Abstract
Quantitative estimation of contrast agent concentration is made possible by spectral CT and material decomposition. There are several approaches to modulate the sensitivity of the imaging system to obtain the different spectral channels required for decomposition. Spectral CT technologies that enable this varied sensitivity include source kV-switching, dual-layer detectors, and source-side filtering (e.g., tiled spatial-spectral filters). In this work, we use an advanced physical model to simulate these three spectral CT strategies as well as hybrid acquisitions using all combinations of two or three strategies. We apply model-based material decomposition to a water-iodine phantom with iodine concentrations from 0.1 to 5.0 mg/mL. We present bias-noise plots for the different combinations of spectral techniques and show that combined approaches permit diversity in spectral sensitivity and improve low concentration imaging performance relative to the those strategies applied individually. Better ability to estimate low concentrations of contrast agent has the potential to reduce risks associated with contrast administration (by lowering dosage) or to extend imaging applications into targets with much lower uptake.
1. INTRODUCTION
Contrast-enhanced computed tomography is an important clinical tool for imaging the three-dimensional structure and function of tissues in the human body. Applications include angiography, perfusion studies, and imaging of contrast-enhanced tissues. In many cases, small changes in contrast agent (e.g. iodine) have important clinical implications. Hepatocellular carcinoma, for example, is known to uptake slightly higher concentrations during the arterial phase relative to healthy liver tissue and it is also known to return to a low-concentration state slightly faster than healthy tissue during the venous phase.1 Sensitivity to low concentrations of iodinated contrast agents in the presence of noise will allow the next generation of CT systems to convey more functional information, e.g. regions of lower uptake, or to conduct similar studies with less contrast injection.
Spectral CT is finding increasing application in contrast-enhanced CT due to its ability to discriminate between contrast and human tissues and its ability to provide quantitative estimates of contrast concentration. Spectral CT uses projection measurements with varied spectral sensitivities to enable material-decomposed image volumes (material density maps). For example, since iodine and soft-tissue have differently-shaped mass attenuation spectra as a function of energy, spectral CT systems can take advantage of this additional information to quantify concentration.
A recent article on the future applications of spectral CT suggests that this technology also shows promise for molecular imaging,2 where drugs or other compounds with various biological function are labeled with a high-atomic-number materials that can be tracked with spectral CT. The concentrations for this type of study could be very low, below the contrast that is detectable with conventional CT in the presence of noise. This is another motivation for designing spectral CT systems for high-sensitivity contrast-enhanced imaging. High contrast-to-noise ratios will enable new applications for functional imaging with spectral CT.
There are various strategies that enable Spectral CT. For example, photon counting detectors are capable of energy discrimination and can effectively break projection data into different spectral channels (i.e., energy bins). Source-switching schemes are also used, wherein the voltage across the x-ray tube alternates between two or more kVp settings.3 In this fashion kV-switching (KVS) provides angular projections with a diversity of incident spectra. Split-filters, which divide the source beam into a low-energy and high-energy half-cone beams, yield two spectral channels for dual-energy CT.4 More recently, a generalization of the split filter idea called a spatial-spectral filter (SSF) has recently been proposed5-7. The SSF consists of a tiled array of k-edge filters which divide the beam into a number of filtered beamlets to provide a diversity of energy channels. See Figure 1 for illustrations of these spectral CT methods.
Figure 1:
Illustrations for kV Switching (left) Dual-Layer Detectors (center) and Spatial-Spectral Filters (right).
We hypothesize that combinations of spectral data acquisition approaches have the potential to increase spectral CT sensitivity to low concentrations of contrast agent. That is, with a greater diversity of spectral channels, one can improve sensitivity over the use of any single individual approach. Many approaches may be combined (e.g. KVS and SSF) into a hybrid acquisitions for potential improvements in performance. Historically, these combinations may have been avoided since it has the potential to greatly complicate data processing. For example, traditional image- or projection-domain processing can be challenged by beam-hardening effects or projection sampling requirements - e.g., combining KVS and SSF makes channelized data with unique spectral sensitivity increasingly sparse and difficult to reconstruct individually. However, recent advances in direct model-based material decomposition (MBMD)8 enable such hybrid acquisitions since all spectral effects for each measurement can be modeled and there is no intermediate processing - material density maps are formed directly from the varied spectral projection data using model-based iterative approaches.
In this work we present a study where we employ a detailed physical model of various spectral CT strategies appropriate for combination on a CT system with an energy-integrating detector including kV-switching, dual-layer detectors (DLD),9 and spatial-spectral filters. We apply MBMD to simulated data using a digital water-iodine phantom, and evaluate performance for individual spectral approaches as well as two- and three-strategy hybrid acquisitions to analyze the impact of spectral diversity on material decomposition performance for contrast-enhanced spectral CT.
2. METHODS
2.1. General Spectral CT Forward Model and Model-Based Material Decomposition
A general forward model for spectral CT with varied spectral measurements and energy-integrating detectors is
| (1) |
| (2) |
where is the mean value of the ith measurement, gi is the corresponding gain, Si(E) is the measurement-dependent system spectral sensitivity (including detector energy sensitivity, filtration, x-ray source spectrum, etc.), aij are projection contributions of the jth voxel to the ith measurement, and μj(E) is the total mass attenuation spectrum of the jth voxel. The latter spectrum is modeled as a weighted sum over k basis materials with material mass attenuation basis functions qk(E) weighted by material densities, ρj,k for each voxel. In practice, we discretize energy spectra as follows
| (3) |
or, compactly, as
| (4) |
This model is extremely general. For example, for spatial-spectral filters we can use a high-resolution voxelized model of the filter tile array and perform a forward projection to establish Si(E) as shown in Figure 2. There is also no restriction to have a small number of discrete spectral channels which allows us to model many combinations of spectral effects as well as calibrate physical non-idealities (e.g. spectral mixing at filter boundaries).
Figure 2:

DLD and SSF hybrid spectral sensitivities.
To conduct MBMD, we seek to minimize the following penalized weighted least-squared objective function:
| (5) |
where wi are the statistical weights, for which we substitute yi (approximate Poisson noise model), and R{·} is a regularization term. In this work we adopt a quadratic penalty with both in-basis and cross-basis penalties10,11 to minimize cross-talk between material basis. Moreover, we seek matched spatial resolution between basis materials, and we control the overall penalty weight via β0. We derive the gradient, ∇Φ, as well as the Hessian, H, of Φ with respect to ρ and we apply the following Newton’s Method iterative update scheme:
| (6) |
where α(n) is a step size chosen in such a way as to satisfy Wolfe’s strong conditions on convergence. The cross-voxel second derivatives contained in H are approximated to be zero. That is,
| (7) |
However, the cross material second derivatives are included to accommodate, in the search direction, the strong anticorrelation between basis materials.
2.2. Simulation Study on Hybrid Acquisitions
To characterize the relative performance of KVS, DLD, and SSF technologies as well as every combination of two or three (e.g. KVS-DLD, or KVS-DLD-SSF), we conducted spectral CT simulations and material decomposition experiments with a 1.0 mm voxelized water/iodine phantom containing a 100.0 mm diameter water tank and six 10.0 mm diameter iodine contrast inserts of concentration 0.10, 0.25, .50, 1.00, 2.50, and 5.00 mg/mL. We modeled a CT system with a source-to-detector distance of 1200 mm, a source-to-axis distance of 600 mm, and 360 views per rotation. For the standard detector, we modeled a flat panel with a 600 μm CsI scintillator and a 1.112 mm pixel pitch. The source spectra were modeled using the TASMICS dataset via the software package SPEKTR 3.0.12
For the three spectral CT strategies, we chose a KVS scheme which involved switching between 70, 100, and 130 kVp settings in that order view-by-view such that each spectral channel is made up of 120 evenly spaced full views. For the DLD system, we replaced the default flat-panel detector model with a Varex prototype dual-layer flat-panel detector which consists of an initial low-energy detector layer with a 175 μm CsI scintillator, followed by a 1.0 mm copper filter for spectral shaping, and a second layer with a 550 μm CsI scintillator for the high-energy channel. For the SSF system we model a source-to-filter distance of 100.0 mm and 3.0 mm filter tiles with a linear translation speed of 0.4 mm per view. The filter tile materials are lead, gold, lutetium, and erbium (chosen for their k-edges). All filters were 0.25 mm thick with the exception of the gold filter tile which was 0.15 mm thick.
For fair comparisons between each individual and combined spectral CT system, we modified tube current so as to normalize the deposited dose, which we computed using a uniform 160.0 mm water cylinder model of the patient. The maximum detector efficiency in the study occurs with the SSF system acting alone and is adjusted to a level of 106 total photon counts per pixel per view. We used a Poisson additive model to simulate noise.
For each acquisition method, we ran 300 iterations of the Newton’s Method material decomposition algorithm. A range of penalty weights, β0, was applied to investigate the bias-noise trade-off for each strategy. That is, for small penalty weights one should see high variance and low bias, and for large penalty weights, low variance and high bias (due to blurring of edges). Noise standard deviation was computed in an 8.0 mm diameter ROI internal to the 10.0 mm diameter insert containing a uniform 5.00 mg/mL of iodine as shown in Figure 3. To compute bias, we run an additional noiseless material decomposition ans we compute RMSE over a 14.0 mm diameter ROI centered on the 5.00 mg/mL insert. This ROI metric is intended to capture the bias due to the blurring of edges and any other errors excluding noise.
Figure 3:

Ground truth numerical phantom. Water (left) Iodine (right)
3. RESULTS
We present iodine decomposition results in Figure 4. The left-hand image shows that concentrations between 0.10 and 5.00 mg/mL of iodine can be reconstructed with spectral CT. However, we are specifically interested in the low-concentration inserts. We show an image comparison between the hybrid method, KVS-DLD-SSF (center), and the individual method, SSF (right). For this image comparison we matched bias metrics and compared contrast-to-noise ratio as a performance metric for the inserts. The contrast-to-noise ratio for the 0.25 mg/mL insert is 4.17 for KVS-DLD-SSF and 1.03 for SSF, and the contrast-to-noise ratio for the 0.10 mg/mL insert is 1.67 for KVS-DLD-SSF and 0.38 for SSF.
Figure 4:

Iodine density maps estimated via MBMD. KVS-DLD-SSF (left) SSF alone (right)
Image quality results across all methods are summarized in the bias-noise plot in Figure 5. As expected, greater regularization strength leads to lower noise and higher bias for all methods. We note that individual methods were not optimized - e.g., a different KVS scheme may be more optimal for water-iodine decomposition. The results illustrate that combinations of two strategies have superior performance as compared to the methods acting individually, and that the three-method combination has best overall performance in the study. This suggests that increased spectral diversity has a positive impact on material decomposition performance.
Figure 5:

Noise-bias curves for spectral CT methods
4. CONCLUSION
The search space for optimal hardware designs and configurations is large and we do not make a claim that the settings for individual methods used here are optimal for water-iodine decomposition. Furthermore, what is optimal for individual methods may not be optimal for the hybrid method which combines them. However, we have illustrated that hybrid acquisition methods that achieve greater spectral diversity have a beneficial impact on material decomposition performance. Specifically, for matched bias, we observed higher sensitivity when combining spectral CT methods than any constituent method acting individually. Thus, hybrid acquisition strategies for spectral CT shows promise for the clinically important task of high-sensitivity imaging in contrast-enhanced CT.
This improvement in contrast-to-noise ratio represents a small step toward the goal of functional imaging with contrast-enhanced spectral CT. Low concentrations that were previously impossible to recover with conventional CT. Future studies will include optimization of individual spectral protocols to further improve low concentration contrast estimates.
ACKNOWLEDGMENTS
This work was supported, in part, by NIH grant R21EB026849.
REFERENCES
- [1].Choi J-Y, Lee J-M, and Sirlin CB, “Ct and mr imaging diagnosis and staging of hepatocellular carcinoma: part ii. extracellular agents, hepatobiliary agents, and ancillary imaging features,” Radiology 273(1), 30–50 (2014). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [2].Taguchi K and Iwanczyk JS, “Vision 20/20: Single photon counting x-ray detectors in medical imaging,” Medical physics 40(10) (2013). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [3].Zou Y and Silver MD, “Analysis of fast kv-switching in dual energy ct using a pre-reconstruction decomposition technique,” in [Medical Imaging 2008: Physics of Medical Imaging], 6913, 691313, SPIE; (2008). [Google Scholar]
- [4].Primak A, Ramirez Giraldo J, Liu X, Yu L, and McCollough CH, “Improved dual-energy material discrimination for dual-source ct by means of additional spectral filtration,” Medical physics 36(4), 1359–1369 (2009). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [5].Tivnan M, Tilley S II, and Stayman JW, “Physical modeling and performance of spatial-spectral filters for ct material decomposition,” in [Medical Imaging 2019: Physics of Medical Imaging], 10948, 109481A, SPIE; (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [6].Tivnan M, Wang W, Tilley II S, Siewerdsen JH, and Stayman JW, “Optimized spatial-spectral ct for multi-material decomposition,” in [Fully 3D], 11072, 1107211, International Society for Optics and Photonics; (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [7].Tivnan M, Wang W, Tilley S, and Stayman J, “Best in physics (imaging): Designing spatial-spectral filters for spectral ct,” in [MEDICAL PHYSICS], 46(6), E407–E407, WILEY 111 RIVER ST, HOBOKEN 07030-5774, NJ USA: (2019). [Google Scholar]
- [8].Tilley S and Stayman J, “Model-based material decomposition for acquisitions with relaxed spectral and spatial sampling,” in [Medical Physics], 45(6), E387–E387 (2018). [Google Scholar]
- [9].Lu M, Wang A, Shapiro E, Shiroma A, Zhang J, Steiger J, and Star-Lack J, “Dual energy imaging with a dual-layer flat panel detector,” in [Medical Imaging 2019: Physics of Medical Imaging], 10948, 1094815, SPIE; (2019). [Google Scholar]
- [10].Wang W, Tilley S II, Tivnan M, and Stayman JW, “Local response prediction in model-based ct material decomposition,” in [Fully3D], 11072, 110720Z, SPIE; (2019). [DOI] [PMC free article] [PubMed] [Google Scholar]
- [11].Wang W, Tivnan M, Gang G, Tilley S, and Stayman J, “Generalized local impulse response prediction in model-based material decomposition of spectral ct,” in [MEDICAL PHYSICS], 46(6), E257–E257, WILEY 111 RIVER ST, HOBOKEN 07030-5774, NJ USA: (2019). [Google Scholar]
- [12].Punnoose J, Xu J, Sisniega A, Zbijewski W, and Siewerdsen J, “spektr 3.0—a computational tool for x-ray spectrum modeling and analysis,” Medical physics 43(8Part1), 4711–4717 (2016). [DOI] [PMC free article] [PubMed] [Google Scholar]