting your figure files are available below the reviewer comments at the end of this letter.

If applicable, we recommend that you deposit your laboratory protocols in protocols.io to enhance the reproducibility of your results. Protocols.io assigns your protocol its own identifier (DOI) so that it can be cited independently in the future. For instructions see: http://journals.plos.org/plosone/s/submission-guidelines#loc-laboratory-protocols

We look forward to receiving your revised manuscript.

Kind regards,

Eric HY Lau, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (if provided):

The Authors have addressed most of the reviewers’ comments satisfactorily, but are expected to address the remaining criticisms by Reviewer #1.

[Note: HTML markup is below. Please do not edit.]

Reviewers' comments:

Reviewer's Responses to Questions

Comments to the Author

1. If the authors have adequately addressed your comments raised in a previous round of review and you feel that this manuscript is now acceptable for publication, you may indicate that here to bypass the “Comments to the Author” section, enter your conflict of interest statement in the “Confidential to Editor” section, and submit your "Accept" recommendation.

Reviewer #1: (No Response)

Reviewer #2: All comments have been addressed

**********

2. Is the manuscript technically sound, and do the data support the conclusions?

The manuscript must describe a technically sound piece of scientific research with data that supports the conclusions. Experiments must have been conducted rigorously, with appropriate controls, replication, and sample sizes. The conclusions must be drawn appropriately based on the data presented.

Reviewer #1: Yes

Reviewer #2: Yes

**********

3. Has the statistical analysis been performed appropriately and rigorously?

Reviewer #1: No

Reviewer #2: Yes

**********

4. Have the authors made all data underlying the findings in their manuscript fully available?

The PLOS Data policy requires authors to make all data underlying the findings described in their manuscript fully available without restriction, with rare exception (please refer to the Data Availability Statement in the manuscript PDF file). The data should be provided as part of the manuscript or its supporting information, or deposited to a public repository. For example, in addition to summary statistics, the data points behind means, medians and variance measures should be available. If there are restrictions on publicly sharing data—e.g. participant privacy or use of data from a third party—those must be specified.

Reviewer #1: No

Reviewer #2: Yes

**********

5. Is the manuscript presented in an intelligible fashion and written in standard English?

PLOS ONE does not copyedit accepted manuscripts, so the language in submitted articles must be clear, correct, and unambiguous. Any typographical or grammatical errors should be corrected at revision, so please note any specific errors here.

Reviewer #1: Yes

Reviewer #2: Yes

**********

6. Review Comments to the Author

Please use the space provided to explain your answers to the questions above. You may also include additional comments for the author, including concerns about dual publication, research ethics, or publication ethics. (Please upload your review as an attachment if it exceeds 20,000 characters)

Reviewer #1: Thank you for the considerable revision you made. This revision is substantially clearer and more complete.

A few comments:

line 249-250 in the discussion: there is a sentence fragment that needs to be removed or completed.

Although I will not "require" it, it would be helpful to provide p-values for comparisons across swab types and/or storage.

Reviewer #2: 1. The study presents the results of original research.

2. Results reported have not been published elsewhere.

3. Experiments, statistics, and other analyses are performed to a high technical standard and are described in sufficient detail.

4. Conclusions are presented in an appropriate fashion and are supported by the data.

5. The article is presented in an intelligible fashion and is written in standard English.

6. The research meets all applicable standards for the ethics of experimentation and research integrity.

7. The article adheres to appropriate reporting guidelines and community standards for data availability.

**********

7. PLOS authors have the option to publish the peer review history of their article (what does this mean?). If published, this will include your full peer review and any attached files.

If you choose “no”, your identity will remain anonymous but your review may still be made public.

Do you want your identity to be public for this peer review? For information about this choice, including consent withdrawal, please see our Privacy Policy.

Reviewer #1: No

Reviewer #2: No

[NOTE: If reviewer comments were submitted as an attachment file, they will be attached to this email and accessible via the submission site. Please log into your account, locate the manuscript record, and check for the action link "View Attachments". If this link does not appear, there are no attachment files.]

While revising your submission, please upload your figure files to the Preflight Analysis and Conversion Engine (PACE) digital diagnostic tool, https://pacev2.apexcovantage.com/. PACE helps ensure that figures meet PLOS requirements. To use PACE, you must first register as a user. Registration is free. Then, login and navigate to the UPLOAD tab, where you will find detailed instructions on how to use the tool. If you encounter any issues or have any questions when using PACE, please email PLOS at figures@plos.org. Please note that Supporting Information files do not need this step.

PLoS One. 2020 Oct 27;15(10):e0241100. doi: 10.1371/journal.pone.0241100.r004

Author response to Decision Letter 1


8 Oct 2020

The authors would like to thank the reviewers and editor for their careful review and consideration of our manuscript titled “A comparison of health care worker-collected foam and polyester nasal swabs in convalescent COVID-19 patients”. Below we provide a point by point response to the reviewer comments below:

• line 249-250 in the discussion: there is a sentence fragment that needs to be removed or completed.

Thank you for catching the error. The sentence fragment has been removed in the resubmitted manuscript.

• Although I will not "require" it, it would be helpful to provide p-values for comparisons across swab types and/or storage.

We have now provided p-values from a Fisher’s Exact test for the three polyester versus foam sensitivity comparisons. The p-values are included in the main text and in Tables 2-4. In Table 2, p-values were not included for the visit specific sensitivity comparisons due to small sample sizes. Two sentences have been added to the Statistical analysis section to explain methods used for p-value calculations.

No additional changes were made to the manuscript. We again thank the reviewers for their constructive feedback which has led to a much improved manuscript.

Attachment

Submitted filename: Response to Reviewers Round 2.docx

Decision Letter 2

Eric HY Lau

9 Oct 2020

A comparison of health care worker-collected foam and polyester nasal swabs in convalescent COVID-19 patients

PONE-D-20-13951R2

Dear Dr. Hart,

We’re pleased to inform you that your manuscript has been judged scientifically suitable for publication and will be formally accepted for publication once it meets all outstanding technical requirements.

Within one week, you’ll receive an e-mail detailing the required amendments. When these have been addressed, you’ll receive a formal acceptance letter and your manuscript will be scheduled for publication.

An invoice for payment will follow shortly after the formal acceptance. To ensure an efficient process, please log into Editorial Manager at http://www.editorialmanager.com/pone/, click the 'Update My Information' link at the top of the page, and double check that your user information is up-to-date. If you have any billing related questions, please contact our Author Billing department directly at authorbilling@plos.org.

If your institution or institutions have a press office, please notify them about your upcoming paper to help maximize its impact. If they’ll be preparing press materials, please inform our press team as soon as possible -- no later than 48 hours after receiving the formal acceptance. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information, please contact onepress@plos.org.

Kind regards,

Eric HY Lau, Ph.D.

Academic Editor

PLOS ONE

Additional Editor Comments (optional):

Reviewers' comments:

Acceptance letter

Eric HY Lau

16 Oct 2020

PONE-D-20-13951R2

A comparison of health care worker-collected foam and polyester nasal swabs in convalescent COVID-19 patients

Dear Dr. Hart:

I'm pleased to inform you that your manuscript has been deemed suitable for publication in PLOS ONE. Congratulations! Your manuscript is now with our production department.

If your institution or institutions have a press office, please let them know about your upcoming paper now to help maximize its impact. If they'll be preparing press materials, please inform our press team within the next 48 hours. Your manuscript will remain under strict press embargo until 2 pm Eastern Time on the date of publication. For more information please contact onepress@plos.org.

If we can help with anything else, please email us at plosone@plos.org.

Thank you for submitting your work to PLOS ONE and supporting open access.

Kind regards,

PLOS ONE Editorial Office Staff

on behalf of

Dr. Eric HY Lau

Academic Editor

PLOS ONE

Associated Data

    This section collects any data citations, data availability statements, or supplementary materials included in this article.

    Supplementary Materials

    S1 Fig. Paired Ct plots: Newly symptomatic.

    Paired Ct plots showing the polyester and foam Ct values for each transport media and target gene combination considered for swabs from patients less than 10 days from symptom onset at the time of sample collection. Swabs collected at the same visit are connected by a black line.

    (TIF)

    S2 Fig. Paired Ct plots: Not newly symptomatic.

    Paired Ct plots showing the polyester and foam Ct values for each transport media and target gene combination considered for swabs from patients at least 10 days from symptom onset at the time of sample collection. Swabs collected at the same visit are connected by a black line.

    (TIF)

    Attachment

    Submitted filename: Response to Reviewers_9_15.docx

    Attachment

    Submitted filename: Response to Reviewers Round 2.docx

    Data Availability Statement

    The data cannot be made publicly available due to patient privacy concerns arising from the small sample size of patients collected in a specific timeframe and region of the country. Deidentified data will still be shared upon reasonable request by contacting the UnitedHealth Group Office of Human Research affairs at ohra_uhg@uhg.com or Tracy Ziolek at Tracy_Ziolek@uhg.com. All legal and ethical questions regarding the sharing of this data can also be directed to the same sources.


    Articles from PLoS ONE are provided here courtesy of PLOS

    RESOURCES