ABSTRACT
BACKGROUND:
Bronchoscopic lung volume reduction (BLVR) by either the endobronchial valve (EBV) or coil (EBC) procedure is recommended for severe emphysematous patients. BLVR applications generally help healthy lung areas ventilate more comfortably by reducing the hyperinflation and improving the contraction capacity of diaphragm.
OBJECTIVES:
Compare our experience with valve and coil BLVR devices.
DESIGN:
Retrospective.
SETTING:
Single tertiary care centre.
PATIENTS AND METHODS:
Demographic data, vital signs, pulmonary function tests (PFTs), the six-minute walking test (6MWT), vital signs, arterial blood gases and complications were recorded.
MAIN OUTCOME MEASURES:
Change in PFTs and completion of the 6MWT.
SAMPLE SIZE:
60 Turkish men with a diagnosis of chronic pulmonary lung disease.
RESULTS:
Clinical and demographic characteristics were similar in patients who underwent EBV and EBC. Thirty (96.8%) EBV patients and 27 (93.1%) of the EBC patients were able to properly complete the PFT before the procedures, but all complied after the procedures. Significant improvement in PFTs were achieved after the procedure and there were no statistically significant differences in post-procedure performance. For the 6MWT, the completion rate improved from 15 (48.4%) to 19 (61.3%) patients in the EBV patients (P=.125) and from 19 (65.5%) to 21 (72.4%) patients in the EBC patients (P=.500). There was no significant difference in completion rates for the walking test for either group (median 32 meters in EBV patients and 37 meters in EBC patients; P=.652). Vital signs and arterial blood gases were similar in the two groups. The rates of complications were similar in both groups.
CONCLUSION:
Endobronchial valves and coils are safe and effective methods for BLVR for patients with severe emphysema.
LIMITATIONS:
Relatively small sample, retrospective design, single-centre retrospective study.
CONFLICT OF INTEREST:
None.
INTRODUCTION
Chronic obstructive lung disease (COPD) is a disease with irreversible airway obstruction that results in relatively resistant pulmonary symptoms.1 Pathologically, alveolar wall damage causes irreversible airway obstruction, loss of elastic recoil and therefore, reduced gas exchange areas.1–3 The emphy sematous phenotype of COPD has a prevalence rate of 1.8% worldwide.2 For terminally emphysemic patients, clinical and life quality are bad.2,3 Emphysema may have a negative effect on lung functions, such as a decline in forced expiratory volume at first second (FEV1), increase in total lung capacity (TLC), functional residual capacity (FRC) and a decrease in carbon monoxide diffusion capacity (DLCO).4 Treatments to reduce mortality and symptoms for COPD are smoking cessation,5 long-term oxygen treatment, bronchodilators and inhalant steroids, pulmonary rehabilitation,6 bronchoscopic lung volume reduction (BLVR), lung volume reduction surgery (LVRS) and transplantation.7
Since 2002, BLRV interventions like the endobronchial valve (EBV) and coil (EBC) have been used to reduce hyperinflation, ameliorate respiratory mechanics, and reduce mortality and morbidity.8 Studies in the following years have shown that BLVR applications improve health-related quality of life, pulmonary functions and increase exercise capacity in emphysematous patients.1,9–12
EBV lung volume reduction provides similar benefits with LVRS but carries fewer risks.3,13,14 EBCs are devices that are placed bronchoscopically into the subsegmental airways. The coils are made up of shape memory nitinol (a nickel-titanium alloy) wire. Unlike the EBV, EBCs are also effectively used in patients with interlobar collateral-ventilation.15,16 Both methods have been effective and safe in short- and long-term clinical studies.17–21 BLVR treatment by either the coil or valve have been recommended by the Global Initiative for Chronic Obstructive Lung Diseases (GOLD).22 We aimed to evaluate and compare the results of EBV and EBC treatments in our patients with emphysema.
PATIENTS AND METHODS
We retrospectively collected all data on BVLR procedures performed on patients between 1 February 2015 and 1 October 2018 (3 years, 8 months). Informed consent was given by all patients after preoperative evaluation for general anaesthesia. All patients were evaluated with pulmonary function tests (PFTs), arterial blood gas analyses and the six-minute walking test (6MWT) before the procedures and then again after 6 months. Approval was obtained from the Ethics Committee of Afyon University of Health Sciences, dated 2019/11.
Inclusion criteria were age older than 40 years, obstructive pulmonary function test with FEV1 >15% and <50%, pulmonary arterial pressure (PAP) < 55 mm Hg, TLC% >100, PaCO2 <60 mm Hg, residual volume (RV) >175%, dyspnea score <2 (evaluated by modified Medical Research Council, mMRC), and the presence of heterogeneous emphysema in candidates for the EBC procedure. Exclusion criteria were the presence of malignant characterized pulmonary nodule, the presence of significant airway pathology, the presence of bronchiectasis in the same lobe, the ability to walk more than 350 meters in the 6MWT, FEV1 <15% and >50%, systolic PAP >55 mm Hg, PaCO2 >60 mm Hg, TLC% <100, RV% >175, dyspnea score <2 (evaluated by modified Medical Research Council, mMRC), previous diagnosis of cancer, the presence of homogeneous emphysema in candidates for the EBC procedure (in that case; we used the ECV procedure), the presence of collateral ventilation for the ECV procedure (when present, the ECV procedure was chosen), or the patient declined to undergo the procedure. Measurements (PFT, arterial blood gases and 6MWT) for all patients were repeated after six months.
Endobronchial valve placement
The EBV was placed into the bronchoscopically targeted segmental or subsegmental bronchi. A Chartis catheter was used for cases in which the fissure integrity was not clearly detected in computed tomography (CT) sections. Three to 5 valves were placed in patients who were determined to have no collateral circulation in the targeted lobe. In the presence of collateral circulation with a Chartis catheter, EBC was preferred over EBV.1,23 Each device is shown in Figure 1.
Figure 1. View of the endobronchial valve in the posterior segment of the right upper lobe (left). Endobronchial coils on chest radiograph (right).
Endobronchial coil placement
Several years after the development of EBV, the coil treatment was developed and applied to heterogeneous emphysematous patients with or without collateral ventilation.1 The coil was placed into the targeted lobe via fiberoptic bronchoscope through an intubation tube. With experience, we began placing coils into some patients who could tolerate bronchoscopy without intubation. Depending on the number of segments in the targeted lobe, 10 to 15 coils were planted into a patient with one lobe per session. In some cases, the coil was preferable to the valve or vice versa, and then the preferred procedure was used in a second session. Control fiberoptic bronchoscopy was performed for all EBV/EBC patients to confirm the position and function of the EBV/EBC procedures at the third and fourth weeks of the procedures. Both procedures took between 30 minutes to one hour and involved a 1-4 night stay in hospital.
Pulmonary rehabilitation program
Four weeks of a comprehensive pulmonary rehabilitation program were prescribed to all candidate patients in the hospital (2 sessions per week), prior to the procedures. After all the EBV/EBC procedures, 4 weeks of pulmonary rehabilitation were also prescribed for all patients.
Statistical analysis
Statistical evaluation was carried out with IBM SPSS for Windows Version 20.0 (IBM SPSS, Armonk, NY). The Kolmogorov-Smirnov test was used to evaluate the distribution of continuous variables. Categorical variables were expressed as numbers and percentages. Chi-square was used to compare group ratios. When the sample size was less than 5, we used Fisher's exact test to compare group ratios. The McNemar test was used to determine the differences between dependent variables (to compare the completion rate of 6MWT, before and after the procedures). We used the kappa coefficient to measure agreement on completion rates of the 6MWT. Continuous variables are expressed using the median (minimum-maximum) values if they were not normally distributed, and they were expressed using mean and standard deviation (SD) if they were normally distributed. The t test was used for all tests of continuous variables when the distribution was normal; otherwise the Mann-Whitney U test was used to compare the two groups. Statistical significance was set at P<.05.
RESULTS
The EBV procedure was applied to 31 patients (51.7%) and the coil to 29 (48.3%) patients (Table 1). Clinical and demographic characteristics of the two groups were similar. Thirty (96.8%) EBV patients and 27 (93.1%) EBC patients cooperated with PFT (P=.606). In the EBV patients, local anesthesia (oropharyngeal lidocaine) accompanied by midazolam sedation was applied in 24 of these patients. The other 7 underwent fiberoptic bronchoscopy through an intubation tube under general anaesthesia. In the EBC patients, local anesthesia was used initially for most patients receiving the coil treatment. There was an improvement in all PFT results in both groups from before to after the procedures and the differences between the groups after the procedure were not statistically significant (Table 2, Figure 2).
Table 1.
Clinical and demographic characteristics of the patients (n=60).
Characteristics | Endobronchial valve (n=31) | Endobronchial coil (n=29) | P value |
---|---|---|---|
Age (years) | 64.5 (9.2) | 68.0 (7.1) | .07 |
Smoking statusa | |||
Current smoker | 17 (54.8) | 13 (44.8) | .480 |
Ex-smoker | 13 (41.9) | 13 (44.8) | |
Non-smoker | 1 (3.2) | 3 (10.3) | |
Height (centimeters) | 166.0 (7.2) | 166.2 (6.6) | .923 |
Weight (kg) | 65.2 (15.2) | 68.9 (14.6) | .344 |
Body mass index (kg/m2) | 23.6 (4.9) | 24.8 (4.4) | .316 |
Data are number (%) or mean (standard deviation).
aFisher's exact test
Table 2.
Change in dyspnea score and pulmonary function tests for the two bronchoscopic volume reduction procedures after application (n=60).
Pulmonary function tests | Endobronchial valve (n=31) | Endobronchial coil (n=29) | P value |
---|---|---|---|
mMRC | −1.0 (−1.0-0) | −1.0 (−2.0-0) | .139 |
FEV1 (L) | 0.25 (−0.18 to 1.35) | 0.16 (−0.13 to 0.81) | .198 |
FEV1 (%) | 9.00 (−9.00 to 25.00) | 7.00 (−7.00 to 17.00) | .476 |
FVC (L) | 0.31 (−0.95 to 2.00) | 0.39 (−0.42 to 1.09) | .497 |
FVC (%) | 7.00 (−10.00 to 24.00) | 6.00 (−9.00 to 29.00) | .466 |
FEV1/FVC | 4.50 (−19.00 to 17.00) | 1.00 (−27.02 to 27.00) | .298 |
RV (mL) | −260.50 (−1295.00 to −45.00) | −245.00 (−490.00 to 62.00) | .813 |
Data are median (min-max) unless indicated otherwise. Statistical comparisons by the Mann-Whitney U test. FEV1: Forced expiratory volume at end of first second, FVC: Forced vital capacity, FEV1/FVC ratio=FEV1%, RV: Residual volume, mMRC: Modified Medical Research Council.
Figure 2. Pre- and post-procedure FEV1 (left) and RV (right) by valve or coil procedure (P<.01 and P<.001 for pre-to-post changes for FEV1 and RV, respectively) (post values are at 6 months after the procedure).
In both EBV and EBC patients, there was an increase in the median distance and duration in the 6MWT. In the comparison of the procedures, there were no statistically significant differences in the distance and duration for the 6MWT and vital sign parameters (blood pressure and oxygen saturation) (Table 3, Figure 3). For the 6MWT, the completion rate improved from 15 (48.4%) to 19 (61.3%) patients in the 6MWT in the EBV patients (P=.125) and from 19 (65.5%) to 21 (72.4%) patients in the EBC patients (P<.500). There were no significant differences in completion rates for either study group (P=.652). For the EBV the kappa coefficient was 0.744, indicating good agreement between observers (P<.001). For the EBC, the kappa coefficient was 0.840, indicating very good agreement (P<.001).
Table 3.
Changes in walking test and other clinical measurements of the effects of the two bronchoscopic volume reduction procedures before and after application (n=60).
Endobronchial valve (n=31) | Endobronchial coil (n=29) | P value | |
---|---|---|---|
Walking test | |||
Distance (meters) | 32.00 (11.00 to 74.00) | 37.00 (−44.00 to 75.00) | .652 |
Duration (seconds) | 11.00 (0.00 to 115.00) | 0.00 (0.00 to 90.00) | .128 |
Pulse rates (bpm) | |||
Before | 14.00 (−2.00 to 28.00) | 8.00 (−27.00 to 23.00) | .508 |
After | 8.00 (−1.00 to 28.00) | 8.00 (−54.00 to 25.00) | .075 |
Blood pressure | |||
Systolic | 0.00 (−20.00 to 10.00) | 0.00 (−20.00 to 15.00) | .108 |
Diastolic | 5.00 (−5.00 to 20.00) | 5.00 (−10.00 to 10.00) | .328 |
Oxygen saturation, % | |||
Before | −7.00 (−36.00 to 2.00) | −7.00 (−30.00 to 0.00) | .241 |
After | −4.00 (−19.00 to 1.00) | −3.00 (−19.00 to −1.00) | .945 |
Data are median (min-max) unless indicated otherwise. Statistical comparisons by the Mann-Whitney U test.
Figure 3. Changes in the six-minute walking test for each patient for the two endobronchial volume reduction procedures after application (P=.652).
In the EBV patients, SBP changed from 119.4 (13.6) to 117.9 (8.3) mm Hg from before to after the procedure (P=.248). DBP changed from 82.4 (9.5) to 85.2 (8.1) from before to after. In the EBC patients, SBP changed from 113.6 (17.9) to 114.3 (10.4) mm Hg from before to after the procedure (P=.841). Changes in arterial blood gases are shown in Table 4.
Table 4.
Changes in arterial blood gases before and after application of the two endobronchial volume reduction procedures.
Endobronchial valve | P value | Endobronchial coil | P value | |||
---|---|---|---|---|---|---|
Before | After | Before | After | |||
pH | 7.40 (0.04) | 7.39 (0.03) | .182 | 7.41 (0.04) | 7.39 (0.03) | .014 |
PO2 | 53.53 (4.91) | 60.51 (4.36) | <.001 | 54.24 (6.93) | 82.49 (12.98) | .182 |
PCO2 | 39.03 (5.25) | 39.22 (3.12) | .756 | 41.38 (6.12) | 41.35 (4.39) | .967 |
Local anesthesia was applied to 24 patients (77.4%) EBV patients and general anesthesia was applied to 7 patients (22.6%). For EBC patients, 26 (89.7%) were under local anesthesia and 3 (10.4%) were under general anesthesia (P=.204). Anatomical localisations for EBV and EBC applied subjects are given in Table 5. The most frequent complications seen after EBV were pneumothorax in 3 patients (9.7%) and cardiac arrhythmia in 2 patients (6.5%). After EBC, 3 patients (10.3%) had pneumonia and 2 (6.9%) patients had COPD exacerbations. All other complications shown in Table 6.
Table 5.
Anatomical localization for the two bronchoscopic volume reduction procedures (n=60).
Location of the applied procedure | Endobronchial valve (n=31) | Endobronchial coil (n=29) | P value |
---|---|---|---|
Left | |||
Upper lobe | 8 (80.0) | 2 (20.0) | .082 |
Lower lobe | 3 (42.9) | 4 (57.1) | .702 |
Right | |||
Upper lobe | 18 (45.0) | 22 (55.0) | .177 |
Lower lobe | 2 (66.7) | 1 (33.3) | .999 |
Data are n (%). Statistical comparisons by Fisher's exact test.
Table 6.
Complications following endobronchial volume reduction procedures.
Total | Endobronchial valve (n=31) | Endobronchial coil (n=29) | P value | |
---|---|---|---|---|
Cardiac arrhythmia | 2 (3.3) | 2 (6.5 | 0 | .492 |
Hypernatremia | 1 (1.7) | 1 (3.2) | 0 | .999 |
Pneumonia | 4 (6.7) | 1 (3.2) | 3 (10.3) | .346 |
Pneumothorax | 4 (6.7) | 3 (9.7) | 1 (3.4) | .613 |
Hemoptysis | 1 (1.7) | 1 (3.2) | 0 (0.0) | .999 |
Expectoration | 1 (1.7) | 1 (3.2) | 0 (0.0) | .999 |
Exacerbation of COPD | 2 (3.3) | 0 | 2 (6.9) | .229 |
Death | 0 | 0 | 0 | |
Total complications | 15 (25) | 9 (29) | 6 (20.6) | .167 |
Data are n (%). Statistical comparisons by Fisher's exact test. COPD: Chronic obstructive pulmonary disease.
DISCUSSION
Airflow limitation is associated with mortality in patients with COPD.11 An FEV1 less than 15-20% has been suggested as an indicator for lung transplantation.24 Due to donor insufficiency and advanced age, very few patients are suitable for lung transplantation.2 Therefore, patients with severe emphysema need different treatment options. Endoscopic valve or coil treatments are suggested treatment options for these patients. EBV treatment is preferred in patients with complete fissure and no collateral ventilation. EBC treatment is preferred in patients whether or not they have an incomplete fis-sure or collateral ventilation.25
Klooster et al reported a 17% increase in FEV1 and a 687 mL decrease in RV in their study, in which they presented 12 months of data for 64 patients treated with EBV.26 Lee et al recently reported a study which showed a 41.5% increase in FEV1 and a 1960 mL decrease in RV.27 In a study conducted by the VENT group, on 171 European and 322 American patients who were treated with EBV, an improvement in FEV1 values was presented.28 Stelvio et al reported a 20.9% improvement in FEV1 in EBV patients with no collateral ventilation.29 The IMPACT group, which included 93 patients with EBV application, reported a 13.7% increase in FEV1.30 Three months later, the same group reported a volume reduction of 1195 mL in the target lobe.31 Slebos et al, at the American Thoracic Society meeting in Washington D.C., presented a study by the IMPACT group on 97 patients and reported that FEV1 increased 32% and RV decreased 14% in their 6-month data on EBV results.17 Schuhmann et al reported a decrease of more than 350 mL in lung volume as a result of evaluation at month 3 following EBV application.32 In another study, 20.7% increase in FEV1 was achieved after the EBV procedure.33
In 2009, EBC was applied to 16 patients in Holland. Six months after the procedure, a 14.9% increase in FEV1 and 0.411% decrease in RV were achieved.34 EBC was applied to 46 patients in the RESET study. Three months after the procedure, a 10.6% increase in FEV1 and a 0.31 L decrease in RV were observed.9 Klooster et al reported an increase of 18.9% in FEV1 and a decrease of 600 mL in RV.10 Deslee et al, 12 months after coil treatment, reported an 110 mL increase in FEV1 and a 710 mL decrease in RV.35 In the REVOLENS study, an increase of 11% in FEV1 and a decrease of 0.37 L in RV was achieved.36 In the RENEW study performed with 315 patients with the coil procedure, 12 months after procedure, a 7% increase in FEV1 and a decrease of 310 mL in RV were achieved.37 Simon et al, in their study of the EBC in 2017, reported an increase in FEV1 from 0.5 L to 0.6 L (0.1 L) and a decrease in RV from 6.1 to 5.6.11 Metin et al, reported a 54.0% increase in FEV1 after the valve procedure, and 44% after the coil procedure.38 In our study, among the patients who underwent EBV, we observed a 380 mL increase in FEV1 (P<.001), and 270 mL decrease in RV (P<.001). A signifi-cant improvement was found in each parameter after application. Among the patients who underwent EBC application before and after the procedure; there was a 260 mL increase in FEV1 and 110 ml in RV. Our results were similar to others in the literature.
In our study, the differences in number of patients who could complete the 6MWT was statistically highly significant for both procedures (P<.001). To our knowledge, this is the first study to evaluate completion rates of 6MWT for both procedures in the English literature. Klooster et al presented 6 MWT results of 12-month data of 64 patients with EBV application. They reported that a 61 meter increase was achieved in 6MWT distance (6MWD) after EBV treatment.26 Lee et al recently reported an increase in 6MWD after EBV treatment.27 Similarly, in other previous studies, there was an improvement in 6MWD from 39.3 meters to 64 meters after EBV treatment.17,28,29,31,33,39 In our study, we observed a 35.4 meter (25.7%) increase after EBV treatment.
On the other hand, Selebos et al reported that 6MWD was increased by 84.4 meters after the EBC procedure.34 In the RESET study, a 63.6 meter increase was achieved in 6MWT 3 months after the EBC procedure.9 In Klooster et al, 6MWT was increased by 61 meters 6 months after coil treatment.10 Deslee et al reported an increase of 51.4 meters in 6MWT 12 months after the coil treatment.35 In the REVOLENS study, which applied the coil over 100 patients, an increase of 21 meters in the 6MWT was achieved.36 In the RENEW study, performed with 315 EBC patients, an increase of 14.6 meters in 6MWT was achieved 12 months after the procedure.37 In our study, we achieved an average increase of 35.2 meters (25.2%) in EBC subjects at 6 months. This improvement was similar to others in the literature.
To our knowledge, there are only two studies that have evaluated arterial blood gas analysis prior to endobronchial treatments.11,27 Post-procedure blood gas values were not available in these studies. In our study; there was a significant improvement in PO2 value following EBV treatment and improvement in pH after both EBV and EBC treatment. Jorrit et al reported that they performed general anesthesia during their EBC procedure.40 Valipour et al reported that they performed sedation and general anesthesia during EBV treatment.30 In our study, 24 patients (77.4%) received only local anesthesia and 7 patients (22.58%) received general anesthesia during EBV treatment. On the other hand, 26 patients (89.7%) received local anesthesia and 3 patients (10.35%) received general anesthesia during the EBC procedure. Although EBC cases are mostly performed under general anesthesia and endotracheal intubation in the current literature, we started to perform the EBC application with sedation without endotracheal intubation without any major complication for our recent cases. Additionally, the rate of the type of anesthesia used in the valve or coil procedure was similar in our study.
Simon et al performed the EBC volume reduction procedure on 21 cases on the right upper lobes (63.6%), to 6 cases on the left upper lobe (18.2%), to 3 cases on the right lower lobe (9.1%), and to 3 cases on the left lower lobe (9.1%).11 Slebos et al performed EBV on the upper lobe in 55% of cases and to the lower lobe in 45% of cases17 Lee et al applied EBV to the right middle lobe valve to all cases.27 Kemp et al applied EBV to 52% of the cases on the left upper lobe, 22% on the left lower lobe, 15% on the right upper lobe, 8% on the right upper and right middle lobe, 3% on the right lower lobe.41 Yu et al performed all volume reduction procedures to the upper lobe.42 In our study, we performed EBV on 18 patients (58.1%) to the right upper lobe and 8 patients (25.8%) to the left upper lobe. In the EBC patients, we performed the EBC to mostly 22 patients (75.85%) to the right upper lobe and 4 patients (13.80%) to the left lower lobe.
As to complications, Welling et al reported pneumonia (11.7%), exacerbation of COPD (9.3%), pneumothorax (4.2%), valve migration or shift (1.5%) and mortality (8%) as complications of EBV.43 Another study by the VENT group found a 4% to 7% complication rates (pneumonia, pneumothorax, hemoptysis, exacerbation of COPD, valve migration, aspiration or expectoration) after EBV treatment.28 Balkissoon et al reported a 25.3% complication rate. Among these complications were pneumothorax, the most prevalent at 17.3%.39 Klooster et al reported pneumothorax as a complication in 22% of patients within 6 months follow-up.26 On the other hand, Lee et al reported no complications after EBV treatment.27 In our study, pneumothorax was the most prevalent complication in 3 patients (9.7%) followed by cardiac arrhythmia in 2 patients (6.5%) after EBV. None of our patients died during follow-up. Simon et al reported exacerbation of COPD in 41.0%, pneumonia in 14.8%, pneumothorax in 5.7%, and mortality in 3.3% as complications following EBC.11 Jorrit et al reported COPD exacerbation in 26.42%, pneumonia in 19.28% and pneumothorax in 6.4% after EBC procedure.40 In our study, we observed 3 patients (10.3%) with pneumonia, 2 patients (6.9%) with COPD exacerbation and 1 patients (3.4%) with pneumothorax as a complication following EBC. As in the EBV procedure, no mortality was seen following EBC. Although similar complications have been reported in the literature, the incidence rates vary. This may depend on the number of patient groups, comorbidities and the experience of the team performing the procedure. To the best of our knowledge, hypernatremia and arrhythmia were not reported in the literature before, so we report these findings as complications for the first time in our study.
The main limitations of this study are its retrospective design and small sample-size. Another limitation is the absence of 9, 12 months or later follow-up results of the patients. Moreover, the absence of comorbidity data for all patients prevented evaluation of potential confounding variables.
In conclusion, endobronchial valves and coils currently appear to be similarly safe and effective bronchoscopic volume reduction procedures. Patient selection is very important. Appropriate patients with no collateral circulation must be chosen for the EBC procedure. In addition, large randomized controlled trials are needed to better define optimal patient selection and find overlooked complications during the endobronchial valve and coil procedures.
Funding Statement
None.
REFERENCES
- 1.Gülşen A.. Bronchoscopic Lung Volume Reduction: A 2018 Review and Update. Turk Thorac J. 2018;19(3):141–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 2.Fellrath JM, Scherer T, Franzen DP, Lovis A, Garnier CV, Plojoux J,. et al. Endobronchial coil therapy in severe emphysema: 6-month outcomes from a Swiss National Registry. J Thorac Dis. 2018;10(Suppl 23):2711–8. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 3.Mansfield C, Sutphin J, Shriner K, Criner GJ, Celli BR.. Patient Preferences for Endo-bronchial Valve Treatment of Severe Emphysema. Journal.copdfoundation.org JCOPDF © 2019. [[AUTHOR: Please provide complete citation]] [DOI] [PMC free article] [PubMed]
- 4.Metin B, Intepe YS, Yıldırım S, Ede H, Karacavus S, Erkoc MF.. Our Endobronchial Valve Applications in Bullous Emphysema. Bozok Med J. 2016;6(3):22–5. [Google Scholar]
- 5.Anthonisen NR, Connett JE, Murray RP.. Smoking and lung function of Lung Health Study participants after 11 years. Am J Respir Crit Care Med. 2002;166:675–9. [DOI] [PubMed] [Google Scholar]
- 6.Nocturnal Oxygen Therapy Trial Group. Continuous or nocturnal oxygen therapy in hypoxemic chronic obstructive lung disease: a clinical trial. Nocturnal Oxygen Therapy Trial Group. Ann Intern Med. 1980;93:391–8. [DOI] [PubMed] [Google Scholar]
- 7.Gordon M, Duffy S, Criner GJ.. Lung volume reduction surgery or bronchoscopic lung volume reduction: is there an algorithm for allocation? J Thorac Dis. 2018;10(Suppl 23):2816–23. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 8.Gompelmann D, Sarmand N, Herth FJ.. Interventional pulmonology in chronic obstructive pulmonary disease. Curr Opin Pulm Med. 2017;23: 261–8. [DOI] [PubMed] [Google Scholar]
- 9.Shah PL, Zoumot Z, Singh S, Bicknell SR, Ross ET, Quiring J,. et al. Endobronchial coils for the treatment of severe emphysema with hyperinflation (RESET): a randomised controlled trial. Lancet Respir Med. 2013;1:233–40. [DOI] [PubMed] [Google Scholar]
- 10.Klooster K, Ten Hacken NH, Franz I, Kerstjens HA, van Rikxoort EM, Slebos DJ.. Lung volume reduction coil treatment in chronic obstructive pulmonary disease patients with homogeneous emphysema: a prospective feasibility trial. Respiration. 2014;88:116–25. [DOI] [PubMed] [Google Scholar]
- 11.Simon M, Harbaum L, Oqueka T, Kluge S and Klose H.. Endoscopic lung volume reduction coil treatment in patients with very low FEV1: an observational study. Ther Adv Respir Dis. 2018;12:1–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 12.Gompelmann D, Eberhardt R, Herth F.. Endoscopic volume reduction in COPD a critical review. Dtsch Arztebl Int. 2014;111:827–33. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 13.Criner GJ, Sue R, Wright S, Dransfield M, Rivas-Perez H, Wiese T,. et al. A Multi-center RCT of Zephyr, Endobronchial Valve Treatment in Heterogeneous Emphysema (LIBERATE). Am J Respir Crit Care Med. 2018198(9):1151–64. [DOI] [PubMed] [Google Scholar]
- 14.Klooster K, Hartman J, Ten Hacken NH, Slebos DJ.. One-year follow-up after endo-bronchial valve treatment in patients with emphysema without collateral ventilation treated in the STELVIO trial. Respiration. 2017;93(2):112–21. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 15.Vijayan VK.. Chronic obstructive pulmonary disease. Indian J Med Res. 2013;137:251–69. [PMC free article] [PubMed] [Google Scholar]
- 16.McDonough JE, Yuan R, Suzuki M, Seyednejad N, Elliott WM, Sanchez PG,. et al. Small-airway obstruction and emphysema in chronic obstructive pulmonary disease. N Engl J Med. 2011;365:1567–75. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 17.Herth FJ, Slebos DJ, Criner GJ, Shah PL.. Endoscopic Lung Volume Reduction: An Expert Panel Recommendation - Update 2017. Respiration. 2017;94:380–8. [DOI] [PubMed] [Google Scholar]
- 18.Ramaswamy A, Puchalski J.. Bronchoscopic lung volume reduction: recent updates. J Thorac Dis. 2018;10(4):2519–27. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 19.Choi M, Lee WS, Lee M, Jeon K, Sheen S, Jheon S,. et al. Effectiveness of bronchoscopic lung volume reduction using unilateral endobronchial valve: a systematic review and meta-analysis. Int J Chron Obstruct Pulmon Dis. 2015;10:703–10. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 20.Snell G, Herth FJ, Hopkins P, Baker KM, Witt C, Gotfried MH,. et al. Bronchoscopic thermal vapour ablation therapy in the management of heterogeneous emphysema. Eur Respir J. 2012;39(6):1326–33. [DOI] [PubMed] [Google Scholar]
- 21.Gompelmann D, Eberhardt R, Herth FJ.. Novel Endoscopic Approaches to Treating Chronic Obstructive Pulmonary Disease and Emphysema. Semin Respir Crit Care Med. 2015;36(4):609–15. [DOI] [PubMed] [Google Scholar]
- 22.Global Initiative for Chronic Obstructive Lung Disease. Global strategy for the diagnosis, management, and prevention of chronic obstructive pulmonary disease 2017.
- 23.Gompelmann D, Eberhardt R, Slebos DJ, Brown MS, Abtin F, Kim HJ,. et al. Diagnostic performance comparison of the Chartis System and high-resolution computerized tomography fissure analysis for planning endoscopic lung volume reduction. Respirology. 2014;19:524–30. [DOI] [PubMed] [Google Scholar]
- 24.Weill D, Benden C, Corris PA, Dark JH, Davis RD, Keshavjee S,. et al. A consensus document for the selection of lung transplant candidates: 2014. An update from the Pulmonary Transplantation Council of the International Society for Heart and Lung Transplantation. J Heart Lung Transplant. 2015;34:1–15. [DOI] [PubMed] [Google Scholar]
- 25.Polke M, Rötting M, Sarmand N, Krisam J, Eberhardt R, Herth FJF,. et al. Interventional therapy in patients with severe emphysema: evaluation of contraindications and their incidence. Ther Adv Respir Dis. 2019;13:1–9. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 26.Klooster K, Ten Hacken NH, Hartman JE, Kerstjens HA, van Rikxoort EM, Slebos DJ.. Endobronchial valves for emphysema without interlobar collateral ventilation. N Engl J Med. 2015;373:2325–35. [DOI] [PubMed] [Google Scholar]
- 27.Lee EG and Rhee CK,. Bronchoscopic lung volume reduction using an endobronchial valve to treat a huge emphysematous bullae: a case report. Lee and Rhee BMC Pulmonary Medicine. 2019;19(1):92. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 28.Aggelou K, Siafakas N.. Medical lung volume reduction for severe emphysema: A review. Respiratory Medicine. 2017;131:141–7. [DOI] [PubMed] [Google Scholar]
- 29.Davey C, Zoumot Z, Jordan S, McNulty WH, Carr DH, Hind MD,. et al. Bronchoscopic lung volume reduction with endobronchial valves for patients with heterogeneous emphysema and intact interlobar fissures (the BeLieVeR-HIFi study): a randomised controlled trial. Lancet. 2015;386:1066–73. [DOI] [PubMed] [Google Scholar]
- 30.Valipour A, Slebos DJ, Herth F, Darwiche K, Wagner M, Ficker JH,. et al. Endobronchial Valve Therapy in Patients with Homogeneous Emphysema. Results from the IMPACT Study. Am J Respir. Crit Care Med. 2016;194(9):1073–82. [DOI] [PubMed] [Google Scholar]
- 31.Herth FJ, Eberhardt R, Gompelmann D, Ficker JH, Wagner M, Ek L,. et al. Radiological and clinical outcomes of using Chartis to plan endobronchial valve treatment. Eur Respir J. 2013;41:302–8. [DOI] [PubMed] [Google Scholar]
- 32.Schuhmann M, Raffy P, Yin Y, Gompel-mann D, Oguz I, Eberhardt R,. et al. Computed tomography predictors of response to endobronchial valve lung reduction treatment. Comparison with Chartis. Am J Respir Crit Care Med. 2015; 191:767–74. [DOI] [PubMed] [Google Scholar]
- 33.Li S, G Wang, Wang C, Jin F, Gao X, Yang H,. et al. The REACH study, a randomized controlled trial assessing the safety and effectiveness of the spiration valve system intrabronchial therapy for severe emphysema. Eur Respir J. 2016;48:OA3013. [Google Scholar]
- 34.Slebos DJ, Klooster K, Ernst A, Herth FJF, Kerstjens HAM.. Bronchoscopic lung volume reduction coil treatment of patients with severe heterogeneous emphysema. Chest 2012;142:574–82. [DOI] [PubMed] [Google Scholar]
- 35.Deslee G, Klooster K, Hetzel M, Stanzel F, Kessler R, Marquette CH,. et al. Lung volume reduction coil treatment for patients with severe emphysema: a European multicentre trial. Thorax. 2014;69:980–6. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 36.Deslée G, Mal H, Dutau H, Bourdin A, Vergnon JM, Pison C,. et al. Lung Volume Reduction Coil Treatment vs Usual Care in Patients With Severe Emphysema: The REVOLENS Randomized Clinical Trial. JAMA. 2016;315:175–84. [DOI] [PubMed] [Google Scholar]
- 37.Sciurba FC, Criner GJ, Strange C, Shah PL, Michaud G, Connolly TA.. et al. Effect of Endobronchial Coils vs Usual Care on Exercise Tolerance in Patients With Severe Emphysema: The RENEW Randomized Clinical Trial. JAMA. 2016;315:2178–89. [DOI] [PubMed] [Google Scholar]
- 38.Metin B, Intepe YS, Kılıc M, Yıldırım S, Cetinkaya E.. A single center comparison of cost analyses of different volume reduction therapy methods in the treatment of emphysema. Cukurova Med J 2017;42(4):721–29. [Google Scholar]
- 39.Balkissoon R.. Journal Club-Endo-bronchial Valve Bronchoscopic Lung Volume Reduction. Chronic Obstr Pulm Dis. 2019;6(1):118–25. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 40.Jorrit B.A.. Welling, Dirk-Jan Slebos. Lung volume reduction with endobronchial coils for patients with emphysema. J Thorac Dis. 2018;10(23):2797–805. [DOI] [PMC free article] [PubMed] [Google Scholar]
- 41.Kemp SV, Slebos DJ, Kirk A,. et al. A Multi-center Randomized Controlled Trial of Zephyr Endobronchial Valve Treatment in Heterogeneous Emphysema (TRANSFORM). Am J Respir Crit Care Med 2017;196(12):1535–43. doi: 10.1164/rccm.201707-1327OC [published Online First: 2017/09/09] [DOI] [PubMed] [Google Scholar]
- 42.Yu H, Wang L, Wu Z, Yang Z.. Status of and prospects for bronchoscopic lung volume reduction for patients with severe emphysema. BioScience Trends. 2016;10(5):344–356. [DOI] [PubMed] [Google Scholar]
- 43.Welling JBA, Hartman JE, Van Rikxoort EM, Ten Hacken NHT, Kerstjens HAM, Klooster K, Slebos DJ.. Minimal important difference of target lobar volume reduction after endobronchial valve treatment for emphysema. Respirology. 2018;23:306–10. [DOI] [PubMed] [Google Scholar]