Skip to main content
. 2020 Nov 5;25(21):5158. doi: 10.3390/molecules25215158

Table 1.

Results of method validation.

Compound Sensitivity Linearity Precision and Accuracy
Within-Run (n = 5) Between-Run (n = 10)
LOD (ng/mL) LQQ (ng/mL) R2 Slope tcal > tcri Expected Conc. (ng/mL) Measured Conc. (ng/mL) Precision (CV%) Accuracy (%) Measured conc. (ng/mL) Precision (CV%) Accuracy (%)
3-NT 0.030 0.100 0.997 168.18 > 1.94 0.275
1.5
2.5
0.261 ± 0.006
1.52 ± 0.07
2.54 ± 0.18
9.04
3.11
2.75
94.9
101.13
101.44
0.264 ± 0.005
1.49 ± 0.07
2.52 ± 0.15
7.26
3.21
2.35
96.0
99.63
100.99
3-CT 0.026 0.096 0.996 175.39 > 1.94 0.275
1.5
2.5
0.279 ± 0.002
1.51 ± 0.11
2.61 ± 0.29
2.49
4.81
4.18
101.5
100.45
104.50
0.279 ± 0.002
1.52 ± 0.09
2.57 ± 0.24
2.51
4.05
3.59
101.5
101.63
102.84
3-BT 0.030 0.098 0.996 170.82 > 1.94 0.268
1.5
2.5
0.239 ± 0.001
1.57 ± 0.09
2.52 ± 0.19
1.52
3.61
2.97
97.5
104.73
100.77
0.270 ± 0.002
1.54 ± 0.08
2.53 ± 0.15
2.10
3.21
2.30
98.3
102.88
101.21

3-NT, 3-Nitro-l-tyrosine; 3-CT, 3-Chloro-l-tyrosine; 3-BT, 3-Bromo-l-tyrosine; LOD, limit of detection; LOQ, limit of quantification; CV, coefficient of variation; R2, regression coefficient; conc., concentration.