Capacitive-Coupling Impedance Spectroscopy Using a Non-Sinusoidal Oscillator and Discrete-Time Fourier Transform: An Introductory Study

Tomiharu Yamaguchi 1, Akinori Ueno 1,*
PMCID: PMC7665133  PMID: 33182456

Abstract

In this study, we propose a new short-time impedance spectroscopy method with the following three features: (1) A frequency spectrum of complex impedance for the measured object can be obtained even when the measuring electrodes are capacitively coupled with the object and the precise capacitance of the coupling is unknown; (2) the spectrum can be obtained from only one cycle of the non-sinusoidal oscillation waveform without sweeping the oscillation frequency; and (3) a front-end measuring circuit can be built, simply and cheaply, without the need for a digital-to-analog (D-A) converter to synthesize elaborate waveforms comprising multiple frequencies. We built the measurement circuit using the proposed method and then measured the complex impedance spectra of 18 resistive elements connected in series with one of three respective capacitive couplings. With this method, each element’s resistance and each coupling’s capacitance were estimated independently and compared with their nominal values. When the coupling capacitance was set to 10 nF or 1.0 nF, estimated errors for the resistive elements in the range of 2.0–10.0 kΩ were less than 5%.

Keywords: capacitive coupling, impedance spectroscopy, non-sinusoidal oscillator, DFT

1. Introduction

Impedance spectroscopy (IS) measures the impedance of targets at various frequencies and assesses their electrical characteristics using spectral analysis. IS is a widely used technique in electrochemistry and biomedical engineering.

In electrochemistry, IS is known as electrochemical impedance spectroscopy (EIS). EIS is used to analyze the characteristics of fuel cells, lithium-ion batteries, and photovoltaic cells [1,2,3,4,5]. It is often used to analyze the response mechanism of chemical sensors and detect response signals in them [6,7,8]. In general, the ionic conduction speed differs in electrode interfaces and the electrolytes of fuel cells and lithium-ion batteries. The former responds to electric fields more slowly than the latter. Therefore, when the frequency of an electric field applied to a measured object is high, the ionic conduction of the electrode interface cannot follow the changes in the alternating field, and, for the most part, there is a bulk impedance component with a fast response speed. Therefore, by measuring the impedance spectrum, each ionic conduction process can be analyzed.

EIS is commonly used to evaluate anti-corrosion properties of metals and to monitor corrosion processes because the corrosion reactions of metals change electrical potential and resistance [9,10]. In addition, it is also applied to the degradation monitoring of anti-corrosive organic coatings [11,12,13,14,15,16,17]. The coatings are degraded or delaminated by corrosion reactions at metal-coating interfaces, which changes their circuit parameters, such as coating resistance and coating capacitance. The coating resistance of anti-corrosive coatings is extremely high, but gradually decreases with their degradation. The coating capacitance depends on the water volume fraction of the coatings [18]. Because most coatings are relatively thick and have low coating capacitance, systems for measuring low capacitance are required for degradation monitoring.

In biomedical engineering, IS is known as bioimpedance spectroscopy (BIS) [19,20,21,22,23]. When measuring impedance for a few frequencies, it is also referred to as bioelectrical impedance analysis (BIA). In a living body, impedance consists of a resistance component dependent on the body’s water content and a capacity component that depends on cell membranes’ capacitance. The body’s composition information, such as its water content, obtained from impedance, is an important indicator of health conditions and their diagnoses. The body’s water content is generally inversely proportional to resistance; thus, low-resistance reflects a high water content in the body. Furthermore, a lower capacitance indicates that there are either not many biological tissues or a high volume of extracellular fluid. Both the BIS and BIA methods can non-invasively determine body compositions and pathologies at relatively low cost; thus, the technique can be used to monitor body fat and body composition [24,25,26].

The non-invasive assessment of body composition is useful for patients with diseases such as heart failure, chronic kidney disease, and cardiorenal syndrome [27]. For a more detailed analysis of body composition in these patients, bioelectrical impedance vector analysis (BIVA) has been proposed [28]. In the BIVA approach, the resistance component R of impedance and reactance component XC are standardized with body height, and the vector of impedance is plotted on an RXC graph. The absolute value of the vector and its phases are then correlated with tissue hydration and body cell mass in soft tissues, respectively. BIVA is also used to analyze the body composition of patients with cancer [29,30], those undergoing dialysis [31,32,33], those with edema [34,35], and those with polyneuropathy, organomegaly, endocrinopathy, M-protein, and skin change syndromes [36].

BIA is also known as impedance microbiology (IM). The impedance of the growth medium for microorganisms depends on microbial concentrations [37]. When the concentration of microorganisms increases, the proteins, carbohydrates, and fats decompose into amino acids, sugars, etc. Because the medium’s impedance is considered a series connection of the medium’s resistance and the electric double layer capacitance formed between the medium and the electrodes, the ionic and polar substances produced change the medium’s impedance. IM has a range of uses and is applied, for instance, to determine microbial concentrations in ice cream [38] and beer [39].

BIS- and BIA-based sensors for detecting biological substances and agents, such as bladder cancer [40], DNA [41], protein [42], and bacteria [43], have also been studied. These sensors can transduce the interaction of antibodies with their object into an electrical signal.

In food engineering, IS is applied to assess the ripeness of fruits and the freshness of fish, with analytical investigations presented in the literature of apples, bananas, and oranges [44,45], in addition to carp and herring [46].

However, in all fields, measured objects are generally bound with a resistive coupling, and sine waves are applied; thus, IS has rarely been applied to the analysis of insulator–electrolyte interfaces or the measurement of the electrical characteristics of resistive elements with an insulating coating.

When electrodes come into contact with thin insulating material (Figure 1a), two capacitive couplings form between the resistive elements of the measured object, which can be expressed with an equivalent circuit (Figure 1b). In such a capacitive-coupling state, if the resistive value can be measured in some IS approach, the scope of application of IS will broaden. Furthermore, because IS measures the impedance of multiple frequencies, sine wave signals with different frequencies are input, and responses are measured. Therefore, IS requires a complex circuit of oscillators that can generate multiple sine wave signals. In addition, measurement times increase because multiple signals are sequentially input. To resolve these two problems, a method has been developed in which signals with multiple frequency components are applied to the measured objects, and impedance is determined using a discrete Fourier transform (DFT) of the obtained signals, a fast Fourier transform (FFT), or a short-time Fourier transform (STFT) [47,48,49,50]. These methods are applied to a wide range of fields, such as lithium ion batteries [51,52], corrosion studies [53], materials [54,55], freezing processes of crops [56], and gene detection [57]; however, no reports of capacitive coupling have been published.

Figure 1.

Figure 1

Schematic model of electrodes capacitively coupled to a conductive material via a thin insulator: (a) cross-sectional diagram; (b) equivalent circuit.

In this study, we demonstrate a new short-time capacitive-coupling IS based on the following ideas:

  • (1)

    By coupling electrodes capacitively to the measured object and by incorporating the resulting couplings into an oscillation circuit, an alternating current is applicable inside the object covered with a thin insulating layer.

  • (2)

    By measuring the amplitude and phase of the object’s current and those of the object’s potential difference resulting from oscillation, even with unknown coupling capacitance, the impedance of the object is measurable.

  • (3)

    By estimating the impedance of the measured object from the amplitude and phase spectrum obtained from the waveform of a few oscillation cycles, the temporal resolution of IS is improved.

  • (4)

    By making the oscillation waveform a non-sinusoidal wave, the fundamental frequency of oscillation and its higher harmonic waves are usable for the analysis. In this manner, the operation to switch frequency of a sinusoidal wave becomes unnecessary.

2. Approach of Capacitive-Coupling IS

2.1. Non-Sinusoidal Oscillator Circuit with Capacitive Couplings

Figure 2 illustrates circuit diagrams for the measurement of impedance shown in Figure 1.

Figure 2.

Figure 2

Circuit diagrams illustrating the measurement of impedance shown in Figure 1: (a) with a sinusoidal voltage source; and (b) in a non-sinusoidal oscillator.

In a circuit featuring the sinusoidal voltage of a single frequency (Figure 2a), an unknown impedance Z˙X is obtained with Equation (1):

Z˙X=V˙12I˙12=V˙12V˙1/Z˙A=V˙12V˙1·Z˙A. (1)

where V˙12 and I˙12 are the complex voltage and complex current of Z˙X, respectively, V˙1 is the complex voltage of Z˙A, and Z˙A is the combined impedance of resistance RA and parasitic capacitance CA. Based on this idea, in the present study, we obtained the frequency characteristics from the voltage signal of a non-sinusoidal oscillator. For the non-sinusoidal oscillator, including the capacitive coupling, we used a Schmitt trigger inverter oscillator. This oscillator does not have high precision for the oscillation frequency, but features a small number of parts and stable oscillation; thus, it makes for a small, cheap, and easy-to-handle system. Figure 2b shows the prepared oscillator. RX is a resistive element, which is the measured object, and CX1 and CX2 are the capacitive-coupling parts of the insulators and electrodes. Because Z˙X=1/jωCX1+RX+1/jωCX2=RX+1/jωCX1CX2/CX1+CX2, the capacitive couplings are considered to be incorporated into the oscillator as a capacitance CX=CX1CX2/CX1+CX2. The resistive elements may also include a capacity component. However, in the present study, we assumed that there is only a resistive component for the sake of simplification. For the Schmitt trigger inverter integrated circuit (IC), we used a readily available 74HC14AP, and the power supply voltage of the IC was 5 V. The resistance for measuring the current i12t is RA. Parasitic capacitance CA combines with RA. The discharge and charge of the capacitor connected on the input side of the Schmitt trigger inverter switched the high and low levels of the voltage vOSCt, which led to the non-sinusoidal oscillation of the voltage v1t and v2t. If parasitic capacitance CA can be ignored, the oscillation frequency f0 can be expressed with the following:

f0=1RF+RA+RXCXlnVHVAVHVP+lnVBVLVNVL, (2)
VA=VN+RA+RXRF+RA+RXVHVL, (3)
VB=VPRA+RXRF+RA+RXVHVL. (4)

where VH is 5 V for the high-level output and 0 V for the low-level output of the Schmitt trigger inverter. In addition, VP and VN are the threshold voltages of the Schmitt trigger inverter, where VP>VN. Equation (2) shows that f0 is determined by the RX and CX of the capacitive-coupling part. Even if f0 is determined, the satisfactory combination of RX and CX is infinite; thus, Equation (2) alone cannot uniquely determine RX and CX.

2.2. Determination of Unknown Capacitance and Resistance in Series Connection

With a normal IS, sine waves with varying frequencies are input into the unknown impedance, and the impedance characteristics are measured and calculated. For example, in Figure 2a, the impedance Z˙Xf at the frequency of f can be calculated with the following equation:

Z˙Xf=V˙12fV˙1f·Z˙AfejθV12fθV1f+θZAf. (5)

where V˙12f and V˙1f are the measured amplitude of V˙12f, V˙1f and the θV12f, θV1f phase. The absolute value Z˙Af and phase θZAf of Z˙Af are theoretically derived. By obtaining frequency–gain characteristics (amplitude spectrum) and frequency–phase characteristics (phase spectrum), IS can be actualized.

If Z˙Xf was incorporated into part of the non-sinusoidal oscillator (Figure 2b), the measured v12t and v1t have periodic waveforms of non-sinusoidal waves with an oscillation frequency f0. Therefore, the Fourier series expansion of one cycle, T0=1/f0, of the observed waveforms can be expressed with Equations (6)–(9) using the complex Fourier coefficients V˙12kf0 and V˙1kf0:

v12t=k=V˙12kf0ej2πkf0t, (6)
V˙12kf0=1T00T0v12tej2πkf0tdt, (7)
v1t=k=V˙1kf0ej2πkf0t, (8)
V˙1kf0=1T00T0v1tej2πkf0tdt. (9)

Therefore, we can assume that v12t and v1t consist of sine waves with a fundamental frequency f0 and its integer multiple kf0. Each frequency component’s coefficient comes from the DFT of one cycle as an amplitude spectrum V˙12kf0 and V˙1kf0 and phase spectrum θV12kf0 and θV1kf0. In Equations (6) and (8), the negative frequency component (k<0) is a complex conjugate of the positive frequency component:

V˙12kf0=2V˙12kf0, (10)
V˙1kf0=2V˙1kf0. (11)

If v1t+T0/2=v1t, v12t and v1t only include the odd-order frequency component.

Furthermore, the impedance Z˙Af=Z˙AfejθZAf can be expressed with the following equation as a continuous function of f:

Z˙Af=11RA2+2πfCA2, (12)
θZAf=tan12πfCARA. (13)

Therefore, for each frequency of nf0 (n=1,2,3,), amplitude and phase spectrums,

Z˙Anf0=11RA2+2πnf0CA2, (14)
θZAnf0=tan12πnf0CARA (15)

are obtained. Thus, the amplitude and phase spectrum, Z˙Xnf0 and θZ12nf0, of unknown impedance Z˙X are obtained with Equations (16) and (17) using the amplitude and phase spectra of V˙12, V˙1, and Z˙A:

Z˙Xnf0=V˙12nf0V˙1nf0·Z˙Anf0, (16)
θZ12nf0=θV12nf0θV1nf0+θZAnf0. (17)

Furthermore, with Equations (16) and (17), the frequency response for the real and imaginary parts of impedance Z˙X, Re(Z˙X) and Im(Z˙X), are obtained with Equations (18) and (19):

Re(Z˙X)=V˙12nf0V˙1nf0·Z˙Anf0cosθV12nf0θV1nf0+θZAnf0, (18)
Im(Z˙X)=V˙12nf0V˙1nf0·Z˙Anf0sinθV12nf0θV1nf0+θZAnf0. (19)

The frequency response of the real and imaginary parts of admittance Y˙X, Re(Y˙X) and Im(Y˙X), can be obtained with Equations (20) and (21) by setting Y˙Xnf0=1/Z˙Xnf0:

Re(Y˙X)=1Z˙Xnf0cosθZ12nf0, (20)
Im(Y˙X)=1Z˙Xnf0sinθZ12nf0. (21)

Next, if we know that Z˙X is a resistive element and capacitive coupling in series, as in Figure 1b, the frequency responses of the real and imaginary parts of the impedance are:

Re(Z˙X)=RX, (22)
Im(Z˙X)=12πnf0CX. (23)

Thus, RX and CX can be estimated by applying the least-squares method to the finite number of spectra of frequency, nf0 (n=1, 2, 3, , N).

2.3. Experimental Method

For the oscillator in Figure 2b, we used multiple high-precision metal film resistors and one multilayer ceramic capacitor to mimic the resistive element RX and the capacitive coupling CX. The RX was changed between 0.20 and 10.0 kΩ using the series connection of resistance of 0.10 or 1.0 kΩ (tolerance of ±0.1% for both). The nominal values of the CX were 0.10 nF, 1.0 nF (tolerance of ±5% for both), or 10 nF (tolerance of ±10%). The RX and CX were wired on a breadboard.

The time waveform data of the voltage were collected using a digital oscilloscope (OWON VDS3104L). v1t and v2t in Figure 2b were measured, and the time waveform data of v12t was obtained from their difference. Then, we transformed DFT for two cycles (CX = 1.0 nF or 10 nF) or four cycles (CX = 0.10 nF) of the time waveform data. The time waveform data was divided by a certain number of data points, and DFT was performed for each cycle. We averaged each DFT result. For the window function of DFT, we used a rectangular window. From the obtained DFT data, we obtained real and imaginary parts for Z˙X following the steps discussed in Section 2.2. RX and CX were estimated using the least-squares method.

3. Results

The oscillation waveforms for the circuit in Figure 2b are shown in Figure 3. The frequency spectra of V˙12 and V˙1 were obtained from the DFT of these waveforms (Figure 4). The RX and CX were 4.0 kΩ and 10 nF, respectively. When CX was charged and v12t increased to about 3.1 V, the output voltage of the Schmitt trigger inverter vOSCt switched from VH to VL. This caused a discharge of CX and v12t was reduced. When v2t decreased to about 2.2 V, vOSCt was switched from VL to VH, and CX was charged again. This discharge and the charge of CX created oscillation. The time it took to charge CX, tC, was 0.663 ms. The time it took to discharge tD was 0.654 ms. Therefore, data for the two cycles were collected in 2.6 ms, enabling analysis in a short period, even when the DFT calculation time was included. Furthermore, tCtD and v1t+T0/2=v1t were mostly satisfied; thus, the frequency components included in the oscillation waveform were odd harmonics 2m1f0(m=1, 2, 3, ).

Figure 3.

Figure 3

Example of the oscillation waveforms of v12t and v1t measured in the circuit of Figure 2b. The RX and CX were set to 4.0 kΩ and 10 nF, respectively. The segment of time with a colored background corresponds to two cycles of the oscillation and was used for the subsequent discrete Fourier transform (DFT) analysis.

Figure 4.

Figure 4

Frequency spectra of: (a) amplitude V˙12; (b) phase θV12; (c) amplitude V˙1; and (d) phase θV1 at 2m1f0 (m=1, 2, 3, , 20 ) Hz. The spectra were obtained using DFT from the two-cycle segment of v12t and v1t in Figure 3. The RX and CX were set to 4.0 kΩ and 10 nF, respectively.

Next, Figure 4 shows odd harmonic components of V˙12 and V˙1 to the 39th order (N=39). Both V˙122m1f0 and V˙12m1f0 contained odd harmonic components with a fundamental frequency of 759 Hz (Figure 4a,c). Because v12t was a waveform close to a triangular wave, the amplitude of V˙122m1f0 was inversely proportional to the square of the harmonic order. By comparison, because v1t is a waveform close to a square wave, the amplitude of V˙12m1f0 was inversely proportional to the harmonic order. Furthermore, the phase of V˙122m1f0 increased as the frequency increased from −2.87 rad, whereas the phase of V˙12m1f0 decreased as the frequency increased from −1.49 rad (Figure 4b,d).

The result of estimating the impedance Z˙X from the obtained frequency spectrum of V˙12 and V˙1 using Equations (16) and (17) is shown in Figure 5. The absolute value and phase of Z˙A necessary to estimate Z˙X were obtained from Equations (14) and (15) (Figure 5a,b). As the frequency increased, the absolute value of the estimated Z˙X, Z˙X, decreased, whereas phase θZX gradually increased from −1.38 rad to −0.0988 rad (Figure 5c,d). This reflects the frequency characteristics of resistor-capacitor (RC) series circuits.

Figure 5.

Figure 5

Frequency spectra of: (a) absolute impedance Z˙A; (b) phase θZA; (c) absolute impedance Z˙X; and (d) phase θZX at 2m1f0 (m=1, 2, 3, , 20 ) Hz.

In addition, the real and imaginary parts of Z˙X and Y˙X were obtained from the absolute value and phase of Z˙X using Equations (18) and (21). A 3D graph with the frequency is shown in Figure 6 [58]. Because RX and CX are connected in series, the real part of Z˙X, Re(Z˙X), equals RX, and remains constant regardless of the frequency. The imaginary part of Z˙X, Im(Z˙X), is inversely proportional to the frequency. The Cole–Cole plot of Z˙X is a straight line in the case of impedance and a semicircle in that of admittance. It showed that the impedance spectrum necessary to analyze EIS and BIS could be obtained using the proposed method. Estimates of the resistance and capacitance are shown in Figure 7. Using the data from the real part of Z˙X, the resistance RX was estimated with Equation (22), and the result was 4.082 kΩ. Furthermore, the capacitance CX can be estimated from the imaginary part of Z˙X, corresponding to Equation (23), which led to 9.395 nF. The errors between these estimates and the nominal values were 2.05% for RX and −6.05% for CX. The estimation error of the CX is within the tolerance of the used capacitor and is considered a reasonable level of accuracy. As such, the proposed method can estimate the resistance of the resistive element and coupled capacitance.

Figure 6.

Figure 6

Figure 6

Three-dimensional perspective plots of: (a) impedance Z˙X; and (b) admittance Y˙X. Three-dimensional DFT data is projected onto each plane. The solid and dashed lines are theoretical curves of Z˙X and Y˙X.

Figure 7.

Figure 7

Frequency spectra of: (a) real part Re(Z˙X); and (b) imaginary part Im(Z˙X) of impedance Z˙X. The DFT data were fitted to Equations (22) and (23) for determining RX and CX (dashed lines).

We also examined whether the resistance and capacitance can be estimated when CX was lowered to 1.0 or 0.10 nF. For an RX from 0.20 to 10.0 kΩ, and CX at 10, 1.0, and 0.10 nF, the resistance and capacitance were estimated (Figure 8). When CX was 10 or 1.0 nF, the error between the resistance and capacitance, and their respective nominal values, was less than 5% in the range of RX from 2.0 to 10.0 kΩ, which is a highly precise estimate. Although the precision of the resistance estimate decreased in the range of RX from 0.40 to 1.0 kΩ, estimation was still possible. The impedance of a living body is generally several hundred Ω, and so the present method can also be applied to BIS. However, when CX was 0.10 nF, the relative error of estimated low-resistance values was high, making estimations difficult.

Figure 8.

Figure 8

RX and CX estimated using oscillation waveforms and DFT: (a) estimated RX; (b) absolute error of estimated RX; (c) estimated CX; (d) relative error of estimated CX.

4. Discussion

With the proposed method, we could estimate RX in the range of 0.90 to 10.0 kΩ and CX of 10 or 1.0 nF with accuracy. If RX is smaller than 0.90 kΩ or CX is 0.10 nF, the estimate was less precise. This likely has an impact on the parasitic impedance Z˙A in wires and circuit boards, in addition to the RA and CA. Thus, to improve the precision for low resistance and capacitance, we determined a more detailed equivalent circuit model of Z˙A from the Cole–Cole plot of Z˙A when CX= 0.10 nF and RX= 0 Ω. When CX was known, Z˙A was obtained with V˙12nf0 and V˙1nf0 with DFT using Equation (24):

Z˙Anf0=V˙1nf0V˙12nf0·12πnf0CXejθV1nf0θV12nf0π2. (24)

Figure 9a shows the Cole–Cole plot of Z˙A. The DFT data creates a capacitive semicircle and is under the notable impact of RA and CA. However, there is a discrepancy between a curve calculated from RA= 20.0 kΩ and CA= 23.2 pF only (RA//CA) and DFT data, indicating that there is parasitic impedance other than RA and CA in Z˙A. As in Figure 9b, when a curve fitting was performed for an equivalent circuit model that considers parasitic resistance RAS and parasitic inductance LA, in addition to RA and CA (RA+LA//RAS+CA), the result was consistent with the DFT data. Each parameter of the equivalent circuit was RA= 20.017 kΩ, RAS= 378.761 Ω, LA= 4.705 mH and CA= 35.416 pF. For RAS, there are possibilities of resistance from the wiring pattern of the circuit board and contact resistance, but because LA is too large as a parasitic inductance, a separate interpretation of LA is needed.

Figure 9.

Figure 9

Equivalent circuit modeling of Z˙A: (a) Cole–Cole plot for CX= 0.10 nF and RX= 0 Ω; (b) equivalent circuit with a stray resistance RAS and stray inductance LA. The symbol n represents the harmonic number of Z˙Anf0. We used pyZwx software to fit the DFT data to the equivalent circuit model [59].

Next, using the obtained circuit parameters, RX and CX were estimated again (Figure 10). When estimated from odd harmonics to the 39th order, by considering both RAS and LA, the precision of the low-resistance estimate was improved. However, there was less improvement in high-resistance values (Figure 10a,b). A similar trend was observed in the estimates of CX(Figure 10c). This indicates that, at higher frequencies, the parasitic capacitance of RX and a breadboard cannot be ignored. Therefore, we limited the frequency for the estimate to odd harmonics of up to 500 kHz (N= 9) and found that the estimate precision was also improved in higher resistance values. When CX was 10 or 1.0 nF, the precision did not change, even taking RAS and LA into consideration. Therefore, by optimizing the equivalent circuit model of Z˙A and keeping the frequency of the higher harmonic wave to 500 kHz or below, estimates of low resistance and low capacitance could be improved. When the frequency of higher harmonic waves is limited, the number of data points for fitting is low when the oscillation frequency is high, which is a problem when estimating a complex circuit’s impedance. However, because Schmitt trigger oscillators can change the oscillation frequency with RF, measurements with multiple RF enable a sufficient number of data points.

Figure 10.

Figure 10

RX and CX estimated using optimized Z˙A: (a) estimated RX; (b) absolute error of estimated RX; (c) relative error of estimated CX.

5. Conclusions

In this study, we proposed a new capacitive-coupling IS that uses a non-sinusoidal oscillator and DFT. To verify this method, we prepared a Schmitt trigger oscillator that incorporated a measured object and a capacitive coupling. This circuit is extremely simple and can be miniaturized. Furthermore, the oscillator’s non-sinusoidal voltage waveform can be DFT-transformed to obtain the fundamental frequency and odd harmonic components. The resistance and capacitance that mimic a resistive element and capacitive-coupling component can be estimated by obtaining impedance from the amplitude of the voltage and phase spectrum. Furthermore, when the coupled capacitance is small, there is more impact from parasitic impedance, such as from circuit boards. This is a factor that reduces the precision of the estimate, but by correcting for parasitic impedance and limiting the order of the higher harmonic waves used for estimates, the estimation precision is improved.

For the practical application of our proposed system, we must analyze more complex circuits consisting of several series or parallel RC circuits. To improve the accuracy of the regression analysis, more DFT data points at low frequencies are necessary to accurately estimate the resistance and capacitance of complex circuits. Although we fixed the resistance RF of the fabricated oscillation circuit at 200 kΩ, this can also easily switch to higher resistance. Therefore, we can acquire non-sinusoidal waveform data quickly at lower oscillation frequencies, which enables more DFT data points to be obtained.

In the future, we will apply the present system to measuring the impedance of various targets, such as a living body, and coating delamination and corrosion processes.

Acknowledgments

The authors thank Crimson Interactive Pvt. Ltd. (Ulatus, www.ulatus.jp) for their assistance in manuscript translation and editing.

Author Contributions

Conceptualization, A.U.; methodology, A.U.; software, T.Y.; validation, T.Y.; formal analysis, T.Y.; investigation, T.Y.; resources, T.Y. and A.U.; data curation, T.Y.; writing—original draft preparation, T.Y. and A.U.; writing—review and editing, T.Y. and A.U.; visualization, T.Y.; supervision, A.U.; project administration, A.U.; funding acquisition, A.U. All authors have read and agreed to the published version of the manuscript.

Funding

This research was supported in part by SECOM Science and Technology Foundation and in part by Private University Research Branding Project supported by MEXT, Japan.

Conflicts of Interest

The authors declare no conflict of interest.

Footnotes

Publisher’s Note: MDPI stays neutral with regard to jurisdictional claims in published maps and institutional affiliations.

References

  • 1.Brear A.R.C., Chown A.L., Burton A.R., Farnum B.H. Electrochemical impedance spectroscopy of metal oxide electrodes for energy applications. ACS Appl. Energy Mater. 2020;3:66–98. doi: 10.1021/acsaem.9b01965. [DOI] [Google Scholar]
  • 2.Encinas-Sánchez V., de Miguel M.T., Lasanta M.I., García-Martín G., Pérez F.J. Electrochemical impedance spectroscopy (EIS): An efficient technique for monitoring corrosion processes in molten salt environments in CSP applications. Sol. Energy Mater. Sol. Cells. 2019;191:157–163. doi: 10.1016/j.solmat.2018.11.007. [DOI] [Google Scholar]
  • 3.Itagaki M., Suzuki S., Shitanda I., Watanabe K. Electrochemical impedance and complex capacitance to interpret electrochemical capacitor. Electrochemistry. 2007;75:649–655. doi: 10.5796/electrochemistry.75.649. [DOI] [Google Scholar]
  • 4.Bonomo M., Naponiello G., Dini D. Oxidative dissolution of NiO in aqueous electrolyte: An impedance study. J. Electroanal. Chem. 2018;816:205–214. doi: 10.1016/j.jelechem.2018.03.058. [DOI] [Google Scholar]
  • 5.Fabregat-Santiago F., Garcia-Belmonte G., Mora-Seró I., Bisquert J. Characterization of nanostructured hybrid and organic solar cells by impedance spectroscopy. Phys. Chem. Chem. Phys. 2011;13:9083–9118. doi: 10.1039/c0cp02249g. [DOI] [PubMed] [Google Scholar]
  • 6.Balasubramani V., Chandraleka S., Subba Rao T., Sasikumar R., Kuppusamy M.R., Sridhar T.M. Review—Recent advances in electrochemical impedance spectroscopy based toxic gas sensors using semiconducting metal oxides. J. Electrochem. Soc. 2020;167:037572. doi: 10.1149/1945-7111/ab77a0. [DOI] [Google Scholar]
  • 7.Valiūnienė A., Rekertaitė A.I., Ramanavičienė A., Mikoliūnait L., Ramanavičius A. Fast Fourier transformation electrochemical impedance spectroscopy for the investigation of inactivation of glucose biosensor based on graphite electrode modified by Prussian blue, polypyrrole and glucose oxidase. Colloids Surf. A Physicochem. Eng. Asp. 2017;532:165–171. doi: 10.1016/j.colsurfa.2017.05.048. [DOI] [Google Scholar]
  • 8.Yamaguchi T., Takisawa M., Kiwa T., Yamada H., Tsukada K. Analysis of response mechanism of a proton-pumping gate FET hydrogen gas sensor in air. Sens. Actuators B Chem. 2008;133:538–542. doi: 10.1016/j.snb.2008.03.021. [DOI] [Google Scholar]
  • 9.Sekine I. Recent evaluation of corrosion protective paint films by electrochemical methods. Prog. Org. Coat. 1997;31:73–80. doi: 10.1016/S0300-9440(97)00020-9. [DOI] [Google Scholar]
  • 10.Bonora P.L., Deflorian F., Fedrizzi L. Electrochemical impedance spectroscopy as a tool for investigating underpaint corrosion. Electrochim. Acta. 1995;41:1073–1082. doi: 10.1016/0013-4686(95)00440-8. [DOI] [Google Scholar]
  • 11.Deflorian F., Fedrizzi L., Rossi S., Bonora P.L. Organic coating capacitance measurement by EIS: Ideal and actual trends. Electrochim. Acta. 1999;44:4243–4249. doi: 10.1016/S0013-4686(99)00139-5. [DOI] [Google Scholar]
  • 12.Amirudin A., Thieny D. Application of electrochemical impedance spectroscopy to study the degradation of polymer-coated metals. Prog. Org. Coat. 1995;26:1–28. doi: 10.1016/0300-9440(95)00581-1. [DOI] [Google Scholar]
  • 13.Mansfeld F., Tsai C.H. Determination of coating deterioration with EIS: I. basic relationships. Corrosion. 1991;47:958–963. doi: 10.5006/1.3585209. [DOI] [Google Scholar]
  • 14.Mansfeld F., Tsai C.H. Determination of coating deterioration with EIS: Part II. development of a method for field testing of protective coatings. Corrosion. 1993;49:726–737. [Google Scholar]
  • 15.Rammelt U., Reinhard G. Application of electrochemical impedance spectroscopy (EIS) for characterizing the corrosion-protective performance of organic coatings on metals. Prog. Org. Coat. 1992;21:205–226. doi: 10.1016/0033-0655(92)87005-U. [DOI] [Google Scholar]
  • 16.Kendig M., Scully J. Basic aspects of electrochemical impedance application for the life prediction of organic coatings on metals. Corrosion. 1990;46:22–29. doi: 10.5006/1.3585061. [DOI] [Google Scholar]
  • 17.Walter G.W. A review of impedance plot methods used for corrosion performance analysis of painted metals. Corros. Sci. 1986;26:681–703. doi: 10.1016/0010-938X(86)90033-8. [DOI] [Google Scholar]
  • 18.Lindqvist S.A. Theory of dielectric properties of heterogeneous substances applied to water in a paint film. Corrosion. 1985;41:69–75. doi: 10.5006/1.3581974. [DOI] [Google Scholar]
  • 19.Grossi M., Riccò B. Electrical impedance spectroscopy (EIS) for biological analysis and food characterization: A review. J. Sens. Sens. Syst. 2017;6:303–325. doi: 10.5194/jsss-6-303-2017. [DOI] [Google Scholar]
  • 20.Davies S.J., Davenport A. The role of bioimpedance and biomarkers in helping to aid clinical decision-making of volume assessments in dialysis patients. Kidney Int. 2014;86:489–496. doi: 10.1038/ki.2014.207. [DOI] [PubMed] [Google Scholar]
  • 21.Kyle U.G., Bosaeus I., De Lorenzo A.D., Deurenberg P., Elia M., Gómez J.M., Heitmann B.L., Kent-Smith L., Melchior J.-C., Pirlich M., et al. Bioelectrical impedance analysis—Part I: Review of principles and methods. Clin. Nutr. 2004;23:1226–1243. doi: 10.1016/j.clnu.2004.06.004. [DOI] [PubMed] [Google Scholar]
  • 22.Kyle U.G., Bosaeus I., De Lorenzo A.D., Deurenberg P., Elia M., Gómez J.M., Heitmann B.L., Kent-Smith L., Melchior J.-C., Pirlich M., et al. Bioelectrical impedance analysis—Part II: Utilization in clinical practice. Clin. Nutr. 2004;23:1430–1453. doi: 10.1016/j.clnu.2004.09.012. [DOI] [PubMed] [Google Scholar]
  • 23.Jaffrin M.Y., Morel H. Body fluid volumes measurements by impedance: A review of bioimpedance spectroscopy (BIS) and bioimpedance analysis (BIA) methods. Med. Eng. Phys. 2008;30:1257–1269. doi: 10.1016/j.medengphy.2008.06.009. [DOI] [PubMed] [Google Scholar]
  • 24.Szuster B., Roj Z.S.D., Kowalski P., Sobotnicki A., Woloszyn J. Idea and measurement methods used in bioimpedance spectroscopy. Adv. Intell. Syst. Comput. 2017;623:70–78. [Google Scholar]
  • 25.Demura S., Sato S., Kitabayashi T. Percentage of total body fat as estimated by three automatic bioelectrical impedance analyzers. J. Physiol. Anthropol. Appl. Hum. Sci. 2004;23:93–99. doi: 10.2114/jpa.23.93. [DOI] [PubMed] [Google Scholar]
  • 26.Cha K., Chertow G.M., Gonzalez J., Lazarus J.M., Wilmore D.W. Multifrequency bioelectrical impedance estimates the distribution of body water. J. Appl. Physiol. 1995;79:1316–1319. doi: 10.1152/jappl.1995.79.4.1316. [DOI] [PubMed] [Google Scholar]
  • 27.Piccoli A. Bioelectric impedance measurement for fluid status assessment. Contrib. Nephrol. 2010;164:143–152. doi: 10.1159/000313727. [DOI] [PubMed] [Google Scholar]
  • 28.Piccoli A., Rossi B., Pillon L., Bucciante G. A new method for monitoring body fluid variation by bioimpedance analysis: The RXc graph. Kidney Int. 1994;46:534–539. doi: 10.1038/ki.1994.305. [DOI] [PubMed] [Google Scholar]
  • 29.Nwosu A.C., Mayland C.R., Mason S., Cox T.F., Varro A., Ellershaw J. The association of hydration status with physical signs, symptoms and survival in advanced cancer—The use of bioelectrical impedance vector analysis (BIVA) technology to evaluate fluid volume in palliative care: An observational study. PLoS ONE. 2016;11:e0163114. doi: 10.1371/journal.pone.0163114. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 30.Toso S., Piccoli A., Gusella M., Menon D., Crepaldi G., Bononi A., Ferrazzi E. Bioimpedance vector pattern in cancer patients without disease versus locally advanced or disseminated disease. Nutrition. 2003;19:510–514. doi: 10.1016/S0899-9007(02)01084-5. [DOI] [PubMed] [Google Scholar]
  • 31.Nescolarde L., Piccoli A., Román A., Núñez A., Morales R., Tamayo J., Doñate T., Rosell J. Bioelectrical impedance vector analysis in haemodialysis patients: Relation between oedema and mortality. Physiol. Meas. 2004;25:1271–1280. doi: 10.1088/0967-3334/25/5/016. [DOI] [PubMed] [Google Scholar]
  • 32.Piccoli A., Italian CAPD-BIA Study Group Bioelectric impedance vector distribution in peritoneal dialysis patients with different hydration status. Kidney Int. 2004;65:1050–1063. doi: 10.1111/j.1523-1755.2004.00467.x. [DOI] [PubMed] [Google Scholar]
  • 33.Pillon L., Piccoli A., Lowrie E.G., Lazarus J.M., Chertow G.M. Vector length as a proxy for the adequacy of ultrafiltration in hemodialysis. Kidney Int. 2004;66:1266–1271. doi: 10.1111/j.1523-1755.2004.00881.x. [DOI] [PubMed] [Google Scholar]
  • 34.Bozzetto S., Piccoli A., Montini G. Bioelectrical impedance vector analysis to evaluate relative hydration status. Pediatr. Nephrol. 2010;25:329–334. doi: 10.1007/s00467-009-1326-3. [DOI] [PubMed] [Google Scholar]
  • 35.Codognotto M., Piazza M., Frigatti P., Piccoli A. Influence of localized edema on whole-body and segmental bioelectrical impedance. Nutrition. 2008;24:569–574. doi: 10.1016/j.nut.2008.02.009. [DOI] [PubMed] [Google Scholar]
  • 36.Nwosu A.C., Morris L., Mayland C., Mason S., Pettitt A., Ellershaw J. Longitudinal bioimpedance assessments to evaluate hydration in POEMS syndrome. BMJ Support. Palliat. Care. 2016;6:369–372. doi: 10.1136/bmjspcare-2015-000991. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 37.Stewart G.N. The changes produced by the growth of bacteria in the molecular concentration and electrical conductivity of culture media. J. Exp. Med. 1899;4:235–243. doi: 10.1084/jem.4.2.235. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 38.Grossi M., Lazzarini R., Lanzoni M., Riccò B. A novel technique to control ice-cream freezing by electrical characteristics analysis. J. Food Eng. 2011;106:347–354. doi: 10.1016/j.jfoodeng.2011.05.035. [DOI] [Google Scholar]
  • 39.Pompei A., Grossi M., Lanzoni M., Perretti G., Lazzarini R., Riccò B., Matteuzzi D. Feasibility of lactobacilli concentration detection in beer by automated impedance technique. MBAA Tech. Q. 2012;49:11–18. doi: 10.1094/TQ-49-1-0315-01. [DOI] [Google Scholar]
  • 40.Chuang C.H., Du Y.C., Wu T.F., Chen C.H., Lee D.H., Chen S.M., Huang T.C., Wu H.P., Shaikh M.O. Immunosensor for the ultrasensitive and quantitative detection of bladder cancer in point of care testing. Biosens. Bioelectron. 2016;84:126–132. doi: 10.1016/j.bios.2015.12.103. [DOI] [PubMed] [Google Scholar]
  • 41.Ma H., Wallbank R.W.R., Chaji R., Li J., Suzuki Y., Jiggins C., Nathan A. An impedance-based integrated biosensor for suspended DNA characterization. Sci. Rep. 2013;3:2730. doi: 10.1038/srep02730. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 42.van Grinsven B., Vandenryt T., Duchateau S., Gaulke A., Grieten L., Thoelen R., Ingebrandt S., De Ceuninck W., Wagner P. Customized impedance spectroscopy device as possible sensor platform for biosensor applications. Phys. Status Solidi A. 2010;4:919–923. doi: 10.1002/pssa.200983305. [DOI] [Google Scholar]
  • 43.Wang Y., Ye Z., Ying Y. New trends in impedimetric biosensors for the detection of foodborne pathogenic bacteria. Sensors. 2012;12:3449–3471. doi: 10.3390/s120303449. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 44.Ibba P., Falco A., Abera B.D., Cantarella G., Petti L., Lugli P. Bio-impedance and circuit parameters: An analysis for tracking fruit ripening. Postharvest Biol. Technol. 2020;159:110978. doi: 10.1016/j.postharvbio.2019.110978. [DOI] [Google Scholar]
  • 45.Chowdhury A., Singh P., Bera T.K., Ghoshal D., Chakraborty B. Electrical impedance spectroscopic study of mandarin orange during ripening. J. Food Meas. Charact. 2017;11:1654–1664. doi: 10.1007/s11694-017-9545-y. [DOI] [Google Scholar]
  • 46.Niu J., Lee J.Y. A new approach for the determination of fish freshness by electrochemical impedance spectroscopy. J. Food Sci. 2000;65:780–785. doi: 10.1111/j.1365-2621.2000.tb13586.x. [DOI] [Google Scholar]
  • 47.Breugelmans T., Tourwé E., Van Ingelgem Y., Wielant J., Hauffman T., Haubrand R., Pintelon R., Hubin A. Odd random phase multisine EIS as a detection method for the onset of corrosion of coated steel. Electrochem. Commun. 2010;12:2–5. doi: 10.1016/j.elecom.2009.10.008. [DOI] [Google Scholar]
  • 48.Min M., Pliquett U., Nacke T., Barthel A., Annus P., Land R. Broadband excitation for short-time impedance spectroscopy. Physiol. Meas. 2008;29:185–192. doi: 10.1088/0967-3334/29/6/S16. [DOI] [PubMed] [Google Scholar]
  • 49.Min M., Parve T., Ronk A., Annus P., Paavle T. Synchronous sampling and demodulation in an instrument for multifrequency bioimpedance measurement. IEEE Trans. Instrum. Meas. 2007;56:1365–1372. doi: 10.1109/TIM.2007.900163. [DOI] [Google Scholar]
  • 50.Darowicki K., Ślepski P. Dynamic electrochemical impedance spectroscopy of the first order electrode reaction. J. Electroanal. Chem. 2003;547:1–8. doi: 10.1016/S0022-0728(03)00154-2. [DOI] [Google Scholar]
  • 51.Lyu C., Liu H., Luo W., Zhang T., Zhao W. A Fast Time Domain Measuring Technique of Electrochemical Impedance Spectroscopy Based on FFT; Proceedings of the 2018 Prognostics and System Health Management Conference; Chongqing, China. 26–28 October 2018; pp. 450–455. [Google Scholar]
  • 52.Zappen H., Ringbeck F., Sauer D.U. Application of time-resolved multi-sine impedance spectroscopy for lithium-ion battery characterization. Batteries. 2018;4:64. doi: 10.3390/batteries4040064. [DOI] [Google Scholar]
  • 53.Orlikowski J., Ryl J., Jarzynka M., Krakowiak S., Darowicki K. Instantaneous impedance monitoring of aluminum alloy 7075 corrosion in borate buffer with admixed chloride ions. Corrosion. 2015;71:828–838. doi: 10.5006/1546. [DOI] [Google Scholar]
  • 54.Valiūnienė A., Baltrūnas G., Valiūnas R., Popkirov G. Investigation of the electroreduction of silver sulfite complexes by means of electrochemical FFT impedance spectroscopy. J. Hazard. Mater. 2010;180:259–263. doi: 10.1016/j.jhazmat.2010.04.023. [DOI] [PubMed] [Google Scholar]
  • 55.Carstensen J., Foca E., Keipert S., Föll H., Leisner M., Cojocaru A. New modes of FFT impedance spectroscopy applied to semiconductor pore etching and materials characterization. Phys. Status Solidi (a) 2008;205:2485–2503. doi: 10.1002/pssa.200824033. [DOI] [Google Scholar]
  • 56.Toyoda K., Tsenkova R. Measurement of freezing process of agricultural products by impedance spectroscopy. IFAC Proc. Vol. 1998;31:89–94. doi: 10.1016/S1474-6670(17)44035-3. [DOI] [Google Scholar]
  • 57.Niedzialkowski P., Slepski P., Wysocka J., Chamier-Cieminska J., Burczyk L., Sobaszek M., Wcislo A., Ossowski T., Bogdanowicz R., Ryl J. Multisine impedimetric probing of biocatalytic reactions for label-free detection of DEFB1 gene: How to verify that your dog is not human? Sens. Actuators B Chem. 2020;323:128664. doi: 10.1016/j.snb.2020.128664. [DOI] [Google Scholar]
  • 58.Macdonald J.R. Impedance spectroscopy and its use in analyzing the steady-state AC response of solid and liquid electrolytes. J. Electroanal. Chem. Interfacial Electrochem. 1987;223:25–50. doi: 10.1016/0022-0728(87)85249-X. [DOI] [Google Scholar]
  • 59.Kobayashi K., Suzuki T.S. Development of impedance analysis software implementing a support function to find good initial guess using an interactive graphical user interface. Electrochemistry. 2020;88:39–44. doi: 10.5796/electrochemistry.19-00058. [DOI] [Google Scholar]