Abstract
The twin-to-twin transfusion syndrome is a severe fetal anomaly appearing in up to 15% of identical twin pregnancies. This anomaly occurs when twins share blood vessels from a common placenta. The complication leads to an unbalanced blood transfusion between both fetuses. A current surgical treatment consists in coagulating the shared vessels using a fetoscope with an embedded laser. Such treatment is very delicate and constraining due to limited vision and size of the insertion area. The rigidity and lack of controllability of the current used instruments add an additional difficulty and limit the choice in insertion site. This letter proposes an improved flexible fetoscope, offering an enhanced laser controllability and higher versatility regarding the location of the insertion site. A better approach angle can therefore be realized. Also, tissue damage may be further reduced. This single-handed controllable active fetoscope is obtained after adaptation of a LithoVue (Boston Scientific, Natick, MA, USA), a commercially available passive flexible ureteroscope. The LithoVue is fitted with a unique lightweight add-on actuation module foreseen of an artificial muscle and a dedicated control system. Experiments in a mixed reality trainer suggested that the proposed fetoscope is compact, ergonomic, and intuitive in use, allowing an adequate control of the flexible end.
Keywords: Medical robots and systems, flexible robots, mechanism design
I. Introduction
T HE twin-to-twin transfusion syndrome (TTTS) is a serious fetal complication occurring in monozygotic twins sharing a single placenta, but separate amniotic sacs. This fetal anomaly, which occurs in 10 to 15% of monochrionic twin pregnancies [1], results from a disproportional blood transfusion from one twin to the other. The donor fetus transfers blood to the recipient fetus, impeding the donor fetus’s proper development. If left untreated, severe TTTS can lead to a near-100% mortality rate. However, endoscopic laser ablation (ELA) of placental vessels allows a trend reversal and can guarantee the survival of at least one twin in 75 to 80% of the cases [1]. This minimally invasive surgery (MIS) consists in inserting a straight fetoscope equipped with a therapeutic laser fiber in the cannula through the maternal abdominal and uterine wall. As such, it provides access into the amniotic sac of the recipient twin. The fetoscope is used to visualize the blood vessels on the placenta surface. By using the embedded laser, the surgeon is able of coagulating the placental vessels responsible for the blood transfusion between twins, as shown in Fig. 1 [2].
Compared to other forms of surgery, the current surgical treatment of TTTS presents important challenges: 1) the vision is blurred due to the turbid nature of the amniotic fluid [3]; 2) the diameter of the incision at the single insertion site must be sufficiently small to avoid iatrogenic preterm premature rupture of membranes (IPPROM) [4]; 3) the instrument workspace is limited due to the considerable thickness of the abdominal wall. Any stress applied on the uterus and fetal membranes may cause premature delivery [5].
The complexity of the ELA is partially caused by the rigid nature of current instruments (rigid and semi-rigid scopes) and their poor controllability. The low dexterity of these instruments severely restricts the number of potential insertion sites. When operating anterior located placentas (Fig. 2b), surgeons are forced to approach the targets under obtuse angles, worsening the visualization and limiting the ablation precision. As a consequence, practitioners tend to apply large forces at the incision site that may lead to tissue damage or cause IPPROM [5].
As an attempt to solve the above-mentioned issues and limitations, several fetoscopes with embedded laser have been reported in the literature. For example, Yamanaka et al. [6] designed a rigid endoscope which allows steering of a laser beam by control of mirror orientation. However, this endoscope presents a limited steering angle and a large external diameter, making it impractical in case of anterior location of the placenta. This mirror approach does not provide a steering of the camera, leading to a fixed vision range. To overcome this limitation and offer larger laser steering angles while keeping the vision range on the laser target, some steerable flexible manipulators have been introduced [3], [7]–[9]. Although their respective workspaces are interesting for TTTS treatment, these instruments only foresee the insertion of a single tool (laser, camera or forceps). An extra instrument is therefore needed for illumination and/or visualization purposes, meaning that extra incisions need to be made. These extra cuts may result in additional complications. Moreover, an additional surgeon is needed for the manipulation of this extra tool. This letter proposes a steerable flexible instrument with large laser and vision steering angle range. This single instrument contains all the necessary tools to perform a TTTS procedure, namely a camera, an illumination fiber and a working channel, allowing the insertion of a coagulation laser while keeping similar external diameter dimensions than proposed fetoscopes in the literature. Moreover, this novel fetoscope has been designed to be comfortably single-handed-controlled, which limits the number of surgeons needed for the surgical procedure.
The layout of this letter is as follows: first, the detailed ELA of placental vessels is described in Section II. Section III addresses the design instrument specifications. An adjustment of an ureteroscope is proposed in Section IV followed by an identification and characterization of the proposed solution. In Section V, a protocol for user experiments is proposed and results are presented. Finally, Section VI concludes the proposed work.
II. Procedure Specifications
An ELA procedure of placental vessels is carried out as follows: initially, a detailed ultrasound inspection of the womb is performed in order to locate the placenta, the amniotic membrane separating the twins, the placental insertion of the umbilical cords and any extra fetal defects if such are present [10]. The most adequate entry site on the maternal abdomen is then chosen for cannula insertion, while avoiding any placental or fetus injuries. The cannula is preferably placed perpendicularly to the shared placenta, in such a way that the fetoscope with embedded laser can properly ablate the vessels. In this way, a good view of the targeted area is available, and the laser energy can be delivered in a focused region and not spread out due to the inclination. Usually, a neodymium-yttrium aluminum garnet (Nd-YAG) or a diode laser is used to ablate the vascular anastomoses [11].
In the womb, the placenta can be situated either on the posterior or on the anterior side. In the case of a posterior position, a straight rigid fetoscope is used (Fig. 2a) because a straight access path from the insertion site to the placenta is available. However, an anterior position of the placenta constraints the surgeon to resort to a semi-rigid curved fetoscope (Fig. 2b). In both cases, when inclining the instrument to ablate targeted vessels, stresses are applied on the uterine wall and on the amniotic sac during the ablation.
The use of a flexible instrument could reduce these stresses, but also provide a single instrument intended to be used in both cases of posterior and anterior positions of the placenta. The external diameter of the currently used fetoscopes varies between 2.3 and 4.0 mm. A new flexible instrument should adhere to these outer dimensions. Petersen et al. linked the increase diameter of the instrument with a higher likelihood of preterm birth [12]. Following this reasoning, the instrument diameter therefore needs to be as small as possible. Yet it must allow insertion of a camera, an illumination system and a therapeutic laser. The maximum bending angle of the flexible device has been fixed to 90°, allowing large freedom of movement for ablation of the selected vessels irrespective of the position of the placenta (Fig. 2) without exerting too much pressure on the uterus and fetal membranes. The bending should also be limited to one direction, as requested by fetal surgeons, in the hope that it would reduce the learning curve somewhat. The total instrument length ought to be somewhere between 188 mm to 306 mm, which is similar to fetoscopes available on the market nowadays [13].
The control method of the flexible tool aims to be singlehanded, meaning that the surgeon can simultaneously manipulate the fetoscope, and with his/her other hand support and adjust the cannula orientation. An intuitive interface would be required such that - without additional mental load - the surgeon could remain at any time in complete control of the instrument. Table I summarizes the main specifications the developed flexible instrument has to fulfill.
Table I. Specifications for an Active Fetoscope - Part 1.
Instrument features | Specifications |
---|---|
Rl: External diameter | 2.3 - 4.0 mm |
R2: Total length | 188 - 306 mm |
R3: Bending angle | 0° - 90°, one direction. |
R4: Embedded tools | Camera, illumination system, therapeutic laser. |
R5: Control type | Single-handed, intuitive operation. |
III. Proposed Design Approach
In order to keep the instrument approval process towards animal and human trials short, an add-on housing and actuation system on a commercially available ureteroscope the LithoVue (Boston Scientific, Natick, USA) (Fig. 3) is proposed. This single-use flexible ureteroscope is a robust medical-grade scope of 680 mm long. Its small external diameter (≤3.23 mm), its large working channel (1.2 mm ID) as well as its integrated light source and chip-on-tip sensor makes the LithoVue a possible candidate for use in fetal surgery, and especially for TTTS treatment. Moreover, the deflection of its distal flexible tip is sufficiently large (max 270° in both directions). The tip is oriented by pushing or pulling a lever situated at the extremity of the ureteroscope handle.
However, the unaltered use of the LithoVue is no option. The flexible nature of the proximal part (part b on Fig. 3) could lead to imprecise ablations and its length is not adapted for fetal intrauterine procedures either. Finally, the current position of the lever does not provide an intuitive and ergonomic operation of the flexible tip (part a on Fig. 3).
A non-ergonomic handle can lead to surgeon’s muscle fatigue, causing their performances to be less efficient [15]. It also prevents him/her from holding and orienting the cannula with his/her other hand, requiring an extra surgeon to perform the task. In order to overcome these issues, this work proposes to turn the LithoVue into an actively controlled system, whereby the control interface is directly integrated in the handle. As the LithoVue itself is a disposable instrument, a drive system integrated in an external housing is foreseen such that a new LithoVue can be easily clicked in and out of the frame. Requirements to which the add-on should satisfy are listed in Table II.
Table II. Specifications for an Active Fetoscope - Part II.
Additional specifications (Add-on system) | |
---|---|
R6: | Modular setup with easy exchange of disposable Litho Vue |
R7: | Enabling intuitive, single-handed operation |
R8: | Lightweight |
R9: | Compact, allowing strong and comfortable handling |
R10: | Precise control of the distal flexible part |
R11: | Easy assembly and disassembly |
In the literature, several works have described systems to drive otherwise manually operated flexible scopes. Ruiter et al. [16] and Fang et al. [17] proposed for example enhanced flexible scopes by fixing motors and couplings on the original instrument, or by transmitting movements using Bowden cables. However, these add-on devices are not compatible with several of the above-mentioned requirements (R6, R8, R9). This letter proposes a compact and lightweight add-on system respecting all the mentioned requirements (R1-R11). The surgeon can orient and insert/retract the scope as before, but can additionally access the distal bending degree of freedom which is controlled by the added drive system.
A. LithoVue Modifications
The Litho Vue is a cable-driven device composed of two antagonistic cables routed along the instrument shaft. At a proximal end, the cables are attached to a driving pulley/handle whereas distally, they are connected to two opposite ends of the distal tip. By pushing/pulling the lever, one cable is tensioned and the other released, allowing the distal flexible tip to bend in one or the other direction. As the surgeons want to limit the instrument to unidirectional motion (R3), one cable has been loosened, making the instrument bend from 0° to 135° in only one direction. Note that this does not restrict the reachable range of orientations as the surgeon can always decide to rotate the instrument about its longitudinal axis. A spring has been foreseen inside the LithoVue ensuring the return to 0° when no load is applied on the cable. In order to meet the requirements of fetal surgery, the non-rigid proximal part has been shortened and rigidified by sliding a rigid tube over that part. After these modifications, the needed maximum force and stroke to move the lever was measured. The force ensuring the return of the flexible tip to a straight configuration (i.e. to 0°) from a position of at least 90° has been estimated to be 0.7 N. The distal tip bends over 90 degrees when adjusting the lever over an angle of 1.2 rad. The spring was chosen to meet these requirements.
B. Add-on Frame and Actuation System
In order to satisfy the requirement in terms of size (compactness), an in-house built artificial muscle has been chosen as the actuator. This actuator is a McKibben muscle. Composed of an expandable bladder surrounded by a meshed structure, the McKibben muscle forms a thin and extremely light weight linear actuator. Once pressurized, the bladder will expand, resulting in an increase of its diameter. The individual strands of the braided mesh stay more or less constant in length. As the bladder and mesh expand in diameter, the muscle thus contracts. McKibben muscles generate relatively large output forces and displacements even for low compact diameter muscles. This actuator can also achieve good positioning accuracy. Moreover, and due to their softness and compliance, McKibben muscles are excellent candidates for instruments used in surgery [18]. A flexible air supply tube is the only connection to a distally positioned proportional valve. The weight of the muscle and the supply tube is negligible. The McKibben muscle is attached at one end to the LithoVue lever. The other end is attached to the distal part of the handle as depicted in Fig. 4. The muscle is therefore placed externally to the patient’s body in order to guarantee the safety of the instrument. Any air leakage that could appear will be situated outside the body. In correspondence to R3, the lever moves 12 mm when the muscle contracts. It was verified that for this, the muscle needs to provide at least 4 N contraction force. A compact housing has been added (Fig. 4) in order to protect the pneumatic actuator. A pair of low-profile buttons were selected and embedded in the external housing. The location was chosen such that it would be easy for the operator to manipulate the scope with one hand and at the same time access the buttons to control the distal tip. By pushing one button or the other, the surgeon commands the pressure inside the muscle. As a result, the position of the lever is adjusted, which through the cable thus controls the bending angle of the distal tip.
IV. Identification & Characterization
The prototype is composed of a helical twisting spring of stiffness 34.2 N.mm/rad. A McKibben muscle of 2mm diameter (in relaxed state) and a length of 160 mm is mounted on the LithoVue. The adapted LithoVue with external housing weighs 255 g (whereas the original one weighs 304 g) and has a total length of 550 mm. The handle is 250 mm and the proximal and distal part measure together 300 mm. A measurement setup has been developed in order to characterize the device.
A. Test-Bench Presentation
Fig. 5a depicts the measurement setup that has been developed to characterize the active LithoVue fetoscope. A computer is used to control the air pressure, the motion of the LithoVue and process the grabbed data. To ensure the control of the elecotrovalve ITV0050-3MN-Q (SMC, Tokyo, Japan), OROCOS, a middleware for real time robot control was used. The proportional valve was controlled via an EtherCat module NI9144 (National Instrument, Texas, USA) and AO DAQ NI9263 (National Instrument, Texas, USA), and supplied by a Cyclon 215 air compressor (CompAir, Wisconsin, USA) that can provide a pressure of maximum 1 MPa. The output of the pneumatic valve is directly linked to the McKibben muscle inside the adapted device. The proportional valve has a response time of 0.1 s when no load is applied, and is supplied with a pressure of 0.6 MPa. For that specific supply pressure, the valve can provide a maximum flow rate of 6 l/min. The corresponding bending angle of the flexible tip is computed from tracking four white markers that were attached at different locations of the distal instrument tip (Fig. 5b). Two markers were placed on the rigid shaft, and two on the extreme tip of the flexible part to determine an axis aligned with the distal tip. The angle formed by these two axes is defined as the bending angle of the instrument (Fig. 5b). The Matlab (The Mathworks, Massachussets, US) function imfind-circles has been used to trace the round marker. The algorithm was tested on a mechanical micrometer device, and could predict a marker position with a 0.09 mm accuracy.
B. System Identification
The characterization of the active LithoVue fetoscope has been conducted by applying 12 pressure cycles. Each cycle has a minimum pressure of 0 MPa and a maximum of 0.5 MPa, which is a safe upper limit for operating this type of muscle. The cycle period was set to 240 s. The pressure has been applied incrementally in order to allow the flexible tip to stabilize. Each increment has an amplitude of 6.25 kPa and a duration of 1.5 s. Several pressurization-depressurization cycles have been performed, and the corresponding bending angle of the instrument has been calculated using the above-mentioned method (Fig. 5b). The relation between the input pressure and the bending angle of the instrument over 12 cycles is depicted on Fig. 6.
A large hysteretic effect is observed whereas the repeatability of this behavior upon full excitation is fairly good. The irregularities perceived on the load curve are due to friction induced by the cable actuation inside the LithoVue. The static friction is also expected to be the reason why the flexible tip never fully returns to a straight 180° position at rest (Fig. 6). From these data, the coordinates of plateaus extremities (constant angle over a range of pressure) can be calculated. These points, called pre-pressurization, represent the pressure from which the system will start bending. Pre-pressurizations for the loading and unloading of the instrument equals respectively at 140 kPa and 175 kPa. The characterization has been conducted with an empty-working channel. Usually, a Nd-YAG or diode laser is used to perform the coagulation. However, these fibers are stiff and not adapted for use in flexible instruments. A study showed that Holmium YAG (Ho-YAG) laser may be used for coagulation [19]. These fibers are very thin and flexible, and their insertion into the working channel has no impact on the instrument output angle.
C. Static Study
The precision (positioning repeatability) of the instrument was evaluated from the 12 actuation cycles and found to be 3.6°. In order to obtain the accuracy of the instrument, a model curve has been plotted and compared to the previous 12 cycles data. This model is composed of the mean angle values taken on 3 pressure loading-unloading cycles (Fig. 6). The accuracy has been evaluated to be in the order of 1.7° with this method. Finally, the drift caused by a long-period loading of the instrument has been measured. The instrument has been pressurized during 20 min., and the position difference between the initial and final position has been found to be −0.9°. This angle decrease under a long period of loading is probably due to fluctuations of the supply pressure.
D. Dynamic Study
In order to evaluate the dynamic performance of the instrument, the maximum time constant in response to an input pressure step has been measured. Fig. 7 depicts the behavior of the flexible tip angle when a step of 0 to 0.5 MPa (the maximum pressure the instrument can handle) is applied. The pressure in the system has also been sensed using the PA-21Y/25BAR pressure sensor (Keller, Winterthur, Switzerland). The right part of the Fig. 7 shows the case of a step of 0.5 MPa to 0 Pa. Throughout the loading of the instrument (pressurization from 0 to 0.5 MPa), the angle response presents a delay of 0.54 s and a time constant of 0.95 s. This delay is mainly due to the switching delay of the used valve as seen on the sensed pressure curve (Fig. 7). Throughout the unloading (depressurization from 0.5 MPa to 0 Pa), the response presents a delay of 0.33 s and a time constant of 0.73 s.
By pressing a button, the surgeon is able to control the bending angle of the flexible tip. At any time, the surgeon will remain responsible for the main device motions. Only the bending of the flexible tip will be controlled via the add-on McKibben muscle. In order to keep the setup simple in the following, it is explored in how far an open-loop feed-forward control can be set to “linearize” the behavior and compensates the hysteresis. By doing so, the pre-pressurization and the non-linear relation between the input pressure and the bending angle can be made to some extent invisible to the surgeon. The open loop control of the actuator is sketched in the block diagram below:
With Kp, the scale factor of the buttons and G(s,a), the transfer function of the instrument. Using the data of dynamic test (Fig. 7), the device behavior can be modeled as a first order system. G(s, a) that takes the following forms depending on the system state; loading (G1 (s, α)) or unloading (G2(s, α)):
(1) |
In (1), τ1 and τ2 are time constants that have been previously calculated and are respectively equal to 0.54 s and 0.73 s. θ-1 and θ2, the time delays, have also been previously calculated for a step of 0.5 MPa amplitude, and are respectively equal to 0.95 s and 0.33 s. F1(α) and F2 (α) are related to the characteristic curve linking the input pressure and the resulting bending angle (Fig. 6). F1(α) and F2 (α) can be modeled as a first order polynomial. The polynomial model has been obtained using the polyfit Matlab function. The characteristic function can therefore be expressed as the following:
In order to check the correct modeling of the instrument, the instrument response to a sine setpoint has been measured under open loop control. Several sinus periods have been tested: 3 s, 5 s and 7 s (Fig. 8). It can be noticed in Fig. 8 that the instrument cannot follow the full input signal for a period of 3 s. For a 5 s period, the expected curve is fairly accurately followed. However, a maximum delay of 1.06 s can be noticed between both curves. The delay upon loading is smaller than the one for unloading in accordance to the results of the step test (Fig. 7). For a period bigger than 5 s, the measured curve presents in addition an overshoot with respect to the expected curve. This can be explained by the fact that the characterization of the instrument has been realized statically. Static friction limits the amplitude of the bending angle in static experiments. Nonetheless, this friction is overcome when the instrument is tested dynamically. Table III summarizes the main features of the active LithoVue fetoscope obtained from the identification and characterization experiments. The means and standard deviations for the accuracy and precision have been taken over 12 cycles. 3 trials have been executed to determine the characteristics of the drift and the dynamic response.
Table III. Adapted Lithovue Features.
Tested features | Characteristic |
---|---|
Identification | |
pre-pressure(pressurisation) | 140kPa |
Static study | |
Accuracy | 0.5°,0.5° |
Dynamic Study | |
τ(loading) | 0.95 s,0.1 s |
However, an important problem resides in the unloading curve. Dynamically, the slope of the measured angle (Fig. 8) is too abrupt. In less than 1 s, the instrument tip relaxes about 60°. The dynamic behavior of the instrument being totally different from the static one complicates a proper open-loop control based on the static model. In order for the surgeon to be able to correctly control the device, two solutions are proposed. First, it would be possible to close the loop by integrating two sensors in the instrument. One at the very end of the flexible tip, and another one in the straight rigid part. However, little space is available inside the flexible tip to place these sensors. On the other hand, a fixation outside the instrument would increase the external diameter. Another viable solution would be to characterize the hysteresis in greater detail and use these improved models in open-loop feed-forward.
A human-in-the-loop experiment was designed using the active LithoVue fetoscope once with the implemented open-loop control, and once with no control at all. The user was asked to orient the flexible tip on several specific angles using the buttons in both operation modes. Beginning at rest position (156°), the flexible tip was then driven to successively reach 90°, 135°, 110° and finally 156°. Once a target is reached, the user was asked to maintain this position for a brief period of time so that it could be recorded. The Kp coefficient was set to 0.007° for both types of control, and a coefficient of 2500 Pa/° was used to convert the angle in an output pressure in the case where the open-loop control was not used. Fig. 9 depicts the results of the experiment. The benefits of the open-loop control are clearly visible. The response is faster when using the implemented control. Also, higher levels of precision are reached. However, switching from 135° to 110° remains an issue since the control is based on a static model. The hysteresis, when driving the tip from 135° to 110° is nevertheless smaller than the one obtained in the static study (Fig. 6).
V. User Experiments
The prototype has been evaluated by two experienced surgeons, a gynecologist and a pediatric surgeon, both with experience in fetal surgery. They were asked to compare two instruments : 1) an original LithoVue of which the proximal part was shortened and made stiffer by adding a rigid shaft and 2) the proposed active LithoVue fetoscope. The surgeons were asked to test both instruments in virtual reality, using a validated simulator [20]. Then, they were asked to fill in a questionnaire to evaluate the ergonomics and the ease of manipulation of these instruments.
A. Detailed Method
A mixed-reality simulator was used to simulate a realistic TTTS treatment [20]. The trainer is composed of physical and virtual components. A 4 cm thick-synthetic body wall phantom renders the interaction between the instrument and the patient’s thick body wall. This body wall provides the surgeon a realistic haptic feeling, generating a natural resistance upon movement of the fetoscope. The inside of the womb is rendered by the means of a VR system. The placenta and the laser coagulation of the tissue is virtually simulated. By pressing a pedal the surgeon can activate the laser. Fig. 10 provides a view on the virtual TTTS surgery. A cannula is first inserted through the body wall phantom of the simulator box to pass the instrument through. An electromagnetic (EM) motion tracking system is installed at the base of the virtual reality setup to track the instrument in real time (Aurora-NDI Medical, Cleveland, USA). A 6 DOF EM sensor is placed in the working channel of the instrument at the distal flexible tip. When the buttons are pressed, the tip of the instrument moves. The tip pose is then measured and sent to the virtual reality system, which renders images of the environment viewed from the instrument tip.
Both surgeons are asked each in turn to perform a laser ablation of the placental anastomoses, typical for TTTS. In the first part of the procedure, the surgeon has to laser selectively some preselected anastomoses seen as targets. He/she then connects these dots by a continuous ablation line (requiring continuous lasering), which represents the second part of the procedure. The surgeon was asked to be as precise as possible. Both tasks are repeated three times by the surgeon. The first time, the task is conducted on a posterior placenta. This part is intended to get the surgeon familiar with the instrument and the simulator. Only the second and third time, respectively conducted on a posterior and anterior placenta are taken into account for the instrument evaluation. After completion of the tasks, the user is asked to fill out a questionnaire to evaluate the perceived weight, compactness and the comfort of the instrument, but also the task difficulty on a 5-point Likert scale.
B. Results
Especially for the anterior positioning of the placenta, which makes the TTTS procedure significantly more complex, both surgeons indicated to make less use of the cannula to orient the laser. They therefore could limit the forces on the body wall. The answers to the questionnaire are summarized in Fig. 11. A score of 5 means the instrument has been evaluated as very comfortable, very lightweight and very compact. It also means that the TTTS task (either with a posterior or anterior positioning of the placenta) has been judged as very easy with the corresponding instruments. The actuation comfort, on the other hand, refers to the comfort felt during actuation of the instrument. A score of 1 means that the instrument is absolutely not comfortable, lightweight and compact, and that the performed tasks were found very difficult. The handling comfort refers to the comfort felt when simply handling the instrument without actuating it.
While only a preliminary validation has been conducted, which prevents from making strong conclusions, one can already observe a trend that the new instrument enhances the manipulation comfort and ease of treating TTTS. The overall weight was still acceptable since the comfort scored well. This tendency shows that even though the instrument control is not fully optimized, the single-handed active button-controlled fetoscope is already preferred for TTTS treatment.
VI. Conclusions & Future Work
An improved flexible fetoscope tailored for TTTS treatment was presented in this Letter. This device, being an adaptation of the passive flexible ureteroscope LithoVue (Boston Scientific, Natick, USA), has shown a good potential for use in TTTS surgery. Tested by two expert surgeons during virtual TTTS treatment, the device could limit variation of forces on the body wall by allowing the surgeon to reduce the use of the cannula to orient the laser. Compared to a passive, lever-actuated fetoscope, the active LithoVue fetoscope was evaluated as being more comfortable and allowing an easier accomplishment of typical TTTS treatment tasks. However, non-ideal behavior such as friction and hysteresis in the drive system negatively affected the steerability. Whereas the users managed to overcome these effects, by modeling and employing this knowledge in a feedforward control scheme, the mental burden of the surgeons may further decrease. This could lead to improved user satisfaction. Whereas current experiments were evaluated qualitatively, a more thorough quantitative evaluation remains for further work.
Acknowledgements
The authors would like to thank clinicians L. Joyeux and A. Inversetti for helping evaluating the fetoscopes.
Contributor Information
Julie Legrand, J. Legrand, M. Ourak, A. Javaux, C. Gruijthuijsen, M. A. Ahmad, and E. Vander Poorten are with the Laboratory of Robot-Assisted Surgery, Department of Mechanical Engineering, KU Leuven, Leuven 3000, Belgium.
Mouloud Ourak, Email: mouloud.ourak@kuleuven.be.
Allan Javaux, Email: allan.javaux@ kuleuven.be.
Caspar Gruijthuijsen, Email: caspar.gruijthuijsen@kuleuven.be.
Mirza Awais Ahmad, Email: mirzaawais.ahmad@ kuleuven.be.
Ben Van Cleynenbreugel, Email: Ben.Vancleynenbreugel@uz.kuleuven.ac.be, B. Van Cleynenbreugel is with the Organ System Unit, KU Leuven, Leuven 3000, Belgium.
Tom Vercauteren, Email: tom.vercauteren@kcl.ac.uk, T. Vercauteren is with the Department of Imaging and Biomedical Engineering, King’s College London, London WC2R 2LS, U.K., and also with the Department of Development and Regeneration, KU Leuven, Leuven 3000, Belgium.
Jan Deprest, Email: Jan.Deprest@uz.kuleuven.ac.be, J. Deprest is with the Department of Development and Regeneration, KU Leuven, Leuven 3000, Belgium.
Sebastien Ourselin, Email: sebastien. ourselin@kcl.ac.uk, Sebastien Ourselin is with the Department of Imaging and Biomedical Engineering, King’s College London, London WC2R 2LS, U.K.
Emmanuel Vander Poorten, Email: Emmanuel.VanderPoorten@kuleuven.be.
References
- [1].Muratore CS, et al. Survival after laser surgery for twin-to-twin transfusion syndrome: When are they out of the woods? J Pediatric Surg. 2009 Jan;44(1):66–69. doi: 10.1016/j.jpedsurg.2008.10.011. [DOI] [PubMed] [Google Scholar]
- [2].Guidance and guidelines. NICE; 2006. Dec, [Accessed on: Jan. 2018]. Intrauterine laser ablation of placental vessels for the treatment of twin-to-twin transfusion syndrome. [Online]. Available: https://www.nice.org.uk/guidance/ipg198. [Google Scholar]
- [3].Harada K, Bo Z, Enosawa S, Chiba T, Fujie MG. Bending laser manipulator for intrauterine surgery and viscoelastic model of fetal rat tissue; Proc Int Conf Robot Automat; 2007. Apr, pp. 611–616. [Google Scholar]
- [4].Devlieger R, et al. Fetal membrane healing after spontaneous and iatrogenic membrane rupture: A review of current evidence. Amer J Obstetrics Gynecol. 2006 Dec;195(6):1512–1520. doi: 10.1016/j.ajog.2006.01.074. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Maradny EE, Kanayama N, Halim A, Maehara K, Terao T. Stretching of fetal membranes increases the concentration of interleukin-8 and collagenase activity. Amer J Obstetrics Gynecol. 1996 Mar;174(3):843–849. doi: 10.1016/s0002-9378(96)70311-3. [DOI] [PubMed] [Google Scholar]
- [6].Yamanaka N, Yamashita H, Masamune K, Chiba T, Dohi T. An endoscope with 2 DOFs steering of coaxial Nd:YAG laser beam for fetal surgery. IEEE/ASME Trans Mechatronics. 2010 Dec;15(6):898–905. [Google Scholar]
- [7].Yamashitaa H, et al. Two-DOFs bending forceps manipulator of 3.5-mm diameter for intrauterine fetus surgery: Feasibility evaluation. Int J Comput Assisted Radiol Surg. 2006 Jun;1:218–220. [Google Scholar]
- [8].Zhang B, Kobayashi Y, Maeda Y, Chiba T, Fujie MG. Development of 6-DOF wire-driven robotic manipulator for minimally invasive fetal surgery; Proc IEEE Int Conf Robot Automat; 2011. May, pp. 2892–2897. [Google Scholar]
- [9].Yao W, Elangovan H, Nicolaides K. Design of a flexible fetoscopy manipulation system for congenital diaphragmatic hernia. Med Eng Phys. 2014 Jan;36(1):32–38. doi: 10.1016/j.medengphy.2013.08.014. [DOI] [PubMed] [Google Scholar]
- [10].Ville Y, Hecher K, Gagnon A, Sebire N, Hyett J, Nicolaides K. Endoscopic laser coagulation in the management of severe twin-to-twin transfusion syndrome. Brit J Obstetrics Gynaecol. 1998 Apr;105(4):446–453. doi: 10.1111/j.1471-0528.1998.tb10132.x. [DOI] [PubMed] [Google Scholar]
- [11].Papanna R, et al. Risk factors associated with preterm delivery after fetoscopic laser surgery for twin twin transfusion syndrome. Ultrasound Obstetrics Gynecol. 2014 Jan;43(1):48–53. doi: 10.1002/uog.13206. [DOI] [PMC free article] [PubMed] [Google Scholar]
- [12].Petersen SG, et al. The impact of entry technique and access diameter on prelabour rupture of membranes following primary fetoscopic laser treatment for twin-twin transfusion syndrome. Fetal Diagnosis Therapy. 2016;40(2):100–109. doi: 10.1159/000441915. [DOI] [PubMed] [Google Scholar]
- [13].Deprest J, et al. Endoscopy in Fetal Medicine. 3rd. Endo-Press; Tuttlingen, Germany: 2015. The eurofoetus and eurotwin2twin projects. [Google Scholar]
- [14].LithoVue single-use digital flexible ureteroscope. Boston Scientific; Marlborough, MA, USA: [Accessed on: Jan. 2018]. [Online]. Available: http://www.bostonscientific.com/en-US/products/Ureteroscopes/LithoVue.html. [Google Scholar]
- [15].Emam TA, Frank TG, Hanna GB, Cuschieri A. Influence of handle design on the surgeon’s upper limb movements, muscle recruitment, and fatigue during endoscopic suturing. Surg Endoscopy. 2001 Jul;15(7):667–672. doi: 10.1007/s004640080141. [DOI] [PubMed] [Google Scholar]
- [16].Ruiter J, Rozeboom E, van der Voort M, Bonnema M, Broeders I. Design and evaluation of robotic steering of a flexible endoscope; Proc 4th IEEE RAS EMBS Int Conf Biomed Robot Biomechatronics; 2012. Jun, pp. 761–767. [Google Scholar]
- [17].Fang C, Cesmeci D, Gumprecht JDJ, Krause E, Strauss G, Lueth TC. A motorized hand-held flexible rhino endoscope in ENT diagnoses and its clinical experiences; Proc 4th IEEE RAS EMBS Int Conf Biomed Robot Biomechatronics; 2012. Jun, pp. 853–858. [Google Scholar]
- [18].De Volder M, Moers AJM, Reynaerts D. Fabrication and control of miniature McKibben actuators. Sens Actuators A, Phys. 2011 Mar;166(1):111–116. [Google Scholar]
- [19].Brenner M, et al. Comparison of Ho:YAG versus Nd:YAG thoracoscopic laser treatment of pulmonary bullae in a rabbit model. J Clin Laser Med Surg. 1997;15(3):103–108. doi: 10.1089/clm.1997.15.103. [DOI] [PubMed] [Google Scholar]
- [20].Javaux A, et al. A mixed-reality surgical trainer with comprehensive sensing for fetal laser minimally invasive surgery. Int J Comput Assisted Radiol Surg. 2018 doi: 10.1007/s11548-018-1822-7. [DOI] [PMC free article] [PubMed] [Google Scholar]