Silicon Photomultipliers as a Low-Cost Fluorescence Detector for Capillary Electrophoresis

Brae V Petersen 1,2, Luke Gallion 1,2, Nancy L Allbritton 1,*
PMCID: PMC7728455  NIHMSID: NIHMS1650489  PMID: 32967426

Abstract

Capillary electrophoresis (CE) is a highly efficient separation method capable of handling small sample volumes (~pL) and low (~yoctomole) detection limits, and as such, is ideal for applications that require high sensitivity such as single-cell analysis.1-3 Low-cost CE instrumentation is quickly expanding but low-cost, open-source fluorescence detectors with ultra-sensitive detection limits are lacking.4-6 Silicon photomultipliers (SiPM) are inexpensive, low-footprint detectors with the potential to fill the role as a detector when cost, size, and customization are important. In this work, we demonstrate the use of a SiPM in CE with zeptomolar detection limits and a dynamic range spanning five orders of magnitude, comparable to photomultiplier detectors. The performance of these detectors was measured using a constant wave excitation laser in an epifluorescence detection configuration. We characterize the performance of the SiPM as a highly sensitive detector by measuring enzyme activity in single cells. This simple, small footprint, and low-cost (<$130) light detection circuit will be beneficial for open-source, portable, and budget-friendly instrumentation requiring high sensitivity.

Graphical Abstract

graphic file with name nihms-1650489-f0001.jpg


Cell heterogeneity has become increasingly recognized as an important factor in many aspects of pathology, immunology, and stem-cell biology.7,8 As a result, a myriad of analytical tools have been developed to evaluate this heterogeneity.2,3 Capillary electrophoresis (CE) with fluorescence detection is one such tool that has been widely utilized in the academic laboratory to analyze single cells, largely due to the high sensitivity and broad dynamic range of fluorescent detectors.2,3 Zeptomole (10−21) detection limits are typical, with some setups capable of achieving yoctomole (10−24) limits of detection (LOD).9 Furthermore, these instruments are well suited for chemical cytometry since CE instruments can handle small (~pL) sample volumes. Therefore single-cell lysates do not experience significant dilution prior to assay. However, CE-based chemical cytometry systems have not seen widespread use clinically, partially due to the high cost, large instrument footprint, and high technical expertise requirements. As such, instrument modifications to address these issues should be explored.

Several fluorescent detection setups have been described for chemical cytometry, many of which have employed photomultiplier tubes (PMTs).9-11 PMTs work using a photocathode housed within a vacuum environment to transduce light into an electrical signal. PMTs have high gain (105−9) and maintain linearity over a wide dynamic range, typically as many as six orders of magnitude.10,12 For these reasons, PMTs are commonly utilized in CE systems where cost is not a driving factor. With the development of low-cost open-source CE setups, there is an unmet desire for an inexpensive, open-source detector that can be customized according to the need of the instrument or project.6

Silicon photomultipliers (SiPM) are significantly less expensive than PMTs but have similar gains (105−6) and have been suggested for use as fluorescence detectors for low-cost CE instruments.5,12 However, applications using SiPMs as a fluorescence detector in analytical instruments have only recently been explored.13-15 SiPMs consist of arrays of micron-sized photodiodes (i.e., 50x50 μm dimensions) that are biased as single-photon avalanche diodes (SPAD).16 Each microcell produces a digital current signal when struck by a photon, with the entire array producing an analog signal at higher light densities.12 Several methods have emerged for deploying a SiPM in experiments for photon counting where a pulsed light source and high bandwidth analog to digital converter (ADC) or oscilloscope is used.17,18 Low frequency or constant power applications have also begun to emerge for projects focused on providing affordable instrumentation.13,19,20 We sought to apply a simple circuit for the SiPM detector for use with a detection apparatus and signal processing workflow designed for on-column epifluorescence detection. By doing so, we found the SiPM can obtain a molar LOD of 10−21 moles and span a linear dynamic range of 5-6 orders of magnitude, thus justifying. SiPM use in affordable chemical cytometry instrumentation and particularly CE-based instruments. The detection circuitry is both low-cost (~$130) and relatively simple, making it ideal for a fluorescence detector for affordable and open-source instrumentation.

EXPERIMENTAL

Materials and Reagents

Sphingosine fluorescein (SF) and sphingosine 1-phosphate fluorescein (S1PF) were purchased from Echelon Biosciences Inc. (Cat. # S-100F and #S-200F, Salt Lake City, UT). Trizma-base, 4-(2-hydroxyethyl)-1-piperazineethanesulfonic acid (HEPES), 1-propanol, potassium chloride, and magnesium chloride were purchased from Millipore Sigma. Tween 20 was purchased from BioRad Laboratories. Glass slides, calcium chloride, sodium phosphate was purchased from Fisher Scientific. Dulbecco’s modified eagle’s medium (DMEM) and fetal bovine serum (FBS) were from ThermoScientific (Waltham, MA).

Instrumentation

For this work, a modular CE detection assembly using epi-illumination was designed and built (Figure 1). The detection window is created by removing the capillary coating 4 cm from the capillary inlet using a flame. The detection window is aligned perpendicular to the light path within the detection objective using 3D printed levers secured with 2M threaded screws to adjust the capillary position. Excitation light from an optical fiber pigtail coupled to a 488 nm constant wave (CW) laser is collimated using a fiber receptacle collimator (OZ Optics, Ottawa, Canada) and directed to the detection objective using a 488 nm dichroic filter. A 0.75 numeric aperture (NA) 40x objective focuses incoming excitation light to the capillary lumen. The emission light is collimated by the objective and passes through a dichroic, a 525 nm bandpass filter, and 6.5 mm iris before striking the detector. The detectors are attached to a four-way optical block and can be exchanged without changing the alignment of the capillary or other aspects of the CE detection assembly. A 3.7 x 13 mm side on PMT (H9306, Hamamatsu, Bridgewater, NJ) was used to provide a comparison to the SiPM detectors. A parts list for the detection assembly is available in the supplemental information (Table S1).

Figure 1.

Figure 1.

CE fluorescence detection assembly and simplified pathway. Diagram (a) and photo (b) of the optical pathway used for testing the detectors. Excitation light from a 488 nm CW laser is collimated by a fiber optic coupler (i). A dichroic beam splitter (iv) reflects the light to a 0.75 NA 40x objective (ii), which then focuses the light onto the capillary (iii). Emission light (from fluorophores within the capillary lumen) travels through the beam splitter (iv), and this emitted light is filtered using a 520 nm low pass filter (v). Stray light is spatially filtered using an iris (vi). A 4-way optic mount (vii) is used to mount the detectors (viii) to the optical path. c) Signal flow chart from SiPM to final electropherogram. A negative current is generated by the SiPM as light strikes the microcells. An inverting transimpedance amplifier converts the current into a voltage using the feedback resistor, Rf. The amplifier output is filtered by a passive low-pass filter then oversampled at 75 kHz using an ADC. The oversampled signal is averaged to 8 Hz and a Butterworth digital filter applied to the 8 Hz data to obtain the final electropherogram (time vs. relative fluorescence)

The custom CE instrument used for this work incorporates characteristics of traditional CE as well as modifications to improve capillary positioning and control for single-cell analysis.11 An overview of the custom instrument is provided in supplemental (Figure S1).

SiPM Detector

SiPM detectors (MICROFC −30035, −30050, and −60035) were procured from OnSemiconductor (Phoenix, AZ). Printed circuit boards (PCB) were designed using an opensource electronics design software suite (KiCad EDA 5.1. http://kicad-pcb.org/, March 6, 2020) and fabricated by BasicPCB (Aurora, Colorado, USA). The SiPM circuit (Figure S2) and parts list (Table S2) are provided in the supplemental along with Gerber files for the PCB fabrication and a complete bill of materials. The manufacturer’s recommendations were followed for assembly. The SiPM chip was negatively biased 1-5 V above the breakdown voltage. The amplifier feedback resistor was selected to achieve the same signal voltage under saturation conditions (470 and 330 Ω for the 9mm2 and 36 mm2 sensors, respectively). Measurements were taken at 1 V intervals. The signal from the detectors was filtered using a passive low-pass hardware filter (5th order elliptical, 1 dB at 1kHz, see Table S1). Oversampling filtering was performed at 75 kHz and averaged to achieve a final sampling rate of 8 Hz to reduce random noise. A 2nd order digital butter filter with a cutoff of 1.5 Hz was applied twice (forwards and backward) to the 8 Hz electropherogram.

Limits of Detection

The signal to noise ratio (SNR) was measured for each of the detectors. The signal was measured from a 41 nM fluorescein solution loaded into the capillary by gravity (3 cm height difference for 3 s). The total injection volume (1.0 nL) was calculated using the volume loaded due to gravity plus the volume loaded due to diffusion and spontaneous fluid displacement.21 To measure spontaneous fluid displacement, the average corrected area from 3 cm gravity injections (n=7) was compared to the average corrected area from 0 cm gravity injections (n=7). The volume due to spontaneous fluid loading was calculated according to Equation 1, where CA is the corrected area for the measured 0 cm and 3 cm injections, and V3cm is the volume loaded into the capillary by gravity:

Vsfd=CA0cmCA3cmCA0cmV3cm (1)

At this volume, the total amount of fluorescein loaded into the capillary was 2.5x10−17 moles. The noise was calculated using the standard deviation across 40 consecutive points (corresponding to a time window of 5 s) of the electropherogram with the least detectable signal. A signal of 3x the noise was used to calculate the molar LOD for each sensor.

Dynamic Range

Fluorescein standards were prepared to create a calibration curve near the saturation point for the detectors. Fluorescein standards were prepared using a 1:3 serial dilution in tris buffer (100 mM, pH 8.1, conductivity 5.3 mS/cm) from 10 μM stock solution. The peak areas were integrated from the filtered electropherograms using a trapezoid approximation function. Corrected areas were calculated by dividing the peak area by the corresponding retention time.

CE Separations

A 50 μm inner diameter, 375 μm outer diameter capillary (#1068150017, Molex, Lisle, IL) with a total length of 45 cm and a detection length of 4 cm was used. Buffer conductivity was measured using a conductivity meter (#4360, Traceable, Webster, TX). At the start of the day, the capillary was conditioned by rinsing for 10 min with 0.1 M HCl, 30 min with 1 M NaOH, followed by 5 min rinses with H2O, and separation buffer. Before each separation, the capillary was rinsed with separation buffer for 30 s. All separations were performed with an electric field strength of 333 V/cm across the capillary. For the fluorescein separations, the separation buffer consisted of 100 mM tris buffered to pH 8.0 with a buffer conductivity of 5.3 mS/cm.

For the sphingolipid separations, a 30 μm inner diameter, 375 μm outer diameter capillary (#1068150013, Molex, Lisle, IL) with a total length of 60 cm and a detection window at 5 cm was used. The separation buffer consisted of 27 mM phosphate at pH 7.4 and 10% n-propanol with a conductivity of 3.0 mS/cm. An electric field strength of 250 V/cm was applied during electrophoresis. SF and S1PF were identified by comparison of their migration times to standards while hexadecenoic acid fluorescein (HAF) was identified from its elution order based prior assays with SF.

Cell Preparation

The K-562 cell line was obtained from the American Type Culture Collection (ATCC) (CRL-1469, ATCC, Manassas, Virginia) and maintained in a humidified atmosphere of 37°C in 5% CO2. Cells were cultured in Dulbecco’s Modified Eagle Medium (DMEM) supplemented with 10% fetal bovine serum (FBS), penicillin (100 units/mL), and streptomycin (100 μg/mL).

SF was loaded into cells by incubating 5 x 105 cells in 100 μL culture media containing freshly diluted 10 μM SF for 30 min, as described previously.21 Cells were stored at 37°C in a 5% carbon dioxide atmosphere during incubation with SF. Cells were centrifuged (700 x g, 2.5 min) and then washed 2 x with 200 μL physiologic cell buffer (135 mM sodium chloride, 5 mM potassium chloride, 1 mM magnesium chloride, 1 mM calcium chloride, and 10 mM HEPES at pH 7.4). An aliquot of 10,000 cells was removed and overlaid onto a glass slide with 600 μL physiologic buffer.

RESULTS AND DISCUSSION

Limits of Detection

SiPM detectors are arrays of micron-sized photodiodes, referred to as microcells, connected in parallel. These microcells are operated in Geiger mode above their breakdown voltage. As a photon strikes a photodiode, a current pulse is produced. At high light levels, hundreds or thousands of photodiodes are simultaneously activated, and their total current can be recorded as a DC analog signal using low-cost electrical components.12,16 While SiPMs can be used in photon counting mode, the excitation light that reached the detector for this setup was too high to employ photon counting. For this reason, only low bandwidth DC measurements from the SiPMs were measured. Of the factors that may influence performance, we hypothesized fill factor would considerably influence SiPM signal strength and sensitivity, as a greater fill factor correlates with a greater probability a photon will strike a photo-responsive area.12 Fill factor increases as the microcell size increases since fewer microcells are needed for the same area and thus require fewer circuit structures (fabricated wires and resistors) for the array. Sensors with different SPAD microcell sizes and total array sizes were tested to gauge which SiPM configuration leads to the greatest SNR.

The LOD of the SiPMs and PMT were compared using a CE detection pathway configured in an epifluorescence format (Figure 1). A feature of this design and housing is that the detector can be easily removed and then re-inserted without modifying the capillary alignment facilitating direct comparison between the different fluorescence detectors. In addition to the three SiPMs, a PMT previously measured to possess zeptomole detection limits was used as a reference detector.21 LODs of three SiPM sensors were measured: a 9 mm2 sensor area with 50 μm microcells, a 9 mm2 sensor area with 35 μm microcells, and a 36 mm2 sensor area with 35 μm microcells. The SNR of each sensor (Figure S3) was used to determine the 3σ limit of detection as a function of the peak height (Figure 2). A sensor with a lower peak height at an equivalent SNR will be able to utilize more of the op-amp and ADC dynamic range. Of the SiPM detectors, the 36 mm2 sensor was the most sensitive, followed by the 50 and 35 μm microcell sized 9 mm2 sensors. The 36 mm2 sensor had the greatest sensor surface area and nearly 4x the number of microcells as the 35 μm 9 mm2 sensor (Table S3), which likely contributed to its improved performance. The 50 μm SPAD microcell size possessed an improved LOD compared to the SiPM with 35 μm microcells. The increased microcell fill factor for the 50 μm (72%) compared to 35 μm microcell array (64%) likely contributed to the improved LOD. In comparison to the PMT, the LOD of the 36 mm2 SiPM (8.5x10−21 mol) was in the same order of magnitude as the 48.1 mm2 PMT (3.9x10−21 mol). A comparison of the noise for each of the detectors (figure S4) shows the same trend as the LOD, with the PMT having a slightly lower noise than the SiPM for equivalent signals.

Figure 2.

Figure 2.

Estimated limit of detection (LOD) from the SNR for fluorescein as a function of signal voltage (mV) for each detector. LOD was estimated as the minimal signal detected at 3x noise. On the x-axis, the signal voltage corresponds to the maximum voltage for the fluorescein peak after subtracting the noise. The moles detected is estimated from the total volume loaded into the capillary (1.0 nL).

Dynamic Range and Signal Saturation

The dynamic range of each sensor was measured using a calibration curve (Figure 3). The bias voltage for each SiPM was chosen by selecting that which provided the best SNR (which also corresponded with the maximum bias voltage of the chips). The gain for the PMT was chosen as the point where SNR began to plateau with increasing signal (Figure 2, 350 mV). Past this gain setting, only slight improvement in LOD could be achieved at significant loss to dynamic range. Standards were prepared to span a concentration range of 10−12 to 10−5 M. As expected, at the highest fluorescein concentrations (or emission intensities), the output from the SiPM detectors reaches a maximum number of cells that are activated (Figure S5). The high gain PMT setting and 9 mm2 50 μm and 36 mm2 35 μm SiPMs reached signal saturation at micromolar fluorescein concentrations. The lower capacitance 9 mm2 (35 μm microcell) SiPM did not show significant peak distortion due to saturation. The shorter deadtime of the smaller 9mm2 sensor likely contributed to its improved dynamic range compared to the larger 36 mm2 SiPM. However, the larger microcells and larger sensor areas provided greater sensitivity.

Figure 3.

Figure 3.

Comparison of the dynamic range for the SiPM and PMT detectors. A fluorescein calibration curve with concentrations from 2.5 μM to 9 pM (n=3) was measured with each detector. SiPM bias voltages were set to their maximum values (29.5 V) while the PMT was set to 300 mV corresponding to its maximum SNR before plateauing (Figure 2).

Single-Cell Analysis

Sphingosine kinase (SK) activity is often increased in a wide range of diseases, including Multiple Sclerosis, Alzheimer’s, and various cancers and leukemias, among others.22,23 SK inhibitors have also been shown to be lethal to cancer cells dependent on high SK activity.24 However, SK signaling can be highly heterogeneous between cells of the same tumor sample.21 We have previously demonstrated that CE can be used to measure SK activity in single cells using the substrate SF.21 The SF fluorescent reporter consists of sphingosine with a fluorescein tag conjugated to the acyl chain, which is converted to S1PF by SK.25 Separation and quantification of the fluorescence of SF and S1PF from a single cell supplies a measure of the activity of SK in that cell.25 The 36 mm2 (35 μm microcell) with a 330 Ω feedback resistance with a bias set to half its maximum (−27.5 V) was used to measure the SK activity in K562 leukemia (Figure 4). The detector was able to detect the main sphingosine fluorescein and the sphingosine-1-phosphate fluorescein products necessary for the assay and demonstrates the ability of the SiPM to be applied to CE based chemical cytometry.

Figure 4.

Figure 4.

Electropherogram trace from a single cell loaded with SF. This assay measures the sphingosine kinase activity by quantifying the amount of reporter SF converted to S1PF by sphingosine kinase. Conversion of S1PF into HAF occurs due to the actions of sphingosine 1-phosphate lyase on S1PF.

CONCLUSION

This work demonstrates that the SiPM can be used as a sensitive and affordable fluorescence detector for CE. The dynamic range required for chemical cytometry spans 3-5 orders of magnitude to match the range of the total amount of reporter loaded into a cell for peptide kinase reporters (10−21 to 10−19 mols) and lipid kinase reporters (10−19 to 10−17 mols).3,21 The SiPM in this configuration has a dynamic range that spans five orders of magnitude while detecting as little as 10−12 M or 10−21 moles of fluorescein. Thus, the SiPM detector possesses comparable performance characteristics as a PMT and is suitable for chemical cytometry assays by CE. Notably, the SiPM and shown here can be assembled and powered for less than 3% of the cost of a traditional PMT (Table S5). Commercially available SiPMs may provide an additional decrease in the noise compared to the detectors assessed here at a slightly increased system cost. An asset of the current SiPM chips is that the detectors can easily be configured according to the application need. This aspect provides flexibility for miniature, portable instrumentation, as well as microfluidic devices in situations where sensitivity, cost, and portability are important factors.21

Supplementary Material

Supplemental

ACKNOWLEDGMENTS

This work is supported by the National Institutes of Health grants R01CA233811, F31CA228223, and F31CA243312. The Authors would also recognize Pete Attayek for his contributions in designing the epi-illumination pathway.

Footnotes

SUPPORTING INFORMATION

The supplemental information includes a description of the custom CE instrumentation, a list of materials for the fluorescence pathway and SiPM detector, and SiPM detector circuit and schematic. Additional data describing the performance of the detector, comparing SNR to detector bias, and traces of the detector’s outputs at low and high intensity fluorescence are included.

REFERENCES

  • (1).Chen D; Dovichi NJ Single-Molecule Detection in Capillary Electrophoresis:  Molecular Shot Noise as a Fundamental Limit to Chemical Analysis. Anal. Chem 1996, 68 (4), 690–696. 10.1021/ac950651r. [DOI] [Google Scholar]
  • (2).Cohen D; Dickerson JA; Whitmore CD; Turner EH; Palcic MM; Hindsgaul O; Dovichi NJ Chemical Cytometry: Fluorescence-Based Single-Cell Analysis. Annual Review of Analytical Chemistry 2008, 1 (1), 165–190. 10.1146/annurev.anchem.1.031207.113104. [DOI] [PubMed] [Google Scholar]
  • (3).Vickerman BM; Anttila MM; Petersen BV; Allbritton NL; Lawrence DS Design and Application of Sensors for Chemical Cytometry. ACS Chem. Biol 2018, 13 (7), 1741–1751. 10.1021/acschembio.7b01009. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (4).Fang X-X; Fang P; Pan J-Z; Fang Q A Compact Short-Capillary Based High-Speed Capillary Electrophoresis Bioanalyzer. ELECTROPHORESIS 2016, 37 (17–18), 2376–2383. 10.1002/elps.201600195. [DOI] [PubMed] [Google Scholar]
  • (5).Casto LD; Do KB; Baker CA A Miniature 3D Printed LED-Induced Fluorescence Detector for Capillary Electrophoresis and Dual-Detector Taylor Dispersion Analysis. Anal. Chem 2019. 10.1021/acs.analchem.8b05824. [DOI] [PubMed] [Google Scholar]
  • (6).Kubáň P; Foret F; Erny G Open Source Capillary Electrophoresis. ELECTROPHORESIS 2019, 40 (1), 65–78. 10.1002/elps.201800304. [DOI] [PubMed] [Google Scholar]
  • (7).Meacham CE; Morrison SJ Tumour Heterogeneity and Cancer Cell Plasticity. Nature 2013, 501 (7467), 328–337. 10.1038/nature12624. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (8).Barker N; Bartfeld S; Clevers H Tissue-Resident Adult Stem Cell Populations of Rapidly Self-Renewing Organs. Cell Stem Cell 2010, 7 (6), 656–670. 10.1016/j.stem.2010.11.016. [DOI] [PubMed] [Google Scholar]
  • (9).Yong Chen D; Dovichi NJ Yoctomole Detection Limit by Laser-Induced Fluorescence in Capillary Electrophoresis. Journal of Chromatography B: Biomedical Sciences and Applications 1994, 657 (2), 265–269. 10.1016/0378-4347(94)00014-X. [DOI] [PubMed] [Google Scholar]
  • (10).Galievsky VA; Stasheuski AS; Krylov SN “Getting the Best Sensitivity from on-Capillary Fluorescence Detection in Capillary Electrophoresis” - A Tutorial. Analytica Chimica Acta 2016, 935, 58–81. 10.1016/j.aca.2016.06.015. [DOI] [PubMed] [Google Scholar]
  • (11).Abraham DH; Anttila MM; Gallion LA; Petersen BV; Proctor A; Allbritton NL Chapter Ten - Design of an Automated Capillary Electrophoresis Platform for Single-Cell Analysis In Methods in Enzymology Allbritton NL, Kovarik ML, Eds.; Enzyme Activity in Single Cells; Academic Press, 2019; Vol. 628, pp 191–221. 10.1016/bs.mie.2019.06.016. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (12).Mazzillo M; Condorelli G; Sanfilippo D; Valvo G; Carbone B; Fallica G; Billotta S; Belluso M; Bonanno G; Cosentino L; Pappalardo A; Finocchiaro P Silicon Photomultiplier Technology at STMicroelectronics. IEEE Transactions on Nuclear Science 2009, 56 (4), 2434–2442. 10.1109/TNS.2009.2024418. [DOI] [Google Scholar]
  • (13).Shin Y-H; Barnett JZ; Gutierrez-Wing MT; Rusch KA; Choi J-W A Portable Fluorescent Sensing System Using Multiple LEDs. In Microfluidics, BioMEMS, and Medical Microsystems XV; International Society for Optics and Photonics, 2017; Vol. 10061, p 100610M 10.1117/12.2261292. [DOI] [Google Scholar]
  • (14).Santangelo MF; Cono SL; Vasquez P; Fallica G; Conoci S; Busacca AC; Pagano R; Sciuto EL; Lombardo S; Libertino S CY5 Fluorescence Measured with Silicon Photomultipliers. In 2014 IEEE Biomedical Circuits and Systems Conference (BioCAS) Proceedings; 2014; pp 284–287. 10.1109/BioCAS.2014.6981718. [DOI] [Google Scholar]
  • (15).Dorosz P; Baszczyk M; Kucewicz W; Mik L Silicon Photomultipliers Applied to the Fluorescence Detection of Biomarkers. Nuclear Instruments and Methods in Physics Research Section A: Accelerators, Spectrometers, Detectors and Associated Equipment 2018. 10.1016/j.nima.2018.10.156. [DOI] [Google Scholar]
  • (16).C-Series SiPM Sensors: Silicon Photomultipliers (SiPM), Low-Noise, Blue-Sensitive. On Semiconductor.
  • (17).Modi M; Turner GC; Podgorski K Two-Photon Imaging with Silicon Photomultipliers. bioRxiv 2019, 717850 10.1101/717850. [DOI] [PubMed] [Google Scholar]
  • (18).Santangelo MF; Sciuto EL; Busacca AC; Petralia S; Conoci S; Libertino S SiPM as Miniaturised Optical Biosensor for DNA-Microarray Applications. Sensing and Bio-Sensing Research 2015, 6, 95–98. 10.1016/j.sbsr.2015.08.003. [DOI] [Google Scholar]
  • (19).Mik Ł; Stẹpień J; Jastrzạb M; Kucewicz W; Sapor M System for Low Intensity Fluorescence Light Measurement Based on Silicon Photomultiplier. In ICSES 2010 International Conference on Signals and Electronic Circuits; 2010; pp 383–386. [Google Scholar]
  • (20).Makkar RL; Aliya SS; Borse V; Srivastava R Design and Development of Portable Fluorescence Reader Using Silicon Photomultiplier (SiPM) Sensor. In Optical Sensing and Detection V; International Society for Optics and Photonics, 2018; Vol. 10680, p 106800B 10.1117/12.2305846. [DOI] [Google Scholar]
  • (21).Dickinson AJ; Hunsucker SA; Armistead PM; Allbritton NL Single-Cell Sphingosine Kinase Activity Measurements in Primary Leukemia. Anal Bioanal Chem 2014, 406 (27), 7027–7036. 10.1007/s00216-014-7974-6. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (22).Powell JA; Wallington-Beddoe CT; Pitson SM Targeting Sphingosine Kinase 1 in Acute Myeloid Leukemia: Translation to Clinic. Int J Hematol Oncol 2017, 6 (2), 31–34. 10.2217/ijh-2017-0011. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (23).Maceyka M; Harikumar KB; Milstien S; Spiegel S Sphingosine-1-Phosphate Signaling and Its Role in Disease. Trends Cell Biol. 2012, 22 (1), 50–60. 10.1016/j.tcb.2011.09.003. [DOI] [PMC free article] [PubMed] [Google Scholar]
  • (24).Cao M; Ji C; Zhou Y; Huang W; Ni W; Tong X; Wei J-F Sphingosine Kinase Inhibitors: A Patent Review. International Journal of Molecular Medicine 2018, 41 (5), 2450–2460. 10.3892/ijmm.2018.3505. [DOI] [PubMed] [Google Scholar]
  • (25).Determination of Sphingosine Kinase Activity for Cellular Signaling Studies ∣ Analytical Chemistry https://pubs.acs.org/doi/full/10.1021/ac702305q (accessed Jul 16, 2020). [DOI] [PMC free article] [PubMed]

Associated Data

This section collects any data citations, data availability statements, or supplementary materials included in this article.

Supplementary Materials

Supplemental

RESOURCES