Abstract
Purpose:
To develop a method for slice-wise dynamic distortion correction for EPI using rapid spatiotemporal B0 field measurements from FID navigators (FIDnavs) and to evaluate the efficacy of this new approach relative to an established data-driven technique.
Methods:
A low-resolution reference image was used to create a forward model of FIDnav signal changes to enable estimation of spatiotemporal B0 inhomogeneity variations up to second order from measured FIDnavs. Five volunteers were scanned at 3T using a 64-channel coil with FID-navigated EPI. The accuracy of voxel shift measurements and geometric distortion correction was assessed for experimentally induced magnetic field perturbations. The temporal signal-to-noise ratio (tSNR) was evaluated in EPI time-series acquired at rest and with a continuous nose touching action, before and after image realignment.
Results:
Field inhomogeneity coefficients and voxel shift maps measured using FIDnavs were in excellent agreement with multi-echo EPI measurements. FID-navigated distortion correction accurately corrected image geometry in the presence of induced magnetic field perturbations, outperforming the data-driven approach in regions with large field offsets. In fMRI scans with nose touching, FIDnav-based correction yielded tSNR gains of 30% in gray matter. Following image realignment, which accounted for global image shifts, tSNR gains of 3% were achieved.
Conclusions:
Our proposed application of FIDnavs enables slice-wise dynamic distortion correction with high temporal efficiency. We achieved improved signal stability by leveraging the encoding information from multi-channel coils. This approach can be easily adapted to other EPI-based sequences to improve temporal SNR for a variety of clinical and research applications.
Keywords: B0 inhomogeneity, dynamic distortion correction, EPI artifacts, FID navigators, field map, functional MRI
Introduction
Single-shot EPI forms the basis of most fMRI studies as it facilitates measurement of BOLD signal fluctuations related to neuronal activity with high temporal resolution. However, given its low effective bandwidth in the phase-encoding direction, EPI is highly sensitive to magnetic field (B0) inhomogeneities that occur at interfaces between materials with different magnetic susceptibilities and due to hardware imperfections. Uncompensated variations in the encoding field produce non-linear geometric distortions and regions of signal “pile-up”, while intra-voxel dephasing creates signal voids in regions with high inhomogeneity. These susceptibility-induced effects increase linearly with field strength, and lead to a loss in BOLD sensitivity (1), as well as difficulties in co-localizing functional data with anatomical information (2).
Active shimming is typically performed prior to imaging to help offset inhomogeneities induced by the subject-specific susceptibility distribution by superimposing an optimal set of low-spatial-order spherical harmonic (SH) fields on the static magnetic field. The EPI readout time can also be reduced by employing ramp sampling and/or parallel imaging acceleration to minimize artifacts, but residual distortions remain a problem (3). Static distortions may be corrected in post-processing by computing voxel displacements from a low-resolution map of field offsets, usually measured from a separate multi-gradient-echo (MGRE) acquisition (4). Alternatively, the displacement field may be estimated from two sets of images acquired with reversed phase encoding directions (“blip-up” and “blip-down”), which exhibit equal magnitude distortions in opposite directions (5). These methods fail, however, to capture dynamic changes that occur over time due to physiological mechanisms such as respiration (6,7), subject motion (8,9), and imperfections in the gradient system, including eddy current effects and gradient heating (10,11). These spatially-dependent field fluctuations result in residual non-linear geometric distortions and signal variations. As most fMRI studies attempt to detect BOLD contrast changes on the order of 1%, even small signal instabilities degrade temporal SNR and reduce sensitivity to detect true neural activity, yielding spurious results (12).
The issues associated with dynamic field perturbations have motivated the development of various strategies to measure B0 fluctuations during fMRI, usually at the cost of modifying the sequence and timing of acquisition. Dynamic spatially-resolved field maps may be estimated from the phase evolution of each EPI volume if a second echo is acquired using accelerated readouts (dynamic off-resonance correction with multi-echo acquisition [DOCMA]) (3); however, acquiring two volumes per TR places a limit on spatial resolution. This spatiotemporal constraint can be avoided by jittering the echo time (13) or alternating the spatial encoding direction (14) for consecutive time points. However, these modifications fundamentally alter the fMRI acquisition, potentially inducing additional BOLD contrast variations, and are not rapid enough to accurately characterize respiration-induced B0 changes (15). Alternatively, coil-dependent phase offsets may be measured in a separate reference scan and subtracted from EPI data, which enables generation of dynamic field maps from single-echo EPI (16,17). In general, data-driven methods are limited by signal dropout, geometric distortions, and phase unwrapping, particularly at ultra-high fields and in highly heterogeneous regions such as air-tissue interfaces and post-operative sites.
Ideally, optimal shim parameters should be calculated for each slice, to enable correction for continuous physiological effects (18). Currently, the only existing solution for real-time field monitoring is external NMR field probes, which allows independent field measurements to be sampled concurrently with the imaging process at discrete locations around the object (19). Multiple studies have demonstrated that physiological and hardware-related temporal B0 variations can be effectively modeled by a series of low-spatial-order expansions (9,20,21). Thus, fitting SH functions to field probe measurements allows correction for field fluctuations that occur due to both hardware imperfections (21) and physiological mechanisms (22). Correction can be performed either retrospectively (23), or prospectively by dynamically updating the scanner shim currents in real time (24). Van Gelderen et al demonstrated real-time shimming capability to compensate for respiration-induced B0 fluctuations (25); however, this approach requires subject-specific training to calibrate phase changes related to chest motion and cannot correct for other sources of field perturbations.
Navigator methods offer a more efficient solution for measuring these low-spatial-order B0 variations by sampling a small additional amount of k-space data, while circumventing the need for additional instrumentation. Double volumetric navigators (DvNavs), based on a pair of highly-accelerated, low-resolution 3D EPI acquisitions interleaved with the host sequence, may also be used to generate a spatially resolved field map for each volume (26); however, the time required for DvNav encoding would place a substantial burden on the fMRI acquisition. Ward et al. first proposed measuring linear B0 inhomogeneities from the phase difference between orthogonal projections with different TEs to enable real-time shimming (8). Cloverleaf navigators combine motion and shim measurement in a single trajectory and measure B0 inhomogeneities directly from shifts in the k-space domain (27). However, as spatial encoding takes time, most navigator solutions increase the minimum TE or decrease acquisition efficiency if interleaved with the host sequence. Furthermore, most of these prior navigator methods cannot resolve higher-order spatial field patterns. Versluis et al. used a combination of one-dimensional navigators and sensitivity encoding to approximate the spatially-resolved B0 field, although this requires explicit knowledge of the coil sensitivity profiles (28).
FID navigators (FIDnavs) measure the signal from each receiver coil without any spatial encoding and thus can be acquired extremely rapidly with minimal impact on the sequence (29). Measuring the phase evolution between two FID signals can compensate for global respiratory-induced frequency fluctuations and system frequency drifts in EPI time series (7,10,30); however, this is insufficient to correct for non-linear geometrical distortions that arise from more localized B0 field variations. In this work, we propose a new method for dynamic distortion correction in EPI time series that leverages the spatial encoding information provided by the multi-channel coil arrays to enable characterization of spatiotemporal B0 fluctuations from FIDnavs embedded in each slice (31,32). This approach overcomes the limitations of previous MR-based solutions by enabling fast calculation of higher-order shim parameters, without the need for spatial encoding, image reconstruction or phase unwrapping algorithms, or explicit calculation of coil sensitivity profiles. In this work we evaluate the accuracy of FIDnav-based voxel shift measurements and assess the improvement in BOLD time-series stability that can be achieved using FID-navigated slice-wise dynamic distortion correction in fMRI. Preliminary results of this study have been presented in abstract form (33).
Methods
B0 field measurements from FIDnavs
As described in Wallace et al. (2020), FIDnavs measure the signal integral over the excited slice of the underlying spin density distribution, modulated by the B0 inhomogeneities and receive coil sensitivities at each spatial location. This may be approximated by the discrete summation of a complex-valued multi-channel reference image with matched contrast properties. Dynamic B0 field changes may be expressed as a series of low-spatial-order basis functions: ΔB0,n(r) = ΔB0,0(r) + β(r)bn, where r denotes the spatial coordinate and n indexes time. A forward model can be generated by simulating the effect of changes in these inhomogeneity coefficients on the complex-valued FIDnav signals (yj,n), with spatial encoding provided by a multi-channel reference image:
[1] |
where yj,n is the measured FIDnav from coil j at acquisition n; each Sj,0 is a 1 × Np vector of the complex pixel intensities of a reference image with Np pixels, measured by coil j; γ is the gyromagnetic ratio (in rad/s); τ is the echo time of the reference image (designed to match the sampling time of the FIDnav); β is an Np × Nb matrix representing the Nb basis functions and bn is an Nb × 1 vector representing the inhomogeneity coefficients at acquisition n. In-plane SH functions up to second order are represented by five inhomogeneity coefficients . These unknown coefficients at each time point can be computed using a non-linear algorithm that minimizes the residual sum-of-squares error between the forward model and the measured FIDnavs at each excitation:
[2] |
Theoretically, the central k-space point of the single-shot EPI trajectory may yield similar information to an FIDnav. However, in practice, gradient delays, eddy currents, and B0 inhomogeneities (which may all be considered as background field gradients) deflect the EPI trajectory from its prescribed path. Thus, in the presence of system imperfections and local B0 offsets, the EPI readout no longer passes through the center of k-space, and a simple forward model cannot accurately predict the measured signal at the assumed center of k-space (Supporting Figure S1). An accurate forward model would require exact knowledge of the trajectory error, which is not available to us because many of these gradient errors vary dynamically. In contrast, the FIDnav is acquired without any gradient encoding and thus the FID signal is not affected by gradient delays or dynamically varying trajectory errors. Changes in the FIDnav signal reflect changes in the local B0 field, as well as eddy currents from preceding gradients, which may be modeled by low-spatial-order SH functions.
Reference image for EPI
The FIDnav signal model requires multi-channel GRE reference image data with matched contrast properties to be acquired for each subject (32). Acquiring two datasets with reversed gradient polarities eliminates phase errors arising from gradient delays and other system imperfections (34). One challenge of this approach is that, to achieve full brain coverage, TRs for single-shot EPI acquisitions are typically on the order of 2–3 seconds. For a matrix size of 64×64 and TR of 2 s, acquisition of reference data for model calibration would take ~4 minutes. In this work, we accelerate reference data acquisition using a segmented EPI reference image with three encoding lines per shot (Fig. 1A), and blip-up and blip-down encoding, to provide self-referencing compensation for the effects of gradient imperfections on the encoded signal. We also assess the minimal resolution requirements for the reference image data to further reduce the time required for forward model generation.
MRI Experiments
All imaging experiments were performed at 3T (MAGNETOM Skyra; Siemens, Erlangen, Germany) using the vendor-supplied 64-channel head coil. Five volunteers were scanned after providing written informed consent in accordance with the institutional IRB-approved protocol. An FIDnav module was inserted into a gradient-echo EPI sequence between each slice-selective excitation and EPI readout (Fig. 1B). A schematic of the experimental design is shown in Figure 2.
Experiment 1: Phantom validation
To validate the accuracy of FIDnav-based B0 measurements in an EPI sequence and assess the minimal requirements for forward model generation, a cylindrical water bottle was scanned with manual adjustment of the scanner shim settings. A segmented EPI reference image was acquired with TE 5 ms, TR 1000 ms, EPI factor 3 and blip-up and blip-down encoding (total acquisition time 22 s). Axial FID-navigated multi-echo EPI scans were acquired with the following parameters: TE1/TE2 16.0/33.8 ms, TR 1000 ms, 10 repetitions, GRAPPA acceleration factor 2. The FIDnav module sampled 64 points in 0.4 ms, centered on TNAV 5 ms. Both reference and FID-navigated scans were acquired with flip angle 90°, receiver bandwidth 2232 Hz/pixel, FOV 192 mm, voxel size 3 mm isotropic, 16 slices with 20% inter-slice gap, centered at the isocenter of the magnet. First and second-order shim settings were manually adjusted up to ±10 μT/m and ±100 μT/m2 in increments of 2.5 μT/m and 25 μT/m2, respectively. EPI images were reconstructed using Dual-Polarity GRAPPA (DPG), which incorporates Nyquist ghost correction into the parallel imaging reconstruction (35). Acquisition of two echoes per TR enabled calculation of spatially-resolved field maps using the DOCMA approach (3). Specifically, the phase evolution between echoes was obtained via a weighted sum over channels of the Hermitian inner product and used to compute ΔB0.
Unwrapped field maps were computed in SPM (Statistical Parametric Mapping v12, Wellcome Trust Centre for Neuroimaging, London, UK, www.fil.ion.ucl.ac.uk/spm/) and smoothed with a 5-mm FWHM kernel. SH functions were fit to the measured pixel-wise field maps in the foreground region, masked using intensity thresholding. Field inhomogeneity coefficients up to second order were estimated from each FIDnav measurement and compared to those measured using multi-echo EPI. This analysis was repeated following retrospective down-sampling of the reference image data to assess the impact of image resolution on the accuracy of model-based FIDnav shim parameter estimation. Field estimates were converted into voxel-shift maps (VSMs):
[3] |
where BWpp is the effective bandwidth per pixel along the phase-encoding direction. VSMs were applied to the reconstructed EPI data using SPM.
Experiment 2: Geometric distortion correction in volunteers
To compare the performance of FIDnav-based VSMs to the reference data-driven method, FID-navigated multi-echo EPI images were acquired in five volunteers with the following conditions: 1) resting position, standard shim settings; 2) applied field offset in the phase-encoding (Y) direction of ±10 μT/m; and 3) with subjects touching their nose. Scan parameters for the FID-navigated multi-echo EPI scans were as follows: TNAV 5 ms, TE1/TE2 18/48.1 ms, TR 2000 ms, receiver bandwidth 1860 Hz/pixel, in-plane resolution 2 mm, GRAPPA factor 2 and 6/8 partial Fourier. Posterior-anterior phase-encoding was used to avoid signal pileup in the frontal cortex (36). Segmented EPI reference images with blip-up and blip-down encoding were acquired for forward model generation with TE 5 ms, TR 2000 ms, receiver bandwidth 2232 Hz/pixel and matrix size 32×32, yielding an in-plane resolution of 7 mm (total acquisition time 44 s). Both sets of scans were acquired with flip angle 70°, FOV 224 mm, 20 slices with 2.5 mm slice thickness and 20% inter-slice gap, and fat suppression. In addition, a multi-echo GRE field map was acquired for static distortion correction (TE1/TE2 4.92/7.38 ms, TR 250 ms, flip angle 25°, receiver bandwidth 800 Hz/pixel, in-plane resolution 2 mm; matched FOV, slice thickness and spacing). An axial T2-weighted fast-spin-echo (FSE) image (TR 7360 ms, TE 78 ms, receiver bandwidth 220 Hz/pixel, in-plane resolution 0.9 mm, with matched FOV, slice thickness and spacing) was acquired to provide an anatomical reference.
Multi-channel EPI images were reconstructed using DPG and combined using sum-of-squares. B0 inhomogeneity coefficients up to second order were measured from FIDnavs in each slice. Estimated field difference maps were superposed onto the static field map and converted into VSMs to correct for geometrical distortions. VSMs estimated using the reference DOCMA method were also applied for comparison. Static, FIDnav-based and DOCMA distortion corrected images were compared to FSE images, assuming these have minimal distortion. In one volunteer, VSMs were also calculated from the first EPI echo image using the method proposed in (17). Structural images were masked using FSL’s BET tool (FMRIB, Oxford, UK, www.fmrib.ox.c.uk/fsl/). Normalized root-mean-square error (NRMSE) was calculated within the brain region for EPI images with induced geometric distortions corrected using each method, relative to a reference EPI image with static distortion correction. Differences in NRMSE were assessed using two-sided paired Wilcoxon rank sum tests with α=0.05.
Experiment 3: Temporal stability of BOLD signal in volunteers
To evaluate the improvement in BOLD signal stability with the proposed approach, fMRI scans were acquired at rest and while the volunteer performed a continuous nose-touching action. Scan parameters were identical to those used in Experiment 2, with 120 EPI volumes acquired in each series, resulting in a total acquisition time of ~4 minutes. Static distortion correction was applied using field measurements from a MGRE acquisition. Dynamic distortion correction was applied by superimposing second-order field changes estimated from FIDnavs on the static field map for each time point (Fig. 2). DOCMA-based correction was also applied for comparison. Distortion-corrected images were aligned to the first image in each series using rigid-body registration in SPM. Temporal SNR (tSNR) was computed as the ratio of the mean and standard deviation of each voxel time-course for the first echo image in each series as a measure of BOLD signal stability for static and dynamic distortion correction, with and without rigid-body realignment (37). Improvement in mean tSNR associated with dynamic distortion correction was characterized by comparing to images reconstructed using the static distortion correction scheme. The relative tSNR gain was calculated as follows:
[4] |
where 〈tsNR(r)〉 denotes the mean tSNR over all voxels within the whole-brain mask or gray matter regions (segmented in SPM). Significance of the one-sided Wilcoxon signed rank test for median tSNR gain above zero was tested.
Functional MRI data analysis was performed on the first echo image series using SPM. Functional images were slice-time corrected; no additional registration was applied as our preprocessing pipeline includes realignment. The data were smoothed spatially with a Gaussian kernel (FWHM 4 mm) to improve SNR. For the nose touching experiment, arm motion events were modeled using a general linear model. These events were modeled as boxcars representing the duration of each block, and were convolved with the canonical hemodynamic response function included with SPM. Differences in brain activation with this condition and rest condition were compared using paired t-tests within SPM. An uncorrected threshold of P<0.001 was considered statistically significant.
Results
Experiment 1: Phantom validation
A comparison between field coefficients measured using the proposed approach and the reference EPI-based method for applied shim changes is shown in Figure 3. FIDnav measurements were computed using a reference image matrix size of 32×32, as this provided a good trade-off between the efficiency of both reference data acquisition and shim parameter computation and the accuracy of field measurements (Supporting Table S1). The proposed approach accurately characterized shim changes, with mean absolute errors of 0.49 ± 0.06 μT/m (first-order) and 1.22 ± 0.37 μT/m2 (second-order) relative to the reference data-driven approach for maximum changes of 10 μT/m and 100μT/m2, respectively. Maps of normalized signal intensity before and after correction for first- and second-order shim changes are shown in Supporting Figure S2.
Experiment 2: Geometric distortion correction in volunteers
Figure 4 shows a comparison of the voxel shifts induced by modifying the shim settings along the phase-encoding direction and by the volunteer touching their nose, computed with the proposed FIDnav-based approach and the reference DOCMA method. Increasing the Y-shim gradient resulted in negative voxel shifts in the frontal brain region (stretching) with respect to the posterior-anterior phase-encoding direction, while nose touching in volunteers led to positive voxel shifts in this region. Difference maps confirm that these two measures are in good agreement. Relative to the spatially-resolved EPI-based VSMs, correction with FIDnav field estimates led to a substantial reduction in the residual voxel shift (Fig. 4C–D). The mean absolute error of voxel shifts reduced from 1.01 ± 0.67 mm to 0.06 ± 0.10 mm (Y-shim) and 0.25 ± 0.25 mm to 0.09 ± 0.17 mm (nose touching) following correction with FIDnav estimates. Figure 5 illustrates geometric distortion correction results in a representative volunteer following an adjustment in Y-shim and during nose touching. Comparison with anatomical image data shows that static distortion correction resulted in residual unwarping errors, which are most easily observed in the frontal brain regions. FIDnav field estimates accurately compensated for geometric distortions induced by both changing shim settings and the nose-touching action, while residual unwarping errors are present in the DOCMA method. Table 1 lists the mean NRMSE values calculated in each EPI image relative to the reference EPI image, which confirms that accuracy of geometric distortion correction is improved following FIDnav correction. Distortion correction results using phase maps derived from single-echo EPI are also shown for comparison in Supporting Figure S3. Both data-driven approaches produced unreliable field estimates in regions of high inhomogeneity, which exhibit large geometric distortions and signal dropout at longer echo times.
Table 1.
NRMSE (%) | Y-Shim | Nose Touching | ||||
---|---|---|---|---|---|---|
Subject No. | Static | FIDnav | DOCMA | Static | FIDnav | DOCMA |
1 | 5.2 | 2.3 | 2.4 | 4.0 | 2.5 | 2.6 |
2 | 5.0 | 1.5 | 1.6 | 1.7 | 1.3 | 1.4 |
3 | 4.8 | 1.4 | 1.5 | 2.2 | 1.4 | 1.3 |
4 | 5.3 | 2.0 | 0.5 | 1.4 | 1.6 | 1.6 |
5 | 6.7 | 2.6 | 6.3 | 3.1 | 1.5 | 1.7 |
Mean | 5.5 | 2.0* | 2.14* | 2.49 | 1.7 | 1.7 |
SD | 1.0 | 0.6 | 1.59 | 1.05 | 0.5 | 0.5 |
Experiment 3: Temporal stability of BOLD signal in volunteers
Figure 6A shows the difference in standard deviation (SD) with static and FIDnav-based dynamic distortion correction for no motion and nose touching, before and after rigid-body realignment. Prior to realignment, the image series with static correction exhibited substantially higher SD than the FIDnav-corrected images for both no motion and nose touching. Rigid-body realignment accounted for global field changes, resulting in highly similar SD maps for static and dynamic distortion correction in the no motion case. Residual differences can be observed in the frontal brain region for nose touching, as changing geometric distortions are accounted for by FIDnav field correction, but not by image realignment. Figure 6B shows a comparison of the signal intensity in this region with static and dynamic distortion correction; signal fluctuations induced by nose touching are successfully removed by FID-navigated correction. The tSNR distribution for each correction scheme is summarized in Table 2. Figure 7 shows the tSNR gains within the whole brain and segmented gray matter across all subjects. Before realignment, FIDnav correction resulted in significant median tSNR increase of 30% and 24% within the brain for no motion and nose touching, respectively. After realignment, tSNR gains for the nose-touching paradigm were 1.6% and 2.8% in whole-brain and gray matter regions, respectively, while there was no gain in tSNR for cases without intentional field perturbations. Results from DOCMA correction were more variable (Fig. 7C).
Table 2.
Without Image Realignment | With Image Realignment | |||||||
---|---|---|---|---|---|---|---|---|
No Motion | Nose Touching | No Motion | Nose Touching | |||||
Static | FIDnav | Static | FIDnav | Static | FIDnav | Static | FIDnav | |
Whole Brain | 51 ± 23 | 64 ± 26 | 45 ± 21 | 53 ± 23 | 67 ± 26 | 65 ± 26 | 56 ± 23 | 57 ± 23 |
Gray Matter | 49 ± 27 | 66 ± 30 | 42 ± 23 | 51 ± 26 | 68 ± 30 | 66 ± 30 | 55 ± 26 | 56 ± 26 |
Maps of brain activation detected during nose touching are displayed in Figure 8 for static and FIDnav-based dynamic distortion correction. In both cases, activation is detected in the motor cortex; however, with static distortion correction, regions of false positive activation are also detected near the edges of the brain due to geometric distortion patterns that correlate with the motor task. These false positive activations are successfully removed with FIDnav-based correction.
Discussion
We have proposed a new approach for slice-wise dynamic distortion correction in fMRI using ultra-short FIDnavs and spatial encoding provided by multi-channel array coils. To our knowledge, this is the first study to perform dynamic distortion correction in EPI using FIDnavs. Our results in phantom and volunteer experiments demonstrate that field changes up to second order can be accurately characterized by FIDnavs embedded in each slice of an EPI sequence using information from a low-resolution reference image. In single-shot EPI, dynamic magnetic field perturbations result in shifts and geometric distortion of the reconstructed images. Comparison with a reference anatomical scan confirms that geometric distortion correction with FIDnavs improves image fidelity in the presence of spatiotemporally-varying magnetic field changes.
In fMRI, field variations give rise to intensity fluctuations within each voxel, adversely impacting the tSNR of the BOLD signal time-course. The dominant impact of field fluctuations in this study was a shift in image space, which was corrected by both FIDnav-based dynamic distortion correction and conventional realignment in image space. However, the nose-touching paradigm also induced higher-order field changes, which could not adequately be compensated by rigid-body realignment, resulting in spurious signal fluctuations and false positive activations. These were successfully removed by the proposed FIDnav correction, with a small gain in tSNR observed across the whole brain region. These results demonstrate that our proposed method can improve BOLD sensitivity by compensating for residual geometric distortions in the reconstructed images.
Overall, the proposed FIDnav-based method was comparable to, or in some cases, outperformed the reference DOCMA approach with respect to both geometric distortion correction and improvement in BOLD signal stability. Reliable phase unwrapping and the invertibility of distortion are both crucial for EPI-based field mapping approaches. We used an accelerated acquisition to minimize ΔTE; however, we found that even with shorter TEs facilitated by acceleration, signal dropout in the second echo led to unreliable phase correction in regions of high inhomogeneity. Figure 5 illustrates that FIDnav correction outperformed DOCMA in the frontal brain cortex, where field inhomogeneity is higher. Although reliable measurement of activation in brain regions with high B0 inhomogeneity remains challenging, we believe that our method represents an improvement over existing techniques for obtaining reliable distortion correction in the presence of dynamic field perturbations. Unwrapping errors led to temporal instability in field measurements in some subjects, which may help explain why DOCMA failed to improve tSNR. While previous studies have demonstrated it is possible to estimate field changes from the phase information in single-echo EPI (16,17), our proposed method has several key advantages over data-driven dynamic distortion correction approaches, as these require reconstruction and phase unwrapping of image data to perform field correction. This limits their suitability and utility for real-time applications or for correcting phase inconsistencies between segments in multi-shot EPI.
Our proposed FIDnav-based solution also has several advantages over other shim navigator approaches. One-dimensional navigator signals have previously been used to measure linear field changes from the phase difference between projections with different TEs in image space (8) or shifts in k-space (27). These methods assume that inhomogeneities along each axis can be measured independently; however, dephasing effects mean the phase evolution of the projection data is not necessarily linearly related to B0 inhomogeneities in that direction (38). Thus, projection data alone is insufficient to resolve field changes in multiple directions without a priori information provided by a reference scan (39). Also, the time required for spatial encoding may substantially increase the minimal TE or limit the frequency of shim updates. Subsequent studies demonstrated that encoding information provided by multi-channel coil arrays allows low-spatial-order B0 information to be recovered from an unencoded FIDnav signal. The first realization of this approach considered the temporal evolution of the FID signal to estimate ΔB0 at each time point (31); however, R2* relaxation and ΔB0-induced dephasing contribute to non-linear phase evolution, limiting application of this technique in vivo. Changes in ΔB0 can be more reliably estimated relative to a reference time point using an FIDnav sampled at the echo time of the reference image, as this approach makes no assumption that the FID signal evolves linearly as a function of ΔB0 (32). An advantage of this method is that no explicit calculation of coil sensitivity profiles is required as this information is inherently encoded within the reference image. As the minimal TE/TR is extended only by the duration of the FIDnav (0.4 ms) with encoding provided by the reference image, our method enables slice-wise dynamic distortion correction without compromising spatiotemporal resolution.
To enable spatially-resolved field measurements from FIDnavs, a multi-channel reference image is required to create a forward model of the FIDnav signals from each receiver coil. We minimized the time needed to acquire this reference data by using a low-resolution (7 mm in-plane) multi-shot EPI sequence, rather than a GRE sequence used in previous work (32). As most fMRI studies already acquire a static field map, model calibration could be integrated in a single step using a multi-echo reference scan. The effect of respiration-induced field perturbations on the reference image is assumed to be minimal at 3T. Future work could investigate reordering reference data segments to minimize sensitivity to phase errors using the fast low-angle excitation echo-planar technique (FLEET) (40). In this work, we estimated field changes up to second order, which enabled validation in a controlled phantom study with modification of the scanner shim settings. With a 64-channel head coil, it is possible to solve for B0 inhomogeneity coefficients up to third order; however, we previously did not find any significant differences between fitting second- and third-order inhomogeneity coefficients at 3T (32), and increasing the number of coefficients may make the system more sensitive to noise. The value of third-order correction may be more evident at higher field strengths, which will be the subject of future work. Like most navigator methods, our proposed approach assumes that the measured field does not change within the EPI readout and thus cannot correct for higher-frequency field changes. Currently field probes are the only method that can simultaneously monitor field changes during the EPI readout itself to compensate for trajectory deviations due to eddy currents and other hardware-related instabilities (21).
We used accelerated, multi-echo EPI to provide a reference VSM for each volume using the DOCMA approach. Acceleration limited the amount of geometric distortion by reducing the total readout time. Adjusting the scanner shim settings and nose touching in volunteers induced small, but measurable distortions in the images. All images were unwarped in SPM using either static or dynamic field estimates and aligned using rigid-body registration. As well as correcting for inter-volume head motion, global frequency changes are also compensated by image realignment. Thus, differences in tSNR following realignment represent residual unwarping errors, slice-wise shifts, realignment errors and BOLD contrast changes.
Motion is another major source of magnetic field perturbations due to the changing orientation of the susceptibility distribution as the head moves within the static magnetic field (9). We minimized head motion using foam padding, and it was assumed that the relative position of the underlying spin density distribution and the coil sensitivities remained fixed. Measurement of rigid-body head pose information using FIDnavs in 2D sequences is non-trivial as it is difficult to model the effect of out-of-plane motion (41). Our proposed approach could be combined with optical tracking (42) or other motion measurement approaches (43,44) to measure motion-induced field changes.
This work presents a proof of principle of the ability of FID navigators to correct for geometric distortions in single-shot EPI at 3T, although this method is readily extendable to other EPI-based sequences and field strengths. The tSNR gains realized with FID-navigated dynamic distortion correction are expected to be even higher at 7T as susceptibility effects scale with field strength. FIDnav-based field measurements are highly applicable in both 2D multi-shot and 3D EPI as the proposed approach does not require formation of an image and minimally impacts the sequence timing. This would enable correction for physiological and hardware-related phase variations between shots, which can corrupt individual image volumes and reduce the signal stability of the BOLD time course (45). Extension to simultaneous multi-slice (SMS) acquisitions is also feasible. In SMS acquisitions, the measured FIDnav represents the sum of the excited spins in each slice and captures shim changes across all simultaneously excited slices. A 3D spherical harmonics model may be used to compute shim changes across the volume from FIDnav measurements following each multi-band excitation, using a forward model generated from multi-slice reference data (Supporting Figure S4).
In this work, higher-order field perturbations were corrected retrospectively by updating the VSM used to unwarp each EPI image. Shim changes measured from FIDnavs could also be used to update the transmit frequency and first-order shim fields in real time. Prospective correction of higher-order field changes is becoming feasible using custom-built shim array coils (18). Real-time shimming would enable correction for the effects of magnetic field perturbations on RF excitation, fat suppression and intra-voxel dephasing, in addition to changing geometric distortions. FID-navigated dynamic distortion correction is expected to translate to improved BOLD sensitivity in fMRI studies, which could help improve localization for presurgical planning in epilepsy patients (46), the reliability of fMRI in patients with Parkinson’s and Alzheimer’s disease, and in research studies utilizing high-resolution fMRI to measure neuronal activation with sub-millimeter spatial resolution (47). This technique could also be translated to other EPI applications, such as perfusion and diffusion-weighted imaging and non-BOLD fMRI.
Conclusion
We have evaluated a new approach for slice-wise dynamic distortion correction using FIDnavs embedded in each slice of a single-shot EPI sequence. FIDnav-based dynamic distortion correction improved geometric accuracy of images acquired with magnetic field perturbations and improved tSNR of the EPI time series by accounting for global and higher-order field fluctuations. Our proposed FIDnav-based approach has several key advantages over data-driven methods: it does not fundamentally alter the acquisition and can be implemented without compromising spatiotemporal resolution. Furthermore it does not require computationally expensive image reconstruction and phase unwrapping, making it more amenable to real-time shimming applications. FID-navigated dynamic distortion correction is widely applicable and is expected to improve BOLD sensitivity in fMRI experiments at field strengths of 3T and above.
Supplementary Material
Acknowledgements
This research was supported in part by NIH grants R01 EB019483, R01 NS079788, R01 NS106030, R01 EB019437, R01 MH111419, R00 EB021349 and by an Early Career Award from the Thrasher Research Fund (#14989). The authors extend thanks to Dr. Benedikt Poser, Maastricht University, the Netherlands, for guidance on implementing the multi gradient echo EPI sequence used in this study.
References
- 1.Deichmann R, Josephs O, Hutton C, Corfield DR, Turner R. Compensation of susceptibility-induced BOLD sensitivity losses in echo-planar fMRI imaging. Neuroimage 2002;135:120–135. [DOI] [PubMed] [Google Scholar]
- 2.Cusack R, Brett M, Osswald K. An evaluation of the use of magnetic field maps to undistort echo-planar images. Neuroimage 2003;18:127–142. [DOI] [PubMed] [Google Scholar]
- 3.Visser E, Poser BA, Barth M, Zwiers MP. Reference-free unwarping of EPI data using dynamic off-resonance correction with multiecho acquisition (DOCMA). Magn. Reson. Med 2012;68:1247–1254. [DOI] [PubMed] [Google Scholar]
- 4.Jezzard P, Balaban RS. Correction for geometric distortion in echo planar images from B0 field variations. Magn. Reson. Med 1995;34:65–73. [DOI] [PubMed] [Google Scholar]
- 5.Andersson JLR, Skare S, Ashburner J. How to correct susceptibility distortions in spin-echo echo-planar images: application to diffusion tensor imaging. Neuroimage 2003;20:870–888. [DOI] [PubMed] [Google Scholar]
- 6.Van de Moortele P-F, Pfeuffer J, Glover GH, Ugurbil K, Hu X. Respiration-induced B0 fluctuations and their spatial distribution in the human brain at 7 tesla. Magn. Reson. Med 2002;47:888–895. [DOI] [PubMed] [Google Scholar]
- 7.Pfeuffer J, Van de Moortele PF, Ugurbil K, Hu X, Glover GH. Correction of physiologically induced global off-resonance effects in dynamic echo-planar and spiral functional imaging. Magn. Reson. Med 2002;47:344–353. [DOI] [PubMed] [Google Scholar]
- 8.Ward HA, Riederer SJ, Jack CR. Real-time autoshimming for echo planar timecourse imaging. Magn. Reson. Med 2002;48:771–780. [DOI] [PubMed] [Google Scholar]
- 9.Liu J, de Zwart JA, van Gelderen P, Murphy-Boesch J, Duyn JH. Effect of head motion on MRI B0 field distribution. Magn. Reson. Med 2018;80:2538–2548. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 10.Foerster BU, Tomasi D, Caparelli EC. Magnetic field shift due to mechanical vibration in functional magnetic resonance imaging. Magn. Reson. Med 2005;54:1261–1267. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 11.Benner T, van der Kouwe AJW, Kirsch JE, Sorensen AG. Real-time RF pulse adjustment for B0 drift correction. Magn. Reson. Med 2006;56:204–209. [DOI] [PubMed] [Google Scholar]
- 12.Murphy K, Birn RM, Bandettini PA. Resting-state FMRI confounds and cleanup. Neuroimage 2013;80:349–359. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Dymerska B, Poser BA, Bogner W, Visser E, Eckstein K, Cardoso P, Barth M, Trattnig S, Robinson SD. Correcting dynamic distortions in 7T echo planar imaging using a jittered echo time sequence. Magn. Reson. Med 2016;76:1388–1399. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Xiang Q-S, Ye FQ. Correction for geometric distortion and N/2 ghosting in EPI by phase labeling for additional coordinate encoding (PLACE). Magn. Reson. Med 2007;57:731–741. [DOI] [PubMed] [Google Scholar]
- 15.Zeller M, Kraus P, Muller A, Bley T, Kostler H. Respiration impacts phase difference-based field maps in echo planar imaging. Magn. Reson. Med 2014;72:446–451. [DOI] [PubMed] [Google Scholar]
- 16.Lamberton F, Delcroix N, Grenier D, Mazoyer B, Joliot M. A new EPI-based dynamic field mapping method: application to retrospective geometrical distortion corrections. J. Magn. Reson. Imaging 2007;26:747–755. [DOI] [PubMed] [Google Scholar]
- 17.Dymerska B, Poser BA, Barth M, Trattnig S, Robinson SD. A method for the dynamic correction of B0-related distortions in single-echo EPI at 7 T. Neuroimage 2018;168:321–331. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Stockmann JP, Wald LL. In vivo B0 field shimming methods for MRI at 7 T. Neuroimage 2018;168:71–87. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Andersen M, Hanson LG, Madsen KH, Wezel J, Boer V, Van Der Velden T, Van Osch MJP, Klomp D, Webb AG, Versluis MJ. Measuring motion-induced B0-fluctuations in the brain using field probes. Magn. Reson. Med 2016;75:2020–2030. [DOI] [PubMed] [Google Scholar]
- 20.Vannesjo SJ, Miller KL, Clare S, Tracey I. Spatiotemporal characterization of breathing-induced B0 field fluctuations in the cervical spinal cord at 7T. Neuroimage 2018;167:191–202. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 21.Kasper L, Bollmann S, Vannesjo SJ, Gross S, Haeberlin M, Dietrich BE, Pruessmann KP. Monitoring, analysis, and correction of magnetic field fluctuations in echo planar imaging time series. Magn. Reson. Med 2015;74:396–409. [DOI] [PubMed] [Google Scholar]
- 22.Bollmann S, Kasper L, Vannesjo SJ, Diaconescu AO, Dietrich BE, Gross S, Stephan KE, Pruessmann KP. Analysis and correction of field fluctuations in fMRI data using field monitoring. Neuroimage 2017;154:92–105. [DOI] [PubMed] [Google Scholar]
- 23.Wilm BJ, Barmet C, Pavan M, Pruessmann KP. Higher order reconstruction for MRI in the presence of spatiotemporal field perturbations. Magn. Reson. Med 2011;65:1690–1701. [DOI] [PubMed] [Google Scholar]
- 24.Duerst Y, Wilm BJ, Dietrich BE, Vannesjo SJ, Barmet C, Schmid T, Brunner DO, Pruessmann KP. Real-time feedback for spatiotemporal field stabilization in MR systems. Magn. Reson. Med 2015;73:884–893. doi: 10.1002/mrm.25167. [DOI] [PubMed] [Google Scholar]
- 25.Van Gelderen P, De Zwart JA, Starewicz P, Hinks RS, Duyn JH. Real-time shimming to compensate for respiration-induced B0 fluctuations. Magn. Reson. Med 2007;57:362–368. [DOI] [PubMed] [Google Scholar]
- 26.Alhamud A, Taylor PA, van der Kouwe AJW, Meintjes EM. Real-time measurement and correction of both B0 changes and subject motion in diffusion tensor imaging using a double volumetric navigated (DvNav) sequence. Neuroimage 2016;126:60–71. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.van der Kouwe AJW, Benner T, Dale AM. Real-time rigid body motion correction and shimming using cloverleaf navigators. Magn. Reson. Med 2006;56:1019–1032. [DOI] [PubMed] [Google Scholar]
- 28.Versluis MJ, Sutton BP, De Bruin PW, Bornert P, Webb A, Van Osch MJ. Retrospective image correction in the presence of nonlinear temporal magnetic field changes using multichannel navigator echoes. Magn. Reson. Med 2012;68:1836–1845. [DOI] [PubMed] [Google Scholar]
- 29.Kober T, Marques JP, Gruetter R, Krueger G. Head motion detection using FID navigators. Magn. Reson. Med 2011;66:135–43. [DOI] [PubMed] [Google Scholar]
- 30.Hu X, Kim S-G. Reduction of signal fluctuation in functional MRI using navigator echoes. Magn. Reson. Med 1994;31:495–503. [DOI] [PubMed] [Google Scholar]
- 31.Splitthoff DN, Zaitsev M. SENSE shimming (SSH): A fast approach for determining B0 field inhomogeneities using sensitivity coding. Magn. Reson. Med 2009;62:1319–1325. [DOI] [PubMed] [Google Scholar]
- 32.Wallace TE, Afacan O, Kober T, Warfield SK. Rapid measurement and correction of spatiotemporal B0 field changes using FID navigators and a multi-channel reference image. Magn. Reson. Med 2020;83:575–589. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Wallace TE, Polimeni JR, Stockmann JP, Hoge WS, Kober T, Warfield SK, Afacan O. Measurement and correction of spatiotemporal B0 fluctuations using an FID-navigated EPI sequence. In: Proceedings 28th Meeting of the ISMRM, Paris, France ; 2020. [Google Scholar]
- 34.Reeder SB, Atalar E, Faranesh AZ, Mcveigh ER. Referenceless interleaved echo-planar imaging. Magn. Reson. Med 1999;94:87–94. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Hoge WS, Polimeni JR. Dual-Polarity GRAPPA for simultaneous reconstruction and ghost correction of echo planar imaging data. Magn. Reson. Med 2016;44:32–44. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.De Panfilis C, Schwarzbauer C. Positive or negative blips? The effect of phase encoding scheme on susceptibility-induced signal losses in EPI. Neuroimage 2005;25:112–121. [DOI] [PubMed] [Google Scholar]
- 37.Triantafyllou C, Polimeni JR, Wald LL. Physiological noise and signal-to-noise ratio in fMRI with multi-channel array coils. Neuroimage 2011;55:597–606. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 38.Splitthoff DN, Zaitsev M. On dephasing effects in complex projection data: implications for rapid B0 estimation. In: International Society for Magnetic Resonance in Medicine Vol. 17 ; 2009. p. 2792. [Google Scholar]
- 39.Dragonu I, Splitthoff DN, Baxan N, Freitag P, Hennig J, Zaitsev M. Shim navigators for accurate detection of the B0 magnetic field inhomogeneities using reference MGE images. In: Proceedings of the 19th Scientific Meeting International Society for Magnetic Resonance in Medicine ; 2011. p. 2690. [Google Scholar]
- 40.Polimeni JR, Bhat H, Witzel T, Benner T, Feiweier T, Inati SJ, Renvall V, Heberlein K, Wald LL. Reducing sensitivity losses due to respiration and motion in accelerated echo planar imaging by reordering the autocalibration data acquisition. Magn. Reson. Med 2016;75:665–679. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Wallace TE, Afacan O, Waszak M, Kober T, Warfield SK. Head motion measurement and correction using FID navigators. Magn. Reson. Med 2019;81:258–274. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 42.Todd N, Josephs O, Callaghan MF, Lutti A, Weiskopf N. Prospective motion correction of 3D echo-planar imaging data for functional MRI using optical tracking. Neuroimage 2015;113:1–12. doi: 10.1016/j.neuroimage.2015.03.013. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 43.Ooi MB, Muraskin J, Zou X, Thomas WJ, Krueger S, Aksoy M, Bammer R, Brown TR. Combined prospective and retrospective correction to reduce motion-Induced image misalignment and geometric distortions in EPI. Magn. Reson. Med 2013;69:803–811. doi: 10.1002/mrm.24285. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 44.Afacan O, Wallace TE, Warfield SK. Retrospective correction of head motion using measurements from an electromagnetic tracker. Magn. Reson. Med 2020;83:427–437. doi: 10.1002/mrm27934. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.van der Zwaag W, Marques JP, Kober T, Glover G, Gruetter R, Krueger G. Temporal SNR characteristics in segmented 3D-EPI at 7T. Magn. Reson. Med 2012;67:344–352. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Cardoso PL, Dymerska B, Bachratá B, Fischmeister FPS, Mahr N, Matt E, Trattnig S, Beisteiner R, Daniel S. The clinical relevance of distortion correction in presurgical fMRI at 7 T. Neuroimage 2018;168:490–498. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 47.Bause J, Polimeni JR, Stelzer J, In M-H, Ehses P, Kraemer-Fernandez P, Aghaeifar A, Lacosse E, Pohmann R, Scheffler K. Impact of prospective motion correction, distortion correction methods and large vein bias on the spatial accuracy of cortical laminar fMRI at 9.4 Tesla. Neuroimage 2020;208. doi: 10.1016/j.neuroimage.2019.116434. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.