Highlights
-
•
Three panels qRT-PCR assays in detecting SARS-CoV-2 was methodologically compared, including the IPBCAMS, the WHO and the CCDC assays.
-
•
Most enrolled assays exhibited Low LoDs of about 10 copies per reaction, except the ORF 1b assay in CCDC assays with a LoD of 100 copies per reaction.
-
•
No cross reaction with other respiratory viruses were observed in all of the three qRT-PCR assays.
-
•
Wide linear detection ranges from 106 to 101 copies per reaction were obtained from all of the three qRT-PCR assays.
-
•
Clinical performance suggested that the CCDC-N assay and IPBCAMS-N assay, and the IPBCAMS-ORF 1b assay performed better.
Keywords: SARS-CoV-2, qRT-PCR, Methodological evaluation, Limit of detection, Reproductivity, Clinical performance
Abstract
Quick and accurate detection of SARS-CoV-2 is critical for COVID-19 control. Dozens of real-time reverse transcription PCR (qRT-PCR) assays have been developed to meet the urgent need of COVID-19 control. However, methodological comparisons among the developed qRT-PCR assays are limited. In the present study, we evaluated the sensitivity, specificity, amplification efficiency, and linear detection ranges of three qRT-PCR assays, including the assays developed by our group (IPBCAMS), and the assays recommended by WHO and China CDC (CCDC). The three qRT-PCR assays exhibited similar sensitivities, with the limit of detection (LoD) at about 10 copies per reaction (except the ORF 1b gene assay in CCDC assays with a LoD at about 100 copies per reaction). No cross reaction with other respiratory viruses were observed in all of the three qRT-PCR assays. Wide linear detection ranges from 106 to 101 copies per reaction and acceptable reproducibility were obtained. By using 25 clinical specimens, the N gene assay of IPBCAMS assays and CCDC assays performed better (with detection rates of 92 % and 100 %, respectively) than that of the WHO assays (with a detection rate of 60 %), and the ORF 1b gene assay in IPBCAMS assays performed better (with a detection rate of 64 %) than those of the WHO assays and the CCDC assays (with detection rates of 48 % and 20 %, respectively). In conclusion, the N gene assays of CCDC assays and IPBCAMS assays and the ORF 1b gene assay of IPBCAMS assays were recommended for qRT-PCR screening of SARS-CoV-2.
1. Introduction
Since the first detection in late 2019, severe respiratory syndrome CoV-2 (SARS-CoV-2) caused Corona Virus Infectious Disease in 2019 (COVID-19) has widely spread in the world. By April 11, 2020, more than 1.7 million patients infected by SARS-CoV-2 has been reported from 185 countries (Dong et al., 2020). Given the quick increase in confirmed cases and asymptomatic infections, there are increasing demands in diagnostic tools for quick and accurate detection of the virus (Corman et al., 2020; Phan, 2020). Several real-time reverse transcription-Polymerase Chain Reaction (qRT-PCR) for the detection of SARS−COV-2 has been developed to meet the demands, including the assays by this group (IPBCAMS [Institute of Pathogen Biology, Chinese Academy of Medical Sciences] assays), and the assays by WHO (WHO assays), and the assays by China CDC (CCDC assays).
Because SARS-CoV-2 usually infected the lower respiratory tract, it is not easy to detect the viral nucleic acids from throat swabs with relatively lower viral load (Zou et al., 2020). Thus, qRT-PCR assays with higher sensitivity and better performance in the detection of SARS-CoV-2 is preferred in aiding the diagnosis of COVID-19 (Corman et al., 2020). However, most of the current available qRT-PCR assays were developed for emergency, a comprehensive methodological comparison among these assays remains unfulfilled.
To comprehensively compare the performance of currently available qRT-PCR assays for detection of SARS-CoV-2, we evaluated the sensitivity, specificity, amplification efficiency, and linear detection ranges among IPBCAMS assays, WHO assays and CCDC assays.
2. Materials and methods
2.1. Nucleic acid extraction
Clinical specimens (throat swabs and sputum) suspected of COVID-19 infection were collected from Jin Yin-Tan hospital. Nucleic acids were extracted from a volume of 200 μl clinical specimens by using NucliSens easyMag apparatus (bioMérieux, MarcyL’Etoile, France) according to the manufacturer’s instructions. A volume of 50 μl total nucleic acid eluate for each specimen was recovered and transferred into a nuclease-free vial and either tested immediately or stored at −80 °C. Clinical specimens from healthy volunteers were applied as negative control. A human house-keeping gene (GAPDH) was employed as internal control. This study was approved by the Ethical Review Board of Institute of Pathogen Biology, Chinese Academy of Medical Sciences & Peking Union Medical College, and the Institutional Review Board of Jin Yin-Tan Hospital.
2.2. Primers and probes
Sequences of primers and probes for the IPBCAMS assays were recently developed (Yiwei et al., 2020), which were designed to exactly matched the genome of SARS-CoV-2 and had low sequence identity to other coronaviruses (SARS-CoV, human CoV 229E/NL63/HKU1/OC43, and Bat SARS-like CoV). The design of the primers and probes followed several principles, including: primer length: 18−25bp; probe length:20−30 bp; melting temperature (Tm) of primers: 55–60 °C, Tm of probes 60−65 °C. Sequences of primers and probes for the WHO assays were obtained from the website of WHO (https://www.who.int/docs/default-source/coronaviruse/protocol-v2−1.pdf?sfvrsn = a9ef618c_2), and those for the CCDC assays were obtained from the website of China CDC (http://www.chinacdc.cn/jkzt/crb/zl/szkb_11803/jszl_11815/202,003/W020200309540843062947.pdf) (Table 1 ). Primers and probes were synthesized by standard phosphoramidite chemistry techniques at Qingke biotechnology Co. ltd (Beijing, China). TaqMan probes were labeled with the molecule 6-carboxy-fluroscein (FAM) at the 5′ end, and with the Blackhole Quencher 1 (BHQ1) at the 3′ end. Optimal concentrations of the primers and probes were determined by cross-titration of serial two-fold dilutions of each primer pairs/probe against a constant amount of purified RNA of SARS-CoV-2. No amplification signal was obtained in all of the three qRT-PCR assays with nucleic acids extracted from clinical specimens of healthy volunteers as template.
Table 1.
Assay | Primer/probe | Sequence (5′-3′) | Genomic location* | Amplicon | |
---|---|---|---|---|---|
N gene assay | IPBCAMS assays | Forward | AACACAAGCTTTCGGCAGAC | 29083−29102 | 195 bp |
Reverse | ACCTGTGTAGGTCAACCACG | 29278−29259 | |||
Probe | CAGCGCTTCAGCGTTCTTCGGAATGTCGC | 29200−29228 | |||
WHO assays | Forward | CACATTGGCACCCGCAATC | 28706−28724 | 127 bp | |
Reverse | GAGGAACGAGAAGAGGCTTG | 28833−28814 | |||
Probe | ACTTCCTCAAGGAACAACATTGCCA | 28753−28777 | |||
CCDC assays | Forward | GGGGAACTTCTCCTGCTAGAAT | 28881−28902 | 98 bp | |
Reverse | CAGACATTTTGCTCTCAAGCTG | 28979−28958 | |||
Probe | TTGCTGCTGCTTGACAGATT | 28934−28953 | |||
ORF 1b gene assay | IPBCAMS assays | Forward | ACGGTGACATGGTACCACAT | 13760−13779 | 215 bp |
Reverse | CTAAGTTGGCGTATACGCGT | 13975−13956 | |||
Probe | TACACAATGGCAGACCTCGTCTATGC | 13804−13829 | |||
WHO assays | Forward | GTGARATGGTCATGTGTGGCGG | 15431−15452 | 99 bp | |
Reverse | CARATGTTAAASACACTATTAGCATA | 15530−15505 | |||
Probe | CAGGTGGAACCTCATCAGGAGATGC | 15470−15494 | |||
CCDC assays | Forward | CCCTGTGGGTTTTACACTTAA | 13342−13362 | 118 bp | |
Reverse | ACGATTGTGCATCAGCTGA | 13460−13442 | |||
Probe | CCGTCTGCGGTATGTGGAAAGGTTATGG | 13377−13404 |
Numbering according to a reference genome of SARS-CoV-2 (MN908947.3).
2.3. qRT-PCR assay
The qRT-PCR assays were performed by using TaqMan Fast Virus 1-Step Master Mix (Thermo Fisher Scientific, MA, USA). Each 20 μl reaction mix contained 5 μl of 4×Fast Virus 1-Step Master Mix, 0.2 μl of 50 μM probe, 0.2 μl each of 50 μM forward and reverse primers, 12.4 μl of nuclease-free water, and 2 μl of nucleic acid extract. Amplifications were carried out in 96-well plates by using Bio-Rad instrument (Bio-Rad CFX96, CA, USA). Thermo-cycling conditions are as follows: 15 min at 50℃ for reverse transcription, 4 min at 95℃ for pre-denaturation, followed by 45 cycles of 15 s at 95℃ and 45 s at 60℃. Fluorescence measurements were taken at 60℃ of each cycle. The threshold cycle (Ct) value was determined by the point at which fluorescence exceeded a threshold limit set at the mean plus 10 stand deviations above the baseline. A result was considered positive if two or more of the SARS-CoV-2 genome targets exhibited positive results (Ct ≤ 35). A result of 35 ≤ Ct ≤ 40 was considered suspected and a repeat test was performed for result confirmation.
2.4. Preparation of RNA transcripts
RNA transcripts for N gene and ORF 1b of SARS-CoV-2 were prepared with a plasmid pEasy-T1 (TransGen Biotech, Beijing, China) with T7 promoter before the multiple cloning sites. The plasmids inserted with viral gene regions of N and ORF 1b were linearized with the restriction enzyme, BamHI, and transcribed in-vitro by using RiboMAX™ Large Scale RNA Production Systems (Promega, WI, USA), respectively. The concentrations of the RNA transcripts were determined by using NanoDrop (Thermo Fisher Scientific, CA, USA).
2.5. Validations of the three qRT-PCR assays
Limit of detection (LoD) of the three qRT-PCR assays were determined through the Probit Regression analysis. The template RNA was diluted from 100 copies per reaction through 50, 20, 10, 5 to 1 copy per reaction. Ten replicates of each dilution were applied for LoD determination. The LoD of a qRT-PCR assay was defined as the lowest detectable dilution of viral RNA transcript with a 95 % probability in the Probit Regression analysis.
Linear detection ranges of the three qRT-PCR assays were determined through 10-fold serial dilutions of the RNA transcripts as template. Fitting curve between the Ct values and quantities of the RNA transcript were applied for evaluation of the detection linearity of the assays. A good linearity was defined with a correlation coefficient ( r2) higher than 0.99 in the fitting curve. Efficiency of the three qRT-PCR assays were evaluated by the slope of the fitting curve, which was defined as 10(−1/slope) – 1.
Reproducibility of the three qRT-PCR assays were assessed by the coefficient of variation of the Ct values of the 10-fold serial diluted RNA transcripts in the intra- and the inter- assay. Triple replicates of each dilution were applied in the intra-assay. The inter-assay consisted of triple replicates of the intra-assay. The coefficient of variation was calculated by the standard deviation of the Ct values of an RNA dilution divided by the mean Ct value of the same RNA dilution.
Nucleic acids of common respiratory viruses, extracted by using a NucliSens easyMag apparatus (bioMérieux, MarcyL’Etoile, France) according to the manufacturer’s instructions, were applied as templates for evaluation of potential cross-reactions of the three qRT-PCR assays, including human coronaviruses (OC43, NL63, 229E, and HKU1), Influenza viruses (A and B), respiratory syncytial virus, parainfluenza virus (1–4), human metapneumovirus, rhinovirus, adenovirus, and bocavirus.
3. Results
3.1. Comparison of the sensitivities, reproducibilities and linear detection ranges of the three qRT-PCR assays
A serial dilution panel of the RNA transcript was tested to determine the LoD of the three qRT-PCR assays, defined as the minimum concentration with detection of 95 % by Probit regression analysis. The 95 % detection limit of the N gene assay were 9.7 copies per reaction (95 % CI 7.4–15.2), 6.6 copies per reaction (95 % CI 4.9–13.1), and 10.5 copies per reaction (95 % CI 7.9–17.1) for the IPBCAMS assay, the CCDC assay, and the WHO assay, respectively. The 95 % detection limit of the ORF 1b gene assay were 27.8 copies per reaction (95 % CI 20.7–48.9), 33.6 copies per reaction (95 % CI 27.1–55.8), and 23.1 copies per reaction (95 % CI 17.6–37.0) for the IPBCAMS assay, the CCDC assay, and the WHO assay, respectively.
The linear detection ranges of the three qRT-PCR assays were determined by using a ten-fold dilution of the RNA transcript as template. Strong linear correlations (Table 2 ) were observed between the Ct values and quantity of RNA transcripts with r2 = 0.9926, 0.9987 in the N gene assay, and r2 = 0.9953, 0.9941 in the ORF 1b assay of IPBCAMS assays and CCDC assays, respectively. Good linear correlations (Table 2) were observed in WHO assays, with r2 = 0.9750 and 0.9897 for the N gene assay and the ORF 1b assay, respectively. These results suggested that all of the three qRT-PCR assays exhibited linear detection ranges from 106 to 101 copies per reaction, while the WHO assays showed lower coefficient of linear correlation.
Table 2.
Assay | Template | Mean Ct values at quantified copy number of RNA transcript |
r2a | Slopeb | Efficiency (%)c | ||||||
---|---|---|---|---|---|---|---|---|---|---|---|
1 × 106 | 1 × 105 | 1 × 104 | 1 × 103 | 1 × 102 | 1 × 101 | ||||||
N gene assay | IPBCAMS assays | RNA transcriptd alone | 17.63 ± 0.09e | 21.99 ± 0.29 | 24.08 ± 0.09 | 28.25 ± 0.13 | 31.00 ± 0.06 | 33.73 ± 0.25 | −3.19 | 105.82 | |
RNA transcript + other viruses | 19.40 ± 0.19 | 22.40 ± 0.04 | 26.38 ± 0.09 | 29.98 ± 0.07 | 32.17 ± 0.28 | 34.51 ± 0.26 | −3.10 | 110.17 | |||
WHO assays | RNA transcript alone | 18.44 ± 0.19 | 22.65 ± 0.27 | 26.78 ± 0.32 | 29.60 ± 0.26 | 32.68 ± 0.15 | 33.97 ± 1.73 | −3.16 | 107.23 | ||
RNA transcript + other viruses | 19.51 ± 0.15 | 24.83 ± 0.36 | 26.59 ± 0.29 | 29.62 ± 0.54 | 32.62 ± 0.70 | 34.19 ± 0.51 | −2.85 | 124.32 | |||
CCDC assays | RNA transcript alone | 17.17 ± 0.09 | 20.71 ± 0.11 | 23.94 ± 0.07 | 27.57 ± 0.20 | 30.37 ± 0.12 | 33.53 ± 0.50 | −3.27 | 102.21 | ||
RNA transcript + other viruses | 18.93 ± 0.16 | 23.79 ± 0.20 | 25.66 ± 0.23 | 29.58 ± 0.52 | 31.92 ± 0.16 | 33.81 ± 0.87 | −2.93 | 119.43 | |||
ORF 1b gene assay | IPBCAMS assays | RNA transcript alone | 18.64 ± 0.14 | 22.20 ± 0.06 | 25.73 ± 0.28 | 28.83 ± 0.37 | 31.90 ± 1.42 | 34.22 ± 1.15 | −3.15 | 107.71 | |
RNA transcript + other viruses | 19.45 ± 0.06 | 22.98 ± 0.13 | 25.88 ± 0.17 | 29.37 ± 0.12 | 32.83 ± 0.40 | 34.65 ± 2.12 | −3.12 | 109.18 | |||
WHO assays | RNA transcript alone | 18.51 ± 0.11 | 21.60 ± 0.10 | 25.05 ± 0.22 | 28.27 ± 0.12 | 30.78 ± 0.09 | 32.57 ± 0.57 | −2.89 | 121.83 | ||
RNA transcript + other viruses | 19.46 ± 0.09 | 22.58 ± 0.13 | 25.75 ± 0.19 | 28.20 ± 0.20 | 30.03 ± 0.70 | 33.04 ± 0.14 | −2.65 | 138.43 | |||
CCDC assays | RNA transcript alone | 18.80 ± 0.31 | 21.96 ± 0.17 | 24.76 ± 0.18 | 28.06 ± 0.26 | 32.47 ± 0.79 | 36.16 ± 2.36 | −3.48 | 93.80 | ||
RNA transcript + other viruses | 18.67 ± 0.04 | 21.54 ± 0.11 | 24.79 ± 0.03 | 28.28 ± 0.04 | 31.09 ± 0.98 | 35.33 ± 0.59 | −3.30 | 100.92 |
r2 represents the correlation of coefficient between Ct values and quantities of RNA transcripts, ranging from 0-1, the value closer to 1 suggests higher linearity of the fitting curve between Ct values and quantities of RNA transcripts.
Slope was generated from the fitting plot between the copy number of RNA transcript and the corresponding Ct value by Excel 2010.
Efficiency = 10(−1/slope) − 1.
“RNA transcript” represents the in vitro transcribed RNA of the corresponding genes of SARS-CoV-2. “other viruses” represents the pooled RNA extracted from 15 human respiratory specimens by using Trizol. “RNA transcript + other viruses” represents a 1:1 (v/v) mixture of these two components.
Values shown are the mean of triplicate samples ± standard deviation.
The reproducibility of the three qRT-PCR assays was assessed by measuring coefficient of variation (CV) of the Ct values in the intra- and inter- assay (Table 3 ). For the N gene assay, the CVs of mean Ct values from 106 to 101 copies of RNA transcript per reaction were 0.20 %–1.33 %, 0.46 %–5.09 %, 0.27 %–1.97 % in intra-assay, and 1.06 %–2.45 %, 0.96 %–7.59 %, 1.00 %–5.51 % in inter-assay of IPBCAMS assay, WHO assay, and CCDC assay, respectively. The N gene assay in WHO assays exhibited relative high CVs with 0.46 %–5.09 % and 0.96 %–7.59 % in the intra- and inter-assay, respectively. For the ORF 1b gene assay, the CVs of mean Ct values were 0.26 %–4.45 %, 0.29 %–1.76 %, 0.71 %–6.52 % in intra-assay, and 2.17 %–5.12 %, 0.30–1.57 %, 2.63 %–4.34 % in inter-assay of IPBCAMS assays, WHO assays, and CCDC assays, respectively.
Table 3.
Assay | Copy number of RNA transcript |
|||||||
---|---|---|---|---|---|---|---|---|
1 × 106 | 1 × 105 | 1 × 104 | 1 × 103 | 1 × 102 | 1 × 101 | |||
N gene assay | IPBCAMS assays | Intra-assay | 0.52* | 1.33 | 0.37 | 0.46 | 0.20 | 1.25 |
Inter-assay | 1.06 | 2.45 | 1.49 | 1.32 | 1.37 | 1.45 | ||
WHO assays | Intra-assay | 1.08 | 1.19 | 1.12 | 0.87 | 0.46 | 5.09 | |
Inter-assay | 7.59 | 2.94 | 2.78 | 6.60 | 0.96 | 3.77 | ||
CCDC assays | Intra-assay | 0.52 | 0.54 | 0.27 | 0.74 | 0.41 | 1.97 | |
Inter-assay | 1.56 | 1.20 | 5.51 | 1.00 | 1.40 | 2.89 | ||
ORF 1b gene assay | IPBCAMS assays | Intra-assay | 0.73 | 0.26 | 1.10 | 1.30 | 4.45 | 3.36 |
Inter-assay | 4.66 | 3.85 | 2.77 | 2.17 | 5.12 | 3.50 | ||
WHO assays | Intra-assay | 0.57 | 0.47 | 0.88 | 0.41 | 0.29 | 1.76 | |
Inter-assay | 1.57 | 0.30 | 0.87 | 0.69 | 0.55 | 1.23 | ||
CCDC assays | Intra-assay | 1.66 | 0.78 | 0.71 | 0.92 | 2.45 | 6.52 | |
Inter-assay | 0.52 | 0.54 | 0.27 | 0.74 | 0.41 | 1.97 |
The coefficient of variation was calculated by the standard deviation of the Ct values of an RNA dilution divided by the mean Ct values of the corresponding RNA dilution.
Because co-infections of respiratory viruses are common, we prepared a mixture of the RNA transcript and a pooled total nucleic acid extract from respiratory specimens (RNA transcript + other extract, v: v = 1:1) as template, to evaluate the effect of co-existed viral nucleic acids on the performance of the assays. The co-existed other viral nucleic acids increased the Ct values of SARS-CoV-2 in most of the qRT-PCR assays, except for the ORF 1b gene assay of the CCDC assays (Table 2). Increased amplification efficiency of SARS-CoV-2 with the co-existed other viral nucleic acids, were observed in all the three qRT-PCR assays (Table 2).
3.2. Comparison of the specificities of the three qRT-PCR assays
To evaluate potential cross-reactions with other human respiratory viruses, the three qRT-PCR assays were examined by using human respiratory samples as templates, which were positive for human coronaviruses (OC43, NL63, 229E, or HKU1), or Influenza viruses (A or B), or respiratory syncytial virus, or parainfluenza virus (1–4), or human metapneumovirus, or rhinovirus, or adenovirus, or bocavirus. No cross reaction was observed in all of the three qRT-PCR assays (data not shown), suggesting high specificities of the three qRT-PCR assays in detecting SARS-CoV-2.
3.3. Assay evaluation with clinical specimens
The three qRT-PCR assays were evaluated with 25 clinical specimens (including 13 throat swabs and 12 sputum) from 25 suspected COVID-19 patients (Table 4 ). SARS-CoV-2 was detected from 92 % (23/25), 60 % (15/25), 100 % (25/25) by the N gene assay, and from 64 % (16/25), 48 % (12/25), 20 % (5/25) of all enrolled clinical specimens by the ORF 1b gene assay in IPBCAMS assays, WHO assays, CCDC assays, respectively (Table 4). With respect to the sputum, SARS-CoV-2 was detected from 100 % (12/12), 75 % (8/12), 100 % (12/12) of specimens by the N gene assay, and from 100 % (12/12), 75 % (8/12), 41.7 % (5/12) of specimens by the ORF 1b gene assay in IPBCAMS assays, WHO assays, CCDC assays, respectively. Referring to the throat swabs, SARS-CoV-2 was detected from 84.6 % (11/13), 53.8 % (7/13), 100 % (12/12) of specimens by the N gene assay, and from 30.8 % (4/13), 30.8 % (4/13), 0% (0/13) of specimens by the ORF 1b gene assay in IPBCAMS assays, WHO assays, CCDC assays, respectively. These results demonstrated that the N gene assay performed better than the corresponding ORF 1b gene assay of all the three qRT-PCR assays, the N gene assay in CCDC assays and ORF 1b gene assay in IPBCAMS assays performed better than the other assays.
Table 4.
Specimen ID | Specimen type | N gene assay |
ORF 1b gene assay |
||||
---|---|---|---|---|---|---|---|
IPBCAMS | WHO | CCDC | IPBCAMS | WHO | CCDC | ||
TS98 | Throat swab | 35.79 | NA | 35.42 | NA | NA | NA |
TS101 | Throat swab | 33.48 | NA | 34.24 | NA | NA | NA |
TS103 | Throat swab | NA | NA | 34.68 | NA | NA | NA |
TS105 | Throat swab | 31.5 | 35.76 | 31.64 | NA | NA | NA |
TS108 | Throat swab | 33.35 | NA | 32.11 | 33.36 | NA | NA |
TS110 | Throat swab | 29.99 | 31.73 | 29.1 | 33.57 | NA | NA |
TS165 | Throat swab | 27.34 | 30.46 | 28.14 | 31.06 | 27.84 | NA |
TS168 | Throat swab | NA | NA | 34.97 | NA | NA | NA |
TS169 | Throat swab | 33.34 | NA | 34.04 | NA | 34.2 | NA |
TS187 | Throat swab | 34.5 | 39.2 | 33.03 | NA | NA | NA |
TS188 | Throat swab | 35.03 | 35.9 | 33.57 | NA | 24.07 | NA |
TS189 | Throat swab | 31.16 | 35.43 | 31.21 | 34.04 | 30.92 | NA |
TS190 | Throat swab | 32.84 | 34.02 | 32.56 | NA | NA | NA |
TY1 | Sputum | 27.35 | 29.44 | 27.6 | 30.98 | 27.33 | NA |
TY2 | Sputum | 29.38 | 31.26 | 29.06 | 32.32 | 28.72 | NA |
TY3 | Sputum | 31.85 | NA | 31.3 | 35.84 | NA | NA |
TY4 | Sputum | 22.99 | 25.57 | 22.08 | 27.42 | 24.12 | 35.99 |
TY6 | Sputum | 25.51 | 27.52 | 25.58 | 29.03 | 25.58 | 41.54 |
TY7 | Sputum | 26.9 | 30.21 | 27.4 | 30.05 | 27.3 | 45.26 |
TY8 | Sputum | 29.21 | 31.87 | 30.06 | 33.65 | 29.84 | NA |
TY9 | Sputum | 26.29 | 28.45 | 26.34 | 30.69 | 26.03 | 46.34 |
XT1 | Sputum | 25.74 | 27.26 | 25.3 | 29.82 | 26.34 | 45.9 |
XT2 | Sputum | 31.57 | NA | 30.95 | 34.19 | NA | NA |
XT3 | Sputum | 31.14 | NA | 32.02 | 35.02 | NA | NA |
XT4 | Sputum | 32.67 | NA | 31.71 | 34.26 | NA | NA |
account (%) of positive | 23 (92 %) | 15 (60 %) | 25 (100 %) | 16 (64 %) | 12 (48 %) | 5(20 %) |
4. Discussion
Rapid and accurate detection of SARS-CoV-2 represents a fast-growing global demand, which could be met by qRT-PCR. However, the current available qRT-PCR assays for SARS-CoV-2 vary in performance, including sensitivity, specificity, reproducibility, linear detection range, etc. Moreover, because the viral load of SARS-CoV-2 in upper respiratory tract is relatively low, reliable qRT-PCR assays for the detection of SARS-CoV-2 are required for accurate diagnosis of COVID-19. We thus compared the performance of three currently wide-applied qRT-PCR assays in the detection of SARS-CoV-2.
Sensitivity is the primary demand in respiratory virus detections (Huang et al., 2018). The three qRT-PCR assays provide LoDs of 6.6–33.6 genomic copies per reaction with a detection range from 106-101 genomic copies per reaction. These results suggested that most of the three qRT-PCR assays provide high sensitivity and wide linear detection range in detecting SARS-CoV-2, except a relative lower sensitivity observed in the ORF 1b gene assay of CCDC assays.
Specificity is also essential in the detection of SARS-CoV-2, because of potential co-infections with other respiratory viruses and high host DNA background in throat swabs (Kim et al., 2013; Touzard-Romo et al., 2020; Wu et al., 2020). We evaluated the specificity of the three qRT-PCR assays with respiratory specimens positive for other common respiratory viruses. No cross reaction was observed, demonstrating high specificity of the three qRT-PCR assays in detection of SARS-CoV-2.
We next evaluated the reproducibility of the three qRT-PCR assays by measuring coefficient of variation (CV) of mean Ct values in intra- and inter- assay (Feng et al., 2018). The N gene assay in IPBCAMS assays and ORF 1b gene assay in WHO assays exhibited relative better reproducibilities with lower intra- and inter- assay CVs, which were not affected by the co-existed nucleic acids of other respiratory viruses.
Efficiency is another key parameter of qRT-PCR, reflecting the binding efficiency of primers & probe to template and the amplification efficiency of the PCR system (Resa et al., 2014). Most of the qRT-PCR assays provided good efficiencies, except an abnormal efficiency of 121.83 % observed in the ORF 1b gene assay of WHO assays. An exceptionally high efficiency indicates an increased risk of false positive (Bilgrau et al., 2016). The co-existed nucleic acids of other respiratory viruses increased the efficiency of all the three qRT-PCR assays, suggesting potential increased risk of cross-reactions between the primers & probe and background nucleic acids.
We finally evaluate the performance of the three qRT-PCR assays with clinical specimens from suspected SARS-CoV-2 infected patients (Zhang et al., 2018). Possibly because of the lower viral load in upper respiratory tract (Zou et al., 2020), the detection rate of SARS-CoV-2 was lower in throat swabs than in sputum by all of the three assays. Meanwhile, the N gene assay performed better than the corresponding ORF 1b gene assay in all of the three qRT-PCR assays. For the N gene assay, IPBCAMS assays and CCDC assays performed better than WHO assays, both of which could detect SARS-CoV-2 from more than 90 % of the suspected specimens. For the ORF 1b gene assay, IPBCAMS assays performed better than WHO assays and CCDC assays, with a detection rate of 64 %.
The results of qRT-PCR assay validations would be more precise with more clinical specimens. Thus, the results of the present study generated from 25 clinical specimens should be limited. Studies enrolled more clinical specimens covering all COVID-19 infected populations were recommended to make more precise validation of qRT-PCR assays for SARS-CoV-2.
In conclusion, we performed methodological evaluations on three widely-applied qRT-PCR assays for the detection of SARS-CoV-2. Although most of the evaluated assays exhibited good sensitivity, specificity, reproducibility and wide linear detection range, performance test with clinical specimens from suspected COVID-19 patients suggested that the N gene assay in IPBCAMS assays and CCDC assays, and the ORF 1b gene assays in IPBCAMS assays were the preferred qRT-PCR assays for accurate detection of SARS-CoV-2.
Data availability
The original data will be available upon request.
CRediT authorship contribution statement
Yan Xiao: Writing - original draft. Zhen Li: Software, Validation. Xinming Wang: Investigation. Yingying Wang: . Ying Wang: Investigation. Geng Wang: Investigation. Lili Ren: Conceptualization, Data curation. Jianguo Li: Writing - review & editing, Supervision.
Declaration of Competing Interest
The authors report no declarations of interest.
Acknowledgements
We would like to thank the clinicians who contributed to sample collection and transportation. This study was funded in part by the Project from the Ministry of Science and Technology of China (2020YFC0841200), the National Major Science & Technology Project for Control and Prevention of Major Infectious Diseases of China (2017ZX10103004, 2018ZX10305409-001-007), the Chinese Academy of Medical Sciences (CAMS) Innovation Fund for Medical Sciences (2020HY320001), the key R&D plan of Shanxi Province (202003D31003/GZ, 202003D31005/GZ) and the non-profit Central Research Institute Fund of Chinese Academy of Medical Sciences (2019PT310029).
References
- Bilgrau A.E., Falgreen S., Petersen A., Kjeldsen M.K., Bodker J.S., Johnsen H.E., Dybkaer K., Bogsted M. Unaccounted uncertainty from qPCR efficiency estimates entails uncontrolled false positive rates. BMC Bioinformatics. 2016;17:159. doi: 10.1186/s12859-016-0997-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Corman V.M., Landt O., Kaiser M., Molenkamp R., Meijer A., Chu D.K., Bleicker T., Brunink S., Schneider J., Schmidt M.L., Mulders D.G., Haagmans B.L., van der Veer B., van den Brink S., Wijsman L., Goderski G., Romette J.L., Ellis J., Zambon M., Peiris M., Goossens H., Reusken C., Koopmans M.P., Drosten C. Detection of 2019 novel coronavirus (2019-nCoV) by real-time RT-PCR. Euro Surveill. 2020:25. doi: 10.2807/1560-7917.ES.2020.25.3.2000045. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Dong E., Du H., Gardner L. An interactive web-based dashboard to track COVID-19 in real time. Lancet Infect. Dis. 2020 doi: 10.1016/S1473-3099(20)30120-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Feng Z.S., Zhao L., Wang J., Qiu F.Z., Zhao M.C., Wang L., Duan S.X., Zhang R.Q., Chen C., Qi J.J., Fan T., Li G.X., Ma X.J. A multiplex one-tube nested real time RT-PCR assay for simultaneous detection of respiratory syncytial virus, human rhinovirus and human metapneumovirus. Virol. J. 2018;15:167. doi: 10.1186/s12985-018-1061-0. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Huang H.S., Tsai C.L., Chang J., Hsu T.C., Lin S., Lee C.C. Multiplex PCR system for the rapid diagnosis of respiratory virus infection: systematic review and meta-analysis. Clin. Microbiol. Infect. 2018;24:1055–1063. doi: 10.1016/j.cmi.2017.11.018. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Kim H.K., Oh S.H., Yun K.A., Sung H., Kim M.N. Comparison of Anyplex II RV16 with the xTAG respiratory viral panel and Seeplex RV15 for detection of respiratory viruses. J. Clin. Microbiol. 2013;51:1137–1141. doi: 10.1128/JCM.02958-12. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Phan T. Novel coronavirus: from discovery to clinical diagnostics. Infect. Genet. Evol. 2020 doi: 10.1016/j.meegid.2020.104211. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Resa C., Magro S., Marechal P., Barranger C., Joannes M., Miszczak F., Vabret A. Development of an efficient qRT-PCR assay for quality control and cellular quantification of respiratory samples. J. Clin. Virol. 2014;60:270–275. doi: 10.1016/j.jcv.2014.03.019. [DOI] [PubMed] [Google Scholar]
- Touzard-Romo F., Tape C., Lonks J.R. Co-infection with SARS-CoV-2 and human metapneumovirus. R I Med J. 2020;103(2013):75–76. [PubMed] [Google Scholar]
- Wu X., Cai Y., Huang X., Yu X., Zhao L., Wang F., Li Q., Gu S., Xu T., Li Y., Lu B., Zhan Q. Co-infection with SARS-CoV-2 and influenza a virus in patient with pneumonia, China. Emerg Infect Dis. 2020;26 doi: 10.3201/eid2606.200299. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Yiwei L., Yingying W., Xinming W., Yan X., Lan C., Li G., Jianguo L., Lili R., Jianwei W. Development of two TaqMan real-time reverse transcription-PCR assays for the detection of severe acute respiratory syndrome coronavirus-2. Biosafety Health. 2020 doi: 10.1016/j.bsheal.2020.07.009. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zhang D., Mao H., Lou X., Pan J., Yan H., Tang H., Shu Y., Zhao Y., Cheng X., Tao H., Zhang Y., Ma X. Clinical evaluation of a panel of multiplex quantitative real-time reverse transcription polymerase chain reaction assays for the detection of 16 respiratory viruses associated with community-acquired pneumonia. Arch. Virol. 2018;163:2855–2860. doi: 10.1007/s00705-018-3921-8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- Zou L., Ruan F., Huang M., Liang L., Huang H., Hong Z., Yu J., Kang M., Song Y., Xia J., Guo Q., Song T., He J., Yen H.L., Peiris M., Wu J. SARS-CoV-2 viral load in upper respiratory specimens of infected patients. N. Engl. J. Med. 2020 doi: 10.1056/NEJMc2001737. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Data Availability Statement
The original data will be available upon request.