A Pumpless Microfluidic Neonatal Lung Assist Device for Support of Preterm Neonates in Respiratory Distress

Mohammadhossein Dabaghi 1, Niels Rochow 2,3, Neda Saraei 4, Gerhard Fusch 2, Shelley Monkman 2, Kevin Da 5, Alireza Shahin‐Shamsabadi 1, John L Brash 1,5, Dragos Predescu 2, Kathleen Delaney 6, Christoph Fusch 1,2,3, P Ravi Selvaganapathy 1,4,
PMCID: PMC7610273  PMID: 33173732

Abstract

Premature neonates suffer from respiratory morbidity as their lungs are immature, and current supportive treatment such as mechanical ventilation or extracorporeal membrane oxygenation causes iatrogenic injuries. A non‐invasive and biomimetic concept known as the “artificial placenta” (AP) would be beneficial to overcome complications associated with the current respiratory support of preterm infants. Here, a pumpless oxygenator connected to the systemic circulation supports the lung function to relieve respiratory distress. In this paper, the first successful operation of a microfluidic, artificial placenta type neonatal lung assist device (LAD) on a newborn piglet model, which is the closest representation of preterm human infants, is demonstrated. This LAD has high oxygenation capability in both pure oxygen and room air as the sweep gas. The respiratory distress that the newborn piglet is put under during experimentation, repeatedly and over a significant duration of time, is able to be relieved. These findings indicate that this LAD has a potential application as a biomimetic artificial placenta to support the respiratory needs of preterm neonates.

Keywords: artificial placenta, microfluidics, oxygenation, piglet models, respiratory distress


Premature neonates with respiratory failure may require respiratory support in the form of extracorporeal blood oxygenation to survive. A modular lung assist device is developed using microfabrication technologies to provide additional gas exchange support over normal breathing in a pumpless fashion. Such technology is designed to extend respiratory support outside the womb in a biomimetic manner similar to the placenta.

graphic file with name ADVS-7-2001860-g010.jpg

1. Introduction

Premature births are associated with significant respiratory morbidity and mortality. Neonates, especially those with very low birthweights and born at less than 28 weeks of gestational age[ 1 , 2 ] have a very low survival rate. In particular, those born earlier than 24 weeks of gestational age[ 3 ] have more than 50% mortality. One of the leading causes of morbidity associated with preterm birth is a respiratory failure caused by a lack of surfactants in immature lungs as well as lung hypoplasia.[ 4 , 5 , 6 , 7 ] Mechanical ventilation is often used for respiratory support of such premature neonates[ 8 ] who are in respiratory failure. Nonetheless, mechanical ventilation is invasive and associated with severe complications such as pulmonary injury, chronic lung disease, and related diseases of prematurity such as retinopathy of prematurity, intraventricular hemorrhage, or necrotizing enterocolitis, which would lead to several long‐term side effects.[ 9 , 10 , 11 ] In late‐preterm and term infants, extracorporeal membrane oxygenation (ECMO) could be an alternative choice of treatment. However, ECMO is invasive and requires central vascular access and, therefore, surgery. Further, the current hollow fiber‐based blood oxygenators that are used in ECMO are not suitable for neonatal application due to their high priming volume (>20 mL), high pressure drop, which requires an external pump for perfusion, and their gas exchange dependency on oxygen.[ 8 , 12 ] Moreover, other complications such as cerebral injury with intracranial hemorrhage or stroke, poor somatic growth, and poor neurodevelopment can be caused by this technology.[ 13 ]

Microfabrication technologies have been employed to fabricate blood oxygenators that can address some of the limitations of current ECMO devices by using a biomimetic architecture similar to the vascular network in the lung in order to enhance gas transfer efficiency, increase effective surface area for transfer as well as reduce shear stress in blood flow, and avoid the formation of blood stagnation zones.[ 14 , 15 ] Several microfluidic blood oxygenators[ 12 , 16 , 17 , 18 , 19 , 20 , 21 , 22 , 23 , 24 , 25 , 26 , 27 , 28 , 29 , 30 , 31 , 32 ] have been introduced aimed at improving gas exchange efficiency, but all of them use external pumps to perfuse the blood through them. An alternate approach is to use the arterio‐venous pressure difference in the body as a natural pumping mechanism to circulate the blood through an oxygenator—a concept known as the artificial placenta. The artificial placenta approach is characterized by i) blood circulating through a passive device pumped solely by the arteriovenous pressure difference of the neonate's heart and ii) being capable of exchanging blood gases with the room air. This concept is physiologically similar to in utero conditions and overcomes some of the issues associated with traditional ECMO devices.[ 8 , 12 , 16 ] Traditionally used hollow fiber ECMO devices have high priming volume and are not suitable for the neonatal application. Only some of the microfluidic oxygenators[ 12 , 16 , 29 , 30 , 31 ] have been explicitly designed for use as an artificial placenta device. Most artificial placenta‐type microfluidic blood oxygenators have only been examined in vitro,[ 12 , 29 , 30 ] and even those[ 16 , 31 ] tested in vivo have not been shown to provide adequate oxygenation in room air to support a suitable animal model in respiratory distress.

Here, we present the first high‐performance, pumpless, neonatal lung assist device that, in in vivo experiments on newborn piglets (of similar size to neonates) has demonstrated its ability to overcome respiratory distress. The lung assist device was assembled from high‐performance microfluidic blood oxygenators (MBOs)[ 30 ] in a parallel configuration to enable sufficient flow of blood through it as it was pumped by the pressure generated solely by the piglet's heart. Additionally, the entire LAD was coated with heparin to reduce its systemic infusion, and the efficacy of this coating was evaluated. We report the in vitro gas exchange performance of the LAD as well as its in vivo performance when the piglet was intentionally put in respiratory distress by placing it under hypoxic conditions. Systemic blood gas measurements, as well as cardiovascular conditions, were monitored during repeated cycles of hypoxic and normoxic conditions over a period of 200 min. Our results showed that this passive pumpless LAD could provide adequate oxygenation in room air to restore near normoxic conditions and has the potential to meet the clinical requirements to support neonates under respiratory distress.

2. Results

The performance of the LAD was demonstrated first with in vitro experiments to ensure that the sizing and the performance would be suitable to provide sufficient oxygenation capacity while operating with the arterio‐venous pressure differential provided by a piglet. Next, the surface modification using heparin is characterized to determine the appropriate protocol for subsequent experiments on the animal model. Finally, the assembled LAD with the surface modification was tested ex vivo on a piglet model by repeatedly placing it under respiratory distress and rescuing it by connecting the LAD to demonstrate that this device can rescue piglets the size of neonatal babies from respiratory distress in a passive manner.

2.1. In Vitro Testing of LAD Using Bovine Blood

The gas exchange performance of the LAD in vitro was investigated in an enriched oxygen environment as well as in room air using heparinized (3 unit mL−1) bovine blood. The oxygen saturation levels (Sa O2) of the blood were adjusted to 58 ± 2% for experiments performed in an enriched oxygen environment and 63 ± 1% for experiments conducted in room air. The desaturated blood was pumped through the device, and the oxygen saturation was measured in the outlet.

The LAD itself was assembled from several MBOs (16 in parallel) with a blood channel height of 180 µm which were made of polydimethylsiloxane (PDMS) (Experimental Section). The LAD was able to fully oxygenate blood (Sa O2 = ≈100%) up to a flow rate of 80 mL min−1 in an enriched oxygen environment and to increase Sa O2 to ≈95% at a blood flow rate of 100 mL min−1 (Figure  1a). The oxygen uptake was calculated by summing the amount of dissolved oxygen and adsorbed oxygen to red blood cells.[ 12 ] The oxygen uptake was between 0.92–5.71 mL min−1 with a linear increase with raising blood flow rates while the LAD was using oxygen as the sweep gas (Figure 1b). Figure 1c displays the normalized oxygen transfer of the LAD when it was exposed to a pure oxygen environment with a maximum value of 95.3 ± 3.0 mL min−1 m−2 (the LAD had an effective surface area of ≈600 cm2). The LAD had a different gas transfer performance in room air as oxygen concentration was lower. Consequently, the oxygen saturation level achieved was lower, and the blood was fully oxygenated only up to a blood flow rate of 20 mL min−1. The outlet saturation level dropped to 87% at a blood flow rate of 60 mL min−1 (Figure 1d). The oxygen uptake linearly increased with blood flow rates and was 1.82 mL min−1 at a blood flow rate of 60 mL min−1 in room air (Figure 1e). Also, the normalized oxygen transfer was calculated for the LAD when it was placed in room air, as shown in Figure 1f. The pressure drop of the LAD (Figure 1g), as well as the LAD with the extracorporeal circuit (Figure 1h), was measured over a range of flow rates from 10–100 mL min−1. Carbon dioxide release data is presented in Figure S1, Supporting Information.

Figure 1.

Figure 1

In vitro test of the LAD with heparinized bovine blood at various blood flow rates: a) oxygen saturation level before and after the LAD in oxygen, b) oxygen uptake in oxygen, c) normalized oxygen transfer of the LAD when it was tested in oxygen environment, d) oxygen saturation level before and after the LAD in room air, e) oxygen uptake in room air, f) normalized oxygen transfer of the LAD when it was tested in room air, g) pressure drops for the LAD (the shaded region indicates the operational pressure drops range for a pumpless operation), and h) the pressure drops for the LAD with the extracorporeal circuit. Data are means ± SD, n = 3. Hematocrit was 38% and 29% for oxygen and air conditions, respectively.

A numerical analysis of the pressure drop was conducted[ 30 , 33 ] to estimate the optimal height range for the blood vascular network to achieve a pumpless operation (Figure S7, Supporting Information). The flow analysis showed that channel height can be reduced to ≈130 µm below which the pressure drop would be significantly high that a pumpless operation may not be possible. As a result, we tested another LAD that had MBOs with the blood channel's height of 130 µm and compared it with the current (180 µm) design. This LAD was tested in vitro and in vivo using only pure oxygen as the sweep gas. Interestingly, both LADs (one with microchannel's height of 130 µm and another one had a microchannel's height of 180 µm) had similar oxygenation performance (Figure S7c, Supporting Information) while they had noticeably different hydraulic resistance. The LAD with SOUs of 130 µm was still able to rescue the animal to nearly 90% systemic oxygenation levels under hypoxic conditions. This was expected as our previous studies[ 29 , 30 , 33 , 34 ] have shown that the oxygen transfer characteristics of an MBO with double‐sided blood channels was not strongly dependent on the height of the device, especially for a height of greater than 100 µm. The LAD with 130 µm channel height however had a lower blood flow rate (60–70 mL min−1) in the corresponding in vivo experiment (Figure S7d, Supporting Information). This comparison showed that even though both the oxygenators performed similarly from an oxygenation perspective, the lower flow resistance of the LAD with MBOs having a channel height of 180 µm was capable of more robust operation as it can withstand higher blood pressure variations while still providing sufficient oxygenation and throughput. It was selected for use in the subsequent long‐term experiment as the optimized configuration (the details of the animal experiment for the LAD with the height of 130 µm are presented in Figures S8–S10, Supporting Information).

2.2. Modification of Blood‐In‐Contact Surfaces with Heparin

PDMS is hydrophobic and is known to adsorb proteins readily from blood,[ 35 ] leading to the activation of blood coagulation, platelet adhesion, and thrombosis. To minimize these effects, all blood‐contacting surfaces were coated with heparin using the set‐up depicted in Figure  2a1. The coating process was optimized using fluorescently labeled heparin, which allowed assessment of the quality and stability of the coating. For instance, the wash time to remove residual heparin was optimized to 15 min by monitoring the fluorescence intensity of the wash eluate (Figure 2a2).

Figure 2.

Figure 2

In vitro experiments to assess the effects of heparin coating on LADs or MBOs. a.1) Experimental setup for coating, a.2) fluorescence intensity of solution samples from washing a FITC heparin‐coated MBO with PBS, b) water contact angles of coated and non‐coated surfaces, c) wide‐angle image of an MBO coated with FITC‐heparin, d) wide‐angle image of an MBO coated with FITC‐heparin after contact with flowing PBS for 20 h, e) fluorescence intensity of solution samples collected during contact with a heparin‐coated MBO for 3 h, f) wide‐angle image of FITC−fibrinogen adsorbed to an uncoated MBO, g) wide‐angle image of FITC−fibrinogen adsorbed to a heparin‐coated MBO, h) fluorescence intensities of uncoated and heparin‐coated surfaces obtained using ImageJ software, i) experimental setup for pressure measurement during plasma circulation, j) representative pressure trace during plasma circulation in coated and uncoated MBOs (n = 3), and k) collected plasma after circulation in the MBOs with coated and uncoated surfaces.

Coating PDMS with heparin decreased the water contact angle from 120° to 66°, rendering the surface moderately hydrophilic and possibly reducing the potential for clot formation on the surface (Figure 2b). More importantly, heparin as an anticoagulant on the surface is expected to inhibit clotting via inhibition of thrombin.

Wide‐angle fluorescence images of the MBO surfaces after heparin coating showed that the coating on the blood vascular network surfaces was uniform (Figure 2c). Subsequently, PBS was perfused in the MBO at a flow rate of 3 mL min−1 for 20 h to investigate the stability of the heparin coating. As seen in Figure 2d, the fluorescence image of FITC‐labeled heparin‐coated inside the MBO showed a similar pattern to the coated MBO before perfusion by PBS without any significant reduction in its intensity. The bonding mechanism of heparin to the PDMS is unknown but could involve hydrogen bonding between hydroxyl groups of heparin and the oxygen atoms in the PDMS[ 36 ] as well as physical adsorption, thus giving a coating which appears to be stable in PBS for at least several hours. Furthermore, the heparin content was determined in PBS samples at 10 min intervals during the first three hours. It appeared that a small heparin quantity was removed initially, but no additional loss occurred over the subsequent three hours period (Figure 2e). These data suggest that the heparin coating was sufficiently stable under the conditions to be used with the oxygenator device.

A solution of FITC‐fibrinogen and PBS buffer (2 mg mL−1) was circulated to coated and non‐coated MBOs to visualize the rate of the protein adsorption on the non‐coated and coated surfaces (Figure 2f,g). The fluorescence intensity of each MBO was quantified using ImageJ software, which showed a significant difference in fibrinogen adsorption between coated and non‐coated MBOs suggesting that the heparin coating was effective in reducing fibrinogen adsorption (Figure 2h).

Antithrombin (AT) is a plasma protein that inhibits several pro‐coagulant factors in the coagulation cascades, most importantly thrombin, and heparin acts as a catalyst in the AT‐thrombin reaction. AT has a specific binding affinity for heparin, and it was expected that circulating AT in plasma would bond to immobilized heparin on the PDMS surface, forming an AT‐heparin complex that is known to inhibit thrombin at a much higher rate than AT alone. As a result, the chance of clot formation would be reduced significantly.

2.3. Heparin‐Coated MBOs Do Not Exhibit Appreciable Clot Formation in Plasma

The formation of micro‐clots can impact device performance. This is especially true for a device with microchannels that could be easily occluded by micro‐clots, resulting in uneven flow distribution, decreased gas transfer rates, and possible device failure.[ 37 , 38 ] To investigate the possible formation of micro‐clots, a plasma clotting assay was performed on both heparin‐coated and uncoated MBOs using recalcified citrated plasma (Figure 2i). The pressure in circuits containing uncoated MBOs increased sharply after 29 ± 3.5 min (by comparison, recalcified plasma stored in a falcon tube clotted after about 20 min). In contrast, no perceptible change in pressure was observed in the loops with coated MBOs up to 150 min, demonstrating the beneficial effect of the heparin coating (Figure 2j). These plasma clots were visible in the samples collected from the loops with uncoated devices, while the plasma in the loops with coated devices was clear (Figure 2k).

2.4. Microfluidic LAD Rescues Hypoxic Piglet from Respiratory Distress

The LAD was tested in vivo exposed to room air or to oxygen‐enriched air using sweep gas on a hypoxic piglet model. Piglets represent the human neonate more closely than any other animal model and were chosen for this experiment. The experimental setup was adapted from Rochow et al.[ 16 ] (Figure  3 ), where the animal was connected arterial‐venously to an extracorporeal circuit that has a short cut loop that returns the blood without any oxygenation and to the LAD in parallel. Initially, when the extracorporeal circuit was connected and perfused by the pressure difference provided by the piglet's heart, the LAD loop stayed closed, while blood flowed through the shortcut line. Under the normoxic condition, this setting was used to ensure that all vital parameters were in steady‐state (termed shortcut normoxic). Hypoxic condition targeting peripheral oxygen saturation (SpO2) of 60% was simulated for the paralyzed piglet by adjusting the ventilator: peak inspiratory pressure (PIP) of 10 cm H2O, positive end‐expiratory pressure (PEEP) of 5 cm H2O, IT of 0.33 s, and RR of 15–19 breaths min−1. After achieving the target SpO2, immediately blood gases and cardiovascular parameters were measured without LAD support, while blood flowed through the shortcut (termed shortcut hypoxic). In order to bring the LAD online, the shortcut valve was closed, and the LAD circuit valve opened, redirecting the blood flow through the LAD loop (termed LAD hypoxic). Figure S2, Supporting Information, shows the sequence of conditions that the piglet was exposed to over the time frame for the in vivo experiment. Initially, baseline measurements were made by exposing the piglet to normoxic conditions with the blood flowing through the shortcut. Then, the hypoxic condition was initiated, and blood flow switched between the shortcut and through the LAD in three cycles each when the LAD is exposed to pure oxygen and air. Multiple blood samples were taken under each of these conditions and at various time points, as shown in Figure S2, Supporting Information. The blood samples were collected from the femoral artery, right atrium, before the LAD, and after the LAD to conduct blood gas analysis. At the end of the experiment, shortcut normoxic condition was established (control), and a final set of blood samples were taken for measurement of blood gas levels to ensure that the piglet recovered to the initial baseline and was not affected by the intervention.

Figure 3.

Figure 3

Experimental setup of the newborn piglet model for testing a pumpless neonatal lung assist device for artificial placenta application.

2.5. Systematic Blood Measurements

It is vital to discern the effect of the extracorporeal circuit on systematic blood parameters such as blood flow rate, heart rate, mean LAD/shortcut pressure, and mean blood pressure. Since the LAD was connected to the systemic circulation (an arteriovenous connection) in a parallel configuration, the resistance to blood flow would be reduced, which would result in a higher systemic blood flow rate (under short cut normoxic condition). When the piglet is made hypoxic, the systemic blood flow rate was found to increase as it compensates for the lower oxygen saturation level in the blood (Figure  4ae). However, when the extracorporeal blood circulation is switched from the shortcut to the LAD, it adds not only oxygen to the blood but also increases the resistance in the parallel extracorporeal circuit. This results in the observed lowering of the systemic blood flow rate when the LAD is connected. Heart rate was found to be elevated under the hypoxic condition to pump blood faster and increase the rate of gas exchange, but it dropped back to normal levels when the LAD became active, indicating that the LAD was successful in alleviating the hypoxic condition (Figure 4bf). The mean blood pressure at the location where the blood emerges extracorporeally (Figure 4cg) was always lower in the shortcut hypoxic condition compared to the LAD hypoxic (the condition that had the LAD in the circuit) confirming that the LAD line achieved lower flow rates. Moreover, mean systemic blood pressure (at the femoral artery) (Figure 4dh) was elevated under the hypoxic conditions and returned to normal levels when oxygenation through the LAD is initiated.

Figure 4.

Figure 4

Effect of extracorporeal bypass on cardiovascular parameters: a) blood flow rate while the LAD was using oxygen as the sweep gas, b) heart rate while the LAD was using oxygen as the sweep gas, c) mean LAD/shortcut pressure while the LAD was using oxygen as the sweep gas, d) mean systemic arterial blood pressure measured at the femoral artery while the LAD was using oxygen as the sweep gas, e) blood flow rate when the LAD was exposed to room air, f) heart rate when the LAD was exposed to room air, g) mean LAD/shortcut pressure when the LAD was exposed to room air, and h) mean systemic arterial blood pressure measured at the femoral artery when the LAD was exposed to room air.

2.6. In Vivo Gas Exchange Testing

At the end of each cycle, blood samples were collected extracorporeally before and after the LAD as well as from the femoral artery and the jugular vein from the right cardiac atrium to determine systemic blood gas levels. The saturation of oxygen (SaO2), the partial pressure of oxygen (pO2), and partial pressure of carbon dioxide (pCO2) were measured in these samples by a blood gas analyzer. Under the shortcut hypoxic condition, the blood flow rate was stable and ranged from 102 to 110 mL min−1. When the LAD (LAD hypoxic condition) was connected, its hydraulic resistance was low enough for it to be perfused solely by the piglet's heart and generate a stable blood flow rate from 65 to 72 mL min−1 (37–41 mL min−1 kg−1), clearly demonstrating that the design was able to support clinically relevant flow rates in a pumpless manner. When the LAD was placed in an enriched oxygen environment (a chamber with a slight oxygen flow of ≈15 L h−1), the LAD was able to fully oxygenate the blood at these relevant flow rates in all experiments. In one instance, it was able to increase the Sa O2 from a low 78% to 100% at a high blood flow rate of ≈70 mL min−1 (Figure  5a). The LAD also could significantly increase the pO2 up to 473 mm Hg (as compared to 68 mm Hg at its inlet), resulting in highly oxygenated blood leaving the LAD outlet (Figure 5b). pCO2 was also decreased at this condition with a maximum drop of 81 mm Hg to 50 mm Hg at a blood flow rate of 70 mL min−1 (Figure 5c).

Figure 5.

Figure 5

Gas exchange at the LAD when connected in vivo to a piglet. Measurements were of the blood at the inlet and outlet of the LAD when it is connected to the piglet and pumped by the arterio‐venous pressure difference: a) Sa O2,b) pO2, c) pCO2 before and after the LAD using oxygen as the sweep gas, and d) Sa O2, e) pO2, f) pCO2 before and after the LAD exposed directly to room air.

When the LAD was exposed to ambient air, the gas exchange performance was lower but still significant. For instance, in one cycle, the LAD was capable of increasing the SaO2 from 73% to 90% at a high blood flow rate of 85 mL min−1 (Figure 5d), demonstrating a clinically relevant performance during pumpless operation. pO2 was significantly enhanced after the LAD, and a maximum value of 66 mm Hg was reached at a blood flow rate of 85 mL min−1 (Figure 5e). pCO2 was also reduced after the LAD (Figure 5f).

The parameters pO2, SaO2, and pCO2 were also measured at the femoral artery and right atrium (mixed deoxygenated blood, venous blood from systemic circulation, and oxygenated blood returning from the LAD; ultrasonography of the right atrium of the heart was performed to assure that the catheter was correctly placed as shown in Figure S3, Supporting Information) to determine the impact of the LAD on systemic oxygenation for the piglet. Measurements were performed when the piglet was under the hypoxic condition. Both systemic pO2 and SaO2 (femoral) increased when the LAD exposed to pure oxygen environment was connected with the hypoxic piglet with a maximum increase of 41 mm Hg to 66 mm Hg in pO2 and 52% to 86% in SaO2 (Figure  6a,b). When the LAD was exposed to room air, the increase in systemic pO2 and SaO2 was lower but an increase of pO2 from 42 mm Hg to 63 mm Hg and Sa O2 from 56% to 84% at a blood flow rate 88 mL min−1 was achieved (Figure 6d,e). pCO2 was decreased in both oxygen environment and room air (Figure 6c,f).

Figure 6.

Figure 6

Gas exchange at the femoral artery and the right atrium when connected in vivo to a piglet. Measurements were of the blood at the femoral artery and the right atrium when the LAD is connected to the piglet and pumped by the arterio‐venous pressure difference a) pO2, b) Sa O2, c) pCO2 for femoral artery using oxygen as the sweep gas, d) pO2, e) Sa O2, f) pCO2 for femoral artery exposed directly to room air, and g) pO2, h) Sa O2, i) pCO2 for right atrium using oxygen as the sweep gas, j) pO2, k) Sa O2, l) pCO2 for right atrium exposed directly to room air.

Similar results were observed when measurements were performed at the right heart (atrium). Systemic pO2 and Sa O2 significantly increased when connected to the LAD, but their values at the atrium (Figure 6g,h) were lower than levels immediately after the LAD as blood in the atrium is a mixture of the venous blood from the systemic circulation, which is usually rich in carbon dioxide and poor in oxygen with the oxygenated blood from the LAD. During each cycle, pCO2 was also elevated after connecting the LAD, which was placed under an oxygen‐rich hood (Figure 6i). When the LAD was exposed to room air, pO2 and Sa O2 measured from right atrium also showed an improvement in oxygen content (Figure 6j,k).

pCO2 was slightly lower after switching from the shortcut to the LAD extracorporeal connection under hypoxic conditions suggesting that the LAD could be more effective in removing carbon dioxide while exposed to room air (Figure 6l). As expected, the oxygen saturation level dropped in both the femoral artery and the right atrium when the piglet was in the hypoxic state, and the LAD was inactive (Figure 6). However, the saturation level significantly increased when the LAD was connected. For instance, when exposed to the hypoxic condition, the oxygen saturation level in the right atrium dropped to 19% but increased to 57% after connecting the LAD (Figure 6h). This effect was observed when the LAD operated in room air with a maximum rise from 31% to 59% in Sa O2 (Figure 6k).

3. Discussion

We tested a new LAD in a newborn piglet model, which was constructed of 16 stacked MBOs and designed to support preterm and term infants with respiratory failure. In this study, the LAD provided significant gas exchange throughout a period of ≈200 min. The current extracorporeal circuit achieved blood flow rates through the LAD was, on average, one‐third of the cardiac output, which was adequate to maintain the piglet cardiovascularly stable throughout the experiment.[ 16 ] Our concept of the LAD includes the features of an artificial placenta, which allows the newborn to continue to breathe while a partial fetal circulation via the umbilical artery, LAD circuit, and the umbilical vein is established.[ 16 ] An important element is that the newborn will continue to breathe while the LAD provides additional gas exchange and allows the lungs to heal. In this way, lifelong side effects due to performing mechanical ventilation at the physiological limit could be prevented. Although our concept of the artificial placenta is designed to be connected to the umbilical vessels,[ 16 ] at this developmental stage, we used the vascular access via carotid artery and jugular vein. This is a compromise of the piglet model. We found that in a one‐day‐old piglet, the lumen of the umbilical vessels became fibrotic. Using a Seldinger wire technique, only catheters not bigger than an outer diameter of 1.2 mm could be placed, which do not provide sufficient blood flow. Other researchers have used lamb models which have umbilical vessels with larger diameters.[ 1 , 2 , 3 , 39 ] Also, they reconnected the artificial placenta immediately after birth, which prevents umbilical vessels from obliteration.

Currently, there are two different approaches to artificial placenta type oxygenation. One concept aims to develop an artificial womb for prolongation of the pregnancy. This would allow developing the infant to maturity while submerged in artificial amniotic fluid. However, the entire gas exchange, like in a womb, would be facilitated by the artificial lung device.[ 1 , 2 , 39 , –41 ] Current studies in this version have used commercial, high priming volume, hollow fiber oxygenator, and an external pump for blood perfusion. Also, a preterm lamb that matches the appropriate gestational age but is heavier (weight up to ≈5 kg) than human neonates (1–2 kg) were used. One of the reasons for the use of lambs was the unavailability of small priming volume, pumpless oxygenators of the kind demonstrated in our study.

The second approach, which is used in this study, is to provide partial oxygenation support to a newborn neonate in respiratory distress to reduce the load on the lungs and allow it to recover while it continues to breathe.[ 16 ] We specifically designed and optimized our microfluidic LAD for human preterm and term neonate resulting in the choice of a piglet model as it has similar weight, blood volume, and physiological characteristics to a human preterm neonate among various animal models. The stackable modular design provides a flexible application from 500 g to 4 kg body weight and facilitates pumpless extracorporeal support with the minimum possible priming volume. Furthermore, the ability to custom fabricates the microfluidic oxygenator allowed us to achieve the specific design criteria for the priming volume, hydraulic resistance, and oxygen uptake suitable for human preterm neonates. By using the piglet of the same birth weight as preterm neonates, we were able to accurately test the extracorporeal system for oxygenation performance. The smaller priming volume of this oxygenator also mitigated the effect of hemodilution when the saline that was used to fill it was mixed with the venous blood. The custom and modular designs were also critical for pumpless operation as hydraulic resistance of the oxygenator could be tailored to produce adequate blood perfusion and oxygen uptake based on the baby's weight, and an external pump could be eliminated.

Since the arteriovenous pressure difference in a neonate is typically between 20–60 mm Hg,[ 12 , 16 , 29 , 30 , 31 ] such a pressure head applied to this LAD would generate a flow rate between 65–86 mL min−1 in this optimized LAD, which meets the requirement of a blood flow rate of 20–30 mL min−1 kg−1 of a neonate typically needed to provide lung assist function (Figure 4ae). In this range of flow rates and under in vitro conditions, this LAD was able to increase the oxygen saturation level from ≈75% to ≈100%, which would be sufficient to fulfill the oxygenation need for these preterm babies.[ 12 , 16 , 29 , 30 , 31 ]

Of note, the hemoglobin of the piglet was at a lower level (hematocrit: 23–27%) in these experiments. A preterm human infant typically has considerably higher hematocrit level (40–60%),[ 41 ] and as a result, the oxygen‐carrying capacity of the blood would allow higher amounts of gas exchange and oxygen delivery compared with the demonstrated capacity in this study. The performance of this LAD connected to the human neonate can be estimated for comparison purposes by calculating its performance at these higher hematocrit levels. Such calculation (as shown in Figure  7 ) should include physiological trends, including a postnatal decrease in hemoglobin content[ 41 ] and a corresponding increase in systemic blood flow rate[ 42 ] during first days of life for which measurements exist.

Figure 7.

Figure 7

An estimation of oxygen transfer for the LAD if it was tested for preterm or term neonates: a) the flow chart for the calculation of the estimated oxygen transfer, b) the relationship between the in vivo blood flow rates and the mean arterial blood pressure when the LAD (MBOs' height = 180 µm) was exposed to room air, c) measured hemoglobin contents for preterm infants with a gestational age of 29–34 weeks (the date was acquired from J. Jopling et. al.[ 41 ]) showing the hematocrit is decreased, d) the measured mean arterial blood pressure for newborns with different weights over 96 h after birth (the date was acquired from J. E. Jones et. al.[ 42 ]), e) the estimation of blood flow rates if the LAD was tested for premature neonates (the mean arterial blood pressure of neonates[ 42 ] were fed into the calibration line calculated in [b]), f) estimated oxygen transfer by the LAD while consuming oxygen or air as the sweep gas for preterm neonates with different birth‐weights over 96 h afterbirths and AP shows the reference values for different weights in an artificial placenta configuration when the LAD is supposed to provide 30% of oxygen consumption.[ 43 ]

The procedure for this calculation is shown in Figure 7a. Briefly, the correlation between mean arterial blood pressure (MABP) and blood flow rates (BFR) in the LAD line was found using the data gathered from the piglet experiment (Figure 7b) and represents the flow characteristics of the LAD using a heart driven operation. Hematocrit levels in neonates vary with gestational age (GA) and have been measured previously[ 41 ] for GA between 29 and 34 weeks and are plotted in (Figure 7c). Similarly, MABP values for different postnatal age (up to 96 h after birth) and various weights (500–1000 g) have also been measured previously[ 42 ] and this data is plotted in Figure 7d. The blood flow rate in the LAD (Figure 7e), if it was attached to human neonates of different body weights and post natal age, was calculated using the data from Figures 7b,d. Next, the change in oxygen saturation level was estimated for each flow rate using in vitro data as the hematocrit in this experiment (≈38%) was closer in range to that of human neonates. Finally, the amount of oxygen transfer (Figure 7f) was calculated for different post natal age and weights, considering the changes in hematocrit level over time (Figure 7c), and the blood flow rate increases (Figure 7e) while the baby is growing. This analysis shows that estimated flow rates in the LAD would be between 35–50 mL min−1 for human neonates. For these flow rates, the in vitro results show that oxygen saturation can be increased atleast 30% or more when oxygen is used as a sweep gas and by 25% when air is used instead. The estimation of oxygen transfer indicates that for preterm neonates with weights from 500 g to 1 kg, over a period of 96 h after birth, the current configuration of the LAD is capable of providing a gas exchange capacity of 4.2 mL min−1 kg−1 using oxygen as the sweep gas or 3.1 mL min−1 kg−1 while the LAD would be exposed to room air. It corresponds to ≈50–75% (≈7 mL min−1 kg−1) of the total oxygen requirement for a resting 1 kg preterm infant at 96 h postnatal age.[ 43 , 44 ] This capacity far exceeds that required for our artificial placenta type oxygenator, which is designed to assist the lungs and provides ≈30% of this total oxygen requirement.[ 8 , 16 ]

The arteriovenous connection used in our approach directs partially oxygenated blood from the lungs into the LAD. The LAD assists by further increasing the oxygen saturation to 100% before returning it back to the body. Calculations show that the LAD is capable of supporting a 1.5 kg neonate by increasing the oxygen saturation of 50 mL min−1 of its blood from ≈75% to 100% in room air, thereby providing 30% of the total oxygen requirement and preventing respiratory distress. Furthermore, the return of oxygenated blood to the right heart also assists in counteracting hypoxic pulmonary vasoconstriction and decrease pulmonary resistance and improve lung perfusion.[ 45 ]

The piglet was cardiovascularly stable throughout the entire in vivo experimentation that occurred over a period of ≈200 min. We did not observe bradycardia or tachycardia or arrhythmia, and the blood pressure was in the desired range. Our LAD was connected in parallel to the systemic circulation and considerable amounts of blood bypassing the systemic circulation extracorporeally. The bypassed blood flow (65–86 mL min−1) reaches up to one‐third of the cardiac output. This clinical condition is similar to a patent ductus arteriosus or a ventricle septum defect and could pose a risk to the heart of high cardiac output failure. Clinical signs would be tachycardia, low systemic blood pressure, and lactacidemia. We found that the blood pressure of the piglet remained stable (Figure 4dh), no significant increase of the heart rate (Figure 4bf) was noted, and no metabolic acidosis could be observed (Figure S6, Supporting Information).

The LAD was able to increase the oxygen saturation of blood to desired target levels for preterm neonates with respiratory failure. Systemic oxygen levels were measured using a pulse oximeter. A typical range of SpO2 is 85–95% for normal neonates, preferably above 90%.[ 46 ]

To achieve a hypoxic condition, we reduced the ventilation rate of the paralyzed piglet and targeted hypoxic SpO2 of 60%. It was expected that adding the LAD will increase the systemic SpO2 to a range of 85–95%. Notably, this target range for SpO2 was achieved in all three cycles when the ventilation gas for the LAD was pure oxygen (results are shown in Figure S5, Supporting Information). Of particular note was in the third cycle, where the SpO2 was increased from 57% (under hypoxic condition) to 91% (upon connection with the LAD). When room air was used as the ventilation gas, the LAD was still able to increase SpO2 during all cycles. However, the SpO2 does not reach above 85% except in the second cycle, where it increased from 69% to 90% (upon the introduction of the LAD). This is because of the excessively low oxygen saturation (42% in cycle 1) induced in the hypoxic condition in cycles 1,3. Even then, the LAD was able to increase systemic levels of SpO2 to ≈80%, which approaches the systemic SpO2 in healthy newborn piglets with a weight of 1.77 ± 0.13 kg varies from 84% to 96%.[ 47 ] Overall, the in vivo experiments demonstrate that the LAD could significantly improve systemic levels of SpO2 either using oxygen or room air as the ventilation gas in a piglet under hypoxic conditions.

The piglet required continuous normal saline infusion at a high rate of more than 15 mL kg−1 h−1 due to significant insensible water losses. There are two main contributing factors: first, the piglet is exposed to a radiant warmer to maintain a body temperature of 39 °C, which promotes evaporation of water. Studies of ventilated ill newborns under a radiant warmer showed that these infants lose 28.04e−1.73(weight in kg) mL kg−1 h−1.[ 48 ] Second, the LAD is composed of PDMS, which is permeable to water. Water from the circulating blood could have evaporated as a result. However, fluid support during the experiments prevented signs of metabolic acidosis, and the hydrogen carbonate remained in the desired range, as well as the urine output was stable, indicating sufficient hydration. As a result, a possible future modification to minimize evaporation and prevent heat loss can be to surround the LAD by humidified air.

We used low dose hydrocortisone and dopamine in this study to allow sufficient stress response, to minimize inflammatory responses, and to provide vascular tone. In our experimental setting, the piglet underwent surgery for the establishment of the tracheostomy, and catheters were placed invasively. Further, exposing the piglet's blood to the extracorporeal tubes and LAD can potentially induce inflammation, including mild capillary leak syndrome. An increase in vascular tone by dopamine, as well as the effects of hydrocortisone on the reduction of inflammation and blood pressure, provide stable conditions for this experiment.

An essential characteristic of the LAD is hemocompatibility because the LAD should not induce hemolysis or blood clot formation. The hematocrit of the piglet remained throughout the experiment at expected levels. There was only a marginal drop of the hematocrit by 4% from 27% to 23%. We estimated the blood loss to be approximately 4 mL kg−1, which could be explained by several blood collections during the study. No side effects of clinically significant anemia, such as tachycardia or acidosis, could be recognized. However, future studies should measure hemolysis.

To prevent blood clotting and to establish hemocompatibility, a key feature of our LAD is the coating of its inner surfaces. In vitro, we performed several experiments to show that heparin was adsorbed on PDMS surfaces and was fairly stable (it could be hypothesized that some of the heparin was absorbed by the PDMS surfaces). This bonding could be explained by the formation of hydrogen bonding or electrostatic interaction between heparin molecules and PDMS chains.[ 36 ] We also stored the plasma samples after performing the plasma clotting assays observing if plasma would be clotted after being exposed to an external surface without heparin coating. We observed that the plasma samples were not clotted even after being exposed to a non‐coated surface for a few hours, suggesting that the heparin was not covalently bonded to PDMS. This was expected and was also reported in another study.[ 36 ] Therefore, in our experiment, we primed the LAD with a normal saline and heparin solution. However, this approach alone was not sufficient to prevent blood clotting, and we used it in conjunction with a typical systemic dose of heparin for systemic anticoagulation. Nevertheless, the coating was important in preventing any deleterious clot formation on the surfaces of the oxygenators and critical for the smooth operation of the LAD in in vivo experiments. In preterm infants, this approach of systemic anticoagulation may not be feasible. Preterm infants are at high risk for intraventricular hemorrhage; in particular, those infants who are born preterm from mothers with chorioamnionitis, HELLP, or preeclampsia. Systemic anticoagulation would elevate this risk significantly. Initial work to achieve the hemocompatibility of the LAD has been done by our group. The coating of inner surfaces of the LAD, the tubes, and the catheter with the novel antithrombin–heparin complex are promising.[ 38 , 49 ] This type of coating could be applied to all internal surfaces to enhance hemocompatibility.

4. Conclusion

In this study, a pumpless LAD using microfluidic blood oxygenators for preterm neonates suffering from respiratory failure was developed. In vitro assessments were done by blood to investigate the gas exchange capacity of this LAD resulting in an acceptable gas transfer in room air and ample oxygenation in an oxygen‐rich environment. A simple coating technique was developed to coat blood‐in‐contact surfaces. The functionality and stability of this coating were evaluated in vitro, suggesting that the coating was effective and stable enough to be used in further animal studies. In the next step, the LAD was tested in a pumpless operation connected to a newborn piglet. Results from the piglet experiments revealed the effectiveness of this LAD in gas exchange without complications and agreed with in vitro results. Further steps in the development of the LAD include the installation of hemocompatibility, biocompatibility, and developing via the umbilical vessels.

5. Experimental Section

The Design of Microfluidic Blood Oxygenator Device

The MBO was constructed entirely out of PDMS comprising three components: i) a blood vascular network that spread the incoming blood into a thin uniform layer suitable for fast gas exchange, enclosed by ii) two thin composite membranes made of polytetrafluoroethylene (PTFE) and PDMS with high permeability for blood gases and sufficient stiffness to withstand the physiological pressure encountered, and (iii) a tapered shaped inlet/outlet to access the blood vascular network. The blood vascular network was a microchamber with a height of 180 µm (or 130 µm) containing an array of 1 × 1 mm square‐shaped micropillars. The design, which is shown in Figure  8a, was optimized to ensure that a uniform blood distribution was achieved. It had a surface area of 29.19 cm2, and a priming volume of 0.52 mL. Tapered inlet and outlet were integrated on top of the blood vascular network to ensure smooth distribution of blood from a cylindrical inlet connection into the vascular network (Figure 8b). The design minimizes shear stress that could lead to hemolysis, platelet activation, and clot formation. Numerical modeling was conducted by COMSOL Multiphysics (COMSOL Multiphysics, V5.3, USA) to show the uniform flow distribution in the blood vascular network without creating any dead zone in the inlet (Figure 8c).

Figure 8.

Figure 8

a) 3D schematic drawing of MBO with tapered inlet/outlet configuration, b) top view of MBO, c) the velocity profile of MBO at a blood flow rate of 5 mL min−1 showing a uniform distribution with a gentle change in flow direction and no dead zone, and d) 3D drawing of the LAD in an unassembled format showing all 3D‐printed components, MBOs, and flow distributors. There are fixed gaps among all MBOs for air to flow convectively. Blood is fed to the LAD from the bottom flow distributor and leaves the LAD from the top one to ensure that no air bubble is trapped inside.

MBO Fabrication

A precisely controlled layer‐by‐layer microfabrication process was developed to manufacture all MBOs (Figure  9ae) and described in the previous publication.[ 30 ] Briefly, a negative mold of MBOs (Figure 9a) was fabricated by conventional photolithography using a negative photoresist (SU8 3035, MicroChem. Corp., Westborough, MA, USA). PDMS (Sylgard 184, Dow Corning, Midland, MI) with a ratio of 10:1 (base to curing agent) was prepared and coated on the mold using a spin‐coater with a speed of 750 RPM for 60 s, followed by being cured on a hot‐plate at 85 °C for 30 min. Another layer of PDMS was spin‐coated (2000 RPM for 40 s) and a porous PTFE membrane (pore size of 1 µm and a porosity of 83%, PTU103001H, Sterlitech, Kent, WA, US) was embedded into the wet PDMS, and cured in an oven at 85 °C (Figure 9b). The blood vascular network was removed from the mold (Figure 9c). Another porous hydrophobic PTFE membrane (pore size of 0.22 µm and porosity of up to 90%, PTU023001, Sterlitech, Kent, WA, US0) was fixed on a hydrophobic substrate, which was made by sticking a piece of PTFE film on a silicon wafer and covered by wet PDMS for ≈15 min to ensure that all pores were filled by PDMS. This substrate with PDMS‐wetted PTFE membrane was spun at 8000 RPM for 60 s and brought into contact with the blood vascular network (Figure 9d). The device was placed on a leveled surface and cured overnight at room temperature. Two inlets with tapered configuration were placed on their designated locations on the top of the blood vascular network being glued by half‐cured PDMS, and they were cured in an oven at 85 °C for an hour (Figure 9e). A scalpel and a biopsy puncher were used to remove residual PDMS inside the inlet and outlet, and the other sides were sealed by a piece of cured PDMS. Figure 9f shows an MBO with the tapered inlet, which was filled with dyed DI water.

Figure 9.

Figure 9

a–e) The microfabrication process of MBOs and f) a close‐view image of an MBO with the tapered inlet configuration filled with dyed DI water.

The Assembly of the Compact Neonatal Lung Assist Device

A single MBO could not provide enough oxygenation to support preterm neonates with respiratory failure with different weights. As a result, the LAD was constructed by assembling 16 MBOs in a parallel configuration (Figure 8d). The frame was 3D‐printed to arrange the MBOs in a compact form and consisted of several fixtures to hold MBOs along with a top and bottom holder to maintain fixed gaps between the MBO holders. Each MBO holder was designed to hold two MBOs at a fixed distance apart from each other to facilitate convective gas flow on all sides of the MBO, naturally. Then, eight such MBO holders were inserted into the appropriate slots in the top and bottom holders to form the compact LAD.

Autodesk Inventor software (2018, USA) was used to draw all components for the compact LAD. All holders were manufactured by a Prusa i3 mk3 3D printer (Partyzánská, Praha Czech Republic), including eight MBO holders, one top and bottom holder, two rack holders, and two flow distributor holders (the bottom one was 3D‐printed by four legs). The flow distributor was cast out of a 3D‐printed mold, which was made by a Formlab Form 2 3D printer (Somerville, MA, USA) using PDMS with the similar mixing ratio and bonded later to a cured 1‐mm‐thick PDMS by means of a flame treatment[ 50 ] and an oxygen plasma activation (1.5 min exposure, an oxygen pressure of 900 mm Torr, Harrick plasma cleaner, Ithaca, NY, USA). For each LAD, at least 32 MBOs were fabricated, and their hydraulic pressure characteristics were quantified. Then, those with similar hydraulic behavior were chosen for LAD assembly.

In Vitro Testing of LAD Using Bovine Blood

Bovine blood was purchased from Lampire Biologicals Laboratories (Bovine 7200807—1L, Pipersville, PA) and heparinized by a concentration of 3 units mL−1 right away after delivery and was stored in a fridge overnight. A hollow fiber oxygenator (PDMSXA‐1.0, PermSelect VR, Ann Arbor, MI, USA) was used to lower the oxygen content of the blood to ≈60% by passing a carbon dioxide/nitrogen gas mixture (5%/95% v/v) into the shell while blood was fed into the fibers using a peristaltic pump (Ismatec model ISM832C, Ismatec, Glattbrugg, Switzerland) at 15–25 mL min−1. The blood was collected in a sealed bottle and stored again in a fridge overnight. Prior to each experiment, the bottle of the blood with a sealed cap was warmed up to room temperature, and an open‐loop circuit was created by pumping (a peristaltic pump was placed in the circuit) the blood from the bottle to the LAD, then to a waste bottle. Blood samples were taken before and after the LAD and analyzed by a point‐of‐care blood gas analyzer (GEM Premier 3000, Instrumentation Laboratory, Lexington, MA, USA). A Complete MicroHematocrit System (StatSpin CritSpin, Norwood, MA, USA) was used to measure the hematocrit level of blood samples as well.

Heparin Coating

Heparin sodium injection USP with 1000 USP units per mL concentration was a high molecular weight polysaccharide and was purchased from Sandoz Canada Inc. It was a sterile, pyrogen‐free solution with high purity derived from the porcine intestinal mucosa. FITC‐labeled heparin was obtained from Creative PEGWorks (HP‐201, Chapel Hill, NC, USA) and had a high molecular weight of 27 kDa. FITC‐labeled heparin was only used for visualizing the quality of coating inside MBOs. Heparin sodium injection USP was used for coating LADs in the animal experiments as well as in the in vivo experiment for systemic heparinization. Generally for the coating, heparin sodium injection USP or FITC‐labeled heparin was circulated in the LADs or MBOs prior to each experiment in a closed‐loop configuration for 4–5 h followed by washing with PBS or normal saline (only for LADs tested in the animal experiments) to remove any uncoated heparin for ≈15 min.

FITC‐Fibrinogen Adhesion Test

Human Fibrinogen Purified FITC Labeled was obtained from Innovative Research Inc. (IHUFBGFITC50MG, Novi, MI, USA) and stored at −20 °C. Prior to each experiment, FITC‐fibrinogen was warmed up to room temperature and was added to PBS buffer with a pH of 7.4 to investigate the antithrombotic property of the non‐coated and coated surfaces. Coated and non‐coated MBOs were perfused for 3 h in separate closed‐loops containing 2 mg mL−1 FITC‐fibrinogen (a total volume of ≈5 mL) and the MBOs were washed with PBS (pH = 7.4) for 30 min with a flow rate of 3 mL min−1. All other components used in the loop for testing MBOs, such as tubes and connectors, were coated with heparin using the same procedure. This was done to ensure that no part of the circuit would induce clot formation.

Plasma Clotting Assays in MBOs

Human plasma was purchased from Innovative Research Inc. (Novi, MI, USA) and prepared by mixing 5 mL of 25 mm CaCl2 with 5 mL human plasma.[ 51 ] The recalcified plasma was transferred immediately to a closed test loop containing the MBO. Care was taken to avoid the formation of air bubbles. The plasma was then circulated in the loop, and clotting was assessed by measuring the pressure change at the inlet of the oxygenator.

Experimental Setup of the Newborn Piglet Model

The newborn piglet was first weighed (1.78 kg) and anesthetized by intraperitoneal injection of sodium pentobarbital (30 mg kg−1, MTC Pharmaceutics, Cambridge, ON, Canada). To monitor the body temperature and maintain it at 39 °C during the experiment, an infant servo control (ISC) probe was placed on the piglet's abdomen. To provide controlled mechanical ventilation, an endotracheal tube (3.5 mm inner diameter, Portex, Keene, NH, USA) was placed via tracheostomy and connected to a Servo 300 ventilator (Siemens, Mississauga, ON, Canada). The ventilator was adjusted in a way to assure that the newborn piglet experienced a physiologic gas exchange at the beginning: PIP of 14 cm H2O, PEEP of 5 cm H2O, inspiratory time (IT) of 0.33 s, and respiratory rate (RR) of 31 breaths min−1. A mixture of oxygen (30%) and nitrogen (70%) was humidified and warmed up to 38 °C for delivering to the piglet (at a flow rate of 8 L min 1). 22‐G angio‐catheters were used to cannulate femoral and abdominal veins for maintenance fluids: sodium pentobarbital (16 mg kg−1), 5% dextrose (80 mg kg−1 day−1), hydrocortisone and dopamine infusion (2 µg kg−1 min−1). The dopamine was adjusted (2–20 µg kg−1 min−1) to maintain the mean systematic blood pressure in a range of 68–80 mmHg. During the experiment, the piglet was paralyzed using a muscle relaxant (Pancuronium) with an initial bolus 0.2 mg kg−1 and a continuous infusion of 20 µg kg−1 min−1.

Two 14‐G 1.1‐inch angio‐catheters (Angiocath) were used to cannulate the left carotid artery and right jugular vein for creating an extracorporeal circuit in which the LAD was attached. This circuit consisted of two loops: 1) shortcut bypass and 2) the LAD bypass. Once one of the bypasses was opened, the other one was closed. A flow sensor (in the mainline) with a monitor and a pressure transducer (in a branching line connected to the mainline) with a blood pressure monitor was installed in the circuit to report the extracorporeal blood flow rate and blood pressure.

While the extracorporeal circuit was connected, a bolus of 400 units kg−1 of heparin as an anticoagulant was delivered to the piglet and maintained at an infusion rate of 30 units kg−1 h−1. Access to the right femoral artery was achieved by insertion of a 3.5 Fr Argyle umbilical catheter. This vascular access was used to measure systemic blood pressure, heart rate, and blood collections. The catheter was perfused by a heparinized normal saline solution with a concentration of 3 units mL−1 at a perfusion rate of 1.5 mL h−1. A pressure transducer with a blood pressure monitor was attached to the catheter to monitor systemic blood pressure during the experiment. Another vascular access was established via the right internal jugular vein using a 3.5 Fr Argyle umbilical catheter, which was advanced under ultrasound into the right atrium of the heart, which allowed measuring the mixed blood (systemic venous blood and LAD oxygenized). Peripheral oxygen saturation was also monitored using a pulse oximeter, which was placed on one of the feet.

The study protocol was approved by the McMaster University Animal Research Ethics Board (AREB #10‐03‐14). The animal was humanely euthanized with Euthanyl Forte (Bimeda‐MTC Animal Health Inc., Cambridge, Ontario, Canada) as approved in the animal utilization protocol.

Statistical Analysis

For in vitro blood experiments, one blood sample was collected before each LAD and three blood samples were collected after the LAD. COMSOL Multiphysics software was used for simulation of blood flow in MBOs. Five coated and non‐coated samples were used to measure CA. To quantify fluorescent intensity in MBOs, ImageJ software was used. FITC‐heparin absorbance measurements were repeated three times for each condition. For plasma clotting experiments, three coated and non‐coated MBOs were tested. For all experiments, data were shown as means ± SD (standard deviation). Two animal experiments were performed: the LAD having MBOs with a height of 130 µm only was tested in pure oxygen in three cycles of LAD hypoxic‐shortcut hypoxic condition; the LAD having MBOs with a height of 180 µm was tested in pure oxygen and room air in three cycles of LAD hypoxic‐shortcut hypoxic conditions.

Conflict of Interest

The authors declare no conflict of interest.

Supporting information

Supporting Information

Acknowledgements

This work was supported by the Natural Sciences and Engineering Research Council of Canada (NSERC) and Canadian Institutes for Health Research (CIHR) through the Collaborative Health Research Program. P.R.S. also acknowledges support from the Canada Research Chairs Program as well as the Discovery Accelerator Supplement grant. C.F. acknowledges support from Jack Sinclair Research Chair position at McMaster University. The authors would like to thank Devon Jones, a graduate student in CAMEF lab, for her assistance in the animal experiments. The authors also acknowledge Erin Warkentin (erin.warkentin@mail.utoronto.ca) for the illustration in Figure 3.

Dabaghi M., Rochow N., Saraei N., Fusch G., Monkman S., Da K., Shahin‐Shamsabadi A., Brash J. L., Predescu D., Delaney K., Fusch C., Selvaganapathy P. R., A Pumpless Microfluidic Neonatal Lung Assist Device for Support of Preterm Neonates in Respiratory Distress. Adv. Sci. 2020, 7, 2001860 10.1002/advs.202001860

References

  • 1. Partridge E. A., Davey M. G., Hornick M. A., McGovern P. E., Mejaddam A. Y., Vrecenak J. D., Mesas‐Burgos C., Olive A., Caskey R. C., Weiland T. R., Han J., Schupper A. J., Connelly J. T., Dysart K. C., Rychik J., Hedrick H. L., Peranteau W. H., Flake A. W., Nat. Commun. 2017, 8, 15112. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 2. Church J. T., Coughlin M. A., Perkins E. M., Hoffman H. R., Barks J. D., Rabah R., Bentley J. K., Hershenson M. B., Bartlett R. H., Mychaliska G. B., J. Pediatr. Surg. 2018, 53, 1896. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 3. Bryner B., Gray B., Perkins E., Davis R., Hoffman H., Barks J., Owens G., Bocks M., Rojas‐Peña A., Hirschl R., Bartlett R., Mychaliska G., J. Pediatr. Surg. 2015, 50, 44. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 4. Stoll B. J., Hansen N. I., Bell E. F., Walsh M. C., Carlo W. A., Shankaran S., Laptook A. R., Sánchez P. J., Van Meurs K. P., Wyckoff M., Das A., Hale E. C., Ball M. B., Newman N. S., Schibler K., Poindexter B. B., Kennedy K. A., Cotten C. M., Watterberg K. L., D'Angio C. T., DeMauro S. B., Truog W. E., Devaskar U., Higgins R. D., JAMA, J. Am. Med. Assoc. 2015, 314, 1039. [Google Scholar]
  • 5. Manktelow B. N., Seaton S. E., Field D. J., Draper E. S., Pediatrics 2013, 131, e425. [DOI] [PubMed] [Google Scholar]
  • 6. Laughon M., O'Shea M. T., Allred E. N., Bose C., Kuban K., Van Marter L. J., Ehrenkranz R. A., Leviton A., ELGAN Study Investigators , Pediatrics 2009, 124, 637.19620203 [Google Scholar]
  • 7. Stoll B. J., Hansen N. I., Bell E. F., Shankaran S., Laptook A. R., Walsh M. C., Hale E. C., Newman N. S., Schibler K., Carlo W. A., Kennedy K. A., Poindexter B. B., Finer N. N., Ehrenkranz R. A., Duara S., Sánchez P. J., O'Shea T. M., Goldberg R. N., Van Meurs K. P., Faix R. G., Phelps D. L., Frantz I. D., Watterberg K. L., Saha S., Das A., Higgins R. D., for the Eunice Kennedy Shriver National Institute of Child Health and Human Development Neonatal Research Network , Pediatrics 2010, 126, 443.20732945 [Google Scholar]
  • 8. Rochow N., Chan E. C., Wu W.‐I., Selvaganapathy P. R., Fusch G., Berry L., Brash J., Chan A. K., Fusch C., Int. J. Artif. Organs 2013, 36, 377. [DOI] [PubMed] [Google Scholar]
  • 9. Philip A. G. S., Neonatology 2012, 102, 1. [DOI] [PubMed] [Google Scholar]
  • 10. Chambers H. M., van Velzen D., Pathology 1989, 21, 79. [DOI] [PubMed] [Google Scholar]
  • 11. Attar M. A., Donn S. M., Semin. Neonatol. 2002, 7, 353. [DOI] [PubMed] [Google Scholar]
  • 12. Wu W.‐I., Rochow N., Chan E., Fusch G., Manan A., Nagpal D., Selvaganapathy P. R., Fusch C., Lab Chip 2013, 13, 2641. [DOI] [PubMed] [Google Scholar]
  • 13. van Heijst A. F. J., de Mol A. C., IJsselstijn H., Semin. Perinatol. 2014, 38, 104. [DOI] [PubMed] [Google Scholar]
  • 14. Federspiel W. J., Henchir K. A., Encyclopedia of Biomaterials and Biomedical Engineering (Eds: Bowlin G. L., Wnek G.), Marcel 49 Dekker, New York, 2004, p. 910.<./bib> [Google Scholar]
  • 15. Lee J. K., Kung H. H., Mockros L. F., ASAIO J. 2008, 54, 372. [DOI] [PubMed] [Google Scholar]
  • 16. Rochow N., Manan A., Wu W.‐I., Fusch G., Monkman S., Leung J., Chan E., Nagpal D., Predescu D., Brash J., Selvaganapathy P. R., Fusch C., Artif. Organs 2014, 38, 856. [DOI] [PubMed] [Google Scholar]
  • 17. Kung M. C., Lee J.‐K., Kung H. H., Mockros L. F., ASAIO J. 2008, 54, 383. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 18. Lee J.‐K., Kung M. C., Kung H. H., Mockros L. F., ASAIO J. 2008, 54, 390. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 19. Kniazeva T., Hsiao J. C., Charest J. L., Borenstein J. T., Biomed. Microdevices 2011, 13, 315. [DOI] [PubMed] [Google Scholar]
  • 20. Hoganson D. M., Pryor H. I., Bassett E. K., Spool I. D., Vacanti J. P., Lab Chip 2011, 11, 700. [DOI] [PubMed] [Google Scholar]
  • 21. Sreenivasan R., Bassett E. K., Hoganson D. M., Vacanti J. P., Gleason K. K., Biomaterials 2011, 32, 3883. [DOI] [PubMed] [Google Scholar]
  • 22. Potkay J. a., Magnetta M., Vinson A., Cmolik B., Lab Chip 2011, 11, 2901. [DOI] [PubMed] [Google Scholar]
  • 23. Kniazeva T., Epshteyn A. A., Hsiao J. C., Kim E. S., Kolachalama V. B., Charest J. L., Borenstein J. T., Lab Chip 2012, 12, 1686. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24. Lo J. H., Bassett E. K., Penson E. J. N., Hoganson D. M., Vacanti J. P., Tissue Eng., Part A 2015, 21, 2147. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25. Kovach K. M., LaBarbera M. A., Moyer M. C., Cmolik B. L., van Lunteren E., Gupta A. S., Capadona J. R., Potkay J. A., Lab Chip 2015, 15, 1366. [DOI] [PubMed] [Google Scholar]
  • 26. Rieper T., Muller C., Reinecke H., Biomed. Microdevices 2015, 17, 86. [DOI] [PubMed] [Google Scholar]
  • 27. Gimbel A. A., Flores E., Koo A., García‐Cardeña G., Borenstein J. T., Lab Chip 2016, 10, 1047. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 28. Thompson A. J., Marks L. H., Goudie M. J., Rojas‐Pena A., Handa H., Potkay J. A., Biomicrofluidics 2017, 11, 024113. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 29. Dabaghi M., Fusch G., Saraei N., Rochow N., Brash J. l., Fusch C., Selvaganapathy P. R., Biomicrofluidics 2018, 12, 044101. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30. Dabaghi M., Saraei N., Fusch G., Rochow N., Brash J. L., Fusch C., Selvaganapathy P. R., Lab Chip 2018, 18, 3780. [DOI] [PubMed] [Google Scholar]
  • 31. Matharoo H., Dabaghi M., Rochow N., Fusch G., Brash J., Fusch C., Selvaganapathy P. R., Biomicrofluidics 2018, 12, 014107. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 32. Thompson A. J., Ma L. J., Plegue T. J., Potkay J. A., IEEE Trans. Biomed. Eng. 2019, 66, 1082. [DOI] [PubMed] [Google Scholar]
  • 33. Dabaghi M., Saraei N., Fusch G., Rochow N., Brash J. L., Fusch C., Selvaganapathy P. R., J. Membr. Sci. 2020, 596, 117741. [Google Scholar]
  • 34. Dabaghi M., Saraei N., Fusch G., Rochow N., Brash J. L., Fusch C., Selvaganapathy P. R., Biomicrofluidics 2019, 13, 034116. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 35. Wu W. I., Sask K. N., Brash J. L., Selvaganapathy P. R., Lab Chip 2012, 12, 960. [DOI] [PubMed] [Google Scholar]
  • 36. Thorslund S., Sanchez J., Larsson R., Nikolajeff F., Bergquist J., Colloids Surf., B 2005, 45, 76. [DOI] [PubMed] [Google Scholar]
  • 37. Potkay J. a., Lab Chip 2014, 14, 4122. [DOI] [PubMed] [Google Scholar]
  • 38. Leung J. M., Berry L. R., Atkinson H. M., Cornelius R. M., Sandejas D., Rochow N., Selvaganapathy P. R., Fusch C., Chan A. K. C., Brash J. L., J. Mater. Chem. B 2015, 3, 6032. [DOI] [PubMed] [Google Scholar]
  • 39. Reoma J. L., Rojas A., Kim A. C., Khouri J. S., Boothman E., Brown K., Grotberg J., Cook K. E., Bartlett R. H., Hirschl R. B., Mychaliska G. B., J. Pediatr. Surg. 2009, 44, 53. [DOI] [PubMed] [Google Scholar]
  • 40. Gray B. W., El‐Sabagh A., Rojas‐Pena A., Kim A. C., Gadepali S., Koch K. L., Capizzani T. R., Bartlet R. H., Mychaliska G. B., ASAIO J. 2012, 58, 148. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 41. Jopling J., Henry E., Wiedmeier S. E., Christensen R. D., Pediatrics 2009, 123, e333. [DOI] [PubMed] [Google Scholar]
  • 42. Jones J. E., Jose P. A., Semin. Perinatol. 2004, 28, 141. [DOI] [PubMed] [Google Scholar]
  • 43. Bauer K., Uhrig C., Sperling P., Pasel K., Wieland C., Versmold H. T., J. Pediatr. 1997, 130, 240. [DOI] [PubMed] [Google Scholar]
  • 44. Bauer J., Maier K., Hellstern G., Linderkamp O., Br. J. Nutr. 2003, 89, 533. [DOI] [PubMed] [Google Scholar]
  • 45. Sylvester J. P. W. J. T., Shimoda L. A., Aaronson P. I., Physiol. Rev. 2012, 92, 367. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 46. Askie L. M., Darlow B. A., Finer N., Schmidt B., Stenson B., Tarnow‐Mordi W., Davis P. G., Carlo W. A., Brocklehurst P., Davies L. C., Das A., Rich W., Gantz M. G., Roberts R. S., Whyte R. K., Costantini L., Poets C., Asztalos E., Battin M., Halliday H. L., Marlow N., Tin W., King A., Juszczak E., Morley C. J., Doyle L. W., Gebski V., Hunter K. E., Simes R. J., JAMA, J. Am. Med. Assoc. 2018, 319, 2190. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 47. Solevåg A. L., Dannevig I., Nakstad B., Saugstad O. D., Neonatology 2010, 98, 64. [DOI] [PubMed] [Google Scholar]
  • 48. Baumgart S., Langman C. B., Sosulski R., Fox W. W., Polin R. A., Clin. Pediatr. (Philadelphia) 1982, 21, 199. [DOI] [PubMed] [Google Scholar]
  • 49. Leung J. M., Berry L. R., Chan A. K. C., Brash J. L., J. Biomater. Sci., Polym. Ed. 2014, 25, 786. [DOI] [PubMed] [Google Scholar]
  • 50. Ghaemi R., Dabaghi M., Attalla R., Shahid A., Hsu H. H., Selvaganapathy P. R., J. Micromech. Microeng. 2018, 28, 087001. [Google Scholar]
  • 51. Meyer dos Santos S., Zorn A., Guttenberg Z., Picard‐Willems B., Kläffling C., Nelson K., Klinkhardt U., Harder S., Biomicrofluidics 2013, 7, 056502. [DOI] [PMC free article] [PubMed] [Google Scholar]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supporting Information


Articles from Advanced Science are provided here courtesy of Wiley

RESOURCES