Concomitant Changes in Ventricular Depolarization and Repolarization and Long-Term Outcomes of Biventricular Pacing

Christoffer Polcwiartek 1,2,3,*, Daniel J Friedman 4, Kasper Emerek 2, Claus Graff 5, Peter L Sørensen 5, Joseph Kisslo 1, Zak Loring 1,6, Steen M Hansen 2, Kristian Kragholm 2, Bhupendar Tayal 2, Svend E Jensen 2,3, Peter Søgaard 2,3, Christian Torp-Pedersen 2,7, Brett D Atwater 1
PMCID: PMC7669630  NIHMSID: NIHMS1635070  PMID: 32901967

Abstract

BACKGROUND:

Biventricular (BiV) pacing increases transmural repolarization heterogeneity due to epicardial to endocardial conduction from the left ventricular (LV) lead. However, limited evidence is available on concomitant changes in ventricular depolarization and repolarization and long-term outcomes of BiV pacing. Therefore, we investigated associations of BiV pacing-induced concomitant changes in ventricular depolarization and repolarization with mortality (i.e., LV assist device, heart transplantation, or all-cause mortality) and sustained ventricular arrhythmia endpoints.

METHODS:

Consecutive BiV-defibrillator recipients with digital pre-and post-implantation electrocardiograms recorded between 2006 and 2015 at Duke University Medical Center were included. We calculated changes in QRS duration and corrected JT (JTc) interval and split them by median values. For simplicity, these variables were named QRSdecreased (≤–12 ms), QRSincreased (>–12 ms), JTcdecreased (≤22 ms), and JTcincreased (>22 ms) and subsequently used to construct four mutually exclusive groups.

RESULTS:

We included 528 patients (median age, 68 years; male, 69%). No correlation between changes in QRS duration and JTc interval was observed (p=0.295). Compared to QRSdecreased/JTcincreased, increased risk of the composite mortality endpoint was associated with QRSdecreased/JTcdecreased (hazard ratio [HR]=1.62; 95% confidence interval [CI]=1.09–2.43), QRSincreased/JTcdecreased (HR=1.86; 95% CI=1.27–2.71), and QRSincreased/JTcincreased (HR=2.25; 95% CI=1.52–3.35). No QRS/JTc group was associated with excess sustained ventricular arrhythmia risk (p=0.400).

CONCLUSION:

Among BiV-defibrillators recipients, QRSdecreased/JTcincreased was associated with the most favorable long-term survival free of LV assist device, heart transplantation, and sustained ventricular arrhythmias. Our findings suggest that improved electrical resynchronization may be achieved by assessing concomitant changes in ventricular depolarization and repolarization.

Keywords: Biventricular Pacing, Cardiac Resynchronization Therapy, Heart Failure, Implantable Cardioverter-Defibrillator, Ventricular Arrhythmias, Ventricular Repolarization

Introduction

Although biventricular (BiV) pacing remains an established therapy for patients with heart failure and left ventricular (LV) dyssynchrony, around 30% of patients derive no benefit.1 Current clinical guidelines on BiV pacing recommend using prolonged QRS duration as a marker of dyssynchrony on the standard 12-lead electrocardiogram (ECG),2 but the change in QRS duration with BiV pacing is inconsistently associated with likelihood of response.3 Novel ECG dyssynchrony markers including R-wave amplitude in V1–V2, vectorcardiographic QRS area, or time from QRS onset to the intrinsicoid deflection in V1, V5, aVL, and aVF have been shown to predict response to BiV pacing better than prolonged QRS duration.48 However, their clinical implementation is limited largely due to the lack of availability during routine clinical practice making post-implantation prognostication using standard 12-lead ECG challenging.

Assessing the absolute change in QRS duration in combination with the concomitant change in ventricular repolarization following BiV pacing is easy and readily available in real time during device implantation and follow-up using standard 12-lead ECG. BiV pacing increases transmural repolarization heterogeneity due to epicardial to endocardial wavefront propagation from the LV lead,911 but data are inconsistent on its effect on common ECG markers of ventricular repolarization including QT interval, QT dispersion, JT interval, and Tpeak to Tend interval.12 Furthermore, while some studies have associated BiV pacing-induced increases in QT interval with excess ventricular arrhythmia burden,1317 others suggest no association.18 Although QT prolongation is common with BiV pacing, and most patients benefit despite developing transmural repolarization heterogeneity,19,20 little is known about the complex interaction between concomitant changes in ventricular depolarization and repolarization and their association with long-term outcomes of BiV pacing.

As such, using a single-center cohort of recipients of BiV implantable cardioverter-defibrillators (ICDs), we investigated long-term risks of mortality and sustained ventricular arrhythmia endpoints associated with acute concomitant changes in QRS duration and corrected JT (JTc) interval. As prolonged QRS duration may contribute to artificial QT prolongation, thus making true QT assessment difficult, we utilized the JT interval as a more valid marker of ventricular repolarization in this patient population.21

Methods

Study Design and Population

This was a single-center retrospective observational cohort study performed at Duke University Medical Center, Durham, NC, United States. Our study was approved by the Duke Institutional Review Board.

Using an institutional data set prepared for the National Cardiovascular Data Registry, first-time recipients of BiV-ICDs were identified between April 01, 2006, and September 30, 2015. We restricted the cohort to patients demonstrating QRS duration ≥120 ms as assessed on digital ECG within 180 days prior to implantation and left ventricular ejection fraction (LVEF) ≤35% as assessed within 365 days prior to implantation. Patients were further required to have a digital follow-up ECG recorded within 90 days following implantation. Patients were excluded if ECG measurements were unmeasurable or they died prior to discharge. If multiple ECGs were available in the allowable pre-implantation and post-implantation period, the ECG closest to the implantation was utilized.

Electrocardiography

Clinically obtained digital ECGs were reanalyzed in the MUSE Cardiology Information System version 8.0.2.10132 using the Marquette 12SL algorithm version 241 (GE Healthcare, WI, USA) and exported in .xml format. QRS morphology was assessed by two authors (DJF and KE) blinded to outcome. P-wave, PR, QRS, and QT onsets and offsets and thereby P-wave duration, PR interval, QRS duration, and QT interval as detected by the 12SL algorithm were over read and manually corrected if needed.

We calculated the JTc interval as the Fridericia-corrected QT interval minus the QRS duration, as done previously.21 Using pre-implantation and post-implantation ECGs, we calculated changes in QRS duration and JTc interval as post-implantation ECG value minus pre-implantation ECG value and split them by median values. This was done to ensure adequate sample sizes for later analyses. For simplicity, these variables were named QRSdecreased (≤–12 ms), QRSincreased (>–12 ms), JTcdecreased (≤22 ms), and JTcincreased (>22 ms) and subsequently used to construct four mutually exclusive groups(QRSdecreased/JTcdecreased, QRSdecreased/JTcincreased, QRSincreased/JTcdecreased, and QRSincreased/JTcincreased).

Echocardiography and Clinical Data

Two-dimensional echocardiography was performed using ImageArena version 4.6 (TomTec Imaging Systems, Unterschleissheim, Germany). We derived LV volumes using a modified Simpson’s triplane method included in the software based on apical four-chamber, two-chamber, and long-axis views. We reported pre-implantation LV volumes including LVEF, left ventricular end-diastolic volume (LVEDV), and left ventricular end-systolic volume (LVESV).

We further reported data on age, sex, race, current ICD indication, prior ICD, prior cardiac arrest, comorbidities, and drug use at discharge. The renal function of patients was assessed using the estimated glomerular filtration rate (eGFR).22

Device Interrogation Data

Implantation, programming, and selection of the device were at the discretion of the treating electrophysiologist. Patients were longitudinally followed using remote patient monitoring or in-clinic device interrogation, and all reports obtained within the first 2 years following implantation were retrospectively reviewed in the electronic medical record. Devices were usually programmed to initially treat ventricular tachycardia (VT) with antitachycardia pacing (ATP) followed by high-voltage shocks if ATP failed. Ventricular fibrillation (VF) was treated with high-voltage shocks. Device electrograms from all treated arrhythmias were manually reviewed by two authors (CP and BDA) and classified as appropriate if therapy was delivered for ventricular arrhythmias meeting the preprogrammed detection criteria or inappropriate if delivered for supraventricular arrhythmias or noise.

Endpoints

Patients were followed from implantation to an incident composite mortality endpoint (i.e., LV assist device, heart transplantation, or all-cause mortality). Endpoint ascertainment was performed on May 24, 2017, using a query of the Duke Enterprise Data Unified Content Explorer that incorporates data from billing claims, hospital records, and the Social Security Death Index.23 We performed an additional analysis using appropriate ICD therapies for sustained ventricular arrhythmias.

Statistical Analysis

Continuous variables were reported as medians with 25th–75th percentiles or means with standard deviations and categorical variables as counts with percentages. Differences in variables were compared using Kruskal-Wallis, one-way analysis of variance, and chi-squared tests as appropriate. For differences in pre-implantation and post-implantation QRS duration and JTc interval, the Wilcoxon rank-sum test was used. We further displayed boxplots and scatterplots for visual comparisons.

Cumulative incidence curves of endpoints by QRS/JTc groups were computed and displayed using the Kaplan-Meier method, and differences were tested using the log-rank test. Cox regression was used to compute hazard ratios (HRs) with 95% confidence intervals (CIs) of the association of QRS/JTc groups with endpoints.

The proportional hazard assumption was tested by displaying cumulative Martingale residuals and was not violated. Interaction testing was performed by introducing an interaction term in a Cox regression model and using a likelihood ratio test to compare this model to one without an interaction term. Specifically, we tested whether the change in JTc interval (decreased vs. increased groups) modified the association of the change in QRS duration (decreased vs. increased groups) with endpoints. Linearity of continuous variables was also assessed using a likelihood ratio test comparing a linear description with a categorical one. Both age and LVEF were observed to violate the linearity assumption and were therefore included as categorical variables based on quartiles.

All models were adjusted for age quartiles, sex, QRS morphology, LVEF quartiles, ischemic cardiomyopathy, atrial fibrillation, hypertension, diabetes, and eGFR <60 mL/min/1.73 m2. We additionally adjusted for current ICD indication, prior ICD, prior cardiac arrest, and amiodarone use when assessing risk of appropriate ICD therapies for sustained ventricular arrhythmias.

Data management and analysis were performed using R version 3.6.3 (R Foundation for Statistical Computing, Vienna, Austria). A p<0.05 was considered statistically significant for all analyses except for during interaction testing where an a prior decision was made to use p<0.01 to account for multiple testing.

Results

Patient Characteristics

A total of 1001 patients underwent first-time BiV-ICD implantation during the study period. After excluding patients with missing ECGs (n=409), pre-implantation QRS duration <120 ms (n=37), LVEF >35% (n=18), unmeasurable post-implantation QRS duration (n=2), and those not surviving to hospital discharge (n=7), 528 patients remained for further analysis.

Patients had a median age of 68 (58–75) years, and most were male (69%), white (71%), had left bundle branch block (LBBB) (64%), and ischemic cardiomyopathy (56%). Median time to pre-implantation ECG was 5 (1–21) days and post-implantation ECG was 1 (1–1) day. Median changes in QRS duration and JTc interval were −12 (−26 to 7) ms and 22 (2–41) ms, respectively. When stratifying by QRS/JTc groups, patients with QRSdecreased/JTcdecreased demonstrated the highest median intrinsic QRS duration and were more often right ventricular (RV) paced. Patients with QRSincreased/JTcdecreased more often had atrial fibrillation and were more often discharged with diuretics. Finally, patients with QRSincreased/JTcincreased demonstrated the highest median intrinsic PR interval (Table 1).

Table 1:

Pre-Implantation Patient Characteristics.

QRSdecreased/JTcdecreased
(n=126)
QRSdecreased/JTcincreased
(n=142)
QRSincreased/JTcdecreased
(n=138)
QRSincreased/JTcincreased
(n=122)
Total
(n=528)
P
Value
Demographics
 Age (years) 66.5 [57.2–74.8] 68.0 [58.0–76.0] 68.0 [59.2–75.0] 68.5 [56.0–75.8] 68.0 [58.0–75.0] 0.471
 Sex (male) 86 (68.3) 91 (64.1) 101 (73.2) 86 (70.5) 364 (68.9) 0.409
 Race 0.967
  Black 22 (22.0) 30 (25.4) 25 (24.8) 27 (28.1) 104 (25.1)
  White 73 (73.0) 84 (71.2) 71 (70.3) 65 (67.7) 293 (70.6)
  Other 5 (5.0) 4 (3.4) 5 (5.0) 4 (4.2) 18 (4.3)
  NA 26 24 37 26 113
 NYHA class III/IV 103 (81.7) 120 (84.5) 121 (87.7) 104 (85.2) 448 (84.8) 0.608
 ICD indication 0.291
  Primary prevention 109 (86.5) 130 (91.5) 116 (84.1) 107 (87.7) 462 (87.5)
  Secondary prevention 17 (13.5) 12 (8.5) 22 (15.9) 15 (12.3) 66 (12.5)
 Prior ICD 32 (25.4) 22 (15.5) 27 (19.6) 27 (22.1) 108 (20.5) 0.229
 Prior cardiac arrest 10 (7.9) 11 (7.7) 6 (4.3) 13 (10.7) 40 (7.6) 0.291
ECG characteristics
 Heart rate (bpm) 75.0 [68.0–84.0] 72.0 [65.2–83.0] 74.0 [65.0–84.0] 70.0 [61.0–81.8] 73.0 [64.0–84.0] 0.376
 P-wave duration (ms) 116.0 [102.0–126.0] 110.0 [102.0–122.0] 118.0 [104.0–130.0] 116.0 [104.5–130.0] 114.0 [103.5–128.0] 0.024
  NA 45 31 52 20 148
 PR interval (ms) 176.0 [152.5–196.0] 174.0 [160.0–198.0] 185.0 [168.0–206.0] 196.0 [178.0–220.0] 182.0 [162.0–206.0] <0.001
  NA 36 21 46 19 122
 QRS duration (ms) 179.0 [160.0–199.5] 170.0 [156.0–185.5] 144.0 [132.0–160.0] 149.0 [136.0–162.0] 160.0 [144.0–180.0] <0.001
 QTc interval (ms) 514 [485.2–543.8] 487.5 [460.2–509.5] 485.5 [464.2–510.0] 469 [451.2–487.8] 486.5 [463.0–513.2] <0.001
 JTc interval (ms) 336 [315.2–354.2] 316.5 [293.5–332.8] 333 [320.0–355.0] 315 [298.2–334.0] 326 [305.0–343.2] <0.001
 QRS morphology <0.001
  LBBB 59 (46.8) 100 (70.4) 90 (65.2) 90 (73.8) 339 (64.2)
  Non-LBBB 19 (15.1) 15 (10.6) 35 (25.4) 29 (23.8) 98 (18.6)
  RV paced 48 (38.1) 27 (19.0) 13 (9.4) 3 (2.5) 91 (17.2)
Echocardiographic characteristics
 LVEF (%) 25.0 [20.0–30.0] 21.5 [20.0–27.0] 25.0 [20.0–30.0] 22.5 [17.0–30.0] 25.0 [20.0–30.0] 0.296
 LVEDV (mL) 202.0 [148.5–246.5] 181.0 [152.2–238.0] 203.0 [169.0–248.0] 209.0 [172.5–258.0] 199.0 [162.0–252.0] 0.466
  NA 31 48 33 27 139
 LVESV (mL) 147.0 [109.0–202.0] 140.0 [109.0–192.0] 158.0 [124.0–205.0] 157.0 [126.5–210.5] 154.0 [118.0–204.0] 0.673
  NA 31 48 33 27 139
Comorbidities
 Ischemic cardiomyopathy 69 (54.8) 77 (54.2) 85 (61.6) 63 (51.6) 294 (55.7) 0.403
 Atrial fibrillation 47 (37.3) 36 (25.4) 64 (46.4) 39 (32.0) 186 (35.2) 0.002
 Hypertension 84 (66.7) 98 (69.0) 103 (74.6) 92 (75.4) 377 (71.4) 0.328
 Diabetes 47 (37.3) 52 (36.6) 58 (42.0) 45 (36.9) 202 (38.3) 0.768
 eGFR (mL/min/1.73 m2) 61.0 [45.0–80.8] 62.0 [46.0–78.0] 57.0 [41.0–70.8] 56.0 [37.0–72.8] 60.0 [42.8–76.0] 0.167
Drug use
 ACEIs or ARBs 97 (77.0) 119 (83.8) 104 (75.4) 91 (74.6) 411 (77.8) 0.239
 Beta-blockers 114 (90.5) 127 (89.4) 117 (84.8) 111 (91.0) 469 (88.8) 0.355
 Diuretics 101 (80.2) 123 (86.6) 126 (91.3) 99 (81.1) 449 (85.0) 0.039
 Amiodarone 23 (18.3) 16 (11.3) 28 (20.3) 25 (20.5) 92 (17.4) 0.146

Values reported as median [25th–75th percentiles] or n(%).p Values based on Kruskal-Wallis and chi-squared tests as appropriate.

Abbreviations: ACEI, angiotensin- converting enzyme inhibitor; ARB, angiotensin II receptor blocker; ECG, electrocardiogram; eGFR, estimated glomerular filtration rate; ICD, implantable cardioverter-defibrillator; JTc, corrected JT; LBBB, left bundle branch block; LVEDV, left ventricular end- diastolic volume; LVEF, left ventricular ejection fraction; LVESV, left ventricular end-systolic olume; NA, not available; NYHA, New York Heart Association; RV, right ventricular.

Device Programming Characteristics

Median time to follow-up device interrogation was 562 (143–703) days. Patients were usually programmed DDD or DDDR (88%), with a median pacing % of 99 (96–100). Median paced and sensed atrioventricular delays were 130 (130–170) ms and 100 (100–120) ms, respectively, and ventricular-ventricular delay was most often simultaneous (62%). Quadripolar leads were implanted in 15% of patients, and AdaptivCRT was programmed on in 11% of patients, which was more common in the QRSdecreased/JTcincreased group. Of note, quadripolar leads and AdaptivCRT programming were not clinically available until later in the study period. Tachycardia therapies did not differ by QRS/JTc groups and were more often programmed as a single-zone device detecting VF at a median rate of 200 (188–200) beats per minute (bpm), with varying detection criteria. For patients programmed with a two-zone or three-zone device, VT was detected at a median rate of 171 (162–181) bpm or 188 (178–197) bpm, respectively (Table 2).

Table 2:

Device Programming Characteristics.

QRSdecreased/JTcdecreased
(n=126)
QRSdecreased/JTcincreased
(n=142)
QRSincreased/JTcdecreased
(n=138)
QRSincreased/JTcincreased
(n=122)
Total
(n=528)
P
Value
Pacing mode 0.052
 DDD or DDDR 105 (84.0) 126 (89.4) 115 (83.3) 112 (94.1) 458 (87.6)
 VVI or VVIR 19 (15.2) 15 (10.6) 20 (14.5) 7 (5.9) 61 (11.7)
 Other 1 (0.8) 0 (0.0) 3 (2.2) 0 (0.0) 4 (0.8)
 NA 1 1 0 3 5
BiV or LV-only pacing (%) 98.3 [94.9–99.3] 99.0 [96.9–99.9] 97.8 [93.6–99.5] 99.0 [96.4–99.7] 98.6 [95.5–99.7] 0.223
 Off 2 (1.6) 4 (2.8) 7 (5.1) 4 (3.3) 17 (3.2) 0.445
 NA 25 17 19 22 83
AT/AF burden (%) 0.1 [0.0–99.4] 0.1 [0.0–1.0] 0.1 [0.0–22.5] 0.1 [0.0–1.0] 0.1 [0.0–1.1] 0.009
 NA 46 38 40 39 163
Lower rate (bpm) 60.0 [50.0–70.0] 60.0 [50.0–60.0] 60.0 [50.0–70.0] 60.0 [50.0–60.0] 60.0 [50.0–60.0] 0.004
 NA 1 1 0 3 5
Upper rate (bpm) 130.0 [120.0–130.0] 130.0 [130.0–130.0] 130.0 [120.0–130.0] 130.0 [120.0–130.0] 130.0 [120.0–130.0] 0.170
 NA 1 5 3 6 15
Paced AV delay (ms) 140.0 [130.0–170.0] 130.0 [130.0–170.0] 130.0 [130.0–170.0] 140.0 [130.0–170.0] 130.0 [130.0–170.0] 0.833
 NA 30 22 29 19 100
Sensed AV delay (ms) 105.0 [100.0–120.0] 100.0 [100.0–120.0] 100.0 [100.0–120.0] 100.0 [100.0–120.0] 100.0 [100.0–120.0] 0.586
 NA 38 28 36 21 123
V-V delay (ms) 0.021
 RV, ≥10 2 (2.5) 6 (5.9) 9 (9.5) 7 (8.1) 24 (6.6)
 LV, 0 56 (70.9) 68 (67.3) 58 (61.1) 40 (46.5) 222 (61.5)
 LV, ≥10 21 (26.6) 27 (26.7) 28 (29.5) 39 (45.3) 115 (31.9)
 NA 47 41 43 36 167
Quadripolar leads 21 (16.7) 18 (12.7) 22 (15.9) 19 (15.6) 80 (15.2) 0.805
AdaptivCRT programming 13 (10.3) 23 (16.2) 7 (5.1) 14 (11.5) 57 (10.8) 0.028
VT therapy zone 1 (bpm) 170.0 [160.5–181.0] 171.0 [165.0–185.0] 171.0 [162.0–180.8] 170.5 [162.0–180.0] 171.0 [162.0–181.0] 0.832
 Off 87 (69.0) 107 (75.4) 91 (65.9) 82 (67.2) 367 (69.5) 0.330
 NA 1 2 1 2 6
VT therapy zone 2 (bpm) 188.0 [173.5–191.0] 183.0 [169.5–192.5] 188.0 [188.0–197.0] 183.5 [166.2–197.0] 188.0 [177.5–197.0] 0.635
 Off 118 (93.7) 136 (95.8) 131 (94.9) 114 (93.4) 499 (94.5) 0.818
 NA 1 2 1 2 6
VF therapy zone (bpm) 200.0 [188.0–200.0] 200.0 [188.0–200.0] 200.0 [188.0–200.0] 200.0 [188.0–200.0] 200.0 [188.0–200.0] 0.424
 NA 1 2 2 2 7
VF detection criteria 0.254
 1.0 s 6 (8.2) 7 (7.3) 2 (2.4) 3 (4.2) 18 (5.5)
 2.5 s 5 (6.8) 6 (6.2) 4 (4.7) 4 (5.6) 19 (5.8)
 12/16 intervals 10 (13.7) 8 (8.3) 13 (15.3) 9 (12.5) 40 (12.3)
 18/24 intervals 16 (21.9) 20 (20.8) 28 (32.9) 28 (38.9) 92 (28.2)
 24/32 intervals 13 (17.8) 19 (19.8) 8 (9.4) 7 (9.7) 47 (14.4)
 30/40 intervals 23 (31.5) 36 (37.5) 30 (35.3) 21 (29.2) 110 (33.7)
 NA 53 46 53 50 202

Values reported as median [25th–75th percentiles] or n(%). p Values based on Kruskal-Wallis and chi-squared tests as appropriate.

Abbreviations: AF, atrial fibrillation; AT, atrial tachycardia; AV, atrioventricular; BiV, biventricular; bpm, beats per minute; CRT, cardiac resynchronization therapy; JTc, corrected JT; LV, left ventricular, NA, not available; RV, right ventricular; VF, ventricular fibrillation; VT, ventricular tachycardia; V-V, entricular-ventricular.

Ventricular Depolarization and Repolarization Changes

Following BiV pacing, median QRS duration decreased (pre, 160 [144–180] ms vs. post, 154 [138–168] ms; p<0.001), and median JTc interval increased (pre, 326 [305–343] ms vs. post, 349 [326–369] ms; p<0.001) (Figure 1A). Around 70% of patients demonstrated post-implantation decrease in QRS duration, and 60% post-implantation increase in JTc interval. Furthermore, we observed that the change in QRS duration did not correlate with the change in JTc interval (p=0.295) (Figure 1B).

Figure 1:

Figure 1:

Boxplots (A) and correlation plot (B) of QRS duration and JTc interval.

Abbreviations: BiV, biventricular; ECG, electrocardiogram; JTc, corrected JT.

Long-Term Outcomes

With a median follow-up of 3 (2–5) years, 47% of patients reached the composite mortality endpoint of LV assist device (3%), heart transplantation (5%), and all-cause mortality (39%). Compared to QRSdecreased/JTcincreased, patients with QRSdecreased/JTcdecreased, QRSincreased/JTcdecreased, or QRSincreased/JTcincreased experienced shorter event-free survival (p<0.001) (Figure 2A). Around 12% of patients received at least one appropriate ICD therapy for sustained ventricular arrhythmias, which did not differ by QRS/JTc groups (p=0.400) (Figure 2B). This was also similar when assessing the average number of appropriate ICD therapies for VT or VF and delivered ATP or shocks (Table 3).

Figure 2:

Figure 2:

Cumulative incidence of long-term mortality (A) and sustained ventricular arrhythmia (B) endpoints by QRS/JTc groups.

Abbreviations: ICD, implantable cardioverter-defibrillator; JTc, corrected JT; LV, left ventricular.

Table 3:

Appropriate Implantable Cardioverter-Defibrillator Therapies For Sustained Ventricular Arrhythmias.

QRSdecreased/JTcdecreased
(n=126)
QRSdecreased/JTcincreased
(n=142)
QRSincreased/JTcdecreased
(n=138)
QRSincreased/JTcincreased
(n=122)
Total
(n=528)
P
Value
Total VT or VF therapies 1.3 (9.6) 0.6 (3.4) 0.6 (3.3) 0.4 (1.7) 0.7 (5.4) 0.614
 Adjusted for person-years follow-up 0.0 (0.0) 0.1 (1.5) 0.0 (0.4) 0.0 (0.0) 0.0 (0.8) 0.544
Total ATP or shock therapies 1.4 (9.9) 0.8 (4.9) 0.7 (3.5) 0.5 (1.9) 0.8 (5.8) 0.637
 Adjusted for person-years follow-up 0.0 (0.0) 0.2 (1.7) 0.1 (0.4) 0.0 (0.0) 0.1 (0.9) 0.518

Values reported as mean (standard deviation). p Value based on one-way analysis of variance test.

Abbreviations: ATP, antitachycardia pacing; JTc, corrected JT; VF, ventricular fibrillation; VT, ventricular tachycardia.

We observed that the change in JTc interval tended to modify the association of the change in QRS duration with the composite mortality endpoint (p=0.011), whereas the interaction analysis did not reach statistical significance with appropriate ICD therapies for sustained ventricular arrhythmias as endpoint (p=0.549).

Following multivariable adjustment, risk of the composite mortality endpoint remained increased for QRSdecreased/JTcdecreased (HR=1.62; 95% CI=1.09–2.43), QRSincreased/JTcdecreased (HR=1.86; 95% CI=1.27–2.71), and QRSincreased/JTcincreased (HR=2.25; 95% CI=1.52–3.35) compared to QRSdecreased/JTcincreased (Figure 3A). No QRS/JTc group was associated with excess risk of appropriate ICD therapies for sustained ventricular arrhythmias (Figure 3B).

Figure 3:

Figure 3:

Multivariable Cox regression of the association of QRS/JTc groups with long-term mortality (A) and sustained ventricular arrhythmia (B) endpoints.

Abbreviations: CI, confidence interval; HR, hazard ratio; ICD, implantable cardioverter-defibrillator; JTc, corrected JT; LV, left ventricular.

Additional Analyses

We designed various additional analyses to test the consistency and robustness of our findings.

  1. We performed an explorative analysis, in which changes in QRS duration and JTc interval were split by tertiles (Q1–Q3) instead of median values (Figure S1, Online Supplementary Material). Particularly, the combination of QRS Q3 (>−2 ms) and JTc Q3 (>34 ms) was associated with highest risk of the composite mortality endpoint (HR=4.12; 95% CI=2.16–7.84) and appropriate ICD therapies for sustained ventricular arrhythmias (HR=5.07; 95% CI=1.24–20.69).

  2. In another explorative analysis, we utilized median values of the change in Fridericia-corrected QT interval (decreased group: ≤12 ms, increased group: >22 ms) and combined these with the prespecified QRS groups (Figure S2AB, Online Supplementary Material). We observed similar trends as in the main analysis.

  3. As the study period may affect findings owing to changes in clinical practice in device implantation and programming, as well as patient selection, we additionally adjusted for implantation year, and findings did not deviate from the main analysis (Figure S3, Online Supplementary Material).

  4. Although the change in QRS duration is highly correlated with intrinsic QRS duration, QRS/JTc groups were still associated with adverse outcome in a multivariable model including intrinsic QRS duration ≥150 ms (Figure S4, Online Supplementary Material).

  5. We additionally adjusted for heart failure drug therapy at discharge including angiotensin-converting enzyme inhibitors or angiotensin II receptor blockers, beta-blockers, and diuretics and observed similar findings to the main analysis (Figure S5, Online Supplementary Material).

  6. We performed subgroup analyses of LBBB (n=339) and non-LBBB or RV-only pacing (n=189) (Figure S6AB, Online Supplementary Material). Despite sample size limitations, we observed similar trends as in the main analysis.

  7. We also performed subgroup analyses of ischemic (n=294) and non-ischemic (n=234) cardiomyopathy (Figure S7AB, Online Supplementary Material). Particularly, QRS/JTc groups were associated with more than a 3-fold increased risk of adverse outcomes in non-ischemic cardiomyopathy.

Discussion

In this study of real-life recipients of BiV-ICDs, we report a series of key findings underscoring the long-term prognostic value of a QRS/JTc risk stratification model derived from the standard 12-lead ECG, which can easily be incorporated in clinical care. First, we observed that 70% of patients demonstrated post-implantation decrease in QRS duration, and 60% post-implantation increase in JTc interval. Of note, narrowing of the QRS duration did not correlate with prolongation of the JTc interval with BiV pacing. Second, QRSdecreased/JTcincreased was associated with the most favorable long-term survival free of LV assist device, heart transplantation, and appropriate ICD therapies for sustained ventricular arrhythmias, which was independent of several critical patient characteristics including intrinsic QRS morphology and duration. Finally, our explanatory analysis revealed that patients demonstrating a decrease in QRS duration of ≤−22 ms and an increase in JTc interval of 11–34 ms with BiV pacing comprised a low-risk group.

As described in animal models and simulation studies, BiV pacing induces a reversed sequence of ventricular depolarization and repolarization due to the epicardial pacing location, thus increasing transmural repolarization heterogeneity and prolonging ventricular repolarization.10, 11 However, the effect of BiV pacing on ECG markers of ventricular repolarization is inconsistently reported,12 and prior studies on the association of BiV pacing-induced concomitant changes in ventricular repolarization with ventricular arrhythmia burden are conflicting.1318 This suggests that BiV pacing has differential effects on arrhythmogenic substrate. The proarrhythmic effect of BiV pacing may occur early due to the reversed direction of the activation of the LV wall, thus prolonging the QT interval and increasing intrinsic transmural dispersion of repolarization and creating a substrate for arrhythmias. Over time, an antiarrhythmic effect may develop particularly among responders who benefit from BiV pacing with reduction of myocardial wall stress and correction of intrinsic global dyssynchrony.19 We observed that QRS/JTc groups were strongly prognostic of LV assist device, heart transplantation, and all-cause mortality rather than appropriate ICD therapies for sustained ventricular arrhythmias. Although the prognostic value of the change in QRS duration or JTc interval as separate markers has been extensively validated, no studies on both markers combined and long-term outcomes of BiV pacing have been performed making comparison of our findings with others difficult.

Concomitant changes in QRS duration and JTc interval following BiV pacing represents a powerful summative marker that incorporates complex information. We speculate that the mechanism behind the favorable long-term event-free survival associated with QRSdecreased/JTcincreased strongly relates to both optimal electrical resynchronization and LV pacing. Therefore, while QRSdecreased is a marker of corrected intrinsic global dyssynchrony, JTcincreased is not necessarily a reflection of proarrhythmia but may also indicate contribution of LV pacing to overall LV activation making non-arrhythmic (e.g., heart failure) endpoints very likely to occur. This is further supported by the fact that prolonged ventricular repolarization is associated with diastolic dysfunction.24 Although JTcincreased portends better outcomes among patients with QRSdecreased, further studies are warranted to explore its role in the QRSincreased group. Of note, patients with QRSincreased/JTcincreased demonstrated by far the highest intrinsic PR interval suggesting overall poor cardiac conduction, and PR prolongation has recently been associated with less intrinsic global dyssynchrony and poor prognosis following BiV pacing.25

We studied only the association of acute concomitant changes in ventricular depolarization and repolarization with long-term outcomes of BiV pacing. Recently, cardiac memory where T-wave changes develop following wide QRS complex rhythms is suggested to be an important time-dependent mechanism strongly associated with hemodynamic improvement.26 Therefore, our finding of a more favorable prognosis for patients with QRSdecreased/JTcincreased may indicate a short-term effect of reverse electrical remodeling away from the configuration during wide QRS complex rhythm to one adapting better to the activation sequence during BiV pacing. However, further studies are warranted to explore long-term effects of QRS/JTc groups derived from serial ECGs on clinical outcomes of BiV pacing.

Considering that a large minority of patients derive no benefit from BiV pacing,1 our study suggests that the combination of concomitant changes in QRS duration and JTc interval to a single marker possess incremental prognostic value and may improve discrimination in identifying patients who are at increased risk of non-response to BiV pacing. Importantly, information on QRS/JTc groups is readily available prior to, during, and following implantation, and further studies are warranted to explore if the groups can assist the electrophysiologist in changing LV lead positions, device programming, or considering an alternative cardiac resynchronization therapy including multipoint pacing, His bundle pacing, or wireless endocardial pacing. Of interest, the latter has recently been associated with a more physiological LV activation, hemodynamic improvement, narrower paced QRS duration, and lesser transmural repolarization heterogeneity, which may reflect an overall improved contribution of LV pacing.27, 28

Limitations

The major limitation of our study is the retrospective observational design and lack of causes of death including arrhythmic death. The underlying mechanism of appropriate ICD therapies is also a limiting factor, thus we could not assess whether sustained ventricular arrhythmias developed due to electrolyte disturbances or premature ventricular contractions. Although we used robust statistical methods to account for differences between groups, we cannot rule out the potential for residual confounding. Specifically, we did not have data on follow-up echocardiography or LV scar burden as assessed on cardiac magnetic resonance imaging, which may contribute to ventricular repolarization abnormalities. Furthermore, we were not able to include a control group to compare the clinical efficacy of BiV-ICD versus ICD only by QRS/JTc groups. In addition, selection bias may have been introduced, in that only half of all implanted patients had follow-up ECG data available. Finally, although the median survival of the cohort was comparable to guideline-forming clinical trials, the single-center study design also limits generalizability of our findings.

Conclusion

Among recipients of BiV defibrillators, QRSdecreased/JTcincreased was associated with the most favorable long-term survival free of LV assist device, heart transplantation, and sustained ventricular arrhythmias. Our findings suggest that improved electrical resynchronization may be achieved by assessing concomitant changes in ventricular depolarization and repolarization.

Supplementary Material

Supplemental files

Funding:

This study was supported by departmental sources, the Augustinus Foundation (18-1436), Danish Physicians′ Insurance (Codan/SEB Pension), Danish Society of Cardiology, Knud Højgaards Foundation (18-02-0981), and Oticon Foundation (18-1463). The funding sources had no influence on study design; collection, analysis, or interpretation of data; writing of the manuscript; or decision to submit the manuscript for publication.

Conflicts of Interest:

CP reported receiving speaking fees from Lundbeck Pharma A/S and research grants from the Danish Heart Foundation and Eva and Henry Frænkel Memorial Foundation.

DJF reported receiving salary support from the National Institutes of Health; research grants from Abbott, Biosense Webster, Boston Scientific, and the National Cardiovascular Data Registry; and educational grants from Abbott, Biotronik, Boston Scientific, and Medtronic.

KE reported receiving research grants from the Augustinus Foundation, Danish Society of Cardiology, Gerda and Hans Hansen’s Foundation, and Knud Højgaards Foundation.

ZL is supported in part by a National Institutes of Health T-32 training grant (5T32HL069749).

SMH reported receiving speaking fees from AstraZeneca and research grants from the Danish Heart Foundation, TrygFonden, and Laerdal Foundation.

KK reported receiving speaking fees from Novartis and research grants from the Laerdal Foundation.

SEJ reported receiving research grants from the Obel Family Foundation.

PS reported receiving research grants from Biotronik and GE Healthcare and acting as adviser to Biotronik.

CTP reported receiving speaking fees from Bayer and research grants from Bayer and Biotronik.

BDA reported receiving speaking fees from Medtronic; research grants from Abbott and Boston Scientific; and acting as adviser to Abbott, Biotronik, and Medtronic.

The remaining authors have nothing to disclose.

Abbreviations

ATP

antitachycardia pacing

BiV

biventricular

bpm

beats per minute

CI

confidence interval

ECG

electrocardiogram

eGFR

estimated glomerular filtration rate

HR

hazard ratio

ICD

implantable cardioverter-defibrillator

JTc

corrected JT

LBBB

left bundle branch block

LV

left ventricular

LVEDV

left ventricular end-diastolic volume

LVEF

left ventricular ejection fraction

LVESV

left ventricular end-systolic volume

RV

right ventricular

VF

ventricular fibrillation

VT

ventricular tachycardia

References

  • 1.Daubert C, Behar N, Martins RP, Mabo P and Leclercq C. Avoiding non-responders to cardiac resynchronization therapy: a practical guide. Eur Heart J. 2017;38:1463–72. [DOI] [PubMed] [Google Scholar]
  • 2.Epstein AE, DiMarco JP, Ellenbogen KA, Estes NA 3rd, Freedman RA, Gettes LS, Gillinov AM, Gregoratos G, Hammill SC, Hayes DL, Hlatky MA, Newby LK, Page RL, Schoenfeld MH, Silka MJ, Stevenson LW and Sweeney MO. 2012 ACCF/AHA/HRS focused update incorporated into the ACCF/AHA/HRS 2008 guidelines for device-based therapy of cardiac rhythm abnormalities: a report of the American College of Cardiology Foundation/American Heart Association Task Force on Practice Guidelines and the Heart Rhythm Society. J Am Coll Cardiol. 2013;61:e6–75. [DOI] [PubMed] [Google Scholar]
  • 3.Korantzopoulos P, Zhang Z, Li G, Fragakis N and Liu T. Meta-analysis of the usefulness of change in QRS width to predict response to cardiac resynchronization therapy. Am J Cardiol. 2016;118:1368–73. [DOI] [PubMed] [Google Scholar]
  • 4.Friedman DJ, Emerek K, Hansen SM, Polcwiartek C, Sorensen PL, Loring Z, Sutter J, Sogaard P, Kisslo J, Graff C and Atwater BD. Non-invasively quantified changes in left ventricular activation predict outcomes in patients undergoing cardiac resynchronization therapy. J Cardiovasc Electrophysiol. 2019;30:2475–83. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 5.Emerek K, Friedman DJ, Sørensen PL, Hansen SM, Larsen JM, Risum N, Thøgersen AM, Graff C, Kisslo J, Søgaard P and Atwater BD. Vectorcardiographic QRS area is associated with long- term outcome after cardiac resynchronization therapy. Heart Rhythm. 2019;16:213–19. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 6.Vereckei A, Szelényi Z, Kutyifa V, Zima E, Szénási G, Kiss M, Katona G, Karádi I and Merkely B. Novel electrocardiographic dyssynchrony criteria improve patient selection for cardiac resynchronization therapy. Europace. 2018;20:97–103. [DOI] [PubMed] [Google Scholar]
  • 7.Maass AH, Vernooy K, Wijers SC, van ‘t Sant J, Cramer MJ, Meine M, Allaart CP, De Lange FJ, Prinzen FW, Gerritse B, Erdtsieck E, Scheerder COS, Hill MRS, Scholten M, Kloosterman M, Ter Horst IAH, Voors AA, Vos MA, Rienstra M and Van Gelder IC. Refining success of cardiac resynchronization therapy using a simple score predicting the amount of reverse ventricular remodelling: results from the Markers and Response to CRT (MARC) study. Europace. 2018;20:e1–e10. [DOI] [PubMed] [Google Scholar]
  • 8.Sweeney MO, van Bommel RJ, Schalij MJ, Borleffs CJ, Hellkamp AS and Bax JJ. Analysis of ventricular activation using surface electrocardiography to predict left ventricular reverse volumetric remodeling during cardiac resynchronization therapy. Circulation. 2010;121:626–34. [DOI] [PubMed] [Google Scholar]
  • 9.Fish JM, Brugada J and Antzelevitch C. Potential proarrhythmic effects of biventricular pacing. J Am Coll Cardiol. 2005;46:2340–7. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 10.Fish JM, Di Diego JM, Nesterenko V and Antzelevitch C. Epicardial activation of left ventricular wall prolongs QT interval and transmural dispersion of repolarization: implications for biventricular pacing. Circulation. 2004;109:2136–42. [DOI] [PubMed] [Google Scholar]
  • 11.Medina-Ravell VA, Lankipalli RS, Yan G-X, Antzelevitch C, Medina-Malpica NA, Medina-Malpica OA, Droogan C and Kowey PR. Effect of epicardial or biventricular pacing to prolong QT interval and increase transmural dispersion of repolarization: does resynchronization therapy pose a risk for patients predisposed to long QT or torsade de pointes? Circulation. 2003;107:740–6. [DOI] [PubMed] [Google Scholar]
  • 12.Duan X and Gao W. Effect of cardiac resynchronization therapy on ventricular repolarization: a meta-analysis. Anatol J Cardiol. 2015;15:188–95. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 13.Cvijić M, Antolič B, Klemen L and Zupan I. Repolarization heterogeneity in patients with cardiac resynchronization therapy and its relation to ventricular tachyarrhythmias. Heart Rhythm. 2018;15:1784–90. [DOI] [PubMed] [Google Scholar]
  • 14.Itoh M, Yoshida A, Fukuzawa K, Kiuchi K, Imamura K, Fujiwara R, Suzuki A, Nakanishi T, Yamashita S, Matsumoto A and Hirata K. Time-dependent effect of cardiac resynchronization therapy on ventricular repolarization and ventricular arrhythmias. Europace. 2013;15:1798–804. [DOI] [PubMed] [Google Scholar]
  • 15.Barbhaiya C, Po JR, Hanon S and Schweitzer P. Tpeak - Tend and Tpeak - Tend /QT ratio as markers of ventricular arrhythmia risk in cardiac resynchronization therapy patients. Pacing Clin Electrophysiol. 2013;36:103–8. [DOI] [PubMed] [Google Scholar]
  • 16.Lellouche N, De Diego C, Akopyan G, Boyle NG, Mahajan A, Cesario DA, Wiener I and Shivkumar K. Changes and predictive value of dispersion of repolarization parameters for appropriate therapy in patients with biventricular implantable cardioverter-defibrillators. Heart Rhythm. 2007;4:1274–83. [DOI] [PubMed] [Google Scholar]
  • 17.Chalil S, Yousef ZR, Muyhaldeen SA, Smith RE, Jordan P, Gibbs CR and Leyva F. Pacing- induced increase in QT dispersion predicts sudden cardiac death following cardiac resynchronization therapy. J Am Coll Cardiol. 2006;47:2486–92. [DOI] [PubMed] [Google Scholar]
  • 18.Dilaveris P, Giannopoulos G, Synetos A, Aggeli C, Raftopoulos L, Arsenos P, Gatzoulis K and Stefanadis C. Effect of biventricular pacing on ventricular repolarization and functional indices in patients with heart failure: lack of association with arrhythmic events. Europace. 2009;11:741–50. [DOI] [PubMed] [Google Scholar]
  • 19.Deif B, Ballantyne B, Almehmadi F, Mikhail M, McIntyre WF, Manlucu J, Yee R, Sapp JL, Roberts JD, Healey JS, Leong-Sit P and Tang AS. Cardiac resynchronization is pro-arrhythmic in the absence of reverse ventricular remodelling: a systematic review and meta-analysis. Cardiovasc Res. 2018;114:1435–44. [DOI] [PubMed] [Google Scholar]
  • 20.Tayal B, Gorcsan J 3rd, Delgado-Montero A, Marek JJ, Haugaa KH, Ryo K, Goda A, Olsen NT, Saba S, Risum N and Søgaard P. Mechanical dyssynchrony by tissue doppler cross- correlation is associated with risk for complex ventricular arrhythmias after cardiac resynchronization therapy. J Am Soc Echocardiogr. 2015;28:1474–81. [DOI] [PubMed] [Google Scholar]
  • 21.Crow RS, Hannan PJ and Folsom AR. Prognostic significance of corrected QT and corrected JT interval for incident coronary heart disease in a general population sample stratified by presence or absence of wide QRS complex: the ARIC study with 13 years of follow-up. Circulation. 2003;108:1985–9. [DOI] [PubMed] [Google Scholar]
  • 22.Levey AS, Stevens LA, Schmid CH, Zhang YL, Castro AF 3rd, Feldman HI, Kusek JW, Eggers P, Van Lente F, Greene T and Coresh J. A new equation to estimate glomerular filtration rate. Ann Intern Med. 2009;150:604–12. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 23.Horvath MM, Winfield S, Evans S, Slopek S, Shang H and Ferranti J. The DEDUCE Guided Query tool: providing simplified access to clinical data for research and quality improvement. J Biomed Inform. 2011;44:266–76. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 24.Wilcox JE, Rosenberg J, Vallakati A, Gheorghiade M and Shah SJ. Usefulness of electrocardiographic QT interval to predict left ventricular diastolic dysfunction. Am J Cardiol. 2011;108:1760–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 25.Atwater BD, Emerek K, Sørensen PL, Hansen SM, Loring Z, Graff C, Polcwiartek C, Kisslo J, Søgaard P and Friedman DJ. PR prolongation predicts inadequate resynchronization with biventricular pacing in left bundle branch block. Pacing Clin Electrophysiol. 2019;42:1477–85. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • 26.Shvilkin A, Huang HD and Josephson ME. Cardiac memory: diagnostic tool in the making. Circ Arrhythm Electrophysiol. 2015;8:475–82. [DOI] [PubMed] [Google Scholar]
  • 27.Reddy VY, Miller MA, Neuzil P, Søgaard P, Butter C, Seifert M, Delnoy PP, van Erven L, Schalji M, Boersma LVA and Riahi S. Cardiac resynchronization therapy with wireless left ventricular endocardial pacing: the SELECT-LV study. J Am Coll Cardiol. 2017;69:2119–29. [DOI] [PubMed] [Google Scholar]