Abstract
We studied the use of powered ankle-foot orthoses (PAFOs) and walking stability of the wearers, focusing on the ankle joint, which is known to play a critical role in gait stability. Recognizing that the subtalar joint is an important modulator of walking stability, we conducted the walking experiment on a treadmill by applying varying assistance techniques to the 2-degree-of-freedom (DOF) PAFO, which has the subtalar joint as the rotating axis, and the commonly used 1-DOF PAFO. The participants were 8 healthy men (mean±SD: height, 174.8±7.1 cm; weight, 69.8±6.5 kg; and age, 29.1±4.8 years) with no history of gait abnormality. Center of pressure (COP) was measured with an in-shoe pressure sensor, and stability was estimated on the basis of the angular acceleration measured with the inertial measurement unit attached to the trunk. The experimental results of the 2-DOF PAFO, with or without assistance, showed a significantly higher stability than those of the 1-DOF PAFO (up to 23.78%, p<0.0326). With the 1-DOF PAFO, the stability deteriorated with the increase in the degree of assistance provided. With the 2-DOF PAFO, this tendency was not observed. Thus, the importance of the subtalar joint was proven using PAFOs. The mean position analysis of the COP during the stance phase confirmed that the COP highly correlated with stability (Pearson correlation coefficient: −0.6607). Thus, we conclude that only the 2-DOF PAFO can maintain walking stability, regardless of the assistance characteristics, by preserving the COP in the medial position through eversion. Awareness regarding the role of the subtalar joint is necessary during the manufacture or use of PAFOs, as lack of awareness could lead to the degradation of the wearer’s gait stability, regardless of effective assistance, and deteriorate the fundamental functionality of PAFO.
Introduction
Exoskeletons, also called wearable suits, are a type of robot designed to be worn by a person to provide functions that assist in human movement [1]. These robots are being actively developed in the field of rehabilitation engineering and are usually used for elderly people or patients who have lost their gait ability. They can be used as rehabilitation exercise devices to improve or relearn walking skills and have been developed and utilized for various other purposes [2–7]. Among these devices, powered ankle-foot orthoses (PAFOs) are used to assist ankle joints [8]. The ankle joint has an important role in walking assistance because it is the rotation axis connecting the calf and foot, which not only exerts the largest torque when walking but also controls the movement of the foot, and acts as an end-effector by directly contacting the ground [9]. The ankle joints consist of talocrural and subtalar joints, which are responsible for propulsion through dorsi/plantar flexion and adaptation to the ground through inversion/eversion, respectively [10].
The users of PAFOs are people who have weakened muscles because of aging or disease. Usually, the fall risk of these individuals is increased due to weakened muscle strength [11], so they supplement their loss of muscle power with assistance of PAFOs to be free from the danger of falls and conduct normal walking. Previous research has focused on the talocrural joint, which is the rotation axis of dorsiflexion that prevents foot drop during the swing phase or of plantar flexion, which is responsible for the push-off during the gait cycle [12–14]. However, the lack of research on the subtalar joint, which keeps the body balanced and stable when walking, can be contradictory to the development of PAFOs.
Stability is defined as resistance to biomechanical changes [15]. It means that the center of mass (COM), which is regarded as the point of action for the resultant perturbation, should be kept in a stable position or trajectory for balance. Balance is defined as the ability to keep the COM in the base of support. Thus, stability can be an index for balance and should be secured for conducting balanced motion [16]. For that reason, postural sway, which is the movement of the COM, which is the trunk in this case [17], is used to measure clinical changes of balance [18–20]. Thus, for maintaining balanced walking, stability should be maintained during the gait cycle, and the subtalar joint is the key element of PAFOs to help wearers maintain the aforementioned functions.
The subtalar joints maintain the postural control of the body in the frontal plane by adjusting the position of the center of pressure (COP) through rotation of the foot [21–23]. The location of the COP is an important indicator of stability during gait, that is, gait stability, and influences the calculation of the extrapolated COM and margin of stability [24, 25]. Therefore, in the absence of the subtalar joints, the position and speed of the COP are limited, resulting in increased postural sway of the frontal plane during the one-leg stance. Hoogveliet et al. demonstrated this through experiments using ankle braces [23]. Thus, if PAFOs do not have a subtalar joint as the rotational axis, it might generate postural problems in wearers.
Another problem results from the absence of a subtalar joint. Bok et al. found that factors related to the deterioration of stability due to aging include the weakening strengths of the muscles responsible for eversion by the subtalar joint and the consequent reduction in range of motion [26]. In other words, the strength of the peroneus longus muscles is weakened due to aging, which prevents these muscles to play their role as an evertor sufficiently during the gait cycle. In fact, the three-dimensional moment of eversion was significantly reduced due to aging in elderly individuals as compared with healthy young individuals [27]. As a result, in elderly people, the walking skills are modified by factors such as increased step width and decreased walking speed to avoid falls, but the reduced mediolateral stability is not avoided. Various factors contribute to this situation, of which the foot tilt strategy (FTS) is the major factor related to the subtalar joint. In the FTS, the evertor generates the stabilizing moment with rotation by the subtalar joint to compensate for the tilting moment resulting from the offset between the foot plantar COP and the projection of the COM [22]. Thus, humans can keep their body in a stable state. When this process cannot be conducted normally, stability would deteriorate, and a fall could occur.
Falls occur when gait stability is lost during walking [11, 16]. This can be confirmed by comparing the gait stability of fall-prone elderly people, healthy elderly people, and young individuals. Liu et al. assessed fall risk by using the inertial measurement unit (IMU) in the three populations [19]. The fall-prone elderly people had significantly greater COP speed, mediolateral and anteroposterior accelerations of the trunk, and maximum Lyapunov exponent. In addition, Howcroft et al. also proved the difference in stability between fall-prone and healthy elderly people during dual-task walking by using a triaxial accelerometer and pressure-sensing insole [20].
These situations show the reason why the subtalar joints should not be excluded from the development of exoskeletons. However, the existing PAFOs have 1 degree of freedom (DOF) with a talocrural joint. This means that they only have a fixed point of action and a constant force direction regardless of the type of actuator, which cannot help wearers keep their balance from external perturbations. Nevertheless, they are still useful because their focus was on the effective assistance for reducing metabolic rate to the maximum. However, if subjects with low stability due to the absence of a subtalar joint are given a certain direction of assistance limited to that subtalar joint, as discussed by Anam and Al-Jumaily, this can be perceived as a perturbation by the robot-human interface [28]. The muscle activity of the ankle might be reduced, and the overall stability is deteriorated, which affects other joints and elements. Therefore, these assistive techniques is unlikely easy to use owing to these shortcomings, as safety is the most important factor considering the characteristics of the subjects.
Therefore, we hypothesized that the presence of a subtalar joint in the PAFO design would improve the wearer’s stability. As the DOF increases, the wearer can achieve the FTS with assistance; through this, direction can be controlled. Thus, we expected that effective force transmission in the optimal direction with high stability is possible. Currently, 2-DOF PAFOs with actuatable talocrural and subtalar joints with pneumatic artificial muscles (PAMs) are available. In this study, by using the existing 2-DOF PAFOs and the additionally manufactured 1-DOF PAFOs, the effect of the subtalar joint in the PAFO and the change in the wearer’s gait stability resulting from various assistance techniques were analyzed to prove our hypothesis using the COP and stabilogram data.
Materials and methods
This study compared the wearer’s stability, measured in unpowered and powered conditions, with the changes in the size and direction of assistance during walking with the 1- and 2-DOF PAFOs. The 2-DOF PAFO used for the experiments was developed by Choi et al. [29], and an additional 1-DOF PAFO lacking a subtalar joint was developed for this study.
The 2-DOF powered ankle-foot orthosis
The 2-DOF PAFO used in this study divides the ankle joint into 2 axes of rotation, as shown in Fig 1A and explained by Choi et al [29]. Spatial formulas for the talocrural and subtalar joints were calculated using the mean of the individual values based on the anthropometric data of Isman and Inman [10]. The robot is interfaced with humans at the feet, calves, and thighs as shown in Fig 1C. Force sensitive resistors (FSRs) are located under the foot to determine the heel strike, and each joint is equipped with an encoder to measure the angle of rotation. Two PAMs are used as the actuator, one acting as the soleus to aid plantar flexion and the other acting as the peroneus longus to assist eversion with plantar flexion as shown in Fig 2. Both PAMs are connected to the solenoid valves, which are controlled by pulse width modulation (PWM) signals and supplied with air pressure, maintained at 6 bar, from the air compressor and regulator. The PAMs and frame are connected by wires, and in the middle of the wires, tensile loadcells are attached to measure the force transmitted. All sensing values are measured on the STMicroelectronics (STM) Nucleo board and transmitted to the STM Discovery board via controller area network (CAN) communication. The discovery board determines the stage of the gait cycle based on the acquired values and sends the certain duty ratio calculated for the contraction of the PAMs to the solenoid valve to implement PAFO assistance. A more detailed explanation is given in the paper by Choi et al. The 1-DOF PAFO was manufactured by simply removing the subtalar joint, as shown in Fig 1B.
Assistance techniques
Various assistance techniques were developed and applied to compare the stability between wearing the 1- and 2-DOF PAFOs. The types of assistance techniques are listed in Table 1. For the reference data, the experiments were conducted in the unpowered state without assistance.
Table 1. Types of assistance technique for PAFOs.
DOF | Duty ratio [%] | Abbr. | PAM1 | PAM2 | |||||
---|---|---|---|---|---|---|---|---|---|
PAM1 | PAM2 | fpeak[N] | tass[%] | tppo[%] | fpeak[N] | tass[%] | tppo[%] | ||
1 | 0 | 0 | 1 DOF UNP | - | - | - | - | - | - |
30 | 0 | 1 DOF A | 59.73 (±25.47) | 21.3 (±6.8) | 47.8 (±3.8) | - | - | - | |
60 | 0 | 1 DOF B | 93.51 (±20.92) | 15.9 (±4.6) | 46.0 (±6.3) | - | - | - | |
90 | 0 | 1 DOF C | 132.42 (±22.29) | 12.1 (±3.4) | 43.6 (±11.0) | - | - | - | |
2 | 0 | 0 | 2 DOF UNP | - | - | - | - | - | - |
30 | 0 | 2 DOF A1 | 60.03 (±19.40) | 21.5 (±9.1) | 48.5 (±4.3) | - | - | - | |
30 | 30 | 2 DOF A2 | 20.76 (±13.40) | 18.9 (±6.9) | 48.5 (±3.4) | 79.06 (±26.06) | 27.8 (±6.6) | 49.6 (±3.8) | |
60 | 0 | 2 DOF B1 | 110.08 (±33.43) | 17.1 (±3.8) | 51.4 (±4.4) | - | - | - | |
60 | 60 | 2 DOF B2 | 57.45 (±38.04) | 17.5 (±5.0) | 48.9 (±4.4) | 76.21 (±24.45) | 23.1 (±4.4) | 49.9 (±5.4) | |
90 | 0 | 2 DOF C1 | 141.97 (±64.99) | 14.0 (±5.5) | 50.3 (±4.2) | - | - | - | |
90 | 90 | 2 DOF C2 | 114.83 (±57.11) | 14.5 (±6.9) | 47.6 (±7.6) | 60.83 (±18.87) | 26.9 (±9.5) | 47.1 (±6.0) | |
90 | 90 | 2 DOF C3 | 82.93 (±61.25) | 17.8 (±8.2) | 48.1 (±5.4) | 107.30 (±22.41) | 16.6 (±4.2) | 49.6 (±3.9) |
fpeak, tass, tppo mean magnitude of peak force, assistance timing and positive power onset, respectively. And the values in parentheses indicate standard deviations.
The 1-DOF powered ankle-foot orthosis
We used a commonly used phase-based controller (P-bc), known as an ankle positive power inspired technique for a 1-DOF PAFO [8]. The P-bc is an assistance technique commonly used with PAFOs that have a PAM as an actuator to maximize the duty ratio of the solenoid valve at a certain timing to provide intensive assistance in the period when the ankle power is positive during the gait cycle [30–33]. The period with positive values occurs when the ankle joint performs plantar flexion. We used the FSR sensor to determine the time point. Among experiments of PAFOs with PAMs, the onset timing of positive power was between 13% and 54% of the gait cycle [33], but high effectiveness of assistance was most frequently achieved between 40% and 50% [32, 34, 35]. We used the stage of the gait cycle, detected by the FSR sensor, to provide 3 duty ratios (30%, 60%, and 90%) at the heel strike and then stopped the pulse at the toe-off time. This results in an onset timing of positive power between 40% and 50% for all the cases are shown in Table 1, which also contains the magnitudes of peak force and assistance timing for each duty ratio, which is explained in detail in Fig 3.
The 2-DOF powered ankle-foot orthosis
The same P-bc was used for comparison with the 1-DOF PAFO. However, in addition to PAM1, the 2-DOF PAFO has PAM2, which acts as the peroneus longus, allowing for a wider variety of controls. As shown in Table 1, the control methods of the P-bcs for the 2 PAMs are classified. The duty ratio of PAM2 does not exceed the size of PAM1, and in both cases, 3 duty ratios were used, and a total of 8 P-bcs were applied.
Experimental protocol
The number of subjects was determined by referring to the number of previously conducted PAFO experiments [33–36]. A total of 8 healthy men (mean ± SD: height, 174.8 ± 7.1 cm; weight, 69.8 ± 6.5 kg; and age, 29.1 ± 4.8 years) with no history of gait abnormality participated in the experiment, which was approved by the institutional review board of Yonsei University (7001988-202003-HR-833-03). Proper written consents were obtained from the participants before the experiment. They were recruited on the campus of Yonsei University regardless of gender. The participants wore a pressure-sensing insole to measure the COP and an IMU to measure the postural sway of the trunk, in addition to the PAFO, as shown in Fig 1C. The subjects walked on the treadmill at a speed of 1.2 m/s, and various assistance techniques were applied. The assistance techniques were applied to the left foot only, and the right PAFO was used without actuation. As with other exoskeleton experiments, we gave the wearer some time to get used to the robot to allow for reliable experimental data measurements and gave the wearers enough rest time between experiments to avoid fatigue. Specifically, before the experiments, they performed treadmill walking two times within 30 minutes to get used to each robot with assistance [35, 37], and 3 minutes was provided between each set of 50 steps. When the experiments with 1-DOF PAFO was finished and before entering the second experiment with 2-DOF PAFO, they took a 10-minute rest to prevent fatigue. The total walking time was around 2 hours, and all the assistance techniques were evaluated thrice over 50 steps each.
Measurements
Center of pressure
We measured the COP using a pressure-sensing insole (MP2512PLUS; Kitronyx Inc, Seoul, Korea) that can be worn inside the shoe and obtained the data at a frequency of 50 Hz. The data were normalized by the gait cycle using the FSR foot sensor, and the average was calculated and represented as a single trajectory. After this, we calculated the average COP position (COPAvg), which is the average of its medial-lateral displacement values as in Fig 4.
Eversion angle
After measuring the angle change during the gait cycle from the rotary encoder attached to the subtalar joint of the PAFO, the average (θAvg) was calculated to determine the degree of quantitative eversion.
Positive power of the PAM
By substituting the rotation angle of the 2 joints into the Rodrigues’ rotation formula used by Choi et al., the change in the length of the PAM can be calculated. The Rodrigues’ rotation formula determines the location at which a point is rotated at an angle with respect to an arbitrary straight line in space [38]. The contraction rate of the PAM can be calculated from the change in length, and the power can be obtained by multiplying the magnitude of the force obtained from the load cell. Normally, the average value of positive power applied to the human body during the gait cycle is proportional to the decrease in electromyography (EMG) activity or metabolic cost [39]. The average positive power is the same as the work applied to the human body. Thus, to compare the difference in the work applied to the human body, we also calculated and showed the average value of the positive area (PAvg) in the power curve for each experiment.
Stabilograms
The stabilograms obtained in this study consisted of the angular acceleration values of the IMU attached to the trunk. According to Kang et al. [17], maintaining the balance of the upper body is important to avoid falling during any form of movement, including walking, and this balance is determined by the movement of the lower limb. Therefore, as in other studies, because the trunk has the smallest change in external perturbation and represents the COM of the whole body, the IMU was attached to it, and stability was judged on the basis of the detected movement [17, 19]. The values obtained from the IMU were converted into Euler angles using a low-pass filter and a complementary filter. We used the Newton difference quotient with backward differences to convert the angles of the roll (medial/lateral) and pitch (anterior/posterior) to angular accelerations. These were represented as a stabilogram on a two-dimensional (2-D) plot, as shown by Seimetz et al. So far, studies have used the maximum radius to compare these stabilograms [40]. However, this comparison is not reliable, and we need stochastic values. Therefore, we applied the Gaussian ellipsoid [41]. As shown in Fig 5, we can calculate the covariance of 2-D angular acceleration data and draw an ellipsoid with vertices of eigenvalue, which means 2 standard deviations. Usually, 1δ, 2δ, and 3δ are used to represent 3 Gaussian ellipsoids containing 39%, 68%, and 99% of the data, respectively. This enables a quantitative and reliable comparison of stabilograms. We compared the overall stability by calculating the root mean square values (δRMS) with two 1δs representing the roll and pitch directions, respectively. So, if stability deteriorated, then the δRMS will be bigger than before.
Statistical analyses
We compared the average COP position (COPAvg), average eversion angle (θAvg), stabilogram (δRMS), and average positive power (PAvg), and calculated the average and standard deviation for each experimental parameter in Table 2. In the case of COPAvg, the data were expanded by interpolating 100 data recorded during the stance phase to increase reliability. The rest of the data were also normalized to the 100 data for the entire gait cycle and compared. To compare the above-mentioned results according to the DOF and assistance techniques, the homogeneity and heteroscedasticity of variance were judged using the f-test between a total of 12 test results, and the difference between the data was verified using a t test for each case. The p value, the result of the t test, is shown for easy comparison between the experiments through the p-table. For comparison between the 12 cases, δRMS was converted to the rate with respect to the unpowered condition with the 2-DOF PAFO, and the average positive power was divided by the participant’s weight. The average eversion angle and COP position were used as they were and included in the S1 Dataset. The level of significance was determined on the basis of the false discovery rate (FDR) [42]. In addition, Pearson correlation coefficients (PCCs) were calculated between PAvg−θAvg, θAvg−COPAvg, COPAvg−δRMS, and PAvg−COPAvg.
Table 2. Experimental results of all the participants.
1 DOF | 2 DOF | Unit | |||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
UNP | A | B | C | UNP | A1 | A2 | B1 | B2 | C1 | C2 | C3 | ||||
δRMS | Avg. | 8.055 | 8.306 | 8.583 | 8.754 | 7.339 | 7.057 | 6.672 | 7.510 | 7.037 | 6.957 | 6.892 | 6.754 | ||
Std. | 1.051 | 1.372 | 1.177 | 1.234 | 1.166 | 1.049 | 0.752 | 1.089 | 0.908 | 0.765 | 0.779 | 0.879 | |||
COPAvg | Avg. | 51.433 | 50.688 | 49.172 | 48.731 | 52.913 | 52.827 | 53.267 | 52.882 | 53.533 | 53.151 | 53.066 | 53.623 | mm | |
Std. | 1.571 | 1.843 | 2.643 | 2.756 | 1.928 | 2.078 | 1.717 | 1.719 | 1.309 | 1.766 | 2.017 | 1.810 | |||
θAvg | Avg. | 0.160 | 0.356 | 1.339 | -0.070 | 1.577 | 0.549 | 1.756 | 2.640 | θ | |||||
Std. | 1.321 | 1.830 | 1.956 | 1.959 | 1.651 | 1.523 | 1.201 | 1.052 | |||||||
PAvg | (1) | Avg. | 0.027 | 0.041 | 0.055 | 0.022 | 0.009 | 0.026 | 0.012 | 0.029 | 0.020 | 0.014 | W/kg | ||
Std. | 0.015 | 0.015 | 0.022 | 0.007 | 0.005 | 0.010 | 0.008 | 0.015 | 0.015 | 0.014 | |||||
(2) | Avg. | 0.019 | 0.015 | 0.013 | 0.023 | ||||||||||
Std. | 0.009 | 0.007 | 0.005 | 0.008 |
The bigger the magnitude of δRMS, the much worse the stability is. And the direction of COPAvg is that when its magnitude increase, it goes toward medial direction. Likewise, for θAvg, the bigger value means that eversion was enhanced.
Results
Center of pressure
The results of the experiments with the 1-DOF PAFO show that the COP trajectories were generally more localized in the lateral direction than those in the experiments with the 2-DOF PAFOs (Table 2). This difference was statistically significant, as shown in Fig 6. Regarding the COPAvg, we found different trends between the 1- and 2-DOF PAFOs. In the 1-DOF PAFO experiment, the larger the duty ratio applied, the more the COP tended to move toward the lateral direction (1-DOF UNP: 51.433, 1-DOF A: 50.688, 1-DOF B: 49.172, 1-DOF C: 48.731). In the case of the 2-DOF PAFO, no significant difference was found between the experiment results, as shown in Table 2 and Fig 6.
Eversion angle
As eversion is a motion only available in the 2-DOF PAFO, we could only obtain data from experiments using the 2-DOF PAFO. We found that the eversion angle was larger in the experiments with the duty ratio of PAM2 (2-DOF A2, 2-DOF B2, 2-DOF C2, and 2-DOF C3) than in the other experiments (eversion angle sheet of the S1 Dataset).
Positive power of PAM
The duty ratio and applied PAvg showed a proportional relationship in PAM1, regardless of the DOF of PAFO. However, in the case of PAM2, the PAvg was determined by the presence or absence of the duty ratio; however, no correlation was observed. For the 2-DOF PAFO, as PAM2 generates the force together with PAM1, PAM1 has a lower power than for the 1-DOF PAFO, even if the duty ratios are the same (average positive power sheet of the S1 Dataset).
Stabilogram
The 1-DOF PAFO experiment results showed that the stability in the roll and pitch directions was lower when assistance was generated by applying the duty ratio than that in the unpowered condition (1-DOF UNP). However, in the 2-DOF PAFO, the highest stability was not achieved in the unpowered condition (2-DOF UNP, δRMS = 7.339). No significant differences or smaller values were observed in comparison with the values in the unpowered conditions among the 2-DOF PAFO experiments except in 2-DOF B1. We found a significant difference depending on the presence of the subtalar joint, shown with pink shadow in Fig 7.
Correlation coefficients
Between the experimental results, the Pearson correlation coefficients (PCCs) are calculated and shown in Table 3. They are interpreted by following a guide for use in medical research [43]. COPAvg−δRMS, PAvg−θAvg (2-DOF PAFO) and PAvg−COPAvg (1-DOF PAFO) have a moderate to high relationship with −0.66 ± 0.15, 0.74 ± 0.10 and −0.71 ± 0.16, respectively. However, θAvg−COPAvg (2-DOF PAFO) has a negligible to low correlation of 0.25 ± 0.20, and PAvg−COPAvg (2-DOF PAFO) has no significant relationship between them.
Table 3. Average correlation coefficient.
Pearson (PCC) | Interpretation | |||
---|---|---|---|---|
Avg. | R2 | SD | ||
COP–Stabilo | −0.66 | 0.44 | 0.15 | Moderate to high negative correlation |
Power–Angle (2) | 0.74 | 0.55 | 0.10 | Moderate to high positive correlation |
Angle–COP (2) | 0.25 | 0.06 | 0.20 | Negligible to low positive correlation |
Power–COP (1) | −0.71 | 0.51 | 0.16 | Moderate to high negative correlation |
Power–COP (2) | 0.18 | 0.03 | 0.18 | Negligible correlation |
Interpretation of correlation was based on the guide for use in medical research [43].
Discussion
In this study, we applied various assistance techniques for PAFOs with 1- and 2-DOF and compared their stabilities when the direction and magnitude of force applied to the wearer was changed. As a result, we found that 2-DOF PAFOs capable of eversion motion with a subtalar joint ensure high stability.
Examination of the stabilogram revealed that most experiments performed with 1-DOF PAFOs demonstrated lower stability (δRMS) than those with 2-DOF PAFOs, as shown in Table 2 and Fig 7. As shown in Fig 8C, wearing the 1-DOF PAFO in the unpowered condition fundamentally provided lower stability than the 2-DOF PAFO, with a difference of 8.89% (Table 2). This comparison is not accurate without taking into account the applied PAvg; however, the difference in δRMS between the experiments with the highest stability among the 2-DOF PAFO experiments (2-DOF A2) and those with the highest stability among the 1-DOF PAFO experiments was 23.78%. As in Fig 7, the comparison of the data between the 1- and 2-DOF PAFO experiments could be achieved on the basis of data with pink shadow, which are significantly different. Notable results were obtained when each DOF was considered separately. In the case of the 1-DOF PAFO experiments, shown in Table 2, the stability worsened when a larger PAvg was applied to the wearer using a larger duty ratio for PAM1. The data with yellow shadow in Fig 7, which represents comparisons between the 1-DOF PAFO experiments, showed a significant difference in 1-DOF B and C as compared with those in the unpowered condition. We used P-bcs as in the other PAFO experiments, and the PAvg applied to the wearer resulted in a reduction in metabolic cost and EMG activity of the muscles involved in ankle movement. Therefore, wearing a 1-DOF PAFO may have some advantages; however, the negative correlation implies that it does not improve stability. Therefore, a stable assistance for balanced walking, which prevent wearers from falls because of the absence of a subtalar joint, could be difficult to achieve.
By contrast, in the case of 2-DOF PAFO, a high PAvg does not cause deterioration of stability and shows no obvious tendencies. Notably, the unpowered condition is also worse than the other cases, as shown in Table 2, and when PAM2 generated force together with PAM1 (2-DOF A1, 2-DOF B1, and 2-DOF C1), the stability was higher in both the roll and pitch directions, regardless of the magnitude of PAvg when compared with the situations in which only PAM1 was actuated. Therefore, applying a high PAvg to the wearer, regardless of DOF, does not reduce the stability; however, other confounding factors may exist.
We analyzed the COPAvg for this purpose. Previous studies showed that indexes based on the COP trajectory play significant roles in the assessment of gait stability, such as the distance between the anteroposterior and mediolateral displacements of the COP, velocity, frequency, or other dimensionless values [44–46]. These values help interpret the balance during gait because the location of the COP under each foot represent the neural control of the muscles related to the ankle joint [47], status of the postural sway during one-leg stance [22], and adjustment of ankle joint moments to change the COM position of the whole body [48]. Among them, the mediolateral movement of COP in the frontal plane can help predict the postural sway in one-leg stance during the stance phase and the risk of injurious falls [44, 49–51]. Thus, we converted the COP trajectory within a sole into the mean value of the mediolateral position during the stance phase, as shown in Table 2.
The result of the experiment conducted with 1-DOF PAFO shows that the greater the PAvg applied, the more the COPAvg was located in the lateral position as compared with that in the unpowered condition. In general, when the trajectory of the COP is located in the lateral direction, the extrapolated COP also increases, which reduces the walking stability [25, 52]. Therefore, increasing the duty ratio at 1-DOF PAFO lowers the stability because the COPAvg is located in a more lateral position. The COPAvg for the 2-DOF PAFO experiments was small but showed a similar trend as those in the 1-DOF PAFO experiments. When actuating PAM2 with PAM1 as mentioned earlier, the stability was higher. The COPAvg was located in a more medial direction. Therefore, we conclude that even when wearing PAFOs, keeping the COP in the medial position can help maintain high stability of the individual. The PCC of COPAvg and δEuclid were −0.6607, supporting our conclusion with strong negative correlations, as shown in Table 3. In addition, the blue and gray dots in Fig 9A represent the results of the 1- and 2-DOF PAFO experiments, respectively.
This raises the question of how PAvg affects stability. We conclude that this effect is caused by the determination of the location of the COP. We calculated the PCC between the PAvg and COPAvg by using the results of the 1-DOF PAFO experiments and demonstrated a strong negative correlation of −0.7149. Therefore, the greater the force, the lower the COP will be in the lateral direction. However, the result differed when the 2-DOF PAFO was used.
The difference between the 1- and 2-DOF PAFOs is the presence of a subtalar joint. Therefore, the eversion angle must be evaluated. We added and actuated PAM2 to enhance the eversion motion and found that when a higher duty ratio was applied to PAM2, its size increased (2-DOF A2, 2-DOF B2, 2-DOF C2, and 2-DOF C3), and the eversion motion became stronger, as shown in Table 2. Moreover, the higher the duty ratio applied, the larger is the PAvg applied by PAM2 to the wearer. We can therefore infer that the greater power PAM2 exerts, the more the eversion can be pulled outward. The PCC between PAvg and θAvg was 0.7441, demonstrating a strong positive correlation (Table 3). Generally, the greater the eversion, the more the COP enters the medial direction. We also calculated the correlation between the two (θAvg−COPAvg), with a value of 0.2480 showing no significant relationship. However, as shown in Table 2, the COP of 2-DOF A2, 2-DOF B2, 2-DOF C2, and 2-DOF C3, in which PAM2 was actuated, moved more in the lateral direction than those of 2-DOF A1, 2-DOF B1, and 2-DOF C1, in which PAM2 was not actuated. From this, we can predict that the COP were not related to each other. However, in Fig 7, we can observe significantly reduced δRMS values between 2-DOF A1–2-DOF A2 (p = 1.37e-5) and 2-DOF B1–2-DOF B2 (p = 2.90e-2), which suggest that PAM2 affects stability. Therefore, as PAM2 was actuated, the eversion motion was stronger, which means that the COP did not move out in the lateral direction and, as a result, did not lead to the deterioration of stability, even though the wearer received assistance.
While PAM2 plays a key role in preserving COP in a stable state through eversion, the primary role of PAM1 is to assist plantar flexion. Unlike in the case of 1-DOF PAFO, the stability did not deteriorate despite the increase in the amount of work that the PAM1 of the 2-DOF PAFO exerted on the wearer. This proves that a high stability can be maintained even if the assistance of the wearer is increased by the addition of DOF and change in the direction of the resultant force to the foot, which is our assumption. However, other existing studies have used 1-DOF PAFO, and in this case, as our study revealed that the greater PAvg applied to the wearer resulted in lower stability of the wearer. Our experiments were conducted in the same environment, and even in this situation, if the differences in DOF and force direction cause such a large change in stability, they are expected to have greater effects in situations involving rough terrains, slopes, and stairs, which require walking outside the laboratory.
The experiments and results of this study have some limitations. First is that we did not measure EMG activity and metabolic cost. These parameters have primarily been analyzed for experimental validation in previous studies that used 1-DOF PAFOs, whose research focus was the effectiveness of assistance. However, in the field of rehabilitation, those values are not important. As rehabilitation involves training to return to a normal state, subjects who use PAFOs for rehabilitation are patients who wanted to walk normally. These would include patients who have to relearn their walking skills and proprioception, which were lost or deteriorated by accidents, diseases, or aging. As mentioned in the Introduction section, the stabilizing moment is essential in the elderly for training their gait skills so as not to lose their stability [53]. Many exercises can aid in balance training, including the one-leg stance, walking on a narrow path, or weight shifting. In these exercises, PAFOs can help subjects train their evertors and increase the range of motion of eversion by controlling the magnitude and direction of the stabilizing moment in stages [54–56]. After repetition of training with PAFO, subjects can achieve proprioception of their foot during various motions. Hence, the results of this study can be considered meaningful as they show that a 2-DOF PAFO is more suitable for stable assistance.
The second limitation is that we did not measure the temporal parameters during experiments such as step width, stride length, and time. These parameters can affect postural sway, balance, and gait stability, so previous studies used them to evaluate gait quality. For example, the margin of stability (MOS) and extrapolated COM, which are related to balance, require the step width to be calculated [57]. However, an in-shoe pressure sensor was used in this paper, so those parameters cannot be determined. Although the COP trajectory in the foot, which was used in this study, is an important factor for attaining gait stability during one-leg stance, the effect of the parameter is as significant as that of gait stability itself, so an accurate analysis is not possible without it. In the Results section, θAvg and COPAvg have a negligible or low positive relationship (0.25 ± 0.20), which cannot be easily explained because when the eversion motion was enhanced, the COP should be in a more medial position normally. This situation can be interpreted when temporal parameters are measured. We can expect some possible answers for the weak relationship similar to when the eversion motion was enhanced; if the step width is wider than before, then the COPAvg cannot be in a medial position as much as expected. Thus, these can help us analyze the series of processes in the function of the PAFO in maintaining the wearer’s stability in more detail.
The third limitation is that all the participants were healthy and had no need for rehabilitation. Elderly people and patients with abnormalities in their walking skills or body elements are the potential wearers of 2-DOF PAFOs when these are applied in rehabilitation. Thus, experiments in these people should be conducted following this research to validate the functions of the 2-DOF PAFO because they may respond differently with 2-DOF PAFOs, although the presence of a subtalar joint and its actuation were proven to be useful with healthy people. For the rehabilitation of people in weakened and unstable states, the optimal design and controller of the assistive device for each participant will be needed, which were not realized in this paper. Two axes of the 2-DOF PAFO were calculated on the basis of the average values of the anatomical data from cadavers [10], so they are not perfect for various sizes of participants. Of course, the existence of a subtalar joint was proved, and its actuation can provide the wearer more effective assistance in terms of stability than the existing 1-DOF PAFO, but the outcome can be improved with adjustment of the misalignment achieved. Solving a misalignment between the joints of the wearable robots and the wearers is difficult, but many research studies have attempted to find solutions. A 4-bar linkage for the knee joint is a representative example. We are now considering this mechanism and the Schmidt coupling joint or something else to solve the problem. This might be the focus of our future work. For the controller, we used P-bc for the PAM actuation, but it is not a notable and adjustable method because it does not conduct a tracking control of the predefined or real-time force profile. It induces the maximum contraction of the PAM with given pressure, generating a force that varies with the circumstances at that time. Although a certain duty ratio can generate similar magnitudes of force during experiments, it cannot be adjustable for various sizes and characteristics of wearers and circumstances. The benefit of PAM2 was proved by experiments, but the existence of PAM2 was newly introduced in this paper, so its optimal controller or force profile has not been developed yet, which can be also a direction for our future work. To develop a controller for PAM2, the relationship of the 2 PAMs should be determined because their effects on the moment and rotation of 2 joints are coupled as shown in Fig 2. However, this is not easily solved because the free body diagram of the ankle joints is too complex owing to their structure, stiffness, and force translation model. Thus, some solutions for the free body diagram can be used regardless of difficulty and obtain the exact effect of the portions of the 2 PAMs on each joint or develop a new controller without them. New methods can be created by locating the COP in the medial position as much as possible or using the chaos theory and maximum Lyapunov exponents as objective variables for optimization such as human-in-the-loop [58–60]. In addition, our results can be used for assistive PAFOs to identify the optimal torque profile to reduce the metabolic cost or EMG, which were not measured in this study, and simultaneously secure the gait stability. This indicates that assistance effectiveness can be obtained and PAFOs can also achieve its fundamental purpose.
Conclusion
We measured the stability of the wearer by applying assistance techniques with various characteristics to 1- and 2-DOF PAFOs and found that the presence of an additional DOF and the direction of force affected stability, regardless of assistance efficiency. Unlike the 1-DOF PAFO, the 2-DOF PAFO has a subtalar joint as the axis of rotation, which enables the eversion motion to preserve COPAvg in the medial position without being affected by the amount of assistance. The location of the COPAvg was directly related to the stability of the trunk, the most inferior segment; thus, stability was also preserved. By contrast, the 1-DOF PAFO, which only has the talocrural joint as the axis of rotation, was unable to alter the eversion. As the applied PAvg increased, the COP moved out in the lateral direction. Consequently, the stability of the trunk deteriorated. This demonstrates the importance of subtalar joints in the use of PAFOs. Further studies are required to prove their importance as indicators of stability and efficiency of assistance.
Supporting information
Data Availability
All relevant data are within the manuscript and its Supporting Information files.
Funding Statement
This study is based on work supported by a grant that Y.S.B received from the National Research Foundation of Korea (NRF) funded by the Korea Government (MSIP) (No. NRF-2017M1A3A3A02016507)(www.nrf.re.kr). The funders had no role in study design, data collection, and analysis, decision to publish, or preparation of the manuscript. And there was no additional external funding received for this study.
References
- 1.Yang CJ, Zhang JF, Chen Y, Dong YM and Zhang Y. A Review of exoskeleton-type systems and their key technologies. Proceedings of the Institution of Mechanical Engineers, Part C: Journal of Mechanical Engineering Science. 2008; 222: 1599–1612. [Google Scholar]
- 2.Farris RJ, Quintero HA and Goldfarb M. Preliminary evaluation of a powered lower limb orthosis to aid walking in paraplegic individuals. IEEE Trans Neural Syst Rehabil Eng. 2011; 19: 652–9. 10.1109/TNSRE.2011.2163083 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Lünenburger L, Colombo G, Riener R and Dietz V. Clinical assessments performed during robotic rehabilitation by the gait training robot Lokomat. In Proceedings of 9th International Conference on Rehabilitation Robotics, Chicago, IL, USA; 2005.
- 4.Esquenazi A, Talaty M, Packel A and Saulino M. The ReWalk powered exoskeleton to restore ambulatory function to individuals with thoracic-level motor-complete spinal cord injury. Am J Phys Med Rehabil. 2012; 91: 911–21. 10.1097/PHM.0b013e318269d9a3 [DOI] [PubMed] [Google Scholar]
- 5.Kawamoto H and Sanka Y. Power Assist Method for HAL-3 using EMG-based Feedback Controller. In Proceedings of 2003 IEEE International Conference on Systems, Man and Cybernetics, Washington, DC, USA. 2003.
- 6.Zoss AB, Kazerooni H and Chu A. Biomechanical design of the Berkeley lower extremity exoskeleton (BLEEX). IEEE/ASME Transactions on Mechatronics. 2006; 11: 128–138. [Google Scholar]
- 7.Choi J, Na B, Jung PG, Rha DW and Kong K. WalkON Suit: A Medalist in the Powered Exoskeleton Race of Cybathlon 2016. IEEE Robotics & Automation Magazine. 2017; 24: 75–86. [Google Scholar]
- 8.Moltedo M, Bacek T, Verstraten T, Rodriguez-Guerrero C, Vanderborght B and Lefeber D. Powered ankle-foot orthoses: the effects of the assistance on healthy and impaired users while walking. J Neuroeng Rehabil. 2018; 15: 86 10.1186/s12984-018-0424-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Matsusaka N. Control of the medial-lateral balance in walking. Acta Orthop Scand. 1986; 57: 555–9. 10.3109/17453678609014793 [DOI] [PubMed] [Google Scholar]
- 10.Isman RE and Inman VT. ANTHROPOMETRIC STUDEIS OF THE HUMAN FOOT AND ANKLE. Bulletin of Prosthetics Research. 1969; 97–129. [Google Scholar]
- 11.Park S and Park J. The Relationship between Dynamic Balance Measures and Center of Pressure Displacement Time in Older Adults during an Obstacle Crossing. The Journal of Korean Society of Physical Therapy. 2011. 23: 1–5. [Google Scholar]
- 12.Galle S, Derave W, Bossuyt F, Calders P, Malcolm P and De Clercq D. Exoskeleton plantarflexion assistance for elderly. Gait Posture. 2017; 52: 183–188. 10.1016/j.gaitpost.2016.11.040 [DOI] [PubMed] [Google Scholar]
- 13.Witte KA, Zhang J, Jackson RW and Collins SH. Design of Two Lightweight, High-Bandwidth Torque-Controlled Ankle Exoskeletons. In proceedings of 2015 IEEE International Conference on Robotics and Automation (ICRA), Seattle, Washington, USA. 2015.
- 14.Kao PC, Lewis CL and Ferris DP. Invariant ankle moment patterns when walking with and without a robotic ankle exoskeleton. J Biomech. 2010; 43: 203–9. 10.1016/j.jbiomech.2009.09.030 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.LeVeau BF. Biomechanics of Human Motion: Basics and Beyond for the Health Professions. SLACK Incorporated; 2011. [Google Scholar]
- 16.Yim-Chiplis PK, Talbot LA. Defining and Measuring Balance in Adults. BIOLOGICAL RESEARCH FOR NURSING. 2000; 1: 321–331. 10.1177/109980040000100408 [DOI] [PubMed] [Google Scholar]
- 17.Kang HG and Dingwell JB. Dynamic stability of superior vs. inferior segments during walking in young and older adults. Gait Posture. 2009; 30: 260–3. 10.1016/j.gaitpost.2009.05.003 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Whitney SL et al. A comparison of accelerometry and center of pressure measures during computerized dynamic posturography: A measure of balance. Gait and Posture. 2011. 33: 594–9. 10.1016/j.gaitpost.2011.01.015 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Liu J, Zhang X and Lockhart TE. Fall risk assessments based on postural and dynamic stability using inertial measurement unit. Saf Health Work. 2012; 3: 192–8. 10.5491/SHAW.2012.3.3.192 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Howcroft J, Kofman J, Lemaire ED and McIlroy WE. Analysis of dual-task elderly gait in fallers and non-fallers using wearable sensors. J Biomech. 2016; 49: 992–1001. 10.1016/j.jbiomech.2016.01.015 [DOI] [PubMed] [Google Scholar]
- 21.Lin WH, Liu YF, Hsieh CC and Lee AJ. Ankle eversion to inversion strength ratio and static balance control in the dominant and non-dominant limbs of young adults. J Sci Med Sport. 2009; 12: 42–9. 10.1016/j.jsams.2007.10.001 [DOI] [PubMed] [Google Scholar]
- 22.Hoogvliet P, van Duyl WA, de Bakker JV, Mulder PGH and Stam HJ. A model for the relation between the displacement of the ankle and the center of pressure in the frontal plane, during one-leg stance. Gait Posture. 1997; 6: 39–49. [Google Scholar]
- 23.Hoogvliet P. Lower extremity movements in frontal plane balance control during one-leg stance. PhD. thesis, Erasmus University, Rotterdam. 1997.
- 24.Hof AL, Gazendam MG and Sinke WE. The condition for dynamic stability. J Biomech. 2005; 38: 1–8. 10.1016/j.jbiomech.2004.03.025 [DOI] [PubMed] [Google Scholar]
- 25.Hof AL. The 'extrapolated center of mass' concept suggests a simple control of balance in walking. Hum Mov Sci. 2008; 27: 112–25. 10.1016/j.humov.2007.08.003 [DOI] [PubMed] [Google Scholar]
- 26.Bok SK, Lee TH and Lee SS. The effects of changes of ankle strength and range of motion according to aging on balance. Ann Rehabil Med. 2013; 37: 10–6. 10.5535/arm.2013.37.1.10 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Liu J, Lockhart TE. Comparison of 3D joint moments using local and global inverse dynamics approaches among three different age groups. Gait & Posture. 2006; 23: 480–485. 10.1016/j.gaitpost.2005.06.011 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Khairul A and Al-Jumaily AA. Active Exoskeleton Control Systems: State of the Art. Procedia Engineering. 2012; 41: 988–994. [Google Scholar]
- 29.Choi HS, Lee CH and Baek YS. Design of a Pneumatic Actuated Ankle-Foot Orthosis which has Talocrural and Subtalar Joint. In Proceedings of 2019 IEEE 16th International Conference on Rehabilitation Robotics (ICORR), Toronto, ON, Canada. 2019. [DOI] [PubMed]
- 30.Koller JR, Remy CD and Ferris DP. Comparing neural control and mechanically intrinsic control of powered ankle exoskeletons. In Proceedings of 2017 International Conference on Rehabilitation Robotics (ICORR), London, UK. 2017. [DOI] [PubMed]
- 31.Asbeck AT, Rossi SMMD, Holt KGand Walsh CJ. A biologically inspired soft exosuit for walking assistance. The International Journal of Robotics Research. 2015; 34: 744–762. [Google Scholar]
- 32.Lee S, Crea S, Malcolm P, Galiana I, Asbeck A and Walsh C. Controlling Negative and Positive Power at the Ankle with a Soft Exosuit. In Proceedings of 2016 IEEE International Conference on Robotics and Automation (ICRA), Stockholm, Sweden. 2016.
- 33.Grimmer M et al. Comparison of the human-exosuit interaction using ankle moment and ankle positive power inspired walking assistance. J Biomech. 2019; 83: 76–84. 10.1016/j.jbiomech.2018.11.023 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 34.Malcolm P et al. Varying negative work assistance at the ankle with a soft exosuit during loaded walking. Journal of NeuroEngineering and Rehabilitation. 14 10.1186/s12984-017-0267-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Malcolm P, Derave W, Galle S and De Clercq D. A simple exoskeleton that assists plantarflexion can reduce the metabolic cost of human walking. PLoS One. 2013; 8: e56137 10.1371/journal.pone.0056137 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Koller JR, Jacobs DA, Ferris DP and Remy CD. Learning to walk with an adaptive gain proportional myoelectric controller for a robotic ankle exoskeleton. J Neuroeng Rehabil. 2015; 12: 97 10.1186/s12984-015-0086-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Gordon KE, Sawicki GS, Ferris DP. Mechanical performance of artificial pneumatic muscles to power an ankle-foot orthosis. Journal of Biomechanics. 39: 1832–1841. 10.1016/j.jbiomech.2005.05.018 [DOI] [PubMed] [Google Scholar]
- 38.Rodrigues O. Des lois geometriques qui regissent les deplacements d’ un systeme dolide dans l’ espace, et de la variation des coordonnees provenant de ces deplacement considerees independent des causes qui peuvent les produire. JOURNAL DE MATHÉMATIQUES PURES ET APPLIQUÉES. 1840; 5: 380–440. [Google Scholar]
- 39.Galle S, Malcolm P, Collins SH and De Clercq D. Reducing the metabolic cost of walking with an ankle exoskeleton: interaction between actuation timing and power. J Neuroeng Rehabil. 2017; 14: 35 10.1186/s12984-017-0235-0 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Seimetz C, Tan D, Katayama R and Lockhart T. A Comparison Between Methods of Measuring Postural Stability: Force Plates versus Accelerometers. Biomedical Sciences Instrumentation. 2012; 48: 386–92. [PMC free article] [PubMed] [Google Scholar]
- 41.Scrucca L, Fop M, Murphy TB and Raftery AE. mclust 5: Clustering, Classification and Density Estimation Using Gaussian Finite Mixture Models. The R Journal. 2016; 8: 289–317. [PMC free article] [PubMed] [Google Scholar]
- 42.Benjamini Y, Hochberg Y. Controlling the False Discovery Rate: A Practical and Powerful Approach to Multiple Testing. 1995. 57: 289–300. [Google Scholar]
- 43.Mukaka MM. Statistics Corner: A guide to appropriate use of Correlation coefficient in medical research. Malawi Medical Journal. 2012. 24: 69–71. [PMC free article] [PubMed] [Google Scholar]
- 44.Biswas A, Lemaire ED, Kofman J. Dynamic gait stability index based on plantar pressures and fuzzy logic. Journal of Biomechanics. 2008. 41: 1574–1581. 10.1016/j.jbiomech.2008.02.009 [DOI] [PubMed] [Google Scholar]
- 45.Piirtola M, Era P. Force Platform Measurements as Predictors of Falls among Older People–A Review. Gerontology. 2006. 52: 1–16. 10.1159/000089820 [DOI] [PubMed] [Google Scholar]
- 46.Qiu H, Xiong S. Center-of-pressure based postural sway measures: Reliability and ability to distinguish between age, fear of falling and fall history. International Journal of Industrial Ergonomics. 2015. 47: 37–44. [Google Scholar]
- 47.Winter DA. Human balance and posture control during standing and walking. Gait and Posture. 1995. 3: 193–214. [Google Scholar]
- 48.Maki BE. Biomechanical approach to quantifying anticipatory postural adjustments in the elderly. Med. & Biol. Eng. & Comput. 1993. 31: 355–362. 10.1007/BF02446688 [DOI] [PubMed] [Google Scholar]
- 49.Bergland A and Wyller TB. Risk factors for serious fall related injury in elderly women living at home. Inj Prev. 2004. 10: 308–313. 10.1136/ip.2003.004721 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 50.Bergland A, Jarnlo GB, Laake K. Predictors of falls in the elderly by location. Aging Clinical and Experimental Research. 2003. 15: 43–50. 10.1007/BF03324479 [DOI] [PubMed] [Google Scholar]
- 51.Mizelle C, Rodgers M, Forrester L. Bilateral foot center of pressure measures predict hemiparetic gait velocity. Gait and Posture. 2006. 24: 356–363. 10.1016/j.gaitpost.2005.11.003 [DOI] [PubMed] [Google Scholar]
- 52.Hof AL, Vermerris SM and Gjaltema WA. Balance responses to lateral perturbations in human treadmill walking. J Exp Biol. 2010; 213: 2655–64. 10.1242/jeb.042572 [DOI] [PubMed] [Google Scholar]
- 53.Willems T, Witvrouw E, Verstuyft J, Vaes P, De Clercq D. Proprioception and muscle strength in subjects with a history of ankle sprains and chronic instability. Journal of Athletic Training. 2002; 37: 487–93. [PMC free article] [PubMed] [Google Scholar]
- 54.Madureira MM, Takayama L, Gallinaro AL, Caparbo VF, Costa RA, Pereira RM. Balance training program is highly effective in improving functional status and reducing the risk of falls in elderly women with osteoporosis: a randomized controlled trial. Osteoporos Int. 2007; 18: 419–25. 10.1007/s00198-006-0252-5 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 55.van Diest M, Lamoth CJ, Stegenga J, Verkerke GJ, Postema K. Exergaming for balance training of elderly: state of the art and future developments. J Neuroeng Rehabil. 2013; 10: 101 10.1186/1743-0003-10-101 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 56.Silsupadol P, Lugade V, Shumway-Cook A, van Donkelaar P, Chou LS, Mayr U, et al. Training-related changes in dual-task walking performance of elderly persons with balance impairment: a double-blind, randomized controlled trial. Gait Posture. 2009; 29: 634–9. 10.1016/j.gaitpost.2009.01.006 [DOI] [PMC free article] [PubMed] [Google Scholar]
- 57.Kim M, Lyness H, Chen T, Collins SH. The Effects of Prosthesis Inversion/Eversion Stiffness on Balance-Related Variability During Level Walking: A Pilot Study. Journal of Biomechanical Engineering. 2020; 142. [DOI] [PubMed] [Google Scholar]
- 58.Wolf A, Swift JB, Swinney HL, Vastano JA. Determining Lyapunov exponents from a time series. Physica D: Nonlinear Phenomena. 1985; 16: 285–317. [Google Scholar]
- 59.Ihlen EAF, Van Schooten KS, Brujin SM, Pijnappels M, Can Dieen JH. Fractional Stability of Trunk Acceleration Dynamics of Daily-Life Walking: Toward a Unified Concept of Gait Stability. Frontiers in Physiology. 2017;8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 60.Zhang J, Fiers P, Witte KA, Jackson RW, Poggensee KL, Atkeson CG, et al. Human-in-the-loop optimization of exoskeleton assistance during walking. Science. 2017; 356: 1280–84. 10.1126/science.aal5054 [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All relevant data are within the manuscript and its Supporting Information files.