Abstract
Purpose:
To develop fast multi-slice apparent T1 (T1app) mapping for accurate cerebral blood flow (CBF) quantification with arterial spin labeling (ASL).
Methods:
Fast multi-slice T1app was measured using a modified inversion recovery echo planar imaging (EPI) sequence with simultaneous application of ASL tagging RF and gradient pulses. The fast multi-slice T1app measurement was compared with the single-slice T1app imaging approach, repeated per slice. CBF was assessed in healthy adult Wistar rats (N = 5) and acute stroke rats 24 hours after a transient middle cerebral artery occlusion (N=5).
Results:
The fast multi-slice T1app measurement was in good agreement with that of a single-slice T1app imaging approach (Lin’s concordance correlation coefficient (CCC) = 0.92). CBF calculated using T1app reasonably accounts for the finite labeling RF duration, while the routine T1-normalized ASL MRI underestimated the CBF, particularly at short labeling durations. In acute stroke rats, the labeling time and the CBF difference (ΔCBF) between the contralateral normal area and the ischemic lesion were significantly correlated when using T1-normalized perfusion calculation (R = 0.844, P = 0.035). In comparison, T1app-normalized ΔCBF had little labeling time dependence based on the linear regression equation of ΔCBF = −0.0247*τ + 1.579 ml/(g∙min) (R = −0.352, P = 0.494).
Conclusions:
Our study demonstrated fast multi-slice T1app imaging, which improves the accuracy and reproducibility of CBF measurement.
Keywords: Arterial spin labeling (ASL), Cerebral Blood Flow (CBF), Magnetization transfer (MT), Stroke, Apparent longitudinal relaxation time (T1app)
Introduction
Cerebral blood flow (CBF) provides hemodynamic characterization that is often informative in neurological disorders. The arterial spin labeling (ASL) MRI allows a noninvasive CBF measurement, serving as a valuable research tool that complements dynamic susceptibility contrast perfusion imaging (1–4). Briefly, ASL MRI inverts the arterial blood signal by simultaneous application of tagging radio-frequency (RF) and gradient pulses, and the signal difference between the label and reference scans is proportional to the CBF (5–7). Continuous ASL (CASL) MRI has been commonly used in the preclinical setting, owing to its non-invasiveness that enables repeated measurement (8–10). However, the ASL tagging pulse not only saturates the arterial blood signal at the labeling plane but also induces off-resonance semisolid magnetization transfer (MT) effect at the imaging volume, particularly so when a volume coil is used for ASL tagging (11–13). Under such conditions, the apparent longitudinal relaxation time is reduced from the intrinsic one, the correction of which is necessary to improve the accuracy of CASL measurement of CBF (14,15).
CBF calculation requires either the longitudinal relaxation time without RF saturation (T1) or that under off-resonance RF saturation (T1app), which yields equivalent results in the case of complete saturation of semisolid macromolecules (12,16–18). However, T1-normalized CBF quantification assumes long RF saturation, which is often not fulfilled to reduce the scan time (19). The use of a single T1app for brain CBF calculation is overly simplistic because of heterogeneous T1app (20). Eng et al. modified the inversion recovery sequence with off-resonance saturation to measure single-slice T1app (21). Because the MT effect in the CASL scan is slice dependent, the T1app mapping has to be repeated per slice and hence time-consuming. Recently, an ultrafast chemical exchange saturation transfer (CEST) magnetic resonance spectroscopy (MRS) technique has been demonstrated in phantoms (22–24), and preliminarily ex vivo and in vivo (25–27). This approach simultaneously applies RF and gradient pulses, encoding the CEST spectrum along the readout dimension. Building on the concept of superfast CEST MRS, we amended the inversion recovery sequence with simultaneous application of RF and gradient pulses that match those in the CASL scan to measure multi-slice T1app, and demonstrated improved CBF measurement in healthy and acute stroke rats.
Theory
The kinetic model for ASL perfusion measurement has been established (5,28). Under the condition that the arterial transit time (ATT) is equal to or larger than the post-label delay (PLD), the arterial blood magnetization is given by I0/λ, where I0 is the tissue magnetization without RF irradiation, and λ is the brain/blood partition coefficient. CBF can be shown to be (4,29),
| (1) |
where Icontrol and Ilabel are control and label signals, respectively, T1b is the arterial blood longitudinal relaxation time, α is the labeling efficiency, δ is the transit time, and τ is the ASL tagging duration. Under the condition of long CASL tagging, the term 1−exp(−τ/T1app) is approximately 1. The CBF is given as (17),
| (2) |
in which and are the steady-state control and label signals, respectively. Zhang et al. derived the steady-state effect of cross-relaxation with macromolecules on perfusion measurements (12). Under the assumption of complete saturation of macromolecules, we have,
| (3) |
| (4) |
where T1int is the intrinsic tissue longitudinal relaxation time, kfor is the exchange rate from the tissue water to the macromolecules, and T1app depends on the RF irradiation (30). We can show that
| (5) |
where T1 is tissue longitudinal relaxation time without RF saturation . Although Eq. (5) quantifies CBF under the condition of a long tagging pulse, Eq. (1) needs to be used when the tagging pulse is not sufficiently long. Importantly, Eq. (1) requires the measurements of I0 and T1app.
Methods
Pulse sequence
Fig. 1 shows the routine single-slice and the proposed fast multi-slice T1app pulse sequences. The single-slice scheme applies an MT pulse at a chosen frequency offset before and during the inversion time (Fig. 1a). This is because CASL MRI induces different MT effects due to their varying distances from the labeling plane, and the single-slice T1app sequence has to be repeated for each slice. The proposed multi-slice T1app sequence applies the same gradient and RF pulses as those of the CASL MRI, resulting in a slice-dependent MT effect identical to that during the CASL scan (Fig. 1b). This modification allows multi-slice T1app-weighting within a single repetition time (TR). The labeling duration and the inversion time are denoted as τlab and τinv, respectively.
Figure 1:

Illustration of the routine single-slice (Fig. 1a) and the proposed fast multi-slice T1app (Fig. 1b) pulse sequences.
Animal preparations
The study has been approved by the local Animal Care and Use Committee. Ten adult male Wistar rats (306 ± 21 g) were examined, including five normal rats and five acute stroke rats 24 hr after 90 min transient middle cerebral artery occlusion (tMCAO). Rats were anesthetized with 1.5–2% isoflurane/air mixture with heart rate, blood oxygen saturation, and rectal temperature continuously monitored. The body temperature was maintained within the physiological range with a circulating warm water jacket. For the tMCAO model, a silicone-coated nylon filament was inserted into the right internal carotid artery to block the origin of the middle cerebral artery under the Laser Doppler Flowmetry monitoring. The filament was withdrawn 90 min post occlusion, after which rats were allowed to recover from the anesthesia. Rats that failed to demonstrate at least 70% CBF reduction from the baseline or suffered from subarachnoid hemorrhage (no CBF recovery after filament withdrawal) were excluded. MRI scans were performed 24 hr after the MCAO.
MRI experiments
All experiments were conducted on a 4.7 T MRI scanner (Bruker Biospec, Billerica, MA) with a cross coil setup (volume transmit coil and surface receiver). We used multi-slice single-shot echo-planar imaging (EPI, 2 mm/slice for 5 slices) with a field of view (FOV) of 20 × 20 mm2 (matrix = 48 × 48). T1-weighted images were acquired using an inversion recovery sequence, with inversion times of 250, 500, 750, 1000, 1500, 2000 and 2750 ms, relaxation delay/echo time (TE) = 6500/28 ms, and the number of averages (NA) = 4 (scan time = 3 min 37 s). The same relaxation delay, inversion time, TE and NA were used for the routine single-slice (scan time =26 min 50 s) and the proposed multi-slice T1app (scan time = 5 min 22 s) sequences, with the RF labeling duration τlab of 5 s. Note that the routine T1app sequence needs to be repeated for each slice, resulting in prolonged scan time. To determine an appropriate RF tagging pulse duration for fast T1app MRI, we also repeated T1app-weighted MRI with an RF tagging pulse of 8 s (scan time = 6 min 46 s). The amplitude modulated (AM)-ASL was performed with the following parameters: repetition time (TR)/TE = 6500/28 ms, B1 = 4.7 μT, labeling gradient strength= 1.5 Gauss/cm, labeling distance = 16 mm, modulation frequency = 250 Hz, and post-labeling duration = 300 ms, and NA = 32 (10). For AM-ASL in normal rats, we set the RF tagging pulse duration to start from 0.5 s and then varied it from 0.6 s to 4.8 s with increments of 0.2 s to evaluate the effect of labeling time on CBF quantification (scan time from 5 min 22 s to 12 min 15 s). For AM-ASL in stroke rats, AM-ASL scans were repeated with labeling time of 0.8, 1.2, 1.5, 2, 3, and 5 s (scan time from 5 min 51 s to 12 min 34s). Also, a fast single-shot isotropic diffusion-weighted MRI was acquired with two b-values of 250 and 1000 s/mm2 (TR/TE = 3250/54 ms, NA = 16, scan time = 1 min 36 s) and T2-weighted images were obtained with two separate EPI scans (TR = 3250 ms, TE=30/100 ms, NA = 16, scan time= 1 min 40 s).
Data analysis
Images were analyzed in Matlab (Mathworks, Natick, MA). T1 and T1app maps were calculated by the least-squares fitting of the signal intensities as functions of the inversion time ( and ) per pixel, in which τinv is the inversion time. There are three free parameters pseudo-I0 or Iss, β, and the parameter of interest, T1 or T1app. Such a three-parameter fitting routine accounts for inversion efficiency, insufficient signal recovery, and post label delay (Appendix). Parametric T1app maps measured from the conventional single-slice and proposed multi-slice methods were denoted as T1app and respectively. T2 and ADC were calculated as and , in which TE1,2 and b1,2 are two echo times and b-values, respectively. We calculated CBF using both Eq. (1) and Eq. (5). The ischemic lesion was defined based on a threshold-based segmentation approach of the T2 map, and the lesion was mirrored along the midline to define the contralateral normal area. We used Lin’s concordance correlation and Bland-Altman plots to evaluate T1app measurements, and the Pearson correlation to analyze CBF measurements. All values were reported as mean ± standard deviation (SD), and P values less than 0.05 were considered statistically significant.
Results
We compared T1app obtained from the routine single-slice method (T1app), repeated per slice and the proposed fast multi-slice T1app method . Fig. 2a shows T1app and maps from a representative normal rat. Quantitative analysis of T1app and across five healthy rats given in Fig. 2b and 2c supports the observation that fast closely matches T1app (Lin’s concordance correlation coefficient (CCC) = 0.92). This finding is also confirmed by the scatter plot and Bland-Altman analysis showing small bias and 98% of the data points lying within ±2SD of the mean difference. The averaged T1app, and T1 values of five slices from five healthy rats were reported in Table S1 (Supporting Information). The frequency offsets relative to the frequency of the labeling plane were matched with the slice positions. Note that T1app is shorter than T1 due to concomitant MT and direct RF saturation effects. The longitudinal relaxations (T1app, and T1) maps also display prominent in-plane and over slice variation.
Figure 2: a).

a) Multi-slice T1app and maps obtained from the conventional single-slice method (repeated for each slice) and the proposed multi-slice method, respectively. b) Lin’s concordance correlation coefficient test between T1app and , per pixel, from five slices of five healthy rats. c) The Bland-Altman analysis of T1app and , per pixel, from five slices of five healthy rats.
Fig. 3 compares T1- and T1app-normalized CBF maps from normal rats. The slice was positioned 2 mm posterior from the bregma. Fig. 3a shows CBF maps from two methods under five representative labeling duration of 1.2, 1.6, 2.0, 3.0, and 4.8 s. It is worth mentioning that in the absence of T1app, the T1-normalized CBF calculation does not correct for the effect of insufficient ASL tagging durations. Fig. 3b shows the mean CBF values across the slice as a function of the RF labeling duration (N = 5). CBF derived from the T1-normalized calculation steadily increased with the labeling time. In comparison, T1app-normalized CBF values showed little dependence on the labeling time. We also investigated the saturation time dependence of CBF measurement. If we assume that the T1-normalized CBF measurement approaches its steady-state method following an exponential correction term, we have CBF = 1.46* (1-exp(−τ/1.17)) ml/(g∙min). In comparison, T1app-normalized CBF calculation remained relatively constant and we have CBF = (0.01*τ+ 1.50) ml/(g∙min) from a linear regression (R=0.17, P=0.051). Although the P-value (0.051) is borderline significant, this dependence is weak. For a typical labeling duration of 2–3 s, this equates to a CBF difference of 0.03 ml/(g∙min), which is no more than 2% of the CBF. Note that there is a substantial difference between T1 and T1app-normalized CBF values, even at the longest labeling duration. The CBF, at the longest saturation time of 4.8 s, was 1.35 and 1.58 ml/(g∙min) for T1- and T1app-normalized calculations, respectively, with a difference of 0.23 ml/(g∙min) (i.e., 14.6%).
Figure 3: a).

a) T1- and T1app-normalized CBF maps with five representative ASL labeling times of 1.2, 1.6, 2.0, 3.0, and 4.8 s. b) The effect of ASL tagging pulse duration on CBF quantification from five healthy rats. The error bars represent the inter-subject standard deviations, which were calculated from the mean values of five healthy rats.
Fig. 4 shows multi-parametric images of a representative stroke rat 24 hours after tMCAO. The ischemic lesion exhibited increased T2 (Fig. 4a), T1 (Fig. 4b), T1app (Fig. 4c), and reduced ADC (Fig. 4l), as expected. Figs. 4d, 4e, 4g, and 4h show T1- and T1app-normalized CBF with ASL tagging RF duration of 1.5 s and 5 s, respectively. The ischemic region exhibited increased CBF when compared to the contralateral normal region, indicating post-ischemic hyperemia (31,32). Fig. 4f (CBF(T1, τ=1.5s)/CBF(T1app, τ=1.5 s)) and Fig. 4i (CBF(T1, τ=5s)/CBF(T1app, τ=5 s)) show the ratio maps of T1- to T1app-normalized CBF maps with the ASL tagging duration of 1.5 s and 5 s, respectively. Fig. 4f shows the ratio in the ischemia region is lower than that in the contralateral normal tissue, and more generally, regions with lower T1app have higher ratios, such as the corpus callosum. This observation is consistent with that T1-normalized CBF is underestimated in regions of long T1app if the steady-state is not fulfilled. In comparison, Fig. 4i shows a relatively uniform CBF ratio throughout the slice, indicating that the underestimation of CBF(T1) was mitigated with a relatively long labeling time (i.e., τ= 5 s). Fig. 4j (CBF(T1, τ=1.5s)/CBF(T1, τ=5 s)) and Fig. 4k (CBF(T1app, τ=1.5s)/CBF(T1app, τ=5 s)) show the ratio maps of CBF with labeling time of 1.5 s to labeling time of 5 s from the T1- and T1app-normalized methods. The results demonstrate that whereas T1-normalized CBF is underestimated when using a short labeling time, T1app-normalized CBF remains reasonably stable with respect to labeling time.
Figure 4:

Multi-parametric images from a representative acute stroke rat after 24 hr of a 90 min tMCAO. a) T2 map, b) T1 map, c) T1app map, d) T1-normalized CBF map with a tagging RF duration of 1.5 s, e) T1app-normalized CBF map with a tagging RF duration of 1.5 s, f) The ratio image of T1- to T1app-normalized CBF maps with a tagging RF duration of 1.5 s, g) T1-normalized CBF with a tagging RF duration of 5 s, h) T1app-normalized CBF map with a tagging RF duration of 5 s, i) The ratio map of T1- to T1app- normalized CBF maps with a tagging RF duration of 5 s, j) The ratio image of T1-normalized CBF maps, with that of tagging RF duration of 1.5 s over that of 5 s, k) The ratio image of T1app-normalized CBF maps, with that of tagging duration of 1.5 s over that of 5 s, and l) ADC map.
Fig. 5 compares the T1- and T1app-normalized CBF between the ischemic lesion and the contralateral normal area as a function of ASL tagging RF duration (N = 5). The regional CBF values determined from the T1-normalized method increased consistently with the ASL tagging duration (Fig. 5a), while T1app-normalized CBF showed little dependence (Fig. 5b). Specifically, CBF (T1, τ=5s) was 1.38±0.42 and 2.44±0.62 ml/(g∙min) for the contralateral normal region and ischemic lesion, respectively, while the corresponding CBF(T1app, τ=5s) was 1.61±0.50 and 3.14±0.74 mg/(g∙min). CBF from the ischemic lesion is greatly increased from the contralateral normal area. Fig. 5c shows a significant positive correlation between the CBF difference (ΔCBF), between the contralateral normal tissue and ischemic lesion, and labeling time using T1-normalized CBF calculation with corresponding linear regression equation of ΔCBF = 0.0956*τ + 0.641 ml/(g∙min) (R=0.844, P=0.035). Fig. 5d demonstrated that T1app-normalized ΔCBF had little labeling time dependence based on the linear regression equation of ΔCBF = −0.0247*τ + 1.579 ml/(g∙min) (R=−0.352, P=0.494).
Figure 5: a).

a) T1-normalized CBF from the contralateral normal tissue and ischemic lesion as a function of the CASL RF tagging duration, b) T1app-normalized CBF from the contralateral normal tissue and ischemic lesion as a function of the ASL RF tagging duration, c) The T1-normalized CBF difference (ΔCBF) between the contralateral normal tissue and ischemic lesion as a function of CASL RF tagging duration, and d) The T1app-normalized ΔCBF as a function of CASL RF tagging duration.
Discussion
Our study developed a fast multi-slice T1app MRI, in good agreement with that using the routine single-slice T1app mapping approach (Fig. 2). Measurements in both healthy and stroke rats demonstrated the advantage of fast T1app-normalized CBF quantification, which is more accurate and robust over T1-normalized CBF mapping. Because the ischemic hyperemia is often associated with edema and blood-brain barrier disruption, accurate CBF quantification is necessary to characterize hyperemia, potentially critical for guiding thrombectomy therapy (33).
McLaughlin et al. derived a general expression for T1app as a function of amplitude and frequency offset of the off-resonance RF irradiation based on a four-compartment exchange model, showing that T1app measurement is important for accurate CBF quantification (18). Our work herein compared T1– and T1app-normalized methods for CBF quantification. The T1–normalized method assumes complete saturation of macromolecular protons, which may not be fulfilled in practice. Under the condition of partial saturation of macromolecular protons, we have the steady-state CBF being , where krev is the reverse MT exchange rate, Im is the unsaturated semisolid macromolecular magnetization (16). It is helpful to point out that although quantitative MT analysis has been well established, the concentration and exchange rate of semisolid macromolecules are not straightforward to map. Therefore, it is not trivial to estimate T1app (34,35). Partial saturation of semisolid macromolecules may also introduce a CBF underestimation using a T1-normalized approach even when the ASL labeling time is sufficiently long. We found T1-normalized CBF is 14.6% lower than T1app-normalized CBF, at the longest RF tagging pulse duration, likely attributable to incomplete semisolid macromolecule saturation. Our study directly measured multi-slice T1app, which allows CBF quantification even in the case of partial MT saturation and finite RF tagging pulse duration.
For the T1app MRI sequence, we assumed that the longitudinal magnetization reaches the steady-state in the presence of RF irradiation before the inversion pulse. To demonstrate this, we performed fast multi-slice T1app mapping with two labeling durations of 5 and 8 s, with their shortest τprep being 2.25 and 5.25 s, respectively. A τprep of 5.25 s should be sufficiently long to achieve steady-state. T1app maps determined from these two labeling durations were in good agreement (Fig. S1, Supporting Information), suggesting that a moderate labeling time of 5 s is sufficient for T1app mapping. It is worth mentioning that several novel pulse sequence without off-resonance RF irradiation before inversion pulse have been proposed for magnetization transfer imaging and T1app determination, and hence the corresponding signal can be modeled using , in which η is the inversion completeness (36,37). Also, the agreement between routine T1app and proposed T1app suggests that the inflow effect introduced by concurrent application of gradient and RF pulses had little impact on fast T1app measurement. This is likely because the inflow effect is much less than that of the tissue. Our study here acquired serial multi-slice EPI readout, which caused a slightly different delay between the end of RF saturation and excitation RF pulse among slices. When we compared the proposed multi-slice with the conventional single-slice T1app method, which had the same short post labeling delay, T1app agreed very well (Fig. 2). This agreement shows that the PLD can be fully accounted for using the three-parameter T1 and T1app fitting (Appendix). Nevertheless, the proposed T1app MRI can be implemented with simultaneous multi-slice (SMS) readout to expedite the acquisition (38,39). Alternative fast imaging readouts, such as the variable flip angle method, could also be used to provide fast multi-slice T1app imaging (40,41).
It’s helpful to discuss the difference between human and rodent brain ASL perfusion imaging. Human ASL imaging has very different transit time and spatial scales from those of rodents. In human imaging, transit times are on the order of 1.5–2 seconds, comparable to T1 and the labeling duration in many experiments (42). As such, blood T1 becomes critical for the quantification of cerebral perfusion (4). Also, the considerable distance between the labeling and imaging planes and frequency offsets in humans makes the MT effect drop rapidly in pseudo-Continuous ASL (pCASL). In comparison, because the transit time is short in the rodent brain, there was little or no post-labeling delay introduced, and also label spends only a short time (about 200 ms) in the blood before entering the tissue (30). Additionally, the short distance between the labeling plane and imaging plane, that is, less than 1 cm, also substantially increased the MT effects. Therefore, CBF quantification strongly depends on tissue T1 and MT in rodent brain perfusion imaging using CASL (18). To summarize, the proposed T1app work improves experimental perfusion imaging.
CONCLUSION
Our study developed a fast multi-slice T1app imaging sequence for improving CBF quantification. The multi-slice T1app approach agreed very well with the single-slice method, which enables accurate CBF measurement in both healthy and acute stroke rats, promising for preclinical acute stroke imaging research.
Supplementary Material
Table S1:Comparison of T1app, and T1 from each slice from the routine single-slice (repeated for each slice), the proposed multi-slice T1app, and multi-slice T1 MRI. Values were reported as mean ± standard deviation (N = 5). The slice position refers to the distance from the bregma.
Figure S1: a) maps obtained from the proposed fast multi-slice T1app MRI sequence, with two RF labeling times of 5 and 8 s. b) The Lin’s concordance correlation coefficient test between maps with labeling time of 5 and 8 s, per pixel, from five slices of five healthy rats. c) The Bland-Altman analysis of obtained with RF labeling times of 5 and 8 s.
Acknowledgments:
This study was supported in part by a grant from NIH R01NS083654 (To Sun). The authors thank Dr. Benjamin Risk for helpful discussions about the statistical analysis.
Appendix
T1app can be numerically solved using the formula
| (A.1) |
with three free parameters to describe the steady-state (Iss), T1app, an inversion efficiency (β ≡ 1 – η), in which η is inversion completeness (η =−1 for complete inversion). This formula, although not explicitly, accounts for the PLD effect. Briefly, the magnetization recovers towards its equilibrium state during the PLD, which can be described as
| (A.2) |
This can be re-written as
| (A.3) |
The same formula also applies to T1 fitting.
REFERENCE
- 1.Golay X, Hendrikse J, Lim TC. Perfusion imaging using arterial spin labeling. Top Magn Reson Imaging 2004;15(1):10–27. [DOI] [PubMed] [Google Scholar]
- 2.Wong EC. An introduction to ASL labeling techniques. J Magn Reson Imaging 2014;40(1):1–10. [DOI] [PubMed] [Google Scholar]
- 3.Wang J, Zhang Y, Wolf RL, Roc AC, Alsop DC, Detre JA. Amplitude-modulated continuous arterial spin-labeling 3.0-T perfusion MR imaging with a single coil: feasibility study. Radiology 2005;235(1):218–228. [DOI] [PubMed] [Google Scholar]
- 4.Alsop DC, Detre JA, Golay X, Günther M, Hendrikse J, Hernandez‐Garcia L, Lu H, MacIntosh BJ, Parkes LM, Smits M. Recommended implementation of arterial spin‐labeled perfusion MRI for clinical applications: A consensus of the ISMRM perfusion study group and the European consortium for ASL in dementia. Magn Reson Med 2015;73(1):102–116. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 5.Williams DS, Detre JA, Leigh JS, Koretsky AP. Magnetic resonance imaging of perfusion using spin inversion of arterial water. Proc Natl Acad Sci USA 1992;89(1):212–216. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 6.Detre JA, Leigh JS, Williams DS, Koretsky AP. Perfusion imaging. Magn Reson Med 1992;23(1):37–45. [DOI] [PubMed] [Google Scholar]
- 7.Alsop DC, Detre JA. Multisection cerebral blood flow MR imaging with continuous arterial spin labeling. Radiology 1998;208(2):410–416. [DOI] [PubMed] [Google Scholar]
- 8.Larkin JR, Simard MA, Khrapitchev AA, Meakin JA, Okell TW, Craig M, Ray KJ, Jezzard P, Chappell MA, Sibson NR. Quantitative blood flow measurement in rat brain with multiphase arterial spin labelling magnetic resonance imaging. J Cereb Blood Flow Metab 2019;39(8):1557–1569. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 9.Thomas DL, Lythgoe MF, van der Weerd L, Ordidge RJ, Gadian DG. Regional variation of cerebral blood flow and arterial transit time in the normal and hypoperfused rat brain measured using continuous arterial spin labeling MRI. J Cereb Blood Flow Metab 2006;26(2):274–282. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Utting JF, Thomas DL, Gadian DG, Helliar RW, Lythgoe MF, Ordidge RJ. Understanding and optimizing the amplitude modulated control for multiple‐slice continuous arterial spin labeling. Magn Reson Med 2005;54(3):594–604. [DOI] [PubMed] [Google Scholar]
- 11.Wolff SD, Balaban RS. Magnetization transfer contrast (MTC) and tissue water proton relaxation in vivo. Magn Reson Med 1989;10(1):135–144. [DOI] [PubMed] [Google Scholar]
- 12.Zhang W, Williams DS, Detre JA, Koretsky AP. Measurement of brain perfusion by volume‐localized NMR spectroscopy using inversion of arterial water spins: Accounting for transit time and cross‐relaxation. Magn Reson Med 1992;25(2):362–371. [DOI] [PubMed] [Google Scholar]
- 13.Pekar J, Jezzard P, Roberts DA, Leigh JS, Frank JA, McLaughlin AC. Perfusion imaging with compensation for asymmetric magnetization transfer effects. Magn Reson Med 1996;35(1):70–79. [DOI] [PubMed] [Google Scholar]
- 14.Fan AP, Jahanian H, Holdsworth SJ, Zaharchuk G. Comparison of cerebral blood flow measurement with [15O]-water positron emission tomography and arterial spin labeling magnetic resonance imaging: A systematic review. J Cereb Blood Flow Metab 2016;36(5):842–861. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Jain V, Duda J, Avants B, Giannetta M, Xie SX, Roberts T, Detre JA, Hurt H, Wehrli FW, Wang DJJ. Longitudinal Reproducibility and Accuracy of Pseudo-Continuous Arterial Spin–labeled Perfusion MR Imaging in Typically Developing Children. Radiology 2012;263(2):527–536. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Roberts DA. Magnetic resonance imaging of perfusion in humans using spin-tagging of arterial water. PhD diss, University of Pennsylvania; 1994. [Google Scholar]
- 17.Zhang W, Silva AC, Williams DS, Koretsky AP. NMR measurement of perfusion using arterial spin labeling without saturation of macromolecular spins. Magn Reson Med 1995;33(3):370–376. [DOI] [PubMed] [Google Scholar]
- 18.McLaughlin AC, Ye FQ, Pekar JJ, Santha AK, Frank JA. Effect of magnetization transfer on the measurement of cerebral blood flow using steady‐state arterial spin tagging approaches: A theoretical investigation. Magn Reson Med 1997;37(4):501–510. [DOI] [PubMed] [Google Scholar]
- 19.Williams DS. Quantitative perfusion imaging using arterial spin labeling. Methods Mol Med 2006;124:151–173. [DOI] [PubMed] [Google Scholar]
- 20.Debacker CS, Daoust A, Köhler S, Voiron J, Warnking JM, Barbier EL. Impact of tissue T1 on perfusion measurement with arterial spin labeling. Magn Reson Med 2017;77(4):1656–1664. [DOI] [PubMed] [Google Scholar]
- 21.Eng J, Ceckler TL, Balaban RS. Quantitative 1H magnetization transfer imaging in vivo. Magn Reson Med 1991;17(2):304–314. [DOI] [PubMed] [Google Scholar]
- 22.Xu X, Lee JS, Jerschow A. Ultrafast scanning of exchangeable sites by NMR spectroscopy. Angew Chem Int Ed 2013;52(32):8281–8284. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 23.Döpfert J, Zaiss M, Witte C, Schröder L. Ultrafast CEST imaging. J Magn Reson 2014;243:47–53. [DOI] [PubMed] [Google Scholar]
- 24.Xu X, Yadav NN, Song X, McMahon MT, Jerschow A, Van Zijl PC, Xu J. Screening CEST contrast agents using ultrafast CEST imaging. J Magn Reson 2016;265:224–229. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 25.Wilson NE, D’aquilla K, Debrosse C, Hariharan H, Reddy R. Localized, gradient‐reversed ultrafast z‐spectroscopy in vivo at 7T. Magn Reson Med 2016;76(4):1039–1046. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Zhou IY, Fuss TL, Igarashi T, Jiang W, Zhou X, Cheng LL, Sun PZ. Tissue Characterization with Quantitative High-Resolution Magic Angle Spinning Chemical Exchange Saturation Transfer Z-Spectroscopy. Anal Chem 2016;88(21):10379–10383. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Liu Z, Dimitrov IE, Lenkinski RE, Hajibeigi A, Vinogradov E. UCEPR: Ultrafast localized CEST‐spectroscopy with PRESS in phantoms and in vivo. Magn Reson Med 2016;75(5):1875–1885. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Buxton RB, Frank LR, Wong EC, Siewert B, Warach S, Edelman RR. A general kinetic model for quantitative perfusion imaging with arterial spin labeling. Magn Reson Med 1998;40(3):383–396. [DOI] [PubMed] [Google Scholar]
- 29.Alsop DC, Detre J. Reduced transit-time sensitivity in noninvasive magnetic resonance imaging of human cerebral blood flow. J Cereb Blood Flow Metab 1996;16(6):1236–1249. [DOI] [PubMed] [Google Scholar]
- 30.Lu H, Leoni R, Silva AC, Stein EA, Yang Y. High-field continuous arterial spin labeling with long labeling duration: Reduced confounds from blood transit time and postlabeling delay. Magn Reson Med 2010;64(6):1557–1566. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Wegener S, Artmann J, Luft AR, Buxton RB, Weller M, Wong EC. The Time of Maximum Post-Ischemic Hyperperfusion Indicates Infarct Growth Following Transient Experimental Ischemia. PLOS ONE 2013;8(5):e65322. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 32.Shen Q, Du F, Huang S, Duong TQ. Spatiotemporal Characteristics of Postischemic Hyperperfusion with Respect to Changes in T1, T2, Diffusion, Angiography, and Blood–Brain Barrier Permeability. J Cereb Blood Flow Metab 2011;31(10):2076–2085. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Campbell BCV, Mitchell PJ, Kleinig TJ, Dewey HM, Churilov L, Yassi N, Yan B, Dowling RJ, Parsons MW, Oxley TJ, Wu TY, Brooks M, Simpson MA, Miteff F, Levi CR, Krause M, Harrington TJ, Faulder KC, Steinfort BS, Priglinger M, Ang T, Scroop R, Barber PA, McGuinness B, Wijeratne T, Phan TG, Chong W, Chandra RV, Bladin CF, Badve M, Rice H, de Villiers L, Ma H, Desmond PM, Donnan GA, Davis SM. Endovascular Therapy for Ischemic Stroke with Perfusion-Imaging Selection. N Engl J Med 2015;372(11):1009–1018. [DOI] [PubMed] [Google Scholar]
- 34.Ewing JR, Jiang Q, Boska M, Zhang ZG, Brown SL, Li GH, Divine GW, Chopp M. T1 and magnetization transfer at 7 Tesla in acute ischemic infarct in the rat. Magn Reson Med 1999;41(4):696–705. [DOI] [PubMed] [Google Scholar]
- 35.Portnoy S, Stanisz GJ. Modeling pulsed magnetization transfer. Magn Reson Med 2007;58(1):144–155. [DOI] [PubMed] [Google Scholar]
- 36.Mangia S, De Martino F, Liimatainen T, Garwood M, Michaeli S. Magnetization transfer using inversion recovery during off-resonance irradiation. Magn Reson Imaging 2011;29(10):1346–1350. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 37.Kim T, Shin W, Kim SG. Fast magnetization transfer and apparent T1 imaging using a short saturation pulse with and without inversion preparation. Magn Reson Med 2014;71(3):1264–1271. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Barth M, Breuer F, Koopmans PJ, Norris DG, Poser BA. Simultaneous multislice (SMS) imaging techniques. Magn Reson Med 2016;75(1):63–81. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Kim T, Shin W, Zhao T, Beall EB, Lowe MJ, Bae KT. Whole brain perfusion measurements using arterial spin labeling with multiband acquisition. Magn Reson Med 2013;70(6):1653–1661. [DOI] [PubMed] [Google Scholar]
- 40.Fram EK, Herfkens RJ, Johnson GA, Glover GH, Karis JP, Shimakawa A, Perkins TG, Pelc NJ. Rapid calculation of T1 using variable flip angle gradient refocused imaging. Magn Reson Med 1987;5(3):201–208. [DOI] [PubMed] [Google Scholar]
- 41.Stikov N, Boudreau M, Levesque IR, Tardif CL, Barral JK, Pike GB. On the accuracy of T1 mapping: searching for common ground. Magn Reson Med 2015;73(2):514–522. [DOI] [PubMed] [Google Scholar]
- 42.Qiu D, Straka M, Zun Z, Bammer R, Moseley ME, Zaharchuk G. CBF measurements using multidelay pseudocontinuous and velocity‐selective arterial spin labeling in patients with long arterial transit delays: Comparison with xenon CT CBF. J Magn Reson Imaging 2012;36(1):110–119. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Table S1:Comparison of T1app, and T1 from each slice from the routine single-slice (repeated for each slice), the proposed multi-slice T1app, and multi-slice T1 MRI. Values were reported as mean ± standard deviation (N = 5). The slice position refers to the distance from the bregma.
Figure S1: a) maps obtained from the proposed fast multi-slice T1app MRI sequence, with two RF labeling times of 5 and 8 s. b) The Lin’s concordance correlation coefficient test between maps with labeling time of 5 and 8 s, per pixel, from five slices of five healthy rats. c) The Bland-Altman analysis of obtained with RF labeling times of 5 and 8 s.
