. 2013 Apr 5;26(5):327–330. [Article in Spanish] doi: 10.1016/S0212-6567(00)78674-4

Evaluación del analizador Glucocard Memory 2 para medir la concentración de glucosa en sangre capilar

Evaluation of the glucocard memory 2 analyser for measuring glucose concentration in capillary blood

T Carrera a, L Bonamusa b, J Benet Travé c, L Almirall b, JM Navarro a,*
PMCID: PMC7675918  PMID: 11100604

Abstract

Objective

To evaluate the analytical reliability and accuracy as well as the practicability of the Glucocard Memory 2 glucose meter, intended to the control of the diabetic patient.

Design

Descriptive, crossover study. To validate an analytical instrument according to guidelines of the Spanish Society of Clinical Biochemistry and Molecular Pathology.

Setting

Primary health care, urban setting.

Participants

Ninety-three blood samples from diabetic patients were used. These samples were selected by a consecutive sampling of the tubes received in the laboratory for the diabetes follow-up protocol.

Measurements and main results

Repeatability of the system was studied analysing the within-run precision at four concentrations of glucose.We obtained coefficients of variation between 2.12% (at 410 mg/dl of glucose) and 4.17% (at 37.2 mg/dl). The linearity study allowed to check experimentally the linear response of the instrument between 27 and 485 mg/dl. The accuracy was evaluated comparing the Glucocard results with the routine procedure of our laboratory (Hitachi 747, GOD-PAP) and calculating the regression parameters with the Passing and Bablok method (y = 1,01 × –2,34) and the intraclass correlation (99%). To evaluate the clinical significance of possible deviations related with the reference laboratory method the «error Grid» analysis was used. This analysis showed that 100% of Glucocard Memory 2 results fell into the clinical accuracy zone. Practicability study showed that the instrument is very simple to use.

Conclusions

Glucocard Memory 2 is a glucose meter intended to the measurement of glucose both on capillary and venous blood that, besides its extreme simplicity of use, shows very good analytical features.

Bibliografía

  • 1.The Diabetes Control and Complications Trial Research Group The effect of intensive treatment of diabetes on the development and progression of longterm complications in isulin-dependent diabetes mellitus. N Engl J Med. 1993;329:977–986. doi: 10.1056/NEJM199309303291401. [DOI] [PubMed] [Google Scholar]
  • 2.Díez C., Vázquez R., Recio F. Evaluación cuantitativa de la linealidad de tres analizadores utilizados para medir la concentración de glucosa en sangre. Química Clínica. 1995;14:86–87. [Google Scholar]
  • 3.American Diabetes Association Self-Monitoring of Blood Glucose. Diabetes Care. 1996;19:62–66. [Google Scholar]
  • 4.Clarke W.L., Cox D., Gonder-Frederic L.A., Carter W., Phol S.L. Evaluating clinical accuracy of systems for self-monitoring of blood glucose. Diabetes Care. 1987;10:622–628. doi: 10.2337/diacare.10.5.622. [DOI] [PubMed] [Google Scholar]
  • 5.Alumà A., Alsina M.J., Armenter C., Bertrán N., Biosca C., Doladé M. Sociedad Española de Bioquímica Clínica y Patología Molecular; Barcelona: 1994. Protocolo de evaluación de analizadores automáticos por el usuario En: Martínez M director. Selección y evaluación de sistemas analíticos; pp. 45–56. [Google Scholar]
  • 6.Passing H., Bablok W. A new biometrical procedure for testing the equality of measurements from two different analytical methods. J Clin Chem Clin Biochem. 1983;21:709–720. doi: 10.1515/cclm.1983.21.11.709. [DOI] [PubMed] [Google Scholar]
  • 7.Jiménez Villa J. Comparación de métodos cuantitativos de medida. FMC. 1994;1:404–410. [Google Scholar]
  • 8.Bland J.M., Altman D.G. Comparing two methods of clinical measurement: a personal history. Int J Epidemiol. 1995;24(Supl 1):7–14. doi: 10.1093/ije/24.supplement_1.s7. [DOI] [PubMed] [Google Scholar]
  • 9.Kroll M.H., Emancipator K. A theorical evaluation of linearity. Clin Chem. 1993;39:405–413. [PubMed] [Google Scholar]
  • 10.Ibarz M., Arce M.A., Mayoral I., Iglesias M., Bergua M. Estudio de diversos aparatos de medición de glucemia capilar. Endocrinología. 1999;46:150–156. [Google Scholar]
  • 11.Medrano A., García JJ. Estudio sobre la exactitud de tres reflectómetros portátiles para la determinación de la glicemia: Reflolux II, Glucometer II y Glucoscot II. Aten Primaria. 1993;12:273–275. [PubMed] [Google Scholar]
  • 12.Henry J.B. Ediciones Científicas y Técnicas; Barcelona: 1993. Diagnóstico y tratamiento clínicos por el laboratorio; pp. 181–182. [Google Scholar]
  • 13.Trajanoski Z., Brunner G.A., Gfrerer R.J., Wach P., Pieber T.R. Accuracy of home blood glucose meters during hypoglycemia. Diabetes Care. 1996;19:1412–1415. doi: 10.2337/diacare.19.12.1412. [DOI] [PubMed] [Google Scholar]
  • 14.Lewis B.D. Laboratory evaluation of the glucocard blood glucose test meter. Clin Chem. 1992;10:2093–2095. [PubMed] [Google Scholar]
  • 15.Vendrell J., Gutiérrez C., Pérez C., Servent D., Richart C. Estudio comparativo de cuatro analizadores utilizados para la medida de la concentración de glucosa en sangre capilar. Quim Clin. 1994;13:202–206. [Google Scholar]
  • 16.Chance J.L., Dai J.L., Jones K.A., Dyer K.L., Nichols J.H. Technical evaluation of five glucose meters with data management capabilities. Am J clin Pathol. 1999;111:547–556. doi: 10.1093/ajcp/111.4.547. [DOI] [PubMed] [Google Scholar]
  • 17.Dedrick R.F., Davis W.K. What do statistics really tell us about the quality of the data from self-monitoring of blood glucose? Diabetic Medicine. 1989;6:267–273. doi: 10.1111/j.1464-5491.1989.tb01159.x. [DOI] [PubMed] [Google Scholar]

Articles from Atencion Primaria are provided here courtesy of Elsevier