Abstract
Background
The collapsibility index of the inferior vena cava (cIVC) has potential for predicting fluid responsiveness in spontaneously breathing patients, but a standardized approach for measuring the inferior vena cava diameter has yet to be established.
The aim was to test the accuracy of different measurement sites of inferior vena cava diameter to predict fluid responsiveness in spontaneously breathing patients with sepsis-related circulatory failure and examine the influence of a standardized breathing manoeuvre.
Results
Among the 81 patients included in the study, the median Simplified Acute Physiologic Score II was 34 (24; 42). Sepsis was of pulmonary origin in 49 patients (60%). Median volume expansion during the 24 h prior to study inclusion was 1000 mL (0; 2000). Patients were not severely ill: none were intubated, only 20% were on vasopressors, and all were apparently able to perform a standardized breathing exercise. Forty-one (51%) patients were responders to volume expansion (i.e. a ≥ 10% stroke volume index increase). The cIVC was calculated during non-standardized (cIVC-ns) and standardized breathing (cIVC-st) conditions. The accuracy with which both cIVC-ns and cIVC-st predicted fluid responsiveness differed significantly by measurement site (interaction p < 0.001 and < 0.0001, respectively). Measuring inferior vena cava diameters 4 cm caudal to the right atrium predicted fluid responsiveness with the best accuracy. At this site, a standardized breathing manoeuvre also significantly improved predictive power: areas under ROC curves [mean and (95% CI)] for cIVC-ns = 0.85 [0.78–0.94] versus cIVC-st = 0.98 [0.97–1.0], p < 0.001. When cIVC-ns is superior or equal to 33%, fluid responsiveness is predicted with a sensitivity of 66% and a specificity of 92%. When cIVC-st is superior or equal to 44%, fluid responsiveness is predicted with a sensitivity of 93% and a specificity of 98%.
Conclusion
The accuracy with which cIVC measurements predict fluid responsiveness in spontaneously breathing patients depends on both the measurement site of inferior vena cava diameters and the breathing regime. Measuring inferior vena cava diameters during a standardized inhalation manoeuvre at 4 cm caudal to the right atrium seems to be the method by which to obtain cIVC measurements best-able to predict patients’ response to volume expansion.
Keywords: Inferior vena cava, Fluids, Fluid responsiveness, Hemodynamic, Sepsis, Severe infection, Ultrasound, Echocardiography, Collapsibility index, Spontaneous breathing
Background
Correcting acute reductions in blood volume through volume expansion (VE) is essential, but excessive VE can increase morbidity and mortality in critically ill patients [1–5]. Despite the importance of accurately predicting the response to VE in cases of acute circulatory failure, its clinical assessment is erroneous in 50% of cases [6–10]. Dynamic parameters have been proposed to predict fluid responsiveness (FR) in critically ill patients [11–13]. Initially developed for mechanically ventilated patients, use of these parameters has been proposed in spontaneously breathing (SB) patients in an effort to avoid unnecessary fluid exposure [14, 15]. Unfortunately, most of these dynamic parameters are unsuitable for SB patients. In fact, under spontaneous breathing, respiratory changes in the stroke volume index (SVI) or their estimates (e.g. respiratory changes in arterial pulse pressure) are either inaccurate as predictors of FR [16], or too complex for routine clinical use [17, 18]. Conversely, passive leg raising-induced change in SVI is an accurate predictor, which has been validated in SB patients [19, 20]; but it requires a tool to measure (or estimate) cardiac output or stroke volume, and may be either technically impossible or unreliable under specific conditions (e.g. insufficient increase in central venous pressure during the procedure [21], pregnancy [22] or intra-abdominal hypertension [23, 24]).
Measuring the diameter of the inferior vena cava (IVC) by ultrasonography is a fast, non-invasive tool for assessment of the blood volumetric status of critically ill patients in the early stages of care [25–27]. In SB patients, and in the absence of a standardized respiratory manoeuvre during ultrasound recordings, IVC collapsibility (cIVC) predicts FR with high specificity but low sensitivity [28–33]. Reversible, non-invasive procedures that magnify the change in intrathoracic and transpulmonary pressure (e.g. a deep inhalation or Valsalva manoeuvre) can improve the accuracy of some parameters to predict FR [17, 18]. In two recent studies, Preau et al. and Bortolotti et al. showed that a standardized, deep inhalation improved the sensitivity of cIVC to predict FR without altering its specificity [31, 32]. The use of cIVC to predict FR is scarce, but has been described in current practice [25–27] and controlled, trial protocols [34, 35]. Different measurement sites have been proposed to measure cIVC, ranging from 0.5 to 4 cm caudal to the IVC–right atrium junction [28–32]. It has been demonstrated that the collapsibility and distensibility of the IVC are both affected by the segment of vein considered [36, 37]. However, a rigorous comparison of the accuracy with which different sites for measuring IVC diameter can predict FR has yet to be published. This study aimed to analyse the effect of the measurement site, and the use of a concomitant, standardized breathing manoeuvre, on the accuracy with which cIVC measurements predict responsiveness to VE in SB patients with sepsis.
Methods
Study design
The present study was conducted post hoc on the ultrasound analyses of all patients included at the Lille University Hospital from two prospective cohorts [31, 32]. The initial objective of these cohorts was to investigate whether the cIVC, measured without (cIVC-ns) or with a standardized inhalation (cIVC-st), predicts FR in septic patients. The first study cohort focused on consecutive patients with regular sinus rhythm and included 90 patients from November 2011 to January 2014 [31]. The second, of identical design, looked at 55 consecutive patients who presented with cardiac arrhythmia (atrial fibrillation or more than 6 extrasystoles per minute) for which the inclusions ran from May 2012 to May 2015 [32].
Patients
Inclusion criteria were as follows: SB, adult patients admitted with sepsis in five intensive care units of the Lille University Hospital [38] with one or more clinical signs of acute circulatory failure for which a physician has prescribed VE. Clinical signs of acute circulatory failure were defined as systemic arterial hypotension (systolic arterial pressure < 90 mm Hg, or a decrease > 40 mm Hg in previously hypertensive patients), oliguria (urine output < 0.5 mL/kg/h over 1 h or more), tachycardia (heart rate > 100 per min), or mottled skin. Exclusion criteria included high-grade aortic insufficiency, transthoracic echogenicity unsuitable for measuring the velocity–time integral of aortic blood flow or IVC diameter, clinical signs of active exhalation, clinical or ultrasonographic evidence of pulmonary oedema due to heart failure [39], pregnancy, or abdominal compartment syndrome [40]. Patients for whom we could not measure IVC diameters in the first 4 cm were also excluded.
Study protocol
Ultrasonographic and clinical data were recorded prospectively immediately before and after VE, which consisted of a 30-min infusion of 500 mL 4% gelatin (Gelofusine® 4%, B. Braun, Melsungen, Germany or Plasmion®, Fresenius-Kabi, Louviers, France). Ultrasound examination was performed in a semi-recumbent position (30–45°) using commercially available Vivid-i and Vivid-S5 (General Electric, Solingen, Germany) with a phased array, low-frequency (2 MHz) transducer, operated by trained physicians (Level 2–3) [41]. The velocity–time integral of aortic blood flow was measured by pulsed-wave Doppler on a five-chamber apical view during unstandardized, spontaneous breathing [42]. The IVC bi-dimensional records were generated using the subcostal long-axis view. Particular attention was made to obtain a satisfactorily long axis section of the vessel at its centre in the sagittal plane. The ultrasound recordings of IVC were carried out over 3 consecutive breathing cycles (without and with a standardized breathing manoeuvre), both before and immediately after VE. The standardized breathing manoeuvre consisted of a brief (< 5 s) and continuous inhalation effort. Patients were verbally coached to generate a minimum buccal pressure from − 5 to − 10 mm H2O without any respiratory resistor. Oral cavity pressures were recorded with commercially available MP101 micromanometers (KIMO Instrument, Montpon, France), connected via a plastic sampling line to an antibacterial filter in series with a S183 mouth end piece (Teleflex Medical, Int’Air medical, Bourg-en-Bresse, France), as previously described [31, 32].
The anonymized recordings were analysed remotely. The SVI and general echocardiographic measurements were performed by echocardiography-trained, intensive care staff who were blind to both the clinical data and IVC measurements. To calculate the SVI, the left ventricular outflow tract (LVOT) velocity–time integral (VTI) was averaged over 15 consecutive cardiac cycles. The SVI equalled the product of the LVOT VTI multiplied by the ratio of the aortic valve area to the body surface area [42]. The aortic valve area was calculated once from the average aortic annulus diameter over three measurements. The body surface area calculated once according to the DuBois Method [43], was considered constant. Relative changes in SVI induced by VE were expressed as a percentage: VE-induced change in SVI (%) = 100 x ((post-VE value−baseline value) / baseline value). Patients were considered to have responded to treatment if VE induced a change in SVI of > 10% (i.e. VE-induced > 10% change in the LVOT VTI). This method is accepted as the reference method to assess the response to VE in SB patients [44, 45]. Moreover, this threshold value appeared to be clinically relevant (i.e. in terms of VE-induced changes in systolic arterial pressure and pulse pressure) and was also at least twice the intra-observer variability in the LVOT VTI measured in our previous studies [29, 30]. The IVC measurements were carried out using EchoPAC PC Software (General Electric Healthcare, Chicago, USA) by different operators, blind to the SVI and general echocardiographic measurements. The measurements of IVC diameters on both inhalation and exhalation were carried out at 5 separate sites: the cavo-atrial junction (site 0), then at 1, 3, 4 and 5 cm (sites 1, 3, 4 and 5, respectively) caudal to this junction, under both unstandardized and standardized breathing conditions (Fig. 1). Measurements of all IVC diameters were realized consecutively from site 0 to 5 for a given patient. The site at 2 cm was excluded a priori in view of its poor reproducibility owing to the entry at this site of the hepatic vein. Diameters were measured perpendicular to IVC walls from the trailing edge to the leading edge of the anterior and posterior wall, respectively. Distances between anatomical levels of the IVC and the cavo-atrial junction were measured parallel to IVC walls. The mean of three measurements was used to calculate the IVC collapsibility index (cIVC) as: ((maximum expiratory diameter—minimum inspiratory diameter) × 100/expiratory diameter).
Statistical analysis
Categorical variables are expressed as counts and percentages. Quantitative variables are reported as medians with interquartile ranges (25th; 75th centiles). Normality of distributions was assessed using histograms and the Shapiro–Wilk test. The use of norepinephrine was compared between responders and non-responders with the Chi-square test. Comparison of the cIVC according to FR and the measurement site (without and with a standardized breathing manoeuvre) were assessed by linear mixed models including FR, measurement site, interaction between FR and measurement site as fixed effects, and the patient as a random effect (in order to take repeated measurements into account). Normality of residuals was checked by plotting QQ-plot of conditional studentized residuals. When the interaction between FR and measurement site was significant, the cIVC according to FR was compared by measurement site using linear contrast applying Bonferroni correction. For each measurement site where a significant result was found, the areas under the receiver operating characteristic curves (ROC) and their 95% confidence intervals (CIs) were calculated. The optimal threshold values were determined by maximizing the Youden Index. Sensitivity and specificity for these values were calculated with their 95% CIs. We also report the threshold values to reach a sensitivity or specificity ≥ 0.90. The correlation between the cIVC and the variation of SVI after VE was assessed with the Spearman’s correlation coefficient. Statistical testing was performed at the two-tailed α level of 0.05. Data were analysed using the SAS software package, release 9.4 (SAS Institute, Cary, NC).
Results
Among the 145 patients from the two prospective cohorts [31, 32], 95 were included at the Lille University Hospital. Among them, 14 were excluded: 2 for unusable corrupted files and 12 for recordings that did not permit all IVC measurements to be taken from site 1 to 4. Eventually, 81 patients with a median age of 64 years old (53; 74) were included. The general characteristics of studied patients are summarized in Table 1 [46].
Table 1.
General data | Total population | Non-responders | Responders |
---|---|---|---|
(n = 81) | (n = 40) | (n = 41) | |
Age (year) | 64 (54; 73) | 63 (51; 76) | 64 (55; 72) |
Sex ratio female/male | 27/54 | 15/25 | 12/29 |
Height (m) | 1.71 (1.65; 1.76) | 1.69 (1.63; 1.75) | 1.72 (1.65; 1.80) |
Weight (kg) | 73 (60; 89) | 70 (60; 85) | 77.5 (60; 90) |
SAPS II | 34 (24; 42) | 37 (28; 48) | 31 (22; 39) |
Pre-existing conditions | |||
Arterial hypertension | 31 (38) | 17 (43) | 14 (34) |
Chronic left ventricular failure | 8 (10) | 4 (10) | 4 (10) |
Chronic right ventricular failure | 3 (4) | 2 (5) | 1 (2) |
Atrial fibrillation | 18 (22) | 8 (20) | 10 (24) |
COPD | 17 (21) | 9 (23) | 8 (20) |
Chronic pulmonary hypertension | 6 (7) | 5 (13) | 1 (2) |
Pulmonary embolism | 2 (3) | 0 (0) | 2 (5) |
Arteriopathy | 9 (11) | 7 (18) | 2 (5) |
Sepsis | |||
Pulmonary | 49 (61) | 24 (60) | 25 (61) |
Skin and soft tissue | 13 (16) | 7 (18) | 6 (15) |
Abdominal | 10 (12) | 5 (13) | 5 (12) |
Urinary | 6 (7) | 3 (8) | 3 (8) |
Other infections | 3 (4) | 1 (3) | 2 (5) |
Hemodynamic parameters | |||
Mean arterial pressure (mm Hg) | 68 (61; 79) | 67(61; 77) | 70 (61; 83) |
Heart rate (beats/min) | 107 (88; 120) | 98 (83; 114) | 109 (102; 126) |
LVOT VTI (cm) | 16.4 (13.2; 19.7) | 17.4 (15.2; 21) | 14.8 (12; 18.4) |
Cardiac index (L/min/m2) | 2.9 (2.3; 3.7) | 2.9 (2.5; 4.1) | 2.9 (2.2; 3.3) |
Norepinephrine | 16 (20) | 13 (33) | 3 (7) |
Norepinephrine µg/kg/min | 0.23 (0.17; 0.36) | 0.25 (0.14; 0.35) | 0.22 (0.19; 0.37) |
VE in last 24 h (mL) | 1000 (0; 2000) | 1000 (0; 2000) | 1000 (30; 1700) |
Oliguria | 39 (48) | 18 (45) | 21 (51) |
Arterial hypotension | 38 (47) | 19 (48) | 19 (46) |
Tachycardia | 57 (70) | 24 (60) | 33 (80) |
Mottled skin | 23 (28) | 15 (38) | 8 (20) |
Values are expressed as counts (%) or medians and interquartile ranges (25th; 75th centiles). COPD chronic obstructive pulmonary disease, SAPS Simplify Acute Physiology Score, LVOT VTI left ventricular outflow tract velocity time integral, VE volume expansion
Among the population, 37 patients (45%) had cardiac arrhythmias, and the median left ventricular ejection fraction before VE was 62% (57; 67). The median inspiratory depression generated without a standardized breathing manoeuvre was lower than when patients were verbally encouraged to perform a standardized inhalation (− 0.5 mm H2O (− 1; 0) and -6 mm H2O (− 7; − 5), p < 0.0001). Intra-patient variability regarding unstandardized and standardized inspiratory depression was 0.4 mm H2O (0; 0.4) and 1.8 mm H2O (1.3; 2.5), respectively. The time needed to perform both IVC recordings (online) and cIVC measurements (blinded, offline) was not assessed in the present study. Intra- and inter-observer variability with regards to the velocity–time integral of aortic blood flow was 3.5% (1.6; 5.4) and 7.9% (5.4; 10.3), respectively. Intra- and interobserver cIVC-st variability was 8.6% (4.4; 12.9) and 9.7% (2.7;17.3), respectively.
For the whole group, VE increased the SVI by 10% (1; 27). Distribution of VE increased in SVI (i.e. VE-induced change in the LVOT VTI) was not normal (p = 0 < 0.001), and patients were divided as follows: 14 with a VE-increase in SVI between -20 and -0.1%, 36 between 0 and 19.9%, 17 between 20 and 39.9%, and 14 between 40 and 59.9%. Forty-one patients (51%) were responders to VE. The VE-induced increase in SVI was 1% (-2; 4) in non-responders and 27% (20; 42) in responders. Twenty-six (65%) non-responders had a VE-related change in LVOT VTI between 0 and 10%, and 27 (66%) responders had a VE-related change in LVOT VTI between 10 and 30%. The general characteristics of the non-responder and responder patients are summarized in Table 1. Non-responders were more likely to receive norepinephrine than responders (33% versus 7%, p = 0.004). The cIVC for each measurement site is presented in Table 2 for non-responders and responders. The relationship between cIVC and FR differed significantly depending on the measurement site in both unstandardized (interaction p = 0.0005) and standardized (interaction p < 0.0001) breathing conditions. For cIVC-ns, no difference was found between responders and non-responders for site 0 (p = 0.086), but a significant difference was found for all other sites (p < 0.0001 for all). With a standardized inhalation manoeuvre, no difference in cIVC-st was found between responders and non-responders at the IVC–right atrium junction (site 0) (p = 0.89), but there was a significant difference at all the other measurement sites: 1 cm (p = 0.0010) and 3, 4 and 5 cm (p < 0.0001). For each site where a difference was found between responders and non-responders, the areas under the ROC curves were calculated (Fig. 2), and optimal thresholds to predict FR are presented in Table 3. Thresholds for a sensitivity or a specificity of 0.9 are given for measurement sites 1, 3, 4 and 5 in Additional file 1. Individual values of cIVC-ns and cIVC-st measured at 4 cm caudal to the right atrium are given in Fig. 3. The accuracy with which cIVC-st measurements at this site predicted FR was significantly better than cIVC-ns, with respective area under ROC curve values of 0.98 [0.97; 1] and 0.85 [0.78; 0.94] (p = 0.0008).
Table 2.
Breathing | Measurement | Non-responders | Responders |
---|---|---|---|
Condition | Site | (n = 40) | (n = 41) |
cIVC-ns (%) | 0 | 19 (8; 28) | 35 (20; 44) |
1 | 22(10; 34) | 41 (24; 53) | |
3 | 11 (3; 18) | 36 (20; 67) | |
4 | 12 (5; 24) | 45 (25; 58) | |
5* | 7 (2; 18) | 29 (17; 56) | |
cIVC-st (%) | 0 | 32 (15; 47) | 41 (28; 51) |
1 | 34 (19; 52) | 54 (42; 65) | |
3 | 26 (11; 42) | 61 (49; 82) | |
4 | 13 (8; 31) | 61 (52; 79) | |
5* | 14 (3; 32) | 55 (32; 72) |
Values are expressed as counts (%) or medians and interquartile ranges (25th; 75th centiles) Measurements of inferior vena cava (IVC) diameters were carried out at five sites: at the IVC-atrium junction (site 0), then at 1 (site 1), 3 (site 3), 4 (site 4) and 5 cm (site 5) caudal to the IVC-atrium junction. cIVC-ns collapsibility index of the IVC in non-standardized breathing conditions, cIVC-st collapsibility index of IVC with standardized breathing manoeuvre. *Measurement site 5 was performed in only 37 non-responders and 39 responders
Table 3.
Breathing condition | Site | ROC [95% CI] | Optimal threshold (%) | Sensitivity | Specificity |
---|---|---|---|---|---|
cIVC-ns | 1 | 0.76 [0.66; 0.86] | > 35* | 0.68 | 0.78 |
3 | 0.82 [0.74; 0.91] | > 20* | 0.76 | 0.78 | |
4 | 0.85 [0.78; 0.94] | > 33* | 0.66 | 0.92 | |
5 | 0.78 [0.68; 0.89] | > 19* | 0.74 | 0.76 | |
cIVC-st | 1 | 0.73 [0.62; 0.84] | > 42* | 0.76 | 0.66 |
3 | 0.80 [0.71; 0.91] | > 49* | 0.76 | 0.85 | |
4 | 0.98 [0.97; 1] | > 44* | 0.93 | 0.98 | |
5 | 0.83 [0.74; 0.92] | > 25* | 0.87 | 0.67 |
Measurements of inferior vena cava (IVC) diameters were carried out at five sites: at the IVC-atrium, then at 1 (site 1), 3 (site 3), 4 (site 4) and 5 cm (site 5) caudal to the IVC-atrium junction. ROC area under the curve of receiver operating characteristics, CI confidence interval, cIVC-ns collapsibility index of the IVC in non-standardized breathing conditions, cIVC-st collapsibility index of IVC with standardized breathing manoeuvre. *Optimal threshold value to predict response to volume expansion
There was a linear correlation between cIVC-st before VE and the VE-induced change in SVI (r = 0.82; p < 0.0001), and between cIVC-ns before VE and the VE-induced change in SVI (r = 0.62; p < 0.0001), at site 4 (Additional file 2).
Discussion
The present study is the first, to our knowledge, to demonstrate that the relationship between cIVC and FR is significantly affected by the measurement site of IVC diameter. Our results confirm that clinicians should avoid measuring cIVC at site 0/1 [36], and suggests that the most reliable measurement site for predicting the cardiac response to VE is at 4 cm caudal (upstream) from the cavo-atrial junction. Moreover, this study is the first to demonstrate a significant impact of a standardized breathing manoeuvre on the accuracy with which cIVC predicts FR. In practical terms, in SB patients, a reduction in IVC diameter of more than 33% during a non-standardized inspiration, measured at 4 cm caudal to the cavo-atrial junction, predicts FR with low sensitivity but great specificity. Thus, in cases of low cIVC values during non-standardized inhalation (cIVC-ns), the response to VE is uncertain and use of a standardized breathing manoeuvre (cIVC-st) is to be recommended. In these cases, a cIVC-st of > 44% predicted FR with both high sensitivity and specificity when IVC diameters were measured 4 cm caudal to the right atrium.
Our study demonstrates that when IVC diameters are measured too close to the right atrium (i.e. ≤ 1 cm), cIVC predicts FR with only low specificity. This phenomenon was accentuated by a deep inhalation breathing manoeuvre. This may be because the diaphragm’s movement generated false-positives in proximal measurement sites, as suggested by Gignon et al. in a physiological study with healthy volunteers [47]. Nevertheless, this motion appears to have less impact at more distal measurement sites. Bortolotti et al. demonstrated that a deep inhalation did not impact cIVC in non-responders to VE when IVC diameters were measured beyond 3 cm caudal to the right atrium [32]. Most of the studies that analysed the accuracy of cIVC-ns to predict FR in SB patients assessed IVC diameters approximately 2–4 cm caudal to the right atrium [30–32]. Accordingly, these studies demonstrated that cIVC-ns predicted FR with good specificity, ranging from 88 to 97%. Moreover, when cIVC was measured between 3 and 4 cm caudal to the IVC–right atrium junction, a deep inhalation manoeuvre did not alter this specificity [31, 32]. Nonetheless, Lanspa et al. managed to predict FR with a specificity of 89% while measuring cIVC during gentle, spontaneous breathing conditions at distances of 0.5–3 cm from the right atrium [29]. The specificity of cIVC in predicting FR may have been preserved in their study by the calmness of the inhalation manoeuvre, which reduced the influence of diaphragm movement on the IVC diameter. To prevent poor specificity, therefore, measurements of cIVC ≤ 1 cm caudal to the right atrium should be avoided for the prediction of FR in SB patients, particularly when a deep inhalation manoeuvre is used. Our results confirm the conclusions from Wallace et al. that clinicians should avoid measuring cIVC at site 0/1 [36].
Our study nevertheless shows that cIVC-ns predicts FR with low sensitivity at all measurement sites of IVC diameter between 0 and 5 cm caudal to the right atrium. Moreover, this study is the first to demonstrate that a simple, standardized, deep inhalation manoeuvre significantly improves the sensitivity of cIVC measurements to predict FR, in particular when IVC diameters are measured beyond 3 cm caudal to the right atrium junction. The fact that cIVC-st gradually increases upstream from the junction of the vessel with the right atrium is counterintuitive physiologically. Indeed, the downstream segments of the IVC should be the most influenced by negative thoracic pressures during inspiration. The IVC can be characterized as a highly compliant, collapsible tube and variations in its diameter during the respiratory cycle can be explained by transmural pressure gradient variations [48, 49]. It is a large, elliptical, intra-abdominal vessel which becomes intra-thoracic in its last portion where it traverses the diaphragm through a fibrous anatomical orifice [50]. This anatomical feature was one of the impetuses for our study, because we suspected that the diaphragm may lower the sensitivity of cIVC in SB patients; the attachment of the diaphragm and the fibrous portion around the vessel may, we hypothesized, result in decreased collapsibility of its downstream segments. Accordingly, Wallace et al. showed, in spontaneously breathing volunteers, that cIVC was significantly lower when recorded closer to the right atrium (cIVC = 20%) than when recorded approximately 4 cm caudal to it (cIVC = 30%, p = 0.03), or at the left renal vein (cIVC = 30%, p = 0.002—approximately 5 cm caudal to the right atrium) [36]. Therefore, measurements ≤ 3 cm caudal to the right atrium should be avoided to prevent poor sensitivity of the prediction of FR in SB patients, even if a standardized breathing manoeuvre is employed.
Overall, our results support the clinical use of cIVC measurements in the sub-costal, long-axis view in the two-dimensional mode, perpendicular to the IVC wall at 4 cm caudal to the right atrium. The rigorous application of this technique guarantees the accuracy of IVC diameter measurements during the respiratory cycle and, may reduce the inter- and intra-observer variability of cIVC as demonstrated by Finnerty et al.[51].
Our study has several limitations. Firstly, this study was performed in a population of adult patients with a median body weight and height of 73 kg (60; 89) and 1.71 m (1.65; 1.76), respectively. Thus, the present results cannot necessarily be generalized to infants or adult patients of extreme size. Secondly, while the post hoc analysis design of two prospective cohorts does not permit us to determine the proportion for whom our preferred measurement site is exploitable, previous studies have demonstrated that IVC diameters could not be assessed in this region in 7% to 15% of SB patients with sepsis [31, 32]. Moreover, we cannot exclude that the admittedly impressive accuracy of cIVC-st (measured 4 cm caudal to the cavo-atrial junction) to predict FR might reflect a selection bias due to the post hoc and monocentric design of the study. Thirdly, IVC measurements were performed without systematic measurement of intra-abdominal pressure. We acknowledge that elevated intra-abdominal pressure may influence IVC diameters, and thus, may alter cIVC accuracy to predict FR [52]. Nevertheless, despite a possible improvement of cIVC diagnostic accuracy, a systematic measurement of bladder pressure would also increase recording complexity and alter measurement feasibility. The influence of intra-abdominal pressure on cIVC accuracy in predicting FR in SB patients has yet to be established. Fourthly, VE duration may have lasted up to 30-min in some patients. Thus, hemodynamic effects may be less marked or have waned at the time of the second echocardiography assessment in the corresponding patients. Eventually, the assessment of our dynamic variables was performed in a select population, since patients were only enrolled with infection-related acute circulatory failure, under none or only a low-dose of norepinephrine, and for whom VE has already been prescribed by the physician in charge. In the present study, non-responders were more likely to be on norepinephrine than responders, despite having received the same amount of fluid. As demonstrated before, the addition of vasopressors, through veno-constriction, increases the stressed volume and may be responsible for a lower prevalence of FR [53]. In addition, patients with active exhalation, abdominal compartment syndrome, pregnancy, obesity or abdominal surgery, all of which could interfere with cIVC accuracy, were excluded from the study. Overall, patients were not severely ill: none were intubated, only 20% were on vasopressors, and all were apparently able to perform a standardized breathing exercise. Overall, patients of the present study were not severely ill, and the results will be more applicable to physicians in emergency or intermediate care departments than in intensive care units. Therefore, our results require confirmation in an unselected, critically ill population.
Conclusions
This rigorous comparison of different measurement sites suggests that the interaction between cIVC and FR is dependent on the way in which IVC diameter is measured: the best diagnostic accuracy for predicting FR is at 4 cm caudal to the cavo-atrial junction. We have also shown that a standardized, short (< 5 s) deep-inhalation manoeuvre improves the accuracy with which cIVC measurements predict patients’ response to VE. The cIVC thus measured is a fast, non-invasive and easy-to-implement tool which could improve clinical management of the acute phase of sepsis in SB patients. Its more widespread implementation in other critical conditions nevertheless requires a similarly rigorous, prospective study.
Supplementary Information
Acknowledgements
None.
Abbreviations
- CI
Confidence Interval
- cIVC
Collapsibility index of the Inferior Vena Cava
- cIVC-ns
Collapsibility index of the Inferior Vena Cava during non-standardized breathing conditions
- cIVC-st
Collapsibility index of the Inferior Vena Cava during a standardized breathing manoeuvre
- COPD
Chronic obstructive pulmonary disease
- FR
Fluid responsiveness
- IVC
Inferior Vena Cava
- LVOT
Left ventricular outflow tract
- SAPS
Simplified Acute Physiology Score
- SB
Spontaneously breathing
- SVI
Stroke Volume Index
- ROC
Receiver Operating Characteristic
- VE
Volume expansion
- VTI
Velocity–time integral
Authors’ contributions
All authors met authorship criteria and participated significantly to the study; in particular, MC, AD, PB, DC, SN, RF and SP involved in conception and design; PB, DC, AD, AR, TD, TO, RF and SP involved in patients’ inclusions; ED, MC and SP designed the statistical analysis; MC, PB, DC, JG, AR, TD and AD acquired data; SP, MH, ED and MC analysed and interpreted data; MC, MH, ED and SP wrote the article; and AR, SN, PB, DC, JG, TD, TO and RF provided critical revisions to the manuscript. All authors read and approved the final manuscript.
Funding
This study was supported by the academic hospital of Lille.
Availability of data and materials
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.
Ethics approval and consent to participate
The creation and analysis of the two cohorts was approved by the regional ethics committee of Nord-Pas-De-Calais, France (2011-A00990-41 and 2011-A01598-33). All subjects gave written, informed consent prior to enrolment.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Footnotes
Publisher's Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Contributor Information
Morgan Caplan, Email: morgan.caplan@chru-lille.fr.
Arthur Durand, Email: arthur.durand@chru-lille.fr.
Perrine Bortolotti, Email: perrine.bortolotti@gmail.com.
Delphine Colling, Email: delphine.colling@gmail.com.
Julien Goutay, Email: julien.goutay@chru-lille.fr.
Thibault Duburcq, Email: thibault.duburcq@chru-lille.fr.
Elodie Drumez, Email: elodie.drumez@chru-lille.fr.
Anahita Rouze, Email: anahita.rouze@chru-lille.fr.
Saad Nseir, Email: saad.nseir@chru-lille.fr.
Michael Howsam, Email: michael.howsam@univ-lille.fr.
Thierry Onimus, Email: thierry.onimus@chru-lille.fr.
Raphael Favory, Email: raphael.favory@chru-lille.fr.
Sebastien Preau, Email: seb.preau@gmail.com.
Supplementary Information
The online version contains supplementary material available at 10.1186/s13613-020-00786-1.
References
- 1.Boyd JH, Forbes J, Nakada T, Walley KR, Russell JA. Fluid resuscitation in septic shock: a positive fluid balance and elevated central venous pressure are associated with increased mortality. Crit Care Med. 2011;39(2):259–265. doi: 10.1097/CCM.0b013e3181feeb15. [DOI] [PubMed] [Google Scholar]
- 2.Sakr Y, Rubatto Birri PN, Kotfis K, et al. Higher fluid balance increases the risk of death from sepsis: results from a large international audit*. Crit Care Med. 2017;45(3):386–394. doi: 10.1097/CCM.0000000000002189. [DOI] [PubMed] [Google Scholar]
- 3.Vincent J-L, Sakr Y, Sprung CL, et al. Sepsis in European intensive care units: results of the SOAP study. Crit Care Med. 2006;34(2):344–353. doi: 10.1097/01.CCM.0000194725.48928.3A. [DOI] [PubMed] [Google Scholar]
- 4.Prowle JR, Echeverri JE, Ligabo EV, Ronco C, Bellomo R. Fluid balance and acute kidney injury. Nat Rev Nephrol. 2010;6(2):107–115. doi: 10.1038/nrneph.2009.213. [DOI] [PubMed] [Google Scholar]
- 5.Silversides JA, Major E, Ferguson AJ, et al. Conservative fluid management or deresuscitation for patients with sepsis or acute respiratory distress syndrome following the resuscitation phase of critical illness: a systematic review and meta-analysis. Intensive Care Med. 2017;43(2):155–170. doi: 10.1007/s00134-016-4573-3. [DOI] [PubMed] [Google Scholar]
- 6.Calvin JE, Driedger AA, Sibbald WJ. The hemodynamic effect of rapid fluid infusion in critically ill patients. Surgery. 1981;90(1):61–76. [PubMed] [Google Scholar]
- 7.Reuse C, Vincent JL, Pinsky MR. Measurements of right ventricular volumes during fluid challenge. Chest. 1990;98(6):1450–1454. doi: 10.1378/chest.98.6.1450. [DOI] [PubMed] [Google Scholar]
- 8.Magder S, Georgiadis G, Cheong T. Respiratory variations in right atrial pressure predict the response to fluid challenge. J Crit Care. 1992;7(2):76–85. doi: 10.1016/0883-9441(92)90032-3. [DOI] [Google Scholar]
- 9.Tavernier B, Makhotine O, Lebuffe G, Dupont J, Scherpereel P. Systolic pressure variation as a guide to fluid therapy in patients with sepsis-induced hypotension. Anesthesiology. 1998;89(6):1313–1321. doi: 10.1097/00000542-199812000-00007. [DOI] [PubMed] [Google Scholar]
- 10.Michard F, Boussat S, Chemla D, et al. Relation between respiratory changes in arterial pulse pressure and fluid responsiveness in septic patients with acute circulatory failure. Am J Respir Crit Care Med. 2000;162(1):134–138. doi: 10.1164/ajrccm.162.1.9903035. [DOI] [PubMed] [Google Scholar]
- 11.Michard F, Teboul J-L. Predicting fluid responsiveness in ICU patients: a critical analysis of the evidence. Chest. 2002;121(6):2000–2008. doi: 10.1378/chest.121.6.2000. [DOI] [PubMed] [Google Scholar]
- 12.Bentzer P, Griesdale DE, Boyd J, MacLean K, Sirounis D, Ayas NT. Will this hemodynamically unstable patient respond to a bolus of intravenous fluids? JAMA. 2016;316(12):1298–1309. doi: 10.1001/jama.2016.12310. [DOI] [PubMed] [Google Scholar]
- 13.Bednarczyk JM, Fridfinnson JA, Kumar A, et al. Incorporating dynamic assessment of fluid responsiveness into goal-directed therapy: a systematic review and meta-analysis. Crit Care Med. 2017;45(9):1538–1545. doi: 10.1097/CCM.0000000000002554. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 14.Coudray A, Romand J-A, Treggiari M, Bendjelid K. Fluid responsiveness in spontaneously breathing patients: a review of indexes used in intensive care. Crit Care Med. 2005;33(12):2757–2762. doi: 10.1097/01.CCM.0000189942.24113.65. [DOI] [PubMed] [Google Scholar]
- 15.de Chaves RCF, Corrêa TD, Neto AS, et al. Assessment of fluid responsiveness in spontaneously breathing patients: a systematic review of literature. Ann Intensive Care. 2018;8(1):21. doi: 10.1186/s13613-018-0365-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 16.Soubrier S, Saulnier F, Hubert H, et al. Can dynamic indicators help the prediction of fluid responsiveness in spontaneously breathing critically ill patients? Intensive Care Med. 2007;33(7):1117–1124. doi: 10.1007/s00134-007-0644-9. [DOI] [PubMed] [Google Scholar]
- 17.Préau S, Dewavrin F, Soland V, et al. Hemodynamic changes during a deep inspiration maneuver predict fluid responsiveness in spontaneously breathing patients. Cardiol Res Pract. 2012;2012:191807. doi: 10.1155/2012/191807. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 18.Monge García MI, Gil Cano A, Díaz Monrové JC. Arterial pressure changes during the Valsalva maneuver to predict fluid responsiveness in spontaneously breathing patients. Intensive Care Med. 2009;35(1):77–84. doi: 10.1007/s00134-008-1295-1. [DOI] [PubMed] [Google Scholar]
- 19.Maizel J, Airapetian N, Lorne E, Tribouilloy C, Massy Z, Slama M. Diagnosis of central hypovolemia by using passive leg raising. Intensive Care Med. 2007;33(7):1133–1138. doi: 10.1007/s00134-007-0642-y. [DOI] [PubMed] [Google Scholar]
- 20.Préau S, Saulnier F, Dewavrin F, Durocher A, Chagnon J-L. Passive leg raising is predictive of fluid responsiveness in spontaneously breathing patients with severe sepsis or acute pancreatitis. Crit Care Med. 2010;38(3):819–825. doi: 10.1097/CCM.0b013e3181c8fe7a. [DOI] [PubMed] [Google Scholar]
- 21.Lakhal K, Ehrmann S, Runge I, et al. Central venous pressure measurements improve the accuracy of leg raising-induced change in pulse pressure to predict fluid responsiveness. Intensive Care Med. 2010;36(6):940–948. doi: 10.1007/s00134-010-1755-2. [DOI] [PubMed] [Google Scholar]
- 22.Marques NR, Martinello C, Kramer GC, et al. Passive leg raising during pregnancy. Am J Perinatol. 2015;32(4):393–398. doi: 10.1055/s-0034-1389089. [DOI] [PubMed] [Google Scholar]
- 23.Mahjoub Y, Touzeau J, Airapetian N, et al. The passive leg-raising maneuver cannot accurately predict fluid responsiveness in patients with intra-abdominal hypertension. Crit Care Med. 2010;38(9):1824–1829. doi: 10.1097/CCM.0b013e3181eb3c21. [DOI] [PubMed] [Google Scholar]
- 24.Beurton A, Teboul J-L, Girotto V, et al. Intra-abdominal hypertension is responsible for false negatives to the passive leg raising test. Crit Care Med. 2019;47(8):e639–e647. doi: 10.1097/CCM.0000000000003808. [DOI] [PubMed] [Google Scholar]
- 25.Boulain T, Boisrame-Helms J, Ehrmann S, et al. Volume expansion in the first 4 days of shock: a prospective multicentre study in 19 French intensive care units. Intensive Care Med. 2015;41(2):248–256. doi: 10.1007/s00134-014-3576-1. [DOI] [PubMed] [Google Scholar]
- 26.Cecconi M, Hofer C, Teboul J-L, et al. Fluid challenges in intensive care: the FENICE study: a global inception cohort study. Intensive Care Med. 2015;41(9):1529–1537. doi: 10.1007/s00134-015-3850-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 27.Preau S, Dewavrin F, Demaeght V, et al. The use of static and dynamic haemodynamic parameters before volume expansion: a prospective observational study in six French intensive care units. Anaesth Crit Care Pain Med. 2016;35(2):93–102. doi: 10.1016/j.accpm.2015.08.003. [DOI] [PubMed] [Google Scholar]
- 28.Muller L, Bobbia X, Toumi M, et al. Respiratory variations of inferior vena cava diameter to predict fluid responsiveness in spontaneously breathing patients with acute circulatory failure: need for a cautious use. Crit Care. 2012;16(5):R188. doi: 10.1186/cc11672. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 29.Lanspa MJ, Grissom CK, Hirshberg EL, Jones JP, Brown SM. Applying dynamic parameters to predict hemodynamic response to volume expansion in spontaneously breathing patients with septic shock. Shock Augusta Ga. 2013;39(2):155–160. doi: 10.1097/SHK.0b013e31827f1c6a. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 30.Airapetian N, Maizel J, Alyamani O, et al. Does inferior vena cava respiratory variability predict fluid responsiveness in spontaneously breathing patients? Crit Care Lond Engl. 2015;19:400. doi: 10.1186/s13054-015-1100-9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 31.Preau S, Bortolotti P, Colling D, et al. Diagnostic accuracy of the inferior vena cava collapsibility to predict fluid responsiveness in spontaneously breathing patients with sepsis and acute circulatory failure. Crit Care Med. 2017;45(3):e290–e297. doi: 10.1097/CCM.0000000000002090. [DOI] [PubMed] [Google Scholar]
- 32.Bortolotti P, Colling D, Colas V, et al. Respiratory changes of the inferior vena cava diameter predict fluid responsiveness in spontaneously breathing patients with cardiac arrhythmias. Ann Intensive Care. 2018;8(1):79. doi: 10.1186/s13613-018-0427-1. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 33.Bortolotti P, Colling D, Preau S. Inferior vena cava respiratory variations: a useful tool at bedside to guide fluid therapy in spontaneously breathing patients. Shock. 2018;49(2):235–236. doi: 10.1097/SHK.0000000000000950. [DOI] [PubMed] [Google Scholar]
- 34.Hernández G, Ospina-Tascón GA, Damiani LP, et al. Effect of a resuscitation strategy targeting peripheral perfusion status vs serum lactate levels on 28-day mortality among patients with septic shock: the ANDROMEDA-SHOCK randomized clinical trial. JAMA. 2019;321(7):654–664. doi: 10.1001/jama.2019.0071. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 35.Kattan E, Ospina-Tascón GA, Teboul J-L, et al. Systematic assessment of fluid responsiveness during early septic shock resuscitation: secondary analysis of the ANDROMEDA-SHOCK trial. Crit Care. 2020;24(1):23. doi: 10.1186/s13054-020-2732-y. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Wallace DJ, Allison M, Stone MB. Inferior vena cava percentage collapse during respiration is affected by the sampling location: an ultrasound study in healthy volunteers. Acad Emerg Med Off J Soc Acad Emerg Med. 2010;17(1):96–99. doi: 10.1111/j.1553-2712.2009.00627.x. [DOI] [PubMed] [Google Scholar]
- 37.Biasucci DG, Cina A, Calabrese M, et al. Size and shape of the inferior vena cava before and after a fluid challenge: a pilot study. Minerva Anestesiol. 2019;85(5):514–521. doi: 10.23736/S0375-9393.18.13041-0. [DOI] [PubMed] [Google Scholar]
- 38.Singer M, Deutschman CS, Seymour CW, et al. The Third International Consensus Definitions for Sepsis and Septic Shock (Sepsis-3) JAMA. 2016;315(8):801–810. doi: 10.1001/jama.2016.0287. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 39.Ponikowski P, Voors AA, Anker SD, et al. 2016 ESC Guidelines for the diagnosis and treatment of acute and chronic heart failureThe Task Force for the diagnosis and treatment of acute and chronic heart failure of the European Society of Cardiology (ESC)Developed with the special contribution of the Heart Failure Association (HFA) of the ESC. Eur Heart J. 2016;37(27):2129–2200. doi: 10.1093/eurheartj/ehw128. [DOI] [PubMed] [Google Scholar]
- 40.Maluso P, Olson J, Sarani B. Abdominal compartment hypertension and abdominal compartment syndrome. Crit Care Clin. 2016;32(2):213–222. doi: 10.1016/j.ccc.2015.12.001. [DOI] [PubMed] [Google Scholar]
- 41.Mayo PH, Beaulieu Y, Doelken P, et al. American College of Chest Physicians/La Société de Réanimation de Langue Française statement on competence in critical care ultrasonography. Chest. 2009;135(4):1050–1060. doi: 10.1378/chest.08-2305. [DOI] [PubMed] [Google Scholar]
- 42.Lewis JF, Kuo LC, Nelson JG, Limacher MC, Quinones MA. Pulsed Doppler echocardiographic determination of stroke volume and cardiac output: clinical validation of two new methods using the apical window. Circulation. 1984;70(3):425–431. doi: 10.1161/01.CIR.70.3.425. [DOI] [PubMed] [Google Scholar]
- 43.Bois DD, Bois EFD. Clinical calorimetry: Tenth paper a formula to estimate the approximative surface area if height and weight be known. Arch Intern Med. 1916;XVII(6–2):863–871. doi: 10.1001/archinte.1916.00080130010002. [DOI] [Google Scholar]
- 44.Cecconi M, De Backer D, Antonelli M, et al. Consensus on circulatory shock and hemodynamic monitoring. Task force of the European Society of Intensive Care Medicine. Intensive Care Med. 2014;40(12):1795–1815. doi: 10.1007/s00134-014-3525-z. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 45.Jozwiak M, Mercado P, Teboul J-L, et al. What is the lowest change in cardiac output that transthoracic echocardiography can detect? Crit Care Lond Engl. 2019;23(1):116. doi: 10.1186/s13054-019-2413-x. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 46.Gall J-RL, Lemeshow S, Saulnier F. A New Simplified Acute Physiology Score (SAPS II) Based on a European/North American Multicenter Study. JAMA. 1993;270(24):2957–2963. doi: 10.1001/jama.1993.03510240069035. [DOI] [PubMed] [Google Scholar]
- 47.Gignon L, Roger C, Bastide S, et al. Influence of diaphragmatic motion on inferior vena cava diameter respiratory variations in healthy volunteers. Anesthesiology. 2016;124(6):1338–1346. doi: 10.1097/ALN.0000000000001096. [DOI] [PubMed] [Google Scholar]
- 48.Takata M, Wise RA, Robotham JL. Effects of abdominal pressure on venous return: abdominal vascular zone conditions. J Appl (1985) 1990;69(6):1961–1972. doi: 10.1152/jappl.1990.69.6.1961. [DOI] [PubMed] [Google Scholar]
- 49.Minten J, Van de Werf F, Aubert AE, Kesteloot H, De Geest H. Influence of transmural pressure on retrograde pressure pulse transmission velocity in the canine superior vena cava. Arch Int Physiol Biochim. 1984;92(4):317–325. doi: 10.3109/13813458409071173. [DOI] [PubMed] [Google Scholar]
- 50.Shin DS, Sandstrom CK, Ingraham CR, Monroe EJ, Johnson GE. The inferior vena cava: a pictorial review of embryology, anatomy, pathology, and interventions. Abdom Radiol (NY) 2019;44(7):2511–2527. doi: 10.1007/s00261-019-01988-3. [DOI] [PubMed] [Google Scholar]
- 51.Finnerty NM, Panchal AR, Boulger C, et al. Inferior vena cava measurement with ultrasound: what is the best view and best mode? West J Emerg Med. 2017;18(3):496–501. doi: 10.5811/westjem.2016.12.32489. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 52.Viellard-Baron A, Evrard B, Repessé X, et al. Limited value of end-expiratory inferior vena cava diameter to predict fluid responsiveness impact of intra-abdominal pressure. Intensive Care Med. 2018;44:197–203. doi: 10.1007/s00134-018-5067-2. [DOI] [PubMed] [Google Scholar]
- 53.Hamzoui O, Georger JF, Monnet X, et al. Early administration of norepinephrine increases cardiac preload and cardiac output in septic patients with life-threatening hypotension. Crit Care. 2010;14(4):R142. doi: 10.1186/cc9207. [DOI] [PMC free article] [PubMed] [Google Scholar]
Associated Data
This section collects any data citations, data availability statements, or supplementary materials included in this article.
Supplementary Materials
Data Availability Statement
The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.