Abstract
Background
The role of drug‐coated balloons (DCBs) in the treatment of de novo coronary lesions is not well established.
Methods and Results
Electronic databases and major conference proceedings were searched for randomized controlled trials that compared DCBs with stents or angioplasty for de novo coronary lesions. The primary outcome was target lesion revascularization. Summary estimates were conducted using random‐effects analysis complemented by several subgroup and sensitivity analyses. A total of 14 randomized controlled trials with 2483 patients were included. At a mean follow up of 12 months, DCBs were associated with no difference in the incidence of target lesion revascularization as compared with alternative strategies (risk ratio [RR], 0.79; 95% CI, 0.35–1.76). There was no difference in treatment effect based on the indication (ie, small‐vessel disease, myocardial infarction, bifurcation, or high bleeding risk) (P interaction=0.22). DCBs were associated with lower target lesion revascularization compared with bare metal stents and similar target lesion revascularization compared with drug‐eluting stents (P interaction=0.03). There was no difference between DCBs and control in terms of major adverse cardiac events, vessel thrombosis, or cardiovascular mortality. However, DCBs were associated with a lower incidence of myocardial infarction (RR, 0.48; 95% CI, 0.25–0.90) and all‐cause mortality (RR, 0.45; 95% CI, 0.22–0.94).
Conclusions
In patients with de novo coronary lesions, use of DCBs was associated with comparable clinical outcomes irrespective of the indication or comparator device. DCBs had a similar rate of target lesion revascularization compared with drug‐eluting stents. A randomized trial powered for clinical outcomes and evaluating the role of DCBs for all‐comers is warranted.
Keywords: coronary artery disease, de novo lesions, drug‐eluting stent, drug‐coated balloon, meta‐analysis, mortality, small vessels
Subject Categories: Catheter-Based Coronary and Valvular Interventions
Nonstandard Abbreviations and Acronyms
- BMS
bare metal stent
- DCB
drug‐coated balloon
- DES
drug‐eluting stents
- MLD
minimum lumen diameter
- MI
myocardial infarction
- TLR
target lesion revascularization
Clinical Perspective
What Is New?
In patients with de novo coronary lesions, drug‐coated balloons were associated with comparable clinical outcomes irrespective of the indication or comparator device.
Drug‐coated balloons had a similar rate of target lesion revascularization compared with drug‐eluting stents.
What Are the Clinical Implications?
These findings suggest the value of drug‐coated balloons as an attractive “leave‐nothing‐behind strategy” for selected patients with de novo coronary lesions provided a satisfactory result is obtained after lesion predilation.
A randomized trial powered for clinical outcomes and evaluating the role of drug‐coated balloons for all‐comers is warranted.
Drug‐eluting stents (DESs), particularly second‐generation, remain the cornerstone management during percutaneous coronary intervention.1 Coronary restenosis as a result of the persistence of the metallic struts within the vessel as well as the need for dual antiplatelet therapy remain major limitations even with the current generation of DESs.2, 3 In this context, drug‐coated balloons (DCBs) offer an attractive therapeutic modality because these devices allow for local delivery of the antiproliferative agent directly into the artery wall with a single balloon inflation without the need for the metallic implant.4 Several randomized trials have established the role of DCBs in treatment of in‐stent restenosis of both DESs and bare metal stents (BMSs),5, 6, 7, 8 and the use of DCBs is currently endorsed by the 2018 European Society of Cardiology guidelines for myocardial revascularization as a class I recommendation for this indication.9
However, the role of DCBs is not as established for de novo coronary lesions.4 Recently, several small‐to‐moderate–sized, randomized trials have evaluated the merits of DCBs for patients with small‐vessel disease,10, 11 high risk of bleeding,12 and myocardial infarction (MI).13, 14 However, most of these individual trials were not powered to assess the differences in clinical outcomes.10, 13, 14 Moreover, the trials that were powered for clinical outcomes were noninferiority trials and did not routinely evaluate angiographic outcomes.11, 12, 13 To address this knowledge gap, we performed a comprehensive systematic review and meta‐analysis of randomized trials to evaluate the impact of DCBs for de novo coronary lesions on angiographic and clinical outcomes.
Methods
The authors declare that all supporting data are available within the article (and in the accompanying supplementary material online).
Data Sources and Search Strategy
Electronic databases, including MEDLINE, Embase, and the Cochrane Register of Controlled Trials, as well as major scientific sessions, were searched without language restriction from inception through November 2019 using the search algorithm in Table S1. The bibliography of the retrieved articles was reviewed. The search was independently performed by 2 authors (I.Y.E., F.A.). The protocol for this meta‐analysis was prospectively registered at the PROSPERO international prospective register of systematic reviews (CRD42019143329),15 and was conducted according to the Preferred Reporting Items for Systematic Reviews and Meta‐Analyses guidelines.16
Selection Criteria and Data Extraction
Trials that randomized patients with obstructive de novo coronary lesions to DCBs versus any comparator were included (ie, DES, BMS, angioplasty only). We excluded trials that electively performed routine BMS placement after DCBs, but included trials that permitted bailout stent placement after DCBs. Clinical and angiographic data from the longest available reported follow‐up time were preferentially used. Observational studies were excluded for inherent risk of bias. Two independent authors (I.Y.E., A.Y.E.) extracted data on study design, sample size, intervention strategies, outcomes, and other study characteristics from the included studies. Discrepancies were resolved by consensus.
Assessment of Quality of Included Studies
The Cochrane Collaboration's tool was used for the assessment of the risk of bias. This consists of 7 points that test for selection, performance, detection, attrition, reporting, and other biases.17 Performance bias (ie, blinding of participants and physicians) was found to be irrelevant due to the interventional nature in both arms. The overall risk of bias for each trial was classified as low, unclear, or high risk, based on whether level of bias in each domain could have resulted in biases in risk estimation.
Outcomes
The primary clinical outcome was target lesion revascularization (TLR). The secondary clinical outcomes included: major adverse cardiac events, as defined by the individual trials (Table S2); target vessel revascularization; MI; vessel thrombosis; cardiovascular mortality; and all‐cause mortality. The following angiographic outcomes were assessed: minimum lumen diameter (MLD); diameter stenosis; late lumen loss; and binary restenosis.
Statistical Analysis
Outcomes were evaluated by an intention‐to‐treat analysis. Random‐effects summary risk ratios were primarily estimated with the DerSimonian and Laird model, because we anticipated a high degree of statistical heterogeneity.18 Summary odds ratios were also estimated with a Peto model as a secondary analysis due to the low incidence of events.19 Statistical heterogeneity was assessed using the Cochrane Q and I 2 statistics.20 Egger's method was used to calculate publication bias.21 Standardized mean differences were used for continuous variables. All P‐values were 2‐tailed, with statistical significance set at 0.05, and CIs were calculated at the 95% level for the overall estimates effect. All analyses were performed using the RStudio software meta package (RStudio, Inc, Boston, MA).
The following prespecified subgroup analyses were performed for the primary outcome (TLR): (1) according to indication; and (2) by comparing DESs versus BMSs. In addition, the following prespecified sensitivity analyses for TLR were also conducted by: (1) excluding trials using the first‐generation DCB, which is no longer available22; (2) excluding trials using angioplasty alone in the control arm; (3) limiting to trials utilizing second‐generation DESs as the control; and (4) excluding trials with high risk of bias. Random‐effects meta‐regression analyses for the primary outcome were prespecified in relation to baseline reference vessel diameter, diabetes mellitus, and proportion of bailout stent placement in the DCBs arm.23 Finally, a sensitivity analysis limited to trials using second‐generation DESs as the control was performed for the angiographic outcomes, and a sensitivity analysis limited to trials that defined MI as spontaneous (ie, not procedure‐related) was also conducted.
Results
Included Studies
The systematic search identified 502 studies after removal of the duplicates, among which 37 were reviewed for eligibility. The final number of records included in this meta‐analysis was 14 trials from 15 reports (Figure 1).10, 11, 12, 13, 14, 22, 24, 25, 26, 27, 28, 29, 30, 31, 32 One trial reported angiographic and clinical outcomes at 6 months26 and reported an extended follow‐up for the clinical outcomes at 36 months.27 A total of 2483 patients were included: 1268 in the DCBs group and 1215 in the control group. The indication for DCBs was small‐vessel disease in 5 trials,10, 11, 22, 24, 25, 26, 27 MI in 3 studies,13, 14, 28 high bleeding risk in 2 trials,12, 29 bifurcational lesions in 2 studies,30, 31 and unspecified de novo lesions in 1 study.32 In the bifurcational lesion trials, 1 trial compared “plain old” balloon angioplasty followed by DCB versus plain old balloon angioplasty alone to the main or side branch,30 whereas the other trial randomized patients with bifurcational lesions to a strategy of side‐branch dilation with DCB versus plain old balloon angioplasty.31 The SeQuent Please paclitaxel‐coated balloon was used by most of the included studies (9 of 14). Only 1 trial tested the Dior paclitaxel‐coated balloon, which is no longer available.22 The control group was exclusively second‐generation DES in 6 trials,10, 11, 14, 24, 28, 32 first‐generation DESs in 2 trials,22, 26 BMSs in 2 trials,12, 29 and plain old balloon angioplasty alone in 3 trials.25, 30, 31 In 1 trial, the control was second‐generation DESs or BMSs, and a subgroup analysis was reported for the outcomes based on the stent type.13 The weighted mean reference vessel diameter was 2.5 mm. Table shows the baseline trial characteristics, follow‐up duration, and interventional strategies. Table S3 summarizes the pertinent patient demographics and trial information. Performance bias was unclear in all the trials. One trial was at high risk for detection bias and unclear for allocation bias,32 otherwise the remainder of the trials were considered to be of high quality (Table S4).
Figure 1. Study search diagram.
Summary of how the systematic search was conducted and eligible studies were identified. DCB indicates drug‐coated balloon.
Table 1.
Characteristics, Interventional Strategies, and Follow‐Up of the Included Trials
Trial (Reference No.) | Year | Indication | Drug‐Coated Balloon Type | Control Group | Patients (n) | Clinical Follow‐Up (months) | Angiographic Follow‐Up (months) | Primary Outcome | Reference Vessel Diameter (mm) | Bailout Stenting in DCB Arm (%) |
---|---|---|---|---|---|---|---|---|---|---|
PICCOLETO II24 | 2019 | Small‐vessel disease | Elutax SV/Emperor | Second‐generation DES | 118/114 | 6 | 6 | Late lumen loss | 2.2/2.2 | 6.8 |
RESTORE CVD10 | 2019 | Small‐vessel disease | Restore | Second‐generation DES | 116/114 | 12 | 9 | Diameter stenosis | 2.4/2.4 | 5.2 |
BASKET‐SMALL 211 | 2019 | Small‐vessel disease | SeQuent Please | Second‐generation DES | 382/376 | 12 | NR | MACE | NR | NR |
Funatsu et al25 | 2017 | Small‐vessel disease | SeQuent Please | POBA | 92/41 | 6 | 6 | TVF | 2.0/2.0 | 2.9 |
BELLO26, 27 | 2012/2015 | Small‐vessel disease | IN.PACT Falcon | First‐generation DES | 90/92 | 36 | 6 | Late lumen loss | 2.4/2.4 | 20.2 |
PICCOLETO22 | 2010 | Small‐vessel disease | Dior | First‐generation DES | 29/31 | 9 | 6 | Diameter stenosis | 2.4/2.4 | NR |
PEPCAD NSTEMI13 | 2019 | Myocardial infarction |
SeQuent Please SeQuent Please Neo |
BMS/second‐generation DES | 104/106 | 9 | NR | Target lesion failure | NR | 7.3 |
REVELATION14 | 2019 | Myocardial infarction | Pantera Lux | Second‐generation DES | 60/60 | 9 | 9 | FFR value | 3.3/3.2 | 18.0 |
Gobic et al28 | 2017 | Myocardial infarction | SeQuent Please | Second‐generation DES | 41/37 | 6 | 6 | Late lumen loss | 2.6/3.0 | 7.3 |
Shin et al29 | 2019 | High bleeding risk | SeQuent Please | BMS | 20/20 | 12 | 9 | Late lumen loss | 3.0/3.2 | NR |
DEBUT12 | 2019 | High bleeding risk | SeQuent Please | BMS | 102/106 | 9 | NR | MACE | NR | 2.0 |
PEPCAD‐BIF30 | 2016 | Bifurcational lesion | SeQuent Please | POBA | 32/32 | 9 | 9 | Late lumen loss | 2.4/2.4 | 0 |
BABILON31 | 2014 | Bifurcational lesion | SeQuent Please | POBA | 52/56 | 24 | 9 | Late lumen loss | 2.3/2.3 | 7.8 |
Nishiyama et al32 | 2016 | Unspecified | SeQuent Please | Second‐generation DES | 30/30 | 8 | 8 | Not specified | 2.9/2.7 | 10.0 |
Results are presented as drug‐coated balloon/control. ACS indicates acute coronary syndrome; BABILON, The Paclitaxel‐Coated Balloon in Bifurcated Lesions Trial; BASKET‐SMALL 2, The Basel Kosten Effektivitäts Trial–Drug‐Coated Balloons versus Drug‐eluting Stents in Small Vessel Interventions; BELLO, Balloon Elution and Late Loss Optimization; BMS, bare metal stent; DCB, drug‐coated balloon; DEBUT, Drug‐Eluting Balloon in Stable and Unstable Angina: A Randomized Controlled Non‐Inferiority Trial; DES, drug‐eluting stent; FFR, fractional flow reserve; MACE, major adverse cardiac events; NR, not reported; PEPCAD‐BIF, Drug eluting balloons as stand alone procedure for coronary bifurcational lesions; PEPCAD NSTEMI, Bare Metal Stent Versus Drug Coated Balloon With Provisional Stenting in Non‐ST–Elevation Myocardial Infarction; PICCOLETO, Paclitaxel‐coated balloon versus drug‐eluting stent during PCI of small coronary vessels; PICCOLETO II, Drug Eluting Balloon Efficacy for Small Coronary Vessel Disease Treatment; POBA, “plain old” balloon angioplasty; RESTORE SVD, Assess the Efficacy and Safety of RESTORE Paclitaxel Eluting Balloon Versus RESOLUTE Zotarolimus Eluting Stent for the Treatment of Small Coronary Vessel Disease; REVELATION, Revascularization With Paclitaxel‐Coated Balloon Angioplasty Versus Drug‐Eluting Stenting in Acute Myocardial Infarction; and TVF, target vessel failure.
Angiographic Outcomes
Routine angiographic follow‐up was performed at a weighted mean of 7 (range, 6–9) months. There was no difference between DCBs and control in terms of MLD (1.9 mm versus 2.0 mm; standardized mean difference, −0.13; 95% CI, −0.32 to 0.06; P=0.17), diameter stenosis (28.0% versus 28.1%; standardized mean difference, 0.22, 95% CI, −6.92 to 7.36; P=0.95), and binary restenosis (13.9% versus 16.3%; RR, 0.83; 95% CI, 0.40–1.71; P=0.61). However, DCBs were associated with lower late lumen loss (0.08 mm versus 0.24 mm; standardized mean difference, −0.17; 95% CI, −0.24 to −0.10; P<0.0001) (Figure 2). There was a significant degree of statistical heterogeneity observed for the angiographic outcomes (I 2 ranged from 60% to 94%), which was explained on the sensitivity analysis limited to trials comparing DCBs with second‐generation DESs (I 2=0% for all the outcomes, except for diameter stenosis where I 2=56%). The findings of the sensitivity analysis were consistent with the main analysis for all angiographic outcomes except for a lower MLD with DCBs (Figure S1).
Figure 2. Summary plots for the angiographic outcomes.
The relative size of the data markers indicates weight of sample size from each study. DCB indicates drug‐coated balloon; MD, mean difference; MLD, minimal lumen diameter; and RR, risk ratio.
Target Lesion Revascularization
The weighted mean follow up for the clinical outcomes was 12 (range, 6–36) months. There was no difference in the incidence of TLR with DCBs compared with control (random effects: 4.6% versus 5.1%; RR, 0.79; 95% CI, 0.35–1.76; P=0.56; fixed effects: OR, 0.91; 95% CI, 0.58–1.44; P=0.69) (Figure 3). There was no evidence of publication bias using Egger's test (P=0.45). The outcome was characterized by moderate heterogeneity (I 2=50%; χ2=22.1; P heterogeneity=0.02). DCBs showed similar TLR compared with control, irrespective of the indication (P interaction=0.22) (Figure 4). The incidence of TLR was similar when DCBs compared with DESs (RR, 1.37; 95% CI, 0.62–3.05; I 2=34%), but DCBs were associated with a lower incidence of TLR compared with BMSs (RR, 0.19; 95% CI, 0.04–1.00; I 2=0%) (P interaction=0.03) (Figure 5). The findings of the prespecified sensitivity analyses for TLR were consistent with the overall analysis: (1) excluding trials that utilized the older generation DCBs (RR, 0.76; 95% CI, 0.35–1.65; I 2=43%; χ2=17.6; P heterogeneity=0.06) (Figure S2); (2) excluding trials using angioplasty alone in the control arm (RR, 0.97; 95% CI, 0.42–2.27; I 2=45%; χ2=14.5; P heterogeneity=0.07) (Figure S3); (3) limited to trials utilizing second‐generation DESs as control (RR, 1.65; 95% CI, 0.65–4.34; I 2=0%; χ2=2.9; P heterogeneity=0.57) (Figure S4); and (4) excluding the trial with high risk of bias (RR, 0.97; 95% CI, 0.45–2.12; I 2=52%; χ2=21.0; P heterogeneity=0.02) (Figure S5). Meta‐regression analysis did not identify a difference in the treatment effect based on baseline reference vessel diameter (P=0.81), diabetes mellitus (P=0.37), and proportion of bailout stent placement (P=0.63).
Figure 3. Summary plot for target lesion revascularization.
The relative size of the data markers indicates weight of sample size from each study. DCB indicates drug‐coated balloon; and TLR, target lesion revascularization.
Figure 4. Subgroup analysis for target lesion revascularization according to indication.
The relative size of the data markers indicates weight of sample size from each study. There was no difference in treatment effect according to the different indications (P interaction=0.22). DCB indicates drug‐coated balloon; and TLR, target lesion revascularization.
Figure 5. Subgroup analysis for target lesion revascularization comparing bare metal and drug‐eluting stents.
The relative size of the data markers indicates the weight of the sample size from each study. Drug‐coated balloon use was associated with lower target lesion revascularization compared with bare metal stents and similar target lesion revascularization compared with drug‐eluting stents (P interaction=0.03). DCB indicates drug‐coated balloon; and TLR, target lesion revascularization.
Secondary Clinical Outcomes
Compared with control, DCBs were associated with no difference in the incidence of target vessel revascularization (6.0% versus 5.3%; RR, 1.21; 95% CI, 0.60–2.44; P=0.59; I 2=52%; χ2=8.3; P heterogeneity=0.08), major adverse cardiac events (6.9% versus 9.1%; RR, 0.83; 95% CI, 0.50–1.36; P=0.46; I 2=53%; χ2=23.3; P heterogeneity=0.02), vessel thrombosis (0.3% versus 1.1%; RR, 0.38; 95% CI, 0.13–1.13; P=0.08; I 2=0%; χ2=0.5; P heterogeneity=0.91), and cardiovascular mortality (1.5% versus 1.5%; RR, 0.90; 95% CI, 0.27–3.00; P=0.86; I 2=56%; χ2=6.8; P heterogeneity=0.08). Importantly, DCBs were associated with a lower incidence of all‐cause mortality (1.2% versus 2.9%; RR, 0.45; 95% CI, 0.22–0.94; P=0.03; I 2=0%; χ2=0.78; P heterogeneity=0.85), and MI (1.1% versus 2.9%; RR, 0.48; 95% CI, 0.25–0.90; P=0.02; I 2=0%; χ2=6.2; P heterogeneity=0.62) (Figures 6 and S6 through S11). In the sensitivity analysis limited to trials that defined MI as spontaneous MI, DCBs were associated with lower incidence of spontaneous MI (RR, 0.49; 95% CI, 0.25–0.96; P=0.04; I 2=0%) (Figure S12). There was no evidence of publication bias for any of the secondary clinical outcomes using Egger's test (all P>0.05).
Figure 6. Forest plots for the clinical outcomes evaluated in this meta‐analysis.
For each comparison, boxes and horizontal lines correspond to the respective point estimate and accompanying 95% CI. DCB indicates drug‐coated balloon; MACE, major adverse cardiac events; and TLR, target lesion revascularization.
Discussion
In this meta‐analysis of 14 randomized trials including 2483 patients with de novo coronary lesions undergoing percutaneous coronary intervention irrespective of indication, we documented that DCBs were associated with similar MLD, diameter stenosis, binary restenosis, and lower late lumen loss compared with control on routine angiographic follow up at a mean of 7 months. These findings were similar when DCBs were only compared with second‐generation DESs (except that DCBs were associated with lower MLD). At a mean of 12 months, DCBs were associated with no difference in the incidence of TLR compared with control. This effect was consistent, regardless of indication (ie, small‐vessel disease, high bleeding risk, MI, or bifurcational lesions), and on multiple sensitivity analyses, including comparing DCBs with second‐generation DESs. DCBs were associated with lower risk of TLR compared with BMS. There was a moderate degree of statistical heterogeneity for TLR, which was partly explained by our subgroup analysis comparing DCBs with DESs versus BMSs, and on the sensitivity analysis limited to second‐generation DESs. DCBs were also associated with no difference in the incidence of target vessel revascularization, major adverse cardiac events, vessel thrombosis, and cardiovascular mortality. Importantly, the incidence of all‐cause mortality and MI (even when spontaneous MI was analyzed separately) was lower with DCBs. However, these findings were based on a small number of trials and the number of events was low, and therefore should be only considered as hypothesis‐generating. Altogether, our findings strongly suggest the value of DCBs as an attractive “leave–nothing‐behind strategy” for selected patients with de novo coronary lesions provided a satisfactory result is obtained after lesion predilation.
DCBs offer the advantage of locally delivering the antiproliferative drug without the need for metal struts, thus directly inhibiting the process of neointimal hyperplasia and negative remodeling.4 Although use of DCBs in patients with in‐stent restenosis has been extensively investigated,9 trials evaluating DCBs for de novo lesions have been small and evaluated specific indications. Our meta‐analysis, including the most recent trials, has demonstrated that DCBs were associated with favorable clinical outcomes irrespective of the indication, even when compared with second‐generation DESs. Although most patients undergoing percutaneous coronary intervention are treated with a second‐generation DES,1 BMSs are still used in a minority of patients, such as those with a high risk of bleeding to minimize the duration of antiplatelet therapy. Our meta‐analysis showed that DCBs represent a reasonable therapeutic strategy for this subset of patients.
Second‐generation DESs may not offer an effective therapeutic strategy in small vessels due to the late lumen loss resulting in late in‐stent restenosis.34 In this challenging setting, several randomized trials have shown that DCBs are noninferior to DESs for major adverse cardiac events.10, 11 By significantly increasing the sample size, the current meta‐analysis has extended our knowledge by showing that DCBs are associated with similar TLR compared with any control, including second‐generation DESs. Moreover, our meta‐regression analysis has shown that there was no difference in treatment effect based on the reference vessel diameter.
One meta‐analysis of randomized trials has raised some concerns about late mortality with DCBs for patients with peripheral artery disease.35 That meta‐analysis was subject to several limitations,36 and the late mortality finding was not replicated in several large observational studies and patient‐level meta‐analysis.37, 38 Our meta‐analysis provides some support for the use of DCBs for coronary lesions. However, the lower mortality seen with DCBs in our meta‐analysis should be interpreted with caution given the limited number of studies that evaluated all‐cause mortality and the low number of events.
Previous meta‐analyses addressed use of DCBs for a specific indication, such as small‐vessel disease or bifurcational lesions.39, 40–41 In addition, those meta‐analyses included observational studies, which are prone to ascertainment and selection biases.39, 40–41 Furthermore, those works did not include the results of several recently published and presented trials.10, 13, 14, 24 The present meta‐analysis only included randomized trials and has provided a comprehensive overview of the angiographic and clinical outcomes of DCBs irrespective of indication. In addition, we performed several subgroup and sensitivity analyses to explore the statistical heterogeneity.
Our meta‐analysis has several limitations. First, although all the included studies used a paclitaxel‐coated balloon, there are several pharmacokinetic differences between the devices. For example, one trial used the first‐generation Drior paclitaxel‐coated balloon, which was shown to be inferior in terms of deliverability and is no longer available. Thus, we performed a sensitivity analysis excluding this trial for the primary clinical outcome. Second, there were differences in the core laboratory assessment of the angiographic outcomes across the trials, which could be a source of the significant heterogeneity noted with these outcomes. However, we observed no heterogeneity for most of the angiographic outcomes on the sensitivity analysis comparing DCBs with second‐generation DESs. Third, we noted a moderate degree of statistical heterogeneity for the primary clinical outcome (ie, TLR). We attempted to mitigate this by using a random‐effects model. In addition, we performed multiple subgroup, sensitivity, and meta‐regression analyses to explore the heterogeneity; however, the number of studies included in some of these subgroup and sensitivity analyses was small, so the findings can only be considered as hypothesis‐generating. Fourth, one of the included trials was at high risk for bias,32 so we performed a sensitivity analysis excluding that trial for TLR. Fifth, despite the extensive subgroup, sensitivity, and meta‐regression analyses conducted, there may be some considerations about clinical and methodologic heterogeneity, because the meta‐analysis included different comparators and the indication for DCBs were variable. Finally, the lack of patient‐level data precluded a careful evaluation for the patient and lesion characteristics that would benefit most from DCBs.
Conclusions
In this meta‐analysis of 14 randomized trials comprising 2483 patients with de novo coronary lesions, DCBs were associated with similar MLD, diameter stenosis, acute lumen gain, binary restenosis, and lower late lumen loss compared with control on routine angiographic follow up. There was no difference in the incidence of TLR between DCBs compared with control. This effect was observed regardless of indication (ie, small‐vessel disease, high bleeding risk, MI, or bifurcational lesions), and was maintained when compared with second‐generation DES alone. Finally, DCBs were associated with lower risk of MI and all‐cause mortality, albeit with a low number of events, so our work should be only considered hypothesis‐generating. Our findings support the need for a randomized trial powered for clinical outcomes evaluating the role of the DCBs in all‐comers.
Sources of Funding
None.
Disclosures
None.
Supporting information
Tables S1–S4 Figures S1–12 References 10–14, 22, 24–27, 29, and 31
(J Am Heart Assoc. 2020;9:e016224 DOI: 10.1161/JAHA.120.016224.)
For Sources of Funding and Disclosures, see page 10.
References
- 1. Palmerini T, Benedetto U, Biondi‐Zoccai G, Della Riva D, Bacchi‐Reggiani L, Smits PC, Vlachojannis GJ, Jensen LO, Christiansen EH, Berencsi K, et al. Long‐term safety of drug‐eluting and bare‐metal stents: evidence from a comprehensive network meta‐analysis. J Am Coll Cardiol. 2015;65:2496–2507. [DOI] [PubMed] [Google Scholar]
- 2. Otsuka F, Vorpahl M, Nakano M, Foerst J, Newell JB, Sakakura K, Kutys R, Ladich E, Finn AV, Kolodgie FD, et al. Pathology of second‐generation everolimus‐eluting stents versus first‐generation sirolimus‐ and paclitaxel‐eluting stents in humans. Circulation. 2014;129:211–223. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3. Mahmoud AN, Barakat AF, Elgendy AY, Schneibel E, Mentias A, Abuzaid A, Elgendy IY. Long‐term efficacy and safety of everolimus‐eluting bioresorbable vascular scaffolds versus everolimus‐eluting metallic stents: a meta‐analysis of randomized trials. Circ Cardiovasc Interv. 2017;10:e005286. [DOI] [PubMed] [Google Scholar]
- 4. Kleber FX, Rittger H, Bonaventura K, Zeymer U, Wöhrle J, Jeger R, Levenson B, Möbius‐Winkler S, Bruch L, Fischer D, et al. Drug‐coated balloons for treatment of coronary artery disease: updated recommendations from a consensus group. Clin Res Cardiol. 2013;102:785–797. [DOI] [PubMed] [Google Scholar]
- 5. Giacoppo D, Alfonso F, Xu B, Claessen BEPM, Adriaenssens T, Jensen C, Pérez‐Vizcayno MJ, Kang DY, Degenhardt R, Pleva L, et al. Paclitaxel‐coated balloon angioplasty vs. drug‐eluting stenting for the treatment of coronary in‐stent restenosis: a comprehensive, collaborative, individual patient data meta‐analysis of 10 randomized clinical trials (DAEDALUS study). Eur Heart J. 2019;ehz594 DOI: 10.1093/eurheartj/ehz594. [Epub ahead of print]. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6. Elgendy IY, Mahmoud AN, Elgendy AY, Cuesta J, Rivero F, Alfonso F. Meta‐analysis comparing the frequency of target lesion revascularization with drug‐coated balloons or second‐generation drug‐eluting stents for coronary in‐stent restenosis. Am J Cardiol. 2019;123:1186–1187. [DOI] [PubMed] [Google Scholar]
- 7. Elgendy IY, Mahmoud AN, Elgendy AY, Mojadidi MK, Elbadawi A, Eshtehardi P, Pérez‐Vizcayno MJ, Wayangankar SA, Jneid H, David Anderson R, et al. Drug‐eluting balloons versus everolimus‐eluting stents for in‐stent restenosis: a meta‐analysis of randomized trials. Cardiovasc Revasc Med. 2019;20:612–618. [DOI] [PubMed] [Google Scholar]
- 8. Siontis GC, Stefanini GG, Mavridis D, Siontis KC, Alfonso F, Pérez‐Vizcayno MJ, Byrne RA, Kastrati A, Meier B, Salanti G, et al. Percutaneous coronary interventional strategies for treatment of in‐stent restenosis: a network meta‐analysis. Lancet. 2015;386:655–664. [DOI] [PubMed] [Google Scholar]
- 9. Neumann FJ, Sousa‐Uva M, Ahlsson A, Alfonso F, Banning AP, Benedetto U, Byrne RA, Collet JP, Falk V, Head SJ, et al. 2018 ESC/EACTS guidelines on myocardial revascularization. Eur Heart J. 2019;40:87–165. [DOI] [PubMed] [Google Scholar]
- 10. Tang Y, Qiao S, Su X, Chen Y, Jin Z, Chen H, Xu B, Kong X, Pang W, Liu Y, et al. Drug‐coated balloon versus drug‐eluting stent for small‐vessel disease: the RESTORE SVD China randomized trial. JACC Cardiovasc Interv. 2018;11:2381–2392. [DOI] [PubMed] [Google Scholar]
- 11. Jeger RV, Farah A, Ohlow MA, Mangner N, Möbius‐Winkler S, Leibundgut G, Weilenmann D, Wöhrle J, Richter S, Schreiber M, et al. Drug‐coated balloons for small coronary artery disease (BASKET‐SMALL 2): an open‐label randomised non‐inferiority trial. Lancet. 2018;392:849–856. [DOI] [PubMed] [Google Scholar]
- 12. Rissanen TT, Uskela S, Eränen J, Mäntylä P, Olli A, Romppanen H, Siljander A, Pietilä M, Minkkinen MJ, Tervo J, et al. Drug‐coated balloon for treatment of de‐novo coronary artery lesions in patients with high bleeding risk (DEBUT): a single‐blind, randomised, non‐inferiority trial. Lancet. 2019;394:230–239. [DOI] [PubMed] [Google Scholar]
- 13. Scheller B, Ohlow MA, Ewen S, Kische S, Rudolph TK, Clever YP, Wagner A, Richter S, El‐Garhy M, Böhm M, et al. Randomized comparison of bare metal or drug‐eluting stent versus drug coated balloon in non‐ST‐elevation myocardial infarction—PEPCAD NSTEMI. EuroIntervention. 2020;15:1527–1533. [DOI] [PubMed] [Google Scholar]
- 14. Vos NS, Fagel ND, Amoroso G, Herrman JR, Patterson MS, Piers LH, van der Schaaf RJ, Slagboom T, Vink MA. Paclitaxel‐coated balloon angioplasty versus drug‐eluting stent in acute myocardial infarction: the REVELATION randomized trial. JACC Cardiovasc Interv. 2019;12:1691–1699. [DOI] [PubMed] [Google Scholar]
- 15. Hutton B, Salanti G, Caldwell DM, Chaimani A, Schmid CH, Cameron C, Ioannidis JP, Straus S, Thorlund K, Jansen JP, et al. The PRISMA extension statement for reporting of systematic reviews incorporating network meta‐analyses of health care interventions: checklist and explanations. Ann Intern Med. 2015;162:777–784. [DOI] [PubMed] [Google Scholar]
- 16. Booth A, Clarke M, Dooley G, Ghersi D, Moher D, Petticrew M, Stewart L. The nuts and bolts of PROSPERO: an international prospective register of systematic reviews. Syst Rev. 2012;1:2. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Higgins JP, Altman DG, Gøtzsche PC, Jüni P, Moher D, Oxman AD, Savovic J, Schulz KF, Weeks L, Sterne JA. The Cochrane Collaboration's tool for assessing risk of bias in randomised trials. BMJ. 2011;343:d5928. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18. DerSimonian R, Laird N. Meta‐analysis in clinical trials. Control Clin Trials. 1986;7:177–188. [DOI] [PubMed] [Google Scholar]
- 19. Yusuf S, Peto R, Lewis J, Collins R, Sleight P. Beta blockade during and after myocardial infarction: an overview of the randomized trials. Prog Cardiovasc Dis. 1985;27:335–371. [DOI] [PubMed] [Google Scholar]
- 20. Higgins JP, Thompson SG, Deeks JJ, Altman DG. Measuring inconsistency in metaanalyses. BMJ. 2003;327:557–560. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21. Egger M, Davey Smith G, Schneider M, Minder C. Bias in meta‐analysis detected by a simple, graphical test. BMJ. 1997;315:629–634. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 22. Cortese B, Micheli A, Picchi A, Coppolaro A, Bandinelli L, Severi S, Limbruno U. Paclitaxel‐coated balloon versus drug‐eluting stent during PCI of small coronary vessels, a prospective randomised clinical trial. The PICCOLETO study. Heart. 2010;96:1291–1296. [DOI] [PubMed] [Google Scholar]
- 23. Thompson SG, Higgins JP. How should meta‐regression analyses be undertaken and interpreted? Stat Med. 2002;21:1559–1573. [DOI] [PubMed] [Google Scholar]
- 24. Cortese B. Drug‐coated balloon vs. drug‐eluting stent for small coronary vessel disease: 6 months primary outcome of the PICCOLETO II randomized clinical trial. Presented at the Transcatheter Cardiovascular Therapeutic meeting, San Francisco, CA, September 2019. [Google Scholar]
- 25. Funatsu A, Nakamura S, Inoue N, Nanto S, Nakamura M, Iwabuchi M, Ando K, Asano R, Habara S, Saito S, et al. A multicenter randomized comparison of paclitaxel‐coated balloon with plain balloon angioplasty in patients with small vessel disease. Clin Res Cardiol. 2017;106:824–832. [DOI] [PubMed] [Google Scholar]
- 26. Latib A, Colombo A, Castriota F, Micari A, Cremonesi A, De Felice F, Marchese A, Tespili M, Presbitero P, Sgueglia GA, et al. A randomized multicenter study comparing a paclitaxel drug‐eluting balloon with a paclitaxel‐eluting stent in small coronary vessels: the BELLO (Balloon Elution and Late Loss Optimization) study. J Am Coll Cardiol. 2012;60:2473–2480. [DOI] [PubMed] [Google Scholar]
- 27. Latib A, Ruparelia N, Menozzi A, Castriota F, Micari A, Cremonesi A, De Felice F, Marchese A, Tespili M, Presbitero P, et al. 3‐year follow‐up of the Balloon Elution and Late Loss Optimization Study (BELLO). JACC Cardiovasc Interv. 2015;8:1132–1134. [DOI] [PubMed] [Google Scholar]
- 28. Gobić D, Tomulić V, Lulić D, Židan D, Brusich S, Jakljević T, Zaputović L. Drug‐coated balloon versus drug‐eluting stent in primary percutaneous coronary intervention: a feasibility study. Am J Med Sci. 2017;354:553–560. [DOI] [PubMed] [Google Scholar]
- 29. Shin ES, Lee JM, Her AY, Chung JH, Eun Lee K, Garg S, Nam CW, Doh JH, Koo BK. Prospective randomized trial of paclitaxel‐coated balloon versus bare‐metal stent in high bleeding risk patients with de novo coronary artery lesions. Coron Artery Dis. 2019;30:425–431. [DOI] [PubMed] [Google Scholar]
- 30. Kleber FX, Rittger H, Ludwig J, Schulz A, Mathey DG, Boxberger M, Degenhardt R, Scheller B, Strasser RH. Drug eluting balloons as stand alone procedure for coronary bifurcational lesions: results of the randomized multicenter PEPCAD‐BIF trial. Clin Res Cardiol. 2016;105:613–621. [DOI] [PubMed] [Google Scholar]
- 31. López Mínguez JR, Nogales Asensio JM, Doncel Vecino LJ, Sandoval J, Romany S, Martínez Romero P, Fernández Díaz JA, Fernández Portales J, González Fernández R, Martínez Cáceres G, et al. A prospective randomised study of the paclitaxel‐coated balloon catheter in bifurcated coronary lesions (BABILON trial): 24‐month clinical and angiographic results. EuroIntervention. 2014;10:50–57. [DOI] [PubMed] [Google Scholar]
- 32. Nishiyama N, Komatsu T, Kuroyanagi T, Fujikake A, Komatsu S, Nakamura H, Yamada K, Nakahara S, Kobayashi S, Taguchi I. Clinical value of drug‐coated balloon angioplasty for de novo lesions in patients with coronary artery disease. Int J Cardiol. 2016;222:113–118. [DOI] [PubMed] [Google Scholar]
- 33. Alfonso F, Elgendy IY, Cuesta J. Drug‐coated balloons versus drug‐eluting stents for in‐stent restenosis: the saga continues. EuroIntervention. 2018;14:1069–1072. [DOI] [PubMed] [Google Scholar]
- 34. Katsanos K, Spiliopoulos S, Kitrou P, Krokidis M, Karnabatidis D. Risk of death following application of paclitaxel‐coated balloons and stents in the femoropopliteal artery of the leg: a systematic review and meta‐analysis of randomized controlled trials. J Am Heart Assoc. 2018;7:e011245 DOI: 10.1161/JAHA.118.011245. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35. Beckman JA, White CJ. Paclitaxel‐coated balloons and eluting stents: is there a mortality risk in patients with peripheral artery disease? Circulation. 2019;140:1342–1351. [DOI] [PubMed] [Google Scholar]
- 36. Freisinger E, Koeppe J, Gerss J, Goerlich D, Malyar NM, Marschall U, Faldum A, Reinecke H. Mortality after use of paclitaxel‐based devices in peripheral arteries: a real‐world safety analysis. Eur Heart J. 2019;ehz698 DOI: 10.1093/eurheartj/ehz698. [Epub ahead of print]. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37. Schneider PA, Laird JR, Doros G, Gao Q, Ansel G, Brodmann M, Micari A, Shishehbor MH, Tepe G, Zeller T. Mortality not correlated with paclitaxel exposure: an independent patient‐level meta‐analysis of a drug‐coated balloon. J Am Coll Cardiol. 2019;73:2550–2563. [DOI] [PubMed] [Google Scholar]
- 38. Megaly M, Rofael M, Saad M, Rezq A, Kohl LP, Kalra A, Shishehbor M, Soukas P, Abbott JD, Brilakis ES. Outcomes with drug‐coated balloons in small‐vessel coronary artery disease. Catheter Cardiovasc Interv. 2019;93:E277–E286. [DOI] [PubMed] [Google Scholar]
- 39. Li M, Guo C, Lv YH, Zhang MB, Wang ZL. Drug‐coated balloon versus drug‐eluting stent in de novo small coronary vessel disease: a systematic review and meta‐analysis. Medicine (Baltimore). 2019;98:e15622. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40. Megaly M, Rofael M, Saad M, Shishehbor M, Brilakis ES. Outcomes with drug‐coated balloons for treating the side branch of coronary bifurcation lesions. J Invasive Cardiol. 2018;30:393–399. [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Tables S1–S4 Figures S1–12 References 10–14, 22, 24–27, 29, and 31