Validation of the clinical applicability of knowledge‐based planning models in single‐isocenter volumetric‐modulated arc therapy for multiple brain metastases

Noriko Kishi 1, Mitsuhiro Nakamura 1,2,, Hideaki Hirashima 1, Nobutaka Mukumoto 1, Keiichi Takehana 1, Megumi Uto 1, Yukinori Matsuo 1, Takashi Mizowaki 1
PMCID: PMC7592973  PMID: 32951337

Abstract

Purpose

To validate the clinical applicability of knowledge‐based (KB) planning in single‐isocenter volumetric‐modulated arc therapy (VMAT) for multiple brain metastases using the k‐fold cross‐validation (CV) method.

Methods

This study comprised 60 consecutive patients with multiple brain metastases treated with single‐isocenter VMAT (28 Gy in five fractions). The patients were divided randomly into five groups (Groups 1–5). The data of Groups 1–4 were used as the training and validation dataset and those of Group 5 were used as the testing dataset. Four KB models were created from three of the training and validation datasets and then applied to the remaining Groups as the fourfold CV phase. As the testing phase, the final KB model was applied to Group 5 and the dose distributions were calculated with a single optimization process. The dose‐volume indices (DVIs), modified Ian Paddick Conformity Index (mIPCI), modulation complexity scores for VMAT plans (MCSv), and the total number of monitor units (MUs) of the final KB plan were compared to those of the clinical plan (CL) using a paired Wilcoxon signed‐rank test.

Results

In the fourfold CV phase, no significant differences were observed in the DVIs among the four KB plans (KBPs). In the testing phase, the final KB plan was statistically equivalent to the CL, except for planning target volumes (PTVs) D2% and D50%. The differences between the CL and KBP in terms of the PTV D99.5%, normal brain, and Dmax to all organs at risk (OARs) were not significant. The KBP achieved a lower total number of MUs and higher MCSv than the CL with no significant difference.

Conclusions

We demonstrated that a KB model in a single‐isocenter VMAT for multiple brain metastases was equivalent in dose distribution, MCSv, and total number of MUs to a CL with a single optimization.

Keywords: k‐fold cross‐validation, knowledge‐based planning, multiple brain metastases, single‐isocenter VMAT

1. Introduction

Brain metastases are the most common intracranial tumors, which are present in approximately 2% of cancer patients at the time of the primary diagnosis, with a prevalence ranging from 15% (small cell lung cancer) to < 0.1% (prostate cancer) and a median overall survival of less than 1 yr. 1 The incidence of brain metastases is observed to increase because of the prevalence of magnetic resonance imaging (MRI), which has improved the rates of detection. Moreover, the development of systemic therapy has improved the survival after the primary diagnosis of brain metastases. The standard treatment strategy for multiple brain metastases is whole‐brain radiotherapy (WBRT); however, WBRT leads to the deterioration of neurocognitive function and patient quality of life. 2

Stereotactic radiosurgery (SRS) is an irradiation technique that requires the precise fixation of the patient, localization of the target, and the application of highly biologically effective radiation doses. 3 Using SRS, the irradiated dose to the normal brain can be reduced, and re‐irradiation can be considered even in post‐SRS patients who experience intracranial recurrence. In patients with 1–4 brain metastases, SRS causes lesser neurocognitive deterioration compared to SRS plus WBRT; additionally, there is no difference in the overall survival between the SRS and SRS plus WBRT. 4 , 5 , 6 Although the clinical advantage of SRS over WBRT is controversial in patients with more than four brain metastases, SRS is considered an effective and safe treatment option, especially in patients with a favorable prognosis. 7

Linac‐based single‐isocenter volumetric‐modulated arc therapy (VMAT) can accomplish clinically equivalent dose distributions to gamma knife radiosurgery, but with a reduced delivery time. 8 However, the treatment planning of single‐isocenter VMAT for multiple brain metastases is generally time‐ and resource‐intensive. When performing the optimization, planners manually determine the optimization parameters; nevertheless, it is cumbersome to identify the optimization parameters due to the presence of multiple target volumes with different sizes and characteristics. In addition, for one target located adjacent to another in the same plane, several optimizations are required to minimize the dose spillage between the target volumes; as there is no clearly defined goal, the process can seem endless.

Recently, knowledge‐based (KB) planning has become available for clinical use as a tool assisting inverse planning. 9 KB planning uses databases constructed according to the anatomical positions and doses of previously treated patients and establishes a machine‐learned model for dose‐volume histogram (DVH) estimation. The model estimates DVHs to create treatment plans for new patients. A recently published review paper showed that, generally, KB models produce plans of comparable quality to those of expert planners, while also reducing the time and effort required to generate plans for various disease sites. 9

When verifying machine‐learned models, the predictive performance of the models should be evaluated on unseen data. 10 The hold‐out validation method is a simple approach for model evaluation where the data are categorized into two subsamples: training and testing. However, this method is prone to subsample bias and is inadequate for small sample sizes. The k‐fold cross‐validation (CV) method is another technique for the evaluation and comparison of machine‐learned models. The k‐fold CV method splits the data into k equal sized subgroups; one subgroup is used as a validation group and the remaining subgroups are used as a training dataset. With k‐fold CV, the whole dataset can be used for both training and validation, and this method is affected by pessimistic bias in a lesser manner, compared to the hold‐out method. In general, k‐fold CV is usually considered as the preferred method because it allows the model to train via multiple train‐test splits providing a better indication of how well the model will perform on unseen data. Most studies on KB planning employed the hold‐out method for prostate/uterine cancer, head and neck cancer, and lung, liver, and primary brain tumors. 11 , 12 , 13 , 14 , 15 To date, however, there has been insufficient literature comparing the performance of KB models through k‐fold CV, or on the application of KB planning to multiple brain metastases.

The purpose of this study was to validate the clinical applicability of KB planning in single‐isocenter VMAT for multiple brain metastases using the k‐fold CV method.

2. Methods

2.1. Patients

This study included 60 consecutive patients with multiple brain metastases treated using single‐isocenter VMAT (28 Gy in five fractions) in our institute from October 2015 to December 2018. The study participants included 29 men and 31 women, with a median age of 64 yr (range: 26–88 yr). In total, there were 317 multiple brain metastases. The patients were divided equally and randomly into five groups (Groups 1–5) using Python scripts (Python version 3.7).

This study was performed in accordance with the Declaration of Helsinki (1975, as revised in 2013). Written informed consent was obtained from all patients before initiating radiotherapy and the Institutional Review Board of the Kyoto University Hospital approved this study (R1446).

2.2. Treatment planning

On the day of CT, the patient’s head was immobilized with a thermoplastic mask in the supine position. Contrast‐enhanced computed tomography (CECT) scans were acquired with the following imaging parameters: a slice thickness of 1 or 1.25 mm (depending on the CT scanner; SOMATOM Definition AS; Siemens Medical Systems, Erlangen, Germany, and LightSpeed RT16; General Electric Medical Systems, Waukesha, WI, USA, respectively), a pixel matrix of 512 × 512 pixels, and a field of view of 400 mm. CECT was combined with gadolinium‐enhanced T1‐weighted MRI with the images obtained within 1 month prior to the CT simulation. The gross tumor volume (GTV) was defined as the visible lesion on both the CT and MRI; no clinical target volume was defined. Planning target volumes (PTVs) were defined by adding a margin of 1–2 mm to the GTV. The median number of PTVs present in any plan was 4 (range: 2–18). The organs at risk (OARs) included the normal brain (whole brain minus the PTVs), brainstem, eyes, lens, optic nerves, chiasm, and skin (defined as a structure cropped 5 mm from the body).

The isocenter was located at the center of all the PTVs. The VMAT plans were created using 3–5 arcs, including 1 full coplanar arc and 2–4 non‐coplanar partial arcs with a couch angle of ± 60°. The collimator angle was manually selected depending on the size and location of the target. Photon beam energies of six flattening filter‐free (FFF) and 10 FFF MV photon beams were used. All the treatment plans were generated the Eclipse planning system (version 13; Varian Medical Systems, Palo Alto, CA, USA), delivered from a TrueBeam STx instrument with a high‐definition 120‐leaf multileaf collimator (MLC) (Varian Medical Systems). Dose calculations were performed using Acuros XB (version 13.7; Varian Medical Systems) with heterogeneity correction and a 1‐mm grid resolution.

Plan normalization was performed in a manner such that at least 99.5% of the prescribed dose (D99.5%) of 28 Gy (five fractions) was generally delivered to each PTV. Plan optimization was performed so that the near‐maximum dose (D2%) to all PTVs was around 40 Gy (Per protocol, 135–150%) and the irradiated dose to the OARs was as low as possible.

2.3. Model creation, evaluation, and selection

The fourfold CV method was used for model creation, evaluation, and selection (Fig. 1). First, the dataset was split into two; Groups 1–4 were used as the training and validation dataset and Group 5 as the testing dataset (Table 1). Next, the model configuration and hyperparameter setting were conducted using RapidPlanTM (version 13.7; Varian Medical Systems). 16 RapidPlanTM is a machine learning system based on the geometric relation of the structures and DVH. As part of the model‐creating process, the dosimetric and geometric data were extracted from the database to establish DVH estimation models using regression techniques. In RapidPlanTM, there are mainly two types of optimization objectives; a fixed objective and a line objective. The former manually adds a fixed upper or lower dose or volume, with a fixed priority or with a generated priority created by the DVH estimation algorithm. Fixed objectives can be used for both OARs and target structures. On the other hand, the latter generates an estimated DVH range only for the OARs, which is also created by the DVH estimation algorithm automatically. Using a line objective, the optimization is performed so that the OAR will receive a dose as low as the estimated DVH range. Therefore, the volumes, doses, and priorities are automatically generated in a line objective.

Fig. 1.

Fig. 1

Schematic of the fourfold cross‐validation (CV) method; (a) dataset splitting, (b) hyperparameter setting of dose‐volume histogram (DVH) estimation models created from three subgroups (Models A ~ D) and model application to the remaining subgroup, and (c) model created based on all training and validation datasets (Model E), and application thereof to the testing dataset. Broken arrows denote model creation and solid arrows denote model application in (b) and (c).

Table 1.

Summary of each Group. The patients were divided randomly into five groups (Groups 1–5). The data of Groups 1–4 were used as the training and validation dataset and those of Group 5 were used as the testing dataset.

Group Role Total number of PTVs

Number of PTVs per patient

[Median (min–max)]

Size (cm3)

[Median (min–max)]

1 Training/Validation 70 6 (2–12) 0.6 (0.1–12.3)
2 Training/Validation 57 3 (2–13) 0.6 (0.1–16.9)
3 Training/Validation 60 5 (2–14) 0.4 (0.1–12.9)
4 Training/Validation 62 4 (2–12) 0.5 (0.1–10.2)
5 Testing 68 4 (2–18) 0.5 (0.2–13.5)

Abbreviation: PTV, planning target volume.

Subsequently, four DVH estimation models were created from the training and validation dataset with the determined hyperparameter setting shown in Table 2: Model A (MA; Groups 2–4), Model B (MB; Groups 1, 3, and 4), Model C (MC; Groups 1, 2, and 4), and Model D (MD; Groups 1–3). MA, MB, MC, and MD were applied to the remaining of the training and validation dataset. In the validation phase, the generalization ability was assessed, which means that the model from training data can reproduce an acceptable outcome when applied to unseen data. The dose‐volume indices (DVIs) were compared among four KB‐generated plans (KBP‐A1: MA applied to Group 1; KBP‐B2: MB applied to Group 2; KBP‐C3: MC applied to Group 3; and KBP‐D4: MD applied to Group 4). A lack of any statistical difference among the DVIs of the four KB plans (KBPs) and a D2% for all PTVs ranging from 130% to 155% of the prescribed dose (clinically acceptable variation), was taken to indicate that the models were not overfitted and had a good generalization performance.

Table 2.

Optimization objectives used in the model.

Structure Type Volume (%) Dose (cGy) Priority
Brain‐PTVs Line Generated Generated Generated
Brainstem Upper Generated 1000 Generated
Line Generated Generated Generated
Eyes Upper Generated 150 Generated
Line Generated Generated Generated
Lens Upper Generated 100 Generated
Optic nerves Upper Generated 200 Generated
Line Generated Generated Generated
Skin Line Generated Generated Generated
PTV Upper 0 3950 300
Upper 10 3850 Generated
Upper 30 3650 Generated
Upper 50 3450 Generated
Upper 70 3300 Generated
Upper 90 3050 200
Lower 0.5 3900 200
Lower 10 3800 Generated
Lower 30 3600 Generated
Lower 50 3400 Generated
Lower 70 3250 Generated
Lower 90 3000 Generated
Lower 99.5 2900 Generated
Lower 100 2850 Generated

The line objective is a type of optimization objective and the values of volume, dose, and priority are automatically generated by RapidPlanTM. Priorities are also generated in this study with respect to some upper or lower objectives.

Abbreviations: Line, line objective; Lower, lower objective; PTV, planning target volume; Upper, upper objective.

Finally, in the testing phase, the accuracy of the created models was focused upon. Model E (ME; Groups 1–4) was constructed using all the training and validation datasets and the same hyperparameter settings as shown in Table 2. ME was applied to Group 5 (independent testing dataset), and a KBP (KBP‐E5) was then generated. A lack of statistical difference between the DVIs extracted from the KBP‐E5 and the CL of each corresponding PTV and each OAR in the same patient was regarded to indicate that the KB models yielded clinically acceptable plans.

2.4. KB‐generated plan optimization and dose calculation

One optimization cycle was performed for each model without the modification of the KB‐generated optimization objectives. Thereafter, the dose distribution was calculated with the same beam arrangement used in the CL and plan normalization was performed.

2.5. Evaluation indices

In the validation phase, the DVIs for the PTV (D2%, D50%, and D99.5%), normal brain (V5 Gy, V10 Gy, V14 Gy, V20 Gy, and V28 Gy), and the maximum dose (Dmax) to all OARs were compared among plans.

In the testing phase, in addition to the DVIs above, the Ian Paddick Conformity Index (IPCI) was also employed to evaluate the dose distributions. 17 The definition of the IPCI is [TVPIV]2/ [TV × PIV], where the TVPIV is the target volume covered by the prescription dose, TV is the target volume, and PIV is the prescription isodose volume. For multiple targets, the modified IPCI (mIPCI) was defined as ([TVPIV_sum]2/ [TVsum × PIVsum]), where TVPIV_sum is the summed target volumes enclosed by an isodose line of the prescription dose, TVsum is the summed volume of all PTVs, and PIVsum is the summed prescription isodose volume. The mIPCI had values in the range 0–1, with the scores approaching 1 when PTVs were conformally covered with isodoses of the prescription dose. The modulation complexity scores for the VMAT plans (MCSv) and the total number of monitor units (MUs) were used as additional metrics. The MCSv was calculated based on the leaf sequence variability (LSV) parameter and aperture area variability. 18 The LSV was defined for each control point considering in each bank the differences in position between adjacent MLC leaves. The MCSv demonstrated values in the range 0–1; additionally, the scores decreased when the modulation increased.

2.6. Statistical analysis

In the validation phase, the Friedman test was performed on the DVIs of all the PTVs and OARs to determine if there is any difference between the KBP‐A1, B2, C3, and D4, followed by the unpaired Wilcoxon signed‐rank test with Bonferroni correction for multiple comparisons as a post hoc test. Meanwhile, in the testing phase, a paired Wilcoxon signed‐rank test was used to evaluate the significant differences between the CL and KBP‐E5 for each corresponding PTVs and OAR in the same patient. The statistical significance was set at P < 0.05. All statistical analyses were performed using R software (version 3.5.1; R Development Core Team, Vienna, Austria).

3. Results

3.1. Validation phase

In the fourfold CV phase, no significant difference was observed in any DVI among the four KBPs. The details of the DVIs for all the PTVs, normal brain, and other OARs between the four KB models are shown in Fig. 2. The PTV D2% values ranged from 130.8% to 153.9% of the prescribed dose, within clinically acceptable variations; thus, it was confirmed that the KB model was not overfitted and had a good generalization performance.

Fig. 2.

Fig. 2

The box‐and‐whisker plots of the dose‐volume indices (DVIs) of (a) the planning target volume (PTV), (b) normal brain, and (c) other organs at risk (OARs): comparison of KBP‐A1, B2, C3, and D4 in the fourfold CV phase. The boxes represent the interquartile range (IQR). Low outliers are below the first quartile minus 1.5 × IQR and high outliers are above the third quartile plus 1.5 × IQR. DXX% = the dose to XX% of the target volume, VXXGy = the volume receiving XX Gy, Dmax = the maximum dose to the volume.

3.2. Testing phase

In the testing phase, the D2% of each PTV in KBP‐E5 (median, 137.2% [interquartile range (IQR), 2.4%]) was significantly higher than that of the corresponding PTVs in the CL (median, 136.3% [IQR, 9.6%]) (P = 0.005). The D50% of each PTV in KBP‐E5 was significantly lower than that of the corresponding PTVs in the CL (median, 118.8% vs. 120.7% [IQR, 3.2% vs.4.6%]) (P = 0.02). The differences of each PTV D99.5%, normal brain (V5 Gy, V10 Gy, V14 Gy, V20 Gy, and V28 Gy), and Dmax to all OARs, were not significant in terms of the CL and KBP‐E5 (> 0.05). Figure 3 shows the scatter plots of the DVIs of the PTV, normal brain, and other OARs for CL and KBP‐E5.

Fig. 3.

Fig. 3

The scatter plots and box‐and‐whisker plots of the DVIs of (a) the PTV, (b) normal brain, and (c) other OARs for clinical plan (CL) and KBP‐E5 in the testing phase can be observed. The variation of the PTV in the KBP‐E5 was smaller than that of PTV in the CL. The plots of all the OARs were along or below the diagonal line (y = x). Abbreviations are the same as in Fig. 2.

The representative dose distributions of the CL and KBP‐E5 are shown in Fig. 4. KBP‐E5 minimized the dose spillage between the target volumes. The representative DVHs of the CL and KBP‐E5 are shown in Fig. 5; the irradiated dose to the OARs was decreased and the DVHs for each PTV was more uniform for KBP‐E5 than for the CL.

Fig. 4.

Fig. 4

The representative dose distributions of the CL and the knowledge‐based plan (KBP) for a patient with 16 PTVs. The modified Ian Paddick Conformity Index (mIPCI) is shown in parentheses. The high‐dose spillage between the target volumes was minimized, as shown by yellow arrows.

Fig. 5.

Fig. 5

The representative DVHs of (a) the CL and (b) KBP‐E5 for a patient with 16 PTVs; the brain PTVs are shown in brown. The dose to the OARs was decreased and the DVHs for each PTV were more uniform for KBP‐E5 than for the CL.

Figure 6 shows the scatter plots of the mIPCI, MCSv, and the total number of MUs for the CL and KBP‐E5. The scatter plots of mIPCI and MCSv were above the diagonal line (y = x), indicating that KBP‐E5 improved dose conformality and decreased modulation complexity compared to the CL [Figs. 6(a) and 6(b)]; however, the differences were not significant (> 0.05). The median values of mIPCI for KBP‐E5 and the CL were 0.808 (range: 0.561–0.916) and 0.774 (range: 0.550–0.876), respectively. The median values of MCSv for KBP‐E5 and the CL were 0.078 (range: 0.040–0.204) and 0.071 (range: 0.042–0.176), respectively. Additionally, KBP‐E5 achieved a lower total number of MUs (median, 2,428.5 MUs [range: 1,836.4–4,054.7 MU]) than the CL (median, 2,532.3 MUs [range: 1,672.5–4,292.9 MU]) [Fig. 6(c)], although the difference was not significant (> 0.05).

Fig. 6.

Fig. 6

The scatter plots and box‐and‐whisker plots of (a) the mIPCI, (b) modulation complexity scores for VMAT plans (MCSv) and (c) monitor units (MUs) in the testing group. The scatter plots of mIPCI and MCSv were above the diagonal line (y = x). mIPCI = modified Ian Paddick Conformity Index. Note: The mIPCI was defined as ([TVPIV_sum]2/ [TVsum × PIVsum]); TVPIV_sum = the sum of target volumes enclosed by an isodose line of the prescription dose, TVsum = the sum of volume of all PTVs, PIVsum = the sum of prescription isodose volume. The mIPCI approaching 1 means that PTVs were conformally covered with the prescribed dose.

4. Discussion

To the best of our knowledge, this is the first study to assess the performance of KBPs in single‐isocenter VMAT for multiple brain metastases using the k‐fold CV method. The number of test samples was almost equivalent to previous reports on KB planning. 11 , 12 , 13 , 14 , 15 , 19 As shown in Table 3, most previous studies employed the hold‐out method; however, as described above, one disadvantage of this method is that the performance evaluation is subject to a higher variance given the smaller dataset, whereas multiple validations can be performed using the k‐fold CV method to conclusively determine the model overfitting. Using the k‐fold CV method, all of our KB models yielded clinically acceptable plans with a single optimization.

Table 3.

Summary of previous reports on knowledge‐based planning. Most of the previous studies employed the hold‐out method.

Disease

Training

samples

Testing

samples

Validation
Chang et al. 12 Nasopharynx 79 20 Hold‐out
Hussein et al. 11 Prostate 40 10 Hold‐out
Uterus 37 10 Hold‐out
Chatterjee et al. 13 Brain 82 21 (GBM) Hold‐out
24 (Others) Hold‐out
Faught et al. 14 Lung 30 20 Hold‐out
Yu et al. 15 Liver 30 and 60* 13 Hold‐out
Babier et al. 19 Oropharynx 216 1 LOOCV
This study

Multiple brain

metastases

48 12 4‐fold CV

Abbreviations: CV, cross‐validation; LOOCV, leave‐one‐out cross‐validation, GBM = Glioblastoma.

*

Note: two models were constructed.

KB planning, a machine‐learning tool for determining the best practice based on past successful treatment plans, creates KB models for improving the treatment plans for future patients. It is important to compare the performance of KB models in terms of generalization. In this study, k‐fold CV was applied for model evaluation and comparison. MA to MD were created with fixed parameters and applied to different patients’ groups for performance evaluation. The KB models were statistically equivalent after the adjustment of learned and fixed optimization parameters, although the interquartile ranges of the DVIs varied among them (Fig. 2). In the testing phase, the DVIs for ME were compared to the CL. We found that the ME generated statistically equivalent plans to the CL with a single optimization, except for PTV D2%. For hyperparameter tuning, which is equivalent to determining both mathematically learned objectives and fixed optimization objectives using RapidPlanTM, there is no clear rule that ensures the best performance. When comparing algorithms, statistical tests such as the McNemar test and Cochran’s Q test are commonly employed; however, it is impossible to obtain the “best” treatment planning algorithm because the definition of “best” varies according to clinical factors, such as the patient’s condition and treatment preferences. Therefore, we used the DVIs of the CL as the “best” outcome measures; the DVIs were compared between the KBPs and the CL for model evaluation.

According to a review paper on KB planning, several researchers achieved comparable, and often improved, VMAT plans using KBPs, while also reducing the planning time and variation in the plan quality. 9 We demonstrated that the PTV D2% for the KBP was significantly higher than that for the CL, while achieving the same dose conformality, and also that the radiation dose to the normal brain for the KBP was similarly low to that of the CL in patients with 2–18 PTVs, with a single optimization. In brain SRS, a higher Dmax of the PTVs is associated with an improved local control of the dose, and a lower irradiated dose to the OARs with higher mIPCI values could decrease the radiation necrosis. According to previous research, a normal brain V14 Gy is a good indicator of radiation necrosis in patients with large metastases after five‐fraction CyberKnife radiotherapy (Accuray, Sunnyvale, CA, USA) 20 ; however, such dose‐volume constraints would not always be applicable, depending on the fractionation, target sizes, and the number of target volumes. During the manual inverse planning of single‐isocenter VMAT SRS for multiple brain metastases, it is difficult to set definitive clinical goals because of such variations. With KB planning, the realistically achievable dose distribution can be predicted and patients can receive high‐quality treatment even with limited time and human resources.

Numerous parameters reveal the plan complexity, such as MCSv, a modulation index, and the plan‐averaged modulation. 18 , 21 , 22 The MCSv was employed in this study as it allows for an effortless comparison to other studies. In the study by Masi et al., 19 the MCSv values were in the range 0.25–0.50 for the conventional VMAT plans. Compared to this, our MCSv values for multiple brain metastases for the CL were extremely low (range: 0.042–0.176), indicating more complex MLC patterns compared to conventional VMAT plans. We found that the MCSv values were slightly higher for the KBP than in the CL [Fig. 6(b)] and fewer MUs were required [Fig. 6(c)]; thus, the KB models provided plans that are less intensity‐modulated than CLs.

Recently, the HyperArc (Varian Medical Systems) has been used clinically for multiple brain metastases. HyperArc plans consist of a maximum of three non‐coplanar arcs, based on four of five possible fixed angular couch positions to be selected between (0°, ±45°, +90°) and (0°, ±45°, −90°), with each arc having a fixed length of 180° and an automatically selected collimator angle. Several studies have demonstrated the superiority of the HyperArc to single‐isocenter VMAT in terms of the conformity and dose falloff. 23 , 24 In this study, the same beam parameters (beam gantry, couch, and collimator angles) used in an approved CL were employed during the testing phase; however, in a clinical situation, the beam arrangement is determined through a process of trial and error, which is time consuming and affects the plan quality. As shown in the present and previous studies, 23 , 24 both KB planning and HyperArc are effective in reducing doses to OARs while maintaining the target coverage for multiple brain metastases. Both methods have advantages and disadvantages with respect to the optimization of parameters including the number of arcs, collimator angle, couch angle, and arc lengths; therefore, an approach combining the advantages of KB planning and HyperArc would provide more conformal dose distributions in a shorter planning time.

There were some notable limitations to this study. First, this study was conducted retrospectively, and the KB models employed were derived from the data of a single institution. To assess the applicability of our model, the clinical data of other institutions, or of different cohorts and beam arrangements, should be used prospectively for external validation. Second, eligible patients can be treated safely without any special effort to decrease the irradiated dose to OARs other than the brain. Patients with a large target volume, prior history of cranial irradiation, or a lesion located in or close to brainstem are treated with other fractionations in our institute; such cases were not included in this study.

5. Conclusions

We employed fourfold CV for the evaluation and selection of KB models. Statistically equivalent models (between the CL and final KB model) were generated after the adjustment of learned and fixed optimization parameters. After confirming the generalization performance of the models, the final KB model was applied to the test group. We demonstrated that the KB model in the single‐isocenter VMAT for multiple brain metastases was equivalent in dose distribution, MCSv, and the total number of MUs to the CL with a single optimization.

Consent for publication

The consent for publication was obtained via our institution’s form.

Authors’ contributions

NK performed the planning study and statistical analysis, and drafted the manuscript. NK, MN, HH, NM, KT, and MU conceived the study, participated in its design and coordination, and helped to draft the manuscript. All authors read and approved the final manuscript.

Funding information

This study was supported in part by JSPS KAKENHI (Grant Nos. 18H02766, 18K07700 and 19K21271).

References

  • 1. Cagney DN, Martin AM, Catalano PJ, et al. Incidence and prognosis of patients with brain metastases at diagnosis of systemic malignancy: a population‐based study. Neuro Oncol. 2017;19:1511–1521. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Kepka L, Tyc‐Szczepaniak D, Osowiecka K, et al. Quality of life after whole brain radiotherapy compared with radiosurgery of the tumor bed: results from a randomized trial. Clin Transl Oncol. 2018;20:150–159. [DOI] [PubMed] [Google Scholar]
  • 3. Shaw E, Kline R, Gillin M, et al. Radiation Therapy Oncology Group: radiosurgery quality assurance guidelines. Int J Radiat Oncol Biol Phys. 1993;27:1231–1239. [DOI] [PubMed] [Google Scholar]
  • 4. Aoyama H, Shirato H, Tago M, et al. Stereotactic radiosurgery plus whole‐brain radiation therapy vs stereotactic radiosurgery alone for treatment of brain metastases: a randomized controlled trial. JAMA. 2006;295:2483–2491. [DOI] [PubMed] [Google Scholar]
  • 5. Chang EL, Wefel JS, Hess KR, et al. Neurocognition in patients with brain metastases treated with radiosurgery or radiosurgery plus whole‐brain irradiation: a randomised controlled trial. Lancet Oncol. 2009;10:1037–1044. [DOI] [PubMed] [Google Scholar]
  • 6. Brown PD, Jaeckle K, Ballman KV, et al. Effect of radiosurgery alone vs radiosurgery with whole brain radiation therapy on cognitive function in patients with 1 to 3 brain metastases: a randomized clinical trial. JAMA. 2016;316:401–409. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 7. Kraft J, Zindler J, Minniti G, et al. Stereotactic radiosurgery for multiple brain metastases. Curr Treat Options Neurol. 2019;21:6. [DOI] [PubMed] [Google Scholar]
  • 8. Thomas EM, Popple RA, Wu X, et al. Comparison of plan quality and delivery time between volumetric arc therapy (RapidArc) and Gamma Knife radiosurgery for multiple cranial metastases. Neurosurgery. 2014;75:409–417. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 9. Ge Y, Wu QJ. Knowledge‐based planning for intensity‐modulated radiation therapy: a review of data‐driven approaches. Med Phys. 2019;46:2760–2775. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10. Raschka S.Model evaluation, model selection, and algorithm selection in machine learning. https://arxiv.org/abs/1811.12808 Last accessed September 23, 2019.
  • 11. Hussein M, South CP, Barry MA, et al. Clinical validation and benchmarking of knowledge‐based IMRT and VMAT treatment planning in pelvic anatomy. Radiother Oncol. 2016;120:473–479. [DOI] [PubMed] [Google Scholar]
  • 12. Chang ATY, Hung AWM, Cheung FWK, et al. Comparison of planning quality and efficiency between conventional and knowledge‐based algorithms in nasopharyngeal cancer patients using intensity modulated radiation therapy. Int J Radiat Oncol Biol Phys. 2016;95:981–990. [DOI] [PubMed] [Google Scholar]
  • 13. Chatterjee A, Serban M, Abdulkarim B, et al. Performance of knowledge‐based radiation therapy planning for the glioblastoma disease site. Int J Radiat Oncol Biol Phys. 2017;99:1021–1028. [DOI] [PubMed] [Google Scholar]
  • 14. Faught AM, Olsen L, Schubert L, et al. Functional‐guided radiotherapy using knowledge‐based planning. Radiother Oncol. 2018;129:494–498. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 15. Yu G, Li Y, Feng Z, et al. Knowledge‐based IMRT planning for individual liver cancer patients using a novel specific model. Radiat Oncol. 2018;13:52. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 16. Fogliata A, Belosi F, Clivio A, et al. On the pre‐clinical validation of a commercial model‐based optimization engine: application to volumetric modulated arc therapy for patients with lung or prostate cancer. Radiother Oncol. 2014;113:385–391. [DOI] [PubMed] [Google Scholar]
  • 17. Paddick I. A simple scoring ratio to index the conformity of radiosurgical treatment plans Technical note. J Neurosurg. 2000;93:219–222. [DOI] [PubMed] [Google Scholar]
  • 18. Masi L, Doro R, Favuzza V, et al. Impact of plan parameters on the dosimetric accuracy of volumetric modulated arc therapy. Med Phys. 2013;40:071718. [DOI] [PubMed] [Google Scholar]
  • 19. Babier A, Boutilier JJ, McNiven AL, et al. Knowledge‐based automated planning for oropharyngeal cancer. Med Phys. 2018;45:2875–2883. [DOI] [PubMed] [Google Scholar]
  • 20. Inoue HK, Sato H, Seto K, et al. Five‐fraction CyberKnife radiotherapy for large brain metastases in critical areas: impact on the surrounding brain volumes circumscribed with a single dose equivalent of 14 Gy (V14) to avoid radiation necrosis. J Radiat Res. 2014;55:334–342. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 21. Li R, Xing L. An adaptive planning strategy for station parameter optimized radiation therapy (SPORT): Segmentally boosted VMAT. Med Phys. 2013;40:050701. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 22. Du W, Cho SH, Zhang X, et al. Quantification of beam complexity in intensity‐modulated radiation therapy treatment plans. Med Phys. 2014;41:021716. [DOI] [PubMed] [Google Scholar]
  • 23. Ohira S, Ueda Y, Akino Y, et al. HyperArc VMAT planning for single and multiple brain metastases stereotactic radiosurgery: a new treatment planning approach. Radiat Oncol. 2018;13:13. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Ruggieri R, Naccarato S, Mazzola R, et al. Linac‐based VMAT radiosurgery for multiple brain lesions: comparison between a conventional multi‐isocenter approach and a new dedicated mono‐isocenter technique. Radiat Oncol. 2018;13:38. [DOI] [PMC free article] [PubMed] [Google Scholar]

Articles from Journal of Applied Clinical Medical Physics are provided here courtesy of Wiley

RESOURCES