Abstract
Background
The left bundle branch pacing (LBBP) makes the ventricular depolarization closer to the physiological state and shortens QRS duration. The purpose of this study is to explore the ventricular systolic mechanical synchronization after LBBP in comparison with traditional right ventricular pacing (RVP) using two‐dimensional strain echocardiography (2D‐STE).
Methods
Thirty‐two patients who received LBBP (n = 16) or RVP (n = 16) from October 2018 to October 2019 and met the inclusion criteria were included in this retrospective study. Electrocardiogram (ECG) characteristics, pacing parameters, pacing sites, and safety events were assessed before and after implantation. Acquisition and analysis of ventricular systolic synchronization were implemented using 2D‐STE.
Results
In RVP group, ECG showed left bundle branch block patterns. At LBBP, QRS morphology was in the form of right bundle branch block, and QRS durations were significantly shorter than that of the RVP QRS (109.38 ± 12.89 vs 149.38 \± 19.40 ms, P < .001). Both the maximum time differences (TD) and SDs of the 18‐segments systolic time to peak systolic strain were significantly shorter under LBBP than under RVP (TD, 66.62 ± 37.2 vs 148.62 ± 43.67 ms, P < .01; SD, 21.80 ± 12.13 vs 52.70 ± 17.72 ms, P < .01), indicating that LBBP could provide better left ventricular mechanical synchronization. Left and right ventricular pre‐ejection period difference was significantly longer in RVP group than in LBBP group (10.23 ± 3.07 vs 39.94 ± 14.81 ms, P < .05), indicating left and right ventricular contraction synchronization in LBBP group being better than in RVP group.
Conclusion
LBBP is able to provide a physiologic ventricular activation pattern, which results in ventricular mechanical contraction synchronization.
Keywords: left bundle branch pacing, right ventricle pacing, systolic synchronization, two‐dimensional strain echocardiography
1. INTRODUCTION
The traditional pacing site at the right ventricular apex or the right side of ventricular septum was demonstrated to have detrimental impact on clinical outcomes due to ventricular mechanical dyssynchrony secondary to electrical dyssynchrony. 1 , 2 Studies have demonstrated the feasibility and clinical benefits of permanent His‐bundle pacing (HBP). 3 Former researchers have found that permanent HBP led to significant narrowing of QRS duration and improvement in the left ventricle (LV) function in patients with reduced LV ejection fraction (LVEF). 4 But His bundle pacing still have several disadvantages, including difficulty of lead implantation, lower R‐wave amplitudes, high and unstable pacing threshold, especially in patients who have conduction block distal to the His bundle. 3 , 5 Upadhyay et al demonstrated that the site of block usually was located within the His or proximal left bundle. 6
Thus, alternative pacing sites have been sought. After penetrating through the membranous atrioventricular septum, the conductive fibers of the left bundle branches spread beneath the endocardium of ventricular septum in a relatively large dimension, 7 which offers an opportunity for pacing the left bundle branch (LBB) in an easier manner. Weijian Huang et al developed a technique for the left bundle branch pacing (LBBP) using a transseptal approach. 8 LBBP has been reported to offer low pacing thresholds and large R waves, and has a lower theoretical risk for the development of distal conduction block due to that the distal conduction system is targeted. 9 Keping Chen et al reported that LBBP had a lower pacing threshold and produced narrower electrocardiogram (ECG) QRS duration compared with the right ventricular pacing (RVP). 10 The purpose of this study is to explore the electrical and mechanical synchrony of the ventricle using two‐dimensional strain echocardiography (2D‐STE) after LBBP in comparison with the traditional RVP.
2. METHODS
2.1. Study design
This study was performed in Beijing Friendship Hospital with patients who were diagnosed as II° atrioventricular block (AVB), III° AVB, or sick sinus syndrome, and indicated for pacing therapy according to 2013 ESC/EHRA Guidelines.
The inclusion criteria were as follows: (a) age > 18 years; (b) II° AVB, III° AVB or I° AVB sick sinus syndrome with ventricular pacing ratio > 70%; (c) LVEF >50%; (d) New York Heart Association (NYHA) score of I or II; and (e) being able to visit the hospital during follow‐up. The exclusion criteria were as follows: (a) moderate to severe valvular diseases; (b) congestive heart failure; (c) acute or old myocardial infarction; (d) history of cardiomyopathy; (e) poor condition of the acoustic window because of emphysema or other reasons; or (f) severe liver, lung or kidney dysfunction.
The protocol was approved by Beijing Friendship Hospital Institutional Review Board, and all patients had submitted written informed consent. From October 2018 to October 2019, 32 patients who received LBBP or RVP and met the criteria mentioned above were included.
2.2. Implantation procedures
Before the pacemaker implantation, the ventricular septal thickness was assessed by Echo. Twelve‐lead ECG and intracardiac electrograms were simultaneously displayed and recorded on a multichannel recorder.
LBBP: The delivery sheath (C315 HIS, Medtronic Inc., Minneapolis, MN) was inserted via the left subclavian vein or axillary vein and placed on the right side of the septum inferior to the septal leaflet of the tricuspid valves about 1 to 1.5 cm from HBP site toward right ventricular apex (RAO 30°). Then, the Select Secure pacing lead (Model 3830 69 cm, Medtronic Inc., Minneapolis, MN) was advanced through the sheath with its tip just beyond the distal part of the sheath for unipolar pacing and local activation potential recording. The sheath and the pacing lead touched the septum and pacing with an output of 5.0 V/0.4 ms was applied, which created ECG QRS morphology of “W” pattern with the notch closer to nadir in lead V1. The pacing lead was then screwed perpendicularly to LV septum (LAO 30°‐45°, Figure 1C). During the lead advancement procedure, the changes in the notch in V1 lead, the fulcrum sign, and impedance changes were observed, and sometimes the sheath angiography was used to determine LBBP lead depth into ventricular septum (Figure 1D). Once ECG QRS morphology during pacing resulted in a pattern of the right bundle branch block (RBBB) or a very narrow QRS complex (<120 ms), the lead had been at or near the left bundle branch and the lead advancement was stopped. Then the test with different output was used to confirm LBB capture. Evidences for direct LBB capture were as follows: (a) pacing morphology of RBBB pattern; (b) recording LBB potential; (c) stimulus‐peak RV5 or RV6 shortening abruptly with increasing output or remaining shortest and constant (<90 ms) at low and high outputs; (d) selective LBBP and non‐selective LBBP; or (e) recording retrograde His potential or anterograde LBB potential during pacing (not routine in clinical practice). 11 The pacing parameters were measured to confirm stable capture threshold and consistent pacing impedance, and then the sheath was removed.
RVP: the pacing lead (models 5076, Medtronic Inc., Minneapolis, MN) was positioned in the right ventricular apex (n = 7, the right ventricular apical pacing [RVAP]) or the right side of ventricular septum (n = 9, right ventricular septal pacing [RVSP]). The lead positions in patients with LBBP and RVP were confirmed with the high‐quality fluoroscopic radiographs. All patients underwent the regular follow‐up postimplantation.
2.3. Programming
Conventional lead‐related parameters including stimulation threshold, impedance, and sensing amplitude were recorded right after implantation and 7 days later. Ventricular pacing rates of all patients were more than 70%. The atrioventricular delay was selected, which corrected bundle branch block and yielded narrower QRS duration. Lead‐related adverse events were not observed during follow‐up.
2.4. Echocardiographic parameters
Echocardiography was performed using VVIq ultrasound (GE Company, USA) with S‐5 transducers 7 days after surgery. Acquisition and analysis of synchronization using 2D‐STE were performed by senior sonographers. Standard apex four‐chamber view, apex three‐chamber view, and apex two‐chamber view were clearly exposed and recorded as well as ECG, and four consecutive cardiac cycles with constant heart rate were collected. The 18‐segment systolic times to peak 2‐D longitudinal systolic strain were recorded for every patient. Then the maximum time difference of systolic times to peak 2‐D strain among the 18 LV segments (2D‐TDmax), and the SDs of systolic times to peak systolic strain (PSS) of the 18 LV segments were calculated. The 6‐segment time differences (TDs) to PSS of the apex four‐chamber view, apex three‐chamber view, and apex two‐chamber view were also calculated, respectively. Both the LV function and synchronization status using 2D‐STE were assessed using the offline software TOMTEC (Tom‐Tec Imaging Systems, Unterschleissheim, Germany). Septal‐to‐posterior wall motion delay (SPWMD) was also recorded (Supplementary Figure S1). Interventricular mechanical delay (IVMD) was used to reflect the synchrony between the left and right ventricles. IVMD was the time difference between the left and right ventricular pre‐ejection periods,which were the periods from the start of QRS wave to the start of aortic valve blood flow or pulmonary valve blood flow (Supplementary Figure S2).
2.5. Statistical analysis
Data were presented as mean ± SD for continuous variables. Quantitative data with normal distribution compared between two groups were evaluated using the Student t test. Quantitative data inconsistent with normal distribution compared between two groups were evaluated using the Wilcoxon rank sum test. Categorical data compared between two groups were evaluated using Fisher exact probabilities method. All P values were two‐tailed, and P values <.05 were considered to indicate statistical significance. All statistical analyses were performed using IBM SPSS Statistics 26.
3. RESULTS
3.1. Baseline characteristics
Thirty‐two patients with good acoustic windows were enrolled from October 2018 to October 2019. Sixteen patients were analyzed in each group. Among these patients (mean age = 71.4 ± 14.4 years in the LBBP group and 73.6 ± 8.9 years in the RVP group; P = .910), 27 were diagnosed with AVB, 6 with sick sinus syndrome, 3 with coronary diseases, 8 with hypertension, 11 with diabetes mellitus, and 2 with paroxysmal atrial fibrillation. No statistical differences were noted in baseline clinical characteristics between the two groups (Table 1).
TABLE 1.
Groups | RVP (n = 16) | LBBP (n = 16) | P value |
---|---|---|---|
Demographics | |||
Age (years) | 73.6 ± 8.9 | 71.4 ± 14.4 | .910 |
Sex (Male/Female) | 5/11 | 7/9 | .716 |
Heart rate (beats/min) | 55.3 ± 13.3 | 58.3 ± 14.4 | .545 |
Comorbidities | |||
Coronary disease (yes/no) | 2/14 | 1/15 | 1.0 |
Hypertension (yes/no) | 4/12 | 4/12 | 1.0 |
Paroxysmal atrial fibrillation (yes/no) | 0/16 | 2/14 | .484 |
Diabetes mellitus (yes/no) | 4/12 | 7/9 | .458 |
Other diseases (yes/no) | 2/14 | 1/15 | 1.0 |
Diagnosis | |||
SN dysfunction (yes/no) | 3/13 | 3/13 | 1.0 |
AVB (yes/no) | 14/2 | 13/3 | 1.0 |
LBBB (yes/no) | 2/14 | 1/15 | 1.0 |
RBBB (yes/no) | 3/13 | 2/13 | 1.0 |
Pacing parameters | |||
Sensing amplitude (mV) | 12.37 ± 3.65 | 9.81 ± 4.08 | .052 |
Pacing impedance (Ω) | 678.38 ± 155.53 | 705.37 ± 133.51 | .545 |
Pacing threshold (V) | 0.78 ± 0.19 | 0.71 ± 0.19 | .283 |
Pacing ECG parameters | |||
Baseline ECG QRS duration (ms) | 107.50 ± 28.17 | 106.25 ± 25.00 | .739 |
Paced ECG ORS duration (ms) | 109.38 ± 12.89 | 149.38 ± 19.40 | <.01 |
Pacing‐R wave peak of V5 lead (ms) | 73.44 ± 7.00 | 97.19 ± 8.36 | <.01 |
Pacing‐R wave peak of V6 lead (ms) | 75.31 ± 7.85 | 102.81 ± 8.56 | <.01 |
Abbreviations: AVB, atrium ventricular block; ECG, electrocardiogram; LBBB, left bundle branch block; LBBP, left bundle branch area pacing; RBBB, right bundle branch block; RVP, right ventricular pacing; SN dysfunction, sinus node dysfunction.
3.2. ECG characteristics, lead parameters, and location
ECG was characterized with wide QRS complex and the pattern of LBBB)or delay during either RVAP or RVSP as shown in Figure 2. The LBB potentials were recorded in 8 of 16 patients who received LBBP. The interval from LBB potential to the beginning of ECG QRS was 21.62 ± 6.28 ms. The pacing spike to R wave peak of V5 and V6 leads were 73.44 ± 7.00 and 75.31 ± 7.85 ms, respectively, in the LBBP group. In all 16 patients receiving LBBP, there were equipotential lines between the pacing spike and the QRS wave, and the pacing spike to the beginning of ECG QRS was 28.48 ± 5.69 ms. In patients receiving RVAP or RVSP, ECG QRSs were immediately initiated by pacing without an equipotential line. ECG QRS duration was much shorter in LBBP group compared with that in RVP group (109.38 ± 12.89 vs 149.38 ± 19.40 ms, P < .01), but there was no significant difference in QRS duration between patients of LBBP group and RVP group at baseline.
There was no significant difference in sensing amplitude, pacing impedance, and capture threshold among LBBP and RVP (Table 1). Pacing thresholds and sensing amplitude measured at time of hospital discharge remained stable compared with that measured at implantation.
In patients with LBBP, fluoroscopy at implantation confirmed the location of the pacing lead in the left bundle branch area (Figure 1A‐D), and postimplantation echocardiography confirmed the location of the pacing lead on the left side of the basal septum (an example shown in Figure 1E‐F). Thus, LBBP in general is left septal subendocardial pacing. The echocardiography estimated distance from the lead tip to the right side of the septum in the LBBP group was 0.8 to 1.5 cm depending on the thickness of the basal ventricular septum.
3.3. LV synchronization status
The LV wall was divided into 18 segments, each of which showed individual systolic time to peak 2‐D longitudinal strain (Figure 3). Postoperatively, 18‐segment maximum time difference to peak 2‐D strain (2D‐TDmax) was significantly shorter in the LBBP group than in the RVP group (66.62 ± 37.24 vs 148.62 ± 43.67 ms, P < .01); while at baseline, 2D‐TDmax did not differ between the two groups (51.69 ± 28.97 vs 50.13 ± 25.66 ms, P = .650). There was no statistical difference for 2D‐TDmax in the LBBP group when comparing with baseline (66.62 ± 37.24 vs 51.69 ± 28.97 ms, P > .05). After pacemaker implantation, the 2D‐TDmax was significantly longer compared with baseline in the RVP (148.62 ± 43.67 vs 50.13 ± 25.66 ms, P < .01). Similar results were obtained when comparing SD of 18‐segment systolic times to peak 2‐D strain between the two groups at baseline and after operation (before pacing 17.57 ± 10.04 vs 13.30 ± 8.14 ms, P = .187; after pacing 21.80 ± 12.13 vs 52.70 ± 17.72 ms, P < .01). There was no statistical difference for SD in the LBBP group when comparing with baseline (21.80 ± 12.13 vs 17.57 ± 10.04 ms, P > .05). After pacemaker implantation, the SD was significantly longer compared with baseline in the RVP (52.70 ± 17.72 vs 13.30 ± 8.14 ms, P < .01). (Table 2).
TABLE 2.
Baseline | P value | After pacemaker implantation | P value | |||
---|---|---|---|---|---|---|
Groups | LBBP, n = 16 | RVP, n = 16 | LBBP n = 16 | RVP n = 16 | ||
EDD | 5.07 ± 0.25 | 5.20 ± 0.25 | .113 | 5.10 ± 0.25 | 5.24 ± 0.22 | .097 |
LVEF (%) | 68.69 ± 3.14 | 66.13 ± 4.50 | .097 | 66.94 ± 1.69 | 65.19 ± 3.43 | .136 |
Average of peak longitudinal 2D‐stain (−%) | 21.36 ± 2.05 | 20.10 ± 4.26 | .720 | 20.59 ± 2.44 | 18.90 ± 5.29 | .474 |
SPWMD (ms) | 36.00 ± 21.48 | 32.50 ± 18.58 | .213 | 43.31 ± 23.27 | 93.56 ± 36.77 | <.01 |
left and right ventricular pre‐ejection period difference (ms) | 11.38 ± 7.79 | 12.88 ± 6.99 | .484 | 10.23 ± 3.07 | 39.94 ± 14.81 | <.01 |
18‐segment maximum time difference to peak 2‐D strain (ms) | 51.69 ± 28.97 | 50.13 ± 25.66 | .650 | 66.62 ± 37.24 | 148.62 ± 43.67 | <.01 |
standard deviation of 18‐segment systolic times to peak 2‐D strain (ms) | 17.57 ± 10.04 | 13.30 ± 8.14 | .187 | 21.80 ± 12.13 | 52.70 ± 17.72 | <.01 |
Anteroseptal segment vs posterior segment (ms) | 23.81 ± 10.44 | 19.81 ± 9.98 | .255 | 22.94 ± 29.20 | 96.63 ± 41.24 | <.01 |
Anterior segment vs inferior segment (ms) | 30.31 ± 8.89 | 25.56 ± 11.56 | .806 | 38.88 ± 37.03 | 115.63 ± 59.43 | <.01 |
Septal segment vs lateral segment (ms) | 40.44 ± 10.45 | 36.5 ± 18.78 | .734 | 47.38 ± 29.89 | 127.31 ± 53.18 | <.01 |
Abbreviations: EDD, end diastolic diameter; LBBP, left bundle branch area pacing; LVEF, left ventricular ejection fraction; RVP, right ventricular pacing; SPWMD, septal to posterior wall motion delay.
The 6‐segment maximum TDs to peak 2‐D longitudinal strain of the apex four‐chamber view, apex three‐chamber view, and apex two‐chamber view were longer in RVP group than those in the LBBP group (Table 2, Figure 3). The TD between the anteroseptal wall and posterior wall was statistically longer in the RVP group than in the LBBP group after pacing (22.94 ± 29.2 vs 96.63 ± 41.24 ms, P < .01). Comparing the TD between the anterior wall and inferior wall and the TD between septum and lateral wall, we can get similar statistical results (38.88 ± 37.03 vs 115.63 ± 59.43 ms, P < .01; 47.38 ± 29.89 vs 127.31 ± 53.18 ms, P < .01). The maximum TD was that between the septal and lateral wall in the RVP group (127.31 ± 53.18 ms). No significant TD difference of the abovementioned ventricle segments was found between the RVP group and the LBBP group at baseline. (Table 2).
Compared with RVP group, the SPWMD was significantly shorter in the LBBP group after pacing (43.31 ± 23.27 vs 93.56 ± 36.77 ms, P < .01) (Supplementary Figure S1). IVMD was significantly longer in the RVP group than in LBBP group (10.23 ± 3.07 vs 39.94 ± 14.81 ms, P < .05), and the result indicated that interventricular contractile synchronization was better in the LBBP group than in the RVP group (Supplementary Figure S2).
The 18‐segment times to peak left ventricular systolic strain exhibiting a bull's eye pattern were used to determine left ventricular synchronization status. The activation sequences in the LBBP group generally started from the basal septum and were conducted to the lateral wall or posterior wall. The LV activation sequence was similar in the right ventricular sepal pacing patient. During selective left branch bundle pacing, the range of 18‐segment times to peak left ventricular systolic strain was from 310 to 330 ms, and LV showed uniform gray levels, which represented good left ventricular synchronization status (Figure 4). During right ventricular sepal pacing, the range of 18‐segment times to peak left ventricular systolic strain was from 330 to 495 ms, and LV showed larger gray‐scale difference, which represented worse left ventricular synchronization status compared with LBBP (Figure 4).
3.4. LV function
Two‐dimensional speckle tracking was performed. Neither systolic strain at baseline (21.36 ± 2.05 vs 20.10 ± 4.26 ms, P > .05) nor after pacemaker implantation (20.59 ± 2.44 vs 18.90 ± 5.29 ms, P > .05) revealed significant differences between the RVP and LBBP groups (Table 2). No significant difference was observed in end‐diastolic diameter and LVEF% between the RVP and LBBP groups.
3.5. Limitations in LBBP therapy
During our study, we found that patients with intraventricular block including left anterior and posterior fascicular block cannot achieve LV contraction synchronization after pacing the left bundle branch (Figure 5).
4. DISCUSSION
Conventional pacing therapy places the pacemaker lead in the right ventricular apex or septum, with ventricular pacing forming a wide QRS wave and unsynchronized contraction. The His bundle‐Purkinje conduction system ensures rapid activation in both ventricles and synchronized contraction. 12 Thus, physiological pacing should utilize the natural His bundle‐Purkinje conduction system. Compared with His bundle pacing, the threshold of LBBP distal to conduction lesion is lower and more stable; at the same time, the LBBP sensing is also significantly better than that of the His bundle. Huang W et al reported that LBBP significantly improved LVEF and LVESV in patients with heart failure and LBBB, especially in six cases with LVEF≤ 40%. 5 Jiang et al demonstrated that patients with typical‐LBBB morphology could benefit more from LBBP for QRS correction. 13 In our study, at LBBP the QRS wave is in the form of RBBB, but the QRS duration is significantly shorter than RVP QRS wave (108.89 ± 12.69 vs 151.00 ± 23.31 ms, P < .01). We fused the right and left ventricular depolarization of three patients in this article to make the LBBP QRS duration and morphology more closer to the normal QRS wave (Figure 2).
Therefore, LBBP makes the ventricular depolarization closer to the physiological state and shortens QRS duration. The ultimate goal is to obtain optimized ventricular systolic mechanical synchronization and thus maintain adequate heart function. Our study in this article aimed to evaluate ventricular systolic mechanical synchronization during permanent LBBP compared with RVP using 2D‐speckle tracking echocardiography. Both 2D‐TDmax and SD of the 18‐segment systolic time to PSS were significantly shorter under LBBP than under RVP, indicating that LBBP could provide better LV mechanical synchronization. As shown in the bull's eye exhibiting the 18‐segments times to peak systolic strain, the LV activation sequences during LBBP were from the septum to the lateral wall, which was consistent with the intrinsic conduction pattern. The earliest activation segments of the LV in the LBBP group were mostly in the basal and middle parts of the septum, and the latest activation segment was the lateral wall. During right ventricular sepal pacing, although the LV activation sequence was similar as the LBBP, the range of 18‐segment times to PSS was wider, and LV showed larger gray‐scale differences, which represents worse left ventricular synchronization status compared with LBBP. We also found that left and right ventricular pre‐ejection period difference which represented left and right ventricular contraction synchronization was significantly longer in the RVP group than in the LBBP group, and the result indicated that left and right ventricular contraction synchronization was better in the LBBP group than in the RVP group. However, the LVEF and average of 18‐segment PSS failed to reveal any significant differences between the LBBP group and the RVP group. This might be due to the limited observation period.
It is precisely because LBBP can bring about synchronization of ventricular contraction that more and more studies report that this technology is currently being used in the treatment of heart failure combined with left bundle branch block. 14 , 15 LBBP was also used in patients of atrioventricular node ablation with persistent atrial fibrillation and implantable cardioverter‐defibrillator therapy in order to improve LV function. 16 In this study, we encountered an interesting case with atrial fibrillation, III° AVB and symptoms of heart failure including dyspnea and decreased activity tolerance. After being upgraded to LBBP using a 3830 pacing lead, the patient's symptoms of heart failure alleviated, and the NYHA heart function class improved from class III to class II about 1 month after operation (Supplementary Figure S3). For those right ventricular apex pacing patients with heart failure, upgrading ventricular lead to LBBP may improve LV systolic mechanical synchronization and alleviate heart failure.
In our study, LBBP is clinically feasible in patients, and pacing parameters were normal during pacemaker implantation and follow‐up. Former researches also got similar results. 17
There were also limitations in LBBP therapy. Because LBBP cannot correct the intraventricular block which was distal to the left bundle branch. Thus, if the widened QRS wave and left ventricular systolic dyssynchrony was caused by intraventricular block, it cannot be corrected by LBBP.
5. CONCLUSIONS
LBBP was able to provide a physiologic ventricular activation pattern almost identical to the intrinsic conduction pattern, which resulted in ventricular mechanical contraction synchronization. Ventricular mechanical contraction synchronization after permanent LBBP helps to maintain good heart function and prevent detrimental impact of RVP on clinical outcomes due to ventricular mechanical dyssynchrony.
CONFLICT OF INTEREST
We declare that there is no conflict of interest in this work.
AUTHOR CONTRIBUTIONS
Beibing Di and Zhijun Sun contributed equally to this study.
Supporting information
Sun Z, Di B, Gao H, Lan D, Peng H. Assessment of ventricular mechanical synchronization after left bundle branch pacing using 2‐D speckle tracking echocardiography. Clin Cardiol. 2020;43:1562–1572. 10.1002/clc.23481
Funding information Beijing Municipal Natural Science Foundation, Grant/Award Number: 7194252; National Science Foundation, Grant/Award Number: IIP‐1624641 CBET‐1218345; Advanced Mammalian Biomanufacturing Innovation Center (AMBIC)
DATA AVAILABILITY STATEMENT
The data used to support the findings of this study are available from Beibing Di (beibingyouxiang@163.com) upon reasonable request.
REFERENCES
- 1. Nahlawi M, Waligora M, Spies SM, Bonow RO, Kadish AH, Goldberger JJ. Left ventricular function during and after right ventricular pacing. J Am Coll Cardiol. 2004;44:1883‐1888. [DOI] [PubMed] [Google Scholar]
- 2. Hussain MA, Furuya‐Kanamori L, Kaye G, Zhijun S, Beibing D, Hui P. The effect of right ventricular apical and nonapical pacing on the short‐ and long‐term changes in left ventricular ejection fraction: a systematic review and meta‐analysis of randomized‐controlled trials. Pacing Clin Electrophysiol. 2015;38:1121‐1136. [DOI] [PubMed] [Google Scholar]
- 3. Vijayaraman P, Chung MK, Dandamudi G, et al. His bundle pacing. J Am Coll Cardiol. 2018;72:927‐947. [DOI] [PubMed] [Google Scholar]
- 4. Shan P, Su L, Zhou X, et al. Beneficial effects of upgrading to his bundle pacing in chronically paced patients with left ventricular ejection fraction <50%. Heart Rhythm. 2018;15:405‐412. [DOI] [PubMed] [Google Scholar]
- 5. Huang W, Su L, Wu S, et al. Long‐term outcomes of his bundle pacing in patients with heart failure with left bundle branch block. Heart. 2019;105:137‐143. [DOI] [PubMed] [Google Scholar]
- 6. Upadhyay GA, Cherian T, Shatz DY, et al. Intracardiac delineation of septal conduction in left bundle‐branch block patterns. Circulation. 2019;139:1876‐1888. [DOI] [PubMed] [Google Scholar]
- 7. Anderson RH, Yanni J, Boyett MR, Chandler NJ, Dobrzynski H. The anatomy of the cardiac conduction system. Clin Anat. 2009;22:99‐113. [DOI] [PubMed] [Google Scholar]
- 8. Huang W, Su L, Wu S, et al. A novel pacing strategy with low and stable output: pacing the left bundle branch immediately beyond the conduction block. Can J Cardiol. 2017;33:1736.e1‐1736.e3. [DOI] [PubMed] [Google Scholar]
- 9. Su L, Wu S, Wang S, et al. Pacing parameters and success rates of permanent his‐bundle pacing in patients with narrow QRS: a single‐Centre experience. Europace. 2019;21:763‐770. [DOI] [PubMed] [Google Scholar]
- 10. Chen K, Li Y, Dai Y, et al. Comparison of electrocardiogram characteristics and pacing parameters between left bundle branch pacing and right ventricular pacing in patients receiving pacemaker therapy. Europace. 2018;0:1‐8. [DOI] [PubMed] [Google Scholar]
- 11. Huang W, Chen X, Su L, Wu S, Xia X, Vijayaraman P. A beginner's guide to permanent left bundle branch pacing. Heart Rhythm. 2019;16:1791‐1796. [DOI] [PubMed] [Google Scholar]
- 12. Tang J, Chen S, Liu L, et al. Assessment of permanent selective his bundle pacing in left ventricular synchronization using 3‐Dspeckle tracking echocardiography. Ultrasound Med Biol. 2019;45:385‐394. [DOI] [PubMed] [Google Scholar]
- 13. Jiang Z, Chang Q, Wu Y, Ji L, Zhou X, Shan Q. Typical BBB morphology and implantation depth of 3830 electrode predict QRS correction by left bundle branch area pacing. Pacing Clin Electrophysiol. 2020;43:110‐117. [DOI] [PubMed] [Google Scholar]
- 14. Zhang W, Huang J, Qi Y, et al. Cardiac resynchronization therapy by left bundle branch area pacing in patients with heart failure and left bundle branch block. Heart Rhythm. 2019;16:1783‐1790. [DOI] [PubMed] [Google Scholar]
- 15. Li X, Li H, Ma W, et al. Permanent left bundle branch area pacing for atrioventricular block: feasibility, safety, and acute effect. Heart Rhythm. 2019;16:1766‐1773. [DOI] [PubMed] [Google Scholar]
- 16. Wang S, Wu S, Xu L, et al. Feasibility and efficacy of his bundle pacing or left bundle pacing combined with Atrioventricular node ablation in patients with persistent atrial fibrillation and implantable Cardioverter‐defibrillator therapy. J Am Heart Assoc. 2019;8:e014253. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17. Chen X, Wu S, Su L, Su Y, Huang W. The characteristics of the electrocardiogram and the intracardiac electrogram in left bundle branch pacing. J Cardiovasc Electrophysiol. 2019;30:1096‐1101. [DOI] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The data used to support the findings of this study are available from Beibing Di (beibingyouxiang@163.com) upon reasonable request.