Abstract
Label-free protein characterization at surfaces is commonly achieved using digestion and/or matrix application prior to mass spectrometry. We report the assignment of undigested proteins at surfaces in situ using secondary ion mass spectrometry (SIMS). Ballistic fragmentation of proteins induced by a gas cluster ion beam (GCIB) leads to peptide cleavage producing fragments for subsequent OrbitrapTM analysis. In this work we annotate 16 example proteins (up to 272 kDa) by de novo peptide sequencing and illustrate the advantages of this approach by characterizing a protein monolayer biochip and the depth distribution of proteins in human skin.
Subject terms: Proteins, Molecular imaging, Mass spectrometry
Label-free protein characterization at surfaces requires digestion or matrix application prior to mass spectrometry. Here, the authors report the assignment of undigested proteins at surfaces by de novo sequencing and apply the methodology to a protein monolayer biochip and for in situ depth profiling of proteins through human skin.
Introduction
Direct analysis of proteins such as in tissue1, at the surface of materials for use in the body2 and biochips3,4, offers insight into the molecular mechanism of disease and aids the development of medicines and medical devices. Label-free methods of imaging proteins are based on mass spectrometry (MS), with the most widely employed being matrix-assisted laser desorption/ionization (MALDI)5. MALDI allows lipids and small peptides to be imaged with high lateral resolution (1.4 μm)6. However, routine analysis of proteins with this technique uses enzymatic digestion and the application of a matrix, which may modify the structure of the sample surface and consequently the spatial distribution of analytes in two and three dimensions7. Other methods used for protein characterization at surfaces are desorption electrospray ionization8 and liquid extraction surface analysis MS9; however, these techniques have limited lateral resolution (>100 μm)10.
Time-of-flight secondary ion MS (ToF-SIMS) is a surface analysis technique able to provide in situ label- and matrix-free chemical information from sample surfaces and in three dimensions (3D) (from a few nanometers to several hundred micrometers into the sample)11. Analysis of proteins in ToF-SIMS has been limited, because they are heavily fragmented by the energetic primary ion beams traditionally applied for imaging, resulting in only single amino acid residue secondary ions which are devoid of any primary structural information. In addition, analysis of macromolecules in ToF-SIMS is limited by the mass-resolving power of the ToF analyser. Information about protein identity, conformation or orientation can be obtained using individual amino acid fragment intensities by statistical analysis as shown by Wagner and Castner12,13. This fingerprinting approach is however limited to comparison of known protein samples and their mixtures14,15. Multi-amino acid fragments and molecular ions of peptides up to 1.6 kDa can also be detected by small polyatomic metal primary ion beams Bi3+ and Au3+16. The emergence of large cluster primary ion source analysis beams, such as C60+ and gas cluster ion beams (GCIBs) such as Arn+, has enabled more information to be derived from larger peptides up to 3 kDa in ToF-SIMS17 and, recently, the detection of whole proteins up to 12 kDa has been reported18. Nonetheless, large proteins have yet to be identified without pre-analysis digestion to constituent peptides.
Recently, the 3D OrbiSIMS instrument which combines a GCIB and an OrbitrapTM analyser has been developed to aid analysis of biological samples19. Here we describe findings illustrating how this combination can be used to achieve in situ label- and matrix-free three-dimensional profiling and mapping of undigested proteins at surfaces.
Results
Peptidic fragments are generated by Argon GCIB
Only single amino acids are observed in the ToF-SIMS Bi3+ spectrum (Fig. 1a and Supplementary Fig. 1) of lysozyme (14 kDa) and peptidic secondary ions resulting from ballistic fragmentation are observed in the 3D OrbiSIMS spectra acquired using an Ar3000+ primary ion beam (Fig. 1b). The mass-resolving power of 240,000 and mass accuracy (<2 p.p.m.) of the OrbitrapTM analyser allowed the assignment of characteristic amino acid neutral losses between peptide fragments of the protein with high certainty enabling protein identification by de novo sequencing illustrated for lysozyme in Supplementary Fig. 2.
We assessed the method using 16 model protein films covering a range of functions and sizes from insulin to fibronectin at 6 and 272 kDa respectively (Fig. 1c). Sequence coverages from 5% (33 of 679 amino acids for transferrin) to 53% (68 of 129 amino acids for lysozyme) can be readily assigned (Fig. 1c and Supplementary Table 2) in the positive polarity spectra. The decrease in the proportion of the protein sequence that is assigned as protein size increases reflects a limit in the number of peptides that can be extracted from the data using this approach. The MS/MS capability of the 3D OrbiSIMS instrument further confirms the sequencing for selected ions, as shown in the example of insulin in Supplementary Fig. 3.
The peptidic secondary ions are derived from ballistic fragmentation of the intact proteins, in contrast to enzymatically digested proteins. Therefore, they cannot be assigned using existing databases employed for peptide fingerprint identification. However, the application of de novo sequencing identification of proteins directly from primary ion-induced fragments is explored herein. Unlike MS/MS of peptides, the peptidic ions observed in the positive SIMS spectra are found to be mostly sodium adducts of neutral protein fragments (Supplementary Tables 10–147). Moreover, in addition to b and y ions, normally seen in collision-induced ionization (CID)20, the spectra obtained with the 20 keV Ar3000+ GCIB also contain a, c and z ions characteristic to other methods of ionization (Supplementary Fig. 4) such as electron capture dissociation (ECD)21. Lastly, in addition to N- and C-terminal ions, a large proportion of the peaks observed in the spectrum are internal fragments of the sequence, not just those derived from the chain ends. Similar to CID fragmentation, cysteine residues participating in disulphide bonds have been found to form R-S and R-SH2 ions resulting from partial side chain cleavage (Supplementary Table 3)22,23. The presence of internal ions, formed when a y ion undergoes further fragmentation into ya, yb, yc and ya-NH3 ions is similar to high-energy collision dissociation24. The proposed structure of an Ar3000+ induced internal fragment ya is presented on an example NAWVA sequence shown in Supplementary Fig. 5a. At the C termini, we detected y ions (typical of CID) and z, z + 1 and z + 2 ions, characteristic of MALDI in-source decay25–27. The fragment structure of a z + 1 ion is proposed on an example RGCRL sequence in Supplementary Fig. 5b. The identified sequences include a number of biologically important binding and active sites listed in Table 1. Full assignments of the identified sequences are presented in Supplementary Tables 10–147.
Table 1.
Functional site | Location in the amino acid sequence | Protein |
---|---|---|
Haem-binding site | H18 and M80 | Cytochrome c |
NAD-binding site | R99 | l-lactate dehydrogenase |
R341 | Alcohol dehydrogenase | |
Substrate-binding site | T248 | l-lactate dehydrogenase |
Carbonate-binding site | T139, T471 | Transferrin |
Metal-binding site | H66 | Alcohol dehydrogenase |
H604 | Transferrin | |
H3, E6, D13, H67, D248 | Bovine serum albumin | |
E8, D10 | Concanavalin A | |
Active site | H74 | Catalase |
Fibrin- and heparin-binding site | C53-R273, T1813-T2083, C2298-G2429 | Fibronectin |
Collagen-binding site | C309-T609 | Fibronectin |
Cell attachment site | V1359-T1632 | Fibronectin |
Charge relay system | C194 | Trypsin |
H57, D102, S195 | Chymotrypsin |
The location of the functional sites is based on UniProt database28.
Analysis of the spectra of similar proteins revealed that both common and distinct fragments can be identified. Bovine serum albumin (BSA) and human serum albumin (HSA) have 76% identical sequences (441 of 584 amino acids) calculated by protein BLAST28. Their spectra contain 32% (1178 of 3709 peaks) common peaks. Common sequences and species-specific sequences were detected in the 3D OrbiSIMS spectra and are presented in Supplementary Fig. 26.
The negative polarity spectra in proteomics can provide additional information to complement the positive polarity29,30. This work focuses on the analysis of positive polarity data in order to place this emerging technology in the context of the widely used CID and MALDI methods. However, analogous fragmentation patterns were observed in the negative polarity spectra acquired with Ar3000+ GCIB and OrbitrapTM, as shown in the example of bovine α-chymotrypsin (Supplementary Tables 148–152). Ions observed in negative polarity are observed as the deprotonated N terminus fragments a, b, c and deprotonated C terminus fragments y, z-H and x ions. These types of ions are consistent with negative polarity ions reported in previous studies for methods such as CID31 or ECD21.
Protein fragments can be used for imaging and mixture deconvolution
The ability to image the proteins with high chemical specificity and high lateral resolution was demonstrated by masking a thin protein film (300 nm) using a transmission electron microscopy grid (Fig. 1b and Supplementary Fig. 6). The 3D OrbiSIMS imaging achieved lateral resolution of 10 μm using the sum of 27 of the characteristic peptidic fragments (Supplementary Fig. 6).
To further test the limits of 3D OrbiSIMS analysis of proteins, we analysed an equimolar mixture of lysozyme and insulin (Fig. 2a). It was possible to assign unique sequences from both lysozyme and insulin, allowing the unambiguous identification of both proteins in the mixture (Supplementary Figs. 7 and 8).
Protein distribution in human skin
The capability of the method to identify and locate proteins within complex 3D biological samples was assessed by depth profiling through human skin (Fig. 2b and Supplementary Fig. 9). We detected three proteins, corneodesmosin, keratin and collagen, known to be abundant in the stratum corneum, epidermal and dermal layers, respectively. Sequences characteristic to these proteins were manually assigned in situ in the skin layers where they are known to be present exclusively or predominantly. The layers of the skin in the depth profile, specifically the stratum corneum, underlying epidermis and dermis, were assigned based on the abundance of a phospholipid marker (PO3−), as described in previous analysis of skin in SIMS by Starr et al.32 (Fig. 2c). Protein locations in the skin layers were based on the immunochemically stained tissue sections gathered in the Protein Atlas33.
Hydroxyproline is an indicator of collagen, as it forms 13.5% of the collagen sequence and is rarely found in other proteins34. The intensity of hydroxyproline, metoxyproline and fragments of hydroxyproline, as well as the PGE sequence all increased with the depth of analysis, reaching a maximum in the dermis region (Supplementary Fig. 10 and Supplementary Table 4). Similarly, the characteristic sequence QHGSG was assigned to corneodesmosin, present in the stratum corneum (Supplementary Fig. 11 and Supplementary Table 5) and SFGGGG sequence assigned to keratin, abundant throughout the stratum corneum and the epidermis (Supplementary Fig. 12 and Supplementary Table 6). The method, together with the knowledge of skin composition, allowed targeted assignment of corneodesmosin, keratin and collagen in situ, which indicates the capability to profile proteins in complex biological systems.
Analysis of a protein monolayer biochip
The method developed here allows the analysis of proteins at surfaces without any pre-treatment, which is particularly advantageous for investigating protein distribution on materials when the amount of the protein on the surface is in pico- or femtomolar range. We assessed the efficacy of the 3D OrbiSIMS in these challenging biological applications by analysing a biochip produced using a method developed by Di Palma et al.35 consisting of a lysozyme monolayer immobilized on a self-assembled monolayer (SAM) of heptakis-(6-deoxy-6-thio)-β-cyclodextrin (TCD) (Fig. 2c). The samples were confirmed to be a monolayer of proteins by ellipsometry and X-ray photoelectron spectroscopy (XPS), as illustrated in Supplementary Fig. 13. Seven diagnostic fragment ions from the N and C terminus of lysozyme were detected in the samples confirming lysozyme immobilization on the biochip. This illustrates that the sensitivity of the analytical technique is sufficient to detect and assign monolayer protein coverage (Supplementary Note 1).
Automated sequence search and protein identification
Computational analysis of the human protein sequences listed in the UniProt database28 showed that 95% of the proteins in the database contain an eight-residue sequence that is unique to each particular protein (Supplementary Fig. 15b). We manually assigned 8- to 13-membered sequences in the spectra of model proteins; therefore, we propose that the GCIB ballistic sequencing method produces sufficient information to be used for protein identification (Supplementary Note 2). Existing protein identification software36 is, however, unsuitable for the data produced by the combination of GCIB and OrbitrapTM; therefore, a set of functions for de novo sequencing was developed in MATLAB. Protein identification is achieved in five steps illustrated in Fig. 3c–f as follows: chemical filtering of the OrbiSIMS data to identify peptides, intensity filtering to remove the background signals, residue identification, sequence search against a protein database, sequence scoring and protein ranking. The script is described in detail in Supplementary Note 3 and the code is available online (https://github.com/guferraz/simsdenovo/). The functions were used to automate sequence assignment of 16 known proteins and allowed unambiguous confirmation of the identity of 6 proteins (Supplementary Table 9). Future development of the script will include expanding a database of protein fingerprints, optimizing data filtering and sequence scoring to aid identification of unknown proteins and indicate the presence of proteins in complex samples.
Discussion
In summary, this work uses the 3D OrbiSIMS to achieve the first matrix- and label-free in situ assignment of undigested proteins at surfaces. A bottom-up de novo sequencing approach demonstrated sufficient sequence coverage for the correct identification of 16 example proteins up to 272 kDa. Beneficial aspects of the method include the capability to image proteins with high chemical specificity, a lateral resolution of 10 μm, sufficient sensitivity to achieve the analysis of a protein monolayer and the accurate assignment of proteins in a mixture. A protein monolayer biochip has been analysed to demonstrate the suitability of the method in the development of biosensors and materials reliant upon adsorbed protein layers. The capability to map proteins in a challenging complex biological environment is demonstrated by targeted assignment of collagen, keratin and corneodesmosin within specific layers of the human skin. Further applications of this approach include obtaining information unavailable by classical methods due to limited solubility of the protein and analysis of protein fouling of medical device materials. The method can be utilized on any instrument with a combined GCIB primary ion beam and analyser with mass-resolving power better than 240,000. An automated spectral search is proposed for assignment of protein sequences in known model protein samples. This work is a step towards adapting conventional bottom-up approach to analysing proteins at surfaces; however, similar to routine proteomic methods, confident untargeted high-throughput identification of proteins remains a challenge for extended studies.
Methods
Sample preparation
Proteins: lysozyme from chicken egg white, α-chymotrypsin from bovine pancreas, insulin solution human, recombinant BSA, horse skeletal muscle myoglobin, l-lactate dehydrogenase from rabbit muscle, human holo-transferrin, concanavalin A from Jack Bean, bovine plasma fibronectin, alcohol dehydrogenase from Saccharomyces cerevisiae, porcine lipase, bovine liver catalase, HSA and cytochrome c from equine heart were purchased from Sigma Aldrich. Porcine pepsin and porcine trypsin were purchased from Promega. Heptakis-(6-deoxy-6-thio)-β-cyclodextrin was purchased from Cyclodextrin-Shop (Tilburg, The Netherlands) to >97% purity. Hydrogen peroxide (H2O2) 30% w/w solution, sulfuric acid (H2SO4) reagent grade 95–95% and N,N-dimethylformamide (DMF) for molecular biology ≥ 99% were purchased from Sigma Aldrich. Amicon Ultra 0.5 mL Centrifugal Filters were purchased from Merck Milipore. Human skin was purchased from Tissue Solutions (https://www.tissue-solutions.com/). Female >40 years of age, ex vivo skin from cosmetic surgery. Insulin, lysozyme and chymotrypsin were purified for the 3D OrbiSIMS analysis. Protein purification was done using Milipore Amicon Ultracentrifuge filter units. The purification was done by following the protocol provided by Milipore: 0.5 mL of 100 μM protein solution in MiliQ water (18 Ω) was placed in the Amicon filter units. The units were centrifuged for 30 min at 14,000 × g. The protein concentrate was recovered by inverting the filter unit in a clean tube and centrifugation for 2 min at 1000 × g. The protein concentration was recovered to 100 μM by refilling the solution with MiliQ water to 0.5 mL. Gold slides were cleaned by ultraviolet and sonication in isopropanol for 10 min, subsequently rinsed with MiliQ water and dried under Argon flow. Purified proteins were spotted and dried onto a gold slide three times to obtain a thick protein film. Protein : peptide mixture was prepared by 1:1 volumetric mixing of 100 μM insulin and 100 μM lysozyme concentrations. The mixture of purified lysozyme and insulin was spotted and dried onto a gold slide three times to obtain a thick protein film. Gold grids for electron transmission microscopy of 200 mesh × 125 μm pitch were purchased from Sigma Aldrich. For the imaging experiment, the bare gold grid was placed on top of the protein film sample. A bare gold slide and a bare gold grid have been analysed as control samples.
For the preparation of protein monolayer biochips, the gold substrates were cleaned by immersion in piranha solution (70% H2SO4, 30% H2O2) at room temperature for 8 min, rinsed with MiliQ water and dried with an argon flow (caution: Piranha solution reacts violently with organic solvents and should be handled with care). The clean gold substrates were immersed for 24 h in 50 μM DMF solutions of heptakis-(6-deoxy-6-thio)-β-cyclodextrin (TCD) for preparation of a self assembled monolayer (SAM). Subsequently, the gold substrates were rinsed with DMF and MiliQ water, and dried under an argon flow. The TCD SAMs were immersed in a 0.05 mM lysozyme in PBS solution for 2 h. Following protein immobilization, the samples were washed with PBS buffer followed by submersion in MiliQ water for 1 min. The samples were then dried under argon and placed in the instrument for the analysis immediately.
Instrument calibration
Calibration of positive mode ToF-SIMS spectra was made using CH3+, C2H3+, C3H5+, C7H7+, and Au+ ions.
Calibration of the Orbitrap analyser was performed on the silver sample, using silver clusters following the method described by Passarelli et al.19. The Bi3+ liquid metal ion gun and the ThermoFisher Tune software were employed for calibration.
Data acquisition
For the acquisition of the 3D OrbiSIMS spectra, a 20 keV Ar3000+ analysis beam of 20 µm diameter, was used as primary ion beam. Ar3000+ with duty cycle set to 4.4% and GCIB current was 218 pA. The Q Exactive depth profile was run on the area of 200 × 200 µm using random raster mode with crater size 280 × 280 µm. The cycle time was set to 400 μs. Optimal target potential varied for different samples, oscillating at approximately +68 V. Argon gas flooding was in operation in order to aid charge compensation, pressure in the main chamber was maintained at 9.0 × 10−7 bar. The spectra were collected in positive polarity, in mass range 150–2250 m/z. The injection time was set to 500 ms. Three separate areas were analysed on each sample and each measurement lasted 30 scans, the total ion dose per measurement was 1.63 × 1011. Mass-resolving power was set to 240,000 at 200 m/z.
For the acquisition of the 3D OrbiSIMS MS/MS, a 20 keV Ar3000+ analysis beam of 20 µm diameter, was used as primary ion beam. Ar3000+ with duty cycle set to 4.4% and GCIB current was 218 pA. The Q Exactive depth profile was run on the area of 200 × 200 µm using random raster mode with crater size 280 × 280 µm. The cycle time was set to 400 μs. The target potential was set to +68 V. Argon gas flooding was in operation in order to aid charge compensation, pressure in the main chamber was maintained at 9.0 × 10−7 bar. The spectra were collected in positive polarity, in mass range 75–1125 m/z. The injection time was set to 500 ms. Three separate areas were analysed on the sample and each measurement lasted 20 scans, the total ion dose per measurement was 1.63 × 1011. Mass-resolving power was set to 240,000 at 200 m/z. Normalized collision energy was set to 35, isolation width was 10 u.
For the acquisition of the 3D OrbiSIMS image, a 20 keV Ar3000+ analysis beam of 20 µm diameter (imaging of the protein mixture) or a 20 keV Ar3000+ imaging beam of 5 µm diameter (imaging of lysozyme film masked with a gold grid) was used as primary ion beam. The 20 µm analysis beam was configured as described in the spectra acquisition section. The 5 µm imaging beam duty cycle set to 37.7% and GCIB current was 18 pA. The Q Exactive images were run on the area of 200 × 200 µm using random raster mode. The cycle time was set to 400 μs. Optimal target potential was set to +58 V. Argon gas flooding was in operation; to aid charge compensation, pressure in the main chamber was maintained at 9.0 × 10−7 bar. The images were collected in positive polarity, in mass range 150–2250 m/z. The injection time was set to 500 ms. Three separate areas were analysed on each sample and each measurement lasted one scan, the total ion dose per measurement was 1.11 × 1012. Mass-resolving power was set to 240,000 at 200 m/z.
For the acquisition of the 3D OrbiSIMS depth profile of human skin, a 20 keV Ar3000+ analysis beam of 20 µm diameter was used as the primary ion beam. Ar3000+ with duty cycle set to 15% and GCIB current was 230 pA. The Q Exactive depth profile was run on the area of 200 × 200 µm using random raster mode with crater size 280 × 280 µm. The cycle time was set to 200 μs. Optimal target potential varied for different samples, oscillating at approximately −384.4 V. The acquisition was run in temperature −170 °C. The depth profile was collected in negative polarity, in mass range 75–1125 m/z. The injection time was set to 500 ms. The measurement lasted 18,307 scans and the total ion dose per measurement was 2.65 × 1013. Mass-resolving power was set to 240,000 at 200 m/z.
For the acquisition of liquid metal ion gun (LMIG) ToF-SIMS image, a 30 keV Bi3+ primary beam was used. LMIG current was 0.05 pA. The ToF image was run on the area of 200 × 200 µm using random raster mode. The cycle time was set to 250 ms. Optimal target potential was set to +58 V. Three separate areas were analysed on each sample and each measurement lasted 15 scans; the total ion dose per measurement was 9.44 × 1010.
The depth of the sputtered material was estimated by the SurfaceLab software as 10 nm per scan and confirmed by profilometry as 300 nm after acquisition of 30 scans. Optical profilometry scans were obtained using a Zeta-20 optical microscope (Zeta Instruments, CA, USA). The scans were acquired in a Z range of 4.6 µm. The number of steps was set to 328, allowing for step size (Z resolution) of 0.014 µm.
The SAM and the SAM with immobilized lysozyme were measured using an Axis-Ultra XPS instrument (Kratos Analytical, UK) with monochromated Al Kα X-ray source, based at the National Physical Laboratory, Teddington, UK. Wide scans were acquired for analysis of organic layer thickness on the gold slide with step size 1000 meV, pass energy 160 eV, dwell time 300 ms, in a range of 1300 to −10 eV.
The thickness of the SAM and the SAM with immobilized lysozyme was measured using J.A. Woollam Co. Inc. alpha-SETM spectroscopic ellipsometer. The CompleteEASE software was employed to determine the thickness values and the calculations were based on a two-phase organic/Au model, in which the organic layer was assumed to be isotropic and assigned a refractive index of 1.50. The thickness reported is the average of three different measurements on SAM or SAM and protein samples, with the errors reported as SD.
Data analysis
Thermo Xcalibur 3.1.66.10 and IonToF SurfaceLab 7.1.116182 were used to process the results and assign the peaks. Xcalibur was used to identify amino acid neutral losses between peaks and SurfaceLab was used to create the peak lists. SurfaceLab was used to measure lateral resolution of ToF-SIMS and 3D OrbiSIMS images. Ellipsometry results were analysed using CompleteEASE 4.06. XPS results were analysed using CasaXPS 2.3.19PR1.0. Protein illustrations with highlighted sequences were generated using PyMolTM 2.3.2. on the structures exported from PDB, referenced for each protein in the Supplementary Information. Proposed chemical structures of observed protein fragments were produced using ChemDraw Professional 16.0. A custom code was written in MATLAB for automated sequence assignment and protein identification. The code is available on Github (https://github.com/guferraz/simsdenovo/) and described in Supplementary Note 3.
Reporting summary
Further information on research design is available in the Nature Research Reporting Summary linked to this article.
Supplementary information
Acknowledgements
We acknowledge the financial support of this work by the Engineering and Physical Sciences Research Council (EPSRC) grant codes: EP/P029868/1, EP/L01646X/1 and EP/P031684/1 and the National Measurement System of the UK Department for Business, Energy & Industrial Strategy.
Author contributions
A.M.K. prepared the samples and analysed the data. A.M.K. and D.J.S. acquired the data. P.M.W. modelled the statistical protein identification. G.F.T. wrote the code and analysed results for automated data analysis. D.J.S., J.W.A., A.G.S., M.R.A. and P.M.W. jointly co-supervised the work and helped with data analysis. A.M.K., G.F.T., P.M.M., P.M.W., J.W.A., A.G.S., M.R.A. and D.J.S. wrote the manuscript.
Data availability
All data have been placed in Nottingham Research Data Management Repository and is available at https://rdmc.nottingham.ac.uk/handle/internal/8604 (10.17639/nott.7070). PDB entries 1HRC, 1AZF, 1WLA, 1EPT, 4PEP, 1AB9, 4W6Z, 1JBC, 4F5S, 1AO6, 5NQB, 3RE8 and 3I40 were used for structural visualization and can be accessed at https://www.rcsb.org/.
Code availability
The code for automated sequence assignment and protein identification is available on Github (https://github.com/guferraz/simsdenovo/) and described in Supplementary Note 3.
Competing interests
The authors declare no competing interests.
Footnotes
Peer review information Nature Communications thanks Satoka Aoyagi and the other, anonymous, reviewers for their contribution to the peer review of this work.
Publisher’s note Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Supplementary information
Supplementary information is available for this paper at 10.1038/s41467-020-19445-x.
References
- 1.Chaurand P, Schwartz SA, Caprioli RM. Assessing protein patterns in disease using imaging mass spectrometry. J. Proteome Res. 2004;3:245–252. doi: 10.1021/pr0341282. [DOI] [PubMed] [Google Scholar]
- 2.Chapman RG, et al. Polymeric thin films that resist the adsorption of proteins and the adhesion of bacteria. Langmuir. 2001;17:1225–1233. doi: 10.1021/la001222d. [DOI] [Google Scholar]
- 3.Mitchell P. A perspective on protein microarrays. Nat. Biotechnol. 2002;20:225–229. doi: 10.1038/nbt0302-225. [DOI] [PubMed] [Google Scholar]
- 4.Mitchell P. In the pursuit of industrial proteomics. Nat. Biotechnol. 2003;21:233–237. doi: 10.1038/nbt0303-233. [DOI] [PubMed] [Google Scholar]
- 5.Kompauer M, Heiles S, Spengler B. Autofocusing MALDI mass spectrometry imaging of tissue sections and 3D chemical topography of nonflat surfaces. Nat. Methods. 2017;14:1156–1158. doi: 10.1038/nmeth.4433. [DOI] [PubMed] [Google Scholar]
- 6.Kompauer M, Heiles S, Spengler B. Atmospheric pressure MALDI mass spectrometry imaging of tissues and cells at 1.4-μm lateral resolution. Nat. Methods. 2016;14:90–96. doi: 10.1038/nmeth.4071. [DOI] [PubMed] [Google Scholar]
- 7.Ryan DJ, Spraggins JM, Caprioli RM. Protein identification strategies in MALDI imaging mass spectrometry: a brief review. Curr. Opin. Chem. Biol. 2019;48:64–72. doi: 10.1016/j.cbpa.2018.10.023. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Rao W, Celiz AD, Scurr DJ, Alexander MR, Barrett DA. Ambient DESI and LESA-MS analysis of proteins adsorbed to a biomaterial surface using in-situ surface tryptic digestion. J. Am. Soc. Mass Spectrom. 2013;24:1927–1936. doi: 10.1007/s13361-013-0737-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Kocurek KI, Griffiths RL, Cooper HJ. Ambient ionisation mass spectrometry for in situ analysis of intact proteins. J. Mass Spectrom. 2018;53:565–578. doi: 10.1002/jms.4087. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Han J, et al. Imaging of protein distribution in tissues using mass spectrometry: an interdisciplinary challenge. Trends Anal. Chem. 2019;112:13–28. doi: 10.1016/j.trac.2018.12.016. [DOI] [Google Scholar]
- 11.Bailey J, et al. 3D ToF-SIMS imaging of polymer multilayer films using argon cluster sputter depth profiling. ACS Appl. Mater. Interfaces. 2015;7:2654–2659. doi: 10.1021/am507663v. [DOI] [PubMed] [Google Scholar]
- 12.Wagner MS, Castner DG. Characterization of adsorbed protein films by time-of-flight secondary ion mass spectrometry with principal component analysis. Langmuir. 2001;17:4649–4660. doi: 10.1021/la001209t. [DOI] [PubMed] [Google Scholar]
- 13.Graham DJ, Castner DG. Multivariate analysis of ToF-SIMS data from multicomponent systems: the why, when, and how. Biointerphases. 2012;7:1–12. doi: 10.1007/s13758-012-0049-3. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Wagner MS, Shen M, Horbett TA, Castner DG. Quantitative time-of-flight secondary ion mass spectrometry for the characterization of multicomponent adsorbed protein films. Appl. Surf. Sci. 2003;203–204:704–709. doi: 10.1016/S0169-4332(02)00795-X. [DOI] [Google Scholar]
- 15.Bernsmann F, Lawrence N, Hannig M, Ziegler C, Gnaser H. Protein films adsorbed on experimental dental materials: ToF-SIMS with multivariate data analysis. Anal. Bioanal. Chem. 2008;391:545–554. doi: 10.1007/s00216-008-2014-z. [DOI] [PubMed] [Google Scholar]
- 16.Solé-Domènech S, Johansson B, Schalling M, Malm J, Sjövall P. Analysis of opioid and amyloid peptides using time-of-flight secondary ion mass spectrometry. Anal. Chem. 2010;82:1964–1974. doi: 10.1021/ac902712f. [DOI] [PubMed] [Google Scholar]
- 17.Yokoyama Y, et al. Peptide fragmentation and surface structural analysis by means of ToF-SIMS using large cluster ion sources. Anal. Chem. 2016;88:3592–3597. doi: 10.1021/acs.analchem.5b04133. [DOI] [PubMed] [Google Scholar]
- 18.Son JG, et al. Ar-gas cluster ion beam in ToF-SIMS for peptide and protein analysis. Biointerphases. 2020;15:021011. doi: 10.1116/6.0000105. [DOI] [PubMed] [Google Scholar]
- 19.Passarelli MK, et al. The 3D OrbiSIMS - label-free metabolic imaging with subcellular lateral resolution and high mass-resolving power. Nat. Methods. 2017;14:1175–1183. doi: 10.1038/nmeth.4504. [DOI] [PubMed] [Google Scholar]
- 20.Steen H, Mann M. The ABC’s (and XYZ’s) of peptide sequencing. Nat. Rev. Mol. Cell Biol. 2004;5:699–711. doi: 10.1038/nrm1468. [DOI] [PubMed] [Google Scholar]
- 21.Zhurov KO, Fornelli L, Wodrich MD, Laskay ÜA, Tsybin YO. Principles of electron capture and transfer dissociation mass spectrometry applied to peptide and protein structure analysis. Chem. Soc. Rev. 2013;42:5014–5030. doi: 10.1039/c3cs35477f. [DOI] [PubMed] [Google Scholar]
- 22.Na S, et al. Characterization of disulfide bonds by planned digestion and tandem mass spectrometry. Mol. Biosyst. 2015;11:1156–1164. doi: 10.1039/C4MB00688G. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Tsai PL, Chen SF, Huang SY. Mass spectrometry-based strategies for protein disulfide bond identification. Rev. Anal. Chem. 2013;32:257–268. doi: 10.1515/revac-2013-0011. [DOI] [Google Scholar]
- 24.Samgina TY, et al. Mass spectrometric de novo sequencing of natural non-tryptic peptides: comparing peculiarities of collision-induced dissociation (CID) and high energy collision dissociation (HCD) Rapid Commun. Mass Spectrom. 2014;28:2595–2604. doi: 10.1002/rcm.7049. [DOI] [PubMed] [Google Scholar]
- 25.Sakakura M, Takayama M. In-source decay and fragmentation characteristics of peptides using 5-aminosalicylic acid as a matrix in matrix-assisted laser desorption/ionization mass spectrometry. J. Am. Soc. Mass Spectrom. 2010;21:979–988. doi: 10.1016/j.jasms.2010.01.018. [DOI] [PubMed] [Google Scholar]
- 26.Asakawa D, Calligaris D, Zimmerman TA, De Pauw E. In-source decay during matrix-assisted laser desorption/ionization combined with the collisional process in an FTICR mass spectrometer. Anal. Chem. 2013;85:7809–7817. doi: 10.1021/ac401234q. [DOI] [PubMed] [Google Scholar]
- 27.Asakawa D, Takahashi H, Iwamoto S, Tanaka K. Fundamental study of hydrogen-attachment-induced peptide fragmentation occurring in the gas phase and during the matrix-assisted laser desorption/ionization process. Phys. Chem. Chem. Phys. 2018;20:13057–13067. doi: 10.1039/C8CP00733K. [DOI] [PubMed] [Google Scholar]
- 28.The UniProt Consortium UniProt. A worldwide hub of protein knowledge. Nucleic Acids Res. 2019;47:D506–D515. doi: 10.1093/nar/gky1049. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Riley NM, et al. The negative mode proteome with activated ion negative electron transfer dissociation (AI-NETD) Mol. Cell. Proteom. 2015;14:2644–2660. doi: 10.1074/mcp.M115.049726. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Tran, T. T. N., Wang, T., Hack, S., Hoffmann, P. & Bowie, J. H. Can collision-induced negative-ion fragmentations of [M-H]- anions be used to identify phosphorylation sites in peptides? Rapid Commun. Mass Spectrom. 25, 3537–3548 (2011). [DOI] [PubMed]
- 31.Lee, J. et al. One-step peptide backbone dissociations in negative-ion free radical initiated peptide sequencing mass spectrometry. Anal. Chem. 85 7044–7051 (2013). [DOI] [PubMed]
- 32.Starr NJ, et al. Enhanced vitamin C skin permeation from supramolecular hydrogels, illustrated using in situ ToF-SIMS 3D chemical profiling. Int. J. Pharm. 2019;563:21–29. doi: 10.1016/j.ijpharm.2019.03.028. [DOI] [PubMed] [Google Scholar]
- 33.Uhlén M, et al. Tissue-based map of the human proteome. Science. 2015;347:1260419. doi: 10.1126/science.1260419. [DOI] [PubMed] [Google Scholar]
- 34.Hofman K, Hall B, Cleaver H, Marshall S. High-throughput quantification of hydroxyproline for determination of collagen. Anal. Biochem. 2011;417:289–291. doi: 10.1016/j.ab.2011.06.019. [DOI] [PubMed] [Google Scholar]
- 35.Di Palma G, et al. Reversible, high-affinity surface capturing of proteins directed by supramolecular assembly. ACS Appl. Mater. Interfaces. 2019;11:8937–8944. doi: 10.1021/acsami.9b00927. [DOI] [PubMed] [Google Scholar]
- 36.Tran NH, Zhang X, Xin L, Shan B, Li M. De novo peptide sequencing by deep learning. Proc. Natl Acad. Sci. USA. 2017;114:8247–8252. doi: 10.1073/pnas.1705691114. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Timofeev VI, et al. X-ray investigation of gene-engineered human insulin crystallized from a solution containing polysialic acid. Acta Crystallogr. F. Struct. Biol. Cryst. Commun. 2010;66:259–263. doi: 10.1107/S1744309110000461. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Lim K, Nadarajah A, Forsytiie EL, Pusey ML. Locations of bromide ions in tetragonal lysozyme crystals. Acta Crystallogr. D. Biol. Crystallogr. 1998;54:899–904. doi: 10.1107/S0907444998002844. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
All data have been placed in Nottingham Research Data Management Repository and is available at https://rdmc.nottingham.ac.uk/handle/internal/8604 (10.17639/nott.7070). PDB entries 1HRC, 1AZF, 1WLA, 1EPT, 4PEP, 1AB9, 4W6Z, 1JBC, 4F5S, 1AO6, 5NQB, 3RE8 and 3I40 were used for structural visualization and can be accessed at https://www.rcsb.org/.
The code for automated sequence assignment and protein identification is available on Github (https://github.com/guferraz/simsdenovo/) and described in Supplementary Note 3.