Abstract
Purpose
To address the challenges posed by fat-water chemical shift artifacts and relaxation rate discrepancies to quantitative susceptibility mapping (QSM) outside the brain, and to generate accurate susceptibility maps of the head-and-neck at 3 and 7 Tesla.
Methods
Simultaneous Multiple Resonance Frequency (SMURF) imaging was extended to 7 Tesla and used to acquire head-and-neck gradient echo images at both 3 and 7 Tesla. Separated fat and water images were corrected for Type 1 (displacement) and Type 2 (phase discrepancy) chemical shift artefacts, and for the bias resulting from differences in T1 and relaxation rates, recombined and used as the basis for QSM. A novel phase signal-based masking approach was used to generate head-and-neck masks.
Results
SMURF generated well-separated fat and water images of the head-and-neck. Corrections for chemical shift artefacts and relaxation rate differences removed overestimation of the susceptibility values, blurring in the susceptibility maps, and the disproportionate influence of fat in mixed voxels. The resulting susceptibility maps showed high correspondence between the paramagnetic areas and the locations of fatty tissues and the susceptibility estimates were similar to literature values. The proposed masking approach was shown to provide a simple means of generating head-and-neck masks.
Conclusion
Corrections for Type 1 and Type 2 chemical shift artefacts and for fat-water relaxation rate differences, mainly in T1, were shown to be required for accurate susceptibility mapping of fatty-body regions. SMURF made it possible to apply these corrections and generate high-quality susceptibility maps of the entire head-and-neck at both 3 and 7 Tesla.
Keywords: chemical shift, fat-water, head-and-neck, QSM, SMURF
1. Introduction
Quantitative susceptibility mapping (QSM) is a method for calculating the distribution of the magnetic susceptibilities of tissue from a fieldmap (ΔB 0)—the difference between the externally applied magnetic field and the demagnetisation field generated by the tissue itself. 1 The phase of the gradient-echo (GRE) signal is used to generate the fieldmap, which is converted to a map of susceptibility by removal of background fields and dipole inversion. Generation of an accurate fieldmap is a crucial step in the process, as errors generally propagate throughout the QSM analysis pipeline and bias susceptibility estimates. 2
QSM provides excellent contrast in the imaging of calcifications, hemorrhages, and iron depositions, as well as tissue microstructure and oxygenation. 3 The technique has found application including the assessment of Alzheimer and Parkinson disease, Multiple Sclerosis (MS), the grading of tumors and monitoring of their growth, 4–6 and the past decade has seen an increase of interest in QSM outside the brain: in the liver, 7–11 breast, 3,12,13 kidney, 14 prostate, 15–17 heart, 18–20 knee, 21 trabecular bone, 22 spine, 23 and the entire head-and-neck region. 24–29 Compared to brain-only QSMs, susceptibility mapping of the head-and-neck allows the visualization of deep gray matter nuclei in the brainstem 30 —potential targets for deep brain stimulation 31,32 —MS lesions 33 in the brainstem, and visualization and assessment of tumors in the facial skull; epi-, naso-, and oropharynx; the base of the tongue and mouth; salivary glands; and possible regional lymph nodes. 34 The presence of fat in these regions, however, generally results in errors in field estimation that originate in several effects. First, because of the circa 3.4 ppm 35 chemical shift difference between fat and water, the fat image is shifted relative to water along the frequency-encoding direction by a number of voxels given by
where rBW/pixel is the receiver bandwidth per pixel and Δf the fat-water chemical shift. This so-called Type 1 (displacement) chemical shift artifact (CSA) leads, in addition to the errors in the spatial distribution of susceptibility sources, also to areas of signal voids, making spatial phase-unwrapping problematic. While QSM generally benefits from the use of higher field strengths, allowing either shorter acquisition times or higher resolution, 36 the increased fat-water chemical shift difference leads to a more pronounced Type 1 CSA. Second, the different precession frequency of fat with respect to water gives rise to an echo time (TE) -dependent phase component (φ) in GRE, defined as:
which does not reflect the tissue susceptibility and also causes destructive interference between water and fat signals—the so-called Type 2 (phase discrepancy) CSA. Last, differences in T1 and relaxation times lead the acquired water and fat signals to be weighted differently, according to the signal equation:
with PD being the proton density and a the flip angle (FA). The proton density ratio of the contributing water-based and fat-based tissues is biased by this signal weighting, which skews field estimates in mixed voxels.
To reduce the effects of Type 1 CSA, high receiver bandwidths are conventionally used, allowing only a partial correction and coming at the price of decreased SNR. To reduce Type 2 CSA, an acquisition at the in-phase TEs can be used, 12,23,24,26 although this restricts the choices of TEs. The effects of relaxation time differences are known to bias the estimation of proton density fat-fraction (PDFF), 37,38 but are generally neglected in QSM. 12,23,24,28 To minimize the bias caused by T1 relaxation rate differences, small FAs have to be used, resulting in poor SNR, or a correction based on known T1 values 38 can be applied. In the case of PDFF mapping of the iron overloaded liver, the differences in are assumed to be negligible 39 due to the shortening of the values being dominated by the presence of iron. It remains to be established if this assumption holds in other cases or if a correction should be applied.
As an alternative to in-phase imaging, the Dixon approach 40 can be used, allowing the simultaneous generation of water-only and fat-only images from a multi-echo acquisition. Unlike in-phase imaging, Dixon allows PDFF maps to be generated simultaneously and the multi-peak fat spectrum to be considered. The Dixon approach is, however, problematic in voxels with fat-fractions close to 50%, 23 which may occur either in mixed tissues such as the parotid glands or, due to partial volume effects, at the edges of water-based and fat-based structures. The Dixon approach also requires a very short echo spacing to achieve robust fat-water separation, 41,42 which is suboptimal for QSM, where the maximal contrast-to-noise ratio (CNR) is achieved at the TE close to the of the tissue. 43 To overcome this problem and also the problem over-smoothing of the fieldmaps inherent to most of the Dixon approaches, two methods have been developed specifically for Dixon-based susceptibility mapping. The first method, simultaneous phase unwrapping and removal of chemical shift (SPURS), 9 simultaneously unwraps the phase and removes chemical shift using single species fitting combined with graph cuts, while the second technique uses two trains of echoes with short and long echo spacings. 44 However, none of the Dixon methods addresses the errors in susceptibility estimates resulting from Type 1 (displacement) CSA. In magnitude Dixon imaging, the separated water and fat images can be corrected for their relative displacement, allowing the generation of chemical shift displacement-free recombined fat-water images. The fieldmap, however, is generally generated as a by-product of the minimization step and, because Type 1 CSA affects the acquired multi-echo signal that is being fitted, it is present in the fieldmaps, to the detriment of QSM. The same is true for the in-phase imaging approach.
We have recently presented a Simultaneous Multiple Resonance Frequency (SMURF 45 ) approach to fat-water imaging based on multi-band principles, which was applied to correct the CSAs in GRE and turbo spin-echo magnitude imaging at 3 T. In SMURF, multi-band pulses are applied together with CAIPIRINHA to allow separate images of fat and water to be generated. Prior to recombination, a complex summation of the fat and water signals, those signals can be corrected for Type 1 and Type 2 CSAs, and, if the relaxation times are known, for the bias resulting from fat-water relaxation rate differences. We hypothesized that the SMURF technique would yield similarly cleanly separated fat and water images at higher field strengths, as both field inhomogeneity and chemical shift difference generally scale with the field. Although a single-peak fat signal model is assumed in generating the spectrally selective radiofrequency (RF) pulses in SMURF, it has been shown that this approximation does not lead to significant bias in QSM outside the brain. 23 Susceptibility values of fatty tissues vary, however, quite widely in the literature, 23 e.g., 0.29 ppm 7 and 0.57 ppm 9 for the fat in the liver, 0.19 ppm 21 for the fat in the knee, and 0.29 ppm 24 for the fatty fascia in the head-and-neck. Nevertheless, in all cases was the fat assessed as being more paramagnetic than the surrounding water-based tissues.
In addition to the complications caused by the presence of fat, QSM outside the brain is also made problematic by the difficulties in generating the tissue masks required by most QSM approaches. Unlike for the brain, there are no standard segmentation tools, and simple magnitude thresholding generally includes some voxels with unreliable phase, giving rise to artifacts in the susceptibility maps, and potentially excludes extended areas of interest where coil sensitivities are low (e.g., in the neck at brainstem at a field strength of 7 T and above). Even these low signal areas often show high coherence of the phase signal, however. This, and the fact that magnitude-based masks include voxels with high magnitude signal but unreliable phase (e.g., veins) have led to the suggestion to generate masks using phase image. 46
The aim of this study was to address the challenges of QSM outside the brain and, using simultaneous fat-water imaging with SMURF, to generate chemical shift and relaxation rate bias-free, high CNR susceptibility maps of the entire head-and-neck region. Since QSM generally benefits from the use of higher field, we extended SMURF to 7 T, although the need for different RF excitation pulses and the modified field variations this a non-trivial translation. Additionally, we aimed to develop a simple but robust approach for generating head-and-neck masks using the signal phase.
2. Methods
Measurements were performed with a 3T Siemens Prisma scanner (syngo MR VE11C, Siemens Healthineers, Erlangen, Germany) and a 7T Siemens MAGNETOM Plus scanner (syngo MR VE12U AP01). The head-and-neck region of 10 healthy volunteers was scanned—five volunteers at 3 T (V1–V5) using a 64-channel Siemens head-and-neck coil and five at 7 T (V6-V10) using a 32-channel head coil (Nova Medical, Wilmington, Massachusetts, USA). The 3T and 7T data were acquired to evaluate the performance of SMURF imaging at each field strength independently and to assess the effectiveness of the corrections for chemical shift and relaxation rate biases at the most commonly used field strengths in QSM, but not with the intention of comparing results between the two field strengths. Written informed consent was provided by all the participants, and the study was approved by the Ethics Committee of the Medical University of Vienna.
2.1. RF pulse design
Least-squares-filtered minimum-phase Shinnar-Le Roux pulses 47–49 were designed using Vespa 50,51 and used to create both bands of two dual-band pulses: one optimized for fat-water imaging at 3 T 45 and one for 7 T. 29 The larger chemical shift difference at 7 T allowed the use of broader and, hence, shorter minimum-phase SLR RF pulses with a duration of 8.0 ms compared to 11.8 ms at 3 T, allowing a first echo time of about 2.5–4.5 ms at 7 T compared to about 4.5–7.0 ms at 3 T.
2.2. Acquisition parameters
2.2.1. 3T measurements
Sagittal 3D multi-echo GRE SMURF images were acquired with anterior-posterior phase-encoding direction, field of view (FOV) = 230 × 230 mm, resolution = 1.3 × 1.3 × 1.3 mm, 128 slices, TE = {6.5, 13.0, 19.5, 26.0, 32.5} ms, repetition time (TR) = 45 ms, FA(water) = 15°, FA(fat) = 27°—the respective Ernst angles assuming T1 values of 400 ms and 1400 ms for fat and water respectively—rBW/pixel = 190 Hz and monopolar read-out. Parallel imaging acceleration of R = 3 and partial Fourier of 7/8 were applied, resulting in total acquisition time of 4 min 53 s.
2.2.2. 7T measurements
Sagittal 3D multi-echo GRE SMURF images were acquired with anterior-posterior phase-encoding direction, FOV = 230 × 230 mm, resolution = 1.0 × 1.0 × 1.0 mm, 160 slices, TE = {3.85, 7.70, 11.55, 15.40, 19.25, 23.10} ms, TR = 35 ms, FA(water) = 12°, FA(fat) = 20°—the respective Ernst angles assuming T1 values of 550 ms and 1700 ms for fat and water respectively—rBW/pixel = 360 Hz, and monopolar readout. Parallel imaging acceleration of R = 3 and partial Fourier of 6/8 were applied, resulting in total acquisition time of 5 min 13 s.
For all volunteers, a low-resolution conventional GRE scan (i.e., with broadband RF excitation) was acquired to calculate the GRAPPA kernel for fat-water separation using parallel imaging reconstruction. The imaging parameters of the prescans were: in-plane resolution = 3.6 × 3.6 mm, TE/TR = 1.5/3.5 ms at 3 T and TE/TR = 3.0/5.4 ms at 7 T, FA = 5°, rBW/pixel = 690 Hz at 3 T and rBW/pixel = 650 Hz at 7 T with the number of slices and slice thickness the same as in the main SMURF scans. To evaluate possible changes to image contrast, signal phase and resulting susceptibility maps from the use of SMURF RF pulses and processing, high-resolution conventional GRE scans, with the same imaging parameters as was used for the SMURF scans, were also acquired and susceptibility maps were calculated as described in the following section (Supporting Figures S1 and S2, which are available online).
2.3. Data analysis
SMURF fat and water images were reconstructed from raw data in MATLAB (Mathworks Inc, Natick, MA) using slice-GRAPPA 52 and coil-combination with ASPIRE. 53 The fat complex images were corrected for Type 1 (displacement) and Type 2 (phase discrepancy) CSAs and for differences in T1 and relaxation rates (Figure 1). Literature values were used for the T1 correction (400/1400 ms; and 550/1700 ms for fat/water at 3 and 7 Tesla, respectively). 54–56 For correction, separate maps of water and fat were generated and the ratio of fat and water signal weighting, in each voxel, was calculated. To avoid transferring noise from this ratio map to the images in a voxel-by-voxel correction, a single correction factor (CF), calculated as the median of the per-voxel ratios of fat and water signal weighting was used. The median was assessed only over voxels for which the fat fraction (FF) was in the range 0.1 ≤ FF < 0.8:
and the resulting CF was used for the correction of all mixed voxels.
Figure 1. Image processing and postprocessing pipeline.
In Step 1, overlapping water and CAIPIRINHA-shifted fat images are unaliased using slice-GRAPPA. In Step 2, the fat image is shifted to reverse chemical shift displacement (Type 1 chemical shift artefact)— see the position before and after the correction relative to the red reference line -, corrected for the chemical shift-related phase evolution (Type 2 chemical shift artefact) and rescaled to correct for the bias caused by T1 and relaxation rate differences. In Step 3, the complex fat and water images are recombined, generating a fat-water image free of chemical shift and relaxation rate discrepancies.
To generate a head-and-neck mask, maps of temporal phase gradient coherence (Weight 2 in Ref [57,58]) and phase combination quality, i.e., Q-metric, 59 were calculated separately for fat and water. The temporal phase gradient coherence weight, calculated as:
with i and j representing two adjacent spatial locations and Ω the wrapping operator, describes how closely the time evolution of phase is to being linear, and the Q-metric, defined as:
with Mc representing the signal acquired by channel c of the receiver array, describes how well the phase offset-corrected signal is matched between the individual receiver channels. Since both of these metrics are low in the voxels with unreliable phase, these voxels tend to be excluded from masks. To exclude voxels with very short relaxation times, the Q-metric of the second echo was used. The fat and water maps of temporal phase gradient coherence (C) were summed to generate a joint fat-water quality map, which was thresholded to exclude voxels with unreliable phase (threshold value of 0.6). The fat and water combination quality (Q) maps were also thresholded (threshold value of 0.98) and combined using an “OR” operation. A joint fat-water mask was generated by combining the maps of temporal phase gradient coherence (C) and the thresholded quality maps (Q) using an “AND” operation (Figure 2). This mask was smoothed (using smoothn in Matlab 60,61 ) and thresholded, to fill in small holes and remove small disconnected groups of voxels for example, in the subcranial region (threshold value of 0.6), eroded and filled again (smoothn with subsequent thresholding at 0.3). All threshold values, as well as the combination and order of individual processing steps, were arrived at empirically on the first datasets and used for all volunteers.
Figure 2. The generation of a head-and-neck mask using signal phase.
Maps of temporal phase coherence (C) and phase combination quality (Q) are calculated, for fat and water separately, thresholded and combined to generate a single mask (C * Q). This mask is smoothed and thresholded to fill in small holes, eroded, and filled again. Note that the phase combination quality of the second echo is used to exclude voxels with very short relaxation times.
To generate susceptibility maps, the multi-echo images were combined over echoes using inverse-variance weighting. 29,62,63 Phase unwrapping was performed using ROMEO. 57,58 Background-field correction was performed with PDF 64 and susceptibility calculation with STAR 65 using the Sepia toolbox. 66
To assess the need for multi-echo acquisitions, including longer TEs, in QSM imaging of the head-and-neck, QSMs generated from all echoes combined and QSMs generated from individual echoes were compared both qualitatively—by visual assessment of the image contrast, noise, and presence of artifacts—and quantitatively—via calculation of CNR within ROIs—given as:
Within the brain, the CNR was assessed over eight ROIs located in deep gray matter structures (ROIsig) relative to the neighboring tissue (ROIbkg) and within the neck, over eight ROIs located in fatty structures (ROIsig) relative to the neighboring water-based tissues (ROIbkg). In both cases, the noise was estimated as the standard deviation of the signal within an ROI located in a homogeneous area of white matter (ROInoise).
To evaluate the effects of the corrections for CSAs and relaxation rate biases, the susceptibility maps generated from SMURF images with and without the individual corrections were compared both qualitatively—by visual assessment of the correspondence between the location of fatty tissues in the magnitude images and the paramagnetic areas in the susceptibility maps—and quantitatively—via the standard deviations within 6 ROIs located in areas of homogeneous tissue (i.e., lymph nodes, neck muscles and parotid glands). Additionally, susceptibility values of the fatty tissue were assessed within one extensive ROI, drawn over several distant fatty-neck tissue areas, and compared over the corrections. The quantitative ROI analysis (assessment of CNR, the variance of the susceptibility estimates, and assessment of the susceptibility of fatty tissues) was performed for three 3T volunteers (V1–V3) and three 7T volunteers (V6–V8), and the mean values over volunteers were calculated and compared over the corrections.
3. Results
There was a very high degree of correspondence between the recombined SMURF fat-water images acquired at the “random” TEs and the conventional images acquired at the same “random” TEs; and between the “random” recombined SMURF fat-water images corrected for the Type 2 CSA and the conventional images acquired at the “in-phase” TEs. This demonstrates that the use of SMURF RF pulses and image reconstruction has no discernible effect on image contrast or signal phase and that the Type 2 CSA correction effectively removed the phase discrepancy between water and fat (Supporting Figures S1 and S2).
The proposed phase-based masking approach, based on the joint fat-water temporal coherence and phase combination quality (Q-metric 59 ) effectively excluded noisy phase voxels but retained the voxels in the neck (Figure 2), where, due to decreased coil sensitivities, the magnitude of the signal was lower (mainly at 7 T, because the transmit efficiency of the 32-channel head coil is low in inferior regions, and receive coil coverage of the neck area is limited).
The susceptibility maps generated from the first echo provided good tissue contrast in the areas of short , such as the neck and around sinuses, but in the brain, the contrast was low and the maps were noisy (Figure 3A, top row). On the other hand, the susceptibility maps generated from the later echoes showed improved contrast and decreased noise in the brain, but artifacts were present in the neck (Figure 3A, second and third rows). The QSMs generated from all echoes combined showed high contrast and low noise over the entire head-and-neck region (Figure 3A, bottom row). This was confirmed by the quantitative analysis; within the brain, the CNR was higher at the later echoes (Figure 3B, top two rows), but within the neck, the later echoes showed very low CNR and high variance, suggesting unreliability of the results (Figure 3B, bottom two rows). Using all echoes combined yielded high CNR in all cases. Within the brain ROIs, the CNR peaked, for the 3T acquisition, between the third and fourth echo (TEs of 19.5 ms and 26.0 ms), and for the 7T acquisition, between the fifth and sixth echo (TEs of 19.3 ms and 23.1 ms).
Figure 3. Comparison of single-echo and multi-echo susceptibility maps (A) and CNR quantification within the brain and the neck (B).
A) 3T QSM images of brain-only (left column) and entire head-and-neck (right column) generated from the first (first row), third (second row), and fifth (third row) echo of the multi-echo acquisition, or from all echoes (bottom row) of the multi-echo acquisition of one 3T volunteer (V1). Note that, while the contrast of the susceptibility maps within the brain visibly increases and the image noise decreases with increasing TE, errors occur for later echoes outside the brain (red arrows). B) Quantification of the CNR within the brain (deep gray matter regions) and within the neck (fatty-neck regions) for the susceptibility maps generated from one respective echo of the multi-echo acquisition. The mean values and standard deviations of the relative CNR values over ROIs are normalized to the CNR of first-echo susceptibility maps and the mean values over the three volunteers (V1–V3 at 3 T and V6–V8 at 7 T) are displayed. While within the brain, the later echoes show higher CNR, in the neck higher CNR is achieved at the earlier echoes. The highest CNR is generally achieved when using all echoes combined.
If no Type 1 (displacement) CSA correction was applied, phase unwrapping errors occurred (Figure 4, third row) in areas where the displacement of fat led to signal voids with unreliable phase (Figure 4, second row), causing artifacts in the susceptibility maps (Figure 4, bottom row).
Figure 4.
Comparison of phase unwrapping performance when applied to the images recombined without (left column) and with (right column) the correction for the Type 1 (displacement) chemical shift artefact. Note the erroneously unwrapped voxels (third row, red arrows) in the areas, where the displacement of fat results in areas of signal voids (first row) with unreliably defined phase (second row), which lead to artifacts in the susceptibility map (bottom row, red arrows).
The SMURF method achieved generally well-separated fat and water images of the head-and-neck region at both 3 and 7 Tesla. For some volunteers, the larger field inhomogeneity at 7 T resulted in suboptimal fat-water separation in the very inferior or caudal area of the neck, manifesting as a local loss of water signal and swap of fat signal to the water image or the other way around (depending on the sign of the local field deviation) (Supporting Figure S3). These were generally easy to identify and located outside the area of interest, however.
Figure 5 illustrates the fat-water separation quality achieved by SMURF and the effects of the individual corrections for CSAs and relaxation rate biases on the magnitude, phase, and QSM. Without Type 2 (phase discrepancy) CSA correction (Figure 5, first column), the phase discrepancy between fat and water led to signal cancellation in mixed voxels—’India ink’ artifact outlining the fat-water borders. Additionally, the susceptibility values in fatty tissues were higher than with the correction applied. Without the Type 1 (displacement) CSA correction (Figure 5, second column), magnitude images contained areas of fat-water signal overlap as well as signal voids, and susceptibility maps at and around the fat-water borders were blurred and contained artifacts resulting from phase unwrapping errors. These were removed by the application of Type 1 CSA correction (Figure 5, center column). Corrections for fat-water relaxation rate bias (Figure 5, right two columns), primarily in T1, removed the fat hyperintensity (the fact that the fat signal was dis-proportionately large compared to the relative proton densities of the fat and water) and, thus, excessive influence on mixed voxels. This resulted in susceptibility maps with a high correspondence between the strongly paramagnetic areas and the locations of fatty tissue structures.
Figure 5.
Illustration of the effects of corrections for chemical shift artefacts and for bias due to fat-water relaxation rate discrepancy, shown on the magnitude, phase and QSM images of one 3T volunteer (V1). The susceptibility maps generated from the images not corrected for chemical shift effects (row 1) are blurred, and the values are much higher than when corrected for the Type 2 CSA (row 2). The susceptibility maps corrected for the Type 1 and Type 2 CSAs clearly depict the paramagnetic fatty areas (row 3). Corrections for the fat-water differences in T1 (row 4) and (row 5) relaxation rates remove the erroneous domination of fat signal in the mixed voxels. Note the small effect of the correction.
The quantitative analysis also showed that the standard deviation of the QSM values within ROIs, which were located in areas of homogeneous tissue, was reduced when the corrections were applied (Figure 6), with the corrections for Type 1 and Type 2 CSAs having the greatest effects. Without the Type 2 (phase discrepancy) CSA correction, the susceptibility estimates of fatty tissue in the neck were significantly higher than with the correction applied (Figure 7): mean values over the three subjects of 0.97 ppm and 0.54 ppm versus 0.38 ppm and 0.39 ppm at 3 and 7 Tesla, respectively. With all corrections, the susceptibility estimates of fatty neck tissue were 0.41 ppm and 0.40 ppm at 3 and 7 Tesla, respectively.
Figure 6.
Standard deviation of QSM values within ROIs located in areas of homogeneous tissue; mean values over three subjects (V1-V3 and V6-V8 at 3 and 7 Tesla, respectively) are shown. In all ROIs, both at 3 T (left) and 7 T (right), the SD of QSM values is highest when no corrections for the CSAs and relaxation rate biases are applied and decreases with each correction. Note the largest effects attributable to CSA corrections, followed by correction of T1 bias.
Figure 7.
Susceptibility estimates of fatty tissue in the neck at 3 T (left) and 7 T (right); mean values over three subjects (V1-V3 and V6-V8 at 3 and 7 Tesla, respectively) are shown. Without correction for the Type 2 (phase discrepancy) chemical shift artefact, the susceptibility estimates are much higher than with the correction, median susceptibilities of 0.97 ppm and 0.54 ppm without corrections versus 0.41 ppm and 0.40 ppm with all corrections applied, at 3 T and 7 T respectively. The variance of the estimates over the ROI is also much higher without the Type 2 CSA correction.
The estimated values of fat and water in mixed voxels were similar (Table 1), resulting in voxel-specific ratios of the signal weightings close to 1. Consequently, the bias correction, in which a correction factor, calculated as a median of these ratios over all mixed voxels was used, had only a modest effect (Figures 5, 8, and 9).
Figure 8.
Susceptibility maps of all 3T volunteers (V1–V5) generated without any correction and with the corrections for chemical shift artefacts and bias due to fat-water relaxation rate discrepancies applied. Without the corrections (column 4), the susceptibility maps are blurred and the values are, compared to literature values, overestimated. The susceptibility maps corrected for the Type 1 and Type 2 CSA clearly depict the paramagnetic fatty areas (column 5, red arrows). Corrections for the fat-water differences in T1 and relaxation rates remove the dominant influence of fat in the mixed voxels (column 6, red arrows).
Figure 9.
Susceptibility maps of all 7T volunteers (V6-V10) generated without any correction and with the corrections for chemical shift artefacts and bias due to fat-water relaxation rate discrepancies applied. Without the corrections (column 4), the susceptibility maps are blurred and the values are, compared to literature values, overestimated. The susceptibility maps corrected for the Type 1 and Type 2 CSA clearly depict the paramagnetic fatty areas (column 5, red arrows). Corrections for the fat-water differences in T1 and relaxation rates remove the dominant influence of fat in the mixed voxels (column 6, red arrows).
Figures 8,9 and Supporting Figures S4–S13 illustrate the effects of the corrections on susceptibility values for all 3T and 7T volunteers, respectively, and demonstrate that chemical shift artefacts and the bias due to fat-water relaxation differences have to be eliminated to generate susceptibility maps without gross artifacts. Without the Type 2 (phase discrepancy) CSA correction (third columns), the susceptibility values of fatty tissues were visibly higher and without the Type 1 (displacement) CSA correction (fourth columns), the susceptibility maps were more blurred around the fat-water borders and contained artifacts resulting from phase unwrapping errors. Corrections for fat-water relaxation rate bias (last two columns), primarily in T1, removed the excessive influence of fat signal on mixed voxels. With all corrections applied, the susceptibility maps showed a high degree of correspondence between the locations of paramagnetic areas and fatty structures. Note that the application of the corrections significantly improved the quality of the susceptibility maps even in the volunteers with only a small amount of fat in the neck.
4. Discussion
The challenges of quantitative susceptibility mapping outside of the brain, in regions that typically contain significant amounts of fat, have been addressed using a recently developed Simultaneous Multiple Resonance Frequency (SMURF) imaging approach combined with corrections for chemical shift and relaxation effects and a novel masking method. This approach led to high-quality susceptibility maps of the entire head-and-neck region, both at 3 and 7 Tesla, with a strong correspondence to the underlying anatomy.
In this study, we adapted the SMURF approach for 7 T, where due to the larger chemical shift difference, shorter RF pulses of 8 ms compared to roughly 12 ms at 3 T, could be used. This allowed a reduction in the minimum TE, roughly 2.5–4.5 ms instead of 4.5–7.0 ms at 3 T (depending on the receiver bandwidth and resolution), and a reduction in echo spacing. The faster-decaying signal at higher field strength could, thus, be captured and high-quality susceptibility maps of the head-and-neck could be generated both at 3 and 7 Tesla. The CNR of deep gray matter structures peaked, for the 3T measurements, between the third (TE = 19.5 ms) and fourth (TE = 26.0 ms) echo of the five echoes acquired and, for the 7T measurements, between the fifth (TE = 19.3 ms) and sixth (TE = 23.1 ms) echo of the six echoes acquired. This demonstrates that the later echoes are required to achieve high CNR in the brain. On the other hand, in the neck the CNR peaked at the earlier echoes, the second echo at both 3 T (TE = 13.0 ms) and 7 T (TE = 7.7 ms). This shows that, to achieve susceptibility maps with high CNR in the cerebrum, cerebellum, brainstem, as well as in the non-brain regions of the neck, both earlier and later echoes have to be acquired and combined with an effective echo combination approach, such as inverse variance weighting. 29,62,63
A similar fat-water separation quality was expected to be achieved at 3 and 7 Tesla as both the field inhomogeneity and the chemical shift scale with field strength. In fact, while at 3 T correct fat-water separation was achieved in all cases over the entire head-and-neck (in addition to the subcutaneous region of the caudal neck area of two volunteers), at 7 T, some fat-water swaps occurred in the very inferior or caudal area of the neck for all volunteers. The head-and-neck represents, however, a challenging region in which the numerous interfaces between bones, air in sinuses, ears, and pharynx and soft tissue create strong susceptibility artifacts, leading to difficulties in shimming. In our previous study, well-separated fat and water images of breast, knee, and abdomen could be generated at 3 T. 45 These results for the head-and-neck suggest that advanced shimming methods may be needed to be applied to 7T SMURF imaging of problematic body regions to exploit the benefits of the inherently higher SNR and shorter required echo trains.
We have shown that the removal of chemical shift artefacts and T1 relaxation differences is crucial to generating susceptibility maps without gross artifacts, and have demonstrated the sizeable effects of Type 2 (phase discrepancy) CSA and of the often neglected corrections for Type 1 (displacement) CSA and for bias due to fat-water T1 differences. Despite the absence of a ground truth against which to compare susceptibility estimates, a reduction in the variance over ostensibly homogeneous tissues and improved agreement with literature values points to the effectiveness of the corrections and the veracity of the results. Without the corrections, the susceptibility maps were blurred at fat-water borders and the values were, compared to those in the literature (0.2–0.6 ppm), overestimated (medians of 0.97 ppm and 0.54 ppm for 3 and 7 Tesla, respectively). With the corrections applied, these artifacts were removed (medians of 0.41 ppm and 0.40 ppm for 3 and 7 Tesla, respectively) and a high degree of correspondence was achieved between the location of fatty tissues and the strongly paramagnetic areas of susceptibility maps. Additionally, the standard deviation of susceptibility values across the ROIs located in areas of homogeneous tissue was reduced, suggesting lower levels of noise and artifacts. Although in-phase imaging 12,23,24,26 or the Dixon approach 9,13,25,44 can be used to eliminate Type 2 (phase discrepancy) CSA in QSM, Type 1 (displacement) CSA and T1 relaxation bias cannot be addressed with those approaches, as explained in the introduction. Chemical shift and relaxation rate effects can be reduced, but not eliminated, using high receiver bandwidths and low flip angles, but this comes at the price of decreased SNR. With SMURF, Type 1 and Type 2 chemical shift artefacts and T1 relaxation bias could be corrected and artifact-free susceptibility maps of the head-and-neck could be generated.
In this work, the T1 bias correction was performed using literature values of fat and water and led to QSMs with better correspondence to the underlying anatomy. Alternatively, and more precisely, T1 mapping could be carried out and measured T1 values could be used for correction. This would eliminate the need for the assumption that T1 values are the same for all water-based and all fat-based tissues present in the mixed voxels, but would require T1 mapping to be performed for both fat and water.
The values of fat and water in mixed voxels were shown to be quite similar, leading to correction factors for the disparate weighting of fat and water signals, which were close to 1. The similarity in fat-water values suggests, given the known differences in their T2 (48 ms and 29 ms, and 46 ms and 23 ms for fat and water at 3 and 7 Tesla, respectively), 55 that the field inhomogeneity contribution dominates over the T2 contribution to . This concurs with observed estimates in mixed voxels (at 3 T, of 16 ms and 14 ms for fat and water respectively, and at 7 T, of 9 ms for both) being much lower than the literature T2 values. Although this disparity is generally increased for later echoes, later echoes contributed little to the combined phase, and thereby also to the QSMs, because of short constants and inverse-variance weighting. The effect of the correction on susceptibility maps was correspondingly small.
The proposed masking approach, which is based on the signal phase, specifically, the product of temporal phase coherence and phase combination quality (Q-metric), was shown to provide a simple means of generating a head-and-neck mask. A similar phase-based approach has been adopted by Stewart et al, 46 who used thresholded spatial phase coherence. In the presence of fat, however, the spatial coherence would be low at the fat-water borders due to the fat-water phase discrepancy resulting from their susceptibility differences. Although this approach could be applied to the fat and water images separately, the values would be low at tissue edges. In both cases, using spatial phase coherence would, thus, lead to the exclusion of voxels at the fat-water border. Similarly, if the temporal phase coherence were to be calculated directly for the combined fat-water image, the resulting temporal phase coherence maps would be low at the fat-water borders, because of the fat-water susceptibility difference leading to a different temporal evolution of fat and water phase signal. The border voxels would be, thus, excluded from the mask even if Type 2 CSA correction or in-phase acquisition would be used (Supporting Figure S14). To avoid excluding these voxels from the mask, the temporal phase coherence maps of the separated fat and water signals had to be calculated and then combined. The proposed approach was shown to generate a mask that included most of the anatomy of interest and excluded noisy voxels. Nevertheless, if masks were to be needed in more problematic areas still, such as the hearing nerve in a study of acoustic neuroma, for instance, SMURF could be combined with a total-field in-version (TFI) QSM approach. 25,67
As demonstrated in this study, SMURF opens up an alternative approach to QSM outside the brain. In addition to the head-and-neck region, SMURF could be also used for chemical shift and relaxation bias-free susceptibility mapping of the liver, kidney, joints, breast, spine, and other organs and regions. Although in-phase imaging 12,23,24,26 and Dixon imaging 9,13,25,44 are often used in these cases to remove Type 2 (phase discrepancy) CSA, we have shown that the correction of Type 1 (displacement) CSA and also of the bias due to fat-water relaxation differences, which cannot be performed with those approaches, is required for accurate susceptibility mapping. Additionally, to reduce shading artifacts, separate susceptibility maps for fat and water could be calculated separately from SMURF images and then combined. 17 This would, however, make the approach more sensitive to masking and would benefit from, and likely be dependent upon, a TFI approach. Recently, a new preconditioned water-fat TFI (wfTFI) algorithm 68 was developed and used to reduce background field-removal artifacts, noise amplification, and streaking artifacts in QSM of water-fat regions. A graph-cut 25 (Dixon style) approach was used to separate water and fat and to generate a fieldmap and an R2 * map, which were then used as the basis for the QSM inversion. Potentially, SMURF could be also combined with wfTFI, which would also allow Type 2 CSA and relaxation bias to be eliminated.
In addition to GRE-based QSM of non-brain regions, SMURF would also bring benefits to QSM of the brain performed with EPI-based sequences. EPI and 3D EPI, in particular, are attracting increasing interest in QSM due to acquisition speed and the possibility to simultaneously map structure and function. 69–74 In EPI, however, the low receiver bandwidth along the phase-encoding direction results in a large chemical shift displacement of the fat signal, which can shift fat from the skull into the brain and subcutaneous fat in the neck such that it overlays inferior brain regions such as the brainstem. Fat-suppression could be used, although the high specific absorption rate (SAR) demand of fat-sat pulses at ultrahigh field (UHF) make fat-water imaging an appealing solution 75 and one that would retain fat information for QSM of non-brain regions and also possibly allow the use of fat images as navigators for prospective motion correction 76 and dynamic B0 shimming. 77
We have shown that CSAs and relaxation rate biases, that adversely affect the reliability of QSM in fatty regions, can be reduced using fat-water imaging with SMURF. This approach was shown to generate high-quality susceptibility maps of the head-and-neck at 3 and 7 Tesla with a strong correspondence to the known anatomy.
5. Conclusions
The challenges of QSM outside of the brain, in regions that typically contain significant amounts of fat, have been addressed with SMURF fat-water imaging with corrections for Type 1 and Type 2 CSAs and for relaxation rate biases. Unlike conventional imaging with broadband excitation, this approach was shown to be required to generate susceptibility maps of fatty regions without gross artifacts and with high correspondence between the paramagnetic areas and the locations of fatty tissues.
Supplementary Material
Table 1. Fat and water values and the weighting bias correction factor.
Median values over all mixed voxels with fat fractions in the range 0.1 ≤ FF < 0.8 are shown. Note the similarity of values of fat and water, resulting in the weighting bias correction factors being close to 1.
| T values of fat and water and weighting bias correction factor | |||
|---|---|---|---|
| 3 Tesla | |||
| Fat | Water | weighting bias correction factor (echo1/echo2/echo3/echo4/echo5) | |
| V1 | 17.72 | 14.80 | 1.06/1.12/1.19/1.26/1.34 |
| V2 | 17.83 | 14.13 | 1.09/1.19/1.30/1.42/1.55 |
| V3 | 16.98 | 15.01 | 1.06/1.12/1.19/1.25/1.33 |
| V4 | 12.01 | 11.77 | 1.04/1.08/1.12/1.17/1.21 |
| V5 | 17.81 | 13.98 | 1.09/1.19/1.30/1.42/1.55 |
| 7 Tesla | |||
| Fat | Water | weighting bias correction factor (echo1/echo2/echo3/echo4/echo5/echo6) | |
| V6 | 7.53 | 8.88 | 0.93/0.86/0.80/0.75/0.69/0.64 |
| V7 | 9.62 | 8.86 | 1.03/1.07/1.10/1.14/1.18/1.22 |
| V8 | 9.12 | 8.58 | 1.02/1.04/1.07/1.09/1.12/1.14 |
| V9 | 7.69 | 8.61 | 0.95/0.90/0.85/0.81/0.76/0.72 |
| V10 | 8.79 | 8.86 | 0.99/0.98/0.98/0.97/0.96/0.95 |
Acknowledgements
This study was funded by the Austrian Science Fund (FWF 31452). SR was additionally supported by the Marie Skłodowska-Curie Action (MS-fMRI-QSM 794298). The financial support by the Austrian Federal Ministry for Digital and Economic Affairs and the National Foundation for Research, Technology and Development is also gratefully acknowledged.
Funding information
Bundesministerium für Digitalisierung und Wirtschaftsstandort; H2020 Marie Skłodowska-Curie Actions, Grant/ Award Number: MS-fMRI-QSM 794298; Österreichische Nationalstiftung für Forschung, Technologie und Entwicklung; Austrian Science Fund, Grant/Award Number: FWF 31452
Footnotes
Simon Daniel Robinson @simon_mri
Data Availability Statement
The data that support the findings of this study, including the ROIs, are openly available in “3T SMURF-based QSM data” and “7T SMURF-based QSM data” https://doi.org/10.7910/DVN/IEFCRA at and https://doi.org/10.7910/DVN/4HBJC6, references [78] and [79], respectively.
References
- 1.Shmueli K, de Zwart JA, van Gelderen P, Li TQ, Dodd SJ, Duyn JH. Magnetic susceptibility mapping of brain tissue in vivo using MRI phase data. Magn Reson Med. 2009;62:1510–1522. doi: 10.1002/mrm.22135. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Schweser F, Robinson SD, de Rochefort L, Li W, Bredies K. An illustrated comparison of processing methods for phase MRI and QSM: removal of background field contributions from sources outside the region of interest. NMR Biomed. 2017;30:e3604. doi: 10.1002/nbm.3604. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Wang Y, Liu T. Quantitative susceptibility mapping (QSM): decoding MRI data for a tissue magnetic biomarker. Magn Reson Med. 2015;73:82–101. doi: 10.1002/mrm.25358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 4.Wang YI, Spincemaille P, Liu Z, et al. Clinical quantitative susceptibility mapping (QSM): Biometal imaging and its emerging roles in patient care: clinical QSM Biometals. J Magn Reson Imaging. 2017;46:951–971. doi: 10.1002/jmri.25693. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Ravanfar P, Loi SM, Syeda WT, et al. Systematic review: Quantitative Susceptibility Mapping (QSM) of brain iron profile in neurodegenerative diseases. Front Neurosci. 2021;15 doi: 10.3389/fnins.2021.618435. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Vinayagamani S, Sheelakumari R, Sabarish S, et al. Quantitative susceptibility mapping: technical considerations and clinical applications in neuroimaging. J Magn Reson Imaging. 2021;53:23–37. doi: 10.1002/jmri.27058. [DOI] [PubMed] [Google Scholar]
- 7.Sharma SD, Fischer R, Schoennagel BP, et al. MRI-based quantitative susceptibility mapping (QSM) and R2* mapping of liver iron overload: comparison with SQUID-based biomagnetic liver susceptometry: QSM and R2*: comparison With SQUID. Magn Reson Med. 2017;78:264–270. doi: 10.1002/mrm.26358. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Sharma SD, Hernando D, Horng DE, Reeder SB. Quantitative Susceptibility Mapping in the abdomen as an imaging biomarker of hepatic iron overload. Magn Reson Med. 2015;74:673–683. doi: 10.1002/mrm.25448. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Dong J, Liu T, Chen F, et al. Simultaneous phase unwrapping and removal of chemical shift (SPURS) using graph cuts: application in quantitative susceptibility mapping. IEEE Trans Med Imaging. 2015;34:531–540. doi: 10.1109/TMI.2014.2361764. [DOI] [PubMed] [Google Scholar]
- 10.Li J, Lin H, Liu T, et al. Quantitative susceptibility mapping (QSM) minimizes interference from cellular pathology in R2* estimation of liver iron concentration. J Magn Reson Imaging. 2018;48:1069–1079. doi: 10.1002/jmri.26019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Jafari R, Sheth S, Spincemaille P, et al. Rapid automated liver quantitative susceptibility mapping. J Magn Reson Imaging. 2019;50:725–732. doi: 10.1002/jmri.26632. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Schweser F, Herrmann K-H, Deistung A, et al. Quantitative magnetic susceptibility mapping (QSM) in breast disease reveals additional information for MR-based characterization of carcinoma and calcification; Proceedings of the 19th Annual Meeting of ISMRM; 2011. #1014. [Google Scholar]
- 13.Dimov AV, Liu T, Spincemaille P, et al. Joint estimation of chemical shift and quantitative susceptibility mapping (chemical QSM) Magn Reson Med. 2015;73:2100–2110. doi: 10.1002/mrm.25328. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Xie L, Dibb R, Cofer GP, et al. Susceptibility tensor imaging of the kidney and its microstructural underpinnings. Magn Reson Med. 2015;73:1270–1281. doi: 10.1002/mrm.25219. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Straub S, Laun FB, Emmerich J, et al. Potential of quantitative susceptibility mapping for detection of prostatic calcifications. J Magn Reson Imaging. 2017;45:889–898. doi: 10.1002/jmri.25385. [DOI] [PubMed] [Google Scholar]
- 16.Straub S, Emmerich J, Schlemmer H-P, et al. Mask-adapted background field removal for artifact reduction in quantitative susceptibility mapping of the prostate. Tomography. 2017;3:96–100. doi: 10.18383/j.tom.2017.00005. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Sato R, Shirai T, Soutome Y, Bito Y, Ochi H. Quantitative susceptibility mapping of prostate with separate calculations for water and fat regions for reducing shading artifacts. Magn Reson Imaging. 2020;66:22–29. doi: 10.1016/j.mri.2019.11.006. [DOI] [PubMed] [Google Scholar]
- 18.Dibb R, Qi Y, Liu C. Magnetic susceptibility anisotropy of myocardium imaged by cardiovascular magnetic resonance reflects the anisotropy of myocardial filament alpha-helix polypeptide bonds. J Cardiovasc Magn Reson. 2015;17:60. doi: 10.1186/s12968-015-0159-4. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Dibb R, Xie L, Wei H, Liu C. Magnetic susceptibility anisotropy outside the central nervous system. NMR Biomed. 2017;30 doi: 10.1002/nbm.3544. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Wen Y, Nguyen TD, Liu Z, et al. Cardiac quantitative susceptibility mapping (QSM) for heart chamber oxygenation. Magn Reson Med. 2018;79:1545–1552. doi: 10.1002/mrm.26808. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Wei H, Dibb R, Decker K, et al. Investigating magnetic susceptibility of human knee joint at 7 Tesla. Magn Reson Med. 2017;78:1933–1943. doi: 10.1002/mrm.26596. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22.Diefenbach MN, Meineke J, Ruschke S, Baum T, Gersing A, Karampinos DC. On the sensitivity of quantitative susceptibility mapping for measuring trabecular bone density. Magn Reson Med. 2019;81:1739–1754. doi: 10.1002/mrm.27531. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Guo Y, Liu Z, Wen Y, et al. Quantitative susceptibility mapping of the spine using in-phase echoes to initialize inhomogeneous field and R2* for the nonconvex optimization problem of fat-water separation. NMR Biomed. 2019;32:e4156. doi: 10.1002/nbm.4156. [DOI] [PubMed] [Google Scholar]
- 24.Karsa A, Punwani S, Shmueli K. An optimized and highly repeatable MRI acquisition and processing pipeline for quantitative susceptibility mapping in the head-and-neck region. Magn Reson Med. 2020;84:3206–3222. doi: 10.1002/mrm.28377. [DOI] [PubMed] [Google Scholar]
- 25.Boehm C, Diefenbach MN, Makowski MR, Karampinos DC. Improved body quantitative susceptibility mapping by using a variable-layer single-min-cut graph-cut for field-mapping. Magn Reson Med. 2021;85:1697–1712. doi: 10.1002/mrm.28515. [DOI] [PubMed] [Google Scholar]
- 26.Buch S, Liu S, Ye Y, Cheng Y-CN, Neelavalli J, Haacke EM. Susceptibility mapping of air, bone, and calcium in the head. Magn Reson Med. 2015;73:2185–2194. doi: 10.1002/mrm.25350. [DOI] [PubMed] [Google Scholar]
- 27.Ruetten PPR, Cluroe AD, Usman A, Priest AN, Gillard JH, Graves MJ. Simultaneous MRI water-fat separation and quantitative susceptibility mapping of carotid artery plaque pre- and post-ultrasmall superparamagnetic iron oxide-uptake. Magn Reson Med. 2020;84:686–697. doi: 10.1002/mrm.28151. [DOI] [PubMed] [Google Scholar]
- 28.Ikebe Y, Ishimaru H, Imai H, et al. Quantitative susceptibility mapping for carotid atherosclerotic plaques: a pilot study. Magn Reson Med Sci. 2020;19:135–140. doi: 10.2463/mrms.mp.2018-0077. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Bachrata B, Eckstein K, Trattnig S, Robinson SD. QSM of the head-and-neck at 7T using simultaneous fat-water imaging with SMURF; Proceedings of the 29th Annual Meeting of ISMRM. Virtual Conference; 2021. #0795. [Google Scholar]
- 30.Deistung A, Schäfer A, Schweser F, et al. High-resolution MR imaging of the human brainstem in vivo at 7 Tesla. Front Hum Neurosci. 2013;7:710. doi: 10.3389/fnhum.2013.00710. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Elias GJB, Loh A, Gwun D, et al. Deep brain stimulation of the brainstem. Brain. 2021;144:712–723. doi: 10.1093/brain/awaa374. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Mazzone P, Vilela Filho O, Viselli F, et al. Our first decade of experience in deep brain stimulation of the brainstem: elucidating the mechanism of action of stimulation of the ventrolateral pontine tegmentum. J Neural Transm. 2016;123:751–767. doi: 10.1007/s00702-016-1518-5. [DOI] [PubMed] [Google Scholar]
- 33.Eskreis-Winkler S, Deh K, Gupta A, et al. Multiple sclerosis lesion geometry in quantitative susceptibility mapping (QSM) and phase imaging. J Magn Reson Imaging. 2015;42:224–229. doi: 10.1002/jmri.24745. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Karsa A. Optimising MRI Magnetic Susceptibility Mapping for Applications in Challenging Regions of the Body. University College London; 2018. [Accessed September 16, 2021]. https://discovery.ucl.ac.uk/id/eprint/10064348/ [Google Scholar]
- 35.Ren J, Dimitrov I, Sherry AD, Malloy CR. Composition of adipose tissue and marrow fat in humans by 1H NMR at 7 Tesla. J Lipid Res. 2008;49:2055–2062. doi: 10.1194/jlr.D800010-JLR200. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Spincemaille P, Anderson J, Wu G, et al. Quantitative Susceptibility Mapping: MRI at 7T versus 3T. J Neuroimaging. 2020;30:65–75. doi: 10.1111/jon.12669. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Reeder SB, Hu HH, Sirlin CB. Proton density fat-fraction: a standardized mr-based biomarker of tissue fat concentration. J Magn Reson Imaging. 2012;36:1011–1014. doi: 10.1002/jmri.23741. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Yang IY, Cui Y, Wiens CN, Wade TP, Friesen-Waldner LJ, McKenzie CA. Fat fraction bias correction using T1 estimates and flip angle mapping. J Magn Reson Imaging. 2014;39:217–223. doi: 10.1002/jmri.24126. [DOI] [PubMed] [Google Scholar]
- 39.Yu H, McKenzie CA, Shimakawa A, et al. Multiecho reconstruction for simultaneous water-fat decomposition and T2* estimation. J Magn Reson Imaging. 2007;26:1153–1161. doi: 10.1002/jmri.21090. [DOI] [PubMed] [Google Scholar]
- 40.Dixon WT. Simple proton spectroscopic imaging. Radiology. 1984;153:189–194. doi: 10.1148/radiology.153.1.6089263. [DOI] [PubMed] [Google Scholar]
- 41.Pineda AR, Reeder SB, Wen Z, Pelc NJ. Cramér-Rao bounds for three-point decomposition of water and fat. Magn Reson Med. 2005;54:625–635. doi: 10.1002/mrm.20623. [DOI] [PubMed] [Google Scholar]
- 42.Reeder SB, McKenzie CA, Pineda AR, et al. Water-fat separation with IDEAL gradient-echo imaging. J Magn Reson Imaging. 2007;25:644–652. doi: 10.1002/jmri.20831. [DOI] [PubMed] [Google Scholar]
- 43.Wu B, Li W, Avram AV, Gho SM, Liu C. Fast and tissue-optimized mapping of magnetic susceptibility and T2* with multi-echo and multi-shot spirals. NeuroImage. 2012;59:297–305. doi: 10.1016/j.neuroimage.2011.07.019. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Liu J, Christiansen SD, Drangova M. Single multi-echo GRE acquisition with short and long echo spacing for simultaneous quantitative mapping of fat fraction, B0 inhomogeneity, and susceptibility. NeuroImage. 2018;172:703–717. doi: 10.1016/j.neuroimage.2018.02.012. [DOI] [PubMed] [Google Scholar]
- 45.Bachrata B, Strasser B, Bogner W, et al. Simultaneous Multiple Resonance Frequency imaging (SMURF): Fat-water imaging using multi-band principles. Magn Reson Med. 2021;85:1379–1396. doi: 10.1002/mrm.28519. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Stewart AW, Robinson SD, O’Brien K, et al. QSMxT: robust masking and artefact reduction for quantitative susceptibility mapping. bioRxiv. :2021.05.05.442850. doi: 10.1002/mrm.29048. Published online May 6, 2021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Shinnar M, Eleff S, Subramanian H, Leigh JS. The synthesis of pulse sequences yielding arbitrary magnetization vectors. Magn Reson Med. 1989;12:74–80. doi: 10.1002/mrm.1910120109. [DOI] [PubMed] [Google Scholar]
- 48.Le Roux P. Exact synthesis of radio frequency waveforms; Proceedings of the 7th Annual Meeting of SMRM; San Francisco, CA, USA. 1988. p. 1049. [Google Scholar]
- 49.Pauly J, Le Roux P, Nishimura D, Macovski A. Parameter relations for the Shinnar-Le Roux selective excitation pulse design algorithm. IEEE Trans Med Imaging. 1991;10:53–65. doi: 10.1109/42.75611. [DOI] [PubMed] [Google Scholar]
- 50.Soher BJ, Semanchuk P, Todd D, Steinberg J. VeSPA: Integrated applications for RF pulse design, spectral simulation and MRS data analysis. Proceedings of the 19th Annual Meeting of ISMRM; Montreal, Quebec, Canada. 2011. #1410. [Google Scholar]
- 51. [Accessed September 16, 2021]. https://vespa-mrs.github.io/vespa.io/
- 52.Setsompop K, Gagoski BA, Polimeni JR, Witzel T, Wedeen VJ, Wald LL. Blipped-Controlled Aliasing in Parallel Imaging (blipped-CAIPI) for simultaneous multi-slice EPI with reduced g-factor penalty. Magn Reson Med. 2012;67:1210–1224. doi: 10.1002/mrm.23097. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 53.Eckstein K, Dymerska B, Bachrata B, et al. Computationally efficient combination of multi-channel phase data from multi-echo acquisitions (ASPIRE) Magn Reson Med. 2018;79:2996–3006. doi: 10.1002/mrm.26963. [DOI] [PubMed] [Google Scholar]
- 54.Gold GE, Han E, Stainsby J, Wright G, Brittain J, Beaulieu C. Musculoskeletal MRI at 3.0 T: relaxation times and image contrast. Am J Roentgenol. 2004;183:343–351. doi: 10.2214/ajr.183.2.1830343. [DOI] [PubMed] [Google Scholar]
- 55.Jordan CD, Saranathan M, Bangerter NK, Hargreaves BA, Gold GE. Musculoskeletal MRI at 3.0T and 7.0T: a comparison of relaxation times and image contrast. Eur J Radiol. 2013;82:734–739. doi: 10.1016/j.ejrad.2011.09.021. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Rooney WD, Johnson G, Li X, et al. Magnetic field and tissue dependencies of human brain longitudinal 1H2O relaxation in vivo. Magn Reson Med. 2007;57:308–318. doi: 10.1002/mrm.21122. [DOI] [PubMed] [Google Scholar]
- 57.Dymerska B, Eckstein K, Bachrata B, et al. Phase unwrapping with a rapid opensource minimum spanning tree algorithm (ROMEO) Magn Reson Med. 2021;85:2294–2308. doi: 10.1002/mrm.28563. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 58. [Accessed September 16, 2021]. https://github.com/korbinian90/ROMEO .
- 59.Robinson S, Grabner G, Witoszynskyj S, Trattnig S. Combining phase images from multi-channel RF coils using 3D phase offset maps derived from a dual-echo scan. Magn Reson Med. 2011;65:1638–1648. doi: 10.1002/mrm.22753. [DOI] [PubMed] [Google Scholar]
- 60.Garcia D. smoothn. [Accessed September 16, 2021]. https://de.mathworks.com/matlabcentral/fileexchange/25634-smoothn .
- 61.Garcia D. Robust smoothing of gridded data in one and higher dimensions with missing values. Comput Stat Data Anal. 2010;54:1167–1178. doi: 10.1016/j.csda.2009.09.020. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 62.Eckstein K, Bachrata B, Hangel G, et al. Improved susceptibility weighted imaging at ultra-high field using bipolar multi-echo acquisition and optimized image processing: CLEAR-SWI. NeuroImage. 2021;237:118175. doi: 10.1016/j.neuroimage.2021.118175. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 63.Quinn MP, Gati JS, Klassen LM, et al. Comparison of multiecho postprocessing schemes for SWI with use of linear and nonlinear mask functions. Am J Neuroradiol. 2014;35:38–44. doi: 10.3174/ajnr.A3584. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 64.Liu T, Khalidov I, de Rochefort L, et al. A novel background field removal method for MRI using projection onto dipole fields (PDF) NMR Biomed. 2011;24:1129–1136. doi: 10.1002/nbm.1670. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 65.Li W, Wang N, Yu F, et al. A method for estimating and removing streaking artifacts in quantitative susceptibility mapping. NeuroImage. 2015;108:111–122. doi: 10.1016/j.neuroimage.2014.12.043. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 66.Chan K-S, Marques JP. SEPIA-Susceptibility mapping pipeline tool for phase images. NeuroImage. 2020;227:117611. doi: 10.1016/j.neuroimage.2020.117611. [DOI] [PubMed] [Google Scholar]
- 67.Liu Z, Kee Y, Zhou D, Wang Y, Spincemaille P. Preconditioned total field inversion (TFI) method for quantitative susceptibility mapping. Magn Reson Med. 2017;78:303–315. doi: 10.1002/mrm.26331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 68.Boehm C, Sollmann N, Meineke J, et al. Preconditioned water-fat total field inversion: application to spine quantitative susceptibility mapping. Magn Reson Med. 2022;87:417–430. doi: 10.1002/mrm.28903. [DOI] [PubMed] [Google Scholar]
- 69.Langkammer C, Bredies K, Poser BA, et al. Fast quantitative susceptibility mapping using 3D EPI and total generalized variation. NeuroImage. 2015;111:622–630. doi: 10.1016/j.neuroimage.2015.02.041. [DOI] [PubMed] [Google Scholar]
- 70.Sun H, Wilman AH. Quantitative susceptibility mapping using single-shot echo-planar imaging: Echo-Planar Quantitative Susceptibility Mapping. Magn Reson Med. 2015;73:1932–1938. doi: 10.1002/mrm.25316. [DOI] [PubMed] [Google Scholar]
- 71.Sun H, Seres P, Wilman AH. Structural and functional quantitative susceptibility mapping from standard fMRI studies: QSM from fMRI studies. NMR Biomed. 2017;30:e3619. doi: 10.1002/nbm.3619. [DOI] [PubMed] [Google Scholar]
- 72.Tourell M, Jin J, Stewart A, et al. Submillimeter, sub-minute quantitative susceptibility mapping using a multi-shot 3D-EPI with 2D CAIPIRINHA acceleration; Proceedings of the 29th Annual Meeting of ISMRM. Virtual Conference; 2021. #0787. [Google Scholar]
- 73.Bachrata B, Eckstein K, Trattnig S, Robinson SD. Considerations in quantitative susceptibility mapping using echo planar imaging; Proceedings of the 26th Annual Meeting of ISMRM; Paris, France. 2018. #4996. [Google Scholar]
- 74.Ehrmann A, Bachrata B, Eckstein K, et al. Fast, high resolution QSM of the brainstem at 7T using super-resolution 2D EPI; Proceedings of the 28th Annual Meeting of ISMRM. Virtual Conference; 2020. #3785. [Google Scholar]
- 75.Seginer A, Furman-Haran E, Goldberg I, Schmidt R. Lipid and water separation through an SMS-like approach in 7T to reduce SAR in brain EPI; Proceedings of the 29th Annual Meeting of ISMRM. Virtual Conference; 2021. #0891. [Google Scholar]
- 76.Skare S, Hartwig A, Mårtensson M, Avventi E, Engström M. Properties of a 2D fat navigator for prospective image domain correction of nodding motion in brain MRI. Magn Reson Med. 2015;73:1110–1119. doi: 10.1002/mrm.25234. [DOI] [PubMed] [Google Scholar]
- 77.Gretch F, Kober T, Waszak M, Marques JP, Gallichan D. High temporal resolution retrospective motion and B0 correction using FIDNavs and segmented FatNavs at 7 T; Proceedings of the 24th Annual Meeting of ISMRM; Singapore. 2016. #4251. [Google Scholar]
- 78.Bachrata B. 3T SMURF-based QSM data. Harvard Dataverse, V1. 2021 doi: 10.7910/DVN/IEFCRA. [DOI] [Google Scholar]
- 79.Bachrata B. 7T SMURF-based QSM data. Harvard Dataverse, V1. 2021 doi: 10.7910/DVN/4HBJC6. [DOI] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data that support the findings of this study, including the ROIs, are openly available in “3T SMURF-based QSM data” and “7T SMURF-based QSM data” https://doi.org/10.7910/DVN/IEFCRA at and https://doi.org/10.7910/DVN/4HBJC6, references [78] and [79], respectively.